Tegam 4040A Instruction Manual

4040A
50 MHz PXI Differential
Instrumentation Amplifier
MODEL 4040A
NOTE: This User’s Manual was as current as possible when this product was manufactured. However, products
are constantly being updated and improved. To ensure you have the latest documentation, refer to
www.tegam.com.
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com
Instruction Manual
PN# 4040A-840
Publication Date: June 2012
REV. E
TABLE OF CONTENTS
TABLE OF CONTENTS
1 INSTRUMENT DESCRIPTION
Instrument Description .............................................................................1-1
Figure 1.1: Block Diagram of the 4040A .................................................1-1
Performance Specifications ........................................................................1-2
Functional Considerations ..........................................................................1-4
Frequency Response Curves ......................................................................1-5
2 PREPARATION FOR USE
Unpacking & Inspection.............................................................................2-1
Safety Information & Precautions ...............................................................2-1
Terms in this Manual ...........................................................................2-1
Terms as Marked on Equipment ............................................................2-1
Grounding the 4040A ...........................................................................2-2
Danger Arising from the Loss of Ground .................................................2-2
Do not Use in Explosive Environments ...................................................2-2
Power Source .....................................................................................2-3
3 OPERATING INSTRUCTIONS
PXI Installation ........................................................................................3-1
Front Panel Description .............................................................................3-2
4 SOFTWARE
5 SERVICE INFORMATION
Cleaning .................................................................................................5-1
Performance Verification Procedure ............................................................5-1
Computer Set-up ................................................................................5-2
CMMR ................................................................................................5-2
Offset Adjustment ...............................................................................5-4
DC Gain .............................................................................................5-6
AC Gain .............................................................................................5-7
Parts Replacement ...................................................................................5-12
Preparation for Calibration or Repair Service ................................................5-12
Expedite Repair and calibration Form ..........................................................5-13
Warranty .................................................................................................5-14
Warranty Limitations ...........................................................................5-14
Statement of Calibration ......................................................................5-14
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com
I
INSTRUMENT DESCRIPTION
SECTION 1
INSTRUMENT DESCRIPTION
The 4040A is a single channel, differential input amplifier capable of high gain and equally high attenuation values with a bandwidth from DC-50 MHz. The Model 4040A is capable of operating with gains of x1, x10, and x100 in conjunction with attenuation by factors of: ÷1, ÷10, and ÷100. These levels of gain and attenuation may be used in combination to condition a differential input signal so that it matches the needed input of a low-level device such as a digitizer or oscilloscope. There is a more in depth explanation contained in the specification section. In addition, the 4040A provides programmable offset, anti-aliasing filters, input impedance and input coupling. Figure 1.1 is a block diagram of the 4040A. The 4040A has been specifically designed to take two signals and make a valid differential measurement by inverting one of the signals and adding the difference, this allows the user to measure signals not referenced to local ground without compromising safety.
Figure 1.1: Block Diagram of the 4040A
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com
1-1
INSTRUMENT DESCRIPTION
Net Gain
Input
Attenuation
Internal
Amplifier
Gain
Peak AC Input
Amplitudes
(V) d per channel
a,b,c,d
Max
Differential
Voltage (V)
w/o clipping
a,b,c,d
Max
Volts to
Chassis
a,b,c
Noise,
Referred to
Input
-3 dB
Bandwidth
÷100
÷100
X1
< 100
< 100
100 V
990 nV/√Hz
20 MHz
÷10
÷10
X1
< 10
< 10
40 V
99 nV/√Hz
20 MHz
÷10
÷100
X10
< 10
< 10
100 V
990 nV/√Hz
50 MHz 1 ÷1
X1
< 1
< 1
4 V
9 nV/√Hz
20 MHz
1
÷10
X10
< 1
< 1
40 V
99 nV/√Hz
50 MHz
1
÷100
X100
< 1
< 1
100 V
990 nV/√Hz
20 MHz
10
÷1
X10
< 0.1
< 0.1
4 V
9 nV/√Hz
50 MHz
10
÷10
X100
< 0.1
< 0.1
40 V
99 nV/√Hz
20 MHz
100
÷1
X100
< 0.01
< 0.01
4 V
9 nV/√Hz
20 MHz
a. DC Coupled, 1 MΩ input b. AC Coupled, 1 MΩ input: DC + Peak AC not to exceed 100 V; Peak AC component not to exceed table c. 5 Vrms max into 50 Ω d. DAC offset adjustment to zero
PERFORMANCE SPECIFICATIONS
The multiple gain and attenuation settings of the 4040A interact to affect the performance in terms of bandwidth and noise. Table 1.1 shows the various combinations and how these settings modify what can be expected from the 4040A.
Table 1.1: Various Combination Possibilities of the 4040A
As an example, a total system gain of X1 can be achieved three different ways depending on the signal requirements through attenuating and amplifying by different amounts. Choosing the best combination involves trade offs in input amplitude, bandwidth and signal to noise ratio. Noise increases with higher levels of attenuation because the noise specification is referred to the input. In relative terms the signal to noise ratio is basically constant.
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com
1-2
INSTRUMENT DESCRIPTION
Input Specifications Value Clarification
Channels Single Channel Differential Input Gains 100, 10, 1, 0.1, 0.01 Maximum Voltage Range ±100 V DC + Peak AC Coupling AC, DC In AC 10 Hz and above Input Impedance 1 MΩ | | 20 pF, 50 Ω ±1%, Selectable Input Voltage Range ±100 V For Gain 0.01 @ 1 MΩ Input Impedance
±10 V For Gain 0.1 @ 1 MΩ Input Impedance ±5 V For Gain 0.1 @ 50 Ω Input Impedance ±1 V For Gain 1 ±0.1 V For Gain 10 ±0.01 V For Gain 100
Connection Type BNC Jacks 50 Ω, Quantity 2
Common Mode Rejection Ratio (CMRR) 77 dB at 60 Hz For Gain setting of x1, x10 and x100 & Attenuation (> 50 dB at 1 MHz) setting of ÷1 57 dB at 60 Hz For Attenuation setting of ÷10 37 dB at 60 Hz For Attenuation setting of ÷100
Total Harmonic Distortion <-60 db @ 1 MHz Output 1 Vp-p into 50 Ω DC Gain Accuracy ± (0.1% |input| + 100 µV) Offset set to 0 for Gain 10, 1, 0.1, 0.01
± (1.5% |input| + 300 µV) Offset set to 0 for Gain 100 (Basic) AC Gain Accuracy ±0.15 dB Attenuation of 1, Gain of 1,
See Figures 1.2, 1.3 and 1.4 for other settings Over-voltage Protection ±100 V DC + Peak AC
(In Any Range) Offset Range 0 to ± Full Scale Full Scale output of ±1 V into 50 Ω,
All Gain Ranges Offset Resolution 38 µV per step 65,535 total DAC steps into 50 Ω Offset Accuracy ±(0.5% of |Setting| + 300 µV) Referenced to 1 V Range Temperature Stability ±(0.01% of |rdg| + 40 µV)/°C All Gain Ranges Noise 9 nV/Hz CMR=±1 V, Gain 10 and 100
Referred to Input for Frequencies >100 Hz Rise Time ≤3.5 ns
Output Specifications Value Clarifications
Maximum Output Voltage ±1 V Single Ended into 50 Ω Connection Type SMB Jack 50 Ω Output Impedance 50 Ω ±1% LP Filter, Cutoff Frequency 100 kHz, 1 MHz Single Pole Filter Bandwidth See Table 1.1 Passband Flatness See Figures 1.2, 1.3 and 1.4 Limits off charts will not exceed ±3 dB for the
for limits Bandwidths listed in Table 1.1
Environmental Specifications Value
Operating Temperature 0 °C to +45 °C, (32 °F to 113 °F) Ambient Storage Temperature -20 °C to +50 °C (-4 °F to +122 °F) Humidity Range <80% RH Non-Condensing Warm-up Time 30 minutes
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com
1-3
INSTRUMENT DESCRIPTION
Functional Considerations
Voltage & Current Limitations
In general, the maximum input voltage of the 4040A should not exceed ±100 V (200 Vp-p) when the
input terminator is set to 1 MΩ. However, some settings will produce incorrect results at lower
voltages. This occurs when the input signal or gain setting is too high and causes the output amplifier to clip the signal. An example of this would be measuring a 10 V signal with the gain set to x100. Mathematically this results in a 1000 V signal which clearly exceeds the 2 Vp-p rating of the output amplifier. See Table 1.1 for a table of appropriate input voltages for given combinations of gain and attenuation. No damage would occur to the 4040A or to a digitizer connected to it if the input range was exceeded, but the signal would be distorted. A signal in excess of 100 V would cause the onboard voltage limiting circuitry to activate and also create distortion. There are other considerations when the 4040A is configured with the 50 Ohm input termination. The input voltage is limited to a maximum of ±5 Vrms. This is due to the power limitations of the terminator which is 2 W. To preserve the signal integrity and bandwidth of the amplifier the 50 Ohm terminator is protected with surface mount single use fuses. Important: Exceeding the rating of the input in this case requires the 4040A to be returned to the factory for service.
Frequency Characteristics
The 4040A is capable of amplifying mill-volt level signals with a bandwidth from DC to 50 MHz. Some combinations of gain and attenuation will limit the bandwidth to 20 MHz. See Table 1.1 for a complete list of settings and the expected bandwidth. In addition, AC coupling rolls off the low frequencies at 10 Hz. Two anti-aliasing low pass filters are included and tuned for 100 KHz and 1 MHz. These are first order filters that roll off at 20 dB per decade.
Offset Adjustment
The 4040A includes a programmable offset adjustment that operates on the signal after the gain stage. The resolution of the adjustment is 38 µV and can offset the signal ±1 V which is the entire range of the output amplifier. It is not able to effectively offset a signal that exceeds the input range of a given gain and attenuation setting.
Selecting Gains and Ranges
The amplitude of the signal being measured is the primary consideration when selecting the appropriate gain and attenuation settings. As an example: A 100 mV signal that is riding on 42 V can be effectively measured with an attenuation setting of ÷100 and a gain setting of x10. This is because the differential operation of the 4040A rejects the 42 V common mode signal and only amplifies the difference producing an output signal of 1 V. If the signal of interest was the 42 V itself the 4040A should be configured with an attenuation of ÷100 and a gain of x1 producing an output signal of 0.42 V. The 4040A attenuates the signal prior to amplification.
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com
1-4
INSTRUMENT DESCRIPTION
Gain Deviation from Nominal (No Attenuation)
(Typical Data and Test Limits)
-0.5
-0.4
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
0.4
0.5
1.0E+03 1.0E+04 1.0E+05 1.0E+06 1.0E+07 1.0E+08
Frequency (Hz)
Gain Error (dB)
1/1 Li mit
10/1 Li mi t
100/1 Li mi t
1/1 POS
1/1 NEG
10/1 POS
10/1 NEG
100/1 POS
100/1 NEG
Figure 1.2: Frequency Response Curves
Figure 1.2 depicts the typical frequency response curves from the nominal gain values with no attenuation applied by the 4040A. The three different black bands, above and below the curves, give the allowable limits for passband flatness for the different gain/attenuation settings. Individual plots are given for both the positive and negative inputs of the 4040A for all gain/attenuation settings. For example, to find the expected passband flatness from 10 kHz to 10 MHz to achieve an inverted gain of 100, the 100/1 NEG plot would be used in conjunction with the solid black limit lines. These lines are used because the first number is for the gain, and the second number is for attenuation, POS for positive, and NEG for negative. The expected passband flatness would be ±0.25 dB from 10 kHz to 6 MHz and +0.25 dB to -3 dB from 6 MHz to 10 MHz.
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com
1-5
INSTRUMENT DESCRIPTION
Gain Deviation from Nominal ÷10 Attenuation
(Typical Data and Test Limits)
-0.5
-0.4
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
0.4
0.5
1.0E+03 1.0E+04 1.0E+05 1.0E+06 1.0E+07 1.0E+08
Frequency (Hz)
Gain Error (dB)
1/10 Li mi t
10/10 Li mi t
100/10 Limi t
1/10 POS
1/10 NEG
10/10 POS
10/10 NEG
100/10 POS
100/10 NEG
Figure 1.3 depicts the typical frequency response curves from the nominal gain values with an attenuation of 10 applied by the 4040A. As with Figure 1.2, the three different black bands, above and below the curves, give the allowable limits for passband flatness for the different gain/attenuation settings. Individual plots are given for both the positive and negative inputs of the 4040A for all gain/attenuation settings. For example, to find the expected passband flatness from 1 kHz to 7 MHz to achieve an inverted gain of 10, the 100/10 NEG plot would be used in conjunction with the solid black limit lines. These lines are used because the first number is for the gain, and the second number is for attenuation, POS for positive, and NEG for negative. The expected passband flatness would be ±0.3 dB
from 1 kHz to 5 MHz and +0.3 dB to -3 dB from 5 MHz to 7 MHz.
Figure 1.3: Frequency Response Curves
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com
1-6
INSTRUMENT DESCRIPTION
Gain Deviation from Nominal ÷100 Attenuation
(Typical Data and Test Limits)
-0.5
-0.4
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
0.4
0.5
1.0E+03 1.0E+04 1.0E+05 1.0E+06 1.0E+07 1.0E+08
Frequency (Hz)
Gain Error (dB)
1/100 Li mi t
10/100 Limi t
100/100 Limi t
1/100 POS
1/100 NEG
10/100 POS
10/100 NEG
100/100 POS
100/100 NEG
Figure 1.4: Frequency Response Curves
Figure 1.4 depicts the typical frequency response curves from the nominal gain values with an attenuation of 100 applied by the 4040A. As with Figure 1.2 & 1.3, the three different black bands, above and below the curves, give the allowable limits for passband flatness for the different gain/attenuation settings. Individual plots are given for both the positive and negative inputs of the 4040A for all gain/attenuation settings. For example, to find the expected passband flatness from 100 kHz to 25 MHz to achieve a gain of 1, the 100/100 POS plot would be used in conjunction with the solid black limit lines. These lines are used because the first number is for the gain, and the second number is for attenuation, POS for positive, and NEG for negative. The expected passband flatness would be ±0.3 dB from 100 kHz to 5 MHz and +0.3 dB to -3 dB from 5 MHz to 20 MHz. 20 MHz is the bandwidth limit for the 4040A in this setting. If 25 MHz is needed a setting of 10/10 or 1/1 can be used as long as the input voltages for those ranges are not exceeded.
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110 • sales@tegam.com
1-7
Loading...
+ 20 hidden pages