Tecno Control TS220EO User Manual

IST-2220.EO03.01/A
(
IST-2220.EO03.01
-
A_TS220EO (30.04.2014).docx
)
Caratteristiche tecniche /
Technical specifications
/
Caractéristiques techniques
TS220EO
Istruzione / User’s Manual / Manuel d’utilisation Pag.1/9
Trasmettitore 4÷20mA di Ossigeno 4÷20mA Oxygen Gas Detector Détecteur 4÷20mA pour gaz Oxygéne
Leggere Attentamente e Conservare questa Istruzione
Please read and keep this manual
Lire avec soin et garder la notice d’istruction
Con Cartuccia Sensore Sostituibile
Inside Replaceable Cartridge Sensor
Avec cartouche capteur échangeable
Modello / Model / Modele Cartuccia Sensore / Cartridge Sensor / Cartouche Capteur
TS220EO ZS EO1
Alimentazione / Power supply / Alimentation
Sensore / Sensor Type / Capteur Elettrochimico / Electrochemical / électrochimique Cartuccia Sensore / Cartridge Sensor / Cartouche capteur Sostituibile / Replaceable / échangeable Uscite / Outputs / Sortie 4 ÷ 20 mA lineare / Linear / linéaire Campo di misura / Standard Range / Champ de mesure Tempo di risposta / Response Time / Temps de réponse T90< 20 secondi / seconds / secondes Ripetibilità / Repeatability / Répétitivité ±1% del segnale / signal Precisione / Accuracy / Precision Deriva a lungo termine Long time drift / Dérive à long terme Temp./umidità di funzionamento / Operation Temp./Humidity Température et hygrométrie de fonctionnement Pressione di funzionamento Operation Pressure / Pression de fonctionnement Limite del Sensore / Sensor limits / Limite échelle Vita media in aria pulita Average Life in fresh air / Vie moyenne en air pur Tempo massimo di immagazzinamento Max Storage Time / Temps maximum de stockage Temperatura / Umidità di immagazzinamento/ Storage Temp-Humidity / Temp. et hygrométrie de stockage Grado di protezione / IP Code / Indice de protection
Dimensioni / Size / Dimensions du boîtier
IT
NOTE SUL GAS RILEVATO .................................................................................................................. 2
FUNZIONAMENTO ................................................................................................................................. 2
INSTALLAZIONE .................................................................................................................................... 2
AVVERTENZE ........................................................................................................................................ 3
VERIFICHE E CALIBRAZIONE .............................................................................................................. 3
EN
NOTES ON THE DETECTED GAS ........................................................................................................ 4
OPERATIONAL DESCRIPTION ............................................................................................................. 4
INSTALLATION ...................................................................................................................................... 5
WARNING ............................................................................................................................................... 5
TEST and CALIBRATION ...................................................................................................................... 5
FR
CARACTÉRISTIQUES DU GAZ ............................................................................................................. 6
FONCTIONNEMENT............................................................................................................................... 6
INSTALLATION ...................................................................................................................................... 7
INSTRUCTIONS ..................................................................................................................................... 7
VÉRIFICATIONS È ETALONNAGE ....................................................................................................... 8
DESCRIZIONE
DESCRIPTION
DESCRIPTION
12÷24Vcc (-10/+15%) 2W
12÷24Vdc(-10/+15%) 2W / 12÷24Vcc(-10/+15%) 2W
0 ÷ 25,0 % O2
± 2 % del segnale in aria / signal in air / du signal en air
<2% segnale mese / signal per month / du signal chaque mois
<5% sulla vita operativa / over Operating Life / sur sa vie
-20 ÷ + 50 °C / 15÷90 % RH
non condensata / non condensed / non condensée
Atmosferica ±10%
Atmospheric±10% / Atmosphérique ±10%
30 % O2
2 anni / years / ans
3 mesi / 3 month / 3 mois
0 ÷ + 20°C / 5 ÷ 95 % RH
non condensata / non condensed / non condensée
IP65
190 x 105 x 83 mm
........................................................................................................................... 2
........................................................................................................................... 4
........................................................................................................................... 6
TECNOCONTROL S.r.l. - Via Miglioli, 47 20090 SEGRATE (MI) ITALY Tel. +39 02 26922890 - Fax +39 02 2133734
http: www.tecnocontrol.it e-mail: info@tecnocontrol.it
Tel.
+39 02
26922890
- Fax
+39 02
2133734
IST-2220.EO03.01/A Istruzione / User’s Manual / Manuel d’utilisation Pag.2/9
DESCRIZIONE
IT
Il TS220EO è un trasmettitore 4÷20mA di Ossigeno con sensore a cella elettrochimica, utilizzato in sistemi centra­lizzati d’allarme per laboratori, industrie e ambienti da proteggere dalla possibile assenza o eccesso di O2 o pre­senza di gas inerti. Il rilevatore è costituito da una custodia che contiene il circuito elettronico e i morsetti di colle­gamento. Nel Portasensore, posto nella parte inferiore della custodia, è alloggiata la “Cartuccia Sensore Sostituibi- le” che contiene l’elemento sensibile e i dati identificativi e di taratura.
Il TS220EO ha un segnale d’uscita (S) 4÷20mA lineare con Fondo Scala di 25,0% O2. Questa uscita va collegata una centrale rilevazione gas. (Vedi Tabella 1). Sul coperchio sono visibili i tasti F1 e F2 per le operazioni di verifica e calibrazione utilizzabili solo tramite codice e i 3 Led che indicano le condizioni d’esercizio:
Led Rosso "ALARM": Segnalazione ottica d’allarme (22,5% O2). Led Verde "ON": Funzionamento normale. Led Giallo "FAULT": Sensore guasto o scollegato o a fondo scala o scaduto.
NOTE SUL GAS RILEVATO
L’ Ossigeno (O2) è un gas inodore, incolore, leggermente più pesante dell'aria (ha densità, relativa all'aria di 1,1). Non è un gas infiammabile ma essendo un forte ossidante reagisce con materiali combustibili e riducenti determi­nando rischi di incendio ed esplosione in quanto ne facilita la combustione.
L’Ossigeno è il 20,95% dell’aria che respiriamo. Se la concentrazione di O2 nell’aria scende sotto il 19,5% si consi­dera carente per la salute nelle persone, se poi la concentrazione scende sotto al 16%, causa problemi respiratori fino a provocare l’asfisia e sotto al 6% causa la morte immediata. Tenere presente che la carenza di ossigeno può essere provocata, da fenomeni di combustione, dall’aumento nell’aria di altri gas (tossici e/o infiammabili) ma an­che da perdite di gas inerti (Azoto, Elio, Argon etc.). A concentrazionei elevate, oltre il 24% l’O2 è irritante per il trat­to respiratorio e può causare effetti sul sistema nervoso centrale, sugli occhi e sui polmoni che possono essere danneggiati dall'inalazione di alte concentrazioni..
FUNZIONAMENTO
Il sensore elettrochimico utilizzato è compensato in temperatura, ma è sensibile alle variazioni d’umidità estreme. Preriscaldo: quando il trasmettitore è alimentato inizia la fase di stabilizzazione del sensore, segnalata dal lam­peggio del Led giallo "FAULT". Dopo circa 60 secondi, il Led giallo si spegne e si accende il Led verde “ON”, che in­dica il normale funzionamento. Dopo questo tempo il sensore è in grado di rilevare ossigeno, ma raggiunge le con­dizioni di stabilità ottimali dopo circa 15 minuti di funzionamento continuo. Funzionamento Normale: deve essere acceso il solo Led Verde (ON).
Il Led Rosso (ALARM) si accende, se la concentrazione di Gas O2 supera il 22,5% oppure scende sotto il 19,5%
(solo se è stato abilitato con il Dip-Switch n.1 posizionato su ON). I Guasti: possibili, sotto elencati, sono indicati dall’accensione del Led giallo (FAULT) e portando l’uscita “S” a 0mA.
Il Led Giallo si accende ogni 4 secondi (con il Led Verde acceso): per avvisare che la “Cartuccia Sensore” ha su-
perato il suo limite teorico di vita e non è più garantito il corretto funzionamento. Il rilevatore continua a funzionare normalmente, ma è necessario, al più presto, sostituire la “Cartuccia Sensore” con una nuova, il tipo da richiedere è indicato a Pag. 1. La procedura di sostituzione è descritta nella documentazione ad essa allegata.
Se il Led Giallo è acceso e il Verde è spento (uscita 0mA): indica più possibilità di guasto. 1) la configurazione dei Dip-Switch non è corretta, verificarne la posizione (Vedi Tabella 2). 2) la “Cartuccia Sensore” è guasta, sostituirla con una nuova. 3) se è installata una “Cartuccia” nuova, o non è collegata correttamente o non è stata montata quella compatibile. Controllare le connessioni con la cartuccia e la compatibilità (vedi a Pag. 1). Eseguite le verifiche spegnere e riaccendere l’apparecchio. Se la condizione persiste sarà necessario sostituire e/o inviare il rilevatore al fornitore per la riparazione.
Se i Led Giallo e il Verde sono accesi (uscita 0mA): indica probabilmente il guasto della “Cartuccia Sensore”. Prima, provare ad eseguire la “Calibrazione in Aria” come descritto nella sezione “Verifiche e Calibrazione”, poi spegnere e riaccendere l’apparecchio ed infine provare a sostituire la “Cartuccia Sensore”. Se la condizione persi­ste sarà necessario sostituire e/o inviare il rilevatore al fornitore per la riparazione.
Se è stato abilitato il Led Rosso e quindi tutti i Led sono accesi (uscita >24,0mA): indica, o il guasto della “Car- tuccia Sensore”, oppure una concentrazione di gas superiore al F.S. (>30% O2). Se non è un allarme reale e la
condizione persiste anche dopo la sostituzione della “Cartuccia” sarà necessario inviare il rilevatore al fornitore per la riparazione.
INSTALLAZIONE
I Trasmettitori vanno installati e posizionati seguendo tutte le norme nazionali vigenti per gli impianti elettrici e le
norme di sicurezza degli impianti.
Posizione: in Fig. 1 sono indicate le dimensioni. Il trasmettitore va installato verticale con il sensore rivolto verso il
basso e va fissato a circa 1,60 m dal pavimento (il gas O2 è circa come l'aria). Collegamenti elettrici (fig. 2): Va utilizzato un cavo schermato a 3 conduttori. La calza va collegata a massa solo dal lato Centrale. la distanza massima, cui può essere installato ogni singolo rivelatore dall'alimentatore, è indicato
Tabella 1 in funzione della sezione del cavo utilizzato e della Centrale Gas utilizzata.
nella Il morsetto d’alimentazione, ( + - S) è ad innesto, ed è necessario sfilarlo per effettuare i collegamenti. Prestare at­tenzione nel reinserirlo dato che è polarizzato.
Nota: I Dip-Switch vanno posizionati prima d’alimentare l’apparecchio. Se si utilizza l’apparecchio con il Dip-
Switch 1 su ON, sarà attivato il funzionamento del Led Rosso d’Allarme (vedi Tabella 1).
TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI )
Tel.
+39 02
26922890
- Fax
+39 02
2133734
IST-2220.EO03.01/A Istruzione / User’s Manual / Manuel d’utilisation Pag.3/9
Importante: terminata l’installazione, per adattare il sensore alle condizioni ambientali, alimentare l’apparecchio,
attendere circa 15÷20 minuti e poi eseguire la “ Calibrazione in Aria (vedi “Verifiche e Calibrazione”).
COMPATIBILITÀ CON ALTRE CENTRALI: Se non si usa una Centrale Gas Tecnocontrol, calcolare la resistenza di carico massima utilizzando il grafico in Fig. 4. La resistenza di carico massima con alimentazione 12Vcc è 50 ohm, mentre con 24Vcc è 400 ohm. Si consiglia di utilizzare un’alimentazione a 24Vcc.
AVVERTENZE
La vita utile del sensore in aria pulita è mediamente 2 anni. Al termine di questo periodo, indicato dallo stru-
mento con un lampeggio del Led Giallo ogni 4 secondi, è necessario sostituire la “Cartuccia Sensore”.
Verifiche Periodiche: si consiglia di eseguire almeno ogni 6÷12 mesi la verifica di funzionamento del rilevatore,
come descritto nel capitolo “Verifiche e Calibrazione”.
Nota: lo strumento non è in grado di rivelare concentrazioni di O2 che avvengono fuori del locale in cui è installato. Limiti di funzionamento: Temperature inferiori a -40°C e superiori a +60°C provocano la rottura del Sensore.
Può danneggiarsi, anche se è esposto a concentrazioni di O2 molto elevate (oltre i 30%). In questi casi il sensore deve essere sostituito. Attenzione: i sensori elettrochimici per Ossigeno contengono una piccola quantità di acetato di potassio, che è corrosivo. Normalmente non si verificano perdite dal sensore, questo può essere causato solo da un'azione mec­canica violenta o un uso oltre i limiti di funzionamento. Il sensore contiene anche piccolo quantità di piombo, ossido di piombo, platino, argento, carbonio e antimonio. Nel caso avvenga un contatto del liquido del sensore, con la pel­le o gli occhi lavare immediatamente con abbondante acqua.
Interferenze con altri gas: La cella elettrochimica utilizzata, ha una buona resistenza ai prodotti d'uso comune
quali spray, ammoniaca, colle o vernici. Questi, però in quantità elevata, potrebbero interferire con il sensore; si consiglia di ventilare il locale, quando s’utilizzano questi prodotti. Considerare che in ambienti particolarmente in- quinati con vapori di solventi organici in quantità molto elevate, possono avvenire falsi allarmi e il sensore deve poi rimanere molto tempo in aria pulita per riportarsi nelle condizioni di funzionamento normale.
VERIFICHE E CALIBRAZIONE
Nota Importante: Tutte le seguenti operazioni vanno eseguite solo da personale esperto e addestrato, in quanto
l’uscita in mA cambiando di valore, provoca l’attivazione dei dispositivi d’allarme della Centrale cui è collegata.
Test Elettrico, Calibrazione in Aria e Regolazione Fine Uscita in mA: per accedere a queste due funzioni è necessario inserire il relativo “Codice” tramite i pulsanti F1 e F2. Per far sì che la pressione su un pulsante sia rico­nosciuta, tenerlo premuto per circa un secondo (finche non si spegne per un attimo il Led Verde). Dopodichè si può passare al pulsante successivo. In caso d’errore basta aspettare circa 10 secondi e la sequenza è automaticamen­te cancellata.
Kit di Taratura e Bombole con Miscela Gas: per la Verifica la miscela da utilizzare è Gas O2 a circa 15%v/v in azoto. È possibile usare bombole ricaricabili ad alta pressione con riduttore di pressione. Inoltre è necessario il kit
di calibrazione Tecnocontrol TC011.
“TEST ELETTRICO”(Codice Test: F2, F2, F1, F1): questa funzione permette di effettuare un test funzionale
dell’apparecchiatura. Dopo aver messo l’impianto in sicurezza ed aver inserito il “Codice Test”, si spengono tutti Led. Quindi si accenderanno in sequenza, i Led, dal Giallo fino al Rosso. L’uscita 4÷20mA rimane invariata. Alla fine, tutti i Led rimarranno accesi per circa 5 secondi, poi la centralina tornerà nelle condizioni di funzionamento normale. È consigliabile eseguire quest’operazione ogni 6-12 mesi in base all’utilizzo.
Nota: Non è possibile eseguire l’operazione se è acceso il Led Rosso.
“CALIBRAZIONE IN ARIA”( Codice: F2, F2, F2, F1, F2, F1): questa funzione permette di tarare il sensore in Aria
a 20,9%. Questa procedura quando viene avviata, procederà automaticamente fino al termine.
AVVISO: per garantire la sicurezza che non avvengano errori d’elaborazione, esiste la rara possibilità che durante la Calibrazione, il Led Giallo si spenga ogni 8 secondi, in questo caso interrompere la procedura, spegnere e riac­cendere l’apparecchio. Ripetere la Calibrazione, se la condizione persiste sarà necessario inviare il rilevatore al fornitore per la riparazione.
Attenzione: Durante la Calibrazione, l’uscita in mA diventerà 0mA.
La Taratura va eseguita esclusivamente in aria pulita (ambiente senza la presenza di gas tossici o altri inquinanti). Con i tasti eseguire il “Codice Calibrazione”. ”. Attendere finché si accendano i Led Giallo e Verde fissi ed il Led Rosso cominci a lampeggiare (il Led Rosso rimarra poi acceso fisso per indicare che la procedura è in funzione).
Attendere che il 1° Led Rosso si spenga (per almeno 2 secondi). A questo punto si possono verificare due casi:
Led Giallo e Verde accesi: la calibrazione è stata eseguita correttamente, dopo 8 secondi l’apparecchio si spe­gne e si riavvia automaticamente in funzionamento normale (vedi capitolo FUNZIONAMENTO “ Preriscaldo”).
Led Giallo acceso: la calibrazione è fallita. Dopo 8 secondi l’apparecchio si spegne e si riavvia automaticamente e ripete la procedura senza reinserire la sequenza. Se la condizione persiste anche dopo la sostituzione della car­tuccia sarà necessario inviare il rilevatore al fornitore per la riparazione.
Importante: Se l’ambiente fosse molto inquinato e ci fosse il dubbio che la concentrazione di Ossigeno non sia 20,9% è possible utilizzare una bombola di Aria con il 20,9% di O2 oppure utilizzare un palloncino per gas (nostro codice PA020) riempito di aria pulita. In questo caso, prima di eseguire il “Codice Calibrazione in Aria” infilare il TC011 sul portasensore, regolare l’afflusso dell’Aria, in modo che il flussometro indichi circa 0,3 l/min (
“REGOLAZIONE FINE USCITA IN mA”(Codice Reg.Fine: F2, F1, F1, F2): questa funzione permette di regolare
l’uscita in mA di ± 0,32mA corrispondente a ± 0,5% O2 ma va utilizzata solo in fase di installazione e se, dopo aver fatto la “CALIBRAZIONE IN ARIA”(pulita), la centrale cui è collegato il TS220EO indica un valore diverso da 20,9%O2.
TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI )
vedi Fig.3).
IST-2220.EO03.01/A Istruzione / User’s Manual / Manuel d’utilisation Pag.4/9
Tel.
+39 02
26922890
- Fax
+39 02
2133734
Dopo aver inserito il “Codice Reg. Fine”, il Led Giallo inizia a lampeggiare, il valore aumenta con il Tasto F1 e dimi­nuisce con F2. Per confermare, tenere contemporaneamente premuti i due tasti F1 ed F2 per 2 secondi e dopo controllare che il Led Giallo rimanga spento.
Questa operazione non può essere usata se è già acceso il Led Giallo, o se la concentrazione di O2 è inferiore a 20,6 o superiore a 21,2 in questo caso va prima fatta la “CALIBRAZIONE IN ARIA”.
“VERIFICA”(non serve codice): anche se il funzionamento del sensore è garantito se viene periodicamente esegui- ta con successo la “Calibrazione in Aria”, è comunque possible simulare la risposta del rilevatore al Gas utilizzando una bombola con miscela di Gas circa al 15% Ossigeno in azoto.
Con i puntali del volmetro sui Test-Point “TESTmA”, (vedi Fig.2) controllare che ci siano circa 174 mV in funziona­mento normale in aria (20,9% O2) [oppure se l’uscita in mA non fosse utilizzata, con i puntali del milliamperometro direttamente sui
morsetti “S e –“ verificare che ci siano 17,4mA circa]. Infilare il TC011 sul portasensore, regolare il riduttore della Bombola in modo che il flussometro indichi circa 0,3 l/min (
vedi Fig. 3). controllare che si raggiunga in 2÷3 minuti un valore di cir-
ca 136mV [ovvero che l’uscita in mA diminuisca fino a circa 13,6 mA]. Terminata la “Verifica”, chiudere la bombola, togliere il
TC011. Attendere 3-5 minuti per tornare alle condizioni di funzionamento normale.
DESCRIPTION
EN
The TS220EO series are a Oxygen three-wire 4÷20mA transmitter employing an electrochemical sensor that find its application in centralized alarm systems for laboratories, industries and environments to be protected from the Oxygen deficiency, leakage or enrichment and also from presence of inert gases. The instruments comprise of a thermoplastic case in which the electronic circuit and the terminals are mounted. The enclosure has downward fac­ing cylindrical sensor housing with inside a “Replaceable Cartridge Sensor”.
The instrument has a 4÷20mA linear output (S) with 25.0 % O2 F.S. This output is connectable to a remote Gas
Central Unit as listed in Table 1. On the front panel there are two F1 e F2 key, using for the Test and Calibration rou­tine, protected by a code, and 3 Led shows the working conditions:
Red LED "ALARM": 22,5% O2 alarm indication Green LED "ON": normal working condition Yellow LED "FAULT": the sensor should be faulty, disconnected, out of scale or expired.
NOTES ON THE DETECTED GAS
Oxygen (O2) is a colourless, odourless gas, a little heavier than air (its density to air is 1.1). Oxygen is not a flam-
mable gas, it is a strong oxidant which reacts with reducing and flamable materials, increasing the risk of fire and explosion. We all need to breathe the oxygen in air to live. Air is made up of several different gases including oxy­gen. Normal ambient air contains an oxygen concentration of 20.9% v/v. When the oxygen level dips below
19.5%v/v, the air is considered oxygen-deficient. Oxygen concentrations below 16%v/v are considered unsafe for humans. Oxygen concentrations below 6%v/v cause death. Considered that the Oxygen decreasing may be caused by combustion, leakage of other gases (toxic and / or flammable) or also displaced by inert gases increas­ing in the environment (nitrogen, helium, argon etc.). A high concentration, more than 24%v/v O2 is irritating to the respiratory tract and can cause effects on the central nervous system, eyes and lungs that can be damaged by the inhalation of high concentrations.
OPERATIONAL DESCRIPTION
The electrochemical sensor is temperature compensated, but is sensitive to extreme humidity variations.
Preheating: when powered, the sensor needs a time of preliminary heating of about 60 seconds. During this peri-
od the yellow LED “FAULT” flashes. After this period, the yellow LED light off, the green LED “ON” illuminates to in­dicate normal functioning. After this period the unit is able to detect gas even if it attains the optimum stability con­ditions after about 15 minutes continual functioning. Normal operation: the green LED “ON” should be light on.
The Red LED (ALARM) illuminates when the O2 concentration attains 22,5 % or drops below 19,5% (only if it be acti-
vate by setting ON Dip-Switch n.1).
Faults: the instrument signal different kind of failures, as listed below. The Yellow LED illuminates, the "S" output falls down to 0 mA
Yellow LED illuminates each 4 seconds (with Green LED activate): this happens when the “Cartridge Sensor” has overcome its theoretical period of life and its correct operation is not longer guaranteed. The detector keeps on op­erating normally but it is necessary to replace, as soon as possible, the “Cartridge Sensor” with a new one. The type to be required is described at Pag. 1. The replacement procedure is described in the attached manual.
Yellow LED activate, Green LED off (0 mA output signal): this signal different kind of faults. 1) The Dip Switch set up is wrong, please verify (see Table 2). 2) The “Cartridge Sensor” is not working, please replace with new one. 3) If a new Cartridge Sensor” is installed or it is not correctly connected or a not compatible one is mounted. Please check the cartridge connections and compatibility (see Pag. 1) these checks are made switching off and on the de­vice. If the condition does not change, it will be necessary to replace the unit and/or send it back to the supplier to repair.
Yellow and Green LED activates (0 mA output signal): this happens when the “Cartridge Sensor” is not working. First try to perform the procedure of “ Calibration in Air” as described in the section “Test and Calibration > Calibra- tion in Air ” then disconnect and connect the unit, finally try to replace a new “Cartridge Sensor”. If the condition is not change, it will be necessary to replace the unit and/or send it back to the supplier to repair.
TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI )
Tel.
+39 02
26922890
- Fax
+39 02
2133734
IST-2220.EO03.01/A Istruzione / User’s Manual / Manuel d’utilisation Pag.5/9
Only with Red set-up “activate” all LED activates (>24 mA output signal): this happens when the “Cartridge
Sensor” is not working or gas concentration is out of scale (higher than 30%v/v O2) If there are not any gas leaks and the condition is not change, it will be necessary to replace the unit and/or send it back to the supplier to repair.
INSTALLATION
The detector must be accurately installed according to the national dispositions in force on the safety of the plants and installation of electric devices. Detector positioning: The
Fig. 2 shows the instrument size. The unit must be positioned vertically with the sensor
downwards. It should be fixed at 1.60 m from the floor (the O2 gas is lighter as air). Electrical Connection (see Fig.2): the maximum distance to install each detector from the Central Unit is show in the Table 1. Normally use a tree wire shielded cable. The terminals (+ - S), are plug-in type, it is necessary to extract them to make the connection. Pay attention when you insert them again, being polarized.
Note: Dip-Switch should be set with instrument powered off. Dip-Switch 1 settled ON activates ALARM Led indica-
tion (see Table 2) (see also “Operational Description > Faults”).
Important: Once installation is completed, only if it is necessary, adjust the sensor to the environment conditions,
switch on the unit, wait about 15÷20 minutes and carry out the ”Calibration in Air” (see ‘Test and Calibration >). COMPATIBILITY WITH OTHER CENTRAL UNITS: In case of a central unit other then Tecnocontrol, please
verify the max load resistor as shown in Fig. 4. The max load resistor is 50 ohm with 12Vdc power supply, while is 400 ohm with 24Vdc power supply. We suggest using 24Vdc power supply.
WARNING
Average life: The sensitive element used in this detector has an excellent stability in time. In fresh air and in nor­mal working condition the sensor's life is about 2 years. After this period the yellow LED “FAULT” flashes every 4 seconds, is necessary replacing the “Cartridge Sensor”.
Periodical testing: we advise to carry out working tests every 6 or 12 months. as explained on page 4 chapter “Tests and Calibration”. Note: the detector is not able to detect Oxygen gas occurring outside the room where it is installed, either inside walls or under the floor.
Sensor operating limits: the electro-chemical cell, in any case, should not be exposed to temperatures lower
than -40° or higher than 60°C. The electrochemical cell should not be exposed to an over-range concentration of 30 % O2. If this condition occurs, we suggest replacing the sensor. Safety: Oxygen sensors contain a small quantity potassium acetate solution which is corrosive. Normally this mate­rial would only leak out as a result of either mechanical damage or an incorrect instrument use beyond the limits of its operational functioning. These sensors also contain small amounts of lead, lead oxide, platinum, silver, carbon and antimony. If the user comes into contact with the oxygen sensor contents the affected area should be washed with a copious supply of water.
Cross sensitivity to other gases: The electrochemical cell employed has a good resistance towards products
such as sprays, detergents, ammonia, glues and paints. However, in case of products containing substances in great quantity, these could interfere with the Sensor and cause false alarms. We recommend ventilating the room when products like these are used. Volatile organic solvents in great quantity cause false alarms, and then the cell needs one or more hours in clean air to recover the normal sensitivity conditions.
TEST and CALIBRATION
PAY ATTENTION: This procedure has to be made with extreme attention and by authorized and trained people; because starting this procedure it will increase mA Output causing the activation of connected alarm devices to the gas Central Unit. Operation Check, Calibration in Air and Fine mA ADJ Output: are different code protected functions. To access these functions is necessary to insert the relevant “Code" through the keys F1s and F2. To have the key pressure recognized, hold pressing it for around a second (until the Green Led doesn't switch off for a moment). Then the next key can be pressed. In case of error all it takes is waiting around 10 seconds and the sequence is automatically erased. The Calibration Check is a function code free.
Calibration Kit, Sample Gas Bottles (for Calibration Check) please, using a mixture about 16%v/v O2 in Nitro-
gen. It is possible to use either the disposable cylinders with adjust valve or the high pressure ones with reduction
gear. Is also necessary the Tecnocontrol Calibration Kit model TC011.
“INSTRUMENT OPERATION CHECK” (Check Code: F2, F2, F1, F1): this function allows to effect a functional test
of the equipment. After having put the system in safety and inserted the "Code Test", all Led are switched off. Then they will switch on in sequence, the Led, from the yellow up to the red. At the end all the Led will remain lighted for around 5 seconds, then the central returns at the conditions of normal operation. It is advisable to perform this op­eration every 6-12 months according to the use.
Note: this function is not working if the red Led have already turned on. “CALIBRATION IN AIR” (Calibration Code: F2, F2, F2, F1, F2, F1): this function allows completely recalibrate the
sensor at 20,9%v/v in air. This procedure when it is launched, will automatically until the end.
Warning: to guarantee that no errors of elaboration happen, the possibility exists that during the Calibration the yel-
low Led switch off every 8 seconds, in this case interrupt the procedure, switch off and switch on the instrument and repeat the Calibration. If condition persists it will be necessary to send the detector to the supplier for the reparation.
Important note: During Calibration routine the mA output indicates 0mA.
TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI )
IST-2220.EO03.01/A Istruzione / User’s Manual / Manuel d’utilisation Pag.6/9
Tel.
+39 02
26922890
- Fax
+39 02
2133734
The Calibration can be done in clean air only (environment without the presence of flammable or other polluting gas). With the
keys perform the “Calibration Code ". Wait until the Yellow and Green Led switch on fix and the red Led starts to flash (the Led Red remains switch on fixed to indicate the calibration working). Wait while the red Led switched off (for more of 2 seconds). At this point we can have two possibilities:
Yellow and Green Led illuminates: the calibration routine
has correctly been performed. The detector automatically
switch off in 8 seconds, and restart into normal operating mode. (see chapter OPERATIONAL DESCRIPTION “Preheating”)
Yellow LED illuminates: the calibration routine has failed. The detector automatically switch off and restart in 8
seconds and repeat the procedure without inserting again the sequence. If condition still persists after the replace­ment of the cartridge it will be necessary to send the detector back to the manufacturer for reparation. Important: If the environment is very polluted and there is doubt that the concentration of oxygen is 20.9%v/v, you can use an air cylinder with 20.9%v/v of O2 or use a gas balloon (our code PA020 ) filled with fresh air. In this case, before executing the "Calibrating Code" insert the TC011 over the sensor holder, adjust the flow of air, so that the flow meter indicates approximately 0.3 l / min (see Fig.3).
"FINE mA ADJ OUTPUT” (ADJ Code: F2, F1, F1, F2): This feature allows you to adjust the mA output of ± 0.32 mA corresponding to ± 0.5% O2, but it should be used only at new installation and if, after the "Calibration in Air" (clean air), the control unit, connected to TS220EO, indicates a value other than 20.9% O2.
After entering the "ADJ Code", the yellow LED starts to flash, the output value increases pressing the F1 key, and decreases pressing F2. Hold down the two F1 and F2 keys simultaneously for 2 seconds to confirm and then check that yellow LED remains off.
This procedure can not be used, if the yellow LED is already on, or if the O2 concentration is less than 20.6% or higher to 21.2%. If this is the case: "Calibration in Air" routine must be done first.
“CALIBRATION CHECKS” (no Code required): although the sensor operation is guaranteed if it is regularly per-
formed successfully "Calibrating in Air", it is possible to simulate the response of the detector using sample gas cyl­inder with a mixture 16%v/v Oxygen in nitrogen.
Verify with a voltmeter connected to the Test-Point “TESTmA” (see Fig.2), the value reach a 174mV value in normal
clean air condition (20.9%v/v).
Then opplying the 16%v/v O2 gas, insert the TC011 over the sensor holder; adjust the sample gas cylinder valve as the flow meter indicates around 0.3 l/mins (see Fig.3), wait for 1÷2 minutes, verify to read a value of about 143 mV. If the result is different, is necessary to recalibrate the sensor (see “Calibration in Air”). Then, close the gas cylin­der and remove TC011, wait 2÷3 minutes, until the instrument restores the normal working conditions.
DESCRIPTION
FR
Le TS220EO sont des transmetteurs (4÷20mA) de Oxygène (O2) avec capteur électrochimique pouvant égale-
ment être utilisé en systèmes centralisés d'alarme pour l’industrie, le tertiaire, et environnements qui doit être protégé par le manque ou l'excès d'O2 ou de la présence de gaz inertes. Le détecteur est constitué par une boîtier contenant le circuit électronique et les borniers de raccordement; dans le porte capteur, placé dans la partie inférieure du boîtier, est logée une "cartouche capteur échangeable" contenant l'élément sensible et les données identificatrices et de réglage. Le TS220EO sont des transmetteurs sur 3 fils avec un signal de sortie (S) 4-20mA linéaire avec Fin D’échelle du 25,0% du gaz O2. Ils sont utilisés avec les centrales de détection gaz Tecnocontrol. Sur le couvercle ils sont, les touches F1 et F2 pour les opérations de vérification et calibrage en utilisant par codes et sont visibles 3 Led qu'ils indiquent les conditions d'exercice:
Signalisations et commandes:
Led Rouge "ALARM": Signal d'alarme ALARM (22,5 % O2) Led Vert "ON": fonctionnement normal. Led Jaune "FAULT": capteur en panne ou déconnecté ou saturé ou échu. Touches F1 et F2 pour les opérations de vérification et calibrage
CARACTÉRISTIQUES DU GAZ
L’Oxygène (O2) est un gaz incolore, inodore, légèrement plus lourd que l’air (sa densité relative à l’air est de 1,1).
Ce n’est pas un gaz inflammable mais étant fortement oxydant, il réagit avec les matériaux combustibles et réduc­teurs en facilitant leur combustion, déterminant ainsi des risques d’incendie et d’explosion.
L’Oxygène constitue 20,95% de l’air que nous respirons. Si la concentration d’O2 dans l’air descend sous 19,5% l’on considère que cette carence est dommageable pour la santé des personnes, si cette concentration descend sous 16%, ce manque cause des problèmes respiratoires jusqu’à provoquer l’asphyxie et sous les 6% entraîner la mort immédiate. Il faut également penser que la carence en oxygène peut être provoquée par des phénomènes de combustion, par l’augmentation dans l’ambiance de la concentration d’autre gaz agissant par oxyprive (CO2, gaz toxiques et/ou inflammables) mais également par des pertes de gaz inertes (azote, hélium, argon etc.). A concen­trations élevées, supérieures à 24% l’oxygène est irritant pour l’appareil respiratoire et peut causer des effets sur le système nerveux central, les yeux ainsi que les poumons qui peuvent être endommagés par l’inhalation de hautes concentrations d’oxygène.
FONCTIONNEMENT
Le capteur électrochimique, est compensé en température mais il est sensible aux variations extrêmes d'humidité.
Préchauffage: a partir de la mise sous tension, le détecteur commence la phase de préchauffage du capteur, si­gnalée par le clignotement de la LED jaune "FAULT". Après environ 60 secondes, la LED jaune s'éteint et la LED
TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI )
IST-2220.EO03.01/A Istruzione / User’s Manual / Manuel d’utilisation Pag.7/9
Tel.
+39 02
26922890
- Fax
+39 02
2133734
verte "ON" s'allume, indiquant le fonctionnement normal Après ce temps le capteur est apte à détecter le gaz, mais il n’atteint les conditions de stabilité optimale qu’après 15 minutes environ de fonctionnement continu.
Fonctionnement Normal: seule la LED Verte (ON) doit être allumée.
Le Led Rouge, ALARM s'allume si la concentration de Gaz dépasse le 22,5%O2 ou bien il descende au dessous
de 19,5% (Abilitation avec Dip-Switch n.1 ON). Dérangement: La centrale signale les anomalies, sous énumérez, en allumant le Led jaune (FAULT), en portant la sortie "S" au 0 mA. Le Led Jaune s'allume chaque 4 secondes, (avec le Led Vert allumé): pour avertir que la "Cartouche Capteur" a dépassé sa limite de vie, et le fonctionnement correct n'est plus garanti. Le détecteur continue à fonctionner nor­malement, mais il est nécessaire, au plus tôt, remplacer la "Cartouche Capteur" avec une nouvelle, le type à de­mander est indiqué en Page 1. La procédure de substitution est décrite dans la documentation jointe. Si le Led Jaune est allumé et le Vert est éteint, (sortie 0 mA): il indique plus de possibilités de dérangement, c'est-à-dire: 1) la configuration des Dip-Switch n'est pas correct, vérifier la position, (Voire Tableau 2). 2) la "Car- touche Capteur" est en panne, la remplacer avec une nouvelle. 3) si une nouvelle "Cartouche" est installée, ou elle n'est pas correctement connectée, elle n’est pas compatible. Contrôler les connexions avec la "Cartouche" et la compatibilité (Voire page 1). Exécutez le contrôle en déconnectant et connectant le détecteur. Si la condition perdure il sera nécessaire de remplacer e/o envoyer le détecteur au fournisseur pour la réparation.
Si les Led Jaune et le Vert sont allumés, (sortie 0 mA): il indique la panne probable de la "Cartouche Capteur". Premier chose, essayer à exécuter le " étalonnage en Air " comme décrit dans la section "Vérification et
Etalonnage", puis dis alimenté et ri alimenté le détecteur, finalement essayer de remplacer la "Cartouche Cap- teur". Si la condition persiste il sera nécessaire de remplacer e/o envoyer le détecteur au fournisseur pour la répa-
ration.
Toutes les LED sont allumées, (sortie >24mA): indication de, ou panne de la "Cartouche Capteur" ou une concentra- tion de gaz supérieur au fond d’échelle (>30% d’O2). S'il n'est pas constaté de fuite de gaz et que la condition per­dure après la substitution de la "Cartouche" il sera nécessaire d'envoyer le détecteur au fournisseur pour réparation.
INSTALLATION
Les détecteurs vont installés et positionné en suivant toutes les règles nationales en vigueur pour les installations électriques et les normes de sûreté des installations.
Montage: en Fig.1 sont indiquées les dimensions. Le détecteur doit être monté en position verticale avec le cap­teur tourné vers le bas.
Positionnement du TS220EO: il doit être fixé à environ 1,60 m du plancher, l’oxygène ayant une densité proche de celle de l’air. Liaisons électriques (Fig.2): La distance maximale à laquelle peu être raccordé la sonde à la centrale est indiqué dans le Tableau 1 en fonction de la section du câble utilisé. Normalement est utilisé un câble à écran, avec 3 con­ducteurs. Raccordez l’écran du câble à la centrale. Les borniers (+ - S) sont de type débrochable, et il est nécessaire de le défiler pour effectuer les connexions. Prêter attention en le les réinsérer étant donné qu'ils sont polarisés.
Remarque: Les Dip-Switch doivent être positionnés avant d'alimenter le détecteur. Si on utilise le détecteur avec le Dip-Switch 1 en position ON, l’indication d’alarme (Led rouge) sera activée (voire Tableau 2) Si on utilise le détecteur avec les Dip-Switch dans une position pas permise l'indication de Dérangement sera activée. Important: Une fois terminée l'installation, pour adapter le capteur aux conditions ambiantes, alimenter le détec­teur, attendre 30÷60 minutes environ et puis exécuter la "étalonnage en Air" (voire "Vérifications et Calibrage").
Comptabilité avec autres centrales: en cas d’utilisation d’une centrale différent de Tecnocontrol, calculer
la résistance de charge maxi en utilisant la charte selon la Fig.4. L’alimentation à 24Vcc est conseillée.
INSTRUCTIONS
La vie utile du capteur en air propre est 2 ans en moyenne. Au terme de cette période, indiqué par l'instrument
par un clignotement de la LED Jaune toutes les 4 secondes, il est nécessaire de remplacer la "Cartouche Capteur".
Vérifications Périodiques: nous conseillons d'exécuter tous les semestres la vérification de fonctionnement du détecteurcomme décrit dans la rubrique: "Vérifications et Etalonnage".
Remarque: Le détecteur n'est pas apte à détecter l’oxygène hors de l’ambiance dans laquelle il est installé.
Limites de fonctionnement: des températures inférieures à -40°C et supérieures à +60°C provoquent la détério-
ration du capteur électrolytique. Il peut également être endommagé s’il est exposé à des concentrations d’O2 très élevées (> 30%). Dans ce cas le capteur doit être remplacé.
Attention: Les capteurs électrochimiques pour oxygène contiennent une petite quantité d’acétate de potasse qui est corrosif. Normalement il ne doit y avoir aucune perte au capteur, dans le cas contraire ceci est du à une action mécanique violente ou à un usage outre-limites de fonctionnement. Le capteur contient également en petites quan­tités: du plomb, de l’oxyde de plomb, du platine, de l’argent, du carbone et de l’antimoine. Dans le cas d’un contact accidentel entre du liquide provenant du capteur et la peau ou les yeux, il est conseillé de laver immédiatement avec de l’eau en abondance. Interférence avec d’autres gaz: La cellule électrochimique utilisée possède une bonne résistance aux produits d’usage courant comme les aérosols, ammoniaque, colles ou vernis; toutefois si ceux-ci sont présents en quantités élevées, ils peuvent interférer avec le capteur; il est conseillé de ventiler l’ambiance après emploi de ces subs­tances. Considérer qu’en ambiances particulièrement polluées par des vapeurs de solvants organiques en quanti­tés élevées, des fausses alarmes peuvent apparaître et que le capteur devra, pour retrouver ses conditions de fonctionnement normales rester longtemps en air propre.
TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI )
IST-2220.EO03.01/A Istruzione / User’s Manual / Manuel d’utilisation Pag.8/9
Tel.
+39 02
26922890
- Fax
+39 02
2133734
VÉRIFICATIONS È ETALONNAGE
Remarque Important: Toutes les opérations suivantes doivent être exécutées seulement par personnel expert et
formé, vu que la sortie en mA en provoquant l'activation des dispositifs d'alarme connectés à la centrale.
Le TS220EO possède trois différentes fonctions protégées par un "Code": Test Électrique, étalonnage en Air et Reglage fin de sortie en mA.
Pour accéder à ces fonctions il est nécessaire d'insérer le "Code" en utilisant les touches F1 et F2. Pour que la pression sur un touche soit reconnue, la tenir pressée durant environ une seconde (jusqu’à ce que s’éteigne un bref instant la LED Verte). Ensuite, l’on peut passer au bouton suivant. En cas d’erreur, il suffit d'attendre environ 10 secondes et la séquence est effacée automatiquement.
Kit d’étalonnage et bouteille avec mélange de gaz titré: pour effectuer les vérifications utiliser comme gaz de l’Oxygène à environ 15%V/V dans l’azote. Il est possible d’utiliser une bouteille rechargeable sous haute pres- sion avec détendeur, il est indispensable d’utiliser le kit d’étalonnage et vérification TC011 de Tecnocontrol.
"TEST ELECTRIC"(Code Test: F2, F2, F1, F1): cette fonction permet d'effectuer un test fonctionnel de l'appareil-
lage. Après avoir mis l'installation en sûreté et avoir inséré le "Code Test", tous Led s'éteignent. Ils s'allumeront ensuite en séquence, les Led, du jaune jusqu'à le rouge. À la fin tous les Led resteront allumé pour environ 5 se­condes, puis le détecteur il reviendra dans les conditions de fonctionnement normal. Il est conseillable d'exécuter cette opération chaque 6-12 mois en base à l’utilisation.
Remarque: Il n'est pas possible d'exécuter l'opération si ils sont allumés déjà la Led Rouge.
ETALONNAGE EN AIR (Code Etalonnage: F2, F2, F2, F1, F2, F1): cette fonction permet l’étalonnage de cap-
teur in Air à 20,9%. Une fois commencée, cette procédure se poursuivra automatiquement jusqu'à à la fin.
Durant la procédure, une rare possibilité existe que la LED Jaune clignote chaque 8 secondes, dans ce cas inter­rompre la procédure puis exécutez le contrôle en coupant et en rétablissant l’alimentation du détecteur. En cas de persistance il sera nécessaire de remplacer et/ou renvoyer le détecteur au fournisseur pour réparation.
Important: Durant l’opération de étalonnage l’appareil portant la sortie "S" au 0mA.
L’étalonnage doit être exécuté en air propre exclusivement (milieu sans la présence de gaz toxiques ou autres). A l’aide des touches exécuter le "Code d’Etalonnage". Attendre que les LED Jaune et Verte s’allument et la Led Rouge clignote. (la LED Rouge reste ensuite en feu fixe pour indiquer que la procédure est en fonction). Attendre que la LED rouge s’éteigne (au moins durant 2 secondes). A ce point, deux cas peuvent se vérifier:
LED Jaune et Verte allumées: l’étalonnage a correctement été exécuté. Apres 8 secondes, l’instrument s’éteint et il se rallume automatiquement en fonctionnement normal. (Voir Chapitre FONCTIONNEMENT "Préchauffage”).
LED Jaune allumée l’étalonnage a échoué. Apres 8 secondes, l’instrument s’éteint et se rallume automatique­ment et répète la procédure sans réinsérer la séquence. Si la condition persiste même après la substitution "Car- touche Capteur", il sera nécessaire d'envoyer la sonde au fournisseur pour réparation.
Important: Si l’ambiance est très polluée et si l’on doute que la concentration d’oxygène soit de 20,9%, il est pos­sible d’utiliser une bouteille d’air avec 20,9% d’ O2 ou bien d’utiliser un ballon pour gaz PA020 rempli d’air propre. Dans ce cas, avant d’exécuter le codeEtalonnage en air” coiffer la tête de détection avec le TC011, régler l’afflux de l’air, de façon à ce que le débitmètre indique environ 0,3 l/min (voir Fig.3). REGLAGE FIN DE SORTIE EN mA (Code: F2, F1, F1, F2) cette function permet le réglage de la sortie en mA de ±0,32 mA qui corresponde à ± 0,5% O2 mais elle doit être utilisée seulement dans la phase d’installation et si, après avoir fait l’Etalonnage en Air (propre) la centrale à la quelle le TS220EO est connecté, indique un valeur diffé­rent de 20,9%O2. Apres avoir inséré le Code, la LED jaune commence à clignoter, la valeur augmente avec la touche F1 et diminue avec F2. Pour confirmer appuyez simultanément les deux touches F1 et F2 pour deus se­conds et contrôlez que la LED jaune reste éteints. Cette opération ne peut pas être utilisée si la LED jaune est déjà allumé, ou si la concentration de O2 est inferieur à 20,6% ou plus grand que 21,2%, dans ce cas il fau faire d’abord l’Etalonnage en Air.
VÉRIFICATION DE L’ETALONNAGE (code non nécessaire): la vérification du fonctionnement correct du détecteur doit être exécuté après l’"Etalonnage en Air", ou après l'installation ou pendant les entretiens périodiques, étant don­né qu’elle constitue l‘unique moyen de contrôler la fonction effective de l’appareil. Le mélange à utiliser est 15%v/v oxygène dans l’azote.
Vérifiez avec le voltmètre sur les points de test "TEST mA" (voir fig.2) si une valeur d’environ 174 mV est présente, en fonctionnement normal en air (20,9% d’O2). Coiffer la tête de détection avec le TC011, régler l’afflux de gaz, de manière à ce que le débitmètre indique 0,3 l/min environ, (voir Fig.3). Attendre 2÷3 minutes environ et contrôler que l’on mesure environ 136 mV.
Un fois terminé la “VERIFICATION”, fermer la bouteille et enlever le TC011. Attendre 3÷5 minutes environ pour revenir aux conditions de fonctionnement normal.
TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI )
IST-2220.EO03.01/A Istruzione / User’s Manual / Manuel d’utilisation Pag.9/9
Tel.
+39 02
26922890
- Fax
+39 02
2133734
Sezione Cavo
Resistenza Cavo
La max distanza cui
può essere install
a-
La max distanza cui può essere installato
0,75 mm2 26 Ω/km 100 m
300 m
1 mm2 20 Ω/km 150 m
400 m
1,5 mm2 14 Ω/km 200 m
500 m
2,5 mm2 8 Ω/km 400 m 800 m
“S1-SET”(Dip
-
Switch)
Led Allarme / Alarm Led / Led d'alarme
1 2 3 4
ON ON ON ON
ON OFF OFF OFF
Size / Dimensions
Fig. 2
- Schema di collegamento
/ Wiring diagram / Schéma
1
2
-
24V
mA
ON
OFF
SET
Test
Uscita
+/-
(12÷24Vdc
)
Centrale rivelazione gas
- S
“S”
Dip Switch
+
m
V
0,3L/min
F1
F2
Tasti /
Key /
Touches
Alimentazione
/ Power Supply / Alimentation
100
300
400
500
50
0
5
200
Vdc
RL=Resistenza di carico
/ Load resistor / Resistance de charge
RL
600
0 10 15 20 25 28
650
10,8 24
Area di funzionamento
350
700
800
Tabella 1 / Table 1 / Tableau 1
Cable Size
Section du câble
[Singolo Conduttore]
Cable Resistance
[Single wire].
Résistance du câble
[Conducteur Unique]
Tabella 2 / Table 2 / Tableau 2
to ogni trasmettitore dalla centralina
Tecnocontrol Modello ID250 e SE148 è:
The maximum distance to install each de-
tector from the Gas Central Unit Model
ID250 and SE148 is:
La maxi distance à laquelle peuvent être
raccordées les sondes à les centrales
ID250 et SE148 est:
ogni trasmettitore dalla centrali Modello
CE100, CE400, CE600, CE700 (CE380UR) è: The max distance to install each detector from
the Gas Central Unit Model
CE100, CE400, CE600, CE700 (CE380UR) is:
La maxi distance à laquelle peuvent être raccor-
dées les sondes à les centrales
CE100, CE400, CE600, CE700(CE380UR) est:
OFF ON
Gas Central Unit
Centrale de détection
RL
1 2 3 4
Fig. 1 – Dimensioni /
Uscita 4÷20 mA
4÷20 mA Output / Sortie en 4÷20 mA
S
-
Output Test
Sortie Test
40÷200mV
+
Alimentazione / Power Supply / Alimentation
Ohm
Allowable oparating region
TESTmA
Fig. 3 - Tester di calibrazione
/ Calibration Tester / Kit de Calibration
TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI )
Fig.4 - Alimentazione / Resistenza di Carico 4÷20mA
Power supply / Load resistance diagram 4÷20mA Alimentation / Résistance de charge 4÷20mA