Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship
for a period of one (1) year from date of shipment.
Keithley Instruments, Inc. warran
probes, cables, software, rechargeable batteries, diskettes, and documentation.
During the warranty period, Keithley Instruments will, at its option, ei
proves to be defective.
To exercise this warranty, write or call your local Keithley Instruments representative, or contact
thley Instruments headquarters in Cleveland, Ohio. You will be given prompt assistance and
Kei
return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs
will be made and the product returned, transportation prepaid. Repaired or replaced products are
warranted for the balance of the original warranty period, or at least 90 days.
ts the following items for 90 days from the date of shipment:
ther repair or replace any product that
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley
Instruments’ express written consent, or misuse of any product or part. This warranty also does not
apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems
arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER W
INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE THE BUYER’S SOLE AND
EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY
FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES
ING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE, EVEN IF KEITHLEY
ARIS
INSTRUMENTS, INC. HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH
DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COST
OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY
PERSON, OR DAMAGE TO PROPERTY.
ARRANTIES, EXPRESSED OR IMPLIED,
OF ITS EMPLOYEES SHALL BE LIABLE
A G R E A T E R M E A S U R E O F C O N F I D E N C E
The print history shown below lists the printing dates of all Revisions and Addenda created
for this manual. The Revision Level letter increases alphabetically as the manual undergoes sub
sequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered
sequentially. When a new Revision is created, all Addenda associated with the previous Revi
sion of the manual are incorporated into the new Revision of the manual. Each new Revision
includes a revised copy of this print history page.
Revision A (Document Number 2440-902-01)................................................................ April 2001
Revision B (Document Number 2440-902-01)........................................................Decmeber 2004
-
-
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand names are trademarks or registered trademarks of their respective holders.
Safety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation.
Although some instruments and accessories would normally be used with non-hazardous voltages, there are
situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety
precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance
information carefully before using the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring
that the equipment is operated within its specifications and operating limits, and for ensuring that operators are
adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper
use of the instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example,
setting the line voltage or replacing consumable materials. Maintenance procedures are described in the user
documentation. The procedures explicitly state if the operator may perform them. Otherwise, they should be
performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly
trained service personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are rated Measurement Category I
and Measurement Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC
60664. Most measurement, control, and data I/O signals are Measurement Category I and must not be directly
connected to mains voltage or to voltage sources with high transient over-voltages. Measurement Category II
connections require protection for high transient over-voltages often associated with local AC mains connections.
Assume all measurement, control, and data I/O connections are for connection to Category I sources unless
otherwise marked or described in the user documentation.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks
or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage
levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect that hazardous
voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that
operators are prevented access and/or insulated from every connection point. In some cases, connections must be
exposed to potential human contact. Product operators in these circumstances must be trained to protect
themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive
part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedancelimited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards,
install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power recep tacle.
Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate
main input power disconnect device must be provided in close proximity to the equipment and within easy reach of
the operator.
04/09
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the
!
circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before:
connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes,
such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line
(earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of
withstanding the voltage being measured.
The instrument and accessories must be used in accordance with specifications and operating instructions, or the
safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and
operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground
connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation
requires the use of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
The symbol on an instrument means caution, risk of danger. The user should refer to the operating instructions
located in the user documenta ti on in all case s where the symbol is marked on the instrument.
The symbol on an instrument means caution, risk of danger. User standard safety precautions to avoid
personal contact with these voltages.
The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The symbol indicates a connection terminal to the equipment frame.
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp
must be properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death.
Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such
damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, rep lacement components in mains circuits - including the power
transformer, test leads, and input jacks - must be purchased from Keithley Instruments. Standard fuses with
applicable national safety approvals may be used if the rating and type are the same. Other components that are
not safety-related may be purchased from other suppliers as long as they are equivalent to the original component
(note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and
functionality of the product). If you are unsure about the applicability of a replacement component, call a Keithley
Instruments office for information.
T o clean an instrument, use a damp cloth or mild, water-based cleaner . Clean the exterior of the instrument only . Do
not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist
of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never
require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected,
the board should be returned to the factory for proper cleaning/servicing.
1-2Performance VerificationModel 2440 5A SourceMeter® Service Manual
Introduction
Use the procedures in this section to verify that Model 2440 accuracy is within the limits
stated in the instrument’s one-year accuracy specifications. You can perform these verifi
cation procedures:
•When you first receive the instrument to make sure that it was not damaged during
shipment.
•To verify that the unit meets factory specifications.
•To determine if calibration is required.
•Following calibration to make sure it was performed properly.
WARNINGThe information in this section is intended for qualified service per-
sonnel only. Do not attempt these procedures unless you are qualified
to do so. Some of these procedures may expose you to hazardous volt
ages, which could cause personal injury or death if contacted. Use
standard safety precautions when working with hazardous voltages.
NOTEIf the instrument is still under warranty and its performance is outside specified
limits, contact your Keithley representative or the factory to determine the cor
rect course of action.
-
-
-
Verification test requirements
Be sure that you perform the verification tests:
•Under the proper environmental conditions.
•After the specified warm-up period.
•Using the correct line voltage.
•Using the proper test equipment.
•Using the specified output signals and reading limits.
Environmental conditions
Conduct your performance verification procedures in a test environment with:
•An ambient temperature of 18-28°C (65-82°F).
•A relative humidity of less than 70% unless otherwise noted.
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-3
Warm-up period
Allow the Model 2440 to warm up for at least one hour before conducting the verification
procedures.
If the instrument has been subjected to temperature extremes (those outside the ranges
stated above), allow additional time for the instrument’s internal temperature to stabilize.
Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified
temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manufacturer.
Line power
The Model 2440 requires a line voltage of 100 to 240V and a line frequency of 50 to
60Hz. Verification tests must be performed within this range.
Recommended test equipment
Table 1-1 summarizes recommended verification equipment. You can use alternate equip-
ment as long as that equipment has specifications at least as good as those listed in Table
1-1. Keep in mind, however, that test equipment uncertainty will add to the uncertainty of
each measurement. Generally, test equipment uncertainty should be at least four times better than corresponding Model 2440 specifications. Table 1-1 lists the specifications of the
recommended test equipment, including maximum allowable uncertainty for alternate test
equipment, which is shown in parentheses.
Performance Verification
1-4Performance VerificationModel 2440 5A SourceMeter® Service Manual
* 90-day, full-range accuracy specifications of ranges required for various measurement points.
** 90-day, ±5°C specifications of nominal resistance values shown. Use actual values for tests. Maximum uncertainty of alternate
test equipment shown in parentheses.
*** Required for verification of 5A current range. Characterize resistor to ±300ppm or better using recommended DMM before ver-
ifying 5A current measurement range.
1Ω resistor characterization
The recommended 1Ω resistor should be characterized to ±300ppm or better before verifying the 5A current measurement range. (You need not characterize the resistor if you are
checking only the 5A current source range.) Use the 4-wire ohms function of the DMM
recommended in
value to calculate the current during the 5A current measurement range test procedure.
Table 1-1 to measure the resistance value, and then use that measured
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-5
Verification limits
The verification limits stated in this section have been calculated using only the Model
2440 one-year accuracy specifications, and they do not include test equipment uncer
tainty. If a particular measurement falls outside the allowable range, recalculate new limits based on Model 2440 specifications and corresponding test equipment specifications.
Example limits calculation
As an example of how verification limits are calculated, assume you are testing the 10V
DC output range using a 10V output value. Using the Model 2440 10V range one-year
accuracy specification of ±(0.02% of output + 1.2mV offset), the calculated output limits
are:
When verifying the resistance measurement accuracy, it will probably be necessary to
recalculate resistance limits based on the actual calibrator resistance values. You can cal
culate resistance reading limits in the same manner described above, but be sure to use the
actual calibrator resistance values and the Model 2440 normal accuracy specifications for
your calculations.
As an example, assume you are testing the 20kΩ range, and that the actual value of the
nominal 19kΩ calibrator resistor is 19.01kΩ. Using the Model 2440 20kΩ range one-year
normal accuracy specifications of ±(0.06% of reading + 3Ω), the recalculated reading
limits are:
7.Press ENTER then EXIT as necessary to return to normal display.
Performing the verification test procedures
Test summary
•DC voltage output accuracy
•DC voltage measurement accuracy
•DC current output accuracy
•DC current measurement accuracy
•Resistance measurement accuracy
If the Model 2440 is not within specifications and not under warranty, see the calibration
procedures in
Section 2 for information on calibrating the unit.
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-7
Test considerations
When performing the verification procedures:
•Be sure to restore factory front panel defaults as previously outlined.
•Make sure that the test equipment is properly warmed up and connected to the
Model 2440 INPUT/OUTPUT jacks. Also be sure that the front panel jacks are
selected with the TERMINALS key.
•Make sure the Model 2440 is set to the correct source range (see below).
•Be sure that the Model 2440 output is turned on before making measurements.
•Be sure the test equipment is set up for the proper function and range.
•Allow the Model 2440 output signal to settle before making a measurement.
•Do not connect test equipment to the Model 2440 through a scanner, multiplexer,
or other switching equipment.
WARNINGThe maximum common-mode voltage (voltage between LO and chas-
CAUTIONThe maximum voltage between INPUT/OUTPUT HI and LO or 4-
Performance Verification
sis ground) is 40V peak. Exceeding this value may cause a breakdown
in insulation, creating a shock hazard.
WIRE SENSE HI and LO is 42V peak. The maximum voltage between
INPUT/OUTPUT HI and 4-WIRE SENSE HI or between INPUT/
OUTPUT LO and 4-WIRE SENSE LO is 5V. Exceeding these voltage
values may result in instrument damage.
Setting the source range and output value
Before testing each verification point, you must properly set the source range and output
value as outlined below.
1.Press either the SOURCE V or SOURCE I key to select the appropriate source
function.
2.Press the EDIT key as required to select the source display field. Note that the cursor will flash in the source field while its value is being edited.
3.With the cursor in the source display field flashing, set the source range to the lowest possible range for the value to be sourced using the up or down RANGE key.
For example, you should use the 10V source range to output a 10V source value.
With a 10V source value and the 10V range selected, the source field display will
appear as follows:
Vsrc:+10.0000V
1-8Performance VerificationModel 2440 5A SourceMeter® Service Manual
4.With the source field cursor flashing, set the source output to the required value
using either:
•The SOURCE adjustment and left and right arrow keys.
•The numeric keys.
5.Note that the source output value will be updated immediately; you need not press
ENTER when setting the source value.
Setting the measurement range
When simultaneously sourcing and measuring either voltage or current, the measure range
is coupled to the source range, and you cannot independently control the measure range.
Thus, it is not necessary for you to set the measure range when testing voltage or current
measurement accuracy.
Compliance considerations
Compliance limits
When sourcing voltage, you can set the SourceMeter to limit current from 10nA to 5.25A.
Conversely, when sourcing current, you can set the SourceMeter to limit voltage from
0.2mV to 42V. The SourceMeter output will not exceed the programmed compliance
limit.
Types of compliance
There are two types of compliance that can occur: “real” and “range.” Depending on
which value is lower, the output will clamp at either the displayed compliance setting
(“real”) or at the maximum measurement range reading (“range”).
The “real” compliance condition can occur when the compliance setting is less than the
highest possible reading of the measurement range. When in compliance, the source out
put clamps at the displayed compliance value. For example, if the compliance voltage is
set to 1V and the measurement range is 2V, the output voltage will clamp (limit) at 1V.
“Range” compliance can occur when the compliance setting is higher than the possible
reading of the selected measurement range. When in compliance, the source output
clamps at the maximum measurement range reading (not the compliance value). For
example, if the compliance voltage is set to 1V and the measurement range is 200mV, the
output voltage will clamp (limit) at 210mV.
-
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-9
Maximum compliance values
The maximum compliance values for the measurement ranges are summarized in
Table 1-2.
Table 1-2
Maximum compliance values
Maximum
Measurement
range
200mV
2V
10V
40V
10μA
100μA
1mA
10mA
100mA
1A
5A
compliance
value
210mV
2.1V
10.5V
42V
10.5μA
105μA
1.05mA
10.5mA
105mA
1.05A
5.25A
Performance Verification
When the SourceMeter goes into compliance, the “Cmpl” label or the units label (i.e.,
“mA”) for the compliance display will flash.
Determining compliance limit
The relationships to determine which compliance is in effect are summarized as follows.
They assume that the measurement function is the same as the compliance function.
•Compliance Setting < Measurement Range = Real Compliance
•Measurement Range < Compliance Setting = Range Compliance
You can determine the compliance that is in effect by comparing the displayed compliance setting to the present measurement range. If the compliance setting is lower than the
maximum possible reading on the present measurement range, the compliance setting is
the compliance limit. If the compliance setting is higher than the measurement range, the
maximum reading on that measurement range is the compliance limit.
1-10Performance VerificationModel 2440 5A SourceMeter® Service Manual
Taking the SourceMeter out of compliance
Verification measurements should not be made when the SourceMeter is in compliance.
For purposes of the verification tests, the SourceMeter can be taken out of compliance by
going into the edit mode and increasing the compliance limit.
NOTEDo not take the unit out of compliance by decreasing the source value or chang-
ing the range. Always use the recommended range and source settings when
performing the verification tests.
Output voltage accuracy
Follow the steps below to verify that Model 2440 output voltage accuracy is within specified limits. This test involves setting the output voltage to each full-range value and measuring the voltages with a precision digital multimeter.
1.With the power off, connect the digital multimeter to the Model 2440 INPUT/
OUTPUT jacks, as shown in
Figure 1-1
Connections for voltage verification tests
Figure 1-1.
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
PEAK
SWEEP
SOURCE
I
5
2440 5A SourceMeter
EDIT
EXIT ENTER
MEAS
FCTN
I
Ω
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
1
REL
RANGE
AUTO
RANGE
®
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V
PEAK
TERMINALS
FRONT/
REAR
Model 2440
Input HI
Digital Multimeter
Input LO
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-11
2.Select the multimeter DC volts measuring function.
3.Set the voltage source protection to NONE. To do so, press CONFIG then
SOURCE V to access the CONFIGURE V-SOURCE menu. Select PROTEC
TION, and set the voltage source protection limit to NONE.
4.Press the Model 2440 SOURCE V key to source voltage, and make sure the source
output is turned on.
5.Verify output voltage accuracy for each of the voltages listed in Table 1-3. For
each test point:
•Select the correct source range.
•Set the Model 2440 output voltage to the indicated value.
•Verify that the multimeter reading is within the limits given in the table.
6.Repeat the procedure for negative output voltages with the same magnitudes as
those listed in
7.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-3
Output voltage accuracy limits
Model 2440
source range
200mV
2V
10V
40V
Table 1-3.
Model 2440 output
voltage setting
200.000mV
2.00000V
10.0000V
40.0000V
Output voltage limits
(1 year, 18°C-28°C)
199.360 to 200.640mV
1.99900 to 2.00100V
9.9968 to 10.0032V
39.9872 to 40.0128V
Performance Verification
-
1-12Performance VerificationModel 2440 5A SourceMeter® Service Manual
Voltage measurement accuracy
Follow the steps below to verify that Model 2440 voltage measurement accuracy is within
specified limits. The test involves setting the source voltage to full-range values, as mea
sured by a precision digital multimeter, and then verifying that the Model 2440 voltage
readings are within required limits.
1.With the power off, connect the digital multimeter to the Model 2440 INPUT/
OUTPUT jacks , as shown in
2.Select the multimeter DC volts function.
3.Set the voltage source protection to NONE. To do so, press CONFIG then
SOURCE V to access the CONFIGURE V-SOURCE menu. Select PROTEC
TION, and set the voltage source protection limit to NONE.
4.Set the Model 2440 to both source and measure voltage by pressing the SOURCE
V and MEAS V keys, and make sure the source output is turned on.
5.Verify output voltage accuracy for each of the voltages listed in Table 1-4. For
each test point:
•Select the correct source range.
•Set the Model 2440 output voltage to the indicated value as measured by the
digital multimeter.
•Verify that the Model 2440 voltage reading is within the limits given in the
table.
Figure 1-1.
-
-
NOTEIt may not be possible to set the voltage source to the specified value. Use the
closest possible setting, and modify reading limits accordingly.
6.Repeat the procedure for negative source voltages with the same magnitudes as
those listed in
7.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-4
Voltage measurement accuracy limits
Model 2440 source
and measure
range*
200mV
2V
10V
40V
Table 1-4.
Source voltage*
200.000mV
10.0000V
40.0000V
2.00000V
Model 2440 voltage reading
limits (1 year, 18°C-28°C)
199.676 to 200.324mV
1.99946 to 2.00054V
9.9978 to 10.0022V
39.991 to 40.009V
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-13
Table 1-4
Voltage measurement accuracy limits
Model 2440 source
and measure
range*Source voltage*
* Measure range coupled to source range when simultaneously sourcing and measuring voltage.
** As measured by precision digital multimeter. Use closest possible value, and modify reading lim-
its accordingly if necessary.
Output current accuracy
Follow the steps below to verify that Model 2440 output current accuracy is within specified limits. The test involves setting the output current to each full-range value and measuring the currents with a precision digital multimeter.
10μA to 1A range accuracy
1.With the power off, connect the digital multimeter to the Model 2440 INPUT/
OUTPUT jacks, as shown in
Figure 1-2
Connections for 10μA to 1A range current verification tests
V
LOCAL
67
DIGITS SPEED
MEAS
I
Ω
1
230
REL
FILTER
89
RECALL
STORE
EDIT
DISPLAY
TOGGLE
POWER
Performance Verification
Model 2440 voltage reading
limits (1 year, 18°C-28°C)
Figure 1-2.
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
PEAK
RANGE
AUTO
RANGE
®
2440 5A SourceMeter
V
4
TRIG
SWEEP
+/-
CONFIG MENU
I
5
SOURCE
EDIT
EXIT ENTER
FCTN
LIMIT
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V
PEAK
TERMINALS
FRONT/
REAR
Model 2440
Digital Multimeter
2.Select the multimeter DC current measuring function.
Input LO
Amps
1-14Performance VerificationModel 2440 5A SourceMeter® Service Manual
3.Press the Model 2440 SOURCE I key to source current, and make sure the source
output is turned on.
4.Verify output current accuracy for the 10μA-1A range currents listed in Table 1-5.
For each test point:
•Select the correct source range.
•Set the Model 2440 output current to the correct value.
•Verify that the multimeter reading is within the limits given in the table.
5.Repeat the procedure for negative output currents with the same magnitudes as
those listed in
Table 1-5.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
5A range accuracy
1.With the power off, connect the digital multimeter and the 1Ω resistor to the Model
2440 INPUT/OUTPUT jacks, as shown in
Figure 1-3
Connections for 5A range current verification tests
SWEEP
I
5
SOURCE
2440 5A SourceMeter
EDIT
EXIT ENTER
RANGE
AUTO
RANGE
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
MEAS
FCTN
I
Ω
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
1
REL
4- WIRE
SENSE
HI
75V
5V
PEAK
PEAK
®
LO
ON/OFF
OUTPUT
INPUT/
OUTPUT
TERMINALS
FRONT/
REAR
Figure 1-3.
75V
PEAK
250V
PEAK
Model 2440
Digital Multimeter
1Ω Resistor
Input HI
Input LO
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-15
2.Select the multimeter DC volts measuring function.
3.Press the Model 2440 SOURCE I key to source current, and make sure the source
output is turned on.
4.Verify output current accuracy for the 5A range. Be sure to:
•Select the 5A source range.
•Set the Model 2440 output current to the correct 5A output value.
•Verify that the multimeter reading is within the 5A range limits given in Table
1-5.
(Since the value of the 1Ω resistor value is assumed to be the same as its nominal
value, the DMM voltage reading is the same as the sourced current.)
5.Repeat the procedure for a negative 5A current output value.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-5
Output current accuracy limits
Model 2440
source range
10μA
100μA
1mA
10mA
100mA
1A
5A*
Model 2440
output current setting
10.0000μA
100.000μA
1.00000mA
10.0000mA
100.000mA
1.00000A
5.00000A
Output current limits
(1 year, 18°C-28°C)
9.9947 to 10.0053μA
99.949 to 100.051μA
0.99946 to 1.00054mA
9.9935 to 10.0065mA
99.914 to 100.086mA
0.99843 to 1.00157A
4.9896 to 5.0104A
Performance Verification
*See separate procedure for 5A range. DMM voltage reading is same as sourced current.
1-16Performance VerificationModel 2440 5A SourceMeter® Service Manual
Current measurement accuracy
Follow the steps below to verify that Model 2440 current measurement accuracy is within
specified limits. The procedure involves applying accurate currents from the Model 2440
current source and then verifying that Model 2440 current measurements are within
required limits.
10μA to 1A range accuracy
1.With the power off, connect the digital multimeter to the Model 2440 INPUT/
OUTPUT jacks, as shown in
2.Select the multimeter DC current function.
3.Set the Model 2440 to both source and measure current by pressing the SOURCE I
and MEAS I keys, and make sure the source output is turned on.
4.Verify measure current accuracy for the 10μA-1A range currents listed in Table
1-6. For each measurement:
•Select the correct source range.
•Set the Model 2440 source output to the correct value as measured by the digital multimeter.
•Verify that the Model 2440 current reading is within the limits given in the
table.
Figure 1-2.
NOTEIt may not be possible to set the current source to the specified value. Use the
closest possible setting, and modify reading limits accordingly.
5.Repeat the procedure for negative calibrator currents with the same magnitudes as
those listed in
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-6.
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-17
5A range accuracy
NOTEThe 1Ω resistor should be characterized to within ±300ppm before verifying the
5A current measurement range. Use the 4-wire ohms function of the DMM to
measure the resistance value, and then use that measured value to calculate the
current during the measurement procedure.
1.With the power off, connect the 1Ω resistor and digital multimeter to the Model
2440 INPUT/OUTPUT jacks, as shown in
2.Select the multimeter DC volts function.
3.Set the Model 2440 to both source and measure current by pressing the SOURCE I
and MEAS I keys, and make sure the source output is turned on.
4.Verify measurement current accuracy for the 5A range as follows:
•Select the 5A source range.
•Set the Model 2440 source output to the correct 5A value as measured by the
digital multimeter.
•Note the DMM voltage reading, and then calculate the current from the voltage reading and characterized 1Ω resistance value as I = V/R, where V is the
DMM voltage reading and R is the characterized resistance value.
•Verify that the Model 2440 current reading is within the 5A limits given in the
Table 1-6.
NOTEIt may not be possible to set the current source to the specified 5A value. Use the
closest possible setting, and modify reading limits accordingly.
Figure 1-3.
Performance Verification
5.Repeat the procedure for a negative 5A current.
6.Repeat the procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select
the rear panel jacks with the front panel TERMINALS key.
Table 1-6
Current measurement accuracy limits
Model 2440 source
and measure range*
10μA
100μA
1mA
10mA
100mA
1A
5A
*Measure range coupled to source range when simultaneously sourcing and measuring current.
** As measured by precision digital multimeter. Use closest possible value, and modify reading limits
accordingly if necessary.
*** Current calculated as follows: I = V/R, where V is the DMM voltage reading, and R is the character-
ized value of the 1Ω resistor.
Source current**
10.00000μA
100.000μA
1.00000mA
10.0000mA
100.000mA
1.00000A
5.00000A
Model 2440 current reading limits
(1 year, 18°C-28°C)
9.9966 to 10.0034μA
99.969 to 100.031μA
0.99967 to 1.00033mA
9.9959 to 10.0041mA
99.939 to 100.061mA
0.99883 to 1.00117A
4.9916 to 5.0084A***
1-18Performance VerificationModel 2440 5A SourceMeter® Service Manual
Resistance measurement accuracy
Use the following steps to verify that Model 2440 resistance measurement accuracy is
within specified limits. This procedure involves applying accurate resistances from a
resistance calibrator and then verifying that Model 2440 resistance measurements are
within required limits.
CAUTIONBefore testing the 2Ω and 20Ω ranges make sure your resistance cali-
brator can safely handle the default test currents for those ranges (see
Model 2440 and calibrator specifications). If not, use the CONFIG
OHMS menu to select the MANUAL source mode, then set the source
current to an appropriate safe value. When using the manual source
mode, total resistance reading uncertainty includes both Source I and
Measure V uncertainty (see specifications), and calculated reading
limits should take the additional uncertainty into account.
If using the Fluke 5450A resistance calibrator, you cannot use the
Auto Ohms mode of the Model 2440 to verify the 2Ω range. The 1A
test current for the 2Ω range of the Model 2440 will damage the cali
brator. On the Model 2440, use the CONFIG OHMS menu to select
the MANUAL source mode, and then set the source (test) current to
100mA.
-
1.With the power off, connect the resistance calibrator to the Model 2440 INPUT/
OUTPUT and 4-WIRE SENSE jacks, as shown in
4-wire connections as shown.
Figure 1-4. Be sure to use the
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-19
Figure 1-4
Connections for resistance accuracy verification
4- WIRE
INPUT/
SENSE
OUTPUT
HI
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
75V
PEAK
SWEEP
I
5
SOURCE
2440 5A SourceMeter
EDIT
EXIT ENTER
MEAS
FCTN
I
Ω
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
1
REL
RANGE
AUTO
RANGE
®
Model 2440
Resistance Calibrator
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V
PEAK
TERMINALS
FRONT/
REAR
Output HISense HI
Output LO
Sense LO
2.Select the resistance calibrator external sense mode.
3.Configure the Model 2440 ohms function for the 4-wire sense mode as follows:
•Press CONFIG then MEAS Ω. The instrument will display the following:
CONFIG OHMS
SOURCE SENSE-MODE GUARD
•Select SENSE-MODE, and then press ENTER. The following will be
displayed:
SENSE-MODE
2-WIRE 4-WIRE
•Select 4-WIRE, and then press ENTER.
•Press EXIT to return to normal display.
4.Press MEAS Ω to select the ohms measurement function, and make sure the source
output is turned on.
5.Verify ohms measurement accuracy for each of the resistance values listed in Table
1-7. For each measurement:
•Set the resistance calibrator output to the nominal resistance or closest available value.
NOTEIt may not be possible to set the resistance calibrator to the specified value. Use
the closest possible setting, and modify reading limits accordingly.
•Select the appropriate ohms measurement range with the RANGE keys.
Performance Verification
1-20Performance VerificationModel 2440 5A SourceMeter® Service Manual
•Verify that the Model 2440 resistance reading is within the limits given in the
table.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT and 4-WIRE
SENSE jacks. Be sure to select the rear panel jacks with the front panel TERMI
-
NALS key.
Table 1-7
Ohms measurement accuracy limits
Calibrator
Model 2440 range
2Ω
20Ω
200Ω
2kΩ
20kΩ
200kΩ
2MΩ
20MΩ
* Nominal resistance value.
** Reading limits based on Model 2440 normal accuracy specifications and nominal resistance values. If
actual resistance values differ from nominal values shown, recalculate reading limits using actual calibra
tor resistance values and Model 2440 normal accuracy specifications. See Verification limits earlier in this
section for details.
resistance*
1.9Ω
19Ω
190Ω
1.9kΩ
19kΩ
190kΩ
1.9MΩ
19MΩ
Model 2440 resistance reading limits**
(1 year, 18°C-28°C)
1.89647 to 1.90353Ω
18.9780 to 19.0220Ω
189.818 to 190.182Ω
1.89837 to 1.90163kΩ
18.9856 to 19.0144kΩ
189.837 to 190.163kΩ
1.89761 to 1.90239MΩ
18.9781 to 19.0219MΩ
-
2
Calibration
Calibration
2-2CalibrationModel 2440 5A SourceMeter® Service Manual
Introduction
Use the procedures in this section to calibrate the Model 2440. These procedures require
accurate test equipment to measure precise DC voltages and currents. Calibration can be
performed either from the front panel or by sending SCPI calibration commands over the
IEEE-488 bus or RS-232 port with the aid of a computer.
WARNINGThis information in this section is intended for qualified service per-
sonnel only. Do not attempt these procedures unless you are qualified
to do so. Some of these procedures may expose you to hazardous volt
ages.
Environmental conditions
Temperature and relative humidity
Conduct the calibration procedures at an ambient temperature of 18-28°C (65-82°F) with
relative humidity of less than 70% unless otherwise noted.
-
Warm-up period
Allow the Model 2440 to warm up for at least one hour before performing calibration.
If the instrument has been subjected to temperature extremes (those outside the ranges
stated above), allow additional time for the instrument’s internal temperature to stabilize.
Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified
temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the
manufacturer.
Line power
The Model 2440 requires a line voltage of 100 to 240V at line frequency of 50 to 60Hz.
The instrument must be calibrated while operating from a line voltage within this range.
Model 2440 5A SourceMeter® Service ManualCalibration2-3
Calibration considerations
When performing the calibration procedures:
•Make sure that the test equipment is properly warmed up and connected to the
Model 2440 front panel INPUT/OUTPUT jacks. Also be certain that the front
panel jacks are selected with the TERMINALS switch.
•Always allow the source signal to settle before calibrating each point.
•Do not connect test equipment to the Model 2440 through a scanner or other
switching equipment.
•If an error occurs during calibration, the Model 2440 will generate an appropriate
error message. See
Appendix B for more information.
WARNINGThe maximum common-mode voltage (voltage between LO and chas-
CAUTIONThe maximum voltage between INPUT/OUTPUT HI and LO or
Calibration cycle
Perform calibration at least once a year to ensure the unit meets or exceeds its
specifications.
NOTECalibration constants are stored in volatile memories of the Model 2440, which
are protected by a replaceable battery when power is off. Typical life for the
battery is approximately 10 years, but the battery should be replaced if the volt
age drops below 2.5V regardless of age. See Section 4 for battery replacement
details.
Calibration
sis ground) is 40V peak. Exceeding this value may cause a breakdown
in insulation, creating a shock hazard.
4-WIRE SENSE HI and LO is 42V peak. The maximum voltage
between INPUT/OUTPUT HI and 4-WIRE SENSE HI or between
INPUT/OUTPUT LO and 4-WIRE SENSE LO is 5V. Exceeding these
voltage values may result in instrument damage.
-
2-4CalibrationModel 2440 5A SourceMeter® Service Manual
Recommended calibration equipment
Table 2-1 lists the recommended equipment for the calibration procedures. You can use
alternate equipment as long that equipment has specifications at least as good as those
listed in the table. For optimum calibration accuracy, test equipment specifications should
be at least four times better than corresponding Model 2440 specifications.
* 90-day, full-range accuracy specifications of ranges required for various measurement points.
** Necessary for calibration of 5A current range. Resistor must be characterized to ±300ppm or better using recommended DMM
before calibrating 5A range.
DC Voltage*
DC Current*
1V:
10V:
100V
10μA:
100μA:
1mA:
10mA:
100mA:
1A:
±5.6ppm
±4.3ppm
±6.3ppm
±25ppm
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
1Ω resistor characterization
The 1Ω resistor must be characterized to ±300ppm or better before calibrating the 5A current range. Use the 4-wire ohms function of the DMM recommended in Table 2-1 to measure the resistance value, and then use that measured value to calculate the current during
the 5A current range calibration procedure.
Model 2440 5A SourceMeter® Service ManualCalibration2-5
Unlocking calibration
Before performing calibration, you must first unlock calibration by entering or sending the
calibration password as explained in the following paragraphs.
Unlocking calibration from the front panel
1.Press the MENU key, and then choose CAL, and press ENTER. The instrument
will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
ß SAVE LOCK CHANGE-PASSWORD
2.Select UNLOCK, and then press ENTER. The instrument will display the
following:
3.Use the up and down RANGE keys to select the letter or number, and use the left
and right arrow keys to choose the position. (Press down RANGE for letters; up
RANGE for numbers.) Enter the present password on the display. (Front panel
default: 002440.)
4.Once the correct password is displayed, press the ENTER key. If the password was
correctly entered, the following message will be displayed:
CALIBRATION UNLOCKED
Calibration can now be executed.
5.Press EXIT to return to normal display. Calibration will be unlocked and assume
the states summarized in
below with calibration unlocked will result in an error +510, “Not permitted with
cal unlocked.”
Table 2-2. Attempts to change any of the settings listed
Calibration
NOTEWith calibration unlocked, the sense function and range track the source func-
tion and range. That is, when :SOUR:FUNC is set to VOLT, the :SENS:FUNC
setting will be 'VOLT:DC'. When :SOUR:FUNC is set to CURR, the
:SENS:FUNC setting will be 'CURR:DC'. A similar command coupling exists
for :SOUR:VOLT:RANG/:SENS:VOLT:RANG and
SOUR:CURR:RANG:SENS:CURR:RANG.
2-6CalibrationModel 2440 5A SourceMeter® Service Manual
Table 2-2
Calibration unlocked states
ModeStateEquivalent remote command
Concurrent Functions
Sense Function
Sense Volts NPLC
Sense Volts Range
Sense Current NPLC
Sense Current Range
Filter Count
Filter Control
Filter Averaging
Source V Mode
Volts Autorange
Source I Mode
Current Autorange
Autozero
Trigger Arm Count
Trigger Arm Source
Trigger Count
Trigger Source
OFF
Source
1.0
Source V
1.0
Source I
10
REPEAT
ON
FIXED
OFF
FIXED
OFF
ON
1
Immediate
1
Immediate
Unlocking calibration by remote
To unlock calibration via remote, send the following command:
:CAL:PROT:CODE '<password>'
:SENS:FUNC:CONC OFF
:SENS:FUNC <source_ function>
:SENS:VOLT:NPLC 1.0
:SENS:VOLT:RANG <source_V_range>
:SENS:CURR:NPLC 1.0
:SENS:CURR:RANG <source_ I_ range>
:SENS:AVER:COUN 10
:SENS:AVER:TCON REPeat
:SENS:AVER:STAT ON
:SOUR:VOLT:MODE FIXED
:SOUR:VOLT:RANG:AUTO OFF
:SOUR:CURR:MODE FIXED
:SOUR:CURR:RANG:AUTO OFF
:SYST:AZERO ON
:ARM:COUNT 1
:ARM:SOUR IMMediate
:TRIG:COUNT 1
:TRIG:SOUR IMMediate
For example, the following command uses the default password:
:CAL:PROT:CODE 'KI002440'
Model 2440 5A SourceMeter® Service ManualCalibration2-7
Changing the password
The default password may be changed from the front panel or via remote as discussed
below.
Changing the password from the front panel
Follow the steps below to change the password from the front panel:
1.Press the MENU key, then choose CAL and press ENTER. The instrument will
display the following:
4.Using the range keys, and the left and right arrow keys, enter the new password on
the display.
5.Once the desired password is displayed, press the ENTER key to store the new
password.
Calibration
Changing the password by remote
To change the calibration password by remote, first send the present password, and then
send the new password. For example, the following command sequence changes the pass
word from the 'KI002440' remote default to 'KI_CAL':
:CAL:PROT:CODE 'KI002440'
:CAL:PROT:CODE 'KI_CAL'
You can use any combination of letters and numbers up to a maximum of eight characters.
NOTEIf you change the first two characters of the password to something other than
“KI”, you will not be able to unlock calibration from the front panel.
-
2-8CalibrationModel 2440 5A SourceMeter® Service Manual
Resetting the calibration password
If you lose the calibration password, you can unlock calibration by shorting together the
CAL pads, which are located on the display board. Doing so will also reset the password
to the factory default (KI002440).
See Section 5 for details on disassembling the unit to access the CAL pads. Refer to the
display board component layout drawing at the end of Section 6 for the location of the
CAL pads.
Viewing calibration dates and calibration count
When calibration is locked, only the UNLOCK and VIEW-DATES selections will be
accessible in the calibration menu. To view calibration dates and calibration count at any
time:
1.From normal display, press MENU, select CAL, and then press ENTER. The unit
will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
2.Select VIEW-DATES, and then press ENTER. The Model 2440 will display the
next and last calibration dates and the calibration count as in the following
example:
NEXT CAL: 02/15/2002
Last calibration: 02/15/2001 Count: 0001
Calibration errors
The Model 2440 checks for errors after each calibration step, minimizing the possibility
that improper calibration may occur due to operator error.
Front panel error reporting
If an error is detected during comprehensive calibration, the instrument will display an
appropriate error message (see
calibration step that caused the error.
Appendix B). The unit will then prompt you to repeat the
Model 2440 5A SourceMeter® Service ManualCalibration2-9
Remote error reporting
You can detect errors while in remote by testing the state of EAV (Error Available) bit (bit
2) in the status byte. (Use the *STB? query to request the status byte.) Query the instru
ment for the type of error by using the appropriate :SYST:ERR? query. The Model 2440
will respond with the error number and a text message describing the nature of the error.
See
Appendix B for details.
-
Front panel calibration
The front panel calibration procedure described below calibrates all ranges of both the
current and voltage source and measure functions. Note that each range is separately cali
brated by repeating the entire procedure for each range.
Step 1: Prepare the Model 2440 for calibration
1.Turn on the Model 2440 and the digital multimeter, and allow them to warm up for
at least one hour before performing calibration.
2.Press the MENU key, then choose CAL and press ENTER. Select UNLOCK, and
then press ENTER. The instrument will display the following:
3.Use the up and down range keys to select the letter or number, and use the left and
right arrow keys to choose the position. Enter the present password on the display.
(Front panel default: 002440.) Press ENTER to complete the process.
4.Press EXIT to return to normal display. Instrument operating states will be set as
summarized in Table 2-2.
-
Calibration
2-10CalibrationModel 2440 5A SourceMeter® Service Manual
Step 2: Voltage calibration
Perform the steps below for each voltage range, using Table 2-3 as a guide.
1.Connect the Model 2440 to the digital multimeter, as shown in Figure 2-1. Select
the multimeter DC volts measurement function.
NOTEThe 2-wire connections shown assume that remote sensing is not used. Remote
sensing may be used, if desired, but it is not essential when using the recom
mended digital multimeter.
Figure 2-1
Voltage calibration test connections
SWEEP
I
5
SOURCE
2440 5A SourceMeter
EDIT
EXIT ENTER
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
MEAS
FCTN
I
Ω
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
1
REL
RANGE
AUTO
RANGE
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
PEAK
®
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V
PEAK
TERMINALS
FRONT/
REAR
-
Model 2440
Input HI
Digital Multimeter
Input LO
2.From normal display, press the SOURCE V key.
3.Press the EDIT key to select the source field (cursor flashing in source display
field), and then use the down RANGE key to select the 200mV source range.
4.From normal display, press MENU.
5.Select CAL, and then press ENTER. The unit will display the following:
8.Note and record the DMM reading, and then adjust the Model 2440 display to
agree exactly with the actual DMM reading. Use the up and down arrow keys to
select the digit value, and use the left and right arrow keys to choose the digit posi
tion (or use the number keys, 0-9, +/-). Note that the display adjustment range is
within ±10% of the present range.
9.After adjusting the display to agree with the DMM reading, press ENTER. The
instrument will then display the following:
V-CAL
Press ENTER to Output +000.00mV
10.Press ENTER. The Model 2440 will source 0mV and at the same time display the
following:
11.Note and record the DMM reading, and then adjust the Model 2440 display to
agree with the actual DMM reading. Note that the display value adjustment limits
are within ±1% of the present range.
12.After adjusting the display value to agree with the DMM reading, press ENTER.
The unit will then display the following:
V-CAL
Press ENTER to Output -200.00mV
13.Press ENTER. The Model 2440 will source -200mV and display the following:
14.Note and record the DMM reading, and then adjust the Model 2440 display to
agree with the DMM reading. Again, the maximum display adjustment is within ±
10% of the present range.
15.After adjusting the display value to agree with the DMM reading, press ENTER
and note that the instrument displays:
V-CAL
Press ENTER to Output -000.00mV
16.Press ENTER The Model 2440 will source -0mV and simultaneously display the
following:
2-12CalibrationModel 2440 5A SourceMeter® Service Manual
17.Note and record the DMM reading, and then adjust the display to agree with the
DMM reading. Once again, the maximum adjustment is within ±1% of the present
range.
18.After adjusting the display to agree with the DMM reading, press ENTER to complete calibration of the present range.
19.Press EXIT to return to normal display, and then select the 2V source range.
Repeat steps 2 through 18 for the 2V range.
20.After calibrating the 2V range, repeat the entire procedure for the 10V and 40V
ranges using
Table 2-3 as a guide. Be sure to select the appropriate source range
with the EDIT and RANGE keys before calibrating each range.
21.Press EXIT as necessary to return to normal display.
Table 2-3
Front panel voltage calibration
Source
range*
Source voltage
0.2V+200.00mV
+000.00mV
-200.00mV
-000.00mV
2V+2.0000V
+0.0000V
-2.0000V
-0.0000V
10V+10.000V
+00.000V
-10.000V
-00.000V
40V+40.000V
+00.000V
-40.000V
-00.000V
**Use EDIT and RANGE keys to select source range.
**Multimeter reading used in corresponding calibration step. See procedure.
8.Note and record the DMM reading, and then adjust the Model 2440 display to
agree exactly with the actual DMM reading. Use the up and down arrow keys to
select the digit value, and use the left and right arrow keys to choose the digit posi
tion (or use the number keys, 0-9, +/-). Note that the display adjustment range is
within ±10% of the present range.
9.After adjusting the display to agree with the DMM reading, press ENTER. The
instrument will then display the following:
I-CAL
Press ENTER to Output +00.000μA
10.Press ENTER. The Model 2440 will source 0mA and at the same time display the
following:
11.Note and record the DMM reading, and then adjust the Model 2440 display to
agree with the actual DMM reading. Note that the display value adjustment limits
are within ±1% of the present range.
12.After adjusting the display value to agree with the DMM reading, press ENTER.
The unit will then display the following:
I-CAL
Press ENTER to Output -10.000μA
13.Press ENTER. The Model 2440 will source -10μA and display the following:
14.Note and record the DMM reading, and then adjust the Model 2440 display to
agree with the DMM reading. Again, the maximum display adjustment is within
±10% of the present range.
15.After adjusting the display value to agree with the DMM reading, press ENTER
and note that the instrument displays:
I-CAL
Press ENTER to Output -00.000μA
16.Press ENTER The Model 2440 will source -0μA and simultaneously display the
following:
17.Note and record the DMM reading, and then adjust the display to agree with the
DMM reading. Once again, the maximum adjustment is within ±1% of the present
range.
-
Model 2440 5A SourceMeter® Service ManualCalibration2-15
18.After adjusting the display to agree with the DMM reading, press ENTER to complete calibration of the present range.
19.Press EXIT to return to normal display, and then select the 100μA source range use
the EDIT and up RANGE keys. Repeat steps 2 through 18 for the 100μA range.
20.After calibrating the 100μA range, repeat the entire procedure for the 1mA through
1A ranges using
Table 2-4 as a guide. Be sure to select the appropriate source
range with the EDIT and up RANGE keys before calibrating each range.
21.After calibrating the 1A range, connect the 1Ω characterized resistor and DMM to
the Model 2440 INPUT/OUTPUT jacks, as shown in
Figure 2-3.
22.Select the DMM DC volts function.
23.Repeat steps 2 through 18 for the 5A range using Table 2-4 as a guide. When entering the DMM reading, use the calculated current as follows: I = V/R, where V is
the DMM voltage reading, and R is the characterized value of the 1Ω resistor.
Figure 2-3
5A range current calibration test connections
4- WIRE
INPUT/
SENSE
OUTPUT
HI
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
75V
PEAK
SWEEP
I
5
SOURCE
2440 5A SourceMeter
EDIT
EXIT ENTER
MEAS
FCTN
I
Ω
230
FILTER
89
RECALL
STORE
V
4
TRIG
LIMIT
+/-
CONFIG MENU
1
REL
RANGE
AUTO
RANGE
®
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V
PEAK
TERMINALS
FRONT/
REAR
Calibration
Model 2440
Digital Multimeter
1Ω Resistor
Input HI
Input LO
2-16CalibrationModel 2440 5A SourceMeter® Service Manual
3.Select LOCK, and then press ENTER. The instrument will display the following
message:
CALIBRATION LOCKED.
Press ENTER or EXIT to continue.
4.Press ENTER or EXIT to return to normal display.
2-18CalibrationModel 2440 5A SourceMeter® Service Manual
Remote calibration
Use the following procedure to perform remote calibration by sending SCPI commands
over the IEEE-488 bus or RS-232 port. The remote commands and appropriate parameters
are separately summarized for each step.
Remote calibration commands
Table 2-5 summarizes remote calibration commands. For a more complete description of
1. Calibration was not unlocked with :CODE command.
2. Invalid data exists. (For example, cal step failed or was aborted.)
3. Incomplete number of cal steps were performed. (For example, omitting a negative full-scale step).
Ranges that calibrated successfully will be saved if calibration is unlocked. Ranges that failed will not be
saved.
Calibration subsystem.
Cal commands protected by password.
Unlock cal; changes password if cal is already unlocked.
(Default password: KI002440.)
Query number of times 2440 has been calibrated.
Save calibration data to EEPROM.*
Lock calibration, inhibit SAVE command operation.
Request cal lock status.
Program calibration year, month, day.
Query calibration year, month, day.
Program calibration due year, month, day.
Query calibration due year, month, day.
Calibrate active measure range. (See Table 2-6 parameters.)
Query measurement cal constants for active range.
Calibrate active source range. (See Table 2-7 parameters.)
Query source cal constants for active range.
Model 2440 5A SourceMeter® Service ManualCalibration2-19
Recommended calibration parameters
The maximum calibration command parameter ranges are: 75% to 150% of full scale for
positive and negative full scale calibration points; ± zero calibration steps have ±50% of
full scale for valid entry ranges. However, for optimum calibration, it is recommended that
you use calibration points within the ranges listed in
each sense range requires three parameters: zero, negative full scale, and positive full
scale. Similarly, each source range requires four parameters: two zero parameters, a posi
tive full-scale parameter, and a negative full-scale parameter.
Note: Parameter steps for each range may be performed in any order, but all four parameter steps for each range must be completed.
For optimum calibration, use parameters within recommended limits.
(negative full scale)
-0.18 to -0.22
-1.8 to -2.2
-9 to -11
-35 to -42
-9E-6 to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-9E-3 to -1E-3
-90E-3 to -110E-3
-0.9 to -1.1
-4.75 to -5.25
Second parameter
(negative zero)
-0.002 to +0.002
-0.02 to +0.02
-0.1 to +0.1
-0.5 to +0.5
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-5E-2 to +5E-2
Third parameter
(positive full scale)
+0.18 to +0.22
+1.8 to +2.2
+9 to +11
+35 to +42
+9E-6 to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+9E-3 to +11E-3
+90E-3 to +110E-3
+0.9 to +1.1
+4.75 to +5.25
Fourth parameter
(positive zero)
-0.002 to +0.002
-0.02 to +0.02
-0.1 to +0.1
-0.5 to +0.5
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-5E-2 to +5E-2
Remote calibration procedure
Step 1: Prepare the Model 2440 for calibration
1.Connect the Model 2440 to the controller IEEE-488 interface or RS-232 port using
a shielded interface cable.
2.Turn on the Model 2440 and the test equipment, and allow them to warm up for at
least one hour before performing calibration.
3.If you are using the IEEE-488 interface, make sure the primary address of the
Model 2440 is the same as the address specified in the program you will be using
to send commands. (Use the MENU key and the COMMUNICATION menu to
access the IEEE-488 address.)
Model 2440 5A SourceMeter® Service ManualCalibration2-21
Step 2: Voltage calibration
1.Connect the Model 2440 to the digital multimeter (see Figure 2-1), and select the
multimeter DC volts function.
2.Send the commands summarized in Table 2-8 in the order listed to initialize voltage calibration. (When the :CAL:PROT:CODE command is sent, the instrument
will assume the operating states listed in
3.Perform the range calibration steps listed in Table 2-9 for all ranges. For each
range:
•Send the :SOUR:VOLT:RANG command to select the source and sense range
being calibrated. For example, for the 2V range, the following command
would be sent:
:SOUR:VOLT:RANG 2
•Program the source to output the negative full-range value using the
:SOUR:VOLT command. For example:
:SOUR:VOLT -2
•Note and record the multimeter reading.
•Use the multimeter reading as the parameter for the :CAL:PROT:SOUR and
•:CAL:PROT:SENS commands. For example, a typical value for the 2V range
would be:
:CAL:PROT:SOUR -1.998
:CAL:PROT:SENS -1.998
•Program the voltage source for 0V output using the :SOUR:VOLT 0.0
command.
•Note the multimeter reading.
•Send the source and sense calibration commands using the multimeter reading
for the parameter. For example:
:CAL:PROT:SOUR 1E-3
:CAL:PROT:SENS 1E-3
•Set the source to the positive full-range value using the :SOUR:VOLT command. For example:
:SOUR:VOLT 2
•Note and record the multimeter reading.
•Send the source and sense commands using the multimeter reading as the
parameter. For example:
:CAL:PROT:SOUR 1.997
:CAL:PROT:SENS 1.997
•Send the SOUR:VOLT 0.0 command to set the source voltage to 0V.
•Note and record the multimeter reading.
•Send the :CAL:PROT:SOUR command using the multimeter reading as the
command parameter. For example:
:CAL:PROT:SOUR -1.02E-3
Table 2-2.)
Calibration
2-22CalibrationModel 2440 5A SourceMeter® Service Manual
Table 2-8
Voltage calibration initialization commands
CommandDescription
*RST
:SOUR:FUNC VOLT
:SENS:CURR:PROT 0.1
:SENS:CURR:RANG 0.1
:SOUR:VOLT:PROT MAX
:SYST:RSEN OFF
:CAL:PROT:CODE 'KI002440'
:OUTP:STAT ON
*Remote sensing may be used if desired, but is not essential when using the recommended digital
multimeter. To use remote sensing, send :SYST:RSEN ON.
Restore GPIB defaults.
Activate voltage source.
Current limit when voltage source is active.
Make sure 1A range is not active.
Maximum allowable source voltage.
Disable remote sensing.*
Unlock cal.
Turn source on.
Table 2-9
Voltage range calibration commands
StepCommand/procedure*Description
1
:SOUR:VOLT:RANGE <Range>
2
:SOUR:VOLT -<Source_value>
3
Take DMM reading.
4
:CAL:PROT:SOUR <DMM_Reading>
5
Check 2440 for errors.
6
:CAL:PROT:SENS <DMM_Reading>
7
Check 2440 for errors.
8
:SOUR:VOLT 0.0
9
Take DMM reading.
10
:CAL:PROT:SOUR <DMM_Reading>
11
Check 2440 for errors.
12
:CAL:PROT:SENS <DMM_Reading>
13
Check 2440 for errors.
14
:SOUR:VOLT +<Source_value>
15
Take DMM reading.
16
:CAL:PROT:SOUR <DMM_Reading>
17
Check 2440 for errors.
18
:CAL:PROT:SENS <DMM_Reading>
19
Check 2440 for errors.
20
:SOUR:VOLT 0.0
21
Take DMM reading.
22
:CAL:PROT:SOUR <DMM_Reading>
*1. Perform complete procedure for each range, where <Range> = 0.2, 2, 10, and 40, and <Source_value> = 0.2, 2, 10, and 40.
2. <DMM_Reading> parameter is multimeter reading from previous step.
3. Use :SYST:ERR? query to check for errors.
Select source range.
Establish negative full-range polarity.
Read actual output value.
Calibrate sense function negative full scale.
Calibrate source function negative full scale.
Set output to 0V.
Read actual output value.
Calibrate sense function negative zero.
Calibration source function negative zero.
Establish positive full-range polarity.
Read actual output value
Calibrate sense function positive full scale.
Calibrate source function positive full scale.
Set output to 0V.
Read actual output value.
Calibrate source positive zero.
Model 2440 5A SourceMeter® Service ManualCalibration2-23
Step 3: Current calibration
1.Connect the Model 2440 to the digital multimeter (see Figure 2-2), and select the
multimeter DC current function.
2.Send the commands summarized in Table 2-10 in the order listed to initialize cur-
rent calibration.
3.Calibrate the 10μA to 1A current ranges using the procedure summarized in Table
2-11. For each range:
•Send the :SOUR:CURR:RANG command to select the source and sense range
being calibrated. For example, for the 1mA range, the command is:
:SOUR:CURR:RANG 1E-3
•Program the source to output the negative full-range value using the
:SOUR:CURR command. For example:
:SOUR:CURR -1E-3
•Note and record the multimeter reading.
•Use the multimeter reading as the parameter for the :CAL:PROT:SOUR and
:CAL:PROT:SENS commands. For example, a typical value for the 1mA
range would be:
:CAL:PROT:SOUR -1.025E-3
:CAL:PROT:SENS -1.025E-3
•Program the current source for 0A output using the :SOUR:CURR 0.0
command.
•Note the multimeter reading.
•Send the source and sense calibration commands using the multimeter reading
for the parameter. For example:
:CAL:PROT:SOUR 1E-6
:CAL:PROT:SENS 1E-6
•Set the source to the positive full-range value using the :SOUR:CURR command. For example, for the 1mA range:
:SOUR:CURR 1E-3
•Note and record the multimeter reading.
•Send the source and sense commands using the multimeter reading as the
parameter. For example:
:CAL:PROT:SOUR 1.03E-3
:CAL:PROT:SENS 1.03E-3
•Send the :SOUR:CURR 0.0 command to set the source current to 0A.
•Note and record the multimeter reading.
•Send the :CAL:PROT:SOUR command using the multimeter reading as the
command parameter. For example:
:CAL:PROT:SOUR 1E-6
Calibration
2-24CalibrationModel 2440 5A SourceMeter® Service Manual
4.Connect the 1Ω resistor and DMM to the Model 2440 INPUT/OUPUT jacks, as
shown in
Figure 2-3. Select the DMM DC volts function.
5.Repeat step 3 for the 5A range using the calculated current as follows: I = V/R,
where V is the DMM voltage reading, and R is the characterized value of the 1Ω
resistor.
Table 2-10
Current calibration initialization commands
CommandDescription
:SOUR:FUNC CURR
:SENS:VOLT:PROT 10
:SENS:VOLT:RANG 10
:OUTP:STAT ON
Select source current mode.
Voltage limit when current source is active.
Make sure 40V range is not active.
Turn source on.
Table 2-11
Current range calibration commands
StepCommand/procedure*Description
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
*1. Perform complete procedure for each range, where <Range> and <Source_value> = 10E-6, 100E-6, 1E-3,
10E-3, 100E-3, 1, or 5.
2. <DMM_Reading> parameter is multimeter reading from previous step.
3. Use :SYST:ERR? query to check for errors.
:SOUR:CURR:RANGE <Range>
:SOUR:CURR -<Source_value>
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2440 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2440 for errors.
:SOUR:CURR 0.0
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2440 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2440 for errors.
:SOUR:CURR +<Source_value>
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2440 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2440 for errors.
:SOUR:CURR 0.0
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Select source range.
Establish negative full-range polarity.
Read actual output value.
Calibrate sense function negative full scale.
Calibrate source function negative full scale.
Set output to 0A.
Read actual output value.
Calibrate sense function negative zero.
Calibration source function negative zero.
Establish positive full-range polarity.
Read actual output value.
Calibrate sense function positive full scale.
Calibrate source function positive full scale.
Set output to 0A.
Read actual output value.
Calibrate source positive zero.
Model 2440 5A SourceMeter® Service ManualCalibration2-25
Step 4: Program calibration dates
Use following commands to set the calibration date and calibration due date:
Note that the year, month, and date must be separated by commas.
Step 5: Save calibration constants
Calibration is now complete, so you can store the calibration constants in EEROM by
sending the following command:
:CAL:PROT:SAVE
NOTECalibration will be temporary unless you send the SAVE command. Also, cali-
bration data will not be saved if (1) calibration is locked, (2) invalid data exists,
or (3) all steps were not completed.
Step 6: Lock out calibration
To lock out further calibration, send the following command after completing the calibration procedure:
Calibration
:CAL:PROT:LOCK
Single-range calibration
Normally, the complete calibration procedure should be performed to ensure that the
entire instrument is properly calibrated. In some instances, however, you may want to cal
ibrate only certain ranges. To do so, complete the entire procedure only for the range(s) to
be calibrated. Keep in mind, however, that you must complete all parameter steps for each
source or sense range. Also be sure to set calibration dates and save calibration after cali
brating the desired range(s).
-
-
2-26CalibrationModel 2440 5A SourceMeter® Service Manual
3
Routine Maintenance
Routine Maintenance
3-2Routine MaintenanceModel 2440 5A SourceMeter® Service Manual
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:NO INTERNAL OPERATOR SERVICABLE PARTS, SERVICE BY QUALIFIED PERSONNEL ONLY.
Introduction
The information in this section deals with routine type maintenance that can be performed
by the operator.
Line fuse replacement
WARNINGDisconnect the line cord at the rear panel, and remove all test leads
connected to the instrument (front and rear) before replacing the line
fuse.
The power line fuse is accessible from the rear panel, just above the AC power receptacle
(see
Figure 3-1).
Figure 3-1
Rear panel
42V
PEAK
4-WIRE
SENSE
WITH FRONT PANEL MENU)
HI
5V
PEAK
LO
INPUT/
!
OUTPUT
IEEE-488
(ENTER IEEE ADDRESS
5V
PK
42V
PEAK
40V
PEAK
5V
PEAK
V, Ω,
GUARD
GUARD
SENSE
CATI
RS232
Model 2440
TRIGGER
LINK
MADE IN
U.S.A.
LINE FUSE
SLOWBLOW
3.15A, 250V
LINE RATING
100-240VAC
50, 60HZ
220VA MAX
OUTPUT
ENABLE
Model 2440 5A SourceMeter® Service ManualRoutine Maintenance3-3
Perform the following steps to replace the line fuse:
1.Carefully grasp and squeeze together the locking tabs that secure the fuse carrier to
the fuse holder.
2.Pull out the fuse carrier, and replace the fuse with the type specified in Table 3-1.
CAUTIONTo prevent instrument damage, use only the fuse type specified in
Table 3-1.
3.Reinstall the fuse carrier.
NOTEIf the power line fuse continues to blow, a circuit malfunction exists and must be
corrected. Refer to the troubleshooting section of this manual for additional
information.
Table 3-1
Power line fuse
Line voltageRatingKeithley part no.
100-240V250V, 3.15A, Slow
Blow 5 × 20mm
FU-106-3.15
Routine Maintenance
3-4Routine MaintenanceModel 2440 5A SourceMeter® Service Manual
4
Troubleshooting
Troubleshooting
4-2TroubleshootingModel 2440 5A SourceMeter® Service Manual
Introduction
This section of the manual will assist you in troubleshooting and repairing the Model
2440. Included are self-tests, test procedures, troubleshooting tables, and circuit descrip
tions. Note that disassembly instructions are located in Section 5, and component layout
drawings are at the end of Section 6.
WARNINGThe information in this section is intended for qualified service per-
sonnel only. Do not perform these procedures unless you are qualified
to do so. Some of these procedures may expose you to hazardous volt
ages that could cause personal injury or death. Use caution when
working with hazardous voltages.
-
Repair considerations
Before making any repairs to the Model 2440, be sure to read the following
considerations.
CAUTIONThe PC-boards are built using surface mount techniques and require
specialized equipment and skills for repair. If you are not equipped
and/or qualified, it is strongly recommended that you send the unit
back to the factory for repairs or limit repairs to the PC-board
replacement level. Without proper equipment and training, you could
damage a PC-board beyond repair.
•Repairs will require various degrees of disassembly. However, it is
recommended that the Front Panel Tests be performed prior to
any disassembly. The disassembly instructions for the Model 2440
are contained in
•Do not make repairs to surface mount PC-boards unless equipped
and qualified to do so (see previous CAUTION).
•When working inside the unit and replacing parts, be sure to
adhere to the handling precautions and cleaning procedures
explained in
•Many CMOS devices are installed in the Model 2440. These static-
sensitive devices require special handling as explained in
Section 5.
•Whenever a circuit board is removed or a component is replaced,
the Model 2440 must be recalibrated. See
calibrating the unit.
Section 5 of this manual.
Section 5.
Section 2 for details on
-
Model 2440 5A SourceMeter® Service ManualTroubleshooting4-3
Power-on self-test
During the power-on sequence, the Model 2440 will perform a checksum test on its
EPROM and test its RAM. If one of these tests fails, the instrument will lock up.
Front panel tests
There are three front panel tests: one to test the functionality of the front panel keys and
two to test the display. In the event of a test failure, refer to
details on troubleshooting the display board.
KEYS test
The KEYS test lets you check the functionality of each front panel key. Perform the following steps to run the KEYS test.
1.Display the MAIN MENU by pressing the MENU key.
2.Select TEST, and press ENTER to display the SELF-TEST MENU.
3.Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4.Select KEYS, and press ENTER to start the test. When a key is pressed, the label
name for that key will be displayed to indicate that it is functioning properly. When
the key is released, the message “No keys pressed” is displayed.
5.Pressing EXIT tests the EXIT key. However, the second consecutive press of EXIT
aborts the test and returns the instrument to the SELF-TEST MENU. Continue
pressing EXIT to back out of the menu structure.
“Display board checks” for
DISPLAY PATTERNS test
The display test lets you verify that each pixel and annunciator in the vacuum fluorescent
display is working properly. Perform the following steps to run the display test:
1.Display the MAIN MENU by pressing the MENU key.
2.Select TEST, and press ENTER to display the SELF-TEST MENU.
3.Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4.Select DISPLAY-PATTERNS, and press ENTER to start the display test. There
are five parts to the display test. Each time a front panel key (except EXIT) is
pressed, the next part of the test sequence is selected. The five parts of the test
sequence are as follows:
•Checkerboard pattern (alternate pixels on) and all annunciators.
•Checkerboard pattern and the annunciators that are on during normal
operation.
•Horizontal lines (pixels) of the first digit are sequenced.
Troubleshooting
4-4TroubleshootingModel 2440 5A SourceMeter® Service Manual
•Vertical lines (pixels) of the first digit are sequenced.
•Each digit (and adjacent annunciator) is sequenced. All the pixels of the
selected digit are on.
5.When finished, abort the display test by pressing EXIT. The instrument returns to
the FRONT PANEL TESTS MENU. Continue pressing EXIT to back out of the
menu structure.
CHAR SET test
The character set test lets you display all characters. Perform the following steps to run the
character set test:
1.Display the MAIN MENU by pressing the MENU key.
2.Select TEST, and press ENTER to display the SELF-TEST MENU.
3.Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4.Select CHAR-SET, and press ENTER to start the character set test. Press any key
except EXIT to cycle through all displayable characters.
5.When finished, abort the character set test by pressing EXIT. The instrument
returns to the FRONT PANEL TESTS MENU. Continue pressing EXIT to back
out of the menu structure.
Principles of operation
The following information is provided to support the troubleshooting tests and procedures
covered in this section of the manual. Refer to the following drawings:
Figure 4-1 shows an overall block diagram of the Model 2440. Circuitry may be divided
into three general areas:
•Analog circuits — includes sourcing circuits such as the DACs, clamps, output
stage, and feedback circuits, as well as measurement circuits such as the A/D
converter.
•Digital circuits — includes the microcomputer that controls the analog section,
front panel, and GPIB and RS-232 ports, as well as associated interfacing circuits.
Model 2440 5A SourceMeter® Service ManualTroubleshooting4-5
•Power supplies — converts the AC line voltage into DC voltages that supply the
power for the digital and analog circuits, and the output stage.
Analog circuits
Figure 4-2 shows the analog circuitry block.
Troubleshooting
4-6TroubleshootingModel 2440 5A SourceMeter® Service Manual
Figure 4-1
Overall block diagram
Analog Section
Display,
Keyboard
Digital
I/O
Trigger
Link
Front
Panel
Controller
Trigger,
Digital
I/O
DACs
Clamps
Feedback
A/D
Converter
Microcomputer
Digital Section
Output
Stage
RS-232
GPIB
Interface
Guard
Buffer
Output
Guard
RS-232 I/O
GPIB I/O
To Analog
Circuits
±15V
+5V±21V
Analog
Power
Supply
To
Output Stage
±58V +5V +12V
Output
Stage
Power
Supply
Line In
Power Supply
To
Digital Circuits
Digital
Power
Supply
Model 2440 5A SourceMeter® Service ManualTroubleshooting4-7
Figure 4-2
Analog circuitry block diagram
V DAC
I DAC
Control
Stage
+21
+58
+
-
Sense
Resistors
Protection
Protection
O
Output
HI
S+
Output
LO
S-
Guard
Out
Guard
Sense
FB
VFB
A/D
IFB
V Clamp
I Clamp
+7
MUX
-21
-58
Error
Amp
VFB
IFB
S
Output
O
Remote
O
D/A converters control the programmed voltage and current, or voltage compliance and
current compliance. Each DAC has two ranges, a 10V full-scale output or a 1V full-scale
output. The DAC outputs are fed to the summing node, FB. Either the V DAC or the
I
DAC has the ability to control the main loop. If the unit is set for SV (source voltage), it
will source voltage until the compliance current is reached (as determined by the I DAC
setting), and the current loop will override the voltage loop. If, however, the unit is set for
SI (source current), it will source current until the compliance voltage is reached (as deter
mined by the V DAC setting), and the voltage loop will override the current loop. A priority bit in the V clamp/I clamp circuit controls these functions.
-
The error amplifier adds open-loop gain and slew-rate control to the system to assure
accuracy and provide a controllable signal for the output stage, which provides the neces
sary voltage and current gain to drive the output. Sense resistors in the HI output lead provide output current sensing, and a separate sense resistor is used for each current range.
The 1A and 5A ranges use 0.2V full-scale for a full-range output, while all other ranges
use 2V output for full-scale current. Voltage feedback is routed either internally or exter
nally.
There are four voltage ranges: 0.2V, 2V, 10V, and 40V. The feedback gain changes only
for the 10V and 40V ranges, resulting in three unique feedback gain values. A multiplexer
directs the voltage feedback, current feedback, reference, or ground signal to the A/D con
verter. An opto-isolated interface provides control signals for both DACs, analog circuit
control, and A/D converter communication to the digital section.
-
Troubleshooting
-
-
4-8TroubleshootingModel 2440 5A SourceMeter® Service Manual
Power supply
Figure 4-3 shows a block diagram of the Model 2440 power delivery system.
Figure 4-3
Power supply block diagram
Line
Neutral
Analog Board
+15V+5V
FFF
Constant Frequency
Low Noise Floating
Switching Supply
D
PFC
+24Vdc
DC/DC
Converter
F
-15V
+24dc
Regulators
D
Output Stage
-58V -21V+21V +58V
+12Vdc
DC
+5Vdc
O
Digital
Circuits
D
The offline switching power supply provides all power for the instrument while providing
universal inputs and power factor correction for the 110/120V line. The digital board runs
off of 5V and 12V supplies derived from the switcher. The +12VD supply is set to pro
-
gram the flash ROM. (See Digital circuitry below.)
A constant-frequency switching supply runs off the +24VD supplies and generates all the
floating and output supply voltages for the analog board: +5V and ±15V, ±21V and ±58V.
Model 2440 5A SourceMeter® Service ManualTroubleshooting4-9
Output stage
Figure 4-4 shows a simplified schematic of the output stage.
Figure 4-4
Output stage simplified schematic
+58V
HI Drive
Q500
+21V
Q504
Q514
Maindrive
O
+15V
300K
-15V
F
F
F
+
-
HI Drive
Q516
Q518
F
Q521
F
Q523
Q525
Troubleshooting
Q507
-21V
Q503
-58V
4-10TroubleshootingModel 2440 5A SourceMeter® Service Manual
The Model 2440 output stage serves two purposes: (1) it converts signals from floating
common to output common, and (2) it provides both voltage and current amplification.
The output stage drive transistors are biased in class B configuration to prevent the possi
bility of thermal runaway with high-current output values.
Output transistors Q518 and Q521 are cascoded with output MOSFETs Q516 and Q523.
All other MOSFETs and transistors are slaves, and the voltages across these devices are
determined by the resistor-capacitor ladder circuits shown. High-current drive capability
is provided by Q500-Q511. Coarse current limits are built into the output stage.
A/D converter
The SourceMeter unit uses a multi-slope charge balance A/D converter with a single-slope
run-down. The converter is controlled by gate array U610. Commands are issued by the
MPU on the digital board through communications opto-isolators to U610, and U610
sends A/D reading data back through opto-isolators to the digital board for calibration and
processing.
Active guard
The Model 2440 has an active guard or “six-wire ohms” circuit used to measure complex
devices. This circuitry provides a low-current (50mA) equivalent of the voltage on output
HI. If the unit is in the SV mode, the low-current equivalent of the source voltage will
appear on the guard terminal. If the unit is in the SI mode, the voltage on output HI is
equal to the source current multiplied by the external resistance value. An equivalent volt
age will be generated by the guard circuit, and a guard sense terminal is provided to sense
around the voltage drop in the guard leads since significant current can flow (50mA).
-
-
Model 2440 5A SourceMeter® Service ManualTroubleshooting4-11
Digital circuitry
Refer to Figure 4-5 for the following discussion on digital circuitry.
The core digital circuitry uses a Motorola 68332 microcontroller running at 16.78MHz.
The memory configuration includes two 256K X 8-bit flash EEPROMs and two 128K X
8-bit RAMs used in parallel to utilize the 16-bit data bus of the MPU. The RAM is battery
backed-up, providing continued storage of data buffer information during power-down
cycles, and flash ROM support allows internal firmware upgrades using either the serial or
GPIB port for downloading new firmware. All calibration constants and the save 0 setup
are stored in a separate serial EEPROM. Setups 1 through 4 are stored in battery backedup RAM.
External communication is provided via GPIB and serial interfaces. A 9914 GPIB
IEEE-488 standard interface IC is used for the GPIB, and a 68332 Queued Serial Module
(QSM) provides the serial UART. For internal communications, the Time Processing Unit
(TPU) is used for serial communications with the front panel display module, and both the
TPU and QSM handle digital-to-analog interfacing.
Figure 4-5
Digital circuitry block diagram
A/D
Control/Data
Reset
2
E PROM
U17
A/D Interface
U9, U25
Voltage Source
Control
ROM
U15, U16
Microprocessor
U3
RAM
U12, U14
16.78MHz
Serial
Interface
U4
GPIB
U6, U13
U20
Trigger
Digital I/O
U23
U7
RS-232 Interface
IEEE-488 Interface
To Display
Board Controller
Trigger
Digital
I/O
Troubleshooting
4-12TroubleshootingModel 2440 5A SourceMeter® Service Manual
Display board circuit theory
U902 is the display microcontroller that controls the VFD (vacuum fluorescent display)
and interprets key data. The microcontroller has four peripheral I/O ports that are used for
the various control and read functions.
Display data is serially transmitted to the microcontroller from the digital board via the
TXB line to the microcontroller PD0 terminal. In a similar manner, key data is serially
sent back to the digital board through the RXB line via PD1. The 4MHz clock for the
microcontroller is generated on the digital board.
DS901 is the VFD (vacuum fluorescent display) module, which can display up to 49 characters. Each character is organized as a 5 × 7 matrix of dots or pixels and includes a long
under-bar segment to act as a cursor.
The display uses a common multiplexing scheme with each character refreshed in
sequence. U903 and U904 are the grid drivers, and U901 and U905 are the dot drivers.
Note that dot driver and grid driver data is serially transmitted from the microcontroller
(PD3 and PC1).
The front panel keys (S901-S931) are organized into a row-column matrix to minimize the
number of microcontroller peripheral lines required to read the keyboard. A key is read by
strobing the columns and reading all rows for each strobed column. Key down data is
interpreted by the display microcontroller and sent back to the main microprocessor using
proprietary encoding schemes.
Model 2440 5A SourceMeter® Service ManualTroubleshooting4-13
Troubleshooting
Troubleshooting information for the various circuits is summarized below.
Display board checks
If the front panel display tests indicate that there is a problem on the display board, use
Table 4-1. See “Principles of operation” for display circuit theory.
Table 4-1
Display board checks
StepItem/componentRequired conditionRemarks
1
Front panel test
2
J1033
3
U902, pin 1
4
U902, pin 43
5
U902, pin32
6
U902, pine 33
Verify that all segments
operate.
+5V, ±5%
Goes low briefly on power
up, and then goes high.
4MHz square wave.
Pulse train every 1 ms.
Brief pulse train when front
panel key is pressed.
Use front panel display test.
Digital +5V supply.
Microcontroller RESET.
Controller 4MHz clock.
Control from main processor.
Key down data sent to main
processor.
Power supply checks
Power supply problems can be checked out using Table 4-2. See “Principles of operation”
for circuit theory on the power supply.
Table 4-2
Power supply checks
StepItem/componentRequired conditionRemarks
1
Line fuse
2
Line power
3
TP502
4
TP503
5
TP504
6
TP505
7
TP507
8
TP508
9
TP510
Check continuity.
Plugged into live receptacle,
power on.
+58V, ±10%
-58V, ±10%
+21V, ±10%
-21V, ±10%
+15V, ±5%
-15V, ±5%
+5V, ±5%
Troubleshooting
Remove to check.
Check for correct power-up
sequence.
Referenced to TP501.
Referenced to TP501.
Referenced to TP501.
Referenced to TP501.
+15VF, referenced to TP500.
-15VF, referenced to TP500.
+5VF, referenced to TP500.
4-14TroubleshootingModel 2440 5A SourceMeter® Service Manual
Digital circuitry checks
Digital circuit problems can be checked out using Table 4-3. See “Principles of opera-
tion” for a digital circuit description.
Table 4-3
Digital circuitry checks
StepItem/componentRequired conditionRemarks
1
Power-on test
2
U3 pin 19
3
U3 pin 7
4
U3 pin 68
5
U3, lines A0-A19
6
U3, lines D0-D15
7
U3 pin 66
8
U4 pin 7
9
U4 pin 8
10
U13 pins 34-42
11
U13 pins 26-31
12
U13 pin 24
13
U13 pin 25
14
U3 pin 43
15
U3 pin 44
16
U3 pin 45
17
U3 pin 47
RAM OK, ROM OK.
Digital common.
+5V
Low on power-up, then goes
high.
Check for stuck bits.
Check for stuck bits.
16.78MHz.
Pulse train during RS-232
I/O.
Pulse train during RS-232
I/O.
Pulse train during IEEE-488
I/O.
Pulses during IEEE-488 I/O.
Low with remote enabled.
Low during interface clear.
Pulse train.
Pulse train.
Pulse train.
Pulse train
Verify that RAM and ROM
are functional.
All signals referenced to
digital common.
Digital logic supply.
MPU RESET line.
Model 2440 5A SourceMeter® Service ManualTroubleshooting4-15
Analog circuitry checks
Table 4-4 summarizes analog circuitry checks.
Table 4-4
Analog circuitry checks
StepItem/componentRequired conditionRemarks
1
TP200*
2
TP201*
3
TP202*
4
TP203*
5
TP213*
6
TP218**
7
TP219**
8
TP214**
9
TP232*
**Measured with respect to FCOM (TP500).
**Measured with respect to OCOM (TP501).
>60V voltage protection
SOURCE +10V
SOURCE +10V (SVMI)
SOURCE +10V
SOURCE +10V
OUTPUT COM
OUTPUT COM
SVMI, OUTPUT ON, 20V
Bench defaults
-13V ±1V
-5V ±0.5V
-10V ±1V
-10.5V ±1V
0V ±0.1V
7V ±0.7V
7V ±0.7V
20V ±0.5V
6.4V ±0.6V
Battery replacement
WARNINGDisconnect the instrument from the power line and all other equip-
ment before changing the battery.
The volatile memories of the Model 2440 are protected by a replaceable battery when
power is off. Typical battery life is approximately 10 years, but the battery should be
replaced if the voltage drops below 2.5V regardless of age. The battery should be sus
pected if the instrument no longer retains buffer data or user-defined operating parameters
such as instrument setups, source memory, and math expressions. If the battery is absent
or totally exhausted, the display will show the “Reading buffer data lost” message shortly
after the Model 2440 is switched on.
The battery is a 3V wafer-type lithium cell, (Keithley part number BA-46), which is
located on the digital board. Replacement of the battery requires removal of the case cover
and analog board assembly. (See
NOTECalibration constants and user defined parameters will be lost when the battery
is replaced. The Model 2440 must be re-calibrated (Section 2) after the battery
is replaced.
-
Troubleshooting
Section 5.)
4-16TroubleshootingModel 2440 5A SourceMeter® Service Manual
Battery replacement precautions
WARNINGThe precautions below must be followed to avoid personal injury.
1.Wear safety glasses or goggles when working with lithium batteries.
2.Do not short the battery terminals together.
3.Keep lithium batteries away from all liquids.
4.Do not attempt to recharge lithium batteries.
5.Observe proper polarity when installing the battery.
6.Do not incinerate or otherwise expose the battery to excessive heat (>60°C).
7.Bulk quantities of lithium batteries should be disposed of as hazardous waste.
Battery replacement procedure
1.Remove the case cover and analog board assembly as covered in Section 5.
2.Locate the battery on the digital board.
3.Carefully unsolder and remove the old battery.
4.Install and solder the new battery in place.
5.Reassemble the instrument, and turn it on. The “Reading buffer data lost” error
message will be displayed.
6.Send the :SYST:MEM:INIT command via remote to perform the following:
•Clear the reading buffer.
•Initialize instrument setups 1 through 4 to present instrument settings.
•Initialize all 100 source memory locations to present instrument settings.
•Delete user math expressions.
Model 2440 5A SourceMeter® Service ManualTroubleshooting4-17
No comm link error
A “No Comm Link” error indicates that the front panel processor has stopped communicating with the main processor, which is located on the digital board. This error indicates
that one of the main processor ROMs may require re-seating in its socket. ROMs may be
reseated as follows:
1.Turn off the power, and disconnect the line cord and all other test leads and cables
from the instrument.
2.Remove the case cover as outlined in Section 5.
3.Remove the analog board assembly as outlined in Section 5.
4.Locate the two firmware ROMS, U15 and U16, located on the digital board. These
are the only ICs installed in sockets. (Refer to the component layout drawing at the
end of
Section 6 for exact locations.)
5.Carefully push down on each ROM IC to make sure it is properly seated in its
socket.
CAUTIONBe careful not to push down excessively, or you might crack the digital
board.
6.Connect the line cord, and turn on the power. If the problem persists, additional
troubleshooting will be required.
Troubleshooting
4-18TroubleshootingModel 2440 5A SourceMeter® Service Manual
5
Disassembly
Disassembly
5-2DisassemblyModel 2440 5A SourceMeter® Service Manual
Introduction
This section explains how to handle, clean, and disassemble the Model 2440. Disassembly
drawings are located at the end of this section.
Handling and cleaning
To avoid contaminating PC board traces with body oil or other foreign matter, avoid
touching the PC board traces while you are repairing the instrument. Motherboard areas
covered by the shield have high-impedance devices or sensitive circuitry where contami
nation could cause degraded performance.
Handling PC boards
Observe the following precautions when handling PC boards:
•Wear cotton gloves.
•Only handle PC boards by the edges and shields.
•Do not touch any board traces or components not associated with repair.
•Do not touch areas adjacent to electrical contacts.
•Use dry nitrogen gas to clean dust off PC boards.
-
Solder repairs
Observe the following precautions when you must solder a circuit board:
•Use an OA-based (organic activated) flux, and take care not to spread the flux to
other areas of the circuit board.
•Remove the flux from the work area when you have finished the repair by using
pure water with clean, foam-tipped swabs or a clean, soft brush.
•Once you have removed the flux, swab only the repair area with methanol, then
blow-dry the board with dry nitrogen gas.
•After cleaning, allow the board to dry in a 50°C, low-humidity environment for
several hours.
Model 2440 5A SourceMeter® Service ManualDisassembly5-3
Static sensitive devices
CMOS devices operate at very high impedance levels. Therefore, any static that builds up
on you or your clothing may be sufficient to destroy these devices if they are not handled
properly. Use the following precautions to avoid damaging them:
CAUTIONMany CMOS devices are installed in the Model 2440. Handle all semi-
conductor devices as being static sensitive.
•Transport and handle ICs only in containers specially designed to
prevent static build-up. Typically, you will receive these parts in
anti-static containers made of plastic or foam. Keep these devices
in their original containers until ready for installation.
•Remove the devices from their protective containers only at a
properly grounded work station. Also, ground yourself with a
suitable wrist strap.
•Handle the devices only by the body; do not touch the pins.
•Ground any printed circuit board into which a semiconductor
device is to be inserted to the bench or table.
•Use only anti-static type desoldering tools.
•Use only grounded-tip solder irons.
•Once the device is installed in the PC board, it is normally ade-
quately protected, and you can handle the boards normally.
Disassembly
Assembly drawings
Use the assembly drawings located at the end of this section to assist you as you disassemble and re-assemble the Model 2440. Also, refer to these drawings for information about
the Keithley part numbers of most mechanical parts in the unit. Assembly drawings
include:
5-4DisassemblyModel 2440 5A SourceMeter® Service Manual
Case cover removal
Follow the steps below to remove the case cover to gain access to internal parts.
WARNINGBefore removing the case cover, disconnect the line cord and any test
leads from the instrument.
1.Remove handle — The handle serves as an adjustable tilt-bail. Adjust its position
by gently pulling it away from the sides of the instrument case and swinging it up
or down. To remove the handle, swing the handle below the bottom surface of the
case and back until the orientation arrows on the handles line up with the orienta
tion arrows on the mounting ears. With the arrows lined up, pull the ends of the
handle away from the case.
2.Remove mounting ears — Remove the screw that secures each mounting ear. Pull
down and out on each mounting ear.
NOTEWhen reinstalling the mounting ears, make sure to mount the right ear to the
right side of the chassis, and the left ear to the left side of the chassis. Each ear
is marked “RIGHT” or “LEFT” on its inside surface.
-
3.Remove rear bezel — To remove the rear bezel, loosen the two screws that secure
the rear bezel to the chassis, then pull the bezel away from the case.
4.Remove grounding screws — Remove the two grounding screws that secure the
case to the chassis. They are located on the bottom of the case at the back.
5.Remove chassis — To remove the case, grasp the front bezel of the instrument,
and carefully slide the chassis forward. Slide the chassis out of the metal case.
NOTETo gain access to the components under the analog board shield, remove the
shield, which is secured to the analog board by a single screw.
Model 2440 5A SourceMeter® Service ManualDisassembly5-5
Analog board removal
Perform the following steps to remove the analog board. This procedure assumes that the
case cover is already removed.
1.Disconnect the front and rear input terminals.
You must disconnect these input terminal connections for both the front and rear
inputs:
•INPUT/OUTPUT HI and LO
•4-WIRE SENSE HI and LO
•V, Ω, GUARD and GUARD SENSE (rear panel only)
Remove all the connections by pulling the wires off the pin connectors, then
remove the ferrite noise filters from the chassis. During re-assembly, use the fol
lowing table to identify input terminals:
Input terminalsFront wire color Rear wire color
INPUT/OUTPUT HI
INPUT/OUTPUT LO
4-WIRE SENSE HI
4-WIRE SENSE LO
V, Ω, GUARD
GUARD SENSE
2.Unplug cables.
•Carefully unplug the ribbon cables at J1027, J1028, and J1029.
•Unplug the ON/OFF cable at J1034.
3.Remove screws.
•Remove the two fastening screws that secure the analog board assembly to the
chassis. These screws are located on the side of the board opposite from the
heat sink.
•Remove the two screws that secure the heat sink to the chassis.
4.Remove analog board assembly.
After all screws have been removed, carefully lift the analog board assembly free
of the main chassis.
5.Disassemble analog board assembly.
•Remove the screws that secure the analog board and heat sink to the analog
board subchassis.
•Carefully remove the heat sink by sliding the clips off the power transistors.
Red
Black
Yellow
Gray
-
-
White/Red
White/Black
White/Yellow
White/Gray
White
Blue/White
-
Disassembly
CAUTIONBe careful not to damage the heat sink insulation layer.
•Remove the analog board from the subchassis.
•Remove the four screws that secure the bottom cover, then remove the cover
from the bottom of the PC board.
NOTEWhen reinstalling the heat sink, make sure that all clips are properly installed
and centered on each pair of output transistors.
5-6DisassemblyModel 2440 5A SourceMeter® Service Manual
Digital board removal
Perform the following steps to remove the digital board. This procedure assumes that the
analog board assembly is already removed.
NOTEIn order to remove the digital board, the display board must first be removed.
1.Remove the IEEE-488, Digital I/O, and RS-232 fasteners.
The IEEE-488, Digital I/O, and RS-232 connectors each have two nuts that secure
the connectors to the rear panel. Remove these nuts.
2.Remove the POWER switch rod.
At the switch, place the edge of a flat-blade screw driver in the notch on the pushrod. Gently twist the screw driver while pulling the rod from the shaft.
3.Unplug cables:
•Unplug the display board ribbon cable.
•Unplug the cables going to the power supply.
•Unplug the rear panel power module cable.
•The fan may need to be removed.
4.Remove digital board.
Slide the digital board forward until it is free of the guide pins, then remove the
board.
During reassembly, replace the board, and start the IEEE-488, Digital I/O, and RS232 connector nuts and the mounting screw. Tighten all the fasteners once they are
all in place and the board is correctly aligned.
Front panel disassembly
Use the following procedures to remove the display board and/or the pushbutton switch
pad.
1.Unplug the display board ribbon cables.
2.Remove the front panel assembly.
This assembly has four retaining clips that snap onto the chassis over four pem nut
studs. Two retaining clips are located on each side of the front panel. Pull the
retaining clips outward and, at the same time, pull the front panel assembly for
ward until it separates from the chassis.
3.Using a thin-bladed screw driver, pry the plastic PC board stop (located at the bottom of the display board) until the bar separates from the casing. Pull the display
board from the front panel.
4.Remove the switch pad by pulling it from the front panel.
-
Model 2440 5A SourceMeter® Service ManualDisassembly5-7
Removing power components
The following procedures to remove the power supply and/or power module require that
the case cover and motherboard be removed, as previously explained.
Power module removal
Perform the following steps to remove the rear panel power module:
1.Remove the analog board.
2.Unplug the cable connecting the power module to the digital board.
3.Disconnect the power module's ground wire. This green and yellow wire connects
to a threaded stud on the chassis with a kep nut.
4.Squeeze the latches on either side of the power module while pushing the module
from the access hole.
WARNINGTo avoid electrical shock, which could result in injury or death, the
ground wire of the power module must be connected to chassis
ground. When installing the power module, be sure to reconnect the
green and yellow ground wire to the threaded stud on the chassis.
Instrument reassembly
Reassemble the instrument by reversing the previous disassembly procedures. Make sure
that all parts are properly seated and secured, and that all connections are properly made.
To ensure proper operation, replace the analog signal wire ferrite noise filters, and
securely fasten the shield.
Disassembly
WARNINGTo ensure continued protection against electrical shock, verify that
power line ground (green and yellow wire attached to the power mod
ule) is connected to the chassis.
Also make sure the two bottom case screws are properly installed to
secure and ground the case cover to the chassis.
-
5-8DisassemblyModel 2440 5A SourceMeter® Service Manual
6
Replaceable Parts
Replaceable Parts
6-2Replaceable PartsModel 2440 5A SourceMeter® Service Manual
Introduction
This section contains replacement parts information and component layout drawings for
the Model 2440.
Parts lists
The electrical parts lists for the Model 2440 are shown tables at the end of this section. For
part numbers to the various mechanical parts and assemblies, use the Miscellaneous parts
list and the assembly drawings provided at the end of
Section 5.
Ordering information
To place an order, or to obtain information concerning replacement parts, contact your
Keithley representative or the factory (see inside front cover for addresses). When order
ing parts, be sure to include the following information:
•Instrument model number (Model 2440)
•Instrument serial number
•Part description
•Component designation (if applicable)
•Keithley part number
-
Factory service
If the instrument is to be returned to Keithley Instruments for repair, perform the
following:
•Call the Repair Department at 1-800-552-1115 for a Return Material Authorization
(RMA) number.
•Complete the service form at the back of this manual, and include it with the instrument.
•Carefully pack the instrument in the original packing carton.
•Write ATTENTION REPAIR DEPARTMENT and the RMA number on the shipping label.
Model 2440 5A SourceMeter® Service ManualReplaceable Parts6-3
Component layouts
The component layouts for the various circuit boards are provided on the following pages.
Drawings include: