TEGAM is a manufacturer of electronic test and measurement equipment for metrology, calibration, and
production test. We also provide repair, calibration, and other support services for a wide variety of test
and measurement equipment including RF power sensor calibration systems, RF attenuation measurement
systems, ratio transformers, arbitrary waveform generators, micro-ohmmeters, LCR meters, handheld
temperature calibrators, thermometers, humidity and temperature control devices, and more.
TEGAM also repairs and calibrates test and measurement equipment formerly manufactured by ElectroScientific Industries (ESI), Gertsch, Keithley Instruments, Lucas Weinschel, and Pragmatic Instruments. A
complete list can be viewed on our Product Service Directory at www.tegam.com
For more information about TEGAM and our products, please visit our website at www.tegam.com
contact one of our customer service representatives at sales@tegam.com or 800-666-1010.
Safety Information & Precautions:
The following safety information applies to both operation and service personnel. Safety precautions and
warnings may be fo und throughout this instruction manual and the equip ment. These warnings may be in
the form of a symbol or a written statement. Below is a summary of these precautions.
Terms in This Manual:
CAUTION statements identify conditions or practices that could result in damage to the equipment or
other property.
WARNING statements identify conditions or practices that could result in per sonal injury or loss of life.
Terms as Marked on Equipment:
CAUTION indicates a personal injury hazard not immediately accessible as one reads the marking, or a
hazard to property including the equipment itself.
DANGER indicates a personal injury hazard immediately accessible as one reads the marking.
Symbols:
As Marked in This Manual:
This symbol denotes where precautionary information may be found.
(Depends on noise floor and resolution of D C instruments)
Calibration Factor Drift
<0.5% per year
10 to 100 MHz in 10 MHz steps
35 to 50 GHz in 1 GHz steps
Thermistor DC Bias Power
Approximately 40 mW (nominal)
Thermistor Resistance at Bias
200 Ohms (set by balancer)
Temperature
Storage
-55° to +75° C (-67° to +167° F)
Purpose and Function
These Coaxial RF Power Standards are designed for the precise measurement of microwave power in the
10 MHz to 50 GHz frequency range. These standards can be configured as a Reference Standard or
Working Standard. The Working RF Power S ta ndard (or Feedthrough RF Power Standard) is a thermistor
mount with power splitter combination employed as a Feedthrough standard for the calibration of
terminating power sensors such as thermoelectric, diode and thermistor power sensors. The Reference RF
Power Standard (or Terminating RF Power Standard) is a ter m inat ing t hermistor mount used for the
calibration of Feedthrough standards and applications that require direct measurement of RF power.
Capabilities
These Coaxial RF Power Standards are designed for the calibration laboratory. These units are highly
accurate, stable with time and temperature , and are designed for use as a standard for the
characterization of calibration factors for other sen sors.
Table 1.1 General Specifications
Power Linearity
Typical Usable Range
Individual Calibration factors are
supplied standard at the following
frequencies. Inquire about
additional points.
Operating
<0.1% from 1 to 10 mW
-20 dBm to +10 dBm (0.01 to 10mW) typical
100 MHz to 2 GHz in 50 MHz steps
2 GHz to 4 GHz in 100 MHz steps
4 to 12.4 GHz in 200 MHz steps
12.75 to 18 GHz in 250 MHz steps
18 to 26 GHz in 1 GHz steps
+/-1.25% from 0.05 to 4.0 GHz
+/-1.5% from 4.20 to 12.0 GHz
+/-2.2% from 12.2 to 17.5 GHz
+/-2.5% from 17.75 to 26.5 GHz
+/-3.0% from 27.0 to 44.0 GHz
+/-4.0% from 45.0 to 50.0 GHz
10.5 cm (4.1 in)
21.7 cm (8.5 in)
33.8 cm (13.3 in)
100 MHz to 2 GHz: 0.0267
2 GHz to 12.4 GHz: 0.1356
12.4 GHz to 18 GHz: 0.2249
18 GHz to 26.5 GHz: 0.3200
26.5 GHz to 50 GHz: 0.3389
+/-1.4% from 0.05 to 4.0 GHz
+/-1.7% from 4.20 to 12.0 GHZ
+/-2.3% from 12.2 to 17.5 GHz
+/-2.6% from 17.75 to 26.5 GHz
+/-3.2% from 27.0 to 44.0 GHz
+/-4.2% from 45.0 to 50.0 GHz
Additional items (required and optional) for operation and maintenance
Required:
• TEGAM Model 1830A, 1806A, 1806, or 1805B1 RF Power Meter
o 1830A Cable (CA-21-15/48)
o 1806A Cable (CA-28-48)
o 1806 Cable (CA-29-48)
o 1805B Cable as heater only (CA-29-48)
• RF Signal Generator
o 10 MHz to 50 GHz
o Minimum power level 17 dBm at 50 GHz
• Torque Wrench
o Torque wrench, 8mm, 8 in-lbs (2510-910-01)
Optional:
• RF Amplifier
• SureCAL Software (P/N SureCAL-PM)
• Rack Mount Kit
o (Single - P/N 1830-910)
o (Dual - P/N 1830-911)
• VNA
o 2-port
o 50 GHz
• TEGAM 1510A (For calibration of TEGAM 2510A)
• Transport Cas e
o 1510 RF Power Standard (1500-910)
o 1830A, F113X & 2510A (2500-910)
• Adapters
o Coax adapter, 2.4mm (F) to 2.92mm (M) with Data (1510-911-01)
o Coax adapter, 2.4mm (F) to 3.5mm (M) with Data (1510-912-01)
o Coax Adapter, 2.4mm (M) to 2.92mm (F) with Data (2510-911-01)
o Coax Adapter, 2.4mm (M) to 3.5mm (F) with Data (2510-912-01)
o Coax Attenuator, 2.4mm (M to F) 30dB with Data (2510-913-01)
1
TEGAM models 2510A and 1510A are designed to be used with TEGAM Models 1830A, 1806A, and 1806. In
addition, the TEGAM Model 1805B can be used to heat the 2510A and 1510A with cable CA-29-48. The
1805B is for heating only and cannot be used for power measurements or biasing of the 2510A or 1510A.
Once you have verified t hat the cause for the Coaxial RF Power Standards malfunction cannot be solved in
the field and the need for repair and calibration service arises, contact TEGAM customer service to obtain
an RMA, (Returned Material Authorization), number. You can contact TEGAM customer service via the
TEGAM website, www.tegam.comStates Only).
The RMA number is unique to your instrument and will help us identify you inst rument and to addre ss the
particular service request by you which is assigned to that RMA number.
Of even importance, a detailed written description of the problem should be attached to the instrument.
Many times repair turnaround is unnecessarily delayed due to a lack of repair instructions or of a detailed
description of the problem.
This description should include inform ation such as measure ment range, and other instrument settings,
type of components being tested, are the symptoms intermittent, conditions that may cause the
symptoms, has anything changed since the last time the instrument was used, etc. Any detailed
information provided to our technicians w il l assist them in identifying and correcting the problem in the
quickest possible manner. Use a copy of the Repair a nd C alibration Service form provided on the next
page.
Once this information is prepared and sent with the instrument to our service department, we will do our
part in making sure that you receive the best possible customer service and turnaround time possible.
or by calling 440.466.6100 (All Locations) OR 800.666.1010 (United
Use this for m to pro vide additional repair information and serv ice instructions. The Completion of this
form and including it with your instrument will ex p edite the processing and repair process.
Repair Instructions:
Evaluation Calibration Only Repair Only Repair & Calibration Z540 (Extra Charge)
Detailed Symptoms:
Include information such as measurement range, instrument settings, type of components being
tested, is the problem intermittent? When is the pro bl em most frequent?, Has anything changed with
the application since the last time the instrument was used?, etc.
TEGAM, Inc. warrants this product to be free from def ects in material and workmanship for a period of
three years from the date of shipment. During this warranty period, if a product proves to be defective,
TEGAM Inc., at its option, will either repa ir the defective product without charge for parts and labo r, or
exchange any product that proves to be defective.
TEGAM, Inc. warrants the calibration of this product for a period of 1 year from date of shipment. During
this period, TEGAM, Inc. will recalibrate any product, which does not conform to the published accuracy
specifications.
In order to exercise this warranty, TEGAM, Inc., must be notified of the defective product before the
expiration of the warranty period. The customer shall be responsible for packaging and shipping the
product to the designated TEGAM service ce nter with shipping charges prepaid. TEGAM Inc. shall pay for
the return of the product to the customer if the shipment is to a location w ithin the country in which the
TEGAM service center is located. The customer shall be responsible for paying all shipping, duties, taxes,
and additional costs if the product is transported to any other locations. Repaire d products are warranted
for the remaining balance of the original warranty, or 90 days, whichever period is longer.
Warranty Limitations
The TEGAM, Inc. warranty does not apply to defects res u lting from unauthorized modification or misuse of
the product or any part. This warra nty does not apply to fuses, batteries, or damage to the instrument
caused by battery leakage.
Statement of Calibration
This instrument has been inspected and tested in accordance with specifications published by TEGAM Inc.
The calibration of this instrument is traceable to the International System of Units (SI) through the
National Institute of Standards and Te chnology (NIST) or other recognized National Metro logy Institutes,
by comparison to equipment and standards maintained in the laboratories of TEGAM Inc.
Document publishing dates may lag produc t changes.
Visit www.tegam.com to download the late st version of this manual.
2510A or 1510A
Technical (Operation and Maintenance) Manual
1510A/2510A-900
Calibration Certificate with Data
N/A
Unpacking & Inspection
Each Coaxial RF Power Standards is put through a series of electrical and mechanical inspections before
shipment to the customer. Upon receipt of your instrument unpack all of the items fro m the shipping
carton and inspect for any damage that may have occurred during transit. Report any damaged items to
the shipping agent. Retain and use the orig ina l p a cking material for reshipment if necess a ry.
Upon Receipt, inspect the carton for the following items:
Table 2.1 Packing List
Mounting
The Model 2510A is shipped with four plastic feet mounted to the bottom cover. Whe n any of these
models are placed on a bench or table, the feet support the instrument. For rack mounting of Model
2510A use rack mount kit, (Single) P/N 1830-910 – (Dual) P/N 1830-911.
The Model 1510A is shipped without any type of rack mounting equipment the new design allows for use
as a terminating sensor and it low weight of 544.3 g (1.2 lb) reduces any stress on connectors.
Only use with Proper Equipment:
Connecting this device to any other type of circ uit or may result in permanent damage to the components.
Additionally, the heater elements used for the internal heater are designed to be used only with the heater
control circuits found in TEGAM Models 1830A, 1820B, 1806A, 1806, and 1805B. Connecting these heater
elements to any other circuit or device m ay result in damage. Contact TEGAM for any questions regarding
instruments that are compatible with these RF Powe r Sta nda rds.
Use in Proper Environment
Normal calibration laboratory practice dictates that the environment should be closely contr olled. This will
minimize errors introduced by temperature and humidity changes. A nominal temperature of +23°C
(+73.4°F) provides a good working condition. A tolerance of ±1°C gives an ideal temperature spread.
Controlled temperatures also stabilize the aging process of the standards.
CAUTION: The RF Power Standards have a specified operating temperature range of +15° to +30° C
(+59° to +86° F). Operating beyond these limits can affect the accuracy of the instruments and damage
internal circuitry.
CAUTION: When an RF Power Standard is to be stored for extended periods , pack the instrument into a
container. Place container in a clean, dry, temperature-controlled location. If instrument is to be stored in
excess of 90 days, place desiccant with items before sealing container. The safe environmental limits for
storage are -55° to +75° C (-67° to +167° F) at less than 95% non-condensing relative humidity.
Theory of RF Measurements with the 2510A and 1510A
Power Measurements
RF power is measured in terms of a power change across the precision resistance leg of a TEGAM 1830A,
1806A, 1806 Bridge Circuit. A digital voltmeter measures voltage across the precision resistance leg which
can be used to determine the power.
The total power applied to the thermistor leg (in the RF Power Standard) of the Type IV Bridge equals the
sum of both DC and RF power. The precision resistor leg only has DC power applied to it. Thus, the RF
power introduced to the thermistor is directly pro portional to the change in DC power across the precision
resistor.
Principle of DC Substitution
The RF Power Standards use the principle of DC substitution to measure RF power. DC substitution refers
to the measurement of RF power according to the amount of DC power that must be substituted for the RF
power in a bolometer in order to cause equivalent thermal effects. Since some of the RF power applied to
the input of the power standard is lost by reflection and other causes before it is applied to the thermistor
element, a calibration factor for the standard is to determine the actual level of RF power. TEGAM Coaxial
RF Power Standards are configured to prov ide a path f or RF energy via a coaxial line. DC Bias is
introduced from the Bias terminals to the thermistors.
Temperature Variable Resistance
Each RF Standard contains a pair of thermistor beads whereby the resistance changes as a function of
temperature. Thermistor bead temperature is a function of the combined DC and RF power a p plied to the
beads and the ambient temperature surrounding the beads. The level of power applied to the beads is
controlled externally. A heating element, controlled by an external controller, provides ambient
temperature stability.
Figure 3.1 Thermistor and Bridge Balancer
Figure 3.1 depicts the thermistor assembly electrical configuration. DC biasing of the dual thermistor
beads to 100 ohms each provides a nominal 50 ohm parallel RF resistance and a 200 ohm series DC
resistance. A DC blocking capacitor and bypass capacitors isolate DC from RF signals. Filtering capacitors
provide low VSWR in the lower end of the frequency range. Application of approximately 40 mW of power
to the thermistor beads produces a 200 ohm DC resistanc e. As the power applied to the thermistor bea ds
increases, their effective resistance values decrease. This is due to the negative temperature coefficient of
the beads. Initially, the beads are DC biased to 200 ohms. Application of RF power increases the power
level present at the beads and causes the effective resistance value of the beads to drop. The bridge
circuit of the power meter reduces DC power until the initial effective resistance value is restored (200
ohms). The total amount of DC power removed is proportional to the amount of RF power that was
introduced. Quantitatively, the total power applied to the thermistor beads equals the sum of the two
types of power.
Power Splitter
The Models 2510A contain a two-element resistive power splitter. The T-shaped divider contains a series
50 ohm resistor in each of the two legs. The test port is an APC 2.4 mm precision female connector. Use
of the splitter in a closed loop configurat ion that a p plies constant power causes the common point
(divider) to become a constant voltage point. This means the source impedance at both splitter output
ports is determined by the 50 ohm resistor and the output connector. This provides very good source
match. In addition, the power is split equally betw een the two ports.
Controlling Thermistor Temperature
Since thermistor mounts are temperature sensitive devices, it is necessary to eliminate or minimize the
effects of changes in the ambient tempera ture. This is accomplished by thermally isolating the mount,
raising its temperature with a heater element to a level higher than the ambient temperature, and
maintaining that level by means of an external temperature controller. The heater element electrical
assembly as illustrated in Figure 3.2. The proper temperature is determined by the characteristics of the
thermistor beads.
Figure 3.2 Simplified of RF Power Standards Internal Heater
The TEGAM RF Power Standards are designed to be employ ed as the fourth arm of a bridge configuration.
These units are designed as the sensing element in RF power calibration and measurement systems.
These systems employ other electronic elements to effect control of the measurement routines. An
example of this type of system is the TEGAM PMX50-001 Power Sensor Calibration System. Proper use of
these standards and the TEGAM PMX50-001 is further documented in other TEGAM Operation and Service
Manuals. The Working RF Power Standard (Model 2510A) are feedthrough mount-splitter combinations are
used to calibrate thermoelectric, diode, and thermistor power sensors. The Reference RF Power Standards
(Model 1510A) are terminating thermistor mounts commonly used as a reference stand a rd fo r ca librating
working standards.
Model 2510A Connector Descriptions
Figure 4.1 TEGAM 2510A Connectors
RF IN connector
APC 2.4 mm female connector connects to signal generator output. The RF power that comes in this
connector is applied equally to the DUT and the power standard. The fact that equal RF power is applied to
both the power standard and the DUT is what allows the d etermination the calibration factor of the DUT.
There is typically 6 dB nominal with a max of 10.5 db of insertion loss in the RF in path of the power
splitter, plus the loss in the standard.
BIAS VOLTAGE/Heater combination connector
A combined connect io n is used to connect the thermistor element to a DC Substitution bridge circuit and
the heater control circuit as is found in the TEGAM Model 1830A, 1806A, and 1806. The bridge operates
on the principal of DC substitution, so only DC voltages and currents are present at these terminals when
connected. In addition, the TEGAM Model 1805B can be used to heat the 2510A and 1510A with cable CA29-48.
SENSOR connector
APC 2.4 mm female coaxial connector provides the RF power to the D UT. The RF power applied to RF IN
connector is applied equally to the DUT and the internal thermistor mount. The fact that equal RF p ower is
applied to both the power standard and the DUT is what allows the calibration factor of the DUT to be
determined.
A combined connect io n i s used to connect the thermistor element to a DC Substitution bri dg e circuit and
the heater control circuit as is found in the TEGAM Model 1830A, 1806A, and 1806. The bridge operates
on the principal of DC substitution, so only DC voltages and currents are present at these terminals when
connected. In addition, the TEGAM Model 1805B can be used to heat the 2510A and 1510A with cable CA29-48.
RF INPUT connector (on the front, not shown)
APC 2.4 mm male coaxial connector where RF power to be measured is applied.
Connecting RF Power Standards
The TEGAM RF Standards are an element of a system and is not a “stand-alone” instrument. The 2510A
and 1510A RF Power Standard can only be connected to the TEGAM 1830A, 1806A, and 1806. Unlike
other TEGAM RF Power Standards, the bias and temperature are connected through one cable. To do this,
simply connect one end of the appropriate cable to the 2510A or 1510A and the other end to the
“SENSOR” and “HEA T ER” connection of the 1830A, 1806A, 1806, and 1805B. Use the appropriate cables
that meet your equipment application
The RF IN on the Working (Feedthrough) Power Standar ds is an APC 2.4 mm connector. An APC 2.4 mm
female coaxial connector is connected to the output of the chosen signal generator, which should be 50
Ohm nominal impedance. A low loss coaxial cable is recommended to a void an excessive power loss.
These Models have an additional female coaxial connector called the “SENSOR” port. This connector is
actually one of the arms of the power splitter (described in Section III Principles of Operation). The DUT is
2
.
2
When connecting to the 1806 or 1806A the FLOAT/GROUND switch should be in the FLOAT (dow n) po s ition.
connected to this port for calibration. When the Wor king RF Power Standard is being calibrated, the
Reference RF Power Standard is connected to this port. This connector is an 2.4 mm female. Proper care,
cleaning, alignment, and torquing of co axial connectors should be practiced to make accurate
measurements, reduce insertion loss and extend the life of th e connectors.
The RF input on the Reference (Terminating) Power Standards (Model 1510A) is a male 2.4 mm connector
that is connected to the RF power source to be measured. Again, proper care, cleaning, alignment, and
torquing of coaxial connectors should be practiced to make accurate measurements, reduce insertion loss,
improve repeatability, and extend the life of the connectors.
Once cabling has been correctly installed, there are no further operator adjustments to be made to the RF
Power Standards. However, allow at least two hours for the heater in the RF Standard to reach its
operating temperature. Additionally, it is recommended that once the BIAS VOLTAGE terminals are
connected, the D C Substitution Bridge should be on for one hour before measurements are taken
3
.
Calculating Calibration Factors
When using a RF Power Meter with a RF Power Sensor to make a RF Power Measurement the user must
know that the measurement is accurate and there is traceability to a known standard. All diode, thermoelectric, and thermistor power sensors have calibration factors associated to particular frequencies that
are used to insure an a ccurate power measurement. Technicians and engine ers use these calibra t io n
factors when making measurements but where do these calibration factors really come from.
Calibration factor of a terminating power sensor, if it is a DC-substitution sensor, relates the change in DC
substituted power to the total RF power incident on the sensor. For this purpose, incident means all of the
RF propagating toward the sensor reference plane, including power that is subsequently reflected. On the
|
=|
signal flow diagram (figure 4.3), the incident power is
.
Calibration factor of a feed-through power calibration setup, again if the monitor is a DC-substitution
sensor, relates the change in DC substituted power in the monitor to the power delivered out of the DUT
port into a load of exactly the nominal charac teristic impedance of the system, or
feed-through DUT port as the output of a leveled generator, then the output into a perfect load is
. If we think of the
.
So for a terminating sensor (using the “M” subscript following Weinschel part numbering):
=
And for a feed-through standard, (using “F” to indicate “feedthrough”):
=
Where:
Calibration factor of the Terminating Mount
k
M =
Calibration factor of the Feed-through Moun t
k
F =
= Power measured terminating mount
P
SubM
= Power measured Feed-through mount
P
SubF
3
When connecting a 1510A or 2510A to an 1806 or 1806A; in addition to a 2 hour warm up time, operator should observe stability
in the bias voltage. Bias voltage stabil ity refers to less than 50uV change per 30 minutes as a general guideline. The error light do es
not apply when connecting the 1510A or 2510A to the 1806A.
In all calibrations, calibration factors are transferred from a terminating reference, to a feed-through
stand, and then into the DUT. In some procedures, this transfer occurs all in one session, a nd in others
time is saved by transferring into the feed-through once and then using that feed-through calibration
factor to calibrate DUT for some time.
In a perfect world, the terminating sensor would present a perfect load, and then by reorganizing the
above two definitions with incident and output power equal, we would have,
=
Because power sensors are always imp erfect loads, additional analysis and sometimes correction is
required.
Figure 4.1 shows a signal flow diagram of two one-port devices connected together. The left-hand port is
the “output”, or “generator” port and is designated on the diagram using the subscript, “g”. The righthand port is the “input”, or “load” port and is des ignated using the subscript “l”. It’s clear from the
diagram that port reflections dominated by the Gamma vectors generally cause the power that the
generator port would output into an ideal load,
We really have to take into account the combination of reflections, or “port match” to come up with a
general understanding of calibration transfer.
Start with the signal flow diagram. In this case, a Generator “g” is represented by an ideal source “s”,
and its port reflection, and the sensor is ter med the load, “l”.
, to not be equal to the power , incident on the load.
Armed with this last result, and calling the Feed-through stand (F subscript) the “generator” (g subscript)
and terminating sensor (M subscript) the “load” (l subsc ript), and substituting in the definitions for cal
factor from earlier, we get the more general equation for transferring between a feed-through and a
terminating sensor:
=
|
1
|
Where:
Calibration factor of the Terminating Mount
k
M =
Calibration factor of the Feed-through Moun t
k
F =
= Power measured terminating mount
P
SubM
= Power measured Feed-through mount
P
SubF
= Gamma Correction full vecto r da t a Terminating Mount
Γ
Μ
= Gamma Correction full vector data Feed-through Mount
Γ
Φ
Now in this general equation, the Gamma terms are the reflection scattering parameter of the respective
port noted in the subscript. Gamma is a complex vector with scalar values denoting the real and
imaginary magnitudes:
= cos + isin
In the general transfer equation, the term,
match” term. Inside the absolute value brackets, however, is a vector subtraction. Expanding out to
make the angles explicit, this becomes:
|
1
cos
|
1 ΓΓ
(
+
|
is the scalar “gamma correction” or “port
)
sin
(
+
)|
(
1
cos
1
, or the “imaginary” component.
(
cos
)
+
+
cos
(
+
))
4-6
(
+
(
+
)
sin
+
(
+
sin
))
(
+
)
Where the i represents
The absolute value, or length of a vector, is g iven by the Pythagorean formula, which is the s quare
root of the square of the magnitudes of the real and imaginary components. It’s convenien t t hat we
are looking for the square of the magnitude, so we don’t have to worry about the square root part.
Our correction term becomes the scalar,
Noting that if we collect the two terms beginning with
(
(
+
The term, cos
becomes,
1 2
(
+
)
+ sin
(
cos
+
(
1 2
)
+
)
is always identically equal to 1, so the final simplified equation
+
cos
Or combining this result with the general transfer equation,
=
(
1 2
If we look at the scalar result of the mismatch ter m, the “1” part is what would happen if at least one
of the ports was “perfect”, or had no reflection. In that case, one of the is zero. The right-most
element has magnitude of
ignored for most connections.
The middle part, 2
multiplied by k, the sensitivity to this change is equal to k, or about 1. In an UNCORRECTED transfer,
this part represents the probable error of the transfer. Since we can’t know the angles in an
uncorrected transfer, we let cos() ta ke its limits of +/- 1, and say that the uncer t ainty of the
uncorrected transfer is 2
probable, but it’s what the industry usually does. Unless the rhos were measured on a scalar analyzer,
we have to use manufacturer’s worst-case values.
Some common power sensor calibration practices do not always use gamma correction. An argument
could probably be made that this was a reasonable practice at lower frequencies. We typically see this at
18 GHz and lower. A value of 0 would be inserted for Γ
, which is typically so much smaller than the middle ter m tha t it can b e
)
cos(+
. That is a little pessimistic, because that is worst-case rather than
, contains the bulk of the impact of port mismatch. Since it is
This would make the formula look like:
=
, we get,
(
cos
cos
Μ
+
)
+
(
+
making that portion of the formula “1”.
)
+ sin
)
+
(
+
)
))
This revision of t he original formula assumes many things that are not ne cessarily true. Things t hat may
not necessarily show up at lower frequencies but will certainly show up at higher frequencies where
connectors change from the very rugged N-type connector to the more sensitive 3.5 mm and 2.4 mm
connectors.
By looking at Figure 4.4 we can get a practical visualization o f the relationship between calibration
factor, gamma and effective efficiency
To ensure accurate measurements and optimal performance of TEGAM products, the microwave coaxial
cable assemblies used in system and test setups must be properly used and maintained. Proper
connections, rout ine i nspection of all cables, and cleaning of t he connectors are extremely important
procedures which can prolong the longevity and accuracy of equipment.
Cable Inspection
Routinely check external cables for signs of cracked insulation, dents, twists, flattening, signs of jacket
abrasion, or other sign s of abuse. Wri nkles in the j acket indi cate that the minimu m bend radi us has been
exceeded. Most often, this occurs near the marker tubes and connectors.
Also inspect the connector interfaces for the following:
• Bent pins (male).
• Bent or missing tines (female).
• Worn or chipped plating.
• Damaged or displaced dielectric inserts.
• Thread damage.
• Folded or mushroomed outer interface rims.
• Mushroomed pin shoulders (male) or tine ends
(female).
• Score lines on pins and outer interface rims visible to the unaided eye.
• Recessed or protruding pins.
It is advisable to clean the connectors prior to inspection to make subtle damage more apparent. If any of
the above is noted, repl ace the a ssemb ly b efore i ts fu rther us e resu lt s in equi pmen t dam age. Al so insp ect
the mating connectors for similar damage.
Inspect the connector interface for signs of debris.
Debris may be in the form of:
• Plating chips or ot her metal particles.
• Dust or dirt.
• Oily films.
• Other miscellaneous foreign particles.
Making Initial Connections
Exercise caution wh en mating cabl es. Poor connecti ons lead to poor system performan ce. They can al so
damage not only the cable assembly, but more significantly, front or rear panel connectors on the
equipment itself which may be more difficult to repair.
Aligning Connectors
Align the center lines of two connectors before actual mating. Male retaining nuts contain a small amount
of necessary play which may make it possible to mate the threads without the pins being properly aligned.
Pin misalignment can damage pins and dielectric inserts.
Mating Connectors
Gently mate the connectors by hand, taking care not to force the coupling nut at the slightest resistance.
It is often possible to feel whether or not the pins are mated. If the coupling nut is difficult to turn, either
the pins are not mated, the coupling nut is cross-threaded, or on e o f t h e con n e c tors has been da m ag ed b y
excess torque.
Never hold a male connector coupling nut stationary while screwing a female connector into it. This
rotation can erode the plating and damage both the outer interface rim as well as the pin. If the pins
become locked, serious damage ca n res ult to both the equipment and the cable assembly.
Over-Torque
Once connectors have been properly mated, apply only the proper amount of torque. Over-torque
damages both connectors involved. Also, a connector which has been damaged by over-torque, in turn,
damages every connector to which it is subsequently mated. It usually leads to poor system performance
as well. Over-torque can cause:
• Bent pins.
• Recessed or protruding pins.
• Recessed or protruding dielectrics.
• Chipped plating.
• Damaged coupling threads.
• Coupling nut retaining ring damage.
• Mushroomed outer interface shells.
• Mushroomed pin shoulders.
Proper Torque of a APC 2.4 mm Hex Nut
To mate a connector of the hex-nut type, always use a torque w rench set to the correct torqu e value.
Tighten the connector slowly until the wrench snaps. Tightening too quickly can cause the wrench to
exceed its set limit. Do not snap the wrench more tha n once as this also causes over-torque.
CLEANING CONNECTOR INTERFACES
Use the following guidelines in cleaning connector interfaces:
•Do not use chlorinated solvents including common tap water. These solvents are extremely
penetrating and sometimes ruin otherwise good devices and assemblies.
• Moisten a cotton swab with isopropyl alcohol. Roll the swab on a paper towel to remove excess.
• Use the moistened cotton swab to wipe away debris. Do not try to dissolve the debris by over-
wetting the swab.
•Repeat the cleaning process using additional swabs as necessary. If metallic particles are
embedded i n the di el ectri c, use a n eyegl a ss and a sh arp pick in an attempt to dislodge them. Swab
again.
•When satisfi ed that the interfaces a re clean , blow th em dry with dry compre ssed air, or p refera bly
dry nitrogen (pressurized spray cans work well). Do not use breath.
•Clean the mating connectors. These may be the source of the debris.
Use the information in this section if you need to return your Power Standard to TEGAM.
Package 2510A/1510A for Shipment
Use the following steps to package the Power Standard fo r sh ipment to TEGAM fo r se rv ice:
1. Fill the EXPEDITE REPAIR & CALIBRATION FORM (found in Section I) and attach it to the Power
Standard. Please be as specific as possible about the natur e of the problem. Send a copy of any or
all of the following information:
o Any error messages that appeared on the power meter display.
o Any information on the performance of the Power Standard.
2. Use the original packaging materials or a strong shipping container. The carton must be both large
enough and strong enough to accommodate the power standard and allow at least three to four
inches on all sides of the power standard for packing material.
3. Seal the shipping container securely with stro ng nylon adhesive tape.
4. Mark the shipping containe r “FRAGILE, HANDLE WI TH CARE” to ensure careful handling.
CAUTION: 2510A/1510A damage can result from using packaging materials other than those specified.
5. Retain copies of all shipping papers.
NOTE: Refer to page 1-6 of the introduction for information on how to obtain a RMA from TEGAM.
When an RF Power Standard is to be stored for extended periods, pack the instrument into a container.
Place container in a clean, dry, temperature-controlled location. If instrument is to be stored in excess of
90 days, place desiccant with items before sealing container. The safe environmental limits for storage are
-55° to +75° C (-67° to +167° F) at less than 95% non-condensing relative humidity.