Tektronix 2440 Service Manual

Model 2440
Service Manual
2440-902-01 Rev. B / December 2004
www.keithley.com
G
A
T
E
E
A
R
AEM
R
RUS
C F
O
IF
NO
ECNED

WARRANTY

Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of one (1) year from date of shipment.
Keithley Instruments, Inc. warran probes, cables, software, rechargeable batteries, diskettes, and documentation.
During the warranty period, Keithley Instruments will, at its option, ei proves to be defective.
To exercise this warranty, write or call your local Keithley Instruments representative, or contact
thley Instruments headquarters in Cleveland, Ohio. You will be given prompt assistance and
Kei return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
ts the following items for 90 days from the date of shipment:
ther repair or replace any product that

LIMITATION OF WARRANTY

This warranty does not apply to defects resulting from product modification without Keithley Instruments’ express written consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER W INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE THE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES
ING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE, EVEN IF KEITHLEY
ARIS INSTRUMENTS, INC. HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COST OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
ARRANTIES, EXPRESSED OR IMPLIED,
OF ITS EMPLOYEES SHALL BE LIABLE

A G R E A T E R M E A S U R E O F C O N F I D E N C E

Keithley Instruments, Inc.
Corporate Headquarters • 28775 Aurora Road • Cleveland, Ohio 44139
440-248-0400 • Fax: 440-248-6168 • 1-888-KEITHLEY (1-888-534-8453) • www.keithley.com
3/07
Model 2440 5A SourceMeter
Service Manual
®
©2001, Keithley Instruments, Inc.
All rights reserved.
Cleveland, Ohio, U.S.A.
Second Printing, December 2004
Document Number: 2440-902-01 Rev. B

Manual Print History

The print history shown below lists the printing dates of all Revisions and Addenda created for this manual. The Revision Level letter increases alphabetically as the manual undergoes sub sequent updates. Addenda, which are released between Revisions, contain important change in­formation that the user should incorporate immediately into the manual. Addenda are numbered sequentially. When a new Revision is created, all Addenda associated with the previous Revi sion of the manual are incorporated into the new Revision of the manual. Each new Revision includes a revised copy of this print history page.
Revision A (Document Number 2440-902-01)................................................................ April 2001
Revision B (Document Number 2440-902-01)........................................................Decmeber 2004
-
-
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc. Other brand names are trademarks or registered trademarks of their respective holders.
Safety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired. The types of product users are: Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring
that the equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained service personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are rated Measurement Category I and Measurement Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC
60664. Most measurement, control, and data I/O signals are Measurement Category I and must not be directly connected to mains voltage or to voltage sources with high transient over-voltages. Measurement Category II connections require protection for high transient over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O connections are for connection to Category I sources unless otherwise marked or described in the user documentation.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance­limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power recep tacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator.
04/09
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the
!
circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with specifications and operating instructions, or the safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard. Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground
connections. If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation
requires the use of a lid interlock. If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation. The symbol on an instrument means caution, risk of danger. The user should refer to the operating instructions
located in the user documenta ti on in all case s where the symbol is marked on the instrument. The symbol on an instrument means caution, risk of danger. User standard safety precautions to avoid
personal contact with these voltages. The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The symbol indicates a connection terminal to the equipment frame.
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans. Before performing any maintenance, disconnect the line cord and all test cables. To maintain protection from electric shock and fire, rep lacement components in mains circuits - including the power
transformer, test leads, and input jacks - must be purchased from Keithley Instruments. Standard fuses with applicable national safety approvals may be used if the rating and type are the same. Other components that are not safety-related may be purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product). If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
T o clean an instrument, use a damp cloth or mild, water-based cleaner . Clean the exterior of the instrument only . Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.

Table of Contents

1 Performance Verification

Introduction ................................................................................ 1-2
Verification test requirements .................................................... 1-2
Environmental conditions ................................................... 1-2
Warm-up period .................................................................. 1-3
Line power .......................................................................... 1-3
Recommended test equipment ................................................... 1-3
1Ω resistor characterization ................................................ 1-4
Verification limits ...................................................................... 1-5
Example limits calculation .................................................. 1-5
Resistance limits calculation ............................................... 1-5
Restoring factory defaults .......................................................... 1-6
Performing the verification test procedures ............................... 1-6
Test summary ...................................................................... 1-6
Test considerations ............................................................. 1-7
Setting the source range and output value .......................... 1-7
Setting the measurement range ........................................... 1-8
Compliance considerations ........................................................ 1-8
Compliance limits ............................................................... 1-8
Types of compliance ........................................................... 1-8
Maximum compliance values ............................................. 1-9
Determining compliance limit ............................................ 1-9
Taking the SourceMeter out of compliance ...................... 1-10
Output voltage accuracy ........................................................... 1-10
Voltage measurement accuracy ............................................... 1-12
Output current accuracy .................................................... 1-13
Current measurement accuracy ................................................ 1-16
Resistance measurement accuracy ........................................... 1-18
10mA to 1A range accuracy ...................................... 1-13
5A range accuracy ..................................................... 1-14
10mA to 1A range accuracy ...................................... 1-16
5A range accuracy ..................................................... 1-17

2Calibration

Introduction ................................................................................ 2-2
Environmental conditions .......................................................... 2-2
Temperature and relative humidity ..................................... 2-2
Warm-up period .................................................................. 2-2
Line power .......................................................................... 2-2
Calibration considerations .......................................................... 2-3
Calibration cycle ................................................................. 2-3
Recommended calibration equipment ................................. 2-4
1Ω resistor characterization ................................................ 2-4
Unlocking calibration ................................................................. 2-5
Unlocking calibration from the front panel ......................... 2-5
Unlocking calibration by remote ......................................... 2-6
Changing the password ............................................................... 2-7
Changing the password from the front panel ...................... 2-7
Changing the password by remote ...................................... 2-7
Resetting the calibration password ............................................. 2-8
Viewing calibration dates and calibration count ........................ 2-8
Calibration errors ........................................................................ 2-8
Front panel error reporting .................................................. 2-8
Remote error reporting ........................................................ 2-9
Front panel calibration ................................................................ 2-9
Step 1: Prepare the Model 2440 for calibration ........... 2-9
Step 2: Voltage calibration ......................................... 2-10
Step 3: Current calibration ......................................... 2-13
Step 4: Enter calibration dates and save calibration . 2-17
Step 5: Lock out calibration ...................................... 2-17
Remote calibration .................................................................... 2-18
Remote calibration commands .......................................... 2-18
Recommended calibration parameters .............................. 2-19
Remote calibration procedure ........................................... 2-20
Step 1: Prepare the Model 2440 for calibration ......... 2-20
Step 2: Voltage calibration ......................................... 2-21
Step 3: Current calibration ........................................ 2-23
Step 4: Program calibration dates .............................. 2-25
Step 5: Save calibration constants .............................. 2-25
Step 6: Lock out calibration ....................................... 2-25
Single-range calibration ............................................................ 2-25

3 Routine Maintenance

Introduction ................................................................................ 3-2
Line fuse replacement ................................................................. 3-2

4 Troubleshooting

Introduction ................................................................................ 4-2
Repair considerations ................................................................. 4-2
Power-on self-test ...................................................................... 4-2
Front panel tests ......................................................................... 4-3
KEYS test ........................................................................... 4-3
DISPLAY PATTERNS test ................................................ 4-3
CHAR SET test ................................................................... 4-4
Principles of operation ............................................................... 4-4
Overall block diagram ........................................................ 4-4
Analog circuits .................................................................... 4-4
Power supply ....................................................................... 4-7
Output stage ........................................................................ 4-8
A/D converter ..................................................................... 4-9
Active guard ........................................................................ 4-9
Digital circuitry ................................................................. 4-10
Display board circuit theory ............................................. 4-11
Troubleshooting ....................................................................... 4-12
Display board checks ........................................................ 4-12
Power supply checks ......................................................... 4-12
Digital circuitry checks ..................................................... 4-13
Analog circuitry checks .................................................... 4-14
Battery replacement ................................................................. 4-14
Battery replacement precautions ....................................... 4-15
Battery replacement procedure ......................................... 4-15
No comm link error .................................................................. 4-16

5Disassembly

Introduction ................................................................................ 5-2
Handling and cleaning ............................................................... 5-2
Handling PC boards ............................................................ 5-2
Solder repairs ...................................................................... 5-2
Static sensitive devices ............................................................... 5-3
Assembly drawings .................................................................... 5-3
Case cover removal .................................................................... 5-4
Analog board removal ................................................................ 5-5
Digital board removal ................................................................ 5-6
Front panel disassembly ............................................................. 5-6
Removing power components .................................................... 5-7
Power module removal ....................................................... 5-7
Instrument reassembly ............................................................... 5-7

6 Replaceable Parts

Introduction ................................................................................ 6-2
Parts lists ..................................................................................... 6-2
Ordering information .................................................................. 6-2
Factory service ............................................................................ 6-2
Component layouts ..................................................................... 6-3

A Specifications

B Command Reference

Introduction ............................................................................... B-2
Command summary ................................................................... B-2
Miscellaneous commands .......................................................... B-3
CODE .......................................................................... B-3
COUNT? ..................................................................... B-3
LOCK .......................................................................... B-4
SAVE .......................................................................... B-4
DATE .......................................................................... B-5
NDUE .......................................................................... B-5
SENSE ........................................................................ B-6
SOURCE ..................................................................... B-7
DATA? ........................................................................ B-9
Detecting calibration errors ....................................................... B-9
Reading the error queue ..................................................... B-9
Error summary .................................................................... B-9
Status byte EAV (Error Available) bit ............................. B-10
Generating an SRQ on error ............................................. B-10
Detecting calibration step completion ..................................... B-11
Using the *OPC? query .................................................... B-11
Using the *OPC command ............................................... B-11
Generating an SRQ on calibration complete .................... B-12

C Calibration Programs

Introduction ............................................................................... C-2
Computer hardware requirements ............................................. C-2
Software requirements ............................................................... C-2
Calibration equipment ............................................................... C-2
General program instructions .................................................... C-3
Program C-1 Model 2440 calibration program ....... C-4
Requesting calibration constants ............................................... C-7
Program C-2 Requesting calibration constants ....... C-7

List of Illustrations

1 Performance Verification

Figure 1-1 Connections for voltage verification tests ........................... 1-10

Figure 1-2 Connections for 10µA to 1A range current verification tests 1-13

Figure 1-3 Connections for 5A range current verification tests ............ 1-14
Figure 1-4 Connections for resistance accuracy verification ................ 1-18

2Calibration

Figure 2-1 Voltage calibration test connections .................................... 2-10
Figure 2-2 10µA to 1A range current calibration test connections ....... 2-13
Figure 2-3 5A range current calibration test connections ..................... 2-15

3 Routine Maintenance

Figure 3-1 Rear panel .............................................................................. 3-2

4 Troubleshooting

Figure 4-1 Overall block diagram ........................................................... 4-5
Figure 4-2 Analog circuitry block diagram ............................................. 4-6
Figure 4-3 Power supply block diagram ................................................. 4-7
Figure 4-4 Output stage simplified schematic ......................................... 4-8
Figure 4-5 Digital circuitry block diagram ............................................ 4-10

List of Tables

1 Performance Verification

Table 1-1 Recommended verification equipment .................................. 1-4
Table 1-2 Maximum compliance values ................................................ 1-9
Table 1-3 Output voltage accuracy limits ............................................. 1-11
Table 1-4 Voltage measurement accuracy limits ................................. 1-12
Table 1-5 Output current accuracy limits ............................................. 1-15
Table 1-6 Current measurement accuracy limits .................................. 1-17
Table 1-7 Ohms measurement accuracy limits ..................................... 1-19

2Calibration

Table 2-1 Recommended calibration equipment .................................... 2-4
Table 2-2 Calibration unlocked states .................................................... 2-6
Table 2-3 Front panel voltage calibration ............................................. 2-12
Table 2-4 Front panel current calibration ............................................. 2-16
Table 2-5 Remote calibration command summary ............................... 2-18
Table 2-6 Recommended :CALibration:PROTected:SENse
Table 2-7 Recommended :CALibration:PROTected:SOURce
Table 2-8 Voltage calibration initialization commands ....................... 2-22
Table 2-9 Voltage range calibration commands ................................... 2-22
Table 2-10 Current calibration initialization commands ........................ 2-24
Table 2-11 Current range calibration commands ................................... 2-24
parameter ranges .................................................................. 2-19
parameter ranges .................................................................. 2-20

3 Routine Maintenance

Table 3-1 Power line fuse ....................................................................... 3-3

4 Troubleshooting

Table 4-1 Display board checks ........................................................... 4-12
Table 4-2 Power supply checks ............................................................ 4-12
Table 4-3 Digital circuitry checks ........................................................ 4-13
Table 4-4 Analog circuitry checks ........................................................ 4-14

6 Replaceable Parts

Table 6-1 Analog board parts list ........................................................... 6-3
Table 6-2 Digital board parts list .......................................................... 6-12
Table 6-3 Display board parts list ......................................................... 6-17
Table 6-4 Miscellaneous parts list ........................................................ 6-18

B Command Reference

Table B-1 Remote calibration command summary ............................... B-2
Table B-2 Recommended :CALibration:PROTected:SENSe
parameter ranges ................................................................... B-7
Table B-3 Recommended :CALibration:PROTected:SOURce
parameter ranges ................................................................... B-8
Table B-4 Calibration errors ................................................................ B-10
1
Performance Verification
Performance Verification
1-2 Performance Verification Model 2440 5A SourceMeter® Service Manual
Introduction
Use the procedures in this section to verify that Model 2440 accuracy is within the limits stated in the instrument’s one-year accuracy specifications. You can perform these verifi cation procedures:
When you first receive the instrument to make sure that it was not damaged during shipment.
To verify that the unit meets factory specifications.
To determine if calibration is required.
Following calibration to make sure it was performed properly.
WARNING The information in this section is intended for qualified service per-
sonnel only. Do not attempt these procedures unless you are qualified to do so. Some of these procedures may expose you to hazardous volt ages, which could cause personal injury or death if contacted. Use standard safety precautions when working with hazardous voltages.
NOTE If the instrument is still under warranty and its performance is outside specified
limits, contact your Keithley representative or the factory to determine the cor rect course of action.
-
-
-
Verification test requirements
Be sure that you perform the verification tests:
Under the proper environmental conditions.
After the specified warm-up period.
Using the correct line voltage.
Using the proper test equipment.
Using the specified output signals and reading limits.
Environmental conditions
Conduct your performance verification procedures in a test environment with:
An ambient temperature of 18-28°C (65-82°F).
A relative humidity of less than 70% unless otherwise noted.
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-3
Warm-up period
Allow the Model 2440 to warm up for at least one hour before conducting the verification procedures.
If the instrument has been subjected to temperature extremes (those outside the ranges stated above), allow additional time for the instrument’s internal temperature to stabilize. Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manu­facturer.
Line power
The Model 2440 requires a line voltage of 100 to 240V and a line frequency of 50 to 60Hz. Verification tests must be performed within this range.
Recommended test equipment
Table 1-1 summarizes recommended verification equipment. You can use alternate equip-
ment as long as that equipment has specifications at least as good as those listed in Table
1-1. Keep in mind, however, that test equipment uncertainty will add to the uncertainty of
each measurement. Generally, test equipment uncertainty should be at least four times bet­ter than corresponding Model 2440 specifications. Table 1-1 lists the specifications of the recommended test equipment, including maximum allowable uncertainty for alternate test equipment, which is shown in parentheses.
Performance Verification
1-4 Performance Verification Model 2440 5A SourceMeter® Service Manual
Table 1-1
Recommended verification equipment
Description Manufacturer/Model Specifications
Digital Multimeter Hewlett Packard
HP3458A
Resistance Calibrator Fluke 5450A Resistance** 1.9Ω:
DC Voltage* 1V:
10V: 100V:
DC Current* 10μA:
100μA: 1mA: 10mA: 100mA: 1A:
19Ω: 190Ω:
1.9kΩ: 19kΩ: 190kΩ:
1.9MΩ: 19MΩ:
±5.6ppm ±4.3ppm ±6.3ppm
±25ppm ±23ppm ±20ppm ±20ppm ±35ppm ±110ppm
±65ppm (±460ppm) ±23ppm (±280ppm) ±10.5ppm (±230ppm) ±8ppm (±200ppm) ±7.5ppm (±195ppm) ±8.5ppm (±200ppm) ±11.5ppm (±180ppm) ±30ppm (±635ppm)
Precision Resistor*** Isotec RUG-Z-1R00-0.1 1Ω, ±0.1%, 100W
* 90-day, full-range accuracy specifications of ranges required for various measurement points.
** 90-day, ±5°C specifications of nominal resistance values shown. Use actual values for tests. Maximum uncertainty of alternate
test equipment shown in parentheses.
*** Required for verification of 5A current range. Characterize resistor to ±300ppm or better using recommended DMM before ver-
ifying 5A current measurement range.
1Ω resistor characterization
The recommended 1Ω resistor should be characterized to ±300ppm or better before veri­fying the 5A current measurement range. (You need not characterize the resistor if you are checking only the 5A current source range.) Use the 4-wire ohms function of the DMM recommended in value to calculate the current during the 5A current measurement range test procedure.
Table 1-1 to measure the resistance value, and then use that measured
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-5
Verification limits
The verification limits stated in this section have been calculated using only the Model 2440 one-year accuracy specifications, and they do not include test equipment uncer tainty. If a particular measurement falls outside the allowable range, recalculate new lim­its based on Model 2440 specifications and corresponding test equipment specifications.
Example limits calculation
As an example of how verification limits are calculated, assume you are testing the 10V DC output range using a 10V output value. Using the Model 2440 10V range one-year accuracy specification of ±(0.02% of output + 1.2mV offset), the calculated output limits are:
Output limits = 10V ± [(10V × 0.02%) + 1.2mV] Output limits = 10V ± (0.002 + 0.0012) Output limits = 10V ± 3.2mV Output limits = 9.9968V to 10.0032V
Resistance limits calculation
-
Performance Verification
When verifying the resistance measurement accuracy, it will probably be necessary to recalculate resistance limits based on the actual calibrator resistance values. You can cal culate resistance reading limits in the same manner described above, but be sure to use the actual calibrator resistance values and the Model 2440 normal accuracy specifications for your calculations.
As an example, assume you are testing the 20kΩ range, and that the actual value of the nominal 19kΩ calibrator resistor is 19.01kΩ. Using the Model 2440 20kΩ range one-year normal accuracy specifications of ±(0.06% of reading + 3Ω), the recalculated reading limits are:
Reading limits = 19.01kΩ ± [(19.01kΩ × 0.06%) + 3Ω] Reading limits = 19.01kΩ ±14.4Ω Reading limits = 18.9956kΩ to 19.0244kΩ
-
1-6 Performance Verification Model 2440 5A SourceMeter® Service Manual
Restoring factory defaults
Before performing the verification procedures, restore the instrument to its factory front panel (bench) defaults as follows:
1. Press the MENU key. The instrument will display the following prompt:
MAIN MENU
SAVESETUP COMMUNICATION CAL ©
2. Select SAVESETUP, and then press ENTER. The unit then displays:
SAVESETUP MENU
GLOBAL SOURCE-MEMORY
3. Select GLOBAL, and then press ENTER. The unit then displays:
GLOBAL SETUP MENU
SAVE RESTORE POWERON RESET
4. Select RESET, and then press ENTER. The unit displays:
RESET ORIGINAL DFLTS
BENCH GPIB
5. Select BENCH, and then press ENTER. The unit then displays:
RESETTING INSTRUMENT
ENTER to confirm; EXIT to abort
6. Press ENTER to restore bench defaults, and note the unit displays the following:
RESET COMPLETE
BENCH defaults are now restored © ß Press ENTER to continue
7. Press ENTER then EXIT as necessary to return to normal display.
Performing the verification test procedures
Test summary
DC voltage output accuracy
DC voltage measurement accuracy
DC current output accuracy
DC current measurement accuracy
Resistance measurement accuracy
If the Model 2440 is not within specifications and not under warranty, see the calibration procedures in
Section 2 for information on calibrating the unit.
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-7
Test considerations
When performing the verification procedures:
Be sure to restore factory front panel defaults as previously outlined.
Make sure that the test equipment is properly warmed up and connected to the Model 2440 INPUT/OUTPUT jacks. Also be sure that the front panel jacks are selected with the TERMINALS key.
Make sure the Model 2440 is set to the correct source range (see below).
Be sure that the Model 2440 output is turned on before making measurements.
Be sure the test equipment is set up for the proper function and range.
Allow the Model 2440 output signal to settle before making a measurement.
Do not connect test equipment to the Model 2440 through a scanner, multiplexer, or other switching equipment.
WARNING The maximum common-mode voltage (voltage between LO and chas-
CAUTION The maximum voltage between INPUT/OUTPUT HI and LO or 4-
Performance Verification
sis ground) is 40V peak. Exceeding this value may cause a breakdown in insulation, creating a shock hazard.
WIRE SENSE HI and LO is 42V peak. The maximum voltage between INPUT/OUTPUT HI and 4-WIRE SENSE HI or between INPUT/ OUTPUT LO and 4-WIRE SENSE LO is 5V. Exceeding these voltage values may result in instrument damage.
Setting the source range and output value
Before testing each verification point, you must properly set the source range and output value as outlined below.
1. Press either the SOURCE V or SOURCE I key to select the appropriate source function.
2. Press the EDIT key as required to select the source display field. Note that the cur­sor will flash in the source field while its value is being edited.
3. With the cursor in the source display field flashing, set the source range to the low­est possible range for the value to be sourced using the up or down RANGE key. For example, you should use the 10V source range to output a 10V source value. With a 10V source value and the 10V range selected, the source field display will appear as follows:
Vsrc:+10.0000V
1-8 Performance Verification Model 2440 5A SourceMeter® Service Manual
4. With the source field cursor flashing, set the source output to the required value using either:
The SOURCE adjustment and left and right arrow keys.
The numeric keys.
5. Note that the source output value will be updated immediately; you need not press ENTER when setting the source value.
Setting the measurement range
When simultaneously sourcing and measuring either voltage or current, the measure range is coupled to the source range, and you cannot independently control the measure range. Thus, it is not necessary for you to set the measure range when testing voltage or current measurement accuracy.
Compliance considerations
Compliance limits
When sourcing voltage, you can set the SourceMeter to limit current from 10nA to 5.25A. Conversely, when sourcing current, you can set the SourceMeter to limit voltage from
0.2mV to 42V. The SourceMeter output will not exceed the programmed compliance limit.
Types of compliance
There are two types of compliance that can occur: “real” and “range.” Depending on which value is lower, the output will clamp at either the displayed compliance setting (“real”) or at the maximum measurement range reading (“range”).
The “real” compliance condition can occur when the compliance setting is less than the highest possible reading of the measurement range. When in compliance, the source out put clamps at the displayed compliance value. For example, if the compliance voltage is set to 1V and the measurement range is 2V, the output voltage will clamp (limit) at 1V.
“Range” compliance can occur when the compliance setting is higher than the possible reading of the selected measurement range. When in compliance, the source output clamps at the maximum measurement range reading (not the compliance value). For example, if the compliance voltage is set to 1V and the measurement range is 200mV, the output voltage will clamp (limit) at 210mV.
-
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-9
Maximum compliance values
The maximum compliance values for the measurement ranges are summarized in
Table 1-2.
Table 1-2
Maximum compliance values
Maximum
Measurement
range
200mV
2V 10V 40V
10μA
100μA
1mA
10mA
100mA
1A
5A
compliance
value
210mV
2.1V
10.5V 42V
10.5μA
105μA
1.05mA
10.5mA
105mA
1.05A
5.25A
Performance Verification
When the SourceMeter goes into compliance, the “Cmpl” label or the units label (i.e., “mA”) for the compliance display will flash.
Determining compliance limit
The relationships to determine which compliance is in effect are summarized as follows. They assume that the measurement function is the same as the compliance function.
Compliance Setting < Measurement Range = Real Compliance
Measurement Range < Compliance Setting = Range Compliance
You can determine the compliance that is in effect by comparing the displayed compli­ance setting to the present measurement range. If the compliance setting is lower than the maximum possible reading on the present measurement range, the compliance setting is the compliance limit. If the compliance setting is higher than the measurement range, the maximum reading on that measurement range is the compliance limit.
1-10 Performance Verification Model 2440 5A SourceMeter® Service Manual
Taking the SourceMeter out of compliance
Verification measurements should not be made when the SourceMeter is in compliance. For purposes of the verification tests, the SourceMeter can be taken out of compliance by going into the edit mode and increasing the compliance limit.
NOTE Do not take the unit out of compliance by decreasing the source value or chang-
ing the range. Always use the recommended range and source settings when performing the verification tests.
Output voltage accuracy
Follow the steps below to verify that Model 2440 output voltage accuracy is within speci­fied limits. This test involves setting the output voltage to each full-range value and mea­suring the voltages with a precision digital multimeter.
1. With the power off, connect the digital multimeter to the Model 2440 INPUT/ OUTPUT jacks, as shown in
Figure 1-1
Connections for voltage verification tests
Figure 1-1.
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
PEAK
SWEEP
SOURCE
I
5
2440 5A SourceMeter
EDIT
EXIT ENTER
MEAS
FCTN
I
Ω
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
1
REL
RANGE
AUTO
RANGE
®
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V PEAK
TERMINALS
FRONT/
REAR
Model 2440
Input HI
Digital Multimeter
Input LO
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-11
2. Select the multimeter DC volts measuring function.
3. Set the voltage source protection to NONE. To do so, press CONFIG then SOURCE V to access the CONFIGURE V-SOURCE menu. Select PROTEC TION, and set the voltage source protection limit to NONE.
4. Press the Model 2440 SOURCE V key to source voltage, and make sure the source output is turned on.
5. Verify output voltage accuracy for each of the voltages listed in Table 1-3. For each test point:
Select the correct source range.
Set the Model 2440 output voltage to the indicated value.
Verify that the multimeter reading is within the limits given in the table.
6. Repeat the procedure for negative output voltages with the same magnitudes as those listed in
7. Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
Table 1-3
Output voltage accuracy limits
Model 2440
source range
200mV
2V 10V 40V
Table 1-3.
Model 2440 output
voltage setting
200.000mV
2.00000V
10.0000V
40.0000V
Output voltage limits
(1 year, 18°C-28°C)
199.360 to 200.640mV
1.99900 to 2.00100V
9.9968 to 10.0032V
39.9872 to 40.0128V
Performance Verification
-
1-12 Performance Verification Model 2440 5A SourceMeter® Service Manual
Voltage measurement accuracy
Follow the steps below to verify that Model 2440 voltage measurement accuracy is within specified limits. The test involves setting the source voltage to full-range values, as mea sured by a precision digital multimeter, and then verifying that the Model 2440 voltage readings are within required limits.
1. With the power off, connect the digital multimeter to the Model 2440 INPUT/ OUTPUT jacks , as shown in
2. Select the multimeter DC volts function.
3. Set the voltage source protection to NONE. To do so, press CONFIG then SOURCE V to access the CONFIGURE V-SOURCE menu. Select PROTEC TION, and set the voltage source protection limit to NONE.
4. Set the Model 2440 to both source and measure voltage by pressing the SOURCE V and MEAS V keys, and make sure the source output is turned on.
5. Verify output voltage accuracy for each of the voltages listed in Table 1-4. For each test point:
Select the correct source range.
Set the Model 2440 output voltage to the indicated value as measured by the digital multimeter.
Verify that the Model 2440 voltage reading is within the limits given in the table.
Figure 1-1.
-
-
NOTE It may not be possible to set the voltage source to the specified value. Use the
closest possible setting, and modify reading limits accordingly.
6. Repeat the procedure for negative source voltages with the same magnitudes as those listed in
7. Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
Table 1-4
Voltage measurement accuracy limits
Model 2440 source
and measure
range*
200mV
2V 10V 40V
Table 1-4.
Source voltage*
200.000mV
10.0000V
40.0000V
2.00000V
Model 2440 voltage reading
limits (1 year, 18°C-28°C)
199.676 to 200.324mV
1.99946 to 2.00054V
9.9978 to 10.0022V
39.991 to 40.009V
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-13
Table 1-4
Voltage measurement accuracy limits
Model 2440 source
and measure
range* Source voltage*
* Measure range coupled to source range when simultaneously sourcing and measuring voltage.
** As measured by precision digital multimeter. Use closest possible value, and modify reading lim-
its accordingly if necessary.
Output current accuracy
Follow the steps below to verify that Model 2440 output current accuracy is within speci­fied limits. The test involves setting the output current to each full-range value and mea­suring the currents with a precision digital multimeter.
10μA to 1A range accuracy
1. With the power off, connect the digital multimeter to the Model 2440 INPUT/ OUTPUT jacks, as shown in
Figure 1-2
Connections for 10μA to 1A range current verification tests
V
LOCAL
67
DIGITS SPEED
MEAS
I
Ω
1
230
REL
FILTER
89
RECALL
STORE
EDIT
DISPLAY
TOGGLE
POWER
Performance Verification
Model 2440 voltage reading
limits (1 year, 18°C-28°C)
Figure 1-2.
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
PEAK
RANGE
AUTO
RANGE
®
2440 5A SourceMeter
V
4
TRIG
SWEEP
+/-
CONFIG MENU
I
5
SOURCE
EDIT
EXIT ENTER
FCTN
LIMIT
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V PEAK
TERMINALS
FRONT/
REAR
Model 2440
Digital Multimeter
2. Select the multimeter DC current measuring function.
Input LO
Amps
1-14 Performance Verification Model 2440 5A SourceMeter® Service Manual
3. Press the Model 2440 SOURCE I key to source current, and make sure the source output is turned on.
4. Verify output current accuracy for the 10μA-1A range currents listed in Table 1-5. For each test point:
Select the correct source range.
Set the Model 2440 output current to the correct value.
Verify that the multimeter reading is within the limits given in the table.
5. Repeat the procedure for negative output currents with the same magnitudes as those listed in
Table 1-5.
6. Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
5A range accuracy
1. With the power off, connect the digital multimeter and the 1Ω resistor to the Model 2440 INPUT/OUTPUT jacks, as shown in
Figure 1-3
Connections for 5A range current verification tests
SWEEP
I
5
SOURCE
2440 5A SourceMeter
EDIT
EXIT ENTER
RANGE
AUTO
RANGE
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
MEAS
FCTN
I
Ω
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
1
REL
4- WIRE
SENSE
HI
75V
5V
PEAK
PEAK
®
LO
ON/OFF
OUTPUT
INPUT/
OUTPUT
TERMINALS
FRONT/
REAR
Figure 1-3.
75V PEAK
250V PEAK
Model 2440
Digital Multimeter
1Ω Resistor
Input HI
Input LO
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-15
2. Select the multimeter DC volts measuring function.
3. Press the Model 2440 SOURCE I key to source current, and make sure the source output is turned on.
4. Verify output current accuracy for the 5A range. Be sure to:
Select the 5A source range.
Set the Model 2440 output current to the correct 5A output value.
Verify that the multimeter reading is within the 5A range limits given in Table
1-5.
(Since the value of the 1Ω resistor value is assumed to be the same as its nominal value, the DMM voltage reading is the same as the sourced current.)
5. Repeat the procedure for a negative 5A current output value.
6. Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
Table 1-5
Output current accuracy limits
Model 2440
source range
10μA
100μA
1mA
10mA
100mA
1A 5A*
Model 2440
output current setting
10.0000μA
100.000μA
1.00000mA
10.0000mA
100.000mA
1.00000A
5.00000A
Output current limits
(1 year, 18°C-28°C)
9.9947 to 10.0053μA
99.949 to 100.051μA
0.99946 to 1.00054mA
9.9935 to 10.0065mA
99.914 to 100.086mA
0.99843 to 1.00157A
4.9896 to 5.0104A
Performance Verification
*See separate procedure for 5A range. DMM voltage reading is same as sourced current.
+ 114 hidden pages