5. Configuration
The device can be configured via DIP switches and jumpers.
Scaling
In modes where a scaling is necessary to represent the correct measurement value
(for example 5 A AC for current transformer) the upper scale value corresponds to
the primary value of the current transformer.
Example: upper scale value = 50 = primary value of the current transformer
Configuration via DIP switches and jumpers if only one measurement mode is
required (Manual mode)
1. Set DIP switch 10 to ON position.
2. Set measurement mode with jumpers at J4-J6 according to table "Measurement
mode”, column “Manual mode".
3. Set upper scale caption for the first measurement mode with DIP switches 1-9
(see “Configuration of the upper scale caption”).
4. Connect supply voltage.
Configuration via DIP switches and jumpers if it is required to switch between
multiple measurement modes via an external switch (Save mode)
The required measurement modes are configured and saved sequentially.
1. Connect supply voltage.
2. Set DIP-switch 10, 11 and 12 to OFF position.
3. Set the measurement mode with jumpers at J4-J6 according to table
"Measurement mode”, column “Save mode".
4. Set DIP switch 10 to ON position.
5. Set upper scale caption for the first measurement mode with DIP switches 1-9
(see “Configuration of the upper scale caption”).
6. Set time base for graphical historic data display (optional, see “Other settings”).
7. Set DIP switch 10 to OFF position.
8. "Saved” appears on the display. The configuration of the first measurement
mode is now completed.
9. When using multiple measurement modes repeat steps 2-8 until all required
measurement modes are configured. The time base does not need to be set
again, as it is used for all measurement modes.
10. In operation the measurement modes can be changed with a switch at J4-J6
according to table "Measurement mode”, column “Save mode".
----- 4 -----