Tegam R1M-A, R1M-AR Operation And Maintenance Manual

Revised August, 2004
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
R1M-A, R1M-AR
MODEL R1M-A, R1M-AR
Operation and Maintenance Manual
PN# R1M-A-900-01
Publication Date: December 2012
REV. C
REPRODUCTION AND DISTRIBUTION OF THIS TECHNICAL MANUAL IS AUTHORIZED FOR US
GOVERNMENT PURPOSES. TEGAM PROPRIETARY INFORMATION.
NOTE: This User’s Manual was as current as possible when this product was manufactured. However, products are constantly being updated and improved. To ensure you have the latest documentation, refer to www.tegam.com
Revised August, 2004
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
TABLE OF CONTENTS
TABLE OF CONTENTS
1. INSTRUMENT DESCRIPTION
Purpose ........................................................................ 1-1
Performance Characteristics ............................................ 1-1
Description of Equipment ................................................ 1-2
List of Items Furnished ................................................... 1-3
Storage and Shipping Requirements ................................. 1-3
2. PREPARATION FOR USE AND INSTALLATION
Unpacking and Inspection ............................................... 2-1
Preparation for Use ........................................................ 2-1
Figure 1: Connection Diagram ..................................... 2-3
3. OPERATING INSTRUCTIONS
Pushbutton Functions ..................................................... 3-1
General Theory of Operation ........................................... 3-2
4. PRINCIPLES OF OPERATION
5. MAINTENANCE
Inspection ..................................................................... 5-1
Cleaning ....................................................................... 5-1
Test Equipment required for Calibration and Repair ............ 5-1
Performance Verification ................................................. 5-2
Calibration .................................................................... 5-3
Troubleshooting ............................................................. 5-4
Table 2: Fault Symptoms and Repair Actions ................. 5-5
Preparation for Shipment ................................................ 5-6
Overhaul Instructions ..................................................... 5-6
Figure 2: Front Panel Controls ..................................... 5-7
Figure 3: Front Panel-Rear View .................................. 5-8
Figure 4: Simplified Block Diagram .............................. 5-9
Figure 5: Power Supply Schematic ............................. 5-10
Figure 6: Power Supply Layout .................................. 5-11
Figure 7a, 7b: Display Board Schematic ...................... 5-12
Figure 8: Display Board Parts Layout .......................... 5-14
Figure 9: System (Side View) Layout ......................... 5-15
Table 3, 4, 5, 6: Parts List ........................................ 5-16
Table 7: Vendor Cage Code Directory ......................... 5-19
6. SERVICE INFORMATION
Preparation for Repair or Calibration Service ..................... 6-1
Expedite Repair and Calibration Form ............................... 6-2
Warranty ...................................................................... 6-3
Revised August, 2004
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-1
INSTRUMENT DESCRIPTION
SECTION 1
INSTRUMENT DESCRIPTION
INTRODUCTION
1.1 Purpose
The Model RIM-A Megohmmeter is a portable instrument used to measure high values of resistance. It is enclosed in a rugged weather-resistant housing designed for severe industrial and military applications. The Model RIM-A may be used to measure resistances between 50 k and 10 GΩ with 50 V test voltage, and between 1 M and 200 GΩ with 100 V, 250 V, or 500 V test voltage.
1.2 Performance Characteristics
This is a digital reading instrument with six resistance ranges and four test voltage ranges.
Range
Full Scale
(Ω)
Resolution
(Ω)
Test Voltage
(V)
1
1 M
1 k
50
2 10 M
10 k
100, 250,
or 500
3
100 M
100 k
4
1 G
1 M
5
10 G
10 M
6
100 G
100 M
Table 1: Specifications
Accuracy: Resistance accuracy on all ranges: ± 5% Test Voltage accuracy on all ranges: ± 3%
Revised August, 2004
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-2
INSTRUMENT DESCRIPTION
1.3 Description of Equipment
Physical: A rugged heavy-duty case is provided to contain and protect the instrument. When closed, a gasket seals the lid to protect the instrument against water and dirt while the instrument is carried through rainstorms or other hazardous conditions. The lid is secured by two latches. A handle is provided for portability. A compartment is provided for storage of test cables and line cord.
Dimensions: 216 mm (8.5) x 228 mm (9) x 152 mm (6). Weight is 2.3 kg (5 lb). Controls and connectors are of a size and spacing such that the instrument may be operated while wearing safety gloves.
Electrical: Input power is from the AC power line (103.5 V to
129 V at 50 Hz or 60 Hz, at less than 1/8 A).
Environmental: This unit will operate over a temperature range from 0 °C to 50 °C, 75% RH non-condensing, up to 3050 m altitude. Withstand functional shock of 40 G for 11 ms. Vibration: 2 G maximum at 5 Hz to 55 Hz.
Front Panel Controls and Displays (See figure 2)
OHMS RANGE Switch is a rotary six position selector switch
used to step through all six ranges.
TEST VOLTAGE RANGE Switch is a rotary four position selector
switch used to step through all four voltage ranges.
POWER Switch is used to turn the power on or off to the
instrument.
TEST pushbutton is a round sealed switch. This pushbutton
switch turns on the power to the unit. A LED is provided to indicate that the test voltage is on.
DISPLAY is a 3 ½ digit LCD, displaying readings from 1.999
to 199.9.
Revised August, 2004
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-3
INSTRUMENT DESCRIPTION
TEST VOLTAGE Jack is recessed to prevent any accidental
encounter.
Three Binding Posts for connection of test leads are marked
SIG (Signal), GUARD and GND (Ground). Proper connections to the resistor under test are described in Section 2.2 below.
WARNING
DO NOT TOUCH THE BINDING POSTS WHEN THEY ARE CONNECTED TO EXTERNAL CIRCUITS. LETHAL VOLTAGES MAY BE PRESENT AT THESE POSTS.
1.4 List of Items Furnished
1 each Model R1M-A with power cord 2 each test cables, one shielded with two alligator clips
and the other with one alligator clip
1 each R1M-A Instruction Manual
1.5 Storage and Shipping Requirements
Standard precautions which apply to electronic test instruments should be followed. A hard mechanical shock, such as from dropping the R1M-A, could damage the liquid crystal display. Care should be taken to prevent damage to associated cables.
The R1M-A should be stored in a relatively dust-free environment. Temperature: -40 °C to +71 °C. Relative humidity: 0 to 100%, non-condensing. Altitude: 4570 m
See Section 5.7 below for shipping requirements.
Revised August, 2004
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
2-1
PREPARATION FOR USE AND INSTALLATION
SECTION 2
PREPARATION FOR USE AND INSTALLATION
2.1 Unpacking and Inspection
Upon receipt, the R1M-A and accessories should be carefully unpacked and removed from the shipping container. Separate the units from the packing material and inspect both the instrument and the accessories for any external damage.
If any dents, broken, or loose parts are seen, do not use
the equipment. Notify the shipping company immediately and follow their instructions as to how to proceed.
Check that all items are present. If any items are missing,
notify the shipper if this is a new instrument. If not new, contact the previous user to locate the missing item(s).
2.2 Preparation for Use
Release the two latches which secure the lid and open the lid. Remove the power cord and the two test cables which are stored in the side compartment. If desired, the lid may then be removed by pushing it to the right. Set the TEST VOLTAGE rotary switch to the desired value, from 50 V to 500 V and set the OHMS RANGE switch to the anticipated range. Note that although the OHMS RANGE switch is marked from 1 M to 100 G, a 100% over-range capability is built into this instrument. Note also, that for resistance values less than 1 M, the only test voltage that should be used is 50 V. No damage will occur if a higher voltage is selected, but the readings may not be accurate. Do not measure resistors less than 50 k with the R1M-A; it may damage the instrument.
WARNING
SHORTING THE ‘TEST’ and ‘SIGNAL’ LEADS TOGETHER WILL
CAUSE INSTRUMENT DAMAGE.
Revised August, 2004
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
2-2
PREPARATION FOR USE AND INSTALLATION
Power Up
Plug the power cord into a source of AC power (103.5 V to 129 V at 50 Hz or 60 Hz). Check that the other end is plugged securely into the power input receptacle on the front panel and turn on the POWER toggle switch.
Connections to DUT
One test cable has a double banana plug at one end and red and black insulated alligator clips at the other end. This is the shielded test cable. Connect the red alligator clip to one end of the resistor under test. Connect the black clip to the guard circuit, if any. If no guard is available, this clip may be left floating (it is connected to the cable shield and to the R1M-A guard circuit at the other end) however, this clip must not be allowed to make accidental contact with any part of the external circuit. Plug the dual-banana plug on this cable into the SIG and GUARD binding posts. Note that the pin marked GROUND should go into the GUARD binding post. If there is no guard available and the black alligator clip at the other end of the test cable is left floating, the operator may choose to orient the dual banana plug so that the GROUND pin is NOT plugged into the GUARD binding post. This will ensure that any accidental contact of the black alligator clip will cause no problem; however, the cable shield will then be floating and the signal may be excessively noisy. The other test cable has a shrouded banana plug on one end and an insulated alligator clip on the other end. This cable carries the high voltage output of the R1M-A to the resistor under test. Connect the alligator clip to the other end of the resistor under test; then plug the shrouded banana plug into the recessed (RED) TEST jack. Please refer to the connection diagram (figure 1) below.
The shorting link on the GROUND binding post may be swung around and connected to the GUARD binding post to ground the R1M-A circuitry. This may be done only if there is no external voltage with respect to ground on the measured resistor or guard.
Revised August, 2004
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
2-3
PREPARATION FOR USE AND INSTALLATION
Figure 1: Connection Diagram
Revised August, 2004
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-1
OPERATING INSTRUCTIONS
SECTION 3
OPERATING INSTRUCTIONS
The R1M-A is designed for bench-top or field operation. Use the cables connecting their banana plug terminations to the appropriate connectors explained in Section 2.2.
WARNING
DO NOT TOUCH ANY ALIGATOR CLIP WHILE DEPRESSING THE TEST PUSHBUTTON. LETHAL VOLTAGES MAY BE PRESENT ON THESE CLIPS.
3.1 Rotary Switch Functions
The two rotary switches manually scroll through the six resistance and four voltage ranges. Full scale on any range may be described as 2,000 (actually 1,999), since this is a 3 ½ digit meter. If the readings are exceeding 1,999 counts, step to the next higher range. If the readings are less than 2,000 counts, step down a range, so that it will read 1,999 counts, or less. The decimal point is located automatically for the correct reading. If the selected range is too low for the value of the resistor under test, the display will show over-range by showing “1” on the display. Similarly, the under-range condition is displayed by showing negative readings on the display. -1” is displayed on the display until the test button is pressed. The POWER switch turns on the power to operate the unit or turns off the power to the unit. Note that when power is turned off, it may take several seconds to discharge circuit capacity to completely blank the display. The black markings seen have no effect and disappear when power is turned back on.
WARNING
TAKING RESISTANCE MEASUREMENTS ON A POWERED OR ENERGIZED COMPONENT COULD CAUSE SERIOUS HARM TO THE OPERATOR AND/OR DAMAGE THE R1M-A. BE SURE THE ITEM TO BE MEASURED IS DISCONNECTED FROM OTHER COMPONENTS AND APPARATUS.
Revised August, 2004
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-2
OPERATING INSTRUCTIONS
3.2 General Theory Of Operation
A 2-terminal measurement method is used to determine the resistance of the item under test. The R1M-A calculates the resistance of the item under test utilizing Ohm's law and displays it on a 3 ½ digit display.
To obtain the best accuracy, allow the R1M-A to warm up for 15 to 30 minutes before making measurements. In any case, a minimum warm-up of 30 seconds is advised.
Depress the TEST pushbutton, which switches on the high voltage. Note that the red LED will light, indicating that high voltage is present on the terminal. Allow for a brief settling time and read the resistance on the digital display. Note that the operating range of resistance values which may be read on any given range extends from 5% to 199.9% of the nominal value of the selected range. Note that readings greater than 100% will take more time. Thus, the 1 MΩ range may be used to measure resistors from 50 k to
1.999 M. Note that the test voltage must be 50 V for resistances less than 1 MΩ. If the resistance measured exceeds the maximum value for the selected range, the display will show blanks; in that case, set the range switch to a higher range. If the resistance measured is less than 5% of the nominal value of the selected range, the display will show a minus sign; if so, switch to a lower range, since the readings may be inaccurate.
If readings are noisy, especially at very high values of resistance, it may be caused by 60 Hz coupled to the test resistance from the power line. Note that at 100 G, an extremely small capacitance will couple many volts into the test resistance from a power line. If so, it may be necessary to interpose a grounded shield plate or to put the unknown resistor into a metal box, connected to the GUARD alligator clip.
To close up the R1M-A, first reattach the lid to the main case, if it has been removed. Then store the power cord and
test cables in the side compartment.
Revised August, 2004
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
4-1
PRINCIPLES OF OPERATION
SECTION 4
PRINCIPLES OF OPERATION
The power supply uses a transformer with two secondary windings. The low-voltage winding is a center-tapped winding. Diodes D9 and D12 provide rectified + power filtered by C26 and regulated at +6 V by D7. Diodes D10 and D11 provide rectified - power, filtered by C25 and regulated at -6 V by D8. The high-voltage winding on T-1 charges C23 through D7 and C24 through D8. Since these two capacitors are connected in series, this circuit functions as a voltage doubler to provide a high voltage nearly equal to twice the peak voltage of the high voltage winding. This high voltage is at least 650 V at low line in order to provide 500 V test voltage with 1 IDA of I x R drop across R37.
R37, a series resistor, is followed by Q1, a high-voltage shunt regulator, to regulate the high voltage supply at the selected value. R38 provides protection for Q1 from excess current caused by external voltages. R6-R9 operate as a four-section voltage divider, dividing the voltage from U2, a
2.5 V precision voltage reference, to 0.25, 0.5; 1.25, and 2.5 V. The voltage selected by S1 A, the ‘TEST VOLTAGE selection switch, is fed into pin 9 of U1, a quad operational amplifier. The high voltage output is divided by R3 and R4 to 1/200 of the high voltage output. This divided voltage is fed to pin 10 of U1, where it is compared with the selected reference voltage. If the divided output is higher than the reference, the output (pin 8) of U1 goes more positive, increasing the base current drive to Q1. Since this is a shunt regulator, the output high voltage decreases, until an equilibrium point is reached. Thus, the high voltage is regulated at the selected value.
When a resistor is not being tested, S3 (the TEST switch) is closed, shorting out the high voltage. This also provides added safety in the event that the external circuit under test contains capacitors. If so, they will be charged during TEST, but discharged automatically as soon as S3 is released.
Loading...
+ 25 hidden pages