Tektronix 020-2494-01 Service

020-2494-01 Optical Connectors Cleaning Kit Instructions
075-0769-01
*P075076901*
075076901
Tape
Pull to
advan ce
Cleaner tape dispenser
Connector ti p
Newly exposed
cleaner tape
Optical Conne ctors Cleaning K it
This kit includes a cleaner for cleaning optical connectors on the products listed below and these instructions for using the cleaner.
This kit supports any product for sampling and real--time oscilloscopes that has an optical connector.
Kit Parts List
Part No. Descri ption
Qty.
1ea. NS Kit Instructions
1ea. NS Cleaner tape w/ dispenser
NS - N o t Saleab le
Cleaning Instructions
To prolong the life of the optical connectors on your module and those connected to your module, observe the following handling and cleaning instructions.
CAUT ION. When using, disconnecting, or
connecting optical connectors, always follow the precautions described in the product manuals to prevent damage from electrosta tic discharg e (ESD).
Small dust particles and oils can easily contaminate optical connectors and reduce or block the signal. Take care to preserv e the integrity of your connectors by keeping them free of contamination.
CAUT ION. To prevent loss of optical power or
damage to the optica l connectors, keep the connectors clean at all times. To reduce the need for cleaning, alwa ys keep the connectors covered with protective caps when not in use. Use only high-quality cleaning supplies that are nonabra sive and leave no residue.
Equipment Required
Use thefollowing items to clean the optical connectors:
Cleaner tape dispenser (included with this kit)
Canned compressed gas (Do not use com-
pressed air from an in-house facility, as oil in the system may contaminate the connector.)
Procedure
To clean the optical connectors, refer to the illustrations and follow thesesteps:
1. Advance the tape by g rasping the tape and pulling in a new section to the w iping surface. (Seethefigureabove-left.)
2. Grasp the connector to be cleaned w ith one hand, being careful not to touch the fiber core.
3. Hold the cleaner tape dispenser firmly with yourotherhand.
4. Wipe the connector tip once across the ex posed tape surface, being careful to use a firm, s teady motion. (See f ig ure above-right. )
Keep thetip nearly perpendicular, but slightly tilted up to the tape surface, for optimal results. When the ti p is kept tilted this way, it will move smoothly and easily across the tape without dragging on it or tearing it.
5. If a piece of lint, dust or any light contaminant is left behind, apply a short burst of compressed gas to the c onnector tip.
6. Carefully attach the connector to the instrument or cable, and check for proper operation.
Copyright © Tektronix, Inc. All rights reserved.