Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 1 year from date of
shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries,
diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be defective.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in Cleveland, Ohio. You will
be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs
will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse of
any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or
problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANT IES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE
BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND
SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF
SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL
AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY
Keithley Instruments, Inc. l 28775 Aurora Road l Cleveland, OH 44139 l 216-248-0400 l Fax: 216-248-6168 l http://www.keithley.com
Keithley Instruments China
Keithley Instruments SARL
Keithley Instruments GmbH
Keithley Instruments, Ltd.
Keitbley Instruments SRL l Viale S. Gimig~~ano 38 l 20146 Milano l 39-2-48303008
Keithley Instruments Far East
Keithley Instruments BV*
Keithley Instruments SA
Keithley Instruments Taiwan
l
Holiday Inn Lido l Office Building 404C l Beijing, 100004 l 8610-4362871 l Fax: 8610-436616s
l
BP 60 l 3 alMe des Garays l 91122 Palaiseau CXdex l 31-6-0115155 l Fax: 31-6-0117726
l
Landsberger St&e 65 l 82110 Germering l 49-89-849307-o l Fax: 49-89-84930759
l
The Minster l 58 Portman Road l Reading, Berkshire RG30 lEA* 44-01734-575666 l Fax: 44-01734-596469
l
KK l Aibido Bldg. l 7-20-2 Nishishinjuku l Shinjuku-ku, Tokyo 160 l 81-3-5389-1964 l Fax: 81-3-5389-2068
Avelingen West 49 l 4202 MS Gorinchem. 31-(0)183-635333 l Fax: 31-(0)183-630821
l
Kriesbachstrasse 4 l 8600 Diibendorfe 41-1-8219444 l Fax: 41-I-8203081
l
1, Ming-Yu First Street l Hsinchu, Taiwan, R.O.C. l 886-35-778462 l Fax: 886-35-778455
Fax: 39-2-48302274
Model 3930A Multifunction Synthesizer
Service Manual
Q 1991, Keithley Instruments, Inc.
Test Instrumentation Group
All rights reserved.
Cleveland, Ohio, U.S.A.
August 1991, First Printing
Document Number: 3930A-902-01 Rev. A
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand and product names are trademarks or registered trader&k of their respective holders.
Safety Precautions
The following safety precautions should be observed before using the Model 3930A Muhifunction Synthesizer and any
associated instruments.
This instzument is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety
precautions required to avoid possible injury. Read over this manual carefully before using the instrument.
Exercise extreme caution when a shock hazard is present at the test circuit. The American National Standards Institute
(ANSI) states that a shock hazard exists when voltage levels greater than 30V rms or 424V peak are present. A good
safety practice is to expect that hazardous voltage is present in any unknown cikuit before measuring.
Inspect the connecting cables and test leads for possible wear, cracks, or breaks before each use.
For maximum safety, do not touch the test cables or any instruments while power is applied to the circuit under test.
Turn off the power and discharge any capacitors before connecting or disconnecting cables from the instrument.
Do not touch any object which could provide a current path to the common side of the circuit under test or power line
(earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
Instrumentation and accessories should not be connected to humans.
HOW TO USE THIS MANUAL
Details procedures to verify that the instrument meets
stated specifications.
Describes basic operating principles for the various cirtits in
the Model 3930A.
Covers fuse replacement, calibration and repair of the instrument, and lists replacement parts.
WARNJNG
The information in this manual is intended for qualified service personnel who can recognize possible shock hazards. Do
not attempt these procedures unless you are qualified to do so.
Analog Board (NP-21022) Calibration Adjustments
Control Board (NJ?-10408) Calibration Adjustments
Co~eciions to Model 197A DMh4
Second Model 3930A Corux!ctions
Co~eciions to PM 6654C Counter
connections to Dv’M
Power Supply Test Point Locations
Limits for Frequency and Duty Cycle Accuracy
Limits for Amplitude Accuracy
Limits for Frequency Response (Sine)
Limits for Frequency Response (Triangle, Sawtooth, Square)
Limits for Total Harmonic Distortion
Limits for DC Voltage Accuracy
The procedures outlined in this section may be used to
verify that the Model 3930A is operating within the limits
stated in the specifications. Performance verification may
be done when the instrument is first received to ensure
that no damage or n&adjustment has occurred during
shipment. Verification may also be performed whenever
there is a question of instrument accuracy, or following
calibration, if desired.
NOTE
If the instrument is still under warranty (Less
than one year from the date of shipment), and
its performance falls outside the specified
range, contact your Keithley representative or
the factory to determine the correct course of
action.
1.2 ENVIRONMENTAL CONDITIONS
All measurements should be made at 18-28°C (65-82W
and at less than 70% relative humidity.
cation procedures. If the instrument has been subject to
extremes of temperature (outside the range specified in
the previous paragraph), additional time should be allowed for internal temperatures to reach normal operating temperature. Typically, it takes one additional hour
to stabilize a unit that is 10°C (18’FQ outside the specified
temperature range.
1.4 LINE POWER
Be sure to set the line voltage switch for the correct voltage. The instrument should be operated at a voltage
within SO% of the line voltage setting and at a frequency
from 48 to 62Hz.
1.5 RECOMMENDED TEST EQUIPMENT
Table l-l lists all the test equipment needed for verification. The procedure for performance verification is based
on using this exact equipment.
NOTE
The verification limits reflect only the accuracy specifications of the Model 3930A. They
do not include test equipment tolerance.
1.3 INITIAL CONDITIONS
The Model 3930A must be turned on and allowed to
warm up for at least one hour before beginning the verifi-
Alternate equipment may be used as long as the substitute equipment has specifications at least as good as
those listed in Table l-l.
The following paragraphs contain the detailed procedures for verifying accuracy specifications of the Model
3930A using the equipment listed in Table l-1. The allowable reading limits in these procedures do not in&de error that could be contributed by this equipment.
These procedures are intended for use only by qualified
personnel using accurate and reliable test equipment. If
the instrument is out of specifications and not under warranty, refer to the calibration information in the service
manual.
ACV; k(O.35% of rdg + 100 counts>
20V range; lkHk-2OOkHz (0.5% of rdg),
2OOkHz-1MHz (0.7%), I-1oMHz (3%), lo-2oMHz (5%)
O.OlHz-12OMHz; time base aging
cl x IO-‘/month; Vp-p measurements
lOHz-11OkHz; 0.001% at full scale; fldB harmonic distortion accuracy from lOH2 to 15.99kHz
5OQ coaxial cable (RG-58C), male BNC connectors, 2ft
(0.6m)
BNC to BNC adapter, 5OQ termination, DC to
25oMH2, VSWR of cl.1
Female BNC connector to double banana
1.6.1
Frequency and Duty Cycle
plug
Accuracy
1. Connect the function synthesizer to a timer/counter
as shown in Figure l-l. Turn on both instruments.
2. Restore factory defaults on the Model 3930A by
pressing SHIFT PRST.
3. Program a 1MHz sine function of 2OVp-p.
4. Set the timer/counter to display the frequency at
Channel A and verify that the frequency reading is
within the limits specified in Table l-2.
5. Program a 1OOHz square wave with a fixed 50% duty
cycle.
6. Set the timer/counter to display the pulse width of
Channel A, and verify that the reading is within
specifications.
7. Modify the frequency and duty cycle settings of the
synthesizer according to Table l-2, and verify the
pulse width readings.
l-2
Model 3930A
Fipure I-I. Connections to Timer/Counter
Performance Verification
SECTTON 1
Model PM 6654C Counter/Timer
Set for 5OQ
input impedance
Table 1-2. Limits for Frequency and Duty Cycle Accuracy
Function
Amplitude Frequency
Sine 2ovp-p MHZ
Square @XD50) 2OVp-p 1OOHZ
1kH.Z
5kHz
1okHz
Square (VAR50) 2OVp-p
1QOHZ
1kHZ
5kHz
1okHz
Allowable Reading
(ltJ°C to 28OC)
999.995lcHz to 1.000005MHz
4.97msec to
497psec to
99.4psec to
49.7psec to
4.98msec to
498psec to
99.6p.sec to
49.8psec to
5.03msec
503psec
100.6psec
50.3pec
5.02msec
502pec
100.4psec
50.2psec
l-3
SECTTON 1
Performance Verification
1.6.2 Amplltude Accuracy
1. Connect the function synthesizer to a DMM as
shown in Figure l-2. Turn on both instruments.
2. Restore factory defaults on the Model 3930A by
pressing SHIFT PRST.
Figure 1-2.
Model 3930A
Connections to Model 297A DMM
I
,
Function
/ output
Model 7051-2 Coaxial
3. Program a 1kHz sine function amplitude of
7.49vp-p.
4: Set the DMM to measure AC volts with autoranging,
and verify that the voltage reading is within the limits specified in Table l-3.
5. Change the function, frequency, and amplitude settings of the synthesizer according to Table l-3, and
verify the voltage
Model .
197A DMM
Cable
readings.,
Pomona 1468
hield
I
Table 1-3.
Function
Amplitude
Sine 7.49vp-p
Limits for Amplitude
Frequency
1kHZ
Square (FXD50)
Triangle
Sawtooth Up
Sawtooth Down
Sine
3ovp-p
1kHZ
1okHz
2okHz
50!&2
Sine lOVp-p lkH2
1okHz
2okHz
501612
Sine
3VP-P
1kl-b
1okHz
2okI-k
5OHz
Accuracy
Allowable Reading
(18OC to 28’0
2.6349Vrms to 26613Vrms
3.558W.n~ to 3.782Vrms
2.0541Vrms to 2.1837Vrms
2.0541Vrn-t~ to 2.1837Vrms
2.0541vrms to 2.1837Vrms
10.5536Vrms to 10.6596Vrms
10.5536Vrms to 10.6596Vrms
10.5536Vrms to 10.6596Vrms
10.5536vrms to
3.5179vrms to 3.5532Vrms
3.5179vrms to 3.5532Vrms
3.5179lhlns to
3.5179vrms to 3.5532Vrms
1.05536vrms to 1.06596Vrms
1.05536Vrms to
1.05536Vn.n~ to 1.06596Vrms
1.05536Vnns to 1.06596Vrms
10.6596Vrms
3.5532Vrms
1.06596Vrms
l-4
Triangle 3ovp-p
Square
Sawtooth Up
Sawtooth Down
1kHZ
8.617OVrms
8.746SVrms
to
1425vrms to 15.15vrms
8.617OVrms to
8.7468Vrms
8.617OVrms to 8.7468Vrms
SECTTONI
Performance Verification
1.6.3
Frequency Response (Sine)
1. Connect the function synthesizer to a wideband AC
DVM as shown in Figure 1-3. Turn on both instruments.
2. Restore factory defaults on the Model 3930A by
pressing SHIFT PRST.
3. Program a lkH2 sine function of BOVp-p.
4. Set the DVM to measure AC volts with autoranging,
and verify that the voltage reading is within the limits specified in Table l-4.
5.
Set the DVM to measure dB and select
a 5052
reference imoedance. Press REL to establish the oresent
voltage ieading as the relative dB reference.
Id 3930A
l
6. Change the frequency setting of the synthesizer according to Table l-4, and verify the subsequent &dB
readings.
7. Program a 1kHz sine function of 3.75Vp-p, and verify that the voltage reading is within limits. Press
REL to establish the reading as the relative dB reference. Verify the fdB readings for the remaining frequencies.
8. Program a 1kHz sine function of 3.74Vp-p, and verify the reading. Establish the present reading as the
relative dB reference. Verify the fdB readings for the
remaining frequencies.
3.75vp-p 1kTdz
1OkHZ -0.ldB to
100kHz -0SdB to +O.ldB
200kHz -0.3dB
35okHz
5OOkHz
7OOkHz -0.3dB to +0.3dB
8OQkHZ
9ookHz -0.5dB
m4J3z
1.2MHz -1.OdB to +0.3dB
3.74vp-p
1kHZ
1OkHZ -0SdB
1001cHz -O.ldB
2OOm -0.3dB to +0.3dB
35okI-k -0.3dB to +0.3dB
500kHz -0.3dB to +0.3dB
7OOkJsz -0.3dB to +03dB
8OOkHz -0.5dB
9ookHz -0.5dB
*I
1MHZ -0.5dB to +0.3dB
1.2MHz -1 .OdB
5.2vrms to 5.4vrms (=REF)
-0.ldB to +O.ldB
-0.3dB to +0.3dB
-0.3dB to +0.3dB
-0.3dB
-0.5dB
-0.5dB
t0
to
to
+0.3dB
+0.3dB
+0.3dB
-0.5dB to +0.3dB
-l.OdB
0.65Vnns to
to
+0.3dB
0.676Vrms (=RJIF)
+O.ldB
to
+0.3dB
-0.3dB
-0.3dB
-0.5dB
-0.5dB
to +0.3dB
to +0.3dB
to +0.3dB
to
+0.3dB
to +0.3dB
0.648vrms to 0.674Vrms (=RJIF)
to +O.ldB
to
+O.ldB
to
+0.3dB
to +0.3dB
to
+0.3dB
l-6
Perhrmance Verification
SECTION 1
1.6.4
Frequency Response (Triangle,
Sawtooth, Square)
1. Connect the function spthesizer to a timer/counter
as shown in Figure l-1. Turn on both instruments.
2. Restore factory defaults on the Model 3930A by
pressing SHJFT PRST.
3. Program a 1kHz triangle wave of 3OVp-p.
4. Set the timer/counter to display the peak-to-peak
voltage at Channel A, and verify that the voltage
Table 1-5.
Function Amplitude Frequency
Triangle 3ovp-p 1kHz
Sawtooth Up
Limits for Frequency Response (Triangle, Sawtooth, Square)
5kHz
1okHz
3OVP-P
1kHz l4.,25Vp-p to
5kHz 0.95 XRECF to 1.05 x REF
1okHz
reading is within the limits specified in Table l-5.
Call this reading REF.
5. Change the frequency to 5kHz. The new reading
should be between 0.97 times the REF reading and
1.03 times the REF reading.
6. Modify the frequency and function settings of the
synthesizer according to Table 1-5, and verify that
the corresponding readings are within the specified
limits.
Allowable Reading
WC to 28OC)
14.25vp-p
0.97xm to 1.03 x REF
0.97xREF to 1.03 x REF
0.95 x RBF to
15.75vpp (=REF,I
.to
15.75vp-p (=REP)
1.05XREF
Sawtooth Down
Square
3ovp-p 1kHz
5kHz 0.95 xREF to 1.05 x REF
lOkJ!Iz 0.95 XRW to 1.05 x REF
3ovp-p 1kHZ
5kHz 0.98 xRRF to 1.02 x REF
1OkHZ 0.98 xRRF to 1.02 x REP
1ookHz 0.98 xREF to 1.02 x REF
14.25vpp to 15.75vp-p (=REF)
14.25vpp to 15.75vp-p (=REFI
l-7
Loading...
+ 38 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.