TDE Instruments Digalox DPM72-AV User guide

Instruction manual
Panel meter Digalox DPM72,
condensing
Failure to comply with these instructions will
Note: the measurement inputs of the device can carry life
device
The device must not be used as the only protective device or protective
Read instruction manual
The device is not intended to protect persons or facilities against harm. Specific
devices must be used to
connecting switches
is at least
when measuring
housing!
the instrument
carrying hazardous
x
all guarantee and warranty.
hazardous voltages must not be connected
carefully before operating the device!
relays
switches must
maximum occurring
must be isolated
flammable substances!
with external
Digalox DPM72-AV
Package contents:
Indoor use non
Panel mounting.
2. Safety instructions
voltages!
When working on the
the device!
shutdown.
(Rev-2015-04)
, non corrosive.
5x jumpers, 1
void
mounting bracket
-threatening
to
When
whose isolation voltage voltage. For example, least 500V.
Do not open the
Do not use
All cables
separators.
guarantee safety (protection
to the terminals J1-J6 only
twice the
250V AC switch
in the presence of explosive or
voltages must be secured
, off-switches, etc).
be used
measurement
for at
----- 1 -----
3. Description
olt AC/DC,
endpoint of the indicator as well as scaling of the analogue
signal and current transformer measurement can be adjusted freely via the DIP
alternatively the option for automatic adaption of the scale
used. The unit can be switched between volt and ampere measurement via an
external switch. Minimum and maximum values are recorded and can be displayed
al switch. Measure values are recorded over a
of 3 minutes up to 14 days. The
history can be activated by an external switch.
mpere, frequency and 5A current transformer
as well as the display of the measuring
Measure mode text
Upper scale caption
Graphical measurement
Digital measurement value
setting
upper
Supply voltage input
configuration
connection external
activating/switching
graphical historic data
max display
Measurement inputs
The measuring modes V are supported. Scale
switch,
optionally using an extern
A
time base
Front
1 2 3 4
endpoint can be
time span
display
Back
1 DIP switch for
mode and 2 3 Jumpers for
mode,
min/ 4
measurement
scale caption
or
switch for
measurement
display,
, auto-scaling
----- 2 -----
4. Electrical Connections
The device may only be operated in one of the connection options shown below.
10-24V
Fuse 250mA
U
s
DPM72 AV/P
Voltage and current
direct measurement
Measurement inputs
10A
I
high
Load
I
low
max.
Supply
U
high
max.
500V
+/L -/N
U
U
s
low
10-24V
Fuse 250mA
U
s
DPM72 AV/P
Voltage direct, current via
current transformer
Measurement inputs
I
low
Current
transformer
I
high
Load
U
Supply
high
max.
500V
+/L
U
U
-/N
s
low
5. Configuration
The device can be configured via DIP switches and jumpers.
Scale
In modes where a scaling is necessary to represent the correct measurement value (for example 5A AC for current transformer) the upper scale value corresponds to the primary value of the current transformer. Example: upper scale value = 50 = primary value of the current transformer
Configuration via DIP switches and jumpers if only one measurement mode is required (Manual mode)
1. Set DIP switch 10 to ON position.
2. Set measurement mode with jumpers at J4-J6 according to table "Measurement mode”, column “Manual mode".
3. Set upper scale caption for the first measurement mode with DIP switches 1-9 (see “Configuration of the upper scale caption”).
4. Connect supply voltage.
----- 3 -----
Configuration via DIP switches and jumpers if it is required to switch between multiple measurement modes via an external switch (Save mode)
The required measurement modes are configured and saved sequentially.
1. Connect supply voltage.
2. Set DIP-switch 10, 11 and 12 to OFF position.
3. Set the measurement mode with jumpers at J4-J6 according to table "Measurement mode”, column “Save mode".
4. Set DIP switch 10 to ON position.
5. Set upper scale caption for the first measurement mode with DIP switches 1-9 (see “Configuration of the upper scale caption”).
6. Set time base for graphical historic data display (optional, see “Other settings”).
7. Set DIP switch 10 to OFF position.
8. "Saved” appears on the display. The configuration of the first measurement mode is now completed.
9. When using multiple measurement modes repeat steps 2-8 until all required measurement modes are configured. The time base must not be set again, as it is used for all measurement modes.
10. In operation the measuring modes can be changed with a switch at J4-J6 according to table "Measurement mode”, column “Save mode".
Configuration of the upper scale caption
The upper scale caption is binary coded using DIP switches 1-9. Possible values are 1 to 500. Switch 1 corresponds to 256, switch 2 corresponds to 128, switch 3 corresponds to 64, etc., switch 9 corresponds to 1. For configuration, proceed as follows:
1. Connect supply voltage.
2. Set DIP switch 10 to ON Position.
3. Set DIP switches 1-9 to OFF position.
4. Start with 1st switch.
5. Set switch to ON position.
6. If the displayed value is greater than the desired value, set the switch back to OFF position.
7. If the displayed value is less than the desired value, leave switch to ON position and move on to the next switch.
8. Repeat steps 5 to 7 until desired value is displayed.
----- 4 -----
Table common switch combinations
Value Combination
(DIP switch 1-9)
10
20
25
ON
1 2 3 4 5 6 7 8 9
ON
1 2 3 4 5 6 7 8 9
ON
1 2 3 4 5 6 7 8 9
10 11 12
10 11 12
10 11 12
Table measurement mode
Measurement
Display Scale Manual mode Save mode
mode
V AC direct Volt
AC
A AC direct Ampere
AC
V DC direct Volt
DC
A DC direct Ampere
DC
AC Frequency Frequency
Hz
5A AC scaled (Current
Ampere
CT transformer) 1A AC scaled (Current
Ampere
CT transformer)
Value Combination
(DIP switch 1-9)
50
100
100
Upper
scale
caption
Upper
scale
caption
ON
1 2 3 4 5 6 7 8 9
ON
1 2 3 4 5 6 7 8 9
ON
1 2 3 4 5 6 7 8 9
10 11 12
10 11 12
10 11 12
ON
1 2 3 4 5 6 7 8 9
J1 J6
ON
1 2 3 4 5 6 7 8 9
J1 J6
ON
1 2 3 4 5 6 7 8 9
J1 J6
ON
1 2 3 4 5 6 7 8 9
J1 J6
ON
1 2 3 4 5 6 7 8 9
J1 J6
ON
1 2 3 4 5 6 7 8 9
J1 J6
ON
1 2 3 4 5 6 7 8 9
J1 J6
Value Combination
(DIP switch 1-9)
ON
1 2 3 4 5 6 7 8 9
ON
1 2 3 4 5 6 7 8 9
ON
1 2 3 4 5 6 7 8 9
ON
1 2 3 4 5 6 7 8 9
J1 J6
ON
1 2 3 4 5 6 7 8 9
J1 J6
ON
1 2 3 4 5 6 7 8 9
J1 J6
ON
1 2 3 4 5 6 7 8 9
J1 J6
ON
1 2 3 4 5 6 7 8 9
J1 J6
ON
1 2 3 4 5 6 7 8 9
J1 J6
ON
1 2 3 4 5 6 7 8 9
J1 J6
10 11 12
10 11 12
10 11 12
10 11 12
10 11 12
10 11 12
10 11 12
200
250
500
10 11 12
10 11 12
10 11 12
10 11 12
10 11 12
10 11 12
10 11 12
10 11 12
10 11 12
10 11 12
----- 5 -----
6. Other settings
The following functions can
jumper
by short
Graphical historic data display
To activate the graphical historic data display
connector J1 is shorted. The unit
stored values within the set time base as a
set to
minutes (3, 15, 30, 60).
be changed by
J1 (interval
For each subsequent
opening and closing the time base
DIP switch 10 must be switched
from ON to OFF position.
Max display connector J2 is
shorted. The display shows the maximum and
minimum values recorded since the last reset.
The values can be reset by opening and closing
connection J2 shortly (interval < 2 sec). The
display shows "minmax reset".
scaling J3 connector is
shorted. The device automatically changes the
upper scale caption depending on the current
measuring value in steps of 10, 100 and the set
connectors J1-J3 by
be activated independently
or switch during operation:
graphical trend.
The time base can be (6, 12, 24, 48, 72) or
The time base can opening and closing first opening and closing base is displayed.
the next setting. In order permanently,
-circuiting
displays the
days (7, 14), hours
alternatingly
< 2 sec). When
J1 the current time
changes to
to save the setting
Min-Max-display
To activate the Min-
Auto scaling
To activate auto-
of top scale caption.
----- 6 -----
7. Mounting
Carefully insert the instrument into the panel cutout. Insert the mounting bracket from the back and push towards the panel until instrument sits tight. Make sure the mounting bracket is snapped into the side of the housing. To ensure IP65 protection (dust and water jet) when mounting in a front panel, use optional gasket.
8. Specifications
DPM72AV
Supply 10-24V AC/DC
Display LCD graphic display 192x160 Pixel
Measuring range voltage ± 500V AC/DC
Accuracy voltage 1%
Measuring range ampere ± 10A AC/DC and 5A AC CT
Accuracy ampere 1%
Measuring range frequency 10-500Hz
Accuracy frequency 0,1Hz
Recording of measurement 3 min to 14 days
Operating temperature 0°C to +50°C
Storage temperature -20°C to +70°C
Front panel 72 mm x 72 mm
Panel cut out 68 mm x 68 mm
Protection front IP65 from the front
9. Cleaning
Observe the safety instructions before cleaning the instrument. Clean instrument with a dry lint-free soft cloth. Do not use solvents.
10. Contact Information TDE Instruments GmbH, Gewerbestraße 8, D-71144 Steinenbronn phone: +49 7157-20801 E-Mail: info@tde-instruments.de Internet: www.tde-instruments.de, www.digalox.com
----- 7 -----
Loading...