Taber 5135, 5155 Operating Instructions Manual

Taber® Rotary Platform
Abrasion Tester
Model 5135 / 5155
Operating Instructions
TABLE OF CONTENTS
WARRANTY...............................................................2
CLAIMS FOR SHORTAGES......................................2
CLAIMS FOR DAMAGES ..........................................2
ICONS........................................................................2
CONTENTS ...............................................................2
INTRODUCTION........................................................3
PRODUCT DESCRIPTION........................................3
PRODUCT FEATURES .............................................3
INSTRUMENT SET-UP .............................................6
SETTING THE TEST PARAMETERS........................8
Setting the Speed 8 Setting the Reface Cycle Default 8 Setting the Counter Display 9 Setting the Test Cycles 9 Setting the Vacuum Level 10 Vacuum Only 10 Vacuum Nozzle Clearance 11 Accessory Receptacle 11
SPECIMEN SIZE .....................................................12
SPECIMEN MOUNTING OPTIONS.........................12
ADJUSTING FOR OPTIMAL PERFORMANCE.......19
Selecting Abradant 19 Mounting the Abrading Wheels 19 Wheel Loading (Auxiliary Weights) 20 Lowering the Abrading Heads 20 Adjusting the Height of the Vacuum Pickup Nozzle 21 Setting the Vacuum Level 21 Modified Vacuum Pick-up Nozzle 21
CHARACTERISTICS OF ABRADING WHEELS......22
Calibrase Wheels 22 Calibrade Wheels 23 Specialty Wheels / Abradants 23
REFACING ABRADING WHEELS...........................24
Refacing Resilient Wheels (Calibrase) 24
Refacing Vitrified Wheels (Calibrade) 26
SIGNIFICANCE AND USE.......................................28
TESTING PROCEDURES .......................................28
METHOD OF EVALUATION ....................................29
SPECIMEN PREPARATION....................................31
TEST PROCEDURES..............................................33
INFLUENCES ON RESULTS...................................40
MAINTENANCE .......................................................40
CALIBRATION / FACTORY SERVICE ....................43
OPTIONAL ACCESSORY INSTRUMENTS.............44
Sample Cutter 5000 44 Wheel Refacer 250 44 Quiet Cabinet 44 Grit Feeder Attachment 155 / 255 45 Multi-Media Abraser 5500 45
Scuffing Head Attachment 45 Haze Kit 46 Arm Height Extension Kit 46 Calibration Verification Kit 46
TEST METHOD REFERENCES ..............................47
CONSUMABLES / ACCESSORIES.........................48
Genuine Taber Wheel Sets (sold in pairs) 48 Wheel Refacing 49 Specimen Mounting 49 Test Accessories 50 Miscellaneous Accessories / Instruments 50
DATA SHEET...........................................................51
1 5135 / 5155 Operating Instructions ver 1.1
WARRANTY
ICONS
Any product or part which, under normal operating conditions in the facility of the original purchaser, proves defective in material or workmanship within one year from the date of shipment by the company, as determined by an inspection by Taber
®
Industries, will be repaired or replaced free of charge provided you promptly return the defective material. All claims must be sent to Taber Industries with transportation charges prepaid, notice of the defect and verification the product has been properly installed, maintained and operated within the limits of rated and normal usage.
Replacement parts will be shipped FOB our plant. The terms of this warranty do not in any way extend to any product or part thereof that has a life, under normal usage, inherently shorter than the one-year indicated above. Said warranty in respect of replacement of defective parts and any such additional warranty or representation expressly made, are in lieu of all other warranties expressed or implied, including any implied warranty of merchantability, or fitness for any particular purpose.
Taber Industries reserves the right to make changes without notice at any time in colors, material, specifications, and models, and to discontinue models.
CLAIMS FOR SHORTAGES
We use extreme care in selection, checking, and packing to eliminate the possibility of error. If a shipping error is discovered:
1. Carefully examine the packing materials and ensure nothing was inadvertently overlooked when the shipment was unpacked.
2. Notify the company you purchased the product from and immediately report the shortage. The materials are packed at the factory, and should be complete if the box has not been opened.
This instruction manual contains several notes and warnings that should be observed carefully by the user. The following icons denote these notes and warnings:
Indicates a NOTE that warrants careful
attention. These notes may detail a step in the procedure or point out a unique feature of the instrument. Please read and follow all notes carefully.
U Indicates a WARNING that warrants careful
attention. These warnings inform the user of any dangers that may cause injury to the operator and/or damage to the instrument. It is imperative that you read and follow all warnings carefully.
CONTENTS
Contents of the shipping container should include the following:
Auxiliary weights (500 gram load)
Auxiliary weights (1000 gram load)
E-100-125 Specimen Holder
E-100-101 Clamp Ring for Specimen
Holder*
Hex Wrench for Clamp Ring
Genuine Taber Calibrase
Genuine Taber Calbrade
S-11 Abrasive Discs (100/pkg)
S-36 Specimen Mounting Card (10 ea)
S-12 Long-Handled Hand Brush
Vacuum Unit with clean-up brush,
suction hose and stainless steel elbow
Power Cords (115V and 230V)
Operating Instructions
* Mounted on the Abraser
®
CS-10
®
H-18
Model
5135
1 set 2 sets 1 set 2 sets
1 2
1 2
1 1 1 pr. 2 pr. 1 pr. 2 pr.
1 pkg. 1 pkg. 1 pkg. 1 pkg.
1 1
1 1
1 1
1 1
Model
5155
3. Claims should be filed within 30 days from shipment.
CLAIMS FOR DAMAGES
Claims for loss or damage in transit should be made promptly and directly to the transportation company.
2 5135 / 5155 Operating Instructions ver 1.1
SECTION I
Understanding the Taber Abraser
INTRODUCTION
The Taber Rotary Platform Abraser Model 5135 and 5155 are durable, precision built test instruments designed to evaluate the resistance of surfaces to rubbing abrasion. Their field of application is varied and includes tests of solid materials, coated surfaces (paint, lacquer, electroplated), plastics, textiles (ranging from sheer silks to heavy upholstery), metals, leather, rubber, linoleum, plus many others. In the hands of a competent research technician, the Abraser is capable of performing reproducible tests, accurate within the variations of quality inherent in the material itself.
axis of the sample. Figure 3 diagrams the
relative positions of the sample and abrading wheels and the direction of their rotation.
One abrading wheel rubs the specimen outward toward the periphery and the other, inward toward the center. The resulting abrasion marks
(Figure 4) form a pattern of crossed arcs over
an area of approximately 30 cm satisfactory for rating most materials.
2
(4.65 inch
2
),
PRODUCT DESCRIPTION
Characteristic rub-wear action of the Abraser is
produced by the contact of a test sample turning
on a vertical axis, against the sliding rotation of
two abrading wheels. The wheels are driven by
Figure 3
the sample in opposite directions about a horizontal axis displaced tangentially from the
Figure 4
An exclusive and important feature of the 5135 and 5155 Abrasers is that the wheels traverse a complete circle on the specimen surface, revealing abrasion resistance at all angles relative to the weave or grain of the material.
PRODUCT FEATURES
The Model 5135 and 5155 Abrasers have many improved features compared to earlier versions of the Taber Rotary Platform Abraser. These are identified with an asterisk (*).
Operator Interface / Display*
The operator interface is an easy to use membrane panel with tactile feel buttons and a 4-line digital display. Simple on-screen instructions allow the operator to change the test parameters and are accessed via the MENU
button. The following test parameters are shown on the screen: test mode, test duration, test cycles (completed or remaining), turntable speed, and vacuum suction level (%).
60 / 72 RPM Turntable Rotation Speed*
Models 5135 and 5155 offer both 60 rpm and 72 rpm turntable rotation speed. This enables the operator to select the speed referenced in the test standard.
NOTE: The turntable speed on previous versions of
the Taber Abraser was dependent on the electrical line frequency. At 60Hz, the turntable speed was 72 rpm and at 50Hz the turntable speed was 60 rpm.
Preset Cycles Buttons*
The operator interface includes four commonly used preset cycles buttons. This enables the operator to automatically program 100, 500, 1000 or 2500 test cycles. The buttons may also be used to “add to” the cycle count by the corresponding incremental values.
Abrading Head Assemblies
Models 5135 and 5155 have independent abrading arms of forged aluminum alloy that pivot on ball bearings. The forward end of the
arm has a quick release wheel-mounting hub to which the abrading wheel is mounted. The arms are precision balanced to ensure exacting results. Each arm applies 250 grams force against the specimen with the wheel mounting assembly in place (exclusive of the mass of the wheels itself).
The outer end of the abrading head assembly allows the auxiliary weight to be mounted concentric with the abrading wheel. A stud on the rear end of the abrading arm may be used to carry a counterweight when it is desired to reduce the wheel load to less than 250 grams when testing delicate materials. It may also be used to hold an abrading wheel as a counterweight to compensate for the mass of the working wheel.
Quick Release Mounting Hub*
An expanding collet hub design with push-button operation permits quick wheel mounting, without the need of a locking nut. A spring loaded, beveled retaining nut provides a positive locking force on the hub retaining lip making certain the wheels remain securely fastened until disengaged. The increased clearance results in a larger viewing area of the test specimen.
4 5135 / 5155 Operating Instructions ver 1.1
Auxiliary Weights
Model 5135 has two pairs of precision stainless steel weights of 250 grams and 750 grams furnished with the testing set. In addition to the mass of the abrading arm itself (250 grams) they provide standard wheel loads of 500 and 1000
grams. Figures 6 – 9 show abrasers with
different loads applied (per wheel).
Available counterweights reduce the 250 gram
load to 125 grams or 75 grams (Figure 9).
Counterweights may be used in conjunction with the standard weights to further expand the range of wheel loads. To offset the weight of the working wheels, wheels may be placed on the counterweight-mounting studs.
NOTE: For simplicity, each weight is marked with the
total load that will be exerted on the wheel.
Figure 9
Vacuum System
Models 5135 and 5155 include a vacuum unit with a heavy-duty motor and life-lubricated sealed bearings. With the vacuum power connection plugged into the rear of the abraser housing, the vacuum will be automatically turned on when a test is commenced and shut off at the end of the test cycles. A flexible rubber suction hose connects it to the Abraser.
A vacuum pick-up nozzle is hinged to an adjustable mount at the rear of the housing. The height above the specimen is set by raising or lowering the precision vacuum adjustment.
Figure 6
Figure 7
Precision Vacuum Nozzle Adjustment*
The height of the vacuum pickup nozzle is adjusted by turning a knob. Each turn represents 1.27 mm (0.05 inch) height adjustment, making it significantly easier to set the proper distance between the nozzle and specimen surface. Located on the top of the instrument next to the vacuum nozzle, the knob is easily accessible.
Specimen Holder
The Model E100-125 specimen holder accommodates most materials less than 6.35
(0.25 inch) thick with a 6.35 mm (0.25 inch)
mm center hole. Rigid, flat specimens are secured using the clamp plate and nut. For flexible specimens, the E100-101 clamp ring is provided in conjunction with the clamp plate and nut. An S-19 rubber pad prevents specimens from slipping during the test.
Specimen Holder Drive Shaft
The drive shaft for the specimen holder projects vertically through the motor well in the Abraser housing. Interchangeable specimen holders are mounted to this shaft.
Figure 8
5 5135 / 5155 Operating Instructions ver 1.1
Accessory Receptacles
The Model 5135 and 5155 Abraser permit control of the vacuum unit and an accessory instrument (wheel refacer or grit feeder) from the main unit. The electrical receptacles are integrated in the rear of the housing and marked
accordingly (Figure 10). The use of the
electrical receptacles is restricted to the specified equipment.
Figure 10
NOTE: The power for the vacuum receptacle is
dependent on the voltage selector switch (115/230V). The power for the accessory receptacle is 115V.
INSTRUMENT SET-UP
Installing Fuses and Setting the Voltage
1. Prior to operating the instrument, the correct fuses must be inserted into the fuse holder (found at the rear of the
instrument). See Figure 11.
115V, insert two (2) T10A, 5 x 20mm fuses [Taber p/n 129735]
230V, insert two (2) T5A, 5 x 20mm fuses [Taber p/n 128561]
NOTE: The factory default is set for 230V operation.
Aluminum Housing
The Abrasers are built with a rugged, compact aluminum housing. This provides sufficient weight and rigidity to ensure the instrument is stable during testing. The overall dimensions of the main units are:
Model 5135 27.9 cm x 40 cm x 25.4 cm
Model 5135 49.5 cm x 35.6 cm x 25.4 cm
(11” x 15.75” x 10”)
(19.5" x 14” x 10”)
Dual Voltage*
The 5135 and 5155 are dual voltage 115/230V, 60/50Hz switchable.
Abrading Wheels
The Taber Abraser is supplied with two types of Genuine Taber abrading wheels, CS-10 and H-
18. For additional details on other available wheels that may be used with this instrument, see Characteristics of Abrading Wheels (page
21).
Auxiliary Port 1*
This feature is not currently used, but is being supplied for future upgrades.
Figure 11
2. Slide the red voltage selector switch to the appropriate voltage setting that the instrument will be operated at.
NOTE: To remove the fuse holder, slide it out of the
switch selector. A small screwdriver may be used to assist in the removal.
U WARNING: Prior to powering-up, ensure the
correct voltage is indicated per the detailed instructions. Failure to set the correct voltage configuration may result in damage to the Rotary Platform Abrasion Tester’s electronic components and will void any warranty.
* Denotes an improved feature
6 5135 / 5155 Operating Instructions ver 1.1
General Set-Up
Figure 12
1. The Abraser unit should be placed on a flat, level and rigid surface.
your convenience. Discard the power cord not used.
5. Turn the instrument on. The ON / OFF power switch is located on the back of the instrument immediately above the cord. The instrument is now ready to operate.
U WARNING: Failure to connect the Rotary
Platform Abrasion Tester to a surge protector or surge suppressor may result in damage to the instrument’s electronic components.
2. Position the vacuum unit in a ventilated area near the Abraser. Remove the motor housing and install the vacuum cartridge filter and filter bag.
3. Insert the stainless steel vacuum hose
elbow into the vacuum hose receptacle found in the rear of the Abraser housing. Connect the flexible rubber suction hose to the vacuum unit and the elbow.
U WARNING: DO NOT place the vacuum in a closed
cabinet without adequate ventilation. Failure to provide sufficient airflow may cause the motor to overheat. Resulting in damage to the vacuum.
Powering-Up
1. Plug the female connector of the Abraser line cord into the rear of the instrument.
2. Attach the male connector of the vacuum power cord into the receptacle marked “VACUUM” at the rear of the Abraser. When operating at 115V, a patch cord is used to connect the vacuum unit to the Abraser.
3. If any accessories are to be used (Grit Feeder or Wheel Refacer), connect the male connector to the receptacle marked “ACCESSORY” at the rear of the Abraser.
4. Connect the instrument to a 115 or 230 volt, 60 or 50-cycle circuit. Two power cords (115V and 230V) are provided for
7 5135 / 5155 Operating Instructions ver 1.1
IMPORTANT
READ PRIOR TO USING INSTRUMENT
To activate the buttons on the membrane switch, press firmly for approximately one (1) second.
To access the MENU features or change the MODE, the turntable must be stopped (the Run or Reface Mode s atus indicated on the t digital display must display “Stopped”).
1-Speed 4-Vacuum
2-Refacing 5-Outlet 3-Display 6-Reset >Select Menu Option
1-Speed 4-Vacuum 2-Refacing 5-Outlet 3-Display 6-Reset >0 Zero key for more
Alternating
Display
Screens
SETTING THE TEST PARAMETERS
Model 5155 Dual Abraser – The numerical
keypad is common to the entire instrument. The digital display shows test cycles specific to each turntable, and common functions to the entire instrument (turntable speed and vacuum suction level).
To change between the left or right turntables, press either the START or STOP buttons on the appropriate table. The last key depressed will determine the side to which the data will be entered, and is indicated by the word “KEYPAD” displayed on the lower corners of the digital display. For example, if the STOP key has been depressed on the left side, any information entered into the keypad will be for the left table. Should the START or STOP key on the right side be depressed, any values that are entered will be for the right turntable.
Setting the Speed
Current Speed = * *1- 60 Cycles/Minute 2- 72 Cycles/Minute >Press 1 or 2
Current Speed = * *1- 60 Cycles/Minute 2- 72 Cycles/Minute >CLEAR key to return
Alternating
Display
Screens
Speed set to
72 Cycles/Minute
The speed of the Rotary Platform Abrasion Tester is reported as CYCLES / MINUTE, where one cycle is defined as a full rotation of the turntable.
 NOTE: The factory default speed is set at 60 RPM.
The speed of the Rotary Platform Abrasion Tester can be selected as 60 or 72 rpm. The tester will store the speed that was last selected in memory and use this as a default until it is changed. Speed is shown on the display.
The following instructions show how to change the speed. To change speed, the instrument must not be in operation (the Run or Reface Mode status indicated on the digital display must display “Stopped”).
1. Press the MENU button to display menu
options (see below for screen displays).
2. From the menu options, press 1 to select the option to change speed.
3. Using the numeric keypad, enter “1” to change to 60 cycles per minute or “2” to enter 72 cycles per minute. The asterisk (*) indicates the current speed selection.
4. Press START to begin testing.
8 5135 / 5155 Operating Instructions ver 1.1
NOTE: After pressing “1” or “2” in step 3, the display
screen will flash “Test speed set to ‘X’ Cycles/Minute”
to confirm your selection and will automatically return to the main display screen.
NOTE: The turntables for the Model 5155 Dual
Abraser will rotate at the same rpm.
Setting the Reface Cycle Default
The Reface Mode allows you to reface the wheels without adding test cycles to the completed (remaining) test cycle count. This convenient feature is ideal for any test that requires the abrading wheels to be refaced at intervals during the test.
The instrument is programmed with a default of 25 cycles. To change this value, follow the instructions below.
1. Press the MENU button to display Menu
Options (see below for screen displays).
2. From the menu options, press 2 to select Refacing.
3. Enter a new value, and press ENTER to save.
3. Using the numeric keypad, enter “1” to display Completed Cycles or “2” for Remaining Cycles. The asterisk (*) indicates the current counter display.
1-Speed 4-Vacuum
2-Refacing 5-Outlet
3-Display 6-Reset >Select Menu Option
Change refacing default cycles: 25 >Enter new total >CLEAR key to return
Change refacing default cycles: 50 >Enter new total >ENTER key to save
Change refacing default cycles: 50 >Enter new total >CLEAR key to cancel
Alternating
Display
Screens
Default number of
Reface Cycles
set to
50 cycles
NOTE: After entering the new value in step 3 and
pressing ENTER, the display screen will flash “Default number of Reface Cycles set to X cycles” to confirm
your selection and will automatically return to the main display screen.
NOTE: If a new value is entered with the numeric
keypad or a Preset Cycles button when the Abraser is in the Reface Mode, the number of reface cycles will change. However, this DOES NOT change the default refacing cycle value.
U WARNING: Do not exceed 50 cycles on an S-11
refacing disc. If more than 50 cycles are required, replace the refacing disc.
Setting the Counter Display
The Rotary Platform Abrasion tester can be programmed to show “Completed Cycles” (count up) or “Remaining Cycles” (count down). The following explains how to change the display:
1. Press the MENU button to display Menu
Options (see below for screen displays).
2. From the menu options, press 3 to select Display.
1-Speed 4-Vacuum 2-Refacing 5-Outlet
3-Display 6-Reset
>Select Menu Option
Current Count Mode=* *1- Completed Cycles 2- Remaining Cycles >Press 1 or 2
Change Count Mode=* *1- Completed Cycles 2- Remaining Cycles >CLEAR key to return
Alternating
Display
Screens
Counting mode set
to display
Remaining Cycles
NOTE: After entering the new value in step 3, the
display screen will flash “Counting mode set to display Remaining (Completed) Cycles” to confirm your
selection and will automatically return to the main display screen.
Setting the Test Cycles
The Abraser can be preset to any number of test
cycles below the maximum value of 50,000.
The total cycle count shown on the display screen defaults to the test duration value that was last entered.
To reset the test cycles, depress the CLEAR key, and then select the desired number of test cycles using one of the two options discussed below. The Abraser turntable will stop automatically at the selected number of test cycles. To begin testing a new specimen, press the CLEAR key. It is not necessary to re-enter the same value of test cycles at the beginning of each test.
Option 1
new value may be entered with the numeric keypad. After the desired number of test cycles has been entered, press the ENTER or START key. If an incorrect value is entered, press the CLEAR key to return to the original value.
Option 2
you conveniently enter values of 100, 500,
: With the Abraser in Run Mode, a
: The PRESET CYCLES buttons let
9 5135 / 5155 Operating Instructions ver 1.1
1000 or 2500. Press the 0 button to reset the cycle count, and then select a preset cycle button(s) to set the new test cycle. A PRESET CYCLES button may be pressed during operation or when a test is stopped, to increase total cycles by that amount.
It is possible to increase (or decrease) test duration beyond the original cycles. To override, the number of turntable revolutions must be a value greater than the completed cycles. Press ENTER to accept the new value.
For the model 5155 Abraser, both turntables can be preset to different values depending on the nature of the test. Prior to entering a new value for the test cycles, ensure the KEYPAD indicator is shown on the correct turntable. To change between the left or right turntables, press either the START or STOP buttons on the appropriate table. The ongoing cycle count for each turntable will be displayed in the corresponding digital readout.
NOTE: An abrasion cycle, or test cycles, is defined
as one complete revolution of the specimen in contact with the abrading wheels.
NOTE: The instrument can be stopped at any time for
inspection and started up again (by depressing the START key) without interrupting the counter reading in the digital display.
NOTE: Entering a new value DOES NOT
automatically zero the completed cycle count. This enables you to reprogram the total test cycles to continue testing beyond the original number of cycles. If you change total cycles to a value less than the completed cycles, you must first press CLEAR to zero the completed cycle count, enter the new value and then press START.
Setting the Vacuum Level
The vacuum furnished with the Model 5135 and 5155 Abrasion Test Sets is equipped with a variable vacuum control. For test control and standardization of vacuum suction, the vacuum level is calibrated to values of 50 – 100%. Changing the vacuum level adjusts the speed of the vacuum unit motor and the corresponding vacuum suction for accurate standardization and control of test procedures.
The vacuum level should be set high enough to lift the abraded particles, but not lift flexible samples. A vacuum level of 100% is programmed into the Abraser. If you require less than 100% vacuum, follow the instructions below. The vacuum level will default to the value that was last entered.
1. Press the MENU button to display Menu
Options (see below for screen displays).
2. From the menu options, press 4 to select Vacuum.
3. Enter a new value between 50 and 100, and press ENTER to save.
1-Speed 4-Vacuum 2-Refacing 5-Outlet 3-Display 6-Reset >Select Menu Option
Change Vacuum Level
>Enter new level >CLEAR key to return
Change Vacuum Level
>Enter new level >ENTER key to save
Change Vacuum Level
>Enter new level >CLEAR key to return
100%
90%
Alternating
Display
Screens
90%
Vacuum Level set to:
90%
NOTE: The factory default vacuum level is set at 100.
U WARNING: Do not use the vacuum pickup for
tests on wet materials. When conducting wet tests, raise the vacuum nozzle, and disconnect the electrical cord to the vacuum unit.
U WARNING: Do not place the v acuum in a closed
cabinet without adequate ventilation. Failure to provide sufficient airflow may cause the motor to overheat, resulting in damage to the vacuum.
Vacuum Only
The VACUUM ONLY key allows you to operate the vacuum unit without the Abraser. This may be necessary to clean the drive shaft and specimen holder cavity or to properly operate the Wheel Refacer.
If only the vacuum is needed, turn the Abraser on then depress the VACUUM ONLY key. The vacuum will operate at the vacuum level last entered.
10 5135 / 5155 Operating Instructions ver 1.1
Vacuum Nozzle Clearance
The height of the vacuum pick-up nozzle is adjustable and may be raised or lowered using
the knob found by the right abraser arm (Figure
13). Each full rotation of the adjustment knob
raises or lowers the nozzle by 1.27 mm
inch)
. To assist with vacuum nozzle adjustments, a calibrated shim may be placed between the test specimen and the nozzle.
(0.05
the start of the Grit Feeder at the same time a test is commenced.
1. Press the MENU button to display Menu
Options (see below for screen displays).
2. From the menu options, press 5 to select Outlet.
3. Using the numeric keypad, enter “1” for Continuous Power or “2” for Power While Testing. The asterisk (*) indicates the current mode.
Figure 13
Many established test methods define the gap between the test sample and vacuum pick-up nozzle. For most flexible materials, the vacuum nozzle should be no closer than 6.35 mm
. For rigid materials, the vacuum nozzle is
inch)
normally specified as a 1.59 mm
3.18 mm
The amount of vacuum suction is influenced by the type of material being tested and height of the nozzle above the specimen surface. The intent of the vacuum system is to remove the abraded particles BUT NOT LIFT THE TEST SPECIMEN.
(0.125 inch) gap.
(0.0625 inch) to
(0.25
U WARNING: When testing flexible materials,
ensure the suction from the vacuum nozzle does not lift the specimen. Any contact between the specimen and vacuum pick-up nozzle during the test, will cause additional wear that may influence the test results.
1-Speed 4-Vacuum 2-Refacing 5-Outlet 3-Display 6-Reset >Select Menu Option
Current Mode=* *1- Continuous Power 2- Power While Test
>Press 1 or 2
Current Mode=* *1- Continuous Power 2- Power While Test >CLEAR key to return
Aux. Outlet Mode
Power While Testing
Set To:
U
WARNING: The use of the electrical receptacles
is restricted to the specified equipment only.
Alternating
Display
Screens
Accessory Receptacle
Incorporated into the Model 5135 and 5155 Abraser is an accessory receptacle to supply power to either the Wheel Refacer or Grit Feeder Attachment.
Through the MENU, you can choose to have continuous power supplied to the receptacle or ONLY while you test. When operating the Wheel Refacer, select continuous power. This allows you to reface wheels at any time, provided the instrument is turned on. However when operating the Grit Feeder, voltage should be supplied ONLY when a test is being conducted. This will enable you to coordinate
11 5135 / 5155 Operating Instructions ver 1.1
SPECIMEN SIZE
The width of the wear path is 12.7 mm (0.5 inch), and located 31.75 mm of the specimen. The size of the specimen may vary depending on the material being evaluated. For most rigid materials, a sample approximately 100 mm
(4.0 inch) square is recommended.
Flexible specimens are typically circular and require the use of the clamp ring. If a mounting card is used, the specimen should be approximately 108 mm no mounting card is used, a 135 mm specimen is required such that the clamp ring will grip the overlapped material.
If the specimen is less than the suggested size, it is possible to position two or more specimens on a mounting card to meet the size requirement. If the edges are not flush together or if the samples are of different thickness, the edges may wear at a different rate than the rest of the specimen.
The standard material thickness that can be evaluated with the Taber Rotary Platform Abraser is 6.35 mm greater than 6.35 mm (0.25 inch) but less than
12.7 mm
(0.5 inch), the S-21 extension nut may
be used. Specimens with a thickness greater than 12.7 mm
(0.5 inch) can be tested using the
arm height extension kit.
(1.25 inch) from the center
(4.25 inch) in diameter. If
(5.25 inch)
(0.25 inch). For materials
Figure 14
Mounting Sheets (S-37 and S-37-1)
These 108 mm (4.25 inch) sheets are coated on both sides with pressure sensitive adhesive overlaid with protective paper that is peeled off just prior to use. Thinner than the S-36 and S­36-1, mounting sheets are used to provide a non-slip attachment of the specimen to the surface of the turntable and are especially useful for materials difficult to drill a center hole (such as glass or ceramics). Mounting sheets are available as square (S-37) or round (S-37-1), and have a test record printed on both sides enabling them to be used as a mounting card.
SPECIMEN MOUNTING OPTIONS
To expand the versatility of the Taber Abraser, mounting options and interchangeable specimen tables are available to evaluate materials not easily tested with the standard configuration. All tables are engineered with a ball plunger and drive pin spring clutch that locks the table onto the motor shaft. This prevents the table from shifting during a test, but allows for it to be easily removed for specimen inspection or to clean the motor shaft cavity. To gain access to the specimen holder, lift the suction nozzle on its hinge and raise the abrading arms up and back beyond their balance point.
Mounting Cards (S-36 and S-36-1)
These 108 mm (4.25 inch) rigid cards are coated on one side with pressure sensitive adhesive. Designed to hold textile and other flexible test specimens tightly to the mounting surface, they prevent wrinkling while undergoing testing. A test record form is printed on the backside of the card for permanent record. Mounting cards are available as square (S-36) or round (S-36-1).
Figure 15
Figure 16
12 5135 / 5155 Operating Instructions ver 1.1
Specimen Holder (E100-125)
The specimen table equipped with the Abraser (E100-125) has been designed to accommodate most materials less than 6.35 mm thick. Unless optional accessories are utilized, a
6.35 mm (0.25 inch) center hole must be punched or drilled in the specimen so it can be securely fastened to the threaded center post.
To secure rigid materials, use the clamp plate and nut. For flexible materials, also use the supplied E100-101 specimen clamp ring. A 108 mm
(4.25 inch) diameter rubber pad (S-19) is
provided to ensure a non-slip surface.
(0.25 inch)
Specimen Holder with Ring Clamp (E140-14)
Rigid specimens that are slightly warped and less than 6.35 mm either dry or wet tested using the E140-14 Ring Clamp Holder. Eight (8) screws are spaced evenly on the clamp plate, and when tightened, will correct for slightly warped conditions. This holder can be furnished with (E140-14-S) or without (E-140-14-NS) a center clamping screw. Specimens should be 137 mm diameter with a 6.35 mm
(0.25 inch) in thickness can be
(5.375 inch) in
(0.25 inch) center hole.
Figure 21
Figure 19
13 5135 / 5155 Operating Instructions ver 1.1
Figure 22
Textile Specimen Holder – Tensioning Type (E140-15)
Textile Holder – Tensioning Type for Wet or Dry Testing (E140-18)
Similar to the standard specimen holder, the E140-15 Tensioning Textile Specimen Holder has a slightly raised wear track. This gives textile and other flexible specimens additional tension when the clamp ring is drawn down over the edge of the holder and the clamp plate is tightened in the center recess. Requires a 134 mm (5.25 inch) diameter specimen with a 6.35 mm
(0.25 inch) center hole. This holder is not
recommended for use with rigid specimens or with mounting cards.
NOTE: For proper mounting, place the specimen on
the holder and draw it taut with the clamp ring. Tighten the ring, and then mount the clamp plate.
The E140-18 holder is designed to give an initial stretch to woven fabrics so that they may be tested while moist with minimum tendency to wrinkle. A raised wear track provides additional tension on the specimen, while a 9.53 mm
inch)
rim helps retain any liquid. Requires a
159mm
inch)
(6.25 inch) specimen with a 6.35 mm (0.25
center hole.
(0.375
NOTE: For proper mounting, place the specimen on
the holder, position the clamp ring and mark the holes. Where each hole was marked, make a small cut such that the screws may be aligned and mounted to the holder. Partially tighten down the clamp ring. Draw the fabric taut under the ring, tighten the screws, and trim off excess material around the edge. Tighten the clamp plate.
NOTE: When the fabric specimen is to be tested at
specific moisture content it must be conditioned to that content before it is mounted on the holder.
Figure 24
Figure 27
Figure 25
14 5135 / 5155 Operating Instructions ver 1.1
Figure 28
Drive Pin Type Holder (E140-19)
Threaded Ring Specimen Holder (E140-21)
The E140-19 Drive Pin Holder is intended for rigid, square specimens that do not have a center-mounting hole. The specimen is adhered to the holder with S-37 or S-37-1 mounting sheets and opposing driving pins prevent it from slipping while the test is being conducted.
NOTE: An example where this holder may be useful
is for glass or ceramic tile, which may be difficult to drill a center hole.
The E140-21 Threaded Ring Specimen Holder incorporates a clamp plate and flanged clamp ring to test slightly warped, rigid specimens. The ring is threaded to the body of the holder, which securely clamps the outer edges of specimens up to 137 mm
(5.375 inch) diameter.
Figure 30
Figure 31
Figure 33
Figure 34
15 5135 / 5155 Operating Instructions ver 1.1
Loading...
+ 37 hidden pages