Tektronix 2440 Service Manual

Model 2440
Service Manual
2440-902-01 Rev. B / December 2004
www.keithley.com
G
A
T
E
E
A
R
AEM
R
RUS
C F
O
IF
NO
ECNED

WARRANTY

Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of one (1) year from date of shipment.
Keithley Instruments, Inc. warran probes, cables, software, rechargeable batteries, diskettes, and documentation.
During the warranty period, Keithley Instruments will, at its option, ei proves to be defective.
To exercise this warranty, write or call your local Keithley Instruments representative, or contact
thley Instruments headquarters in Cleveland, Ohio. You will be given prompt assistance and
Kei return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
ts the following items for 90 days from the date of shipment:
ther repair or replace any product that

LIMITATION OF WARRANTY

This warranty does not apply to defects resulting from product modification without Keithley Instruments’ express written consent, or misuse of any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER W INCLUDING ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE THE BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES
ING OUT OF THE USE OF ITS INSTRUMENTS AND SOFTWARE, EVEN IF KEITHLEY
ARIS INSTRUMENTS, INC. HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COST OF REMOVAL AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
ARRANTIES, EXPRESSED OR IMPLIED,
OF ITS EMPLOYEES SHALL BE LIABLE

A G R E A T E R M E A S U R E O F C O N F I D E N C E

Keithley Instruments, Inc.
Corporate Headquarters • 28775 Aurora Road • Cleveland, Ohio 44139
440-248-0400 • Fax: 440-248-6168 • 1-888-KEITHLEY (1-888-534-8453) • www.keithley.com
3/07
Model 2440 5A SourceMeter
Service Manual
®
©2001, Keithley Instruments, Inc.
All rights reserved.
Cleveland, Ohio, U.S.A.
Second Printing, December 2004
Document Number: 2440-902-01 Rev. B

Manual Print History

The print history shown below lists the printing dates of all Revisions and Addenda created for this manual. The Revision Level letter increases alphabetically as the manual undergoes sub sequent updates. Addenda, which are released between Revisions, contain important change in­formation that the user should incorporate immediately into the manual. Addenda are numbered sequentially. When a new Revision is created, all Addenda associated with the previous Revi sion of the manual are incorporated into the new Revision of the manual. Each new Revision includes a revised copy of this print history page.
Revision A (Document Number 2440-902-01)................................................................ April 2001
Revision B (Document Number 2440-902-01)........................................................Decmeber 2004
-
-
All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc. Other brand names are trademarks or registered trademarks of their respective holders.
Safety Precautions
The following safety precautions should be observed before using this product and any associated instrumentation. Although some instruments and accessories would normally be used with non-hazardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired. The types of product users are: Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring
that the equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained service personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are rated Measurement Category I and Measurement Category II, as described in the International Electrotechnical Commission (IEC) Standard IEC
60664. Most measurement, control, and data I/O signals are Measurement Category I and must not be directly connected to mains voltage or to voltage sources with high transient over-voltages. Measurement Category II connections require protection for high transient over-voltages often associated with local AC mains connections. Assume all measurement, control, and data I/O connections are for connection to Category I sources unless otherwise marked or described in the user documentation.
Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30V RMS, 42.4V peak, or 60VDC are present. A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 volts, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance­limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, make sure the line cord is connected to a properly grounded power recep tacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator.
04/09
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the
!
circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: connecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
The instrument and accessories must be used in accordance with specifications and operating instructions, or the safety of the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories, as defined in the specifications and operating information, and as shown on the instrument or test fixture panels, or switching card.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard. Chassis connections must only be used as shield connections for measuring circuits, NOT as safety earth ground
connections. If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation
requires the use of a lid interlock. If a screw is present, connect it to safety earth ground using the wire recommended in the user documentation. The symbol on an instrument means caution, risk of danger. The user should refer to the operating instructions
located in the user documenta ti on in all case s where the symbol is marked on the instrument. The symbol on an instrument means caution, risk of danger. User standard safety precautions to avoid
personal contact with these voltages. The symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
The symbol indicates a connection terminal to the equipment frame.
If this symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be properly disposed of according to federal, state, and local laws.
The WARNING heading in the user documentation explains dangers that might result in personal injury or death. Always read the associated information very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could damage the instrument. Such damage may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans. Before performing any maintenance, disconnect the line cord and all test cables. To maintain protection from electric shock and fire, rep lacement components in mains circuits - including the power
transformer, test leads, and input jacks - must be purchased from Keithley Instruments. Standard fuses with applicable national safety approvals may be used if the rating and type are the same. Other components that are not safety-related may be purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product). If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
T o clean an instrument, use a damp cloth or mild, water-based cleaner . Clean the exterior of the instrument only . Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.

Table of Contents

1 Performance Verification

Introduction ................................................................................ 1-2
Verification test requirements .................................................... 1-2
Environmental conditions ................................................... 1-2
Warm-up period .................................................................. 1-3
Line power .......................................................................... 1-3
Recommended test equipment ................................................... 1-3
1Ω resistor characterization ................................................ 1-4
Verification limits ...................................................................... 1-5
Example limits calculation .................................................. 1-5
Resistance limits calculation ............................................... 1-5
Restoring factory defaults .......................................................... 1-6
Performing the verification test procedures ............................... 1-6
Test summary ...................................................................... 1-6
Test considerations ............................................................. 1-7
Setting the source range and output value .......................... 1-7
Setting the measurement range ........................................... 1-8
Compliance considerations ........................................................ 1-8
Compliance limits ............................................................... 1-8
Types of compliance ........................................................... 1-8
Maximum compliance values ............................................. 1-9
Determining compliance limit ............................................ 1-9
Taking the SourceMeter out of compliance ...................... 1-10
Output voltage accuracy ........................................................... 1-10
Voltage measurement accuracy ............................................... 1-12
Output current accuracy .................................................... 1-13
Current measurement accuracy ................................................ 1-16
Resistance measurement accuracy ........................................... 1-18
10mA to 1A range accuracy ...................................... 1-13
5A range accuracy ..................................................... 1-14
10mA to 1A range accuracy ...................................... 1-16
5A range accuracy ..................................................... 1-17

2Calibration

Introduction ................................................................................ 2-2
Environmental conditions .......................................................... 2-2
Temperature and relative humidity ..................................... 2-2
Warm-up period .................................................................. 2-2
Line power .......................................................................... 2-2
Calibration considerations .......................................................... 2-3
Calibration cycle ................................................................. 2-3
Recommended calibration equipment ................................. 2-4
1Ω resistor characterization ................................................ 2-4
Unlocking calibration ................................................................. 2-5
Unlocking calibration from the front panel ......................... 2-5
Unlocking calibration by remote ......................................... 2-6
Changing the password ............................................................... 2-7
Changing the password from the front panel ...................... 2-7
Changing the password by remote ...................................... 2-7
Resetting the calibration password ............................................. 2-8
Viewing calibration dates and calibration count ........................ 2-8
Calibration errors ........................................................................ 2-8
Front panel error reporting .................................................. 2-8
Remote error reporting ........................................................ 2-9
Front panel calibration ................................................................ 2-9
Step 1: Prepare the Model 2440 for calibration ........... 2-9
Step 2: Voltage calibration ......................................... 2-10
Step 3: Current calibration ......................................... 2-13
Step 4: Enter calibration dates and save calibration . 2-17
Step 5: Lock out calibration ...................................... 2-17
Remote calibration .................................................................... 2-18
Remote calibration commands .......................................... 2-18
Recommended calibration parameters .............................. 2-19
Remote calibration procedure ........................................... 2-20
Step 1: Prepare the Model 2440 for calibration ......... 2-20
Step 2: Voltage calibration ......................................... 2-21
Step 3: Current calibration ........................................ 2-23
Step 4: Program calibration dates .............................. 2-25
Step 5: Save calibration constants .............................. 2-25
Step 6: Lock out calibration ....................................... 2-25
Single-range calibration ............................................................ 2-25

3 Routine Maintenance

Introduction ................................................................................ 3-2
Line fuse replacement ................................................................. 3-2

4 Troubleshooting

Introduction ................................................................................ 4-2
Repair considerations ................................................................. 4-2
Power-on self-test ...................................................................... 4-2
Front panel tests ......................................................................... 4-3
KEYS test ........................................................................... 4-3
DISPLAY PATTERNS test ................................................ 4-3
CHAR SET test ................................................................... 4-4
Principles of operation ............................................................... 4-4
Overall block diagram ........................................................ 4-4
Analog circuits .................................................................... 4-4
Power supply ....................................................................... 4-7
Output stage ........................................................................ 4-8
A/D converter ..................................................................... 4-9
Active guard ........................................................................ 4-9
Digital circuitry ................................................................. 4-10
Display board circuit theory ............................................. 4-11
Troubleshooting ....................................................................... 4-12
Display board checks ........................................................ 4-12
Power supply checks ......................................................... 4-12
Digital circuitry checks ..................................................... 4-13
Analog circuitry checks .................................................... 4-14
Battery replacement ................................................................. 4-14
Battery replacement precautions ....................................... 4-15
Battery replacement procedure ......................................... 4-15
No comm link error .................................................................. 4-16

5Disassembly

Introduction ................................................................................ 5-2
Handling and cleaning ............................................................... 5-2
Handling PC boards ............................................................ 5-2
Solder repairs ...................................................................... 5-2
Static sensitive devices ............................................................... 5-3
Assembly drawings .................................................................... 5-3
Case cover removal .................................................................... 5-4
Analog board removal ................................................................ 5-5
Digital board removal ................................................................ 5-6
Front panel disassembly ............................................................. 5-6
Removing power components .................................................... 5-7
Power module removal ....................................................... 5-7
Instrument reassembly ............................................................... 5-7

6 Replaceable Parts

Introduction ................................................................................ 6-2
Parts lists ..................................................................................... 6-2
Ordering information .................................................................. 6-2
Factory service ............................................................................ 6-2
Component layouts ..................................................................... 6-3

A Specifications

B Command Reference

Introduction ............................................................................... B-2
Command summary ................................................................... B-2
Miscellaneous commands .......................................................... B-3
CODE .......................................................................... B-3
COUNT? ..................................................................... B-3
LOCK .......................................................................... B-4
SAVE .......................................................................... B-4
DATE .......................................................................... B-5
NDUE .......................................................................... B-5
SENSE ........................................................................ B-6
SOURCE ..................................................................... B-7
DATA? ........................................................................ B-9
Detecting calibration errors ....................................................... B-9
Reading the error queue ..................................................... B-9
Error summary .................................................................... B-9
Status byte EAV (Error Available) bit ............................. B-10
Generating an SRQ on error ............................................. B-10
Detecting calibration step completion ..................................... B-11
Using the *OPC? query .................................................... B-11
Using the *OPC command ............................................... B-11
Generating an SRQ on calibration complete .................... B-12

C Calibration Programs

Introduction ............................................................................... C-2
Computer hardware requirements ............................................. C-2
Software requirements ............................................................... C-2
Calibration equipment ............................................................... C-2
General program instructions .................................................... C-3
Program C-1 Model 2440 calibration program ....... C-4
Requesting calibration constants ............................................... C-7
Program C-2 Requesting calibration constants ....... C-7

List of Illustrations

1 Performance Verification

Figure 1-1 Connections for voltage verification tests ........................... 1-10

Figure 1-2 Connections for 10µA to 1A range current verification tests 1-13

Figure 1-3 Connections for 5A range current verification tests ............ 1-14
Figure 1-4 Connections for resistance accuracy verification ................ 1-18

2Calibration

Figure 2-1 Voltage calibration test connections .................................... 2-10
Figure 2-2 10µA to 1A range current calibration test connections ....... 2-13
Figure 2-3 5A range current calibration test connections ..................... 2-15

3 Routine Maintenance

Figure 3-1 Rear panel .............................................................................. 3-2

4 Troubleshooting

Figure 4-1 Overall block diagram ........................................................... 4-5
Figure 4-2 Analog circuitry block diagram ............................................. 4-6
Figure 4-3 Power supply block diagram ................................................. 4-7
Figure 4-4 Output stage simplified schematic ......................................... 4-8
Figure 4-5 Digital circuitry block diagram ............................................ 4-10

List of Tables

1 Performance Verification

Table 1-1 Recommended verification equipment .................................. 1-4
Table 1-2 Maximum compliance values ................................................ 1-9
Table 1-3 Output voltage accuracy limits ............................................. 1-11
Table 1-4 Voltage measurement accuracy limits ................................. 1-12
Table 1-5 Output current accuracy limits ............................................. 1-15
Table 1-6 Current measurement accuracy limits .................................. 1-17
Table 1-7 Ohms measurement accuracy limits ..................................... 1-19

2Calibration

Table 2-1 Recommended calibration equipment .................................... 2-4
Table 2-2 Calibration unlocked states .................................................... 2-6
Table 2-3 Front panel voltage calibration ............................................. 2-12
Table 2-4 Front panel current calibration ............................................. 2-16
Table 2-5 Remote calibration command summary ............................... 2-18
Table 2-6 Recommended :CALibration:PROTected:SENse
Table 2-7 Recommended :CALibration:PROTected:SOURce
Table 2-8 Voltage calibration initialization commands ....................... 2-22
Table 2-9 Voltage range calibration commands ................................... 2-22
Table 2-10 Current calibration initialization commands ........................ 2-24
Table 2-11 Current range calibration commands ................................... 2-24
parameter ranges .................................................................. 2-19
parameter ranges .................................................................. 2-20

3 Routine Maintenance

Table 3-1 Power line fuse ....................................................................... 3-3

4 Troubleshooting

Table 4-1 Display board checks ........................................................... 4-12
Table 4-2 Power supply checks ............................................................ 4-12
Table 4-3 Digital circuitry checks ........................................................ 4-13
Table 4-4 Analog circuitry checks ........................................................ 4-14

6 Replaceable Parts

Table 6-1 Analog board parts list ........................................................... 6-3
Table 6-2 Digital board parts list .......................................................... 6-12
Table 6-3 Display board parts list ......................................................... 6-17
Table 6-4 Miscellaneous parts list ........................................................ 6-18

B Command Reference

Table B-1 Remote calibration command summary ............................... B-2
Table B-2 Recommended :CALibration:PROTected:SENSe
parameter ranges ................................................................... B-7
Table B-3 Recommended :CALibration:PROTected:SOURce
parameter ranges ................................................................... B-8
Table B-4 Calibration errors ................................................................ B-10
1
Performance Verification
Performance Verification
1-2 Performance Verification Model 2440 5A SourceMeter® Service Manual
Introduction
Use the procedures in this section to verify that Model 2440 accuracy is within the limits stated in the instrument’s one-year accuracy specifications. You can perform these verifi cation procedures:
When you first receive the instrument to make sure that it was not damaged during shipment.
To verify that the unit meets factory specifications.
To determine if calibration is required.
Following calibration to make sure it was performed properly.
WARNING The information in this section is intended for qualified service per-
sonnel only. Do not attempt these procedures unless you are qualified to do so. Some of these procedures may expose you to hazardous volt ages, which could cause personal injury or death if contacted. Use standard safety precautions when working with hazardous voltages.
NOTE If the instrument is still under warranty and its performance is outside specified
limits, contact your Keithley representative or the factory to determine the cor rect course of action.
-
-
-
Verification test requirements
Be sure that you perform the verification tests:
Under the proper environmental conditions.
After the specified warm-up period.
Using the correct line voltage.
Using the proper test equipment.
Using the specified output signals and reading limits.
Environmental conditions
Conduct your performance verification procedures in a test environment with:
An ambient temperature of 18-28°C (65-82°F).
A relative humidity of less than 70% unless otherwise noted.
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-3
Warm-up period
Allow the Model 2440 to warm up for at least one hour before conducting the verification procedures.
If the instrument has been subjected to temperature extremes (those outside the ranges stated above), allow additional time for the instrument’s internal temperature to stabilize. Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manu­facturer.
Line power
The Model 2440 requires a line voltage of 100 to 240V and a line frequency of 50 to 60Hz. Verification tests must be performed within this range.
Recommended test equipment
Table 1-1 summarizes recommended verification equipment. You can use alternate equip-
ment as long as that equipment has specifications at least as good as those listed in Table
1-1. Keep in mind, however, that test equipment uncertainty will add to the uncertainty of
each measurement. Generally, test equipment uncertainty should be at least four times bet­ter than corresponding Model 2440 specifications. Table 1-1 lists the specifications of the recommended test equipment, including maximum allowable uncertainty for alternate test equipment, which is shown in parentheses.
Performance Verification
1-4 Performance Verification Model 2440 5A SourceMeter® Service Manual
Table 1-1
Recommended verification equipment
Description Manufacturer/Model Specifications
Digital Multimeter Hewlett Packard
HP3458A
Resistance Calibrator Fluke 5450A Resistance** 1.9Ω:
DC Voltage* 1V:
10V: 100V:
DC Current* 10μA:
100μA: 1mA: 10mA: 100mA: 1A:
19Ω: 190Ω:
1.9kΩ: 19kΩ: 190kΩ:
1.9MΩ: 19MΩ:
±5.6ppm ±4.3ppm ±6.3ppm
±25ppm ±23ppm ±20ppm ±20ppm ±35ppm ±110ppm
±65ppm (±460ppm) ±23ppm (±280ppm) ±10.5ppm (±230ppm) ±8ppm (±200ppm) ±7.5ppm (±195ppm) ±8.5ppm (±200ppm) ±11.5ppm (±180ppm) ±30ppm (±635ppm)
Precision Resistor*** Isotec RUG-Z-1R00-0.1 1Ω, ±0.1%, 100W
* 90-day, full-range accuracy specifications of ranges required for various measurement points.
** 90-day, ±5°C specifications of nominal resistance values shown. Use actual values for tests. Maximum uncertainty of alternate
test equipment shown in parentheses.
*** Required for verification of 5A current range. Characterize resistor to ±300ppm or better using recommended DMM before ver-
ifying 5A current measurement range.
1Ω resistor characterization
The recommended 1Ω resistor should be characterized to ±300ppm or better before veri­fying the 5A current measurement range. (You need not characterize the resistor if you are checking only the 5A current source range.) Use the 4-wire ohms function of the DMM recommended in value to calculate the current during the 5A current measurement range test procedure.
Table 1-1 to measure the resistance value, and then use that measured
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-5
Verification limits
The verification limits stated in this section have been calculated using only the Model 2440 one-year accuracy specifications, and they do not include test equipment uncer tainty. If a particular measurement falls outside the allowable range, recalculate new lim­its based on Model 2440 specifications and corresponding test equipment specifications.
Example limits calculation
As an example of how verification limits are calculated, assume you are testing the 10V DC output range using a 10V output value. Using the Model 2440 10V range one-year accuracy specification of ±(0.02% of output + 1.2mV offset), the calculated output limits are:
Output limits = 10V ± [(10V × 0.02%) + 1.2mV] Output limits = 10V ± (0.002 + 0.0012) Output limits = 10V ± 3.2mV Output limits = 9.9968V to 10.0032V
Resistance limits calculation
-
Performance Verification
When verifying the resistance measurement accuracy, it will probably be necessary to recalculate resistance limits based on the actual calibrator resistance values. You can cal culate resistance reading limits in the same manner described above, but be sure to use the actual calibrator resistance values and the Model 2440 normal accuracy specifications for your calculations.
As an example, assume you are testing the 20kΩ range, and that the actual value of the nominal 19kΩ calibrator resistor is 19.01kΩ. Using the Model 2440 20kΩ range one-year normal accuracy specifications of ±(0.06% of reading + 3Ω), the recalculated reading limits are:
Reading limits = 19.01kΩ ± [(19.01kΩ × 0.06%) + 3Ω] Reading limits = 19.01kΩ ±14.4Ω Reading limits = 18.9956kΩ to 19.0244kΩ
-
1-6 Performance Verification Model 2440 5A SourceMeter® Service Manual
Restoring factory defaults
Before performing the verification procedures, restore the instrument to its factory front panel (bench) defaults as follows:
1. Press the MENU key. The instrument will display the following prompt:
MAIN MENU
SAVESETUP COMMUNICATION CAL ©
2. Select SAVESETUP, and then press ENTER. The unit then displays:
SAVESETUP MENU
GLOBAL SOURCE-MEMORY
3. Select GLOBAL, and then press ENTER. The unit then displays:
GLOBAL SETUP MENU
SAVE RESTORE POWERON RESET
4. Select RESET, and then press ENTER. The unit displays:
RESET ORIGINAL DFLTS
BENCH GPIB
5. Select BENCH, and then press ENTER. The unit then displays:
RESETTING INSTRUMENT
ENTER to confirm; EXIT to abort
6. Press ENTER to restore bench defaults, and note the unit displays the following:
RESET COMPLETE
BENCH defaults are now restored © ß Press ENTER to continue
7. Press ENTER then EXIT as necessary to return to normal display.
Performing the verification test procedures
Test summary
DC voltage output accuracy
DC voltage measurement accuracy
DC current output accuracy
DC current measurement accuracy
Resistance measurement accuracy
If the Model 2440 is not within specifications and not under warranty, see the calibration procedures in
Section 2 for information on calibrating the unit.
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-7
Test considerations
When performing the verification procedures:
Be sure to restore factory front panel defaults as previously outlined.
Make sure that the test equipment is properly warmed up and connected to the Model 2440 INPUT/OUTPUT jacks. Also be sure that the front panel jacks are selected with the TERMINALS key.
Make sure the Model 2440 is set to the correct source range (see below).
Be sure that the Model 2440 output is turned on before making measurements.
Be sure the test equipment is set up for the proper function and range.
Allow the Model 2440 output signal to settle before making a measurement.
Do not connect test equipment to the Model 2440 through a scanner, multiplexer, or other switching equipment.
WARNING The maximum common-mode voltage (voltage between LO and chas-
CAUTION The maximum voltage between INPUT/OUTPUT HI and LO or 4-
Performance Verification
sis ground) is 40V peak. Exceeding this value may cause a breakdown in insulation, creating a shock hazard.
WIRE SENSE HI and LO is 42V peak. The maximum voltage between INPUT/OUTPUT HI and 4-WIRE SENSE HI or between INPUT/ OUTPUT LO and 4-WIRE SENSE LO is 5V. Exceeding these voltage values may result in instrument damage.
Setting the source range and output value
Before testing each verification point, you must properly set the source range and output value as outlined below.
1. Press either the SOURCE V or SOURCE I key to select the appropriate source function.
2. Press the EDIT key as required to select the source display field. Note that the cur­sor will flash in the source field while its value is being edited.
3. With the cursor in the source display field flashing, set the source range to the low­est possible range for the value to be sourced using the up or down RANGE key. For example, you should use the 10V source range to output a 10V source value. With a 10V source value and the 10V range selected, the source field display will appear as follows:
Vsrc:+10.0000V
1-8 Performance Verification Model 2440 5A SourceMeter® Service Manual
4. With the source field cursor flashing, set the source output to the required value using either:
The SOURCE adjustment and left and right arrow keys.
The numeric keys.
5. Note that the source output value will be updated immediately; you need not press ENTER when setting the source value.
Setting the measurement range
When simultaneously sourcing and measuring either voltage or current, the measure range is coupled to the source range, and you cannot independently control the measure range. Thus, it is not necessary for you to set the measure range when testing voltage or current measurement accuracy.
Compliance considerations
Compliance limits
When sourcing voltage, you can set the SourceMeter to limit current from 10nA to 5.25A. Conversely, when sourcing current, you can set the SourceMeter to limit voltage from
0.2mV to 42V. The SourceMeter output will not exceed the programmed compliance limit.
Types of compliance
There are two types of compliance that can occur: “real” and “range.” Depending on which value is lower, the output will clamp at either the displayed compliance setting (“real”) or at the maximum measurement range reading (“range”).
The “real” compliance condition can occur when the compliance setting is less than the highest possible reading of the measurement range. When in compliance, the source out put clamps at the displayed compliance value. For example, if the compliance voltage is set to 1V and the measurement range is 2V, the output voltage will clamp (limit) at 1V.
“Range” compliance can occur when the compliance setting is higher than the possible reading of the selected measurement range. When in compliance, the source output clamps at the maximum measurement range reading (not the compliance value). For example, if the compliance voltage is set to 1V and the measurement range is 200mV, the output voltage will clamp (limit) at 210mV.
-
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-9
Maximum compliance values
The maximum compliance values for the measurement ranges are summarized in
Table 1-2.
Table 1-2
Maximum compliance values
Maximum
Measurement
range
200mV
2V 10V 40V
10μA
100μA
1mA
10mA
100mA
1A
5A
compliance
value
210mV
2.1V
10.5V 42V
10.5μA
105μA
1.05mA
10.5mA
105mA
1.05A
5.25A
Performance Verification
When the SourceMeter goes into compliance, the “Cmpl” label or the units label (i.e., “mA”) for the compliance display will flash.
Determining compliance limit
The relationships to determine which compliance is in effect are summarized as follows. They assume that the measurement function is the same as the compliance function.
Compliance Setting < Measurement Range = Real Compliance
Measurement Range < Compliance Setting = Range Compliance
You can determine the compliance that is in effect by comparing the displayed compli­ance setting to the present measurement range. If the compliance setting is lower than the maximum possible reading on the present measurement range, the compliance setting is the compliance limit. If the compliance setting is higher than the measurement range, the maximum reading on that measurement range is the compliance limit.
1-10 Performance Verification Model 2440 5A SourceMeter® Service Manual
Taking the SourceMeter out of compliance
Verification measurements should not be made when the SourceMeter is in compliance. For purposes of the verification tests, the SourceMeter can be taken out of compliance by going into the edit mode and increasing the compliance limit.
NOTE Do not take the unit out of compliance by decreasing the source value or chang-
ing the range. Always use the recommended range and source settings when performing the verification tests.
Output voltage accuracy
Follow the steps below to verify that Model 2440 output voltage accuracy is within speci­fied limits. This test involves setting the output voltage to each full-range value and mea­suring the voltages with a precision digital multimeter.
1. With the power off, connect the digital multimeter to the Model 2440 INPUT/ OUTPUT jacks, as shown in
Figure 1-1
Connections for voltage verification tests
Figure 1-1.
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
PEAK
SWEEP
SOURCE
I
5
2440 5A SourceMeter
EDIT
EXIT ENTER
MEAS
FCTN
I
Ω
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
1
REL
RANGE
AUTO
RANGE
®
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V PEAK
TERMINALS
FRONT/
REAR
Model 2440
Input HI
Digital Multimeter
Input LO
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-11
2. Select the multimeter DC volts measuring function.
3. Set the voltage source protection to NONE. To do so, press CONFIG then SOURCE V to access the CONFIGURE V-SOURCE menu. Select PROTEC TION, and set the voltage source protection limit to NONE.
4. Press the Model 2440 SOURCE V key to source voltage, and make sure the source output is turned on.
5. Verify output voltage accuracy for each of the voltages listed in Table 1-3. For each test point:
Select the correct source range.
Set the Model 2440 output voltage to the indicated value.
Verify that the multimeter reading is within the limits given in the table.
6. Repeat the procedure for negative output voltages with the same magnitudes as those listed in
7. Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
Table 1-3
Output voltage accuracy limits
Model 2440
source range
200mV
2V 10V 40V
Table 1-3.
Model 2440 output
voltage setting
200.000mV
2.00000V
10.0000V
40.0000V
Output voltage limits
(1 year, 18°C-28°C)
199.360 to 200.640mV
1.99900 to 2.00100V
9.9968 to 10.0032V
39.9872 to 40.0128V
Performance Verification
-
1-12 Performance Verification Model 2440 5A SourceMeter® Service Manual
Voltage measurement accuracy
Follow the steps below to verify that Model 2440 voltage measurement accuracy is within specified limits. The test involves setting the source voltage to full-range values, as mea sured by a precision digital multimeter, and then verifying that the Model 2440 voltage readings are within required limits.
1. With the power off, connect the digital multimeter to the Model 2440 INPUT/ OUTPUT jacks , as shown in
2. Select the multimeter DC volts function.
3. Set the voltage source protection to NONE. To do so, press CONFIG then SOURCE V to access the CONFIGURE V-SOURCE menu. Select PROTEC TION, and set the voltage source protection limit to NONE.
4. Set the Model 2440 to both source and measure voltage by pressing the SOURCE V and MEAS V keys, and make sure the source output is turned on.
5. Verify output voltage accuracy for each of the voltages listed in Table 1-4. For each test point:
Select the correct source range.
Set the Model 2440 output voltage to the indicated value as measured by the digital multimeter.
Verify that the Model 2440 voltage reading is within the limits given in the table.
Figure 1-1.
-
-
NOTE It may not be possible to set the voltage source to the specified value. Use the
closest possible setting, and modify reading limits accordingly.
6. Repeat the procedure for negative source voltages with the same magnitudes as those listed in
7. Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
Table 1-4
Voltage measurement accuracy limits
Model 2440 source
and measure
range*
200mV
2V 10V 40V
Table 1-4.
Source voltage*
200.000mV
10.0000V
40.0000V
2.00000V
Model 2440 voltage reading
limits (1 year, 18°C-28°C)
199.676 to 200.324mV
1.99946 to 2.00054V
9.9978 to 10.0022V
39.991 to 40.009V
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-13
Table 1-4
Voltage measurement accuracy limits
Model 2440 source
and measure
range* Source voltage*
* Measure range coupled to source range when simultaneously sourcing and measuring voltage.
** As measured by precision digital multimeter. Use closest possible value, and modify reading lim-
its accordingly if necessary.
Output current accuracy
Follow the steps below to verify that Model 2440 output current accuracy is within speci­fied limits. The test involves setting the output current to each full-range value and mea­suring the currents with a precision digital multimeter.
10μA to 1A range accuracy
1. With the power off, connect the digital multimeter to the Model 2440 INPUT/ OUTPUT jacks, as shown in
Figure 1-2
Connections for 10μA to 1A range current verification tests
V
LOCAL
67
DIGITS SPEED
MEAS
I
Ω
1
230
REL
FILTER
89
RECALL
STORE
EDIT
DISPLAY
TOGGLE
POWER
Performance Verification
Model 2440 voltage reading
limits (1 year, 18°C-28°C)
Figure 1-2.
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
PEAK
RANGE
AUTO
RANGE
®
2440 5A SourceMeter
V
4
TRIG
SWEEP
+/-
CONFIG MENU
I
5
SOURCE
EDIT
EXIT ENTER
FCTN
LIMIT
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V PEAK
TERMINALS
FRONT/
REAR
Model 2440
Digital Multimeter
2. Select the multimeter DC current measuring function.
Input LO
Amps
1-14 Performance Verification Model 2440 5A SourceMeter® Service Manual
3. Press the Model 2440 SOURCE I key to source current, and make sure the source output is turned on.
4. Verify output current accuracy for the 10μA-1A range currents listed in Table 1-5. For each test point:
Select the correct source range.
Set the Model 2440 output current to the correct value.
Verify that the multimeter reading is within the limits given in the table.
5. Repeat the procedure for negative output currents with the same magnitudes as those listed in
Table 1-5.
6. Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
5A range accuracy
1. With the power off, connect the digital multimeter and the 1Ω resistor to the Model 2440 INPUT/OUTPUT jacks, as shown in
Figure 1-3
Connections for 5A range current verification tests
SWEEP
I
5
SOURCE
2440 5A SourceMeter
EDIT
EXIT ENTER
RANGE
AUTO
RANGE
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
MEAS
FCTN
I
Ω
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
1
REL
4- WIRE
SENSE
HI
75V
5V
PEAK
PEAK
®
LO
ON/OFF
OUTPUT
INPUT/
OUTPUT
TERMINALS
FRONT/
REAR
Figure 1-3.
75V PEAK
250V PEAK
Model 2440
Digital Multimeter
1Ω Resistor
Input HI
Input LO
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-15
2. Select the multimeter DC volts measuring function.
3. Press the Model 2440 SOURCE I key to source current, and make sure the source output is turned on.
4. Verify output current accuracy for the 5A range. Be sure to:
Select the 5A source range.
Set the Model 2440 output current to the correct 5A output value.
Verify that the multimeter reading is within the 5A range limits given in Table
1-5.
(Since the value of the 1Ω resistor value is assumed to be the same as its nominal value, the DMM voltage reading is the same as the sourced current.)
5. Repeat the procedure for a negative 5A current output value.
6. Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
Table 1-5
Output current accuracy limits
Model 2440
source range
10μA
100μA
1mA
10mA
100mA
1A 5A*
Model 2440
output current setting
10.0000μA
100.000μA
1.00000mA
10.0000mA
100.000mA
1.00000A
5.00000A
Output current limits
(1 year, 18°C-28°C)
9.9947 to 10.0053μA
99.949 to 100.051μA
0.99946 to 1.00054mA
9.9935 to 10.0065mA
99.914 to 100.086mA
0.99843 to 1.00157A
4.9896 to 5.0104A
Performance Verification
*See separate procedure for 5A range. DMM voltage reading is same as sourced current.
1-16 Performance Verification Model 2440 5A SourceMeter® Service Manual
Current measurement accuracy
Follow the steps below to verify that Model 2440 current measurement accuracy is within specified limits. The procedure involves applying accurate currents from the Model 2440 current source and then verifying that Model 2440 current measurements are within required limits.
10μA to 1A range accuracy
1. With the power off, connect the digital multimeter to the Model 2440 INPUT/ OUTPUT jacks, as shown in
2. Select the multimeter DC current function.
3. Set the Model 2440 to both source and measure current by pressing the SOURCE I and MEAS I keys, and make sure the source output is turned on.
4. Verify measure current accuracy for the 10μA-1A range currents listed in Table
1-6. For each measurement:
Select the correct source range.
Set the Model 2440 source output to the correct value as measured by the dig­ital multimeter.
Verify that the Model 2440 current reading is within the limits given in the table.
Figure 1-2.
NOTE It may not be possible to set the current source to the specified value. Use the
closest possible setting, and modify reading limits accordingly.
5. Repeat the procedure for negative calibrator currents with the same magnitudes as those listed in
6. Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
Table 1-6.
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-17
5A range accuracy
NOTE The 1Ω resistor should be characterized to within ±300ppm before verifying the
5A current measurement range. Use the 4-wire ohms function of the DMM to measure the resistance value, and then use that measured value to calculate the current during the measurement procedure.
1. With the power off, connect the 1Ω resistor and digital multimeter to the Model 2440 INPUT/OUTPUT jacks, as shown in
2. Select the multimeter DC volts function.
3. Set the Model 2440 to both source and measure current by pressing the SOURCE I and MEAS I keys, and make sure the source output is turned on.
4. Verify measurement current accuracy for the 5A range as follows:
Select the 5A source range.
Set the Model 2440 source output to the correct 5A value as measured by the digital multimeter.
Note the DMM voltage reading, and then calculate the current from the volt­age reading and characterized 1Ω resistance value as I = V/R, where V is the DMM voltage reading and R is the characterized resistance value.
Verify that the Model 2440 current reading is within the 5A limits given in the
Table 1-6.
NOTE It may not be possible to set the current source to the specified 5A value. Use the
closest possible setting, and modify reading limits accordingly.
Figure 1-3.
Performance Verification
5. Repeat the procedure for a negative 5A current.
6. Repeat the procedure using the rear panel INPUT/OUTPUT jacks. Be sure to select the rear panel jacks with the front panel TERMINALS key.
Table 1-6
Current measurement accuracy limits
Model 2440 source
and measure range*
10μA
100μA
1mA
10mA
100mA
1A 5A
*Measure range coupled to source range when simultaneously sourcing and measuring current.
** As measured by precision digital multimeter. Use closest possible value, and modify reading limits
accordingly if necessary.
*** Current calculated as follows: I = V/R, where V is the DMM voltage reading, and R is the character-
ized value of the 1Ω resistor.
Source current**
10.00000μA
100.000μA
1.00000mA
10.0000mA
100.000mA
1.00000A
5.00000A
Model 2440 current reading limits
(1 year, 18°C-28°C)
9.9966 to 10.0034μA
99.969 to 100.031μA
0.99967 to 1.00033mA
9.9959 to 10.0041mA
99.939 to 100.061mA
0.99883 to 1.00117A
4.9916 to 5.0084A***
1-18 Performance Verification Model 2440 5A SourceMeter® Service Manual
Resistance measurement accuracy
Use the following steps to verify that Model 2440 resistance measurement accuracy is within specified limits. This procedure involves applying accurate resistances from a resistance calibrator and then verifying that Model 2440 resistance measurements are within required limits.
CAUTION Before testing the 2Ω and 20Ω ranges make sure your resistance cali-
brator can safely handle the default test currents for those ranges (see Model 2440 and calibrator specifications). If not, use the CONFIG OHMS menu to select the MANUAL source mode, then set the source current to an appropriate safe value. When using the manual source mode, total resistance reading uncertainty includes both Source I and Measure V uncertainty (see specifications), and calculated reading limits should take the additional uncertainty into account.
If using the Fluke 5450A resistance calibrator, you cannot use the Auto Ohms mode of the Model 2440 to verify the 2Ω range. The 1A test current for the 2Ω range of the Model 2440 will damage the cali brator. On the Model 2440, use the CONFIG OHMS menu to select the MANUAL source mode, and then set the source (test) current to 100mA.
-
1. With the power off, connect the resistance calibrator to the Model 2440 INPUT/ OUTPUT and 4-WIRE SENSE jacks, as shown in 4-wire connections as shown.
Figure 1-4. Be sure to use the
Model 2440 5A SourceMeter® Service Manual Performance Verification 1-19
Figure 1-4
Connections for resistance accuracy verification
4- WIRE
INPUT/
SENSE
OUTPUT
HI
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
75V
PEAK
SWEEP
I
5
SOURCE
2440 5A SourceMeter
EDIT
EXIT ENTER
MEAS
FCTN
I
Ω
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
1
REL
RANGE
AUTO
RANGE
®
Model 2440
Resistance Calibrator
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V PEAK
TERMINALS
FRONT/
REAR
Output HI Sense HI
Output LO
Sense LO
2. Select the resistance calibrator external sense mode.
3. Configure the Model 2440 ohms function for the 4-wire sense mode as follows:
Press CONFIG then MEAS Ω. The instrument will display the following:
CONFIG OHMS
SOURCE SENSE-MODE GUARD
Select SENSE-MODE, and then press ENTER. The following will be displayed:
SENSE-MODE
2-WIRE 4-WIRE
Select 4-WIRE, and then press ENTER.
Press EXIT to return to normal display.
4. Press MEAS Ω to select the ohms measurement function, and make sure the source output is turned on.
5. Verify ohms measurement accuracy for each of the resistance values listed in Table
1-7. For each measurement:
Set the resistance calibrator output to the nominal resistance or closest avail­able value.
NOTE It may not be possible to set the resistance calibrator to the specified value. Use
the closest possible setting, and modify reading limits accordingly.
Select the appropriate ohms measurement range with the RANGE keys.
Performance Verification
1-20 Performance Verification Model 2440 5A SourceMeter® Service Manual
Verify that the Model 2440 resistance reading is within the limits given in the table.
6. Repeat the entire procedure using the rear panel INPUT/OUTPUT and 4-WIRE SENSE jacks. Be sure to select the rear panel jacks with the front panel TERMI
-
NALS key.
Table 1-7
Ohms measurement accuracy limits
Calibrator
Model 2440 range
2Ω
20Ω
200Ω
2kΩ
20kΩ
200kΩ
2MΩ
20MΩ
* Nominal resistance value.
** Reading limits based on Model 2440 normal accuracy specifications and nominal resistance values. If
actual resistance values differ from nominal values shown, recalculate reading limits using actual calibra tor resistance values and Model 2440 normal accuracy specifications. See Verification limits earlier in this section for details.
resistance*
1.9Ω
19Ω
190Ω
1.9kΩ
19kΩ
190kΩ
1.9MΩ
19MΩ
Model 2440 resistance reading limits**
(1 year, 18°C-28°C)
1.89647 to 1.90353Ω
18.9780 to 19.0220Ω
189.818 to 190.182Ω
1.89837 to 1.90163kΩ
18.9856 to 19.0144kΩ
189.837 to 190.163kΩ
1.89761 to 1.90239MΩ
18.9781 to 19.0219MΩ
-
2
Calibration
Calibration
2-2 Calibration Model 2440 5A SourceMeter® Service Manual
Introduction
Use the procedures in this section to calibrate the Model 2440. These procedures require accurate test equipment to measure precise DC voltages and currents. Calibration can be performed either from the front panel or by sending SCPI calibration commands over the IEEE-488 bus or RS-232 port with the aid of a computer.
WARNING This information in this section is intended for qualified service per-
sonnel only. Do not attempt these procedures unless you are qualified to do so. Some of these procedures may expose you to hazardous volt ages.
Environmental conditions
Temperature and relative humidity
Conduct the calibration procedures at an ambient temperature of 18-28°C (65-82°F) with relative humidity of less than 70% unless otherwise noted.
-
Warm-up period
Allow the Model 2440 to warm up for at least one hour before performing calibration.
If the instrument has been subjected to temperature extremes (those outside the ranges stated above), allow additional time for the instrument’s internal temperature to stabilize. Typically, allow one extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manufacturer.
Line power
The Model 2440 requires a line voltage of 100 to 240V at line frequency of 50 to 60Hz. The instrument must be calibrated while operating from a line voltage within this range.
Model 2440 5A SourceMeter® Service Manual Calibration 2-3
Calibration considerations
When performing the calibration procedures:
Make sure that the test equipment is properly warmed up and connected to the Model 2440 front panel INPUT/OUTPUT jacks. Also be certain that the front panel jacks are selected with the TERMINALS switch.
Always allow the source signal to settle before calibrating each point.
Do not connect test equipment to the Model 2440 through a scanner or other switching equipment.
If an error occurs during calibration, the Model 2440 will generate an appropriate error message. See
Appendix B for more information.
WARNING The maximum common-mode voltage (voltage between LO and chas-
CAUTION The maximum voltage between INPUT/OUTPUT HI and LO or
Calibration cycle
Perform calibration at least once a year to ensure the unit meets or exceeds its specifications.
NOTE Calibration constants are stored in volatile memories of the Model 2440, which
are protected by a replaceable battery when power is off. Typical life for the battery is approximately 10 years, but the battery should be replaced if the volt age drops below 2.5V regardless of age. See Section 4 for battery replacement details.
Calibration
sis ground) is 40V peak. Exceeding this value may cause a breakdown in insulation, creating a shock hazard.
4-WIRE SENSE HI and LO is 42V peak. The maximum voltage between INPUT/OUTPUT HI and 4-WIRE SENSE HI or between INPUT/OUTPUT LO and 4-WIRE SENSE LO is 5V. Exceeding these voltage values may result in instrument damage.
-
2-4 Calibration Model 2440 5A SourceMeter® Service Manual
Recommended calibration equipment
Table 2-1 lists the recommended equipment for the calibration procedures. You can use
alternate equipment as long that equipment has specifications at least as good as those listed in the table. For optimum calibration accuracy, test equipment specifications should be at least four times better than corresponding Model 2440 specifications.
Table 2-1
Recommended calibration equipment
Description Manufacturer/Model Specifications
Digital Multimeter Hewlett Packard
HP3458A
Precision Resistor** Isotec RUG-Z-1R00-0.1 1Ω, ±0.1%, 100W
* 90-day, full-range accuracy specifications of ranges required for various measurement points.
** Necessary for calibration of 5A current range. Resistor must be characterized to ±300ppm or better using recommended DMM
before calibrating 5A range.
DC Voltage*
DC Current*
1V: 10V: 100V
10μA: 100μA: 1mA: 10mA: 100mA: 1A:
±5.6ppm ±4.3ppm ±6.3ppm
±25ppm ±23ppm ±20ppm ±20ppm ±35ppm ±110ppm
1Ω resistor characterization
The 1Ω resistor must be characterized to ±300ppm or better before calibrating the 5A cur­rent range. Use the 4-wire ohms function of the DMM recommended in Table 2-1 to mea­sure the resistance value, and then use that measured value to calculate the current during the 5A current range calibration procedure.
Model 2440 5A SourceMeter® Service Manual Calibration 2-5
Unlocking calibration
Before performing calibration, you must first unlock calibration by entering or sending the calibration password as explained in the following paragraphs.
Unlocking calibration from the front panel
1. Press the MENU key, and then choose CAL, and press ENTER. The instrument will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
ß SAVE LOCK CHANGE-PASSWORD
2. Select UNLOCK, and then press ENTER. The instrument will display the following:
PASSWORD:
Use ß, ©, ¹, ƒ, ENTER or EXIT.
3. Use the up and down RANGE keys to select the letter or number, and use the left and right arrow keys to choose the position. (Press down RANGE for letters; up RANGE for numbers.) Enter the present password on the display. (Front panel default: 002440.)
4. Once the correct password is displayed, press the ENTER key. If the password was correctly entered, the following message will be displayed:
CALIBRATION UNLOCKED
Calibration can now be executed.
5. Press EXIT to return to normal display. Calibration will be unlocked and assume the states summarized in below with calibration unlocked will result in an error +510, “Not permitted with cal unlocked.”
Table 2-2. Attempts to change any of the settings listed
Calibration
NOTE With calibration unlocked, the sense function and range track the source func-
tion and range. That is, when :SOUR:FUNC is set to VOLT, the :SENS:FUNC setting will be 'VOLT:DC'. When :SOUR:FUNC is set to CURR, the :SENS:FUNC setting will be 'CURR:DC'. A similar command coupling exists for :SOUR:VOLT:RANG/:SENS:VOLT:RANG and SOUR:CURR:RANG:SENS:CURR:RANG.
2-6 Calibration Model 2440 5A SourceMeter® Service Manual
Table 2-2
Calibration unlocked states
Mode State Equivalent remote command
Concurrent Functions Sense Function Sense Volts NPLC Sense Volts Range Sense Current NPLC Sense Current Range Filter Count Filter Control Filter Averaging Source V Mode Volts Autorange Source I Mode Current Autorange Autozero Trigger Arm Count Trigger Arm Source Trigger Count Trigger Source
OFF Source
1.0 Source V
1.0 Source I 10 REPEAT ON FIXED OFF FIXED OFF ON 1 Immediate 1 Immediate
Unlocking calibration by remote
To unlock calibration via remote, send the following command:
:CAL:PROT:CODE '<password>'
:SENS:FUNC:CONC OFF :SENS:FUNC <source_ function> :SENS:VOLT:NPLC 1.0 :SENS:VOLT:RANG <source_V_range> :SENS:CURR:NPLC 1.0 :SENS:CURR:RANG <source_ I_ range> :SENS:AVER:COUN 10 :SENS:AVER:TCON REPeat :SENS:AVER:STAT ON :SOUR:VOLT:MODE FIXED :SOUR:VOLT:RANG:AUTO OFF :SOUR:CURR:MODE FIXED :SOUR:CURR:RANG:AUTO OFF :SYST:AZERO ON :ARM:COUNT 1 :ARM:SOUR IMMediate :TRIG:COUNT 1 :TRIG:SOUR IMMediate
For example, the following command uses the default password:
:CAL:PROT:CODE 'KI002440'
Model 2440 5A SourceMeter® Service Manual Calibration 2-7
Changing the password
The default password may be changed from the front panel or via remote as discussed below.
Changing the password from the front panel
Follow the steps below to change the password from the front panel:
1. Press the MENU key, then choose CAL and press ENTER. The instrument will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES ©
ß SAVE LOCK CHANGE-PASSWORD
2. Select UNLOCK, then enter the password. (Default: 002440.)
3. Select CHANGE-PASSWORD, and then press ENTER. The instrument will dis­play the following:
New Pwd: 002440
Use ß, ©, ¹, ƒ, ENTER or EXIT.
4. Using the range keys, and the left and right arrow keys, enter the new password on the display.
5. Once the desired password is displayed, press the ENTER key to store the new password.
Calibration
Changing the password by remote
To change the calibration password by remote, first send the present password, and then send the new password. For example, the following command sequence changes the pass word from the 'KI002440' remote default to 'KI_CAL':
:CAL:PROT:CODE 'KI002440' :CAL:PROT:CODE 'KI_CAL'
You can use any combination of letters and numbers up to a maximum of eight characters.
NOTE If you change the first two characters of the password to something other than
“KI”, you will not be able to unlock calibration from the front panel.
-
2-8 Calibration Model 2440 5A SourceMeter® Service Manual
Resetting the calibration password
If you lose the calibration password, you can unlock calibration by shorting together the CAL pads, which are located on the display board. Doing so will also reset the password to the factory default (KI002440).
See Section 5 for details on disassembling the unit to access the CAL pads. Refer to the display board component layout drawing at the end of Section 6 for the location of the CAL pads.
Viewing calibration dates and calibration count
When calibration is locked, only the UNLOCK and VIEW-DATES selections will be accessible in the calibration menu. To view calibration dates and calibration count at any time:
1. From normal display, press MENU, select CAL, and then press ENTER. The unit will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
2. Select VIEW-DATES, and then press ENTER. The Model 2440 will display the next and last calibration dates and the calibration count as in the following example:
NEXT CAL: 02/15/2002
Last calibration: 02/15/2001 Count: 0001
Calibration errors
The Model 2440 checks for errors after each calibration step, minimizing the possibility that improper calibration may occur due to operator error.
Front panel error reporting
If an error is detected during comprehensive calibration, the instrument will display an appropriate error message (see calibration step that caused the error.
Appendix B). The unit will then prompt you to repeat the
Model 2440 5A SourceMeter® Service Manual Calibration 2-9
Remote error reporting
You can detect errors while in remote by testing the state of EAV (Error Available) bit (bit
2) in the status byte. (Use the *STB? query to request the status byte.) Query the instru ment for the type of error by using the appropriate :SYST:ERR? query. The Model 2440 will respond with the error number and a text message describing the nature of the error. See
Appendix B for details.
-
Front panel calibration
The front panel calibration procedure described below calibrates all ranges of both the current and voltage source and measure functions. Note that each range is separately cali brated by repeating the entire procedure for each range.
Step 1: Prepare the Model 2440 for calibration
1. Turn on the Model 2440 and the digital multimeter, and allow them to warm up for at least one hour before performing calibration.
2. Press the MENU key, then choose CAL and press ENTER. Select UNLOCK, and then press ENTER. The instrument will display the following:
PASSWORD:
Use ß, ©, ¹, ƒ, ENTER or EXIT.
3. Use the up and down range keys to select the letter or number, and use the left and right arrow keys to choose the position. Enter the present password on the display. (Front panel default: 002440.) Press ENTER to complete the process.
4. Press EXIT to return to normal display. Instrument operating states will be set as summarized in Table 2-2.
-
Calibration
2-10 Calibration Model 2440 5A SourceMeter® Service Manual
Step 2: Voltage calibration
Perform the steps below for each voltage range, using Table 2-3 as a guide.
1. Connect the Model 2440 to the digital multimeter, as shown in Figure 2-1. Select the multimeter DC volts measurement function.
NOTE The 2-wire connections shown assume that remote sensing is not used. Remote
sensing may be used, if desired, but it is not essential when using the recom mended digital multimeter.
Figure 2-1
Voltage calibration test connections
SWEEP
I
5
SOURCE
2440 5A SourceMeter
EDIT
EXIT ENTER
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
MEAS
FCTN
I
Ω
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
1
REL
RANGE
AUTO
RANGE
4- WIRE
INPUT/
SENSE
OUTPUT
HI
75V
PEAK
®
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V PEAK
TERMINALS
FRONT/
REAR
-
Model 2440
Input HI
Digital Multimeter
Input LO
2. From normal display, press the SOURCE V key.
3. Press the EDIT key to select the source field (cursor flashing in source display field), and then use the down RANGE key to select the 200mV source range.
4. From normal display, press MENU.
5. Select CAL, and then press ENTER. The unit will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES ©
ß SAVE LOCK CHANGE-PASSWORD
Model 2440 5A SourceMeter® Service Manual Calibration 2-11
6. Select EXECUTE, and then press ENTER. The instrument will display the follow­ing message:
V-CAL
Press ENTER to Output +200.00mV
7. Press ENTER. The Model 2440 will source +200mV and simultaneously display the following:
DMM RDG: +200.0000mV
Use ß, ©, ¹, ƒ, ENTER, or EXIT.
8. Note and record the DMM reading, and then adjust the Model 2440 display to agree exactly with the actual DMM reading. Use the up and down arrow keys to select the digit value, and use the left and right arrow keys to choose the digit posi tion (or use the number keys, 0-9, +/-). Note that the display adjustment range is within ±10% of the present range.
9. After adjusting the display to agree with the DMM reading, press ENTER. The instrument will then display the following:
V-CAL Press ENTER to Output +000.00mV
10. Press ENTER. The Model 2440 will source 0mV and at the same time display the following:
DMM RDG: +000.0000mV
Use ß, ©, ¹, ƒ, ENTER, or EXIT.
11. Note and record the DMM reading, and then adjust the Model 2440 display to agree with the actual DMM reading. Note that the display value adjustment limits are within ±1% of the present range.
12. After adjusting the display value to agree with the DMM reading, press ENTER. The unit will then display the following:
V-CAL
Press ENTER to Output -200.00mV
13. Press ENTER. The Model 2440 will source -200mV and display the following:
DMM RDG: -200.0000mV
Use ß, ©, ¹, ƒ, ENTER, or EXIT.
14. Note and record the DMM reading, and then adjust the Model 2440 display to agree with the DMM reading. Again, the maximum display adjustment is within ± 10% of the present range.
15. After adjusting the display value to agree with the DMM reading, press ENTER and note that the instrument displays:
V-CAL
Press ENTER to Output -000.00mV
16. Press ENTER The Model 2440 will source -0mV and simultaneously display the following:
DMM RDG: +000.0000mV
Use ß, ©, ¹, ƒ, ENTER or EXIT.
-
Calibration
2-12 Calibration Model 2440 5A SourceMeter® Service Manual
17. Note and record the DMM reading, and then adjust the display to agree with the DMM reading. Once again, the maximum adjustment is within ±1% of the present range.
18. After adjusting the display to agree with the DMM reading, press ENTER to com­plete calibration of the present range.
19. Press EXIT to return to normal display, and then select the 2V source range. Repeat steps 2 through 18 for the 2V range.
20. After calibrating the 2V range, repeat the entire procedure for the 10V and 40V ranges using
Table 2-3 as a guide. Be sure to select the appropriate source range
with the EDIT and RANGE keys before calibrating each range.
21. Press EXIT as necessary to return to normal display.
Table 2-3
Front panel voltage calibration
Source range*
Source voltage
0.2V +200.00mV +000.00mV
-200.00mV
-000.00mV
2V +2.0000V
+0.0000V
-2.0000V
-0.0000V
10V +10.000V
+00.000V
-10.000V
-00.000V
40V +40.000V
+00.000V
-40.000V
-00.000V
**Use EDIT and RANGE keys to select source range.
**Multimeter reading used in corresponding calibration step. See procedure.
Multimeter
voltage reading**
___________ mV ___________ mV ___________ mV ___________ mV
___________ V ___________ V ___________ V ___________ V
___________ V ___________ V ___________ V ___________ V
___________ V ___________ V ___________ V ___________ V
Model 2440 5A SourceMeter® Service Manual Calibration 2-13
Step 3: Current calibration
Perform the steps below for each current range, using Table 2-4 as a guide.
1. Connect the Model 2440 to the digital multimeter, as shown in Figure 2-2. Select the multimeter DC current measurement function.
Figure 2-2
10μA to 1A range current calibration test connections
4- WIRE
INPUT/
SENSE
OUTPUT
HI
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
75V
PEAK
SWEEP
I
5
SOURCE
2440 5A SourceMeter
EDIT
EXIT ENTER
MEAS
FCTN
I
Ω
230
FILTER
89
STORE
LIMIT
RECALL
V
4
TRIG
+/-
CONFIG MENU
1
REL
RANGE
AUTO
RANGE
®
Model 2440
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V PEAK
TERMINALS
FRONT/
REAR
Input LO
Calibration
Digital Multimeter
Amps
2. From normal display, press the SOURCE I key.
3. Press the EDIT key to select the source display field, and then use the down RANGE key to select the 10μA source range.
4. From normal display, press MENU.
5. Select CAL, and then press ENTER. The unit will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES ©
ß SAVE LOCK CHANGE-PASSWORD
6. Select EXECUTE, and then press ENTER. The instrument will display the follow­ing message:
I-CAL
Press ENTER to Output +10.000μA
2-14 Calibration Model 2440 5A SourceMeter® Service Manual
7. Press ENTER. The Model 2440 will source +10μA and simultaneously display the following:
DMM RDG: +10.00000μA Use ß, ©, ¹, ƒ, ENTER, or EXIT.
8. Note and record the DMM reading, and then adjust the Model 2440 display to agree exactly with the actual DMM reading. Use the up and down arrow keys to select the digit value, and use the left and right arrow keys to choose the digit posi tion (or use the number keys, 0-9, +/-). Note that the display adjustment range is within ±10% of the present range.
9. After adjusting the display to agree with the DMM reading, press ENTER. The instrument will then display the following:
I-CAL
Press ENTER to Output +00.000μA
10. Press ENTER. The Model 2440 will source 0mA and at the same time display the following:
DMM RDG: +00.00000μA
Use ß, ©, ¹, ƒ, ENTER, or EXIT.
11. Note and record the DMM reading, and then adjust the Model 2440 display to agree with the actual DMM reading. Note that the display value adjustment limits are within ±1% of the present range.
12. After adjusting the display value to agree with the DMM reading, press ENTER. The unit will then display the following:
I-CAL
Press ENTER to Output -10.000μA
13. Press ENTER. The Model 2440 will source -10μA and display the following:
DMM RDG: -10.00000μA
Use ß, ©, ¹, ƒ, ENTER, or EXIT.
14. Note and record the DMM reading, and then adjust the Model 2440 display to agree with the DMM reading. Again, the maximum display adjustment is within ±10% of the present range.
15. After adjusting the display value to agree with the DMM reading, press ENTER and note that the instrument displays:
I-CAL
Press ENTER to Output -00.000μA
16. Press ENTER The Model 2440 will source -0μA and simultaneously display the following:
DMM RDG: +00.00000μA
Use ß, ©, ¹, ƒ, ENTER, or EXIT.
17. Note and record the DMM reading, and then adjust the display to agree with the DMM reading. Once again, the maximum adjustment is within ±1% of the present range.
-
Model 2440 5A SourceMeter® Service Manual Calibration 2-15
18. After adjusting the display to agree with the DMM reading, press ENTER to com­plete calibration of the present range.
19. Press EXIT to return to normal display, and then select the 100μA source range use the EDIT and up RANGE keys. Repeat steps 2 through 18 for the 100μA range.
20. After calibrating the 100μA range, repeat the entire procedure for the 1mA through 1A ranges using
Table 2-4 as a guide. Be sure to select the appropriate source
range with the EDIT and up RANGE keys before calibrating each range.
21. After calibrating the 1A range, connect the 1Ω characterized resistor and DMM to the Model 2440 INPUT/OUTPUT jacks, as shown in
Figure 2-3.
22. Select the DMM DC volts function.
23. Repeat steps 2 through 18 for the 5A range using Table 2-4 as a guide. When enter­ing the DMM reading, use the calculated current as follows: I = V/R, where V is the DMM voltage reading, and R is the characterized value of the 1Ω resistor.
Figure 2-3
5A range current calibration test connections
4- WIRE
INPUT/
SENSE
OUTPUT
HI
EDIT
DISPLAY
TOGGLE
POWER
V
LOCAL
67
DIGITS SPEED
75V
PEAK
SWEEP
I
5
SOURCE
2440 5A SourceMeter
EDIT
EXIT ENTER
MEAS
FCTN
I
Ω
230
FILTER
89
RECALL
STORE
V
4
TRIG
LIMIT
+/-
CONFIG MENU
1
REL
RANGE
AUTO
RANGE
®
ON/OFF
OUTPUT
75V
5V
PEAK
PEAK
LO
250V PEAK
TERMINALS
FRONT/
REAR
Calibration
Model 2440
Digital Multimeter
1Ω Resistor
Input HI
Input LO
2-16 Calibration Model 2440 5A SourceMeter® Service Manual
Table 2-4
Front panel current calibration
Multimeter
Source range* Source current
current reading**
10μA +10.000μA
+00.000μA
-10.000μA
-00.000μA
100μA +100.00μA
+000.00μA
-100.00μA
-000.00μA
1mA +1.0000mA
+0.0000mA
-1.0000mA
-0.0000mA
10mA +10.000mA
+00.000mA
-10.000mA
-00.000mA
100mA +100.00mA
+000.00mA
-100.00mA
-000.00mA
1A +1.0000A
+0.0000A
-1.0000A
-0.0000A
___________ μA ___________ μA ___________ μA ___________ μA
___________ μA ___________ μA ___________ μA ___________ μA
___________ mA ___________ mA ___________ mA ___________ mA
___________ mA ___________ mA ___________ mA ___________ mA
___________ mA ___________ mA ___________ mA ___________ mA
___________ A ___________ A ___________ A ___________ A
5A*** +5.0000A
+0.0000A
-5.0000A
-0.0000A
*Use EDIT and RANGE keys to select source range.
**Multimeter reading used in corresponding calibration step. See procedure.
***Current calculated as follows: I = V/R, where V is the DMM reading, and R is the
characterized value of the 1Ω resistor.
___________ A ___________ A ___________ A ___________ A
Model 2440 5A SourceMeter® Service Manual Calibration 2-17
Step 4: Enter calibration dates and save calibration
NOTE For temporary calibration without saving new calibration constants, proceed to
“Step 5: Lock out calibration”.
1. From normal display, press MENU.
2. Select CAL, and then press ENTER. The Model 2440 will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES ©
ß SAVE LOCK CHANGE-PASSWORD
3. Select SAVE, and then press ENTER. The instrument will display the following message:
SAVE CAL Press ENTER to continue; EXIT to abort calibration sequence.
4. Press ENTER. The unit will prompt you for the calibration date:
CAL DATE: 2/15/2001
Use ß, ©, ¹, ƒ, ENTER, or EXIT.
5. Change the displayed date to today's date, and then press the ENTER key. Press ENTER again to confirm the date.
6. The unit will then prompt for the calibration due date:
NEXT CAL: 2/15/2002
Use ß, ©, ¹, ƒ, ENTER, or EXIT.
7. Set the calibration due date to the desired value, and then press ENTER. Press ENTER again to confirm the date.
8. Once the calibration dates are entered, calibration is complete, and the following message will be displayed:
CALIBRATION COMPLETE
Press ENTER to confirm; EXIT to abort
9. Press ENTER to save the calibration data (or press EXIT to abort without saving calibration data.) The following message will be displayed:
CALIBRATION SUCCESS
Press ENTER or EXIT to continue.
10. Press ENTER or EXIT to complete process.
Calibration
Step 5: Lock out calibration
1. From normal display, press MENU.
2. Select CAL, and then press ENTER. The Model 2440 will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES ©
ß SAVE LOCK CHANGE-PASSWORD
3. Select LOCK, and then press ENTER. The instrument will display the following message:
CALIBRATION LOCKED.
Press ENTER or EXIT to continue.
4. Press ENTER or EXIT to return to normal display.
2-18 Calibration Model 2440 5A SourceMeter® Service Manual
Remote calibration
Use the following procedure to perform remote calibration by sending SCPI commands over the IEEE-488 bus or RS-232 port. The remote commands and appropriate parameters are separately summarized for each step.
Remote calibration commands
Table 2-5 summarizes remote calibration commands. For a more complete description of
these commands, refer to Appendix B.
Table 2-5
Remote calibration command summary
Command Description
:CALibration
:PROTected
:CODE '<password>'
:COUNT? :SAVE :LOCK :LOCK? :DATE <y>,<m>,<d> :DATE? :NDUE <y>,<m>,<d> :NDUE? :SENSe <nrf>
:DATA?
:SOURce <nrf>
:DATA?
* Calibration data will not be saved if:
1. Calibration was not unlocked with :CODE command.
2. Invalid data exists. (For example, cal step failed or was aborted.)
3. Incomplete number of cal steps were performed. (For example, omitting a negative full-scale step). Ranges that calibrated successfully will be saved if calibration is unlocked. Ranges that failed will not be saved.
Calibration subsystem. Cal commands protected by password. Unlock cal; changes password if cal is already unlocked.
(Default password: KI002440.) Query number of times 2440 has been calibrated. Save calibration data to EEPROM.* Lock calibration, inhibit SAVE command operation. Request cal lock status. Program calibration year, month, day. Query calibration year, month, day. Program calibration due year, month, day. Query calibration due year, month, day. Calibrate active measure range. (See Table 2-6 parameters.) Query measurement cal constants for active range. Calibrate active source range. (See Table 2-7 parameters.) Query source cal constants for active range.
Model 2440 5A SourceMeter® Service Manual Calibration 2-19
Recommended calibration parameters
The maximum calibration command parameter ranges are: 75% to 150% of full scale for positive and negative full scale calibration points; ± zero calibration steps have ±50% of full scale for valid entry ranges. However, for optimum calibration, it is recommended that you use calibration points within the ranges listed in each sense range requires three parameters: zero, negative full scale, and positive full scale. Similarly, each source range requires four parameters: two zero parameters, a posi tive full-scale parameter, and a negative full-scale parameter.
Table 2-6
Recommended :CALibration:PROTected:SENse parameter ranges
Table 2-6 and Table 2-7. Note that
-
Sense range
0.2V 2V 10V 40V
10μA 100μA 1mA 10mA 100mA 1A 5A
Note: Parameter steps for each range may be performed in any order, but all three parameter steps for each
range must be completed. For optimum calibration, use parameters within recommended limits.
First parameter
(zero)
-0.002 to +0.002
-0.02 to +0.02
-0.1 to +0.1
-0.5 to +0.5
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-5E-2 to +5E-2
Second parameter
(negative full scale)
-0.18 to -0.22
-1.8 to -2.2
-9 to -11
-35 to -42
-9E-6 to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-9E-3 to -11E-3
-90E-3 to -110E-3
-0.9 to -1.1
-4.75 to -5.25
Third parameter
(positive full scale)
+0.18 to +0.22
+1.8 to +2.2
+9 to +11
+35 to +42
+9E-6 to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+9E-3 to +11E-3
+90E-3 to +110E-3
+0.9 to +1.1
+4.75 to +5.25
Calibration
2-20 Calibration Model 2440 5A SourceMeter® Service Manual
Table 2-7
Recommended :CALibration:PROTected:SOURce parameter ranges
Second parameter
Source range
0.2V 2V 10V 40V
10μA 100μA 1mA 10mA 100mA 1A 5A
Note: Parameter steps for each range may be performed in any order, but all four parameter steps for each range must be completed.
For optimum calibration, use parameters within recommended limits.
(negative full scale)
-0.18 to -0.22
-1.8 to -2.2
-9 to -11
-35 to -42
-9E-6 to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-9E-3 to -1E-3
-90E-3 to -110E-3
-0.9 to -1.1
-4.75 to -5.25
Second parameter
(negative zero)
-0.002 to +0.002
-0.02 to +0.02
-0.1 to +0.1
-0.5 to +0.5
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-5E-2 to +5E-2
Third parameter
(positive full scale)
+0.18 to +0.22
+1.8 to +2.2
+9 to +11
+35 to +42
+9E-6 to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+9E-3 to +11E-3
+90E-3 to +110E-3
+0.9 to +1.1
+4.75 to +5.25
Fourth parameter
(positive zero)
-0.002 to +0.002
-0.02 to +0.02
-0.1 to +0.1
-0.5 to +0.5
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-5E-2 to +5E-2
Remote calibration procedure
Step 1: Prepare the Model 2440 for calibration
1. Connect the Model 2440 to the controller IEEE-488 interface or RS-232 port using a shielded interface cable.
2. Turn on the Model 2440 and the test equipment, and allow them to warm up for at least one hour before performing calibration.
3. If you are using the IEEE-488 interface, make sure the primary address of the Model 2440 is the same as the address specified in the program you will be using to send commands. (Use the MENU key and the COMMUNICATION menu to access the IEEE-488 address.)
Model 2440 5A SourceMeter® Service Manual Calibration 2-21
Step 2: Voltage calibration
1. Connect the Model 2440 to the digital multimeter (see Figure 2-1), and select the multimeter DC volts function.
2. Send the commands summarized in Table 2-8 in the order listed to initialize volt­age calibration. (When the :CAL:PROT:CODE command is sent, the instrument will assume the operating states listed in
3. Perform the range calibration steps listed in Table 2-9 for all ranges. For each range:
Send the :SOUR:VOLT:RANG command to select the source and sense range being calibrated. For example, for the 2V range, the following command would be sent:
:SOUR:VOLT:RANG 2
Program the source to output the negative full-range value using the :SOUR:VOLT command. For example:
:SOUR:VOLT -2
Note and record the multimeter reading.
Use the multimeter reading as the parameter for the :CAL:PROT:SOUR and
:CAL:PROT:SENS commands. For example, a typical value for the 2V range would be:
:CAL:PROT:SOUR -1.998 :CAL:PROT:SENS -1.998
Program the voltage source for 0V output using the :SOUR:VOLT 0.0 command.
Note the multimeter reading.
Send the source and sense calibration commands using the multimeter reading for the parameter. For example:
:CAL:PROT:SOUR 1E-3 :CAL:PROT:SENS 1E-3
Set the source to the positive full-range value using the :SOUR:VOLT com­mand. For example:
:SOUR:VOLT 2
Note and record the multimeter reading.
Send the source and sense commands using the multimeter reading as the parameter. For example:
:CAL:PROT:SOUR 1.997 :CAL:PROT:SENS 1.997
Send the SOUR:VOLT 0.0 command to set the source voltage to 0V.
Note and record the multimeter reading.
Send the :CAL:PROT:SOUR command using the multimeter reading as the command parameter. For example:
:CAL:PROT:SOUR -1.02E-3
Table 2-2.)
Calibration
2-22 Calibration Model 2440 5A SourceMeter® Service Manual
Table 2-8
Voltage calibration initialization commands
Command Description
*RST :SOUR:FUNC VOLT :SENS:CURR:PROT 0.1 :SENS:CURR:RANG 0.1 :SOUR:VOLT:PROT MAX :SYST:RSEN OFF :CAL:PROT:CODE 'KI002440' :OUTP:STAT ON
*Remote sensing may be used if desired, but is not essential when using the recommended digital
multimeter. To use remote sensing, send :SYST:RSEN ON.
Restore GPIB defaults. Activate voltage source. Current limit when voltage source is active. Make sure 1A range is not active. Maximum allowable source voltage. Disable remote sensing.* Unlock cal. Turn source on.
Table 2-9
Voltage range calibration commands
Step Command/procedure* Description
1
:SOUR:VOLT:RANGE <Range>
2
:SOUR:VOLT -<Source_value>
3
Take DMM reading.
4
:CAL:PROT:SOUR <DMM_Reading>
5
Check 2440 for errors.
6
:CAL:PROT:SENS <DMM_Reading>
7
Check 2440 for errors.
8
:SOUR:VOLT 0.0
9
Take DMM reading.
10
:CAL:PROT:SOUR <DMM_Reading>
11
Check 2440 for errors.
12
:CAL:PROT:SENS <DMM_Reading>
13
Check 2440 for errors.
14
:SOUR:VOLT +<Source_value>
15
Take DMM reading.
16
:CAL:PROT:SOUR <DMM_Reading>
17
Check 2440 for errors.
18
:CAL:PROT:SENS <DMM_Reading>
19
Check 2440 for errors.
20
:SOUR:VOLT 0.0
21
Take DMM reading.
22
:CAL:PROT:SOUR <DMM_Reading>
*1. Perform complete procedure for each range, where <Range> = 0.2, 2, 10, and 40, and <Source_value> = 0.2, 2, 10, and 40.
2. <DMM_Reading> parameter is multimeter reading from previous step.
3. Use :SYST:ERR? query to check for errors.
Select source range. Establish negative full-range polarity. Read actual output value. Calibrate sense function negative full scale.
Calibrate source function negative full scale.
Set output to 0V. Read actual output value. Calibrate sense function negative zero.
Calibration source function negative zero.
Establish positive full-range polarity. Read actual output value Calibrate sense function positive full scale.
Calibrate source function positive full scale.
Set output to 0V. Read actual output value. Calibrate source positive zero.
Model 2440 5A SourceMeter® Service Manual Calibration 2-23
Step 3: Current calibration
1. Connect the Model 2440 to the digital multimeter (see Figure 2-2), and select the multimeter DC current function.
2. Send the commands summarized in Table 2-10 in the order listed to initialize cur- rent calibration.
3. Calibrate the 10μA to 1A current ranges using the procedure summarized in Table
2-11. For each range:
Send the :SOUR:CURR:RANG command to select the source and sense range being calibrated. For example, for the 1mA range, the command is:
:SOUR:CURR:RANG 1E-3
Program the source to output the negative full-range value using the
:SOUR:CURR command. For example:
:SOUR:CURR -1E-3
Note and record the multimeter reading.
Use the multimeter reading as the parameter for the :CAL:PROT:SOUR and :CAL:PROT:SENS commands. For example, a typical value for the 1mA range would be:
:CAL:PROT:SOUR -1.025E-3 :CAL:PROT:SENS -1.025E-3
Program the current source for 0A output using the :SOUR:CURR 0.0 command.
Note the multimeter reading.
Send the source and sense calibration commands using the multimeter reading for the parameter. For example:
:CAL:PROT:SOUR 1E-6 :CAL:PROT:SENS 1E-6
Set the source to the positive full-range value using the :SOUR:CURR com­mand. For example, for the 1mA range:
:SOUR:CURR 1E-3
Note and record the multimeter reading.
Send the source and sense commands using the multimeter reading as the parameter. For example:
:CAL:PROT:SOUR 1.03E-3 :CAL:PROT:SENS 1.03E-3
Send the :SOUR:CURR 0.0 command to set the source current to 0A.
Note and record the multimeter reading.
Send the :CAL:PROT:SOUR command using the multimeter reading as the command parameter. For example:
:CAL:PROT:SOUR 1E-6
Calibration
2-24 Calibration Model 2440 5A SourceMeter® Service Manual
4. Connect the 1Ω resistor and DMM to the Model 2440 INPUT/OUPUT jacks, as shown in
Figure 2-3. Select the DMM DC volts function.
5. Repeat step 3 for the 5A range using the calculated current as follows: I = V/R, where V is the DMM voltage reading, and R is the characterized value of the 1Ω resistor.
Table 2-10
Current calibration initialization commands
Command Description
:SOUR:FUNC CURR :SENS:VOLT:PROT 10 :SENS:VOLT:RANG 10 :OUTP:STAT ON
Select source current mode. Voltage limit when current source is active. Make sure 40V range is not active. Turn source on.
Table 2-11
Current range calibration commands
Step Command/procedure* Description
1 2 3 4 5 6 7 8
9 10 11 12 13 14 15 16 17 18 19 20 21 22
*1. Perform complete procedure for each range, where <Range> and <Source_value> = 10E-6, 100E-6, 1E-3,
10E-3, 100E-3, 1, or 5.
2. <DMM_Reading> parameter is multimeter reading from previous step.
3. Use :SYST:ERR? query to check for errors.
:SOUR:CURR:RANGE <Range> :SOUR:CURR -<Source_value> Take DMM reading. :CAL:PROT:SOUR <DMM_Reading> Check 2440 for errors. :CAL:PROT:SENS <DMM_Reading> Check 2440 for errors. :SOUR:CURR 0.0 Take DMM reading. :CAL:PROT:SOUR <DMM_Reading> Check 2440 for errors. :CAL:PROT:SENS <DMM_Reading> Check 2440 for errors. :SOUR:CURR +<Source_value> Take DMM reading. :CAL:PROT:SOUR <DMM_Reading> Check 2440 for errors. :CAL:PROT:SENS <DMM_Reading> Check 2440 for errors. :SOUR:CURR 0.0 Take DMM reading. :CAL:PROT:SOUR <DMM_Reading>
Select source range. Establish negative full-range polarity. Read actual output value. Calibrate sense function negative full scale.
Calibrate source function negative full scale.
Set output to 0A. Read actual output value. Calibrate sense function negative zero.
Calibration source function negative zero.
Establish positive full-range polarity. Read actual output value. Calibrate sense function positive full scale.
Calibrate source function positive full scale.
Set output to 0A. Read actual output value. Calibrate source positive zero.
Model 2440 5A SourceMeter® Service Manual Calibration 2-25
Step 4: Program calibration dates
Use following commands to set the calibration date and calibration due date:
:CAL:PROT:DATE <year>, <month>, <day> (Calibration date) :CAL:PROT:NDUE <year>, <month>, <day> (Next calibration due date)
Note that the year, month, and date must be separated by commas.
Step 5: Save calibration constants
Calibration is now complete, so you can store the calibration constants in EEROM by sending the following command:
:CAL:PROT:SAVE
NOTE Calibration will be temporary unless you send the SAVE command. Also, cali-
bration data will not be saved if (1) calibration is locked, (2) invalid data exists, or (3) all steps were not completed.
Step 6: Lock out calibration
To lock out further calibration, send the following command after completing the calibra­tion procedure:
Calibration
:CAL:PROT:LOCK
Single-range calibration
Normally, the complete calibration procedure should be performed to ensure that the entire instrument is properly calibrated. In some instances, however, you may want to cal ibrate only certain ranges. To do so, complete the entire procedure only for the range(s) to be calibrated. Keep in mind, however, that you must complete all parameter steps for each source or sense range. Also be sure to set calibration dates and save calibration after cali brating the desired range(s).
-
-
2-26 Calibration Model 2440 5A SourceMeter® Service Manual
3
Routine Maintenance
Routine Maintenance
3-2 Routine Maintenance Model 2440 5A SourceMeter® Service Manual
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING: NO INTERNAL OPERATOR SERVICABLE PARTS, SERVICE BY QUALIFIED PERSONNEL ONLY.
Introduction
The information in this section deals with routine type maintenance that can be performed by the operator.
Line fuse replacement
WARNING Disconnect the line cord at the rear panel, and remove all test leads
connected to the instrument (front and rear) before replacing the line fuse.
The power line fuse is accessible from the rear panel, just above the AC power receptacle (see
Figure 3-1).
Figure 3-1
Rear panel
42V
PEAK
4-WIRE SENSE
WITH FRONT PANEL MENU)
HI
5V
PEAK
LO
INPUT/
!
OUTPUT
IEEE-488
(ENTER IEEE ADDRESS
5V PK
42V
PEAK
40V
PEAK
5V
PEAK
V, Ω,
GUARD
GUARD
SENSE
CATI
RS232
Model 2440
TRIGGER
LINK
MADE IN
U.S.A.
LINE FUSE
SLOWBLOW
3.15A, 250V
LINE RATING
100-240VAC
50, 60HZ
220VA MAX
OUTPUT ENABLE
Model 2440 5A SourceMeter® Service Manual Routine Maintenance 3-3
Perform the following steps to replace the line fuse:
1. Carefully grasp and squeeze together the locking tabs that secure the fuse carrier to the fuse holder.
2. Pull out the fuse carrier, and replace the fuse with the type specified in Table 3-1.
CAUTION To prevent instrument damage, use only the fuse type specified in
Table 3-1.
3. Reinstall the fuse carrier.
NOTE If the power line fuse continues to blow, a circuit malfunction exists and must be
corrected. Refer to the troubleshooting section of this manual for additional information.
Table 3-1
Power line fuse
Line voltage Rating Keithley part no.
100-240V 250V, 3.15A, Slow
Blow 5 × 20mm
FU-106-3.15
Routine Maintenance
3-4 Routine Maintenance Model 2440 5A SourceMeter® Service Manual
4
Troubleshooting
Troubleshooting
4-2 Troubleshooting Model 2440 5A SourceMeter® Service Manual
Introduction
This section of the manual will assist you in troubleshooting and repairing the Model
2440. Included are self-tests, test procedures, troubleshooting tables, and circuit descrip tions. Note that disassembly instructions are located in Section 5, and component layout drawings are at the end of Section 6.
WARNING The information in this section is intended for qualified service per-
sonnel only. Do not perform these procedures unless you are qualified to do so. Some of these procedures may expose you to hazardous volt ages that could cause personal injury or death. Use caution when working with hazardous voltages.
-
Repair considerations
Before making any repairs to the Model 2440, be sure to read the following considerations.
CAUTION The PC-boards are built using surface mount techniques and require
specialized equipment and skills for repair. If you are not equipped and/or qualified, it is strongly recommended that you send the unit back to the factory for repairs or limit repairs to the PC-board replacement level. Without proper equipment and training, you could damage a PC-board beyond repair.
Repairs will require various degrees of disassembly. However, it is
recommended that the Front Panel Tests be performed prior to any disassembly. The disassembly instructions for the Model 2440 are contained in
Do not make repairs to surface mount PC-boards unless equipped
and qualified to do so (see previous CAUTION).
When working inside the unit and replacing parts, be sure to
adhere to the handling precautions and cleaning procedures explained in
Many CMOS devices are installed in the Model 2440. These static-
sensitive devices require special handling as explained in
Section 5.
Whenever a circuit board is removed or a component is replaced,
the Model 2440 must be recalibrated. See calibrating the unit.
Section 5 of this manual.
Section 5.
Section 2 for details on
-
Model 2440 5A SourceMeter® Service Manual Troubleshooting 4-3
Power-on self-test
During the power-on sequence, the Model 2440 will perform a checksum test on its EPROM and test its RAM. If one of these tests fails, the instrument will lock up.
Front panel tests
There are three front panel tests: one to test the functionality of the front panel keys and two to test the display. In the event of a test failure, refer to details on troubleshooting the display board.
KEYS test
The KEYS test lets you check the functionality of each front panel key. Perform the fol­lowing steps to run the KEYS test.
1. Display the MAIN MENU by pressing the MENU key.
2. Select TEST, and press ENTER to display the SELF-TEST MENU.
3. Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4. Select KEYS, and press ENTER to start the test. When a key is pressed, the label name for that key will be displayed to indicate that it is functioning properly. When the key is released, the message “No keys pressed” is displayed.
5. Pressing EXIT tests the EXIT key. However, the second consecutive press of EXIT aborts the test and returns the instrument to the SELF-TEST MENU. Continue pressing EXIT to back out of the menu structure.
“Display board checks” for
DISPLAY PATTERNS test
The display test lets you verify that each pixel and annunciator in the vacuum fluorescent display is working properly. Perform the following steps to run the display test:
1. Display the MAIN MENU by pressing the MENU key.
2. Select TEST, and press ENTER to display the SELF-TEST MENU.
3. Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4. Select DISPLAY-PATTERNS, and press ENTER to start the display test. There are five parts to the display test. Each time a front panel key (except EXIT) is pressed, the next part of the test sequence is selected. The five parts of the test sequence are as follows:
Checkerboard pattern (alternate pixels on) and all annunciators.
Checkerboard pattern and the annunciators that are on during normal operation.
Horizontal lines (pixels) of the first digit are sequenced.
Troubleshooting
4-4 Troubleshooting Model 2440 5A SourceMeter® Service Manual
Vertical lines (pixels) of the first digit are sequenced.
Each digit (and adjacent annunciator) is sequenced. All the pixels of the selected digit are on.
5. When finished, abort the display test by pressing EXIT. The instrument returns to the FRONT PANEL TESTS MENU. Continue pressing EXIT to back out of the menu structure.
CHAR SET test
The character set test lets you display all characters. Perform the following steps to run the character set test:
1. Display the MAIN MENU by pressing the MENU key.
2. Select TEST, and press ENTER to display the SELF-TEST MENU.
3. Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4. Select CHAR-SET, and press ENTER to start the character set test. Press any key except EXIT to cycle through all displayable characters.
5. When finished, abort the character set test by pressing EXIT. The instrument returns to the FRONT PANEL TESTS MENU. Continue pressing EXIT to back out of the menu structure.
Principles of operation
The following information is provided to support the troubleshooting tests and procedures covered in this section of the manual. Refer to the following drawings:
Figure 4-1Overall block diagram Figure 4-2Analog circuitry block diagram Figure 4-3Power supply block diagram Figure 4-4Output stage simplified schematic Figure 4-5Digital circuitry block diagram
Overall block diagram
Figure 4-1 shows an overall block diagram of the Model 2440. Circuitry may be divided
into three general areas:
Analog circuits — includes sourcing circuits such as the DACs, clamps, output stage, and feedback circuits, as well as measurement circuits such as the A/D converter.
Digital circuits — includes the microcomputer that controls the analog section, front panel, and GPIB and RS-232 ports, as well as associated interfacing circuits.
Model 2440 5A SourceMeter® Service Manual Troubleshooting 4-5
Power supplies — converts the AC line voltage into DC voltages that supply the power for the digital and analog circuits, and the output stage.
Analog circuits
Figure 4-2 shows the analog circuitry block.
Troubleshooting
4-6 Troubleshooting Model 2440 5A SourceMeter® Service Manual
Figure 4-1
Overall block diagram
Analog Section
Display,
Keyboard
Digital
I/O
Trigger
Link
Front
Panel
Controller
Trigger,
Digital
I/O
DACs
Clamps
Feedback
A/D
Converter
Microcomputer
Digital Section
Output
Stage
RS-232
GPIB
Interface
Guard Buffer
Output
Guard
RS-232 I/O
GPIB I/O
To Analog
Circuits
±15V
+5V ±21V
Analog
Power
Supply
To
Output Stage
±58V +5V +12V
Output
Stage
Power
Supply
Line In
Power Supply
To
Digital Circuits
Digital
Power
Supply
Model 2440 5A SourceMeter® Service Manual Troubleshooting 4-7
Figure 4-2
Analog circuitry block diagram
V DAC
I DAC
Control
Stage
+21
+58
+
-
Sense
Resistors
Protection
Protection
O
Output
HI
S+
Output
LO
S-
Guard
Out
Guard
Sense
FB
VFB
A/D
IFB
V Clamp
I Clamp
+7
MUX
-21
-58
Error Amp
VFB
IFB
S
Output
O
Remote
O
D/A converters control the programmed voltage and current, or voltage compliance and current compliance. Each DAC has two ranges, a 10V full-scale output or a 1V full-scale output. The DAC outputs are fed to the summing node, FB. Either the V DAC or the I
DAC has the ability to control the main loop. If the unit is set for SV (source voltage), it will source voltage until the compliance current is reached (as determined by the I DAC setting), and the current loop will override the voltage loop. If, however, the unit is set for SI (source current), it will source current until the compliance voltage is reached (as deter mined by the V DAC setting), and the voltage loop will override the current loop. A prior­ity bit in the V clamp/I clamp circuit controls these functions.
-
The error amplifier adds open-loop gain and slew-rate control to the system to assure accuracy and provide a controllable signal for the output stage, which provides the neces sary voltage and current gain to drive the output. Sense resistors in the HI output lead pro­vide output current sensing, and a separate sense resistor is used for each current range. The 1A and 5A ranges use 0.2V full-scale for a full-range output, while all other ranges use 2V output for full-scale current. Voltage feedback is routed either internally or exter nally.
There are four voltage ranges: 0.2V, 2V, 10V, and 40V. The feedback gain changes only for the 10V and 40V ranges, resulting in three unique feedback gain values. A multiplexer directs the voltage feedback, current feedback, reference, or ground signal to the A/D con verter. An opto-isolated interface provides control signals for both DACs, analog circuit control, and A/D converter communication to the digital section.
-
Troubleshooting
-
-
4-8 Troubleshooting Model 2440 5A SourceMeter® Service Manual
Power supply
Figure 4-3 shows a block diagram of the Model 2440 power delivery system.
Figure 4-3
Power supply block diagram
Line Neutral
Analog Board
+15V +5V
FF F
Constant Frequency
Low Noise Floating
Switching Supply
D
PFC
+24Vdc
DC/DC
Converter
F
-15V
+24dc
Regulators
D
Output Stage
-58V -21V +21V +58V
+12Vdc
DC
+5Vdc
O
Digital
Circuits
D
The offline switching power supply provides all power for the instrument while providing universal inputs and power factor correction for the 110/120V line. The digital board runs off of 5V and 12V supplies derived from the switcher. The +12VD supply is set to pro
-
gram the flash ROM. (See Digital circuitry below.)
A constant-frequency switching supply runs off the +24VD supplies and generates all the floating and output supply voltages for the analog board: +5V and ±15V, ±21V and ±58V.
Model 2440 5A SourceMeter® Service Manual Troubleshooting 4-9
Output stage
Figure 4-4 shows a simplified schematic of the output stage.
Figure 4-4
Output stage simplified schematic
+58V
HI Drive
Q500
+21V
Q504
Q514
Maindrive
O
+15V
300K
-15V
F
F
F
+
-
HI Drive
Q516
Q518
F
Q521
F
Q523
Q525
Troubleshooting
Q507
-21V
Q503
-58V
4-10 Troubleshooting Model 2440 5A SourceMeter® Service Manual
The Model 2440 output stage serves two purposes: (1) it converts signals from floating common to output common, and (2) it provides both voltage and current amplification. The output stage drive transistors are biased in class B configuration to prevent the possi bility of thermal runaway with high-current output values.
Output transistors Q518 and Q521 are cascoded with output MOSFETs Q516 and Q523. All other MOSFETs and transistors are slaves, and the voltages across these devices are determined by the resistor-capacitor ladder circuits shown. High-current drive capability is provided by Q500-Q511. Coarse current limits are built into the output stage.
A/D converter
The SourceMeter unit uses a multi-slope charge balance A/D converter with a single-slope run-down. The converter is controlled by gate array U610. Commands are issued by the MPU on the digital board through communications opto-isolators to U610, and U610 sends A/D reading data back through opto-isolators to the digital board for calibration and processing.
Active guard
The Model 2440 has an active guard or “six-wire ohms” circuit used to measure complex devices. This circuitry provides a low-current (50mA) equivalent of the voltage on output HI. If the unit is in the SV mode, the low-current equivalent of the source voltage will appear on the guard terminal. If the unit is in the SI mode, the voltage on output HI is equal to the source current multiplied by the external resistance value. An equivalent volt age will be generated by the guard circuit, and a guard sense terminal is provided to sense around the voltage drop in the guard leads since significant current can flow (50mA).
-
-
Model 2440 5A SourceMeter® Service Manual Troubleshooting 4-11
Digital circuitry
Refer to Figure 4-5 for the following discussion on digital circuitry.
The core digital circuitry uses a Motorola 68332 microcontroller running at 16.78MHz. The memory configuration includes two 256K X 8-bit flash EEPROMs and two 128K X 8-bit RAMs used in parallel to utilize the 16-bit data bus of the MPU. The RAM is battery backed-up, providing continued storage of data buffer information during power-down cycles, and flash ROM support allows internal firmware upgrades using either the serial or GPIB port for downloading new firmware. All calibration constants and the save 0 setup are stored in a separate serial EEPROM. Setups 1 through 4 are stored in battery backed­up RAM.
External communication is provided via GPIB and serial interfaces. A 9914 GPIB IEEE-488 standard interface IC is used for the GPIB, and a 68332 Queued Serial Module (QSM) provides the serial UART. For internal communications, the Time Processing Unit (TPU) is used for serial communications with the front panel display module, and both the TPU and QSM handle digital-to-analog interfacing.
Figure 4-5
Digital circuitry block diagram
A/D
Control/Data
Reset
2
E PROM
U17
A/D Interface
U9, U25
Voltage Source
Control
ROM
U15, U16
Microprocessor
U3
RAM
U12, U14
16.78MHz
Serial
Interface
U4
GPIB
U6, U13
U20
Trigger
Digital I/O
U23
U7
RS-232 Interface
IEEE-488 Interface
To Display
Board Controller
Trigger
Digital
I/O
Troubleshooting
4-12 Troubleshooting Model 2440 5A SourceMeter® Service Manual
Display board circuit theory
U902 is the display microcontroller that controls the VFD (vacuum fluorescent display) and interprets key data. The microcontroller has four peripheral I/O ports that are used for the various control and read functions.
Display data is serially transmitted to the microcontroller from the digital board via the TXB line to the microcontroller PD0 terminal. In a similar manner, key data is serially sent back to the digital board through the RXB line via PD1. The 4MHz clock for the microcontroller is generated on the digital board.
DS901 is the VFD (vacuum fluorescent display) module, which can display up to 49 char­acters. Each character is organized as a 5 × 7 matrix of dots or pixels and includes a long under-bar segment to act as a cursor.
The display uses a common multiplexing scheme with each character refreshed in sequence. U903 and U904 are the grid drivers, and U901 and U905 are the dot drivers. Note that dot driver and grid driver data is serially transmitted from the microcontroller (PD3 and PC1).
The front panel keys (S901-S931) are organized into a row-column matrix to minimize the number of microcontroller peripheral lines required to read the keyboard. A key is read by strobing the columns and reading all rows for each strobed column. Key down data is interpreted by the display microcontroller and sent back to the main microprocessor using proprietary encoding schemes.
Model 2440 5A SourceMeter® Service Manual Troubleshooting 4-13
Troubleshooting
Troubleshooting information for the various circuits is summarized below.
Display board checks
If the front panel display tests indicate that there is a problem on the display board, use
Table 4-1. See “Principles of operation” for display circuit theory.
Table 4-1
Display board checks
Step Item/component Required condition Remarks
1
Front panel test
2
J1033
3
U902, pin 1
4
U902, pin 43
5
U902, pin32
6
U902, pine 33
Verify that all segments operate. +5V, ±5% Goes low briefly on power up, and then goes high. 4MHz square wave. Pulse train every 1 ms. Brief pulse train when front panel key is pressed.
Use front panel display test.
Digital +5V supply. Microcontroller RESET.
Controller 4MHz clock. Control from main processor. Key down data sent to main processor.
Power supply checks
Power supply problems can be checked out using Table 4-2. See “Principles of operation” for circuit theory on the power supply.
Table 4-2
Power supply checks
Step Item/component Required condition Remarks
1
Line fuse
2
Line power
3
TP502
4
TP503
5
TP504
6
TP505
7
TP507
8
TP508
9
TP510
Check continuity. Plugged into live receptacle, power on. +58V, ±10%
-58V, ±10% +21V, ±10%
-21V, ±10% +15V, ±5%
-15V, ±5% +5V, ±5%
Troubleshooting
Remove to check. Check for correct power-up sequence. Referenced to TP501. Referenced to TP501. Referenced to TP501. Referenced to TP501. +15VF, referenced to TP500.
-15VF, referenced to TP500. +5VF, referenced to TP500.
4-14 Troubleshooting Model 2440 5A SourceMeter® Service Manual
Digital circuitry checks
Digital circuit problems can be checked out using Table 4-3. See “Principles of opera-
tion” for a digital circuit description.
Table 4-3
Digital circuitry checks
Step Item/component Required condition Remarks
1
Power-on test
2
U3 pin 19
3
U3 pin 7
4
U3 pin 68
5
U3, lines A0-A19
6
U3, lines D0-D15
7
U3 pin 66
8
U4 pin 7
9
U4 pin 8
10
U13 pins 34-42
11
U13 pins 26-31
12
U13 pin 24
13
U13 pin 25
14
U3 pin 43
15
U3 pin 44
16
U3 pin 45
17
U3 pin 47
RAM OK, ROM OK.
Digital common.
+5V Low on power-up, then goes high. Check for stuck bits. Check for stuck bits.
16.78MHz. Pulse train during RS-232 I/O. Pulse train during RS-232 I/O. Pulse train during IEEE-488 I/O. Pulses during IEEE-488 I/O. Low with remote enabled. Low during interface clear. Pulse train. Pulse train. Pulse train. Pulse train
Verify that RAM and ROM are functional. All signals referenced to digital common. Digital logic supply. MPU RESET line.
MPU address bus. MPU data bus. MPU clock. RS-232 RX line.
RS-232 TX line.
IEEE-488 data bus.
IEEE-488 command lines. IEEE-488 REN line. IEEE-488 IFC line. D_ADDATA D_DATA D_CLK D_STB
Model 2440 5A SourceMeter® Service Manual Troubleshooting 4-15
Analog circuitry checks
Table 4-4 summarizes analog circuitry checks.
Table 4-4
Analog circuitry checks
Step Item/component Required condition Remarks
1
TP200*
2
TP201*
3
TP202*
4
TP203*
5
TP213*
6
TP218**
7
TP219**
8
TP214**
9
TP232*
**Measured with respect to FCOM (TP500).
**Measured with respect to OCOM (TP501).
>60V voltage protection SOURCE +10V SOURCE +10V (SVMI) SOURCE +10V SOURCE +10V OUTPUT COM OUTPUT COM SVMI, OUTPUT ON, 20V Bench defaults
-13V ±1V
-5V ±0.5V
-10V ±1V
-10.5V ±1V 0V ±0.1V 7V ±0.7V 7V ±0.7V 20V ±0.5V
6.4V ±0.6V
Battery replacement
WARNING Disconnect the instrument from the power line and all other equip-
ment before changing the battery.
The volatile memories of the Model 2440 are protected by a replaceable battery when power is off. Typical battery life is approximately 10 years, but the battery should be replaced if the voltage drops below 2.5V regardless of age. The battery should be sus pected if the instrument no longer retains buffer data or user-defined operating parameters such as instrument setups, source memory, and math expressions. If the battery is absent or totally exhausted, the display will show the “Reading buffer data lost” message shortly after the Model 2440 is switched on.
The battery is a 3V wafer-type lithium cell, (Keithley part number BA-46), which is located on the digital board. Replacement of the battery requires removal of the case cover and analog board assembly. (See
NOTE Calibration constants and user defined parameters will be lost when the battery
is replaced. The Model 2440 must be re-calibrated (Section 2) after the battery is replaced.
-
Troubleshooting
Section 5.)
4-16 Troubleshooting Model 2440 5A SourceMeter® Service Manual
Battery replacement precautions
WARNING The precautions below must be followed to avoid personal injury.
1. Wear safety glasses or goggles when working with lithium batteries.
2. Do not short the battery terminals together.
3. Keep lithium batteries away from all liquids.
4. Do not attempt to recharge lithium batteries.
5. Observe proper polarity when installing the battery.
6. Do not incinerate or otherwise expose the battery to excessive heat (>60°C).
7. Bulk quantities of lithium batteries should be disposed of as hazardous waste.
Battery replacement procedure
1. Remove the case cover and analog board assembly as covered in Section 5.
2. Locate the battery on the digital board.
3. Carefully unsolder and remove the old battery.
4. Install and solder the new battery in place.
5. Reassemble the instrument, and turn it on. The “Reading buffer data lost” error message will be displayed.
6. Send the :SYST:MEM:INIT command via remote to perform the following:
Clear the reading buffer.
Initialize instrument setups 1 through 4 to present instrument settings.
Initialize all 100 source memory locations to present instrument settings.
Delete user math expressions.
Model 2440 5A SourceMeter® Service Manual Troubleshooting 4-17
No comm link error
A “No Comm Link” error indicates that the front panel processor has stopped communi­cating with the main processor, which is located on the digital board. This error indicates that one of the main processor ROMs may require re-seating in its socket. ROMs may be reseated as follows:
1. Turn off the power, and disconnect the line cord and all other test leads and cables from the instrument.
2. Remove the case cover as outlined in Section 5.
3. Remove the analog board assembly as outlined in Section 5.
4. Locate the two firmware ROMS, U15 and U16, located on the digital board. These are the only ICs installed in sockets. (Refer to the component layout drawing at the end of
Section 6 for exact locations.)
5. Carefully push down on each ROM IC to make sure it is properly seated in its socket.
CAUTION Be careful not to push down excessively, or you might crack the digital
board.
6. Connect the line cord, and turn on the power. If the problem persists, additional troubleshooting will be required.
Troubleshooting
4-18 Troubleshooting Model 2440 5A SourceMeter® Service Manual
5
Disassembly
Disassembly
5-2 Disassembly Model 2440 5A SourceMeter® Service Manual
Introduction
This section explains how to handle, clean, and disassemble the Model 2440. Disassembly drawings are located at the end of this section.
Handling and cleaning
To avoid contaminating PC board traces with body oil or other foreign matter, avoid touching the PC board traces while you are repairing the instrument. Motherboard areas covered by the shield have high-impedance devices or sensitive circuitry where contami nation could cause degraded performance.
Handling PC boards
Observe the following precautions when handling PC boards:
Wear cotton gloves.
Only handle PC boards by the edges and shields.
Do not touch any board traces or components not associated with repair.
Do not touch areas adjacent to electrical contacts.
Use dry nitrogen gas to clean dust off PC boards.
-
Solder repairs
Observe the following precautions when you must solder a circuit board:
Use an OA-based (organic activated) flux, and take care not to spread the flux to other areas of the circuit board.
Remove the flux from the work area when you have finished the repair by using pure water with clean, foam-tipped swabs or a clean, soft brush.
Once you have removed the flux, swab only the repair area with methanol, then blow-dry the board with dry nitrogen gas.
After cleaning, allow the board to dry in a 50°C, low-humidity environment for several hours.
Model 2440 5A SourceMeter® Service Manual Disassembly 5-3
Static sensitive devices
CMOS devices operate at very high impedance levels. Therefore, any static that builds up on you or your clothing may be sufficient to destroy these devices if they are not handled properly. Use the following precautions to avoid damaging them:
CAUTION Many CMOS devices are installed in the Model 2440. Handle all semi-
conductor devices as being static sensitive.
Transport and handle ICs only in containers specially designed to prevent static build-up. Typically, you will receive these parts in anti-static containers made of plastic or foam. Keep these devices in their original containers until ready for installation.
Remove the devices from their protective containers only at a properly grounded work station. Also, ground yourself with a suitable wrist strap.
Handle the devices only by the body; do not touch the pins.
Ground any printed circuit board into which a semiconductor
device is to be inserted to the bench or table.
Use only anti-static type desoldering tools.
Use only grounded-tip solder irons.
Once the device is installed in the PC board, it is normally ade-
quately protected, and you can handle the boards normally.
Disassembly
Assembly drawings
Use the assembly drawings located at the end of this section to assist you as you disassem­ble and re-assemble the Model 2440. Also, refer to these drawings for information about the Keithley part numbers of most mechanical parts in the unit. Assembly drawings include:
Front panel assembly — 2440-040
Analog board/heat sink/shield assembly — 2440-050
Chassis assembly — 2440-051
Chassis/analog board assembly — 2440-052
Final chassis assembly — 2440-053
5-4 Disassembly Model 2440 5A SourceMeter® Service Manual
Case cover removal
Follow the steps below to remove the case cover to gain access to internal parts.
WARNING Before removing the case cover, disconnect the line cord and any test
leads from the instrument.
1. Remove handle — The handle serves as an adjustable tilt-bail. Adjust its position
by gently pulling it away from the sides of the instrument case and swinging it up or down. To remove the handle, swing the handle below the bottom surface of the case and back until the orientation arrows on the handles line up with the orienta tion arrows on the mounting ears. With the arrows lined up, pull the ends of the handle away from the case.
2. Remove mounting ears — Remove the screw that secures each mounting ear. Pull
down and out on each mounting ear.
NOTE When reinstalling the mounting ears, make sure to mount the right ear to the
right side of the chassis, and the left ear to the left side of the chassis. Each ear is marked “RIGHT” or “LEFT” on its inside surface.
-
3. Remove rear bezel — To remove the rear bezel, loosen the two screws that secure
the rear bezel to the chassis, then pull the bezel away from the case.
4. Remove grounding screws — Remove the two grounding screws that secure the
case to the chassis. They are located on the bottom of the case at the back.
5. Remove chassis — To remove the case, grasp the front bezel of the instrument,
and carefully slide the chassis forward. Slide the chassis out of the metal case.
NOTE To gain access to the components under the analog board shield, remove the
shield, which is secured to the analog board by a single screw.
Model 2440 5A SourceMeter® Service Manual Disassembly 5-5
Analog board removal
Perform the following steps to remove the analog board. This procedure assumes that the case cover is already removed.
1. Disconnect the front and rear input terminals. You must disconnect these input terminal connections for both the front and rear inputs:
INPUT/OUTPUT HI and LO
4-WIRE SENSE HI and LO
•V, Ω, GUARD and GUARD SENSE (rear panel only) Remove all the connections by pulling the wires off the pin connectors, then remove the ferrite noise filters from the chassis. During re-assembly, use the fol lowing table to identify input terminals:
Input terminals Front wire color Rear wire color
INPUT/OUTPUT HI INPUT/OUTPUT LO 4-WIRE SENSE HI 4-WIRE SENSE LO V, Ω, GUARD GUARD SENSE
2. Unplug cables.
Carefully unplug the ribbon cables at J1027, J1028, and J1029.
Unplug the ON/OFF cable at J1034.
3. Remove screws.
Remove the two fastening screws that secure the analog board assembly to the
chassis. These screws are located on the side of the board opposite from the heat sink.
Remove the two screws that secure the heat sink to the chassis.
4. Remove analog board assembly. After all screws have been removed, carefully lift the analog board assembly free of the main chassis.
5. Disassemble analog board assembly.
Remove the screws that secure the analog board and heat sink to the analog board subchassis.
Carefully remove the heat sink by sliding the clips off the power transistors.
Red
Black
Yellow
Gray
-
-
White/Red White/Black White/Yellow White/Gray White Blue/White
-
Disassembly
CAUTION Be careful not to damage the heat sink insulation layer.
Remove the analog board from the subchassis.
Remove the four screws that secure the bottom cover, then remove the cover from the bottom of the PC board.
NOTE When reinstalling the heat sink, make sure that all clips are properly installed
and centered on each pair of output transistors.
5-6 Disassembly Model 2440 5A SourceMeter® Service Manual
Digital board removal
Perform the following steps to remove the digital board. This procedure assumes that the analog board assembly is already removed.
NOTE In order to remove the digital board, the display board must first be removed.
1. Remove the IEEE-488, Digital I/O, and RS-232 fasteners. The IEEE-488, Digital I/O, and RS-232 connectors each have two nuts that secure the connectors to the rear panel. Remove these nuts.
2. Remove the POWER switch rod. At the switch, place the edge of a flat-blade screw driver in the notch on the push­rod. Gently twist the screw driver while pulling the rod from the shaft.
3. Unplug cables:
Unplug the display board ribbon cable.
Unplug the cables going to the power supply.
Unplug the rear panel power module cable.
The fan may need to be removed.
4. Remove digital board. Slide the digital board forward until it is free of the guide pins, then remove the board. During reassembly, replace the board, and start the IEEE-488, Digital I/O, and RS­232 connector nuts and the mounting screw. Tighten all the fasteners once they are all in place and the board is correctly aligned.
Front panel disassembly
Use the following procedures to remove the display board and/or the pushbutton switch pad.
1. Unplug the display board ribbon cables.
2. Remove the front panel assembly. This assembly has four retaining clips that snap onto the chassis over four pem nut studs. Two retaining clips are located on each side of the front panel. Pull the retaining clips outward and, at the same time, pull the front panel assembly for ward until it separates from the chassis.
3. Using a thin-bladed screw driver, pry the plastic PC board stop (located at the bot­tom of the display board) until the bar separates from the casing. Pull the display board from the front panel.
4. Remove the switch pad by pulling it from the front panel.
-
Model 2440 5A SourceMeter® Service Manual Disassembly 5-7
Removing power components
The following procedures to remove the power supply and/or power module require that the case cover and motherboard be removed, as previously explained.
Power module removal
Perform the following steps to remove the rear panel power module:
1. Remove the analog board.
2. Unplug the cable connecting the power module to the digital board.
3. Disconnect the power module's ground wire. This green and yellow wire connects to a threaded stud on the chassis with a kep nut.
4. Squeeze the latches on either side of the power module while pushing the module from the access hole.
WARNING To avoid electrical shock, which could result in injury or death, the
ground wire of the power module must be connected to chassis ground. When installing the power module, be sure to reconnect the green and yellow ground wire to the threaded stud on the chassis.
Instrument reassembly
Reassemble the instrument by reversing the previous disassembly procedures. Make sure that all parts are properly seated and secured, and that all connections are properly made. To ensure proper operation, replace the analog signal wire ferrite noise filters, and securely fasten the shield.
Disassembly
WARNING To ensure continued protection against electrical shock, verify that
power line ground (green and yellow wire attached to the power mod ule) is connected to the chassis.
Also make sure the two bottom case screws are properly installed to secure and ground the case cover to the chassis.
-
5-8 Disassembly Model 2440 5A SourceMeter® Service Manual
6
Replaceable Parts
Replaceable Parts
6-2 Replaceable Parts Model 2440 5A SourceMeter® Service Manual
Introduction
This section contains replacement parts information and component layout drawings for the Model 2440.
Parts lists
The electrical parts lists for the Model 2440 are shown tables at the end of this section. For part numbers to the various mechanical parts and assemblies, use the Miscellaneous parts list and the assembly drawings provided at the end of
Section 5.
Ordering information
To place an order, or to obtain information concerning replacement parts, contact your Keithley representative or the factory (see inside front cover for addresses). When order ing parts, be sure to include the following information:
Instrument model number (Model 2440)
Instrument serial number
Part description
Component designation (if applicable)
Keithley part number
-
Factory service
If the instrument is to be returned to Keithley Instruments for repair, perform the following:
Call the Repair Department at 1-800-552-1115 for a Return Material Authorization (RMA) number.
Complete the service form at the back of this manual, and include it with the instru­ment.
Carefully pack the instrument in the original packing carton.
Write ATTENTION REPAIR DEPARTMENT and the RMA number on the ship­ping label.
Model 2440 5A SourceMeter® Service Manual Replaceable Parts 6-3
Component layouts
The component layouts for the various circuit boards are provided on the following pages. Drawings include:
Analog board — 2440-120
Display board — 2400-110
Digital board — 2440-140
Table 6-1
Analog board parts list
Keithley
Circuit Designation Description
C200-C203, C205-C210, C225,
C226, C231, C232 C204,C411, C608, C609 CAP, .01UF, 10%, 50V, CERAMIC C-491-.01 C211, C212, C213, C214, C301,
C302 C215-C222, C611, C331, C332,
C333, C334 C223, C224, C227, C229, C234-
C236, C239, C240 C230, C233 CAP, 7.5P, 10%, 100V, CERAMIC C-452-7.5P C237, C238, C242, C243, C248,
C249, C254, C255 C241 CAP, 1000PF, 1%, 50V, CERAMIC C-347-1000P C246, C336 CAP, 3300PF, 10%, 100V, CERAMIC C-430-3300P C247 CAP, .01, 5%, 50V, NPO C-514-.01 C250, C251, C272, C297, C298,
C299, C300, C524 C258, C259, C260-C263, C269-
C271, C273-C282 C267, C268 CAP, 1000P, 10%, 100V, CERAMIC C-451-1000P C283 CAP, 100P, 10%, 100V, CERAMIC C-451-100P C284-C291, C305-C310, C321,
C322, C548, C558 C296, C330, C400 CAP, 270PF, 2.5%, 630V, POLYPROPYLENE C-405-270P C311 CAP, 100UF, 20%, 16V, TANTALUM C-504-100 C312-C319 CAP, 10PF, 5%, 100V, CERAMIC C-372-10P C320 CAP, 33PF, 5%, 100V, CERAMIC C-465-33P C323, C324, C325, C326, C603,
C604, C615 C327 CAP, 390P, 10%, 100V, CERAMIC C-451-390P
CAP, .1UF, 10%, 25V, CERAMIC C-495-.1
CAP, 1000PF, 10%, 50V, MONO CERAMIC C-452-1000P
CAP, 100PF, 5%, 100V, CERAMIC C-465-100P
CAP, 33PF, 5%, 100V, CERAMIC C-465-33P
CAP, .1UF, 10%, 25V, CERAMIC C-495-.1
CAP, .1UF, 10%, 25V, CERAMIC C-495-.1
CAP, .1UF, 10%, 25V, CERAMIC C-495-.1
CAP, .1UF, 10%, 25V, CERAMIC C-495-.1
CAP, 47P, 5%, 100V, CERAMIC C-465-47P
Part
No.
Replaceable Parts
6-4 Replaceable Parts Model 2440 5A SourceMeter® Service Manual
Table 6-1 (continued)
Analog board parts list
Keithley
Circuit Designation Description
Part No.
C328 CAP, 22P, 2%, 200V, CERAMIC C-531-22P C329 CAP, 220PF, 10%, 100V, CERAMIC C-451-220P C409, C410 CAP, .15UF, 20%, 50V, CERAMIC C-418-.15 C412 CAP, 1UF, 20%, 35V, TANTALUM C-494-1 C501, C503 CAP, 3.3, 20%, 250V, ALUM ELECTROLYTIC C-400-3.3 C502, C504, C505, C506, C550,
CAP, .01U, 10%, 500V, CERAMIC C-497-.01
C551, C552, C553 C512, C514, C256, C257 CAP, 22PF, 10%, 100V, CERAMIC C-451-22P C513 CAP, 560PF, 2.5%, 630V, POLYPROPYLENE C-405-560P C515, C541 CAP, .1UF, 20%, 50V, CERAMIC C-418-.1 C522, C523, C526, C527 CAP, 3300P, 10%, 500V, CERAMIC C-497-3300P C525, C542, C549, C561, C564,
CAP, .1UF, 10%, 25V, CERAMIC C-495-.1
C585 C528, C533, C567, C568, C569,
CAP, 3300P, 10%, 500V, CERAMIC C-497-3300P
C570, C663, C664 C534, C535 CAP, 47U, 20%, 50V, ALUM C-579-47 C543 CAP, 10PF, 10%, 1000V, CERAMIC C-64-10P C544, C545, C546, C547 CAP, .01U, 10%, 500V, CERAMIC C-497-.01 C556, C557, C559, C560, C562,
CAP, 22UF, 20%, 25V, TANTALUM C-440-22
C563, C616 C565, C566, C581, C582, C600,
CAP, .1UF, 10%, 25V, CERAMIC C-495-.1
C601, C602, C605 C583, C584 CAP, 1000PF, 20%, 50V, CERAMIC C-418-1000P C606, C607, C613, C614, C617,
CAP, .1UF, 10%, 25V, CERAMIC C-495-.1
C618, C619, C620 C610, C612 CAP, 2200P, 10%, 100V, CERAMIC C-430-2200P C621, C622, C623, C624, C625,
CAP, .1UF, 10%, 25V, CERAMIC C-495-.1
C650, C651, C652 C659-C662 CAP, .1UF, 10%, 25V, CERAMIC C-495-.1 CR200-CR205, CR207, CR208,
DIODE, SWITCHING, MMBD914 RF-83
CR210-CR213 CR206, CR209, CR222, CR223,
DIODE, 1N3595 RF-43
CR230-CR233 CR216-CR221, CR235, CR236,
DIODE, SWITCHING, MMBD914 RF-83
CR239, CR518 CR224, CR225 SCHOTTKY BARRIER RECTIFIER RF-154 CR237, CR523 DIODE, DUAL SWITCHING, BAV99L RF-82 CR400, CR401 DIODE, 1N3595 RF-43 CR517, 519 SWITH MODE POWER RECTIFIER RF-155
Model 2440 5A SourceMeter® Service Manual Replaceable Parts 6-5
Table 6-1 (continued)
Analog board parts list
Keithley
Circuit Designation Description
CR522, CR524, CR525 DIODE, SWITCHING, MMBD914 RF-83 CR600 DIODE, DUAL HSM-2822T31 RF-95 CR601 DIODE, SCHOTTKY, SD103C RF-113 CR602 SCHOTTKY DIODE RF-125 HS1, HS2, HS3, HS4, HS5, HS6,
HS7, HS17, HS19 J1027, J1036 CONN, HEADER STRAIGHT SOLDER PIN CS-368-16 J1028 CONNECTOR, HEADER STRAIGHT SOLDER
J1029 CONN, MALE, 5-PIN (MOLEX 42491) CS-784-5 J1034, J1037, J1038 LATCHING HEADER, FRICTON, SGL ROW CS-724-3 K200-K205 N.C RELAY, 1 FORMB, AQV214S RL-176 K206, K207, K208 FORM 2A2B POLARIZED RELAY, S28B-5V RL-207 K211, K500 RELAY, SURFACE MOUNT RL-188 K215 NON LATCHING RELAY RL-242 L201, L500, L501 CHOKE, SHIELD BEAD CH-52 L600, L601, L602 FERRITE CHIP, 600 OHM, BLM32A07 CH-62 L603 FERRITE CHIP, 600 OHM, BLM32A07 CH-62 Q200, Q201, Q202, Q203, Q204,
Q205, Q206, Q207 Q208, Q210, Q212, Q214, Q216,
Q218 Q209, Q211, Q213, Q215, Q217,
Q219 Q220 TRANS, NPN SILICON, MJE340 TG-209 Q221 TRANS, PNP POWER, MJE350 TG-210 Q222, Q224 TRANS, N-CHAN JFET, SST4393 TG-263 Q223, Q229, Q234-Q239, Q244,
Q245, Q246, Q255 Q225 TRANS, NPN SILICON, MJD340 TG-373 Q226 TRANS, PNP SILICON, MJD350 TG-374 Q227, Q257, Q501, Q510, Q515,
Q517, Q519, Q522 Q228, Q247, Q248, Q520, Q601 TRANS, PNP, MMBT3906L TG-244 Q240, Q241 TRANS, N CHANNEL JFET, SNJ132199 TG-294 Q242, Q243 TRANS, N-CHAN MOSFET, TN254ON8 TG-274 Q249, Q528 TRANS, NPN SILICON, BC846BL, TG-278 Q250, Q529 TRANS, PNP, BC856BL TG-287
HEAT SINK HS-55
PIN
TRANS, N CHANNEL JFET, SNJ132199 TG-294
TRANS, NPN CATV SILICON, MPSH17 TG-221
TRANS, CURR REG, CDLL5313 TG-321
TRANS, N-MOSFET, VN0605T TG-243
TRANS, NPN, MMBT3904 TG-238
Part No.
CS-368-14
Replaceable Parts
6-6 Replaceable Parts Model 2440 5A SourceMeter® Service Manual
Table 6-1 (continued)
Analog board parts list
Keithley
Circuit Designation Description
Part No.
Q256, Q258, Q401, Q404, Q405,
TRANS, N-MOSFET, VN0605T TG-243
Q410, Q411, Q420 Q402, Q409, Q259 TRANS, N-CHAN DMOS, FET TN2504N8 TG-261 Q403, Q406, Q407, Q408 TRANS, N-CHAN JFET, SST109 TG-266 Q412, Q414 TRANS, N-MEGAFET, RFD14N05LSM TG-267 Q415, Q416 TRANS, HEXFET POWER MOSFET, IRFR5305 TG-377 Q417, Q419 TRANS, N-CHANNEL MOSFET,
TG-339
RFD16N05LSM Q418, Q542, Q230 TRANS, P-CHAN, MOSFET, TP0610T TG-259 Q500, Q503 TRANS, POWER MOSFET, IRF3415S TG-372 Q502, Q540, Q541, Q602-Q605 TRANS, N-MOSFET, VN0605T TG-243 Q504, Q514, Q516, Q518 N-CHANNEL POWER MOSFET TG-349 Q507, Q521, Q523, Q525 P-CHANNEL POWER MOSFET TG-348 Q524, Q534, Q535, Q537, Q538,
TRANS, NPN, MMBT3904 TG-238
Q600 Q526, Q527, Q530, Q531 TRANS, N-CHANNEL, MOSFET, SI9410DY TG-302 Q532, Q536 TRANS, N-CHANNEL, FET, IRF630 TG-214 Q533, Q539 TRANS, P-CHANNEL, 200V FET, IRF9630 TG-215 R117-R128, R133-R136, R141-
RES, 10K, 5%, 250MW, METAL FILM R-376-10K
R144 R129, R155, R211, R224-R229,
RES, 10K, 1%, 100MW, THICK FILM R-418-10K
R234-R237 R130 RES, 10K, 1%, 100MW, THICK FILM R-418-10K R137-140, R145-149, R713, R714 RES, 10K, 5%, 250MW, METAL FILM R-376-10K R150, R151, R389, R390, R451,
RES, 1K, 1%, 100MW, THICK FILM R-418-1K
R520, R531, R559 R152, R656 RES, 4.99K, .1%, .125W, THIN FILM R-456-4.99K R153, R154, R240, R243, R248,
RES, 4.99K, 1%, 100MW, THICK FILM R-418-4.99K
R251, R253, R254 R200, R201, R203, R232, R238 RESISTOR NETWORK, 10K-10K TF-236 R202, R204, R252, R255, R256,
RES, 100K, 1%, 100MW, THICK FILM R-418-100K
R259, R260, R263 R205, R310, R485, R552, R595,
RES, .0499, 1%, 100MW, THICK FILM R-418-.0499
R598, R723, R724 R207 RES, 20K, .1%, .125W, THIN FILM R-456-20K R208, R210 RES, 768, 1%, 100MW, THICK FILM R-418-768 R209, R268, R269, R270, R271,
RES, 1M, 1%, 100MW, THICK FILM R-418-1M
R672 R213 RES NET, 5K, .1%, 668A TF-243-5K
Model 2440 5A SourceMeter® Service Manual Replaceable Parts 6-7
Table 6-1 (continued)
Analog board parts list
Keithley
Circuit Designation Description
Part No.
R216, R221, R223, R274, R275,
R276, R277, R278 R217 RES, 121K, 1%,100MW, THICK FILM R-418-121K R218 RES, 80.6K, 1%, 100MW, THICK FILM R-418-80.6K R219, R220, R222, R230, R231,
R373, R374, R616 R241, R242, R249, R250, R653,
R664 R244-R247, R281, R282, R284-
R286, R288, R301 R257, R258, R261, R262, R265,
R266, R311, R312 R264, R267, R364, R370, R371,
R372, R402 R272, R273, R347, R348, R511,
R512, R589 R279 RES, 20K, 1%, 100MW, THICK FILM R-418-20K R280, R516, R523 RES, 4.02K, 1%, 100MW, THICK FILM R-418-4.02K R283, R287, R297, R299, R349,
R350, R351, R352 R289, R303, R681, R684 RES, 20K, .1%, .125W, THIN FILM R-456-20K R290, R309, R336, R340, R343,
R344, R353, R355 R291, R293, R682, R685 RES, 42.7K, .1%, .125W, THIN FILM R-456-42.7K R292, R338, R527, R528, R533,
R535, R544, R546 R294, R295, R296 RES, 100K, .1%, .125W, THIN FILM R-456-100K R298, R300 RES, 150K, 1%, 100MW, THICK FILM R-418-150K R302 RES, .0499, 1%, 100MW, THICK FILM R-418-.0499 R304, R305, R306, R548 RES, 33, 5%, 250MW, METAL FILM R-376-33 R307, R413, R652, R665 RES, 357, 1%, 100MW, THICK FILM R-418-357 R308, R379-R387, R391, R392 RES, 475, 1%, 100MW, THICK FILM R-418-475 R313, R314, R315, R316, R317,
R334, R335, R337 R318, R563, R358 RES, 15K, 1%, 100MW, THICK FILM R-418-15K R332, R423, R430, R604, R606,
R436 R339, R357, R366, R367, R424,
R425, R426, R429
RES, 20K, 1%, 100MW, THICK FILM R-418-20K
RES, 49.9K, 1%, 100MW, THICK FILM R-418-49.9K
RES, 140K, 1%, 100MW, THICK FILM R-418-140K
RES, 10K, 1%, 100MW, THICK FILM R-418-10K
RES, 4.99K, 1%, 100MW, THICK FILM R-418-4.99K
RES, 100K, 1%, 100MW, THICK FILM R-418-100K
RES, 249K, 1%, 100MW, THICK FILM R-418-249K
RES, 30.1K, 1%, 100MW, THICK FILM R-418-30.1K
RES, 1K, 1%, 100MW, THICK FILM R-418-1K
RES, 100K, 1%, 100MW, THICK FILM R-418-100K
RES, 4.99K, 1%, 100MW, THICK FILM R-418-4.99K
RES, 499, 1%, 100MW, THICK FILM R-418-499
RES, 4.99K, 1%, 100MW, THICK FILM R-418-4.99K
Replaceable Parts
6-8 Replaceable Parts Model 2440 5A SourceMeter® Service Manual
Table 6-1 (continued)
Analog board parts list
Keithley
Circuit Designation Description
Part No.
R341, R342, R505 RES, 13K, 1%, 100MW, THICK FILM R-418-13K R345 RES, 316K, 1%, 100MW, THICK FILM R-418-316K R346 RES, 392K, 1%, 100MW, THICK FILM R-418-392K R354, R437 RES, 560K, 5%, 250MW, METAL FILM R-376-560K R356, R359, R361, R362, R365,
RES, 1K, 1%, 100MW, THICK FILM R-418-1K
R369, R418, R419 R363 RES NET, VARIOUS, .1%, .1W TF-283 R368, R212 RES, 332, 1%, 100MW, THICK FILM R-418-332 R375, R376, R576, R578, R536,
RES, 10, 10%, 100MW, THICK FILM R-418-10
R537 R388, R393-R399, R411, R412,
RES, 357, 1%, 100MW, THICK FILM R-418-357
R699 R403, R404, R416, R417, R459,
RES, 100K, 1%, 100MW, THICK FILM R-418-100K
R488, R560 R420, R421, R456, R458, R460,
RES, 1K, 1%, 100MW, THICK FILM R-418-1K
R462, R466, R472 R422 RES, 4.7K, 5%, 250MW, METAL FILM R-376-4.7K R431, 432 RES, 3.01K, 1%, 100MW, THICK FILM R-418-3.01K R433, R434 RES, 1K, 1%, 100MW, THICK FILM R-418-1K R435, R504, R590, R673, R688,
RES, 10K, 1%, 100MW, THICK FILM R-418-10K
R689, R716, R715, R438, R439, R592, R617, R622,
RES, 10K, 1%, 100MW, THICK FILM R-418-10K
R654, R655, R687 R450 RES, 200K, .1%, 1/10W, METAL FILM R-263-200K R452 RES, 21.93K, .1%, 1/10W, METAL FILM R-263-21.93K R453 RES, 1.82K, .1%, 1/10W, METAL FILM R-263-1.82K R454, R455, R467, R468, R469,
RES, 100, .1%, 1/10W, METAL FILM R-263-100
R470, R471 R457, R461, R463, R473, R479,
RES, 30.1K, 1%, 100MW, THICK FILM R-418-30.1K
R697, R698 R475 RES, 1, 1%, 1/5W, WIRE R-501-1 R478, R481, R489, R502, R510,
RES, 1K, 1%, 100MW, THICK FILM R-418-1K
R558, R675, R704 R482, R484, R487 RES, 6.04K, 1%, 100MW, THICK FILM R-418-6.04K R483, R691, R695, R721 RES, 2K, 1%, 100MW, THICK FILM R-418-2K R490 RES, 205, .1%, 1/10W, METAL FILM R-263-205 R503, R545, R587 RES, 15K, 1%, 100MW, THICK FILM R-418-15K R508, R509, R506, R507 RES, 150K, 5%, 250MW, METAL FILM R-376-150K
Model 2440 5A SourceMeter® Service Manual Replaceable Parts 6-9
Table 6-1 (continued)
Analog board parts list
Keithley
Circuit Designation Description
Part No.
R517, R524, R564, R565, R566,
R567 R521, R658 RES, 2M, 1%, 100MW, THICK FILM R-418-2M R525, R526 RES, 1.21K, 1%, 100MW, THICK FILM R-418-1.21K R530, R532, R534, R712 RES, 1K, 1%, 100MW, THICK FILM R-418-1K R538, R540, R568, R569 RES, .5, 5%, 1W, METAL FILM R-444-.5 R539, R541, R542, R543 RES, .25, 1%, IN, METAL R-444-.25 R547, R588 RES, 100K, 1%, 100MW, THICK FILM R-418-100K R553, R554 RES, 75K, 1%, 100MW, THICK FILM R-418-75K R561, R562, R600, R500, R515,
R722 R570, R571, R573, R581, R705 RES, 249, 1%, 100MW, THICK FILM R-418-249 R572, R574, R580, R582 RES, 200K, 1%, 100MW, THICK FILM R-418-200K R575, R579, R614, R615 RES, 2.21K, 1%, 100MW, THICK FILM R-418-2.21K R586 RES, 10, 5%, 125MW, METAL FILM R-375-10 R601, R603, R605, R608, R626,
R702, R703 R602, R607 RES, 150, 5%, 250MW, METAL FILM R-376-150 R609 RES NET TF-245 R610 RES, 1.28M, .1%, 1/8W, METAL FILM R-176-1.28M R611, R621 RES, 475, 1%, 100MW, THICK FILM R-418-475 R612 RES, 5.11K, 1%, 100MW, THICK FILM R-418-5.11K R613, R624, R659, R660 RES, 100, 1%, 100MW, THICK FILM R-418-100 R618 RES, 34K, 1%, 100MW, THICK FILM R-418-34K R619 RES, 4.75K, 1%, 100MW, THICK FILM R-418-4.75K R620 RES, 82.5, 1%, 100MW, THICK FILM R-418-82.5 R623 RES, 10, 10%, 100MW, THICK FILM R-418-10 R650 RES NET, 9K-1K, MICRO DIVIDER TF-246-2 R651 RES NET, 20K, .1%, .2W, MICRO DIVIDE TF-250 R657, R692, R696 RES, 1M, 1%, 100MW, THICK FILM R-418-1M R666 RES NET, 3K, 12K, 0.1%, 100MW, THIN FILM TF-237-2 R670 RES, 1K, 1%, 100MW, THIN FILM R-438-1K R671 RES, .0499, 1%, 100MW, THICK FILM R-418-.0499 R680 RES, 8.87K, 1%, .1W, THIN FILM R-438-8.87K R683, R686 RES, 1K, .1%, .125W, THIN FILM R-456-1K R690, R694 RES, 100K, 5%, 250MW, METAL FILM R-376-100K R700, R701 RES, 49.9K, 1%, 100MW, THICK FILM R-418-49.9K R706, R707, R708, R709, R501,
R514, R529,
RES, 12.1, 1%, 1/4W, METAL FILM R-391-12.1
RES, 100K, 1%, 100MW, THICK FILM R-418-100K
RES, 4.99K, 1%, 100MW, THICK FILM R-418-4.99K
RES, 1K, 1%, 100MW, THICK FILM R-418-1K
Replaceable Parts
Loading...