Tegam 911A, 912A, 931A, 932A Operation Manual

911A, 912A
Models:
911A, 912A
Thermocouple Thermometers
Manual Part Number: 911-900, Rev. B
Published July 2017, Geneva, OH
HERMOCOUPLE THERMOMETERS
Operation Manual
rev B
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
Notices
death.
CAUTION denotes a h az ard the unit or other equipment.
NOTICES
Copyright Notice
© TEGAM, Inc., 2017 No part of this manual may be reproduced in any
form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from TEGAM, Inc. as governed by United States and international copyright laws.
This Manual
Part Number:
911-900 Revision B, July 2017 Supersedes: Rev. A, November 2016
Published by:
TEGAM, Inc. 10 TEGAM Way Geneva, OH 44041
Disclaimer and Manual Revis ions :
THE MATERIAL CONTAINED IN THIS USER MANUAL, AND ANY COMPUTER SOFTWARE ASSOCIATED WITH THIS USER MANUAL OR THE PRODUCTS COVERED BY IT, ARE PROVIDED AS IS, AND ARE SUBJECT TO CHANGE, WITHOUT NOTICE, IN FUTURE REVI S IONS .
This User Manual was current at the time of publication. However, TEGAM is dedicated to a process of continual product improvement, and the products covered by this User Manual, and any associated computer software, are su bject to periodic functional and design updates. Please visit tegam.com for the most current product documentation.
U.S. Government Rights
This computer software and/or technical data is TEGAM proprietary information developed exclusively at private expense. Computer software and technical data rights grant ed to th e federal government include only those rights customarily provided to the public, pursuant to FAR 12.211 (Technical data) and FAR 12.212 (Computer software) for the federal government, and DFARS 252.227-7015 (Technical data – Commercial items) and DFARS 227-7202-3 (Rights in commercial computer software or commercial computer software documentation) for the Department of Defense. Except as
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
explicitly permitted by the foregoing, reproduction for non-governmental use of the information or illustrations contained in this computer software and technical data is not permitted.
Compliance
Safety Notice Symbols and Terms
Safety Notices denote hazards. They indicate an operating procedure, instruction, or practice that, if not correctly performed or followed, could result in damage to equipment, or injury or death to personnel . Do not proceed beyond a Safe ty Notice until all conditions and instructions are fully understood and complied with.
Safety Notices Symbols:
ii
sales@tegam.com
WARNING
CAUTION
REMINDER denotes important information about instrument functions, menus, and measurements.
WARNING denotes an imminent hazard that could result in injury to personnel or
that could result in damage to
Table of Contents
TABLE OF CONTENTS
1. Instrument Description................................................................................................................................ 1-1
Specifications ..................................................................................................................................... 1-1
Optional Accessories and Or d e ring Information ........................................................................ 1-2
TEGAM Family of Thermometers ..................................................................................................... 1-3
2. Preparation for Use ..................................................................................................................................... 2-1
General Information ......................................................................................................................... 2-1
Feature Overview .............................................................................................................................. 2-1
Safety Notices and Information ...................................................................................................... 2-2
Unpacking and Inspection .............................................................................................................. 2-4
Battery Installation and Replacement .......................................................................................... 2-4
Making Your First Temperature Measurement ............................................................................. 2-5
3. Operating Instructions ................................................................................................................................ 3-1
Keypad Functions .............................................................................................................................. 3-1
LCD Display ......................................................................................................................................... 3-1
Setup Menu ........................................................................................................................................ 3-3
View Mode s and Sta t istics ............................................................................................................... 3-4
Auto-Power Off .................................................................................................................................. 3-5
Backlight and Backlight Timeout .................................................................................................... 3-6
Hold Function ..................................................................................................................................... 3-6
Trend Indicators ................................................................................................................................. 3-6
Battery Indicator ................................................................................................................................ 3-6
Probe Offset ....................................................................................................................................... 3-7
Open Lead Detection Enable/Disable (SN xxxxxx and above only) ....................................... 3-8
Clear Function ................................................................................................................................... 3-9
Invalid Measuremen t Indic ations ................................................................................................... 3-9
4. Service Information ..................................................................................................................................... 4-1
Inspection and Cleaning ................................................................................................................. 4-1
Calibration .......................................................................................................................................... 4-1
4.2.1 Verification Procedure ................................................................................................................ 4-1
4.2.2 Alignment Procedure .................................................................................................................. 4-2
Troubleshooting ................................................................................................................................. 4-6
Diagnostic Routines and Error Codes ............................................................................................ 4-7
Memory Sterilization .......................................................................................................................... 4-7
Preparation for Calibration or Repair Service .............................................................................. 4-8
Expedite Repair & Calibration Form .............................................................................................. 4-9
Warranty....................................................................................................................................................... 4-10
Warranty Limitations ................................................................................................................................... 4-10
Statement of Calibration .......................................................................................................................... 4-10
A. Required Equipment.................................................................................................................................... A-i
B. Expanded Instrument Uncertainties ..........................................................................................................B-i
C. Instrument Verification Data Sheet ........................................................................................................... C-i
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
iii
sales@tegam.com
Instrument Description
G
:
Basic Accuracy
±(0.04% |rdg| + 0. 3 °C )1
Conformity
ITS-90
Temperature
K
-200 to 1372
-328 to 2502
73 to 1645
J
-210 to 1200
-346 to 2192
63 to 1473
T
-250 to 400
-418 to 752
23 to 673
E
-250 to 1000
-418 to 1832
23 to 1273
Connector Type
One (1) Mini-TC (911A)
Two (2) Mini-TC (912A)
Probe Zero Func ti o n
Resolution 0.1 °C/°F/K
Four (4) digit LCD, with Temperature, Units, Function, Trend, Polarity, Battery, and Decimal Indicators
Display Backlight
Four (4) LED Backlight with 30-second timeout
Display Resolution
(-99.9 to 999.9 °)
1 ° (< -99.9 °, > 999.9 °)
Reading Rate
Battery Type
Battery Life
Battery Indicato r
Four (4) Stage Battery Charge Indicator
Reading Range
Keypad
Eight (8) momentary switches with audible and tactile feedback
Clock
Elapsed Statistics Run Time
Instrument retains last selected:
Input Current
Maximum Common Compliance
CE (2014/30/EU) / RoHS2 (2011/65/EU)
ESD
IEC 61000-4 2:2009, Class B

1. INSTRUMENT DESCRIPTION

Specifications

ENERAL SPECIFICATIONS
Ranges
Display
Statistics
Power Cycle Configuration Retention
°C °F K
3 / Second for Readings and Trend Indicators 3 AA (IEC LR6, ANSI 15) Alkaline 2000 Hours Typical
Minimum Reading Maximum Reading Average Reading
Sensor Type
Temperature Unit
Offset Values
±50 nA
Standard Deviation T1–T2 (912A only)
Mode Voltage
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
42 V peak to earth
1-1
sales@tegam.com
1 V p-p between T1 and T2
Instrument Description
EN 55022:2010+A1:2015,
E
:
Standards
MIL-PRF-28800F, Class 2
UL 60079-0 § 26.4.2
Operating Temp
-20 to 55 °C
-4 to 131 °F
For specification varia nce s du e to a mbi en t op e r atin g
<10 °C (50 °F): Non-condensing
Altitude
0 to 4600 m
0 to 15,092 ft
Vibration
Random 10 – 500 Hz, 0.03 g2/Hz
Shock
30g Half Sine
Drop
4 Drops from 1 m to Conc re t e
Storage Temp
-40 to 71 °C
-40 to 159 °F
P
:
Dimensions
193 x 84 x 28 mm
7.6 x 3.3 x 1.1 in
Weight (incl.
1
For complete instrument accuracies, see the Expanded Instrument Uncertainty charts in Appendix B of
this manual.
P
M
D
911-910
Tilt Stand/Magnet/Hanger
911-911
Foam-Filled Hard Carry Case
See TEGAM Temperature Probe Selection Guide at
Printed Manual
911-900
Operation Manual
Manual
Chinese, Dutch, French, German, Japanese, Korean, and
Service Options
Calibration with Statement of Traceability
EMC
NVIRONMENT
Temperature Coefficient
Humidity
HYSICAL CHARACTERISTICS
Batteries)
Class A; EN 61000-4 3:2006+A2:2010, 10 V/m (80 MHz to 1 GHz)
temperature, see the Expanded Instrument Unc ertainty charts in Appendix temperatures not shown in Appendix B, accuracies shall be interpolated linearly.
10 to 30 °C (50 to 86 °F): 5 to 95% RH 30 to 40 °C (86 to 104 °F): 5 to 85% RH 40 to 55 °C (104 to 131 °F): 5 to 60% RH
911A: 300.9 g (10.6 oz.) 912A: 303.2 g (10.7 oz.)
B of this manual. For ambient operating
MIL-PRF-28800F, Class 2

Optional Accessories and Ordering Information

RODUCT
Accessories
Translations
ODEL
ESCRIPTION
tegam.com for available temperature probes
Spanish (download at tegam.com)
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
1-2
Instrument Description
911A
Thermocouple Thermometer, Single Input
912A
Thermocouple Thermome te r, Dual Input
931A
Data Thermometer, Single Input
932A
Data Thermometer, Dual Input

TEGAM Family of Thermometers

Thermocouple
Thermometers
Data
Thermometers
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
1-3
sales@tegam.com
Preparation for Use

2. PREPARATION FOR USE

General Information

The TEGAM 911A and 912A Thermocouple The rmo met ers ar e hig h-accuracy handheld digital thermometers that provide accura te tem pera ture re adi ngs in a wide range of manu factu ring and service applications. Thes e full-featured, durable, and versatile instruments sim pli fy the process of temperature me as ure ment through the intui tive user-interface. T he y are compatible with the fo ur most popular NIST tracea bl e thermocouple types: E, J, K, and T.

Feature Overview

Keypad with audible and tactile feedback
1
2000-hour battery life
Four (4) digit dual LCD with LED Backlight
Four (4) NIST-traceable thermocouple types: E, J, K, and T
Comprehensive real-time statistics: MIN, MAX, AVG, RNG, STDEV , and T1-T2
Easy to clean
Probe offset function to minimize probe error
0.1° / 1 ° display resolution
°F, °C, and K temperature units
Reading HOLD mode
Conforms to ITS-90 thermocouple tables
Durable: Meets MIL-PRF-28800F, Class 2 requirements
Optional Tilt Stand/Magnet/Hanger
User-friendly operation
Retains measurement parameters, even when turned off
Self-diagnostic routine to identify fault conditions
Low battery and open sensor indications
2
1
Typical battery life under normal use conditions in laboratory environment. Continuous or repeated use of
features such as the backlight, or use or storage at high or low temperature extremes may reduce battery life.
2
T1-T2 is available on model 912A only.
2-1
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
Preparation for Use
could expose users to elevated potentials.
Never use this instrument or any temperature probe or sensor inside a microwave oven.
damage or personal injury could result from exceeding a probe’s maximum temperature rating.

Safety Notices and Information

Read this Operation Manual thoroughly before using the instrument to become familiar with its
Visually inspect instrument before using. Do not use if unit appears damaged or with any part of the
MAINTENANCE INSTRUCTIONS WITHIN THIS MANUAL ARE FOR USE BY QUALIFIED SERVICE
PERSONNEL ONLY. DO NOT ATTEMPT TO SERVICE THIS UNIT UNLESS YOU ARE QUALIFIED TO DO SO.
Disconnect all temperature probes and turn the unit off before removing the battery cover.
Never connect thermocouple leads to any source where more than 42 Volts (peak) could exist
between the lead and ground. If it is necessary to make measurements of an object at ele vated
electrical potential, the user is responsible for obtaining and properly using a probe that provides adequate
insulation between the surface with elevated potential and the thermocouple wiring.
Always disconnect probe leads before opening the battery door or the instrument housing.
Internal circuits can present a shock hazard if leads are connected to a source of elevated potential.
Do not use this instrument if the housing, probe wiring, probe, or probe handle are damaged
or distorted. Housings and wire insulation are part of the personnel protection system, and if damaged
Never use or store this product with batteries installed, or change b atteries, in an environment
where explosive or flammable vapors or dust suspensions may exist. For thermocouple
thermometers suitable for use in explosive environments, see TEGAM’s 921A or 922A Intrinsically Safe
Thermometers.
Do not attempt to recharge alkaline batteries.
Do not put batteries into bags designed to protect parts from electrostatic discharge (ESD).
These bags are specially designed with metal shielding which can short circuit a battery.
Do not expose batteries to extreme heat or fire. Observe all regional laws and regulations when
disposing batteries.
operations and capabilities.
case removed.
WARNING
SHOCK HAZARD
EXPLOSION H AZ ARD
BURN HAZARD
Do not touch a temperature probe sheath that has been exposed to toxic substances or
extremely high or low temperatures.
Do not attempt to measure temperatures beyond the range of the temperature probe. Probe
Safety Notices and Information continued on next page . . . .
2-2
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
Preparation for Use
RISK OF INSTRUMENT DAMAGE
care to avoid static discharge when handling the instrument or connected probes.
Do not use when AC or DC voltages in excess of 1V exist between thermoco uple channels
(on instruments with more than one channel). Excessive voltage could result in an incorrect
reading, or in more extreme cases, a blown fuse that will result in incorrect readings and need for
repair.
Only replace batteries with size AA (IEC LR6, ANSI 15). Observe proper polarity when installing
batteries. Do not mix old and new batteries.
Do not apply voltages across thermocouple leads in excess of normal thermocouple voltage
for the selected range. Excessive input voltage could result in blown fuse, component damage, or
fire. Application of excessive voltage is not covered by the warranty.
Avoid making sharp bends in probe or sensor lead wires. Bending lead wires at sharp angles can
damage the wire and cause probe failure.
When using both thermometer inputs and a voltage differential exists between the two
measurement points, at least one probe should be electrically insulated. If not, a ground-loop
current can flow through the thermocouple leads causing measurement error or instrument damage.
Static discharge through a connected temperature probe may cause instrument damage. Use
RISK OF INCORRECT READING
CAUTION
2-3
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
Preparation for Use

Unpacking and Inspection

Each instrument is electrically and mechanically inspected before s hipm e nt . Upon receiving your new TEGAM T herm ocou ple Therm ome ter, unpack all items from the shipping container and check for any obvious damage that may h a ve occurred during trans i t. Use the original packing materials if reshipment is necessary.
If any dents, broken, or loose parts are seen, do no t u se the equipment. Notify TEG AM immediately.
Check that all items are present. If any items are mis sing , notify TEGAM immediately. The following items are included with every new instrument:
One (1) Thermocouple Thermometer;
One (1) Quick Start Guide;
Statement of Traceability;
Three (3) AA, 1.5 V batteries; and
Optional accessories (if purchased).

Battery Installation and Replacement

Three (3) AA 1.5 V batteries are supplied with the instrument, but not installed. Read the following battery replacement instructions before attempting to install or remove the batteries.
CAUTION
Always turn the instrument off and disconnect any input connections before replacing the batteries. Re-install the battery compartment cover before resuming use of the instrument.
CAUTION
CAUTION
The battery compartment is sealed with a rubber gasket. Use care to not damage the gasket when removing or installing the battery compartment cover.
Remove the batteries when storing the instrument for an extended period of time or in a high temperature environment to prevent battery leakage and possible damage to the instrument.
All measurement parameters may be reset to factory default if batteries are removed while the instrument is powered on. Always turn the instrument off before changing batteries.
To install or replace batteries: Required Tools: Phillips Head Screwdriver
1. Identify the battery comp a rtm e nt l o cated on the back of t he instrument (see Figure 1 below);
2. Remove the two (2) battery compartme nt re tainin g screw s;
3. Remove the battery compartment cover;
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
2-4
sales@tegam.com
Preparation for Use
Retaining Screws
Battery
Gasket
4. If present, carefully remo ve old batteries being caref ul to not damage the battery contacts;
5. Observing proper polarity, install three (3) new, AA alkaline (IEC LR6, ANSI 15) batteries;
6. Re-install the battery cover and two (2) retaining screws;
7. At initial power on after battery replacement, allow approximately 30 second s for instrument to stabilize.
Figure 1: Battery Installation

Making Your First Temperature Measurement

TEGAM’s 900 Series Thermocouple Thermometers are designed for easy operation, while still providing a feature-ri ch experience via the intuitive user interfa ce .
To get started maki ng temperature measurements r i gh t away, follow these steps:
1. Perform Section 2.5, Ba ttery Insta ll ati on and Repla cem ent;
2. Connect a compatible temperature probe to t he Channel 1 and/or Channel 2 input connector located at the to p of th e i nstr umen t;
3. The instrument will immediately display a temperature measurement for the
4. Set the desired measurement parameters as follows:
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
To ensure best measurement accuracy, allow several minutes for the thermocouple probe and connector to thermally stabilize af ter connection to the instrument.
connected channels. However, to ensure valid and best accuracy measurements, continue to Step 4 below;
sales@tegam.com
2-5
Preparation for Use
a. Enter the Set up Menu by pressing , hold the key down for
approximately 1.5 seconds, and then releas e it;
b. The active thermocouple type is flas hi ng on the display. Use to
select the thermocouple t yp e of the connected temperat ure probe (E, J, K, or T);
c. Momentarily (do not hold) press to save your selection a nd move to
the next parameter;
d. The active temperatu re unit is flashing on the display. Use to
select the desired temperature uni t (°C, ° F, or K);
e. Momentarily press to save your selec tion and mo ve to the next
parameter;
f. Channel 1 probe o f fset value is flashing on the display. If the
temperature probe’s o ff set value is known, press to set the Channel 1 probe offset to the prob e ’s offse t v alue. See Section 3.10, Probe Offset, for more information.
g. Momentarily press to save your se l e c ti on and move to Channel 2
probe offset (if equipped);
h. If desired, repeat Step (f) above for Channel 2;
i. Momentarily press to save your selection and exit the setup menu .
Congratulations! You ’re now ready to make a cc ura t e an d reli abl e temperature measurements, wherever and whenever you may need to.
We know you are eager to begin using your new thermometer, but this overview is just the beginning. Please take a moment to familiarize yourself with this Operation Manual to learn about all the features and bene fit s of your new TEGAM Thermocouple Thermometer.
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
2-6
sales@tegam.com
Operating Instructions
While in Setup Menu, save current value and step to next parameter

3. OPERATING INSTRUCTIONS

Keypad Functions

The instrument keypad is an eight (8) key, sealed membrane keypad. Each key provides audible and tactile us er f ee db a ck w he n pr e s sed. Key functions are descri bed in Figure 2 below.
The , , , and keys have multiple functions which can be accessed by momentarily pressing the key, or alte rn ati vely, by pressing and holding the ke y for approximately 1.5 seconds. Through out this Operation Manual, the press a nd h ol d sequence
(1.5s)
is indicated by the key designator followed by the subscript (1.5s). For i ns t an ce, indicates that the key should be pressed and held for 1. 5 seconds , the n rele ased to
access the desired function.
Power instrument ON or OFF
Disable auto-power OFF
(1.5s)
Enter instrument Setup Menu
(1.5s)
Toggle display backlight
Disable backlight 30-second timeout
(1.5s)
Hold currently displayed measurement
Reset all statistics currently stored in memory
(1.5s)
While in Setup Menu, discard all unsaved changes and exit menu
Cycle through view modes and statistics
While in Setup Menu, save changes and exit menu
While in Setup Menu, advance or reverse selected setting
Figure 2: Keypad Button Functional Description

LCD Display

The display is a la r ge, ea s y to r ea d, dual LCD display, with an LED backlight for clear viewing in low-light conditions. It simulta neousl y dis pla ys tempe rat ure mea sureme nts f or Channel 1 and Channel 2, curre nt therm oc ouple type and tempe rat ure uni t, trend indi ca tors for both Channel 1 and Channel 2, and a battery voltage indicator.
In Statistics View, the display substitu te s the Channel 2 tempe rature measurement wit h the active statistic result, and displays an active statistic mode indicator and the elapsed time of
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
3-1
Operating Instructions
T1 and/or T2 OFFSET is active1
The active thermocouple type
Channel 1 temperature measurement
T1-T2 temperature measurement indicator2
the current statisti c session. See Figure 3 below for fur the r de scription of each display indicator.
1 HOLD function is active
2
3
4 Remaining battery life
5 Channel 1 indicator
6 Channel 1 trend indicators
7
8
9 Channel 2 indicator2
10 Channel 2 trend indicators2
Channel 2 temperature measurement
2
11
T1-T2 measurement result2, or active statistic result
12 Active temperature unit
13 When viewing statistics, time elapsed since start of statistics collection
14 Active statistic
1
T2 Probe Offset available on mo del 912A only.
2
Models 912A only.
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
Figure 3: LCD Display Description
3-2
sales@tegam.com
Operating Instructions
The applied temperature is gre ate r than th e ma xim um tem per atur e for the selected thermocouple type
The applied temperature is less than the minimum temperature for the selected thermocouple type
The LCD can display error information about the current measurement, as s hown in Figure
4.
D
ISPLAY DESCRIPTION
OPEn No thermocouple prob e is connected
-Or-
-Ur-
Figure 4: LCD Error Indications

Setup Menu

Key designators followed by (1.5s), e.g. pressed and held for 1.5 seconds, then release d to a ccess th e desired function.
Measurement settings are configured in the Setup Menu. Pr e ss
(1.5s)
, indicate that the key should be
(1.5s)
to access the Setup Menu. From within the Setup Menu, press to step through the user-definable parameters and the keys to advance or reverse the selected value for the activ e p a r amet e r. The
active parameter value will flash on the display.
Press to save a setting and step to the next parameter. Press to save a setting and exit the Setup Menu. Press to di s re g a rd unsaved changes and e xit the Setup Menu. If no
key is pressed for 10 seconds, the current configur ati o n i s saved and the instrument wi l l exit the Setup Menu.
Figure 5 lists the user-definable parameters and the available values for each parameter. To set a parameter value:
(1.5s)
1. Press
to enter
the Setup Menu;
2. Press to cycle through parameters as shown in Figure 5 until
PARAMETER AVAILABLE VAL UES
Thermocouple Type E, J, K, T
Temperature Units °C, °F, K
T1 Probe Offset ±0.1 ° increment s the desired parameter is reached;
3. To change the value o f the current parameter, press
;
T2 Probe Offset1 ±0.1 ° increments
Open Wire Detection ON, OFF
1
T2 Probe Offset available on model 912A only.
Figure 5: Setup Menu Parameters and Values
3-3
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
Operating Instructions
D INDICATOR
Current Channel 1 measurement – current Channel 2 measurement
Minimum temperature recorded during curre nt session
Maximum temperature recorded during current session
Average of all temperatures recorded during current session
Standard deviation of al l tem pe rat ures recorded during the current sessi o n1.
=
(−µ)
2
4. To save the current param ete r value and cycle to the next p ar ameter , pre ss
(1.5s)
;
5. To save the current parameter value a nd exi t the Setu p Men u, pre ss ;
6. To disregard changes made to the current parameter value and exi t t he Setup Menu, press .
If no key is pressed for 10 seconds, the instrument will save the current configuration and exit the Setup Menu.

View Modes and Statistics

The instrument fea tur e s multiple view modes including a variety of real-time statistics, all available at the touch of a button. Figure 6 below describes e a ch view mode.
VIEW MODE
T1–T2 T1-T2
Minimum MIN
Maximum MAX
Average AVG
Range RNG Maximum - Minimum
Standard Deviation STDEV
1
Standard Deviation is calculated using the population formula:
Press to change view m odes. For each mode, the active measurement or statistic result is displayed on the se c on d l i ne of the display.
The T1-T2 view mode displays th e current Channel 1 measureme nt mi n us the current Channel 2 measurement. The display indicates T1-T2 at the left side of the display. If either channel is not conne cted to a probe, or the current measurement on either channel is over ­or under-range, T1-T2 view mode is not available.
When viewing statistics, the active sta tistic is indicated di rectly below the resul t. The elapsed time of the curre nt st a ti stics session is displayed i n the l ow e r-left corner of the display.
ISPLAY
DESCRIPTION
Figure 6: View Modes and Statistics
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
3-4
Operating Instructions
results and initiate a new statistics session.
M
C
S
Statistics are calculate d co ntin uousl y, begi nni ng whe n the inst rume nt is powered on or when
(1.5s)
is pressed. To pause statistics collection temporarily, press . To resume statistics
collection, press again. It is important to no te that changing parameter values or temperature probes will invalidate
(1.5s)
the current statistics session. When using statistic s, al w ays begin by pressing
to
delete existing sta ti s tics data and initiate a n e w statistics session.
Press to step throug h the avail abl e st atis tic s. S tati stic s are di splay ed i n t he or der shown in Figure 7 below. For du al-channel models, the LCD T1 or T2 indicators ar e l it to identify the channel’s statistics currently being displayed.
When using statistics, always begin by pressing
(1.5s)
to clear existing statistics
The first line of the display indicates the current Channel 1 temperature, regardless of which view mode or channel’s statistic is currently displayed.
ODEL
HANNEL
TATISTIC VIEW SEQUENCE
911A T1 MIN MAX AVG RNG STDEV
T1 MIN MAX AVG RNG STDEV
912A
T2 MIN MAX AVG RNG STDEV
Figure 7: Statistics Sequence
If the instrument records invalid measurement data during the statistics session such as an over-range, under-range, or open input value, ———— will be displayed for each affected statistic result.
To return to the acti ve measurement mode, press repeatedly to step through the remaining view modes, or cycle power.

Auto-Power Off

Key designators followed by (1.5s), e.g. pressed and held for 1.5 seconds, then release d to a ccess th e desired function.
(1.5s)
, indicate that the key should be
To conserve battery life, the instrument automatically turns off if no key is pressed for 20 minutes. To disable this feature, press
(1.5s)
. The remaining battery life indicator will flash
once, indicating auto-power off is disabled.
3-5
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
Operating Instructions
BARS
APPROX. BATTERY LIFE
Auto-power off will remain disa bl e d u n ti l i ns tr ument power is cycled. At nex t p ow e r on, auto-power off returns to the defaul t en able d condi ti on.

Backlight and Backlight Timeout

The instrument includes an LED backlight feature to ensure measurement data can be easily read in low-light conditions. To activate the backlight, pres s .
Once the backlight is activated, it will automatically turn off after 30 seconds if no key is
(1.5s)
pressed to preserve battery life. To disable the backlight timeout feature, press backlight will flash to indicate the timeout feature has been disabled. To re-enable the
backlight timeout feature, turn the backlight off then on by pressing twice.
. The

Hold Function

Press to hold the c urrent reading and/or statistics result, and to pause statistics accumulation. HOLD is displayed at the top-left of the LCD display. New measurements are not displayed, trend i ndi c a to rs are n ot ref reshed, and statistics are n ot cal c ul a te d while the hold function is active.
To disable the hold function and resume n or m al o pe r a tion and s tati sti c s dat a accum ul atio n, press again.

Trend Indicators

Trend indicators provide a visual representation of th e mea sureme nt ’s s tabi li ty, an d separate indicators are provided for each channel. An up arrow indicates th at the curre nt measurement is trending upwards, while a down arrow i ndicates the measurement is trending downwards. Neit her ar ro w i s vi si bl e w hen the measurement is stable. For best accuracy, always allow the mea sure men t to s tabi li ze bef ore eval uati ng or rec ordi ng the measured temperature.

Battery Indicator

Battery depletion or battery replacement will reset all measurement param eters to their default value s and deletes all existing statistics data. Af te r b a t te r y r eplacement, set measurement parameters as required.
The battery voltage indicator provides a visual re presentation of approximate remaining battery life. It is loca ted at the to p-right of the display.
The battery voltage i ndicator uses three b ar s to represent remaining b a tt er y life. Figure 8 shows the approximate battery life for each bar.
At zero (0) bars, the instrument will momentarily display bATT and initiate a shutdown sequence. T o prevent disruption of
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
3 100% - 50% 2 50% - 20% 1 20% - 5% 0 0% - Shutdown Ini ti a ted
Figure 8: Battery Voltage Indicator
3-6
Operating Instructions
Probe offset rounding errors may occur if temperature units are changed while a programmed probe offset after changing temperature units.
Neither trend indic a to r is
has stabilized.
the measurement process and statisti cs col lec tion , the ba tteri es sho uld be repla ced bef ore the battery voltage indicator reaches zero (0) bars. See Section 2.5, Battery Installati on a nd Replacement.

Probe Offset

The probe offset feature c ompe ns a tes for temperature probe error s, significantly improving overall measurement un ce r t ai nty . Pro be offset can be set for Channel 1 and 2 individually. Once set, the probe offset i s auto mati cal ly appli ed to al l subs eque nt me asur ement s and statistics on the o f fset channel.
Current statistics will be invalidated after changing settings such as probe offset. Press
(1.5s)
to delete existing statistics data and ini ti ate a new sta ti s tic s s ess i on .
probe offset is active. When using a probe offset, verify and if necessary correct th e
To set the probe offset when using an un-calibrated t e mpe ra t ure probe:
1. Connect the temperature probe to Channel 1 or Channel 2 (as desire d) of the instrument;
2. Place the probe into a kno w n te mp e r at ure ref e re nc e such as a thermowell or ice
3. Allow the temperatu re p r obe to stabilize in the ice bath or thermowel l b y observing
4. Press
3
;
bath
the instrument trend indicators for the appropriate channel;
(1.5s)
to enter the Setup Me n u ;
displayed when the temperature measurement
5. Press three (3) times to cycle to the Channel 1 Offset parameter;
6. Observe the current Channel 1 temperat ure mea sureme nt di spla yed on the top measurement line of the display, and current offset value displayed on t he se cond line of the display;
7. Press to set the offset in 0.1 ° increments until the displayed temperature equals the known temperature reference value;
8. Press to save the offset value and proceed to Channel 2 offset (912A only), or press to save the o f fset value and exit the Setup Menu.
a. Alternativel y, to di sregard the new offset v al ue and exit the Setup M e nu,
press .
3
Probe offset measurement using an ice bath or thermowell should only be performed by personnel trained
and qualified in the use of such instruments and related metrology methods.
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
3-7
Operating Instructions
9. OFFSET is displayed at the top-left of the LCD display.
To set the probe of fset when using a calibr a ted temperature probe with a known offse t:
1. Press
2. Press three (3) times to cycle to the Channel 1 Offset parameter;
3. Observe the current o f fse t v alue displayed on the second line of the display;
4. Press to set the offset in 0.1 ° increments until the displayed offset value equals the calibrated probe offset value;
5. Press to save the offset value and proceed to Channel 2 offset (912A only), or press to save the offset value and exit the Se t up M e nu.
6. OFFSET is displayed at the top-left of the LCD display.
(1.5s)
to enter the Setup Me n u ;
a. Alternatively, to disregard the new offset value and exit the Setup Menu,
press .

Open Lead Detection Enable/Disable (SN 001201 and above only)

Open Lead Detection allows the unit to detect i f a t hermocouple probe is connected to the thermometer. This fe atur e is no t com pati bl e wit h some thermocouple calibrators and can result in measurement instability.
Disabling Open Lead D etection in these situati ons c an significantly improve reading stability. Once disabled, Open Lead Detection will remain disabled until changed by following the below steps, or the instrument is powered off.
If no thermocouple probe is connec ted an d Open Le ad Dete ction i s dis abled, the unit will not indicate OPEn and may display erratic readings.
To change the Open Le a d Detection setting:
1. Press
2. Press four (4) times for 911A, five (5) times for 912A, to cycle to the Open Lead Detection parameter;
3. Press to change the Open Lead Detection setting;
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
(1.5s)
to enter the Setup Me n u ;
a. OLd is displayed o n Line 1 of th e LC D, and the curre nt Op en Le ad
Detection status is displayed on Line 2.
a. ON indicates t hat Open Lead Detection is enabled;
3-8
sales@tegam.com
Operating Instructions
and TEGAM Cloud to upload saved data bef ore per formi ng t his acti on.
The current measurement or statistic is over-range for the selected thermocouple type
The current measurement or statistic is under-range for the selected thermocouple type
4. Press VIEW or SET to save the Open Lead Detection Setting and exit the Setu p
b. OFF indicates tha t Open Lead Detection is disabled;
Menu.
a. Alternatively, to disregard the Op e n Lead Detection settin g a nd e xi t th e
Setup Menu, press CLR.

Clear Function

From active measurem en t or hold mode, press begin a new statistics session. The LCD display will indicate CLr to confirm the actio n, and return to active measuremen t mode.
(1.5s)
to clear the statistics regi sters a nd
Pressing internal memory. To prevent data loss, connect to the Thermometer Link mobi le ap p
From the Setup Menu, p re s s to disregard change s to the cur rent p arame te r val ue and exit the Setup Menu.
(1.5s)
deletes all measurement data currently saved in the instrument’s

Invalid Measurement Indications

The LCD display indicates when a measurement or statistic is invalid, as shown in Figure 9 below.
I
NDICATION DESCRIPTION
- Or -
- Ur -
OPEn No probe is connected or the probe senso r is faul ty
———— Cannot compute a valid statistical resul t
Figure 9: Invalid Measurement Indications
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
3-9
sales@tegam.com
Service Information
PARAMETER
VALUE
Thermocouple Type
As Desired
Temperature Units
°C
Offset Ch. 1
0.0 °C
Offset Ch. 2 (912A only)
0.0 °C
As Needed (see Section 3.11, Open Lead Detection Enable/Disable)
Figure 10: UUT Verification Parameter Settings

4. SERVICE INFORMATION

Inspection and Cleaning

To extend the life o f t he instrument, inspect an d cl ean the instrument regularly. Inspect the instrument for any signi ficant abrasions, cuts, cracks, dents, or other signs o f dam age on the case, keypad, and di splay lens. Inspect the connectors for breaks, dirt, o r co rr osion. Ensure all screws are securely f astened, and i f equi pped, that the tilt stand/magnet/hanger is in good conditio n and locks into positi on p ro pe rly.
With all screws securel y f a ste ned and the battery compartmen t cov er i n place, use a damp cloth or towel to w i pe down the instrument. U se care to avoid scratching the display lens. Mild, non-abrasive detergents ma y be used providing the instrument is then wiped down with a clean damp cloth or towel.

Calibration

4.2.1 Verification Procedure

1. This procedure shall be performed within environmental conditions of 23 ±1 °C and 5% to 95% RH.
2. The unit under test (“UUT”) shall be acclimated to the controlled environment for a minimum of four (4) hours.
3. The equipment listed i n Appendix A is requi red to fully verify the UUT to the expanded instrument uncertainties specified in Appendix B.
4. Refer to Appendix C for standard measurement points and tole rance s for each thermocouple type.
5. One, several, or all the available thermocouple types may be verified as necessary. In the steps below, use the appropriate Thermocouple Cable and Thermocouple Calibrator settings as appropriate for the desired thermocouple type.
6. For two channel UUTs, bot h ch a nn el s ma y be verified concurrently.
7. Set the UUT parameters as show n in Figure 10 below. Refer to Section 3.3, Setup Menu as necessary for UUT parameter setup instructions.
Open Lead Detection Enable/Disable
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
4-1
sales@tegam.com
Service Information
8. Connect the miniature the rmocou pl e conne c tor of the Th erm ocoupl e Cable to Channel 1 of the UUT.
a. For two channel UUTs using the Spli t T he rm ocouple Cable, connect the
miniature thermocoupl e connector of one leg to th e UUT Channel 1 input, and the miniature thermocouple con nector of the other leg to the Channel 2 input.
9. Connect the opposite end of the Thermocouple Cable (or the single-connector end of the Split Thermocouple Cable) to the Thermocouple Calibrator thermocouple output.
10. Set the Thermocouple Calibrator thermocouple type to the desi re d thermocouple type.
11. Allow at least five minutes for this connection to stabilize.
12. Set the calibrator to output to the fi rs t St a ndard Value in Appendix C for the desired thermocouple type.
13. Record the UUT measurement in the Reading column of Appendix C for the appropriate Standard Value.
14. Record the cable correc tion value for the The rmo coupl e cable in the Cabl e Correction column of Appendix C.
15. Subtract the Cable Correction value from t he Reading and record the result as the Corrected Reading (Reading – Cable Correction = Corrected Reading) in Appendix C.
16. Compare the Corrected Reading to the tol erances stated in the 2 -Sigma Tolerance column of Appendix C to determine whether t he UUT measurement is wi t hi n published specifications.
17. Repeat Steps 12 through 16 for each remaining Standard Value in Appendix C for the current thermocouple ty pe.
18. Repeat Steps 7 through 17 for each desired thermocouple type.
19. If Open Lead Detection was disable d in Step 7 ab ove , en able t he fea ture as s hown in Section 3.11, Open Lead Detection Enable/Disable.

4.2.2 Alignment Procedure

Preparation
1. This procedure shall be performed within environmental conditions of 23 ±1 °C and 5% to 95% RH.
2. The unit under test (“UUT”) shall be acclimated to the controlled environment for a minimum of four (4) hours.
3. The equipment listed i n Appendix A above i s r e quired to align the UU T to the expanded instrument uncertainties specified in Appendix B.
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
4-2
sales@tegam.com
Service Information
PARAMETER
VALUE
Thermocouple Type
Type E1
Temperature Units
°C
Offset Ch. 1
0.0 °C
Offset Ch. 2 (912A only)
0.0 °C
1
Other thermocouple types may be used in this procedure as desired. For instance, if the
in Appendix B assume alignment using Type E.
Figure 11: UUT Alignment Parameter Settings
Figure 12: Alignment A cc es s Hole Location
4. Remove the UUT battery door hou sing to exp ose the alignme nt acc ess hol e.
5. Press UUT to turn t he UUT on.
6. Set the UUT parameters as shown in Figure 11 below. Refer to Section 3.3, Setup Menu as necessary for UUT parameter setup instructions.
UUT is used primarily with Type J applications, Cold Junction Compensation may be aligned using Type J. Note however that the expanded instrument uncertainties provided
7. Insert the Straighte ne d Paper Clip through the alignment access hole and gently press the calibration enable switch located on the circuit board. See Figure 12 for location.
Voltage Gain and Offset Alignment
8. The UUT display will indicate as follows:
a. Line 1: CAL1 b. Line 2: mV portion of Channel 1 voltage reading c. Line 3: nV portion of C ha nn e l 1 voltage reading
9. Connect the miniature thermocouple connector of the Copper Mini-TC Cable to the Channel 1 input of the UUT.
a. For two channel UUTs using the Spli t C op pe r M i ni -TC Cable, connect one
miniature thermocouple c onne cto r to t he Ch annel 1 inp ut of the UU T, and the other connector to the Channel 2 input.
10. Connect the opposite end of the Copper M i ni -TC Cable (or Split Coppe r Mini-TC Cable) to the appropriate output connecto rs of the DC Volt age Sou rc e.
11. Allow at least three minu tes fo r the conne c tion s to temp er atur e stab il ize before proceeding.
CAUTION
Do not apply voltages greater than 80 mV DC to the UUT inputs. Voltages greater than 80 mV may damage the instrument.
4-3
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
Service Information
CHANNEL
UUT DISPLAY
APPLIED VOLTAGE (mV)
CAL1
-10
CAL2
(75
CAL3
-10
CAL4
(30 2 CAL5
-10
CAL6
(75
CAL7
-10
CAL8
(30
Figure 13: Gain Alignment Values
12. Set the DC Voltage Source to output the first Applied Voltage value in Figure 13 below.
13. Allow the DC Voltage so ur ce output to stabilize befo re proceeding.
14. The UUT will display the current voltage reading.
15. Allow the UUT displayed voltage to stabilize before proceeding.
16. Press UUT to automa tical l y adj ust the UUT vol tage re adi ng to the Ap pli ed Voltage, ±0.001 mV.
a. If the UUT displayed voltage is not within ±0.001 mV of the Applied
Voltage, press until the UUT displayed voltage is within ±0.001 mV, adjusting as close to the Applied Voltage as possi ble.
17. Press UUT . The display will change to rES1 [2, 3 …] showing the actual measured value saved in the previous step.
18. Press UUT again. This will increment to the next CAL v al ue.
19. Repeat Steps 12 through 18 for eac h remaining value in Figure 13 below. For single channel UUTs, omit Channel 2 values.
a. For two channel UUTs using the singl e -ended Copper Mini-TC Ca ble,
disconnect the cable from the UU T Ch annel 1 inp ut, a nd reconnect the cable to the Channel 2 input after compl eti ng CAL4 in Figure 13 below. Repeat Step 11.
20. Remove the copper cable from the DC V ol t age Source and UUT.
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
1
4-4
sales@tegam.com
Service Information
Cold Junction Compensation Alignment
4
21. Connect one end o f the Type E
Thermocouple Cable to the UUT Channel 1 input.
a. For two channel UUTs using the Split T hermocouple Cable, connect the
miniature thermocouple connector of one leg to the UUT Channel 1 input, and the mini ature the rmo coupl e connector of the other leg to the Channel 2 input.
22. Connect the opposite end of the Thermocouple Cable (or the single-connector end of the Split Thermocouple Cable) to the Thermocoupl e Cal ibr ator the rmoco uple output.
23. The UUT display will indicate as follows:
a. Line 1: CALA b. Line 2: tempera tu re in °C c. Line 3: temperature in tenths of °C (out to 1 μ or 0.000001 °C)
24. Set the Thermocouple Calibrator thermocouple type to Type E
5
.
25. Set the calibrator to outpu t 0.0 °C.
26. Allow at least five minutes for this connection to stabilize.
27. Press UUT to set the UUT display equ al to the Thermocouple Cable calibrated correction value ± 0.02 °C.
28. Press UUT .
29. The display will c h an ge to rESA showing the a ctual measured value s a ved in the previous steps.
30. For single channel UUTs, skip to Step 36 below.
31. For two channel UUTs, con ti n ue wi t h Step 32.
a. For two channel UUTs using the singl e -ended Thermocouple Cable,
disconnect the cable from the UU T Ch annel 1 inp ut, a nd re conne ct the cable to the Channel 2 input. Repeat Step 26.
32. Press UUT .
33. The UUT display will indicate as follows:
a. Line 1: CALb b. Line 2: tempera tu re in °C
4
If substituting another thermocouple type, use the appropriate Thermocouple Cable for the selected
thermocouple type.
5
If substituting another thermocouple type, set the Thermocouple Calibrator as appropriate for the selected
thermocouple type.
4-5
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
Service Information
Statistics)
Set probe offset to correct value (see Section 3.10, Probe Offset)
Observe display tre nd i nd i c a to rs
Indicators)
Set the thermocouple type as
Menu)
When sourcing from a Lead Detection is enabled.
See Section 3.11, Open Lead disable.
power
Figure 14 continued on next page . . .
34. Repeat Steps 27 and 28.
35. The display will c h an ge to rESb showing the a c tu al measured value saved i n the
36. Press UUT to save the current alignment v al ue s a nd return the UUT to no r mal
c. Line 3: temperature in tenths of °C (out to 1 μ or 0.000001 °C)
previous steps.
operation.

Troubleshooting

TEGAM’s digital handheld thermomet ers ar e desig ned an d buil t to provi de ye ars of uninterrupted use. In the event the instrument malfunctions o r doe s not perform as expected, helpful trou bl esho otin g ti ps a re pr ovi ded b elow. Figure 14 below lists some of the more common issues and t hei r re s ol u ti on s.
S
YMPTOM DESCRIPTION RESOLUTION
Unexpected reading on Line 2 of Display
Unexpected or Erroneous Measurement
Unresponsive
Statistics View Mode is active
Probe offset is active
Temperature probe has not stabilized
Instrument is set to the wro ng thermocouple type for the attached probe
thermocouple simulator, Ope n
Hold Mode is active
Static discharge through connected probes
Press to cycle through statistics views until active measurement is displayed (see
Section 3.4
for connected temperature probe
and wait for stable me as urement (see Section
appropriate for the attached probe (see Section
Detection Enable/Disable to
Press , and verify that the HOLD indicator is not a cti ve (se e Section 3.7, Hold Function)
Press to cycle instrument
View Modes and
3.8 Trend
3.3, Setup
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
4-6
sales@tegam.com
Service Information
Shuts down
power on
Replace batteries (see Section
SYMPTOM DESCRIPTION RESOLUTION
unexpectedly or will not
Batteries are low or depleted
Figure 14: Common Troubleshooting Issues
2.5, Battery Installation and Replacement)

Diagnostic Routines and Error Codes

The instrument momentarily activates all display annunciators and segments during s ta rt up to allow for visual inspection of the LCD. Observe the LCD and verify all segments activate.
Internal diagnosti c ro utines are also execute d during startup. If any diagnostic routine detects a malfunction, an error will be displayed as shown in Figure 15 below.
E
RROR CODE DESCRIPTION
Err ADC Analog to digital conver ter error
Err CJC Cold junction compensation e rror
Err FLSH Flash me m ory error
Err InP Stuck key or other keypad error
Figure 15: Diagnostic Routine Error Codes

Memory Sterilization

To erase all locally stored measureme n t data and r e se t a cc umulated statistics, press See Section 3.12, Clear Function for instructions.
Instrument parameters will be retained. Refer to Section 3.3, Setup Menu to set i n s trument parameters as desired.
(1.5s)
.
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
4-7
sales@tegam.com
Service Information

Preparation for Calibration or Repair Service

Once you have verified that the cause of the malfunction cannot be solved in the field and the need for repair and calibration service arises, contact TEGAM customer service to obtain an RMA (Returned Material Authorization) number. You can contact TEGAM customer service v ia the TEGAM website, www.tegam.com 1010 (United States Only).
The RMA number is unique to your instrument and will help us ident ify you instrument and to address the particular service request by you which is assigned to that RMA number.
Of even greater importance, a detailed written description of the problem should be attached to the instrument. Many times repair turnaround is unnecessarily delayed due to a lack of repair instructions or a detailed description of the problem.
This description should include information such as measurement ra nge an d other instrument settings at the time of the malfunction, type of components being tested, frequency of the symptoms (intermittent or continuous), conditions that may cause the symptoms, changes to the test setup or operating environment that may affect the instrument, etc. Any detailed information provided to our technicians will assist them in identifying and correcting the problem in the quickest possible manner. Use a copy of the Repair and Calibration Service form provided on the next page.
Once this information is prepared and sent with the instrument to our service department, we will do our part to make sure that you receive the best possible customer service and turnaround time possible.
or by calling 440-466-6100 (All Locations) or 800-666-
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
4-8
sales@tegam.com
Service Information
RMA#:
Instrument Serial Number:
Company:
Technical Contact:
Phone Number:
Additional

Expedite Repair & Calibration Form

Use this form to provide additional repair information and service instructions. The completion of this form and including it with your instrument will expedite the processing and repair process.
Model #:
Contact Info:
Service Instructions:
Evaluation Calibration Only Repair Only Repair & Calibration ISO 17025 Calibration with Data
Detailed Symptoms:
Include information such as measurement range, instrument settings, type of components being tested, is the problem intermittent? When is the problem most frequent?, has anything changed with the application since the last time the instrument was used?, etc.
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
4-9
sales@tegam.com
Service Information

Warranty

TEGAM, Inc. warrants this product to be free from defects in material and workmanship for a period of three (3) years from the date of shipment. During this warranty period, if a product proves to be defective, TEGAM Inc., at its option, will either repa ir t he defective product without charge for parts and labor, or exchange any product that proves to be defective.
TEGAM, Inc. warrants the calibration of this product for a period of two (2) years from date of shipment. During this period, TEGAM, Inc. will recalibrate any product, which does not conform to the published accuracy specifications.
In order to exercise this warranty, TEGAM, Inc., must be notified of the defective product before the expiration of the warranty period. The customer shall be responsible for packaging and shipping the product to the designated TEGAM service center with shipping charges prepaid. TEGAM Inc. shall pay for the return of the product to the customer if the shipment is to a location within the country in which the TEGAM service center is located. The customer shall be responsible for paying all shipping, duties, taxes, and additional costs if the product is transported to any other locations. Repaired products are warranted for the remaining balance of the original warranty, or 90 days, whichever is greater.

Warranty Limitations

The TEGAM, Inc. warranty does not apply to defects resulting from una u t horized modification or misuse of the product or any part. This warranty does not apply to fuses, batteries, or damage to the instrument caused by battery leakage.
The foregoing warranty of TEGAM is in lieu of all other warranties, expressed or implied. TEGAM specifically disclaims any implied warranties of merchantability or fitness for a particular purpose. In no event will TEGAM be liable for special or c onsequential damages. Purchaser’s sole and exclusive remedy in the event any item fai ls t o comply with the foregoing express warranty of TEGAM shall be to return the item to TEGAM; sh ipping charges prepaid and at the option of TEGAM obtain a replacement item or a refund of the purchase price.

Statement of Calibration

This instrument has been inspected and tested in accordance with specifications published by TEGAM, Inc.
TEGAM, Inc. certifies the above listed instrument has been inspected and calibrated and meets or exceeds all published specifications and has been calibrated using standards whose accuracies are traceable to the International System of Units (SI) through t he National Institute of Standards and Technology (NIST) or other recognized National Metrology Institutes.
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
4-10
sales@tegam.com
Appendices
T
SIGMA)
DC Voltage Source
DC Voltage Output
± (13 ppm of output + 0.66 µV)
N/A
-250 to -100 °C
650 to 1000 °C
± 0.33 °C
± 0.14 °C
± 0.47 °C
± 0.20 °C
-210 to -100 °C
760 to 1200 °C
± 0.18 °C
± 0.15 °C
± 0.25 °C
± 0.22 °C
-200 to -100 °C
1000 to 1372 °C
± 0.22 °C
± 0.27 °C
± 0.31 °C
± 0.38 °C
-250 to -150 °C
120 to 400 °C
± 0.42 °C
± 0.09 °C
± 0.59 °C
± 0.13 °C
each leg of the Split Thermocouple Cable.
male miniature thermocouple copper connectors.
Straightened Paper Clip
Required to access the calibration enable switch. Any rigid wire, approximately 0.8 mm in diameter, may be used.

A. REQUIRED EQUIPMENT

EQUIPMENT FUNCTION RANGE
-10 to 75 mV
S
PECIFICATION
(2-SIGMA)
HERMOCOUPLE
CABLE CALIBRATION UNCERTAINTY (2-
Thermocouple Calibrator
Thermocouple Cables
Copper Mini-TC Cable
Thermocouple Type E
Thermocouple Type J
Thermocouple Type K
Thermocouple Type T
Calibrated Thermocouple Cables with recorded correction values shall be used throughout this procedure to adhere to the expanded instrument uncertainties provided in Appendix B. Thermocouple Cables must be calibrated to the accuracies listed in the Thermocouple Calibrator Specification column of Appendix A, or better.
Thermocouple Cables are only required for each desired thermocouple type. The thermocouple cables must be terminated at one end with a male miniature
thermocouple connector for connection to the UUT. The opposite end should be terminated as appropriate for the thermocouple calibrator.
For two channel UUTs, a split or “Y” cable may be used, terminated with two (2) male miniature thermocouple connectors. Correction values must be established for
Copper Mini-TC Cable is required for Voltage Gain and Offset alignment only. This cable does not require calibration.
One end shall be terminated with a male miniature thermocouple connector for connection to the UUT. The opposite end shall be terminated with copper connections appropriate for the DC Voltage Source.
For two channel UUTs, a split or “Y” cable may be used, terminated with two (2)
-100 to -25 °C
-25 to 350 °C 350 to 650 °C
-100 to -30 °C
-30 to 150 °C 150 to 760 °C
-100 to -25 °C
-25 to 120 °C 120 to 1000 °C
-150 to 0 °C 0 to 120 °C
± 0.11 °C ± 0.09 °C ± 0.11 °C
± 0.11 °C ± 0.09 °C ± 0.11 °C
± 0.12 °C ± 0.11 °C ± 0.17 °C
± 0.16 °C ± 0.11 °C
± 0.15 °C ± 0.13 °C ± 0.15 °C
± 0.15 °C ± 0.13 °C ± 0.16 °C
± 0.17 °C ± 0.15 °C ± 0.25 °C
± 0.23 °C ± 0.15 °C
A-i
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
Appendices

B. EXPANDED INSTRUMENT UNCERTAINTIES

Thermocouple Type E
B-i
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com
Appendices
Thermocouple Type J
B-ii
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com
Appendices
Thermocouple Type K
B-iii
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com
Appendices
Thermocouple Type T
B-iv
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ● sales@tegam.com
Appendices
T
TYPE
R
C (°C)
C R
2-S T °C)
-250
2.1
-95
0.4
0
0.3
995
0.5
-210
1.1
-95
0.4
0
0.3
995
0.6
1200
1.1
-200
1.1
-95
0.4
0
0.3
995
0.6
1372
1.2
-250
2.2
-95
0.5
0
0.3
400
0.3

C. INSTRUMENT VERIFICATION DATA SHEET

HERMO
-COUPLE
E
J
K
T
STANDARD VALUE (°C)
EADING (°C)
ABLE OFFSET
ORRECTED EADING (°C)
IGMA
OLERANCE
C-i
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com
GENEVA, OHIO 44041
CAGE Code: 49374
WEB: http://www.tegam.com
TEGAM INC.
10 TEGAM WAY
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
sales@tegam.com