Techkon SpectroDens User Manual

Page 1
Manual
Spectro-Densitometer
SpectroDens
Software
SpectroConnect
Page 2
TECHKON manuals, technical documentation and programs are copyrighted. Reproduction, translation or transfer to an electronic medium – in whole or in parts – is prohibited.
TECHKON software is the intellectual property of TECHKON GmbH. Purchase of the software grants the user a license for the use on one computer only. Programs may only be copied for back-up purposes.
TECHKON can not be held liable for any damages that may occur by the use of TECHKON products.
The TECHKON team takes great care in producing this manual. Nev­ertheless, we can not fully guarantee the complete absence of errors. TECHKON GmbH and the authors can not be held liable neither for any possibly incorrect statements nor their consequences.
Products that are not made by TECHKON GmbH are mentioned for information purposes only and such statements do not represent trademark infringement. All registered trademarks are recognized.
TECHKON products and product names are registered internationally under several intellectual patents and trademarks.
Specications can be subject to change without notice.
Version 4.3, March 2018
You can reach us at the following address: TECHKON GmbH Wiesbadener Str. 27 • D-61462 Königstein/Germany T +49 (0) 6174 / 92 44 50 • F +49 (0) 6174 / 92 44 99 info@techkon.com • www.techkon.com
Page 3
Welcome
We welcome you among the worldwide community of users of TECHKON products. We are happy that you have selected this high-quality measurement instrument. It will be a valuable tool for your day-to-day quality control tasks. With this manual we invite you to learn how to use SpectroDens.
The manual is divided into three chapters:
CHAPTER 1: How to use TECHKON SpectroDens
CHAPTER 2: Description of the measurement functions
CHAPTER 3: How to use the Windows software SpectroConnect
For the rst steps we recommend to read chapter 1; in particular paragraph 1.4.
You will be fascinated that after only 15 minutes you will know how to work with SpectroDens. The second chapter will show you detailed insight into the measure­ment functions and chapter 3 is about the Windows software SpectroConnect, which is part of the package.
Please get the device registered by using the registration card, which you will nd in the appendix of this manual. That way we can keep you updated about product news.
Please visit us as well on internet at www.techkon.com. You will nd useful information about the complete product range and new software versions.
Do you have any suggestions for improvements or do you require information that goes beyond the contents of this manual? We will be glad to hear from you. Your sug­gestions or questions make an important contribution to the continuous optimization of our documentation and products.
Your TECHKON Team
Page 4
Content
CHAPTER 1: How to use TECHKON SpectroDens
1.1 Product description ............................................................................................................................ 5
1.2 Packing list ................................................................................................................................................ 8
1.3 Design of SpectroDens ..................................................................................................................... 9
1.4 Working with SpectroDens .......................................................................................................... 11
1.5 Care and maintenance ....................................................................................................................19
CHAPTER 2: Description of the measurement functions
2.1 Measurement functions of SpectroDens Basic ................................................................33
2.2 Additional functions of SpectroDens Advanced ............................................................ 51
2.3 Additional functions of SpectroDens Premium ............................................................. 63
2.4 Storage of measurement values .............................................................................................. 77
CHAPTER 3: How to use the Windows software SpectroConnect
3.1 Software description ....................................................................................................................... 80
3.2 Installation .............................................................................................................................................. 80
3.3 Overview / Software module: Compare colors ...............................................................81
3.4 Software module: SpectroDens settings ............................................................................82
3.5 Device update and device upgrade ...................................................................................... 83
3.6 Software module: Export ............................................................................................................. 84
3.7 Software module: ColorCatcher............................................................................................... 85
3.8 Software module: PhotoLab ...................................................................................................... 86
3.9 Software module: MediaWedge .............................................................................................. 89
3.10 Software module: Transfer curve ..............................................................................................91
3.11 Software module: ISO-Report ....................................................................................................93
3.12 Software module: SpectroChart .............................................................................................. 94
3.13 Software module: SpectroCheck ............................................................................................ 95
APPENDIX:
Specications Manufacturer certicate EC-Declaration of Conformity Registration card
Page 5
5
CHAPTER 1 – How to use TECHKON SpectroDens
Chapter 1
How to use TECHKON SpectroDens
1.1 Product description
With the new SpectroDens generation TECHKON continues the success story of the SpectroDens hand-held devices in the long term. The dictum during product development was: Pursue proven concepts and integrate new meaningful tech­nologies. The result sets a new standard.
A measurement device for all applications
SpectroDens is an all-purpose measurement device, that is universally used for qual­ity control in the printing industry as well as for other tasks, where colors have to be judged critically. SpectroDens is exceptionally suitable, no matter which application. Thanks to the individually adjustable display functions, you can quickly set up the device for your particular job. The solid aluminum unibody case makes the device a reliable tool even in a harsh industrial environment. Direct positioning of the measure­ment head on the measurement eld ensures a secure and quick measurement. In ad­dition to single measurements you can easily carry out scan measurements. Tracking wheels on the bottom of the device make it possible to scan very fast up to 200 color patches at once on a print control strip or test chart.
Two devices in one
As a spectro-densitometer, SpectroDens combines the qualities of a highly accurate spectro-photometer and an easy-to-use densitometer. Measurement is spectral, i. e. the entire color information – the spectral ngerprint of the color – is precisely regis­tered. The data is converted into measurement data for descriptive analysis and dis­play:
Densitometric data is widely used in quality control during the printing process. Colorimetric data is used in the evaluation of print proofs, the creation of color pro-
les in color management as well as the color matching and formulation of inks.
Densitometry / Density measurement
A push of a button provides you with solid density as well as all useful additional in­formation such as dot gain, dot area, gray balance and print contrast. In the automatic measurement mode the device always displays the relevant information instantly. You can even quickly and easily produce complete printing curves. The spectral measure-
Page 6
6
Manual TECHKON SpectroDens
ment technology calculates not only density values for CMYK, but also exact data for the density of spot colors.
Colorimetry
All standard colorimetric functions are displayed clearly. Even the entire CIE L*a*b* color circle appears in the high resolution color display. As a result understanding CIE L*a*b* values becomes very easy and clear. In the color library 25.000 reference colors and complete digital color books can be saved.
Standardized measurement
Thanks to its ISO-compliant measuring head SpectroDens works strictly according to the standards valid for the graphic industry. Select dierent status lters for density measurement in the device settings. A special technical feature is the polarization lter which can be switched on and o by a push of a button and makes SpectroDens par­ticularly good at evaluating the Ugra/Fogra media wedge.
Using the latest LED technology SpectroDens provides D50 illumination and there­fore fullls the M0, M1, M2, M3 measuring conditions in accordance to ISO 13655. The lifetime of the LED illumination is nearly unlimited. Moreover the new design of the direction-independent measuring head provides reliable measurements of a wide range of print media.
Software SpectroConnect
The supplied Windows-software TECHKON SpectroConnect provides the connection between SpectroDens and the PC via the micro-USB connector or via the optionally available WLAN-module. Measurement data can thus be comfortably processed on the PC and transferred, for example, to Microsoft Excel™.
Use the color library to load complete digital color books quickly and easily into the device. An import of color values in CXF® format is possible as well. SpectroConnect is also the connecting module for other applications, e. g. programs for the production of printing curves, for colorimetric quality control or for checking the compliance of printing products to ISO standards.
Page 7
7
CHAPTER 1 – How to use TECHKON SpectroDens
7
Three dierent performance packages
We supply SpectroDens in three types of performance packages: The model Basic with all density functions and the Advanced version which adds essential colorimetric functions and a color library. The fully equipped agship model SpectroDens Premi- um meets all demands with regard to quality control. All SpectroDens devices can also be used for scans and feature a brilliant color display. Optionally data transmission by means of a WLAN module is possible. SpectroDens can be upgraded post-purchase by program upload to higher versions (charged service).
SpectroDens Basic
Automatic density Density CMYK (with reference measurements)  Dot area  Dot
gain
Print contrast Gray and color balance Ink trapping Printing curve Density
spectrum
Spectral density for spot colors Trend function ExPresso mini  Slur/dou-
bling
Dot area to Yule-Nielsen for printing plates  LED technology provides mea-
surement conditions M0, M1, M2, M3 according to ISO 13655
Spot Colour Tone Value
(SCTV)
Spot Colour Tone Value (SCTV) Printing curve  Average
SpectroDens Advanced
Same functions as SpectroDens Basic and additionally:
CIE L*a*b*  ΔE*a*b*  CIE L*C*h*ab  CIE XYZ  CIE color circle  ΔE*cmc ΔE*CIE94 ΔE*CIE2000  Remission spectrum  InkCheck: Color control of spot
colors
Color library with up to 20 color books and a total number of 25.000 reference
values
Memory capacity for 3.000 sample values and 300 reference values G7™
SpectroDens Premium
Same functions as SpectroDens Advanced and additionally:
Ugra/Fogra media wedge evaluation ISO-Check: Color control according to ISO
126 47
CIE L*u*v*  CIE L*C*h*uv  CIE xyY  DIN Lab99  Metamery index  Whiteness
Yellowness Pass/fail tolerance Opacity OBA-Check
Page 8
8
Manual TECHKON SpectroDens
1.2 Packing list
Contents of delivery
Measurement device SpectroDens Charging console with integrated
absolute white standard
AC adapter with universal plugs
for Europe, UK and USA
This manual with ISO 9000 compliant
certicate
Manufacturer certicate Micro-USB cable for PC connection CD with Windows software
TECHKON SpectroConnect
Contents of delivery of SpectroDens
Optional accessories
Exchangeable apertures Print control strips TECHKON TCS
(the digital print control strips TECHKON TCS can be downloaded free of charge at www.techkon.com)
Device upgrades for Basic- or Advanced models to higher functionality Color reference SpectroCheck WLAN module
Replacement parts
Charging console with white standard AC adapter Rechargeable battery
You will nd the technical specications in the appendix of this manual.
Page 9
9
CHAPTER 1 – How to use TECHKON SpectroDens
1.3 Design of SpectroDens
SpectroDens is a solidly designed measurement device which is very functional and easy to use. It is compact in size; the buttons can be reached with the right index nger when the device is held with the right hand. The LCD display is inclined towards the user, so the display information can be read clearly.
SpectroDens can be easily positioned on the measurement sample.
The device incorporates a high-resolution spectral sensor which analyzes the color spectrum of the measured sample. The spectral data is the base information for cal­culating densitometric and colorimetric data which is displayed. Densitometric values can be solid density, dot area, dot gain or trapping values. Colorimetric values are typi­cally CIE L*a*b*- or ΔE*-color information.
The USB-connector is for linking with a PC. The USB-cable with a Micro-USB plug, which ts into the connector at the back of the device, is part of the package.
Tracking wheels
Device label with
serial number
USB port for PC connection
Navigation buttons
Right hand grip zone
Measurement aperture
Measurement head
Green measurement button (also for switching on)
Device display
Page 10
10
Manual TECHKON SpectroDens
The charging console incorporates an absolute white standard, which should not be touched but should always be kept clean.
The above image shows how to connect the plug of the AC adapter with the charging console. The AC adapter can be used universally for 100 – 240 V and 47 – 63 Hz.
There are three adapter plugs for dierent countries. To change an adapter plug, just unlock the lock, pull the plug o the AC adapter and push in the new one. Make sure it is connected rmly.
Connect the mains plug to the power socket. The socket must be located near the facility and must be easily accessible. Do not use the power plug if the power cord is damaged.
Always use the original TECHKON SpectroDens AC adapter. The use of other power sources is hazardous.
AC adapter with three dierent mains plugs Device in the state of charging
UK
USA
Europe
1. Absolute white standard
2. Charging contacts for wireless, inductive battery charging
3. Low voltage connector for the plug of the AC adapter
1.
2.
3.
Page 11
11
CHAPTER 1 – How to use TECHKON SpectroDens
1.4 Working with SpectroDens
Switching on and o
SpectroDens is switched on by pressing the green measurement button. It is immedi­ately ready to measure. In this state power consumption is very low.
If no button is pressed within one minute, the device will switch o automatically. Switching it on again with the green measurement button will show on the display the last measurement done. There is no button to turn SpectroDens o.
Charging
The battery status is displayed as a battery symbol in the upper right corner of the LC-display, when the unit is turned on.
When the battery turns empty, the battery symbol turns to red. In this state several measurements still can be obtained; but it is time to put it on the charging console to avoid a total discharge and damage of the battery.
Charging starts automatically by placing the device on the charging console. Please keep the charging contacts on the console and under the device always clean. Five seconds after placing the device on the console the display will show an animated battery symbol. When the symbol displays a full battery, the charging is completed and the battery is fully charged.
Up to 10.000 measurements can be carried out with one battery charge. Completely charging an empty battery will take approximately four hours.
SpectroDens has a regulated charging management. This means it will power-charge an empty battery, so the device can be used again after a few minutes. On the other hand, there is no overcharging of full batteries.
Disposal instructions: Please dispose of the battery only in a designated recycling cen­ter.
You can park the device on the charging console when not in use. This way it is guar­anteed to have always an operational measurement instrument at hand.
Page 12
12
Manual TECHKON SpectroDens
How to measure
After switching on SpectroDens, it is instantly ready to measure. Just position the mea­surement aperture on top of the measurement sample and press shortly with the in­dex nger the green measurement button. Within one second the resulting measure­ment value will appear on the display of the device.
Please ensure that the device always has a rm stand on a at surface. There must not be a distance between the measurement head and the sample where light could pass through, because this can inuence the measurement. On the bottom of the device there are tracking wheels which allow to drive easily and fast to the measurement point.
The color of the background material underneath the measurement sample can have an inuence on the measurement result. Dierent technical standards describe which backing material to use.
In the printing industry the following procedure is widely used:
white backing for proofs and black backing for prints.
TIP:
By pressing the red Escape button and one of the Arrow buttons simultaneously (see p. 13), SpectroDens switches directly between the densitometric and the colorimetric measurement function that was last used.
Page 13
13
CHAPTER 1 – How to use TECHKON SpectroDens
The center of the display shows the current measurements.
1. Reference measurement
2. Sample measurement
3. Battery and charging status
4. Stack of paper icon for paper white calibration; CAL icon for absolute white calibration
5. DELTA-mode (= for comparisons)
6. Average function
7. Color library
8. Device settings 9 Measurement conditions
10. Measurement functions
Device settings with the display menu
Thanks to the interactive device menu, settings are very easy to adjust. The display is divided logically into a center and a frame. The center displays the measurement results. It is framed by interactive “soft keys” and status information displays. Pressing shortly on one of the arrow buttons will move an active cursor from “soft key” to “soft key” around the frame. An active “soft key” is indicated by a frame.
By keeping an arrow button pressed, the cursor travels from “soft key” to “soft key” at a constant speed until the arrow button is depressed again. By pressing the enter but­ton a function behind a “soft key” is selected and a selectable menu appears. Within such a menu again the navigation is done via the arrow buttons and the selections are made with the enter button. Some menu items or “soft keys” might be shaded in appearance and can not be selected, because they are only active in a higher perfor­mance package (Advanced or Premium) of SpectroDens.
Pressing the red, round escape button will always bring you one step back in the menu or will close a menu window. It can be compared to the escape button or undo­function on a PC. It will always bring you back to the measurement mode.
1. Escape button
2. Arrow buttons
3. Enter button
1. 2. 3.
4.
5.
6.
7.
8.
10. 9.
1.
2.
3.
Page 14
14
Manual TECHKON SpectroDens
2. Sound will activate an acoustic signal after each measurement.
3. With activated Automatic Paper White Calibration SpectroDens will always perform automatically a paper white calibration when a measurement on a paper white is made.
We recommend to activate this function only when the substrate to be printed is often changed.
1. The SpectroDens INFO screen informs about the type of device (Basic, Advanced or Premium), the software- and hardware­version as well as the serial number. It also indicates whether a WLAN module is installed and informs about device checks.
Menu: Device settings
The “soft key” with the cogwheel symbol opens the window for the device settings.
Within this window you can navigate with the arrow buttons and select with the enter button. The red escape button brings you back to the measurement mode.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
Page 15
15
CHAPTER 1 – How to use TECHKON SpectroDens
4. In the measurement function Automatic Density, the device recognizes %-patches,
switches automatically to the display of the dot area and shows the calculated values.
The automatic switch to %-values can also be turned o using the device setting % Measurement Automatic Density to avoid switching caused by uctuation. When the function is deactivated, the device will always show the density values and the dot area will not be calculated automatically.
5. SpectroDens can be connected to a WLAN network to send information via wireless data transmission to the PC. The WLAN module is available as optional accessory.
6. Active Functions allows the user to per­sonalize the device by disabling those of the measurement functions that are not currently needed. So the handling is concentrated on the essentials. If necessary, the inactive mea­surement functions are simply reactivated.
7. Sets the menu Language of the device.
8./9. All measurement values get a Date- and Time-stamp. This information will be visible when the data is transmitted to the PC soft­ware SpectroConnect.
10. For the measurement of densities, the Dis- play format CMYK can be selected. The device can display two or three decimals. The information in the chosen format is car­ried out in all density modes.
Menu: Device settings
11./12. The functions Density adjustment (SLOPE) and Absolute white reference val­ues are for the internal device adjustment.
The device is correctly adjusted by the fac­tory and changes of these settings should be made only in special cases (see page 20 f.: Set­ting of the Density adjustment (SLOPE)).
13. A Reset puts the device to factory settings.
Attention ! All stored measurement values and a recently made calibration will be de­leted and should therefore be saved on the PC if necessary, by using the Windows soft­ware SpectroConnect.
Only the factory-set calibration values are still valid. A reset may be useful, if SpectroDens does not work properly.
14. The information of Device Voltages is for device diagnosis by service personnel.
Page 16
16
Manual TECHKON SpectroDens
Menu: Measurement functions
When the “soft key” in the left section of the lower bar is activated, a window opens where you can select the measurement functions. The measurement functions listed below are described in detail in chapter 2.
Index of the measurement functions:
1. Automatic Density p. 33
2. Densities CMYK p. 36
3. Dot Area p. 37
4. Dot Gain/Contrast p. 38
5. SCTV p. 39
6. Trapping p. 40
7. Printing Curve p. 41
8. Printing Curve SCTV p. 42
9. Printing Plate p. 43
10. Slur and Doubling p. 45
11. Density Graph p. 47
12. ExPresso Mini p. 48
13. CIE L*a*b* p. 51
14. CIE Color Circle p. 53
15. G7™ p. 54
16. Remission Curve p. 59
17. InkCheck p. 60
18. CIE XYZ p. 62
19. CIE L*C*h* ab p. 62
20. ISO-Check p. 63
21. CIE L*u*v* p. 67
22. CIE L*C*h* uv p. 67
23. CIE xyY p. 67
24. DIN Lab99 p. 67
25. Metamerism-Index p. 68
26. Whiteness/Yellowness p. 69
27. PASS/FAIL p. 70
28. Media Wedge p. 71
29. Scan p. 72
30. Opacity p. 74
31. OBA-Check p. 75
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18
19.
20.
21.
22.
23.
24.
25.
26.
27.
28.
29.
30.
31.
Page 17
17
CHAPTER 1 – How to use TECHKON SpectroDens
Menu: Measurement conditions
When the “soft key” in the right section of the lower bar is activated, a window opens where you can set the measurement conditions which dene the parameters of the measurement, e. g. which lter will be applied. The measurement conditions 4. – 18. are described in detail together with the corre­sponding measurement functions in chapter 2.
1. The device has an integrated Polarizing l- ter, which can be activated or deactivated by a push of a button. Polarizing lters compen­sate for measurement dierences between glossy and non-glossy surfaces (e. g. freshly printed and dried sheets). The technical stan­dards in Europe recommend the use of such a lter for densitometric measurements. On the other hand, colorimetric measurements (e. g. L*a*b*-measurements) should be taken with­out a polarizing lter.
By selecting Auto from the menu, this will be taken into account and there will be automat­ically the right lter setting.
2. There are two possibilities for dening the White Reference: Density measurements usually are referred to the white value of the paper printed on. Colorimetric measurements
refer to an absolute white reference which is supplied as a ceramic tile on the charging console.
By selecting Auto from the menu, there will always be the right setting according to the measurement mode which is selected.
3. Depending on the regional technical stan­dards dierent Density Filter settings are required. SpectroDens as a powerful spectro­photometer has the great advantage that any lter specication can be exactly calculated out of the spectral data. There is no limitation known from conventional lter based densi­tometers with xed lters built in.
In Europe lter ISO E is to be selected as the standard lter.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
Page 18
18
Manual TECHKON SpectroDens
Menu: Calibration
Prior to every series of measurements a calibration should be carried out. After a cali­bration the device is long term stable, however it is recommended to perform a cali­bration before starting a new measurement task just to be sure the device is working properly.
For density measurements the calibration is carried out on paper white. This will en­sure that the density values are in direct ratio to the ink thickness which is applied on the paper. When there is a change in the paper type, a new calibration has to be done. Only in a few cases density measurements are referenced to an absolute white standard.
A calibration is as easy to carry out as a measurement: Select the “soft key” Calibration and place the device with the measurement head on top of a blank area of the paper sheet. The calibration can be started now by pressing either the enter button or the green measurement button.
After one second there will appear a window with the message that the calibration has been done. The device is now ready for measurements. The message window will disappear with the next measurement.
During the calibration process always take care that the instrument is on a at, stable surface and that the material backing the paper is the same as for the following mea­surements (white, gray or black). When the paper sample is printed on both sides, it is recommended to use a black backing material in order to avoid mismeasurements due to color shining through the back of the opaque paper.
Page 19
19
CHAPTER 1 – How to use TECHKON SpectroDens
1.5 Care and maintenance
SpectroDens is a highly-precise optical instrument. It is designed to work in harsh, industrial environments. However, it should be handled with care. Avoid mechanical shocks, heat, dusty or humid environments! Keep it always at a safe place, for example in the carrying case.
Cleaning
The measurement head with the optical system is sealed against dust and dirt. Take care that the visible, open measurement aperture is always free of dust. You can clean the measurement aperture with oil-free, clean compressed air and a brush used for cleaning photographic equipment. For easy cleaning the aperture can be taken o from the measurement head by unscrewing three screws with a special TORX T6 screwdriver (which is part of the package of a spare aperture set).
Clean the device casing and the display window only with a soft cloth and a non­aggressive plastic cleaner. Never use alcohol or chemically aggressive solvent-based cleaners which can destroy the surfaces. The same procedure is valid for the white standard which is integrated in the charging console. If the white standard should be defective, it has to be replaced completely.
Should the display window show small scratches, these can be polished out with spe­cial display polish greases as they are e. g. available for mobile phones and other elec­tronic equipment as well.
Please do not stick any labels at the bottom of the device. This could lead to faulty measurements, because the direct contact – which is important for the correct opti­cal eld of depth – might not be maintained. The battery charge can be aected by a label as well.
Page 20
20
Manual TECHKON SpectroDens
Device adjustment
Should SpectroDens do not work properly, rst check, if all instrument settings and measurement conditions are set correctly. Especially the settings of the Density ad- justment (SLOPE) and the Absolute white reference values inuence the measure- ment result.
Setting of the Density adjustment (SLOPE)
With the SLOPE correction values it is possible to adjust SpectroDens to the density values of other devices.
Activate the “soft key” with the cogwheel icon and select the menu item Density adjust- ment (SLOPE).
Select the menu item CYAN.
Enter the Cyan value, you nd on the calibra­tion chart. By pressing the arrow buttons the value can be increased or decreased.
In the same way enter the data given by the calibration chart for Magenta, Yellow and Black.
Page 21
21
CHAPTER 1 – How to use TECHKON SpectroDens
Select the menu item Make SLOPE calibra­tion.
Carry out a paper white calibration as de­scribed in the device display.
Now measure consecutively the solid patches of Cyan, Magenta, Yellow and Black.
A correct measurement is indicated by a green dot behind the respective color.
The SLOPE correction values are shown.
A reset of the SLOPE correction values is done by activating the menu item Reset SLOPE coecients. You reset the device back to its original factory settings.
Page 22
22
Manual TECHKON SpectroDens
Setup of the SpectroDens WLAN connection
To be able to use the WLAN function in SpectroDens, a WLAN module has to be installed. The module is available as accessory from TECHKON. It is a small electronic board which can be inserted into the instrument easily.
Assembly of the WLAN module
SpectroDens WLAN module
To insert the module the casing has to be opened.
Lift the upper case carefully and put the two casing parts side by side on the table.
Page 23
23
CHAPTER 1 – How to use TECHKON SpectroDens
The slot into which the WLAN module has to be inserted is located on the main board.
Due to its reverse battery protection the board cannot be inserted in a wrong way.
After the WLAN module is inserted in its slot, connect the battery cable again. The plug has to clip into the socket.
When putting the casing together, make sure the WLAN module ts properly in its com­partment.
Tighten carefully all four screws on the bot­tom of the device. Now the instrument is ready to use.
Page 24
24
Manual TECHKON SpectroDens
In the upper left corner of the device display the WLAN symbol will be shown when acti­vated.
The color of the symbol depends on the sta­tus. Gray: Searching Blue: Connection established
To be able to use the WLAN connection, the function has to be activated in the settings of the instrument.
Notice: To save power, it is recommended to turn o WLAN if not required.
After a successful connection with the soft­ware SpectroConnect, the WLAN information can be checked in the SpectroDens INFO window.
Conguration of the device
Page 25
25
CHAPTER 1 – How to use TECHKON SpectroDens
Setup of the WLAN connection
Start the software SpectroConnect, select SpectroDens settings, turn on the instrument and connect it with the USB cable to the PC.
The symbol to the right of the serial number indicates the type of connection; in this example it is an USB connection.
The WLAN settings folder is available after connecting the instrument via USB cable to the PC. To establish a WLAN connection, please start the WLAN wizard.
Page 26
26
Manual TECHKON SpectroDens
The WLAN wizard will guide you trough the setup process. Make sure you have all information about your WLAN network available.
Two types of WLAN connections are possible: Infrastructure and Ad-Hoc. Infrastructure should be selected if a WLAN network is already installed in the working area. Ad-Hoc allows a direct connection to a PC using its WLAN module or any WLAN USB stick.
Infrastructure
Select Infrastructure and click on the Next button.
Page 27
27
CHAPTER 1 – How to use TECHKON SpectroDens
The wizard will show all WLAN networks which are available.
Select your network and click on the Next button.
Should your network not be listed you can try again by clicking on Search again using SpectroDens.
Please enter the IP-Address, the netmask and the password of your WLAN network in order to connect. You can enable DHCP if your network is supporting this function. Then a free IP-Address will be used automatically.
After entering the information, please click on the Next button.
Page 28
28
Manual TECHKON SpectroDens
SpectroDens is now looking for the WLAN network and if available will establish a connection.
The gray color of the symbol is indicating the search function. Once the WLAN network is found and the connection is established, the color of the symbol will change to blue.
SpectroDens is now connected to the WLAN network and ready to be used wirelessly.
Page 29
29
CHAPTER 1 – How to use TECHKON SpectroDens
The settings will be applied and shown in the WLAN settings.
By activating the function Automatically search for WLAN SpectroDens the connection will be established automatically when used next time.
Now you can disconnect the USB cable. The connection via WLAN will be indicated by the symbol next to the serial number. The blue color of the symbol indicates an active connection.
Page 30
30
Manual TECHKON SpectroDens
Ad-Hoc connection
To establish an Ad-Hoc network select this function during the wizard setup.
The IP-Address, the netmask and the name (SSID) of the network are predened in this case and will be shown for your information.
Click on the Next button to continue.
Page 31
31
CHAPTER 1 – How to use TECHKON SpectroDens
Use the software of your WLAN network card or WLAN USB stick to setup the WLAN connection to your device
Select the WLAN network named after the serial number of your device.
Conrm the settings and check the connection.
Once the connection with your network card or WLAN USB stick is established you have to go back and click on Repeat search to connect SpectroConnect to the WLAN network.
A successful connection will be indicated by the blue WLAN symbol.
Continue in the same way as described for the infrastructure network (see p. 29).
Page 32
32
Manual TECHKON SpectroDens
Device information and safety instructions
The use of this device in a dierent way than recommended by TECHKON GmbH can endanger the device safety.
WARNING: This device may not be used in potentially explosive atmospheres.
Please protect your eyes and do not look directly into the optics when the device is switched on.
Operating conditions: Ambient temperature: 10 to 35 °C, max. 30 to 85 % rel. humidity (non-condensing); storage temperature: -20 to 50 °C.
Reset
A device reset sets the device back to factory settings. All calibration data (including SLOPE values), stored measurement values, reference-, target-, and tolerance-values will be deleted.
If the reset procedure in the settings window is not possible, a hardware reset has to be carried out. The battery plug has to be unplugged for a couple of seconds and then to be connected again.
Warranty
The warranty for TECHKON products is 24 months starting with the date of purchase. Consumable and wearing parts are excluded from this. The invoice is the certicate of warranty. The warranty is invalid if the damage is caused by inadequate use of the device.
Should a TECHKON product do not work according to the specication, please contact us before sending us the device. In most cases we can solve the problem over the phone or via E-mail.
Inspection intervals
TECHKON SpectroDens is maintenance free. We recommend to validate the com­plete functionality of the devices in a 24 months time interval in the TECHKON ser-
vice center. Please send the device always securely in the carrying case with complete accessories. For a at fee the device will be cleaned, checked and re-calibrated. In case a repair or exchange of components should be necessary we will inform you in advance.
Page 33
33
CHAPTER 2 – Measurement functions
In this example the measurement has been taken on a Cyan patch. SpectroDens recog­nizes the type of process color and displays the solid density value of Cyan, which is dis­played with the “C” symbol. Values for Cyan (C), Magenta (M), Yellow (Y) and Black (K = key color) can be shown.
The density value is a value without a unit. It corresponds to the amount of ink applied on the paper surface. The higher the density value the higher the thickness of the ink layer. The correct density value is dependent on the printing parameters (paper, ink, printing process). Usually, density values are between
1.00 and 2.00.
The second measurement has been taken on a Magenta patch. The former Cyan measure­ment still gets displayed. Measurements of a specic process color will be overwritten as soon as a new measurement of this specic color will be made.
Chapter 2
Description of the measurement functions
2.1 Measurement functions of SpectroDens Basic
Automatic Density
SpectroDens Basic comprises all functions related to density measurement. A very practical feature is the Automatic Density mode. By a push of the green measure­ment button the relevant measurement values are displayed. Depending on the type of measurement patch this can be:
Solid density D Dot area in % Gray balance
K
C M Y SLUR C+MKC M Y
1
80%40%
BAL
2
K
C M Y SLUR C+MKC M Y
1
80%40%
BAL
2
Page 34
34
Manual TECHKON SpectroDens
For the display of dot area (or dot percent­age) in %, two measurements have to be made: First the solid density patch and then the %-patch, which is in this example a 80 %­patch. The resulting dot percentage is in this case 91.9 %; the referring dot gain is 11.9 %:
91.9 % - 80 % = 11.9 %
Should a solid density value be displayed instead of the desired dot percentage value, then the density measured rst was signi­cantly higher than the second one. In this case we recommend to delete the displayed value by pressing the red escape button and to measure again.
When making a measurement on a gray balance patch, the referring density values for all four process colors are shown in a bar graph. The values for Cyan, Magenta and Yel­low should be in close range to each other to ensure a neutral gray balance without a color hue.
It is important that the gray balance patch, on which the measurement is taken, comprises the right %-values to achieve a neutral gray when printed correctly. The process standard ISO 12647 for oset print denes the values as: C = 50 %, M = 40 % and Y = 40 %.
The value for Black indicates how dark the gray color is. The higher the value for Black (= K) the darker the sample.
1. Measurement 2. Measurement
A gray balance patch is composed of dierent portions of Cyan, Magenta and Yellow. It does not include black ink.
K
C M Y SLUR C+MKC M Y
1
80%40%
BAL
2
M YKC M Y 80%40%
0KCBAL
BAL C+M BAL
BALM Y
S/D
Page 35
35
CHAPTER 2 – Measurement functions
In DELTA Mod e (= comparison mode) the measured values are compared to reference values. The mode is activated by navigating to the “soft key” with the triangle symbol and selecting it.
In the example on the left side a Yellow patch was measured.
The measured solid density is:
Y = 1.25
This is 0.20 less than the set reference (= tar­get) density of Y = 1.45. The measured value is outside the limits of the set tolerance of ± 0.06. Therefore it is marked with a cross symbol in the round icon.
A checkmark symbol would be shown in case of a measurement within the tolerance limits.
The setting of reference values and toler- ances is done in the menu under the item measurement conditions.
In the Density Reference Setup up to ve data sets (REF 1 – 5) for reference (= target) densities can be edited. Reference densities recommended for the ve dened paper classes are already pre-set.
The tolerances dene the lower and upper limits around the reference values. In this ex­ample for Yellow:
1.45 + 0.06 = 1.51
1.45 - 0.06 = 1.39
K
C M Y SLUR C+MKC M Y180%40% BAL
2
Page 36
36
Manual TECHKON SpectroDens
In this example, a solid Magenta patch was measured. The solid density is:
M = 1.47
But why are densities for the other colors shown as well ? Shouldn’t they have a value of 0 ? No. These are the so-called partial densi- ties. No printing ink is 100 % spectrally pure. There will always be partial colors besides the actual main color.
This function also makes comparison be­tween measurements and reference values. In this example the reference value was not taken from the reference memory but was measured just before by putting the “soft key” to Reference:
1. First select via “soft key” Reference and make a measurement on a patch which will be the future reference patch.
2. Now activate via “soft key“ Sample and make a measurement on a sample patch. A triangle symbol in front of the measurement value indicates that delta values (dierence between reference and sample) are shown. Future sample measurements will be com­pared to the once measured reference values as well.
3. Now activate the DELTA Mode by select­ing the triangle symbol.
The function Densities CMYK is especially suitable for the measurement of multicolored composed CMYK colors, e. g. in logos, to nd out the reason for possible color shifts.
For information about the ∆Dmax-function see page 47.
Densities CMYK
In the measurement function Densities CMYK all four partial densities for Cyan, Ma­genta, Yellow and Black are displayed.
Patch for reference measurement
Patch for sample measurement
Page 37
37
CHAPTER 2 – Measurement functions
Dot Area
For the function Dot Area (= dot percentage) two measurements have to be carried out. First a measurement on a solid patch and then a measurement on a %-patch (= screen patch). The dot area value is calculated with the Murray-Davies formula.
1. First place the measurement head above a solid patch and start the measurement by pressing the green button. The solid density value will be displayed.
2. Now make a second measurement on a %-patch and the dot area in % will be shown.
In this case a measurement has been taken on a 40 %-patch.
The resulting dot gain can be easily calculated: 59 % - 40 % = 19 %
The dot gain value gives the information to what degree a point is enlarged (or de­creased) during each process step of the re­production process chain.
The double arrow on the left side of the de­vice display indicates which measurement has to be done next.
Should several dot area values refer to the same solid density, this value has to be mea­sured only once and can be saved for con­secutive %-measurements. After a solid den­sity measurement, just keep the enter button pushed for approximately 3 seconds until the density value is framed. Now an unlimited number of %-measurements can be made. Deactivating is done by keeping the enter button pressed again.
K
C M Y SLUR C+MKC M Y
1
80%40%
BAL
2
1. 2.
Page 38
38
Manual TECHKON SpectroDens
Dot Gain / Contrast
The function Dot Gain / Contrast is closely related to the previously described func­tion of dot area. Here also, rst a solid density measurement is carried out and then %-measurements follow.
In the lower part of the device display there is shown which patch has to be measured next.
In this case there just has been measured a solid density patch with a density of 1.84.
The double arrow points to the 80 meaning an 80 %-patch has to be measured next.
After the last %-measurement has been done (in this case the 80 %-patch) additionally to the Dot Gain value the Printing Contrast will be calculated and displayed.
The round escape button will always lead back to the previous step in case a wrong measurement has been made.
The setting for the %-patches to be measured is done in Dot Gain Setup.
Up to three percentage values can be edited. The TECHKON TCS print control strips include two dierent %-patches; 80 % and 40 %.
K
C M Y SLUR C+MKC M Y180%40% BAL
2
1. 2. 3.
Page 39
39
CHAPTER 2 – Measurement functions
SCTV
The Spot Colour Tone Value (SCTV) function, which is dened in ISO 20654:2017, pro­vides a precise colorimetric quality inspection for spot colors. SpectroDens supports SCTV for single tone value measurements and the SCTV printing curve (see page 42).
1. First place the measurement head above a solid patch and start the measurement by pressing the green button. The solid density value will be displayed.
2. Now make a second measurement on a %-patch and the dot area in % will be shown.
In this case a measurement has been taken on a 80 %-patch.
The resulting dot gain can be easily calculated:
81.2 % - 80 % = 1.2 %
The dot gain value gives the information to what degree a point is enlarged (or de­creased) during each process step of the re­production process chain.
The double arrow on the left side of the de­vice display indicates which measurement has to be done next.
Should several dot area values refer to the same solid density, this value has to be mea­sured only once and can be saved for con­secutive %-measurements. After a solid den­sity measurement, just keep the enter button pushed for approximately 3 seconds until the density value is framed. Now an unlimited number of %-measurements can be made. Deactivating is done by keeping the enter button pressed again.
1. 2.
80%40%S2S10BALS2S180% 40%K C M Y S2S1KKC M YKC M Y
Page 40
40
Manual TECHKON SpectroDens
Trapping
With the feature Trapping there can be checked, how good two layers of ink which are printed over each other represent the resulting mixed color.
There are no standard values for the trapping measurement. The values which can be obtained depend on the printing process and particularly on the type of ink and paper used. Typical values for standardized print with Paper Class 1 (glossy, coated) are: C+M > 60 %, M+Y > 72 % and C+Y > 85 %.
SpectroDens oers extended trapping calculation according to Professor Ritz and Fe­lix Brunner, which ensure a better comparison between C+M, M+Y and C+Y measure­ments. An alternative determination how well two separate colors overprint can be checked with a colorimetric function, such as the L*C*h*-feature, where the L*- and C*-values give the information of the printable gamut.
Three measurements are necessary: Two measurements on each of the two separate process colors and nally one measurement on the resulting overprinted patch.
Here in this example rst Cyan and second Magenta has been measured. The nal mea­surement is done on a blue patch as the overprint of Cyan and Magenta. The trapping value is 73.2 %. The higher the value the bet­ter the overprint of two colors.
Additionally to the trapping value in % there are the densities for the separate color patches displayed. The densities are always displayed with the lter channel of the sec­ond printed ink; in this case Magenta. This explains why the density value for the rst printed ink Cyan is so low. SpectroDens will recognize automatically the order of colors as they are printed. Switching the rst and sec­ond measurement therefore will lead to the same result.
K
C M Y SLUR C+MKC M Y
1
80%40%
BAL
2
1. 2. 3.
Page 41
41
CHAPTER 2 – Measurement functions
Printing Curve
The measurement function Printing Curve displays a complete transfer curve, which describes how screen dots (and tonal values) are transferred during the several pro­cess steps of the reproduction process.
First, make the required settings in the Print- ing Curve Setup:
1. Tolerance range can be set according to standard values (in this case the ISO Paper Class 1/2 was selected)
2. Dot (screen) frequency (only relevant for the display of tolerance values)
3. Increments of the wedge to be measured, e. g. in 5 % steps
In the rst step a 100 %-(solid) patch is mea­sured. SpectroDens recognizes automatically the color and displays the solid density. In this case D = 1.50 for Cyan.
The double arrow in the upper left corner in­dicates that the next patch to be measured will be a 90 %-patch. Now all other pre-set %-patches can be measured on the measure­ment wedge.
Step wedges with separation bars (as shown) can be measured by scanning. Just keep the green start-button pressed during the whole scanning procedure.
1.
2.
3.
1.
2.
Page 42
42
Manual TECHKON SpectroDens
Now the corresponding current dot area is indicated in the lower right section of the dis­play. In this case: + 7.3 % dot gain is displayed at 90 % dot percentage. The shaded area close to the curve shows the tolerance range. The curve should be within this area.
A wrong measurement can be easily deleted by the red escape button and the device will be ready again to read the patch correctly.
When using the scan function we recom­mend to use a patch width of 5 x 5 mm (min. 4 x 4 mm) with the standard aperture and a patch width of 4 x 4 mm (min. 3 x 3 mm) with the exchangeable 1,5 mm aperture.
Printing Curve SCTV
The measurement function Printing Curve SCTV displays a complete transfer curve of a spot color, which describes how screen dots (and tonal values) are transferred during the several process steps of the reproduction process.
The measurement of the Printing Curve
SCTV is carried out analogous to the regular Printing Curve (described on page 41 f.).
Only the calibration of the device is done on paper as well as on the absolute white stan­dard.
Page 43
43
CHAPTER 2 – Measurement functions
Printing Plate
Beside energy and focus of the laser, dot gain is the most important information which has to be checked in plate exposure. Special plate measuring instruments working with video analysis have been developed for this application; for example TECHKON SpectroPlate. Compared to densitometers and colorimeters they oer big advantages. Nevertheless it is basically possible to make measurements with a densitometer on a printing plate. In doing so one should regard the following: Variations in the coating of the printing plates may result in strong inuences on the measurements, especially in the crucial range under 10 %. Variations of 0.02 D can cause errors up to 6 %. The values measured with a densitometer do not correspond with the geometric dot area, due to the fact that the densitometric measurement process is subject to optical inuences, which increase the measurement value. Therefore densitometers measure always a higher value than video analysis devices. But the measurement value of the densitometer can be taken as a reference value for the tonal value adjustment.
For the measurement of the zero point (1.), the solid patch (2.) and the screen patch (3.) you should select areas on the printing plate which are located as close as possible to each other.
1.
2.
3.
The measurement feature Printing Plate measurement allows reading values directly on printing plates. The process is comparable to reading on printed paper.
Before taking a measurement it is important, that the rubber coating is washed o the printing plate, because otherwise light gets lost and the measurement result will be distorted. In addition a correction factor, the so-called Yule-Nielsen factor, can be set according to the material characteristics of the printing plate. Usually the plate manufacturer will furnish the information about the factor for the dierent plate types.
Page 44
44
Manual TECHKON SpectroDens
The setting of the Yule-Nielsen factor is done in the settings window of the measurement conditions. If there is no information about the factor available from the documentation furnished with the plates, we recommend to keep the value in a neutral range at n = 1.00.
The measurement is performed as known from measuring on printed paper:
1. First, calibrate on a light, not imaged area of the printing plate, as it would be a paper white calibration.
2. Now measure a dark, 100 %-imaged patch. SpectroDens will select one of the four CMYK channels where the contrast signal is highest depending on the color shade of the plate. In this case the Cyan channel gave the highest signal, because the plate is blue/violet.
3. Now the measurement on a %-patch is carried out and displayed. Additional %-mea­surements can be easily made. It is not neces­sary to measure the solid density again, be­cause it is already saved, which is indicated by the frame around the value.
C 40%
1. 2. 3.
Page 45
45
CHAPTER 2 – Measurement functions
Slur and Doubling
With the measurement function Slur and Doubling printing deciencies, caused by inexact, geometrical transfer of the image from the plate to the paper can by evalu­ated.
Slurring means that the printed dot deforms from the transfer of the printing plate to the rubber blanket or from the rubber blanket to the paper. The eect of doubling will be present when a printed dot is transferred a second time with less intensity geo­metrically slightly oset. This ghost image is caused by ink being re-transferred to the rubber blanket which should have stayed completely on the printed paper.
To make the slur and doubling measurement possible, it is necessary to have on the color bar vertical and horizontal lined patches with identical line thickness as it is the case with the print control strip TECHKON TCS.
Three measurements have to be carried out:
1. The rst measurement on the solid patch; in this case of Magenta.
2. The second measurement on the horizon­tal lined patch of the same process color.
After the measurement of the horizontal lined patch the dot area of the patch is calculated and stored (see the following gure).
K C K C M Y
BAL
M+Y
S/D
1.
K C K C M Y
BAL
M+Y
S/D
2.
Page 46
46
Manual TECHKON SpectroDens
0 % is the best achievable factor, meaning the dot transfer is perfect, the horizontal and vertical lines of the measurement patches have the same thickness and that nei­ther slur nor doubling eects are present. Values up around 5 % still are acceptable. %-values higher than 10 % indicate signicant geometrical errors in print due to slur or doubling eects. There are no standard values for the maximum allowable %-factor.
3. The third measurement on a vertical lined patch. After the measurement of the vertical lines again the dot area of the patch is calcu­lated and stored (see the following gure).
As a result the so-called slur/doubling factor in % is displayed.
K C K C M Y
BAL
M+Y
S/D
3.
Page 47
47
CHAPTER 2 – Measurement functions
Density Graph
The function Density Graph enables the exact display of density values for any spot color. Usually the standards dene density lter characteristics only for the process colors Cyan, Magenta, Yellow and Black. Thanks to the spectral measurement of SpectroDens, a virtual spectral density characteristic for any spot color can be ob­tained. The great advantage is that the process control of spot colors can now be as easily achieved as known from the four process colors.
In the following example a measurement has been made on a blue spot color.
The density graph is derived from the re­mission spectrum and represents the den­sity spectrum for wavelengths from 400 to 700 nm. The peak of the curve displays where the spectral density value has to be taken.
In this case the spectral density for the blue color is D = 1.57 at a peak wavelength of 580 nm.
Only spectral densities relating to the same peak wavelength (Dmax) can be compared.
SpectroDens displays the spectral density automatically as well in the Densities CMYK measurement mode.
The density values for CMYK show to what ex­tent the spot color can as well be registered by the CMYK-color related standard density lters.
Here in this case, the blue color is close to a Cyan, therefore the density value in the Cyan channel is quite high.
Page 48
48
Manual TECHKON SpectroDens
ExPresso Mini
The measurement function ExPresso Mini allows the fast control of the printing qual­ity during the printing process. The last ten consecutive measurements of a printing color can be taken on a print control strip of a printed sheet for the ink zone control. In addition ExPresso Mini provides the opportunity to get an overview of the stability of the print run, by measuring the last ten printed sheets.
Advanced functions for the fast and comprehensive control of the printing quality during the printing process features the software TECHKON ExPresso 4.0 in connec­tion with the automatic measurement system TECHKON SpectroDrive.
ExPresso Mini can display the last ten mea­surements per color.
When, after the tenth measurement, the dis­play is lled up with measurement values, the already existing values shift with every new measurement one step to the left, and the rst value falls away.
The reference line marks the average value of all measurements.
The tolerance area is dened with ± 0.10 den­sity.
The average value is recalculated after every measurement.
Page 49
49
CHAPTER 2 – Measurement functions
In ExPresso Mini the Delta Mode can be ac- tivated.
Taken measurements will then be compared to reference densities, pre-set in the Density Reference Setup.
The reference data and the result will be dis­played. The reference line is always the refer­ence value and the corresponding tolerance.
The gure describes the situation inside and outside of the tolerance and the situation outside of the graphic space.
Page 50
50
Manual TECHKON SpectroDens
The measurement conditions M0, M1, M2, M3 according to ISO 13655 -2009
Many types of paper contain optical brighteners or pigments to improve the visual performance of the paper. The brighteners and pigments are used to make the paper look brighter and whiter. But for the accurate color adjustment they are quite a chal­lenge, because the type of ambient light inuences the color perception in most cases to a large extent. To be able to evaluate these printing substrates precisely when mea­suring, new standards have been dened concerning the measurement conditions.
The following measurement conditions are selectable in the measurement parameter menu of the SpectroDens color measurement device.
Measurement condition M0 is used when the printing substrate contains no or only little optical brighteners. The measurement light corresponds to the standard illumina­tion type A and the UV component of the source, which activates the optical brighten­ers, is not specied. Almost every color measurement device in the graphic industries works so far with this measurement condition. This measurement condition should also be used, when measurement values have to be compliant to older devices.
Measurement condition M1 was dened to activate optical brighteners with a speci­ed component of UV light, to be able to evaluate the eect of the brighteners more precisely. The measurement light conforms to the illumination type D 50 with a speci­ed UV component.
Measurement condition M2 describes the measurement without UV light. In older devices this measurement condition was realized using an UV-Cut lter. The compari­son of the measurement value M1 to M2 shows clearly how strong the eect of optical brighteners is.
Measurement condition M3 is mainly used for densitometric measurements and only to a less extent for colorimetric measurements. It describes the use of polarization lters which reduce surface reections to minimize measurement deviations between wet and dry printing sheets.
Page 51
51
CHAPTER 2 – Measurement functions
The thematic separation between densito­metric and colorimetric functions is displayed by a dotted line within the measurement function window. CIE stands for the interna­tional color institute Commission Internatio­nale de l’Eclairage which publishes the stan­dards for colorimetry.
The most popular color system in the printing industry is the CIE L*a*b* color space. Every color is precisely described by three values: The L*-value stands for the luminance and can have values between 0 (a theoretical, abso­lute Black) and 100 (a theoretical, ideal White). The a*-value describes the color value on the Green/Red-axis (- a*: Green, + a*: Red) and the b*-value the color value on the Blue/Yellow­axis (- b*: Blue, + b*: Yellow).
Colorimetric L*a*b*-measurement on a blue sample. The blue color has a low luminance of L = 47.21 and a low portion of Red.
2.2 Additional functions of SpectroDens Advanced
CI E L* a*b *
The previously described measurement functions all were based on densitometry.
Thanks to the spectral performance of SpectroDens it is possible to calculate and dis­play colorimetric values as well. Measurements in colorimetry modes have the advan­tage of an absolute description of color based on characteristic values.
L* = 0
L* = 100
C*
h*
h* =
0h* = 180
L* = 50
Page 52
52
Manual TECHKON SpectroDens
In colorimetry the white reference is absolute in contrary to densitometry where the white is usually referenced to paper white. Before making colorimetric measurements, an absolute white calibration has to be carried out by placing SpectroDens on the charging console and calibrating on the round, white ceramic standard.
Colorimetric measurements can only be com­pared when the measurement conditions are the same:
Pollter: o White reference: Absolute Illuminant: D50 Observer: 2°
When selecting Auto in the polarizing lter setting, there will be automatically the polar­izing lter inactivated when making colori­metric measurements.
The color distance ∆E* describes how close two dierent colors match. A value of 0 means that they are identical. ∆E*-color dierences in the range of 5 result in con­siderably perceptible deviations when reference and sample are close together and observed at the same time.
For measurements of the color distance ∆E* of two dierent colors, activate the D ELTA Mode rst by selecting the “soft key” with the triangle icon.
Now two measurements are made:
1. Reference measurement
2. Sample measurement
The color distance ∆E* is displayed with two digits. Additionally the components ∆L*, ∆a* and ∆b* can be seen.
Color comparison:
Reference Sample
Page 53
53
CHAPTER 2 – Measurement functions
Select the CIE Color Circle Setup function out of the list of measurement conditions. Then choose L*a*b* or L*C *h *.
As a result the CIE Color Circle feature dis­plays L*a*b*, L*C*h* or L*C*H* values.
The measurement function CIE Color Circle, (as well as CIE XYZ and Remission Curve), can be operated in absolute mode with DELTA ON or in comparison mode with DELTA OFF.
CIE Color Circle
The function CIE Color Circle shows the L*a*b*-color values as well. Optionally L*C*h*­values can be displayed. Additionally there is a graphic display and a color is shown as a spot within the color circle.
Page 54
54
Manual TECHKON SpectroDens
G7™
The measurement function G7™ supports the calibration of presses and proofs ac­cording to the method developed by the IDEAlliance (International Digital Enterprise Alliance).
The G7™ mode can be selected out of the list of measurement functions.
In the pop-up window of the measurement conditions the item G7™ Setup can be cho­sen. Now target values can be edited and printing conditions can be selected.
Edit the target values for SC, HR and HC for K and CMY. For Auto Set a*/b* use half a*/b* values of measured paper as target value.
You will nd a comprehensive description of the SCCA function on page 59.
For the comparison to the right ISO reference values the used paper type must be selected.
Page 55
55
CHAPTER 2 – Measurement functions
In order to take correct measurements an ab­solute white calibration must be performed and will be conrmed in the display when done.
A measurement of the paper white is neces­sary for correct calculations.
Now the instrument is ready to measure. The selected paper type and backing are indicat­ed in the display.
When measuring solid inks the values are compared to the ISO reference values. The ∆E*-value is indicated and the symbol shows if the value is within or out of tolerance.
The recognized color and L*a*b*-value of the measurement are displayed as well.
The selected paper type is always indicated.
Page 56
56
Manual TECHKON SpectroDens
The ∆E*-value is calculated as long as the dif­ference is in a reasonable range.
If the dierence exceeds a limit, then only the measured density value is displayed without comparison.
This procedure works for all CMYK colors.
The selected paper type is always indicated.
The color detection works for trapping patch­es as well. The recognized type of patch is dis­played in the printing order of the two colors.
The measured value is compared with the ISO reference and the dierence is displayed as ∆E*.
If paper white or any other undened patch is measured then all four densities are displayed simultaneously.
Page 57
57
CHAPTER 2 – Measurement functions
If a gray balance patch is measured all four densities are displayed in a bar chart.
If the value is close to the SC_cmy “neutral density” target it will be displayed with rec­ommended adjustments.
SC stands for “Shadow Contrast”.
The HR_cmy is detected as well. The recom­mended adjustments are shown by the ar­rows and numbers.
HR stands for “Highlight Range”.
As soon as the value is close to the HC_cmy “neutral density” target it is recognized and displayed.
HC stands for “Highlight Contrast”.
The recommended density adjustments for black patches of SC_k, HR_k and HC_k are displayed in the same way.
Page 58
58
Manual TECHKON SpectroDens
1. Detected color
2. Measured density
3. ∆E*-dierence compared to ISO reference
4. Indication in/out tolerance
5. Paper type
1. Detected color
2. Direction of adjustment
3. Recommended adjustment
4. Detected patch type
5. Measured value
6. Paper type
1. Color
2. Recommended adjustment
3. Adjustment as bar graph
4. Direction of adjustment
5. Detected patch type
6. Dierences of HR_cmy and a*b* values compared to target values
7. Paper type
1. Indicator for neutral density ND CMY
2. Direction of adjustment (+/-)
3. Recommended adjustment
4. Detected patch type
5. Measured value
6. Paper type
1.
2.
3.
4.
5.
1. 2. 3.
4. 5.
6.
1.3.2.
4.
5.
6.
7.
1. 2. 3.
4. 5.
6.
Page 59
59
CHAPTER 2 – Measurement functions
Remission Curve
The function Remission Curve oers the complete spectral graph.
Remission values are shown in 10 nanometer steps in the spectral range of 400 to 700 nm. A remission value has a range between 0 (no remission, absolute darkness) and 1 (total re­mission of the emitted light).
The remission curve is the most exact and to­tal description of a color.
SCCA reference value correction
The practice of following print standards like ISO 12647 that pre-denes the combination of ink, paper and target reference colors for commercial printers to use in production is in­valuable. But if the paper does not fall within the specied target values the accuracy of the press run is greatly diminished. The more out of spec the paper is the worse the color match will be.
One member of the IDEAlliance Print Prop­erties & Colorimetric Committee, David Mc­Dowell, developed a formula that recalculates the data set with new aim points based on the specic paper color being used on press. It is called SCCA (Substrate Corrected Colori­metric Aims).
Techkon became aware of this new specica­tion and implemented the ability to measure the paper and based on those measurements and using the SCCA formula calculates new aim points right on the instrument. This fea­ture will allow printers to quickly and easily achieve an acceptable color match regardless of the presence of optical brightening agents in their substrates.
Page 60
60
Manual TECHKON SpectroDens
InkCheck
The measurement function InkCheck combines in a very easy way the advantages of colorimetry with the familiar densitometry. Thanks to this helpful function, working with the process standard ISO 12647 is simply achieved. The colorimetric deviations are reported and additionally recommendations for amounts of ink supplied in terms of density values are given.
In contrast to the measurement function ISO-Check (see p. 63) the measurement function InkCheck is not only able to check the CMYK print ink, but also spot colors, provided that the appropriate spot colors were chosen as references in advance. In ad­dition to the color books which are included in the contents of delivery, for example the HKS-color books, you can create your own color books for spot colors and use them as a reference.
The recommendations for amounts of ink supplied in terms of density values for spot colors are based on CIE L*a*b* measurements.
First, a color book which holds the reference color data has to be selected in the reference menu. The reference values for the 5 paper classes dened in the ISO standard are pre-set in the device memory. The data set “Proof” is for measurements on white backing and the data set “Print” for measurements on black backing.
Notice: If they can not be found in the device memory, they can be uploaded with the Win­dows software SpectroConnect.
First, a calibration on paper white has to be made. Position the device on the paper white and conrm the request to measure the pa­per white by pressing the green measure­ment button.
Page 61
61
CHAPTER 2 – Measurement functions
Now the device is ready for measurements on solid color patches. In this case a yellow patch was measured.
Density trend: It is recommended to increase the density by 0.11.
Color deviation: The color distance to the ref­erence is 4.8 ∆E*. After following the density trend it can be as low as approx. 1.4 ∆E*.
A closer color match will be limited because the paper and ink characteristics do not allow to get closer to the reference standard.
Please note that the density values are ob­tained without the polarizing lter activated and therefore are lower as values obtained with polarizing lter. Please ensure that the right measurement conditions are set.
Page 62
62
Manual TECHKON SpectroDens
CIE XYZ
The CIE XYZ will display standard color values.
CIE L*C*h* ab
The CIE L*C*h* ab display is closely related to the CIE L*a*b* display (see p. 51). Instead of Cartesian coordinates, polar coordinates are used. The value C* stands for Chroma and the angular value h* for hue.
L*C*h*-values are very popular, because they describe colors in a simple way. The further the color is located from the center of the color circle, the more saturated and vivid it appears.
XYZ standard color values are (beside L*a*b*­values) a dierent way of describing the color values. They are not as practical to use and therefore not as widely used in the printing industry.
Page 63
63
CHAPTER 2 – Measurement functions
2.3 Additional functions of SpectroDens Premium
ISO-Check
The measurement function ISO-Check provides the fast check of a print with regard to the compliance with the ISO standard values ISO 12647-2. The L*a*b*-values, the dot gain and the dot spread of the printing inks CMYK are documented.
First the ISO Check Setup in the menu sec­tion of the measurement conditions has to be set.
You can select the control strip and the pa­per type. If you choose the TECHKON print control strip you can select two dot values additionally.
ISO 12647-2 denes the following standards:
Dot frequency 150 lpi (60 L/cm)
Dot values 80 % (75 %) and 40 % (50 %)
In the next step the paper white must be measured. When the message for measuring the paper white is displayed, please conrm it by pressing the green measurement button.
When the ∆L*a*b*-values are marked with checkmarks, the paper white is in accordance with the ISO-standards.
A special ISO/PSO control strip (as shown) can also be measured by scanning. Just keep the green start-button pressed during the whole scanning procedure.
......
Page 64
64
Manual TECHKON SpectroDens
When the solid density is measured, the com­pliance with the tolerances is checked.
In this example the measurement has been taken on a black solid patch.
At the same time the recommendation how to adjust the density is calculated.
The smallest possible ∆E*-value is calculated in advance.
The measurement function ISO-Check recog­nizes, whether the given tolerances can be achieved with the set of print ink in use. If this is not the case, another set of print ink must be used.
Page 65
65
CHAPTER 2 – Measurement functions
After the measurement of the solid patch, you can measure the %-patch, which you have dened in the setup.
The dot gain is checked in regard to its con­formity to the ISO-standards.
Concerning the midtone the spread will be displayed additionally (CMY).
The deviation of the colors CMY is allowed to reach ± 4 % in the middle-tone of 40 % per ink . But to avoid that the gray balance is disturbed too much, the spread limits the permitted dif­ference between the highest and the lowest tonal value to 5 %.
Page 66
66
Manual TECHKON SpectroDens
By measuring the trapping only the variation from the reference will be displayed.
The ISO standards dene no tolerances for trapping, only informative reference values are given.
But it is recommended, not to exceed a ∆E*­value of 10 in the secondary colors, which corresponds with the sum of the permitted ∆E*-deviations of the primary colors.
Instead of scanning the TECHKON strip the measurement function ISO-Check can also capture the ECI/bvdm Gray Control Strip ver­sion tvi 10 (i1) and (M i1). When using the scan function we recom­mend to use a patch width of 5 x 5 mm (min. 4 x 4 mm) with the standard aperture and a patch width of 4 x 4 mm (min. 3 x 3 mm) with the exchangeable 1,5 mm aperture.
ECI Gray Control Strips: ECI/bvdm tvi 10 (i1) and M i1
ECI/bvdm
tvi 10 • v2
ECI/bvdm tvi 10 (i1) • 6 mm • Control strip with tone values in 10 percent steps for print process control in accordance with ISO 12647 • www.bvdm.org • www.eci.org
Stop cmy cy cm my Paper 100 10 90 20 80 30 70 40 60 50 50 60 40 70 30 80 20 90 10 100 100 10 90 20 80 30 70 40 60 50 50 60 40 70 30 80 20 90 10 100 Start
cmy 70 k 70 cmy 50 k 50 cmy 30 k 30 Stop cmy 70 Paper k 70 my cmy 50 cm k 50 cy cmy 30 cmy k 30 100 20 80 40 60 60 40 80 20 100 100 20 80 40 60 60 40 80 20 100 < Start
ECI/bvdm Gray Control Strip (M i1) • FOGRA30 • ISO 12647-2 gray balance condition ‘CIELAB black ink’ • O set on uncoated yellowish paper (PT 5) • ISO 12647-2:2004/Amd 1
• Reference FOGRA30.txt (www.fogra.org) • www.eci.org, www.bvdm.org
ECI/bvdm
GrayCon M i1 • v2
FOGRA30
Page 67
67
CHAPTER 2 – Measurement functions
CIE L*u*v*, CIE L*C*h* uv, CIE xyY and DIN Lab99
The Premium-version of SpectroDens enables the display of additional color spaces, which might be used mainly in industries outside printing: CI E L* u* v*, CIE L*C*h* uv, CIE xyY and DIN Lab99.
Dierent formulas for calculating the color distance ∆E* can be selected as well. The Euclidean color distance ∆E* a*b* is the most widely used, although it is criticized that for certain areas in the color space the visual per­ception of color distances does not match the ∆E* a*b* color distance accordingly.
The ISO-Check function features the use of the SCCA reference value correction (select­able in the ISO-Check setup).
You will nd a comprehensive description of the SCCA function on page 59.
Page 68
68
Manual TECHKON SpectroDens
Metamerism-Index
Metamerism is the visual phenomenon that two colors match under certain light con­ditions and dier under dierent light conditions.
Colorimetry oers the possibility to detect this eect by measuring the color distance between the two color samples under two exactly dened light sources. The result of this operation is the Metamerism-Index.
For the determination of the Metamerism-In- dex there are two measurements to be made.
1. A reference measurement, which refers to illuminant condition D65.
2. A sample measurement on the sample patch.
As a result the color distances in ∆E* a*b* for illuminant conditions A, C and D50 are shown.
Identical color impression under light condition 1:
Color deviation under light condition 2 caused by metamerism:
Page 69
69
CHAPTER 2 – Measurement functions
Whiteness / Yellowness
Whiteness- and Yellowness-indices are quality parameters important for the paper
industry. SpectroDens Premium will display the whiteness and yellowness indices ac­cording to the following standards: Whiteness: E 313, Berger, Stensby, TP 425 – Yellow­ness: E 313, D 1925.
Whiteness index: Values are usually between 90 and 120 for “white” paper. The higher the value, the “whit­er” (= the brighter) the paper. Articial opti­cal brighteners in the paper will lead to high values as well.
Yellowness index: Positive value: the paper has a yellow cast Negative value: the paper has a blue cast
Whiteness Berger, Stensby
Paper Whiteness is dened by how closely a surface matches the properties of a perfect reecting diuser, i.e. an ideal reecting sur­face that neither absorbs nor transmits light, but reects it at equal intensities in all direc­tions. For the purposes of this standard, the color of such a surface is known as preferred white. Whiteness is closely related to our vi­sual perception of the color of paper. Some­times these variations in whiteness are subtle. For instance two pieces of paper that fall at dierent ends of the spectrum on the CIE whiteness index may appear white by them­selves however when juxtaposed; their varia­tions are revealed. The type of light paper is viewed under can aect how it looks to the human eye along with other elements like optical brightening agents (OBAs) used in paper production. For example, when OBAs are applied to a dull sheet of paper it will look brighter in natural light and dull in articial light. When there are little to no OBAs applied the paper will look brighter in articial lighting when compared to its counterpart with larger amounts of OBAs. The Techkon SpectroDens uses the Stensby and Berger formula’s for quantifying the whiteness of a print substrate.
Page 70
70
Manual TECHKON SpectroDens
PASS / FAIL
The PASS / FAIL function shows at a glance if a measured sample is in or out a set toler- ance compared to a reference value.
First, set the allowable limits of the ∆E*-toler­ance.
Select in the menu section of the measure­ment conditions the item PASS / FAIL toler- ance and edit the desired tolerance value by pressing the arrow buttons.
Every sample measurement is compared to the reference.
If the color distance is within the set tolerance (in this case 1.24 < 3.0), PASS will be displayed.
For out of tolerance measurements, accord­ingly FAIL will be shown accompanied by an acoustic signal of three short beeps.
Brightness T (TAPPI 452)
Like whiteness, paper brightness is also a measurement of the amount of light reected from paper. Unlike whiteness, it is focused on a narrow wavelength of blue light as opposed to all wavelengths in the spectrum. The higher the number on the brightness scale, the more light a particular sheet reects. Brightness is just one part of the equation when determining the visual aspects of paper. Two sheets with identical brightness can look drastically dierent because shade and whiteness are not taken into account.
In general, there are two measurement systems that are widely used for determining paper brightness. One (used in North America) is the TAPPI scale. The other (used primarily in other parts of the world) is the ISO system. Some papers can have a rating higher than 100 on the brightness scale. This is due to optical brightening agents present in the paper. When these are applied, the paper can actually reect more light than is coming from the source. An easy way to remember this is the brighter, the whiter.
Yellowness CIE
Yellowness is the degree that the color of a surface is shifted from preferred white (or colorless) towards yellow.
Page 71
71
CHAPTER 2 – Measurement functions
First, a Color book has to be selected which contains the Fogra reference data. If not found in the device memory, the color book which is valid for the paper class to be mea­sured, can be uploaded from the Windows software SpectroConnect.
Additional settings have to be made in the Media Wedge Setup window. The settings for ∆E*-tolerances are according to Fogra specications.
By activating Automatic Measurement the device will make consecutive measurements automatically with Measurement Interval to be set. Starting and stopping a measurement series is done by pressing the green measure­ment button.
The reading of the media wedge is from left to right and from rst to second row. The measurement begins with patch A1, then A2, etc.. Wrong measurements can be corrected by pressing the red escape button and re­measuring. After all measurements have been performed, the total result will be displayed.
Media Wedges with separation bars can also be measured by scanning. Just keep the green start-button pressed during the whole scanning procedure.
SpectroDens recognizes the version 2 as well as the version 3 of the media wedge by the dierent number of reference values. The Windows software SpectroConnect allows to analyze the measured data in detail and to print a report (see p. 89 f.).
Media Wedge
The Ugra / Fogra Media Wedge CMYK is a control element for checking the color con­sistency of digital proof systems issued by the printing institute Fogra. This practical function is for measuring and analyzing the patches of this control element.
Page 72
72
Manual TECHKON SpectroDens
Scan
The measurement function Scan provides the scanning of individual control elements.
The scanning procedure is very easy. Choose the measurement function SCAN in the device menu. Then place the measurement device on the print control strip. Take care that the measurement aperture is positioned on the rst measurement patch (= start position) and that the device edge is aligned to the print control strip (see picture).
Now press and hold the measurement button, move the device in constant speed along the print control strip until you reach the last patch to be measured (= stop position) and release the measurement button. If the measurement was carried out successfully a short acoustic signal can be heard and all color patches are displayed in the device. Using the SpectroConnect Export module measurement data can nally be exported in any other Windows application e. g. Microsoft Excel™, Word™ or other programs which can handle color data, e.g. a RIP calibration software.
Examples of control elements which can be scanned:
ECI Gray Control Strips: ECI/bvdm tvi 10 (i1) and M i1
TECHKON ISO/PSO Checker
ECI/bvdm
tvi 10 • v2
ECI/bvdm tvi 10 (i1) • 6 mm • Control strip with tone values in 10 percent steps for print process control in accordance with ISO 12647 • www.bvdm.org • www.eci.org
Stop cmy cy cm my Paper 100 10 90 20 80 30 70 40 60 50 50 60 40 70 30 80 20 90 10 100 100 10 90 20 80 30 70 40 60 50 50 60 40 70 30 80 20 90 10 100 Start
cmy 70 k 70 cmy 50 k 50 cmy 30 k 30 Stop cmy 70 Paper k 70 my cmy 50 cm k 50 cy cmy 30 cmy k 30 100 20 80 40 60 60 40 80 20 100 100 20 80 40 60 60 40 80 20 100 < Start
ECI/bvdm Gray Control Strip (M i1) • FOGRA30 • ISO 12647-2 gray balance condition ‘CIELAB black ink’ • O set on uncoated yellowish paper (PT 5) • ISO 12647-2:2004/Amd 1 • Reference FOGRA30.txt (www.fogra.org) • www.eci.org, www.bvdm.org
ECI/bvdm
GrayCon M i1 • v2
FOGRA30
Page 73
73
CHAPTER 2 – Measurement functions
SCAN function with strip detection
The SpectroDens enhanced scanning feature includes:
Dene the color bar by uploading a color library to the SpectroDens or by scanning a color bar with the instrument.
Advanced pattern recognition algorithm’s that automatically detect that the color bar being scanned is correct, has the correct number of patches and the colors are in the right sequence.
If the incorrect number of patches are scanned or the scanning speed is too fast or too slow the SpectroDens will display an error message informing you.
No guide bar required. The 4 wheels keep straight alignment while scanning.
Easily export scanned measurement data to other software applications automatically with SpectroConnect software (included with the instrument at no cost).
It is possible to scan a sequence of 200 patch­es.
Just select Strip detection o in the SCAN measurement function.
Page 74
74
Manual TECHKON SpectroDens
Opacity
Opacity refers to the transparency or opaqueness of paper. Papers with lower opac­ity tend to let text and images show through more than papers with higher opacity. Opacity is expressed as a percentage in paper. For example, paper with 98 % opacity means that 98 % of light is not allowed to pass through the paper.
Select the Opacity function from the mea­surement functions menu.
Place the sheet of paper to be a measured on a black backing.
In the next step take a measurement of the same sheet of paper after you have placed it on a stack of white paper (minimum 50 sheets).
Now the measurement device displays the result of the opacity measurement.
For further measurements just press the enter button and repeat the whole procedure.
Page 75
75
CHAPTER 2 – Measurement functions
OBA amount no OBA faint low moderate high
∆B 0 to 1 1 to 4 4 to 8 8 to 14 14 to 25*
∆CIEb* 0 -0,3 to -2,0 -2,0 to -3,7 -3,7 to -6,2 -6,2 to -10,4*
OBA-Check
Optical Brightening Agents (= OBAs) which are used more and more in the paper pro­duction, have a signicant eect not only on how we perceive the paper under dif­ferent illuminants, but also on the results of the printing process. The measurement function OBA-Check provides precise information about the intensity of the optical brightening agents used and the shade of the paper.
Choose the function OBA-Check from the menu measurement functions.
First you have to switch to the „soft-key“ CAL to carry out a calibration on the absolute white standard.
SpectroDens performs the OBA-Check ap­plying the measurement conditions M1 and M2, i. e. with and without stimulation through UV light (see p. 50). The dierence between the two CIEb*-values leads to the nal mea­surement result.
The amount of optical brightening agents according to ∆B (ISO 15397) and its categori­zation according to ISO 12647-2:
(0 ≤ ∆B < 1) no OBA
(0 ≤ ∆B < 4) OBA faint
(4 ≤ ∆B < 8) OBA low
(8 ≤ ∆B < 14) OBA moderate
(14 ≤ ∆B < 25*) OBA high
The table shows the relation between ∆B­and CIEb*-values for the practical application.
* and beyond
Page 76
76
Manual TECHKON SpectroDens
Average values
From a series of single measurements Average values can be calculated and dis- played. This function may be used when the surface of the sample has no even distri­bution of color.
By selecting the “soft key” with the average symbol a pop-up window will appear where Average measurement can be set, started and cancelled.
Here the Number of measurements to be calculated to one average value is set. The maximum number of single measurements is 9.
Now activate Start Average measurement.
The central window indicates how many sin­gle measurements have been taken.
After the last single measurement has been taken, the resulting average value will be displayed and the procedure can start again from the beginning.
Activating the menu item Stop Average mea- surement you quit this measurement mode.
Page 77
77
CHAPTER 2 – Measurement functions
2.4 Storage of measurement values
The versions SpectroDens Advanced and Premium include a Color library. This is a digital memory for storing color values, which can be compared or exchanged with
the Windows software SpectroConnect.
There are three dierent sections where color values can be stored:
Reference memory for up to 300 reference values
Sample memory for up to 3000 sample values
Color book memory for up to 20 color books with a total number of 25000
reference values
Reference memory
To access the Reference memory, you have to select Reference in the display. A mem­ory capacity for 300 references is at your disposal. After the 300th color value is stored, the message indicates that the memory is full.
By pressing the enter button with the device in Reference mode a pop-up window opens. Within the window navigation is done with the arrow keys.
1. With Save the last reference measurement is stored.
2. Select shows a list of stored color values. Within the list, values can be selected and dis­played by using the arrow- and enter-buttons.
1.
2.
3.
4.
5.
6.
1.
2.
Page 78
78
Manual TECHKON SpectroDens
Sample memory
The Sample memory has almost the same functions as the reference memory. Here up to 3000 sample color values can be stored. However the sample memory is a ring memory, i. e. maximum 3000 samples can be stored before the rst sample will be overwritten by a new measurement.
The pop-up window opens by pressing the enter button with the device in Sample mode.
3. By activating Save automatically all future reference measurements will be stored auto­matically.
4. By activating Search automatically Refer- ence Mem the closest matching reference color when making a sample measurement will be displayed. This feature is active in the sample mode with DELTA Mode on.
5. Delete will remove selected color values from the list.
6. Delete all will remove all values in the list.
1.
2.
3.
4.
5.
1.
2.
4.
5.
3.
3.
5.
6.
Page 79
79
CHAPTER 2 – Measurement functions
The “soft key” with the color book symbol opens the window for the settings of the color book memory.
1. Here a Color book is selected from the list.
2. With Load Reference from Color book a single color value from a color book is loaded into the reference memory.
3. Search for closest color determines the closest color match in a color book compared to a sample value.
Color book memory
Additionally to storing reference and sample color values, complete Color books can be managed. A color book is a folder which can hold up to 25000 single color values. An example are the HKS® spot color collections which are included in SpectroDens with the software package SpectroConnect. New color books can easily be edited and transferred into the device. Up to 20 color books can be stored in SpectroDens; there is no limitation of storing color books in the Windows software SpectroConnect.
1.
2.
3.
Page 80
80
Manual TECHKON SpectroDens
Chapter 3
How to use the Windows software SpectroConnect
3.1 Software description
The supplied Windows software TECH­KON SpectroConnect allows to trans­fer measurement data to the PC and to make device specic settings via the PC. The measurement values can be displayed on the computer monitor. Color information is displayed, colors can be compared and data can be exported into other software applications, e. g. Microsoft Excel™.
When using SpectroDens Basic the device has to be connected via the USB cable to the PC when working with SpectroConnect. When performing a measurement with the device, the value will be displayed immediately on the computer screen.
The models SpectroDens Advanced and Premium have an internal memory for up to 300 reference values, 3000 sample values and 20 “digital” color books, which can contain up to 25000 color values. When connecting the device to the PC via the USB cable, all stored values will be uploaded to the software and displayed on the comput­er screen. Of course, the device can be operated as usual when connected to the PC.
The software requires a computer with a free USB port and a completely installed Microsoft Windows 7, 8 or 10 operating system.
3.2 Installation
It is important, to carry out the following steps in the right order, to make sure that the USB device driver will be installed properly.
1. Make sure that the device is NOT connected to the PC. Insert the SpectroConnect CD into the CD drive of the running computer. You will nd the CD at the back of this manual.
2. The installation routine will start automatically. Follow the steps of the installation, until it is completed.
3. Now, after the installation was nished successfully you can connect the device with the Micro-USB cable to the computer.
Page 81
81
CHAPTER 3 – Software SpectroConnect
3.3 Overview / Software module: Compare colors
The application will be installed in the Windows program les section in the folder TECHKON GmbH / TECHKON SpectroConnect. After starting the program the software module Compare Colors will appear. Like all the SpectroConnect software modules it is divided into four logical segments: On the left side you see the memory content of the connected device, in the center the active program module, on the right side the color library which is stored on the PC and in the section at the bottom a bar which lists all available and selectable program modules.
1. The menu bar includes functions known from standard Windows applications. They comprise: New, Open, Save, Print and Quit.
2. The window Samples shows samples which are stored in the device. Only in Spec­troDens Advanced and Premium available.
3. The window References shows stored ref­erences. Only in SpectroDens Advanced and Premium available.
4. The window Color books displays the col­or books which are stored in the color library of the device. Per Drag and Drop you can
copy references and complete color books from here.
5. In this section, the appropriate measure­ment device can be selected.
6. Window of the active program module
7. This bar shows the list of available program modules. They can be selected from this bar directly via mouse click and the application will appear in the middle of the display.
After launching SpectroConnect it is pre-set to display the module Compare colors.
Software symbol SpectroConnect
1.
2.
3.
6.
7.
4.
5.
Page 82
82
Manual TECHKON SpectroDens
3.4 Software module: SpectroDens settings
In this module device specic settings can be made.
1. SpectroDens Info: All settings, which can be made at the device can also be made with this software feature.
2. Lock device functions: Device functions can be locked and unlocked with this feature. Locked functions will appear in shaded gray on the LCD display.
3. Device update: New device software can be uploaded into the device. There are two dierent types of les to upload: Device update: loads the new version of device software Device upgrade: upgrades to SpectroDens Advanced or Premium (must be purchased)
4. Device screenshot: This shows the device display on the computer screen. All measurement buttons can be remotely operated by the mouse pointer. The screenshot can be saved. This is a very useful feature for product trainings.
5. WLAN settings: Window for the conguration of the WLAN connection (see p. 22 .)
1. 2. 3. 4. 5.
Page 83
83
CHAPTER 3 – Software SpectroConnect
3.5 Device update and device upgrade
By the application of this module SpectroDens can be loaded with a new internal rmware which is selected by Device update or SpectroDens can be expanded in functionality by Device upgrade.
The SpectroDens Basic model can be extended via data upload to Advanced- or Premium- functionality without the need to send the instrument to the TECHKON service center. Of course, the same procedure applies for upgrades from Advanced to Premium.
The procedure is explained in the program window:
Page 84
84
Manual TECHKON SpectroDens
3.6 Software module: Export
Measurement data can be exported in any other Windows application e. g. Microsoft Excel™, Word™ or other programs which can handle color data, e.g. a RIP calibration software.
SpectroDens works like the keyboard: Pressing the measurement button will place the measurement value automatically at the location where the cursor is.
1. Here is dened, which data will be trans­mitted, how it will be formatted and what type of spacing between the measurement data will be. All the settings can be saved and recalled later.
2. The Export window can also be used for testing data transmissions.
This example shows the transfer of CMYK density values.
1.
2.
Page 85
85
CHAPTER 3 – Software SpectroConnect
3.7 Software module: ColorCatcher
The program module ColorCatcher is available in the models SpectroDens Advanced and Premium. Based on a L*a*b*-measurement and referring to selected ICC proles this program module displays the conversion into the color models L*C*h*, RGB and CMYK. Furthermore an automatic search for the closest matching color of a pre-set or self created color book is carried out. The lower part of the module window informs you about the exact color deviations between the sample and the recommended color of the color book.
1. Selected ICC proles
2. Measured sample color
3. Calculation of L*a*b*-, RGB-, L*C*h*- and CMYK-values based on selected ICC proles for the current sample.
4. Shows the closest match to the current sample out of the selected color book.
5. Calculation of L*a*b*-, RGB-, L*C*h*- and CMYK-values based on selected ICC proles.
6. Self created color book
7. Colors from a color book
8. Variations and visible dierences
1.
2.
3.
4.
6.
7.
8.
5.
Page 86
86
Manual TECHKON SpectroDens
3.8 Software module: PhotoLab
TECHKON PhotoLab is an optionally available program module for evaluating and displaying L*a*b* color data graphically. Color dierences between sample and refer­ence colors can be analyzed. PhotoLab is only available in the SpectroDens Premium version.
Furthermore, color values and complete color books from the device memory or out of the PC color library can be selected and be moved by drag-and-drop with the mouse pointer directly to the center of the display where the L*a*b* color circle is.
A pop-up window will ask, if the values should be used as sample or as reference values.
4. A mouse click on the program symbol will start the plug-in PhotoLab within the SpectroConnect environment.
1. In the upper section of the module win­dow, the color dierences between sample and reference are shown numerically.
2. Reference: viewnder with tolerance circle Sample: bold point The current selected value is marked in white.
3. Color samples and references are directly imported into PhotoLab by making measure­ments with SpectroDens having it connected to the PC via the USB-cable.
1.
2.
3.
4.
Page 87
87
CHAPTER 3 – Software SpectroConnect
1. Samples and references can be selected di­rectly with a mouse click or by choosing from the “pull-down” menu.
Delete and renaming is done via the menu.
2. Dierent modes of the L*a*b*-display can be selected:
The modes Bargraph and Quality Certi-
cation Chart show the trend of ∆L*-, ∆a*- and
∆b*-values of a measurement series.
The Table Vi ew shows all values numeri-
cally.
In the Job window job relevant text data
can be edited.
3. A measurement value highlighted in red indicates an out-of-tolerance sample.
4. The black circle line indicates the tolerance limit of the ∆E*-region. If the black pointer reaches into the outer segment of the circle, the tolerance has been exceeded.
5. Enlarged view of the tolerance circle
6. Display of the color components ∆L*, ∆a* and ∆b*.
1.
2.
3.
4.
5.
6.
Page 88
88
Manual TECHKON SpectroDens
1. General settings: This is the window for setting the dimensions of the L*a*b* color circle and the ∆L*a*b* color deviation.
2. Tolerances: In this window the tolerances for the color de­viations are dened and the size of the toler­ance circle is set.
3. Mean values: It can be appropriate to take several single measurements on one sample which then are averaged to an average value. The number of measurements which result in the calculation of the average value is dened in this window.
PhotoLab settings
When you open the pop-up window PhotoLab settings in the menu bar of the pro­gram module PhotoLab, you can select and edit three areas.
Page 89
89
CHAPTER 3 – Software SpectroConnect
3.9 Software module: Media Wedge
The program module Media Wedge serves the fast, colorimetric analysis of the Ugra / Fogra Media Wedge. This digital control element is available from Fogra, the Ger­man “Graphic Technology Research Association” (www.fogra.org). The test element is placed at the border of a proof. The software evaluates the color quality printed in colorimetric terms.
Color dierences between the Fogra target values and the measured samples will be clearly represented in this program module.
The software module Media Wedge is only available in the SpectroDens Premium version.
1. Manual selection of the media wedge ver­sion. (You will nd the version number on the printed test element.) The software supports the Ugra / Fogra Media Wedge in version 3.0 as well.
2. With SpectroDens Premium there is a me­dia wedge feature within the device as well.
Measurements can be made remotely with­out a connection to the PC. When the device
gets connected later, the measured data will be uploaded.
3. Selection of reference values
4. Overview of measurement results
5. A short protocol can be printed with a Dymo label printer on a self-adhesive label which can be stuck on the proof. Alternatively a detailed report of the measurement values can be printed (see p. 90).
5.
1.
3.
4.
2.
Page 90
90
Manual TECHKON SpectroDens
Media Wedge evaluation: Label and details
After the import of the media wedge a short protocol will be generated and can be printed on a self-adhesive label with a Dymo label printer.
The software supports the Ugra / Fogra Media Wedge in version 3.0 as well. The picture shows the section Details.
Page 91
91
CHAPTER 3 – Software SpectroConnect
3.10 Software module: Transfer curve
Using the software module Transfer curve you can fast and easily determine transfer characteristics on prints and check the compliance of the dot gain according to ISO 12647 targets or individual standards. The software calculates at the same time com­pensation values for the adjustment during plate exposure.
Values for CMYK as well as spot colors can be analyzed. Measurement results can be displayed as graphics or charts. Furthermore several curves can be compared and av­erage values can be calculated.
Any step wedge can be evaluated. The design of the step wedge can be dened in 1 % steps by using the window Reference values.
Page 92
92
Manual TECHKON SpectroDens
In the section Target values the target values can be edited. Furthermore settings for the assignments of the targets can be dened.
Page 93
93
CHAPTER 3 – Software SpectroConnect
3.11 Software module: ISO-Report
The software module ISO-Report allows a fast and comprehensive documentation according to the ISO standard. The ISO-Report can be printed out. ISO-Report is only available in the SpectroDens Premium version.
Page 94
94
Manual TECHKON SpectroDens
3.12 Software module: SpectroChart
The software module SpectroChart allows the fast and easy evaluation of test charts like for example ECI 2002, IT8.7/3, Fogra27L, Fogra39L and others. Related target templates in .ttg-format are deposited in the folder User Public Public Documents TECHKON GmbH TECHKON SpectroConnect.
The measurement results are saved in ISO-format. This way they can be imported into common ICC prole generator software.
In the program section Job (1.) the job is specied. Here you also choose the target you wish to use. Additionally you can determine its patch width and the amount of patches per row (2.).
The resulting nal size of the target will be calculated automatically and then displayed to you. Alternatively you can simply use the presets of a target template or else choose one of the pre-dened paper formats.
When the button Create Ti le (3.) is used, the target can be saved as a ti le on the PC, where you can initiate to get it printed. Additionally the “Job” section can be used to save the job and to start the export of the measurement values.
In the program section Measurement (4.) the current measurement result is displayed clearly. Furthermore here you can dene and activate a ∆E value as tolerance.
1. 4.
2.
3.
Page 95
95
CHAPTER 3 – Software SpectroConnect
3.13 Software module: SpectroCheck
TECHKON SpectroCheck (optionally available) is a color reference used to check the measuring accuracy of the SpectroDens advice.
Connect SpectroDens and SpectroCheck via the USB cable with the computer. The program module SpectroCheck will now lead you step by step through the checking procedure of the measurement device. The information, whether the device corre­sponds to the technical specications, can also be printed out.
Note: The checking of the TECHKON SpectroDens has to be made with 3.0 mm stan­dard aperture.
Page 96
Specications can be subject to change without notice. Specications valid for standard device with 3 mm aperture. All mentioned trademarks and copyrights are recognized. SpectroPlate, SpectroDens, SpectroJet, SpectroDrive, DENS, SpectroCheck , InkCheck and TECHKON are registered trademarks of TECHKON GmbH.
Specications
Measurement geometry 0° : 45° optics according to ISO 5-4
Spectral range 400 to 700 nm in 10 nm steps, spectral resolution 10 nm,
pixel distance sensor < 3 nm
Measurement aperture 3 mm round standard; 1.5 mm round optional
Light source LED, provides measurement conditions M0, M1, M2, M3
according to ISO 13655
Polarization lter Twice linear crossed, switched on and o per button release
Measurement time Approx. 1 sec. per measurement; max. 12 seconds in scan mode
White reference Absolute and relative; absolute white standard integrated and
protected in charging console
Illumination types A, C, D50, D65, F 2/7/11
Standard observer 2°, 10°
Density lter DIN 16536, DIN 16536 NB, ISO/ANSI T, ISO/ANSI I, ISO E; spectral
density Dmax
Density measurement range 0.00 – 2.50 D
Repeatability 0.01 D; 0.03 CIE ∆E*a*b*
Inter-instrument agreement 0.01 D; 0.3 CIE ∆E*a*b*
Display Color LC backlight display, anti-reective, 320 x 240 pixels
Power supply Rechargeable LiFePO4 battery, regulated recharge via charging
console with AC adapter, 100 – 240 V, 47 – 63 Hz, approx. 10000
measurements per battery charge, battery level control
Communication port USB; WLAN module optional
Weight 495 grams
Dimensions 62 x 50 x 185 mm (approx. 2.4 x 2.0 x 7.3 inches)
System requirements for TECHKON software:
Windows 7, 8 or 10; 32- and 64-bit, minimum: IBM-compatible PC with Intel Core Duo
processor or comparable processor, 4 GB RAM, 2 USB ports
Performance packages SpectroDens Basic, Advanced and Premium: see page 7
Accessories, spare parts and contents of delivery: see page 8
Appendix
Page 97
Manufacturer certicate
applicable for ISO 9000 documentation
Device: Spectro-Densitometer TECHKON SpectroDens
Serial Number:
Manufacturer: TECHKON GmbH • Wiesbadener Str. 27 • D-61462 Königstein
Telephone: +49 (0)6174 9244 50 • Telefax: +49 (0)6174 9244 99 info@techkon.com • www.techkon.com
Certication: The device is compliant with EU directive 2014/53/EU concerning
the electromagnetic compatibility EMC and is provided with the CE label. The device is RoHS compliant (class 9). Japanese Radio Law and Japanese Telecommunications Business Law Compliance. This device is granted pursuant to the Japanese Radio Law (電波
) and the Japanese Telecommunications Business Law (電気 通信事業法). This device should not be modied (otherwise the
granted designation number will become invalid).
The device complies with the Japanese Type Approval:
Notice: The supplied AC adapter is according to regulations UL, IP 40, IEC
950 and VDE EN-EC10. The device is to be used only with the origi­nal TECHKON SpectroDens AC adapter.
Maintenance: The device is maintenance free. The measurement aperture has to
be kept clean from dust. It can be cleaned with clean, compressed air and an optics brush.
We recommend a functionality check-up every 24 months at the
TECHKON service center, which includes the issue of a new Manu­facturer certicate.
R 202-LSE115
T D 16-0031 202
Page 98
Warranty: The warranty for TECHKON products is 24 months starting with the
date of purchase. The warranty is invalid if the damage is caused by improper use of the device. Only original TECHKON spare parts and accessories are to be used.
Recycling: The device is according to §14 ElektroG registered under the EAP
no.: DE 98280049. Devices for disposal can be sent directly to the manufacturer.
Calibration: The integrated spectral sensor is calibrated by a white calibration.
After performing a white calibration with the supplied absolute white standard integrated in the charging console, the device is long term stable. We recommend to make a white calibration be­fore every measurement series, to ensure the device is calibrated correctly.
The remission values documented on the back of the charging
console are derived from a ceramic white standard, which is refer­enced to measurements taken and certied by the National Insti­tute of Standard and Technology (NIST).
Standards: The calculation of colorimetric values is according to the formulas
and tables of ISO 13655:2009 and publications of CIE.
The calculation of densitometric values is according to the formu-
las and tables of ISO 5-3.
ETSI EN301-489-1 EMC standard for radio equipment and services
V2.1.1 (2017- 02)
ETSI EN301-489-17 EMC standard for radio equipment: specic con-
ditions for broadband data transmission systems V3.1.1 (2017-02) ETSI EN 300 328 RF spectrum use V2.1.1/2016-11
Place Date Signature
D-614 62 Königstein
Page 99
EU-Konformitätserklärung
EC-Declaration of Conformity
Déclaration de Conformité de la CE
Dichiarazione di conformità CE
Hersteller: TECHKON GmbH
Manufacturer / constructeur / costruttore
Adresse: Wiesbadener Str. 27
Address / adresse / indirizzo
erklärt, dass die Produkte:
Declares that the products / déclare, que le produits / si dichiara che il prodotto
Typ: SpectroDens
Type / tipo
Verwendungszweck: Farbmessungen
Intended use / utilisation prévue / Color measurements / mesure de la uso previsto couleur / misurazione del colore
bei bestimmungsgemäßer Anwendung den grundlegenden Anforderungen ge­mäß EU-Richtlinie 2004/108/EC entspricht und dass die folgenden Normen ange­wandt wurden:
complies with the essential requirements of the 2004/108/EC Directive, if used for its intended use and that the following standards has been applied: / répond aux exigences essentielles de l'article 3 de la directive 2004/108/EC, prévu qu’il soit utilisé selon sa destination, et qu’il répond aux standards suivants: / soddisfa tutti i requisiti della direttiva 2004/108/EC qualora venga utiliz­zato per l’uso previsto e che le seguenti norme siano applicate:
angewendete Norm: EN 55022:2006
Applied standard / standard appliqué / norma applicata issue / édition pubblicato
D-61462 Königstein
Ausgabe: 2008-05 +A1: 20 07
EN 55024 Ausgabe: 2003- 01 1998+A1:2001+A2:2003
ETSI EN301-489-1 EMC standard for radio equipment and services V2.1.1 (2017-02) ETSI EN301-489-17 EMC standard for radio equipment: specic conditions for broadband data transmission systems V3.1.1 (2017-02) ETSI EN 300 328 RF spectrum use V2.1.1/2016-11
Page 100
Name:
Company:
Department / Job title:
Address:
City / ZIP-Code:
Country:
Telephone:
Telefax:
E-mail:
Your TECHKON-Dealer:
Device Serial number:
(Label on the bottom
of the device)
TECHKON Registration card
Please send me information about
the entire TECHKON product range
Please put my E-mail address on the
mailing list for the TECHKON-Newsletter
TECHKON GmbH
Wiesbadener Str 27