Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship
for a period of one (1) year from date of shipment.
Keithley Instruments, Inc. warran
probes, cables, software, rechargeable batteries, diskettes, and documentation.
During the warranty period, Keithley Instruments will, at its option, ei
proves to be defective.
To exercise this warranty, write or call your local Keithley Instruments representative, or contact
thley Instruments headquarters in Cleveland, Ohio. You will be given prompt assistance and
Kei
return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs
will be made and the product returned, transportation prepaid. Repaired or replaced products are
warranted for the balance of the original warranty period, or at least 90 days.
ts the following items for 90 days from the date of shipment:
ther repair or replace any product that
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley
Instruments’ express written consent, or misuse of any product or part. This warranty also does not
apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems
arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER W
INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE THE BUYER’S SOLE AND
EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY
FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES
ING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE, EVEN IF KEITHLEY
ARIS
INSTRUMENTS, INC. HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH
DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COST
OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY
PERSON, OR DAMAGE TO PROPERTY.
ARRANTIES, EXPRESSED OR IMPLIED,
OF ITS EMPLOYEES SHALL BE LIABLE
A G R E A T E R M E A S U R E O F C O N F I D E N C E
The print history shown below lists the printing dates of all Revisions and Addenda created
for this manual. The Revision Level letter increases alphabetically as the manual undergoes sub
sequent updates. Addenda, which are released between Revisions, contain important change information that the user should incorporate immediately into the manual. Addenda are numbered
sequentially. When a new Revision is created, all Addenda associated with the previous Revi
sion of the manual are incorporated into the new Revision of the manual. Each new Revision
includes a revised copy of this print history page.
Revision A (Document Number 2440-902-01)................................................................ April 2001
Revision B (Document Number 2440-902-01)........................................................Decmeber 2004
-
-
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc.
Other brand names are trademarks or registered trademarks of their respective holders.
Safety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation.
Although some instruments and accessories would normally be used with non-hazardous voltages, there are
situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety
precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance
information carefully before using the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring
that the equipment is operated within its specifications and operating limits, and for ensuring that operators are
adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper
use of the instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example,
setting the line voltage or replacing consumable materials. Maintenance procedures are described in the user
documentation. The procedures explicitly state if the operator may perform them. Otherwise, they should be
performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly
trained service personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are rated Measurement Category I
and Measurement Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC
60664. Most measurement, control, and data I/O signals are Measurement Category I and must not be directly
connected to mains voltage or to voltage sources with high transient over-voltages. Measurement Category II
connections require protection for high transient over-voltages often associated with local AC mains connections.
Assume all measurement, control, and data I/O connections are for connection to Category I sources unless
otherwise marked or described in the user documentation.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks
or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage
levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect that hazardous
voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that
operators are prevented access and/or insulated from every connection point. In some cases, connections must be
exposed to potential human contact. Product operators in these circumstances must be trained to protect
themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive
part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedancelimited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards,
install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power recep tacle.
Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate
main input power disconnect device must be provided in close proximity to the equipment and within easy reach of
the operator.
04/09
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the
!
circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before:
connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes,
such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line
(earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of
withstanding the voltage being measured.
The instrument and accessories must be used in accordance with specifications and operating instructions, or the
safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and
operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard.
Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground
connections.
If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation
requires the use of a lid interlock.
If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation.
The symbol on an instrument means caution, risk of danger. The user should refer to the operating instructions
located in the user documenta ti on in all case s where the symbol is marked on the instrument.
The symbol on an instrument means caution, risk of danger. User standard safety precautions to avoid
personal contact with these voltages.
The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The symbol indicates a connection terminal to the equipment frame.
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp
must be properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death.
Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such
damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test cables.
To maintain protection from electric shock and fire, rep lacement components in mains circuits - including the power
transformer, test leads, and input jacks - must be purchased from Keithley Instruments. Standard fuses with
applicable national safety approvals may be used if the rating and type are the same. Other components that are
not safety-related may be purchased from other suppliers as long as they are equivalent to the original component
(note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and
functionality of the product). If you are unsure about the applicability of a replacement component, call a Keithley
Instruments office for information.
T o clean an instrument, use a damp cloth or mild, water-based cleaner . Clean the exterior of the instrument only . Do
not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist
of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never
require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected,
the board should be returned to the factory for proper cleaning/servicing.
1-2Performance VerificationModel 2440 5A SourceMeter® Service Manual
Introduction
Use the procedures in this section to verify that Model 2440 accuracy is within the limits
stated in the instrument’s one-year accuracy specifications. You can perform these verifi
cation procedures:
•When you first receive the instrument to make sure that it was not damaged during
shipment.
•To verify that the unit meets factory specifications.
•To determine if calibration is required.
•Following calibration to make sure it was performed properly.
WARNINGThe information in this section is intended for qualified service per-
sonnel only. Do not attempt these procedures unless you are qualified
to do so. Some of these procedures may expose you to hazardous volt
ages, which could cause personal injury or death if contacted. Use
standard safety precautions when working with hazardous voltages.
NOTEIf the instrument is still under warranty and its performance is outside specified
limits, contact your Keithley representative or the factory to determine the cor
rect course of action.
-
-
-
Verification test requirements
Be sure that you perform the verification tests:
•Under the proper environmental conditions.
•After the specified warm-up period.
•Using the correct line voltage.
•Using the proper test equipment.
•Using the specified output signals and reading limits.
Environmental conditions
Conduct your performance verification procedures in a test environment with:
•An ambient temperature of 18-28°C (65-82°F).
•A relative humidity of less than 70% unless otherwise noted.
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-3
Warm-up period
Allow the Model 2440 to warm up for at least one hour before conducting the verification
procedures.
If the instrument has been subjected to temperature extremes (those outside the ranges
stated above), allow additional time for the instrument’s internal temperature to stabilize.
Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified
temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manufacturer.
Line power
The Model 2440 requires a line voltage of 100 to 240V and a line frequency of 50 to
60Hz. Verification tests must be performed within this range.
Recommended test equipment
Table 1-1 summarizes recommended verification equipment. You can use alternate equip-
ment as long as that equipment has specifications at least as good as those listed in Table
1-1. Keep in mind, however, that test equipment uncertainty will add to the uncertainty of
each measurement. Generally, test equipment uncertainty should be at least four times better than corresponding Model 2440 specifications. Table 1-1 lists the specifications of the
recommended test equipment, including maximum allowable uncertainty for alternate test
equipment, which is shown in parentheses.
Performance Verification
1-4Performance VerificationModel 2440 5A SourceMeter® Service Manual
* 90-day, full-range accuracy specifications of ranges required for various measurement points.
** 90-day, ±5°C specifications of nominal resistance values shown. Use actual values for tests. Maximum uncertainty of alternate
test equipment shown in parentheses.
*** Required for verification of 5A current range. Characterize resistor to ±300ppm or better using recommended DMM before ver-
ifying 5A current measurement range.
1Ω resistor characterization
The recommended 1Ω resistor should be characterized to ±300ppm or better before verifying the 5A current measurement range. (You need not characterize the resistor if you are
checking only the 5A current source range.) Use the 4-wire ohms function of the DMM
recommended in
value to calculate the current during the 5A current measurement range test procedure.
Table 1-1 to measure the resistance value, and then use that measured
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-5
Verification limits
The verification limits stated in this section have been calculated using only the Model
2440 one-year accuracy specifications, and they do not include test equipment uncer
tainty. If a particular measurement falls outside the allowable range, recalculate new limits based on Model 2440 specifications and corresponding test equipment specifications.
Example limits calculation
As an example of how verification limits are calculated, assume you are testing the 10V
DC output range using a 10V output value. Using the Model 2440 10V range one-year
accuracy specification of ±(0.02% of output + 1.2mV offset), the calculated output limits
are:
When verifying the resistance measurement accuracy, it will probably be necessary to
recalculate resistance limits based on the actual calibrator resistance values. You can cal
culate resistance reading limits in the same manner described above, but be sure to use the
actual calibrator resistance values and the Model 2440 normal accuracy specifications for
your calculations.
As an example, assume you are testing the 20kΩ range, and that the actual value of the
nominal 19kΩ calibrator resistor is 19.01kΩ. Using the Model 2440 20kΩ range one-year
normal accuracy specifications of ±(0.06% of reading + 3Ω), the recalculated reading
limits are:
7.Press ENTER then EXIT as necessary to return to normal display.
Performing the verification test procedures
Test summary
•DC voltage output accuracy
•DC voltage measurement accuracy
•DC current output accuracy
•DC current measurement accuracy
•Resistance measurement accuracy
If the Model 2440 is not within specifications and not under warranty, see the calibration
procedures in
Section 2 for information on calibrating the unit.
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-7
Test considerations
When performing the verification procedures:
•Be sure to restore factory front panel defaults as previously outlined.
•Make sure that the test equipment is properly warmed up and connected to the
Model 2440 INPUT/OUTPUT jacks. Also be sure that the front panel jacks are
selected with the TERMINALS key.
•Make sure the Model 2440 is set to the correct source range (see below).
•Be sure that the Model 2440 output is turned on before making measurements.
•Be sure the test equipment is set up for the proper function and range.
•Allow the Model 2440 output signal to settle before making a measurement.
•Do not connect test equipment to the Model 2440 through a scanner, multiplexer,
or other switching equipment.
WARNINGThe maximum common-mode voltage (voltage between LO and chas-
CAUTIONThe maximum voltage between INPUT/OUTPUT HI and LO or 4-
Performance Verification
sis ground) is 40V peak. Exceeding this value may cause a breakdown
in insulation, creating a shock hazard.
WIRE SENSE HI and LO is 42V peak. The maximum voltage between
INPUT/OUTPUT HI and 4-WIRE SENSE HI or between INPUT/
OUTPUT LO and 4-WIRE SENSE LO is 5V. Exceeding these voltage
values may result in instrument damage.
Setting the source range and output value
Before testing each verification point, you must properly set the source range and output
value as outlined below.
1.Press either the SOURCE V or SOURCE I key to select the appropriate source
function.
2.Press the EDIT key as required to select the source display field. Note that the cursor will flash in the source field while its value is being edited.
3.With the cursor in the source display field flashing, set the source range to the lowest possible range for the value to be sourced using the up or down RANGE key.
For example, you should use the 10V source range to output a 10V source value.
With a 10V source value and the 10V range selected, the source field display will
appear as follows:
Vsrc:+10.0000V
1-8Performance VerificationModel 2440 5A SourceMeter® Service Manual
4.With the source field cursor flashing, set the source output to the required value
using either:
•The SOURCE adjustment and left and right arrow keys.
•The numeric keys.
5.Note that the source output value will be updated immediately; you need not press
ENTER when setting the source value.
Setting the measurement range
When simultaneously sourcing and measuring either voltage or current, the measure range
is coupled to the source range, and you cannot independently control the measure range.
Thus, it is not necessary for you to set the measure range when testing voltage or current
measurement accuracy.
Compliance considerations
Compliance limits
When sourcing voltage, you can set the SourceMeter to limit current from 10nA to 5.25A.
Conversely, when sourcing current, you can set the SourceMeter to limit voltage from
0.2mV to 42V. The SourceMeter output will not exceed the programmed compliance
limit.
Types of compliance
There are two types of compliance that can occur: “real” and “range.” Depending on
which value is lower, the output will clamp at either the displayed compliance setting
(“real”) or at the maximum measurement range reading (“range”).
The “real” compliance condition can occur when the compliance setting is less than the
highest possible reading of the measurement range. When in compliance, the source out
put clamps at the displayed compliance value. For example, if the compliance voltage is
set to 1V and the measurement range is 2V, the output voltage will clamp (limit) at 1V.
“Range” compliance can occur when the compliance setting is higher than the possible
reading of the selected measurement range. When in compliance, the source output
clamps at the maximum measurement range reading (not the compliance value). For
example, if the compliance voltage is set to 1V and the measurement range is 200mV, the
output voltage will clamp (limit) at 210mV.
-
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-9
Maximum compliance values
The maximum compliance values for the measurement ranges are summarized in
Table 1-2.
Table 1-2
Maximum compliance values
Maximum
Measurement
range
200mV
2V
10V
40V
10μA
100μA
1mA
10mA
100mA
1A
5A
compliance
value
210mV
2.1V
10.5V
42V
10.5μA
105μA
1.05mA
10.5mA
105mA
1.05A
5.25A
Performance Verification
When the SourceMeter goes into compliance, the “Cmpl” label or the units label (i.e.,
“mA”) for the compliance display will flash.
Determining compliance limit
The relationships to determine which compliance is in effect are summarized as follows.
They assume that the measurement function is the same as the compliance function.
•Compliance Setting < Measurement Range = Real Compliance
•Measurement Range < Compliance Setting = Range Compliance
You can determine the compliance that is in effect by comparing the displayed compliance setting to the present measurement range. If the compliance setting is lower than the
maximum possible reading on the present measurement range, the compliance setting is
the compliance limit. If the compliance setting is higher than the measurement range, the
maximum reading on that measurement range is the compliance limit.
1-10Performance VerificationModel 2440 5A SourceMeter® Service Manual
Taking the SourceMeter out of compliance
Verification measurements should not be made when the SourceMeter is in compliance.
For purposes of the verification tests, the SourceMeter can be taken out of compliance by
going into the edit mode and increasing the compliance limit.
NOTEDo not take the unit out of compliance by decreasing the source value or chang-
ing the range. Always use the recommended range and source settings when
performing the verification tests.
Output voltage accuracy
Follow the steps below to verify that Model 2440 output voltage accuracy is within specified limits. This test involves setting the output voltage to each full-range value and measuring the voltages with a precision digital multimeter.
1.With the power off, connect the digital multimeter to the Model 2440 INPUT/
OUTPUT jacks, as shown in
Figure 1-1
Connections for voltage verification tests
Figure 1-1.
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
PEAK
SWEEP
SOURCE
I
5
2440 5A SourceMeter
EDIT
EXIT ENTER
MEAS
FCTN
I
Ω
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
1
REL
RANGE
AUTO
RANGE
®
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V
PEAK
TERMINALS
FRONT/
REAR
Model 2440
Input HI
Digital Multimeter
Input LO
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-11
2.Select the multimeter DC volts measuring function.
3.Set the voltage source protection to NONE. To do so, press CONFIG then
SOURCE V to access the CONFIGURE V-SOURCE menu. Select PROTEC
TION, and set the voltage source protection limit to NONE.
4.Press the Model 2440 SOURCE V key to source voltage, and make sure the source
output is turned on.
5.Verify output voltage accuracy for each of the voltages listed in Table 1-3. For
each test point:
•Select the correct source range.
•Set the Model 2440 output voltage to the indicated value.
•Verify that the multimeter reading is within the limits given in the table.
6.Repeat the procedure for negative output voltages with the same magnitudes as
those listed in
7.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-3
Output voltage accuracy limits
Model 2440
source range
200mV
2V
10V
40V
Table 1-3.
Model 2440 output
voltage setting
200.000mV
2.00000V
10.0000V
40.0000V
Output voltage limits
(1 year, 18°C-28°C)
199.360 to 200.640mV
1.99900 to 2.00100V
9.9968 to 10.0032V
39.9872 to 40.0128V
Performance Verification
-
1-12Performance VerificationModel 2440 5A SourceMeter® Service Manual
Voltage measurement accuracy
Follow the steps below to verify that Model 2440 voltage measurement accuracy is within
specified limits. The test involves setting the source voltage to full-range values, as mea
sured by a precision digital multimeter, and then verifying that the Model 2440 voltage
readings are within required limits.
1.With the power off, connect the digital multimeter to the Model 2440 INPUT/
OUTPUT jacks , as shown in
2.Select the multimeter DC volts function.
3.Set the voltage source protection to NONE. To do so, press CONFIG then
SOURCE V to access the CONFIGURE V-SOURCE menu. Select PROTEC
TION, and set the voltage source protection limit to NONE.
4.Set the Model 2440 to both source and measure voltage by pressing the SOURCE
V and MEAS V keys, and make sure the source output is turned on.
5.Verify output voltage accuracy for each of the voltages listed in Table 1-4. For
each test point:
•Select the correct source range.
•Set the Model 2440 output voltage to the indicated value as measured by the
digital multimeter.
•Verify that the Model 2440 voltage reading is within the limits given in the
table.
Figure 1-1.
-
-
NOTEIt may not be possible to set the voltage source to the specified value. Use the
closest possible setting, and modify reading limits accordingly.
6.Repeat the procedure for negative source voltages with the same magnitudes as
those listed in
7.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-4
Voltage measurement accuracy limits
Model 2440 source
and measure
range*
200mV
2V
10V
40V
Table 1-4.
Source voltage*
200.000mV
10.0000V
40.0000V
2.00000V
Model 2440 voltage reading
limits (1 year, 18°C-28°C)
199.676 to 200.324mV
1.99946 to 2.00054V
9.9978 to 10.0022V
39.991 to 40.009V
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-13
Table 1-4
Voltage measurement accuracy limits
Model 2440 source
and measure
range*Source voltage*
* Measure range coupled to source range when simultaneously sourcing and measuring voltage.
** As measured by precision digital multimeter. Use closest possible value, and modify reading lim-
its accordingly if necessary.
Output current accuracy
Follow the steps below to verify that Model 2440 output current accuracy is within specified limits. The test involves setting the output current to each full-range value and measuring the currents with a precision digital multimeter.
10μA to 1A range accuracy
1.With the power off, connect the digital multimeter to the Model 2440 INPUT/
OUTPUT jacks, as shown in
Figure 1-2
Connections for 10μA to 1A range current verification tests
V
LOCAL
67
DIGITS SPEED
MEAS
I
Ω
1
230
REL
FILTER
89
RECALL
STORE
EDIT
DISPLAY
TOGGLE
POWER
Performance Verification
Model 2440 voltage reading
limits (1 year, 18°C-28°C)
Figure 1-2.
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
PEAK
RANGE
AUTO
RANGE
®
2440 5A SourceMeter
V
4
TRIG
SWEEP
+/-
CONFIG MENU
I
5
SOURCE
EDIT
EXIT ENTER
FCTN
LIMIT
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V
PEAK
TERMINALS
FRONT/
REAR
Model 2440
Digital Multimeter
2.Select the multimeter DC current measuring function.
Input LO
Amps
1-14Performance VerificationModel 2440 5A SourceMeter® Service Manual
3.Press the Model 2440 SOURCE I key to source current, and make sure the source
output is turned on.
4.Verify output current accuracy for the 10μA-1A range currents listed in Table 1-5.
For each test point:
•Select the correct source range.
•Set the Model 2440 output current to the correct value.
•Verify that the multimeter reading is within the limits given in the table.
5.Repeat the procedure for negative output currents with the same magnitudes as
those listed in
Table 1-5.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
5A range accuracy
1.With the power off, connect the digital multimeter and the 1Ω resistor to the Model
2440 INPUT/OUTPUT jacks, as shown in
Figure 1-3
Connections for 5A range current verification tests
SWEEP
I
5
SOURCE
2440 5A SourceMeter
EDIT
EXIT ENTER
RANGE
AUTO
RANGE
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
MEAS
FCTN
I
Ω
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
1
REL
4- WIRE
SENSE
HI
75V
5V
PEAK
PEAK
®
LO
ON/OFF
OUTPUT
INPUT/
OUTPUT
TERMINALS
FRONT/
REAR
Figure 1-3.
75V
PEAK
250V
PEAK
Model 2440
Digital Multimeter
1Ω Resistor
Input HI
Input LO
Model 2440 5A SourceMeter® Service ManualPerformance Verification1-15
2.Select the multimeter DC volts measuring function.
3.Press the Model 2440 SOURCE I key to source current, and make sure the source
output is turned on.
4.Verify output current accuracy for the 5A range. Be sure to:
•Select the 5A source range.
•Set the Model 2440 output current to the correct 5A output value.
•Verify that the multimeter reading is within the 5A range limits given in Table
1-5.
(Since the value of the 1Ω resistor value is assumed to be the same as its nominal
value, the DMM voltage reading is the same as the sourced current.)
5.Repeat the procedure for a negative 5A current output value.
6.Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-5
Output current accuracy limits
Model 2440
source range
10μA
100μA
1mA
10mA
100mA
1A
5A*
Model 2440
output current setting
10.0000μA
100.000μA
1.00000mA
10.0000mA
100.000mA
1.00000A
5.00000A
Output current limits
(1 year, 18°C-28°C)
9.9947 to 10.0053μA
99.949 to 100.051μA
0.99946 to 1.00054mA
9.9935 to 10.0065mA
99.914 to 100.086mA
0.99843 to 1.00157A
4.9896 to 5.0104A
Performance Verification
*See separate procedure for 5A range. DMM voltage reading is same as sourced current.
Loading...
+ 114 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.