that in all previously published material. Specifications and price change privileges reserved.
T ektronix, Inc., P.O. Box 500, Beaverton, OR 97077
TEKTRONIX and TEK are registered trademarks of T ektronix, Inc.
WARRANTY
T ektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship
for a period of one (1) year from the date of shipment. If a product proves defective during this warranty period, T ektronix,
at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in
exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the
warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for
packaging and shipping the defective product to the service center designated by T ektronix, with shipping charges prepaid.
T ektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the
T ektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any
other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. T ektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting
from attempts by personnel other than T ektronix representatives to install, repair or service the product; b) to repair
damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction
caused by the use of non-T ektronix supplies; or d) to service a product that has been modified or integrated with other
products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR
IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO
REP AIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO
THE CUSTOMER FOR BREACH OF THIS WARRANTY . TEKTRONIX AND ITS VENDORS WILL NOT BE
LIABLE FOR ANY INDIRECT , SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE
OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH
DAMAGES.
The Tektronix 1503C Metallic-cable Time-Domain Reflectometer (MTDR) is a
cable test instrument that uses radar principles to determine the electrical
characteristics of metallic cables.
The 1503C generates a half-sine wave signal, applies it to the cable under test, and
detects and processes the reflected voltage waveform. These reflections are
displayed in the 1503C liquid crystal display (LCD), where distance measurements
may be made using a cursor technique. Impedance information may be obtained
through interpreting waveform amplitude.
The waveform may be temporarily stored within the 1503C and recalled or may be
printed using the optional dot matrix strip chart recorder, which installs into the
front-panel Option Port.
The 1503C may be operated from an AC power source or an internal lead-acid
battery that supply a minimum of five hours operating time (see the Specifications
appendix for specifics).
Options available for the 1503C are explained in the Options and Accessories
chapter of this manual.
Standards, Documents,
and References Used
Changes and History
Information
1503C MTDR User Manual
Terminology used in this manual is in accordance with industry practice.
Abbreviations are in accordance with ANSI Y1.1–19722, with exceptions and
additions explained in parentheses in the text. Graphic symbology is based on ANSI
Y32.2–1975. Logic symbology is based on ANSI Y32.14–1973 and manufacturer’s
data books or sheets. A copy of ANSI standards may be obtained from the Institute
of Electrical and Electronic Engineers, 345 47th Street, New York, NY 10017.
Changes that involve manual corrections and/or additional data will be incorporated
into the text and that page will show a revision date on the inside bottom edge.
History information is included in any diagrams in gray.
vii
General Information
Installation and Repacking
Unpacking and InItial
Inspection
Power Source and Power
Requirements
Before unpacking the 1503C from its shipping container or carton, inspect for signs
of external damage. If the carton is damaged, notify the carrier. The shipping carton
contains the basic instrument and its standard accessories. Refer to the replaceable
parts list in the Service Manual for a complete listing.
If the contents of the shipping container are incomplete, if there is mechanical
damage or defect, or if the instrument does not meet operational check requirements,
contact your local T ektronix Field Office or representative. If the shipping container
is damaged, notify the carrier as well as Tektronix.
The instrument was inspected both mechanically and electrically before shipment.
It should be free if mechanical damage and meet or exceed all electrical
specifications. Procedures to check operational performance are in the Performance
Checks appendix. These checks should satisfy the requirements for most receiving
or incoming inspections.
The 1503C is intended to be operated from a power source that will not apply more
than 250 volts RMS between the supply conductors or between either supply
conductor and ground. A protective ground connection, by way of the grounding
conductor in the power cord, is essential for safe operation.
The AC power connector is a three-way polarized plug with the ground (earth) lead
connected directly to the instrument frame to provide electrical shock protection. If
the unit is connected to any other power source, the unit frame must be connected
to earth ground.
viii
Power and voltage requirements are printed on the back panel. The 1503C can be
operated from either 115 VAC or 230 VAC nominal line voltage at 45 Hz to 440 Hz,
or a battery pack.
Further information on the 1503C power requirements can be found in the Safety
Summary in this section and in the Operating Instructions chapter.
1503C MTDR User Manual
General Information
Repacking for Shipment
When the 1503C is to be shipped to a T ektronix Service Center for service or repair,
attach a tag showing the name and address of the owner, name of the individual at
your firm who may be contacted, the complete serial number of the instrument, and
a description of the service required. If the original packaging is unfit for use or is
not available, repackage the instrument as follows:
1. Obtain a carton of corrugated cardboard having inside dimensions that are at
least six inches greater than the equipment dimensions to allow for cushioning.
The test strength of the shipping carton should be 275 pounds (102.5 kg). Refer
to the following table for test strength requirements:
SHIPPING CARTON TEST STRENGTH
Gross Weight (lb)Carton Test Strength (lb)
0 – 10200
11 – 30275
31 – 120375
121 – 140500
141 – 160600
2. Install the front cover on the 1503C and surround the instrument with
polyethylene sheeting to protect the finish.
3. Cushion the instrument on all sides with packing material or urethane foam
between the carton and the sides of the instrument.
4. Seal with shipping tape or an industrial stapler.
If you have any questions, contact your local Tektronix Field Office or
representative.
1503C MTDR User Manual
ix
General Information
Contacting Tektronix
Phone1-800-833-9200*
AddressTektronix, Inc.
Department or name (if known)
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
Web sitewww.tektronix.com
Sales support1-800-833-9200, select option 1*
Service support1-800-833-9200, select option 2*
Technical supportEmail: support@tektronix.com
1-800-833-9200, select option 3*
1-503-627-2400
6:00 a.m. – 5:00 p.m. Pacific time
*This phone number is toll free in North America. After office hours, please leave a
voice mail message.
Outside North America, contact a Tektronix sales office or distributor; see the
Tektronix web site for a list of offices.
x
1503C MTDR User Manual
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it. To avoid potential hazards, use this
product only as specified.
Only qualified personnel should perform service procedures.
To Avoid Fire or
Personal Injury
Power Source
Use Proper Power Cord. Use only the power cord specified for this product and
certified for the country of use.
Use Proper V oltage Setting. Before applying power, ensure that the line selector is
in the proper position for the power source being used.
This product is intended to operate from a power source that will not apply more than
250 volts RMS between the supply conductors or between the supply conductor and
ground. A protective ground connection, by way of the grounding conductor in the
power cord, is essential for safe operation.
Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
The standard power cord (161-0288-00) is rated for outdoor use. All other optional
power cords are rated for indoor use only.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
1503C MTDR User Manual
Replace Batteries Properly. Replace batteries only with the proper type and rating
specified.
Recharge Batteries Properly. Recharge batteries for the recommended charge
cycle only.
Use Proper AC Adapter. Use only the AC adapter specified for this product.
Do Not Operate Without Covers. Do not operate this product with covers or panels
removed.
Use Proper Fuse. Use only the fuse type and rating specified for this product.
Avoid Exposed Circuitry. Do not touch exposed connections and components
when power is present.
xi
General Safety Summary
Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
Do Not Operate in an Explosive Atmosphere.
Symbols and Terms
T erms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
T erms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear onthe product:
xii
Battery Recycling
CAUTION
Refer to Manual
WARNING
High Voltage
Double
Insulated
Protective Ground
(Earth) Terminal
This product contains a Nickel Cadmium (NiCd) battery, which must be recycled
or disposed of properly. For the location of a local battery recycler in the U.S. or
Canada, please contact:
The 1503C front panel is protected by a watertight cover, in which the standard
accessories are stored. Secure the front cover by snapping the side latches outward.
If the instrument is inadvertently left on, installing the front cover will turn off the
POWER switch automatically.
The carrying handle rotates 325° and serves as a stand when positioned beneath the
instrument.
The 1503C can be stored in temperatures ranging from –62° C to +85° C if a battery
is not installed. If a battery is installed and the storage temperature is below –35°
C or above +65° C, the battery pack should be removed and stored separately (see
1503C Service Manual for instructions on removing the battery). Battery storage
temperature should be between –35° C to +65° C.
In the field, the 1503C can be powered using the internal battery . For AC operation,
check the rear panel for proper voltage setting. The voltage selector can be seen
through the window of the protective cap. If the setting differs from the voltage
available, it can be easily changed. Simply remove the protective cap and select the
proper voltage using a screwdriver.
1503C MTDR User Manual
REMOVE
CAP TO
SELECT
VOLTAGE
REMOVE
CAP TO
REPLACE
FUSE
Voltage
Selector
Line Fuse
AC Power
Cord Receptacle
Figure 1–1: Rear Panel Voltage Selector, Fuse, AC Receptacle
1–1
Operating Instructions
The 1503C is intended to be operated from a power source that will not apply more
than 250 V RMS between the supply conductors or between either supply conductor
and ground. A protective ground connection by way of the grounding conductor in
the power cord is essential for safe operation.
The AC power connector is a three-way polarized plug with the ground (earth) lead
connected to the instrument frame to provide electrical shock protection. If the unit
is connected to any other power source, the unit frame must be connected to an earth
ground. See Safety and Installation section.
CAUTION. If you change the voltage selector, you must change the line fuse to the
appropriate value as listed near the fuse holder and in the table below.
Care of the Battery Pack
Battery Charging
FUSE RATING
250 VNOMINAL RANGE
0.3 A T115 VAC (90 – 132 VAC)
0.15 A T230 VAC (180 – 250 VAC)
CAUTION. Read these instructions concerning the care of the battery pack. They
contain instructions that reflect on your safety and the performance of the
instrument.
The 1503C can be powered by a rechargeable lead-gel battery pack that is accessible
only by removing the case from the instrument. When AC power is applied, the
battery pack is charged at a rate that is dependent on the battery charge state.
The battery pack will operate the 1503C for a minimum of eight continuous hours
(including making 30 chart recordings) if the LCD backlight is turned off.
The battery pack will charge fully in 16 hours when the instrument is connected, via
the power cord, to an AC power source with the instrument turned off. The
instrument may be turned on and operated while the batteries are charging, but this
will increase the charging time. For longest battery life, a full charge is preferred
over a partial charge.
VOLTAGE RATING
1–2
For maximum capacity , the batteries should be charged within a temperature range
of +20° C to +25° C. However, the batteries can be charged within a temperature
range of 0° C to +40° C and operated in temperatures ranging from –10° C to +55° C.
1503C MTDR User Manual
Operating Instructions
CAUTION. Do not charge battery pack below 0° C or above +40° C. Do not
discharge battery pack below –10° C or above +55° C. If r emoving the battery pack
during or after exposure to these extreme conditions, turn the instrument off and
remove the AC power cord.
The battery pack should be stored within a temperature range of –35° C to +65° C.
However, the self-discharge rate will increase as the temperature increases.
If the instrument is stored with the battery pack installed, the battery pack should
be charged every 90 days. A fully charged battery pack will lose about 12% of its
capacity in three to four months if stored between +20° C and +25° C.
Low Battery
If the battery is low, it will be indicated on the LCD (bat/low). If this is the case,
protective circuitry will shut down the 1503C within minutes. Either switch to AC
power or work very fast. If the instrument is equipped with a chart recorder, using
the recorder will further reduce the battery level, or the added load might shut down
the instrument.
Protection circuits in the charger prevent deep discharge of the batteries during
instrument operation. The circuits automatically shut down the instrument
whenever battery voltage falls below approximately 10 V. If shutdown occurs, the
batteries should be fully recharged before further use.
Low Temperature
Operation
1503C MTDR User Manual
NOTE. Turn the POWER switch off after instrument shutdown to prevent continued
discharge of the batteries.
When the instrument is stored at temperatures below –10° C, voids might develop
in the liquid crystal display (LCD). These voids should disappear if the instrument
is placed in an ambient temperature w +5° C for 24 hours.
1–3
Operating Instructions
When operating the 1503C in an environment below +10° C, a heater will activate.
The element is built into the LCD module and will heat the display to permit normal
operation. Depending on the surrounding temperature, it might take up to 15
minutes to completely warm the crystals in the LCD. Once warmed, the display will
operate normally.
Preparing to Use the 1503C
Check the power requirements, remove the front cover, and you are ready to test
cables. The following pages explain the front-panel controls.
11
12
13
14
15
Tektronix
MENU
VIEW
INPUT
VIEW
STORE
VIEW
DIFF
STORE
INPUT PROTECTED
400 V PEAK MAX
CABLE
134567
ac0.00 ft
O
N
O
F
F
O
F
F
O
F
F
50 Ω
IMPEDANCE
1 avg
NOISE FILTERVERT SCALEDIST/DIV
HORZ
SET REF
VERT
2
1503C
0.00 db
METALLIC TDR
CABLE TESTER
1 ft
2 ns
.3
.4
POSITION
POSITION
PULSE WIDTH
Vp
.5
.6
.03
.7
.02
.8
.01
.9
.04
.05
.06
.07
.08
.09
.00
POWER
(PULL ON)
910
8
Figure 1–3: 1503C Front-Panel Controls
CAUTION. Do not connect to circuits or cables with live voltages gr eater than 400 V
peak. Voltages exceeding 400 V might damage the 1503C front-end circuits.
1–4
1503C MTDR User Manual
Display
Power
TypeCursorWaveform
Operating Instructions
Front-Panel to Cursor
Distance Window
Front-Panel Controls
IMPEDANCE
View Input
Indicator
View Store
Indicator
View Difference
Indicator
Store
Indicator
ac
O
N
O
F
F
O
F
F
O
F
F
50 W2 ns
Selected
Impedance
1 avg
Selected
Noise Filter
0.00 dB5000 ft
SelectedSelectedSelected
Vertical Scale Distance per
Division
0.00 ft
Pulse Width
Figure 1–4: Display and Indicators
1. CABLE: A female BNC connector for attaching a cable to the 1503C for
testing.
2. IMPEDANCE: A four-position rotary switch that selects the output impedance
of the cable test signal. Available settings are 50, 75, 93, and 125 Ohms. The
selected value is displayed above the control on the LCD.
Grid
NOISE FILTER
VERT SCALE
DIST/DIV
1503C MTDR User Manual
3. NOISE FILTER: If the displayed waveform is noisy, the apparent noise can
be reduced by using noise averaging. A veraging settings are between 1 and 128.
The time for averaging is directly proportional to the averaging setting chosen.
A setting of 128 might take the instrument up to 35 seconds to acquire and
display a waveform. The first two positions on the NOISE FILTER control are
used for setting the vertical and horizontal reference points. The selected value
or function is displayed above the control on the LCD.
4. VERT SCALE: This control sets the vertical gain, displayed in dB, or the
vertical sensitivity, displayed in mr per division. Although the instrument
defaults to dB, you may choose the preferred mode from the SetupMenu. The
selected value is displayed above the control on the LCD.
5. DIST/DIV: Determines the number of feet (or meters) per division across the
display . The minimum setting is 1 ft/div (0.25 meters) and the maximum setting
is 5000 ft/div (1000 meters). The selected value is displayed above the control
on the LCD.
1–5
Operating Instructions
A standard instrument defaults to ft/div. A metric instrument (Option 05)
defaults to m/div, but either may be changed temporarily from the menu. The
default can be changed by changing an internal jumper (see 1503C ServiceManual and always refer such changes to qualified service personnel).
.3
.4 .5
POWER
(PULL ON)
Vp
.03
.6
.7
.02
.8
.01
.9.00
PULSE WIDTH
n
POSITION
o
n
o
POSITION
.04 .05
6. Vp: The two Velocity of Propagation controls are set according to the
.06
.07
.08
.09
propagation velocity factor of the cable being tested. For example, solid
polyethylene commonly has a Vp of 0.66. Solid polytetraflourethylene (T eflon
) is approximately 0.70. Air is 0.99. The controls are decaded: the left control
is the first digit and the right control is the second digit. For example, with a Vp
of 0.30, the first knob would be set to .3 and the second knob to .00.
7. POWER: Pull for power ON and push in for power OFF . When the front cover
is installed, this switch is automatically pushed OFF.
8. PULSE WIDTH: This is a five-position rotary switch that selects the pulse
width of the cable test signal. The available settings are: 2, 10, 100, 1000
nanoseconds, and AUTO. The selected value is displayed on the LCD adjacent
to the control. The AUTO setting sets the pulse width according to the distance
registered at the right side of the LCD. The selected value is displayed to the left
of this control on the LCD.
n
9.
POSITION: This is a continuously rotating control that positions the
o
displayed waveform vertically, up or down the LCD.
n
o
10.
POSITION: This is a continuously rotating control that moves a vertical
cursor completely across the LCD graticule. In addition, the waveform is also
moved when the cursor reaches the extreme right or left side of the display. A
readout (seven digits maximum) is displayed in the upper right corner of the
LCD, showing the distance from the front panel BNC to the current cursor
location.
1–6
MENU
VIEW
INPUT
VIEW
STORE
VIEW
DIFF
STORE
11. MENU: This pushbutton provides access to the menus and selects items chosen
from the menus.
12. VIEW INPUT: When pushed momentarily, this button toggles the display of
the waveform acquired at the CABLE connector. This function is useful to stop
displaying a current waveform to avoid confusion when looking at a stored
waveform. This function defaults to ON when the instrument is powered up.
13. VIEW STORE: When pushed momentarily, this button toggles the display of
the stored waveform.
14. VIEW DIFF: When pushed momentarily , this button toggles the display of the
current waveform minus the stored waveform and shows the difference between
them.
15. STORE: When pushed momentarily, the waveform currently displayed will be
stored in the instrument memory . If a waveform is already stored, pushing this
1503C MTDR User Manual
Menu Selections
Operating Instructions
button will erase it. The settings of the stored waveform are available from the
first level menu under View Stored Waveform Settings.
There are several layers of menu, as explained below.
Main Menu
The Main Menu is entered by pushing the MENU button on the front panel.
1. Return to Normal Operations puts the instrument into normal operation
mode.
2. Help with Instrument Controls explains the operation of each control. When
a control or switch is adjusted or pushed, a brief explanation appears on the
LCD.
3. Cable Information has these choices:
a. Help with Cables gives a brief explanation of cable parameters.
b. Velocity of Propagation V alues displays a table of common dielectrics and
their Vp values. These are nominal values. The manufacturer’s listed
specifications should be used whenever possible.
c. Impedance Values displays impedances of common cables. In some cases,
these values have been rounded off. Manufacturer’s specifications should
be checked for precise values.
d. Finding Unknown Vp Values describes a procedure for finding an
unknown Vp.
4. Setup Menu controls the manner in which the instrument obtains and displays
its test results.
1503C MTDR User Manual
a. Acquisition Control Menu has these choices:
i.Max Hold Is: On/Off. Turn Max Hold on by pushing MENU then
STORE. In this mode, waveforms are accumulated on the display . Max
Hold can be deactivated by pushing STORE or the mode exited by
using the Setup Menu.
ii. Pulse Is: On/Off. T urns the pulse generator of f so the 1503C does not
send out pulses.
iii. Single Sweep Is: On/Off. This function is much like a still camera; it
will acquire one waveform and hold it.
b. Vertical Scale Is: dB/mr. This offers you a choice as to how the vertical
gain of the instrument is displayed. You may choose decibels or millirho.
When powered down, the instrument will default to decibels when powered
back up.
1–7
Operating Instructions
c. Distance/Div Is: ft/m. Offers you a choice of how the horizontal scale is
displayed. You may choose from feet per division or meters per division.
When powered up, the instrument will default to feet unless the internal
jumper has been moved to the meters position. Instructions on changing this
default are contained in the 1503C Service Manual.
d. Light Is: On/Off. This control turns the electroluminescent backlight
behind the LCD on or off.
5. Diagnostics Menu lists an extensive selection of diagnostics to test the
operation of the instrument.
NOTE. The Diagnostics Menu is intended for instrument repair and calibration.
Proper instrument setup is important for correct diagnostics results. Refer to the
1503C Service Manual for more information on diagnostics.
a. Service Diagnostics Menu has these choices:
i.Sampling Efficiency Diagnostic displays a continuous efficiency
diagnostic of the sampling circuits.
ii. Noise Diagnostic measures the internal RMS noise levels of the
instrument.
iii. Impedance Diagnostic tests the output impedance circuits in the
instrument.
iv. Offset/Gain Diagnostic reports out-of-tolerance steps in the program-
mable gain stage. This can help a service technician to quickly isolate
the cause of waveform distortion problems.
v. RAM/ROM Diagnostics Menu performs tests on the RAM (Random
Access Memory) and the ROM (Read Only Memory).
vi. Timebase Is: Normal - Auto Correction / Diagnostic - No
Correction. When in Normal - Auto Correction, the instrument
compensates for variations in temperature and voltage. This condition
might not be desirable while calibrating the instrument. While in
Diagnostic - No Correction, the circuits will not correct for these
variations.
b. Front Panel Diagnostics aids in testing the front panel.
c. LCD Diagnostics Menu has these choices:
i.LCD Alignment Diagnostic generates a dot pattern of every other
pixel on the LCD. These pixels can be alternated to test the LCD.
1–8
ii. Response Time Diagnostic generates alternate squares of dark and
light, reversing their order. This tests the response time of the LCD and
1503C MTDR User Manual
Operating Instructions
can give an indication of the effectiveness of the LCD heater in a cold
environment.
iii. LCD Drive Test Diagnostic generates a moving vertical bar pattern
across the LCD.
iv. Contrast Adjust allows you to adjust the contrast of the LCD. It
generates an alternating four-pixel pattern. The nominal contrast is set
internally . When in Contrast Adjust mode, VERT SCALE is used as the
contrast adjustment control. This value ranges from 0 to 255 units and
is used by the processor to evaluate and correct circuit variations caused
by temperature changes in the environment.
d. Chart Diagnostics Menu offers various tests for the optional chart
recorder.
i.LCD Chart allows adjusting the number of dots per segment and the
number of prints (strikes) per segment.
ii. Head Alignment Chart generates a pattern to allow mechanical
alignment of the optional chart recorder.
6. View Stored W aveform Settings displays the instrument settings for the stored
waveform.
7. Option Port Menu contains three items. T wo items allow configuration of the
option port for communicating with devices other than the optional chart
recorder and one item test the option port.
a. Option Port Diagnostic creates a repeating pattern of signals at the option
port to allow service technicians to verify that all signals are present and
working correctly.
b. Set Option Port Timing allows adjustment of the data rate used to
communicate with external devices. The timing rate between bytes can be
set from about 0.05 to 12.8 milliseconds.
c. Option Port Debugging Is Off/On. Off is quiet, On is verbose. This
chooses how detailed the error message reporting will be when communicating with an external device.
It is possible to connect the instrument to a computer through a parallel interface
with a unique software driver. Because different computers vary widely in
processing speed, the instrument must be able to adapt to differing data rates
while communicating with those computers. With user-developed software
drivers, the ability to obtain detailed error messages during the development can
be very useful. For more information, contact your T ektronix Customer Service
representatives. They have information describing the option port hardware and
software protocol and custom development methods available.
1503C MTDR User Manual
1–9
Operating Instructions
The SP-232, a serial interface product, also allows for connection of the 1503C
to other instrumentation, including computers, via the option port. SP-232 is an
RS-232C-compatible interface. For more information, contact your Tektronix
Customer Service Representative. They can provide you with additional details
on the hardware and software protocol.
8. Display Contrast (Software Version 5.02 and above)
a. Press the MENU button firmly once. If the display is very light or very dark,
you might not be able to see a change in the contrast.
b. T urn the VER TICAL SCALE knob slowly clockwise to darken the display
or counterclockwise to lighten the display . If you turn the knob far enough,
the contrast will wrap from the darkest to lightest value.
c. When the screen is clearly readable, press the MENU button again to return
to normal measurement operation. The new contrast value will remain in
effect until the instrument is turned off.
Test Preparations
The Importance of Vp
(Velocity of Propagation)
Vp of Various Dielectric
Types
Vp is the speed of a signal down the cable given as a percentage of the speed of light
in free space. It is sometimes expressed as a whole number (e.g., 66) or a percentage
(e.g., 66%). On the 1503C, it is the percentage expressed as a decimal number (e.g.,
66% = .66). If you do not know the velocity of propagation, you can get a general
idea from the following table, or use the Help with Cables section of the CableInformation menu. You can also find the Vp with the procedure that follows using
a cable sample.
NOTE. If you do not know the Vp of your cable, it will not prevent you fr om finding
a fault in your cable. However, if the Vp is set wrong, the distance readings will be
affected.
All Vp settings should be set for the cable under test, not the supplied jumper cable.
DielectricProbable Vp
Jelly Filled.64
Polyethylene (PIC, PE, or SPE).66
PTFE (Teflon R) or TFE.70
Pulp Insulation.72
Foam or Cellular PE (FPE).78
Semi-solid PE (SSPE).84
Air (helical spacers).98
1–10
1503C MTDR User Manual
n
o
Impedance of Various
Cable Types
Finding an Unknown Vp
Operating Instructions
50 W75 W93 W125 W
RG-4RG-6/URG-7/URG-23/U
RG-8/URG-11/URG-22/URG-63/U
RG-9/URG-12/URG-62/URG-79/U
RG-58/URG-13/URG-71/URG-89/U
RG-62/URG-59/URG-111/UFlat Lead
RG-81RG-124/UTwisted PairTwisted Pair
RG-93RG-140/U
RG-142B/URG-179/U
RG-225/U75 Video
RG-303B/U
RG-316/U
RG-393/U
Vertebrae Helix
1. Obtain a known length of cable of the exact type you wish to test. Attach the
cable to the CABLE connector on the front panel.
2. Pull POWER on.
3. Turn the DIST/DIV to an appropriate setting (e.g., if trying to find the Vp of a
three-foot cable, turn the DIST/DIV to 1 ft/div).
4. Turn the
POSITION control until the distance reading is the same as the
known length of this cable.
5. Turn the Vp controls until the cursor is resting on the rising portion of the
reflected pulse. The Vp controls of the instrument are now set to the Vp of the
cable.
The following three illustrations show settings too low, too high, and correct for a
sample three-foot cable.
ac3.00 ft
O
N
O
F
F
O
F
F
O
F
F
Figure 1–5: Vp Set at .30, Cursor Beyond Reflected Pulse (Setting Too Low)
1503C MTDR User Manual
1–11
Operating Instructions
ac3.00 ft
O
N
O
F
F
O
F
F
O
F
F
Figure 1–6: Vp Set at .99, Cursor Less Than Reflected Pulse (Setting Too High)
ac3.00 ft
O
N
O
F
F
Cable Test Procedure
Distance to the Fault
O
F
F
O
F
F
Figure 1–7: Vp Set at .66, Cursor on Rising Edge of Reflected Pulse (Set Correctly)
2. If you know approximately how long the cable is, set the DIST/DIV
appropriately (e.g., 20-ft cable would occupy four divisions on the LCD if
5 ft/div was used). The entire cable should be displayed.
1–12
1503C MTDR User Manual
Operating Instructions
ac0.00 ft
O
N
O
F
F
O
F
F
O
F
F
Figure 1–8: 20-ft Cable at 5 ft/div
If the cable length is unknown, set DIST/DIV to 5000 ft/div and continue to decrease
the setting until the reflected pulse is visible. Depending on the cable length and the
amount of pulse energy absorbed by the cable, it might be necessary to increase the
VERT SCALE to provide more gain to see the reflected pulse.
The best pulse width is dependent on the cable length. A short pulse can be
completely dissipated in a long cable. Increasing the pulse width will allow the
reflected pulse to be more visible when testing long cables. AUTO will select the
pulse width for you, depending on the distance on the right side of the LCD.
CABLE LENGTH
SUGGESTED PULSESUGGESTED ft/div
0 to 100 ft2 ns10 ft/div
51 to 500 ft10 ns50 ft/div
501 to 5000 ft100 ns500 ft/div
5001 to 50,000 ft1000 ns5000 ft/div
When the entire cable is displayed, you can tell if there is an open or a short.
Essentially, a drop in the pulse is a short and a rise in the pulse is an open. Less
catastrophic faults can be seen as hills and valleys in the waveform. Bends and
kinks, frays, water, and interweaving all have distinctive signatures.
ac0.00 ft
O
N
O
F
F
Short
O
F
F
O
F
F
Figure 1–9: Short in the Cable
1503C MTDR User Manual
1–13
Operating Instructions
n
o
n
o
ac20.00 ft
O
N
O
F
F
O
F
F
O
F
F
Open
Figure 1–10: Open in the Cable
3. To find the distance to the fault or end of the cable, turn the
POSITION
control until the cursor rests on the leading edge of the rising or falling reflected
pulse (see Figure 1–10). Read the distance in the distance window in the upper
right corner of the display.
A more thorough inspection might be required. This example uses a longer cable:
4. When inspecting a 455-foot cable, a setting of 100 ft/div allows a relatively fast
inspection. If needed, turn VERT SCALE to increase the gain. The higher the
gain, the smaller the faults that can be detected. If noise increases, increase the
NOISE FILTER setting.
ac455.00 ft
O
N
O
F
F
O
F
F
O
F
F
Figure 1–11: 455-ft Cable
5. Change DIST/DIV to 20 ft/div . The entire cable can now be inspected in detail
on the LCD. Turn the
POSITION control so the cursor travels to the far right
side of the LCD. Keep turning and the cable will be “dragged” across the
display.
1–14
1503C MTDR User Manual
Loading...
+ 98 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.