Tek-Trol 1100A-2-1-025B-1-S-150-1-E, 1100A-2-1-040A-1-S-150-1-E, 1100A-2-1-050A-1-S-150-1-E, 1100A-2-1-015B-1-S-150-1-E, 1100A-2-1-015A-1-S-150-1-E User manual

...
Page 1
Technology Solutions
EK-C
T
OR 1100A
Coriolis Mass Flowmeter
Instruction Manual
Document Number: IM-1100A
www.GlobalTestSupply.com
Page 2
Technology Solutions
Instruction Manual Tek-Cor 1100A
Table of Contents
1 Safety Instructions ......................................................................................................... 3
1.1 Intended Use .....................................................................................................................3
1.2 Certification .......................................................................................................................3
1.3 Safety Instructions from the Manufacturer .........................................................................3
1.3.1 Disclaimer ................................................................................................................................ 3
1.3.2 Product Liability and Warranty ............................................................................................... 3
1.3.3 Information Concerning the Documentation .......................................................................... 3
1.4 Safety Precautions .............................................................................................................3
1.5 Packaging, Transportation and Storage ...............................................................................5
1.5.1 Packaging ................................................................................................................................. 5
1.5.2 Transportation......................................................................................................................... 5
1.5.3 Handling .................................................................................................................................. 5
1.5.4 Storage .................................................................................................................................... 6
1.5.5 Nameplate ............................................................................................................................... 6
2 Product Description ....................................................................................................... 8
2.1 Principle of Operation ........................................................................................................8
2.2 Models ..............................................................................................................................9
2.2.1 U-Shaped ................................................................................................................................. 9
2.2.2 Micro-bend Shaped ................................................................................................................. 9
2.2.3 Triangle Shaped ..................................................................................................................... 10
2.3 Technical Specifications ................................................................................................... 11
2.4 Flow Ranges..................................................................................................................... 12
2.4.1 Flow Ranges for Liquid (U-Shaped) ....................................................................................... 12
2.4.2 Flow Range for Liquid (Micro-bend Shaped) ......................................................................... 12
2.4.3 Flow Range for Gas ................................................................................................................ 12
2.4.4 Flow Range of Volume of Air under STP Conditions ............................................................. 13
2.4.5 Flow Rate Factor .................................................................................................................... 13
2.4.6 Accuracy, Basic Error and Repeatability ................................................................................ 14
2.4.7 Density Measuring ................................................................................................................ 14
2.4.8 Temperature Measuring ....................................................................................................... 14
2.5 Electrical Specifications .................................................................................................... 15
2.5.1 Current Output ...................................................................................................................... 15
2.5.2 Pulse Output .......................................................................................................................... 15
2.5.3 Low Flow Cut-off ................................................................................................................... 15
2.5.4 RS 485 Output ....................................................................................................................... 15
2.5.5 Power Consumption .............................................................................................................. 15
2.6
2.7 Dimensional Drawings ..................................................................................................... 17
2.8 Model Chart..................................................................................................................... 20
Environmental Specification ............................................................................................. 16
2.6.1
Environment Vibration .......................................................................................................... 16
2.6.2 Environment Temperature .................................................................................................... 16
2.6.3 Enclosure Grade .................................................................................................................... 16
2.7.1 Outline Dimensions for U-Shaped ......................................................................................... 17
2.7.2 Outline Dimensions for Micro-Bend Shaped ........................................................................ 18
2.7.3 Outline Dimensions for Triangle Shaped............................................................................... 19
2.7.4 Weights ................................................................................................................................. 19
www.GlobalTestSupply.com
1
Page 3
Technology Solutions
Instruction Manual Tek-Cor 1100A
1 Safety Instructions
1.1 Intended Use
This measuring device is intended to measure the mass flow rate. At the same time, the system also measures fluid density and fluid temperature. These parameters are then used to calculate other variables such as volumetric flow.
1.2 Certification
Tek-Cor 1100A has UL, CE, NTEP, and HMO certifications.
1.3 Safety Instructions from the Manufacturer
1.3.1 Disclaimer
The manufacturer will not be held accountable for any damage that happens by using its product, including, but not limited to direct, indirect, or incidental and consequential damages. Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale. The manufacturer has the right to modify the content of this document, including the disclaimer, at any time for any reason without prior notice, and will not be answerable in any way for the possible consequence of such changes.
1.3.2 Product Liability and Warranty
The operator shall bear authority for the suitability of the device for the specific application. The manufacturer accepts no liability for the consequences of misuse by the operator. Wrong installation or operation of the devices (systems) will cause the warranty to be void. The respective Terms and Conditions of Sale, which forms the basis for the sales contract shall also apply.
1.3.3 Information Concerning the Documentation
To prevent any injury to the operator or damage to the device it is essential to read the information in this document and the applicable national standard safety instructions. This operating manual contain all the information that is required in various stages, such as product identification, incoming acceptance and storage, mounting, connection, operation and commissioning, troubleshooting, maintenance, and disposal.
1.4 Safety Precautions
You must read these instructions carefully prior to installing and commissioning the device. These instructions are an important part of the product and must be kept for future reference. Only by observing these instructions, optimum protection of both personnel and the environment, as well as safe and fault-free operation of the device can be ensured.
3 Installation................................................................................................................... 21
3.1 Basic Requirement for Installation .................................................................................... 21
3.1.1 Position Selection .................................................................................................................. 21
3.1.2 Maximum Length of the Cable .............................................................................................. 21
3.1.3 Closing of Valve before Zero-Point Calibration ..................................................................... 22
3.2 Installation Direction ....................................................................................................... 22
3.2.1 Horizontal and Vertical Installation ....................................................................................... 22
3.2.2 Difficult Installation Location for Liquid and Gas Measurement .......................................... 23
3.2.3 Flow Direction ....................................................................................................................... 25
3.3 Mounting......................................................................................................................... 25
3.3.1 Sensor Mounting ................................................................................................................... 26
3.3.2 Transmitter Mounting ........................................................................................................... 27
3.4 Wiring ............................................................................................................................. 28
3.4.1 Basic Requirements ............................................................................................................... 28
3.4.2 Junction Box .......................................................................................................................... 28
3.4.3 Cable Connection .................................................................................................................. 29
3.4.4 Earthing ................................................................................................................................. 29
3.4.5 Power Line Wiring ................................................................................................................. 30
4 Electrical Connections .................................................................................................. 31
4.1 Basic Requirement ........................................................................................................... 31
4.2 Power Cable .................................................................................................................... 31
4.3 Current Output Wiring ..................................................................................................... 32
4.4 Pulse Output Wiring ......................................................................................................... 32
4.5 RS 485 Output Wiring ....................................................................................................... 33
5 Operation .................................................................................................................... 34
5.1 Display ............................................................................................................................ 34
5.2 Configuration Parameter .................................................................................................. 36
5.2.1 Measuring Unit ...................................................................................................................... 36
5.2.2 Reset Totalizer ....................................................................................................................... 36
5.2.3 Low Flow Cut-off ................................................................................................................... 36
5.2.4 Current Output ...................................................................................................................... 36
5.2.5 Pulse/Frequency Output ....................................................................................................... 37
5.2.6 RS 485 Output ....................................................................................................................... 37
5.2.7 First Menu ............................................................................................................................. 37
5.2.8 Oil and Water Content Analysis ............................................................................................ 38
5.2.9 Zero Calibration ..................................................................................................................... 38
5.3
5.4 Menu Tree ....................................................................................................................... 40
Calibration ....................................................................................................................... 39
5
.3.1 Zero Calibration ..................................................................................................................... 39
5.3.2 Flow Calibration .................................................................................................................... 39
6 Pressure Loss ............................................................................................................... 42
7 Troubleshooting ........................................................................................................... 44
7.1 Diagnostic Tool ................................................................................................................ 44
7.2 Checking the Sensor Connection ....................................................................................... 44
www.GlobalTestSupply.com
2
Page 4
Technology Solutions
Instruction Manual Tek-Cor 1100A
1 Safety Instructions
1.1 Intended Use
This measuring device is intended to measure the mass flow rate. At the same time, the system also measures fluid density and fluid temperature. These parameters are then used to calculate other variables such as volumetric flow.
1.2 Certification
Tek-Cor 1100A has UL, CE, NTEP, and HMO certifications.
1.3 Safety Instructions from the Manufacturer
1.3.1 Disclaimer
The manufacturer will not be held accountable for any damage that happens by using its product, including, but not limited to direct, indirect, or incidental and consequential damages. Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale. The manufacturer has the right to modify the content of this document, including the disclaimer, at any time for any reason without prior notice, and will not be answerable in any way for the possible consequence of such changes.
1.3.2 Product Liability and Warranty
The operator shall bear authority for the suitability of the device for the specific application. The manufacturer accepts no liability for the consequences of misuse by the operator. Wrong installation or operation of the devices (systems) will cause the warranty to be void. The respective Terms and Conditions of Sale, which forms the basis for the sales contract shall also apply.
1.3.3 Information Concerning the Documentation
To prevent any injury to the operator or damage to the device it is essential to read the information in this document and the applicable national standard safety instructions. This operating manual contain all the information that is required in various stages, such as product identification, incoming acceptance and storage, mounting, connection, operation and commissioning, troubleshooting, maintenance, and disposal.
1.4 Safety Precautions
You must read these instructions carefully prior to installing and commissioning the device. These instructions are an important part of the product and must be kept for future reference. Only by observing these instructions, optimum protection of both personnel and the environment, as well as safe and fault-free operation of the device can be ensured.
www.GlobalTestSupply.com
3
Page 5
Technology Solutions
Instruction Manual Tek-Cor 1100A
For additional information that are not discussed in this manual, contact the manufacturer
Warnings and Symbols Used
The following safety symbol marks are used in this operation manual and on the instrument.
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury
CAUTION
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices.
NOTE
Indicates that operating the hardware or software in this manner may damage it or lead to system failure.
www.GlobalTestSupply.com
4
Page 6
Technology Solutions
Instruction Manual Tek-Cor 1100A
1.5 Packaging, Transportation and Storage
1.5.1 Packaging
The original package consists of
1. Tek-Cor 1100A Coriolis Mass Flowmeter
2. Documentation
1.5.2 Transportation
Avoid impact shocks to the device and prevent it from getting wet during
transportation.
Verify local safety regulations, directives, and company procedures with respect
to hoisting, rigging, and transportation of heavy equipment.
Transport the product to the installation site using the original manufacturer’s
packing whenever possible.
www.GlobalTestSupply.com
5
Page 7
Technology Solutions
Instruction Manual Tek-Cor 1100A
1.5.3 Handling
Use a sling to lift and carry the instrument.
Do not lift the instrument using the flange bolt holes.
Do not lift the instrument on the transmitter housing and on the transmitter stem.
WARNING
The centre of gravity of the instrument is higher than the suspension points of the sling. There is a risk of injury if the instruments slips. Fulfil the local safety instructions while lifting the instrument
1.5.4 Storage
If this product is to be stored for a long period of time before installation, take the
following precautions:
Store your product in the manufacturer’s original packing used for shipping.
Storage location should conform to the following requirements:
o Free from rain and water o Free from vibration and impact shock o At room temperature with minimal temperature and humidity
variation
Before storing a used flowmeter remove any fluid from the flowmeter line completely. Properties of the instrument can change when stored outdoors.
www.GlobalTestSupply.com
6
Page 8
Technology Solutions
Instruction Manual Tek-Cor 1100A
1.5.5 Na
The nameplate lists the order number and other important information, such as design details and technical data
meplate
NOTE
Check the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
Name Plate of Tek-Cor 1100A Sensor
Name Plate of Tek-Cor 1100A Transmitter
www.GlobalTestSupply.com
7
Page 9
Technology Solutions
Instruction Manual Tek-Cor 1100A
2 Product Description
This section covers the reference and specification data, as well as ordering information.
2.1 Principle of Operation
The Coriolis measuring principle refers to the effect that a moving mass has on a body in a rotating frame of reference. The moving mass exerts an apparent force on the body, causing a deflection. This force is called the Coriolis force. It does not act directly on the body, but on the motion of the body. This principle is used in Coriolis flowmeters. A Coriolis flowmeter consists of two parallel tubes that are made to oscillate using a magnet. These oscillations are recorded by sensors fitted at the inlet and outlet of each tube. In a no-flow state, the oscillations are synchronized, since there is no mass exerting any force on the tubes. Any fluid flowing through the tubes generates Coriolis forces, causing the tubes to twist in proportion to the mass flow rate of the fluid.
www.GlobalTestSupply.com
8
Page 10
Technology Solutions
Instruction Manual Tek-Cor 1100A
2.2 Models
U-Shaped sensor (Size 1½” – 8”)
The Tek-Cor 1100A Series Coriolis Mass Flowmeters are available in the following three configurations:
2.2.1 U-Shaped
These flowmeters are comprised of two U-shaped tubes, a magnet and coil assembly, and sensors at the inlet and outlet of the tubes. Coriolis forces exerted by the flowing medium are used to determine the mass flow rate and density of the medium.
2.2.2 Micro-bend Shaped
These flowmeters are comprised of two U-shaped tubes in a casing with a considerably smaller radius than the conventional U-Shaped Coriolis flowmeters. The smaller radius ensures a more compact instrument with significantly lower pressure differential values compared to other flowmeters.
Micro-bend Shaped sensor (Size ½” – 8”)
www.GlobalTestSupply.com
9
Page 11
Technology Solutions
Instruction Manual Tek-Cor 1100A
2.2.3 Triangle Shaped
These flowmeters are the most compact in our range of Coriolis Mass Flowmeters, designed specifically to provide optimum performance in low-flow applications. It comprises a single flow tube which is considerably smaller in size than the conventional U-Shaped flowmeters.
Triangle Shaped sensor (Size ½” – 1”)
www.GlobalTestSupply.com
10
Page 12
Technology Solutions
Instruction Manual Tek-Cor 1100A
2.3 Technical Specifications
Service Liquids and gas Size U-Shaped: 1 ½" to 8"
Micro-bend Shaped: ½" to 8" Triangle Shaped: ½" to 1"
Accuracy 0.1% or 0.2% or 0.5% Density measuring range Density Accuracy 0.001 g/cm3 Density basic error ±0.02 lb/gal (±0.002 g/cm3) Density Repeatability ±0.01 lb/gal (± 0.001 g/cm3) Repeatability For 0.1% accuracy: ±0.05%
Temperature limits Direct mount: -58 °F to 257 °F (-50 °C to 125 °C)
Power requirements 18 to 28 VDC or 85 to 220 VAC Output signal 4-20 mA and pulse, Optional: HART or Modbus RS485 Process Connection As per DIN or ANSI Enclosure rating NEMA 4X (IP67) Agency approvals UL, CE, NTEP, and HMO
2 to 30 lb/gal (0.2 to 3 g/cm3)
For 0.2% accuracy: ±0.1% For 0.5% accuracy: ±0.25%
Remote mount: -58 °F to 392 °F (-50 °C to 200 °C)
www.GlobalTestSupply.com
11
Page 13
Technology Solutions
Instruction Manual Tek-Cor 1100A
2.4 Flow Ranges
Size (inch)
Allowable Flow
Normal Flow Range
Normal Flow Range
(lb/min)
1 ½”
12 - 1175
74 - 1175
56 - 1175
2”
19 - 1837
110 - 1837
92 - 1837
3”
52 - 5144
221 - 5144
203 - 5144
4”
74 - 7340
552 - 7348
441 - 7348
6”
184 - 18371
1287 - 18371
1103 - 18371
8”
368 - 36743
2572 - 36743
2573 - 36743
Size (inch)
Allowable Flow
Normal Flow Range
Normal Flow Range
(lb/min)
½”
1 – 110
8 – 110
6 – 110
1”
3 – 293
23 – 293
15 – 293
1 ½”
9 – 881
89 – 881
42 – 881
2”
19 – 1837
184 – 1837
92 – 1837
3”
30 – 4409
294 – 4409
294 – 4409
4”
56 – 7348
552 – 7348
368 – 7348
6”
184 – 18371
1838 – 18371
919 – 18371
8”
368 – 36743
3674 – 36743
18372 – 36743
Size (Inch)
Measurable Flow Range
(lb/min)
Flow Range with 0.5%
(lb/min)
1/2”
1 - 110
3 - 110
1”
2 - 293
8 - 293
1 ½”
12 - 1176
30 - 1176
2”
19 - 1837
46 - 1837
3”
26 - 5144
129 - 5144
4”
37- 7348
184 - 7348
6”
92 - 18371
460 - 18371
2.4.1 Flow Ranges for Liquid (U-Shaped)
Range
for Accuracy 0.1%
(lb/min)
2.4.2 Flow Range for Liquid (Micro-bend Shaped)
Range
(lb/min)
for Accuracy 0.1%
(lb/min)
(lb/min)
for Accuracy 0.2%,
0.5%
for Accuracy 0.2%,
0.5%
2.4.3 Flow Range for Gas
www.GlobalTestSupply.com
12
Page 14
Technology Solutions
Instruction Manual Tek-Cor 1100A
2.4.4 Flow Range of Volume of Air under STP Conditions
Size (Inch)
Initial flow
Initial flow
Flow range with
Flow range with
1/2”
12.5
8
62.5 – 2500
40 – 1585
1”
33.33
21
166.7 – 6666.7
106 – 4226
1 ½”
133.33
84
666.7 – 26666.7
423 – 16906
2”
208.33
132
1041.7 – 41666.7
661 – 26416
3”
583.33
370
2916.7 – 116666.7
1849 - 73966
4”
833.33
528
4166.7 – 166666.7
2642 - 105665
6”
2083.33
1321
10416.7 – 416666.7
6604 – 264164
Size (inch)
½”
1”
1 ½”
2”
3”
4”
6”
8”
Flow rate factor
0.362
1.558
1.634
3.535
8.338
15.89
26.15
58.84
2.4.5 Flow Rate Factor
The flow rate of the Tek-Cor 1100A must be checked according to the formula below:
Medium flow rate =
(Nm3/h)
(Scfm)
accuracy 0.5%
(Nm3/h)
     
  
accuracy 0.5%
(Scfm)
NOTE
The gas flow velocity is usually much higher than liquid when measured by a flowmeter, so there is noise caused by gas medium and tube wall of flowmeter under a high-speed gas flow and if the noise becomes loud enough, the signal of flowmeter will be influenced, so use the Tek-Cor 1100A Coriolis flowmeter for gaseous medium at velocity less than 1/3rd of the sound velocity (360 fps).
Use Tek-Cor 1100A for gas with a pressure less than 29 PSI (0.2 MPa).
www.GlobalTestSupply.com
13
Page 15
Technology Solutions
Instruction Manual Tek-Cor 1100A
2.4.6 Accuracy, Basic Error and Repeatability
Accuracy
0.1%
0.2%
0.5%
Basic Error
±0.10%
±0.20%
±0.50%
Repeatability
±0.05%
±0.10%
±0.25%
Accuracy is calculated based on the water measurement under the conditions of 68oF to 77oF (20oC to 25oC) and 14.5 PSI to 29 PSI (0.1 MPa to 0.2 MPa)
Density Range
2 – 30 lb/gal (0.2 to 3 g/cm3)
Basic Error
±0.02 lb/gal (±0.002 g/cm3)
Repeatability
±0.01 lb/gal (±0.001 g/cm3)
Temperature Range
Direct Mount
-58 oF to 257 oF (-50 oC to 125 oC)
Remote Mount
-58 oF to 392 oF (-50 oC to 200 oC)
Basic Error
≤±32 oF (≤±1 oC)
2.4.7 Density Measuring
2.4.8 Temperature Measuring
www.GlobalTestSupply.com
14
Page 16
Technology Solutions
Instruction Manual Tek-Cor 1100A
2.5 Electrical Specifications
Output Range
4-20 mA
Resolution
0.000244 mA
Basic Error
0.2% FS
Temperature Influence
±0.005 FS/oF
External resistor should be 250 Ω to 600 Ω
Output Range
0 – 10 kHz
Resolution
0.152 kHz
Basic Error
0.075% FS
Temperature Influence
±0.005 FS/oF
Capability of output range is 12 kHz
2.5.1 Current Output
A passive current output of 4-20mA can be configured to indicate the mass flow, volume flow, or density.
2.5.2 Pulse Output
Active pulse output can be configured to indicate the mass flow, volume flow, or density.
2.5.3 Low Flow Cut-off
When the mass flow value is lower than the value of low flow cut-off, the Tek-Cor 1100A Coriolis Mass Flowmeter shows the output of flow rate as zero, and the totalizer stops accumulating. The value of low flow cut-off is usually set to be 1% of the maximum flow rate.
2.5.4 RS 485 Output
RS485 output is compatible with the RTU mode of the Modbus protocol. For details, contact Tek-Trol.
2.5.5 Power Consumption
The normal power consumption for flowmeter is 10 W, and the maximum power consumption is 15 W.
www.GlobalTestSupply.com
15
Page 17
Technology Solutions
Instruction Manual Tek-Cor 1100A
2.6 Environmental Specification
Frequency Range
10 to 2000 Hz
Maximum Acceleration
2 g
Maximum Vibration cycles
50
Working Temperature
-40 oF to 131 oF (-40 oC to 55 oC)
Storage Temperature
-68 oF to 158 oF (-20 oC to 70 oC)
2.6.1 Environment Vibration
2.6.2 Environment Temperature
2.6.3 Enclosure Grade
The enclosure grade for the Tek-Cor 1100A Coriolis Mass Flowmeter is IP67.
www.GlobalTestSupply.com
16
Page 18
Technology Solutions
Instruction Manual Tek-Cor 1100A
2.7 Dimensional Drawings
2.7.1 Outline Dimensions for U-Shaped
Size
1 ½”
(0.13 ft)
2”
(0.16 ft)
3”
(0.26 ft)
4”
(0.32 ft)
6”
(0.49 ft)
8”
(0.65 ft)
L
L1
H
H1
≤300# (4 MPa) ≥600# (6.3 MPa) Integrated Remote
20.47”
(1.70 ft)
21.96”
(1.83 ft)
30.70”
(2.55 ft)
36.22”
(3.01 ft)
43.30”
(3.60 ft)
53.70”
(4.47 ft)
21.53”
(1.79 ft)
23.14”
(1.92 ft)
31.81”
(2.65 ft)
37.32”
(3.11 ft)
44.88”
(3.74 ft)
55.51”
(4.62 ft)
18.5”
(1.54 ft)
21.65”
(1.80 ft)
27.95”
(2.32 ft)
33.85”
(2.82 ft)
41.33”
(3.44 ft)
45.66”
(3.80 ft)
25.98”
(2.16 ft)
27.95”
(2.32 ft)
40.94”
(3.41 ft)
44.88”
(3.74 ft)
59.84”
(4.98 ft)
65.15”
(5.42 ft)
11.02”
(0.91 ft)
11.41”
(0.95 ft)
12.59”
(1.04 ft)
13.77”
(1.14 ft)
14.96”
(1.24 ft)
16.53”
(1.37 ft)
8.26”
(0.68 ft)
8.66”
(0.72 ft)
9.84”
(0.82 ft)
11.02”
(0.91 ft)
12.20”
(1.01 ft)
13.77”
(1.14 ft)
www.GlobalTestSupply.com
17
Page 19
Technology Solutions
Instruction Manual Tek-Cor 1100A
2.7.2 Outline Dimensions for Micro-Bend Shaped
L
Size
≤300# (4 MPa) ≥600# (6.3 MPa) Integrated Remote
L1
H1
H
½”
(0.04 ft)
1”
(0.08 ft)
1 ½”
(0.13 ft)
2”
(0.16 ft)
3”
(0.26 ft)
4”
(0.32 ft)
6”
(0.49 ft)
8”
15.74”
(1.31 ft)
19.68”
(1.64 ft)
23.62”
(1.96 ft)
31.49”
(2.62 ft)
35.43”
(2.95 ft)
44.48”
(3.70 ft)
55.51”
(4.62 ft)
70.86”
16.29”
(1.35 ft)
21.10”
(1.75 ft)
24.96”
(2.08 ft)
32.59”
(2.71 ft)
36.53”
(3.04 ft)
45.51”
(3.79 ft)
57.08”
(4.75 ft)
72.59”
11.02”
(0.91 ft)
14.17”
(1.18 ft)
18.11”
(1.50 ft)
25.19”
(2.09 ft)
27.55”
(2.29 ft)
33.85”
(2.82 ft)
47.24”
(3.93 ft)
57.08”
7.24”
(0.60 ft)
9.84”
(0.82 ft)
11.81”
(0.98 ft)
16.14”
(1.34 ft)
19.29”
(1.60 ft)
25.98”
(2.16 ft)
35.43”
(2.95 ft)
46.06”
11.41”
(0.95 ft)
11.81”
(0.98 ft)
12.20”
(1.01 ft)
12.59”
(1.04 ft)
13.77”
(1.14 ft)
14.56”
(1.21 ft)
15.74”
(1.31 ft)
16.53”
8.66”
(0.72 ft)
9.05”
(0.75 ft)
9.44”
((0.78 ft)
9.84”
(0.82 ft)
11.02”
(0.91 ft)
11.41”
(0.95 ft)
12.99”
(1.08 ft)
13.77”
(0.65 ft)
(5.90 ft)
(6.04 ft)
(4.75 ft)
(3.83 ft)
www.GlobalTestSupply.com
(1.37 ft)
(1.14 ft)
18
Page 20
Technology Solutions
Instruction Manual Tek-Cor 1100A
2.7.3 Outline Dimensions for Triangle Shaped
Size (inch)
½”
1”
1 ½”
2”
3”
4”
6”
8”
U-shaped, lb (kg)
- - 67
(30 kg)
89
(40 kg)
220
(100 kg)
419
(190 kg)
717
(325 kg)
1182
(536 kg)
Micro-bend shaped, lb (kg)
27
(12 kg)
33
(15 kg)
55
(25 kg)
84
(33 kg)
172
(78 kg)
298
(135 kg)
629
(285 kg)
948
(430 kg)
Triangle shaped, lb (kg)
29
(13 kg)
38
(17 kg)
- - - - -
-
Size
½”
(0.04 ft)
1”
(0.08 ft)
2.7.4 Weights
L
L1
H
H1
≤300# (4 MPa) ≥600# (6.3 MPa) Integrated Remote
7.08”
(0.59 ft)
7.87”
(0.65 ft)
7.63”
(0.63 ft)
7.96”
(0.81 ft)
13.77”
(1.14 ft)
18.03”
(1.50 ft)
11.41”
(0.95 ft)
15.74”
(1.31 ft)
10.23”
(0.85 ft)
11.02”
(0.91 ft)
7.48”
(0.62 ft)
8.26”
(0.68 ft)
www.GlobalTestSupply.com
19
Page 21
Technology Solutions
Instruction Manual Tek-Cor 1100A
2.8 Model Chart
www.GlobalTestSupply.com
20
Page 22
Technology Solutions
Instruction Manual Tek-Cor 1100A
3 Installation
Cable Model
Cable Specification
Max. Length
Special Nine-Core Cable
Special
985 ft. (300 m)
Current Power Line
18AWG (0.8 mm2)
985 ft. (300 m)
RS485 Communication Cable
22AWG (0.35 mm2)
985 ft. (300 m)
This section covers instructions on installation and commissioning. Installation of the device must be carried out by trained, qualified specialists authorized to perform such works.
CAUTION
When removing the instrument from hazardous processes, avoid direct contact with the fluid and the meter
All installation must comply with local installation requirements and local electrical code
3.1 Basic Requirement for Installation
3.1.1 Position Selection
The flowmeter should be placed away from interference sources (such as a pump), which may cause vibration in the pipe. If the flowmeters are used in series along the same line, care must be taken to avoid the mutual influence due to vibration resonance. The distance between sensors should be more than 7 ft. (2 m).
When installing the sensor, pay attention to the expansion and contraction of the process pipeline due to temperature changes. It is strongly recommended that the sensor not be installed near an expansion joint of the process pipeline. Otherwise, the pipe expansion and contraction of the pipeline will cause transverse stress,
which will affect the flowmeter’s zero and will affect the measurement accuracy.
The flowmeter should be placed at least 16 ft. (5 m) away from industrial electromagnetic interference sources, such as large power motors and transformers.
The sensor should be placed in the position where its measuring tube is always filled with fluid and pressure is maintained at the outlet, thus it should be placed in a position lower than the pipeline.
Tek-Cor 1100A does not require straight pipe upstream or downstream. However, if more than one mass flow transmitters are installed in the same pipe, ensure the length of pipe between any two flowmeters is more than 2 meters.
3.1.2 Maximum Length of the Cable
www.GlobalTestSupply.com
21
Page 23
Technology Solutions
Instruction Manual Tek-Cor 1100A
3.1.3 Closing of Valve before Zero-Point Calibration
It is necessary to do zero-point calibration after the installation is completed. Close the downstream stop valve before zero-point calibration, and then close the upstream stop valve.
3.2 Installation Direction
3.2.1 Horizontal and Vertical Installation
The Tek-Cor 1100A should be installed in the orientation that can ensure the measuring tube is filled with the medium being measured.
For horizontal installation, the measuring tube should be installed downside of the pipeline when the process medium is liquid or slurry (Picture 1); upside of the pipeline when the process medium is gas (Picture 2). For vertical installation, the measuring tube should be installed besides the pipeline when the process medium is liquid or slurry or gas (Picture 3).
www.GlobalTestSupply.com
22
Page 24
Technology Solutions
Instruction Manual Tek-Cor 1100A
3.2.2 Difficult Installation Location for Liquid and Gas Measurement
Pipe reduction or extension should be avoided directly before or after the
The flowmeter must not be mounted at the highest point of the tubing (A) if gas bubbles are expected, or at the lowest point (B) if solid particles are expected. Also, the meters must not be mounted in a drop-line near the open end (C), to avoid the flowmeter becoming empty.
NOTE
flowmeter
Avoid any control valves, or orifices, or any sound generator near the sensor
www.GlobalTestSupply.com
23
Page 25
Technology Solutions
Instruction Manual Tek-Cor 1100A
Avoid direct installation at the sensor side as this can cause measurement errors.
For installing 6” or larger size flowmeter, support the sensors using the rubber
connectors as the buffer.
buffer
www.GlobalTestSupply.com
24
Page 26
Technology Solutions
Instruction Manual Tek-Cor 1100A
3.2.3 Flow Direction
Transmitter
Sensor
The flow arrow in front of the flowmeter indicates the direction of flow, so install the Tek-Cor 1100A accordingly. Otherwise, the transmitter may not display the mass flow accurately. If the process medium is liquid or slurry, the flow direction is down­to-up; if the process medium is gas, the flow direction can be either down-to-up or up-to-down.
3.3 Mounting
There are 2 types of mounting that can be done on Tek-Cor 1100A Coriolis Mass Flowmeter:
1. Sensor Mounting
2. Transmitter Mounting
www.GlobalTestSupply.com
25
Page 27
Technology Solutions
Instruction Manual Tek-Cor 1100A
3.3.1 Sensor Mounting
If the medium contains solid particles, mount the meter as shown in Position A, and in all other cases mount the meter as shown in Position B. Vertical Mounting would be the recommended, if the medium contains gas bubbles and solid particles.
Liquid Gas
Recommended Horizontal Installation Vertical Mounting
Liquid Gas
Recommended Horizontal Installation Vertical Mounting
www.GlobalTestSupply.com
26
Page 28
Technology Solutions
Instruction Manual Tek-Cor 1100A
3.3.2 Transmitter Mounting
The Tek-Cor 1100A can be mounted directly or remotely with the help of a nine-core cable.
Install the sensor and the transmitter in the pipeline. The transmitter can be mounted with 90° rotation according to the requirement of sensor installation as shown in the picture above.
www.GlobalTestSupply.com
27
Page 29
Technology Solutions
Instruction Manual Tek-Cor 1100A
3.4 Wiring
3.4.1 Basic Requirements
If the sensor of the Tek-Cor 1100A Coriolis Mass Flowmeter is installed integrally with the transmitter, it is easy to power up the transmitter. If the sensor of Tek-Cor 1100A is installed separately from the transmitter then the transmitter is connected with the sensor through special nine-core cable.
3.4.2 Junction Box
If the sensor and the transmitter are installed separately, ensure that the sensor and transmitter have been respectively matched with junction box for connecting the special nine-core cable.
CAUTION
Cut off power before connecting the cables. The power input voltage must match the voltage specified in the junction box of the transmitter. The earth connector must be well connected with earth wire to ensure its intrinsic safety performance.
www.GlobalTestSupply.com
28
Page 30
Technology Solutions
Instruction Manual Tek-Cor 1100A
3.4.3 Cable Connection
Line No.
Line Colour
Function
1
Brown
Left coil +
2
Red
Left coil -
3
Orange
Right coil +
4
Yellow
Right coil -
5
Green
Driving coil +
6
Blue
Driving coil -
7
Gray
Temperature +
8
White
Temperature -
9
Black
Temperature compensation
A nine-core cable is used when the sensor and the transmitter are connected separately. The following picture shows a nine-core cable.
3.4.4 Earthing
The sensor and the transmitter must be earthed correctly, otherwise there can be a measurement error and the sensor may not work. If the pipeline is connected to the ground, the transmitter can be earthed through the pipeline; if the pipeline is not connected with the ground, the transmitter should be earthed independently. The grounding terminal is provided at the external part of the transmitter and must be grounded when put into service.
www.GlobalTestSupply.com
29
Page 31
Technology Solutions
Instruction Manual Tek-Cor 1100A
For direct mounting, protective earth terminal is connected as shown below:
Transmitter’s side
For remote mounting, protective earth terminal is connected as shown below:
Transmitter’s side
3.4.5 Power Line Wiring
The transmitter can be supplied with 18 to 28 VDC (standard) or 85 to 220 VAC
2
(optional). The power line more than 0.00124 in2 (0.8 mm
) is recommended and the maximum length of power line should be 985 ft. (300 m). For Tek-Cor 1100A of size 6” (DN150 mm) or larger, the drive-amplifier needs to be supplied with separate power.
www.GlobalTestSupply.com
30
Page 32
Technology Solutions
Instruction Manual Tek-Cor 1100A
4 Electrical Connections
85 to 220 VAC
Power consumption: Normal 10 W, Maximum 15 W
18 to 28 VDC
Power consumption: Normal 10 W, Maximum 15 W
4.1 Basic Requirement
The transmitter can be connected to the AC or the DC power.
4.2 Power Cable
The power cable should be a two-core cable and 20-gauge minimum. The maximum length of the power cable is 985 ft. (300 m).
AC Power Wiring
DC Power Wiring
www.GlobalTestSupply.com
31
Page 33
Technology Solutions
Instruction Manual Tek-Cor 1100A
4.3 Current Output Wiring
4-20 mA output can be configured to display mass flow, volume flow, density or water ratio.
The cable should be two-core cable and 24-gauge minimum.
The factory default current output is passive current output.
Current Output Wiring
The outer wiring of passive current output is as the figure show below:
4.4 Pulse Output Wiring
Active pulse output can be configured to mass flow, volume flow, density or water ratio.
The cable should be two-core cable and 24-gauge minimum. The maximum length of output line is 492 ft. (150 m).
Pulse Output Wiring
www.GlobalTestSupply.com
32
Page 34
Technology Solutions
Instruction Manual Tek-Cor 1100A
4.5 RS 485 Output Wiring
RS 485 output is compatible to RTU mode of MODBUS protocol. The maximum length of output line is 985 ft. (300 m).
RS 485 Output Wiring
www.GlobalTestSupply.com
33
Page 35
Technology Solutions
Instruction Manual Tek-Cor 1100A
5 Operation
No.
Notes
1
E key: enter
2
key: move cursor or return
3
key: Page down
4
Light indication for working status
5
Two line LED display
Indication of title
Indication of numeric
Indication of units
This section covers operation techniques and guidelines.
Data setting can be done by HART-Communication or with the 3 keys on the front panel. The following section describes how to use the three panel keys.
5.1 Display
Components of Tek-Cor 1100A display:
LED with 2-line indication
3 photoelectric keys
Use the operation panel of the transmitter to set the configuration parameters such as the zero calibration, cut-off value of low flow and output range of current frequency, etc.
www.GlobalTestSupply.com
34
Page 36
Technology Solutions
Instruction Manual Tek-Cor 1100A
Key
Measurement State
Menu State
Function State
Data State
E
Show the Measure-
Next Menu
Select Function
Change
Character
Return to the Previous
Return to the
State
Select Function
Move the Cursor
Enter the Menu
Confirm and Save
Save the Input,
Back to the Function Menu
ment Results and State on Page1/2/3. Page Down to Menu State.
Number
Change Unit
Change
Screen
Upper-Level Menu.
to the Right Press this Key to Return to the Measurement
the Function
then Go
NOTE
Operation point of the photoelectric key is located right behind the glass panel.
It is better to operate the photoelectric key in vertical direction, rather than
horizontal direction.
www.GlobalTestSupply.com
35
Page 37
Technology Solutions
Instruction Manual Tek-Cor 1100A
5.2 Configuration Parameter
Review or set the configuration parameters according to the following indications
(Press to go to the next page and press to move the position of cursor)
WARNING
If you forget your password, you must call a Tek-Trol engineer to reset it.
5.2.1 Measuring Unit
5.2.2 Reset Totalizer
5.2.3 Low Flow Cut-off
5.2.4 Current Output
Set the Flow Rate for 20 mA
www.GlobalTestSupply.com
36
Page 38
Technology Solutions
Instruction Manual Tek-Cor 1100A
Set Current Output Signal
5.2.5 Pulse/Frequency Output
Set Pulse Equivalent
Set Pulse Signal
5.2.6 RS 485 Output
5.2.7 First Menu
The screen automatically displays the content chosen in “First menu” if no key operation is being done for 128 seconds.
www.GlobalTestSupply.com
37
Page 39
Technology Solutions
Instruction Manual Tek-Cor 1100A
5.2.8 Oil and Water Content Analysis
5.2.9 Zero Calibration
www.GlobalTestSupply.com
38
Page 40
Technology Solutions
Instruction Manual Tek-Cor 1100A
5.3 Calibration
The Tek-Cor 1100A Coriolis Mass Flowmeter does not need field calibration because it is calibrated before delivery. Each Tek-Cor 1100A has its own instrumental coefficient, including one flow coefficient and four density coefficients, which are high density D1, high period K1, low density D2, and low period K2. These are present on the sensor. The direct mount instrument comes preset with the instrument coefficients and the user cannot change this.
5.3.1 Zero Calibration
Zero calibration provides the reference point for the flowmeter. It is necessary to conduct zero calibration whenever the Coriolis Mass Flowmeter installation is performed. For zero calibration, perform the following steps after installation-
1. Warm-up the flowmeter for 30 minutes.
2. After the warm-up time is complete, continue the flow through the flowmeter till the temperature of the flowmeter is same as the working temperature of the fluid.
3. After the flowmeter has reached the required temperature, close the downstream valve and make the fluid pass through the flowmeter under normal temperature, density, and pressure and then close the upstream valve to ensure that the sensor is full of liquid during the process of zero calibration.
4. Do the zero calibration of the flowmeter.
NOTE
Each zero calibration takes 30 seconds and must repeated at least 10 times.
5.3.2 Flow Calibration
The mass measured by Tek-Cor 1100A is calculated from the multiplication of the
detected signals’ time difference between two circuits and flow calibration factor. If
the accuracy is not up to grade after long-term service, modify the flow calibration factor according to the following formula:
=
=
× ( +
×
where,
K1 = New flow calibration factor K0 = Old flow calibration factor M = Total mass flow of the mass flowmeter Mt = Total mass flow of tested flowmeter
(
−
)
)
www.GlobalTestSupply.com
39
Page 41
Technology Solutions
Instruction Manual Tek-Cor 1100A
5.4 Menu Tree
www.GlobalTestSupply.com
40
Page 42
Technology Solutions
Instruction Manual Tek-Cor 1100A
www.GlobalTestSupply.com
41
Page 43
Technology Solutions
Instruction Manual Tek-Cor 1100A
6 Pressure Loss
Pressure loss in Tek-Cor 1100A Coriolis Mass Flowmeter can be checked on the following Pressure Loss Charts (including pressure loss, flow, and viscosity parameters). When the viscosity is between two adjacent pressure loss lines, the pressure loss can be calculated with following formula:
P = 


 󰇛  󰇜
∆P1 = Pressure at the upstream side ∆P2 = Pressure at the downstream side ∆P = Total differential pressure ∆     
Following is the pressure loss chart for different flowmeter sizes:
www.GlobalTestSupply.com
42
Page 44
Technology Solutions
Instruction Manual Tek-Cor 1100A
www.GlobalTestSupply.com
43
Page 45
Technology Solutions
Instruction Manual Tek-Cor 1100A
7 Troubleshooting
LED Indication
Working Condition
Green light
Normal operation
Red light
Error
Loop
Line Color
Sensor Port
Normal Resistance
Range
Left Coil
Brown, Red
1,2
(60-75) Ω
Right Coil
Orange, Yellow
3,4
(60-75) Ω
Drive Coil
Blue, Green
5,6
(6-30) Ω
Temperature
Gray, White
7,8
(75-175) Ω
Temperature
Gray, Black
7,9
(75-175) Ω
This section provides troubleshooting techniques for most common operating problems.
7.1 Diagnostic Tool
You can detect the fault in the flowmeter with the help of the LED indicator and LCD displays. Different colours of the LED lights indicate the working condition of the flowmeter, while the LCD displays shows the self-diagnostic error indication of the transmitter, which is useful for detecting malfunctions.
In addition, use handheld digital multi-meter when testing the static resistance values and cables of the sensor.
7.2 Checking the Sensor Connection
When testing a malfunction of the flowmeter, first test the coils resistance according to the Table below and check if their values are within the normal range.
www.GlobalTestSupply.com
44