Technology & Calibration Inc. TechCal Owner's Manual

Owner’s Manual
Technology & Calibration, Inc. - Houston, Texas
www.techcal.com
Operating Instructions
1) Thoroughly check instrument at the time of unpacking to inspect for any damage that may have occurred during shipment.
2) Mount the unit as level as possible.
4) Select the desired rotation speed on the chart drive.
5) Install chart by unlocking the chart hub.
6) Raise the Pen Lifter.
7) Slide the chart paper between the pens and the chart plate securing it in the tabs.
8) Lock the Hub and lower the Pen Lifter.
9) Verify that the pen (s) are in in contact with the chart paper and have sufficient ink.
10) Apply zero pressure and adjust the pen for chart zero.
11) Turn on the pressure to the unit.
12) Upon completion of test, turn off the Pressure.
13) Turn off the Chart Drive.
14) Lift pen (s) from Chart Paper.
15) Replace Pen covers to keep pens from drying out.
* If the recorder is suspected to be out of tolerance or not reading correctly please contact an authorized repair facility or the manufacture for repair and/or recalibration.
* Each Tech Cal Recorder is supplied with NIST Traceable Certifications. It is required
that occasionally the recorder be recalibrated to ensure it accuracy.
General Specifications: Case Polyethylene Copolymer, Chemical Resistant, Light Weight & Durable Temperature Range: -10°F to 210°F
Window Lexan, Polypropylene, Impact Resistant
Chart Drive ‘C’ Battery Operated, UL and cUL Listed For Hazardous Locations, Temperature Range: -40°F to 160°F
Chart Size 12-inch Diameter (American Meter Style)
Chart Rotation 60 Min, 96 Min., 12 HR, 1,2,7,8,14,16,31 & 32 Day
Number of Pens 1-3 (one for each element)
Pen Style Disposable, Clip-On
Element Type Pressure-Helical (bourdon), Stainless Steel Temperature-Helical (bourdon) Class V Mercury Filled
Range Pressure 0 thru 30,000 PSI,
Temperature 0 - 250 Deg. F
Accuracy +/- .5% of Full Scale
Parts List
Item Part # Description
1 4P143 Foot 2 4P136 Pen Lifter Assembly 3 4P113 Soft Grip Top Handle 4 4P131 Case Connector, 1/4” NPTF 5 4P121 Pen Mount Assembly
6.1 4P101 500 PSI Static Pressure Element
6.2 4P102 1K PSI Static Pressure Element
6.3 4P103 2K PSI Static Pressure Element
6.4 4P104 3K PSI Static Pressure Element
6.5 4P105 5K PSI Static Pressure Element
6.6 4P106 10K PSI Static Pressure Element
6.7 4P107 15K PSI Static Pressure Element
6.8 4P108 20K PSI Static Pressure Element
6.9 4P109 25K PSI Static Pressure Element
6.10 4P110 30K PSI Static Pressure Element
7.1 4P140 1st From Chart, Blue
7.2 4P141 2nd From Chart, Red
7.3 4P142 3rd From Chart, Green 8 4P801 Link Assembly 9 4P127 Pen Arm, Long Slot 10 4P139 Riser, Chart Plate 11 4P137 Riser, Temp. System 12 4P138 Riser, Chart Drive
Item Part # Description
13 4P130 Hub, G190 14 4P125 Pen Shaft Arm Assembly, 90 Deg 15 4P135 Range Adjustment Arm
16.1 4P122 1st From Chart, Shaft Assembly.
16.2 4P123 2nd From Chart, Shaft Assembly.
16.3 4P124 3rd From Chart, Shaft Assembly. 17 4P128 Chart Plate 18 4P120 Back Plate 19 4P144 Lexan Window 20 4P161 Ambient Temp. Sys., 0-150 Deg.F
21.1 4P111 0-150Deg F, Temp. Sys. /10’ Armor Cap
21.2 4P112 0-200Deg F, Temp. Sys. /10’ Armor Cap
21.3 4P164 50-250Deg F, Temp. Sys./10’ Armor Cap 22 4P163 Pen / Battery Holder
23.1 4P116 Black Case
23.2 4P117 Orange Case
23.3 4P118 Yellow Case
23.4 4P119 Silver Case 24 4P129 Chart Drive, Sonceboz 0820 25 4P158 Pipe Mount, 2” Slip-On Not Shown 4P165 Spool, Temp. Cable Not Shown 4P166 Panel Mount Brackets Not Shown 4P167 Lower Connection Plate
Warranty
Tech Cal Recorders are warranted against defective mate­rials or workmanship for 12 months following the date of shipment. Technology & Calibration, Inc.’s liability is lim­ited to repair or replacement at the factory. No obliga­tion can be assumed for recorders subjected to mis-
use or unauthorized repairs.
Technology & Calibration, Inc.
ISO 9001 Cered
Manufactured By:
Houston, Texas
888-546-6506 Toll Free
www.techcal.com
© 2007 Technology & Calibration, Inc.