Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a
period of 1 year from date of shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables,
rechargeable batteries, diskettes, and documentation.
During the warranty period, we will, at our option, either repair or replace any product that proves to be
defective.
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in
Cleveland, Ohio. You will be given prompt assistance and return instructions. Send the product, transporta-
tion prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation
prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at
least 90 days.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse of any product or part. This warranty also does not apply to fuses, software, nonrechargeable batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING
ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE BUYERS SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY
DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE
USE OF ITS INSTRUMENTS AND SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN
ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES
SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL AND INSTALLATION, LOSSES
SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
Test Instrumentation Group, Keithley Instruments, Inc. l 28775 Aurora Road * Cleveland, Ohio 44139 l (216) 248-0400 l Fax: (216) 248-6168
AUSTRIA:
FRANCE:
GERMANY:
GREAT BRITAIN:
ITALY
JAPAN:
NETHERLANDS:
SWITZERLAND:
TAIWAN:
Keithley Instruments GesmbH l Rosenhiigelstrasse lKeithley Instruments SARL l 3 All& des Garays l B.l? 60 l 91121 Palaiseau C&+x l 0160-1151-55 l Fax: 01-60-11-77-26
Keithley Instruments GmbH l Landsberger Sh. 65 l D-8034 Gemwring l 089-849307-O l Fax: 089-84930759
Keithley Instruments, Ltd. l The Minster l 58 Portman Road l Reading, Berkshire RG3 IEA l 07X-575666 l Fax 0734-596-l69
Keithley Instruments SRL l Viale S. Gimignano 38 l 20146 Milano l 02-&8303008 l Fax: 0248302274
Keithley Instruments Far Fast KK l Sumiyoshi 24 Bldg., Room 201 l 2-X-2Sumiyoshi-cho l Naka-ku, Yokohama 231 l 81-45-201-2246 l Fax: 81-45-201-2247
Keithley Instruments BV l Avelingen West -19 l 4202 MS Gorinchem l Postbus 559 l 4200 AN Gcxinchrm l 01830-35333 l Fax: 01830-30821
Keithley Instruments SA l Kriesbachstrasse 4 l 8600 Dilbendorf l 01-821-9444 l Fax: 01-820-3081
Keithley Instruments Taiwan l 3rd Floor, Spring Plaza 6 l Section 3, Min Chuan East Road l Taipei, R.O.C. l 8R6-2-501-7065 l Fax: S866-2-509-5329
7 l A-1120 Wien .0222-804-6548 . Fax: 0222-804-3597
hstructionManual
Model 251
I-V Test Software
61990, Keithley Instruments, Inc.
Test Instrumentation Group
All rights reserved.
Cleveland, Ohio, U.S.A.
Document Number: 251-901-01 Rev. B
AI1 Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other btand and product names are trademarks or registered trademarks of their respective holders
Model 251 I-V Test Software
OVERVIEW
INSTRUMENTS CONTROLLED: Model 236/237/238 Source Measure
Units (SMU),Model213QuadVoltageSource,Model707MahixSwitch.
Model 2361 Trigger Controller is required for 2 or more SMUs.
TESTS: Control instruments to source and measure current or voltage.
plots, 707 downloads, data save/acport, prompts and display, for attendedorunattendedoperation. Printingorplottingcanbedoneoff-line
(after acquisition).
MEASUREMENT CAPACITY LIMITS: A curve in the Curve Family Test
may contain up to 1000 points. Up to six source arrays and up to six
measurement arrays can be created by each test. Each array is limited to
4000 Points. The maximum number of arrays and tests is limited only by
the available disk storage space.
HELP: On-line, context sensitive, with index.
SYSTEM REQUIREMENTS
MINIMUM HOST COMFUTERr IBM compatible 80386,640kb Conven-
tional RAM (53Ok or 542720 bytes free) plus 1Mb of LIM expanded
memory, hard drive (2MB free), 720kb 3v2 inch or 1.2Mh 5v, inch floppy
drive, Intel 80387 math coprocessor, EGA color monitor and graphics
card, parallel Port.
OFERATING SYSTEM: M!3-DOS or PC-DOS 3.2 kninhum).
GRAPHICS ADAPTER EGA or VGA color, VGA monochrome.
COMPATIBLE PRINTERS: IBM Graphics Printers, Epson or compatible,
HP LaserJet, IBM Proprinter, and Okidata Microline 182.
COMPATIBLE PLOTTERS HP7440, HP7470, HP7475.
COMFATIBLE MOUSE: Microsoft or Logitech.
The Test Library provides default settings for the Test, 707 Matrix, GPIB
Control, Data Display and Graphic Analysis screens.
FET TESTS: Curve Family, Threshold Voltage, Transconductance, Break-
down, and Leakage.
BIPOLAR TESTS: Curve Family, Gummel Plot, Current Gain, Break-
down, and Leakage (linear and pulsed sweeps).
DIODE TESTS: I-V Curve, Reverse Breakdown, Zener Voltage.
COMPONENT TESTS: Capacitor Leakage, Resistor Voltage Coefficient.
EXAMPL.E GPIB CONTROL: 213 Quad Voltage Source, 707 Switching
Matrix and hot chuck.
TEST SEQUENCES: Bipolar tests, multiple diodes.
MATERIALS PROVIDED
Instruction Manual
Diskettes: 720kb 3tis inch and 1.2Mb 5v, inch containing installation,
program, test library, and sample data.
Specifications subject to change without notice.
INTRODUCTION
This addendum to the Model 251 Instruction Manual is being provided to supply you with the
latest information in the least possible time. Please incorporate the following information into the
manual.
LOADING AND SAVING FILES
An enhancement to the Model 251 I-V Test Software is its ability to perform file operations from
the CURVE FAMILY TEST and DC SOURCE-MEASURE TEST forms. You no longer have to return to the main menu to load/save data, test setup, instrument configuration, and 707 files, or
import/export ASCII and Lotus formatted files. (Loading sequence files remains a main menu operation.)
This load/save enhancement is implemented with function keys. The 4% key loads (and imports) files and the <F3> key saves (and exports) file.
When the cl% key is pressed from the CURVE FAMILY TEST or DC SOURCE-MEASURE TEST
form, the following pop-up is displayed:
Load Data File
I
Load Test File
Load Inst. Cf. File
Load 707 File
hn~ortm File
Imbort Lotus File 1
Use the cursor to select the desired operation and press <Enter>. When loading test setup and data
files, you are reminded that loading overwrites old test parameters and data. If you continue, the
next screen prompts for a file name in the usual manner. After the tile is loaded or imported you
are returned to the appropriate test form.
When the <F3> key is pressed from the CURVE FAMILY TESTor DC SOURCE-MEASURE TEST
form, the following pop-up is displayed:
Save Data File
Save Test File
Save Inst. Cfg. File
Save 707 File
Import ASCII File
Import Lotus File
Use the cursor to select the desired operation and press <Enter>. The next screen prompts for a file
name in the usual manner. After the file is saved or exported you are returned to the appropriate
test form.
SEQUENCER DELAY COMMAND
A DELAY co
mmand has been added to the test sequencer. It forces the sequencer to wait a specified interval before processing the next command. This allows for thermal settling between tests
or sending GPIB commands to other instruments. The DELAY command has the form:
DELAY: delay-value
where delay-value may be specified in seconds or dd:hh:mrn:ss for days, hours, minutes, and seconds (expressed in positive integers). There is a maximum of three colons in the delay-value field,
plus one that must follow the DELAY command.
During the time delay, the screen displays the seconds remaining. You can press <Eso to stop the
delay. The following examples show legal and illegal forms:
DELAY: 0 zero seconds
DELAY: illegal value
DELAY: : zero seconds
DELAY:3::120:100 3 days, 120 minutes, 100 seconds
Table of Contents
SECTION 1
1.1
1.2
1.3
1.4
1.5
1.6
1.6.1
1.6.2
1.6.3
1.7 COMPUTERREQ
1.7.1 ComputerHardware
1.7.2 IEEE-488Interfaces..
1.7.3 Printers and Plotters
- General Information
INTRODU~ON
FEATURES
WARRANTYINFORMATION
MANUALADDENDA
sAFETysyMBoLsANDTERMs
UNFACKINGANDINSPECTION..
Printing or Plotting from a Graphics Screen
Printing from a Display List Screen
Printing or Plotting from Other Screens
Using GRAPHICS.COM for Graphics Screen Dump to Printer
Model 2361 Trigger Controller Front and Rear Panels
Single Unit Connections with Model 8006 Test Fixtures
Single Unit Connections with Model 8002A Test Fixture (Local Sense only)
Multiple Unit Signal and Interlock Connections - Local Sense
Multiple Unit Signal and Interlock Connections - Remote Sense
Multiple Unit Trigger and Bus Connections
Example Signal Connections with Model 213
Example SignaI Connections with Model 7072
Model 213 DIP Switch Factory Defaults
Copy Protection Block
Example of MODEL25KFG File
Curves per Family versus Points per Curve
Sample Single Channel Test Parameters
Sample MuItichannel Test Parameters
DC SOURCE-MEASURE Test Form
Sample DC Source-Measure Test Parameters
GENERAL GPIB OUTPUT Form
Model 236/237/238 Serial PoII Byte Format
Uploading and Storing Setups
Leading and Downloading Setups
CheckingStatusofModel213
This section contains overview information for the
Model 251 I-V Test Software system and is arranged as
follows:
1.2 Features
1.3 Warranty Information
1.4 Manual Addenda
1.5 Safeiy!i$nbols and Terms
--- --r
1.7 Computer Requirements
‘0 ..a... a...y~~.a.“‘S
1.2 FEATURES
Model 251 I-V Test Software is a program that controls a
system of instruments designed to measure I-V characteristics of semiconductors. Model
to support the following instruments:
l
Model 236/237/238 Source Measure Units
e Model 2361 Trigger Controller
l
Model 213 Quad Voltage Source
l
Model 707 Switching Matrix
251 can
be configured
Other Keithley and non-Keithley instruments are supported with user-programmed device-dependent commands.
Key Model 231 features include:
l
The menu-driven software allows easy data collection
of parameters from semiconductors and other components. No computer programming knowledge is necessary to operate the system
l
The data is displayed on the computer monitor and
can be printed or plotted on a hardcopy peripheral.
l
craphi& analysis caiabilit& i&cl&e m&kers, line
slope, zooming, and smoothing.
l
A library of tests is included for resistors, capacitors,
diodes, bipolar transistors, and FEIS. Test setup and
sample data files are included on disk.
l
A test sequencing procedure lets you chain multiple
tests without operator interventions.
l
Full Model 213 support provides an easy method of interfacing a four-channel voltage bias source.
l
Model 707 support for uploading and downloading of
up to 100 setups expands the capability to run multiple
tests on multiple devices.
l
The Model 2361 Trigger Controller that is available
simplifies trigger connections and optimizes trigger
response times among the source measure units.
l-l
SECTION 1
General Information
1.3 WARRANTY INFORMATION
Warranty information is located on the inside front cover
of this instruction manual. Should you require warranty
service, contact your Keithley representative or the fac-
tory for further information.
1.4 MANUAL ADDENDA
Any improvements or changes concerning Model 251 I-V
Test Software or this instruction manual will be explained on a separate addendum supplied with the documentation. Please be sure to note these changes and incorporate them into the manual before operating the system.
1.5 SAFETY SYMBOLS AND TERMS
The following safety symbols and terms may be found on
the instruments or used in this manual:
A
The ’
should consult the operating instructions in the associated manual.
symbol on an instrument indicates that you
A CAUTION heading outlines dangers that could damage an instrument. Such damage may invalidate the warranty.
1.6 UNPACKING AND INSPECTION
1.6.1
Upon receiving the Model 251, carefully unpack it from
the shipping carton and inspect all items for any obvious
physical damage. Report any such damage to the shipping agent at once. Save the original packing carton for
possible future reshipment.
1.6.2
The following items are included with every Model 251
order:
Model 251 distribution disks (XM-byte, 5-l/4” and
72OK-byte, 3-l /2”)
Copy protection block
Model 251 Instruction Manual
Unpacking Procedure
Shipment Contents
The t symbol on an instrument shows that high voltage
may be present on the terminal(s). Use standard safety
precautions to avoid personal contact with these volt-
ages.
A WARNING heading in this and other manuals cautions against possible hazards that could lead to personal
injury or death. Always read the associated information
very carefully before performing the indicated procedure.
NOTE
Model 251 distribution disks are available
upon request in the 36OK-byte, 5-l/4” format.
Contact Keithley’s Applications department.
1.6.3 Instruction Manual
If an additional instruction manual is required, order the
manual package, 251-901-00. The manual package includes an instruction manual and any applicable addenda.
l-2
General Informafion
SECTION 2
1.7 COMPUTER REQUIREMENTS
The following paragraphs discuss minimum and recommended computer requirements, supported interface
cards, printers, and plotters.
1.7.1 Computer Hardware
Model 251 I-V Test Software is intended to run on an IBM
Personal System/2 (80386) or IBM compatible 80386
computer. Table l-l summarizes the computer configuration.
NOTE
Although Model 251 software will run on an
80286-based computer, an 80386based computer is recommended for speed considerations.
1.7.2
The computer must be equipped with a suitable IEEE-488
interface so that it can communicate with the instruments. IEEE-488 interfaces supported by Model 251 software are listed below:
To obtain hard copies of your curves, it will be necessary
for you to connect a suitable printer (serial or parallel) or
plotter (serial or IEEE-4881 to your computer. Compatible
printers and plotters for Model 251 software are as follows:
Printers
IBM Graphics Printer, Epson or compatible
Disk drives
Monitor/graphics card
Mouse
Operating system
Requirements
IBM compatible 80386 with parallel port
64OKB conventional memory (53OKB free), 1MB of LlM expanded memory
Intel 80387
Hard drive, 72OKB 3-l /2” or 1.2MB 5-l /4” floppy drive
EGA or VGA color, VGA monochrome
Microsoft or Lotitech mouse
MS-DOS or PC-“m 3.2
Plotters
HP 7440
HP 7470
HP 7475
Computer Hardware Requirements
l-3
SECTION 2
Getting Stc[&cJ
2.1 INTRODUCTION
5eciion 2 contains introductory information to help you
get your system up and
Section 3 contains more detailed information on using
the Model 251 I-V Test Software.
5ection 2 is organized as follows:
22 Hardware Configuration: Details the system hardware contiguration, cable connections, and trigger controller.
2.3 System Power-up: Covers the power-up procedure
for a typical system, environmental conditions, and
warm-up periods.
2.4 Computer Hardware and Software Installation:
Outlines installing the IEEE-488 interface and Model 251
software, including configuring the software for a graphics adapter, printer, and plotter.
2.5 Operation Overview: Describes the overall structure
of the Model 251 I-V Test Software and details configuring the software for the IEEE-488 interface and instruments.
2.6 System Checkout: Gives the procedure for checking
out the system to ensure that everything is working properly, including
running sample tests.
running as quickly as possible.
2.2 HARDWARE CONFIGURATION
2.2.1
An overall block diagram of a system for running Model
251 software is shown in Figure 2-l. This configuration
with two source measure units, a quad voltage source,
and a switching card and mainframe, is a typical system
with switching, yours may differ. The various system
components perform the following functions:
Model 236/237/238 Source Measure Units - Each
source measure unit can source voltage and simultaneously measure current, or source current and simultane
ously measure voltage. One unit is necessary to run twoterminal device tests, while two units are required to run
bipolar transistor and FET tests. If a substrate bias is required, a third source measure unit, a quad voltage
source, or a current or voltage source will be necessary.
Model 2361 Trigger Controller - A Model 2361 Trigger
Controller is a bus-programmable instrument that is required for all tests where more than one source measure
unit is configured to measure (including synchronized
sweeps). The unit can handle the triggering needs of up
to six source measure units. It not only simplifies system
trigger connections, but it also coordinates the trigger
schemes needed for ah tests. The front and rear panels of
the Model 2361 are shown in Figure 2-2. Refer to Appendix C for more information on the Model 2361.
System Block Diagram
2-l
SECTION 2
Getting Started
Model 213 Quad Voltage Source - The Model 213 is a
quad output digital to analog converter interface for the
IEEE-488 bus. Each port can be programmed for full scale
output of *lV, i5V, or flOV.
Model 707 Switching Matrix - The Model 707 is a programmable switch for connecting signal paths in a matrix
or multiplex topology. It is for large-scale switching applications (up to 576 crosspoints per mainframe). Plug-in
cards are available for general purpose and special purpose switching.
Model 8006 Component Test Fixture -The test fixture is
the interface between the DUT (device under test) and
the instruments. The Model 8006 includes sockets for axial-component, TO packages, and 24-pin DIP package devices. A Model 8007 Semiconductor Test Fixture or a user-supplied test fixture could also be used.
Computer - An IBM F’S/2 (or compatible), or a
386-based computer, equipped with an IEEE-488 interface card, controls the test sys tern.
SECTION 2
Getting Started
Figwe2-2.
m i
I I
II
I
HI I
L
MO&l 251 sonwara
0
0
Q&l
IBM AT, P.Y2 (or compafibb).
ora336+wdoompufsrtih
fdOl@S:
4. subosse~ Of Model 707 swwing Maftfx.
0
Cwnputer
IEEE-488 Inferfaa
1. signal conlwctiotls mada wifh kdel707fJ-TRX Tfiax cabbs.
2 1-r connecUonr male wifh Model 7OWCoax Cab!+%
3. IEEE-483 contwclionr mada wifh Model 7007 Inferfax C&k
(
Semiconductor Matrix Card
I -1
IN OUT
>
IN OUT
Block Diagram of Test System with Model 251 Software
Signal connections between 236/237/238,7072 and
8006or8002A
Interlock
co~eCtions
between 236/237/238 and
8006or8OOL4
COM~C~~OIIS
2361; unguarded signal
between 236/237/238 and
COMWtiOIlS
Unguarded signal connections between
236/237/238,7072 and coax cables
IEEE-488 instrument bus
2.2.2
System Connections
Table 2-l summarizes the necessary cables for a test sys-
tem with Model 236/237/238 Source Measure Units, a
Model 2361 Trigger Controller, a Model 707 Switching
Mainframe with a Model 7072 Semiconductor Matrix
Card, and a Model 8006 or 8002A Test Fixture.
24
Using Figure 2-3 and Figure 2-4 as a guide for a single
source measure unit, and Figure 2-5 through Figure 2-7
for multiple source measure units, connect the equipment together as shown. For test systems using a Model
213 Quad Voltage Source or a Model 7072 Semiconductor
Matrix Card, see Figure 2-8 and Figure 2-9.
SECTION 2
Getting Started
WARNING
All equipment should be turned off when
making connections.
WARNING
Safety interlocks must be
used
to protect the
operator from potentially lethal voltages
that could result in injury or death. Keithley
source measure units are designed to be used
with test fixtures that incorporate a safety interlock switch, such as the Keithley Model
8006 Test Fixture. By using the interlock, the
source measure unit cannot source current or
voltage when the lid of the test fixture is
open. Refer to paragraph 2.3 in the Model
23612371238 Operatois Manual for more information on interlock usage for the Model
8006 and custom test fixtures.
WARNING
The test fixtures must be connected to safety
earth ground using 818 AWG or larger wire.
To simplify test connections, most examples in this manual use local sensing for the source measure units. Local
sensing requires connecting only two signal cables between the source measure unit and the test fixture (OUTF’UT HI and OUTPUT LO). Also, the appropriate mpers between the test fixture signal panel and the compo-
nent test module must be installed.
To optimize accuracy, you should use remote sensing in
any of the examples (remote sensing is recommended for
optimal accuracy with currents above ImA). To use remote sensing,
connect
SENSE HI and SENSE LO to the
test fixture, and add the necessary jumpers to the DUT.
See Section 2 of the Model 236/237/238 Operator’s Man-
ual for complete details on test connections.
WARNING
With remote sensing enabled, an open sense
lead will result in lethal voltages appearing
at OUTPUT HI and, GUARD. This voltage
can cause injury
or
death, and damage external circuitry. Always make sure that the
sense leads are properly connected before
enabling remote sense. NEVER change connections with power applied. Be sure to al-
ways discharge and/or disconnect external
power sources.
2-s
SECTION 2
A. Local Sense
source Measure Unit 1
I I
7007 IEEE468
l%tM
To IEEE438 Interfisca
of Computer
Figure 2-3.
2-6
I
Source Measure Unit
6. Remote Sense
Single Source Measure Unit Connections with Model 8006 Test Fixtures
To IEEE- Intedaa
of czoqwter
8002A Test Fixture
1
------ 1
I-
Source Measure Unit
Not.: Setfbttunswflohfor
sounx nmasum opnraffon.
WARMNQ : Safe oaftmfbn rwuinr the use of
ctewbsafatveafthamu
.-.
_ _..__. -
700;7lE&E48Ei
SECTION 2
Getting Started
e g;;ol-488
of Computer
Figure24
Single Source Measure Unit Connections with Model 8002A Test Fixture @ma1 Sense only)
2-7
SECTION 2
Getting Started
WARNINQ : Safetv interbxks
-SImdadJunlm
Not. : Jumpers am shown for
03mmon-emitter curves
urkg local SeXI.
+707&TRx
Figure 2-5.
Triax Cabkr
Sourca Measure Unit #2
1
t-
236-lLC-3
klmdookcabls
Transistor
Under -rest
Source Measure Unit #3
SocketPimuts: 1 -Etaso (Unit 1)
Multiple Source Measure Unit Signal and Interlock Connections - Local Sense
3 - yieco (Unit 2)
I
Notei To ptwent ground loops.
only on* some measure
unit Shou!d have the Qmund
link connect&g OUTPUT LO
to earth amund. usuallv the
unit thatsupplies pow& to
the base (or gate) of a
remiconduotor.
2-8
SECTION 2
Gefting Sfarfed
Model 8008
23WLC-3
InterId C&k
Source Measure Unit 81
Test Fixture
terminals
common-emitter curves
I
-8MdrrdJurpwr
--Jurprr
r-z
using mmte sense.
?gure 2-6.
IL
236-lLC-3
InMock C&la
Source Measure4 Unit x3
Multiple Source Measure Unit Signal and Interlock Connections - Remote Sense
ScdwtPinout8: 1 =Base
NOW: To prevent gmurd loop.
; = h&o (Unit 2)
I
only ona souma masure
unit slwkl havcr tlw gmund
link mnneoGng OUTPUT LO
I0 earth gmund. usualy me
unk that supplii paver to
(Unit 1)
2-9
SECTION 2
Getting Started
7051 BNC cables
r
~IIIII
II
Model 213 Quad Voltage Source
To IEEE-488
lnterfaca
\
of Computer
Modal 2361 Trigger Controller
I
Source Measure Unit
7007 IEEE-488
cables
I
I
Source Measure Unit
I
r
I !
Sourca Maasum Unlt
Notee : 1. Conneot TFttGGER OUT of source
messura unit to CHANNEL IN
of t@Qer c0ntroiler.
2 Conneot CHANNEL OUT of tr&er
controller to TRIGGER IN of
soura measure unft.
Figure 2-7.
2-10
Model 707 Switching Matrix
Multiple Unit Trigger and Bus Connections
WARNING : Safety inbxla
must be used to pmtect
the operator fmm potenlial
lelhal voltages.
SECTlON 2
Getting Started
Source hhasure Unit Wl
I
‘707ETRx
Tliax oables
Source Measure Unit Y2
,
I
I
I
Model 213 Quad Voltaae Source
Banana Cables -
Noto : Jumpers
common-emitter curves
- StsMard Jumrws
-Gw&mtJwrpen
\ ,.
/
MOSFET
Under Test
Socket Pinouts : 1 I Drain
Note: To plwrNlt ground loops.
2=Gate
I
3 Substrate l?A’f)213)
4-soulcfl
only one sour03 measure
unit should have the ground
link connectiq OUTPUT LO
to earth gmund. usually +Jw
unit
that supplies power to
the base (or gate) of
semiconduotor.
am
shown for
using local sense.
(Unit 2)
a
Figure 2-8.
Example Signal Connections with Model 213
2-11
SECTION 2
Getting Started
Model 8006 Test Fixture
terminals
X18
lyaied
707BTAX
Trlax Cables
II
Source
I I,
@ .’
A
Measure Unit #3
c- z3wLc-3
lntedookceblr
I
!
--JIMQMIX
-auudedJ~mpm
I
Noto: Jumpers am shown
fXd$b3~NBS USit-kJ
!?gure 2-9.
2-12
Example Signal Connections with Model 7072
SECTION 2
Gettim Started
2.3 SYSTEM POWER-UP
2.3.1
The Model 236/237/238 Source Measure Units are de-
signed to operate from 105-125V or 210~250V line power
ranges. A special transformer may be instaIled (at the factory) for 90-1lOV and 195~235V ranges. A slide switch on
the rear panel indicates the present operating voltage. If
the line voltage needs to be changed, refer to the Model
236/237/23S Operator’s Manual.
Both the Model 2361 Trigger Controller and Model 213
Quad Voltage Source operate from 90-125V or 180-250V
line power ranges. The present operating voltage is indicated on the rear panel. If the line voltage needs to be
changed, a slide switch is located inside the unit (refer to
Appendix B or the appropriate instruction manual).
The Model 707 Switching Matrix operates from 105125V
or 210-250V line power ranges. It can be rewired at the
factory for the 90-1lOV and 180-220V ranges. A slide
switch on rear panel indicates the selected line voltage.
To change the selection, refer to the Model 707 Switching
Matrix Instruction Manual.
Instrument Power Requirements
WARNING
As a safety precaution, it is recommended
that all instruments in a rack be connected to
a single power outlet strip. The strip should
be wired to a panel-mounted emergency
stop switch within easy access of the user. A
red push-pull mushroom switch should be
mounted to the top front of the instrument
rack. The switch must be UL recognized and
have sufficient break current capacity for the
load. To specify the break current of the
switch, add up the total VA of the rack instruments and divide by the line voltage.
Multiply the total by 125% to yield the minimum break current specification for an inductive load.
2.3.3 Environmental Conditions
For maximum accuracy, all measuring instruments and
the test fixture must be operated at an ambient temperature between 18 and 28°C at a relative humidity less than
70%. (See specifications in Models 236/237/238 Operatofs Manual.)
2.3.4
Warm-up Period
CAUTION
Do not attempt to operate an instrument on a
supply voltage outside the allowed range, or
instrument damage may occur.
2.3.2
Each instrument should be connected to a grounded AC
outlet using the supplied AC power cord or the equivalent.
Power Connections
WARNING
Each instrument must be connected to a
grounded outlet to ensure protection from
electric shock Failure to use a grounded outlet and a 3-wire power cord may result in personal injury or death because of electric
shock.
The system can be used immediately when all instru-
ments are
tem accuracy, all instruments should be turned on and allowed to warm up for at least one hour before use.
2.3.5
Follow the general procedure below to power up a Model
251 I-V Test system:
1. Connect the instruments together as outlined in
2 Examine the rear panel DIP switches on the Model
3. Turn on each instrument by pressing in its power
first turned on, however, to achieve rated sys-
Power-up Procedure
paragraph 2.2.3.
2361 and Model 213. Set switches to desired
IEEE-488 bus addresses. The factory default setting
(15) of the Model 2361 is shown in Figure 2-2. The default setting (9) of the Model 213 is shown in
Figure 2-10.
switch. Verify that each instrument goes through its
normal power-up routine, as described below.
2-13
SECTION 2
Geffing Sfarfed
OX16
-0
1x0 -6
0x4 -0
0x2 -0
IX1 + -1
IEEEJ89Address I 9
i
Figure 2-10. Model 213 DIP Switch Facto y Defaults
-
self test. If it is not, cycle power on the instrument. (Complete details on the power-up sequence are in the Model
236/237/238 Operator’s Manual.)
The instrument will operate at a line frequency of 50 or
6OHz. To check the line frequency setting, press the
MENU button and rotate the knob until the line frequency is displayed. The alternate line frequency can be
selected by displaying it with the SELECT buttons and
then pressing ENTER. To exit from the menu, press the
MENU button.
To change the IEEE-438 bus address, press the MENU
button and rotate the knob until the bus address is displayed. Key in the desired address and press ENTER. Examples in this manual will assume unit #l at address 16,
unit #2 at address 17, and unit #3 at address 18. To exit
from the MENU, press the MENU button.
WARNING
In a multiple source measure unit test system, make sure all the Model 236/237/23t3s remain powered up. Turning off one or more
Model 236l237I2389 may result in erratic in-
terlock operation. For example, if any unit is
turned off, an operational Model 236/237/238
may not go into standby when the lid of the
system test fixture is opened. This condition
exposes the operator to potentially lethal
voltages that could result in injury or death.
Model 2361 Power-up
The instrument first turns on all LEDs. If the self-test di-
agnostic passes, all LEDs but the one for ROWER turn off,
and the unit is ready for normal operation. If the self-test
diagnosticfails,theLEDswillindicatetheproblem,asexplained in Appendix C.
Model 236/237/238 Power-up
The source measure units perform a self-test at powerup, then display a model number, software revision, and
IEEE-488 address.
Model 213 Power-up
All the front panel LEDs should light for approximately
one second while the unit performs an internal ROM and
RAM self-test. At the end of this self-test, all indicators
should turn off except POWER Flashing LEDs or the ERROR LED indicate failure of the self-test. Refer to the
Model 213 Instruction Manual for details.
Model 707 Power-up
The Model 707 will perform a power-up self-test to check
ROM, RAM, card configuration, stored setups, master/
slave loop, indicators, and displays. It will then show the
software revision level and IEEE-488 bus address. (See
the Model 707 Instruction Manual for a listing of error
messages.) When the self-test has completed, the Model
707 is configured with:
l
All relays opened.
l
RELAYS indicator lit.
l
RELAYsTErtoOoo.
. MEMORYSTEl’toOOl.
The MANUAL TRIGGER LED on the front panel should
be blinking when the source measure unit completes its
2-14
To change the IEEE-488 bus address, press the MENU
key until the present address is displayed. Use the data
entry keys to key in a new value, then press ENTER. This
action will also exit menu mode.
Loading...
+ 298 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.