Tegam 3525 Instruction Manual

MODEL 3525
GENERAL-PURPOSE, PROGRAMMABLE LCR METER
Instruction Manual
PN# 3525-900-01CD
Publication Date: February 2009
REV. E
Table of Contents
TEGAM INC.
MODEL 3525
GENERAL-PURPOSE, PROGRAMMABLE
LCR METER
MODEL 3525
Instruction Manual
PN# 3525-900-01CD
Publication Date: February 2009
REV. E
NOTE: This user’s manual was as current as possible when this product was manufactured. However, products are constantly being updated and improved. Because of this, some differences may occur between the description in this manual and the product received.
Model 3525 LCR Meter Instruction Manual
a
Table of Contents
I INSTRUMENT DESCRIPTION............................................................................1-1
Instrument Description ........................................................................... 1-2
Feature Overview................................................................................... 1-2
Accessories Sheet .................................................................................. 1-4
Performance Specifications...................................................................... 1-5
Measurement Ranges......................................................................... 1-5
Table 1.1 - Absolute Impedance Range Limits Chart...............................1-5
Display A & B – Basic Accuracies ......................................................... 1-6
Table 1.2 – Basic Accuracy for Display A............................................... 1-6
Table 1.3 – Basic Accuracy for Display B............................................... 1-6
Calculating Actual Measurement Accuracies........................................... 1-7
Table 1.4 – Speed Coefficient.............................................................. 1-7
Table 1.5 – Amplitude Coefficient ........................................................ 1-7
Instrument Specifications........................................................................ 1-8
II PREPARATION FOR USE ..................................................................................2-1
Unpacking & Inspection........................................................................... 2-2
Safety Information & Precautions .............................................................2-2
Terms in this Manual ......................................................................... 2-2
Terms as Marked on Equipment........................................................... 2-2
Symbols........................................................................................... 2-3
Grounding the Equipment...................................................................2-3
Danger Arising from the Loss of Ground ............................................... 2-3
Use the Proper Fuse........................................................................... 2-4
Do not Use in Explosive Environments..................................................2-4
Do not Operate without Covers ........................................................... 2-4
Servicing Safety Summary ...................................................................... 2-4
Do not Service Alone ......................................................................... 2-4
Use Care when Servicing with Power On............................................... 2-4
Power Source ................................................................................... 2-4
Line Voltage Selection.............................................................................2-5
Table 2.1 Operating Voltage Ranges.................................................. 2-5
III QUICK START INSTRUCTIONS.........................................................................3-1
Power the Unit.......................................................................................3-2
Factory Default Settings..........................................................................3-2
Table 3.1 - Factory Default Settings..................................................... 3-2
Instrument Setup...................................................................................3-3
Display A ......................................................................................... 3-3
Display B ......................................................................................... 3-3
Measurement Speed .......................................................................... 3-3
Test Frequency ................................................................................. 3-3
Test Voltage Level............................................................................. 3-3
Circuit Mode ..................................................................................... 3-3
Range.............................................................................................. 3-3
Trigger ............................................................................................ 3-3
Performing an Open and Short Circuit Null................................................. 3-4
Measurement Tips.................................................................................. 3-4
Model 3525 LCR Meter Instruction Manual
b
Table of Contents
IV OPERATING INSTRUCTIONS............................................................................4-1
Basic Operation...................................................................................... 4-2
Default Parameters................................................................................. 4-2
Front Panel Description ........................................................................... 4-3
Figure 4.1a Front Panel Layout ......................................................... 4-3
Figure 4.1b Measurement & Parameter Control Area............................ 4-4
Figure 4.1c Comparator Controls Area ............................................... 4-5
Rear Panel Description............................................................................ 4-6
Figure 4.2 Rear Panel...................................................................... 4-6
Selecting the Measurement Parameters..................................................... 4-7
Table 4.1a – Display A Description of Measurement Functions.................. 4-7
Table 4.1b – Display B Description of Measurement Functions..................4-7
Setting the Measurement Speed...............................................................4-8
Table 4.2 – 3525 Typical Measurement Times........................................4-8
Setting the Measurement Frequency ......................................................... 4-9
Setting the Measurement Signal Level.......................................................4-9
Setting the Equivalent Circuits Mode......................................................... 4-10
Setting the measurement Range .............................................................. 4-11
Table 4.3 Measurement Range Characteristics .................................... 4-11
Setting the Trigger Mode.........................................................................4-13
Using the Comparator Functions...............................................................4-14
Overview ......................................................................................... 4-14
Table 4.4 – Operation of GO/NO-GO Comparator ................................... 4-14
Setting the High and Low Limit Values.................................................. 4-15
Preset Panel Function..............................................................................4-18
Overview ......................................................................................... 4-18
Setting Panel Numbers....................................................................... 4-18
Recalling Panel Numbers .................................................................... 4-18
Zero Correction...................................................................................... 4-19
Overview ......................................................................................... 4-19
Figure 4.3 – Terminal Connections for Zero Correction............................ 4-19
Open Circuit Correction......................................................................4-20
Short Circuit Correction......................................................................4-20
Audible Beeper ...................................................................................... 4-21
Overview ......................................................................................... 4-21
Setting the Beep Sound...................................................................... 4-21
Connection Methods ............................................................................... 4-22
Figure 4.4 – 3 Terminal Connection...................................................... 4-22
Figure 4.5 – 5 Terminal Connection...................................................... 4-23
V PROGRAMMING & INTERFACING.....................................................................5-1
Interfacing to the 3525 ........................................................................... 5-2
Front Panel Connections..........................................................................5-2
Figure 5.1 – Front Panel Measurement Connections................................ 5-2
Rear Panel ............................................................................................ 5-3
Figure 5.2 – Rear Panel Interfaces....................................................... 5-3
Model 3525 LCR Meter Instruction Manual
c
Table of Contents
V PROGRAMMING & INTERFACING cont’d
I/O Control Connector.............................................................................5-4
Compatible Connectors ...................................................................... 5-4
I/O Pin Outputs.................................................................................5-4
Figure 5.3 – I/O Control Connector Pin Orientation................................. 5-4
Table 5.1 Pin Designations............................................................... 5-5
I/O Pin Descriptions...........................................................................5-5
Table 5.2 BCD Panel Settings ........................................................... 5-6
Table 5.3 – Total Go & No Go Functional Relationships............................ 5-7
I/O Control Connector Electrical Schematic ........................................... 5-8
Figure 5.4 – I/O Control Connector Schematic....................................... 5-8
I/O Timing Diagrams .............................................................................. 5-9
Figure 5.5 – I/O Control Connector Timing Diagram ............................... 5-9
Approximate Measurement Times............................................................. 5-9
Table 5.4 – Approximate Measurement and Signal Processing Times......... 5-9
Analog Signal Measurement Time ........................................................ 5-10
Table 5.5a – Analog Signal Measurement Time...................................... 5-10
Measurement Speed .......................................................................... 5-10
Table 5.5b Total Measurement Speed ................................................ 5-10
RS232C Communications ........................................................................ 5-11
Figure 5.6 – RS-232C Connector ......................................................... 5-11
Table 5.6 – RS-232C DTE Pin Configurations......................................... 5-11
RS-232C Settings.............................................................................. 5-12
Remote Commands ................................................................................ 5-13
Program Messages ............................................................................ 5-13
The Header ...................................................................................... 5-13
The Parameter.................................................................................. 5-13
Rules for Remote Commands .............................................................. 5-14
Rules for Commands and Parameters................................................... 5-14
Condition & State Registers ................................................................ 5-15
Commands List.................................................................................5-18
Table 5.7 RS-232C & GPIB Command Set .......................................... 5-18
Command Reference.......................................................................... 5-19
VI SERVICE INFORMATION..................................................................................6-1
Warranty .............................................................................................. 6-2
Warranty Limitations .............................................................................. 6-2
Statement of Calibration ......................................................................... 6-2
Contact Information ............................................................................... 6-2
Repair Parts .......................................................................................... 6-3
Changing the Power Fuse ........................................................................ 6-3
Preparation for Calibration or Repair Service .............................................. 6-4
Expedite Repair & Calibration Form...................................................... 6-5
Model 3525 Verification Procedure............................................................6-6
VI APPENDIX .......................................................................................................A-1
Model 3525 LCR Meter Instruction Manual
d
Instrument Description
INSTRUMENT DESCRIPTION
Model 3525 LCR Meter Instruction Manual
PREPARATION FOR USE
QUICK START INSTRUCTIONS
OPERATING INSTRUCTIONS
PROGRAMMING & INTERFACING
SERVICE INFORMATION
APPENDIX
1-1
Model 3525 LCR Meter Instruction Manual Instrument Description
Instrument Description
Model 3525,TEGAM’s innovative solution for cost sensitive LCR applications, does not compromise quality or performance. Its flexible design allows a diverse range of testing applications including manual or automated testing of capacitors, inductors, coils, resistors, thermoelectric cooling devices, materials, piezo-electric or other sensors or components possessing AC impedance characteristics.
Up to 9 impedance parameters are easily viewed on each of the 3525’s two 4 ½ digit, LED displays. Display A provides accurate and repeatable readings of Inductance (L), Capacitance (C), Resistance (R), and Impedance (|Z|). Display B measures values for Dissipation Factor (D or Tan δ), Quality Factor (Q), Phase Angle (θ), Measurement Voltage (V) and Measurement Current (I). All parameters are easily monitored, selectable from the front panel, and may be measured as Series or Parallel equivalent circuits.
The 3525’s compact size is unprecedented and allows side-by-side mounting in standard 19” racks. A built-in comparator function makes the 3525 the ideal choice for manual component checking. Whether you are verifying component values for QA or manufacturing, the built in GO­NO GO comparator reduces test time and increases productivity with an audible beeper that may be set to alarm when a pass or fail state has been measured. A front panel lock feature prevents accidental changes on instruments settings.
Model 3525 is shipped from the factory with standard Kelvin Klips, RS-232C port, and I/O connector. There are a variety of optional accessories available, including GPIB Interface, BCD Output, Kelvin Klip Leads, Kelvin Tweezers, Radial Sorting Fixture, and SMT Chip Test Fixture. See the Accessories Sheet on page 1-4 for complete descriptions.
Feature Overview
The Model 3525 was designed with the right balance of features to keep cost down while providing optimal performance for most basic LCR applications. Below is a summary of the instrument’s main features.
0.08% Basic Accuracy with 4.5 digit resolution.
The 3525 is designed to perform basic impedance measurements with up to 0.08% basic accuracy. Measurement accuracies are dependent upon a number of parameters and test conditions. See Performance Specifications to determine the actual measurement accuracies.
High Speed Measurement
With a 1kHz test frequency a reading can be obtained in as little as 15 mS when the instrument is operating in fast mode.
Manual or Remote Operation The 3525 is equipped for manual or remote operation. Standard interfaces include a RS­232C and I/O port. GPIB and BCD interfaces are available as optional accessories.
1-2
Model 3525 LCR Meter Instruction Manual Instrument Description
Feature Overview cont’d:
Dual High Intensity Displays
The 3525 is equipped with dual LED displays for easy viewing. Display A measures 4½ digits of Inductance (L), Capacitance (C), Resistance (R), or Impedance (|Z|) while Display
B indicates Dissipation Factor (D or Tan δ), Quality Factor (Q), Phase Angle (θ), Measurement Voltage (V), or Measurement Current (I).
Programmable GO/NO-GO Comparator with Audible Buzzer
Each unit has an internal comparator function that can be set by the user. Comparator Outputs result from the values of Display A, Display B, or a combination of the two. The comparator outputs may be read from the front panel, buzzer, the I/O connector located on the rear panel, or queried from the remote interfaces.
99 Preset Storage Locations
Panel Settings 01-99 are user-definable and may be programmed manually or via communication interfaces. These settings are stored in non-volatile RAM locations and may be recalled from the front panel or any of the rear interfaces.
Less than ½ Rack Width
Two 3525 units are able to be rack mounted, side-by-side, in a standard 19” rack using a third party rack mount kit, without modifying the instrument cases.
Manual or Auto Ranging
In AUTO range, the instrument automatically selects from 10 different impedance ranges. The selected range is dependent upon the DUT’s absolute impedance, |Z|. In the manual range mode, the user may select from any of these ten ranges through the front panel.
Voltage and Current Monitors
Test Voltage (Vx) and Current (Ix) can be measured at the DUT terminals and are viewable on Display B to three significant digits.
Open and Short Circuit Zero Correction
Parasitic impedance of the test cables and/or test fixture may be cancelled out by performing the 3525’s auto zero functions.
Front Panel Lock The 3525 is equipped with a front panel lock-out feature to prevent accidental changes to instrument settings.
Dual Trigger Modes Trigger instructions may be executed either internally or externally. Internal triggering allows the instrument to run continuously while external triggering requires instruction from the front panel, I/O connector, RS-232C, or GPIB interfaces.
Series or Parallel Equivalent Measurements AC parameters may be measured as Series or Parallel equivalents.
1-3
Model 3525 LCR Meter Instruction Manual Instrument Description
3525 Accessories Sheet
2005B - Chip Tweezers (5ft)
This four-terminal tweezer set makes solid connections to chip components in manual sorting applications. Capacity of jaws is
12.7mm (0.5 in.). The 2005B Chip Component Tweezer Set includes a
1.5m (5 ft) cable for connection to the 3525. Contact tips are replaceable. P/N 47422.
47422 Chip Tweezer Replacement Kit
Tweezer tips are intended to last 100,000 to 500,000 operations. An optional tip replacement kit includes 12 replacement tips, 2 screws and 1 wrench.
3511 - Chip Test Fixture
Available for performing three terminal measurements on surface mount devices. Connects directly to the front panel of the 3525. Use the 3511 for medium and high impedance measurements.
3525-900-01CD – Instruction Manual
Printed copy of the 3525 users manual.
1-4
3510-Radial Lead Adapter
This sorting fixture allows for efficient four-wire measurement of leaded parts. The test fixture features spring action contacts for easy insertion and removal of test components.
47454 – Kelvin Klips
Kelvin Klips allow solid four-terminal, Kelvin connections to leaded components. The jaws are assembled with hardened gold-plated, beryllium copper, which ensures low contact resistance, low thermal emf to copper, high corrosion resistance, and long service life. An alligator clip is provided allowing connection of a passive guard. The assembly includes a 5 ft (1.5m) cable for connection to the 3525.
NOTE: Under certain measurement conditions, Kelvin Klips can cause a loss of measurement accuracy. Fixtures 3510, 3511, or 2005B chip tweezers are recommended for the following component values:
C<100pF; L < 100μH; R > 1MΩ
CBL-3102-BNC-BNC Cable
BNC to BNC, RG 58/U, coaxial cable for use at the trigger inputs and outputs of the 3525. Cable length is a minimum of 1 meter (3.28 ft) long.
KK100- Kelvin Klip™ Rebuild Kit
Kelvin Klip™ replacements for construction or repair of Kelvin Klip leads.
GPIB (IEEE-488) Cables
1583-3 – 1-meter GPIB buss cable 1583-6 – 2-meter GPIB buss cable 1583-9 – 3-meter GPIB buss cable
3502 – BCD Interface
BCD output for interface to handlers and other automated systems.
3501 – GPIB Interface
GPIB (IEEE-488)-compatible interface card. Field or factory installation.
740565 – CABLE NULL
740565-6 – 6’ cable, null mod, 9-9 740565-10 – 10’ cable, null mod, 9-9
Model 3525 LCR Meter Instruction Manual Instrument Description
Performance Specifications
The advertised specifications of the Model 3525 are valid under the following conditions:
1. The instrument’s calibration must be verified using the methods and intervals as described
in the Service Information Section of this user’s manual.
2. The instrument must be in an environment which does not exceed the limitations as
defined under “Environmental” in the Instrument Specifications section.
3. The unit is allowed to warm up for a period of at least 30 minutes before measurements
are taken. A warm-up period of 60 minutes is recommended after exposure to or storage in a high humidity, non-condensing environment.
4. Use only TEGAM-manufactured test leads and sorting fixtures with this device. Modification
of existing test cables or fixtures may compromise the accuracy of your readings. Always perform an open and short zero null procedure before taking measurements.
Measurement Ranges
Table 1.1: Absolute Impedance Range Limits Chart
Range No. |Z| Measurement Range
1 0.0100Ω ~ 0.1999Ω
2 0.1800Ω ~1.9999Ω
3 1.800Ω ~ 19.999Ω
4 20.00Ω ~ 199.99Ω
5 0.2000kΩ ~ 1.9999kΩ
6 2.000kΩ ~ 19.999kΩ
7 20.00kΩ ~ 199.99kΩ
8 0.1800MΩ ~ 1.9999MΩ
9 2.000MΩ ~ 19.999MΩ
10 18.0MΩ ~ 199.99MΩ
The measurement range is determined by the absolute impedance, |Z| of the DUT. The absolute impedance includes the vector sum of L, C, and R values.
1-5
Model 3525 LCR Meter Instruction Manual Instrument Description
Display A & B - Basic Accuracies
Table 1.2: Basic Accuracy for Display A
Range L C R Z
1 ±(0.90+30/fL1)% ±(0.60+1.5fC1)% ±(1.00+0.21/R1)% ±(1.00+0.15/Z1)%
2 ±2.10% ±2.10% ±2.10% ±1.80%
3 ±0.39% ±0.39% ±0.39% ±0.35%
4 ±0.10% ±0.10% ±0.10% ±0.08%
5 ±0.09% ±0.09% ±0.09% ±0.08%
6 ±0.13% ±0.13% ±0.13% ±0.11%
7 ±0.16% ±0.16% ±0.16% ±0.14%
8 ±0.34% ±0.34% ±0.34% ±0.30%
9 ±(0.17+1.17fL2)% ±(0.17+30/fC2)% ±(0.15+0.20R2)% ±(0.15+0.16Z2)%
10 ±(2.00+1.00fL2)% ±(1.7+30/fC2)% ±(2.00+0.16R2)% ±(2.00+0.11Z2)%
Table 1.3: Basic Accuracy for Display B
Range D θ
1 ±(0.002+0.0015/Z1) ±(0.10+0.09/Z1)°
2 ±0.0179 ±1.00°
3 ±0.0034 ±0.18°
4 ±0.0016 ±0.08°
5 ±0.0011 ±0.05°
6 ±0.0016 ±0.08°
7 ±0.0020 ±0.10°
8 ±0.0036 ±0.19°
9 ±(0.002+0.0015Z2) ±(0.10+0.09Z2)°
10 ±(0.012+0.0014Z2) ±(0.70+0.08Z2)°
IMPORTANT ACCURACY NOTES:
Basic Precision Test Conditions:
Measurement Speed: SLOW
Measurement Signal Level: 1V
Open and Closed Circuit Zero Correction
Performed
Units:
Z
- DUT absolute impedance [Ω]
1
Z
- DUT absolute impedance [MΩ]
2
C
- Capacitance [mF]
1
C
- Capacitance [pF]
2
L
- Inductance [μH]
1
L
- Inductance [kH]
2
f - Test Frequency [kHz]
Other Conditions
Q – Quality factor based on 1/D
C - Accuracy is for when D0.1
L - Accuracy is for when D0.1
R - Accuracy is for when θ≤
The basic accuracy formulas in Tables 1.2 & 1.3 are based on the assumption that the measurement speed is Slow, and the measurement signal level is 1V. The exact accuracies will vary according to the measurement speed and measurement signal level. The formula for calculating the exact measurement accuracy is found in the next section:
1-6
Model 3525 LCR Meter Instruction Manual Instrument Description
Calculating Actual Measurement Accuracies
NOTE: All measurement accuracies assume that open and short circuit zero corrections have
been performed.
Measurement Accuracy = Basic Accuracy X Speed Coefficient X Level Coefficient + 2 Counts
Table 1.4: Speed Coefficient
FAST NORM SLOW
Speed Coefficient
Table 1.5: Amplitude Coefficient
1V 500mV 50mV Amplitude Coefficient 1 1.5 2
NOTE: The 3525 will produce a reading in both range 1 and range 10 using a 50mV test voltage. These readings are for reference use only and are not within any specified accuracy limits.
When measuring C and L, if the condition D0.1 is not satisfied, use Z and θ for accuracy calculations.
See below for an example calculation of this condition:
< Using Z and θ for accuracy calculation > When C=20nF, D=0.5, Measurement Frequency = 1kHz, Signal Level = 1V, and Measurement Speed is SLOW:
1. Solve for Z and θ (The |Z| or θ measured values of the 3525 can also be used.)
θ=tan-1 (1/D)=63.43° Z=(1/ωC)×(1/sinθ)=8.897kΩ
2. Find the accuracy value using the charts for Z and θ
From Z, the measurement range will be range 6. In range 6, the accuracy of Z =±0.11%, and the accuracy of θ =±0.08°. From this accuracy, the maximum and minimum values of Z and θ should be calculated.
: 8.907kΩ Z
Z
max
: 63.51° θ
θ
max
min
min
3. From the max and min values, find the range of C and D.
= 1/(Z
C
max
C
= 1/(Z
min
D
= 1/tan θ
max
= 1/tan θ
D
min
×ω×sinθ
min
×ω×sinθ
max
= 0.5017 (0.0017)
min
= 0.4983 (-0.0017)
max
4. The Accuracy of C =±0.18%, D=±0.0017
3 1.5 1
: 8.887kΩ : 63.35°
) = 2.0036nF (+0.18%)
min
) = 1.9964nF (-0.18%)
max
1-7
Model 3525 LCR Meter Instruction Manual Instrument Description
Instrument Specifications
Measurement Parameters
L (Inductance)
C (Capacitance)
R (Resistance)
|Z| (Impedance)
D (Dissipation Factor / Tan δ)
Q (Quality Factor)
θ (Phase Angle in Degrees)
V (Inter-Terminal Voltage)
I (Inter-Terminal Current)
Measurement Range
L 1.6000uH ~ 199.99kH
C 0.9400pF ~ 199.99mF
R 0.0100 ~ 199.99MΩ
|Z| 0.0100 ~ 199.99MΩ
D 0.0001 ~ 19.999
Q 0.5 ~ 199.99
θ –180.00 ~ 180.00°
V 0.00V ~ 1.00V
I 0.00mA ~ 10.00mA
Basic Precision
0.08% (See Measurement Range and Precision for Exact Accuracies)
Measurement Frequency
1kHz, 120Hz (Frequency Precision: ±0.01% or less)
Output Impedance
100Ω ±10Ω
Measurement Signal Level
50mV, 500mV, 1V (Setting Accuracy: ±10%±10mV)
Maximum Short-Circuit Current
10mA
Measurement Ranges
The measurement range is based on the value of |Z|. Parameters other than |Z|, such as L, C, R, D, Q, & θ are
calculable values.
There are ten ranges from 0.1Ω ~ 100MΩ
Auto or Manual Range Modes
Equivalent Circuit Modes
Auto or Manually selected Parallel or Series Equivalent Circuit Measurement Modes
1-8
Model 3525 LCR Meter Instruction Manual Instrument Description
Instrument Specifications cont’d:
Displays A & B
4½ Digit, High Visibility LED Displays
Measurement & Measurement Speed
Measured parameter values are determined by the voltage and current amplitudes & phases detected by the
instrument’s circuitry and/or the average of calculated values.
Total measurement times are dependent on the measurement frequency, integration speed, comparator settings
and correction factor values.
There are three user-selectable integration speeds to choose from. These are Fast, Normal, and Slow.
Typical Measurement Times:
MODE
MEASUREMENT
FREQUENCY
FAST
NORMAL
SLOW
Trigger Functions
Internal Trigger – Instrument runs in continuous mode.
External Trigger – Executed by the front panel push button, I/O connector TTL input, RS-232C, or GPIB
commands
Measurement Terminal
5-Terminal (BNC with Guard) Kelvin Configuration
Auto Zero Correction
Open Circuit Correction 1kΩ
Short Circuit Correction 1kΩ
Absolute Comparator
Functions
Display A
Display B
Display A & Display B
Output
s
Front Panel LED
Audible Buzzer
External I/O Connector (HI,LO,GO,TOTAL GO,BUSY,END)
RS-232C or GPIB Interface
Front Panel Settings
Up to 99 user-definable front panel settings
System Reset – Initializes the instrument to factory default settings
120 Hz 1kHz
40mS 90mS
360mS
15mS 50mS
250mS
1-9
Model 3525 LCR Meter Instruction Manual Instrument Description
Instrument Specifications cont’d:
Key Lock Function
Disables all front panel keys except the manual trigger key.
Audible Buzzer
User selectable to GO, NO-GO, or OFF operation
Remote Interface
Control I/O Connector (TTL Open Collector; Optically Isolated) – Standard Accessory
RS-232C Interface – Standard Accessory
GPIB Interface – Optional Accessory
BCD Interface – Optional Accessory
Environmental
Operating Conditions: from 32° to 104°F (0° to 40°C), <80% RH; (Non-Condensing)
NOTE: Accuracy errors will double when the unit is operated outside of the specified environmental conditions.
Storage Conditions: from 14° to 131°F (–10° to 55°C), <80% RH; (Non-Condensing)
AC Power Requirements:
100/120/220/240 VAC ± 10%; 50/60 Hz
20VA ±10%
Fuse:
For 100/120 VAC Operation; Use 1A @ 250V, fast acting, TEGAM PN#FU-3102-220
For 220/240 VAC Operation; Use ½A @ 250V, fast acting, TEGAM PN#49743
3525 Dimensions:
Depth: 6.69” (17.0 cm)
Width: 7.87” (20.0 cm)
Height: 3.94” (10.0 cm)
Weight:
5.51 lb (2.5kg)
Standard Accessories:
Users Manual CD Version – PN# 3525-900-01CD
Standard AC Power Cord – PN#161006600
Kelvin Klips - PN#47454
Optional Accessories:
Radial Lead Adapter – PN#3510
Chip Tweezers – PN#2005B
Chip Test Fixture – PN#3511
BCD Interface – PN#3502
GPIB (IEEE-488 Interface) – PN#3501
1-10
Preparation for Use
INSTRUMENT DESCRIPTION
PREPARATION FOR USE
QUICK START INSTRUCTIONS
OPERATING INSTRUCTIONS
PROGRAMMING & INTERFACING
SERVICE INFORMATION
APPENDIX
Model 3525 LCR Meter Instruction Manual 2-1
Model 3525 LCR Meter Instruction Manual Preparation for Use
Unpacking & Inspection:
Each 3525 is put through a series of electrical and mechanical inspections before shipment to the customer. Upon receipt of your instrument, unpack all of the items from the shipping carton and inspect, them for any damage that may have occurred during transit. Report any damaged items to the shipping agent. Retain and use the original packing materia l f o r reshipment if necessary.
Upon receipt, inspect the carton for the following items: Model 3525 General Purpose, Programmable LCR Meter
Model 3525 User’s Manual CD Kelvin Klips™ PN#47454
!
Safety Information & Precautions:
The following safety information applies to both operation and service personnel. Safety precautions and warnings may be found throughout this instruction manual and on the equipment. These warnings may be in the form of a symbol or a written statement. Below is a summary of these precautions.
Terms in This Manual:
CAUTION statements identify conditions or practices that could result in damage to the equipment or other property.
WARNING life.
statements apply conditions or practices that could result in personal injury or loss of
Terms as Marked on Equipment:
CAUTION indicates a personal injury hazard not immediately accessible as one reads the marking, or a hazard to property including the equipment itself.
DANGER
indicates a personal injury hazard immediately accessible as one reads the marking.
2-2
Model 3525 LCR Meter Instruction Manual Preparation for Use
!
Safety Information & Precautions Cont’d:
Symbols:
As Marked in This Manual:
!
This symbol denotes where precautionary information may be found.
As Marked on Equipment:
!
CAUTION – Risk of Danger
DANGER – Risk of Electric Shock
Earth Ground Terminal
l
O
On
Off
Frame or Chassis Terminal
Earth Ground Terminal
Alternating Current
Grounding the Equipment
This product is grounded through the grounding conductor of the power cord. WARNING: properly wired receptacle before using this instrument. The proper grounding of this instrument is essential for safety and optimizing instrument operation.
Danger Arising from Loss of Ground
WARNING: in the instrument. Under these conditions all accessible parts, including insulating parts such as keypads and buttons, could develop a hazardous voltage and put the user at risk.
To avoid electrical shock or other potential safety hazards, plug the power cord into a
If the connection to ground is lost or compromised, a floating potential could develop
2-3
Model 3525 LCR Meter Instruction Manual Preparation for Use
!
Use the Proper Fuse
To avoid fire hazard, use only the correct fuse type as specified for the AC power supply in the “Instrument Description” or “Service Information” sections of this manual. Please note that the fuse rating for 100/120-volt operation is different than the rating for 200/240-volt operation. Information about the proper fuse type is also printed on the rear panel of the instrument.
Refer fuse replacement to qualified service personnel.
Do Not Use in Explosive Environments
WARNING:
Do not Operate Without Covers
WARNING: missing panels or covers could result in personal injury.
The 3525 is not designed for operation in explosive environments.
This device should be operated with all panels and covers in place. Operation with
FOR QUALIFIED SERVICE PERSONNEL ONLY
Servicing Safety Summary:
!
Do Not Service Alone
Do not perform service or adjustment on this product unless another person capable of rendering first aid is present.
Use Care When Servicing with Power On or Off
Dangerous voltages may exist at several points in this product. To avoid personal injury or damage to this equipment, avoid touching exposed connections or components while the power is on. Assure that the power is off by unplugging the instrument when removing panels, soldering, or replacing components. WARNING: present throughout the instrument even when the instrument is in the OFF state. Always unplug the instrument and wait 5 minutes before accessing internal components.
Power Source
This product is intended to connect to a power source that will not apply more than 250V RMS between the supply conductors or between either supply conductor and ground. A protective ground connection by way of the grounding conductor in the power cord is essential for safe operation.
The instrument power source is electronically con trolled, meaning that there is power
2-4
Model 3525 LCR Meter Instruction Manual Preparation for Use
Line Voltage Selection:
!
CAUTION: DO NOT APPLY POWER TO THE INSTRUMENT BEFORE READING THIS SECTION: Unless otherwise specified, the Model 3525 is delivered from TEGAM with its power supply set for
120V, 60 Hz operation. However, the 3525 design allows it to operate under 100/120/220/240V @ 50/60 operation. It is strongly recommended that the line voltage, frequency setting, and fuse type be verified before powering the unit.
First, determine the operating voltage that the instrument will be supplied. Also, verify that the supply voltage does not fall outside of the allowable ranges in the table below:
Table 2.1 – Operating Voltage Ranges
Switch Position Voltage Range
100 90V ~ 110V 120 108V ~ 132V 220 198V ~ 242V 240 216V ~ 250V
The following procedure describes the steps necessary to change the 3525 power settings from factory default settings to a new setting.
1. Verify that there is no power connected to the unit.
2. Change the line selector switch located on the rear panel of the 3525 to the desired operating voltage by using a small screwdriver to turn the voltage selection switch to the desired position.
Refer to the figure below:
With your screwdriver, turn the desired voltage number to the black arrow. For example, the setting here is for 240V.
3. The fuse size may need to be changed as a result of the increased or decreased voltage operation. Please refer to the “Service Information” section of this manual (page 6-3) for detailed fuse information. The rear panel of the 3525 will also contain information regarding fuse sizes and types.
2-5
Quick Start Instructions
INSTRUMENT DESCRIPTION
PREPARATION FOR USE
QUICK START INSTRUCTIONS
OPERATING INSTRUCTIONS
PROGRAMMING & INTERFACING
SERVICE INFORMATION
APPENDIX
Model 3525 LCR Meter Instruction Manual 3-1
Model 3525 LCR Meter Instruction Manual Quick Start Instructions
Proper use of the Model 3525 will maximize measurement accuracy and productivity. The Quick Start section is designed to provide speedy setup for general-purpose applications. It is highly recommended that the user read Chapter 4 for applications involving high accuracy or requiring optimal performance of the instrument.
Power the Unit
!
CAUTION
The power supply of the Model 3525 is designed for 50-60 Hz operation and a voltage range of 100-240 VAC. Review the line voltage selection procedure on page 2-5 before proceeding. Make sure that the line voltage setting matches the supply voltage and frequency.
Power the unit and allow at least 30 minutes for the unit to warm up. Make sure that the safety precautions on pages 2-3 and 2-4 have been reviewed and understood. Verify that the environmental conditions, listed on pages 1-10, are met.
Factory Default Settings
The 3525 is shipped from the factory with instrument settings as follows:
Table 3.1 – Factory Default Settings
Parameter Setting Parameter Setting 1 Range AUTO 8 Communication *RS232 or GPIB 2 Trigger Internal 9 Key Lock UN-LOCKED 3 Speed FAST 10 4 Display A Capacitance 11 5 Display B Dissipation Factor 12 6 Trigger Internal 13 7 Range Auto 14
* When a device clear command is sent via communications interface, this value does not change.
These settings can be recalled by sending a device clear command, *RST via RS232, or GPIB interface.
NOTE: When the 3525’s power is cycled, the instrument’s settings will return to the state that existed at the time the power was last turned off. You can initialize the 3525 to factory default settings by pressing the [MANU] key while powering the unit.
Comparator/Buzzer OFF
Frequency 1kHz
Voltage Level 1V
Circuit Mode AUTO
Manual/Remote Remote
3-2
Model 3525 LCR Meter Instruction Manual
Quick Start Instructions
Instrument Setup
Setup of the Model 3525 for general-purpose use is easy. Once the unit has been allowed to warm up, review the front panel and note the changes that need to be made in order to perform your required measurements. Below are Quick Setup instructions to adjust the main measurement parameters of the 3525. Refer to Chapter IV, Operating Instructions, for a comprehensive description of all the measurement options and instrument settings, including the ones listed below:
Display A
To select L, C, R or |Z| value to be indicated on Display A, make sure that the comparator function
is disabled by pressing the [ON] key until all of the LED’s in the JUDGMENT field are off. In the FUNCTION field, press the [A] key until the desired parameter’s annunciator is lit.
Display B
To select D, Q, θ, V or I value to be indicated on Display B, make sure that the comparator
function is disabled by pressing the [ON] key until all of the LED’s in the JUDGMENT field are off. In the FUNCTION field, press the [B] key until the desired parameter’s annunciator is lit.
Measurement Speed
In order to select FAST, NORMAL, OR SLOW mode, press the [SPD] key until the desired measurement rate is illuminated.
Test Frequency
In order to choose a 1kHz or 120 Hz test frequency, press the [FREQ] key until the desired frequency LED is illuminated.
Test Voltage Level
Select either 1V, 500mV, or 50 mV test voltage by pressing the [LEV] key until the desired voltage is selected.
Circuit Mode
The default equivalent circuit mode is AUTO. Refer to Chapter IV for information on series or parallel equivalent circuit measurements.
Range
The default range setting is AUTO. Refer to Chapter IV for more information on manual range selection.
Trigger
Press the [INT] key to select internal (continuous) triggering mode. Press the [MANU] key to select manual or external triggering options. Trigger the unit from the front panel by pressing the [MANU] key.
3-3
Model 3525 LCR Meter Instruction Manual Quick Start Instructions
Performing an Open and Short Circuit Null
Once you have verified that all of the instrument settings are correct, the next step is to perform a zero correction.
1. Assuming that Kelvin Klip™ leads are being used, orient the leads in a similar position that
they will be connected to the DUT. Open the leads (refer to page 4-19 for proper lead orientation.)
2. Press the [OPEN] key in the Zero field of the front panel. Allow the unit to step through the
open circuit null procedure.
3. Then short the Kelvin Klip™ leads together. Press the [SHORT] key and allow the
instrument to step through the short zero adjustment.
You have completed the Zero Correction procedure and are now ready to take basic impedance measurements. Refer to Chapter IV, Operating instructions for additional details or information regarding the use of this instrument.
Measurement Tips
Here are some general tips for users who are relatively new to performing LCR measurements:
Unless otherwise specified, use 120Hz for capacitance measurements and use 1kHz for
testing inductors.
The 3525 will take the most accurate reading at SLOW measurement speed and 1V test
voltage level.
Orientation of test leads is important for repeatable measurements especially at high
frequencies. Try to keep lead orientation the same when making precision measurements.
Keep metallic or other objects away from test leads including hands when performing zero
correction or impedance measurements.
When measuring components with values that fall to the outer limits or outside of a
particular range, put the instrument in manual range to eliminate range switching by locking into one range. This increases repeatability and measurement accuracy.
When testing capacitors, make sure that the capacitor is completely discharged before
connection to the measurement terminals of the 3525.
Use the Guard Terminal for high impedance measurements to minimize leakage errors and
noise.
3-4
Operating Instructions
INSTRUMENT DESCRIPTION
PREPARATION FOR USE
QUICK START INSTRUCTIONS
OPERATING INSTRUCTIONS
PROGRAMMING & INTERFACING
SERVICE INFORMATION
APPENDIX
Model 3525 LCR Meter Instruction Manual 4-1
Model 3525 LCR Meter Instruction Manual Operating Instructions
Please read and understand Chapter III, Quick Start Instructions before proceeding with this section.
Basic Operation
The Model 3525 is a highly versatile product, designed for use in many different applications. There are ideal configurations of the 3525 for each measurement application. These configurations optimize accuracy, measurement speed, and flexibility. This section is designed to give the user a comprehensive description of the parameters and operating modes available from the Model 3525. The user will be exposed to additional topics that will enhance the integration of the Model 3525 into their application. It is highly recommended that the user read this section thoroughly if the intended application requires high precision and maximum throughput.
Default Parameters
Model 3525 is shipped from TEGAM fully calibrated, tested, and preset to factory default settings to accommodate general-purpose manual operation. Chapter III, Quick Start Instructions, contains specific information about factory default settings and some general tips for new users. Please read and understand Chapter III, Quick Start Instructions, before proceeding.
4-2
Model 3525 LCR Meter Instruction Manual Operating Instructions
Front Panel Description
Figure 4.1a: Front Panel Layout
DISPLAY A
COMPARATOR
STATUS
DISPLAY
COMPARATOR
OPERATIONS PANEL
MEASUREMENT TERMINALS
DISPLAY B
PARAMETER
SELECTION KEYS
PARAMETER MODE
INDICATORS
DISPLAY A
4-1/2 Digit LED Display indicates measured values for L, C, R, and |Z|.
DISPLAY B
4-1/2 Digit LED Display indicates measured values for D, Q, θ, V, and I.
Parameter Mode Indicators
Displays active parameter modes and settings. Comparator Status Display Displays judgment results for display A, Display B or both. Results are based on comparator settings and measurement results.
Parameter Selection & Control Keys
Use these keys for selecting measurement modes and parameters.
Comparator Operations Panel
Used for enabling the comparator, programming high and low limits and for preset panel selection.
Measurement Terminals:
A 5-terminal scheme is used: HFORCE – High potential terminal for sourcing measurement current. HSENSE - The "high" terminal for voltage sensing. LSENSE - The "low" terminal for voltage or sensing LFORCE - Low potential terminal for sourcing measurement current. GUARD - The guard terminal. Use this connection to cancel any stray capacitance and conductance associated with high impedance measurements.
4-3
Model 3525 LCR Meter Instruction Manual Operating Instructions
R
R
Front Panel Description cont’d:
Figure 4.1b: Measurement & Parameter Control Area
MEASUREMENT SPEED
DISPLAY A OR B
PARAMETER SELECTION
TEST FREQUENCY
COUNT SETTING
Measurement Speed Key
Use this key to select either FAST, NORMAL, or SLOW measurement speed.
Test Frequency Key
Use this key to select either 120Hz or 1kHz measurement frequency.
Measurement Signal Level Key
Press this key to select 1V, 500mV, or 50mV test signal amplitude.
Equivalent Circuits Mode Key
Press this key to select Series, Parallel, or AUTO equivalent circuit measurements.
Range Mode Key
Press the AUTO key to toggle between AUTO and Manual ranging modes.
Internal Trigger Key
Press this key to select continuous internal triggering of the instrument.
Manual Trigger Key
Press the manual trigger key to select manual or external triggering from the front panel or remote interface. Consequent pressing of the [MANU] key will trigger the instrument.
MEASUREMENT SIGNAL AMPLITUDE LEVEL
EQUIVALENT MEASURMENT CIRCUIT
AUTO OR MANUAL RANGE
INTERNAL TRIGGER
CURSO
POSITIONING
ENTER KEY
ANGE UP/DOWN
MANU
EXTERNAL TRIGGER SELECTION
MANUAL TRIGGER AND
SELECTION
POWER ON/OFF
DISPLAY A Selection Key
Press [A] to select the active measurement parameter for Display A.
DISPLAY B Selection Key
Press [B] to select the active measurement parameter for Display B.
Count Setting Keys
Pressing the [▲] or [▼] keys allows the user to change count, decimal place or units settings during comparator setup.
Cursor Positioning Keys
The [►] or [◄] key allows movement of the display cursor when programming comparator settings.
Enter Key
Enters new parameters or modes into instrument memory.
Range Down key
In manual range mode, this key decrements the active measurement range number.
Range Up Key
In manual range mode, this key increments the active measurement range number.
Power Switch
Toggles the unit power either on or off.
4-4
Model 3525 LCR Meter Instruction Manual Operating Instructions
T
T
Front Panel Description cont’d:
Figure 4.1c: Comparator Controls Area
PANEL SELECTION
COMPARATOR LOW LIMIT
COMPARATOR HIGH LIMIT
Panel Selection
Use this key to recall panel presets from 00-99
Comparator Low Limit
Press this key to access the comparator low limit programming mode.
Comparator High Limit
Press this key to access the comparator high limit programming mode.
SHORT
Comparator ON/OFF
Pressing this key will either enable or disable comparator operation.
Open Circuit Null
With the test leads open, pressing this key will initiate the open circuit zeroing process.
Short Circuit Null
With the test leads shorted, pressing this key will initiate the short circuit zeroing process.
COMPARATOR
ON/OFF
OPEN CIRCUI
NULL
SHORT CIRCUI
NULL
4-5
Loading...
+ 66 hidden pages