Tegam R1L-BIR1 Instruction Manual

R1L-BIR1
MODEL R1L-BIR1
Instruction Manual
PN# R1L-BIR1-900
Publication Date: May 2019
REPRODUCTION AND DISTRIBUTION OF THIS TECHNICAL MANUAL IS AUTHORIZED FOR US
NOTE: This User’s Manual was as current as possible when this product was manufactured. However, products are constantly being updated and improved. To ensure you have the latest documentation, refer to www.tegam.com
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
GOVERNMENT PURPOSES. TEGAM PROPRIETARY INFORMATION.
REV. B
sales@tegam.com
TABLE OF CONTENTS
TABLE OF CONTENTS
1. INSTRUMENT DESCRIPTION
Purpose ............................................................... 1-1
Performance Characteristics Table 1, Figure 1 ........... 1-1
Description of Equipment ....................................... 1-2
Figure 2: Front Panel Controls ............................. 1-3
List of Items Furnished .......................................... 1-5
Storage and Shipping Requirements ........................ 1-5
Safety Information and Precautions ......................... 1-5
2. PREPARATION FOR USE AND INSTALLATION
Unpacking and Inspection ...................................... 2-1
Preparation for Use................................................ 2-1
3. OPERATING INSTRUCTIONS
Pushbutton Functions ............................................ 3-1
General Theory of Operation ................................... 3-2
4. PRINCIPLES OF OPERATION
Principles of Operation ........................................... 4-1
5. MAINTENANCE
Inspection ............................................................ 5-1
Cleaning .............................................................. 5-1
Calibration ........................................................... 5-2
Troubleshooting .................................................... 5-6
Table 2: Fault Symptoms and Repair Actions ........ 5-6
Preparation for Shipment ....................................... 5-7
Overhaul Instructions ............................................ 5-8
Figure 3: Parts Layout ........................................ 5-9
Figure 4, 5: Digital, Analog Section .................... 5-10
Table 3: Parts List – Main Board Assembly .......... 5-12
Table 4: Parts List – Final Assembly ................... 5-14
Table 5: Parts List – Front Panel Assembly .......... 5-15
Table 6: Parts List – Battery Board Assembly ...... 5-17
Table 7: Vendor Cage Code Directory ................. 5-18
6. SERVICE INFORMATION
Preparation for Repair or Calibration Service ............. 6-1
Expedite Repair and Calibration Form ...................... 6-2
Warranty .............................................................. 6-3
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
INSTRUMENT DESCRIPTION
Full
(Ω)
Test
Peak (A)
1
1.999 m
1 µ
1.4
2
19.99 m
10 µ
140 m
3
199.9 m
100 µ
14 m
4
1.999
1 m
1.4 m
5
19.99
10 m
140 µ
SECTION 1
INSTRUMENT DESCRIPTION

1.1 Introduction

The TEGAM Model R1L-BIR1 bond meter is a purpose built, rechargeable battery powered portable instrument for ground bond measurements. The R1L-BIR1 comes with push-pin probes and Kelvin alligator clips, all housed within a ruggedized case. It can accurately make measurements with a resolution of 1 µΩ. It has full scale ranges from 2 mΩ to 20 Ω. The resistance reading is displayed on a 3½ digit liquid crystal display on the front panel. An over-range condition is indicated as a "1" followed by three blank digits (the decimal point is also displayed in its normal position for the range selected).

1.2 Performance Characteristics

This is a range-selection instrument, with a rotary 5 range selector switch.
Range
Scale
Resolution
(Ω)
Current
Reactive loads not to exceed 2mH or 100µF with a minimum of 5µΩ per µH of series resistance. See figure 1 below.
Table 1: Specifications
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-1
INSTRUMENT DESCRIPTION
Figure 1 Accuracy: Accuracy on all ranges: ± (0.25% of reading + 1 count), when powered from the batteries only.

1.3 Description of Equipment

Physical: The circuitry is enclosed in a rugged case.
Dimensions: 14.1 in (358 mm) W x 10.6 in (269 mm) D x 6.1 in (155 mm) H. Weight with accessories is 4.45 kg (9.8 lb). Controls and connectors are of a size and spacing such that the instrument may be operated while wearing safety gloves.
Electrical: The R1L-BIR1 is powered by an internal battery consisting of three rechargeable NiCad ‘C’ cells (2800 mAh ­3500 mAh) with a built-in charger. Input power for the charger is 90 VAC to 250 VAC at 50 Hz or 60 Hz. Charging power is supplied via a removable line cord; one end of which plugs into a line filter on the front of the R1L-BIR1, and the other end being a standard three-pin grounded line plug. Fuse: ¼ A 3AG Slow blow, TEGAM PN#FU-202
Input Protection: The R1L-BIR1 will not be damaged by a signal of 1 Vp-p applied across any pair of input terminals.
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-2
INSTRUMENT DESCRIPTION
Environmental: This unit will operate over a temperature range from 0 °C to 50 °C, 75% RH non-condensing, up to 3050 m altitude.

Front Panel Controls and Displays (See figure 1)

RANGE Switch. The rotary 5 range selector switch may be used
to step through all five ranges.
POWER Switch. Used to turn the power on or off to the
instrument.
DISPLAY is a 3 ½ digit, displaying readings from 1.999 to
199.9.
Four binding posts for connection of test leads are marked +I,
+E, -E, and -I. Proper connections to the resistor under test are described in Section 3.2 below.
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-3
INSTRUMENT DESCRIPTION
Figure 2: Front Panel Controls
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-4
INSTRUMENT DESCRIPTION
!

1.4 List of Items Furnished

1 each Model R1L-BIR1 with power cord 3 each rechargeable NiCad battery cells (P/N: BA-110) 1 each R1L-BIR1 Instruction Manual (P/N: R1L-BIR1-900) 2 each Kelvin Clip Cables, one end with a dual banana
plug and the other end with a gold plated Kelvin clip (P/N: KTL-100)
2 each Kelvin Probes, one end with a dual banana plug
and the other end with a Kelvin Probe with replaceable tips (8 each) (P/N: BCP-10)

1.5 Storage and Shipping Requirements

Standard precautions which apply to electronic test instruments should be followed. A hard mechanical shock, such as from dropping the R1L-BIR1, could damage the liquid crystal display. Care should be taken to prevent damage to associated cables. Remove the battery cells for inactive storage of 30 days or more.
Temperature: -40 °C to +71 °C. Relative humidity: 0 to 100%, non-condensing. Altitude: 4570 m See Section 5.6 below for shipping requirements.
1.6 Safety Information and Precautions
Grounding the Equipment
This product is grounded through the grounding conductor of the power cord. To avoid electrical shock or other potential safety hazards, plug the power cord into a properly wired receptacle before using this instrument. The proper grounding of this instrument is essential for safety and optimizing instrument operation.
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-5
INSTRUMENT DESCRIPTION
Danger Arising from Loss of Ground
If the connection to ground is lost or compromised while charging the batteries, a floating potential could develop in the instrument. Under these conditions all accessible parts, including insulating parts such as knobs and terminals could develop a hazardous voltage and put the user at risk.
Use the Proper Fuse
To avoid fire hazard, use only the correct fuse type as specified for the AC power supply in the “Description of Equipment” or “Maintenance” sections of this manual.
Do Not Use in Explosive Environments

The R1L-BIR1 microohmmeter is not designed for operation in explosive environments.

10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-6
PREPARATION FOR USE AND INSTALLATION
WARNING
DO NOT CONNECT THE AC LINE CORD PRIOR TO
WILL PRESENT A POSSIBLE SHOCK HAZARD.
CAUTION
BE SURE TO OBSERVE THE PROPER POLARITY WHEN INSERTING THE BATTERY CELLS OR DAMAGE MAY RESULT.
SECTION 2
PREPARATION FOR USE AND INSTALLATION

2.1 Unpacking and Inspection

Upon receipt, the R1L-BIR1 and accessories should be carefully unpacked and removed from the shipping container. Separate the units from the packing material and inspect both the instrument and the accessories for any external damage.

If any dents, broken, or loose parts are seen, do not use the
equipment, and notify TEGAM immediately.
Check that all items are present. If any items are missing,
notify TEGAM immediately.
Note that the instrument is shipped with its battery
removed. Unwrap the 3 cells which comprise the battery.

2.2 Preparation for Use

Battery Installation

Be sure the power line cord is disconnected. Check that the Power Switch is in the OFF position. Remove the ten screws around the R1L-BIR1’s top (control/display) panel. Carefully lift the unit out of its cabinet. Place the unit upside down on a surface which will not mar or damage the control/display panel.
REMOVING THE UNIT FROM ITS CASE OR WHILE THE UNIT IS OUT OF ITS CASE, OR ELSE THE MAINS VOLTAGE
Insert the battery cells, one at a time, in accordance with the
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
2-1
PREPARATION FOR USE AND INSTALLATION
WARNING
VIA THE GROUND PIN OF THE POWER CABLE, IF THIS
POWER CABLE IS SHORTED TO THE PANEL, THE PANEL
; SO DO
NOT TOUCH THE PANEL WHEN AC POWER IS CONNECTED.
polarity indications marked on both the battery cell and the battery holder. Insert the positive end of the battery cells first. After installing the cells, reinstall the instrument in its case and secure the ten screws.
Power Up
Plug the power cord into the R1L-BIR1 front panel receptacle, and plug the other end into a standard AC outlet. For battery operation (with charged battery cells), this step is omitted. Turn on the POWER rotary switch. Note that this switch turns on both line and battery power.
Charging the Battery Cells
Prior to battery operation, it is recommended the R1L-BIR1 Microohm/Bond Meter be plugged into AC power for at least 14 hours to charge the battery cells. To charge the battery cells, the Power Switch must be turned ON, and the Range Switch should be set to 20 . With the Range Switch set to lower ranges, the charge rate decreases and the batteries will be discharged, especially on the 2 mΩ range, even with no resistance connected to the instrument. (However, the discharge rate in this range will permit operation for at least 14 hours with no resistance connected.)
ALTHOUGH THE FRONT PANEL IS NORMALLY GROUNDED
CONNECTION IS FAULTY AND IF THE HIGH SIDE OF THE
COULD POSSIBLY HAVE MAINS VOLTAGE ON IT
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
2-2
OPERATING INSTRUCTIONS
WARNING
MAY BE PRESENT AT THESE POSTS.
SECTION 3
OPERATING INSTRUCTIONS

The R1L-BIR1 is designed for field or bench-top operation. Use appropriate cables connecting their terminations to the four binding posts marked +I, +E, -E, and -I.

DO NOT TOUCH THE BINDING POSTS WHEN THEY ARE CONNECTED TO EXTERNAL CIRCUITS. LETHAL VOLTAGES

3.1 Rotary Switch Functions

Operation of the Model R1L-BIR1 is simple. There are only two switches.
The RANGE rotary switch manually scrolls through the five ranges.
Full scale on any range may be described as 2,000 counts (actually 1,999), since this is a 3½ digit meter. If the readings are exceeding 1,999 counts, step to the next higher range. If the readings are less than 2,000 counts, step down a range, so that it will read 1,999 counts, or less. The decimal point is located automatically for the correct reading.
If the selected range is too low for the value of the resistor under test, the display will show over-range.
The POWER rotary switch turns on the power to operate the unit or turns off the power to the unit. The appearance of the display will show that power is on; absence of the display indicates that either power is off, batteries are discharged, or there is some malfunction. Note that when power is turned off, it may take several seconds to discharge circuit capacity to completely blank the display. The black markings seen have no effect and disappear when power is turned back on.
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-1
OPERATING INSTRUCTIONS
WARNING
FROM OTHER COMPONENTS AND APPARATUS.

3.2 General Theory Of Operation

Measurement Principle
A 4-terminal measurement method is used to determine the resistance of the item under test. This method allows the accurate measurement of low values of resistance in spite of the other resistances present in the connection leads. A known current is supplied to the item under test via two wire leads connected to the +I and -I binding posts on the R1L-BIR1 Microohm/Bond Meter front panel. Two other wire leads connected to the +E and -E binding posts on the R1L-BIR1 front panel sense the voltage developed across the resistance under test due to the supplied known current. The R1L-BIR1 then calculates the resistance of the item under test utilizing Ohm's law, and displays it on a 3 ½ digit display.
Kelvin Clip Cable Description
A Kelvin clip consists of two opposing jaws of an alligator style clip which are insulated from each other, so one jaw may be used to connect a source of current to the resistance under test, and the other jaw used to measure the voltage at the test point. The two jaws are spring loaded to clamp onto a cylinder of ½ inch diameter or less.
OPERATING THE R1L-BIR1 MICROOHM/BOND METER AND TAKING RESISTANCE MEASUREMENTS ON A POWERED OR ENERGIZED COMPONENT COULD CAUSE SERIOUS HARM TO THE OPERATOR AND/OR DAMAGE THE R1L-BIR1. BE SURE THE ITEM TO BE MEASURED IS DISCONNECTED
If the battery cells have not been installed, refer to section
2.2. Battery only operation will result in the most stable and accurate readings. The R1L-BIR1 may be operated while charging the batteries, however it will not conform to specifications in this configuration.
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-2
Loading...
+ 27 hidden pages