WARRANTY
Measurement Specialties, Inc. accelerometers are warranted during a period of one year from date of shipment to original purchaser to be free from defects in material and workmanship. The liability of Seller under this warranty is limited to replacing or repairing any instrument or component thereof which is returned by Buyer, at his expense, during such period and which has not been subjected to misuse, neglect, improper installation, repair, alteration, or accident. Seller shall have the right to final determination as to the existence and cause of a defect. In no event shall Seller be liable for collateral or consequential damages. This warrant is in lieu of any other warranty, expressed, implied, or statutory; and no agreement extending or modifying it will be binding upon Seller unless in writing and signed by a duly authorized officer.
RECEIVING INSPECTION
Every Measurement Specialties, Inc. accelerometer is carefully inspected and is in perfect working condition at the time of sh ipment. Each accelerometer should be checked as soon as it is received. If the unit is damaged in any way, or fails to operate, a c laim should immediately be filed with the transportation company.
SERVICE CONCERNS
If a Measurement Specialties, Inc. instrument requires service, first contact the nearest Measurement Specialties, Inc. representative. They may be able to solve the problem without returning the unit to the factory. If it is determined that factory service is required, call Customer Service at the regional headquarters for an RMA number before return.
RETURNS
All units being returned to the factory require an RMA (Return Material Authorization) number before they will be accepted. This number may be obtained by calling Customer Service at the regional headquarters with the following information; model number(s), quantity, serial number(s), and symptoms of the problem, if being returned for service. You must include the original purchase order number if under warranty.
RECALIBRATION SERVICES
The Vibration Sensors Design Center and its two manufacturing facilities in China and France offer factory re-calibration services for Piezoresistive, Piezoelectric and Integrated Electronics Piezoelectric (IEPE, ISOTRON, ICP, etc.) accelerometers. NIST (US), DKD (Germany), COFRAC (France) traceable calibration services on sensitivity at 100 Hz (102 or 120 Hz in Europe) and full frequency sweeps are offered. Contact the regional headquarters for pricing information.
TE CONNECTIVITY SENSORS/// MODEL 4020 & 4030 ACCELEROMETER 05/2016
Page 3
MODEL 4020 & 4030 ACCELEROMETER
Page 3
2x ¼ or M6 mounting screws Torque to 18 lb-in (2.0 Nm)
DESCRIPTION
The Model 4020 & 4030 are low noise, signal conditioned DC accelerometers packaged in a durable molded housing with an integral cable assembly. The accelerometers are offered in ±2g & ±6g ranges with a nominal 0-200Hz bandwidth and an operating temperature range of ­40°C to +85°C. The model 4020 is a dual axis configuration while model 4030 is a triaxial configuration.
INSTALLATION
The model 4020 & 4030 accelerometers are designed to be screw mounted. The following guidelines should be followed when screw mounting these accelerometers.
The mounting surface should be clean and free of any residue or foreign material.  The mounting surface should be smooth and flat.  Torque screws to recommended limits using steel washers under the heads of the mounting screws. Use manual torque wrench (do not
use electric tools).
TE CONNECTIVITY SENSORS/// MODEL 4020 & 4030 ACCELEROMETER 05/2016
Page 4
MODEL 4020 & 4030 ACCELEROMETER
Page 4
WIRING
The accelerometers are designed to be operated from 5-30Vdc excitation and provide a 0.5 to 4.5V full scale output. The output is DC-coupled and should be used in single-ended mode. The sensor and electronics are fully contained in an internal Faraday cage which is connected to the cable shield (floating with respect to circuit ground) for RFI protection.
A 2.5Vdc bias will be present on the output leads and the output of the accelerometer will be 0.5-4.5V full scale. The accelerometer should be connected to the interface circuitry as detailed below.
Biaxial model 4020 wiring diagram
Triaxial model 4030 wiring diagram
TE CONNECTIVITY SENSORS/// MODEL 4020 & 4030 ACCELEROMETER 05/2016
Page 5
MODEL 4020 & 4030 ACCELEROMETER
Page 5
Range (g)
±2
±6
Self Test Output Change (mV)
X = +210 ±90
X = +70 ±30
Y = -210 ±90
Y = -70 ±30
Z = -340 ±190
Z = -110 ±65
FULL SCALE
RATING
(g-pk)
SENSITIVITY
(mV/g)
±5% AMPL
PASSBAND
(Hz)
PASSBAND
NOISE
(µVrms)
SPECTRAL NOISE
(µg-rms/√Hz)
DYNAMIC RANGE
(dB)
2
1000
200
600
51
67 6 333
200
240
42
75
CABLE ROUTING
The model 4020 & 4030 accelerometers incorporate a PVC jacketed cable with an integral shield. Th e cable assembly should be properly secured at regular intervals during testing. It is recommended to use clamps, wax, or tape to secure the cable to minimize cable motion that can add noise to the output signal. The initial attachment should be within two to three inches of the accelerometer.
Avoid routing cables near high-voltage wires and also ground the shield at the signal conditioner to minimize ground loops.
SELF-TEST OPERATION
The model 4020 & 4030 accelerometers incorporate a self-test feature. This feature is activated by grounding the self-test lead (gray). The corresponding shift in bias output is shown in table below.
The self-test bias shift takes place no matter the orientation of the accelerometer. Sensitivity is unaffected. For normal operation the gray lead floats and is not connected.
NOISE FLOOR SPECIFICATIONS
The model 4020 & 4030 accelerometers incorporate a LP filtered output for a high output signal. The noise specifications for this accelerometer are detailed in table below.
TE CONNECTIVITY SENSORS/// MODEL 4020 & 4030 ACCELEROMETER 05/2016
Page 6
MODEL 4020 & 4030 ACCELEROMETER
Page 6
te.com/sensorsolutions
Measurement Specialties, Inc., a TE Connectivity company.
Measurement Specialties (MEAS), American Sensor Technologies (AST), TE Connectivity, TE Connectivity (logo) and EVERY CONNECTION COUNTS are trademarks. All other logos, products and/or company names referred to herein might be trademarks of their respective owners.
The information given herein, including drawings, illustrations and schematics which are intended for illustration purposes only, is believed to be reliable. However, TE Connectivity makes
no warranties as to its accuracy or completeness and disclaims any liability in connection with its use. TE Connectivity‘s ob ligations shall only be as set forth in TE Connectivity‘s Standard
Terms and Conditions of Sale for this product and in no case will TE Connectivity be liable for any incidental, indirect or consequential damages arising out of the sale, resale, use or misuse of the product. Users of TE Connectivity products should make their own evaluation to determine the suitability of each such product for the specific application.
© 2016 TE Connectivity Ltd. family of companies All Rights Reserved.
NORTH AMERICA
EUROPE
ASIA
Measurement Specialties, Inc., a TE Connectivity Company 1000 Lucas Way Hampton, VA 23666 United States Tel: +1 757 766 1500 Email: customercare.hmpt@te.com
MEAS Deutschland GmbH a TE Connectivity Company Impasse Jeanne BENOZZI CS 83 163, 31027 Toulouse Cedex 3, France Tel: +33 (0) 582 08 22 00 Email: customercare.tlse@te.com
Measurement Specialties China Ltd., a TE Connectivity Company No. 26, Langshan Road Shenzhen High-tech Park (North) Nanshan District, Shenzhen 518057 China Phone: +86-755-3330-5088 Email: customercare.shzn@te.com
FREQUENCY RESPONSE CURVES
The typical frequency response curves for the model 4030 accelerometers are illustrated below.
Measurement Specialties, Inc., a TE Connectivity Company Vibration Sensors Design Center 32 Journey, Suite 150 Aliso Viejo, CA 92656, USA Tel: +1 949 716 7324
TE CONNECTIVITY SENSORS/// MODEL 4020 & 4030 ACCELEROMETER 05/2016
Loading...