200 V Protection Module with 1 A Clamp
Description
The Model 2600B-PM-1 200 V Protection Module with 1 A Clamp is a stand-alone module that protects certain
lower-voltage source-measure units (SMUs) that are part of a testing configuration from damage by voltage
sources that are greater than 220 V.
The maximum DC current is 100 mA, and the clamp prevents < 100 µs transient current spikes up to 1 A.
Figure 1: Model 2600B-PM-1 protection module
The protection module is designed for use with the following lower-voltage SMUs:
• Model 2635B Single-Channel System SourceMeter Instrument
• Model 2636B Dual-Channel System SourceMeter Instrument
The protection module prevents damage due to transient overcurrent conditions. It is intended for use in
applications where a device breakdown or other potential failure could connect the high voltage output of the
1000 V power supply or SMU instrument, such as the Keithley Instruments Model 2410 or 237.
071322002 / June 2015 *PPA071322002* 1

Model 2600B-PM-1
To prevent damage to the Model 2600B-PM-1, in the case of device failure, the current from all power
sources, including SourceMeter instruments and power supplies, should be limited to 100 mA DC or
less.
The Model 2600B-PM-1 protection module must be connected to protective earth (safety
ground) using the supplied green and yellow ground cables (Model CA-568). Connection of the
ground cable to protective earth is necessary for safety.
Figure 2: Model 2290-PM-200 grounding
The standard triaxial IN connectors are used to connect the SMU. Maximum clamped voltage on the IN
connectors is 240 VDC.
• The center conductor of the LO connector inner shield is connected to LO. The outer shield (shell) is chassis
ground.
• The center conductor of the HI connector is HI and the inner shield is guard. The outer shield (shell) is chassis
ground.
There are two high-voltage coaxial connectors used for connection to the device under test (DUT).
• The OUT connector is intended for use in applications involving a 1000 V power supply or SMU instrument,
such as the Keithley Instruments Model 2410 or 237.
2 071322002 / June 2015