Takara Belmont Bel-Cypher_Pro_Installation_Manual_Ver3.8_0612 Bel-Cypher Pro Digital Panoramic Radiograph Installation Manual

B03-I216E Bel-Cypher Pro
Installation Manual
Digital Panoramic Radiograph
Bel-Cypher Pro
Notice
The classification is shown as follows According to the type of protection against electric shock. : Class According to the degree of protection against electric shock. : Type B applied part
TAKARA BELMONT CORPORATION.
B03-I216E Bel-Cypher Pro
cedures
read and understood before attempting to install
CAUTION
This manual provides information and instruction for the installation, assembly, and certification pro for the “Bel-Cypher Pro” X-Ray. The instructions contained in this book should be thoroughly
“Bel-Cypher Pro” unit. After the installation is completed, file this manual and refer back to it when per-
B03-I216E Bel-Cypher Pro
CONTENTS
1. Introduction 01 1 / 1
2. Warning 02 1 / 9
3. Pre-Installation Instructions 03 1 / 1
4. Specification 04 1 / 1
5. Name of each parts and Dimension of equipment 05 1 / 2
6. Checking of each components and accessories 06 1 / 2
7. Installation Procedure 07 1 /20
8. Methods to install a TWAIN driver in imaging software 08 1 /20
9. Post-Installation Instructions 09 1 / 5
10. Contact Information 10 1 / 1
11. Revised record of this manual
11 1 / 1
B03-I216E Bel-Cypher Pro
1. Introduction
1. Observe “Warning” and “Prohibition” matters in this Installation Manual.
2. Read this Installation Manual thoroughly to prevent an accident or trouble.
3. If you have any unclear matters in installation, reconfirm it by reading this Installation Manual.
4. After installation, read Operation Manual to understand operation procedures.
5. Discharge Be sure to observe Installation Manual. If accidents or troubles of the equipment happen due to im-
proper installation, we cannot be responsible for those accidents or troubles.
6. Repair and repair parts supply Repair and repair parts supply is available for 10 years from discontinued date.
7. mark means “ Attention, consult accompanying documents “.
01-1
B03-I216E Bel-Cypher Pro
Indicates an imminently hazardous situation, which, if not avoided, may result in death, serious injury and/or property
rdous situation, which, if not avoided, could result in death, serious injury and/or property
ded, may result in minor or moderate injury, and/or partial
2. Warning
WARNING
Always conform to the safety work standards to assure safety for workers and other people concerned. Repair work for internal parts of the equipment involves high risk. This should be strictly conducted by an au­thorized service personnel only.
NOTE
The warning messages in this Installation Manual are defined and classified as follows:
DANGER
damage such as total loss of equipment and fire when the instructions are not followed.
WARNING
Indicates a potentially haza damage such as total loss of equipment and fire when the instructions are not followed.
CAUTION
Indicates a potentially hazardous situation, which, if not avoi damage to property and loss of computer data when the instructions are not followed.
NOTE
Indicates information helpful for use of the equipment.
02-1
B03-I216E Bel-Cypher Pro
WARNING
Precautions on use (and safety and hazard prevention) of medical electric/electronic equip­ment
1. DO NOT allow the person without expertise to use the equipment.
2. Pay attention to the following when installing the equipment.
(1) Install the equipment to avoid splashing water. (2) Do not install the equipment to the place harmful to it due to atmospheric pressure, temperature,
humidity, airflow, direct sunlight, dust, salinity and air including sulfur.
(3) Pay attention to the safety conditions such as inclination, vibration and impact (including those
in transportation). (4) Do not install the equipment to the place where chemicals are stored or gas generates. (5) Pay attention to commercial frequency of power, its voltage and allowable current
(or power consumption). (6) Connect the equipment to the ground properly.
3. Check the following before using the equipment.
(1) Check contact of switches, polarity, dial settings and indicators. Check that the equipment operates
normally.
(2) Check that the equipment is connected to the ground perfectly. (3) Check that all cords are connected properly and perfectly. (4) Pay attention to parallel use with other equipment as it may cause erroneous diagnosis or danger. (5) Recheck the external circuit connected directly to a patient. (6) Check the battery power.
4. Pay attention to the following when using the equipment.
(1) Avoid exceeding the time and the radiation dose rate necessary for diagnosis. (2) Always monitor no abnormality in all equipments in use and the condition of a patient. (3) When any abnormality is observed in the equipment or the patient, take appropriate action such as
to stop the equipment in the condition safe for the patient.
(4) Do not allow a patient to touch the equipment.
5. Pay attention to the following after using the equipment.
(1) After returning operation switches and dials to their original position before use in accordance
with predetermined procedure, turn off the power.
(2) Do not pull cords when removing them from the equipment. (3) Be careful for the following items for storage.
I) Store the equipment where water is not splashed on it.
II) Do not store the equipment to the place harmful to it due to atmospheric pressure, temperature,
humidity, airflow, direct sunlight, dust, salinity and air including sulfur.
III) Pay attention to the safety condition such as inclination, vibration and impact (including
those in transportation).
IV) Do not store the equipment to the place where chemicals are stored or gas generates.
(4) After cleaning accessories, cords, and terminals, put them in order. (5) Keep the equipment clean for the next use.
6. When the equipment fails, DO NOT attempt to repair it by yourself. Indicate that the equipment is in failure and consign its repair to special engineer.
7. DO NOT modify the equipment.
8. Maintenance and inspection
(1) Be sure to perform regular inspection of equipment and its accessories. (2) When resuming to use the equipment that was not used for a long time, be sure to check that
the equipment operates normally and safely before use.
02-2
B03-I216E Bel-Cypher Pro
DANGER
This equipment is an electrical instrument. Avoid contact with water, which may cause electric shock and damage to the equipment.
This equipment is an electrical instrument incorporating high voltage circuits. Removing the equipment cover in the installation work may cause electric shock. Before opening the cover, ensure to turn OFF the power, and remove the electrical power plug from the outlet.
Installation work can contain hazardous works; allow only qualified engineers with suitable education and training to perform installation work.
Installation work can contain hazardous works; the installation engineer should instruct any other person not to enter into a scope of installation work.
02-3
B03-I216E Bel-Cypher Pro
This equipment has units which are more than 20kg. Heavy units should be brought up by more than 2
er source from the other electric
WARNING
This X-ray Unit may be dangerous to patient and operator unless safe exposure factors and operating in­structions are observed.
Follow the local zoning authority's regulations regarding X-ray room that you are going to install Bel-Cypher Pro.
The floor should have strength of 650N/ 1 0 0 cm
people.
Do not put things in area where equipment functions.
In order to avoid influence by other electric equipment, use separated pow equipment.
Those who install X-ray apparatus should wear X-ray protector apron.
2
and should be rigid.
Pay attention when Sliding Unit is moves up and down.
LASER RADIATION, DO NOT STARE INTO A BEAM, CLASS 1 LASER PRODUCT 1.Laser Beam is applied. For safety, instruct patient not to look at the laser beam.
2.Before the beam is lightened, lower Frankfort Plane Beam to bottom. 3.Do not set the beam to patient's eyes.
CAUTION
Do not turn the Arm unit by hand. It might cause a trouble of the equipment.
02-4
B03-I216E Bel-Cypher Pro
WARNING
Environmental Conditions of EMC (Electromagnetic Compatibility)
The equipment satisfies the EMC(Electromagnetic Compatibility) standard below:
IEC 60601-1-2:2007
Classification of EMI in Accordance with IEC 60601-1-2:2007 Group 1,Class B
1. The medical electrical equipment needs special precautions regarding EMC and needs to be in­stalled and put into service according to the EMC information provided in the Operation manual.
2. The portable and mobile RF communications equipment can affect medical electrical equipment.
3. The use of accessories, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of the equipment as replacement parts for in­ternal components, may result in increased emissions or decreased immunity of the equipment.
4. The cable that suits EN60601-1-2:2007 of standards emission and the requirement of the immuni-
ty are recorded below.
1) Power cable (4.5m)
2) Exposure Switch cable (10m)
3) LAN cable(type braided shield) for connection with personal computer (10m)
5. The equipment should not be used adjacent to or stacked with other equipment.
6. The equipment should be used only in the specified type of shielded location.
7. Please use the LAN Cable specified by EC representative.
8. Essential performance
1) Accuracy of x-ray tube loading factor.
2) Reproducibility of radiation output.
02-5
B03-I216E Bel-Cypher Pro
Guidance and manufacturer’s declaration
electromagnetic emissions
WARNING
The model Bel-Cypher Pro is intended for use in the electromagnetic environment specified be­low. The customer or the user of the model Bel-Cypher Pro should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC61000-3-2
Voltage fluctuations/ Flicker emissions IEC61000-3-3
Group 1
Class B
Not Applicable
Not Applicable
The model Bel-Cypher Pro uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
The model Bel-Cypher Pro is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
02-6
B03-I216E Bel-Cypher Pro
Guidance and manufacturer’s declaration
– electromagnetic imm
unity
WARNING
The model Bel-Cypher Pro is intended for use in the electromagnetic environment specified below. The customer or the user of the model Bel-Cypher Pro should assure that it is used in such an en­vironment.
Immunity test
Electrostatic discharge (ESD) IEC61000-4-2
Electrical fast transient/burst IEC61000-4-4
Surge IEC61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines IEC61000-4-11
Power frequency (50/60Hz) magnetic field IEC61000-4-8
IEC 60601 test level ±6 kV contact ±8kV air
±2 kV for power supply lines ±1 kV for input/output lines ±1kV differential mode ±2kV common mode
5 Ut (95 dip in Ut) for 0.5 cycle 40 Ut (60 dip in Ut) for 5 cycles 70 Ut (30 dip in Ut) for 25 cycles 5 Ut (95 dip in Ut) for 5 sec
Compliance level ±6 kV contact ±8kV air
±2 kV for power supply lines ±1 kV for input/output lines ±1kV differential mode ±2kV common mode
5 Ut (95 dip in Ut) for 0.5 cycle 40 Ut (60 dip in Ut) for 5 cycles 70 Ut (30 dip in Ut) for 25 cycles 5 Ut (95 dip in Ut) for 5 sec
Electromagnetic environment
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30.
Mains power quality should be that of a typical commercial or hospital environment.
Mains power quality should be that of a typical commercial or hospital environment
Mains power quality should be that of a typical commercial or hospital environment
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of a typical location in a typical commercial or hospital environment
NOTE Ut is the a.c. mains voltage prior to application of the test level.
- guidance
02-7
B03-I216E Bel-Cypher Pro
Guidance and manufacturer’s declaration
– electromagnetic
i
mmunity
Immunity test
IEC 60601
test
Compliance
Electromagnetic environment
The model Bel-Cypher Pro is intended for use in the electromagnetic environment specified be­low. The customer or the user of the Bel-Cypher Pro should assure that it is used in such an en­vironment.
WARNING
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
3 Vrms
150kHz to
80MHz
3 V/m
80MHz to
2.5 GHz
level
level
3 Vrms
3 V/m
- guidance
Portable and mobile RF communications equipment should be used no closer to any part of the Bel-Cypher Pro, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d =1,2
P
d =1,2P 80 MHz to 800 MHz
d =2,3
800 MHz to 2.5 GHz
P
where P is the maximum output power rating of the transmitter in watts(W) according to the transmitter manufacturer and d is the recommended separation distance in me­tres(m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, ashould be less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following sym­bol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) tele-
phones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Bel-Cypher Pro is used exceeds the appli­cable RF compliance level above, the Bel-Cypher Pro should be observed to verify normal op­eration. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Bel-Cypher Pro.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
02-8
B03-I216E Bel-Cypher Pro
Recomme
nde
d separation distances between
R
ated maximum
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
WARNING
portable and mobile RF communications equipment and the Bel-Cypher Pro
The Bel-Cypher Pro is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Bel-Cypher Pro can help prevent elec­tromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Bel-Cypher Pro as recommended below, accord­ing to the maximum output power of the communications equipment.
output power of
transmitter
W
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.72 1 1.20 1.20 2.30
10 3.79 3.79 7.27
100 12.00 12.00 23.00 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the
P
transmitter, where the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
is the maximum output power rating of the transmitter in watts(W) according to
d
=1,2P
m
d
=1,2P
d
=2,3P
02-9
B03-I216E Bel-Cypher Pro
3. Pre-Installation Instructions
1. Manuals
1) Installation Manual of Bel-Cypher Pro
2) Operation Manual of Bel-Cypher Pro
2. Measuring instruments and tools
2-1. Measuring instruments
1) Digital Multi Meter with an accuracy of 1%, capable of measuring 150VAC and 20mA DC, and capable of indicating true RMS value within one second
2) Fluorescent Screen
2-2. Tools
1) Philips Head Screwdrivers (Small and Big)
2) Slotted Head Screwdrivers (Small, Anti-Static type)
3) Nut Drivers (M6, M5, M4 and M3)
4) Ratchet wrench
5) Allen keys
6) Cutting Nippers
7) Long nose nippers
8) Hammer
9) Electric power drill (for making a hall on the wall and the floor)
10) Drill bit 8.5mm = 21/64”(which can drill in wall and a floor )(Driving core bar typed)
11) Drill bit 12.5mm = 31/64”(which can drill in wall and a floor )(Driving anchor)
12) Blower or dust separator
2-3. Others
1) Alcohol for disinfection
2) A rag
3) Cleaning solution
3. User supplied Network Card (NIC) and PC.
NOTE: IN GENERAL, MAJOR STRUCTURAL MODIFICATIONS ARE NOT REQUIRED,
HOWEVER THE FLOOR ON WHICH Bel-Cypher Pro IS PLACED SHOULD BE ABLE TO SUPPORT 600 lbs. DEAD LOAD.
3 Electrical Requirements
1) Power Supply Bel-Cypher Pro X-Ray operates on a power supply of 120 VAC. A three wire GROUNDED circuit,
separately connected to the central distribution panel with an over current protection device rated for 15 amperes. Recommended wire size is 12 AWG. But if the wire run distance is to exceed 50 feet, 10 AWG is required. For wire run distance in excess of 75 feet, up to 125 feet, 8 AWG is required.
2) All connections, workmanship and materials used must comply with the national Electric Code and
local codes.
03-1
B03-I216E Bel-Cypher Pro
4. Specifications
General name Digital Panoramic Radiograph Model Bel-Cypher Pro
Power Voltage
Power Capacity 9A (80kV 8mA) High Tension Generator
X-ray Tube Voltage
X-ray Tube Current
Exposure method Manual X-ray Tube D-052SB (Toshiba) X-ray Tube Focal Spot 0.5 mm Total Filtration 2.5mmAl (minimum)
Radiographic Mode
Exposure Time
Magnification Ratio
Image receiver Digital Sensor Beam for patient positioning 3 beams Dimension W:850mm x D:1,135mm x H:2,275mm Weight Approximately 231 lb(105 kg)
Environmental condition for Operation
Temperature : 41~95F (535℃) Humidity : 3085% Pressure : 7001060 hpa
Environmental condition for Storage Temperature : 14~140F (1060℃) Humidity : 1095% Pressure : 7001060 hpa
Environmental condition for Transportation Temperature : 14~140F (1060℃) Humidity : 1095% Pressure : 7001060 hpa
120Vac 60Hz 1φ Acceptable voltage regulation: ±10%
High Tension Generator (100kHz)
60kV to 80kV (1kV step) Tolerance 54kV~88Kv(±10%of 60kV 80kV
2 to 8mA (1mA step) Tolerance 1.6mA~9.6mA(±20%of 2mA 8mA
Panorama: (Adult, Child), Panoramic Bitewing (Adult, Child)
TMJ: Panorama: 10 sec. Panoramic Bitewing:4.6 sec.
TMJ: 2.5 sec. (x 4) Panorama: 1.2 to 1.3, Panoramic Bitewing: 1.2 to 1.3 TMJ: 1.2
04-1
B03-I216E Bel-Cypher Pro
5. Name of each part and Dimensions of equipment
1.Names of each parts of equipment
Arm Unit
Rotation Unit
Control Unit
Image
detector
X-Ray source Unit
Rest Unit
Emergency Stop
Switch
Sliding Unit
Stand
Exposure Switch
Power Switch
Emergency Stop
Switch (Option)
05-1
B03-I216E Bel-Cypher Pro
2Dimensions of equipment
05-2
B03-I216E Bel-Cypher Pro
6. Parts and Accessories for equipment
Parts
No. Number
Stand + Sliding Unit
1 1 set.
Wall bracket
2 1 pce.
3 Fixing screw for Wall bracket
Rotation Unit
4 1 set.
5 Fixing bolt for Rotation Unit
Rest Unit
6 1 set.
7 Fixing screw for Rest Unit
8 Rotation Unit Cover
9 Fixing Screw for Rotation Unit Cover
10 Sliding Unit Cover
11 Fixing Screw for Sliding Unit Cover A 2 pcs.
Exposure Switch
12 1 pce.
Exposure Switch Cable
13 1 pce.
LAN Cable
14 1 pce.
Emergency Stop Switch (Option)
15 1 pce.
16 Anchor (Driving anchor) (A) GA-8M 4 pcs.
17 Anchor (Driving core bar typed) (B)
18 Coach screw (C)
19 Spcialized rod for Driving anchor 1 pce.
Collar(B-1)
20 4 pcs.
21 Bolt to use with Driving anchor 4 pcs.
22 Cover Cap for hexagon head bolt
Sponge(S-1)
23 1 pce.
24 Steel s pacer (Hexagonal) 2 pce.
Hexagon socket head cap screw with spring and flat washer M6×12
Metal washer d7×D22×t1.6
Metal washer d9×D26×t2.3
Hexagon head bolt (grade 10.9) M8×25
Flat washer φ8
Spring washer φ8
Taper pin with thread
Hexagon socket head cap screw with spring and flat washer M5×12
Taper pin with thread
Rotation Unit Cover A
Rotation Unit Cover B
Rotation Unit Cover C
Cross recessed head screw with captive washer M3×10
Binding head screw (White) M3×8
Sliding Unit Cover A
Sliding Unit Cover B
Cross recessed head screw with captive washer M3×20
Wall : C8-50 2 pcs.
Floor : C8-50 4 pcs.
Wall : φ8×40 2 pcs.
Floor : φ8×65 4 pcs.
GA-25H
Hexagon socket countersunk head screw M8×23
Wall : Plastic Cap
Floor : Plastic Cap
BSF-328E
2 pcs.
2 pcs.
2 pcs.
4 pcs.
4 pcs.
4 pcs.
2 pcs.
4 pcs.
2 pcs.
1 pce.
1 pce.
1 pce.
6 pcs.
4 pcs.
1 pce.
1 pce.
2 pcs.
4 pcs.
06-1
B03-I216E Bel-Cypher Pro
No. Number
Accessories
1 Temporal support(Left and right) 1 pair
2 Chinrest 1 pce.
3 Bite Block 2 pcs.
4 Ear Rods(Left and right) 1 pair
5 CD containing TW AIN Driver 1 pce.
06-2
B03-I216E Bel-Cypher Pro
low
7. Installation Procedure
There are some hole/screw hole on this device, because of sharing some parts with other device. Describe about hole/screw hole which uses in this device below.
1. Remove packing material cover for Pillar, then remove pillows shown in be drawing they are used for pillars at installation.
Two pillows
Put vinyl in the packing or waste cloth on the pillow for protecting scratch of Pillar. Then, put the Pillar on the pillows (Sliding Unit is down side).
Top view
Side view
Pillow
Put waste cloth on packing vinyl on the pillows to avoid scratch.
2. Wire power cable, X-ray Exposure Switch and LAN Cable between relay box and PC.
Pillar without upper cover is shown.
Take out pillows and put them on the floor as shown below.
Floor
Pillow
07-1
B03-I216E Bel-Cypher Pro
1
2.1. Remove the cover (A) of relay box.
Remove biding head screws (white) which fixed the cover.
Relay box cover (A)
2.2. Remove covers in left.
07-2
B03-I216E Bel-Cypher Pro
2.3. Connect X-ray Exposure Switch Cable and LAN Cable.
LAN Cable from PC
Cable fixed base
Please make sure to fix a cable to cable fixed base with a cable tie after connecting X-ray Exposure Switch and LAN Cable, so the cable don’t come off.
2.4. Attach relay box by using biding head screws (white).
X-ray Exposure Switch Cable
Emergency Stop Switch Cable (Option)
CAUTION
Please pay attention to attach a cover when connecting the cable.
Minimum bending radius of LAN Cable is 30mm.
CAUTION
Please make Bending radius of LAN Cable longer than 30mm In case of Bending radius is shorter than 30mm, it will be the cause for damage of LAN Cable and fail­ure loose connection.
07-3
B03-I216E Bel-Cypher Pro
3. Attach wall bracket to the stand.
Hexagon socket head cap screw with spring and flat washer M6 x 12 : 2 pcs. Metal washer d7 x D22 x t1.6 : 2 pcs.
Hexagon socket head cap screw with spring and flat washer
Wall bracket
4. Reference Drawing for Equipment Installation.
Metal washer
07-4
B03-I216E Bel-Cypher Pro
5. Please go to provision 6, if you have not free-standing base.
Please go to provision 7, if you have free-standing base.
6. In case of installing by not using free-standing base.
6.1. Put the stand on the installation position.
6.2. To fix the flange base to the floor with an anchor (Driving anchor) (A) or anchor (Driving c
ore bar typed) (B) or coach screw (C).
Flange base
Hexagon socket countersunk head screw M8×23
Collar(B-1)
Anchor (A)
A
Nut
Washerφ9
Collar(B-1) *1
B
C
Plastic Cap
Coach screw φ8×65
Washerφ9
Collar(B-1)*1
*1)Flat side must be upper side.
[Construction case, an example of using Anchor (Driving anchor) (A).]
(1). Make a hole of drilling depth (39mm) by using a drill which is φ12.5 at the position for fixing an
equipment.
Perforation marking.
Drill
(2). Clean up inside of the hole.
Anchor (B)
Remove chips from a hole by using a blower or dust separator.
07-5
B03-I216E Bel-Cypher Pro
(3). Insert an Anchor (Driving anchor) (A).
(4). Drive an anchor into the ground by using the specialized driving rod.
(5). Fix flange base with hexagon socket countersunk head screw.
Collar(B-1)
[Construction case, an example of using Anchor (Driving core bar typed) (B).]
(1). Make a hole of 30mm deep at the position for fixing equipment with a drill of φ8.5.
07-6
Perforation marking.
Drill
Hexagon socket countersunk head screw
Hammer
Specialized rod for Driving anchor
Flange base
B03-I216E Bel-Cypher Pro
(2). Clean up inside of the hole.
(3). Attach a nut (M8) to a screw of Anchor (B).
(4). Insert until nut of Anchor (B) has reached the floor.
(5). Drive pin in until the edge of it has reached the edge of screw.
Remove chips from a hole by using a blower or dust separator.
5 or 6 of turns should be seen.
Hammer
07-7
B03-I216E Bel-Cypher Pro
(6). Remove a nut.
[Construction case, an example of using coach screw.]
After making a pilot hole with a drill which is smaller than φ4, fix it with coach screw.
7. In case of installing by using free-standing base.
7.1. Put stand on the free-standing base, then fix it by using 4 pieces of flat head bolts.
Hexagon socket countersunk head screw M8 x 23 : 4psc.
Stand
M8
M8
Hexagon socket countersunk head screw
Collar(B-1)
Free-standing base
07-8
B03-I216E Bel-Cypher Pro
7.2. Move free-standing base to installing position, then fix it at 4 locations with anchor or coac
h screw. Please refer to provision 4 about construction of the floor.
8. Adjust installation position for an equipment and fixing position (the distance between The s
tand and a wall) for a wall bracket.
8.1. In case of fixing wall bracket with Anchor (B).
8.1.1 Loosen a hexagon socket head cap screw with spring and flat washer for adjusting fixin
g position for a wall bracket.
8.1.2 Put a wall bracket on the wall, and then have 2 hexagon socket head cap screws with
spring and flat washer tightened temporarily. (not final tightening)
8.1.3 Drive Anchor (B) in
Please refer to the provision 6 [Construction case, an example of using Anchor (B)].
8.1.4 Fix a wall bracket with a nut.
Metal washer d9 x D26 x t2.3
8.1.5 Attach a Plastic Cap to a nut.
8.1.6 Tighten a hexagon socket head cap screw with spring and flat washer.
Plastic Cap
Anchor(B)
Nut
Metal washer d7 x D22 x t1.6
Hexagon socket head cap screw with spring and flat washer
Metal washer d7 x D22 x t1.6
Wall bracket
Anchor
Metal washer d9 x D26 x t2.3
Hexagon socket head cap screw with spring and flat washer
07-9
B03-I216E Bel-Cypher Pro
8.2. In case of fixing a wall bracket with Coach screw.
8.2.1 Loosen a hexagon socket head cap screw with spring and flat washer for adjusting fixin
g position for a wall bracket.
8.2.2 Put a wall bracket on the wall, and then have 2 hexagon socket head cap screws with
spring and flat washer tightened temporarily.(not final tightening)
8.2.3 Fix a wall bracket with a Coach screw. Please refer to the provision 6 [Construction case, an example of using Coach screw].
8.2.4 Attach a Plastic Cap to the Coach screw.
8.2.5 Tighten a hexagon socket head cap screw with spring and flat washer.
Coach screw
Plastic Cap
Metal washer d7 x D22 x t1.6
Metal washer d9 x D26 x t2.3
Plastic Cap
9. Mounting Rotation Unit on Sliding Unit
9.1. Hook to the U-notch at Sliding Unit.
9.2. Insert taper pins with thread with a hammer.
9.3. Tighten and Fix hexagon head bolts (grade 10.9).
Hexagon head bolt (grade 10.9) M8×25 : 4pcs. Spring washer φ8 : 4pcs. Flat washer φ8 : 4pcs.
Taper pin with thread
Sliding Unit
Hexagon socket head cap screw with spring and flat washer
Wall bracket
Metal washer
d9 x D26 x t2.3 Coach screw φ8 x 40
Rotation Unit
07-10
B03-I216E Bel-Cypher Pro
10. Remove Carriage Holder from Rotation Unit
10.1. Take the carriage holder off the Rotation Unit.
Nut with spring washer φ8 : 4pcs. Hexagon socket head cap screw with spring and flat washer M6 x 12 : 4pcs.
Nut with spring washer
11. Remove the protection plate for Median Beam
Cross recessed head screw with captive washer M3 x 6 : 2pcs.
Protection plate for Median Beam
Hexagon socket head cap screw with spring and flat washer
Cross recessed head screw with captive washer
07-11
B03-I216E Bel-Cypher Pro
12. Connect relay connector in Rotation Unit to the Connector in Sliding Unit. Please make sure to check mark of connector and then connect it.
12.1. Relay connector in Rotation Unit.
Mark of connector : CN1CNUMCNPSHBCN5_6_7
CN5_6_7CN11
Switching HUB
Mark of connector : CNLAN1FPD CN18
12.2. Relay connector in Sliding Unit.
Mark of connector : CNS5CN13CN14
07-12
B03-I216E Bel-Cypher Pro
13. Attach Rest Unit of Sliding Unit.
13.1. Align positioning pins of Rest Unit ASSY to positioning holes of Sliding Unit.
13.2. Attach Chinrest Unit by hexagon socket head cap screw with spring and flat washer.
Hexagon socket head cap screw with spring and flat washer M5 x 12 : 4pcs.
Hexagon socket head cap screw with spring and flat washer
07-13
Positioning hole
Hexagon socket head cap screw with spring and flat washer
Rest Unit
B03-I216E Bel-Cypher Pro
14. Attach Rotation Unit Cover C of Rotation Unit.
14.1. Slowly rotate arm to 90 degrees position.
Cross recessed head screw with captive washer M3 x 10 : 2pcs.
Rotation Unit Cover C
Cross recessed head screw with captive washer
Cross recessed head screw with captive washer
07-14
B03-I216E Bel-Cypher Pro
15. Attach Rotation Unit Cover B of Rotation Unit.
Cross recessed head screw with captive washer M3 x 10 : 4pcs.
Cross recessed head screw with captive washer
Cross recessed head screw with captive washer
07-15
Rotation Unit Cover B
B03-I216E Bel-Cypher Pro
16. Attach Sliding Unit Cover A of Sliding Unit.
16.1. Connect relay connector of Sliding Unit Cover A.
Mark of connector : CN11P
16.2. Fix cover A of Sliding Unit by cross recessed head screw with captive washer.
Cross recessed head screw with captive washer M3 x 20 : 2pcs.
CN11P
07-16
Cross recessed head screw with captive washer
Sliding Unit Cover A
These screw holes do not use.
B03-I216E Bel-Cypher Pro
17. Put Frankfurt plane knob to Sliding Unit.
Frankfurt plane knob
Sliding Unit Cover A
18. Attach Sliding Unit Cover B of Sliding Unit.
18.1. Insert the upper side of Sliding Unit Cover B into underside of Sliding Unit Cover A,
then the underside of Sliding Unit Cover B will be inserted in pin and fixed.
07-17
Sliding Unit Cover A
These screw holes do not use.
Sliding Unit Cover B
Insert in pin
B03-I216E Bel-Cypher Pro
19. Put the Rotation Unit Cover A on Rotation Unit.
19.1. Attach steel spacer (hexagonal) on Rotation Unit.
Steel spacer (hexagonal) BSF-328E : 2pcs.
19.2. Hang the tip of the cover at a hook at rotation unit.
19.3. Fix Rotation Unit Cover A by binding head screws.
Binding head screw (white) M3 x 8 : 4pcs.
07-18
Binding head screw (white)
steel spacer (hexagonal)
Rotation Unit Cover A
Cover tip
Hook
B03-I216E Bel-Cypher Pro
20. Paste sponge(S-1)
Sponge(S-1) should be pasted on relay Box cover(B) with double sided seal for making less gap from sliding unit as possible.
*If you hear sound of rubbing sponge when confirming operation check, then paste sponge (S-1) on the position that is not conspicuous sound.
07-19
Relay box cover (B)
Sponge(S-1)
Sliding Unit
B03-I216E Bel-Cypher Pro
21. Insert accessories to Rest Unit.
Confirm if the desorption of accessories is possible smoothly.
Accessories should be attached on each radiographic mode, so please refer to the following tabl
e.
Accessories
Temporal support
Chinrest
Bite Block
Ear Rods
Temporal support (Left)
Panorama Panoramic Bitewing TMJ
Chinrest
Ear Rods (Left)
Radiographic Mode
Bite Block
Temporal support (Right)
Ear Rods (Right)
07-20
B03-I216E Bel-Cypher Pro
8Procedures to install a TWAIN driver in imaging software
1. Set IP address
1) OpenControl Panel
2) ChooseNetwork and Internet
3) ChooseNetwork and Sharing Center
4) ChooseChange adapter settings
08-1
B03-I216E Bel-Cypher Pro
5) Right click LAN Connection that is connected to Bel-Cypher Pro. Choose property. If you
use the LAN card supplied with Bel-Cypher Pro, open activeLocal Area Connection X
6) Double click Internet Protocol Version 4(TCP/IPv4)inThe connection uses the fol­lowing items:
7) ChooseUse the following IP address
8) Set IP address and subnet mask as follows
You need to set following values first, even though you change IP address later
IP address192.168.2.33 Subnet mask:255.255.255.0
9) Enter only IP address and Subnet Mask. Click OK to close all windows
08-2
B03-I216E Bel-Cypher Pro
2. Set Jumbo Packet
1) Go toControl Panel」→「System and Security」→「Device Manager
2) Right click LAN Connection that is connected to Bel-Cypher Pro. Then choose property. If you use the NIC supplied with Bel-Cypher Pro, right click and choose property.
3) Property screen of NIC comes up.
4) Choose Advancedtab
5) Choose Jumbo PacketinProperty
6) Set the value as9014 Bytes
(4)
(6)
(5)
(7)
7) Click OKto close property screen and device manager screen
08-3
B03-I216E Bel-Cypher Pro
3. Uninstall DCAM-API and Twain Driver
If you setup new PC, please go to next section.
3.1 Bel-Cypher Pro If you used an old version of DCAM-API and Bel-Cypher Pro Twain Driver in the PC, you have to remove it before installing newer version.
1) Open “Control Panel”.
2) Click “Uninstall a program”.
3) Uninstall “DCAM-API Driver for GigE Cameras” and “Bel-Cypher Pro”.
4) Open the properties of the network connected to the device.
5) If “eBUS Universal Pro For Ethernet Driver” and “GigE Vision Filter Driver” exists, please uninstall it.
6) Execute batch file “deleteDCAMpath.vbs” in install Disc.
08-4
B03-I216E Bel-Cypher Pro
7) The confirmation message window is displayed. Click “OK”.
8) After completion, the message window is displayed. Click “OK”.
9) Restart the PC.
3.2 Bel-Cypher N If you used Bel-Cypher N, you have to remove DCAM-API for Bel-Cypher N in the PC, before installing it for Bel-Cypher Pro.
1) Open “Control Panel”.
2) Click “Uninstall a program”.
3) Uninstall “DCAM-API Driver for GigE Cameras”.
4) Open the properties of the network connected to the device.
08-5
B03-I216E Bel-Cypher Pro
5) If “eBUS Universal Pro For Ethernet Driver” and “GigE Vision Filter Driver” exists, please uninstall it.
6) Execute batch file “deleteDCAMpath.vbs” in install Disc.
7) The confirmation message is displayed. Click “OK”.
8) If the confirmation message is displayed. Click “OK”.
08-6
B03-I216E Bel-Cypher Pro
9) After completion, the message window is displayed. Click “OK”.
10) Restart the PC.
NOT E : If error message was displayed during running batch file, please take the following pro-
cedure.
1) Hold windows key + R. Type “regedit” and click “OK”.
2) Open the registry editor.
3) Delete “HKEY_LOCAL_MACHINE¥SYSTEM¥ControlSet001¥Services¥WINIO”.
4) Restart PC.
5) Execute batch file again.
08-7
B03-I216E Bel-Cypher Pro
4. Installation of Bel-Cypher Pro Twain Driver and DCAM-API
1) Turn on PC.
2) Set Bel-Cypher Pro TWAIN Driver of Install Disc. If the driver is not installed automatically, then run “Setup.bat” stored in a CD
3) The Visual C++ Runtime Setup screen such like Fig. 4-1 is displayed. Click “Install”
Fig 4-1 Visual C++ Run time Setup screen
4) Check “I agree” and Click “Install” such like Fig. 4-2 is displayed. Click “Install”
Fig 4-2 Visual C++ Run time Install screen
5) The dialogue such like Fig. 4-3 is displayed. Click “Close”.
Fig 4-3 Installation Complete screen
08-8
B03-I216E Bel-Cypher Pro
6) After the installation was completed, the dialogue such like Fig. 4-4 Bel-Cypher Pro TWAIN Setup Wizard starts. Click “Next”
Fig 4-4 Bel-Cypher Pro TWAIN Setup Wizard
7) The “Select install folder” dialogue such like Fig. 4-5 is displayed. Click “Next”
Fig 4-5 “Select Installation Folder” screen
8) The dialogue such like Fig. 4-6 is displayed. Click “Next”
Fig 4-6 Confirm Installation screen
08-9
B03-I216E Bel-Cypher Pro
9) The dialogue such like Fig. 4-7 is displayed. Bel-Cypher Pro TWAIN is installed.
Fig4-7 Installing Bel-Cypher Pro TWAIN screen
10) The dialogue such like Fig. 4-8 is displayed. Click “Close”
Fig 4-8 Installation Complete screen
13) Click “Install”.
08-10
B03-I216E Bel-Cypher Pro
14) Click “Next”.
15) Check “I accept…” and click “Next”.
08-11
B03-I216E Bel-Cypher Pro
16) Click “Install”.
17) Click “Finish”
16) Installation is completed
08-12
B03-I216E Bel-Cypher Pro
5. Setting of Firewall
1) Open Control Panel.
2) Open “Windows Firewall” from Control Panel.
08-13
B03-I216E Bel-Cypher Pro
3) Click “Allow a program or feature through Windows Firewall”.
4) Click “Change settings”.
08-14
B03-I216E Bel-Cypher Pro
5) Click “Allow another program”.
6) Click the “Browse” and select radiograph software (C:¥Bel-CypherPro¥app¥NeoExpGroup¥NeoExp.exe).
08-15
B03-I216E Bel-Cypher Pro
7) Click “Add”.
8) Check “Private” and ”Public” and click “OK”.
9) Click OK to finish Firewall setting.
08-16
B03-I216E Bel-Cypher Pro
6. Select image processing type.
1) Open “C¥Bel-CypherPro¥app¥NeoExpGroup¥2DRecon¥BelCypherPro¥
PanoramaSetting.ini”.
2) Select image processing type of the initial image displayed after irradiation.
0 : Standard 1 ; Type1 2 : Type2 3 : Type3 4 : Type4
example of image processing type
0 : Standard
1 : Type1 2 : Type2
3 : Type3 4 : Type4
08-17
B03-I216E Bel-Cypher Pro
7. Install Bel-Cypher Pro and Bel-Cypher N Twain Driver. You can operate Bel-Cypher Pro and Bel-Cypher N with one PC by changing connection of LAN cable. If you use only Bel-Cypher Pro, please skip this procedure.
1) Install Bel-Cypher N Twain Driver and DCAM-API. Refer to Installation Manual of Bel-Cypher N.
2) Uninstall only DCAM-API for Bel-Cypher N. Refer to section 8.3.2 in this manual.
3) Install Bel-Cypher Pro Twain Driver and DCAM-API. Refer to section 8.4-6 in this man­ual.
NOT E : When you change connection of LAN cable to another device, you have to change IP ad-
dress according to connected device.
Open the properties of the network connected to the device, and change IP address of
Internet Protocol Version 4(TCP/IPv4) before you use another device.
Bel-Cypher Pro : IP address : 192.168.2.33 Subnet mask : 255.255.255.0 Bel-Cypher N : IP address : 169.254.87.101 Subnet mask : 255.255.0.0
08-18
B03-I216E Bel-Cypher Pro
8. Firmware update of Bel-Cypher Pro. After Bel-Cypher Pro Twain Driver is installed, if installer includes a newer firmware data, firmware will be automatically updated at the first connection with Bel-Cypher Pro.
1) If there is a newer firmware data, firmware update will automatically start when software is launched for the first time. Please wait while below message is displayed.
2) After completion, the message window is displayed. Click “OK”.
3) After update finished, please restart Bel-Cypher Pro.
08-19
B03-I216E Bel-Cypher Pro
CAUTION
Recommended specifications of Personal Computer
Recommended PC specifications CPU Video Controller Memory OS Windows 7 Professional 64-bit, English Version
Other
Minimum PC specifications CPU Video Controller Memory OS Windows 7 Professional 64-bit, English Version
Other
Use a LAN cable of Category 5e or better
Recommended specifications of the image display monitor resolution Grayscale Tones 256 tones Display Colors 16.77 million : supported 8-bit
Intel Core i7-6600U (2.6GHz) or later NVIDIA GeForce 930M or more, or equivalent 16GB or more
Windows 8 Professional 64-bit, English Version Windows 10 Professional 64-bit, English Version
Ethernet ×1 (1000BASE-T)
Intel Core i5-6300U (2.4GHz) or later Intel HD Graphics 520 or more, or equivalent 8GB or more
Windows 8 Professional 64-bit, English Version Windows 10 Professional 64-bit, English Version
Ethernet ×1 (1000BASE-T)
1920×1080
PC and monitor are required to meet the international IEC60950-1 or IEC60601-1 standard.
If you run software that heavily loads the PC, transfer & acquisition of images could be interrupted. Please schedule Windows updates or virus scan updates to be performed outside clinic hours.
08-20
B03-I216E Bel-Cypher Pro
9. Post-Installation Instructions
1. Check listed items by referring to the Operation Manual.
1) Measurement of Feedback voltage and tube current
Make sure the main power is off. Remove covers from Rotation Unit. (Figure 1) Cover radiation aperture with lead Turn on the power of PC that is connected to Bel-Cypher Pro. Hit Ctrl + Alt + T simultaneously to g
et into Test mode.
Set exposure condition by referring the Operation Manual.
1) Exposure Orbit Panorama
2) Manual Exposure
3) Tube Voltage: 70kV
4) Tube Current: 8mA
Irradiate X-ray, then measure the feedback voltage between CP11 & CP20 (Tube voltage),and
between CP12 & CP20 (Tube current). (Figure 2)
INV 80(PFC)A BOARD
09-1
Figure 2
B03-I216E Bel-Cypher Pro
2. Confirmation of movement of X-ray irradiation.
2.1. Check listed items by referring to the Operation Manual.
2.2. Confirm the operation with X-ray
2.2. 1. Cover radiation aperture with lead.
2.2. 2. Set exposure condition by referring the Operation Manual.
1) Exposure Orbit Panorama
5) Manual Exposure
6) Tube Voltage: 60kV
7) Tube Current: 2mA
2.2.3. Irradiate X-Ray and confirm the operation.
09-2
B03-I216E Bel-Cypher Pro
3. Check listed items by referring to the Operation Manual
1. Power
1) Measurement of Input Power Voltage Vac
2) Does 1) meet the rating description on the Controller plate?
3) Rating Values of the Circuit Protector on the Rotation Unit
4) Are there any problems when the power plug is insert­ed?
5) Does the Power Code have a scratch or a crack?
Voltage V Current A
Heat Allophone Off-flavor
OK NG
OK NGOK NGOK NG
OK NG
09-3
B03-I216E Bel-Cypher Pro
4. Operation
1) After Power On, does the main body have a problem? Heat
Allophone Off-flavor
2) After depressing the "RESET" key, is “READY” displayed?
3) Does the main body move with up / down switch?
4) Does up / down operation of main body have a prob-
5) Does Sliding Unit stop at the highest and the lowest position?
6)
6-2) Dose Forward / backward operation of Focus Beam on ARM work
Allophone
lem?
Positioning Beams in Panorama 6-1) Are all positioning beams turned on by depressing forward / back-
ward switch of Focus Beam?
by depressing forward / backward switch?
OK NG OK NG OK □ NG
OK NG
OK NG
OK NG
OK NG
OK NG
OK NG
09-4
B03-I216E Bel-Cypher Pro
Beams in T.M.J. 7-1) Are all positioning beams turned on by depressing forward / back-
7)
ward switch of Focus Beam?
7-2) Dose Forward / backward operation of Focus Beam on ARM work
by depressing forward / backward switch?
Are all positioning beams turned off automatically when forward / back-
8)
ward switch of Focus Beam are not depressed for 2 minute?
9) Check in Panorama mode 9-1) Does equipment work normally with exposure op-
eration at 0kV, 0mA, at Panoramic mode?
Allophone Vibration
10) Check in T.M.J. 10-1) Does equipment work normally with exposure op­eration at 0kV, 0mA, at TMJ mode?
Allophone Vibration
10-2) After the first exposure, Does Rotation Arm Unit return to start
position automatically.
10-3) After the second exposure, Does Rotation Arm Unit stop at the
end position.
5. Externals
1) Are there scratches or cracks?
2) Are all covers secured with screws?
3) Check in the Save Image dialog.
3-1)In case of “PresetButtonEnabled=1” in
BEL_CYPHER_PRO_Save.ini, Preset buttons are displayed.
3-2)Image becomes dark when you push “Dark” button.
3-3)Image becomes normal when you push “Normal” button.
3-4)Image become bright when you push “Bright” button.
3-5)In case of “PresetButtonEnabled=0” in
BEL_CYPHER_RRO_Save.ini,Preset buttons are NOT displayed.
OK NG
OK NG
OK NG
OK NG OK NG
OK NG OK NG
OK NG
OK NG
OK NG
OK NG
OK NG
OK NG
OK NG
OK NG
OK NG
09-5
B03-I216E Bel-Cypher Pro
10. Contacting information
Belmont Equipment
A Division of TAKARA BELMONT USA, Inc. 101 Belmont Drive Somerset, NJ 08873 Toll Free (800) 223-1192 Toll Free Fax (800) 280-7504
www.belmontequip.com
TAKARA COMPANY, CANADA, LTD.
2706 South Sheridan Way Mississauga, Ontario, Canada L5J 2M4 Toll Free (800) 268-5351 Fax (905) 822-6203
www.takarabelmont.ca
10-1
B03-I216E Bel-Cypher Pro
11. Revision data of this manual
This manual was created on Sept 2016.
Revised in March 2017
Document number B03-I216E Bel-Cypher Pro
11-1
Loading...