The material contained in this manual, and in the online help for the software used to support this instrument, is believed adequate for the intended use of the instrument. If the instrument or procedures are used
for purposes other than those specified herein, confirmation of their suitability must be obtained from TA
Instruments. Otherwise, TA Instruments does not guarantee any results and assumes no obligation or liability. TA Instruments also reserves the right to revise this document and to make changes without notice.
TA Instruments may have patents, patent applications, trademarks, copyrights, or other intellectual property covering subject matter in this document. Except as expressly provided in written license agreement
from TA Instruments, the furnishing of this document does not give you any license to these patents, trademarks, copyrights, or other intellectual property.
TA Instruments Operating Software, as well as Module, Data Analysis, and Utility Software and their associated manuals and online help, are proprietary and copyrighted by TA Instruments. Purchasers are granted
a license to use these software programs on the module and controller with which they were purchased.
These programs may not be duplicated by the purchaser without the prior written consent of TA Instruments. Each licensed program shall remain the exclusive property of TA Instruments, and no rights or
licenses are granted to the purchaser other than as specified above.
TA Instruments can accept no liability for loss or damage, however caused, arising from the faulty or
incorrect use of its products.TA Instruments shall not be liable for any damages caused by interactions
between exogenous materials (e.g. chemicals) and parts of the instrument. This includes interactions of
gaseous, liquid or solid materials with, for instance, ampoule surfaces and/or parts of the instrument. It also
includes gases or vapors leaking from ampoules (e.g. originating from chemical reactions producing
gaseous substances), with subsequent cause of damage to the instrument.
This manual uses NOTES, CAUTIONS, and WARNINGS to emphasize important and critical
instructions. In the body of the manual these may be found in the shaded box on the outside of the page.
NOTE: A NOTE highlights important information about equipment or procedures.
CAUTION: A CAUTION emphasizes a procedure that may damage equipment or cause loss of data if not
followed correctly.
A WARNING indicates a procedure that may be hazardous to the operator or to the
environment if not followed correctly.
Regulatory Compliance
Safety Standards
EMC Directive
This instrument has been tested to meet the European Electromagnetic Compatibility Directive (EMC
Directive, 2004/108/EC). The Declaration of Conformity for your instrument lists the specific standards to
which the unit was tested.
The instrument was designed specifically as a test and measuring device. Compliance to the EMC directive
is through IEC 61326-1 Electrical equipment for measurement, control and laboratory use - EMC requirements (1998).
As noted in the IEC 61326-1, the instrument can have varying configurations. Emissions may, in non-typical applications, exceed the levels required by the standard. It is not practical to test all configurations, as
the manufacturer has no control over the user application of the instrument.
Immunity Testing
The instrument was tested to the requirements for laboratory locations.
Emission Testing
The instrument fulfills the limit requirements for Class A equipment but does not fulfill the limit
requirements for Class B equipment. The instrument was not designated to be used in domestic
establishments.
Page 4Nano ITC Getting Started Guide
Low Voltage Directive (Safety)
In order to comply with the European Low Voltage Directive (2006/95/EC), this equipment has been
designed to meet IEC 1010-1 (EN 61010-1) standards. To comply with requirements in the USA, this
instrument has been tested to the requirements of UL61010a-1.
WARNING: If this instrument is used in a manner not intended or specified in this manual, the
protection provided by the instrument may be impaired.
DANGER: High voltages are present in this instrument. Maintenance and repair of
internal parts must be performed only by TA Instruments or other qualified service
personnel.
Safety
High voltages are present in this instrument. Maintenance and repair of internal parts must be performed
only by TA Instruments or other qualified service personnel.
Electrical Safety
You must unplug the instrument before doing any maintenance or repair work; voltages as high as
125/250 VAC are present in this system.
Lifting the Instrument
The Nano ITC is not a portable instrument. In order to avoid injury, particularly to the back, please follow
this advice:
WARNING: Use appropriate care when unpacking or moving the instrument. It may be too
heavy for some individuals working alone to handle safely.
Index ................................................................................................................................................... 43
Buret Position Functions ........................................................................................... 41
Page 8Nano ITC Getting Started Guide
Chapter: 1
Introducing the Nano ITC
Overview
There are three ways in which a calorimeter may be designed. Heat measurements may be based on the
following:
•A temperature rise measured in a system of known heat capacity, (ΔT)
•The measured change in power (typically resistance heating) required to maintain a system at a
constant temperature (power compensation), and
•A direct measure of the heat flowing between the system and large heat sink maintained at a constant
temperature (heat flow)
Each method (ΔT, power compensation, and heat flow) has its advantages and disadvantages. The TA
Instruments Nano Isothermal Titration Calorimeter (ITC) uses a differential power compensation design
for maximum sensitivity and responsiveness.
The Nano ITC is available in three configurations: the Standard Volume model with 1-mL measurement
®
cells made of either 24K gold or Hastelloy
Standard Volume ITCs shipped before September 2009 were labeled “Nano ITC
incorporates second generation technology featuring enhanced baseline stability and increased sensitivity.
alloy, and the Low Volume model with 190-µL gold cells.
2G
”. The Nano ITC
Instrument Models Covered in this Guide
This guide covers Nano ITC instrument models 5300, 5301, and 5302.
Nano ITC Getting Started GuidePage 9
The Nano Isothermal Titration Calorimeter (ITC)
Buret assembly
The Nano ITC (shown in the figure below) consists of the measuring unit (calorimeter block and two nonremovable reaction vessels), the buret assembly, which includes the stirring system, and a cleaning accessory. With the exception of the power on/off switch located on the back of the calorimeter unit, all functions of the Nano ITC are controlled remotely by the computer through the USB connection.
2G
Figure 1 Nano ITC
.
Applications
Batch/Incremental Titration
In incremental or continuous titration, one of the reactants is placed in a syringe or buret external to the
reaction vessel. If individual, repeated injections are made, incremental titration takes place (as seen in the
example below); if only one injection is made, it is continuous injection calorimetry.
Figure 2 Incremental titrations.
Page 10Nano ITC Getting Started Guide
The baseline data, i.e. heat flow in the regions before and after each titrant pulse, shows the power required
A
B
to maintain a zero temperature difference between the sample and reference cells.
The baseline in this region is a function of heating by stirring. The baseline is used to calculate the area or
the heat from each pulse in the reaction vessel during the titration or batch reaction. The thermogram
constructed from the integrated peak areas is then used for data analysis.
Titration/Data Analysis
A single titration calorimetric experiment yields heat data as a function of the ratio of the concentrations of
the reactants. Titration data, in the form of heat change versus volume of titrant added, can be examined for
both analytical (thermometric titrimetry) and thermodynamic (titration calorimetry) information.
Other corrections must be made to the heat data to account for heat effects associated with titrant dilution
and any temperature difference between titrant and titrate solutions. These corrections are most easily
accomplished by performing a blank titration experiment and subtracting the blank heat data from the
experimental thermogram.
In the case of quantitative reaction of added titrant, the analysis of the thermogram is quite simple. All
peak areas will be the same (with the possible exception of the last peak) and ΔH calculated from the incremental heat and the number of moles of titrant added per increment. The titrant concentration is calculated
from the total heat divided by the ΔH for the reaction.
Calculation of Equilibrium Constants
The equilibrium constant for a given reaction may be simultaneously determined with the enthalpy change,
if the magnitudes of K and ΔH for the overall reaction taking place in the calorimeter are within certain
limits. The family of curves presented in the figure below shows that increased overall curvature of the
thermogram is generated with decreasing values of the association constant, K
eq
.
Figure A below shows the effects of varying magnitudes of the enthalpy change ΔH. Figure B shows the
effects of varying the equilibrium constant K.
Figure 3 Calculation of equilibrium constants.
Nano ITC Getting Started GuidePage 11
System Components
TED-controlled
block
Thermal
shield
Reference
cell
Sample cell
Control heater
DT
•Nano Isothermal Titration Calorimeter
•Personal computer (optionally available from TA Instruments)
•ITCRun and NanoAnalyze software
•Power cord
•Getting Started Guide (this manual)
•Data Collection and Analysis Software
•1 each 2.5-mL filling syringe with 16-gauge, 8-inch long needle (Standard Volume ITC)
•1 each 0.5-mL filling syringe with 16-gauge, 8-inch long needle (Low Volume ITC)
•1 each 100-µL and 250-µL syringes (with Nano ITC Standard Volume)
•1 each 50-µL syringe (with Nano ITC Low Volume)
•1 each buret drive
•USB cable
The components that make up the Nano ITC system are briefly described in the following sections.
Measuring Unit
The measuring unit includes the calorimeter block and two non-removable reaction vessels (sample and
reference cells). Access tubes extend downward from inside the buret mounting cavity on the top of the
calorimeter. The access tubes serve as conduits for the filling syringe, titrant delivery, and reference needle.
They also provide for titrant equilibration and as a thermal barrier to the environment outside the
calorimeter.
Figure 4 Nano ITC measuring unit.
The Nano ITC utilizes a differential power compensation design. Semiconducting thermoelectric devices
(TED) are used for temperature control and to detect temperature differences between the sample and
Page 12Nano ITC Getting Started Guide
reference cells. A proportional/integral/derivative (PID) control loop uses a control heater on the sample
Platinum
Access Tube
Teflon Bushings on
Needle to Dampen
Stirring Noise
24K Gold
Reaction Vessel
Titrant Delivery
Needle
Stir Paddle
Titrant Exit Point
cell to maintain a zero temperature difference between the sample and reference cells. The power required
to maintain this zero difference is used as the calorimeter signal and is monitored as a function of time. If a
reaction that produces heat occurs in the sample cell, the heat required to maintain the zero difference
decreases by the amount of heat supplied by the reaction, resulting in a peak in the thermogram.
A calibration heater located on the outside of the sample cell is used to provide precisely controlled heat
pulses for electrical calibrations, and to verify instrument performance.
The entire measuring unit is encased within an insulated air-tight canister which has been purged on a vacuum pump and filled with dry nitrogen at the factory. This is to prevent possible condensation and
evaporation of moisture around the unit which would create excessive baseline noise.
CAUTION: The purge port valve on the back of the Nano ITC should remain in the closed position at all times to maintain the integrity of the nitrogen purge.
NOTE: Purging of the canister is not a routine maintenance operation; contact TA Instruments before
proceeding.
Reaction Vessel
The calorimeter uses two matched reaction vessels with options of 1-mL gold, 190-µL gold, or
®
1-mL Hastelloy
match as closely as possible the thermal properties of the sample cell. Accordingly, a reference needle is
placed inside the reference cell during operation to correspond to the titrant needle in the sample cell.
. The vessels are accessed through platinum tubes. The reference cell is constructed to
Figure 5 Sample cell assembly.
CAUTION: Extreme care should be taken not to bend the syringe needle, because this would
impair proper stirring and possibly damage the reaction vessel.
Nano ITC Getting Started GuidePage 13
Syringe/Stirrer
Plunger position
indicator
Buret handle
Syringe plunger
Rotating buret
shaft
Graduated
syringe barrel
Knurled syringe
mounting knob
Reference cell
with reference
needle
Sample cell
with syringe
needle/stirrer
Nano ITC Standard Volume systems include two syringes of 100 and 250 µL capacities. Two buret
syringes are provided with 100 µL and 250 µL capacities. The only difference in dimension between the
two is the inner diameter of the syringe barrel; the needles are identical in order to maintain the thermal
and mechanical properties.
The Nano ITC Low Volume system includes one 50-µL syringe. This syringe uses a shorter needle. To
avoid possible damage, do not use the larger syringes with the Low Volume ITC. If you feel a stiff resistance or if the buret handle does not readily slip fully into place in the ITC, do not force it. Remove the
buret and check to see if the correct size syringe is installed. You can verify the feel of the normal
resistance by installing a buret with no syringe in place.
The titrant syringe needle also functions as the stirrer and extends down into the reaction vessel from the
top when the buret is mounted. The needle is balanced for optimum stirring efficiency. It has two Teflon
bushings to help dampen stirring noise and ensure that the needle spins true within the cell access tube (see
the figure below).
Figure 6 Orientation of buret, syringe, needles, and cells during experiments.
CAUTION: To avoid possible damage, do not use the 100 or 250 µL syringes with the Nano ITC
Low Volume instrument. Syringes are shipped with warning labels which may, if desired, be
affixed to the ITC by the customer to serve as a reminder.
Page 14Nano ITC Getting Started Guide
Loading...
+ 32 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.