Tektronix 2337 Instruction Manual

Tfektronix
COMMITTED TO EXCELLENCE
Tektronix, Inc. P.O. Box 500 Beaverton, Oregon 97077
070-4119-00
Product Group 40
2337
OSCILLOSCOPE
OPERATORS
INSTRUCTION
Serial Number
________________
MANUAL
First Printing SEP 1981 Revised MAY 1983
TABLE OF CONTENTS
Page
LIST OF ILLUSTRATIONS.................................................iii
LIST OF TABLES....................................................................v
OPERATORS SAFETY SUMMARY
....................................
vi
INTRODUCTION....................................................................1
PREPARATION FOR USE
....................................................
2
LINE VOLTAGE SELECTION
.......................................
2
LINE FUSE
........................................................................
4
POWER CORD....................................................................5
CONTROLS, CONNECTORS, AND INDICATORS
..........
6
POWER AND DISPLAY....................................................6
VE R TICAL.........................................................................7
HORIZONTAL..................................................................10
ATRIG GER....................................................................12
BTRIGGER
................................................
15
REAR PANEL..................................................................18
Page
OPERATING CONSIDERATIONS
....................................
20
GRATICULE................................. 20
GROU NDIN G.................................................................20
SIGNAL CONNECTIONS...............................................21
INPUT COUPLING CAPACITOR
PRECHARGING..............................................................21
INSTRUMENT COOLING
............................................
22
INSTRUMENT FA MILIARIZATIO N
...............................
23
INTRODUCTION
...........................................................
23
EQUIPMENT REQUIRED
............................................
23
NORMAL SWEEP DISPLAY..........................................24
DISPLAYING A SIGNAL...............................................26
OPERATOR'S CHECKS AND ADJUSTMENTS
..............
36
INTRODUCTION
...........................................................
36
TRACE ROTATION
......................................................
36
PROBE COMPENSATION
............................................
37
ASTIGMATISM..............................................................38
VERTICAL GAIN CHECK
............................................
39
@
2337 Operators
l
TABLE OF CONTENTS (cont)
Page
BASIC APPLICATIONS.......................................................40
NONDELAYED MEASUREMENTS
.............................
40
AC Peak-to-Peak Voltage
..........................................
40
Instantaneous DC Voltage..........................................42
Algebraic A d d ition
....................................................
43
Common-Mode Rejection..........................................45
Time Duration............................................................46
Frequency
...................................................................
48
Rise Tim e
...................................................................
48 Time Difference Between Two Time-Related Pulses
....................................................
50
Phase Difference
.........................................................
51
Amplitude Comparison
...............................................
53
Time Comparison.......................................................55
DELAYED-SWEEP MAGNIFICATION
........................
57
Magnified Sweep Runs After Delay
...........................
57
Pulse Jitter Time Measurement
..................................
59
Triggered Magnified Sweep
.......................................
60
DELAYED-SWEEP TIME MEASUREMENTS
..............
60
Time D uration............................................................61
Page
Rise T im e....................................................................61
Time Difference Between Repetitive Pulses
..........
63 Time Difference Between Two Time-Related Pulses
....................................................
65
DMM DISPLAYS AND MEASUREMENTS
................
67
Resistance....................................................................67
RMS A C ......................................................................67
DC Volts......................................................................68
OPTIO NS..............................................................................69
OPTION 0 3 ......................................................................69
SPECIFICATION
.................................................................
70
ACCESSORIES......................................................................99
STANDARD ACCESSORIES INCLUDED
...................
99
OPTIONAL POWER CORDS
..........................................
99
APPENDIX A .......................................................................100
CHANGE INFORMATION
ii
2337 Operators
@
LIST OF ILLUSTRATIONS
Figure Page
The 2337 Oscilloscope...........................................................................................................................viii
1 LINE VOLTAGE SELECTOR switch, line fuse, and power c ord
......................................................
3
2 Optional power cords.................................................................................................................................5
3 Power and display controls and indicators.............................................................................................7
4 Vertical controls, connectors, and indicators and calibrator output....................................................8
5 Horizontal controls and indicator..........................................................................................................10
6 A TRIGGER controls, connector, and indicator.................................................................................13
7 DMM and B TRIGGER controls, connectors, and LCD readout........................................................16
8 Rear-panel connectors............................................................................................................................19
9 Graticule measurement m arkings..........................................................................................................20
10 Initial setup for instrument familiarization procedure.........................................................................27
11 Probe compensation..................................................................................................................................38
12 Vertical display accuracy.......................................................................................................................39
13 Peak-to-peak waveform voltage.............................................................................................................41
14 Instantaneous voltage measurement.....................................................................................................43
15 Algebraic addition...............................................................................< .................................................45
16 Common-mode rejection.........................................................................................................................46
17 Time duration...........................................................................................................................................47
@ 2337 Operators **
Fig 18
19
20
21
22
23 24 25 26 27
LIST OF ILLUSTRATIONS (cont)
Page
Rise tim e...................................................................................................................................................49
Time difference between two time-related pulses................................................................................51
Phase difference......................................................................................................................................53
High-resolution phase difference.............................................................................................................53
Delayed-sweep magnification..................................................................................................................58
Pulse jitter................................................................................................................................................59
Time duration using delayed sweep....................................................................................................... 62
Rise time, differential time method....................................................................................................... 63
Time difference between repetitive pulses...........................................................................................64
Time difference between two time-related pulses
................................................................................
66
2337 Operators
LIST OF TABLES
Table Page
1 Line Voltage and Fuse Selection..............................................................................................................4
2 Option 03 Line Voltage and Fuse Selection..........................................................................................4
3 DMM Autoranging................................................................................................................................... 17
4 Equipment Required for Instrument Familiarization Procedure........................................................24
5 Electrical Characteristics.........................................................................................................................72
6 Environmental Characteristics...............................................................................................................95
7 Physical Characteristics........................................................................................................................... 97
8 Option Electrical Characteristics............................................................................................................98
@
2337 Operators
v
OPERATORS SAFETY SUMMARY
The general safety information in this part o f the summary is for both operating and servicing personnel. Specific warnings
and cautions w ill be found throughout the manual where they apply and do not appear in this summary.
Terms in This Manual
CAUTION statements identify conditions or practices that could result in damage to the equipment or other property.
WARNING statements identify conditions or practices that could result in personal injury or loss of life.
Terms as Marked on Equipment
CAUTION indicates a personal injury hazard not
immediately accessible as one reads the markings, or a
hazard to property, including the equipment itself.
DANGER indicates a personal injury hazard immedi ately accessible as one reads the marking.
Symbols in This Manual
A
This symbol indicates where applicable
cautionary or other information is to be found. For maximum input voltage see Table 5.
Symbols as Marked on Equipment
DANGER High voltage.
Protective ground (earth) terminal.
ATTENTION Refer to manual.
VI
2337 Operators
@
Power Source
This product is intended to operate from a power source that does not apply more than 250 volts rms between the supply conductors or between either supply conductor and ground. A protective ground connection by way of the grounding conductor in the power cord is essential for safe operation.
Grounding the Product
This product is grounded through the grounding con ductor of the power cord. To avoid electrical shock, plug the power cord into a properly wired receptable before connecting to the product input or output terminals. A
protective ground connection by way of the grounding
conductor in the power cord is essential for safe operation.
Danger Arising From Loss of Ground
Upon loss of the protective-ground connection, all
accessible conductive parts (including knobs and controls that may appear to be insulating) can render an electric shock.
Use the Proper Power Cord
Use only the power cord and connector specified for
your product.
Use only a power cord that is in good condition.
For detailed information on power cords and con
nectors see Figure 2.
Use the Proper Fuse
To avoid fire hazard, use only a fuse of the correct type,
voltage rating and current rating as specified in the parts
list for your product.
Do Not Operate in Explosive Atmospheres
To avoid explosion, do not operate this product in an explosive atmosphere unless it has been specifically certified for such operation.
Do Not Remove Covers or Panels
To avoid personal injury, do not remove the product
covers or panels. Do not operate the product without the covers and panels properly installed.
@
2337 Operators
Vll
@
INTRODUCTION
The TEKTRONIX 2337 Oscilloscope is a rugged, light
weight, dual-channel, 100-MHz instrument with a general-
purpose digital multimeter (DMM). It also features a
compact crt that provides sharply defined traces. The
vertical system supplies calibrated deflection factors from 5 mV per division to 5 V per division. Sensitivity can be increased to at least 2 mV per division by the variable VOLTS/DIV VAR control. Trigger circuits enable stable triggering over the full bandwidth of the vertical system. The horizontal system provides calibrated sweep speeds from 0.5 s per division to 50 ns per division, along with delayed-sweep features, thus accommodating accurate
relative-time measurements. A X10 magnifier circuit extends the maximum sweep speed to 5 ns per division when the SEC/DIV switch is set to 0.05 ps.
A 3 1/2-digit LCD (liquid crystal display) readout enables rapid measurement of time difference between any two points on the oscilloscope display.
The DMM portion measures dc voltage, resistance, and true rms ac voltage. Measurement values are displayed on
the LCD readout, together with polarity, overrange, and function indicators.
The instrument is shipped with the following standard
accessories:
2 Probe packages
1 Accessory pouch 1 Operators manual 1 Service manual 1 Accessory pouch, zip lock 1 Crt filter, clear plastic
2 1,0-A AGC fast-blow fuses
1 0.5-A AGC fast-blow fuse 1 Pair test leads
For part numbers and further information about both standard and optional accessories, refer to the "Acces sories" page at the back of this manual. Your Tektronix representative or local Tektronix Field Office can also provide accessories information.
@
2337 Operators
1
PREPARATION FOR USE
Refer to the Safety Summary at the front of this manual for power source, grounding, and other safety considera tions pertaining to the use of the 2337. Before connecting the instrument to a power source, read the following infor mation, then verify that the LINE VOLTAGE SELECTOR switch is properly set for the ac power source being used and that the proper power-input fuse is installed.
This instrument may be damaged if operated with the LINE VOLTAGE SELECTOR switch set for the wrong applied ac power input source voltage or if the wrong line fuse is installed.
LINE VOLTAGE SELECTION
The 2337 operates from either a 115-V or a 230-V nominal ac power input source with a line frequency ranging from 48 Hz to 440 Hz. Before connecting the power cord to a power input source, verify that the LINE VOLTAGE SELECTOR switch, located on the rear panel (see Figure 1), is set for the correct nominal ac power input source voltage. To convert the instrument for operation from one line-voltage range to the other, move the LINE VOLTAGE SELECTOR switch to the correct nominal ac source voltage position (see Table 1). If your instrument is equipped with Option 03 (100-V/200-V Power Trans former), use Table 2. The detachable power cord may have
to be changed to match the power source outlet.
2
2337 Operators
@
LINE VOLTAG E
SELECTOR SW ITC H
DETACHAB LE
POW ER CO RD
____________
CO RD-SE T-
SEC UR ING
CLAM P
LINE
FUSE
4119-02
Figure 1. LINE VO LTAG E SELECTOR switch, line fuse, and power cord.
@
2337 Operators
3
Table 1
Line Voltage and Fuse Selection
LINE FUSE
Line Voltage
Selector Switch
Voltage
Position
Range
Fuse Data
115 V Nominal
100 to 132 V
1.0 A, 250 V, Fast-blow
230 V Nominal
200 to 250 V
0.5 A, 250 V, Fast-blow
Table 2
Option 03 Line Voltage and Fuse Selection
Line Voltage
Selector Switch Voltage
Position Range
Fuse Data
100 V Nominal
90 to 115 V
1.0 A, 250 V, Fast-blow
200 V Nominal
180 to 230 V
0.5 A, 250 V, Fast-blow
To verify that the instrument power-input fuse is of proper value for the nominal ac source voltage, perform the following procedure:
1. Press in the fuse holder cap and release it with a slight counterclockwise rotation.
2. Pull the cap (with the attached fuse inside) out of the fuse holder.
3. Verify proper fuse value (see Tables 1 and 2).
4
2337 Operators
@
POWER CORD
This instrument has a detachable, three-wire power cord with a three-contact plug for connection to both the power source and protective ground. Its power cord is secured to the rear panel by a cord-set-securing clamp. The plug protective-ground contact connects (through the power- cord protective grounding conductor) to the accessible metal parts of the instrument. For electrical-shock pro
tection, insert this plug into a power source outlet that has
a properly grounded protective-ground contact.
Instruments are shipped with the required power cord
as ordered by the customer. Available power cord options are illustrated in Figure 2, and their part numbers are listed on the "Accessories" page at the back of this manual.
Contact your Tektronix representative or local Tektronix
Field Office for additional power-cord information.
NOTE
See APPENDIX A at the back of this manual for fur
ther power input information. Figure 2. Optional power cords.
Plug
Configuration
Usage
Nominal
Line-
Voltage
(AC)
Reference Standards
Option #
North
American
120 V/
15A
120 V
ANSI C73.11*h NEMA 5-15-P IEC 83c
Standard
Universal
Euro
240V/
10-16A
240V
CEE (7), II, IV,
V lld
IEC 83c
A1
UK
240V/
13A
240V
BS 136
IEC 83c
A2
&
Australian
240V/
10A
240V
ASC 112t A3
North
American
240V /
15A
240V
ANSI C73.201 NEMA 6-15-P IEC 83c
A4
aANSIAmerican National Standards Institute bNEM ANational Electrical Manufacturer's Association CIECInternational Electrotechnical Commission dCEEInternational Commission on Rules for the Approval of
Electrical Equipment
eBSBritish Standards Institution fASStandards Association of Australia 2931-05
@
2337 Operators
5
CONTROLS, CONNECTORS, AND INDICATORS
This section of the manual will familiarize the operator with the location and function of instrument controls, connectors, and indicators.
POWER AND DISPLAY
Refer to Figure 3 for location of items 1 through 8.
Q POWER SwitchTurns instrument power on and off.
Press in for ON; press again for OFF.
(T ) FOCUS ControlAdjusts for optimum display
definition.
(T ) ASTIG ControlScrewdriver control used in con
junction with the FOCUS control to obtain a well- defined display over the entire graticule area. It does
not require readjustment during normal operation
of the instrument.
(T ) INTEN ControlDetermines the brightness of the crt
display (has no effect when BEAM FIND switch is pressed in).
(T ) BEAM FIND SwitchWhen held in, compresses the
display to within the graticule area and provides a visible viewing intensity to aid in locating off-screen displays.
(T ) TRACE ROTATION ControlScrewdriver control
used to align the crt trace with the horizontal grati cule lines.
Internal GraticuleEliminates parallax viewing error between the trace and graticule lines. Rise-time amplitude measurement points are indicated at the left edge of the graticule.
(T ) SERIALand Mod SlotsTheSERIALsIot is imprinted
with the instrument's serial number. The Mod slot contains the option number that has been installed in the instrument.
6 2337 Operators @
VERTICAL
Refer to Figure 4 for location of items 9 through 19.
(^T) AMPL CAL ConnectorProvides a 0.2 V, positive
going square-wave voltage (at approximately 1 kHz) that permits the operator to compensate voltage probes and to check oscilloscope vertical operation.
It is not intended to verify time-base calibration.
® CH 1 OR X and CH 2 OR Y ConnectorsProvide for
application of external signals to the inputs of the vertical deflection system or for an X-Y display. In the X-Y mode, the signal connected to the CH 1 OR X connector provides horizontal deflection, and the signal connected to the CH 2 OR Y connector provides vertical deflection.
@ Input Coupling Switches (AC-GND-DC)-Select the
method of coupling input signals to the vertical deflection system.
ACInput signal is capacitively coupled to the
vertical amplifier. The dc component of the
input signal is blocked. Low-frequency lim it
(3 dB point) is approximately 10 Hz.
@
2337 Operators
7
Figure 3. Power and display controls and indicators.
Figure 4. Vertical controls, connectors, and indicators and calibrator output.
GNDThe input of the vertical amplifier is grounded to provide a zero (ground) reference voltage display (does not ground the input signal). Allows precharging the input coupling capacitor.
DCAll frequency components of the input signal are coupled to the vertical deflection system.
(j? ) CH 1 VOLTS/DIV and CH 2 VOLTS/DIV Switches-
Select the vertical deflection factor in a 1-2-5 sequence. VAR control must be in detent to obtain a calibrated deflection factor.
IX PROBEIndicates the deflection factor selected when using either a IX probe or coaxial cable.
10X PROBEIndicates the deflection factor selected when using a 10X probe.
(l? ) VAR ControlsProvide continuously variable uncal
ibrated deflection factors between the calibrated settings of the VOLTS/DIV switches when rotated clockwise out of the detent position. Channel 1 VOLTS/DIV VAR control is inoperative when X-Y VERTICAL MODE is selected.
8
2337 Operators
@
M4) UNCAL IndicatorLED illuminates to indicate that
either Channel 1 or Channel 2 VOLTS/DIV VAR control is out of calibrated detent (vertical deflection factor is uncalibrated).
VERTICAL MODE SwitchesFive push-button switches that select the mode of operation for the vertical amplifier system.
CH 1Selects only the Channel 1 input signal for display.
ALTThe display alternates between Channel 1 and Channel 2 input signals. The alternation occurs during retrace at the end of each sweep. This mode is useful for viewing both vertical input signals at sweep speeds from 0.2 ms per division to 0.05 ps per division.
CHOPThe display switches between the Channel
1 and Channel 2 input signals during the sweep. The switching rate is approximately 500 kHz. This mode is useful for viewing both Channel 1 and Channel 2 inputs at sweep speeds from 0.5 ms per division to 0.5 s per division.
ADDSelects the algebraic sum of the Channel 1 and Channel 2 input signals for display.
CH 2Selects only the Channel 2 input signal for display.
AUTOPress in both ALT and CHOP buttons.
The A Sweep circuitry automatically selects the
most useful switching method (ALT or CHOP) for dual displays.
X-YPress in both CH 1 and CH 2 buttons. The
X-signal is applied through the Channel 1 input connector, and the Y-signal is applied through the Channel 2 input connector.
(j? ) CH 2 INVERT SwitchInverts Channel 2 display
when button is pressed in. Push button must be pressed in a second time to release it and regain a noninverted display.
(l7 ) POSITION ControlsDetermine the vertical position
of the displays on the crt. When X-Y VERTICAL MODE is selected, the Channel 2 POSITION control moves the display vertically (Y-axis), and the Hori zontal POSITION control moves the display hori zontally (X-axis).
@
2337 Operators
9
(T?) BW LIMIT SwitchLimits the bandwidth of the
vertical amplifier to approximately 20 MHz when
pressed in. Push button must be pressed a second time to release it and regain full 100-MHz bandwidth operation. Provides a method for reducing inter ference from unwanted high-frequency signals when viewing low-frequency signals.
(fiT) TRIG VIEW Switch-Press in and hold this push
button to display a sample of the signal present in
the A Trigger amplifier (for all A TRIGGER SOURCE switch settings except VERT MODE). All other signal displays are removed while the TRIG VIEW push button is held in.
HORIZONTAL
Refer to Figure 5 for location of items 20 through 26.
(20) B DELAY TIME POSITION-A TIME POSITION
ControlsSelect the amount of delay time between start of the A Sweep and start of the B Sweep. Delay time is variable to at least 10 times the A SEC/DIV switch setting. The B DELAY TIME POSITION
(outer knob) controls the reference point when the
10
2337 Operators
4119 06
@
Figure 5. Horizontal controls and indicator.
B TRIGGER SOURCE switch is set to either A TIME or RUN AFTER DLY. The A TIME (inner knob) controls the time-measurement point only when the
B TRIGGER SOURCE switch is set to A TIME. When the time-measurement point is to the left of the
reference point, the LCD readout indicates a negative time difference.
(2?) A AND B SEC/DIV SwitchesSelects the sweep
speed for the A and B Sweep generators in a 1-2-5 sequence. The A SEC/DIV switch sets the time
between the B Sweeps (delay time). For calibrated sweep rates, the TIME (PULL) VAR control must be
in the calibrated detent (fully clockwise position).
A SEC/DIVThe A Sweep speed is shown between the two black lines on the clear plastic skirt. This switch also selects the delay time (used in conjunction with the B DELAY TIME POSI
TION control) for delayed sweep operation.
B SEC/DIVThe B Sweep speed is set by pulling the inner knob and rotating it to a setting shown by the white line scribed on the knob. The B Sweep circuit is used for delayed sweep operation only.
(22) TIME (PULL) VAR ControlProvides continuously
variable, uncalibrated A Sweep speeds between SEC/
DIV switch settings to at least 2.5 times the cali brated setting (extends slowest sweep speed to at least 1.25 s per division). To operate this control, pull out the VAR knob and rotate it counterclockwise out of the detent.
(2?) UNCAL Indicator LEDIlluminates to indicate that
the A Sweep speed is uncalibrated when the TIME
(PULL) VAR control is rotated out of the calibrated
detent.
(24) HORIZ MODE Switches-Three push-button switches
that select the mode of operation for the horizontal deflection system.
AHorizontal deflection is provided by the A Sweep generator at a sweep speed determined by the setting of the A SEC/DIV switch.
A INTENHorizontal deflection is provided by the A Sweep generator at a speed determined by the A SEC/DIV switch. The B Sweep generator provides an intensified zone on the display. The
@
2337 Operators
11
length of the intensified zone is determined by the setting of the B SEC/DIV switch. The location of the intensified zone is determined by the setting of the B DELAY TIME POSITION-A TIME POSITION controls.
BHorizontal deflection is provided by the B Sweep generator at a sweep speed determined by the setting of the B SEC/DIV switch. The start of the B Sweep is delayed from the start of the A Sweep by a time determined by the settings of the A SEC/DIV switch and the B DELAY TIME POSITION-A TIME POSITION controls.
(25) X10 MAG SwitchWhen pressed in, increases the
displayed sweep speed by a factor of 10. Extends fastest sweep speed to 5 ns per division. Push button must be pressed in a second time to release it and regain the XI sweep speed.
(26) POSITION ControlPositions the display hori
zontally in all modes. Provides both coarse and fine control action. Reverse the direction of rotation to actuate fine positioning feature. When X-Y VER TICAL MODE is selected, the Horizontal POSITION control moves the display horizontally (X-axis).
A TRIGGER
Refer to Figure 6 for location of items 27 through 34.
(27) SLOPE SwitchSelects the slope of the signal that
triggers the sweep.
+ (plus)When push button is released out, sweep is triggered from the positive-going slope of the trigger signal.
(minus)When push button is pressed in, sweep is triggered from the negative-going slope of the trigger signal.
(2?) LEVEL ControlSelects the amplitude point on the
trigger signal at which the sweep is triggered. The
LEVEL control is usually adjusted for the desired display after trigger SLOPE, COUPLING, and SOURCE switch settings have been selected.
(29) Trigger Mode SwitchesThree push-button switches
that determine the trigger mode for the A Sweep.
AUTOPermits triggering on waveforms with repetition rates down to approximately 10 Hz.
12 2337 Operators
@
Sweep free runs and provides a baseline trace either in the absence of an adequate trigger signal or when the repetition rate of the trigger signal
is below approximately 10 Hz.
NORMSweep is initiated when an adequate
trigger signal is applied. In the absence of a trigger
signal, no baseline trace will be present.
SGL SWPPress in the spring-return push button momentarily to arm the A Sweep circuit for a single sweep display. This mode operates the same as NORM, except only one sweep is displayed for each trigger signal. Another single sweep cannot be displayed until the SGL SWP push button is momentarily pressed in again to reset the A Sweep circuit. This mode is useful for displaying and photographing either nonrepetitive signals or signals that cause unstable conventional displays
(e.g., signals that vary in amplitude, shape, or
time).
(30) TRIG'D-READY Indicator LEDIlluminates when
either AUTO or NORM Trigger Mode is selected to
indicate that the A Sweep is triggered (TRIG'D).
When SGL SWP Trigger Mode is selected, the LED
______
illuminates to indicate that the trigger circuit is
armed (READY) for a single sweep display. Pig<
@ 2337 Operators
13
ire 6. A TRIGGER controls, connector, and indicator.
^3?) SOURCE SwitchDetermines the source of the trig- (33^
ger signals coupled to the input of the trigger circuit.
VERT MODEThe internal trigger source is determined by the signals selected for display by
the VERTICAL MODE switches.
CH 1The signal applied to the CH 1 input connector is the source of the trigger signal.
CH 2The signal applied, to the CH 2 input connector is the source of the trigger signal.
LINEProvides a trigger signal from a sample of the ac-power-source waveform. This trigger source is useful when channel input signals are time related (multiple or submultiple) to the frequency
of the power-input source voltage.
EXTPermits triggering on signals applied to the External Trigger Input connector (A EXT).
EXT-MO—External trigger signals are attenuated by a factor of 10.
(32) A EXT ConnectorProvides a means of applying
external signals to the trigger circuit.
14
2337 Operators
(COUPLING SwitchDetermines the method used to
couple the trigger signal to the input of the trigger circu it.
ACSignals above 20 Hz are capacitively coupled, blocking any dc components of the signal. Signals below 20 Hz are attenuated.
LF REJSignals are capacitively coupled. The
dc component is blocked, and signals below
approximately 50 kHz are attenuated. This
position is useful for providing a stable display of the high-frequency components of a complex waveform.
HF REJSignals are capacitively coupled. The dc component is blocked, and signals below approximately 20 Hz and above approximately 50 kHz are attenuated. This position is useful for providing a stable display of the low-frequency components of a complex waveform.
DCAll components of the signal are coupled to the A Trigger circuitry. This position is useful for displaying low-frequency or low-repetition-rate
signals.
@
(34) TRIG HOLDOFF (PUSH) VAR Control-Provides
continuous control of holdoff time between sweeps. This control improves the ability to trigger on aperi odic signals (such as complex digital waveforms) and increases the minimum holdoff time to at least
2.5 times at any sweep speed.
B TRIGGER
Refer to Figure 7 for location of items 35 through 41.
(35) LEVEL ControlSelects the amplitude point on the
trigger signal at which the sweep is triggered. This control is usually adjusted for the desired display
after Trigger SLOPE and SOURCE switch settings
have been selected.
(36) SOURCE SwitchDetermines the mode of operation
for the B Sweep and the signal source for the B
Trigger.
A TIMEProvides two intensified zones on the crt trace for differential time measurements. The time difference between the two intensified zones
is determined by the B DELAY TIME POSITION
and the A TIME POSITION controls. Time differ-
@ 2337 Operators
ence is displayed on the LCD readout in seconds
(s), milliseconds (ms), or microseconds (ps). The LCD readout will display UNCAL when the TIME (PULL) VAR control is out of calibrated detent. With the HORIZ MODE set to A INTEN,
alternation of the reference intensified zone occurs
at the end of each sweep. With the HORIZ MODE set to B, the start of the B Sweep alternates
between the setting of the reference intensified
zone and the setting of the measurement inten
sified zone.
RUNS AFTER DLYThe B Sweep starts immedi ately after the delay time selected by the DELAY TIME POSITION control and is independent of
the B Trigger signal.
VERT MODEAllows the internal trigger source to be determined by the vertical mode of operation.
CH 1—The signal applied to the CH 1 input con nector is the source of the trigger signal.
CH 2The signal applied to the CH 2 input
connector is the source of the trigger signal.
EXTPermits triggering on signals applied to the
External Trigger Input (B EXT) connector.
15
Figure 7. DMM and B TRIGGER controls, connectors, and LCD readout.
16
2337 Operators
@
(3?) SLOPE SwitchSelects the slope of the signal that
triggers the sweep.
+ (plus)Sweep is triggered on the positive-going portion of the trigger signal.
(minus)Sweep is triggered on the negative
going portion of the trigger signal.
(38) B EXT ConnectorProvides a means of introducing
external signals into the B Trigger Generator.
(39) RANGE/FUNCTION SwitchSelects the function
and the range of values for the DMM when not in the A TIME mode. The LCD readout will indicate the function in which the DMM is operating by displaying AC V, V, or 12.
Each DC, RMS AC, and 12 setting, except the 500V DC and the 350V RMS AC, consists of two internal
(upper and lower) ranges. The internal ranges are fully autoranging within the DMM RANGE selected. See Table 3 for the LCD readout values at which autoranging will occur for each DMM RANGE selected.
Table 3
DMM Autoranging
DOWNRANGE UPRANGE
When LCD When LCD Reads Less Reads More
RANGE
FUNCTION
Than
Than
500 V
DC
NA
NA
200 V
DC 10.0 V
19.99 V
2 V
DC
0.100 V
199.9 mV
350 V
RMS AC
NA
NA
200 V
RMS AC
10.0 V 19.99 V
2 V
RMS AC
0.100 V
199.9 mV
20 M
12
1.00 M12 1.999 M12
200 K
12
1.00 K12
1.999 K12
2 K
12 0.100 K12 199.912
12Measures resistance from 0 12 to 20 M12 in three ranges.
RMS ACMeasures true rms value of the input signal (20 Hz to 20 kHz) from 0 V to 350 V in
three ranges. The LCD readout will display
REV JAN 1982
2337 Operators
17
UNCAL when the peak-to-peak input voltage is greater than three times the RMS AC range selected.
DCMeasures dc voltage from 0 to 500 V in three ranges.
A TIMEDisconnects the DMM from the LCD readout and allows the instrument to display a digital readout of time difference between any two points on the oscilloscope display.
The maximum safe input voltage is 500 V (dc +
peak ac) between the + (positive) and (negative)
inputs or between the + and inputs to ground.
(40) DMM INPUT ConnectorsTwo banana jacks provide
positive (red) and negative (black) inputs for voltage and resistance measurements.
(4?) Readout-Consists of a 3 1/2-digit LCD unit which is
used to display measurements selected by the RANGE/FUNCTION switch. No polarity indication
is displayed for positive values. Negative polarity indication is automatic for negative values. Decimal point indication is automatic. In an overrange con dition, a "1" will be displayed on the left side of the LCD readout, followed by three blank digits and a decimal point.
REAR PANEL
Refer to Figure 8 for location of items 42 and 43.
(42) GND ConnectorProvides direct connection to
instrument chassis ground.
(43) EXT Z AXIS INPUT Connector-Provides a means of
connecting external signals to the Z-Axis amplifier to intensity modulate the crt display. Applied signals do not affect display waveshape. Signals with fast rise time and fall time provide the most abrupt intensity change. Positive-going signals decrease
the intensity, and a 5-V p-p signal will produce
noticeable modulation. Z-axis signals must be time- related to the display to obtain a stable presentation on the crt.
18
2337 Operators
@
EXT I AX IS IN PUT IQ Kft 25V PEAK MAX POSITIVE GOMG W U T DECREASES IN TEN SITY 5V P P CAUSES NOTKEABIE MODULATIO N AT NORMAL INTENSITY
CAUTION
OO NOT REMOVE COVERS REFER SERVICING TO OUAUEIEO PERSONNEL
FOR CONTINUED FIRE PROTECTION. REPIACE ONLY W ITH S PEOK O TYPE AMO RATED FUSE DISCONNECT POWER INPUT BEFORE REPLACING FUSE
NOMINAL RANGE FUSE gO V
i;$v_
a w a o v o-s a fa s t
POWER MAX WATTS 80
MAX VA n
FREQ 48440Hf
TEKTRONIX. INC. BEAVERTON OREGON. U S A
4119-09
Figure 8. Rear-panel connectors.
@
2337 Operators
19
OPERATING CONSIDERATIONS
This section contains basic operating information and techniques that should be considered before attempting any measurements.
GRATICULE
The graticule is internally marked on the faceplate of the crt to enable accurate measurements without parallax error (see Figure 9). It is marked with eight vertical and ten
horizontal major divisions. In addition, each major division is divided into five subdivisions. The vertical deflection factors and horizontal timing are calibrated to the graticule so that accurate measurements can be made directly from the crt. Also, percentage marks for the measurement of rise and fall times are located on the left side of the graticule.
GROUNDING
The most reliable signal measurements are made when
the 2337 and the unit under test are connected by a
1ST OR LEFT 11TH OR R IGHT
VERT ICA L VERT ICAL GRATIULE GRATICULE
LINE L IN E /
/ )
( . . .
-
-
-
it-
RISE A ND
FALL TIME
ME ASURE ME NT
PERC EN TA GE j'JT *
MA RKE RS ' aK* 1
CENTER
HOR IZON TAL
To a i GRATICULE
K / A L | m r
ICULE LINE
NE 411 9 10
Figure 9. Graticule measurement markings.
20
2337 Operators
@
common reference (ground lead) in addition to the signal
lead or probe. The probe's ground lead provides the best grounding method for signal interconnection and ensures the maximum amount of signal-lead shielding in the probe cable. A separate ground lead can also be connected from
the unit under test to the oscilloscope GND connector
located on the rear panel.
SIGNAL CONNECTIONS
Probes
Generally, probes offer the most convenient means of connecting an input signal to the instrument. They are shielded to prevent pickup of electromagnetic interference, and the supplied 10X probe offers a high input impedance that minimizes circuit loading. This allows the circuit under test to operate with a minimum of change from the normal condition of the circuit when measurements are being made.
Coaxial Cables
Cables may also be used to connect signals to the input connectors, but they may have considerable effect on the accuracy of a displayed waveform. To maintain the original
frequency characteristics of an applied signal, only high- quality, low-loss coaxial cables should be used. Coaxial cables should be terminated at both ends in their char acteristic impedance. If this is not possible, use suitable impedance-matching devices.
INPUT COUPLING CAPACITOR
PRECHARGING
When the input coupling switch is set to GND, the input signal is connected to ground through the input coupling capacitor in series with an 800-2 resistor to form a pre charging network. This network allows the input coupling capacitor to charge to the average dc-voltage level of the signal applied to the probe. Thus, any large voltage transients that may accidentally be generated will not be applied to the amplifier input when input coupling is switched from GND to AC. The precharging network also provides a measure of protection to the external circuitry by reducing the current levels that can be drawn from the external circuitry during capacitor charging.
The following procedure should be used whenever the probe tip is connected to a signal source having a different
@ 2337 Operators
21
dc level than that previously applied, especially if the dc-
level difference is more than 10 times the VOLTS/DIV
switch setting:
4. Connect the probe tip to the signal source.
1. Set the AC-GND-DC switch to GND before connect
ing the probe tip to a signal source.
5. Wait several seconds for the input coupling capacitor
to charge.
NO TE
The outer shells of the A EXT, CH 1 OR X, and CH 2 OR Y connectors are attached to the 2337 chassis ground.
6. Set the AC-GND-DC switch to AC. The display will remain on the screen, and the ac component of the signal can be measured in the normal manner.
2. Touch the probe tip to the oscilloscope chassis
ground.
INSTRUMENT COOLING
3. Wait several seconds for the input coupling capacitor
to discharge.
To maintain adequate instrument cooling, the ventila
tion holes on both sides of the equipment cabinet must
remain free of obstructions.
22
2337 Operators
@
INSTRUMENT FAMILIARIZATION
INTRODUCTION
The procedures in this section are designed to assist you in quickly becoming familiar with the 2337. They provide information which demonstrates the use of all the controls, connectors, and indicators and will enable you to effi ciently operate the instrument.
Before proceeding with these instructions, verify that the LINE VOLTAGE SELECTOR switch is placed in the proper position and that the correct line fuse is installed for the available ac-power-input source voltage. Refer to the "Preparation for Use" instructions in this manual for this information. Verify that the POWER switch is OFF (push button out), then plug the power cord into the ac- power-input-source outlet.
If during the performance of these procedures an
improper indication or instrument malfunction is noted, first verify correct operation of associated equipment. Should the malfunction persist, refer the instrument to qualified service personnel for repair or adjustment.
EQUIPMENT REQUIRED
The equipment listed in Table 4, or the equivalent, is
required to complete these familiarization procedures for
the 2337.
@
2337 Operators 23
NORMAL SWEEP DISPLAY
Table 4
Equipment Required for Instrument
Familiarization Procedure
First obtain a Normal Sweep Display (baseline trace),
using the following procedure.
Description
Minimum Specification
Calibration Generator
Standard-amplitude accuracy, ±0.25%; signal amplitude, 2 mV to 50 V; output signal, 1-kHz square wave. Fast-rise
repetition rate, 1 to 100 kHz; rise time, 1 ns or less; signal amplitude, 100 mV to 1 V; aberrations, ±2%.
Dual-Input
Coupler
Connectors, bnc female-to-dual-bnc male.
Cable
(2 required)
Impedance, 50 £2; length, 42 in.;
connectors, bnc.
Adapter
Connectors, bnc female-to-bnc female.
Termination
Impedance, 50 £1; connectors, bnc.
1. Preset the instrument front-panel controls as follows:
Display
INTEN
ASTIG FOCUS
Vertical (both CH 1 and
AC-GND-DC VOLTS/DIV VOLTS/DIV VAR
VERTICAL MODE CH 2 INVERT BW LIMIT
POSITION
Fully counterclockwise (minimum)
Midrange
Midrange
CH 2 if applicable)
AC 50 m (IX) Calibrated detent (fully
counterclockwise)
Select CH 1 Off (push button out) Not limited (push button
out) Midrange
24
2337 Operators
@
Horizontal
A AND B SEC/DIV
TIME (PULL) VAR
HORIZ MODE X10 MAG POSITION B DELAY TIME
POSITION
A TIME POSITION
Locked together at 0.5 ms Pull out the VAR knob and
set it to the calibrated detent
(fully clockwise), then push
in the VAR knob. Select A Off (push button out) Midrange
Fully counterclockwise
Midrange
B Trigger
SLOPE + (up)
LEVEL Midrange
SOURCE A TIME
DMM
RANGE/FUNCTION A TIME
2. Press in the POWER switch button (ON) and allow
the instrument to warm up for 20 minutes.
3. Adjust the INTEN control for desired display
brightness.
A Trigger
4. Adjust the Vertical and Horizontal POSITION
controls to center the trace on the screen.
SLOPE
LEVEL Trigger Mode COUPLING SOURCE TRIG HOLDOFF
(PUSH) VAR
+ (push button out) Midrange Select AUTO AC VERT MODE
Fully clockwise and pushed in
NOTE
Normally, the resulting trace will be parallel with the center horizontal graticule line and should not require adjustment. If trace alignment is required, see the " Trace Rotation" adjustment procedure under "Operator's Checks and Adjustments."
@
2337 Operators
25
DISPLAYING A SIGNAL
After obtaining a Normal Sweep Display (baseline trace),
you are now ready to connect an input signal and display it on the crt screen.
1. Connect the calibration generator standard-amplitude
output to both the CH 1 and CH 2 inputs as shown in
Figure 10.
2. Set the calibration generator for a standard-amplitude
1-kHz square-wave signal and adjust its output to obtain
a vertical display of 4 divisions.
3. Adjust the Channel 1 POSITION control to center
the display vertically on the screen.
4. Adjust the A TRIGGER LEVEL control for a stable
triggered display.
NOTE
The A TRIGGER TRIG'D-READY light should
illuminate to indicate that the A Sweep is triggered.
5. Rotate the FOCUS control between its maximum clockwise and counterclockwise positions. The display should become blurred on either side of the optimum control setting.
6. Set the FOCUS control for a sharp, well-defined display over the entire trace length. If a well-focused display cannot be obtained, see the "Astigmatism" adjust ment procedure under "Operator's Checks and Adjust ments." The ASTIG adjustment is used in conjunction with the FOCUS control to initially obtain a well-defined display over the entire trace length. Once set, the ASTIG control usually requires little or no adjustment.
7. Move the display off the screen using the Channel 1
POSITION control.
8. Press in and hold the BEAM FIND push button; the display should reappear on the screen. Adjust the Channel 1 and Horizontal POSITION controls to center the trace both vertically and horizontally. Release the
BEAM FIND button; the display should remain within the
viewing area.
9. Adjust the INTEN control counterclockwise until
the display disappears.
26 2337 Operators
@
CALIBRATION
GENERATOR
4119-11
Figure 10. Initial setup for instrument familiarization procedure.
@
2337 Operators 27
10. Press in and hold the BEAM FIND push button;
the display should reappear. Release the BEAM FIND
button and adjust the INTEN control to desired display brightness.
Using the Vertical Section
1. Set the Channel 1 AC-GND-DC switch to GND.
2. Adjust the trace to the center horizontal graticule
line.
3. Set the Channel 1 AC-GND-DC switch to DC.
4. Observe that the bottom of the display remains at
the center horizontal graticule line (ground reference).
5. Set the Channel 1 AC-GND-DC switch to AC.
6. Observe that the display is centered approximately
at the center horizontal line.
7. Set the CH 1 VOLTS/DIV switch to 0.1 and observe
that a 2-division vertical display appears.
8. Rotate the CH 1 VOLTS/DIV VAR control fully clockwise. Note that the vertical UNCAL light illuminates when the VOLTS/DIV VAR control is out of the calibrated detent.
9. Observe that maximum vertical deflection occurs when the VOLTS/DIV VAR control is fully clockwise.
10. Rotate the CH 1 VOLTS/DIV VAR control fully
counterclockwise to its calibrated detent.
11. Select CH 2 VERTICAL MODE and perform preceding steps 1 through 10 using Channel 2 controls. Performance should be similar to Channel 1.
12. Set both Channel 1 and Channel 2 AC-GND-DC switches to DC. Ensure that both CH 1 and CH 2 VOLTS/ DIV switches are set to 0.1 for 2-division displays.
13. Select ADD VERTICAL MODE and observe that the resulting display is 4 divisions in amplitude. Both Channel 1 and Channel 2 POSITION controls should move the display. Recenter the display on the screen.
28 2337 Operators
14. Press in the CH 2 INVERT push button to invert the
Channel 2 signal.
15. Observe that the display is a straight line, indicating
that the algebraic sum of the two signals is zero.
16. Set the CH 2 VOLTS/DIV switch to 50 m (IX).
17. Observe the 2-division display, indicating that the
algebraic sum of the two signals is no longer zero.
18. Press in the CH 2 INVERT push button again to
release it. Observe a noninverting display having a 6-division
signal amplitude.
19. Select CH 1 VERTICAL MODE and set CH 1
VOLTS/DIV switch to 50 m (IX).
20. Adjust the Channel 1 POSITION control to center
the display on the screen.
21. Set the A TRIGGER SOURCE switch to CH 1.
22. Press in and hold the TRIG VIEW push button. Observe the Channel 1 trigger signal that is present in the A Trigger amplifier.
NOTE
When in the TRIG VIEW mode, a trigger signal is displayed on the crt screen for every A TRIGGER COUPLING switch position and for every A TRIG
GER SOURCE switch position except VERT MODE.
23. Still holding in the TRIG VIEW push button, rotate
the A TRIGGER LEVEL control between its maximum
clockwise and counterclockwise positions.
24. Observe that maximum signal amplitude and most stable display occurs when the displayed signal is at the approximate center of the graticule area.
25. Release the TRIG VIEW button and set the A
TRIGGER COUPLING switch to AC. Note that a normal
display is regained.
26. Set both Channel 1 and Channel 2 AC-GND-DC
switches to GND.
@
2337 Operators
29
27. Position Channel 1 trace two divisions above the
center graticule line.
28. Select CH 2 VERTICAL MODE and position the Channel 2 trace two divisions below the center graticule line.
29. Select ALT VERTICAL MODE and rotate the A SEC/DIV switch throughout its range. The display will alternate between channels at all sweep speeds. This mode is most useful for sweep speeds from 0.05 ns to 0.2 ms per division.
30. Select CHOP VERTICAL MODE and rotate the A SEC/DIV switch throughout its range. A dual-trace display will be presented at all sweep speeds, but unlike the ALT
mode, both Channel 1 and Channel 2 signals are displayed for each sweep speed on a time-shared basis. This mode is most useful for sweep speeds from 0.5 ms to 0.5 s per division.
31. Set the A TRIGGER SOURCE switch to VERT
MODE.
32. Observe the switching between Channel 1 and Channel 2 at the higher sweep speeds (indicated by the segmented trace).
33. Select AUTO VERTICAL MODE by simultaneously pressing in both ALT and CHOP push buttons.
34. Rotate the A SEC/DIV switch throughout its range. Note that a dual-trace display is present at all sweep speeds.
35. Select CH 1 VERTICAL MODE and set Channel 1
AC-GND-DC switch to DC. Recenter the display on the
screen.
Using the Horizontal Section
1. Return the A SEC/DIV switch to 0.5 ms and note
the display for future comparison in step 3.
2. Set the A SEC/DIV switch to 5 ms and press in the
X10 MAG push button.
3. Observe that the display is similar to that obtained in step 1.
30 2337 Operators
@
4. Rotate the Horizontal POSITION control throughout
its range. Observe that the display can be positioned off
the screen horizontally with X I0 MAG selected.
11. Observe that the sweep rate is approximately 2.5
times slower than in step 9, as indicated by more cycles
displayed on the screen.
5. Press in the X10 MAG push button again to release
it for X I sweep.
12. Return the TIME (PULL) VAR control to its cali
brated detent (fully clockwise) and push in the knob.
6. Return the A and B SEC/DIV switches to 0.5 ms.
Using the A Trigger Section
7. Rotate the TRIG HOLDOFF (PUSH) VAR knob out
of detent to its maximum counterclockwise position.
1. Rotate the A TRIGGER LEVEL control between its maximum clockwise and counterclockwise positions. The display will free-run near each lim it of rotation. Observe that the TRIG'D-READY light illuminates only when the display is triggered.
8. Observe that the crt trace becomes dimmer as the
holdoff between sweeps is increased.
2. Adjust the A TRIGGER LEVEL control for a stable
display.
9. Return the TRIG HOLDOFF (PUSH) VAR control to its calibrated detent (fully clockwise). Note the display for future comparison in step 11.
3. Press in the A TRIGGER SLOPE push button to the (minus) position. Observe that the display starts on the negative-going slope of the applied signal.
10. Pull out the TIME (PULL) VAR knob and rotate the control out of detent to its maximum counterclockwise position.
4. Press in the A TRIGGER SLOPE push button again
to release it to the + (plus) position. Observe that the
@ 2337 Operators
31
display starts on the positive-going slope of the applied signal.
5. Select the A TRIGGER NORM mode.
6. Rotate the A TRIGGER LEVEL control between its maximum clockwise and counterclockwise positions. Observe that the display is presented when correctly triggered.
7. Set the A TRIGGER COUPLING switch to DC and the A TRIGGER SOURCE switch to VERT MODE.
8. Rotate the Channel 1 POSITION control until the display becomes unstable. Note that changing the vertical POSITION control setting affects the dc trigger level with the A TRIGGER SOURCE switch set to VERT MODE.
9. Adjust the Channel 1 POSITION control to center the display on the screen.
10. Set the A TRIGGER COUPLING switch to AC and
the A TRIGGER SOURCE switch to CH 1.
11. Remove the calibration signal from the CH 1 input
connector.
12. Press in the A TRIGGER SGL SWP push button
momentarily for single-sweep operation.
13. Observe that the TRIG'D-READY light illuminates, indicating that the A trigger circuit is armed (READY) for a single-sweep display. No display should be present on the crt screen.
14. Reconnect the calibration signal to the CH 1 input connector. A single sweep of the applied signal should appear on the screen. When the TRIG'D-READY light is out, another single sweep cannot be displayed until the SGL SWP button is pressed in again to reset the A Trigger circuit.
15. Select the A TRIGGER AUTO mode and set the A TRIGGER SOURCE switch to EXT.
16. Remove the calibration signal from the CH 2 input connector and connect it to the A EXT input connector.
32 2337 Operators
@
17. Set the Channel 1 VOLTS/DIV switch to 0.5 (IX) and adjust the output of the calibration generator to pro vide a 4-division display. Adjust the A TRIGGER LEVEL control for a stable display.
18. Set the A TRIGGER SOURCE switch to EXT -5- 10.
19. Observe that adjustment of the A TRIGGER LEVEL control provides a triggered display over a narrower
range than in preceding step 17, indicating trigger-signal
attenuation.
Using the Delayed-Sweep Controls
1. Remove the calibration signal from the A EXT input
connector and connect it to the CH 2 input connector.
2. Set the A TRIGGER SOURCE switch to VERT MODE and adjust the A TRIGGER LEVEL control for a stable display.
3. Set B SEC/DIV switch to 0.1 ms and select the A INTEN HORIZ MODE. Ensure that the B TRIGGER SOURCE switch is set to A TIME.
4. Observe that two intensified zones, approximately
1 division in length, appear on the display.
5. Rotate both the B DELAY TIME POSITION control (reference point) and the A TIME POSITION control (measurement point) throughout their ranges; the inten
sified zones move continuously across the display as the controls are rotated.
6. Observe that the time difference between the start of the reference intensified zone and the start of the measurement intensified zone is displayed on the LCD Readout.
7. Select the B HORIZ MODE and observe that the
intensified zones, previously viewed with A INTEN selected,
are displayed on the crt screen.
8. Rotate both the B DELAY TIME POSITION and the A TIME POSITION controls and observe that both displays move continuously across the crt screen.
@
2337 Operators
33
9. Select the A INTEN HORIZ MODE and set the
B TRIGGER SOURCE switch to RUNS AFTER DLY.
Instead of two intensified zones, there is now only one intensified zone. This intensified zone can be continuously positioned on the screen by adjusting only the B DELAY
TIME POSITION control. The LCD readout is disabled.
Using the B Trigger Section
1. Set the B SEC/DIV switch to 1 ms and set the B
TRIGGER SOURCE switch to VERT MODE.
2. Rotate the B DELAY TIME POSITION control throughout its range and observe that the intensified zone appears to jump between positive slopes of the display.
3. Set the B TRIGGER SLOPE switch to - (minus) and observe that the intensified portion begins on the negative slope. Return the B DELAY TIME POSITION control to its fully counterclockwise position.
4. Rotate the B TRIGGER LEVEL control and observe that the intensified zone of the display disappears when this control is out of triggerable range. Readjust the B TRIGGER LEVEL control for a stable display.
5. Select the B HORIZ MODE and rotate the B DELAY TIME POSITION control clockwise. Observe that the length of the display decreases.
6. Operation with the B TRIGGER SOURCE switch set to CH 1, CH 2, or EXT is similar to operation with the B TRIGGER SOURCE switch set to VERT MODE.
Using the X-Y Mode
1. Set both the CH 1 and CH 2 VOLTS/DIV switches
to 1 (IX) and adjust the generator output to provide a
5-division display.
2. Select X-Y VERT MODE by simultaneously pressing in the CH 1 and CH 2 push buttons.
3. Increase the INTEN control setting until two dots are displayed diagonally. This display can then be posi tioned horizontally with the Horizontal POSITION control and vertically with the Channel 2 POSITION control. Note that the dots are separated by 5 horizontal divisions and 5 vertical divisions.
34 2337 Operators
@
4. Set both the CH 1 and CH 2 VOLTS/DIV switches
to 2 (IX). Note that the dots are now separated by 2 1/2
horizontal divisions and 2 1/2 vertical divisions.
5. Select CH 1 VERTICAL MODE and adjust the
INTEN control for normal brightness.
Using the Z-Axis Input
1. Disconnect the dual-input coupler from the CH 2 input connector and connect a bnc female-to-bnc female adapter to the disconnected end of the coupler.
2. Connect a 42-inch, 50-£2 bnc cable from the Z-AXIS INPUT connector (located on the rear panel) to the dual input coupler via the bnc female-to-bnc female adapter.
3. Set the Channel 1 VOLTS/DIV switch to 1 (IX) and adjust the output of the calibration generator to provide a 5-division display.
4. Observe that the positive peaks of the waveform are
blanked, indicating intensity modulation (adjust INTEN
control as necessary).
5. Disconnect the 50-J2 bnc cable from the Z-AXIS
INPUT connector and disconnect the dual-input coupler
from the CH 1 input connector.
Using the Bandwidth Limit Switch
1. Connect a fast-rise + output calibration signal through a 42-inch, 50-£2 cable and a 50-£2 termination to the CH 1 input connector.
2. Set the CH 1 VOLTS/DIV switch to 50 (IX) and
adjust the calibration generator output to provide a 4-
division display.
3. Set the A SEC/DIV switch to 0.5 [is and adjust the
calibration generator fast-rise + output signal frequency to
1 MHz. Adjust the generator amplitude to provide approxi
mately 5 cycles of the displayed signal.
4. Press in the BW LIMIT push button and observe the rounding-off of the front corners of the display. This indicates a decrease in the frequency response of the
vertical amplifier.
@
2337 Operators
35
OPERATOR'S CHECKS AND ADJUSTMENTS
INTRODUCTION TRACE ROTATION
To verify the operation and accuracy of your instru
ment, perform the following check and adjustment pro
cedures before making a measurement. If adjustments are
required beyond the scope of these operator's checks and adjustments, refer the instrument to a qualified service technician for calibration. Before proceeding with these
instructions, verify that the LINE VOLTAGE SELECTOR switch is placed in the proper position and that the correct
line fuse is installed for the ac-power-in put source voltage to be used. Refer to the "Preparation for Use" information
in this manual for procedures relating to ac-power-in put source voltage and fuse selection. Verify that the POWER switch is OFF (push button out), then plug the power cord
into the ac-power-input source outlet. Push in the POWER switch (ON) to apply power to the instrument and allow sufficient time for warm-up before starting these checks and adjustments. Warm-up time required to meet all the
instrument's specification is 20 minutes.
1. Preset instrument controls and obtain a Normal
Sweep Display (refer to "Instrument Familiarization").
2. Use the Channel 1 POSITION control to move the
baseline trace to the center horizontal graticule line.
NOTE
Normally, the resulting trace will be parallel to the
center horizontal graticule line, and the Trace
Rotation adjustment should not be required.
3. If the resulting trace is not parallel to the center horizontal graticule line, use a small-bladed screwdriver to adjust the TRACE ROTATION control (see Figure 3) and align the trace with the center horizontal graticule line.
36
2337 Operators @
PROBE COMPENSATION
Misadjustment of probe compensation is one of the sources of measurement error. Most attenuator probes are equipped with compensation adjustments. To ensure optimum measurement accuracy, always compensate the oscilloscope probe before making measurements. Probe compensation is accomplished as follows:
1. Preset instrument controls and obtain a Normal
Sweep Display (refer to "Instrument Familiarization").
2. Connect the two 10X probes (supplied with the
instrument) to the CH 1 and CH 2 input connectors.
3. Set both VOLTS/DIV switches to 0.5 m and set
both AC-GND-DC switches to DC.
4. Select CH 1 VERTICAL MODE and insert the tip of the probe connected to the Channel 1 input connector to the AMPL CAL output.
5. Using the approximately 1-kHz AMPL CAL square- wave signal as the input, obtain a display of the signal (refer to "Instrument Familiarization").
6. Set the A SEC/DIV switch to display several cycles of the AMPL CAL signal. Use the Channel 1 POSITION control to vertically center the display.
7. Check the waveform presentation for overshoot and rolloff (see Figure 11). If necessary, adjust the probe compensation for flat tops on the waveforms. Refer to the instructions supplied with the probe for details of com pensation adjustment.
8. Select CH 2 VERTICAL MODE and connect the
Channel 2 probe tip to the AMPL CAL output.
9. Use the Channel 2 POSITION control to vertically
center the display and repeat step 7 for the Channel 2
probe.
10. Disconnect the probes from the instrument.
@ 2337 Operators 37
Figure 11. Probe compensation.
2. Set:
Channel 1 AC-GND-DC DC
INTEN Visible display
3. Connect a 10X probe to the Channel 1 input con
nector and connect the probe tip to the AMPL CAL output.
4. Adjust the Channel 1 POSITION control to center
the display on the screen.
5. Adjust the A TRIGGER LEVEL control for a stable
display of the AMPL CAL signal.
6. Slowly adjust the FOCUS control to its optimum setting (best defined display). If the ASTIG adjustment is correctly set, all portions of the trace will come into sharpest focus at the same position of the FOCUS control.
ASTIGMATISM
1. Preset instrument controls and obtain a Normal
Sweep Display (refer to "Instrument Familiarization").
NOTE
The setting of the ASTIG adjustment should be
correct for any display. However, it may be necessary
to reset the FOCUS control slightly when the INTEN control setting is changed.
38 2337 Operators
@
7. If focusing is not uniform over the entire graticule 6. Adjust the Channel 1 POSITION control to vertically area, use a small-bladed screwdriver to adjust the ASTIG center the display, control (see Figure 3).
8. Since the ASTIG and FOCUS adjustments interact, repeat steps 6 and 7 for the best defined display over the entire crt graticule area.
VERTICAL GAIN CHECK
1. Preset instrument controls and obtain a Normal
Sweep Display (refer to "Instrument Familiarization").
2. Set:
AC-GND-DC (both) DC
3. Connect a 10X probe to the Channel 1 input con nector and connect the probe tip to the AMPL CAL output.
4. Adjust the INTEN control for desired brightness and adjust the FOCUS control for best defined display.
5. Adjust the A TRIGGER LEVEL control for a stable
display of the AMPL CAL signal.
7. Check for a vertical display amplitude of 4 divisions
±0.2 division (3.8 to 4.2 divisions). See Figure 12.
8. Repeat steps 3 through 7 using Channel 2.
0
0
r
t '
4 DIV
±0.2 DIV
1
i-
411 5-1 8
@
2337 Operators
39
Figure 12. Vertical display accuracy.
BASIC APPLICATIONS
After becoming familiar with all the capabilities of the 2337 Oscilloscope, the operator can then adopt a con venient method for making a particular measurement. The following information describes the recommended pro
cedures and techniques for making basic measurements with your instrument. When a procedure first calls for presetting instrument controls and obtaining a Normal Sweep Display, refer to the "Instrument Familiarization" section and perform steps 1 through 4 under "Normal Sweep Display."
NONDELAYED MEASUREMENTS
AC Peak-to-Peak Voltage
To perform a peak-to-peak voltage measurement, use the
following procedure:
NO TE
This procedure may also be used to make voltage
measurements between any two points on the waveform.
1. Preset instrument controls and obtain a Normal
Sweep Display.
2. Apply the ac signal to either vertical-channel input connector and set the VERTICAL MODE switch to display the channel used.
3. Set the appropriate VOLTS/DIV switch to display about five divisions of the waveform, ensuring that the VOLTS/DIV VAR control is in the calibrated detent.
4. Adjust the A TRIGGER LEVEL control to obtain a stable display.
5. Set the A SEC/DIV switch to a position that displays several cycles of the waveform.
6. Vertically position the display so that the negative peak of the waveform coincides with one of the horizontal
graticule lines (see Figure 13, Point A).
40
2337 Operators
@
7. Horizontally position the display so that one of the
positive peaks coincides with the center vertical graticule
line (see Figure 13, Point B).
8. Measure the vertical deflection from peak to peak
(see Figure 13, Point A to Point B).
NOTE
If the amplitude measurement is critical or if the
trace is thick (as a result o f hum or noise on the signal), a more accurate value can be obtained by measuring from the top of a peak to the top of a
valley. This w ill eliminate trace thickness from the measurement.
9. Calculate the peak-to-peak voltage, using the follow
ing formula:
vertical VOLTS/DIV probe
Volts (p-p) = deflection x switch x attenuation
(divisions) setting factor
EXAMPLE: The measured peak-to-peak vertical deflec tion is 4.6 divisions (see Figure 13) with a VOLTS/DIV switch setting of 0.5, using a 10X probe.
Substituting the given values:
Volts (p-p) = 4.6 div x 0.5 V/div x 10 = 23 V.
Operators 41
Figure 13. Peak-to-peak waveform voltage.
NOTE
Instantaneous DC Voltage
To measure the dc level at a given point on a waveform,
use the following procedure:
1. Preset instrument controls and obtain a Normal
Sweep Display.
2. Apply the signal to either vertical-channel input connector and set the VERTICAL MODE switch to display the channel used.
3. Verify that the VOLTS/DIV VAR control is in the
calibrated detent and set the AC-GND-DC switch to GND.
4. Vertically position the baseline trace to the center
horizontal graticule line.
5. Set the AC-GND-DC switch to DC. If the waveform moves above the centerline of the crt, the voltage is positive. If the waveform moves below the centerline of the
crt, the voltage is negative.
If using Channel 2, ensure that the CH 2 INVERT
switch is in its noninverting mode (push button out).
6. Set the AC-GND-DC switch to GND and position the baseline trace to a convenient reference line, using the Vertical POSITION control. For example, if the voltage to be measured is positive, position the baseline trace to the bottom graticule line. If a negative voltage is to be
measured, position the baseline trace to the top graticule line. Do not move the Vertical POSITION control after this reference line has been established. The ground refer ence line can be checked at any later time by switching the AC-GND-DC switch to GND.
7. Set the AC-GND-DC switch to DC.
8. If the voltage-level measurement is to be made with
respect to a voltage level other than ground, apply the reference voltage to the unused vertical-channel input connector. Then position its trace to the reference line.
9. Adjust the A TRIGGER LEVEL control to obtain
a stable display.
42
2337 Operators
@
10. Set the A SEC/DIV switch to a position that displays 12. Calculate the instantaneous voltage, using the follow-
several cycles of the signal. ing formula:
11. Measure the divisions of vertical deflection between the reference line and the desired point on the waveform at which the dc level is to be determined (see Figure 14).
Instantaneous
Voltage
vertical , .
deflection x P° ant^
, .. . . , (+ or - )
(divisions)
VOLTS/DIV
x switch x
setting
probe
attenuation
factor
EX A M P LE : The measured vertical deflection from the
reference line is 4.6 divisions (see Figure 14), the wave form is above the reference line, the VOLTS/DIV switch is set to 2, and a 10X attenuator probe is being used.
Substituting the given values:
Instantaneous Voltage = 4.6 div x (+1) x 2 V/div
x 10 = 92 V.
Algebraic Addition
With the VERTICAL MODE switch in the ADD posi
tion, the waveform displayed is the algebraic sum of the
signals applied to the Channel 1 and Channel 2 inputs
@
2337 Operators
43
Figure 14. Instantaneous voltage measurement.
(CH 1 + CH 2). If the CH 2 INVERT push button is pressed in, the waveform displayed is the difference between the
signals applied to the Channel 1 and Channel 2 inputs
(CH 1 CH 2). The total deflection factor in the ADD mode is equal to the deflection factor indicated by either VOLTS/DIV switch (when both VOLTS/DIV switches are
set to the same deflection factor). A common use for the
ADD mode is to provide a dc offset for a signal riding on
top of a high dc level.
The following general precautions should be observed
when using the ADD mode.
a. Do not exceed the input voltage rating of the oscillo
scope.
b. Do not apply signals that exceed the equivalent of
about eight times the VOLTS/DIV switch settings, since large voltages may distort the display. For example, with a VOLTS/DIV switch setting of 0.5, the voltage applied to that channel should not exceed approximately 4 volts.
c. Use Channel 1 and Channel 2 POSITION control
settings which most nearly position the signal on each
channel to midscreen when viewed in either CH 1 or CH 2 VERTICAL MODE. This ensures the greatest dynamic range for ADD mode operation.
d. To attain similar response from each channel, set
both the Channel 1 and Channel 2 AC-GND-DC switches to the same position.
EXAMPLE: Using the graticule center line as 0 V,
the Channel 1 signal is at a 3-division, positive dc level
(see Figure 15A).
1. Multiply 3 divisions by the VOLTS/DIV switch
setting to determine the dc-level value.
2. To the Channel 2 input connector, apply a negative dc level (or positive level, using the CH 2 INVERT switch) whose value was determined in step 1 (see Figure 15B).
3. Select ADD VERTICAL MODE to place the resultant display within the operating range of the vertical POSITION controls (see Figure 15C).
44
2337 Operators
@
/
7
7
!
7 7
1 1 1
POSITIVE LEVE L
1
,i. .
1
1
:
__
__
90
___
NEGA TIVE OFFSET
1
;
___
___
(A) CHANNEL 1 SIGNAL
WITH 3 D IVIS IONS OF POSITIVE DC LEVEL.
(B) CHANNEL 2 DISPLAY
WITH 3 DIVISIO NS OF NEGA TIVE OFFSET.
(C) RESULTANT DISPLAY
46 5 /D M -0 -1
Figure 15. Algebraic addition.
Common-Mode Rejection
The ADD mode can also be used to display signals that contain undesirable frequency components. These undesir able components can be eliminated through common-mode
rejection. The precautions given under the preceding "Algebraic Addition" procedure should be observed.
EXAMPLE: The signal applied to the Channel 1 input
connector contains unwanted ac-input-power-source
frequency components (see Figure 16A). To remove the undesired components, use the following procedure:
1. Preset instrument controls and obtain a Normal Sweep Display.
@
2337 Operators
45
CH 1 SIGNAL WITH UNWANTED LINE FREQUENCY '«
COMPONENT *
CH 2 SIGNAL ,,
FROM LINE
FREQUENCY
SOURCE
(INVERTED)
tot
9C
SIGNAL WITH
LINE FREQUENCY
COMPONENT
CANCELED
OUT
K
.-*r
*
>
y
S
*
* *»
\
\
>
_
_
r
....
(A) C
H 1 AND CH 2 SIGNALS.
....
'1 .
__
__
....
(8) RESULTANT SIGNAL. 173S-19
2. Apply the signal containing the unwanted line- frequency components to the Channel 1 input.
3. Apply a line-frequency signal to the Channel 2 input.
4. Select ALT VERTICAL MODE and press in the CH 2 INVERT push button.
5. Adjust the Channel 2 VOLTS/DIV switch and VAR
control so that the Channel 2 display is approxi mately the same amplitude as the undesired portion of the Channel 1 display (see Figure 16A).
6. Select ADD VERTICAL MODE and slightly readjust
the Channel 2 VOLTS/DIV VAR control for maxi
mum cancellation of the undesired signal component (see Figure 16B).
Time Duration
To measure time between two points on a waveform, use
the following procedure:
1. Preset instrument controls and obtain a Normal
Sweep Display.
2337 Operators
@
46
Figure 16. Common-mode rejection.
2. Apply the signal to either vertical-channel input connector and set the VERTICAL MODE switch to display the channel used.
EXAMPLE: The distance between the time-measurement points is 8 divisions (see Figure 17), and the A SEC/DIV
switch is set to 2 ms. A magnification factor of 1 is used.
3. Adjust the A TRIGGER LEVEL control to obtain a stable display.
4. Set the A SEC/DIV switch to display one complete period of the waveform. Ensure that the TIME (PULL) VAR control is in the calibrated detent.
5. Position the display to place the time-measurement points on the center horizontal graticule line (see Fig ure 17).
6. Measure the horizontal distance between the time-
measurement points.
7. Calculate time duration, using the following formula:
horizontal A SEC/DIV
distance x switch
Time _ (divisions)
_______
setting
Duration magnification factor
Substituting the given values:
Time Duration = 8.3 div x 2 ms/div = 16.6 ms
Figure 17. Time duration.
@
2337 Operators
47
Frequency
The frequency of a recurrent signal can be determined
from its time-duration measurement as follows:
1. Measure the time duration of one waveform cycle using the preceding "Time Duration" measurement procedure.
2. Calculate the reciprocal of the time-duration value to
determine the frequency of the waveform.
EXAMPLE: The signal in Figure 17 has a time duration of 16.6 ms.
Calculating the reciprocal of time duration:
Frequency =
----------
: =
-----
= 60 Hz
time duration 16.6 ms
Rise Time
Rise-time measurements use the same methods as time
duration, except that the measurements are made between
the 10% and 90% points on the leading edge of the wave form (see Figure 18). Fall time is measured between the 90% and 10% points on the trailing edge of the waveform.
1. Preset instrument controls and obtain a Normal
Sweep Display.
2. Apply the signal to either vertical-channel input connector and set the VERTICAL MODE switch to display the channel used.
3. Set the A TRIGGER SLOPE switch to + (plus). Use
a sweep-speed setting that displays several complete cycles
or events (if possible).
4. Set the VOLTS/DIV switch and VAR control for an exact 5-division display.
5. Adjust vertical positioning so that the zero reference of the waveform touches the 0% graticule line and the top
48
2337 Operators
@
of the waveform touches the 100% graticule line (see
Figure 18).
Substituting the given values in the formula:
6. Set the A SEC/DIV switch for a single-waveform display, with the rise time spread horizontally as much as possible.
Rise Time
5 div x 1 yus/div
1
= 5 /is
7. Horizontally position the display so the 10% point on the waveform intersects the second vertical graticule
line (see Figure 18, Point A).
8. Measure the horizontal distance between the 10% and 90% points and calculate the time duration using the following formula:
horizontal SEC/DIV
distance x switch
_. (divisions) setting Rise Time =
-------------
: -
------------
magnification factor
EXAMPLE: The horizontal distance between the 10% and 90% points is 5 divisions (see Figure 18), and the A SEC/DIV switch is set to 1 /ts/division. A magnification factor of 1 is used.
Figure 18. Rise time.
@
2337 Operators
49
Time Difference Between Two Time-Related Pulses
The calibrated sweep speed and dual-trace features of the 2237 allow measurement of the time difference between two separate events. To measure time difference, use the following procedure:
1. Preset instrument controls and obtain a Normal
Sweep Display.
2. Set the A TRIGGER SOURCE switch to CH 1.
3. Set both AC-GND-DC switches to the same position,
depending on the type of coupling desired.
4. Using either probes or cables with equal time delays, connect a known reference signal to the Channel 1 input and the comparison signal to the Channel 2 input.
5. Set both VOLTS/DIV switches for 4- or 5-division displays.
6. Select either ALT or CHOP VERTICAL MODE, depending on the frequency of input signals (or select AUTO VERTICAL MODE if automatic selection is desired).
7. If the two signals are of opposite polarity, press in the CH 2 INVERT push button to invert the Channel 2 display (signals may be of opposite polarity due to 180° phase difference; if so, note this for use later in the final calculation).
8. Adjust the A TRIGGER LEVEL control for a stable display.
9. Set the A TIM E/DIV switch to a sweep speed which provides three or more divisions of horizontal separation between the reference points on the two displays. Center each of the displays vertically (see Figure 19).
10. Measure the horizontal difference between the two signal reference points and calculate the time difference using the following formula:
SEC/DIV horizontal
switch x difference
Time _ setting
______
(divisions)
Difference magnification factor
EXAMPLE: The A SEC/DIV switch is set to 50 /is,
the X10 MAG switch is pressed in and the horizontal
50
2337 Operators
@
difference between waveform measurement points is 4.5 divisions.
Substituting the given values in the formula:
Time _ 50 ps/div x 4.5 div _
Difference 10
Phase Difference
In a similar manner to "Time Difference," phase com parison between two signals of the same frequency can be made using the dual-trace feature of the 2337. This method of phase difference measurement can be used up to the frequency lim it of the vertical system. To make a phase comparison, use the following procedure:
CHA NN EL 1 (REFERENC E) CH ANN EL 2
Figure 19. Time difference between two time-related pulses.
1. Preset instrument controls and obtain a Normal Sweep Display, then set the A TRIGGER SOURCE switch to CH 1.
2. Set both AC-GND-DC switches to the same position,
depending on the type of coupling desired.
3. Using either probes or coaxial cables with equal time delays, connect a known reference signal to the Channel 1 input and the unknown signal to the Channel 2 input.
4. Select either ALT or CHOP VERTICAL MODE, depending on the frequency of the input signals (or select AUTO VERTICAL MODE if automatic selection is desired). The reference signal should precede the comparison signal
in time.
@ 2237 Operators
51
5. If the two signals are of opposite polarity, press in the CH 2 INVERT push button to invert the Channel 2 display.
6
. Set both VOLTS/DIV switches and both VAR
controls so the displays are equal in amplitude.
graticule line (50% of rise time) and calculate the phase difference using the following formula:
Phase
Difference
horizontal
difference
(divisions)
horizontal
graticule
calibration
(deg/div)
7. Adjust the A TRIGGER LEVEL control for a stable display.
EXAMPLE: The horizontal difference is 0.6 division with a graticule calibration of 45° per division as shown in Figure 20.
8
. Set the A SEC/DIV switch to a sweep speed which
displays about one full cycle of the waveforms.
9. Position the displays and adjust the TIME (PULL) VAR control so that one reference-signal cycle occupies exactly 8 horizontal graticule divisions at the 50% rise-time points (see Figure 20). Each division of the graticule now represents 45° of the cycle (360° t 8 divisions), and the horizontal graticule calibration can be stated as 45° per division.
10. Measure the horizontal difference between corre
sponding points on the waveforms at a common horizontal
Substituting the given values into the phase difference
formula:
Phase Difference = 0.6 div x 45° /div = 27°
More accurate phase measurements can be made by using the X I0 MAG function to increase the sweep rate without changing the SEC/DIV TIME (PULL) VAR control setting.
EXAMPLE: If the sweep rate were increased 10 times
with the magnifier (X10 MAG push button in), the magnified horizontal graticule calibration would be 45° per division divided by 10 (or 4.5° per division).
52
2237 Operators
@
CH ANN EL 1 CHA NN EL 2
(REFER EN CE) / (LA GGING )
j
'
(IT)
-
/
/
IT i I
V
1 1 1 1
8 D IVIS IO NS
(36 0°)
ME ASU RE
TIM E FROM
® TO ®
HORIZO NTAL
DIF FER ENCE
411921
CH ANN EL 1 (REFER EN CE )
CHA NNE L 2
MEA SURE
TIM E FR OM
® TO ®
4119-22
Figure 20. Phase difference.
Figure 21. High-resolution phase difference.
Figure 21 shows the same signals illustrated in Figure 20, but magnifying the displays results in a horizontal difference of 6 divisions between the two signals.
Substituting the given values in the phase difference
formula:
Phase Difference = 6 div x 4.5°/div = 27°
Amplitude Comparison
In some applications it may be necessary to establish
a set of deflection factors other than those indicated by
the VOLTS/DIV switch settings. This is useful for com
paring unknown signals to a reference signal of known amplitude. To accomplish this, a reference signal of known
amplitude is first set to an exact number of vertical divisions
by adjusting the VOLTS/DIV switch and VAR control.
53@
2237 Operators
Unknown signals can then be quickly and accurately compared with the reference signal without disturbing the setting of the VOLTS/DIV VAR control. The procedure is as follows.
5. Disconnect the reference signal and apply the unknown signal to be measured to the same channel input. Adjust the VOLTS/DIV switch to a setting that provides sufficient vertical deflection to make an accurate measure ment. Do not readjust the VOLTS/DIV VAR control.
1. Preset instrument controls and obtain a Normal
Sweep Display.
2. Apply the reference signal to either vertical channel
input and set the VERTICAL MODE switch to display the
channel used.
6
. Establish an arbitrary deflection factor, using the
following formula:
Arbitrary vertical VOLTS/DIV
Deflection = conversion x switch
Factor factor setting
3. Set the amplitude of the reference signal to an exact number of vertical divisions by adjusting the VOLTS/DIV 7. Measure the vertical deflection of the unknown signal switch and VOLTS/DIV VAR control. in divisions and calculate its amplitude using the following
formula:
4. Establish a vertical conversion factor, using the following formula (reference signal amplitude must be known):
Unknown
Signal
Amplitude
arbitrary
deflection
factor
vertical
x deflection
(divisions)
Vertical
Conversion
Factor
reference signal amplitude (volts)
vertical VOLTS/DIV deflection x switch (divisions) setting
EXAMPLE: The reference signal amplitude is 30 volts,
with a VOLTS/DIV switch setting of 5 and the VOLTS/
DIV VAR control adjusted to provide a vertical deflec
tion of exactly 4 divisions.
54
2237 Operators
@
Substituting these values in the vertical conversion factor
formula:
Vertical
Conversion
Factor
30 V
4 div x 5 V/div
1.5
Continuing, for the unknown signal the VOLTS/DIV switch setting is 1 and the peak-to-peak amplitude spans five vertical divisions. The arbitrary deflection factor is then determined by substituting values in the formula:
reference signal (e.g., on assembly line test) may be easily and accurately measured with the 2337. To accomplish this, a reference signal of known time duration is first set to an exact number of horizontal divisions by adjusting the A SEC/DIV switch and the TIME (PULL) VAR control. Unknown signals can then be compared with the reference signal without disturbing the setting of the TIME (PULL) VAR control. The procedure is as follows:
1. Set the time duration of the reference signal to an exact number of horizontal divisions by adjusting the A SEC/DIV switch and TIME (PULL) VAR control.
Arbitrary
Deflection = 1.5 x 1 V/div = 1.5 V/div
Factor
2. Establish a horizontal conversion factor, using the following formula (reference signal time duration must be known):
The amplitude of the unknown signal can then be deter mined by substituting values in the unknown signal ampli tude formula:
Amplitude = 1.5 V/div x 5 div = 7.5 V
Horizontal
Conversion
Factor
reference signal time
duration (seconds)
horizontal A SEC/DIV
distance x switch
(divisions) setting
Time Comparison
In a similar manner to "Amplitude Comparison",
repeated time comparisons between unknown signals and a
3. For the unknown signal, adjust the A SEC/DIV switch to a setting that provides sufficient horizontal deflec tion to make an accurate measurement. Do not readjust the TIME (PULL) VAR control.
@
2237 Operators
55
4. Establish an arbitrary deflection factor, using the Substituting the given values in the horizontal conver-
following formual: sion factor formula:
Arbitrary
horizontal ASEC/DIV
Horizontal
2.19 ms
Deflection
= conversion x
switch
Conversion =
Factor
factor
setting
Factor
8
div x
0.2
ms/div
5. Measure the horizontal distance of the unknown signal in divisions and calculate its time duration using the following formula:
Continuing, for the unknown signal the A SEC/DIV switch setting is 50 /is, and one complete cycle spans 7 horizontal divisions. The arbitrary deflection factor is then determined by substituting values in the formula:
Time
Duration
arbitrary
deflection x
factor
horizontal
distance
(divisions)
Arbitrary
Deflection = 1.37 x 50/is/div = 68.5/is/div
Factor
6
. Frequency of the unknown signal can then be deter
mined by calculating the reciprocal of its time duration.
The time duration of the unknown signal can then be computed by substituting values in the formula:
Time _ gg 5 x 7 div = 480 /is/div
Duration
EXAMPLE: The reference signal time duration is
2.19 ms, the A SEC/DIV switch setting is 0.2 ms, and the TIME (PULL) VAR control is adjusted to provide a horizontal distance of exactly 8 divisions.
The frequency of the unknown signal is then calculated:
Frequency = | = 2.083 kHz
H 1 480/is
56
2237 Operators
@
DELAYED-SWEEP MAGNIFICATION
The delayed-sweep feature of the 2337 can be used to provide higher apparent magnification than is provided by the X I0 MAG function. Apparent magnification occurs as
a result of displaying a selected portion of the A trace at a faster sweep speed (B Sweep speed). The A SEC/DIV switch setting determines how often the B trace will be displayed. Since the B Sweep can occur only once for each A Sweep, the A Sweep time duration sets the amount of
time elapse between succeeding B Sweeps.
The intensified zone is an indication of both the location
and length of the B Sweep interval within the A Sweep
interval. Positioning of the intensified zone (i.e., setting
the amount of time between start of the A Sweep and start of the B Sweep) is accomplished with the B DELAY TIME
POSITION control. With either the A INTEN or the B HORIZ MODE selected and with the B TRIGGER
SOURCE switch set to RUNS AFTER DLY, the B DELAY TIME POSITION control provides continuously variable positioning of the start of the B Sweep. The range of this control is sufficient to place the B Sweep interval at any
location within the A Sweep interval. When the A INTEN HORIZ MODE is selected, the B SEC/DIV switch setting
determines the B Sweep speed and concurrently sets the
length of the intensified zone on the A trace.
Using delayed-sweep magnification may produce a
display with some slight horizontal movement (pulse jitter).
Pulse jitter includes not only the inherent uncertainty of
triggering the delayed sweep at exactly the same trigger
point each time, but also jitter that may be present in the input signal. If pulse jitter needs to be measured, use the "Pulse Jitter Time Measurement" procedure which
follows the discussion of "Magnified Sweep Runs After
Delay."
When looking at a signal having pulse jitter, it is often useful to be able to hold that signal stationary on the screen. Operating the B Sweep in a triggered mode will produce a more stable display, since the delayed display is triggered at the same level each time. This capability is
provided by the 2337 and is explained in the "Triggered Magnified Sweep" discussion.
Magnified Sweep Runs After Delay
The following procedure explains how to operate the B Sweep in a nontriggered mode and to determine the resulting apparent magnification factor.
1. Preset instrument controls and obtain a Normal
Sweep Display.
@
2237 Operators
57
2. Set the B TRIGGER SOURCE switch to RUNS
AFTER DLV.
3. Apply the signal to either vertical-channel input
connector and set the VERTICAL MODE switch to display
the channel used.
4. Set the appropriate VOLTS/DIV switch to produce a
display of approximately 5 divisions in amplitude and
center the display.
5. Set the A SEC/DIV switch to a sweep speed which
displays at least one complete waveform cycle.
6
. Select the A INTEN HORIZ MODE and adjust the B DELAY TIME POSITION control to position the intensified zone to the portion of the display to be mag
nified (see Figure 22A).
7. Set the B SEC/DIV switch to a setting which inten
sifies the full portion of the A trace to be magnified. The
intensified zone will be displayed as the B trace. The start
of the intensified zone remains as previously positioned in step 6.
58 2237 Operators
@
Figure 22. Delayed-sweep magnification.
8
. Select the B HORIZ MODE to magnify the inten
sified portion of the A Sweep (see Figure 22B).
2. Referring to Figure 23, measure the difference
between Point A and Point B in divisions and calculate
the pulse jitter time using the following formula:
9. The apparent sweep magnification can be calculated
from the following formula:
Pulse
Jitter
Time
horizontal
difference x (divisions)
B SEC/DIV
switch setting
Apparent
Delayed Sweep
Magnification
A SEC/DIV switch setting
B SEC/DIV switch setting
EXAMPLE: Determine the apparent magnification of a display with an A SEC/DIV switch setting of 0.1 ms and a B SEC/DIV switch setting of 1
ms-
Substituting the given values:
Apparent 1 x 10
4
s
=
-----------------------------
= 102 = 100
Magnification 1x10s
Pulse Jitter Time Measurement
1. Perform steps 1 through 8 of the preceding "Mag
nified Sweep Runs After Delay" procedure.
ME ASUR E
TIM E FROM
® TO ®
4119-24
Figure 23. Pulse jitter.
@
2237 Operators
59
Triggered Magnified Sweep
The following procedure explains how to operate the B Sweep in a triggered mode and to determine the resulting apparent magnification factor.
1. Perform steps 1 through 8 of the preceding "Mag
nified Sweep Runs After Delay" procedure.
2. Set the B TRIGGER SOURCE switch to the same position to which the A TRIGGER SOURCE switch is set. If external triggering is used, connect the required signal to
the B EXT input connector.
NOTE
The B D ELAY TIME POSITION control w ill not
provide continuously variable delay when the B
TRIGGER SOURCE switch is set to VERT MODE,
CH 1, CH 2, or EXT. Adjustment o f the B DELAY
TIME POSITION control will still position the B Sweep with respect to the A Sweep, bu t the B Sweep will occur only when properly triggered. The inten sified zone seen in the A INTEN HORIZ MODE display will move from trigger point to trigger po in t as the B DELAY TIME POSITION control is rotated.
3. Adjust the B TRIGGER LEVEL control so the inten sified zone on the A trace is stable. If an intensified zone cannot be obtained, see step 4.
4. Inability to intensify the desired portion of the trace
indicates that the signal does not meet the triggering requirements. If the condition cannot be remedied either by using the B Sweep triggering controls or by increasing display amplitude, you should do the following in the sequence indicated: lower the VOLTS/DIV switch setting,
set the B TRIGGER SOURCE switch to EXT and trigger the B Sweep externally, and then reset the VOLTS/DIV switch for the desired display amplitude.
5. The apparent magnification factor can be calculated from the same formula shown in step 9 of the "Magnified Sweep Runs After Delay" procedure.
DELAYED-SWEEP TIME
MEASUREMENTS
Operating the 2337 Oscilloscope with HORIZ MODE set to either A INTEN or B will permit time measurements to be made with a greater degree of accuracy than can
be attained with HORIZ MODE set to A. With the B
TRIGGER SOURCE switch and RANGE/FUNCTION
60
2237 Operators
@
switch set to A TIME, two intensified zones are displayed
on the crt. The intensified zones consist of a reference point (moved by rotating the B DELAY TIME POSITION control) and a measurement point (moved by rotating the A TIME POSITION control). The time difference between
the reference point and the measurement point is displayed
on the LCD readout. The following procedures describe how these measurements are accomplished.
Time Duration
1. Preset instrument controls and obtain a Normal
Sweep Display.
2. Apply the signal to either vertical-channel input connector and set the VERTICAL MODE switch to display the channel used.
3. Set the appropriate VOLTS/DIV switch to produce a display of approximately 5 divisions in amplitude and center the display.
4. Select the A INTEN HORIZ MODE and set the A SEC/DIV switch to a sweep speed that displays the entire portion of the waveform for which time duration is to be measured.
5. For the most accurate measurements, set the B SEC/
DIV switch to the fastest sweep speed that provides a
usable (visible) intensified zone.
6. Adjust the B DELAY TIME POSITION control to move the start of the reference point so that it just touches the intersection of the signal and the center horizontal graticule line (see Figure 24A, Point A).
7. Adjust the A TIME POSITION control to move the start of the measurement point along the center horizontal graticule line to the beginning of the next waveform cycle
(see Figure 24A, Point B).
8. Press in the B HORIZ MODE push button and adjust the A TIME POSITION control to superimpose the wave forms as shown in Figure 24B.
9. Read the time duration (or waveform period) on the
LCD readout.
Rise Time
Rise-time measurements use the same methods as time
duration, except that the measurements are made between
@
2237 Operators
61
8
r
\
V
3 S
r
b
P
u r c m m ru a c
OR TIONS OF
WAVEFORM H ___1___
1 J
___
__
(B) M A G NIFIED B TRACE
62
Figure 24. Time duration using delayed sweep.
9 25
the 10% and 90% points on the leading edge of the wave
form. Fall time is measured between the 90% and 10% points on the trailing edge of the waveform.
1. Preset instrument controls and obtain a Normal
Sweep Display.
2. Apply the signal to either vertical-channel input connector and set the VERTICAL MODE switch to display the channel used.
3. Set the appropriate VOLTS/DIV switch and VAR
control for an exact 5-division display.
4. Vertically position the trace so that the zero refer ence of the waveform touches the 0% graticule line and the top of the waveform touches the 100% graticule line (see
Figure 25).
5. Set the A SEC/DIV switch for a single-waveform display, with the rise time spread horizontally as much as possible. Ensure that the A SEC/DIV TIME (PULL) VAR control is in the calibrated detent.
2237 Operators
@
6. Horizontally position the display so the 10% point
on the waveform intersects the third vertical graticule line.
7. Select the A INTEN HORIZ DISPLAY and set the
B SEC/DIV switch to the fastest sweep speed that provides
a usable (visible) intensified zone.
Figure 25. Rise time, differential time method.
8. Adjust the B DELAY TIME POSITION control to move the start of the reference point (left-hand edge) until it just touches the intersection of the signal and the
10% graticule line (see Figure 25, Point A).
9. Adjust the A TIME POSITION control to move the start of the measurement point (left-hand edge) until it just touches the intersection of the signal and the 90% graticule line (see Figure 25, Point B).
10. Read the rise time on the LCD readout.
Time Difference Between Repetitive Pulses
1. Preset instrument controls and obtain a Normal
Sweep Display.
2. Apply the signal to either vertical-channel input connector and set the VERTICAL MODE switch to display the channel used.
3. Set the appropriate VOLTS/DIV switch to produce a display of approximately 5 divisions in amplitude and vertically center the display.
@
2237 Operators
63
4. Set the A SEC/DIV switch to display the measure
ment points of interest within the graticule area.
5. Select the A INTEN HORIZ MODE and set the
B SEC/DIV switch to the fastest sweep speed that provides
a usable (visible) intensified zone.
6. Adjust the B DELAY TIME POSITION control to move the reference point to the first pulse (see Figure 26A, Point A).
7. Adjust the A TIME POSITION control to move the measurement point to the second pulse (see Figure 26A, Point B).
8. Press in the B HORIZ MODE push button and adjust the A TIME POSITION control to superimpose the wave forms as shown in Figure 26B.
'1 4 1
TIME
.
r DI
FFE
3EIVCE-
f- i
1
i
>
\
7
v
\
/
7 "
** 1N 1 tN S I M b U
ZONE
(A) A TRACE
VERTICAL
REFERENCE LINE
X
I
5UP
S)AI
:R I
VID(£
VIPOL
i....
(B) B TRACE
(4 11 5-2 4 ) 4 11 7-26
9. Read the time difference on the LCD readout.
64 2237 Operators
(®
Figure 26. Time difference between repetitive pulses.
Time Difference Between Two Time-Related Pulses
1. Preset instrument controls and obtain a Normal
Sweep Display.
2. Using probes or cables having equal time delays, apply the reference signal to the Channel 1 input and apply the comparison signal to the Channel 2 input.
3. Set both VOLTS/DIV switches to produce a display
of either 4 or 5 divisions in amplitude.
4. Select either ALT or CHOP VERTICAL MODE,
depending on the input frequencies (or select AUTO VERTICAL MODE if automatic selection is desired).
5. Set the A SEC/DIV switch to display the measure
ment points of interest within the graticule area.
6. Select the A INTEN HORIZ MODE and set the
B SEC/DIV switch to the fastest sweep speed that provides
a usable (visible) intensified zone.
7. Adjust the B DELAY TIME POSITION control to
move the reference point to the reference signal (see
Figure 27A, Point A).
8. Adjust the A TIME POSITION control to move the measurement point to the comparison signal (see Figure 27A, Point B).
9. Press in the B HORIZ MODE push button and adjust the B DELAY TIME POSITION and the A TIME POSITION controls to superimpose the reference- and comparison-signal leading edges as shown in Figure 27B.
10. Read the time difference on the LCD readout.
(P>
2237 Operators
65
Figure 27. Time difference between two time-related pulses.
66 2237 Operators
@
DMM DISPLAYS AND
MEASUREMENTS
Before connecting the DMM test leads to a unit-under-
test, select the desired measurement RANGE and
FUNCTION. When the value of the quantity to be
measured is unknown, select the highest range first.
Decrease the range setting until the display reads between 10% and 100% of the full-range value.
Resistance
The DMM may be damaged if it is operating in the resistance mode (SI function selected) and more than 500 V is applied between the + (red) and - (black) inputs.
1. Rotate the RANGE/FUNCTION switch to any
SI range.
The maximum safe input voltage is 500 V (dc + peak ac) between the + (positive) and (negative) inputs or between the + and inputs to ground.
If an overrange condition exists, a "1 " will be displayed
on the left side of the LCD readout, followed by three blank digits and a decimal point. Should this occur, select
the next higher RANGE value.
The 500 V DC and 350 V RMS AC RANGE will not
indicate an overrange condition when the input signal exceeds the maximum allowable input voltage.
2. Connect the + (red) and (black) leads across
the unknown resistance.
3. Select the best SI range to obtain maximum resolu
tion without overranging; read the resistance value on the
LCD readout.
RMS AC
If the LCD readout exceeds 350 V, immediately disconnect the test leads from the unit-under-test to prevent possible instrument damage.
2237 Operators
67
When the peak voltage being measured is greater than three times the RMS AC range selected, UNCAL will be displayed on the LCD readout.
1. Rotate the RANGE/FUNCTION switch to RMS AC
and select 350 V.
2. Connect the (negative) lead to the reference point (usually a ground or test point) and connect the + (positive) lead to the unknown voltage to be measured.
3. Observe the LCD readout. If necessary, select the
next lower value range to obtain maximum resolution without overranging.
DC Volts
If the LCD readout exceeds 500 V, immediately disconnect the test leads from the unit-under-test to prevent possible instrument damage.
1. Rotate the RANGE/FUNCTION switch to DC and
select 500 V.
2. Connect the (negative) lead to the reference point
(usually a ground or test point) and connect the + (positive) lead to the unknown voltage to be measured.
3. Observe the LCD readout. If necessary, select the
next lower value range to obtain maximum resolution without overranging.
68
2337 Operators
OPTIONS
INTRODUCTION
There is presently only one option available for the
2337. A brief description of this option is given in the following discussion. For further information about instrument options, see your Tektronix Catalog or contact
your Tektronix Field Office or representative.
OPTION 03
Option 03 (100-V/200-V Power Transformer) permits operation of the instrument from either a 100-V or a 200-V nominal ac-power-input source at a line frequency from
48 Hz to 440 Hz. This option does not affect the basic
instrument operating instructions presented in this manual.
@
2337 Operators
69
SPECIFICATION
The following electrical characteristics (Table 5) are valid for the 2337 when it has been calibrated at an ambient temperature between +20°C and +30°C, has had a warmup period of at least 20 minutes, and is operating at an ambient temperature between 15°C and +55°C
(unless otherwise noted).
Items listed in the "Performance Requirements" column are verifiable qualitative or quantitative limits, while items listed in the "Supplemental Information" column are either
explanatory notes, calibration setup descriptions, per formance characteristics for which no absolute limits are specified, or characteristics that are impractical to check.
Environmental characteristics of the 2337 are given in Table 6. All environmental tests performed meet the requirements of MlL-T-28800B, Type III, Class 3 equip ment, except where otherwise noted.
Physical characteristics of the instrument are listed in Table 7, and option electrical characteristics are presented in Table 8.
70
2337 Operators
@
Table 5
Electrical Characteristics
Characteristics
Performance Requirements
Supplemental Information
VERTICAL DEFLECTION SYSTEM
Deflection Factor
Range
5 mV per division to 5 V per division in a 1, 2, 5 sequence.
Accu racy
±3% on all ranges when VOLTS/DIV is calibrated at 5 mV per division; add
0.05% per °C deviation from 2C.
Uncalibrated (VAR) Range
Continuously variable between VOLTS/ DIV switch settings. Reduces deflection factor at least 2.5 to 1 on all VOLTS/DIV
switch settings.
Reduces deflection factor to at least 2 mV per division with VOLTS/DIV switch set to 5 mV.
Frequency Response
6-division reference signal from a 25-£2 source; centered vertically, with VOLTS/DIV VAR control in cali brated detent.
15°C to +40°C
Dc to at least 100 MHz.
Reduces to 88 MHz at 2 mV per division.3
Performance Requirement not checked in Service Manual.
@
2337 Operators
71
Table 5 (cont)
Characteristics
Performance Requirements
Supplemental Information
VERTICAL DEFLECTION SYSTEM (cont)
Frequency Response (cont)
+4 C to +55° C
Dc to at least 85 MHz.3
Reduces to 70 MHz at 2 mV per division.3
Ac Coupled Lower 3 dB Point
IX Probe
10 Hz or less.3
10X Probe
1 Hz or less.3
Step Response
5-division reference signal, dc coupled at all deflection factors, from a 25-J2
source; centered vertically with
VOLTS/DIV VAR control in cali brated detent. BW LIMIT push button must be out for full bandwidth
operation.
Performance Requirement not checked in Service Manual.
<S>
72
2337 Operators
Table 5 (cont)
Characteristics
Performance Requirements
Supplemental Information
VERTICAL DEFLECTION SYSTEM (cont)
Step Response (cont)
Rise Time (5 mV per division to 5 V per division)
15° C to +40° C 3.5 ns or less.
Rise time is calculated from the
formula:
, . 350
Rise Time (ns = . ,
BW (MHz)
+40° C to +55° C 4.15 ns or less.3
Aberrations
Positive-Going Step (Excluding ADD Mode)
5 mV per division to 0.2 V per division
+3%, 3%, 3% p-p or less.
Negative-Going Step
Add 2% to all positive-going step specifications; checked at 5 mV per division.
Performance Requirement not checked in Service Manual.
@ 2337 Operators
73
Table 5 (co nt)
Characteristics
Performance Requirements
Supplemental Information
VERTICAL DEFLECTION SYSTEM (cont)
Aberrations (cont)
ADD Mode
Add 4% to all positive-going step specifications; checked at 5 mV per division.
Position Effect
Total aberrations less than +5%, 5%, 5% p-p; checked at 5 mV per division.
Temperature Effect Add 0.15% per °C deviation to aber
rations specifications from 25 C.
Common-Mode Rejection Ratio
At least 10 to 1 at 50 MHz for common mode signals of 6 divisions or less.
VAR control adjusted for best CMRR at 10 mV per division at 50 kHz;
checked at 10 mV per division.
Channel 2 Invert Trace Shift Less than 0.4 division from center screen
when switching from normal to inverted.
74
2337 Operators
@
Table 5 (cont)
Characteristics
Performance Requirements Supplemental Information
VERTICAL DEFLECTION SYSTEM (cont)
Input Gate Current
1C to +30°C
0.5 nA or less.
0.1-division trace shift when moving Input Coupling switch from GND to
AC at 5 mV per division.
+30° C to +55° C
4.0 nA or less.3
0.8-division trace shift when moving Input Coupling switch from GND to
AC at 5 mV per division.
Attenuator Isolation (CH 1 to CH 2)
At least 100 to 1.
With one vertical input set at 0.5 V per division, apply 4-V p-p 25-MHz signal;
set the other vertical input to 10 mV per division. Check for less than 4 divisions of signal.
POSITION Control Range
At least +12 and -12 divisions from graticule center.
Performance Requirement not checked in Service Manual.
@
2337 Operators
75
Table 5 (cont)
Characteristics
Performance Requirements Supplemental Information
VERTICAL DEFLECTION SYSTEM (cont)
Step Attenuator Balance
Less than or equal to 0.2-division trace shift when rotated from 5 mV per division to 5 V per division.
Double for each 10°C deviation
from 2C.
Chop Frequency
275 kHz ±30%.
Input Characteristics
Resistance 1 M ft ±2%.a
Capacitance
20 pF ±10%.a
Maximum Input Voltage
DC Coupled 400 V (dc + peak ac) or
500 V p-p ac at 1 kHz or less.3
AC Coupled
400 V (dc + peak ac) or
500 V p-p ac at 1 kHz or less.3
Performance Requirement not checked in Service Manual.
76
2337 Operators
@
Table 5 (cont)
Characteristics Performance Requirements
Supplemental Information
TRIGGER SYSTEM
Sensitivity
With VOLTS/DIV VAR control in calibrated detent.
In E XTt 10, multiply input require
ments by 10.
A TRIGGER
AC Coupled Signal
0.3 division internal or 50 mV external from 20 Hz to 20 MHz; increasing to
1.1 divisions internal or 150 mV external
at 100 MHz.
LF REJ Coupled Signal
0.3 division internal or 50 mV external from 50 kHz ±10 kHz to 20 MHz;
increasing to 1.1 divisions internal or 150 mV external at 100 MHz.
Attenuates signals below 50 kHz ±10 kHz (-3 dB at 50 kHz).
HF REJ Coupled Signal
0.3 division internal or 50 mV external from 20 Hz ±4 Hz to 50 kHz ±10 kHz.
Attenuates signals below 20 Hz ±4 Hz
and above 50 kHz ±10 kHz (3 dB at 20 Hz and 50 kHz).
DC Coupled Signal
0.3 division internal or 50 mV external from dc to 20 MHz; increasing to
1.1 divisions internal or 150 mV external
at 100 MHz.
@
2337 Operators
77
Table 5 (cont)
Characteristics
Performance Requirements
Supplemental Information
TRIGGER SYSTEM (cont)
Sensitivity (cont)
B TRIGGER (Ac Coupled Signal)
0.3 division internal or 50 mV external from 30 Hz to 20 MHz; increasing to
1.1 divisions internal or 150 mV external
at 100 MHz.
Trigger Jitter
0.2 division or less at 5 ns per division (X10 MAG on) with 100 MHz applied
and at the rated trigger sensitivity.
VOLTS/DIV VAR control must be in calibrated detent.
External Trigger Inputs
Maximum Input Voltage
400 V (dc + peak ac) or 500 V p-p ac at 1 kHz or less.3
Input Resistance
1 MJ2 ±10%.a
Input Capacitance
20 pF ±30%.a
LEVEL Control Range
EXT
At least ± 1 V, 2 V p-p.
Performance Requirement not checked in Service Manual.
78
2337 Operators
@
Table 5 (con t)
Characteristics
Performance Requirements
Supplemental Information
TRIGGER SYSTEM (cont)
LEVEL Control Range (cont)
EXT -r 10
At least ± 10 V, 20 V p-p.a
Trigger View (A Trigger)
Deflection Factor
EXT
100 mV per division ±40%.
EXT t 10
1 V per division ±40%.
Centering of Trigger Point
Within 1 division of center screen.
Bandwidth
To at least 80 MHz. 4-division reference signal from a 25-J2
source; centered vertically.
Delay Difference 3 ns ±2 ns. 5-division signal with 5-ns rise time or
less from 25-S2 source, centered vertically; equal cable length from signal source to vertical channel and external trigger inputs, terminated in 50 at each input.
Performance Requirement not checked in Service Manual.
@
2337 Operators
79
Table 5 (cont)
Characteristics Performance Requirements
Supplemental Information
HORIZONTAL DEFLECTION SYSTEM
Sweep Rate
Calibrated Range
A Sweep
0.5 s per division to 0.05 /is per division in a 1, 2, 5 sequence. X10 MAG extends maximum sweep speed to 5 ns per
division.
B Sweep
50 ms per division to 0.05 jus per division in a 1, 2, 5 sequence. X10 MAG extends maximum sweep speed to 5 ns per division.
Accuracy
+20° C to +30° C
Unmagnified
Magnified
Accuracy specification applies over the full 10 divisions with X10 MAG on and off. Exclude the first and last 40 ns of the sweep on all sweep speeds with X10 MAG on and off.
±2%
±3%
15°C to +55°C
±3%a
±4%a
Performance Requirement not checked in Service Manual.
80
2337 Operators
@
Table 5 (cont)
Characteristics
Performance Requirements
Supplemental Information
HORIZONTAL DEFLECTION SYSTEM (cont)
Linearity
±5%.
Over any 2-division portion of the full
10 divisions, displayed at all sweep speeds. Exclude the first and last dis played divisions of the 5- and 10-ns per division sweep speeds with X10 MAG on.
Variable Range (VAR)
Continuously variable between calibrated settings of the SEC/DIV switches.
Extends maximum A Sweep speed to at least 1.25 s per division.
A Sweep Length
10.5 to 11.5 divisions.
Checked at 1 ms per division.
A Trigger Holdoff (VAR)
At least 2.5 times the minimum holdoff at any sweep speed.3
Magnifier Registration ±0.2 division from graticule center (X10
MAG on to X10 MAG off).
POSITION Control Range
Start of sweep must position to right of graticule center. End of sweep must position to left of graticule center.
Checked at 1 ms per division.
Performance Requirement not checked in Service Manual.
@
2337 Operators
81
Table 5 (cont)
Characteristics
Performance Requirements
Supplemental Information
HORIZONTAL DEFLECTION SYSTEM (cont)
Differential Time Measurement Accuracy
+15°C to +35° C ±1% of reading ±1 count.
Exclude delayed operation when knobs are locked at any sweep speed
or when the A SEC/DIV switch is at
either 0.1 /us per division or 0.05 /us per division. Exclude the first 0.25
division on all A Sweep speeds.
15°C to +55°C
±2.5% of reading ±1 count.8
Delay Time Jitter
±0.005% of 10 times the A SEC/DIV
switch setting (less than one part in
20,000) over the full delay time range.
X-Y OPERATION
Deflection Factor Range
5 mV per division to 5 V per division in a 1, 2, 5 sequence.
No X-axis variable.
Performance Requirement not checked in Service Manual.
82
2337 Operators
@
Table 5 (cont)
Characteristics
Performance Requirements
Supplemental Information
X-Y OPERATION (cont)
Bandwidth
X-Axis
Dc to at least 2 MHz.
Y-Axis
Dc to at least 100 MHz.
Input Characteristics
Resistance
1 M fi ±2%.a
Capacitance
20 pF ±10%.a
Phase Difference Between X- and Y-Axis Amplifiers
< 3° from dc to 200 kHz.
Accuracy
X-Axis
0°C to +4 C
±5% of indicated deflection.
15° C to +55° C
±8% of indicated deflection.3
Performance Requirement not checked in Service Manual.
@
2337 Operators
83
Table 5 (cont)
Characteristics
Performance Requirements
Supplemental Information
CALIBRATOR
Waveshape
Positive-going square wave.
Duty Cycle
50% ±10%.
Output Voltage
0°C to +40° C
0.2 V±1%.
15°C to +55°C 0.2 V±1.5%.a
Repetition Rate
1kHz ±25%.
Output Impedance
200 £2 ±1%.
Performance Requirement not checked in Service Manual.
84
2337 Operators
@
Table 5 (cont)
Characteristics
Performance Requirements
Supplemental Information
Z-AXIS INPUT
Sensitivity 5 V p-p signal referenced to ground causes
noticeable modulation of display at normal intensity.
Positive-going signal decreases intensity; negative-going signal increases intensity.
Usable Frequency Range
Dc to 20 MHz.
Input Resistance
10 kJ2 ±6%.
Input Capacitance
Less than 15 pF.
Maximum Input Voltage
±25 V (dc + peak ac) for dc to 10 MHz.a
For frequencies greater than 10 MHz, use the following formula to calculate the maximum input voltage.3
V (dc + peak ac) = ^ -
f (MHz)
Input Coupling
Dc.
Performance Requirement not checked in Service Manual.
@
2337 Operators
85
Loading...