Teguar TMT-4391-08 Series Startup Manual

Medical Computer TMT-4391-08 Series Startup Manual
TMT-4391-08 Appearance
Left: Front View Right: Rear View
For more information on this and other Teguar products, please visit our website at:
http://www.teguar.com
For technical support and service, please visit our support website at:
http://www.teguar.com/support
This manual is for TMT-4391-08 Series 3rd Edition
Description of Parts
A: Front Camera B: Power LED C: Light Sensor D: Speaker E: micro-SIM & micro-SD Card Door F: Programmable Key (Default Barcode scanner) G: Pogo Pin (USB3.0) H: Micro USB Door I: Micro HDMI Door J: Power Button K: Volume Key (Up/Down) L: Audio Jack (Headset Combo) M: Extension Module N: NFC O: Rear Camera with LED Flash P: Battery Cover
Easy Setup
Power On Computer
1. Model: TMT-4391-08 Rating 5 Vdc / 3.0A
2. This product is intended to be supplied by a UL Listed Power Adapter (Phihong / MM18M-59A) or DC power source, rated 5 Vdc, 3A min. for model TMT-4391-08, and Tma 45 degree C. Please contact Teguar for further information and assistance.
3. The product is equipped with shipping mode for battery protected and power saving, please charge the embedded battery of the computer: Connect the Micro USB adaptor on the TMT-4391-08 computer. (“H” on the description of parts.) Please charge for at least one hour when you use this computer for the first time.
4. Push the Power button (“J” on description of parts.) for 2~3 seconds to start the computer.
5. The product is equipped with a 8” Capacitive T ouch Panel. Use finger , capacitive stylus, or nitrile gloves to touch the following Active Area to operate the computer.
6. The product is equipped with one LED indicator for battery status. The following shows LED status for different power states: a) The orange LED blinks when internal battery
capacity is below 10% to warn the user to charge.
b) The Blue LED lights up when internal battery is
being charged.
c) The Green LED light s up when internal b attery
has been fully charged.
7. While the computer is running, push the power button for 1 second will disable LCD backlight for power saving. Push the power button again will enable the backlight again; press the power button for 10 seconds while computer is running, the system will shut down.
8. Environment:
Operating Temperature: -10°C ~ +50 °C Operating Humidity : 10% ~ 90%@30°C
non-condensed
Storage/Transportation Temperature:
-40°C ~ +80°C (Internal lab test)
Storage/Transportation Humidity :
5% ~ 95%@30°C non-condensed
Humidity consider to 25°C, 48 hrs Number and type of Means of Protection :
MOPP Altitude 3000m
Atmospheric pressure 700-1013 hPa (for
operating); 500-1013 hPa (for Storage/Transportation)
IP level : IPX0
9. There will be “TEGUAR” l ogo upside down while first boot up. This scenario will be disappear after 2
nd
boot up and won’t influence functions and
usage.
Intended Use
The TMT-4391-08 is intended for integration with hospital system. It is designed for general purpose for hospital environment. For data collection and display for reference. It should not be used for life-supporting system.
EMC Table
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The model TMT-4391-08 is intended for use in an electromagnetic environment as specified below. The customer or the user of the TMT-4391-08 should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environmental Guidance
The model TMT
-4391-08 uses
RF RF emissions
energy only for its internal func-
Group 1
tion. Therefore, its RF emissions
CISPR 11
are very low and are not likely to
cause any interference in nearby
electronic equipment.
RF emissions
Class B
The model TMT
-4391-08 is
suitable
CISPR 11
for use in all establishments,
Harmonic emissions
Class A
including domestic establishments
IEC 61000-3-2
and those directly connected to
the public low-voltage power sup-
Voltage fluctuations/
ply network that supplies buildings
flicker emissions
Not applicable
used for domestic purposes.
IEC 61000-3-3
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the TMT-4391-08
TMT-4391-08 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the model TMT-4391-08 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the model TMT-4391-08 as recommended below, according to the maximum output power of the communications equipment.
Rated Maximum
Output Power of
Separation Distance According to Frequency of Transmitter
Transmitter
m
W
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d = 1,2 P
d = 1,2 P
d = 2,3 P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation in the table above applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
TMT-4391-08 is intended for use in the electromagnetic environment specified below. The customer or the user of the model TMT-4391-08 sho uld assur e that it is used in such an environment.
Immunity Test
IEC 60601 Test
Compliance
Electromagnet ic Environmental
Level
Level
Guidance
Electrostatic dis-
Floors should be wood, con-
±8 kV
contact
±8 kV
contact
crete or ceramic tile. If floors
charge (ESD)
are covered with synthetic
±
15 kV
air
±15 kV air
IEC 61000-4-2
material, the relative humidity
should be at least 30%.
Electrical fast
±
2 kV for
power
±
2 kV for
power
supply lines
supply lines
Main power quality should be
transient/bur st
that of a typical commercial or
±1 kV for
input/out-
±1 kV for
input/
IEC 61000-4-4
hospital environment.
put lines
output lines
±
1 kV
line(s) to
±
1 kV
line(s) to
Main power quality should be Surge
line(s)
line(s)
that of a typical commercial or
IEC 61000-4-5
±
2 kV
line(s) to
±
2 kV
line(s) to
hospital environment.
earth
earth
<5%
UT
(>95% dip
<5% UT
(>95% dip in
in UT) for 0,5 cycle
UT) for 0,5
Main power quality should be cycle
40% UT
that of a typical commercial or
Interruptions and
(60% dip in UT)
40%
UT
(60%
hospital environment. If the voltage varia-
for 5 cycles
dip in UT) for 5
user of the model TMT-4391-08
tions on power
cycles
requires continued operation
supply input lines
70% UT
during main power interrup-
(30% dip in UT)
70%
UT
(30%
tion, it is recommended that
IEC 61000-4-11
for 25 cycles
dip in UT) for
the model TMT
-4391-08
be 25 cycles
powered from an uninterrupt-
<5% UT
<5% UT
ible power supply or a battery.
(>95% dip in UT)
for 5 sec
(>95% dip in
UT) for 5 sec
Power frequency
Power frequency magnetic
fields should be at levels
(50/60 Hz)
3 A/m
3 A/m
characteristic of a typical
magnetic field
location in a typical commer-
IEC 61000-4-8
cial or hospital environment.
NOTE UT is the A.C. main voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The model TMT-4391-08 is intended for use in the electromagnetic environment specified below. The customer or the user of the model TMT-4391-08 should assure that it is used in such an environment.
Immunity
IEC 60601 Test
Compliance
Electromagnet ic Environmental
test
Level
Level
Guidance
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the model
TMT-4391-08, including cables, than
the recommended separation dis-
tance calculated from the equation
applicable to the frequency of the
transmitter.
Recommende d Separation Dis-
tance
d = 1,2 P
d =
1,2 P
80 MHz to 800 MHz
Conducted
3 Vrms
RF
d = 2,3 P 800 MHz to 2,5 GHz
150 kHz
to
80
IEC 61000-4-6
MHz
Vrms
where P is the maximum output
Radiated RF
3 V/m
V/m
power rating of the transmitter in
watts (W) according to the transmit-
80 MHz to 2,5
IEC 61000-4-3
ter manufacturer and d is the rec-
GHz
ommended separation distance in
meters (m).
Field strengths from fixed RF trans-
mitters, as determined by an elec-
tromagnetic site survey,
a
should
be
less than the compliance level in
each
frequency
range. b
Interference may occur in t he vicin-
ity of equipment marked with the fol-
lowing symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.