Tek-Trol 1300C-025S-W-R-M-1-B User manual

Page 1
Technology Solutions
EK-V
T
OR 1300C
Vortex Mass Flowmeter
Instruction Manual
Document Number: IM-1300C
www.GlobalTestSupply.com
Page 2
NOTICE
Read this manual before working with the product. For personal and system safety, and for optimum product performance, make sure you thoroughly understand the contents before installing, using, or maintaining this product.
© COPYRIGHT Tek-Trol LLC 2021
No part of this publication may be copied or distributed, transmitted, transcribed, stored in a retrieval system, or translated into any human or computer language, in any form or by any means, electronic, mechanical, manual, or otherwise, or disclosed to third parties without the express written permission. The information contained in this manual is subject to change without notice.
www.GlobalTestSupply.com
Page 3
Table of Contents
1 Safety Instructions .................................................................................................... 3
1.1 Intended Use ............................................................................................................... 3
1.2 Certification ................................................................................................................. 3
1.3 Safety Instructions from the Manufacturer ................................................................... 3
1.3.1 Disclaimer .......................................................................................................................... 3
1.3.2 Product Liability and Warranty ......................................................................................... 3
1.3.3 Information Concerning the Documentation .................................................................... 3
1.4 Safety Precautions ....................................................................................................... 3
1.5 Packaging, Transportation and Storage ......................................................................... 4
1.5.1 Packaging ........................................................................................................................... 4
1.5.2 Transportation................................................................................................................... 4
1.5.3 Storage .............................................................................................................................. 5
1.5.4 Nameplate ......................................................................................................................... 5
2 Product Description .................................................................................................. 6
2.1 Introduction ................................................................................................................ 6
2.2 Measuring Principle ..................................................................................................... 6
2.3 Specifications ............................................................................................................... 8
2.4 Dimensional Drawings ................................................................................................. 8
2.5 Model Chart............................................................................................................... 13
3 Installation.............................................................................................................. 14
3.1 General Notes on Installation ..................................................................................... 14
3.2 Safety Precautions ..................................................................................................... 14
3.3 Installation Condition ................................................................................................. 15
4 Electrical Installation ............................................................................................... 18
4.1 Terminal Board .......................................................................................................... 18
4.2 Wiring for a 5-Terminal Board .................................................................................... 19
4.2.1 Wiring for a 3-Wire Pulse Output .................................................................................... 19
4.2.2 Wiring for a 2-Wire HART with 4 to 20mA ...................................................................... 20
4.2.3 Wiring for RS485 ............................................................................................................. 20
4.3 Wiring for a 12-Terminal Board .................................................................................. 20
4.3.1 Wiring for a 3-Wire Pulse Output .................................................................................... 20
4.3.2 Wiring for a 3-Wire HART with 4 to 20mA ...................................................................... 21
4.3.3 Wiring for a 4-Wire HART with 4 to 20mA ...................................................................... 22
4.3.4 Wiring for RS485 ............................................................................................................. 22
5 Operation ............................................................................................................... 23
5.1 Display ...................................................................................................................... 23
5.1.1 Introduction to the Multi-Functional LCD Display .......................................................... 23
5.1.2 Units of the Variable Displayed ....................................................................................... 24
5.1.3 Three Button Setting ....................................................................................................... 24
5.1.4 Total Flow Display ........................................................................................................... 25
5.1.5 Modes .............................................................................................................................. 25
5.2 Parameter Setting ...................................................................................................... 26
5.2.1 Code Setting .................................................................................................................... 26
5.2.2 Digit Setting ..................................................................................................................... 27
5.2.3 Setting Address ............................................................................................................... 28
5.2.4 Example of Setting .......................................................................................................... 32
5.2.5 Password Setting Instruction .......................................................................................... 32
5.3 Communication using RS485 MODBUS ....................................................................... 33
5.4 Bluetooth Access........................................................................................................ 39
www.GlobalTestSupply.com
Page 4
5.4.1 How to download the application ................................................................................... 39
5.4.2 Access the application ..................................................................................................... 39
5.4.3 Utilize the application ..................................................................................................... 39
6 Maintenance ........................................................................................................... 40
6.1 How to Rotate Transmitter Head ................................................................................ 40
6.2 Replace a Transmitter Circuit Board ............................................................................ 40
6.3 Replace the Pressure Sensor ....................................................................................... 41
7 Troubleshooting ...................................................................................................... 41
7.1 Troubleshooting Table ............................................................................................... 41
7.2 Self-Diagnostic Messages ........................................................................................... 43
www.GlobalTestSupply.com
Page 5
1 Safety Instructions
1.1 Intended Use
Tek-Vor 1300C is primarily used to measure volumetric flow rate of gas, steam, and liquids. It is suitable for SIP and CIP process in food, beverage, and pharmaceutical industries. It is also used in water and wastewater industry. The volumetric flow rate, temperature, pressure and density of any liquid, gas or steam are measured as an option. From these parameters, Tek-Vor 1300C can calculate the mass flow rate.
1.2 Certification
General purpose IP67/NEMA 4X
1.3 Safety Instructions from the Manufacturer
1.3.1 Disclaimer
The manufacturer will not be held accountable for any damage that happens by using its product, including, but not limited to direct, indirect, or incidental and consequential damages. Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale. The manufacturer has the right to modify the content of this document, including the disclaimer, at any time for any reason without prior notice, and will not be answerable in any way for the possible consequence of such changes.
1.3.2 Product Liability and Warranty
The operator shall bear authority for the suitability of the device for the specific application. The manufacturer accepts no liability for the consequences of misuse by the operator. Wrong installation or operation of the devices (systems) will cause the warranty to be void. The respective Terms and Conditions of Sale, which forms the basis for the sales contract shall also apply.
1.3.3 Information Concerning the Documentation
To prevent any injury to the operator or damage to the device it is essential to read the information in this document and the applicable national standard safety instructions. This operating manual contain all the information that is required in various stages, such as product identification, incoming acceptance and storage, mounting, connection, operation, and commissioning, troubleshooting, maintenance, and disposal.
1.4 Safety Precautions
You must read these instructions carefully prior to installing and commissioning the device. These instructions are an important part of the product and must be kept for future reference. Only by observing these instructions, optimum protection of both personnel and the environment, as well as safe and fault-free operation of the device can be ensured.
For additional information that are not discussed in this manual, contact the manufacturer
Warnings and Symbols Used
The following safety symbol marks are used in this operation manual and on the instrument.
www.GlobalTestSupply.com
Page 6
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices.
NOTE
Indicates that operating the hardware or software in this manner may damage it or lead to system failure.
1.5 Packaging, Transportation and Storage
1.5.1 Packaging
The original package consists of
1. Tek-Vor 1300C Vortex Flowmeter
2. Documentation
NOTE
Unpack and check the contents for damages or signs of rough handling. Report damage to the manufacturer immediately. Check the contents against the packing list provided.
1.5.2 Transportation
Avoid impact shocks to the device and prevent it from getting wet during transportation.
www.GlobalTestSupply.com
Page 7
Verify local safety regulations, directives, and company procedures with respect to hoisting, rigging, and transportation of heavy equipment.
Transport the product to the installation site using the original manufacturer’s packing whenever possible.
1.5.3 Storage
If this product is to be stored for a long period of time before installation, take the following precautions:
Store your product in the manufacturer’s original packing used for shipping.
Storage location should conform to the following requirements:
o Free from rain and water o Free from vibration and impact shock o At room temperature with minimal temperature and humidity variation
Before storing a used flowmeter remove any fluid from the flowmeter line completely. Properties of the instrument can change when stored outdoors.
1.5.4 Nameplate
The nameplate lists the order number and other important information, such as design details and technical data
Note
Check the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
www.GlobalTestSupply.com
Page 8
2 Product Description
This section covers the reference and specification data, as well as ordering information.
2.1 Introduction
Tek-Vor 1300C Vortex Flowmeter (also called a Vortex Shedding Flowmeter) is a versatile instrument that calculates the mass flow, volumetric flow rate, temperature, and pressure and density of any liquid, gas, or steam through a pipeline.
2.2 Measuring Principle
This flowmeter operates on the principle of Karman Vortex Street, any medium passing through the pipeline flows around the bluff body and sheds a series of alternating vortices on each side of the body. This phenomenon is referred to as Vortex Shedding. These vortices shed downstream of the bluff body and dissipate as they flow further. This pattern of vortices is called a Karman Vortex Street (also called a Von Karman Vortex Street).
A Vortex Flowmeter primarily consists of a bluff body, a sensor assembly, and a transmitter. A bluff body or a shedder is nothing but a non-streamlined object or a barrier placed perpendicular to the axis of the pipeline, around which the medium flows.
Calculation of volumetric flow rate
The frequency of the vortices, i.e., the number of vortices shed per second, is directly proportionate to the velocity of the medium. This Vortex Shedding Frequency is used to calculate the mass flow as well as the volume flow. The sensor assembly records the pressure and velocity oscillations generated on each side of the bluff body by the vortices and generate a digital linear output signal. The Vortex Shedding Frequency is calculated using the following formula:
𝒇 = 𝑺𝒕 ×
𝑽
𝒅
Where, f = Frequency of Vortex Shedding
www.GlobalTestSupply.com
Page 9
St = Strouhal Number V = Flow Velocity d = Width of the Bluff Body
The Vortex Shedding frequency is directly proportional to the velocity of any given bluff body diameter.
𝒇 = 𝒌 × 𝑽
Where, k = A constant for all fluids on the given design of flowmeter
Hence,
𝑽 =
𝒇
𝒌
Then the volumetric flow rate can be calculated by using the formula:
𝒒𝒗= 𝑨 ×
𝒇
𝒌
Where, A = Area of flowmeter bore
NOTE
Strouhal’s Number is constant across a wide range of Reynold’s number
www.GlobalTestSupply.com
Page 10
2.3 Specifications
2.4 Dimensional Drawings
Size and Dimension for Wafer Type
www.GlobalTestSupply.com
Page 11
D
(Size)
In.
(mm)
K
(Pipe
OD)
In.
(mm)
L (Pipe
length)
In.
(mm)
W
(Flange
screw
hole
distance)
In.
(mm)
C (Flange
thickness)
In.
(mm)
M (Screw
hole
diameter)
In.
(mm)
N
(Screw
qty.)
Meter
height
In.
(mm)
Flange
OD
In.
(mm)
1"
(25)
3.58" (91)
2.55" (65)
3.93" (100)
0.70" (18)
0.51" (13)
4
11.35"
(288.5)
5.11" (130)
1½"
(40)
3.30" (84)
2.55" (65)
4.72" (120)
20" (0.78)
0.51" (13)
4
11.63"
(295.5)
5.70" (145)
2"
(50)
3.7" (94)
2.55" (65)
5.19" (132)
0.86" (22)
0.66" (17)
4
11.85" (301)
6.29" (160)
3"
(80)
4.27" (120)
2.55" (65)
6.29" (160)
0.94" (24)
0.66" (17)
6
12.44" (316)
7.55" (192)
4"
(100)
5.51" (140)
3.54" (90)
7.48" (190)
0.94" (24)
0.66" (17)
8
12.87" (327)
9.05" (230)
6"
(150)
7.48" (190)
2.55" (65)
9.44" (240)
1.10" (28)
0.82" (21)
8
13.89" (353)
11.02"
(280)
8"
(200)
9.44" (240)
3.34" (85)
11.65" (296)
1.10" (28)
0.82" (21)
12
14.88" (378)
13.18"
(335)
10"
(250)
11.41" (290)
3.93" (100)
13.93" (354)
1.10" (28)
0.82" (21)
12
15.90" (404)
15.94"
(405)
12"
(300)
13.38" (340)
4.72" (120)
16.22" (412)
1.18" (30)
0.82" (21)
12
16.88" (429)
18.11"
(460)
Size and Dimension for Flanged Type (150# ANSI)
www.GlobalTestSupply.com
Page 12
D
(Size)
In.
(mm)
K (pipe
OD)
In.
(mm)
L (Pipe
length)
in. (mm)
W
(Flange
screw
hole
distance)
In.
(mm)
C (Flange
thickness)
In.
(mm)
M (Screw
hole
diameter)
In.
(mm)
N
(Screw
qty.)
Meter height
In.
(mm)
1"
(25)
4.33" (110)
7.08" (180)
3.12"
(79.4)
0.57"
(14.7)
0.62" (16)
4
11.63"
(295.5)
1½" (40)
4.92" (125)
7.08" (180)
3.87"
(98.4)
0.70"
(17.9)
0.62" (16)
4
11.90"
(302.5)
2"
(50)
5.90" (150)
7.08" (180)
4.75"
(120.7)
0.76"
(19.5)
0.74"
(19)
4
12.08" (307)
3"
(80)
7.48" (190)
7.87" (200)
6"
(152.4)
0.95"
(24.3)
0.74" (19)
4
12.86" (326)
4"
(100)
9.05" (230)
8.66" (220)
7.5"
(190.5)
0.95"
(24.3)
0.74" (19)
8
13.22" (336)
6"
(150)
11.02" (280)
8.66" (220)
9.5"
(241.3)
1.01"
(25.9)
0.86" (22)
8
14.17" (360)
8"
(200)
13.58" (345)
8.66" (220)
11.75"
(298.5)
1.14" (29)
0.86"
(22)
8
15.15" (385)
10"
(250)
15.94" (405)
9.84" (250)
14.25" (362)
1.20"
(30.6)
0.98" (25)
12
16.24"
(412.7)
12"
(300)
19.09" (485)
11.81" (300)
17"
(431.8)
1.26"
(0.62)
0.98" (25)
12
17.53"
(445.4)
Size and Dimension for Multi-Variable Wafer Type
www.GlobalTestSupply.com
Page 13
Size
In.
(mm)
K (pipe
OD)
In.
(mm)
L (Pipe
length
) In.
(mm)
W
(Flang
e
screw
hole distan ce) In.
(mm)
C
(Flang
e
thickn
ess) In.
(mm)
M
(Screw
hole
dia-
meter)
In.
(mm)
N
(Screw
qty.)
Meter
height
In.
(mm)
Flange
OD
In.
(mm)
H2 conde nsatio n pipe
height
In.
(mm)
L1 conde nsatio n pipe
length
In.
(mm)
1"
(25)
3.58" (91)
2.55" (65)
3.93"
(100)
0.70" (18)
0.51" (13)
4
11.35"
(288.5)
5.11"
(130)
5.74"
(146)
5.35"
(136)
1½" (40)
3.30" (84)
2.55" (65)
4.72"
(120)
20"
(0.78)
0.51" (13)
4
11.63"
(295.5)
5.70"
(145)
6.21"
(57.8)
5.07"
(129)
2"
(50)
3.7" (94)
2.55" (65)
5.19"
(132)
0.86" (22)
0.66" (17)
4
11.85" (301)
6.29" (160)
6.61" (168)
4.88" (124)
3"
(80)
4.27" (120)
2.55"
(65)
6.29" (160)
0.94"
(24)
0.66"
(17)
6
12.44" (316)
7.55" (192)
8.11" (206)
4.29" (109)
4"
(100)
5.51" (140)
3.54"
(90)
7.48" (190)
0.94"
(24)
0.66"
(17)
8
12.87" (327)
9.05" (230)
8.54" (217)
3.89"
(99)
6"
(150)
7.48" (190)
2.55"
(65)
9.44" (240)
1.10"
(28)
0.82"
(21)
8
13.89" (353)
11.02" (280)
10.55" (268)
3.89"
(99)
8"
(200)
9.44" (240)
3.34"
(85)
11.65" (296)
1.10"
(28)
0.82"
(21)
12
14.88" (378)
13.18" (335)
12.51" (318)
3.89"
(99)
10"
(250)
11.41" (290)
3.93" (100)
13.93" (354)
1.10"
(28)
0.82"
(21)
12
15.90" (404)
15.94" (405)
14.52" (369)
3.89"
(99)
12"
(300)
13.38" (340)
4.72" (120)
16.22" (412)
1.18"
(30)
0.82"
(21)
12
16.88" (429)
18.11" (460)
16.49" (419)
3.89"
(99)
Size and Dimension for Multi-Variable Flanged Type (150# ANSI)
www.GlobalTestSupply.com
Page 14
D
(Size)
in.
(mm)
K
(Pipe
OD)
in.
(mm)
L
(Pipe
length)
in.
(mm)
W
(Flange
screw
hole
distance)
in.
(mm)
C
(Flange
thickness)
in.
(mm)
M (Screw
hole
diameter)
in.
(mm)
N
(Screw
qty.)
Meter height
in.
(mm)
Flange
OD
in.
(mm)
H2
condensa
tion pipe
height
in.
(mm)
1"
(25)
4.33" (110)
7.08" (180)
3.12"
(79.4)
0.57"
(14.7)
0.62" (16)
4
11.63"
(295.5)
6.71"
(170.5)
5.35"
(136)
1½" (40)
4.92" (125)
7.08" (180)
3.87"
(98.4)
0.70"
(17.9)
0.62" (16)
4
11.90"
(302.5)
7.28" (185)
5.07" (129)
2"
(50)
5.90" (150)
7.08" (180)
4.75"
(120.7)
0.76"
(19.5)
0.74" (19)
4
12.08" (307)
7.95" (202)
4.88" (124)
3"
(80)
7.48" (190)
7.87" (200)
6"
(150)
0.95"
(24.3)
0.74" (19)
4
12.86" (326)
9.48" (241)
4.29" (109)
4"
(100)
9.05" (230)
8.66" (220)
7.5"
(190.5)
0.95"
(24.3)
0.74" (19)
8
13.22" (336)
10.66" (271)
3.89" (99)
6"
(150)
11.02" (280)
8.66" (220)
9.5"
(241.3)
1.01"
(25.9)
0.86" (22)
8
14.17" (360)
12.59" (320)
3.89" (99)
8"
(200)
13.58" (345)
8.66" (220)
11.75"
(298.5)
1.14" (29)
0.86" (22)
8
15.15" (385)
14.86"
(377.5)
3.89" (99)
10"
(250)
15.94" (405)
9.84" (250)
14.25"
(362)
1.20"
(30.6)
0.98" (25)
12
16.24"
(412.7)
17.13"
(435.2)
3.89" (99)
12"
(300)
19.09" (485)
11.81" (300)
17"
(431.8)
1.26"
(0.62)
0.98" (25)
12
17.53"
(445.4)
19.99"
(507.9)
3.89" (99)
www.GlobalTestSupply.com
Page 15
2.5 Model Chart
Example
Tek-Vor 1300C
050S W R
M 1 B
Tek-Vor 1300C-050S-W-R-M-1-B
Series
Tek-Vor 1300C
Vortex Mass Flowmeter
Size
015S
1/2", +/- 1.0% Accuracy, Standard Vortex Meter
025S
1", +/- 1.0% Accuracy, Standard Vortex Meter
040S
1-1/2", +/- 1.0% Accuracy, Standard Vortex Meter
050S
2", +/- 1.0% Accuracy, Standard Vortex Meter
080S
3", +/- 1.0% Accuracy, Standard Vortex Meter
100S
4", +/- 1.0% Accuracy, Standard Vortex Meter
150S
6", +/- 1.0% Accuracy, Standard Vortex Meter
200S
8", +/- 1.0% Accuracy, Standard Vortex Meter
250S
10", +/- 1.0% Accuracy, Standard Vortex Meter
300S
12", +/- 1.0% Accuracy, Standard Vortex Meter
015M
1/2", +/- 1.0% Accuracy, Multivariable Vortex Meter
025M
1", +/- 1.0% Accuracy, Multivariable Vortex Meter
040M
1-1/2", +/- 1.0% Accuracy, Multivariable Vortex Meter
050M
2", +/- 1.0% Accuracy, Multivariable Vortex Meter
080M
3", +/- 1.0% Accuracy, Multivariable Vortex Meter
100M
4", +/- 1.0% Accuracy, Multivariable Vortex Meter
150M
6", +/- 1.0% Accuracy, Multivariable Vortex Meter
200M
8", +/- 1.0% Accuracy, Multivariable Vortex Meter
250M
10", +/- 1.0% Accuracy, Multivariable Vortex Meter
300M
12", +/- 1.0% Accuracy, Multivariable Vortex Meter
Process Connection
W
Wafer (Comes with two 150# ANSI flange adaptors)
F
150# ANSI Flange
T
300# ANSI Flange
Output
I
4-20mA, Pulse
H
4-20mA, Pulse, HART (only available for direct mount)
R
4-20mA, Pulse, Modbus RS485
Process
Temperature S
302 Degrees F (150 Degrees C) (MV or Standard Vortex)
M
482 Degrees F (250 Degrees C) (MV or Standard Vortex)
H
662 Degrees F (350 Degrees C) (Standard Vortex Only)
Electronics 1
Direct Mount
2
Remote Mount (comes with 15 ft. of cable)
Diagnostics
B
Bluetooth
www.GlobalTestSupply.com
Page 16
3 Installation
Vibration
The flowmeter should not be installed at a location where it could have strong vibration. If the mounting pipeline could have heavy vibration, the pipeline should be held steady by some support racks.
All screws and bolts should be tightened Make sure there is no leakage point on the connection
The process pressure should not be higher than the flowmeter’s rated pressure
This section covers instructions on installation and commissioning. Installation of the device must be carried out by trained; qualified specialists authorized to perform such works.
CAUTION
When removing the instrument from hazardous processes, avoid direct contact with the fluid and the meter
All installation must comply with local installation requirements and local electrical code
3.1 General Notes on Installation
Ambient Temperature
Please avoid installing the flowmeter at a location where temperature could dramatically change. If the meter is under heavy heat radiation, please implement effective heat insulation and venting method.
Atmosphere
Please do not install the flowmeter at a location where the atmosphere contains a high level of corrosive substance. If the meter cannot be installed at a better location, please make sure there is enough venting.
CAUTION
Once the meter is under pressure, please do not screw the bolts and screws
3.2 Safety Precautions
For person and equipment safety, please observe below provision:
Before installation, please read this manual properly, check the safety requirements for flowmeter, relevant equipment, and environment.
www.GlobalTestSupply.com
Page 17
Install and maintain flowmeter by person who has the knowledge of flowmeter.
Install flowmeter sensor and its pipe correctly, make sure the seal and safety, liquid pressure
shall be no more than the maximum working pressure on nameplate.
Prevent electric shock accident.
Lifting equipment for flowmeter should confirm to safety provision.
3.3 Installation Condition
NOTE
When the upstream pipeline is a T-type and the flowmeter and the valve downstream of the meter is shut, fluid will flow toward pipe B, however the meter may still show a reading, as it may be detecting a pulsating pressure. In this case, please close the upstream valve V1.
Standard Installation
Installation for the Pipe Reducer
Installation for the Pipe Expander
www.GlobalTestSupply.com
Page 18
Installation for Single Bend Pipeline
Installation for Double Bend Pipeline
Installation when Valve is at Downstream
Installation when Valve is at Upstream
Installation when Temperature and Pressure Sensors are at Downstream
www.GlobalTestSupply.com
Page 19
Installation when Roots Blower, Piston Blower, or Compressor are at Upstream
Installation of a Flowmeter after a Piston Pump
Installation when T-type Pipeline is at Upstream
www.GlobalTestSupply.com
Page 20
4 Electrical Installation
This section covers all electrical connection requirement. Electrical connection of the device must be carried out by trained; qualified specialists authorized to perform such work by the installation site.
WARNING
Connect all electrical cables when the power is switched off. If the device does not have switch-off elements, then, overcurrent protection devices, lightning protection and/or energy isolating devices must be provided by the customer.
The device must be grounded to a spot in accordance with regulations in order to protect personnel against electric shocks.
NOTE
When using the measuring device in hazardous areas, installation must comply with the corresponding national standards and regulations and the Safety Instructions or Installation or Control Drawings.
4.1 Terminal Board
The Tek-Vor 1300C has two different terminal boards; a 5-terminal board for standard models and a 12-terminal board for multi-variable models.
www.GlobalTestSupply.com
Page 21
Power supply
V +
V -
Pulse output
RS485 communication
A+
B-
Current
I+
I-
Temperature sensor
RTD1, RTD2, RTD3
Pressure Sensor
P+
P-
On the above boards, V+ and V- are for power. is the pulse output terminal. A, B are “+” and -terminals for RS485 Modbus communication. I+ and I- are “+” and “–“ for 3-wire or 4-wire 4 to 20 mA. RT1, RT2, RT3 are for separate RTD. P+, P- are for pressure transmitters. The Tek- Vor 1300C multi–variable version has a built in RTD and pressure sensor, so there is no requirement to wire for temperature or pressure compensation.
4.2 Wiring for a 5-Terminal Board
4.2.1 Wiring for a 3-Wire Pulse Output
A 3-wire pulse output requires a power source of 13.5 to 42VDC. VFM uses a current pulse output with 50% duty ratio. If the pulse-receiving instrument requires a voltage pulse, add a
resistor between “ ” and “V-”; the resistance should be within 500 ohms to 1000 ohms, and the power consumption should be no less than 0.5W.
*Note:- Pulse Output: Active (Pulse, V-), 4-20 mA output: Passive (I+, I-),
Remarks for Pulse Output: With/Without 250 ohms resistor across Pulse &V- terminals, Remarks for 4-20 mA output: External 24VDC to I+.connect PLC/Multimeter between I- and -Ve of power supply, Diagrams for Pulse: Active, Diagrams for 4-20mA:
www.GlobalTestSupply.com
Page 22
4.2.2 Wiring for a 2-Wire HART with 4 to 20mA
When there is no temperature and pressure compensation and the power source is 24VDC, the maximum load for 4 to 20mA analog is 500 ohms. When there is temperature and pressure compensation and the power source is 24VDC, the maximum load for 4 to 20mA analog is 400 ohms. When using a HART communicator, add a 250 ohms load resistor.
4.2.3 Wiring for RS485
4.3 Wiring for a 12-Terminal Board
4.3.1 Wiring for a 3-Wire Pulse Output
A 3-wire pulse output requires a power source of 13.5 to 42VDC. VFM uses a current pulse output with 50% duty ratio. If the pulse receiving instrument requires a voltage pulse, add a
www.GlobalTestSupply.com
Page 23
resistor between “ and “V-”; the resistance should be within 500Ω to 1000Ω, and the power consumption should be no less than 0.5W.
4.3.2 Wiring for a 3-Wire HART with 4 to 20mA
When the power source is 24VDC, the maximum load for 4 to 20mA analogue is 500 ohms.
www.GlobalTestSupply.com
Page 24
4.3.3 Wiring for a 4-Wire HART with 4 to 20mA
When the power source is 24VDC, the maximum load for 4 to 20mA analogue is 500 ohms.
4.3.4 Wiring for RS485
www.GlobalTestSupply.com
Page 25
5 Operation
This section covers operation techniques and guidelines.
5.1 Display
The Tek-Vor 1300C Vortex Flowmeter provides local settings and a display panel. It can display several variables on this multifunctional LCD display. It has 3 buttons.
5.1.1 Introduction to the Multi-Functional LCD Display
The Tek-Vor 1300C Vortex Flowmeter has a display to indicate “Frequency”, “Flow Rate”, and “Total flow”. The Tek-Vor 1300C Multi-variable version or a standard Tek-Vor 1300C working
with RTD and pressure transmitters can also indicate other variables such as temperature/pressure/density/mass flow”.
The LCD display has 2 areas to display the content - the upper row and the lower row. The upper row displays the flow rate/mass flow/standard flow rate. Immediately below the upper row is displayed the unit of the variables in the upper row. The lower row display indicates other variables, such as frequency/ pressure/ temperature/ density/ total flow/ velocity. Immediately below the lower row is displayed the unit of the variables in lower row.
The Tek-Vor 1300C Multi-variable version and the Standard version with temperature and pressure compensation, can calculate and display the mass flow of both saturated steam and superheated steam.
www.GlobalTestSupply.com
Page 26
The Tek-Vor 1300C multi-variable version and the standard version with temperature and pressure compensation can display variables such as temperature/ pressure/ density. Use the “Switch” button to switch to the next variable and it will display for 30 seconds.
Below is a sample of the temperature being displayed. The lower row can also consistently display a variable through the settings. The default variable displayed on the lower row is total flow. You can also set the lower row to display several variables in circular turn.
5.1.2 Units of the Variable Displayed
Subject
Variable
Unit
Circular Display
Code
TOTAL
Total Flow
m/s (ft/sec)
01
TEMP
Temperature
OF(O
C)
02
PRES
Pressure
MPa or kPa
03
FREQ
Frequency
Hz
04
DENS
Density
kg/m3
05
5.1.3 Three Button Setting
Tek-Vor 1300C vortex flowmeter has three buttons on the top of the display, which are
used as L-R button, used as U-D button, used as Enter button.
www.GlobalTestSupply.com
Page 27
The “U-D” button is used to switch the displaying content up and down, the “L-R” button can
Working Mode Setting Mode
be used for the left and right digits of total flow. The “Enter” button” displays the entire digits of total flow directly.
5.1.4 Total Flow Display
The Tek-Vor 1300C can display 9 digits to the left of decimal point and 3 digits to the right of it. When there is more than 6 digits, the total flow reading will be displayed twice; first it will display digits on right, and then second the digits on left. You can use the “L-R” button to switch between the right and left digits. The left digits will be displayed with “x1000”
To check the right digits now, press the “L-R button”, the display will be
According to the above images, the total flow is 569864.581 kg.
5.1.5 Modes
The Tek-Vor 1300C Vortex Flowmeter have three different Modes:
Calibration Mode
When under the Setting mode you can set the flowmeter while the flowmeter is still processing, so setting will not have effect on the measuring parameters. The calibration of the flowmeter has been finished in manufacturer’s laboratory before delivery, including temperature and pressure calibration along with the setting of the high-limit and the low-limit of 4 to 20mA stimulation output.
www.GlobalTestSupply.com
Page 28
5.2 Parameter Setting
The Tek-Vor 1300C Vortex Flowmeter has Digit and Code Setting functions. Use the Code Setting function to set parameters such as fluid type, compensation type, and output signal. Use the Digit Setting function to set parameters related to a number such as pipe size, flow range factor.
NOTE
Tek-Vor 1300C Vortex Flowmeters have been set according to requirement before delivery. Please do not change the setting unless it is necessary and under correct instruction
5.2.1 Code Setting
To enter the Code Setting in working status, press the “Enter” button, and the “U-D” button at the same time.
When in the code setting function, the upper row will display the reference number for the code setting, and the lower row will display the contents of this parameter. The digit that is flashing is the digit under the setting. In the diagram below, this is shown as C01=02, which means the fluid type is liquid.
When in the code setting function, the user can now use “L-R” to select which digits on the display are to be set, and “R-D” to switch the digit to between 0 and 9. When “Enter” is pressed, it will set the lower row. Press Enter again to check if the setting is available. If the
www.GlobalTestSupply.com
Page 29
setting is available, the setting made will be cancelled and the display will not flash. If this happens, press “L-R” or “U-D” to set it again. When the display is not flashing, press “Enter” to save and go to the next setting. If the user wants to quit the code setting function, hold down “Enter” and press “U-D” at the same time.
5.2.2 Digit Setting
To enter the Digit Setting function in working status, hold down “Enter” and press “L-R” at the same time.
When in the Digit Setting function, the upper row will display the reference number of the digital setting and the lower row will display the contents of this parameter. In the diagram below, the digit that is flashing is the digit under the setting. This means D001=1.60000; the maximum pressure is 1.6 (unit according to other settings).
When in the Digit Setting function, the user can use “L-R” to select which digit on the display
panel are to be set and use “R-D” to switch the digit to between 0 and 9. The first press of “Enter” sets the lower row. Press “Enter” again to check if the setting is available. If the setting
is available, the setting made will be cancelled and the display will not flash. If this happens, press “L-R” or “U-D” to set it again. When the display is not flashing, press “Enter” to save and go to the next setting. If want to quit the digital setting function, again hold down “Enter” and press “U-D” at the same time.
www.GlobalTestSupply.com
Page 30
5.2.3 Setting Address
Code Setting Address
NOTE
If the unit of flow rate is changed or the measurement is changed from the flow rate to mass flow, users can reset the total flow to 0 or record the current total flow: Total flow= (number of times over total flow) * (maximum display of total flow) + (Current total flow reading)
Code Setting Address
01 Fluid 01 Gas
02 Density
05 Output 01 Pulse
Item Code Description of Code
02 Liquid 00 Volume Flow Display, No
Compensation
01 Density Preset 02 Pressure Compensation (for
03 Temperature Compensation
04 Temperature and Pressure
05 ρ=A+BP (Pressure
06 ρ=A+BT (Temperature
07 AGA-NX-19 to Calculate
08 Temperature and Pressure
09 AGA-8 to Calculate
02 4 to 20mA or HART at 4 to
03 200-1000HZ Frequency
Density Compensation
Saturated Steam Pressure Not Larger than 20MPa
(for Saturated Steam)
Compensation (for Superheated Steam)
Compensation)
Compensation)
Compressibility Factor
Compensation to get for Normal Condition Flow Rate of Gas
Compressibility Factor
20mA
Output, Set Output in C06
www.GlobalTestSupply.com
Page 31
04 Frequency Output for Total
Flow, Set Factor in D013
06 200-1000Hz Output
Parameter
07 Damping 01 to
08 Instrument Number 00 to
09 Baud Rate 01 1200 No Parity 1 Stop Bit
10 Time Unit for Flow
Rate
11 Mass Unit 01 Kg
12 Volume Unit for
Flow Rate
13 Pressure Unit 01 MPa
00 Flow Rate 01 Temperature 02 Pressure
1 to 99 Seconds
99
For Modbus 99 00 to 15
02 1200 Even Parity 1 Stop Bit 03 2400 No Parity 1 Stop Bit 04 2400 Even Parity 1 Stop Bit 05 4800 No Parity 1 Stop Bit 06 4800 Even Parity 1 Stop Bit 07 9600 No Parity 1 Stop Bit 08 9600 Even Parity 1 Stop Bit 09 19200 No Parity 1 Stop Bit 10 19200 Even Parity 1 Stop Bit 11 1200 Odd Parity 1 Stop Bit 12 2400 Odd Parity 1 Stop Bit 13 4800 Odd Parity 1 Stop Bit 14 9600 Odd Parity 1 Stop Bit 15 19200 Odd Parity 1 Stop Bit 16 38400 No Parity 1 Stop Bit 17 38400 Even Parity 1 Stop Bit 18 38400 Odd Parity 1 Stop Bit 19 57600 No Parity 1 Stop Bit 20 57600 Even Parity 1 Stop Bit 21 57600 Odd Parity 1 Stop Bit 22 115200 No Parity 1 Stop Bit 23 115200 Even Parity 1 Stop Bit 24 115200 Odd Parity 1 Stop Bit 00 /s 01 /min 02 /h
02 Ton 03 Lb 01 m3 02 L 03 ft3 04 US gal 05 UK gal
For HART Communication
www.GlobalTestSupply.com
Page 32
02
kPa
03
Psi
04
Bar
14
Temperature Unit
01 OC
02 OF
03
K
15
Right Digits Number for Total Flow
00 to 05
00:No Right Digits for Total Flow
01 to 051 to 5 Right Digits for Total Flow
16
1st Row Display Parameter 01
Flow Rate
02
Percentage of Flow Rate to Flow Range
17
Lower Row Display Parameter
00
No Display
01
Total Flow
02
Temperature
03
Pressure
04
Density
05
Frequency
18
Density Unit 01
Kg/m3
02
lb/ft3
30
Time Space for Circle Display
00 to 30
1 to 301 to 30 Seconds Between the Display of Different Parameter
38
Sequence of Float (under RS485 Communication)
01
LL_LH_HL_HH
02
HH_HL_LH_LL
03
LH_LL_HH_HL
04
HL_HH_LL_LH
47
Password Function 0
Off 1 On
48
Set Password 0
Keep the Password
1
Change the Password
49
Spectrum Analyzing Checking
0
Working Status
12
Spectrum Analyzing Checking
50
Total Flow Reset 0
Reset Total Flow to 0
1
Default
55
Times of Over Total Flow
00 to 99
For Reading Only
60
Restore to Backup Date
6
Restore to Backup Date
61
Save Setting Backup
16
Save Current Setting for Backup
www.GlobalTestSupply.com
Page 33
Digit Setting Address
Digit Setting Address
Item
Code
Description of Code
001
Max Pressure
[-99999, 999999]
Max Input/Output Pressure
002
Min Pressure
[-99999, 999999]
Min Input/Output Pressure
003
Max Temperature
[-99999, 999999]
Max Input/Output Temperature
004
Min Temperature
[-99999, 999999]
Min Input/Output Temperature
005
Pre-set Density
[0999999]
When C02=01, the Flowmeter will use this Density, Unit According to Setting
008
K Factor
[0999999]
K Factor According to Calibration Result, unit is Pulses/Litre. Flow=3.6xFreq/K
009
Max Flow Rate
[0999999]
Unit is same as Flow Rate, Max/Min Flow Rate of 4 to 20mA and 200 to 1000Hz Output
010
Min Flow Rate
[0,999999] 013
Pulse Factor for Total Flow
[0999999]
Used when Freq. Output of Total Flow
014
Ambient Pressure
[0999999]
Unit According to Setting
015
Pipe Size
[0999999]
Unit is mm
021
Cut off Small signal
[0,999999]
Unit is Hz
030
Specific Density of Compressibility Factor
[0.550.90]
For Calculation of Compressibility Factor of Natural Gas
031
Mol % of N2 and H2
[0,0.1]
For Calculation of Compressibility Factor of Natural Gas
032
Mol % of CO2
[0,0.3]
For Calculation of Compressibility factor of Natural Gas
033
Higher Heating Value
[20,48]
KJ/Mol, for Calculation of Compressibility Factor of Natural Gas
www.GlobalTestSupply.com
Page 34
NOTE
Maximum frequency output=10KHz. The pulse factor for total flow should be set properly according to the current total flow
5.2.4 Example of Setting
For Tek-Vor 1300C Vortex Flowmeter, measure gas in a 2” (DN50) pipe; K factor= 7.802P/L, density pre-set, mass flow display unit is kg/h. 4 to 20mA output with a flow range of 0 to 4000kg/hr.
Code Setting
Address
Code
Description
01
01
Gas
02
01
Density Pre-set
05
02
4 to 20mA Analog Output
Digit Setting
005
2.0000
Density=2
008
7802
K Factor=7.802 P/L
009
4000
Flow Rate of 20mA
010
0
Flow Rate of 4mA
015
50
Pipe Size=50mm
5.2.5 Password Setting Instruction
Select the code setting, set C47=01, confirm and then enter the password setting interface:
To set a new password, enter the correct password twice; the password will only become active if both entries are correct; otherwise users have to enter the password again. If the power is cut off during a password setting process, the password will revert to 2000 as default. When a password becomes active, users will have to enter the correct password before the flowmeter can be set. If the user enters 3 times consecutively, the display will revert to the normal display.
www.GlobalTestSupply.com
Page 35
5.3 Communication using RS485 MODBUS
Interface Regulation
1. The communication interface should be RS485, the range of Baud rate should be 1200 to 115200
2. The wiring terminal for communication is “A” and “B”. Refer to the Electrical Installation Section for Wiring Terminal information
3. The communication should comply with MODBUS-RTU statute
The combination of a communication signal: Address code - function code – date segment – CRC calibration code. The distance between two characters should not be longer than one character, or it will be considered the beginning of a new message or the end of an old message. The message is combined with hexadecimal arrays.
a. Address of the Displayed Date
Register Address
Usage
Nature
Date Type
0-1
Flow Rate
Read Only
Float
2-3
Frequency
Read Only
Float
4-5
Reserved
Read Only
Float
6-7
Pressure
Read Only
Float
8-9
Temperature
Read Only
Float
10-11
Density
Read Only
Float
12-13
Reserved
Read Only
Float
14-15
Reserved
Read Only
Float
16-17
Reserved
Read Only
Float
18-19
Reserved
Read Only
Float
20-21
Reserved
Read Only
Float
22-23
Reserved
Read Only
Float
24-25
Total Flow
Read Only
Float
The displayable date including flow rate, frequency, pressure, temperature, density and total flow, if the meter does not have density compensation, then the reading of pressure and temperature will both be 0. Using function code 03 according to the address above and shifting can read the date of the parameters in above chart.
www.GlobalTestSupply.com
Page 36
b. Addresses of Code Setting
Resister
Usage
Range
Nature
Date Type
1000
Fluid Type C01
1-2
Read Only
Short
1001
Density Compensation C02
0-9
Read/Write
Short
1004
Output C05
1-4
Read/Write
Short
1005
200-1000Hz Output Parameter C06
1-3
Read/Write
Short
1006
Damping C07
1-99
Read/Write
Short
1007
Instrument Number C08
Hart (0 to 15) MB (1 to 99)
Read
Short 1008
Baud Rate C09
1-24
Read
Short
1009
Unit of Time C10
0-2
Read/Write
Short 1010
Mass Unit C11
1-3
Read/Write
Short
1011
Volume Unit C12
1-5
Read/Write
Short
1012
Pressure Unit C13
1-4
Read/Write
Short
1013
Temperature Unit C14
1-3
Read/Write
Short
1014
Right Digits Number for Total Flow C15
0-5
Read/Write
Short 1015
1st Row Display Parameter C16
1-2
Read/Write
Short
1016
Lower Row Display Parameter C17
0-5
Read/Write
Short 1017
Density Unit C18
1-2
Read/Write
Short
1029
Time Space for Circle Display C30
0-30
Read/Write
Short
1030
First Parameter Displayed in Circle Display C31
0-5
Read/Write
Short
1031
Second Parameter Displayed in Circle Display C32
0-5
Read/Write
Short
www.GlobalTestSupply.com
Page 37
1032
Third Parameter Displayed in Circle Display C33
0-5
Read/Write
Short
1033
Fourth Parameter Displayed in Circle Display C34
0-5
Read/Write
Short 1034
Fifth Parameter Displayed in Circle Display C35
0-5
Read/Write
Short
1035
C36
0-1
Read/Write
Short
1036
C37
0-10
Read/Write
Short
1037
Sequence of Float C38
1-4
Read/Write
Short
1046
Password Function C47
0-1
Read
Short
1047
Set Password C48
0-1
Read
Short
1048
Spectrum Analyzing Checking C49
0-12
Read/Write
Short 1049
Total Flow Reset to 0 C50
0-1
Read/Write
Short 1050
C51
0-0
Read/Write
Short
1051
C52
0-99
Read/Write
Short
1052
C53
0-0
Read/Write
Short
1053
C54
0-0
Read/Write
Short
1054
Times of Over Total Flow C55
0-0
Short
1059
Restore to Backup Date C60
0-99
Read/Write
Short 1060
Save Setting Backup C61
0-99
Read/Write
Short
Use the function codes 04 and 06 to access the address for code setting above.
www.GlobalTestSupply.com
Page 38
c. Addresses of Digit Setting
Resister
Usage
Restriction of modification
Nature
Date type
2000 to 2001
D001 Max Pressure
-1e5 to 1e6
Read/Write
Float
2002 to 2003
D002 Min Pressure
-1e5 to 1e6
Read/Write
Float
2004 to 2005
D003 Max Temperature
-1e5 to 1e6
Read/Write
Float
2006 to 2007
D004 Min Pressure
-1e5 to 1e6
Read/Write
Float
2008 to 2009
D005 Density
0 to 1e6
Read/Write
Float
2014 to 2015
D008 K Factor
0 to 1e6
Read/Write
Float
2016 to 2017
D009 Max Flow rate
0 to 1e6
Read/Write
Float
2018 to 2019
D010 Min Flow rate
0 to 1e6
Read/Write
Float
2024 to 2025
D013 Factor for Total Flow Output
0 to 1e6
Read/Write
Float
2026 to 2027
D014 Ambient Pressure
0 to 1e6
Read/Write
Float 2028 to 2029
D015 Pipe Size
0 to 1e6
Read/Write
Float
2040 to 2041
D021 Cut off Small Signal
0 to 1e6
Read/Write
Float
2058 to 2059
D030 Specific Density
[0.550.90]
Read/Write
Float
2060 to 2061
D031 mol% of N2 and H2
[0,0.1]
Read/Write
Float
2062 to 2063
D032 mol% of CO2
[0,0.3]
Read/Write
Float
2064 to 2065
D033 Higher Heating Value
[20,48]
Read/Write
Float
The chart above indicates the register address, usage of the register, restriction of modification, read/write nature, and date type. The registers above are all holding registers; the supporting function codes are 03,04,06, and 16.
www.GlobalTestSupply.com
Page 39
Command
Function codes 03 and 04 are the codes supported for reading the registers. Function code 06 is for writing one register. Function code 16 is for writing multiple registers. Function code 06 is only supported for writing a short date. Function code 16 is supported for writing both a short date and a float date.
a. Function Code 03 – Read Register
Request
Response
01: Address
01: Address
03: Function Code
03: Function Code
00: Register Address Higher
04: Quantity of Bit
00: Register Address Lower (display the address)
80: Date 1 00: Register Number Higher
04: Date 2
02: Register number lower
80: Date 3
CRCL: CRC Parity code lower
80: Date 4
CRCH: CRC parity code higher
CRCL: CRC Parity code lower
CRCH: CRC parity code higher
NOTE
To read a float date, both the quantity of the register address and its value have to be even, or response will be “error:
Function code 04 – Same as function code 03
b. Function Code 06 – Write One Register
Request
Response
01: Address
01: Address
06: Function code
06: Function code
00: Register address higher
00: Register address higher
01: Register address lower (code setting address)
01: Register address lower 00: Value higher
00: Value higher
04: Value lower
04: Value lower
CRCH: CRC parity code higher
CRCH: CRC parity code higher
CRCL: CRC Parity code lower
CRCL: CRC Parity code lower
www.GlobalTestSupply.com
Page 40
c. Function Code 16 – Write Multiple Registers
Request
Response
01: Address
01: Address
10H: Function code
10H: Function code
00: Register address higher
00: Register address higher
01: Register address lower (digital setting address)
01: Register address lower 00: Quantity of register higher
00: Quantity of register higher
02: Quantity of register lower
02: Quantity of register lower
04: Quantity of values
CRCH: CRC parity code higher
86H: Value 1
CRCL: CRC Parity code lower
00: Value 2
00: Value 3
48H: Value 4
CRCH: CRC parity code higher
CRCL: CRC Parity code lower
NOTE
Function code 16 is supported to write both a short date and a float date. However, for a float date, the first register address and the quantity of the registers must both be even, or writing will not be allowed
Calculation of CRC Parity Code
Request
Response
01: Address
N1 CRC=0FFFFH is Initial Value
10: Function code
N2 XOR Operation the CRCL and N1
00: Register Address Higher
N3 CRC Move 1 Bit Right, if Move Out is 1 Bit
01: Register Address Lower
N4 CRC=CRC XOR A001H
00: Register Quantity Higher
N5 if Move Out is 0, CRC=CRC
04: Register Quantity Lower
N6 Move Right for 8 Times to Finish the N1 Calculation
04: Date Quantity
N7 …
80: Date 1
N8 XOR Operation the CRCL and N11
04: Date 2
N9 CRC move 1 bit Right, if Move Out is 1 Bit
80: Date 3
N10 CRC=CRC XOR A001H
80: Date 4
N11 if Move Out is 0, CRC=CRC
CRCL: CRC Parity Code Lower
Move Right for 8 Times to Finish the N11 Calculation
CRCH: CRC Parity Code Higher
Get the CRC Calibration Value
www.GlobalTestSupply.com
Page 41
5.4 Bluetooth Access
5.4.1 How to download the application
Visit Apple’s application store
Search “Tek-trol Bluetooth” or “Tek-Bluetooth” to find our app profile
Download the application
5.4.2 Access the application
Open application
Enter login information (ID: 20000000007, Verify code:341234)
This will take you to “device list” where your phone will automatically sync with your product (you
must be close enough)
5.4.3 Utilize the application
Follow prompts and menu to navigate through the many features of our Bluetooth application!
www.GlobalTestSupply.com
Page 42
6 Maintenance
This section covers maintenance techniques and guidelines.
6.1 How to Rotate Transmitter Head
Before rotating the transmitter, please take out the tightening screw under the transmitter
Rotate transmitter head up to 180o and tighten the screw
6.2 Replace a Transmitter Circuit Board
Make sure the power is off before replacing the transmitter
Remove the front cover
Loose the 4 screws on the circuit board
Remove all the plugs on the circuit board, then remove the circuit board
Put the new circuit board in and put the plug on
Tighten the 4 screws on the board, fix the front cover back on
www.GlobalTestSupply.com
Page 43
6.3 Replace the Pressure Sensor
Make sure the pressure valve on the meter is off before replacing the pressure transmitter
Loosen the nut that is holding the pressure sensor
Remove the front cover, loosen the circuit board and remove the wiring of the pressure
sensor
Remove the enclosure
Loosen the gasket and sealing nut
Slowly take out the pressure sensor and the wire
After installing the pressure sensor, follow the reverse process of instruction
7 Troubleshooting
This section provides troubleshooting techniques for most common operating problems.
7.1 Troubleshooting Table
Error
Reason
Troubleshooting
Repair
No Display
Power Supply Failure
Test the voltage on the power source with a universal meter
Re-wire the power or use a new power
Power is Not Wired
Test the voltage on the power source with a universal meter
Wire the power Cable if Broken
Check for a break off point on the cable
Check the cable and re-wire
Wrong Wiring
Check if wired to the correct terminal
Re-wire
Displayed Flow Rate is 0 While there is Flow in the Pipe
Flow Rate is Lower than the Meter's Lower Limit
Increase the flow rate to check
Increase the flow rate or replace with a new proper flowmeter
The Flow Rate of Small Signal Cut-off Function is too High
Check the small signal cut off setting
Set the small signal cut off to a proper value
www.GlobalTestSupply.com
Page 44
Energy Threshold Value is too High
Check if the Energy threshold value is too high in spectrum analyzing checking mode
Set the Energy threshold value to a proper value
Transmitter Function Failure
Replace the transmitter with another transmitter of same type to check
Replace the transmitter
Sensor is Damaged
Increase the flow rate to check first, then install the transmitter to another flowmeter in same type to check.
Replace the sensor Pipeline Blocked, or Sensor Jammed.
If all above possibilities are eliminated, please check the pipe line and installation.
Re-install the flowmeter
The Flowmeter has Flow Reading with No Flow in the Pipe
Power Frequency Interference
Check the frequency display on meter is stable at the value that same as the power frequency
Re-wire the meter with shielded cable according to requirement.
The Flow Rate Reading Fluctuates Significantly
High Voltage Instrument or High Frequency Interference close to the Flowmeter
Check if there is high voltage instrument or high frequency interference close to the flowmeter
Re-locate the flowmeter Heavy Vibration on the Pipeline
Locate the vibration on the pipeline by touching it with hand
Re-locate the flowmeter
Valve not Closed Properly
- Flow Leaks into the Pipe
Check pressure and check if valve is closed and sealed
Repair the valve The Gasket and the Pipe are not Concentric
Check the position of the gasket
Re-install the gasket
The Flowmeter Pipe Body and the Pipeline are not Concentric
Check if the flowmeter pipe body and the pipe line are not concentric
Re-install the meter
Straight Pipe Length not enough or the Inner Diameter of Flowmeter Pipe Body do not Match the Pipeline
Check the straight pipe length and the diameter of the pipeline
Re-locate the flowmeter Heavy Vibration on the Pipeline
Locate the vibration on the pipeline by touching it with hand
Tighten the pipeline where the flowmeter is installed
Fluid has not Filled the Pipeline fully
Check the fluid status and the location of the meter.
Re-locate the flowmeter
Two Phases Flow
Check if there is two­phase flow according to
If the fluid is liquid­solid two-phase flow,
www.GlobalTestSupply.com
Page 45
the pressure and temperature of the fluid.
need to install a filter at upstream of the flowmeter. If the fluid is liquid-gas two phase flow, need to install a getter at upstream of the flowmeter.
Transmitter Failure
Replace the transmitter with another transmitter of same type to check
Replace the transmitter
A Big Difference Between the Flow Reading and the Process Flow Rate
There is big difference between the flow reading and the process flow rate
Check the density compensation devices and the setting
Fix density compensation
Wrong Estimated Flow Rate Before use of the Meter
Use another flowmeter to confirm the actual flow rate
Setting incorrect
Check the settings of meter K factor, upper and lower limit of flow rate
Set the meter correctly
NOTE
Select the Code Setting, set C49=12. Press the “U-D” button to check the current energy of the vortex flow signal and vibration signal. E1 is the energy of the vortex flow signal; set the energy threshold value lower than the displayed value. E.1 is the energy of vibration; set the energy threshold value lower than the displayed value. Set above value in D017 (energy threshold of the vortex flow signal) and D018 (energy threshold of vibration), then set C49 back to 00
7.2 Self-Diagnostic Messages
Error code
Problem
Repair
Err-003
Temperature Sensor Disconnected
Check Temperature Sensor
Err-004
Pressure Sensor Disconnected
Check Pressure Sensor Err-005
About to Over Total Flow
This is a Reminder Message
Err-006
Display Value Over Limit
The Value is Over the Physical Limit of the Display
www.GlobalTestSupply.com
Page 46
Err-011
Superheated Steam Temperature is Over Limited
Reduce the Steam Temperature
Err-012
Superheated Steam Pressure is Over Limited
Reduce the Steam Pressure
Err-013
Button is Pressed and Hold for too Long Time
Check the Button Circuit Err-014
Reset Code Setting Failed
Check EEPROM
Err-015
Reset Digit Setting Failed
Check EEPROM
Err-016
Read Total Flow Error
Check EEPROM
Err-017
Temperature Calibration Setting is Wrong
Check the Record of Temperature Calibration
Err-018
Pressure Calibration Setting is Wrong
Check the Record of Pressure Calibration
Err-020
Flow Rate Limit Setting is Incorrect
Check the Flow Rate Limit Setting
Err-021
Temperature Limit Setting is Incorrect
Check the Temperature Limit Setting
Err-022
Pressure Limit Setting is Incorrect
Check the Pressure Limit Setting
Err-023
Communication Connection Error
Check the Communication Link
Err-024
Setting is Incorrect when Using aga_nx_19 to Calculate the Compressibility Factor
Check if the Setting for Compressibility Factor is Correct
Err-025
Frequency Output for Total Flow is over Limit
Reset the Total Flow Frequency Output Factor
Err-026
3V Power Source Failure
Check the Circuit Board
www.GlobalTestSupply.com
Page 47
TEKMATION LLC reserves the right to change the designs and/or materials of its products without notice. The contents of this publication are the property
of TEKMATION and cannot be reproduced by any other party without written permission. All rights reserved. Copyright © 2021 TEKMATION LLC
TEKMATION LLC DOC#TEK/PO/MNL/211110/1300C/r00.1
Tek-Trol is a fully owned subsidiary of TEKMATION LLC. We oer our customers a comprehensive range of products and solutions for process, power and oil & gas industries. Tek-Trol provides process measurement and control products for Flow, Level, Temperature & Pressure Measurement, Control Valves & Analyzer systems. We are present in 15 locations globally and are known for our knowledge, innovative solutions,
reliable products and global presence.
Flow | Level | Temperature | Pressure | Valves | Analyzers | Accessories | TekValSys
www.GlobalTestSupply.com
Loading...