Battery Installation and Replacement
Three (3) AA 1.5 V batteries are supplied with the instrument, but not installed. Read the following battery replacement instructions before attempting to install or remove the batteries.
To install or replace batteries:
Required Tools: Phillips Head Screwdriver
1. Identify the battery compartment located on the back of the instrument;
2. Remove the two (2) battery compartment retaining screws;
3. Remove the battery compartment cover;
4. If present, carefully remove old batteries being careful to not damage the battery contacts;
5. Observing proper polarity, install three (3) new, AA alkaline (IEC LR6, ANSI 15) batteries;
6. Re-install the battery cover and two (2) retaining screws;
7. At initial power on after battery replacement, allow approximately 30 seconds for instrument to stabilize.
Initial Power ON
8. The instrument will initially display every segment on the LCD for 2 seconds as a test. An internal hardware, memory and battery self-test is performed during this time.
9. Upon completing the internal tests, the instrument will immediately display the Source and Read mode last user settings and battery indicator.
10. Set the desired measurement parameters as follows:
a. Enter the Setup Menu by pressing , hold the key down for
approximately 1.5 seconds, and then release it;
Key designators followed by
(1.5s), e.g. (1.5s), indicate that the key should be pressed and held for 1.5 seconds, then released
to access the desired function.
Battery Installation
Digital Temperature Calibrator/ Thermometer 940A/945A
sales@tegam.com
940A-901 Rev B
b. The active thermocouple type is flashing on the display. Use or
to select the desired thermocouple type;
c. Momentarily (do not hold) press to save your selection and move to
the next parameter;
d. The active temperature unit is flashing on the display. Use to
select the desired temperature unit (°C, °F, or mV);
e. Momentarily press to save your selection and move to the next
parameter;
f. Read Channel offset value is flashing on the display. If the temperature
probe’s offset value is known, press to set the Channel 2 probe offset to the probe’s offset value.
g. Momentarily press to save your selection and move to Open Lead
Detection, press to toggle on/off;
h. Momentarily press to save your selection and move to Source on/off;
See SETUP MENU CHOICES FOR °C AND °F” and/or “SETUP MENU CHOICES FOR mV” in the table below.
i. To save the current parameter value and exit the Setup Menu, press ;
j. To disregard changes made to the current parameter value and exit the
Setup Menu, press .
k. If parameter 2, “Temperature and Voltage Units” is set as either “°C” or
“°F”, the remaining parameter choices available are in the table below under “Setup Choices for °C and °F”. If parameter 2 is set to “mV”, the
remaining parameter choices available are in the table below under “Setup Choices for mV”.