Tektronix 237-TRX-T User manual

Keithley Instruments
Model 237-TRX-T (CS-737)
28775 Aurora Road Cleveland, Ohio 44139 1-800-833-9200
tek.com/keithley
Three-Slot Triaxial Tee Adapter Information

Description

The Model 237-TRX-T (CS-737) is a 3-slot plug triaxial to dual 3-lug socket tee adapter. You can use the 237-TRX-T adapter to connect multiple inputs to SMUs and electrometers. It also allows connection of external feedback elements to electrometers.
Figure 1: Model 237-TRX-T (CS-737) 3-Slot Triaxial Tee Adapter
To maintain maximum insulation resistance, keep the connectors free of contaminants (such as dirt or oil). If the connections become contaminated, clean them thoroughly with methanol. Allow them to dry completely before use.

Characteristics

Working voltage
Maximum working current
Operating environment
Contact resistance
Insulation resistance
Materials
600 V 1300 V
1 A
0 °C to 50 °C, up to 70% relative humidity at ≤35 °C
< 0.5 Ω
15
10 relative humidity)
Body: Brass with nontarnish finish and Teflon Contacts: Hardened beryllium copper
center conductor to inner shield
PEAK
center conductor and inner shield to outer shell
PEAK
PEAK
Ω, center conductor to inner shield (500 V test voltage, 23 °C at < 40%
®
insulation
PA-305 Rev. D September 2022
*PPA-305D*
1
Model 237-TRX-T (CS-737) Three-Slot Triaxial Tee Adapter Information
he dimensions are shown in the following figure.
T
Figure 2: 237-TRX-T (CS-737) adapter dimensions
The schematic of the 237-TRX-T is shown in the following figure.
Figure 3: 237-TRX-T (CS-737) schematic
The outer shell of the triaxial connector is for protection from voltage on the center and inner shield conductors. Make sure the outer shell is always connected to earth ground or a properly grounded chassis.
Never touch or change the connections when power is applied to the adapter. Always turn off test system power and discharge all capacitors before connecting or disconnecting the adapter.
To prevent voltages from being exposed or connections from shorting together, make sure both ends of the adapter are properly connected before applying voltage.
Failure to do so may result in equipment damage, personal injury, or even death due to electric shock.
2 PA-305 Rev. D September 2022
Loading...
+ 2 hidden pages