200 V Protection Module with 1 A Clamp

Description

The Model 2600B-PM-1 200 V Protection Module with 1 A Clamp is a stand-alone module that protects certain lower-voltage source-measure units (SMUs) that are part of a testing configuration from damage by voltage sources that are greater than 220 V.
The maximum DC current is 100 mA, and the clamp prevents < 100 µs transient current spikes up to 1 A.
Figure 1: Model 2600B-PM-1 protection module
The protection module is designed for use with the following lower-voltage SMUs:
Model 2635B Single-Channel System SourceMeter Instrument
Model 2636B Dual-Channel System SourceMeter Instrument
The protection module prevents damage due to transient overcurrent conditions. It is intended for use in applications where a device breakdown or other potential failure could connect the high voltage output of the 1000 V power supply or SMU instrument, such as the Keithley Instruments Model 2410 or 237.
071322002 / June 2015 *PPA071322002* 1
Model 2600B-PM-1
To prevent damage to the Model 2600B-PM-1, in the case of device failure, the current from all power sources, including SourceMeter instruments and power supplies, should be limited to 100 mA DC or less.
The Model 2600B-PM-1 protection module must be connected to protective earth (safety ground) using the supplied green and yellow ground cables (Model CA-568). Connection of the ground cable to protective earth is necessary for safety.
Figure 2: Model 2290-PM-200 grounding
The standard triaxial IN connectors are used to connect the SMU. Maximum clamped voltage on the IN connectors is 240 VDC.
The center conductor of the LO connector inner shield is connected to LO. The outer shield (shell) is chassis ground.
The center conductor of the HI connector is HI and the inner shield is guard. The outer shield (shell) is chassis ground.
There are two high-voltage coaxial connectors used for connection to the device under test (DUT).
The OUT connector is intended for use in applications involving a 1000 V power supply or SMU instrument, such as the Keithley Instruments Model 2410 or 237.