Tegam 3550 Instruction Manual

Page 1
TEGAM Inc.
Model 3550
42.0Hz-5.00MHz
Fully Programmable
LCR Meter
© Copyright 2004, TEGAM, Inc. All rights reserved.
NOTE: This user’s manual was as current as possible when this product was manufactured. However, products are constantly being updated and improved. Because of this, some differences may occur between the descriptions in this manual and the product received. Please refer to www.tegam.com for future updates of this manual.
Model 3550 Programmable LCR Meter Instruction Manual a
MODEL 3550
Instruction Manual
Publication Date April 2004
REV. B
Page 2
Model 3550 Programmable LCR Meter Instruction Manual b
Page 3
Table of Contents
I INSTRUMENT DESCRIPTION............................................................................1-1
Instrument Description ........................................................................... 1-1
Feature Overview ................................................................................... 1-1
3550 Options and Accessories .................................................................. 1-4
Performance Specifications ...................................................................... 1-5
Table 1.1 – Model 3550 Measurement Range Limits ............................... 1-5
Table 1.2 Model 3550 Measurement Frequencies ................................. 1-7
Table 1.3 Model 3550 Measurement Amplitudes .................................. 1-7
Table 1.4 Model 3550 Amplitude Accuracies ....................................... 1-8
Table 1.5 Model 3550 Measurement Ranges ....................................... 1-8
Table 1.6 Typical Analog Measurement Times ..................................... 1-9
Table 1.7a Absolute Comparator Operation......................................... 1-11
Table 1.7b Percentage Comparator Operation ..................................... 1-11
Table 1.8 – Numeric Displays and Annunciators ..................................... 1-11
Formulas and Measurement Accuracy........................................................ 1-13
Table 1.9 – Series and Parallel Equivalent Circuit Measurements .............. 1-14
II PREPARATION FOR USE ..................................................................................2-1
Unpacking and Inspection........................................................................ 2-2
Safety Information and Precautions .......................................................... 2-2
Servicing Safety Summary ...................................................................... 2-4
Line Voltage and Fuse Selection ............................................................... 2-5
III QUICK START INSTRUCTIONS .........................................................................3-1
Power the Unit ....................................................................................... 3-2
Instrument Settings................................................................................ 3-2
Table 3.1 Factory Default Settings .................................................... 3-2
Table 3.2 Control Command Summary............................................... 3-3
Table 3.3 Shift Function Summary .................................................... 3-4
Instrument Setup for Basic Measurements ................................................. 3-5
Table 3.4 Settings Summary ............................................................ 3-5
Setting the Absolute Comparator.............................................................. 3-6
Setting the Percent Comparator ............................................................... 3-7
IV OPERATING INSTRUCTIONS............................................................................4-1
Basic Operation...................................................................................... 4-2
Default Parameters................................................................................. 4-2
Precautions ........................................................................................... 4-2
Measurement Tips .................................................................................. 4-2
Making Measurements that are Sensitive to Voltage and Current.............. 4-2
Measuring Earth Grounded Test Pieces ................................................. 4-2
Measurement Circuit Modes ................................................................ 4-3
Table 4.1 – Relationship Between Series and Parallel Equivalent Circuits ... 4-3
Negative Capacitance or Inductance Readings ....................................... 4-4
Zero Correction of Parasitic Impedance ................................................ 4-4
Table 4.2 – Short and Open Null Connections ........................................ 4-4
Model 3550 Programmable LCR Meter Instruction Manual c
Page 4
Table of Contents
IV OPERATING INSTRUCTIONS CONT’D...............................................................
Measurement Tips ..................................................................................
Test Lead Requirements..................................................................... 4-5
Figure 4.1a Four-Wire Kelvin Measurement ........................................ 4-5
Figure 4.1b – Proper Application of Four Wire Kelvin Measurement ........... 4-5
Measurement Signal Levels................................................................. 4-6
Series or Parallel Circuit Modes............................................................ 4-6
Auto range Switching Threshold .......................................................... 4-6
Display Resolution ............................................................................. 4-6
AC Resistance of Coils with Cores ........................................................ 4-6
Front Panel Description ........................................................................... 4-7
Figure 4.2 – Front Panel Layout........................................................... 4-7
Power Switch.................................................................................... 4-8
Circuit Mode Panel............................................................................. 4-8
Trigger Panel .................................................................................... 4-8
Zero Panel ....................................................................................... 4-9
Range Selection Panel........................................................................ 4-9
Shift Key.......................................................................................... 4-10
Frequency Set Mode........................................................................ 4-10
Test Voltage Set Mode..................................................................... 4-11
Test Current Set Mode..................................................................... 4-11
Averaging Set Mode ........................................................................ 4-12
Key Lock Mode ............................................................................... 4-12
Comparator buzzer Function............................................................. 4-13
Trigger Delay Time.......................................................................... 4-13
Comparator Functions Mode ............................................................. 4-14
Defining Spot Frequencies for Zero Adjustments ................................. 4-14
RS-232C Settings Mode ................................................................... 4-15
Correction Frequency Limit Setting Mode............................................ 4-16
Spot Correction Mode (Open Correction) ............................................ 4-16
Spot Correction Mode (Short Correction) ............................................ 4-16
Unknown Terminals Panel................................................................... 4-17
Table 4.3 Unknown Terminals........................................................... 4-17
Comparator Limit Set Panel ................................................................ 4-17
The Guard Terminal ........................................................................... 4-17
Remote Interface Status LEDs............................................................. 4-18
Panel Selector Switch......................................................................... 4-18
Panel Display LED.............................................................................. 4-18
Display A Parameter Selection Key....................................................... 4-18
Display A ......................................................................................... 4-18
Bin Display LED................................................................................. 4-18
Comparator State Annunciators........................................................... 4-19
Sample LED (Trigger Status)............................................................... 4-19
Comparator ON/OFF Switch ................................................................ 4-19
Display B Parameter Selection Key....................................................... 4-19
Display B ......................................................................................... 4-19
Model 3550 Programmable LCR Meter Instruction Manual d
Page 5
Table of Contents
IV OPERATING INSTRUCTIONS CONT’D...............................................................
Front Panel Description Cont’d ............................................................
Display C Measurement Panel ............................................................. 4-20
Table 4.4 CV/CC Indicators .............................................................. 4-21
Display C ......................................................................................... 4-21
Enter Key......................................................................................... 4-21
Rear Panel Description ............................................................................ 4-22
Figure 4.3 – Rear Panel Layout............................................................ 4-22
Making Accurate Measurements ............................................................... 4-23
Connections to the Device Under Test .................................................. 4-23
3-Terminal Measurement.................................................................... 4-23
4 or 5-Terminal Measurements............................................................ 4-23
Test Fixture Selection ........................................................................ 4-24
Table 4.4 – Impedance and Frequency Limitations of Kelvin Klip Leads ..... 4-24
Table 4.5 – Impedance and Frequency Limitations of the SMD Test Fixture 4-25
Zero Corrections ............................................................................... 4-26
Table 4.6 Zero Correction Limits ....................................................... 4-26
Figure 4.4 – Terminal Connections for Zero Correction............................ 4-27
Equivalent Circuits............................................................................. 4-28
Table 4.7 Equivalent Circuit Relationships .......................................... 4-29
Operating the 3550 ................................................................................ 4-30
Preparation ...................................................................................... 4-30
How to Perform a Zero Correction........................................................ 4-30
Measuring L, C, and R........................................................................ 4-31
Measuring |Z| and |Y|........................................................................ 4-33
Measuring D, Q, Rs, Rp, G, X, B, and LJ................................................. 4-34
The Comparator..................................................................................... 4-35
Important Notes for the Absolute and Comparator Functions ................... 4-35
Absolute Value Settings...................................................................... 4-36
Percent Value Settings ....................................................................... 4-37
Bin Operation ................................................................................... 4-39
Figure 4.5 – Sequenced Bin Sorting ..................................................... 4-39
Figure 4.6 – Nested Bin Sorting........................................................... 4-40
Outputting the Comparator’s Results.................................................... 4-41
Audible Buzzer Operation ................................................................... 4-41
Model 3550 Programmable LCR Meter Instruction Manual e
Page 6
f
Table of Contents
V PROGRAMMING AND INTERFACING ................................................................5-1
Interfacing to the 3550 ........................................................................... 5-2
Control I/O Connector............................................................................. 5-2
Inputs ............................................................................................. 5-2
Figure 5.1 – Input Schematic .............................................................. 5-2
Outputs ........................................................................................... 5-3
Figure 5.2 – Output Schematic ............................................................ 5-3
Timing Diagrams............................................................................... 5-4
Figure 5.3a Test Busy Operation ....................................................... 5-4
Figure 5.3b Measure End Operation................................................... 5-4
Figure 5.3c GO/NO-GO Operation ..................................................... 5-4
Table 5.1 Connector Pin Designations ................................................ 5-5
Table 5.2 Control I/O Function Summary ........................................... 5-6
Sample Diagram of External Connections to Sequencer........................... 5-7
RS-232C ............................................................................................... 5-8
Interface Cable ................................................................................. 5-8
Input Format .................................................................................... 5-8
Table 5.3a Command Summary........................................................ 5-11
Table 5.3b Command Summary cont’d .............................................. 5-12
Table 5.3c Command Summary cont’d............................................... 5-13
Table 5.4 - Comparator Instructions..................................................... 5-14
Data Output Format for Measurement Settings ...................................... 5-15
Table 5.5a – Data Output Code Functions ............................................. 5-16
Table 5.5b – Data Output Code Functions cont’d .................................... 5-17
Monitor Output Format....................................................................... 5-18
Table 5.6c Monitor Output Code Functions.......................................... 5-21
Table 5.6b – Monitor Output Code Functions cont’d ................................ 5-22
RS-232C Hardware and Computer Settings ........................................... 5-23
Sample Programs for RS-232C ............................................................ 5-23
The GPIB Interface (Optional) .................................................................. 5-28
General Description of GPIB................................................................ 5-28
Primary Specifications of GPIB ............................................................ 5-28
Table 5.7 GPIB Functions ................................................................. 5-28
Installation of the GPIB Board ............................................................. 5-29
Figure 5.4a – Rear View ..................................................................... 5-29
Figure 5.4b – Side View ..................................................................... 5-30
Figure 5.4c – Top View....................................................................... 5-30
Front Panel Status LEDs ..................................................................... 5-31
Local Key Switch ............................................................................... 5-31
GPIB Switches .................................................................................. 5-31
Delimiter Output Formats ................................................................... 5-31
Output Status Bits for Service Requests................................................ 5-31
Figure 5.5 – GPIB Status Bits.............................................................. 5-31
Additional Notes for Using the GPIB Interface ........................................ 5-31
Programming Examples...................................................................... 5-33
Model 3550 Programmable LCR Meter Instruction Manual
Page 7
Table of Contents
VI SERVICE INFORMATION..................................................................................6-1
Warranty .............................................................................................. 6-2
Warranty Limitations .............................................................................. 5-2
Statement of Calibration ......................................................................... 6-2
Contact Information ............................................................................... 6-2
Repair Parts .......................................................................................... 6-3
Troubleshooting .................................................................................... 6-4
Preparation for Repair or calibration Service............................................... 6-5
Expedite Repair and calibration Form ........................................................ 6-6
Performance Verification Procedure........................................................... 6-7
Model 3550 Calibration Adjustment Procedure ............................................ 6-12
VII APPENDIX .......................................................................................................A.1
Setting the Constant Voltage Mode ........................................................... A.3
To Enable the Constant Voltage Mode................................................... A.3
Calculating the maximum Allowable Constant Voltage Setting.................. A.3
Setting the Constant Current Mode ........................................................... A.4
Calculating the maximum Allowable Constant Current Setting.................. A.4
Basic Measurement Accuracy Tables ......................................................... A.5
Table A.1 Basic Accuracy Table ......................................................... A.5
Table A.2 – Basic Accuracy Table [0.01V to 0.04V]................................. A.6
Table A.3 – Basic Accuracy Table [0.05V to 0.10V]................................. A.7
Table A.4 – Basic Accuracy Table [0.11V to 0.20V]................................. A.8
Table A.5 – Basic Accuracy Table [0.21V to 0.45V]................................. A.9
Table A.6 – Basic Accuracy Table [0.46V to 1.00V]................................. A.10
Table A.7 – Basic Accuracy Table [1.01V to 5.00V]................................. A.11
LC Impedance Calculation Chart ............................................................... A.12
Error Codes ........................................................................................... A.13
Model 3550 Programmable LCR Meter Instruction Manual g
Page 8
Model 3550 Programmable LCR Meter Instruction Manual h
Page 9
Instrument Description
INSTRUMENT DESCRIPTION
PREPARATION FOR USE
QUICK START INSTRUCTIONS
OPERATING INSTRUCTIONS
PROGRAMMING & INTERFACING
SERVICE INFORMATION
APPENDIX
Page 10
Model 3550 LCR Meter Instruction Manual Instrument Description
Instrument Description
The Model, 3550, is a high-speed LCR Meter capable of measuring Inductance (L), Capacitance(C), Resistance (R), Absolute Impedance (|Z|), Absolute Admittance (|Y|), Dissipation factor (D), Quality (Q), Serial Equivalent Resistance (Rs), Parallel Equivalent Resistance (Rp), Conductance (G), Reactance (X), Susceptance (B), and Phase Angle (LJ). It has a wide measurement range, and includes full adjustability of Measurement Frequency and Measurement Signal Level. It also features Automatic adjustment for Zero Correction, Range Switching, and Series and Parallel Mode Switching.
The basic comparator function of the 3550 is extended to allow sorting of components to up to 10 different bins. Up to 9 panels worth of measurement settings can be stored in the instrument’s non-volatile memory for convenient recall. The instrument can perform a complete measurement including a comparator judgment in as little as 40ms.
The 3550 has standard RS-232C and Control I/O Connector Interfaces with the availability of an optional GPIB Interface. The Control I/Os are optically isolated thus making the instrument ideal for noise immunity.
Feature Overview
The 3550 Programmable LCR Meter offers a complete solution for a wide variety of specialized LCR measurement applications. Listed below are some of the features.
High Speed
The measurement time, including data display, zero correction, and comparator function is approximately 40 milliseconds.
Extended Frequency Range
The Measurement Frequency can be adjusted between 42.0Hz - 5.00MHz.
Programmable Test Voltage and Current
The test signal level can be adjusted between 10mV to 5.00V in the constant voltage mode and from 10µA to 99.99mA in the constant current mode.
Voltage & Current Monitor
The voltage at the test piece terminals and the current, which is flowing through the test piece, can be displayed up to 3 significant digits.
1-2
Page 11
Model 3550 LCR Meter Instruction Manual Instrument Description
Feature Overview cont’d:
1
/2 Digits Resolution
4
"DISPLAY A" (L, C, R, |Z|, |Y|), and "DISPLAY B" (D, Q, R up to 4
1
/2digits.
GO/NO-GO COMPARATOR
For L, C, R, |Z|, |Y|, D, Q, R
, RP, G, X, B, and LJ, upper and lower limit values can be set
S
for 9 items (10 categories), for up to 9 panels of settings.
Series or Parallel Equivalent Circuit Mode
When in "Auto" mode, Parallel or Series Equivalent Circuit measurements are automatically determined by the active measurement range. In manual range, the Parallel or Series Equivalent Circuit mode are user selectable.
, RP, G, X, B, LJ) can both display
S
Front Panel Key Lock
The Key Switch on the front panel can be set to lock the instrument settings for protection against accidental changes.
Automatic Zero Correction of Parasitic Impedance
Performing Open and Short Circuit Zeroing cancel the offset errors caused by Stray Capacitance, Stray Conductance, Residual Inductance, and Residual Resistance.
Non-Volatile Memory
All measurement settings such as the Zero Correction Values, Comparator Limits, Panel Settings etc. are stored in non-volatile memory and are retained in the event of power loss.
Remote Operation
The TEGAM Model 3550 provides standard RS-232C and I/O Control Connector interfaces for remote operation. There is an optional GPIB (IEEE-488) interface available for purchase.
1-3
Page 12
Model 3550 LCR Meter Instruction Manual Instrument Description
3550 Options and Accessories
2005B - Chip Tweezers (5ft)
This four-terminal tweezer set makes solid connections to chip components in manual sorting applications. Capacity of jaws is 12.7mm (0.5 in.). The 2005B Chip Component Tweezer Set includes a 1.5m (5 ft) cable for connection to the 2150/2160. Contact tips are replaceable. P/N 47422.
47422 Chip Tweezer Replacement Kit
Tweezer tips are intended to last 100,000 to 500,000 operations. An optional tip replacement kit includes 12 replacement tips, 2 screws and 1 wrench.
3510-Radial Lead Adapter
This sorting fixture allows for efficient four-wire measurement of leaded parts. The test fixture features spring action contacts for easy insertion and removal of test components.
47454 – Kelvin Klips
Kelvin Klips allow solid four-terminal, Kelvin connections to leaded components. The jaws are assembled with hardened gold-plated, beryllium copper, which ensures low contact resistance, low thermal emf to copper, high corrosion resistance, and long service life. An alligator clip is provided allowing connection of a passive guard. The assembly includes a 5 ft (1.5m) cable for connection to the 3525.
KK100- Kelvin Klip™ Rebuild Kit
Kelvin Klip™ replacements for construction or repair of Kelvin Klip leads.
GPIB (IEEE-488) Cables
1583-3 – 1-meter GPIB buss cable 1583-6 – 2-meter GPIB buss cable 1583-9 – 3-meter GPIB buss cable
3505 – GPIB Interface
3550-900-01 – User’s Manual­Hard Copy
3511 - SMD Test Fixture
Available for performing three terminal measurements on surface mount devices. Connects directly to the front panel of the 3550. Use the 3511 for medium and high impedance measurements.
1-4
NOTE: Under certain measurement conditions, Kelvin Klips can cause a loss of measurement accuracy. Fixtures 3510, 3511, or 2005B chip tweezers are recommended for the following component values:
C<100pF; L < 100µH; R > 1Mƻ
Page 13
Model 3550 LCR Meter Instruction Manual Instrument Description
Performance Specifications
The advertised specifications of the model 3550 are valid under the following conditions:
1. The instrument must be verified and/or adjusted using the methods and intervals as described in the calibration section of this user’s manual.
2. The instrument must be in an environment, which does not exceed the limitations as defined under “Environmental” in the Miscellaneous Specifications section.
3. The unit is allowed to warm up for a period of at least 30 minutes before measurements are taken. A warm-up period of 60 minutes is recommended after exposure to or storage in a high humidity, (non-condensing), environment.
4. Only TEGAM-manufactured Kelvin Klips™, Tweezers and other test fixtures are used with this device during measurements.
Measurement Parameters:
Inductance (L), Capacitance(C), Resistance(R), Impedance (|Z|), Admittance (|Y|), Dissipation Factor (D), Quality (Q), Equivalent Series Resistance (R Resistance (R
), Conductance (G), Reactance (X), Susceptance (B), and Phase Angle (LJ).
P
), Equivalent Parallel
S
Measurement Ranges
Table 1.1 – Model 3550 Measurement Range Limits
Parameter Low Limit High Limit
L 320.0nH (At: 10ƻ 5MHz) 0.7500MH(At: 199.99Mƻ 42Hz) (LJ=90°) C 0.160pF (At: 100kƻ 5MHz) 0.037F(At: 100mƻ 42Hz) (LJ=-90°) R 0.01mƻ 199.99Mƻ |Z| 0.01mƻ 199.99Mƻ |Y| 5.000nS(199.99Mƻ) 100.00S(100mƻ) D 0.0001 9.999 Q 0.1 1999.9 R
S
R
P
G 5.000nS(199.99Mƻ) 100.00S(100mƻ) X 0.01mƻ 199.99Mƻ B 5.000nS(199.99Mƻ) 100.00S(100mƻ) LJ -180.00° 180.00°
0.01mƻ 199.99Mƻ
0.01mƻ 199.99Mƻ
NOTE: The Measurement Range is dependent on the Measurement Frequency. The figures in parentheses are DUT impedance values.
1-5
Page 14
Model 3550 LCR Meter Instruction Manual Instrument Description
Performance Specifications cont’d:
Basic Accuracy
0.1%
Display Specifications:
x Numeric Display (DISPLAY A)
1
L, C, R, |Z|, |Y| (4
x Numeric Display (DISPLAY B) D, Q, R
, RP, G, X, B, LJ (41/2 digits)
S
x Numeric Display (DISPLAY C)
Frequency, Voltage, Current, Terminal Voltage, Terminal Current, PANEL (3 digits)
x Unit Prefix Display (DISPLAY A) p, n, µ, m, k, M
/2 digits)
x Unit Prefix Display (DISPLAY B) n, µ, m, k, M
x Unit Display (DISPLAY A)
F, H, ƻ, S
x Unit Display (DISPLAY B)
ƻ, S, deg
x Unit Display (DISPLAY C)
Hz, kHz, MHz, V, mA
Switch Indicator Lamps
x RANGE – AUTO; UP; DOWN x CIRCUIT MODE – AUTO; SER; PRL x SHIFT x DISPLAY C – V; I; CV/CC x TRIGGER - INT; MAN/EXT x SAMPLE
1-6
Page 15
Model 3550 LCR Meter Instruction Manual Instrument Description
Performance Specifications cont’d:
Measurement Frequency
Table 1.2 – Model 3550 Measurement Frequencies
Frequency Range Resolution
42.0Hz ~99.9Hz 0.1Hz Steps 100Hz~999Hz 1Hz Steps
1.00kHz~9.99kHz 10Hz Steps
10.0kHz~99.9kHz 100Hz Steps 100kHz~999kHz 1kHz Steps
1.00MHz~5.00MHz 10kHz Steps
Measurement Frequency Accuracy
±0.01%
Measurement Signal Level
Table 1.3 – Model 3550 Measurement Amplitudes
Frequency Range Test Voltage
Range
42.0Hz~1.00MHz 0.01~5.00Vrms 0.01~99.99mArms
1.01MHz~5.00MHz 0.05~1.00Vrms 0.05~20.00mArms
Test Current Range
Open Terminal Voltage Mode
x Increment/Resolution 0.01V Steps x Max. Shorting Current 99.99mA (Dependent on Measurement Frequency)
Fixed Voltage Mode
x Increment/Resolution 0.01V Steps x Max. Shorting Current 99.99mA(Dependent on Measurement Frequency)
Fixed Current Mode
x Increment/Resolution 0.01mA Steps x Max. Shorting Current 99.99mA(Dependent on Measurement Frequency)
1-7
Page 16
Model 3550 LCR Meter Instruction Manual Instrument Description
Performance Specifications cont’d:
Monitor Functions
Voltage Monitor 0.00V~5.00V Current Monitor 0.00mA~99.99mA
Settings and Monitor Accuracy
Table 1.4 – Model 3550 Amplitude Accuracies
Frequency Range Test Voltage
Accuracy
42.0Hz~4.00MHz ±(10%+10mV) ±(10%+10µA)
4.01MHz~5.00MHz ±(20%+10mV) ±(20%+10µA)
Signal Source Impedance
50ƻ ±10%
Test Current
Accuracy
Measurement Circuit Modes
x Auto/Manual Circuit Selection x Parallel Equivalent Circuit Mode x Series Equivalent Circuit Mode
Range
Measurement Range is determined by the DUT’s absolute impedance |Z|. There are 9 total ranges selectable in either AUTO or MANUAL modes. Measurement parameters (L, C, R, etc.) are calculated from actual test voltage and current values.
Table 1.5 – Model 3550 Measurement Ranges
Range 1 2 3 4 5 6 7 8 9 AUTO Upper
Limit
|Z|
100mƻ 10ƻ 100ƻ 1kƻ 10kƻ 100kƻ 1Mƻ 10Mƻ 100mƻ~10Mƻ
1-8
Page 17
Model 3550 LCR Meter Instruction Manual Instrument Description
Performance Specifications cont’d:
Measurement Time
Open Terminal Voltage Mode
When the Measurement Frequency is 10kHz, no averaging, the range is held constant, RS-232C is OFF and the Comparator is ON, the shortest measurement time is 18ms.
Typical Analog Measurement Times
Conditions: Averaging= 1, range held, RS-232C is OFF, and Comparator ON:
Table 1.6 – Typical Analog Measurement Times
Test Frequency Typical Analog Measurement Time
42Hz 624ms Max.
120Hz 221ms Max.
1kHz 31ms Max.
10kHz 18ms Max.
100kHz 30ms Max.
1MHz 25ms Max. 5MHz 21ms Max.
About the Typical Analog Measurement Time
The Typical Analog Measurement Time is dependent upon the Measurement Frequency. For ranges
42.0Hz~750Hz, and 10.1kHz~19.0kHz, Typical Analog Measurement Time will exceed 40 ms. For other frequencies, 40ms will not be exceeded. (Conditions: Averaging = 1, range held, RS-232C is OFF, and Comparator ON)
Fixed Voltage/Fixed Current Mode
The measurement time can be up to 4X the time of the Open Terminal Voltage Mode.
Average Mode
The user can define from 1~100 averages.
1-9
Page 18
Model 3550 LCR Meter Instruction Manual Instrument Description
Performance Specifications cont’d:
Measurement Terminals
x 2 – BNC Source Terminals x 2 – BNC Detect Terminals x 1 – Guard Banana/Binding Post
Parasitic Impedance Correction
Open Correction 1kƻ or higher Impedance Short Correction Less than 1kƻ Impedance
Programmable Delay Time
Measurements will be started after a trigger is detected from the front panel, Control I/O Connector or RS232C interface. The delay time setting is user-definable and ranges from 0~10 seconds in 1ms increments.
GO/NO-GO Limits Comparator/Binning
Absolute Value
x Absolute Settings must fall within the Measurement Ranges of DISPLAY A and DISPLAY B. x Comparator Settings must be applied to existing preset panels. x Up to 9 programmable bins (1-9) may be defined. Bin 0 is reserved for measured values
that fall outside of the limits.
x User may program the High or Low limits of DISPLAY A and DISPLAY B to be ignored.
Percent Value
x Percent Settings must fall within the Measurement Ranges of DISPLAY A and DISPLAY B. x Limit Values of 00000~999.99 % for both DISPLAY A and DISPLAY B x Comparator Settings must be applied to existing panels. x Up to 9 programmable bins (1-9) may be defined. Bin 0 is reserved for measured values
that fall outside of the limits.
x User may program the High or Low limits of DISPLAY A and DISPLAY B to be ignored.
1-10
Page 19
Model 3550 LCR Meter Instruction Manual Instrument Description
Performance Specifications cont’d:
Comparator/Binning: Absolute Limits Operation
Table 1.7a – Absolute Comparator Operation
Operation Function
Bins 1~9 "LOW LIMIT" Measurement Value "HIGH LIMIT" LO Measurement Value < "LOW LIMIT" HI "HIGH LIMIT" < Measurement Value TOTAL GO When both DISPLAY A and DISPLAY B are not LO or HI
Comparator/Binning: Percent Limits Operation
*The location of the decimal point affects the % comparator/binning values.
Table 1.7b – Percentage Comparator Operation
Operation Function
Bins 1~9 "LOW LIMIT" Measurement Value "HIGH LIMIT" LO Measurement Value < "LOW LIMIT" HI "HIGH LIMIT" < Measurement Value TOTAL GO When both DISPLAY A and DISPLAY B are not LO or HI
Displays
Table 1.8 – Numeric Displays and Annunciators
Display Display Description & Range
Numeric DISPLAY A -19999 ~ 19999 (4-1/2 digits) Numeric DISPLAY B -19999 ~19999 (4-1/2 digits) Unit Prefix Annunciators for DISPLAY A p, n, µ, m, k, M Unit Prefix Annunciators for DISPLAY B n, µ, m, k, M Unit Annunciators for DISPLAY A F, H, ƻ, S Unit Annunciators for DISPLAY B ƻ, S, deg Comparator Limit LOW & HIGH Pushbuttons Comparator Enabled ON Pushbutton Comparator State LO-HI (For DISPLAY A & B) Comparator Total GO TOTAL GO LED
BIN 0-9 Seven Segment LED Display PANEL 0-9 Seven Segment LED Display GPIB Status SRQ, LTN, TLK, RMT
1-11
Page 20
Model 3550 LCR Meter Instruction Manual Instrument Description
Performance Specifications cont’d:
Digital Interfaces
RS-232C Standard interface GPIB Optional Interface Control I/O Standard Interface Connector
External Control Signals
CONTROL PANEL/EXT, EXT TRIGGER, PANEL No. (Isolated)
External Output Signals
MEASURE END, ERROR, DISPLAY A HI, DISPLAY A LO, DISPLAY A GO, DISPLAY B LO, DISPLAY B HI, DISPLAY B GO, TOTAL GO, TOTAL NO-GO, BIN (Optically Isolated)
Audible Buzzer
Can be set to turn ON with GO or NO-GO comparator state; or it can be disabled.
Panel Presets
9 panel presets including comparator settings can be programmed into non-volatile memory.
Front Panel Key Lock
Protects accidental changing of front panel settings all panel keys except the Manual Trigger are disabled.
Environmental
Operating Temperature = 5°C~40°C (41°F~104°F) <80%RH Non-Condensing
o Accuracy specifications are based on temperature conditions within 23°C±5°C
(73.4°F±9°F).
o For temperatures exceeding 23°C±5°C and still within the operating temperature,
the margin of error will be 2 times larger.
Power Source
x Operating Voltage=100/120/220/240V AC ±10%(MAX250V) x Power Frequency=50/60Hz x Maximum Power Usage=40VA
Dimensions
Width: 9.84” (250mm) Height: 5.83” (148mm) Depth: 15.74” (400mm)
Weight
Approximately 15.4lb (7kg)
1-12
Page 21
Model 3550 LCR Meter Instruction Manual Instrument Description
Formulas and Measurement Accuracy
Parameters and Formulas
Generally speaking, the characteristics of circuit components can be evaluated in terms of impedance (Z). The 3550 LCR Meter measures the voltage and current components of the circuit in relation to the alternating signal of the measurement frequency, and uses these values to calculate impedance (Z) and phase angle (LJ). Displaying impedance (Z) in a complex notation as below allows us to determine the following values.
I
Z
Z
jX
Z=R+jX LJ = TAN |Z| =
-1
X/R
R2 +X
V
Z =V / I
2
IMAGINARY NUMBERS
ș
R
REAL NUMBERS
Z: Impedance (ƻ) LJ: Phase Angle (deg)
R: Resistance (ƻ) X: Reactance (ƻ) |Z|: Absolute Value of Impedance (ƻ)
Depending on the properties of the DUT, Admittance (Y), which is the inverse of Impedance (Z), may be used to calculate parameter values instead. Admittance (Y), like Impedance (Z) can be represented in complex notation in order to calculate the following values.
Y=G+jB ĭ= TAN |Y|=
-1
(B*G)
G2 +B
2
Y: Admittance (S)
REAL NUMBERS
jB
G
ij
G: Conductance (S) B: Susceptance (S) |Y|: Absolute value of Admittance (S)
IMAGINARY NUMBERS
Y
1-13
Page 22
Model 3550 LCR Meter Instruction Manual Instrument Description
Formulas and Measurement Accuracy cont’d:
The 3550 calculates the various measurement values using the inter-terminal voltage (V) applied to the DUT (device under test) terminals, the current (I) that occurs in relation to the voltage, the Phase Angle (LJ) of (V) and (I), and the Angular Velocity (ǔ) of the measurement frequency. These factors are used in the formulas below to determine values for the measurements made by the TEGAM 3550.
Note: The Phase Angle is based on Impedance (Z). In order to base the Phase Angle on Admittance (Y), add a "-" sign to the Impedance value to negate it. Thus, ij for Admittance will be ij=-LJ.
Table 1.9 – Series and Parallel Equivalent Circuit Measurements
Item Series Equivalent Circuits Parallel Equivalent Circuits
Z
|Z| = V/I( )
R2 +X
2
|Y| = 1/|Z|( )
R
X
Y
RS = ESR = | |Z|cosLJ | RP = EPR = | 1/(|Y|cosLJ) | = 1/G
X = |Z| sinLJ
G B
L
C D Q
*** ij is the Phase Angle of Admittance: ij = -LJ
, CS, RSrepresent the L, C, R measurements for series equivalent circuits.
L
S
, CP, RPrepresent the L, C, R measurements for parallel equivalent circuits.
L
P
LS = X/ǔ LP = 1/(ǔ*B)
CS = 1/(ǔ*X) CP = B/ǔ
D = 1/tanLJ = 1/Q
Q = tanLJ = 1/D
G2 +B
2
|Y|cosLJ ***
B = |Y|sinLJ ***
1-14
Page 23
Model 3550 LCR Meter Instruction Manual Instrument Description
Measurement Accuracy
The impedance of the DUT is either the actual measured value or a value, which is derived from the formulas below:
If LJ=90° ĺ |Z| = ǔL
If LJ= -90° ĺ |Z| = 1/ǔC
If LJ=0° ĺ |Z| = R
Voltage Level Accuracy Limit
See the Measurement Accuracy Tables located in the Appendix. The tables contain formulas for calculating the Voltage Level Accuracies.
1-15
Page 24
Model 3550 LCR Meter Instruction Manual Instrument Description
1-16
Page 25
Preparation for Use
INSTRUMENT DESCRIPTION
PREPARATION FOR USE
QUICK START INSTRUCTIONS
OPERATING INSTRUCTIONS
PROGRAMMING & INTERFACING
SERVICE INFORMATION
APPENDIX
Model 3550 LCR Meter Instruction Manual 2-1
Page 26
Model 3550 LCR Meter Instruction Manual Preparation for Use
Unpacking & Inspection:
Each 3550 is put through a series of electrical and mechanical inspections before shipment to the customer. Upon receipt of your instrument unpack all of the items from the shipping carton and inspect for any damage that may have occurred during transit. Report any damaged items to the shipping agent. Retain and use the original packing material for reshipment if necessary.
Upon Receipt, inspect the carton for the following items:
Model 3550 General Purpose, Programmable LCR Meter Model 3550 User’s Manual CD Power Cord
!
Safety Information & Precautions:
The following safety information applies to both operation and service personnel. Safety precautions and warnings may be found throughout this instruction manual and the equipment. These warnings may be in the form of a symbol or a written statement. Below is a summary of these precautions.
Terms in This Manual:
CAUTION statements identify conditions or practices that could result in damage to the equipment or other property.
WARNING life.
Terms as Marked on Equipment:
CAUTION indicates a personal injury hazard not immediately accessible as one reads the marking, or a hazard to property including the equipment itself.
DANGER
statements apply conditions or practices that could result in personal injury or loss of
indicates a personal injury hazard immediately accessible as one reads the marking.
2-2
Page 27
Model 3550 LCR Meter Instruction Manual Preparation for Use
!
Safety Information & Precautions Cont’d:
Symbols:
As Marked in This Manual:
!
This symbol denotes where precautionary information may be found.
As Marked on Equipment:
!
Caution – Risk of Danger
Danger – Risk of Electric Shock
Earth Ground Terminal-
On
l
O
Grounding the Equipment
This product is grounded through the grounding conductor of the power cord. WARNING: properly wired receptacle before using this instrument. The proper grounding of this instrument is essential for safety and optimizing instrument operation.
Off
Chassis Terminal
Alternating Current
Earth Ground Terminal / Guard
To avoid electrical shock or other potential safety hazards, plug the power cord into a
Danger Arising from Loss of Ground
WARNING: in the instrument. Under these conditions all accessible parts, including insulating parts such as keypads and buttons could develop a hazardous voltage and put the user at risk.
If the connection to ground is lost or compromised, a floating potential could develop
2-3
Page 28
Model 3550 LCR Meter Instruction Manual Preparation for Use
!
Use the Proper Fuse
To avoid fire hazard, use only the correct fuse type as specified for the AC power supply in the “Preparation for Use”” or “Service” sections of this manual. Please note that the fuse rating for 100 & 120-volt operation is different than the rating for 200 & 240-volt operation. Information about the proper fuse type is also printed on the rear panel of the instrument.
Refer fuse replacement to qualified service personnel.
Do Not Use in Explosive Environments
WARNING: The 3550 is not designed for operation in explosive environments.
Do not Operate Without Covers
WARNING: missing panels or covers could result in personal injury.
This device should be operated with all panels and covers in place. Operation with
FOR QUALIFIED SERVICE PERSONNEL ONLY
Servicing Safety Summary:
!
Do Not Service Alone
Do not perform service or adjustment on this product unless another person capable of rendering first aid is present.
Use Care When Servicing with Power On or Off
Dangerous voltages may exist at several points in this product. To avoid personal injury or damage to this equipment, avoid touching exposed connections or components while the power is on. Assure that the power is off by unplugging the instrument when removing panels, soldering, or replacing components. WARNING: present throughout the instrument even when the instrument is in the OFF state. Always unplug the instrument and wait 5 minutes before accessing internal components.
Power Source
This product is intended to connect to a power source that will not apply more than 250V RMS between the supply conductors or between either supply conductor and ground. A protective ground connection by way of the grounding conductor in the power cord is essential for safe operation.
The instrument power source is electronically controlled meaning that there is power
2-4
Page 29
Model 3550 LCR Meter Instruction Manual Preparation for Use
Line Voltage & Fuse Selection:
!
CAUTION: DO NOT APPLY POWER TO THE INSTRUMENT BEFORE READING THIS SECTION:
Unless other wise specified, the Model 3550 is delivered from TEGAM with its power supply set for 120V, 60 Hz operation. However, the 3550 design allows it to operate under 100, 120V, 220 or 240V @ 50/60 operation. It is strongly recommended that the line voltage and fuse size be verified before powering the unit.
First, determine the supply voltage that the instrument will be operating under and verify that the supply voltage does not fall outside of the allowable ranges in the table below:
LINE VOLTAGE RANGE FUSE SIZE 100/120V 90~132VAC 1A 220/240V 198~250VAC 0.5A
Make sure that the proper fuse size is installed.
Next, verify that the jumper block in the rear panel of the instrument is securely plugged into the correct position. There are four positions that the block may be set to. Set the jumper block to point to one of the four voltage settings. You can remove the jumper block by pulling it from the jumper socket.
120V240V
220V 100V
Unit set for 120VAC Operation 50/60Hz
The instrument is ready for power up. Proceed to “Quick Start Instructions” for continued operation.
2-5
Page 30
2-6
Page 31
Quick Start Instructions
SPECIFICATIONS
PREPARATION FOR USE
QUICK START INSTRUCTIONS
OPERATING INSTRUCTIONS
PROGRAMMING & INTERFACING
SERVICE INFORMATION
APPENDIX
Model 3550 LCR Meter Instruction Manual 3-1
Page 32
Model 3550 LCR Meter Instruction Manual Quick Start Instructions
The Model 3550 is a versatile product, which can be used in many different configurations. Because of its ability to measure a large number of impedance parameters under a dynamic range of test frequencies, amplitudes, and configurations, its is highly recommended that the entire “Operating Instructions” Section of this manual be reviewed to insure proper use of this equipment.
This Quick Start section is designed to give the user a general instruction set for the speedy setup and measurement of impedance values where arbitrary measured values are needed and accuracy is not critical. Whenever additional information is applicable, a reference will be made to other parts of this manual so that the user, at their discretion, can decide whether or not to pursue additional information.
Power the Unit
The power supply of the Model 3550 is designed for 50-60 Hz operations and a voltage range of 90-250 VAC. It is assumed that the “Preparation for Use” section of this manual has been read an understood and the line voltage and fuse settings have been verified to be correct.
Power the unit by depressing the pushbutton located on the lower right corner of the front panel. Allow at least 30 minutes for the unit to warm up and verify that the unit will be operated in its specified operating environment described on page 1-12.
Instrument Settings
Factory Default Settings
Before performing an actual impedance measurement, there are a number of test parameters, which must be defined. The factory default settings can be used for most general-purpose measurements.
The 3550 is shipped from the factory with default settings as follows:
Table 3.1 – Factory Default Settings
Parameter Setting Parameter Setting
1 DISPLAY A Capacitance 9 Communications *RS232 2 DISPLAY B Dissipation Factor 10 Trigger INT 3 DISPLAY C Frequency 11 PANEL 0 4 Comparator Data Cleared 12 Circuit Mode AUTO 5 Comparator OFF 13 Range AUTO 6 Zero Correction Data Cleared 14 LOCK OFF 7 Test Frequency 1kHz 15 Test Voltage 1.00 V 8 Measurement Mode Open Terminal
Voltage
* not storable in a setup location.
At any time during use, you may recall these default settings by pressing the [MAN/EXT] key while powering the unit.
NOTE: When the unit is reset to factory defaults, user-defined comparator settings and zero correction factors will be lost.
3-2
Page 33
Model 3550 LCR Meter Instruction Manual Quick Start Instructions
Table 3.2a – Control Command Summary
1 Change measure item of DISPLAY A A
Commands
Display
2 Change measure item of DISPLAY B B
3 Show measure frequency on DISPLAY C FREQ
4 Show terminal voltage on DISPLAY C V
5 Show measure. current on DISPLAY C I
6 Change panel number 0-9
Basic Commands
7 Change measurement frequency FREQ ĺ DISPLAY C's UP and DOWN
8 Change measurement voltage V ĺ DISPLAY C's UP and DOWN
9 Carry out open correction OPEN
10 Carry out short correction SHORT
11 Change to auto range RANGE's AUTO
12 Change to auto circuit mode CIRCUIT MODE's AUTO
13 Change to fixed voltage mode V ĺ CV/CC
14 Change the fixed voltage setting

15 Change to fixed current mode I ĺ CV/CC
16 Change the fixed current setting
Applied Commands
17 Raise the range RANGE's UP
18 Lower the range RANGE's DOWN
19 Parallel equivalent circuits mode PRL
20 Serial equivalent circuits mode SER
21 Measure with continuous trigger INT
22 Measure with manual (single shot) trigger MAN/EXT ĺ MAN/EXT ĺ MAN/EXT ĺ
23 Set comparator lower limit LOW
24 Set comparator upper limit HIGH
25 Turn comparator on ON
26 Move cursor forward in setting mode
27 Move cursor backward in setting mode
28 Set measurement frequency SHIFT ĺ 0 ĺ Frequency Value ĺ ENTER
Commands
Shift
29
30 Set current for fixed current mode SHIFT ĺ 2 ĺ Current Value (mA) ĺ ENTER
31 Set number of measurements SHIFT ĺ 3 ĺ Number Setting ĺ ENTER
Set voltage for measurement or fixed voltage mode
When V and CV/CC LED are flashing, DISPLAY C's UP and DOWN
When I and CV/CC LED are flashing DISPLAY C's UP and DOWN
SHIFT ĺ 1 ĺ Voltage Value (V) ĺ ENTER
3-3
Page 34
Model 3550 LCR Meter Instruction Manual Quick Start Instructions
Table 3.2b – Control Command Summary cont’d
32 Release Key Lock SHIFT ĺ 4 ĺ 0 ĺ ENTER
33 Lock Key SHIFT ĺ 4 ĺ 1 ĺ ENTER
34 No buzzer for decision SHIFT ĺ 5 ĺ 0 ĺ ENTER
35 Buzzer for GO decision ON SHIFT ĺ 5 ĺ 1 ĺ ENTER
36 Buzzer for NO GO decision ON SHIFT ĺ 5 ĺ 2 ĺ ENTER
37 Set start delay time SHIFT ĺ 6 ĺ delay time (ms) ĺ ENTER
38 To use absolute value comparator SHIFT ĺ 7 ĺ 0 ĺ ENTER
Shift Commands
39 To use percent comparator SHIFT ĺ 7 ĺ 1 ĺ ENTER
40 Set 1st spot correction frequency SHIFT ĺ 8 ĺ 1 ĺ frequency value ĺ ENTER
41 Set 2nd spot correction frequency SHIFT ĺ 8 ĺ 2 ĺ frequency value ĺ ENTER
42 Set 3rd spot correction frequency SHIFT ĺ 8 ĺ 3 ĺ frequency value ĺ ENTER
43 Set low frequency of correction SHIFT ĺ - ĺ 1 ĺ frequency value ĺ ENTER
44 Set high frequency of correction SHIFT ĺ - ĺ 2 ĺ frequency value ĺ ENTER
45 Do spot or open correction SHIFT ĺ OPEN
46 Do spot or open correction SHIFT ĺ SHORT
47 Change RS-232C baud rate SHIFT ĺ 9 ĺ 1 ĺ Setting ĺ ENTER
48 Change RS-232C data length setting SHIFT ĺ 9 ĺ 2 ĺ Setting ĺ ENTER
49 Change RS-232C parity setting SHIFT ĺ 9 ĺ 3 ĺ Setting ĺ ENTER
50 Change RS-232C stop bit setting SHIFT ĺ 9 ĺ 4 ĺ Setting ĺ ENTER
3-4
Table 3.3 - Shift Function Summary
SHIFT 0 Set measurement frequency
SHIFT 1
SHIFT 2 Set current for fixed current mode
SHIFT 3 Set number of measurements
SHIFT 4 Set key lock
SHIFT 5 Set buzzer
SHIFT 6 Set start delay time
SHIFT 7 Set comparator mode
SHIFT 8 Set spot correction frequency
SHIFT 9 Set RS-232C communication parameters
SHIFT - Set correction frequency limits
SHIFT OPEN Do spot open correction
SHIFT SHORT Do spot short correction
Set voltage for measure voltage/fixed voltage mode
Page 35
Model 3550 LCR Meter Instruction Manual Quick Start Instructions
Instrument Setup for Basic Measurements
To perform basic measurements with the Model 3550, perform the steps below: Refer to “Operating Instructions” for more details about specific commands or functions of the 3550.
NOTE: The numerals in parentheses reference the Control Command Tables on the previous pages. The letters in parentheses reference the Settings Summary below:
Perform an open and closed circuit zero correction {Normal correction (9,10), Frequency
limit correction (F), or Spot correction (E)} on the test leads or fixture.
Set DISPLAY A measurement parameters (1) Set DISPLAY B measurement parameters (2) Set measurement frequency (7 or 28) Set measurement mode {Voltage (8 or 29), Constant Voltage (A), or Constant Current (B)} Set the comparator limits and operation if necessary {Absolute Comparator (C) or Percent
Comparator (D)}
Set the trigger mode {Continuous (21) or One-Shot (22) trigger}
Table 3.4 - Settings Summary
To use fixed voltage
A
mode To use fixed current
B
mode
To use absolute
C
value comparator
To use percent
D
comparator
To use spot
E
correction
To use frequency
F
limit correction
Set voltage for fixed voltage mode (14 or 29) Change to fixed voltage mode (13) Set current for fixed current mode (16 or 30) Change to fixed current mode (15) Select absolute value comparator (38) Set comparator lower limit value (23) Set comparator upper limit value (24) Turn comparator ON (25) For comparator setting details, see absolute value
comparator setting instructions.
Select percent comparator (39) Set comparator lower limit value (23) Set comparator upper limit value (24) Turn comparator ON (25) For comparator setting details, see percent
comparator setting instructions.
Set spot correction frequency (40~42) Carry out spot-open correction (45) Carry out spot-short correction (46) Set upper frequency limit for correction (44) Set lower frequency limit for correction (43) Carry out open correction (9) Carry out short correction (10)
3-5
Page 36
Model 3550 LCR Meter Instruction Manual Quick Start Instructions
Setting the Absolute Comparator
To set the Absolute Value Comparator, following the procedure below:
1. Select either the upper or lower comparator limit. To input an upper limit, press [HIGH] To input a lower limit, press [LOW]
2. Input the BIN No.
After the [HIGH] or [LOW] key is pressed, the BIN LED will begin to flash. Use keys [1]~[9] to designate the number.
3. Input a limit value for DISPLAY A After selecting the bin number, use the cursor to highlight DISPLAY A.
Enter the 5-digit upper or lower value using the keys [1]~[9]. When this is complete, the decimal point will flash.
Press [DP] to move the decimal point to the position of the setting value.
Use the cursor [Ź] to highlight the unit. Here, each time [UNIT] is pressed, the unit will change. Designate the unit for the Nominal value.
4. Input a limit value for DISPLAY B When the setting for DISPLAY A is complete, use the cursor [Ź] to highlight DISPLAY B. Enter the 5 digits, decimal point and unit as done for DISPLAY A.
If the setting is a negative value, use the cursors keys, [Ź] or [Ż] to highlight the 1 of DISPLAY B.
Press [-] to negate the value.
5. Finalize the Comparator Settings If defining an upper limit, press [HIGH] to complete the setting. If defining a lower limit, press [LOW] to complete the setting.
To input settings for other bins, repeat the steps above.
6. Invalidating the DISPLAY A or DISPLAY B Comparator settings.
To invalidate the settings, take the cursor, [Ź] or [Ż], to any digit location in DISPLAY A or DISPLAY B. Then press [IGNORE] to invalidate the settings of DISPLAY A or DISPLAY B.
NOTES:
The comparator can only be set using panels [1]~[9]. [0] is not used. The above procedure applies only to the Absolute Value Comparator. The procedure for
setting the Percent Comparator is different. Before attempting to set the Absolute Comparator limits, verify that the 3550 is in the Absolute Comparator Mode.
After setting the limit values for the comparator, the measurement parameters for
DISPLAY A and DISPLAY B cannot be changed. The measurement parameters must be defined before the comparator is set.
For DISPLAY B, there are no units for D,Q, or LJ.
st
digit
3-6
Page 37
Model 3550 LCR Meter Instruction Manual Quick Start Instructions
Setting the Percent Comparator
To set the Percent Comparator, follow the procedure below.
The upper and lower limit for the nominal value are the same. For the threshold values, different values for the upper and lower limit must be set. In order to proceed, the 3550 must be in the percent comparator mode.
1. Select the Upper or Lower Limit to Enter the % Comparator Mode
To input the Nominal value, press the [HIGH] key. (The [LOW] key may also be pressed to enter the % Comparator Set Mode)
2. Select the Nominal Value Input Mode To input the Nominal Value, press [0] when the first digit of DISPLAY C is flashing.
3. Input the BIN No.
After selecting the Nominal Value Input Mode, use the cursor [Ź] to highlight the bin number, and then press [1]~[9] to select the desired bin.
4. Set the Nominal Value for DISPLAY A
After inputting the bin number, use the cursor, [Ź] to highlight DISPLAY A. Enter a 5 digit Nominal Value for DISPLAY A using keys [0]~[9]. (NOTE: The 1
st
digit, is restricted to only a “1” or “0” setting.) When this is complete, the decimal point will flash. Press [D.P] to move the decimal point to the desired position. Then, use the cursor [Ź] to highlight the unit. Press the [UNIT] key until the required unit annunciator is lit. This designates the measurement unit for the Nominal value.
5. Input the Nominal Value for DISPLAY B When the setting for DISPLAY A is complete, use the cursor [Ź] to highlight DISPLAY B. Enter the 5 digits, decimal point and unit just as in DISPLAY A.
If the setting is to be a negative value, use cursors [Ź] or [Ż] to highlight the 1
st
DISPLAY B then press [-] to negate the value.
6. Completing the Nominal Value Setting Press the [HIGH] key to finalize the Nominal Value Setting for the % Comparator.
(If the [LOW] key was pressed to enter into the % Comparator Mode then press it instead of the [HIGH] key to exit the mode and finalize the settings.)
7. Set an Upper or Lower Limit for the % Threshold Values
To input an upper limit threshold value, press the [HIGH] key. Press the [LOW] key to set the lower limit threshold value. Then, highlight the first digit of DISPLAY C and press [1].
digit of
3-7
Page 38
Model 3550 LCR Meter Instruction Manual Quick Start Instructions
Setting the Percent Comparator
8. Input the BIN No.
After selecting the threshold value input mode, use the cursor key, [Ź] to highlight the bin LED, and then press [1]~[9] to select the desired bin.
9. Input the Threshold Value for DISPLAY A After entering the bin number, use the cursor key [Ź] to highlight DISPLAY A.
Enter the 5-digit threshold value for DISPLAY A using keys [0]~[9]. (The setting range is from 000.00%~199.99%)
10. Input the Threshold Value for DISPLAY B
Enter the 5-digit threshold value for DISPLAY B using keys [0]~[9]. (The setting range is
000.00%~199.99%)
11. Finalizing the Setting To finalize an upper limit, press [HIGH] or press [LOW] to finalize a low limit setting. NOTE: Whether a high or low limit %comparator setting is defined is determined in step 7.
To input settings for other bins, return to Step 1 and repeat the process.
12. Invalidating the DISPLAY A or DISPLAY B Percent Comparator settings
To invalidate the settings, for threshold values, simply enter the Threshold Percent Comparator Set Mode and move the cursor to any segment in DISPLAY A or DISPLAY B. Then press [IGNORE] to invalidate the % comparator settings for that DISPLAY.
NOTES:
The comparator can only be set using panels 1~9. 0 cannot be used. This procedure is only for the percent comparator setting. Note that the procedure for the
absolute value comparator is different.
Make sure the comparator mode is set for “Percent Mode” otherwise the procedure will not
work correctly.
After setting the limit values for the comparator, the measurement parameters for DISPLAY
A and DISPLAY B cannot be changed. This means that the measurement parameters must be defined before the comparator settings.
For DISPLAY B, there are no units for D,Q, or LJ.
3-8
Page 39
Operating Instructions
INSTRUMENT DESCRIPTION
PREPARATION FOR USE
QUICK START INSTRUCTIONS
OPERATING INSTRUCTIONS
PROGRAMMING & INTERFACING
SERVICE INFORMATION
APPENDIX
Model 3550 LCR Meter Instruction Manual 4-1
Page 40
Model 3550 LCR Meter Instruction Manual Operating Instructions
Basic Operation
The Model 3550 is a fully programmable LCR meter, designed for use in many different low and high frequency applications. There are ideal configurations of the 3550 for each type of measurement. These configurations optimize test conditions while enhancing accuracy and measurement speed. In order to maximize the effectiveness the 3550, the user should have a thorough understanding of the instrument’s operation along with a basic knowledge of LCR measurement techniques. This section will provide the user with the necessary information to make accurate and repeatable measurements.
Default Parameters
Each unit is delivered from the factory with predefined settings intended for general-purpose LCR measurement and ease of use. Chapter III, Quick Start Instructions, contains information on factory default settings.
!
CAUTION:
NEVER APPLY DIRECT CURRENT TO THE MEASUREMENT TERMINALS
Never apply direct voltage to the measuring terminals. When testing capacitors, always make sure that they are fully discharged otherwise damage to the instrument will result.
NEVER MEASURE COMPONENTS IN A POWERED CIRCUIT
If you are measuring components, which are integrated into circuit boards, always check that power is disconnected from the circuit before taking measurements. When measuring transformers, always verify that no induced voltage is present before making connections.
Measurement Tips
Making Measurements that are Sensitive to Voltage and Current
The measured values of many components such as core-filled coils, layered ceramic capacitors etc. are dependent upon the applied level of test current or voltage. In applications where the test signal must remain constant, it is recommended that the range of the 3550 be manually set.
In the event that AUTO Range must be used, then operation of the 3550 in the “Constant Current” or “Constant Voltage” Mode will maintain signal levels while allowing range adjustments.
Measuring Earth-Grounded Test Pieces
The L-side (L
FORCE
& L point of the internal amplifiers. The L-side is a high-impedance point with zero electrical potential. These terminals cannot be directly connected to Earth Ground. Measurement of devices with one lead tied to earth ground is not recommended.
) of the measurement terminals is set up to be the virtual grounding
SENSE
4-2
Page 41
Model 3550 LCR Meter Instruction Manual Operating Instructions
R
R
R
R
Measurement Tips cont’d:
Measurement Circuit Modes
When the Measurement Circuit Mode is set to "AUTO", the Parallel Mode (PRL) or the Serial Mode (SER) is automatically selected by the instrument based on the Measurement Range. The measurement range is determined by the instrument by the magnitude of |Z| with no regards to D or Q of the DUT.
Thus, the measured values of a component will vary depending on whether series or parallel equivalent circuit modes are used. The relationship between series and parallel equivalent circuit measurements is illustrated in Table 4.1.
Table 4.1: Relationship Between Series and Parallel Equivalent Circuit Values
Equivalent Circuit
Mode
(SER)
Ls
L
(PRL)
G
L
(SER)
C
S
C
(PRL)
G
C
As a general rule, when measuring components where the reactive component is high, >10kƻ then the parallel equivalent mode should be used as parallel leakage becomes more significant to the measurement. For low reactance components, <10ƻ the series equivalent mode should be used since lead resistance becomes more significant to the reading than parallel leakage. For measurements of component values that fall somewhere in between, follow the manufacturer’s recommendations.
Dissipation
Factor: D
R
D =
D =ZLPG
D =ZCSR
D =
=
Z
L
S
=
=
G
=
Z
C
P
1 Q
1 Q
1 Q
1 Q
Conversion Formula
P LD1L
S L
L
P C
C
S CD1C
2
1
2
D1
1
2
D1
2
2
G
D
S
R
D
P
G
D
S
D
R
P
1
2
D1
2
1
2
G
D1
2
1
2
D1
2
1
2
G
D1
4-3
Page 42
Model 3550 LCR Meter Instruction Manual Operating Instructions
Measurement Tips cont’d:
Negative Capacitance or Inductance Readings
If a capacitor is connected to the 3550 while in the inductance measurement mode or if an inductor is connected to the 3550 while in the capacitance mode, a negative sign will appear in the display. The negative sign in the capacitance mode indicates an inductive measurement. The negative sign in the inductance mode indicates a capacitive measurement.
If the resonant frequency of a DUT is reached, a “-“ sign will appear in the display.
Zero Correction of Parasitic Impedance
"Open Correction" is for the purpose of compensating for the stray capacitance and stray conductance of cables or test fixtures. Executing an "Open Correction" procedure on the test leads of fixture before measurement of high impedance DUTs improves measurement accuracy.
"Short Correction" compensates for the residual inductance and resistance of cables or test fixtures. Performing a "Short Correction" procedure on cables or test fixture before measurement of low impedance DUTs improves measurement accuracy.
If the condition or position of a cable or test fixture changes, you MUST carry out Zero Correction to compensate for the change. When performing a Zero Correction, keep hands and metallic objects away from the cable or fixture.
Connection
Terminal
Only
5-Terminal
Kelvin
Klips™
Table 4.2: Short and Open Null Connections
Open Correction Short Correction
Correct Incorrect Correct Incorrect
G L H
G L H
G L H G L H G L H
G L H G L H G L H
L
H
F
F
L
H
S
S
L
H
F
L
H
S
Insulator
F
S
L
F
H
F
H
L
S
S
L
H
L
F
S
H
S
F
4-4
Page 43
Model 3550 LCR Meter Instruction Manual Operating Instructions
T
Measurement Tips cont’d:
Test Lead Requirements
Four-wire Kelvin-type cables or fixtures must be used with the 3550 in order to obtain accurate impedance measurements. The Kelvin measurement technique allows significant accuracy advantage over the two-wire method. This is because it virtually eliminates lead resistance and inductance and reduces stray capacitance between the source and sense leads. Two of the four conductors are designated as source leads. These source leads provide the precision test current that will be referenced in making the resistance measurement. Since current is the same throughout a series circuit, the lead resistance of the test leads will not have any effect on the level of reference current.
The other two conductors are designated as voltage sense leads. These leads originate from a high impedance, volt measurement circuit. When these leads are terminated at the points of contact, an exact resistance reading is calculated by the 3550’s microprocessor. The series lead resistance of the voltage sense leads is negligible due to the high impedance of the voltage measurement circuitry within the 3550.
Figure 4.1a: Four-Wire Kelvin Measurement
Z
S
VOLTAGE
REFERENCE
CURRENT SOURCE
H
FORCE/LFORCE
SENSE
H
SENSE/LSENSE
Z
S
V
CONTACT
POINTS
Z
DUT
Figure 4.1b: Proper Application of Four Wire Kelvin Measurement
H
S
H
L
L
F
S
F
L
F
L
S
HSH
F
DUT DUT
CORRECT INCORREC
4-5
Page 44
Model 3550 LCR Meter Instruction Manual Operating Instructions
Measurement Tips cont’d:
Measurement Signal Levels
Measurement signal levels are adjustable between 0.01V~5.00V. It is recommended that the voltage level be kept at a high value (usually 1V or larger) to avoid potential stability problems associated with test frequency, device impedance and noise.
Series or Parallel Circuit Modes
The "CIRCUIT MODE" should always be "AUTO" mode, unless there is a special circumstance that requires otherwise. The 3550 is in the AUTO circuit mode when the pushbutton’s LED is illuminated. Manually setting the circuit mode should only be done in special circumstances where series or parallel measurement is necessary even if some discrepancy is known to occur.
Auto Range Switching Threshold
When using Auto Range for measurement, the relationship between the switching direction and indicated switching values are as follows: (When Dissipation Factor: D=0)
Range UP When|Z|19999 counts Range DOWN When|Z| 17999 counts
As the dissipation factor, D increases then these threshold values will change.
Display Resolution
The absolute impedance, |Z| of the DUT determines the range of the 3550. All subsequent measurements will display with a resolution that is consistent with the instrument’s current range setting.
AC Resistance of Coils with Cores
In most cases the AC resistance of a coil with no core at low frequencies is the same as the DC resistance. This is not the case for coils with high permeability cores such as ferrite or iron. Higher permeability will promote core losses from an AC signal causing the resistance measurement to appear larger than it would be for a DC measured resistance value. Therefore, an inductance measurement taken at 42Hz (the lowest frequency setting of the 3550) will not necessarily reflect a DC equivalent resistance value.
4-6
Page 45
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description
Front Panel Layout
15
17
12
10
11
8
6
14
16
18 20
22
21
2
9
Except for the "POWER" switch, a buzzer will be sounded whenever the function of a switch is changed.
7
13 19
Figure 4.2: Front Panel layout
34 5
1
4-7
Page 46
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
1 - [POWER] Switch – [POWER]
With the power switch in the down state, power is applied to the 3550 and in the up state power is off. When the power is turned on, the remaining operational switches will retain the same status as they had before the power was previously turned off. If the power is turned on while simultaneously pushing the [MAN/EXT] Key, the machine will return to its factory default settings.
2 "CIRCUIT MODE" Panel - [AUTO] [SER] [PRL] This panel contains the keys used for setting the Series or Parallel Equivalent Circuit Mode. When verifying incoming component values it is usually good practice to consult the manufacturer for the recommended test mode. In general, parallel mode is used when the reactance of a device is large. Parallel mode should be used when a capacitor’s leakage or an inductor’s core losses significantly affects the reading. Series mode is recommended for small reactance measurements in capacitors and inductors where series lead resistance must be considered and leakage and core losses become negligible.
x [AUTO] – When this key is illuminated, the instrument is in AUTO equivalent circuit mode.
This means that the 3550 will automatically select either series or parallel equivalent circuit mode based on the impedance of the DUT. It does not necessarily mean that the correct equivalent circuit mode will be selected.
x [SER] – Series equivalent circuit mode is typically desirable for low reactance
measurements. This is because Rs or series resistance becomes more significant than parallel resistance to the measured value.
x [PRL] – Parallel equivalent circuit mode is often more effective for measurement of
impedances with high reactive components. This is because in a capacitor or inductor with a high reactance, the parallel leakage is more significant to the measured value.
3 "TRIGGER" Panel - [INT] [MAN/EXT] Select either an internal or external trigger by pressing either the [INT] or [MAN/EXT] key. The active trigger mode of the 3550 is indicated by an illuminated pushbutton.
x [INT] - When the [INT] key is pressed, the 3550 operates in the continuous trigger mode.
The instrument automatically takes readings at regular intervals.
x [MAN/EXT] – Depressing the [MAN/EXT] key will put the 3550 into manual one-shot trigger
mode, which requires the [MAN/EXT] key to be pressed for each reading.
4-8
Page 47
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
4 “ZERO” Panel – [OPEN] [SHORT]
Performing OPEN and SHORT corrections before making an impedance reading greatly improves measurement accuracy. Refer to page 4-4, “Zero Correction of Parasitic Impedance” for connection diagrams and other details regarding these procedures.
x [OPEN] - The OPEN circuit zero correction procedure compensates for stray capacitance
and conductance between test leads/fixture. Pressing the [OPEN] key initiates the open circuit zero correction procedure. Make sure that the test fixture is in the open state before performing the OPEN circuit zero correction procedure.
x [SHORT] – The SHORT circuit zero correction procedure compensates for residual
resistance and inductance in the test leads/fixture. Pressing the [SHORT] key initiates this procedure. Make sure that the test fixture is shorted before performing the SHORT circuit zero correction procedure.
There are three variations of zero correction procedures:
1) Standard Zero Correction: corrects all frequencies. This process requires several minutes to step through all of the test frequencies. See 4 (p. 4-9) for details.
2) Spot Zero Correction, allows up to three user-defined frequencies to be corrected: see [SHIFT]ĺ[8] (p. 4-14) to define spot frequencies; [SHIFT] ĺ[OPEN] (p. 4-16) for Open circuit spot correction; & [SHIFT] ĺ[SHORT] (p. 4-16) for short circuit spot correction.
3) Correction Frequency Limit Mode: allows the user to define a range of test frequencies to be zero corrected during a standard zero correction. Refer to [SHIFT] ĺ[ - ] (p. 4-16) and 4 (p. 4-9) for more details.
5 "RANGE" Selection Panel – [AUTO] [DOWN] [UP] The Range panel is used to select either automatic or manual measurement ranges. If AUTO range is used to find the correct measurement range and the 3550 is then switched to MANUAL range, the 3550 will retain the current measurement range.
x [AUTO] – Pressing the [AUTO] key will toggle between Auto and Manual Range Modes.
When the 3550 is in the AUTO mode, the pushbutton will be illuminated. When the instrument is in the manual mode the AUTO pushbutton is not illuminated.
x [DOWN] – In the manual mode, pressing the [DOWN] key will lower the range of the
instrument. The [DOWN] key will flash to indicate when the current range is under ranged and needs to be lowered for a reading to take place. When the [DOWN] key is pressed, then an audible beeper will sound to indicate that the range has been lowered.
x [UP] - In the manual mode, pressing the [UP] key will raise the current range of the 3550.
The [UP] key will flash to indicate when the current range is over ranged and needs to be raised in order for a reading to take place. When the [UP] key is pressed, then an audible beeper will sound to indicate that the range has been raised.
4-9
Page 48
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
6 [SHIFT] Key
The [SHIFT] key has dual functionality. Its primary purpose is to allow access to the secondary functions of the numeric keys. Its secondary function allows the user to switch from Remote communications Mode (GPIB or RS-232C Operation) to Local (Front Panel) Operation.
Below is a summary of the secondary functions accessible by pressing the [SHIFT] key and then a corresponding function key. The secondary function keys are [0]~[9], [OPEN], [SHORT] and [-]. For descriptive convenience, notation such as [SHIFT] ĺ [0] etc will be used to indicate that first the [SHIFT] Key should be pushed, and then the [0]" Key.
[SHIFT] ĺ [0]: Frequency Set Mode
The 3550 test frequency settings may be programmed from 42.0Hz to 5.00 MHz with a resolution of three significant digits. If DISPLAY C is set to indicate test frequency and the 3550 is in the Measurement mode, the frequency can be adjusted changed by pressing the [UP] and [DOWN] keys in the DISPLAY C panel.
Pressing [SHIFT] ĺ [0] will cause the 3550 to enter the frequency set mode.
Entering the Frequency Values The most significant digit in DISPLAY C will flash indicating that a new value must be
entered. Press a key from [0]~[9] to set the value of the first digit and to proceed to set the next digit value. Once again, press a key from [0]~[9] and proceed to the next digit.
Set the Decimal Point Location Once values for all of the three digits have been defined, the decimal point will start to
blink indicating that the decimal point needs to be positioned. Press the [DP] key and notice that the position of the decimal point will change. Continue to press the [DP] key until the decimal point is positioned correctly.
Select the Frequency Units
Press either the [CURSOR Ź] or [CURSOR Ż] keys repeatedly until one of the frequency units annunciators (Hz, kHz, or MHz) begins to blink. This indicates that the frequency units mode is now active. Press the [UNIT] key until the desired frequency units annunciator starts to blink.
Verify and Set the Frequency The user can go back and edit any of the previous settings by pressing the [CURSOR Ź] or [CURSOR Ż] keys to access the settings. Verify that all frequency settings are correct and then press the [ENTER] key. This will store the new frequency settings and return the instrument back to measurement mode.
4-10
Page 49
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
[SHIFT] ĺ [1]: Test Voltage Set Mode
*See the Appendix for detailed information concerning the measurement ranges for constant current or constant voltage modes and for instructions for setting either of these modes.
The 3550 test voltage values may be modified if the open circuit voltage or constant voltage setting is displayed on DISPLAY C. Constant voltage mode is indicated when the [CV/CC] and [V] pushbuttons are illuminated. The 3550 is in open circuit voltage mode when only the [V] key is lit. The Model 3550 has programmable test voltages from 0.01V~5.00V for test frequencies of
42.0Hz~1.00MHz. For test frequencies above 1.00MHz, the voltage can be adjusted between 0.01V and 1.00V. Test voltages are adjustable in steps of 0.01V and are displayed in DISPLAY C to three significant digits. Refer to the Appendix to calculate the maximum voltage range of the constant voltage mode. If DISPLAY C is set to indicate test voltage and the 3550 is in the Measurement mode, the voltage amplitude can be incremented or decremented by pressing the [UP] and [DOWN] keys in the DISPLAY C panel. If the [UP] and [DOWN] keys are held down the 3550 will increase the speed at which the voltage is incremented or decremented. If more dramatic changes are required, then the test voltage must be manually entered by accessing the 3550’s voltage set mode.
Pressing [SHIFT] ĺ [1] will cause the 3550 to enter the voltage set mode. Below are
instructions to set the test voltage.
Enter the Voltage Values
The most significant digit in DISPLAY C will flash indicating that a new value must be
entered. Press a key from [0]~[9] to set the value of the first digit and to proceed to set the next digit value. Once again, press a key from [0]~[9] and proceed to the next digit.
Verify and Set the Test Voltage Once all three digits have been entered, double-check the final voltage value. The values are changed by pressing the [CURSOR Ź] or [CURSOR Ż] keys and entering a new value. Press the [ENTER] key to save the voltage setting and to return back to the measurement mode. If a programmed voltage exceeds the instrument’s sourcing capability, the 3550 will beep three times and not allow the illegal value to be saved if the [ENTER] key is pressed.
[SHIFT] ĺ [2]: Test Current Set Mode
*See the Appendix for detailed information concerning the measurement ranges for constant current or constant voltage modes and for instructions for setting either of these modes.
The 3550 test current values may be modified if the open circuit current or constant current setting is displayed on DISPLAY C. Constant current mode is indicated when the [CV/CC] and [I] pushbuttons are illuminated. The 3550 is in open circuit current mode when only the [I] key is lit.
Programmable test currents are available from 0.01mA~99.99mA for test frequencies of
42.0Hz~1.00MHz. For test frequencies above 1.00MHz, the current can be adjusted between 0.01mA and 20.00mA. The 3550’s test current is adjustable in 0.01mA steps and is displayed to three significant digits. Refer to the Appendix for instructions to calculate the maximum test current available in the constant current mode for a given impedance.
4-11
Page 50
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
[SHIFT] ĺ [2]: Test Current Set Mode cont’d:
If DISPLAY C is set to indicate test current and the 3550 is in the Measurement mode, the current amplitude can be incremented or decremented by pressing the [UP] and [DOWN] keys in the DISPLAY C panel. If the [UP] and [DOWN] keys are held down the instrument will increase the speed at which the current is incremented or decremented. The test current can be manually entered by accessing the instrument’s current set mode. Below are instructions to manually set the test current.
Enter the Current Values Press [SHIFT] ĺ [2] to enter the current set mode. The most significant digit in DISPLAY C
will flash indicating that a new value must be entered. Press a key from [0]~[9] to set the value of the first digit and to proceed to set the next digit value. Once again, press a key from [0]~[9] and proceed to the next digit.
Verify and Set the Test Current Once all three digits have been entered, double-check the final current value. Any of the digits may be accessed and modified by pressing the [CURSOR Ź] or [CURSOR Ż] keys and entering a new value. Press the [ENTER] key to save the current setting and to return back to the measurement mode. The 3550 will beep three times and will not return to measurement mode if a programmed test current exceeds the instrument’s sourcing capability.
[SHIFT] ĺ [3]: Averaging Set Mode
You can define the number of measurements the 3550 makes before it produces a readable measurement. Averaging is used to stabilize the output reading and to increase the accuracy of a reading. To set the number of averages per reading, press [SHIFT] ĺ [3]. The active average setting will be displayed on DISPLAY B. The first digit will flash indicating that one of the numeric keys, [0]~[9] must be pressed to set the value of the first digit. Once a value has been entered then the next digit will flash. Enter the desired values for the three digits and press the [ENTER] key to store the new setting. The number of averages may be set from 1~100 readings. If an illegal value is attempted the instrument will beep three times when the [ENTER] key is pressed to indicate that the attempted value is outside of the instrument’s acceptable range.
[SHIFT] ĺ [4]: Key Lock Mode
In the Key Lock mode, all front panel keys except the [SHIFT]" Key, [4] Key, and (in the case that the Measurement Trigger is set to Manual) the [MAN/EXT] Key will be locked. The Key lock mode is used to protect the 3550 settings from unintentional changes. To access the Key Lock Mode settings, press the [SHIFT] ĺ [4] keys. A single digit, either a “1” or a “0” will be displayed in DISPLAY B. Set the digit to “1” to lock the front panel or to “0’ to unlock the front panel keys. Press the [ENTER] key to finalize the setting.
4-12
Page 51
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
[SHIFT] ĺ [5]: Comparator Buzzer Function
Press the [SHIFT] ĺ [5] keys to enter the Buzzer Settings Mode. The active setting, 0~2 will blink in DISPLAY B.
Buzzer Function Value
Buzzer OFF 0 GO = Buzzer ON 1 NO-GO = Buzzer ON 2
Press either [0], [1] or [2]. The new entry will appear in DISPLAY B. Press [ENTER] to finalize the setting and to return to the measurement mode.
[SHIFT] ĺ [6]: Trigger Delay Time
The trigger delay time is programmable from 0~10,000 milliseconds (10s). It is used to add a time delay immediately after the detection of a trigger signal from either the [MAN/EXT] key, external control device (I/O Control Connector), RS-232C or GPIB Interfaces. In the Trigger Delay Time Settings Mode, the active setting for the trigger delay time will appear flashing on DISPLAY B. The default setting for the Trigger Delay Time is “00000 ms” The active flashing number can be changed by highlighting the active digits by pressing the [CURSOR Ź] or [CURSOR Ż] keys and entering a new value by pressing the [0]~[9] keys. When all the necessary digits have been changed press the [ENTER] Key to finalize the settings and return to Measurement mode.
4-13
Page 52
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
[SHIFT] ĺ [7]: Comparator Functions Mode
Press [SHIFT] ĺ [7] to access the Comparator Functions Mode. Either a flashing “0” or a “1” will appear on DISPLAY B. The meaning of the numbers is as follows:
0: Absolute Value Comparator Function 1: Percent Comparator Function
To change the flashing digit, press either [0] or [1] and press [ENTER] to store the new setting and return to the measurement mode.
[SHIFT] ĺ [8]: Defining Spot-Correction Frequencies for Zero Adjustments
The spot-correction frequency mode allows the user to identify up to three specific test frequencies where open and short-circuit, zero adjustments may be performed. This feature condenses the default zero adjustment procedure by limiting the zero adjustments to user-defined frequency points instead of stepping through the entire range of test frequencies (42Hz~5MHz). Performing spot-correction frequency zero adjustment saves time by eliminating the zero adjustment of unused test frequencies.
Select the Spot Frequency Number Press [SHIFT] ĺ [8] to enter the spot-correction frequency editing mode. The BIN LED will display a flashing “1”, “2”, or “3”. Press either the [1], [2], or [3] keys to select the spot frequency you would like to set. Entering the Spot Frequency Value Once a key is pressed, DISPLAY C will display “---“ indicating that a spot frequency must be defined. Press the [0]~[9] keys to enter the spot frequency digits. Next, press the [DP] key to position the decimal point. Press either of the [CURSOR Ź] or [CURSOR Ż] keys to highlight the frequency units. Press the [UNIT] key until the desired frequency units annunciator starts to blink. Press the [ENTER] key to finalize the spot frequency setting.
To disable the Spot Correction Frequency, make sure that you are in the spot frequency editing mode (press [SHIFT] ĺ [8]) use the [CURSOR] keys to highlight DISPLAY C. When this is done, press the [IGNORE] key to erase the current spot frequency setting.
4-14
Page 53
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
[SHIFT] ĺ [9]: RS-232C Settings Mode
For the RS-232C, the baud rate, data length, parity and stop bit can be defined by the user. Press the [SHIFT] ĺ [9] key to enter into the RS-232C setup mode.
After entering the RS-232C setup mode, you will notice that the BIN LED will begin to blink. Press a key from [1] ~[4] to select the RS-232C parameter that you wish to set.
[1]: Set the Baud Rate Default=5, 9600BPS [2]: Set the Data Length Default=1, 8 Bits
[3]: Set the Parity Default=0, No Parity
[4]: Set the Stop Bit Default=0, 1 Stop Bit
Notice that after selecting a parameter from 1-4, DISPLAY B will show a number that corresponds to the parameter setting as summarized in the tables below: Use the tables to determine the appropriate setting for the RS-232C communications. Use the [CURSOR Ź] or [CURSOR Ż] keys to navigate between the RS-232C parameter and the setting on DISPLAY B. Press the [ENTER] key to finalize the settings and return to measurement mode.
BIN Display "1" Baud Rate Settings
DISPLAY B Settings
0 300bps
1 600bps
2 1200bps
3 2400bps
4 4800bps
5 9600bps
6 19.2kbps
7 38.4kbps
RS-232C Settings
BIN Display "2" Data Length Settings
DISPLAY B Settings
0 7bits
1 8bits
RS-232C Settings
BIN Display "3" Parity Settings
DISPLAY B Settings
0 No parity
1 Odd parity
2 Even parity
RS-232C Settings
BIN Display "4" Stop Bit Settings
DISPLAY B Settings
0 1bit
1 2bit
RS-232C Settings
4-15
Page 54
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
[SHIFT] ĺ [-]: Correction Frequency Limit Setting Mode
As an alternative to zero correcting the entire frequency range 42Hz~5MHz, this function allows the user to set the correction frequency range in terms of upper and lower limits. Press [SHIFT] ĺ [-] to enter the Correction Frequency Setting Mode. The “BIN” LED will be flashing. Press either [1] to edit the low limit correction frequency or [2] to edit the high limit correction frequency. Use the [CURSOR Ź] or [CURSOR Ż] keys to navigate between the “BIN” LED, DISPLAY C Digits, Decimal Point and Frequency Units. To change the frequency value indicated on DISPLAY C, simply press the [0]~[9] keys. Press the [DP] key to change the position of the decimal point and use the [UNIT] key to select Hz, kHz, or MHz units. Press the [ENTER] key in order to finalize the new setting and to return to measurement mode.
NOTE: While in the Spot Correction Frequency Mode, the Spot Correction Frequency high or low limit setting can be enabled or disabled. This is done by pressing the [IGNORE] key.
[SHIFT] ĺ [OPEN]: Spot Correction Mode (Open Correction)
The spot correction Mode is used for making open and short-circuit frequency corrections. Pressing the [SHIFT] ĺ [OPEN] keys will allow an open circuit, spot frequency correction to take place. The open circuit correction is performed only on the three spot frequencies that are defined by pressing [SHIFT] ĺ [8], (see Defining Spot-Correction Frequencies for Zero Adjustments). Note that Spot Frequency Zero Corrections have precedence over Normal Zero Corrections. This means that the open and closed-circuit, zero correction factors for the spot frequencies will be stored and retained in the instrument’s memory and will be written over unless an additional spot frequency correction procedure is performed. Performing a standard zero correction procedure will write over the zero adjustment factors of all test frequencies except the three user-defined, spot correction frequencies.
[SHIFT] ĺ [SHORT]: Spot Correction Mode (Short Correction)
Pressing [SHIFT] ĺ [SHORT] initiates the Short-Circuit, Spot Frequency Correction process. The offset adjustment affects only the three user-defined Spot Correction Frequencies discussed in the, Defining Spot-Correction Frequencies for Zero Adjustments section ([SHIFT] ĺ [8]). Like the Open Circuit Spot Frequency Correction Mode, Short Circuit Spot Corrections have precedence over Normal Short Circuit Zero Corrections. Closed circuit, spot correction factors will be stored and retained in the instrument’s memory and cannot be written over unless an additional spot frequency correction procedure is performed. Performing a standard zero correction procedure will write over the zero adjustment factors of all test frequencies except the three user-defined, spot correction frequencies.
4-16
Page 55
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
7 “UNKNOWN" Terminals Panel
Four BNC-type terminals plus one GUARD are available for making shielded or unshielded Kelvin­type measurement connections to the DUT. Their designations and functions are summarized below:
Table 4.3 – Unknown Terminals
Connection
Measurement Function
Type/Description
BNC L BNC L BNC H BNC H Binding Post GUARD Five-wire shielded measurements
Low Side Current Source
FORCE
Low Side Voltage Sense
SENSE
High Side Current Source
FORCE
High Side Voltage Sense
SENSE
8 "COMPARATOR LIMIT SET" Panel - [CURSOR Ź] [CURSOR Ż] [LOW] [HIGH] When programming comparator settings, the keys in the Comparator Limit Set Panel are used to define the DISPLAY A & DISPLAY B Comparator values. The two left most keys are the [CURSOR Ź] or [CURSOR Ż] keys. In the comparator set mode, these cursor keys allow navigation between the present BIN number (1-9) and DISPLAY A and DISPLAY B comparator value settings.
If a value is defined for either DISPLAY A or DISPLAY B (by pressing digits [0]~[9] while the DISPLAY digits are blinking.) then the functionality of the [CURSOR Ź] or [CURSOR Ż] keys extends to allow navigation of the comparator value’s decimal point location and selection of the DISPLAY A and DISPLAY B units of measure.
The Comparator's High and Low Limit Values for DISPLAY A and DISPLAY B can be set by pressing the [LOW] and [HIGH] Keys while in the measurement mode. Pressing either of these keys will put the 3550 into the Comparator Limits Set Mode. Once in the Comparator Limits Set Mode, the BIN# and values for DISPLAY A and DISPLAY B comparator settings can be defined by using the secondary functions (blue text) of the keys located in the CIRCUIT MODE, RANGE, DISPLAY C, and DISPLAY A-B Panels. These secondary functions include [0]~[9], [-], [D.P.], [UNIT], and [IGNORE] commands.
NOTE: Only panels 1~9 can have comparator limit values assigned to them.
9 The "GUARD" Terminal
This terminal is connected to the chassis of the 3550. When measuring High-Impedance devices, connect the device’s shield to this binding post.
4-17
Page 56
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
10 Remote Interface Status LEDs – “SRQ” “LTN” “TLK” “RMT”
These LEDs indicate the status of remote communications using the 3550’s standard RS-232C and optional GPIB (IEEE-488.2) interfaces All four annunciators are used for GPIB communications. If the RS-232C is the controller, then only the "RMT" LED is used.
11 "PANEL" Selector Switch
Pressing the [0-9] key in the Panel section allows the user to select a preset panel setting from 0-
9. Each consequent press of the [0-9] key will increment to the next preset panel number. The active instrument settings are automatically updated to the panel number indicated on the seven segment LED (Panel Display LED). If the instrument settings are changed while a panel number (0-9) is active and the panel key is pressed then the existing settings will be stored into the panel location indicated by the Panel Display LED. Only panels 1~9 can have comparator limit values assigned to them.
12 "PANEL" Display LED
The PANEL Display LED is a seven segment LED that displays the active panel setting. Press the [0-9] key located in the Panel Section to increment through panel settings 0-9.
13 "DISPLAY A" Parameter Selection Key - [A] Press this key to select the measurement parameter of DISPLAY A. Measurements of L, C, R, |Z| and |Y| are available.
14 "DISPLAY A"
Measured values of L, C, R, |Z|, and |Y| are shown in this panel. The measurements values range from 0000 to 19999 and are displayed with 4
1
/2- digit resolution. In addition, units of measure are annunciated above and beside the display for reading convenience. To change the parameters shown by DISPLAY A, press the [A] key located in the DISPLAY A-B instrument panel until the desired measurements are shown.
15 "BIN" Display LED
The BIN Display LED is a seven segment LED that displays the active sorting bin number of the Comparator. When the Comparator is ON, this LED displays which BIN Number the present measurement is associated with. "1" through "9" are for OK products, while "0" is reserved for defects such as open circuit conditions.
4-18
Page 57
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
16 Comparator State Annunciators
Located in the Comparator Panel there are three total comparator state annunciators. The annunciators become active when the comparator is on. The comparator is turned on by pressing the [ON] key in the Comparator Panel. See below for a summary of the annunciator functions.
Annunciator Functional Description
DISP-A LO-HI
DISP-B LO-HI
TOTAL GO
Indicates when the measured value of DISPLAY A is equal to or outside of the Low or High Comparator setting. Indicates when the measured value of DISPLAY B is equal to or outside of the Low or High Comparator setting. Indicates when the measured value of DISPLAY A and DISPLAY B are within the High and Low limits of the Low or High Comparator setting. If DISPLAY A comparator is disabled then the TOTAL GO DISPLAY B determines condition only. If DISPLAY B comparator is disabled then the TOTAL GO condition is determined by DISPLAY A only.
17 Sample LED (Trigger Status)
This LED displays the measurement status. When the TRIGGER mode is set to "INT", this LED will flash continuously. When the TRIGGER mode is set to "MAN/EXT", this LED will flash every time a measurement is made.
18 "COMPARATOR ON/OFF" Switch – [ON] This switch enables or disables the Comparator. When the switch is in the ON state (LED on), the BIN Display LED, Comparator State Annunciators, and Comparator functions are enabled.
19 "DISPLAY B" Parameter Selection Key - [B] Press this key to select the measurement parameter of DISPLAY B. Measurements of D, Q, RS, RP, G, X, B and LJ are available.
20 "DISPLAY B"
Measured values of D, Q, RS, RP, G, X, B and LJ are shown in this panel. The measurements values are displayed with 4
1
/2- digit resolution. In addition, units of measure are annunciated above and beside the display for reading convenience. To change the parameters shown by DISPLAY B, press the [B] key located in the DISPLAY A-B instrument panel until the desired annunciator is lit.
4-19
Page 58
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
21 DISPLAY C Measurement Panel – [FREQ] [ V ] [ I ] [CV/CC] [DOWN] [UP]
The keys on the DISPLAY C Measurement panel are used to set the measurement frequency, constant voltage, voltage or constant current mode, test signal amplitude and the measurement that will be shown on DISPLAY C.
[FREQ] – Test Frequency Select Key
When [FREQ] is pressed, the test frequency will be shown on DISPLAY C. The displayed test frequency setting can be adjusted using the [DOWN] and [UP] Keys to increment or decrement the existing test frequency setting. To define the test frequency value, decimal point location and units refer to the [SHIFT] ĺ [0]: Frequency Set Mode section.
[V] – Show Test Voltage on DISPLAY C
When the [V] key is pressed the test voltage will be shown on DISPLAY C. Pressing either the [UP] or [DOWN] keys will adjust the test voltage. The test voltage, decimal point and units can be set by using the [SHIFT] ĺ [V] key sequence.
The Model 3550 has programmable test voltages from 0.01V~5.00V for test frequencies of
42.0Hz~1.00MHz. For test frequencies above 1.00MHz, the voltage can be adjusted between 0.01V and 1.00V. Test voltages are adjustable in steps of 0.01V and are displayed in DISPLAY C to three significant digits.
When in the constant voltage mode, and the [UP] or [DOWN] keys are pressed, DISPLAY C will momentarily show the open circuit test voltage setting before displaying the constant voltage applied to the actual test device terminals. If either the [UP] or [DOWN] key is pressed and held, the voltage setting will display and the setting will either scroll up or down (depending on whether the [UP] or [DOWN] key is pressed.
[I] – Show Test Current on DISPLAY C
When [I] is pressed, the test current flowing through the DUT will be displayed. Pressing either the [UP] or [DOWN] keys will adjust the test current level. The test current, decimal point and units can be set by using the [SHIFT] ĺ [I] key sequence.
Programmable test currents are available from 0.01mA~99.99mA for test frequencies of
42.0Hz~1.00MHz. For test frequencies above 1.00MHz, the current can be set between
0.01mA and 20.00mA. The 3550’s test current is adjustable in 0.01mA steps and is displayed to three significant digits.
When in the constant current mode, and the [UP] or [DOWN] keys are pressed, DISPLAY C will momentarily show the open circuit test voltage setting before displaying the actual current flowing through the test device. If either the [UP] or [DOWN] key is pressed and held, the voltage setting will display and the setting will either scroll up or down (depending on whether the [UP] or [DOWN] key is pressed.
4-20
Page 59
Model 3550 LCR Meter Instruction Manual Operating Instructions
Front Panel Description cont’d:
Test Signal Modes - [CV/CC]
Model 3550 is capable of three different measurement modes. These are Open Circuit Test Voltage Mode, Constant Voltage Mode, and Constant Current Mode. The active measurement mode can be confirmed by looking at the [V], [I], and [CV/CC] LED Keys in the DISPLAY C panel. The relationship between the operation modes and the LED displays are as follows:
Table 4.4 – CV/CC Indicators
Measurement Mode
[V] LED
[I] LED
[CC/CV] LED
Open Circuit Test Voltage Mode ON OFF OFF
Constant Voltage Mode ON OFF ON
Constant Current Mode OFF ON ON
Set Voltage Mode
Measurements are taken using the open circuit Test Voltage. The test voltage varies with current. Use this mode for ordinary test objects that possess an overall impedance value with real and imaginary components.
Constant Voltage Measurement Mode
This makes the measurement while keeping the voltage of the measurement terminals at a fixed level. For example, this mode can be used for testing inductors where terminal voltage will vary. Using Constant Voltage mode can help stabilize a reading.
Constant Current Measurement" Mode
This makes the measurement stable by keeping the current passing through the DUT at a fixed level. For example, this mode can be used for testing capacitors where test current continuously varies with applied voltage.
22 "DISPLAY C"
Displays Frequency (Hz, kHz, and MHz), Open Circuit Test Voltage (V), Terminal Voltage (Constant V), and DUT Test Current (Constant mA) based on the settings of the DISPLAY C Measurement Panel. The measurement values are displayed with 4
1
/2- digit resolution. Units of measure are
annunciated beside the display for reading convenience.
23 ENTER Key – [ENTER] Used in the various programming modes to finalize user-defined instrument settings. For example, the [ENTER] key is pressed to finalize any of the [SHIFT] ĺ [0~9] commands and to return back to the measurement mode.
4-21
Page 60
Model 3550 LCR Meter Instruction Manual Operating Instructions
Rear Panel Description
Figure 4.3 - Rear Panel Layout
Line Voltage
Selector Plug
Fuse Rating
Chart
Fuse Input
VAC Input
GND Terminal
Line Voltage Selector Plug – The Model 3550 is
capable of 100/120/220/240 VAC 50/60 Hz Operation. To change the line voltage, remove the plug by pulling it from the rear panel. Put the plug back in to the socket with the arrow pointing to one of four supply voltage settings.
Fuse Input – Use 3AG Type Fuses, ¼” X 1¼” (6.35X31.75mm)
Fuse Rating Chart - See Rear Panel or Preparation for Use Section for fuse ampere ratings. Use fast acting fuses only.
VAC Input – 100/120/220/240, power input. An internal line filter is included in this input.
GND Terminal – The GND binding post provides a direct connection to the instrument’s chassis and the grounded connector of the VAC input plug.
GPIB Board
Slot
RS-232C Port
Cooling Fan
Ventilation
Control I/O
Port
Cooling Fan Ventilation Holes – To prevent
overheating, keep these ventilation holes free of obstructions. Periodically inspect these holes to assure they are free of obstructions.
Control I/O Port – The Control I/O Port has optically isolated TTL inputs and outputs used for remote operation. Inputs include BCD Panel Selection and trigger inputs. Outputs include A, B and total Comparator outputs, End of measurement, BIN, Busy and Error Outputs. The control connector also includes pins for external power connections.
RS-232C Port – Standard RS-232C Port.
GPIB Board Slot – Slot provided for factory installed
GPIB (IEEE-488.2) option PN# 3505.
4-22
Page 61
Model 3550 LCR Meter Instruction Manual Operating Instructions
L
L
T
L
L
L
L
T
Making Accurate Measurements:
Connections to the Device Under Test (DUT)
The relationship between the connections schemes and impedance, |Z| ranges are shown in Table 5-1. It is extremely important, for measurement accuracy purposes, to observe appropriate connection methods for the subject test piece and measurement range.
Range 1 2 3 4 5 6 7 8 9
Impedance Mode
Connection Type
Low Impedance (Series Equivalent Mode)
5 Terminal Connection
High Impedance (Parallel Equivalent Mode)
3 Terminal Connection
3-Terminal Measurement
GUARD
GUARD
FORCE
FORCE
SENSE
SENSE
H
H
SENSE
SENSE
DU
4 or 5-Terminal Measurements
GUARD
FORCE
SENSE
H
DU
SENSE
H
H
H
FORCE
FORCE
FORCE
This measurement method is used for testing devices having high impedance. In general, components with large inductance, low capacitance, or high resistance fall into this category. The major advantage to 3-terminal measurements is that the influence of stray capacitance and conductance between the test leads and nearby conductors becomes negligible.
The 5-terminal measurement method can be used for all impedance ranges and has significant advantage over the 3-terminal measurement method. The advantage is that in addition to canceling the effects of stray capacitance and conductance between the measurement leads and close proximity conductors, the residual inductance and resistance of the test lead are bypassed by placing the voltage sense points of contact directly at the DUT terminals. A four-terminal measurement is the same as a five-terminal measurement except that the guard connection is not used.
4-23
Page 62
Model 3550 LCR Meter Instruction Manual Operating Instructions
Making Accurate Measurements:
Test Fixture Selection
Model 3510 Radial/Axial Lead Adapter
The Model 3510 Radial/Axial Lead adapter is a general-purpose accessory used for testing components with radial or axial leads. Components are tested by plugging the leads directly into the adapter sockets. The Model 3510 is designed for 5-terminal measurements and is capable of measuring a wide range of impedances.
Model 47454 Kelvin Klips™
Model 47454 Kelvin Klip Leads are designed for 5­wire, Kelvin measurement of all types of component styles. They may be used for both low and high impedance measurements. Kelvin Klips are particularly convenient for low impedance test devices or components with short leads, however, the measurement frequency and impedance range are limited as shown in Table 4.4. Exceeding the recommended measurement ranges may result in erroneous and inaccurate readings.
Table 4.4 - Impedance and Frequency Limitations of Kelvin Klip Leads
Frequency Measurement Range 42Hz ~ 10kHz 100mƻ ~ 1Mƻ
10.1kHz ~100kHz 1ƻ ~ 100kƻ 101kHz ~ 1.00MHz 10ƻ ~ 10kƻ
NOTE: Additional test voltage and current restrictions of the 3550 unrelated to test fixtures apply for measurements exceeding 1MHz.
4-24
Page 63
Model 3550 LCR Meter Instruction Manual Operating Instructions
Making Accurate Measurements cont’d:
Test Fixture Selection cont’d
Model 3511 Surface Mount Device (SMD) Test Fixture
The 3511 is a test fixture designed to perform 3-wire impedance measurements of surface mount devices. This adapter is suitable for measurement of chip components with middle to high impedance ranges. The measurement frequency and range limitations are summarized in table 4.5 below Exceeding the recommended measurement ranges may result in erroneous and inaccurate readings. The signal level limitations of the instrument must also be considered when making measurements above 1MHz.
Table 4.5 - Impedance and Frequency Limitations of the SMD Test Fixture
Frequency Measurement Range 42Hz ~ 100kHz 10ƻ ~ 1Mƻ 101kHz ~1.00MHz 10ƻ ~ 10kƻ
1.01MHz ~ 5.00MHz
10ƻ ~ 10kƻ (10ƻ Minimum)
Model 2005B Chip Component Tweezers
2005B is designed for manual sorting of chip components from DC to 1MHz. Four-terminal tweezers make solid connections on chips up to
0.50 in. (12.7mm) long. The set includes a 5 ft. (1.5m) cable for connection to the 3550
47422 – Chip Tweezer Rebuild Kit
Tweezer tips are intended to last from 100,000 to 500,000 operations. This tip replacement kit includes 12 replacement tips, 2 screws, and 1 wrench.
4-25
Page 64
Model 3550 LCR Meter Instruction Manual Operating Instructions
Making Accurate Measurements cont’d:
Zero Corrections
Residual Impedance and Leakage of test fixtures exists for all types of test fixtures. To maintain the performance accuracy of the 3550 it is highly recommended that a zero correction be performed whenever a change in test fixture takes place. For test fixtures such as Kelvin Klip Leads where the positioning may vary, extra effort must be made to perform the zero correction while positioning the leads in the exact location, as the measurement will be taken.
When performing a standard zero correction, the instrument will step through frequencies from 42Hz to 5MHz. This takes a considerable amount of time and may not be desired in certain applications especially when the actual component test will take place at a single frequency point. To overcome this, the 3550 has the ability to perform spot frequency corrections where zero corrections are performed on up to three user defined frequency set points. The spot frequency zero correction data is stored in the instrument’s non-volatile RAM and will overwrite any correction data obtained from a standard zero correction procedure. A standard zero correction cannot write over data from a spot frequency zero correction. The spot frequency setting must be disabled in order for the standard zero correction to be able to write over all frequency points. Refer to “Defining Spot-Correction Frequencies for Zero Adjustments”, [SHIFT] ĺ [8] in the front panel description to learn how to enable or disable the spot frequency settings.
Table 4.6 - Zero Correction Limits
Zero Correction Correction Correction Limit
Open Stray Capacitance & Stray Conductance (C0)&(G0) >1kƻ
Short Residual Inductance & Residual Resistance (L0)&(R0) < 1kƻ
If during OPEN correction the Impedance is less than 1kƻ, or during SHORT correction the Impedance is 1kƻ or higher, a buzzer will go off to warn that correction cannot be made, and the system will reset without making the correction.
There are three methods of defining what frequencies a zero correction will be performed on. They are Standard, Frequency Limit, and Spot Frequency Zero Correction.
Standard Zero Correction
Refer to Section 4 , “ZERO” Panel – for instructions to perform a standard open and short circuit correction. This procedure will perform a zero correction on all test frequencies, 42Hz~5MHz except for those designated as spot correction frequencies.
Frequency Limit Zero Correction
For zero correction, [SHIFT] ĺ [-], Correction Frequency Limit Setting Mode allows the user to define the upper and lower frequency limits instead of using the default 42Hz~5MHz range.
Spot Frequency Zero Correction
Refer to [SHIFT] ĺ [OPEN] and [SHIFT] ĺ [SHORT] sections for instructions to perform a spot frequency open and short zero correction procedure. After performing the zero correction procedure, perform a “before and after” measurement on a component with an accurately known value to confirm that a zero correction has taken place. A calibrated capacitance or inductance standard would be ideal (but not required) for this type of verification.
4-26
Page 65
Model 3550 LCR Meter Instruction Manual Operating Instructions
Making Accurate Measurements cont’d:
Zero Corrections cont’d
NOTE: The parasitic impedance of test adaptors and measurement cables differ from cable to
cable. Whenever a test adaptor or cable is changed, make sure that zero correction is performed again.
Figure 4.4 - Terminal Connections for Zero Correction
Connection
Open Correction Short Correction
Normal Error Normal Error
LHGLHGLHGLHG
BNC Terminals
LHGLHGLHGLHG
5-Terminal Correction
Kelvin Klip™ Correction
H
L
F
F
L
H
s
s
H
L
F
F
L
H
s
s
L
F
H
F
L
s
H
s
L
F
H
F
L
s
H
s
Use Figure 4.4 as a reference when performing short-circuit and open-circuit zero corrections.
4-27
Page 66
Model 3550 LCR Meter Instruction Manual Operating Instructions
Making Accurate Measurements cont’d:
Equivalent Circuits
The DUT can be described in terms of Impedance (Z) or Admittance (Y). The relationships of the various characteristics are as follows:
Z = R + jX
Z : Impedance R : Equivalent Resistance X : Reactance (jX) Y : Admittance G : Conductance B : Susceptance (jB)
Y = G - jB =
RXG
D = =
1
Q =
D
1
Z
B
D : Loss Coefficient
XL = ZL = 2SfL
XC = =
1
Z
C
2
1
S
fC
Q : Quality ș : Phase Angle f : Measurement Frequency Hz Ȧ : Measurement Frequency rad/s
T
= tan-1 = tan
X R
B
-1
G
The measurable quantities of the 3550 are L, C, R, Z, Y, D, Q, G, Rp, Rs, X, B, and LJ. The measured value of these parameters will change in value as the test frequency is increased or decreased. These measured parameters are said to be dependent upon test frequency.* Two examples are capacitive reactance, which is inversely proportional to frequency and inductive reactance, which is inversely proportional to frequency. Because of the direct relationship to test frequency, related quantities such as Dissipation Factor and Quality Factor is also frequency dependent.
*In measurement applications, the measured R and G values will change with frequency. True R
and G are not theoretically dependent to frequency but in LCR measurement applications, parallel leakage and series losses can affect the measured R and G value of the component. Compensation for these errors is the reason for Series and Parallel Equivalent Circuit Measurement Modes.
In general, parallel leakage becomes significant in highly reactive (small capacitors or large inductors) measurements and series resistance is more significant in low reactance (large capacitors and small inductors) measurements. This is the reasoning behind the use of series or
parallel equivalent circuit measurement modes.
If the dissipation factor, D is small then the difference between Series Equivalent and Parallel Equivalent Circuit Measurement Mode measurement will be small. As the dissipation factor increases, the difference between parallel and series equivalent measurements will increase. Thus, precautions must be taken when performing measurements on components with high dissipation factors.
4-28
Page 67
Model 3550 LCR Meter Instruction Manual Operating Instructions
R
R
Making Accurate Measurements cont’d:
Equivalent Circuits cont’d
Table 4.7 Illustrates the effects of leakage and series resistance on component measurements. It also provides some additional formulae to illustrate the relationship that dissipation factor has on a device’s impedance measurement.
Table 4.7 – Equivalent Circuit Relationships
Equivalent
(SER)
(PRL)
L
(SER)
C
(PRL)
Circuit
Mode
Ls R
G
L
P
Cs R
G
C
P
Dissipation Factor
(D) Conversion Formula
S
LPG
CSR
P
1
Q
1
Q
1
Q
1
Q
R
D = =
Z
L
Z
D = =
D = =
Z
G
D = =
Z
C
P LD1L
S L
L
P C
C
S CD1C
2
1
2
D1
1
2
D1
2
2
D
G
S
R
D
P
G
D
S
D
R
P
1
2
D1
2
1
2
G
D1
2
1
2
D1
2
1
2
G
D1
The Series Equivalent Circuit Mode of the 3550 can be used for testing devices where series impedance has a great effect on the measurement. Parallel Equivalent Circuit Mode is used for testing devices with high reactance where parallel leakage becomes more significant. The formulae in Table 4.7 can be used to convert readings in one equivalent circuit mode to the other.
Note that the measured value of D (or Q) will not be different whether it is measured in Series or Parallel Equivalent circuit mode.
4-29
Page 68
Model 3550 LCR Meter Instruction Manual Operating Instructions
Operating the 3550
Preparation
Check the Power Source Voltage. The voltage range of this machine is displayed on the rear of the machine. The power voltage may be selected using the switching plug on the rear panel. See “Preparation for Use” for more setup details.
When the [POWER] Switch is turned ON the settings of the machine will be same as when the power was previously turned off. In order to reset the 3550, simply press [MAN/EXT] at the same time the [POWER] key is pressed. Resetting will clear the zero correction factors and return the instrument settings to factory defaults. A table summarizing the factory default settings can be found in Chapter 3, Quick Start Instructions.
How to Perform a Zero Correction
Determine whether test leads, an adapter or some other fixture will be used to make measurements and connect it to the front panel BNC terminals. Perform a Standard, Frequency Limit, or Spot Frequency Zero Correction for both open and closed circuit conditions. To perform a standard zero correction, follow the instructions below:
a) With the tips of the test leads or the adapter in an "open" condition, push the [OPEN]
switch. DISPLAY A will show “OP–57” to indicate that an open correction is being performed and “57” will indicate what step frequency the instrument is adjusting for. The instrument will count down from “57” to “1”. Do not touch or adjust the test leads/fixture while the zero correction procedure is taking place.
b) Once the “OPEN” correction is completed the unit will return to the measurement mode.
Connect the test leads/fixture in the "shorted" position (see Figure 4.X) and press the [SHORT] key. The instrument will execute the short circuit correction procedure.
During the Open or Short Circuit Zero Correction process, if the Correction Value min/max threshold is exceeded, a warning buzzer will sound 3 times to warn that the correction could not be done. If this happens, lower the parasitic Impedance, and re-do the correction. If the test leads/fixture continue to fail the zero correction, double check that the operating frequency range of the leads/fixture have not been exceeded. See “Test Fixture Selection” to determine whether the test frequency range of your test leads/fixture have been exceeded and whether a Frequency Limit or Spot Frequency Zero Correction need to be performed instead of a Standard Zero Correction. Frequency Limit and Spot Frequency Zero Corrections allow the user to define the specific frequencies to be open and short compensated. Frequency Limit and Spot Frequency Corrections can also be used to shorten the total zero correction time by ignoring unused test frequencies.
If the proper frequency correction has been performed and the test leads/fixture continues to fail zero correction, have the leads/fixture tested by an authorized service person.
!
Detailed instructions for performing "open " or "short” circuit zero corrections are found in the “Measurement Tips” and in “Front panel Description” sections of this manual. Make sure that front panel operation and LCR measurement technique is thoroughly understood before proceeding further. LCR measurement can be quite complex and the slightest overlooked detail could greatly affect measurement accuracy. To maximize measurement accuracy it is highly recommended that this chapter be understood in its entirety.
4-30
Page 69
Model 3550 LCR Meter Instruction Manual Operating Instructions
Operating the 3550 cont’d:
Measuring Inductance (L), Capacitance (C) and Resistance (R)
1) Always perform a zero correction procedure if the condition of the test leads/fixture has changed. This procedure must be performed before making a measurement. Refer to “How to Perform a Zero Correction” for more details.
2) Press the [A] key and select "L", "C" or "R" on DISPLAY A.
3) Press the [FREQ] key and adjust the frequency using the [UP] or [DOWN] keys or by manually entering a new value for test frequency. See [SHIFT] ĺ [0]: “Frequency Set Mode” for more details on manually setting the test frequency.
4) Press the [V] key, and adjust the measurement voltage using the [UP] or [DOWN] keys or by manually entering a new value for test frequency. See [SHIFT] ĺ [1]: “Test Voltage Set Mode” for more details on manually setting the test frequency.
5) Press the [AUTO] key in the RANGE panel and verify that the “AUTO” LED is on. This puts the 3550 in AUTO range.
NOTE: When repeatedly measuring DUTs with similar values selecting the manual mode allows the measurement time to be shortened by eliminating the time needed for switching the ranges. See "RANGE" Selection Panel” for more details.
6) Press the [AUTO] key in the CIRCUIT MODE panel and verify that the “AUTO” LED is on.
7) Press the [INT] key in the TRIGGER panel and verify that the “INT” LED is on. This will continuously trigger the 3550 and update the reading shown on DISPLAY at a rate determined by measurement frequency, delay time, number of averages, and the time taken to auto range. If MAN/EXT trigger mode is selected the reading will not be updated until a new trigger is received from the front panel or remote interface.
8) Connect the DUT to the measurement terminals. The CIRCUIT MODE and RANGE will be automatically selected by the 3550. The L, C, or R reading will be shown on DISPLAY A.
4-31
Page 70
Model 3550 LCR Meter Instruction Manual Operating Instructions
Operating the 3550 cont’d:
Measuring Inductance (L), Capacitance (C) and Resistance (R) cont’d:
Notes of Caution for Inductance (L) Measurement
For low inductance measurements, 5 Terminal Measurements are required to minimize
series resistance and inductance errors. Margins of error for 3 terminal measurements on low inductances will add significant error to the reading.
When performing measurements on small coils with a core, it is possible that the measured
inductance value will vary with a change in test signal level. Take extra caution to match the test signal level of other machines in order to obtain matching results.
Keep metal objects away from the DUT when performing measurements. The presence of
metal objects in close proximity can affect the measured value. Especially in coils with air or other low permeability cores.
Notes of Caution for Capacitance (C) Measurement
For large capacitance measurements, 5 Terminal Measurements are required to minimize
errors caused by series resistance and inductance. Margins of error for 3 terminal measurements on large capacitances will add significant error to the reading.
!

Always make sure that capacitors are completely discharged before connecting to the 3550 measurement terminals otherwise erroneous readings or damage to equipment could result.
Notes of Caution for Resistance (R) Measurement
For low resistance measurements, 5 Terminal Measurements are required to minimize
series resistance and inductance errors. Margins of error for 3 terminal measurements on low resistance will add significant error to the reading.
4-32
Page 71
Model 3550 LCR Meter Instruction Manual Operating Instructions
Operating the 3550 cont’d:
Measuring Impedance (|Z|) and Admittance (|Y|)
1) Always perform a zero correction procedure if the condition of the test leads/fixture has changed. This procedure must be performed before making a measurement. Refer to “How to Perform a Zero Correction” for more details.
2) Press the [A] key in the “DISPLAY A-B panel to select "|Z|" or "|Y|".
3) Press the [FREQ] key and adjust the frequency using the [UP] or [DOWN] keys or by manually entering a new value for test frequency. See [SHIFT] ĺ [0]: “Frequency Set Mode” for more details on manually setting the test frequency. The measurement frequency will be displayed on DISPLAY C.
4) Press the [V] key, and adjust the measurement voltage using the [UP] or [DOWN] keys or by manually entering a new value for test frequency. See [SHIFT] ĺ [1]: “Test Voltage Set Mode” for more details on manually setting the test frequency. The measurement voltage will be displayed on DISPLAY C.
5) Press the [AUTO] key in the RANGE panel and verify that the “AUTO” LED is on. This puts the 3550 in AUTO range.
NOTE: When repeatedly measuring DUTs with similar values selecting the manual mode allows the measurement time to be reduced by eliminating the time needed for switching the ranges. See "RANGE" Selection Panel” for more details.
6) Press the [INT] key in the TRIGGER panel and verify that the “INT” LED is on. This will continuously trigger the 3550 and update the reading shown on DISPLAY A at a rate determined by measurement frequency, delay time, number of averages, and the time taken to auto range. If MAN/EXT trigger mode is selected the reading will not be updated until a new trigger is received from the front panel or remote interface.
7) Connect the DUT to the measurement terminals. The RANGE will be automatically selected by the 3550. The |Z| or |Y| reading will be shown on DISPLAY A.
4-33
Page 72
Model 3550 LCR Meter Instruction Manual Operating Instructions
Operating the 3550 cont’d:
Measuring: Loss Coefficient (D), Quality (Q), Equivalent Series Resistance (R Equivalent Parallel Resistance (R
), Conductance (G), Reactance (X), Susceptance (B),
P
),
S
and Phase Angle (ș)
1) Always perform a zero correction procedure if the condition of the test leads/fixture has changed. This procedure must be performed before making a measurement. Refer to “How to Perform a Zero Correction” for more details.
2) Press the [B] key in the “DISPLAY A-B panel to select D, Q, R
, RP, G, X, B, or LJ.
S
3) Press the [FREQ] key and adjust the frequency using the [UP] or [DOWN] keys or by manually entering a new value for test frequency. See [SHIFT] ĺ [0]: “Frequency Set Mode” for more details on manually setting the test frequency. Frequency will be displayed on DISPLAY C.
4) Press the [V] key, and adjust the voltage using the [UP] or [DOWN] keys. See [SHIFT] ĺ [1]: “Test Voltage Set Mode” for more details on manually setting the test frequency. The measurement voltage will be displayed on DISPLAY C.
5) Press the [AUTO] key in the RANGE panel and verify that the “AUTO” LED is on. This puts the 3550 in AUTO range.
NOTE: Selecting the manual mode allows the measurement time to be reduced by eliminating the time needed for switching the ranges. See "RANGE" Selection Panel” for more details.
6) Press the [INT] key in the TRIGGER panel and verify that the “INT” LED is on. This will continuously trigger the 3550 and update the reading shown on DISPLAY A at a rate determined by measurement frequency, delay time, number of averages, and the time taken to auto range. If MAN/EXT trigger mode is selected the reading will not be updated until a new trigger is received from the front panel or remote interface.
8) Connect the DUT to the measurement terminals. The CIRCUIT MODE and RANGE will be automatically selected by the 3550. The D, Q, R
, RP, G, X, B, or LJ reading will be shown on
S
DISPLAY B.
Notes for Measuring Reactance & LJ
On occasion a “-“ sign will be observed occupying the same display as the most significant
digit “1”. The minus sign indicates a negative phase angle from a capacitive reactance, X
.
C
This is normal functionality in order to maximize the number of significant digits for an upper-range measurement.
Display B indicating 16.775ƻ XC Value Display B indicating 16.775ƻ XL Value
4-34
Page 73
Model 3550 LCR Meter Instruction Manual Operating Instructions
The Comparator
Important Notes for the Absolute and Percent Comparator Functions
Comparator settings can only be defined and stored into panels 1-9. Before a comparator setting can be defined, a panel, 1-9 must be recalled.
This is done by pressing the [0-9] key until the desired panel is active while the 3550 is in the measurement mode.
Settings for DISPLAY A, DISPLAY B and CIRCUIT MODE must be finalized in the panel setting, [1­9] before defining comparator upper and lower limits. This is because these values cannot be changed after the comparator high and low limits have been set.
If any of the DISPLAY A, DISPLAY B, or CIRCUIT MODE keys, ([A][B][AUTO][SER][PRL]) are pressed after the comparator limits have been set, the [LOW] and [HIGH] LEDs of the comparator will start flashing to indicate an illegal operation. If either the [LOW] or [HIGH] keys are pressed afterwards then all of the comparator settings will be cleared. This will be indicated by “_____” appearing on the display. Pressing any other keys on the front panel will return the instrument to measurement mode without affecting the comparator settings.
The 3550 offers a dual-function comparator meaning that for each BIN there are two comparator settings that may be defined. The DISPLAY A setting allows a high and low limit to be defined for L, C, R, |Z|, and |Y| measurements while DISPLAY B allows limits to be set for D, Q, RS, RP, G, X, B, or LJ. The comparator is equipped to offer an output based on the measurement results of DISPLAY A only, DISPLAY B only or the combined measurement results of both DISPLAY A and DISPLAY B.
The comparator may be used in absolute or percent mode. In the absolute mode, the comparator high and low limits are defined by entering absolute value high or low limits. This means that the high and low limits are entered as actual magnitudes and units of measure. The percent mode requires that the high and low limits are defined in terms of a percentage of a nominal value. The following sections provide instruction for setting the high and low limits in the Absolute or Percent Comparator modes.
During high and low limit setup, if a numeric key is pressed that will create a limit value, which falls outside of the measured parameter’s range, the instrument will disable the functionality of that key.
Refer to the following sections for additional information regarding the 3550 comparator functions.
[SHIFT] [SHIFT] 8 - Comparator Limit Set Panel 18 – Comparator ON/OFF Switch
ĺ [5] –Setting the comparator buzzer function. ĺ [7] – Absolute or Percentage Comparator Function
4-35
Page 74
Model 3550 LCR Meter Instruction Manual Operating Instructions
The Comparator cont’d:
Absolute Value Settings
NOTE: A condensed version of this procedure can be found in the “Quick Start Section”
1) Make sure that panel 1~9 is selected. Do this by pressing the [0~9] key until the desired panel # appears on the Panel LED. Press [SHIFT] ĺ [7] to access the Comparator Functions Mode. Either a flashing “0” or a “1” will appear on DISPLAY B. The meaning of the numbers is as follows:
0: Absolute Value Comparator Function 1: Percent Comparator Function
To change the flashing digit, press [0] ĺ [ENTER] to call the absolute comparator mode and return to the measurement mode.
2) In the measurement mode, press the [LOW] key in the Comparator Limit Set Panel. Make sure
that a panel number from 1~9 is selected otherwise the comparator [HIGH] and [LOW] keys will be disabled. The [LOW] LED will light. The BIN segment LED will flash the numeral “1”. Press a
numeric key [1]~[9] to define the bin number.
th
3) Press the [CURSOR Ź] once. This will cause the 5 comparator low limit value using the [0-9] keys and the [CURSOR Ź] or [CURSOR Ż] keys to navigate between the 1
st
and 5th digits.After digits 1 through digit 5 has been set, press the
digit if DISPLAY A to flash. Type in the
[CURSOR Ź] key until the decimal point begins to flash. Press the [DP] key until the decimal point is in the desired position. Then press the [CURSOR Ź] key once again to set the units of measure. Units are defined by pressing the [UNITS] key until the desired annunciator illuminates.
6) After the units for the low limit (DISPLAY A) have been selected, press the [CURSOR Ź] key.
th
The 5
digit of DISPLAY B will begin to blink indicating that a value can be entered for the high
comparator limit.
7) Enter the values for the DISPLAY B low comparator limit in the same manner as performed for the DISPLAY A low limit.
NOTE: If the comparator function for DISPLAY A or DISPLAY B is not required, press the [IGNORE] key while the display is active and the comparator for that display will be disabled. A disabled display will show a “-----“ in place of a numerical value.
8) To set the high value limit for the comparator press the [HIGH] key located in the Comparator Limit Set panel. The comparator high limit set mode is indicated when the [HIGH] LED lights. Repeat steps 3-7 to set the comparator’s high limit values.
9) This completes setting the values for BIN #1. To set the high and low limits of the remaining BINs return to step #2. When the BIN LED is flashing select a BIN number from 1-9 by pressing the [1]-[9] keys and continue with the procedure.
NOTE: The BIN functionality is expressed as:
4-36
LOWBINHIGH
Page 75
Model 3550 LCR Meter Instruction Manual Operating Instructions
The Comparator cont’d:
Percent Value Settings
NOTE: A condensed version of this procedure can be found in the “Quick Start Section”
Set the Nominal Value
1) Make sure that panel 1~9 is selected. Do this by pressing the [0~9] key until the desired panel # appears on the Panel LED. Press [SHIFT] ĺ [7] to access the Comparator Functions Mode. Either a flashing “0” or a “1” will appear on DISPLAY B. The meaning of the numbers is as follows:
0: Absolute Value Comparator Function 1: Percent Comparator Function
To change the flashing digit, press [1] ĺ [ENTER] to call the percent comparator mode and return to the measurement mode.
2) In the measurement mode, press the [LOW] key in the Comparator Limit Set Panel. Make sure
that a panel number from 1~9 is active otherwise the comparator [HIGH] and [LOW] keys will be disabled. The [LOW] LED will light. This will cause the 4
a “0”. A flashing “0” indicates the Nominal Value Set Mode. This is where the nominal value of the measured value will be entered. High and low limits are calculated as a percent of this value. A flashing “1” indicates the High and Low Percent Limits Set Mode. The High and Low Percent Limits Mode requires a [HIGH] and [LOW] comparator setting. The Nominal Value Set Mode requires only one setting. Press the [0] key to set the instrument to Nominal Value Set Mode.
th
digit if DISPLAY C to flash either a “1” or
3) Press the [CURSOR Ź] once. The BIN LED in the Comparator Panel will flash a numeral “1”. Press a numeric key [1]~[9] to set the active bin number. BIN 0 cannot be used with the comparator function. Then press the [CURSOR Ź] key once again so that the 5
th
digit of DISPLAY A
starts to flash.
4) Enter the nominal digit value of the % comparator setting. After a value has been assigned for each digit the decimal point will start to blink. Use the [DP] key to position the decimal point and then press the [CURSOR Ź] key to set the unit prefix of measure. Press the [UNITS] key to select the desired prefix for the DISPLAY A, nominal value setting.
NOTE: Units of measure in DISPLAY A or DISPLAY B cannot be changed once the comparator has been enabled. The [IGNORE] function and the equivalent circuit mode setting are also locked out.
5) After the prefix units for the nominal value have been selected, press the [CURSOR Ź] key. The
th
digit of DISPLAY B will begin to blink indicating that a value can be entered for the DISPLAY B
5 measurement. Enter a nominal value setting for DISPLAY B then press the [CURSOR Ź] key.
6) The 4
th
digit of DISPLAY C will display a blinking “0”. Press the [1] key to enter the High and
Low Percent Limits Mode.
4-37
Page 76
Model 3550 LCR Meter Instruction Manual
Operating Instructions
The Comparator cont’d:
Percent Value Settings cont’d
7) To input an upper limit threshold value, press the [HIGH] key. Press the [LOW] key to set the lower limit threshold value. Then, highlight the first digit of DISPLAY C and press [1].
8) After selecting the threshold value input mode, use the cursor key, [Ź] to highlight the bin LED, and then press [1]~[9] to select the desired bin.
9) After entering the bin number, use the cursor key [Ź] to highlight DISPLAY A. Enter the 5-digit threshold value for DISPLAY A using keys [0]~[9]. (The setting range is from 000.00%~199.99%)
10) Enter the 5-digit threshold value for DISPLAY B using keys [0]~[9]. (The setting range is
000.00%~199.99%)
11) To finalize an upper limit, press [HIGH] or press [LOW] to finalize a low limit setting.
NOTE: Step 7 determines whether a high or low limit %comparator setting is defined be sure that both a low and high limit have been defined. To input settings for other bins, return to Step 1 and repeat the process.
12. To invalidate the settings, for threshold values, simply enter the Threshold Percent Comparator Set Mode and move the cursor to any segment in DISPLAY A or DISPLAY B. Then press [IGNORE] to invalidate the % comparator settings for that DISPLAY.
NOTES:
The comparator can only be set using panels 1~9. 0 cannot be used. This procedure is only for the percent comparator setting. Note that the procedure for the
absolute value comparator is different.
Make sure the comparator mode is set for “Percent Mode” otherwise the procedure will not
work correctly.
After setting the limit values for the comparator, the measurement parameters for DISPLAY
A and DISPLAY B cannot be changed. This means that the measurement parameters must be defined before the comparator settings.
For DISPLAY B, there are no units for D,Q, or LJ.
NOTE: The BIN functionality is expressed as:
Refer to Figures 4.5 and 4.6 for an illustration of BIN operation.
4-38
LOWBINHIGH
Page 77
Model 3550 LCR Meter Instruction Manual Operating Instructions
The Comparator cont’d:
BIN Operation
Comparator operation may be disabled for DISPLAY A or DISPLAY B by pressing the [IGNORE] key while in the high or low limit set mode. A disabled DISPLAY is indicated by “-----“ in place of the limit value being shown on the display.
Sequenced Sorting Configuration
There are two common multi category BIN settings for sorting components. The first is Sequenced Sorting as illustrated in Figure 4-5. In this setting, the 3550 comparator will sort components based on their limit values where each BIN# will define a unique range of measured values. Any number of bins may be used for Sequenced Sorting. If there is no matching BIN found for the component value to fall under, then the component will be placed in BIN 0.
BIN 1
BIN 2
BIN 3
BIN 7
BIN 8
BIN 9
L1
Figure 4.5 Sequenced BIN Sorting
LOWBINHIGH
1
H1
2
L2 H2
L3 H3
3
L7
7
H7
L8 H8
8
L9 H9
9
4-39
Page 78
Model 3550 LCR Meter Instruction Manual Operating Instructions
The Comparator cont’d:
BIN Operation cont’d
Nested Sorting Configuration
Nested sorting is typically used for tolerance-based sorting where an acceptable % or value range is defined for each BIN #. First, the 3550 evaluates the measured value according to the defined range of BIN 1. If the measured falls within the range of BIN 1 then the component is sorted into BIN 1. If the measured value falls outside of the BIN 1 range then it is evaluated for BIN 2. If the component value falls within the BIN 2 range then it is sorted into BIN 2. If not, then the 3550 proceeds the evaluation to BIN 3 and so on. If the measured value does not fall within any defined ranges then the component will be sorted into BIN 0, the reject BIN.
BIN 1
BIN 2
BIN 3
BIN 8
BIN 9
L1
L1
9
8
Figure 4.6 Nested BIN Sorting
LOWBINHIGH
1
H1
2
H1
3
L1
3
L1
L1
2
H1
8
H1
9
H1
4-40
Page 79
Model 3550 LCR Meter Instruction Manual Operating Instructions
The Comparator cont’d:
Outputting the Comparator’s Results
There are two ways to determine the state of the comparator. The LEDs of the Comparator Panel, and the optically isolated, open collector outputs of the Control I/O Connector on the rear of the instrument.
Comparator values may be entered and sorted into BINs 1~9. When a measured value falls within ant of the ranges of BINs 1~9, the TOTAL-GO LED will light. When a measured value falls outside of the defined 1~9 BIN ranges, then either the “LO” or “HI” LEDs on the Comparator Panel will illuminate (depending on whether the measured value is higher or lower than the BIN settings). Measured values outside of the comparator range will be sorted to BIN 0.
For instructions regarding the pin connections of the Control I/O Connector, refer to “Programming and Interfacing”
Audible Buzzer Operation
When the [SHIFT] "TOTAL GO" condition. If it is set as "2", the buzzer will sound with "TOTAL NO-GO" condition. If the buzzer function is set to "0", then the buzzer will be disabled. See [SHIFT] Buzzer Function for more information on setting the comparator-state buzzer function.
ĺ [5]: Comparator Buzzer Function is set to "1", the buzzer will sound with a
ĺ [5]: Comparator
4-41
Page 80
4-42
Page 81
Programming and Interfacing
INSTRUMENT DESCRIPTION
PREPARATION FOR USE
QUICK START INSTRUCTIONS
OPERATING INSTRUCTIONS
PROGRAMMING & INTERFACING
SERVICE INFORMATION
APPENDIX
3550 LCR Meter Instruction Manual 5-1
Page 82
3550 LCR Meter Instruction Manual Programming and Interfacing
Interfacing to the 3550
This section provides detailed information about the 3550 electrical interfaces and their functionality. It will provide all of the necessary information required to integrate the 3550 easily into a working test application. A Control I/O connector is available for optically coupled TTL interface to PLCs and other control hardware. The 3550 has a standard RS232 communications port and an optional GPIB interface for software-based applications.
Control I/O Connector
Located on the rear panel, the Control I/O Connector offers an optically isolated, TTL-level interface for integration into PLC and other hardware controlled applications. These terminals require a positive 5V~24V signal across the 3~8 Pins (these are the anode connections for external trigger, panel enable/disable, and the four panel BCD inputs) and a common potential across pins 28~33. A constant current diode is in series with the 5~24V supply to limit the current flowing through the photo diode. There is no need for a current limiting resistor for the inputs or outputs of the Control I/O Connector.
Inputs
There are 3 groups of inputs available from the Control I/O Connector:
CONTROL PANEL/EXT (Pins 3 & 28) – This input works in conjunction with the front panel
trigger or the rear panel External Trigger Input. When this input is true (current flows through pins 3 & 28) the rear panel External Trigger is enabled. If the Control PANEL/EXT input is false, then the triggering must be performed from the front panel.
External Trigger (Pins 4 & 29) – Used with the CONTROL PANEL/EXT input. The 3550 will
take measurements when the CONTROL PANEL/EXT input AND
External Trigger are true.
Panel BCD Inputs, [Pins 5~8 (Anode) and Pins 30~33 (Cathode)]. - A BCD setting of 0~9 is
required. Settings other than 0~9 will automatically default to panel 9.
The inputs of the Control I/O Connector are optically Isolated. The figure below is a representation of a typical input.
Figure 5.1 – Input Schematic
Fixed Current
Photo
Coupler
Diode
10 mA
3~8
28~33
5-2
Page 83
3550 LCR Meter Instruction Manual Programming and Interfacing
Control I/O Connector cont’d:
Outputs
The 3550 has 5 data output lines available for external monitoring. Each of the data outputs are optically isolated and current limited through the use of a constant current diode as shown in the illustration below:
Figure 5.2 – Output Schematic
Photo
Coupler
+5V
330ƻ
Fixed Current
Diode
1, 26 VCC (EXT)
10mA
13~20, 38~44
2, 27 EMITTER COMMON (EXT)
5-3
Page 84
3550 LCR Meter Instruction Manual Programming and Interfacing
Control I/O Connector cont’d:
Timing Diagrams
Figure 5.3a – Test Busy Operation
MEASURE
START
MEASURE
END
TEST
BUSY
GO/NO-GO
DATA OUT
MEASURE
END
MEASURE
END
GO/NO-GO
Figure 5.3b – Measure End Operation
1ms
Figure 5.3c - GO/NO-GO Operation
5-4
Page 85
3550 LCR Meter Instruction Manual Programming and Interfacing
Control I/O Connector cont’d:
Table 5.1 - Connector Pin Designations
PIN NO.
Signal Name
PIN NO.
Signal Name
1 EXT POWER 26 EXT POWER
2 EMITTER COMMON 27 EMITTER COMMON
3* CONTROL PANEL/EXT (A) 28* CONTROL PANEL/EXT (K)
4* EXTERNAL TRIGGER (A) 29* EXTERNAL TRIGGER (K)
5* PANEL 23 (A) 30* PANEL 23 (K)
6* PANEL 22 (A) 31* PANEL 22 (K)
7* PANEL 21 (A) 32* PANEL 21 (K)
8* PANEL 20 (A) 33* PANEL 20 (K)
9 34
10 35
11 36
12 37
13 DISPLAY A HNG 38 DISPLAY B HNG
14 DISPLAY A LNG 39 DISPLAY B LNG
15 DISPLAY A GO 40 DISPLAY B GO
16 TOTAL GO 41 TOTAL NG
17 BIN 2
18 BIN 2
3
1
42 BIN 2
43 BIN 2
2
0
19 TEST BUSY 44 MEASURE END
20 ERROR 45
21 46
22 47
23 48
24 49
25 50
* : Input Pin (A) and (K) are the Anode(A) and Cathode(K) of the photo coupler LED Output is the Open Collector of the transistor of the photo coupler
5-5
Page 86
3550 LCR Meter Instruction Manual Programming and Interfacing
Control I/O Connector cont’d:
Table 5.2 - Control I/O Function Summary
Item PIN No. Functions
0 1 2 3 4 5 6 7 8 9
5, 30 0 0 0 0 0 0 0 0 1 1
*PANEL
*CONTROL PANEL/EXT
6, 31 0 0 0 0 1 1 1 1 0 0 7, 32 0 0 1 1 0 0 1 1 0 0 8, 33 0 1 0 1 0 1 0 1 0 1
EXT PANEL
3, 28
ON OFF
*EXTERNAL TRIGGER
DISPLAY A
GO/NO-GO
DISPLAY B
GO/NO-GO
TOTAL
GO/NO-GO
BIN No.
TEST BUSY 19
4, 29
14
13
15
39
38
40
16
41
17 42 18 43
OFF
ON (5~15ms)
OFF
Low NG High NG GO
0 1 1
1 0 1
1 1 0
Low NG High NG GO
0 1 1
1 0 1
1 1 0
GO NO-GO
0 1
1 0
0 1 2 3 4 5 6 7 8 9
1 1 1 1 1 1 1 1 0 0 1 1 1 1 0 0 0 0 1 1 1 1 0 0 1 1 0 0 1 1 1 0 1 0 1 0 1 0 1 0
READY (OFF) BUSY (ON)
MEASURE
END
**ERROR
*: Input Pin **: The ERROR signal indicates a GO/NO-GO decision error. When the range is wrong, or the fixed current/voltage setting value has not been reached, "0" is outputted. "1" for the Output Pins = C-E Open, "0" = C-E Short
44
20
5-6
Test End ON (1ms)
ON OFF
0 1
Page 87
3550 LCR Meter Instruction Manual Programming and Interfacing
Control I/O Connector cont’d:
Figure 5.4 - Sample Diagram of External Connections to Sequencer
CONTROL PANEL
3
28
4
29
EXT (A)
CONTROL PANEL EXT (K)
EXTERNAL TRIGGER (A)
EXTERNAL TRIGGER(K)
5~24 VDC
OUTPUT UNIT
INPUT UNIT
EXT
1, 26
POWER
TOTAL GO
16
0
BIN 2
43
EMITTER COMMON
2, 27
25, 50
GND (Inner Circuit)
Note: The diagram above provides examples of typical input and output connections as required for interface to a sequencer. Similar connections apply to the remaining terminals of the I/O Connector.
5-7
Page 88
3550 LCR Meter Instruction Manual Programming and Interfacing
RS-232C
Interface Cable
The interface is based on RS-232C. The 3550 is designed to act as the DTE side, so please use a cross cable when connecting with a personal computer or other controller unit.
Input Format
Remote In-Commands
To activate the RS-232C, send the "REN" command. This will change the 3550 from Panel Mode to Remote Mode.
Entering Program Codes
Program codes will be sent from the controller unit to the 3550 as a string of ASCII characters. To change the panel number, "PAN*" should be sent. (If a Panel Number is not relayed, then the currently displayed panel will be re-set by the command)
e.g.) "PAN0 FA0 R0 M0 S0 E0"
Notice that a space is necessary between each of the commands. If a space does not precede a command, then the subsequent command/s will be ignored.
Setting Limit Values
Below is an example of how the high and low limits of a BIN are set for DISPLAY A and DISPLAY B.
e.g.) "BIN1 CAL+1.0000e-06 CAH+1.5000E-06" "BIN1 CBL+0.0100e00 CBL+0.1000e00"
The order of LOW and HIGH values may be reversed. Programming LOW or HIGH alone is also permitted.
e.g.) "BIN1 CAH+1.5000e-06 CAH+1.0000e-06" "BIN1 CAL+1.0000e00"
To invalidate a specific BIN’s high or low limit value replace the polarity and numerical data with a series of nine dashes as expressed below:
e.g.) "BIN1 CAL---------"
This command will invalidate the comparator’s DISPLAY A, low limit value.
5-8
Page 89
3550 LCR Meter Instruction Manual Programming and Interfacing
RS-232C
Input Format cont’d
Setting the Measurement Voltage
The voltage setting range is from 0.01V through 5.00V, in 0.01V increments. However, for frequencies between 1.00MHz~5.00MHz, the range is limited to 0.01V~1.00V.
Units: Volts Resolution: 3 digits
e.g.) "LEV1.00E0" (Set the Measurement Voltage to 1.00V) "LEV1.52E0" (Set the Measurement Voltage to 1.52V)
Values, which are less than the 0.01V steps, will be automatically truncated.
e.g.) "LEV1.248E00" (Set the Measurement Voltage to 1.24V)
The 3550 will automatically reassign voltage levels if the absolute maximum levels are exceeded. As shown above, any voltage defined with less than 0.01V resolution will automatically be truncated to 0.01V resolution. Voltage levels exceeding 5.00V will automatically become 5.00V. If the test frequency is greater than 1.00MHz, voltages defined that exceed 1.00V will become 1.00V.
Setting Measurement Currents
For frequencies from 42.0Hz to 1.00MHz, the test current setting ranges from
0.01mA~99.99mA in 0.01mA steps. If the test frequency is between 1.01MHz~5.00MHz, the current setting are limited to 0.01mA~20.00mA.
Unit: mA Significant Digits: 3 digits
e.g.) "LEI1.00E0" (Set the Measurement Current to 1.00mA) "LEI15.20E0" (Set the Measurement Current to 15.20mA)
Like the voltage settings, any values less than the 0.01mA resolution will be automatically truncated.
e.g.) "LEV1.549E00" (Set the Measurement Current to 1.54mA)
The 3550 will automatically reassign test current levels if the absolute minimum or maximum levels are exceeded. As shown above, any current level defined with less than
0.01mA resolution will automatically be truncated to a value with 0.01mA resolution. Defined current levels exceeding 99.99mA will automatically become 99.99mA. If the test frequency is greater than 1.00MHz, test current will be limited to 20.00mA.
5-9
Page 90
3550 LCR Meter Instruction Manual Programming and Interfacing
RS-232C
Input Format cont’d
Setting the Measurement Frequency
The setting range for Measurement Frequency is between 42.0Hz~5.00MHz.
Unit: Hz Significant Digits: 3 digits.
e.g.) "FRE42.0E0" ~ “FRE99.9E0" 42.0Hz ~ 99.9Hz "FRE100E0" ~ "FRE999E0" 100Hz ~ 999Hz "FRE1.00E3" ~ "FRE9.99E3" 1.00kHz ~ 9.99kHz "FRE10.0E3" ~ "FRE99.9E3" 10.0kHz ~ 99.9kHz "FRE100E3" ~ "FRE999E3" 100kHz ~ 999kHz "FRE1.00E6" ~ "FRE5.00E6" 1.00MHz ~ 5.00MHz
Because the 3550 uses exponential number formatting for the test frequency, variations of the above format may be used. To set 100kHz, the following methods are also valid:
e.g.) "FRE1.00E5" OR- "FRE10.0E4"
(Set the test frequency to 100kHz)
Setting the Number of Averages per Reading
To change the number of measurements to be averaged use the "V*" command. * can be replaced by values between 1~100
e.g.) "V55" (Set the number of averages to 55 samples)
Errors in defining the number of samples to be averaged for a reading are addressed as follows:
Settings outside the range of 1~100, are considered program errors.
e.g.) The values 0 or 101 would be considered invalid values for the average parameter.
5-10
Page 91
3550 LCR Meter Instruction Manual Programming and Interfacing
RS-232C
Table 5.3a – Command Summary
Function
Panel
DISPLAY A Measurement
Function
DISPLAY B Measurement
Function
Measurement Range
Program
Code
*PAN0 ~ PAN9
FA0 *FA1 FA2 FA3 FA4
*FB0 FB1 FB2 FB3 FB4 FB5 FB6 FB7
*R0 R1 R2 R3 R4 R5 R6 R7 R8 R9
Content Remarks
Panel Number
DISPLAY A Measurement Item Change
DISPLAY B Measurement Item Change
Auto Range Settings Range1 Range2 Range3 Range4 Range5 Range6 Range7 Range8 Range9
Measurement Item "L" Measurement Item "C" Measurement Item "R" Measurement Item "|Z|" Measurement Item "|Y|"
Measurement Item "D" Measurement Item "Q" Measurement Item "R Measurement Item "R Measurement Item "G" Measurement Item "X" Measurement Item "B" Measurement Item "LJ"
AUTO RANGE 100mƻRange 1ƻRange 10ƻRange 100ƻRange 1kƻRange 10kƻRange 100kƻRange 1MƻRange 10MƻRange100MƻRange
"
S
"
P
Measurement Mode
Measurement Frequency FRE Measurement Frequency Settings "42.0E0"~"5.00E6"
Measurement Mode
Change
Measurement Signal
Voltage LEV
Measurement Signal
Current LEI
*M0 M1 M2
*JV JCV JCC
Auto Mode Settings Series Mode Settings Parallel Mode Setting
Voltage Measurement Fixed Voltage Measurement Fixed Current Measurement
Measurement Signal Voltage Settings
Fixed Current Mode Current Settings
AUTO MODE SER MODE PRL MODE
Open Terminal Voltage Measurement Mode Fixed Voltage Measurement Mode
Fixed Current Measurement Mode
"0.01E0"~"5.00E0"
"0.01"~"99.99"
5-11
Page 92
3550 LCR Meter Instruction Manual Programming and Interfacing
RS-232C
Table 5.3b – Command Summary cont’d
Function
Service Request
Measurement Start-Up
Trigger
Trigger T0 Trigger Command Only Valid for "E3"
Initial Settings
Zero Offset
Command Lock
Measurement
Conditions Monitor MON
Measurement Data
Averaging V Averaging Times
Data Output REN Measurement Data Output
Correction Frequency
Limit
Program
Code
*S0 S1
*E0 E1 E2 E3
I0 I1
Z0 Offset Clear Clear Correction Value
Z1 ZERO OPEN
Z2 ZERO SHORT
Z3 Spot OPEN
Z4 Spot SHORT
Z5 Spot Correction Clear Correction Value Spot Clear *L0 L1
WCLI0
WCLI1 Effectively Low Frequency
WCHI0 Invalid High Frequency limit
WCHI1
WCL
OFF ON
Internal Trigger External Trigger TRIGGER EXT "T0"Trigger
Clear Panel Clear All Panels
Lock OFF Lock ON
Measurement Conditions Monitor
Invalid Low Frequency limit
limit
Effectively Low Frequency limit
Low Frequency limit setting
Content
Remarks
SRQ Settings for GPIB
Address Command GET Start Start by "MAN" Listener-Address Designation Start
Measurement Terminal OPEN -ZERO Correction Measurement Terminal SHORT-ZERO Correction OPEN-ZERO Spot Correction SHORT-ZERO Spot Correction
Set 1~100
Set so that data output will be through GPIB/RS­232C
"WCL42.0E0" ~ "WCL5.00E6"
WCH High Frequency limit setting "WCH42.0E0" ~ "WCH5.00E6"
O0 Output OFF
Data Output Destination
Selection
O1 RS232C Only
O2 GPIB Only
O3 RS232C/GPIB
5-12
GPIB only valid with Installed Option
Page 93
3550 LCR Meter Instruction Manual Programming and Interfacing
RS-232C
Table 5.3c – Command Summary cont’d
Function
Soft Version VER
Spot Correction Input Number
Spot Correction Frequency Limit
Program
SPX
SPFXXX Set Spot Frequency “SPF42.0E0” ~ “SPF5.00E6”
Code
Content
Information on the Software Version
Spot Correction Select from 1 to 3
“SP1”
Remarks
*1) * indicates initial setting *2) If a non-program code is entered, it will be considered invalid, and no setting will be made. In this case, the previous setting will be unchanged.
NOTE 1: After receiving “L1”, only “L0”, “T0” and GET can be received. NOTE 2: As soon as “MON” is sent, make sure that the system is ready to receive. NOTE 3: Settings will vary according to the measurement range and measurement frequency. Please refer to the range charts. NOTE 4: After the termination of zero offset, the following indications are shown on the computer. “Correct OK” - Zero correction has executed properly “Correct Error Impedance Under” – The open circuit impedance is less than 1kƻ, which is beyond correction limits. “Correct Error Impedance Over” The short circuit impedance has exceeded 1kƻ, which is beyond correction limits.
5-13
Page 94
3550 LCR Meter Instruction Manual Programming and Interfacing
RS-232C
Table 5.4 - Comparator Instructions
Function Program Code Content Remarks
Comparator
Comparator Mode
Bin Number BIN Sorting Category BIN No.(19)
Absolute Value Setting Limit Values
% Setting Limit Values
*1) * indicates a initial setting
NOTE 1: As soon as "LIM" is sent, make sure that the system is ready to receive signals.
*C0 Comparator OFF
C1 Comparator ON
CP0 Absolute Value Setting Absolute Value Setting Comparator
CP1 % Setting % Setting Comparator
CALXXXX CAHXXXX CBLXXXX CBHXXXX
CAPXXXX CBPXXXX PALXXXX PBLXXXX PAHXXXX PBHXXXX
DISPLAY A LOW DISPLAY A HIGH DISPLAY B LOW DISPLAY B HIGH
DISPLAY A Std. Value DISPLAY B Std. Value DISPLAY A Min. % DISPLAY B Min. % DISPLAY A Max. % DISPLAY B Max. %
XXXX: ±0.0000e00 ~ ±1.9999e08 Disabled Setting = "---------"
XXXX: ±0.0000e00 ~ ±1.9999e08XXXX: ±0.0000e00 ~ ±2.000e02 Disabled Setting = "---------"
5-14
Page 95
3550 LCR Meter Instruction Manual Programming and Interfacing
RS-232C
Data Output Format for Measurement Settings
Output Format Sample
0000 _ P1 _ 1.00 E03 Hz _ OV _ +1.0795 E00 V _ +1.0771 E-03 A,
N R P _+1.0022 E03 R, GO, N _ A0 _ -1.1874 E03 R, GO, BIN1, GO
1 2 3 4 5 6 7 8 9 10 11 12
13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28
1 Indexing Number 15 Circuit Mode
There is no circuit mode on Z and Y, alternatively, "-" is outputted.
2 Panel Number 16 Data Temp. Value (A) 3 Measurement Frequency Temp. Value 17 Data Exponent (A) 4 Measurement Frequency Exponent 18 Data Unit (A) 5 Measurement Frequency Unit 19 Comparator Result (A) 6 Measurement Mode 20 Status (B) 7 Test Piece Voltage Temp. Value 21 Function (B) 8 Test Piece Voltage Exponent 22 Data Temp. Value (B)
9 Test Piece Voltage Unit 23 Data Exponent (B) 10 Test Piece Current Temp. Value 24 Data Unit (B) 11 Test Piece Current Exponent 25 Comparator Result (B) 12 Test Piece Current Unit 26 Bin Number 13 Status (A) 27 Comparator Result (TOTAL) 14 Function (A) 28 Delimiter
NOTES:
For the Comparator Result, if the Comparator is OFF, 2 "_" symbols (ASCII Code 5Fh)
will be outputted instead of a result.
For the Bin No., if the Comparator is OFF, 4 "_" symbols (ASCII Code 5Fh) will be
outputted.
In the Format Sample, the "_” denotes a "space" (ASCII code 20h).
5-15
Page 96
3550 LCR Meter Instruction Manual Programming and Interfacing
RS-232C
Table 5.5a - Data Output Code Functions
Number Output Code Content
1
2
3
4
5
6
7 8
9
10 11
12
13
14
15
16
17
**** Index Number (0001~9999)
P* Panel Number (0~9)
**.* *.** ***. E+** Measurement Frequency Exponent
Hz Measurement Frequency Unit
OV Open Voltage Mode CV Fixed Voltage Mode CC Fixed Current Mode
+*.****E±**
V Test Piece Voltage Unit
+*.****E-**
A Test Piece Current Unit O Over Range
N Measurement Data Normal U Under Range E Error L Inductance Measurement (L) C Capacity Measurement (C) R Resistance Measurement (R)
Z Impedance Measurement (|Z|) Y Admittance Measurement (|Y|) S P +1.9999 Over Range, Measurement Data Abnormal ±*.**** Under Range, Measurement Data Normal Measurement Data Exponent E-12 p (10
E-09 n (10-9) E-06 ư (10-6)
E-03 m (10-3) E+00 (100) E+03 k (103) E+06 M (106)
Measurement Frequency Temp. Value
e.g.) 42.0Hz ĺ "42.0": 100Hz ĺ "100.": 1.00kHz ĺ "1.00"
e.g.) "E+00" ĺ Hz: "E+03" ĺ kHz: "E+06" ĺ MHz
Test Piece Voltage Temp. Value Exponent e.g.) 1.00V ĺ "1.0000E+00"
0.10V ĺ "100.00E-03"
Test Piece Voltage Temp. Value Exponent e.g.) 10.00mA ĺ "10.000E-03"
0.01mA ĺ"10.000E-06"
Series Mode Measurement Parallel Mode Measurement
-12
)
5-16
Page 97
3550 LCR Meter Instruction Manual Programming and Interfacing
RS-232C
Table 5.5b - Data Output Code Functions cont’d:
Number Output Code Content
DISPLAY A Display Unit
18
19
20
21
22
23
24
25
H For Inductance Measurement F For Capacity Measurement R For Resistance, Impedance Measurement S For Admittance Measurement LO LOW NO-GO
GO GO HI HIGH NO-GO __ Comparator OFF (ASCII Code 5fh, outputted 2 times)
O N
D0 Loss Coefficient Measurement (D) Q0 Quality Measurement (Q) RS Equivalent Series Resistance Measurement(RS) RP Equivalent Parallel Resistance Measurement (RP) G0 Conductance Measurement (G) X0 Reactance Measurement (X) B0 Susceptance Measurement (B) A0 Phase Angle Measurement (LJ) +1.9999 Over Range, Measurement Data Abnormal
±*.**** Under Range, Measurement Data Normal Measurement Data Exponent E-09 n (10-9) E-06 µ (10-6) E-03 m (10-3)
E+00 (100) E+03 k (103)
E+06 M (106) R For Equivalent Parallel + Series Resistance Measurement, and also Reactance
S For Conductance and Susceptance Measurement _ For Loss Coefficient, Quality Measurement, Phase Angle Measurement LO LOW NO-GO GO GO HI HIGH NO-GO __ Comparator OFF (ASCII Code 5Fh, outputted 2 times)
Over Range Measurement Data Normal
Measurement
26
27
28
BIN* BIN Number Decision Result ( *:09)
GO TOTAL GO NG TOTAL NO-GO
__ Comparator OFF (ASCII Code 5Fh, 2times)
CR+L
F
If Comparator is OFF, then "____"(ASCII Code 5Fh, 4times)
Delimiter. For RS-232C, only CR+L
F
5-17
Page 98
3550 LCR Meter Instruction Manual Programming and Interfacing
RS-232C
Monitor Output Format
Output in Relation to Monitor Command "MON"
For all 10 panels, (0~9) the PANEL No., DISPLAY A Function, DISPLAY B Function, CIRCUIT MODE, RANGE, Measurement Signal Frequency and Measurement Signal Voltage will be outputted and divided by spaces. At the end, whether or not there is a Limit Value will be indicated as well.
Output Format Sample:
P0 _ L _ D0 _ R0 _ A _ 1.00E+03 _ OV _ 1.0000E+00V _ V1 _ ­P1 _ L _ D0 _ R0 _ S _ 1.00E+03 _ OV _ 1.0000E+00V _ V1 _ ­P2 _ L _ Q0 _ R1 _ S _ 1.00E+03 _ OV _ 1.0000E+00V _ V1 _ ­P3 _ L _ Q0 _ R2 _ S _ 10.0E+03 _ OV _ 1.0000E+00V _ V1 _ * P4 _ L _ D0 _ R3 _ S _ 10.0E+03 _ OV _ 1.0000E+00V _ V1 _ * P5 _ L _ D0 _ R4 _ S _ 10.0E+03 _ CV _ 5.0000E-01V _ V1 _ ­P6 _ L _ D0 _ R5 _ S _ 100.E+03 _ OV _ 1.0000E+00V _ V1 _ ­P7 _ L _ D0 _ R6 _ S _ 100.E+03 _ CC _ 1.0000E+00V _ V1 _ ­P8 _ L _ D0 _ R7 _ S _ 5.00E+06 _ OV _ 1.0000E+00V _ V1 _ ­P9 _ L _ D0 _ R7 _ S _ 2.21E+06 _ OV _ 1.0000E+00V _ V1 _ -
1 2 3 4 5 6 7 8 9 10 11 12 13
1 - Panel Number 2 - DISPLAY A Function 3 - DISPLAY B Function 4 - Range Number 5 - Measurement Circuit Mode 6 - Measurement Signal Frequency Temp. Value 7 - Measurement Signal Frequency Exponent 8 - Measurement Mode 9 - Measurement Signal Voltage Temp. Value 10 - Measurement Signal Voltage Exponent 11 - Measurement Data Averaging 12 - Comparator (*:ON -:OFF) 13 - Delimiter
NOTE: In the Output Format Sample, the "_" denotes a space (ASCII Code 20h).
5-18
Page 99
3550 LCR Meter Instruction Manual Programming and Interfacing
RS-232C
Monitor Output Format
Output in Relation to Limit Monitor Command "LIM"
Absolute Value Settings Mode This will output the PANEL No., BIN No., LOW LIMIT, and HIGH LIMIT for BIN No.1No.9 in the
order of DISPLAY A and then DISPLAY B. However, if "LIM" is sent when the Panel Number is"0", the system will announce an error.
Output Format Sample
P1_BIN1_CP0_+9.9000E-03H/+10.100E-03H_+1.0000E+00R/+10.000E+00R P1_BIN2_CP0_+9.8000E-03H/+10.200E-03H_+10.000E+00R/+100.00E+00R P1_BIN3_CP0_+9.7000E-03H/+10.300E-03H_+100.00E+00R/+1.0000E+03R P1_BIN4_CP0_------------/------------_------------/-----------­P1_BIN5_CP0_------------/------------_------------/-----------­P1_BIN6_CP0_------------/------------_------------/-----------­P1_BIN7_CP0_------------/------------_------------/-----------­P1_BIN8_CP0_------------/------------_------------/-----------­P1_BIN9_CP0_------------/------------_------------/------------
1 14 18 15 16 17 15 16 17 13
1 - Panel Number 13 - Delimiter 14 - BIN Number 15 - Low Limit Setting Value 16 - Divider for High/Low Settings (2Fh) 17 - High Limit Setting Value 18 - Comparator Mode
NOTES:
Places with no setting will be filled with "------------" (ASCII Code 2Dh×12) D, Q, and LJ will have a "_"(ASCII Code 5Fh) for their "Unit" In Output Format Sample, the "_" denotes a space (ASCII Code 20h)
5-19
Page 100
3550 LCR Meter Instruction Manual Programming and Interfacing
RS-232C
Monitor Output Format
Output in Relation to Limit Monitor Command "LIM"
% Setting Mode
Output Format Sample
P1_BIN1_CP1_+10.000E-03H_005.00%/006.00%_+1.0000E00R_010.00%/010.00% P1_BIN2_CP1_+10.000E-03H_010.00%/010.00%_+1.0000E00R_020.00%/020.00% P1_BIN3_CP1_+10.000E-03H_020.00%/020.00%_+1.0000E00R_030.00%/030.00% P1_BIN4_CP1_+0.0000E-00H_-------/-------_+0.0000E00R_-------/------­P1_BIN5_CP1_+0.0000E-00H_-------/-------_+0.0000E00R_-------/------­P1_BIN6_CP1_+0.0000E-00H_-------/-------_+0.0000E00R_-------/------­P1_BIN7_CP1_+0.0000E-00H_-------/-------_+0.0000E00R_-------/------­P1_BIN8_CP1_+0.0000E-00H_-------/-------_+0.0000E00R_-------/------­P1_BIN9_CP1_+0.0000E-00H_-------/-------_+0.0000E00R_-------/-------
1 14 18 19 20 16 21 19 20 16 21 13
1 - Panel Number 13 - Delimiter 14 - BIN Number 16 - Divider for High and Low Settings (2Fh) 18 - Comparator Mode 19 - Standard (Std.) Value 20 - Low Limit Value Setting 21 - High Limit Value Setting
NOTES:
Places with no setting will be filled with "-------" (ASCII Code 2Dh×7) D, Q, and LJ will have a "_"(ASCII Code 5Fh) for their "Unit" In Output Format Sample, the "_" denotes a space (ASCII Code 20h)
5-20
Loading...