Digital Clinical Thermometer for household use 【For Oral Use Only】
Thank you for purchasing this Digital Clinical Thermometer. To use this
Digital Clinical Thermometer correctly and safely, please read all of the
instruction manual provided by TDK carefully and keep in mind the
before using this Product.
After reading the manual, please keep it at hand so that others can refer
it at any time.
* Specialized smartphone application is available. Visit the following site
to confirm.
https://product.tdk.com/info/en/products/biosensor/biosensor/dbt/index.html
* The information in the Instruction Manual (except for product
specification and design) may be changed without prior notice.
Illustrations in this instruction manual are
schematic figures of this Product.
● Make sure that the temperature is taken at the base of tongue.
● Please refrain from using the Product after eating or drinking.
● Please refrain from using the Product after exercise.
Body Temperature Transfer Symbol
Shows the state ready to transfer body temperature
data to your smartphone.
Previous Value Memory Symbol
Shows the body temperature value measured last time
Shows the remaining battery level in four increments.
Shows a temperature value that is measured by the
predictive function.
This section lists important precautions that users of this product (and anyone else) should observe in order to avoid injury to you
and any person and damage to property.
Be sure to observe the following pr ecautions
Do not disassemble, repair or modify this Product.
( Attempting to do so may cause trouble or damage of this Product and result an incorrect measurement. )
Use a battery designated in the Manual.
(Using an undesignated battery may cause injury and disaster due to heat generation, rupture, leakage, or ignition of the battery.)
Set the battery in this Product correctly so that “
+” and “-” are in their proper positions.
( Incorrect setting of the battery may cause injury and disaster due to heat generation, rupture, leakage, or ignition of the battery.)
Do not allow children to use, and keep this Product out of reach from small c hildren.
If you allow this Product to use small children, this Product may cause their injury.
Do not use this Product for any purposes other than taking your body temperature.
(Using this Product on other living things such as your pets or other animals may cause their injury.)
Do not take your body temperature anywhere other than in-oral (armpit, ea r, etc.).
(Using this Product in areas other than the mouth may cause your injury or result an incorrect measurement.)
This Product is exclusively designed for oral use.
Do not share this Product among multiple persons.
(Attempting to do so could cause infectious disease.)
Do not bite or bite off the measurement area of this Product.
(Attempting to do so may cause an accident such as accidental swallowing.)
In case of accidental swallowing, consult with a medical doctor immediately.
Do not heat the battery or place the battery in a fire.
(Attempting to do so may cause injury due to battery rupture or similar.)
Do not attempt to self-diagnose or self-treatment based on the measurem ent result.
(Attempting to do so may cause worsening of symptoms.)
Please consult with a doctor when seeking medical attention.
Keep batteries or the battery cover out of reach from small children.
(To prevent small children from swallowing batteries or the battery cover accidentally.)
If this Product is swallowed, consult with a medical doctor immediately.
Do not take your body temperature by using this Product that has been sto red outside the range of the
storage environment temperatures.
(Storage outside the range of the storage environment temperatures may cause trouble of this Product and result an incorrect
measurement.)
Do not use this Product in a place where a strong static discharge or elect romagnetic radiation occurs, or
near a magnet.
(Such usage may cause trouble or damage to this Product and result an incorrect measurement.)
Do not drop, step on or otherwise apply a strong shock to this Product or the storage case.
(Such usage may cause trouble or damage to this Product and result an incorrect measurement.)
Be careful so that fluid or water does not enter into this Product.
(This Product is not designed as fully waterproof. Entry of water into this Product may cause trouble of this Product.)
TDK warrants that the Product shall be free from defects in materials and workmanship when used normally in accordance with
TDK’s instruction manual for a period of one (1) year from the date of original retail purchase by the end-user purchaser
("Warranty Period").
If you find any defect in materials and workmanship in your Product within the Warranty Period, please contact the call center and
provide something to certify the date of your purchase. TDK will replace the Product without charge.
Call Center: https://product.tdk.com/info/en/products/biosensor/biosensor/dbt/contact.html
This Warranty does not apply: (a) to consumable parts, such as batteries; (b) to cosmetic damage, including but not limited to
scratches and dents; (c) to damage caused by operating the Product outside TDK’s instruction manual; (d) to damage caused by
accident, abuse, misuse, fire, earthquake or other external cause; or (e) to defects caused by normal wear and tear or otherwise
due to the normal aging of the Product.
TO THE EXTENT PERMITTED BY LAW, THIS WARRANTY AND THE REMEDIES SET FORTH ARE EXCLUSIVE AND IN LIEU
OF ALL OTHER WARRANTIES, REMEDIES AND CONDITIONS, WHETHER ORAL, WRITTEN, STATUTORY, EXPRESS OR
IMPLIED. TDK DISCLAIMS ALL STATUTORY AND IMPLIED WARRANTIES, INCLUDING W ITHOUT LIMITATION,
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND WARRANTIES AGAINST
HIDDEN OR LATENT DEFECTS, TO THE EXTENT PERMITTED BY LAW.
EXCEPT AS PROVIDED IN THIS WARRANTY AND TO THE MAXIMUM EXTENT PERMITTED BY LAW, TDK SHALL NOT BE
RESPONSIBLE FOR DIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY BREACH
OF WARRANTY OR CONDITION, OR UNDER ANY OTHER LEGAL THEORY, INCLUDING BUT NOT LIMITED TO, LOSS OF
USE, LOSS OF OPPORTUNITY, OR LOSS OF, DAMAGE TO, COMPROMISE OR CORRUPTION OF DATA (INCLUDING ANY
DATA STORED IN THE PRODUCT).
Please carefully read this instructions manual.
Digital Clinical Thermometer
DC 3V (CR2032 battery × 1 pc)
Approx. 1 year
( under the usage condition of room
temperature 23ºC,
measurement for 7 minutes
1 time/day, data transfer 1 time/day )
Four-digit with ºC display
in increments of 0.01ºC
±0.10ºC (38.01 to 42.00ºC)
±0.05ºC (35.00 to 38.00ºC)
±0.10ºC (33.00 to 34.99ºC)
(measured in a constant temperature
water tank at a normal room
(compatible with predictive
temperature measurement and
actual temperature measurement)
Relative humidity: 30 to 85% RH
(Without condensation)
Relative humidity: 30 to 90% RH
(Without condensation )
Approx. 20g (including battery)
Approx.
46(W) × 99.5(L) × 11(T) mm
(installed in the product):1pc

Take your body temperature as soon as you wake up and while you are still in bed.
( Measurement after getting up or a meal may result an incorrect measurement. )
Whenever you measure your body temperature by using this Product, please measure the
same point in your mouth.
Measurement should be conducted by using this Product at the same time every morning to the
extent possible.
1. Take Out this Product from the Storage Case. Push the Operation Button at least
One Second.
• The power turns on and the display indicates all symbols with a beeper sound.
• Display the previous value.
• Next, show ambient temperature.
At this time, "Lo" is displayed when the ambient temperature is lower than 33℃
.
When this Product Displays “Lo” or the ambient temperature,
Place the Measurement Area on Either the Right or Left Side of
the Base of Your Tongue w ith the Display of this Product upwardly.
Hold this Product with Your Tongue and Shut Your Mouth.
Hold this Product by Your Hand and make sure the Measuring Point
in the mouth does not move.
How to Take Your Temperature
• When a beeper sound goes off in approximately one minute and “ ºC” is indicated
on the display, a predictive measurement is successfully completed.
Check your body temperature measurement result ( Predictive temperature value ).
• Measurement in an incorrect manner may result an incorrect measurement.
4. Turn Off the Power by Pushing the Operation Button for at least One Second.
• The power is automatically shut down after a certain period of time lapse even if the operation
button is not pushed.
5. Put this Product into the Storage Case.
* If you want to measure your body temperature strictly, we recommend that you should take an
actual temperature.
Keep on taking your body temperature after the beeper sound in step 3 above.
The actual temperature measurement is successfully completed when the beeper sound goes off
once again after approximately 5 minutes lapse.
Check the result of the actual measurement and continue with steps 4 and 5 above.
Transfer your body temperature data in this Product to your smartphone
・If “ ” is displayed, data transfer to your smartphone is available.
■ Delete your body temperature data in this Product
・Push the operation button of this Product at least 10 seconds while this Product is powered on.
Upon pushing the operation button even after this Product is powered off, four “ ” are
displayed and your body temperature data in this Product is deleted. If you release the operation
button before four “ ” are displayed, the data will not be deleted.
Data transfer in progress
Use a dry, soft cloth when removing stains from the product.
Darker stains may be removed by using a soft cloth slightly moistened with water or alcohol.
Please observe the following precautions. Cleaning this Product without obser ving the precautions may cause the
trouble of this Product.
• The thermometer is not designed for fully waterproof. When removing stains, please b e careful to prevent water
and alcohol from entering into this Product and keep them away from the displa y.
• Do not soak the measurement area in alcohol or water at than 50ºC or higher to sterilize it. Do not clean the
measurement area by using ultrasonic cleaning.
Dry this Product b y us ing a dry cloth before storing it in the storage case.
If you do not intend to use this Product for a long period, be sure to remove the battery before s torage.
Storage in the places below may cause the trouble of this Product.
• Dusty places. Places subjected to direct sunlight, high-temperature or high- humidity. Places subjected to water or
● For support of this Product, please check our following site.
https://product.tdk.com/info/en/products/biosensor/biosensor/dbt/index.html
How to Replace the Battery
Upon pushing the operation button, the remaining battery is indicated in four levels
such as
“ ” (battery level low). Replace the battery promptly
in accordance with the following procedures when
Measurement is impossible when
“ ” is indicated. *Enclosed battery when you
purchase this Product is for the temporary use and it may run out faster than usual.
■
Battery Replacement Procedure (CR2032 lithium battery × 1 pc)
1. Insert a Pointed Object, Such as a Tooth Pick, into the Hole on the Back Side
of this Product (1), Slide Out the Battery Cover by Using the Pointed Object (2),
and Pull Out the Battery Cover (3).
2. Remove the Old Battery (4).
3. Wait for About at least approximately 10 Seconds to Prevent Malfunction of this
Product, And Then Set the New Battery with the “+” Side Upwardly (5).
※ Do not insert a battery directly into this Product. It may cause trouble
of this Product.
4. Attach the Battery Cover So That the cover is Inserted into this Product (6).
5. Push the Operation Button and Check that “ ” is Displayed.
Precautions in Replacing a Battery
Replacing the battery deletes your body temperature data stored in this Product. Transfer the temperature data to
your smartphone before replacing the battery.
When disposing of a used battery, please follow all applicable laws and regulations in which it is being used.
Keep batteries and the battery cover away from small children to prevent their accidental swallowing.
TOSHIBA HOKUTO ELECTRONICS CORPORATION
1975, 23-chome, Minami 5-jodori, Asahikawa, Hokkaido 078-8335, JAPAN
Scattering in
measurement
values
Placed the measurement area in
different positions for each
measurement or opened your
mouth during measurement.
See the “ For Correct Measurement ” and ” How To Take Your Temperature ”.
A trouble has occurred in this
Product during predictive
temperature measurement.
See the “ For Correct Measurement ”and ” How to Take Your Temperature ”.
A Predictive measurement value is not correctly estimated if the measurement
area is taken out of your mouth during predictive temperature measurement.
Keep on measurement ( for actual temperature measurement ) or turn off the
power of this Product once and take your body temperature again after at least
30 seconds lapse.
“Er.11 to 19” is
displayed.
Measurement has not been
conducted in the correct manner.
See the “ For Correct Measurement ” and ” How To Take Your
Temperature ”.Turn off the power of this Product once and take your body
temperature again after at least 30 seconds lapse.
The thermometer has been stored
in a place where the temperature
is higher than +40ºC.
Take your body temperature again after placing this Product in a +10 to +40ºC
environment for at least one minute.
The thermometer has been stored
in a place where the temperature
is lower than +10ºC.
Take your body temperature again after placing this Product in a +10 to +40ºC
environment for at least one minute.
The measurement result
exceeded 42ºC.
See the “ For Correct Measurement ”and ” How To Take Your Temperature ”.
The measurement result was
33ºC or less.
See the “ For Correct Measurement ”and ” How To Take Your Temperature ”.
The thermometer has failed.
Please visit our website for FAQ
Nothing is displayed
on the display.
• The battery has run out.
• The thermometer has failed.
Replace the old battery with a new one. If the power does not turn on even after
the battery has been replaced, please visit our website for FAQ
The power does not
be turned on.
Battery of this Product is not
properly seated.
If power cannot be turned on, remove the battery once and insert it again.
3-9-1 Shibaura, Minato-ku, Tokyo 108-0023, JAPAN
“Lo” < 33℃ ≦ambient temperature
Power Off (Blank screen )
Whole symbols on the screen
Previous value is displayed
Under taking a temperature
Completion of
Predictive temperature
measurement
Completion of
actual temperature
measurement
Completion of data deletion