Tek-Trol 1700B-C-4-1-T-2-1-8-1 User manual

Page 1
Technology Solutions
TEK-T HERMAL 1700B
Thermal Mass Flow Meter
Instruction Manual
Document Number: IM-1700B
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
Instruction Manual Tek-Thermal 1700B
Table of Contents
1 Safety Instructions ...................................................................................................... 3
1.1 Intended Use ................................................................................................................. 3
1.2 Certification ................................................................................................................... 3
1.3 Safety Instructions from the Manufacturer .................................................................... 3
1.4 Safety Precautions ......................................................................................................... 4
1.5 Packaging, Transportation, and Storage ......................................................................... 5
1.6 Nameplate..................................................................................................................... 6
2 Product Description .................................................................................................... 7
2.1 Introduction .................................................................................................................. 7
2.2 Measuring Principle ....................................................................................................... 7
2.3 Specifications ................................................................................................................ 8
2.4 Dimensional Drawings ................................................................................................... 8
2.5 Model Chart .................................................................................................................. 9
Disclaimer .......................................................................................................................... 3
Product Liability and Warranty ......................................................................................... 3
Information Concerning the Documentation .................................................................... 3
Packaging ........................................................................................................................... 5
Transportation ................................................................................................................... 5
Storage .............................................................................................................................. 5
3 Installation ................................................................................................................ 10
3.1 Selecting the Installation Location ................................................................................ 10
Standard Installation ....................................................................................................... 11
Installation for bend pipeline is at upsteram or downstream ........................................ 11
Installation for bend pipeline that may cause turbulence in the upstream or
downstream ................................................................................................................................. 11
Installation when valves or pressure controller or any other device may cause
turbulence in the upstream or downstream of the flowmeter ................................................... 12
Installation for pipe expander ......................................................................................... 12
Installation for pipe reducer ........................................................................................... 12
3.2 Requirement on Insertion Direction .............................................................................. 13
On a horizontal pipeline, normal air or gas ..................................................................... 13
On a horizontal pipeline, high humidity air or wet natural gas ...................................... 14
On a vertical pipeline, when the density of the gas is higher than air ............................ 14
3.3 Installation Procedure .................................................................................................. 15
Nut Sleeve Insertion ........................................................................................................ 15
4 Electrical Connection ................................................................................................ 18
4.1 Grounding of Measuring Device .................................................................................... 18
4.2 Power and Signal Connection ....................................................................................... 19
Wiring for 3-Wire Pulse Output ...................................................................................... 19
Wiring for 3-Wire HART with 4 to 20 mA ........................................................................ 20
Wiring for 4-Wire HART with 4 to 20 mA ........................................................................ 20
Wiring for RS485 ............................................................................................................. 21
5 Operation ................................................................................................................. 22
5.1 Local Operation and Display ......................................................................................... 22
Multi-functional LCD Display ........................................................................................... 23
Total Flow on Display ...................................................................................................... 24
5.2 Parameter Setting ........................................................................................................ 25
Code Setting .................................................................................................................... 25
01www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
Instruction Manual Tek-Thermal 1700B
5.2.1.1 Code Setting Address .......................................................................................................... 26
5.2.2.1 Digital Setting Address ........................................................................................................ 30
5.3 Bluetooth Access .......................................................................................................... 32
6 Communication ........................................................................................................ 34
6.1 RS485 Modbus ............................................................................................................. 34
6.2 Address of the Displayed Date ...................................................................................... 34
6.3 Address of Code Setting ................................................................................................ 34
7 Maintenance ............................................................................................................. 36
7.1 Sensor Cleaning ............................................................................................................ 36
7.2 Remove transmitter circuit board ................................................................................. 36
8 Troubleshooting ........................................................................................................ 37
8.1 Self-Diagnostic Functions .............................................................................................. 38
Digital Setting .................................................................................................................. 29
Example of Setting .......................................................................................................... 31
Password Setting ............................................................................................................. 31
How to download the application ................................................................................... 32
Access the application ..................................................................................................... 33
Utilize the application ..................................................................................................... 33
02www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
www.tek-trol.com
NOTICE
Read this manual before working with the product. For personal and system safety, and for optimum product performance, make sure you thoroughly understand the contents before installing, using, or maintaining this product. For technical assistance, contact Customer Support 796 Tek-Drive Crystal Lake, IL 60014 USA Tel: +1 847 857 6076, +1 847 655 7428
© COPYRIGHT Tek-Trol LLC 2016
No part of this publication may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or translated into any human or computer language, in any form or by any means, electronic, mechanical, manual, or otherwise, or disclosed to third parties without the express written permission. The information contained in this manual is subject to change without notice.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
Instruction Manual Tek-Thermal 1700B
1 Safety Instructions
1.1 Intended Use
Tek-Thermal 1700B is specially designed for air and N2 applications, such as compressed air, venting air, aeration, process protection N2, and combustion air.
1.2 Certification
General purpose IP67/NEMA 4X
1.3 Safety Instructions from the Manufacturer
Disclaimer
The manufacturer will not be held accountable for any damage that happens by using its product, including, but not limited to direct, indirect, or incidental and consequential damages. Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale. The manufacturer has the right to modify the content of this document, including the disclaimer, at any time for any reason without prior notice, and will not be answerable in any way for the possible consequence of such changes.
Product Liability and Warranty
The operator shall bear authority for the suitability of the device for the specific application. The manufacturer accepts no liability for the consequences of misuse by the operator. Wrong installation or operation of the devices (systems) will cause the warranty to be void. The respective Terms and Conditions of Sale, which forms the basis for the sales contract shall also apply.
Information Concerning the Documentation
To prevent any injury to the operator or damage to the device it is essential to read the information in this document and the applicable national standard safety instructions. This operating manual contain all the information that is required in various stages, such as product identification, incoming acceptance and storage, mounting, connection, operation and commissioning, troubleshooting, maintenance, and disposal.
03www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
Instruction Manual Tek-Thermal 1700B
1.4 Safety Precautions
You must read these instructions carefully prior to installing and commissioning the device. These instructions are an important part of the product and must be kept for future reference. Only by observing these instructions, optimum protection of both personnel and the environment, as well as safe and fault-free operation of the device can be ensured.
For additional information that are not discussed in this manual, contact the manufacturer
Warnings and Symbols Used
The following safety symbol marks are used in this operation manual and on the instrument.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in
th or serious injury
dea
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result i
n
minor or moderate injury. It may also be used to alert against unsafe practices.
NOTE
Indicates that operating the hardware or software in this manner may damage it or lead to system failure.
04www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
Instruction Manual Tek-Thermal 1700B
1.5 Packaging, Transportation, and Storage
Packaging
The original package consists of
1. Tek-Thermal 1700B Thermal Mass Flowmeter
2. Documentation
NOTE
Unpack and check the contents for damages or signs of rough handling. Report damage to the manufacturer immediately. Check the contents against the packing list provided.
Transportation
Avoid impact shocks to the device and prevent it from getting wet during transportation.
Verify local safety regulations, directives, and company procedures with respect to hoisting, rigging, and transportation of heavy equipment.
Transport the product to the installation site using the original manufacturer’s
packing whenever possible.
Storage
If this product is to be stored for a long period of time before installation, take the
following precautions:
Store your product in the manufacturer’s original packing used for shipping.
Storage location should conform to the following requirements:
o Free from rain and water o Free from vibration and impact shock o At room temperature with minimal temperature and humidity
variation
Before storing a used flowmeter remove any fluid from the flowmeter line completely. Properties of the instrument can change when stored outdoors.
05www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
Instruction Manual Tek-Thermal 1700B
1.6 Nameplate
The nameplate lists the order number and other important information, such as design details and technical data
Check the device nameplate to ensure that the device is delivered according to your order. Chec
NOTE
k for the correct supply voltage printed on the nameplate.
06www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 9
Instruction Manual Tek-Thermal 1700B
2 Product Description
This section covers the reference and specification data, as well as ordering information.
2.1 Introduction
Tek-Thermal 1700B Thermal Mass Flowmeters, also known as the thermal dispersion or immersible mass flowmeters, are the precision instruments to measure mass flow of fluid flowing in a closed channel like a pipe or duct. Tek-Thermal 1700B is specially designed for air and nitrogen (N2) applications. It has the more compact design. It has a small enclosure and thin insertion probe tube. It is used in high-pressure applications.
2.2 Measuring Principle
The Tek-Thermal 1700B measures the gas mass flow based on Thermal Diffusion theory. It has two RTD (Resistance temperature Detector) sensors. One sensor measures the velocity of gas flow (Rh) and the other sensor detects the temperature shift of gas flow (Rmg). When there are two RTDs in the gas flow, Rh will be heated and Rmg will sense the temperature difference. As the velocity of gas flow increases, more heat will dissipate and so the temperature read by Rh will decrease.
Tek-Thermal 1700B Thermal Mass Flowmeter design is based on constant power measuring method, thus the RTD is heated on a consistent power and will be more durable and stable. Due to that, Tek-Thermal 1700B has less problem of zero-off which may cause a function failure of RTD due to overheating the term.
07www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
Instruction Manual
Media capability
Air, Nitrogen (N2)
Pipe diameter
1" to 20"
Flow velocity range
Tek-Thermal 1700B
2.3 Specifications
Accuracy ±0.5% of Reading Maximum Pressure 232 PSI Temperature of medium Power supply 13.5 to 42 VDC or 85 to 265 VAC Response time 1 second Output 4 to 20mA and Pulse standard Communication Modbus RS485 or HART optional Display Indication Mass flow, Volume flow in normal condition, Total flow,
Ingress Protection IP65
2 to 200 ft/sec or 1 to 100 ft/sec or 3 to 300 ft/sec
-40°F to 302°F (-40°C to 150°C)
Temperature of medium, Velocity
2.4 Dimensional Drawings
(Ø0.03)
0.47"
9.64"
(0.80 ft)
G 1/2
16.22"
(1.35 ft)
6548.1
(0.23 ft)
2.83"
(0.38 ft)
(0.22 ft)
2.75"
(0.41 ft)
4.96"
4.60"
08www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 11
Instruction Manual Tek-Thermal 1700B
2.5 Model Chart
Example
Series
Process Connection
Insertion Probe
Range
Transmitter
Material
Pressure
Output
Power Supply
Diagnostics
Tek-Thermal 1700B
Tek-Thermal 1700B
C
C
T 2 1 8 1 B
4 1
4
1
2
3
T
R
Tek-Thermal 1700B-C-4-1-T-2-1-8-1-B
Thermal Mass Flowmeter
Insertion (NPT)
10" Probe Length, 0.5" Diameter, 1" to 20" Pipe Diameter
2 to 200 ft/sec
1 to 100 ft/sec
3 to 300 ft/sec
Direct Mount
Remote Mount (comes with 16.40 ft of cable)
2
1
7
8
316 SS
232 PSI Max Pressure
Pulse, 4-20 mA, HART
Pulse, 4-20 mA, RS485
1
13.5-42 VDC
85-265 VAC
2
B Bluetooth
09www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 12
Instruction Manual Tek-Thermal 1700B
3 Installation
This section covers instructions on installation and commissioning. Installation of the device must be carried out by trained, qualified specialists authorized to perform such works.
CAUTION
When re
moving the instrument from hazardous processes, avoid direct
contact with the fluid and the meter
All installation must comply with local installation requirements and local electrical code
3.1 Selecting the Installation Location
The flowmeter is designed to withstand the severe environmental conditions. However, to ensure stable and accurate operation, observe the following precautions when selecting an installation location.
Ambient Temperature
Avoid locations subject to wide temperature variations or a significant temperature gradient. If the location is exposed to radiant heat from plant
equipment’s, provide adequate thermal insulation and/or ventilation.
Ambient Atmosphere
Avoid installing the transmitter in a corrosive atmosphere. If the transmitter must be installed in a corrosive atmosphere, there must be adequate ventilation as well as measures to prevent intrusion or stagnation of rain water in conduits.
Shock and Vibration
Select an installation site with minimum shock and vibration (although the transmitter is designed to be relatively resistant to shock and vibration).
10www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 13
Instruction Manual Tek-Thermal 1700B
Standard Installation
Installation for bend pipeline is at upsteram or downstream
Installation for bend pipeline that may cause turbulence in the upstream or downstream
11www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 14
Instruction Manual Tek-Thermal 1700B
Installation when valves or pressure controller or any other device may cause turbulence in the upstream or downstream of the flowmeter
Installation for pipe expander
Installation for pipe reducer
12www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 15
Instruction Manual Tek-Thermal 1700B
3.2 Requirement on Insertion Direction
On a horizontal pipeline, normal air or gas
13www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 16
Instruction Manual Tek-Thermal 1700B
On a horizontal pipeline, high humidity air or wet natural gas
On a vertical pipeline, when the density of the gas is higher than air
14www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 17
Instruction Manual Tek-Thermal 1700B
3.3 Installation Procedure
Nut Sleeve Insertion
1. Use cutting torch or sharp cutting tool to tap into the pipe. The pipe opening must be at least 0.5 inch (0.04 ft) in diameter (Do not insert sensor probe through a smaller hole)
2. Remove the burrs from the tap and sharp edges may causes flow profile disturbance that can affect the accuracy of flowmeter
3. Weld the MNPT ½” socket (Part No.1) on the open hole. The correct insertion depth places the centreline of the sensor access hole in probe at the pipe line’s centreline
4. Bolt the ½” ball (isolation) valve (Part No.2) with FNPT threads on both ends of the socket. Seal the thread connection with thread sealant. The lever on the ball valve should be up when the valve is open.
15www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 18
Instruction Manual Tek-Thermal 1700B
5. Insert the sensor probe through the ball valve (isolation valve) into the pipe. The
correct insertion depth places the sensor at the pipe’s centreline.
6. Calculate the insertion depth of the sensor probe,
Where,
A=Inner diameter of the pipeline
B=Thickness of the pipe line C=The distance between the top of the pipeline and the upper end of the nut when the nut is fixed
7. connect the sleeve (Part No.3) on the meter and the ball valve (isolation valve), seal the thread connection part with thread sealant. Tighten the nut (Part No.4) with wrench.
 =
+  + 
2
16www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 19
Instruction Manual Tek-Thermal 1700B
8. Adjust the direction of the flowmeter, make sure the direction mark on the probe (Part No.5) is pointing to the direction as the flow goes.
NOTE
If flow rate higher than 66 lb.ft/sec (90Nm/s) or pipe size larger than 8”(DN400),
ll require 19mm diameter probe, mounted in 1” ball valve and 1” socket and
wi
22mm hole
17www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 20
Instruction Manual Tek-Thermal 1700B
4 Electrical Connection
This section covers the all electrical connection requirement. Electrical connection of the device must be carried out by trained, qualified specialists authorized to perform such work by the installation site.
Connect all electrical cables when the power is switched off. If the device does
The device must be grounded to a spot in accordance with regulations in
WARNING
not have switch-off elements, then, overcurrent protection devices, lightning protection and/or energy isolating devices must be provided by the customer.
order to protect personnel against electric shocks.
NOTE
When using the measuring device in hazardous areas, installation must comply with the corresponding national standards and regulations and the Safety Instructions or Installation or Control Drawings.
4.1 Grounding of Measuring Device
Tek-Thermal 1700B Thermal Mass Flowmeter the power supply of signal processing circuit is transferred from outside power supply by an isolation type DC-DC transmitter with advanced grounding technology. The field frequency interference can be isolated. The “V” terminal of power supply should not be connected to the ground. When Tek-Thermal 1700B is used in the strong interference environment, the shell should be clean and connect to the earth through the cable, so that interference can be eliminated.
18www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 21
Instruction Manual
V +
Power Supply +
V -
Power Supply -
Pulse Output
A+
RS485 Communication +
B-
RS 485 Communication -
I+
Current +
I-
Current -
P+
Pressure Transmitter +
P-
Pressure Transmitter -
Tek-Thermal 1700B
4.2 Power and Signal Connection
It has AC and DC power options. Supply power must be 13.5 to 42 VDC and 85 to 265 VAC.
Wiring for 3-Wire Pulse Output
3-wire pulse output requires a power source of 13.5 to 42 VDC. Tek-Thermal 1700B use a current pulse output with 50% duty ratio. Add resistor between and “V-”, the resistance should be within 500Ω to 1000Ω, and power consumption should be no less than 0.5W
19www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 22
Instruction Manual Tek-Thermal 1700B
Wiring for 3-Wire HART with 4 to 20 mA
Wiring for 4-Wire HART with 4 to 20 mA
20www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 23
Instruction Manual Tek-Thermal 1700B
Wiring for RS485
21www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 24
Instruction Manual Tek-Thermal 1700B
5 Operation
This section covers operation techniques and guidelines.
5.1 Local Operation and Display
Tek-Thermal 1700B provide local display and setting, can display several variables on the local multifunctional LCD display. It has 3 buttons used for configuration. The local display enables you to read all important parameters directly at the measuring point and configure the device using the function matrix. The LCD display indicates output. Tek-Thermal 1700B has three buttons are placed on the top of the display, which are
used as L-R button, used as U-D button, used as Enter button.
Pressing the L-R button will move the menu selection to the left/right, or the data entry field to the left/right if updating a menu value
Pressing the U-D button will move the menu selection up/down, or increment the data entry field if you are updating a menu value
Pressing enter button to save the current value
All the “Digital setting” and “Code Setting” of the flowmeter is made through these 3 buttons.
22www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 25
Instruction Manual Tek-Thermal 1700B
Multi-functional LCD Display
Tek-Thermal 1700B Thermal Mass Flowmeter has a display to indicate temperature, flow rate, total flow and more parameters.
The LCD display has 2 areas to display the content, the upper row, the lower row. The upper row displays the flow rate, mass flow, and standard flow rate. Below the upper row displays the unit of the variable displayed in the upper row. The lower row displays other variables, such as frequency, pressure, temperature, density and total flow. And below the lower row displays the unit of the variable displayed in the lower row.
23www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 26
Instruction Manual Tek-Thermal 1700B
Tek-Thermal 1700B Multi-Variable version can also display temperature. Users can switch the parameter displayed on LCD by using the buttons and the next parameter will be displayed for 30 seconds. Users can also fix the lower row to display a parameter consistently. You can also set the lower row to display several variables in circular turn.
Total Flow on Display
The lower row of the flowmeter displays total flow. Tek-Thermal 1700B can display 9 digits left to decimal point and 3 digits right to the decimal point. When more than six digit displays on LCD, the total flow reading will be displays two times. First displays the right digits and then displays the left digits. The left digits display with a
mark “x1000”. By using L-R button you can switch the left and the right digits on LCD.
To check the right digits now, press the “L-R button”, the display will be
According to picture, the total flow is 569864.581 kg
24www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 27
Instruction Manual Tek-Thermal 1700B
5.2 Parameter Setting
The Tek-Thermal 1700B has three modes as below
Working Mode
Setting Mode
Calibration Mode
Code Setting
In working mode, press the Enter button and U-D button at the same time to entre in the code setting
In the code setting first row displays the reference number and lower row displays the content of this parameter
Use the “L-R” button to move to the next digit and use “U-D” to cycle through the number. Once the correct value has been entered, hit the “Enter” button to
save current value.
The C01=02, that is set the pre-heating time to 2 seconds.
Press the “Enter button and U-D button at the same time to exit the code
setting menu
25www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 28
Instruction Manual
Code setting address
Item
Code
Description of code
C01
Start-up time
01 to 99
Set the start-up time within 1
C02
Density Type
00
Display standard volume flow rate
01
Density pre-set, display mass flow
02
Display actual flow rate
C03
Pulse Output Type
00
pulse output off
01
Frequency output
02
Pulse equivalent output
C06
Pulse output parameter
00
Flow rate
01
Temperature
02
Pressure
03
Total Flow
C07
Damping
01 to 16
1 to 16 seconds
C08
Instrument number
00 to 99
For Modbus
00 to 15
For HART communication
C09
Baud rate
01
1200 no parity 1 stop bit
02
1200 even parity 1 stop bit
03
2400 no parity 1 stop bit
04
2400 even parity 1 stop bit
05
4800 no parity 1 stop bit
06
4800 even parity 1 stop bit
07
9600 no parity 1 stop bit
08
9600 even parity 1 stop bit
09
19200 no parity 1 stop bit
10
19200 even parity 1 stop bit
11
1200 odd parity 1 stop bit
12
2400 odd parity 1 stop bit
13
4800 odd parity 1 stop bit
14
9600 odd parity 1 stop
15
19200 odd parity 1 stop bit
16
38400 no parity 1 stop bit
17
38400 even parity 1 stop bit
18
38400 odd parity 1 stop bit
19
57600 no parity 1 stop bit
20
57600 even parity 1 stop bit
21
57600 odd parity 1 stop bit
22
115200 no parity 1 stop bit
23
115200 even parity 1 stop bit
24
115200 odd parity 1 stop bit
Tek-Thermal 1700B
5.2.1.1 Code Setting Address
to 99 seconds Liquid
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
26www.tek-trol.com
Page 29
Instruction Manual
C10
Time unit for flow rate
00
/s
01
/min
02
/h
C11
Mass unit
01
Kg
02
Ton
03
Lb
C12
Volume unit for flow rate
01
m3
02 L 03
ft3
04
US gal
05
UK gal
C13
Pressure unit
00
kPa (Absolute Pressure)
01
kPa (Gauge pressure, minus sign for negative pressure)
02
KPa (Gauge pressure)
03
MPa (Absolute pressure)
04
MPa (Gauge pressure, minus sign for negative pressure)
05
MPa (Gauge pressure)
06
PSI (Absolute pressure)
07
PSI (Gauge pressure, minus sign for negative pressure)
08
PSI (Gauge pressure)
09
Bar (Absolute pressure)
10
Bar (Gauge pressure, minus sign for negative pressure)
11
Bar (Gauge pressure)
C14
Temperature unit
01 ℃ 02 ℉ 03
K
C15
Right digits number for
00 to 05
00:No right digits for total flow
01 to 05:1 to 5 right digits for total flow
C25
Password for setting 00
Password protection off
01
Password protection on
C26
Fixed max flow
00
Fixed max flow off
011
Fixed max flow on. (If the flow
will show the reading as D039)
C31
Time space for circle
0 to 30
00:circle display off
1 to 30:1 to 30 seconds
different parameter
C32
First parameter
00 to 05
00:circle display off
01 to 05:see chart 4.1 C33
Second parameter
00 to 05
00:circle display off
Tek-Thermal 1700B
total flow
rate measured is over the setting of D039, the display
display
between the display of
displayed in circle
Find Quality Products Online at: sales@GlobalTestSupply.com
display
27www.tek-trol.com
www.GlobalTestSupply.com
Page 30
Instruction Manual
displayed in circle display
01 to 05:see chart 4.1
C36
Last parameter displayed
00 to 05
Same as above
C40
The date sequence of IEEE
01
Lower 16 bits address in front,
address in front
02
Higher 16 bits address in front,
address in front
03
Lower 16 bits address in front,
address in front
04
Higher 16 bits address in front,
address in front
C41
Shape of pipeline
00
Round pipeline (D: D010)
01
Rectangle pipeline (L:D018 W:D019)
C47
Set password 00
Keep the password
01
Change the pass word
C49
Status
00
Working status
01
4 to 20mA output calibration statues
02
Flow rate calibration statues
03
Temperature calibration statues
04
Pressure calibration statues
C50
Total flow reset 00
Reset total flow to 0
01
Default
C60
Restore to backup date
06
Restore to backup date
C61
Save setting backup
16
Save current setting for backup
C80
Version No. of hardware
Read Only
C81
Version No. of software
Read Only
C82
Calibration date
Read Only
C83
Number of errors
Read Only
C84
Version No. of HART
Read Only
C86
Communication interface 00
Modbus Read Only
01
HART Read Only
C87
Pressure sensor type
00
No pressure sensor Read Only
01
Piezo Read Only
02
Piezo Read Only
03
4 to 20mA Read Only
04
0 to 10mA Read Only
C88
Product ID No.
Read Only
Tek-Thermal 1700B
in circle display
format float of Modbus
the lower 8 bits in the 16 bits
the Higher 8 bits in the 16 bits
the higher 8 bits in the 16 bits
the lower 8 bits in the 16 bits
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
28www.tek-trol.com
Page 31
Instruction Manual Tek-Thermal 1700B
Digital Setting
To enter in to the digital setting, press “Enter” and “L-R” button at the same time.
The first row displays the reference number of digital setting and lower row display the content of this parameter
Use the “L-R” button to move to the next digit and use “U-D” to cycle through the number. Once the correct value has been entered, hit the “Enter” button to
save the current value
In below picture, D001=1.60000, Max flow is 1.6 (unit according to another setting)
Press the “Enter” button and “U-D” button at the same time to exit the code setting menu
29www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 32
Instruction Manual Tek-Thermal 1700B
5.2.2.1 Digital Setting Address
Digital setting address D001 Max flow rate
D002 Min flow rate
D003 Max frequency
D004 Min frequency
D005 Cut off small
D008 K factor [0 to 999999] D009 Density [0 to 999999] Unit is Kg/m³ D010 Pipe size [0 to 3281 Unit is feet D011 Max
D012 Min temperature
D013 Max pressure
14 Min pressure
D015 Ambient
D017 Equivalent of
Item Code Description of code
[-99999999999] [-99999999999]
0.5 to 10000 The frequency output of
output
0.5 to 10000 The frequency output of
output
[-99999999999]
signal
-99999 to 999999 The max frequency when temperature (Unit as set in C14
-99999 to 999999 The min frequency when (Unit as set in C14
-99999 to 999999 The max frequency when (Unit as set in C13
-99999 to 999999 The min frequency when (Unit as set in C13
0 to 999999 Unit according to setting
pressure
0 to 999999 Set the equivalent that
pulse output
Unit is same as flow rate, Max/min flow rate of 4 to 20mA and 200 to 1000Hz output
max flow, unit is Hz
min flow, unit is Hz Unit is same as flow rate
temperature is output by frequency, 1000Hz for instance. The max current when temperature is output by current, 20mA for instance
temperature is output by frequency, 200Hz for instance. The min current when temperature is output by current, 4mA for instance
temperature is output by frequency, 1000Hz for instance. The max current when temperature is output by current, 20mA for instance.
temperature is output by frequency, 200Hz for instance. The min current when temperature is output by current, 4mA for instance.
one pulse output stands for Unit as C06
30www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 33
Instruction Manual
D018
Length of
pipeline
D019
Width of
pipeline
D030
Temperature of
condition
0 to 20
Default temperature of
D033
Reference flow
0 to 999999
When measured flow rate
the reference flow rate
Code setting
Address
Code
Description
C02
01
Density pre-set
C03
01
Frequency output
C06
00
Output parameter is flow rate
Digital setting
D009
1.2930
Density=1.293, air
D008
1.000
Factor=1
D001
5000
The flow rate for 1000Hz output
D002
0
The flow rate for 200Hz output
D010
150
Pipe size=150mm
D003
1000
Max output frequency
D004
200
Min output frequency
Tek-Thermal 1700B
0 to 3281 Unit is feet
rectangle
0 to 3281 Unit is feet
rectangle
standard
rate
standard condition is 20 °C
is over calibrated max flow, will display/output
Example of Setting
For insertion connection, measure air in 6” (DN0.5) pipe, density pre-set, unit displayed in kg/hr, flow range of 0 to 4000kg/hr, 200 to 1000Hz output
Password Setting
In code setting, set C47 to 01.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
31www.tek-trol.com
Page 34
Instruction Manual Tek-Thermal 1700B
To set the new password, use “U-D” button to cycle through numbers and “L-R” button to move to the next digit. The user must have entered the password twice. If power is off once the password has been entered, hit “Enter” button to save the current password value.
5.3 Bluetooth Access
How to download the application
Visit Apple’s application store
Search “Tek-trol Bluetooth” or “Tek-Bluetooth” to find our app profile
Download the application
32www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 35
Instruction Manual Tek-Thermal 1700B
Access the application
Open application
Enter login information (ID: 20000000007, Verify code:341234)
This will take you to “device list” where your phone will automatically sync with your
product (you must be close enough)
Utilize the application
Follow prompts and menu to navigate through the many features of our bluetooth
application!
33www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 36
Instruction Manual
Register address
Usage
Nature
Date type
0 to 1
Flow rate
Read only
Float
2 to 3
Frequency
Read only
Float
4 to 5
Pressure
Read only
Float
6 to 7
Total Flow
Read only
Float
Tek-Thermal 1700B
6 Communication
This section covers guidelines for setting up RS485 Modbus communication
6.1 RS485 Modbus
The communication interface should be RS485, the range of Baud rate should be 1200
to 115200. The wiring terminal is “A” and “B”. The communication should comply with
MODBUS-RTU. The combination of a communication signal: Address code - function code – date segment – CRC calibration code. The distance between two characters should not be longer than one character, or it will be considered as the beginning of a new message or the end of an old message. The message is combined with hexadecimal arrays.
6.2 Address of the Displayed Date
The displayable dates including flow rate, pressure, temperature and total flow, if the meter is not multi-variable version, then the reading of pressure and temperature will both be 0. The date of the parameters in above chart can be read by using function code 03 according to the address above and shifting.
6.3 Address of Code Setting
Register Usage Range Nature Date type 1000 Pre-heating time
(warm up) C01
1001 Density
compensation C02
1002 Method of pulse
output
1004 Output C05 0 to 2 Read/Write Short
1005 Pulse output
parameter 1006 Damping C07 1 to 16 Read/Write Short 1007 Instrument
number C08 1008 Baud rate C09 1-24 Read Short 1009 Unit of time C10 0-2 Read/Write Short 1010 Mass unit C11 0-2 Read/Write Short 1011 Volume unit C12 0-6 Read/Write Short 1012 Pressure unit
C13 1013 Temperature 0-2 Read/Write Short
1 to 99 Read only Short
0 to 2 Read/Write Short
0 to 2 Read/Write Short
0 to 3 Read/Write Short
Hart (0 to 15) MB (1 to 99)
0-11 Read/Write Short
Read Short
34www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 37
Instruction Manual
unit C14
1014
Right digits
flow C15
0-5
Read/Write
Short 1030
Time of circle display C31
0-30
Read/Write
Short
1031
First parameter
circle display C32
1-5
Read/Write
Short
1032
Second
circle display C33
0-5
Read/Write
Short 1033
Third parameter
circle display C34
0-5
Read/Write
Short
1034
Fourth
circle display C35
0-5
Read/Write
Short 1035
fifth parameter
circle display C36
0-5
Read/Write
Short 1039
Sequence of float C40
1-4
Read/Write
Short
1040
Shape of pipeline C41
0-1
Read/Write
Short
1042
The array length
C43
5-50
Read/Write
Short
1048
Status C49
0-10
Read/Write
Short
1049
Total flow reset to 0 C50
0-1
Read/Write
Short
Tek-Thermal 1700B
number for total
displayed in
parameter
displayed in
displayed in
parameter
displayed in
displayed in
of moist control
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
35www.tek-trol.com
Page 38
Instruction Manual Tek-Thermal 1700B
7 Maintenance
This section covers maintenance techniques and guidelines.
7.1 Sensor Cleaning
The sensor is insensitive to the small amount of dirt, but due to continuous use in the dirty environment will need to do the periodic cleaning.
Follow these instructions while cleaning the sensor
Disconnect power supply from electronics
Remove the sensor carefully from pipe
Clean the sensor with water alcoholic solutions. Use soft brush or cotton bug for
this purpose
Mount the sensor in pipe and switch on the power supply
7.2 Remove transmitter circuit board
Disconnect the power supply from electronics
Remove the front cover of transmitter
Loosen the 4 screws on the circuit board
Remove all connections on circuit board and remove the board
Place new circuit board and plug it
Tighten the screws and place front cover
36www.tek-trol.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 39
Instruction Manual
Problem
Cause
Solution
Velocity
Very erratic or non-uniform flow
Follow installation requirements
Flowmeter installed with less
downstream of the sensor
Follow installation requirements
Insertion sensor probe not mounted securely
Sensor probe must be mounted securely without vibration
Earthing circuit loop
Check if wiring to the correct terminal
Liquid contains in fluid
Install a water filter in upstream of the meter
Sensor failure
Return to factory for evaluation
The measuring flow is
Sensor is not inserted to the middle of the pipeline
Please fix the sensor in the middle of the pipeline
The flowmeter is not vertical
pipeline
Re-install the meter and make the
of the pipe line
No flow measured
The limit of small signal cut off is set too high
Set the small signal cut off limit to a lower and proper value
The flow rate in pipeline is
Check if there is high voltage
flowmeter
The flow rate in pipeline is
of the meter
Sense the vibration on the Flow profile distortion
Try to find another location for
Sensor failure
Return to factory for evaluation
Transmitter failure
Return to factory for evaluation
Extremely turbulent flow
Do not place the meter near a
Tek-Thermal 1700B
8 Troubleshooting
This section provides troubleshooting techniques for most common operating problems.
Before troubleshooting check the following:
Wiring of power supply
Meter’s flow direction
measurement is erratic or fluctuating
than required minimum pipe diameters upstream and
too high or too low
to the center line of the
smaller than the lower limit of the meter
larger than the higher limit
probe vertical to the center line
instrument or high-frequency interference close to the
pipeline by touch it with hand
the meter
ventilator static mixer or valve
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
37www.tek-trol.com
Page 40
Instruction Manual
Error code
Problem
Repair
Err-003
Temperature sensor disconnected
Check Temperature sensor Err-004
Pressure sensor disconnected
Check pressure sensor
Err-005
About to over total flow
This is a reminding message
Err-006
Display value over limit
The value is over the physical limit of the display
Err-013
Button is pressed and hold for too long time
Check the button circuit Err-014
Reset code setting failed
Check EEPROM
Err-015
Reset digital setting failed
Check EEPROM
Err-016
Read total flow error
Check EEPROM
Err-017
Temperature calibration setting is the wrong
Check the record of temperature calibration
Err-018
pressure calibration setting is the wrong
Check the record of pressure calibration
Err-020
Flow rate limit setting is incorrect
Check the flow rate limit setting
Err-023
Communication connection error
Check the communication link Err-026
Pulse equivalent is set too high
Reset D017
Err-027
Pulse equivalent is set too low
Reset D017
Tek-Thermal 1700B
8.1 Self-Diagnostic Functions
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
38www.tek-trol.com
Page 41
www.tek-trol.com
Tek-Trol LLC Tek-Trol Solutions BV Tek-Trol Middle East FZE
796 Tek Drive Crystal Lake, IL 60014,
USA
Sales: +1 847-655-7428
Support: +1 847-857-6076 Email: tektrol@tek-trol.com www.tek-trol.com
Florijnstraat 18, 4879 AH Etten-Leur,
Netherlands
Sales: +31 76-2031908
SAIF Zone, Y1-067, PO BOX No.
21125, Sharjah, UAE
Sales: +971-6526-8344
TEKMATION LLC reserves the right to change the designs and/or materials of its products without notice. The contents of this publication are the property
of TEKMATION and cannot be reproduced by any other party without written permission. All rights reserved. Copyright © 2020 TEKMATION LLC
TEKMATION LLC DOC#/TEK/MNL/200606/1700B/r02
Tek-Trol is a fully owned subsidiary of TEKMATION LLC. We oer our customers a comprehensive range of products and solutions for process, power and oil & gas industries. Tek-Trol provides process measurement and control products for Flow, Level, Temperature & Pressure Measurement, Control Valves & Analyzer systems. We are present in 15 locations globally and are known
for our knowledge, innovative solutions, reliable products and global presence.
Flow | Level | Temperature | Pressure | Valves | Analyzers | Accessories | TekValSys
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com