Tegam 110A Instruction Manual

Page 1
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
110A
Safety Voltmeter
Instruction Manual
PN# 110A-901-01E
Publication Date: January 2018
REV. F
NOTE: This User’s Manual was as current as possible when this product was manufac tured. However , products are constantly being update d and improved. To ensure you have the latest documentation, refer to www.tegam.com
Page 2
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
110A
Page 3
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
TABLE OF CONTENTS
TABLE OF CONTENTS
1
INSTRUMENT DE
SCRIPT
IO
N
INTR
ODUCTION ...................................................... 1-1
SPECIFICATIONS .................................................... 1-
2
ACCESSORIES SHEET ................................................. 1-4
2
PREPARATION FOR USE
UNPACKING AND INSPECTION
................................. 2-1
PREPARATION FOR USE
............................................ 2-
1
OPERATION ........................................................... 2-1
OPERATING CONDITIONS ........................................ 2-2
FRONT PANEL ........................................................ 2-3
3
SAFETY AND MAINTENANCE
SAFETY SYMBOLS AND TERMS
................................. 3-1
OPTIONAL SAFETY TESTS ........................................ 3-3
MAINTENANCE INFORMATION .................................. 3-5
Test Lead Replacement ..................................... 3-5
FUNCTIONALITY CHECKS ........................................ 3-6
PERFORMANCE VERIFICATION ................................. 3-7
BATTERY INSTALLATION/REPLACEMENT .................... 3-8
CLEANING INSTRUCTIONS ...................................... 3-9
4
SERVICE INFORMATION
PREPARATION FOR CALIBRATION OR REPAIR ............ 4-1
EXPEDITE REPAIR & CALIBRATION FORM ................... 4-2
WARRANTY ............................................................ 4-3
Page 4
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-1
INSTRUMENT DESCRIPTION
SECTION 1
INSTRUMENT DESCRIPTION
INTRODUCTION
The Model 110A Safety Voltmeter is the only single switch, single range, digital voltmeter that automatically measures AC & DC volts to 1000 V for voltage troubleshooting, line clearing, lock out/tag out procedures and anywhere hazardous voltages may be present. The 110A has a 1 MΩ current limiting resistor in each lead to protect the user against arc-flash potential.
Model 110A meets Category IV 600 V and Category III 1000 V Overvoltage protection under IEC-61010-1.
SPECIFICATIONS
Page 5
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-2
INSTRUMENT DESCRIPTION
FUNCTIONS: VDC/VAC, with automatic selection. When both DC and AC input voltages are present, the voltage type with the higher magnitude is selected and displayed.
RANGES: ±1000 VDC . 1000 VAC (45-500 Hz, average responding, calibrated in RMS of a sine wave).
RESOLUTION: 1 V ACCURACY: ±(0.2% rdg + 1 volt), DC, 50-60 Hz, 64 to 82 °F. MAXIMUM ALLOWABLE INPUT: 1000 volts continuous. Overload
annunciator (OL) turns on above 999 volts. OVERVOLTAGE TEST: High Pot tested at 2500 VAC for < 5 sec. NORMAL MODE REJECTION: AC Rejection (VDC Mode): > 40 dB.
DC Rejection (VA C Mode): 40 dB typ. (for DC voltage < 10% peak AC voltage).
READING RATE: 2.5 readings/second INPUT RESISTANCE: 10 M(8 Min volt meter, 1 Mi n each
attached input probe) DISPLAY: 0.5” LCD (3 ½ digits), polarity, and annunciators for
VAC, VDC, OL, Low Battery. INPUT PROBES: Permanently attached to vol tmeter. S heathed tips
with retracting sleeves. PVC insulated leads, 36” long.
POWER: 9 V alkaline or carbon-zinc battery (NEDA 1604). BATTERY LIFE: 200 hours typical, alkaline.
ENVIRONMENTAL LIMITS FOR OPERATING: –10 to 150 °F, less
Page 6
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-3
INSTRUMENT DESCRIPTION
than 80% R.H. up to 95 °F. Reduce R.H. li mit by 1.7% per °F abov e 95 °F.
ENVIRONMENTAL LIMITS FOR STORAGE: –30 to 150 °F, less than 90% R.H. up to 95 °F. Reduce R H limit by 1.7% per °F above 95 °F.
ENVIRONMENTAL/TIME LIMITS TO ACCURACY: 64° to 82 °F, 80% R.H., 1 year.
TEMPERATURE COEFFICIENT: From 64 °F to 82 °F; included in accuracy specifications. Below 64 °F and above 82 °F; less than
0.05 times applicable specifications per °F. LOW-BATTERY INDICATOR: Display indicates ‘BAT’ when less
than 10% of life remains. AUTOMATIC TURN-OFF: Voltmeter turns off after 10 minutes of
operation. Turns off below 6.5 V battery voltage.
SIZE: 6.3” x 2.7” x 1.2” WEIGHT: 12 oz. CONSTRUCTION: Heavy-duty ABS plastic housing. INCLUDED ACCESSORIES: Battery, Operating Instructions.
ACCESSORIES SHEET
Page 7
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-4
INSTRUMENT DESCRIPTION
1104 - Leather Utilit y Belt Carrying Case
The 1104 case protects your voltmeter and probes from broken
displays and other damages that can occur. This case has a 2 in. belt loop. It is long lasting, super durable, and is made from top-grain cowhide with riveted sea ms .
1204 - Cordura Nylon lineman’s Carrying Cas e
The 1204 case prote cts your voltme ter and saves you time beca use th e meter can stay in the case during use. It features a window to access the ON/OFF switch, a ret ainer that holds the meter in place, spac e for probe storage and a 22 in. shoulder strap. It also comes with a belt clip and can be worn on a utility belt. Size- 5 ½ x 7
½ x 1 ½ in. (W x H x
D).
Page 8
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
1-5
INSTRUMENT DESCRIPTION
12501 - Alligator Clip Adapter
The 12501 probe adapter kit allows you to convert your test leads to alligator clips. The kit includes two screw-on probe adapters and two heavy duty screw in alligator clips with insulating rubber boots. These alligator clips open to ½ in. and fit most ter minals and bus bars.
12502 - Universal Adapter Kit
The 12502 probe adapt er kit includ es two screw -on probe adapters and interchangeable screw-in accessories including two alligator clips with insulating rubber boots, two banana plugs, two needle tips, two spade
lugs and two heavy duty tips. With this kit you can adapt your
voltmeter to the many different situations you encounter on the job.
PN 110-404
Replacement probe/test-lead set (1 red, 1 black) for Model 110A.
Page 9
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
2-1
PREPARATION FOR USE
SECTION 2
PREPARATION FOR USE
UNPACKING AND INSPECTION
Each instrument is inspected both mechanically and electrically before shipment. Upon receiving your instrument unpack all items from the shipping container and check for any obvi ous damage that may have occurred during transit. Report any damage to the shipping agent. Retain and use the original packing materials if
reshipment is necessary.
PREPARATION FOR U SE
Each instrument is supplied with a 9 V battery. See page 3-8 of thi s manual for battery installation instructions.
OPERATION
The Model 110A digital voltmeter (DVM) is easy to use. There is only one control (located on the side of the instrument) that turns the instrument ON/OFF (Figure 1).
After turn-on, the instrument automatically selects either VDC or VAC function. VDC is selected if the magnitude of a DC input voltage is greater than the peak value of any AC voltage present. Otherwise the VAC function is enabled. VAC and VDC annunciators in the liquid crystal display indicate the selected function. Observe
the maximum allowable input (1000 V continuous).
After approximately 10 minutes, the voltmeter turns off automatically to conserve battery life. To turn the voltmeter on
Page 10
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
2-2
PREPARATION FOR USE
again, return the ON/OFF switch to the OFF position, and then back
to the ON position.
When battery voltage drops below 6.5 V, the voltmeter turns off until a new battery is installed (Refer to Battery Installation Section).
OPERATING CONDITIONS
In accordance with the Standard for Electrical Measuring and Test Equipment IEC-61010-1, the TEGAM Model 110A Safety Voltmeter is designed for safe operation under the following conditions:
• Indoor use, or outdoor use in where the instrument and probes remain dry (i.e. protected from exposure to rain, fluids, or condensing humidity).
• Altitudes up to 2000 m above mean sea-level.
• Temperature: –23 °C to 65 °C
• Relative humidity: 80% R.H. up to 35 °C, decreasing linearly to 0% R.H. at 60 °C.
• Power source: 9 V transistor battery (NEDA 1604), alkaline or carbon-zinc type. The voltmeter automatically turns off before battery discharge can cause malfunction.
• Pollution: Degree 2 per IEC 664.
• Installation Category IV (up to 600 V) / Category III (up to 1000 V)
Page 11
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
2-3
PREPARATION FOR USE
Figure 1: Model 110A Front Panel
Minus Displayed
(Positive Polarity Implied)
Liquid Crystal
3 ½ Digital Display
Low Battery Indicator
AC Volts Annunciator
DC Volts Annunciator
Overload Indicator
(Input > 999 Volts)
ON/OFF Switch
(Automatic turn-off after 10 minutes) (Automatic turn-off when battery voltage < 6.5 volts)
Probe
Holder
Color
Coded Probes Red (+)
Black (-)
Retractable
Probe Sheaths
NOTE:
In operate position, install probe with “TEGAM” marking facing to right. To store probe, remove from probe-holder, rotate “TEGAM” marking toward rear, and reinsert
Page 12
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-1
SAFETY AND MAINTENANCE
SECTION 3
SAFETY AND MAINTENANCE
SAFETY SYMBOLS AND TERMS
The symbol on the instrument denotes that the user should
refer to the operating instructions.
The symbol on the instrument denotes that 750 V or more may be present on the terminal(s).
The WARNING used in this manual explains dangers that could result in personal injury or death.
The CAUTION used in this manual explains hazards that could damage the instrument.
SAFETY PRECAUTIONS
Fundamental safe work practices recommend testing any voltmeter on a known low-energy source before and after use.
WARNING
The following safety precautions should be observed before
operating the Model 110A DVM.
Les précautions de sécurité suivantes doivent être respectées avant
d'utiliser le modèle 110A DVM.
1.This instrument is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read over the
Page 13
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-2
SAFETY AND MAINTENANCE
manual carefully before operating this instrument.
2. Exercise extreme caution when a shock hazard is present at the instrument’s input. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30 Vrms or 42.4 Vpeak are present. A good safety practice is to expect that a hazardous voltage is present in any unknown circuit before measuring.
3. Inspect the test leads for possible wear, cracks or breaks before each use. If any defects are found, replace immediately with P/N 110-404 probe/test-lead set. Refer to Maintenance Information section.
4. For optimum safety do not touch the test leads or the instrument while power is applied to the circuit under test. Turn the power off and discharge all capacitors, before connecting or disconnecting the instrument.
5. Do not touch any objects which could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface, capable of withstanding the voltage being measured.
6. Exercise extreme caution when testing high energy power circuits (AC power lines, etc.).
7. Do not exceed the instrument’s maximum allowable input as defined in the specifications and printed on the front panel of the instrument.
Page 14
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-3
SAFETY AND MAINTENANCE
OPTIONAL SAFETY TESTS
TEGAM, Inc. recommends periodic safety inspections for all high voltage equipment. With the Model 110A, users may wish to check probe resistance and instrument over-voltage capability in order to find safety hazards (actual or potential) due to product damage and/or wear.
WARNING
The information presented in this section is intended for
use by qualified personnel only. To reduce the risk of
electric shock, do not perform any servicing unless you are
qualified to do so.
AVERTISSEMENT
L'information présentée dans cette section est destiné à
être utilisé par du personnel qualifié. Pour réduire le risque
de choc électrique, ne p a s ef fectuer l'entretien, sauf si vous
êtes qualifié pour le f a ire.
WARNING
Disconnect probes from external circuits before performing
safety tests.
AVERTISSEMENT
Débrancher les sondes de circuits externes avant d'effectuer les
tests de sécurité.
Probe Resistance Check
Open the DVM case (see Battery Installation/Replacement section) and measure the resistance between each probe tip and the circuit­board eyelet terminating the respect ive probe lead-wire. Each probe should measure 1 MΩ ± 1%. An excessively hi gh or low resistance measurement indicates probe failure due to electrical overload or
Page 15
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-4
SAFETY AND MAINTENANCE
mechanical wear. Replace defective probes immediately. (Refer to Maintenance Information section.) Close DVM case.
Differential HI-POT Test
Apply a momentary (< 5 seconds) 2500 VAC hi-pot between the tips of the two test probes. There should be no voltage breakdown across the probes or instrument. Units failing this test should be returned to TEGAM for diagnosis and repair.
Common Mode HI-POT Test
Wrap the test leads and the instrument in a single sheet of aluminum foil. About 1” -2” of the tips of each probe should extend outside the foil wrap. Apply a momentary (< 5 seconds) 2500 VAC hi-pot between one probe tip and the foil wrap. No breakdown should be observed. Repeat this test for the other probe tip. Units failing this test should be returned to TEGAM for diagnosis and repair.
Page 16
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-5
SAFETY AND MAINTENANCE
MAINTENANCE INFORMATION
This section contains information needed to maintain your instrument. The following information is included: probe replacement procedure, functionality checks, performance
verification, and battery installation/replacement.
WARNING
The information presented in this section is intended for
use by qualified personnel only. To reduce risk of electric
shock, do not perform any servicing unless you are
qualified to do so.
AVERTISSEMENT
L'information présentée dans cette section est destiné à
être utilisé par du personnel qualifié. Pour réduire les
risques de choc électrique, ne pas effectuer l'entretien,
sauf si vous êtes qualifié pour le faire.
Test Lead Replacement
WARNING
Replacing test leads with anything other than a TEGAM P/N
110-404 probe/test-lead set will degrade accuracy, reduce
meter overload protection, and pose a serious safety
hazard to the user.
AVERTISSEMENT
Remplacement essai conduit avec d'autres d'une TEGAM P
/ N 110-404 sonde / test-tête ensemble tout va dégrader
la précision, réduire mètre protection contre les
surcharges, et poser un risque grave pour la sécurité de
l'utilisateur.
Page 17
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-6
SAFETY AND MAINTENANCE
WARNING
Disconnect probes from external circuits and turn the
instrument off before removing the bottom cover. Reinstall
the cover before resuming use of the instrument.
AVERTISSEMENT
Débrancher les sondes de circuits externes et éteindre
l'appareil avant de retirer le couvercle inférieur. Remettre le
couvercle avant de reprendre l'utilisation de l'instrument.
Open the voltmeter case (see Battery Installation/Replacement section), and cut the (2) strain relief cable ties that secure the test­leads to the circuit board. Next, remove the circuit board from the top case by unscrewing the hexagonal spacer in the center of the board. Use a small soldering iron (< 75 W) to remove the test-leads from the circuit board eyelets. Pull the old probe/test-lead set f ree of the voltmeter case.
To install the new lead set, push the leads through the top case holes, and solder into the appropriate eyelets (red-probe connects to +eyelet, black-probe connects to –eyelet). To secure the test leads, use the cable ties supplied with the replacement lead set. Allow a lead length of 23-28 mm (0.9-1.1 inches) between the circuit-board eyelets and the respective cable ties. Use a cable-tie tool with a tension setting of 35-55 N (8-12 lb.) to install the cables
ties.
Reinstall the circuit board into the top case. Check that the fish paper barrier at the switch opening is securely attached to th e top case. Before the bottom case is reattached, be sure that the switch actuator cover is in place.
Confirm meter operation (refer to Performance Verification, page 3-
7).
Page 18
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-7
SAFETY AND MAINTENANCE
FUNCTIONALITY CHECKS
1. ZERO Switch on the DVM, and touch the probe tips together. Meter should
read 0 VDC or 0 VAC. An occasional reading flicker to 1 V is normal.
2. DC VOLTS Touch the black probe to the negative terminal of a fresh 9 volt
transistor battery (alkaline or carbon-zinc), and touch the red probe to the positive terminal. A reading of 9-10 VDC should be displ ayed. Reverse the probes. The same reading except for a minus polarity
should be displayed.
3. AC VOLTS
Connect the DVM probes to the hot and neutral outlets of a standard 120 VAC power receptacle. Under normal line conditions, a reading of 120 ± 10 VAC should be display ed.
PERFORMANCE VER I FIC ATI O N
Equipment needed:
• 750 VDC voltage source with 0.05% accuracy or better
• 750 VAC, 60 Hz voltage source with 0.05% accuracy or better
• Ohmmeter with 0.1% accuracy or better at 10 M.
Turn on the DVM, and use the voltage sources and the following table to verify that readings are within specifications.
Input
Allowable Reading
Shorted
0 ± 1
+750 VDC
750 ± 3 VDC
–750 VDC
–750 ± 3 VDC
750 VAC
750 ± 3 VAC
Page 19
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-8
SAFETY AND MAINTENANCE
TABLE 1: Verification Summary
Keep the DVM turned on, and measure the resistance between the probe tips. A measurement of 9.89-10.11 MΩ confirms integrity of the high voltage resistors inside the probes and meter.
If DVM readings and input resistance are out of speci fication, check the probes further for possible failure. Refer to Probe Resistance Check section.
Page 20
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-9
SAFETY AND MAINTENANCE
BATTERY INSTALLATION/REPLACEMENT
WARNING
Disconnect probes from external circuits and turn the
instrument off before removing the bottom cover. Reinstall
the cover before resuming use of the instrument.
AVERTISSEMENT
Débrancher les sondes de circuits externes et éteindre l'appareil avant
de retirer le couvercle inférieur. Rem et t re le couvercle avant de
reprendre l'utilisation de l'instrument.
1.Place the unit facing down on a bench or other similar surface and remove the screws from the bottom cover.
2.Separate the bottom cover from the rest of the instrument by grasping the top of the case (just above display) and carefully lifting it away from the display.
BATTERY TERMINALS
BATTERY
Figure 2: Battery Installation
3. Remove the old battery.
4. Place the new battery in the battery compartment. Be sure to
Page 21
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-10
SAFETY AND MAINTENANCE
observe the proper polarity (refer to Figure 2).
5. Reinstall the bottom cover before resuming use of the instrument.
CLEANING INSTRUCTIONS
Before cleaning a Model 110A voltmeter, disconnect the voltmeter probes from electrical sources, and turn off the voltmeter. Use a soft non-abrasive cloth dampened with a water-detergent solution to clean outside surfaces of the voltmeter. Squeeze excess moisture from the cloth so that no liquid residue is left on the voltmeter. Do not allow moisture to accumulate on any surface or penetrate case openings or crevices. Do not attempt to clean the inside of the voltmeter. Allow 24 hours for the voltmeter to dry at room­temperature before resuming operation.
Page 22
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
3-11
SAFETY AND MAINTENANCE
Page 23
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
4-1
SERVICE INFORMATION
SECTION 4
SERVICE INFORMATION
PREPARATION FOR CALIBRATION OR RE P AI R SE RVICE
Once you have verified that the cause for 110A malfunction cannot be solved in the field and the need for repair and calibration service arises, contact TEGAM customer service to obtain an RMA, (Returned Material Authorization), number. You can contact TEGAM customer service via the TEGAM website, www.tegam.com
or by calling 440.466.6100 (All Locations) or 800.666.1010 (United States Only).
The RMA number is unique to your instrument and will help us identify you instrument and to address the particular service request by you which is assigned to that RMA number.
Of even greater importance, a detailed written description of the problem should be attached to the instrument. Many times repair turnaround is unnecessarily delayed due to a lack of repair instructions or of a detailed description of the problem.
This description should include information such as measurement range, and other instrument settings, type of components being tested, are the symptoms intermittent?, conditi ons that may cause the symptoms, has anything changed since the last time the instrument was used?, etc. Any detailed information provided to our technicians will assist them in identifying and correcting the problem in the quickest possible manner. Use a copy of the Repai r and Calibration Service form provided on the next page.
Once this information is prepared and sent with the instrument to our service department, we will do our part in making sure that you receive the best possible customer service and turnaround time possible.
Page 24
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
4-2
SERVICE INFORMATION
EXPEDITE REPAIR & CALIBRATION FORM
Use this form to provide additional repair information and service instructions. The Completion of this form and including it with your
instrument will expedite the processing and repair process.
RMA#:
Instrument Model #:
Serial Number:
Company:
Technical Contact:
Phone Number:
Additional
Contact Info:
Repair Instructions:
Evaluation Calibration Only Repair Only
Repair & Calibration Z540 (Extra Charge)
Detailed Symptoms:
Include information such as measurement range, instrument settings, type of components being tested, is the problem intermittent? When is the problem most frequent?, Has anything changed with the application since the last time the instrument was used?, etc.
Page 25
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
4-3
SERVICE INFORMATION
WARRANTY
TEGAM, Inc. warrants this product to be free from defects in material and workmanship for a period of one year from the date of shipment. During this warranty period, if a product proves to be defective, TEGAM Inc., at its option, will either repair the defective product without charge for parts and labor, or exchange any product that proves to be defective.
TEGAM, Inc. warrants the calibration of this product for a period of one year from date of shipment. During this period, TEGAM, Inc. will recalibrate any product, which does not conform to the published accuracy specifications.
In order to exercise this warranty, TEGAM, Inc., must be notified of the defective product before the expiration of the warranty period. The customer shall be responsible for packaging and shipping the product to the designated TEGAM service center with shipping charges prepaid. TEGAM Inc. shall pay for the return of the product to the customer if the shipment is to a location within the country in which the TEGAM service center is located. The customer shall be responsible for paying all shipping, duties, taxes, and additional costs if the product is transported to any other locations. Repaired products are warranted for the remaining balance of the original warranty, or 90 days, whichever period is longer.
Page 26
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
4-4
SERVICE INFORMATION
Warranty Limitations
The TEGAM, Inc. warranty does not apply to defects resulting from unauthorized modification or misuse of the product or any part. This
warranty does not apply to fuses, batteries, or damage to the instrument caused by battery leakage.
The foregoing warranty of TEGAM is in lieu of all other warranties, expressed or implied. TEGAM specifically disclaims any implied warranties of merchantability or fitness for a particul ar purpose. In no event will TEGAM be liable for special or consequential damages. Purchaser’s sole and exclusive remedy in the event any item fails to comply with the foregoing express warranty of TEGAM shall be to return the item to TEGAM; shipping charges prepaid and at the option of TEGAM obtain a replacement item or a refund of the purchase price.
Statement of Calibration
This instrument has been inspected and tested in accordance with specifications published by TEGAM Inc. The accuracy and cal ibration of this instrument are traceable to the National Institute of Standards and Technology through equipment, which is calibrated at planned intervals by comparison to certified standards maintained in the laboratories of TEGAM Inc.
Contact Information
TEGAM INC.
10, TEGAM WAY
GENEVA, OHIO 44041
WEB: http://www.tegam.com
Page 27
10 TEGAM WAY • GENEVA, OHIO 44041 • 440-466-6100 • FAX 440-466-6110
sales@tegam.com
4-5
SERVICE INFORMATION