067-0539-00 CALIBRATION FIXTURE
Input Normalizer 24 pF IBNCI
REF.
NO.
A
PART NO.
067-0539-00
SERIAL MODEL NO. a
EFF. DISC.
T
Y.
DESCRIPTION
An Input Time Constant Normalizer for signal inputs
having a
nominal 1 megohm x 24 pF RC time constant and
having BNC connectors.
DATA SHEET NO.
062-0836-00
APRIL 1967
Copyright © 1967,
Tektronix, Inc.
All Rights Reserved.

OSCILLOSCOPE INPUT NORMALIZERS
Input Normalizers are used for checking or adjusting the input capacitance of oscilloscopes,
or
oscilloscope plug-in units, where high-impedance
attenuator probes may be used. In a sense they
are dummy probes. Input capacitance should be
adjusted for each setting of the input attenuator
switch, so that you will not need to readjust an
input probe each time the attenuator setting is
changed.
Different oscilloscope types have different nominal
values of input capacitance, so require different
input normalizers. The Instruction Manual for each
Tektronix Oscilloscope identifies which normalizer
should be used, if any, and how to use it.
An instruction manual may refer to an Input Normalizer as a Capacitance Standardizer, a Time
Constant Standardizer, or as an RC Standardizer.
The name was changed to Normalizer to reduce
confusion about whether they should be considered
a measurement Standard. The Tektronix part number was changed at the same time.
Input Normalizers are not intended to be used to
determine whether input capacitance is withincertain limits, or whether the RC product of input
resistance and input capacitance is within certain
limits. Therefore, they are not intended to be
measurement Standards. Furthermore, it is not
advisable to attempt to maintain the input capacitance or the input RC time constant on different
scopes to tight tolerances with the hope of obviating the need to check or adjust a probe when
it is used with a different oscilloscope. Proper
probe adjustment should always be checked each
time a probe is connected to a different oscilloscope or plug-in unit. When this practice is
observed, fewer measurement errors will~ made
than can be prevented by a rigorous proiam to
maintain speclfic tolerance limits. Nor
11 it ad-
visable to rely on a known input capacitanc1, measured at one frequency, to calculate the in~t loading effects on a signal having a different fre!uency.
Most Tektronix Input Normalizers never med to be
adjusted after they leave the factory. Performance
is tested each time one is used, and nirrnall Y
should be considered adequate as long asihe adjustable input capacitors in the oscillosco~rs have
sufficient range.
If readjustment should be necessary, the [11ternal
resistor should first be measured to be sure it
is in tolerance. Then the internal variablecapacitor may be adjusted to produce a square-cirnered
flat-topped response to a square-wave s!plal on
an oscilloscope it is designed to
be usid with,
in the same way as described in the insttuctions
for adjusting scope input capacitance. The risponse
of the scope to square waves should first
to be normal, and the input capacitanci to the
Ii? known
scope should be known to be within about!ive per
cent of its nominal value at the attenuator setting
used. It is normal for input capacitanceto vary
slightly from one setting
to the next to correspond
with slight variations of input resistance. A Tektronix Type
130 L-C Metermaybeusedto J11easure
the scope input capacitance. Measurement should
be made while the scope is operating. Whm using
the Type
130, to avoid a small possibility of a
measurement error due to overdriving tbe input,
it is a good idea to use an attenuator setting thatcorresponds to a deflection factor of aboilt
1 volt
per division.
1
A

MECHANICAL PAR.TS LIST-067-0539-00
i
I
©--1·
FIG. & TEKTRONIX
;~~EX PART NO.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
131-0126-00
210-0962-00
202-0095-00
214-0271-00
213-0075-00
352-0045-00
200-0752-00
213-0088-00
131-0428-00
132-0081-00
166-0217-00
214-0109-01
358-0072-00
134-0044-00
SERIAL/K>DEL
EFF DISC
Q
T
y
1 CONNECTOR, coaxial, BNC, single contact
1
WASHER, bevel, plastic
1 OOX, normalizer
1 BUTTON, plastic
2 SCREW, set, 4-40 x 3/32 inch, HSS
1 HOLDER, insulating cap
1 COVER, normalizer
- mounting hardware: (not included w/cover)
2 SCREW, thread forming, #4 x 1/4 inch, PHS
1 ASSEMBLY, connector, BNC
assembly includes:
1 NUT, BNC
1 TUBE, spacer, insulator, 5/8 inch long
1 PIN, contact, male
1 BUSHING, insulator, 0.323 inch long
1 PLUG, probe, BNC
DESCRIPTION
A
2

ELECTRICAL PARTS LIST-067-0539-00
i-
i
Cl
2-8 pF
I
I
I
I
I 1 Mn I
~
INPUT¥------t1t-,._ _ _, __ ,.._--1._.---~ OUTPUT
Rl I
1/2 W
I
I
I
I
I
~
I I
I
I
I
: ______
C2
~P~
___ ,
I
I
Values are fixed unless marked Variable.
TEKTRONIX
CKT. NO. PART NO.
Cl
C2
Rl
3
281-0060-00
281-0511-00
323-0481-00
SERIAL/MODEL NO.
EFF DISC
Capacitors
Resistor
2-8 pF, Var
22 pF
1 Mn
DESCRIPTION
Cer
Cer
1/2 W
Pree
500
V 10%
1%
A