SETTING THE TEST PARAMETERS........................8
Setting the Speed 8
Setting the Reface Cycle Default 8
Setting the Counter Display 9
Setting the Test Cycles 9
Setting the Vacuum Level 10
Vacuum Only 10
Vacuum Nozzle Clearance 11
Accessory Receptacle 11
DATA SHEET...........................................................51
1 5135 / 5155 Operating Instructions ver 1.1
WARRANTY
ICONS
Any product or part which, under normal
operating conditions in the facility of the original
purchaser, proves defective in material or
workmanship within one year from the date of
shipment by the company, as determined by an
inspection by Taber
®
Industries, will be repaired
or replaced free of charge provided you promptly
return the defective material. All claims must be
sent to Taber Industries with transportation
charges prepaid, notice of the defect and
verification the product has been properly
installed, maintained and operated within the
limits of rated and normal usage.
Replacement parts will be shipped FOB our
plant. The terms of this warranty do not in any
way extend to any product or part thereof that
has a life, under normal usage, inherently
shorter than the one-year indicated above. Said
warranty in respect of replacement of defective
parts and any such additional warranty or
representation expressly made, are in lieu of all
other warranties expressed or implied, including
any implied warranty of merchantability, or
fitness for any particular purpose.
Taber Industries reserves the right to make
changes without notice at any time in colors,
material, specifications, and models, and to
discontinue models.
CLAIMS FOR SHORTAGES
We use extreme care in selection, checking, and
packing to eliminate the possibility of error. If a
shipping error is discovered:
1. Carefully examine the packing materials and
ensure nothing was inadvertently overlooked
when the shipment was unpacked.
2. Notify the company you purchased the
product from and immediately report the
shortage. The materials are packed at the
factory, and should be complete if the box has
not been opened.
This instruction manual contains several notes
and warnings that should be observed carefully
by the user. The following icons denote these
notes and warnings:
Indicates a NOTE that warrants careful
attention. These notes may detail a step in
the procedure or point out a unique feature
of the instrument. Please read and follow
all notes carefully.
U Indicates a WARNING that warrants careful
attention. These warnings inform the user
of any dangers that may cause injury to the
operator and/or damage to the instrument.
It is imperative that you read and follow all
warnings carefully.
CONTENTS
Contents of the shipping container should
include the following:
• Auxiliary weights (500 gram load)
• Auxiliary weights (1000 gram load)
• E-100-125 Specimen Holder
• E-100-101 Clamp Ring for Specimen
Holder*
• Hex Wrench for Clamp Ring
• Genuine Taber Calibrase
• Genuine Taber Calbrade
• S-11 Abrasive Discs (100/pkg)
• S-36 Specimen Mounting Card (10 ea)
• S-12 Long-Handled Hand Brush
• Vacuum Unit with clean-up brush,
suction hose and stainless steel elbow
• Power Cords (115V and 230V)
• Operating Instructions
* Mounted on the Abraser
®
CS-10
®
H-18
Model
5135
1 set 2 sets
1 set 2 sets
1 2
1 2
1 1
1 pr. 2 pr.
1 pr. 2 pr.
1 pkg. 1 pkg.
1 pkg. 1 pkg.
1 1
1 1
1 1
1 1
Model
5155
3. Claims should be filed within 30 days from
shipment.
CLAIMS FOR DAMAGES
Claims for loss or damage in transit should be
made promptly and directly to the transportation
company.
2 5135 / 5155 Operating Instructions ver 1.1
SECTION I
Understanding the Taber Abraser
INTRODUCTION
The Taber Rotary Platform Abraser Model 5135
and 5155 are durable, precision built test
instruments designed to evaluate the resistance
of surfaces to rubbing abrasion. Their field of
application is varied and includes tests of solid
materials, coated surfaces (paint, lacquer,
electroplated), plastics, textiles (ranging from
sheer silks to heavy upholstery), metals, leather,
rubber, linoleum, plus many others. In the
hands of a competent research technician, the
Abraser is capable of performing reproducible
tests, accurate within the variations of quality
inherent in the material itself.
axis of the sample. Figure 3 diagrams the
relative positions of the sample and abrading
wheels and the direction of their rotation.
One abrading wheel rubs the specimen outward
toward the periphery and the other, inward
toward the center. The resulting abrasion marks
(Figure 4) form a pattern of crossed arcs over
an area of approximately 30 cm
satisfactory for rating most materials.
2
(4.65 inch
2
),
PRODUCT DESCRIPTION
Characteristic rub-wear action of the Abraser is
produced by the contact of a test sample turning
on a vertical axis, against the sliding rotation of
two abrading wheels. The wheels are driven by
Figure 3
the sample in opposite directions about a
horizontal axis displaced tangentially from the
Figure 4
An exclusive and important feature of the 5135
and 5155 Abrasers is that the wheels traverse a
complete circle on the specimen surface,
revealing abrasion resistance at all angles
relative to the weave or grain of the material.
PRODUCT FEATURES
The Model 5135 and 5155 Abrasers have many
improved features compared to earlier versions
of the Taber Rotary Platform Abraser. These
are identified with an asterisk (*).
Operator Interface / Display*
The operator interface is an easy to use
membrane panel with tactile feel buttons and a
4-line digital display. Simple on-screen
instructions allow the operator to change the test
parameters and are accessed via the MENU
button. The following test parameters are
shown on the screen: test mode, test duration,
test cycles (completed or remaining), turntable
speed, and vacuum suction level (%).
60 / 72 RPM Turntable Rotation Speed*
Models 5135 and 5155 offer both 60 rpm and 72
rpm turntable rotation speed. This enables the
operator to select the speed referenced in the
test standard.
NOTE:The turntable speed on previous versions of
the Taber Abraser was dependent on the electrical
line frequency. At 60Hz, the turntable speed was 72
rpm and at 50Hz the turntable speed was 60 rpm.
Preset Cycles Buttons*
The operator interface includes four commonly
used preset cycles buttons. This enables the
operator to automatically program 100, 500,
1000 or 2500 test cycles. The buttons may also
be used to “add to” the cycle count by the
corresponding incremental values.
Abrading Head Assemblies
Models 5135 and 5155 have independent
abrading arms of forged aluminum alloy that
pivot on ball bearings. The forward end of the
arm has a quick release wheel-mounting hub to
which the abrading wheel is mounted. The arms
are precision balanced to ensure exacting
results. Each arm applies 250 grams force
against the specimen with the wheel mounting
assembly in place (exclusive of the mass of the
wheels itself).
The outer end of the abrading head assembly
allows the auxiliary weight to be mounted
concentric with the abrading wheel. A stud on
the rear end of the abrading arm may be used to
carry a counterweight when it is desired to
reduce the wheel load to less than 250 grams
when testing delicate materials. It may also be
used to hold an abrading wheel as a
counterweight to compensate for the mass of
the working wheel.
Quick Release Mounting Hub*
An expanding collet hub design with push-button
operation permits quick wheel mounting, without
the need of a locking nut. A spring loaded,
beveled retaining nut provides a positive locking
force on the hub retaining lip making certain the
wheels remain securely fastened until
disengaged. The increased clearance results in
a larger viewing area of the test specimen.
4 5135 / 5155 Operating Instructions ver 1.1
Auxiliary Weights
Model 5135 has two pairs of precision stainless
steel weights of 250 grams and 750 grams
furnished with the testing set. In addition to the
mass of the abrading arm itself (250 grams) they
provide standard wheel loads of 500 and 1000
grams. Figures 6 – 9 show abrasers with
different loads applied (per wheel).
Available counterweights reduce the 250 gram
load to 125 grams or 75 grams (Figure 9).
Counterweights may be used in conjunction with
the standard weights to further expand the range
of wheel loads. To offset the weight of the
working wheels, wheels may be placed on the
counterweight-mounting studs.
NOTE:For simplicity, each weight is marked with the
total load that will be exerted on the wheel.
Figure 9
Vacuum System
Models 5135 and 5155 include a vacuum unit
with a heavy-duty motor and life-lubricated
sealed bearings. With the vacuum power
connection plugged into the rear of the abraser
housing, the vacuum will be automatically turned
on when a test is commenced and shut off at the
end of the test cycles. A flexible rubber suction
hose connects it to the Abraser.
A vacuum pick-up nozzle is hinged to an
adjustable mount at the rear of the housing. The
height above the specimen is set by raising or
lowering the precision vacuum adjustment.
Figure 6
Figure 7
Precision Vacuum Nozzle Adjustment*
The height of the vacuum pickup nozzle is
adjusted by turning a knob. Each turn
represents 1.27 mm (0.05 inch) height adjustment,
making it significantly easier to set the proper
distance between the nozzle and specimen
surface. Located on the top of the instrument
next to the vacuum nozzle, the knob is easily
accessible.
Specimen Holder
The Model E100-125 specimen holder
accommodates most materials less than 6.35
(0.25 inch) thick with a 6.35 mm (0.25 inch)
mm
center hole. Rigid, flat specimens are secured
using the clamp plate and nut. For flexible
specimens, the E100-101 clamp ring is provided
in conjunction with the clamp plate and nut. An
S-19 rubber pad prevents specimens from
slipping during the test.
Specimen Holder Drive Shaft
The drive shaft for the specimen holder projects
vertically through the motor well in the Abraser
housing. Interchangeable specimen holders are
mounted to this shaft.
Figure 8
5 5135 / 5155 Operating Instructions ver 1.1
Accessory Receptacles
The Model 5135 and 5155 Abraser permit
control of the vacuum unit and an accessory
instrument (wheel refacer or grit feeder) from the
main unit. The electrical receptacles are
integrated in the rear of the housing and marked
accordingly (Figure 10). The use of the
electrical receptacles is restricted to the
specified equipment.
Figure 10
NOTE:The power for the vacuum receptacle is
dependent on the voltage selector switch (115/230V).
The power for the accessory receptacle is 115V.
INSTRUMENT SET-UP
Installing Fuses and Setting the Voltage
1. Prior to operating the instrument, the
correct fuses must be inserted into the
fuse holder (found at the rear of the
instrument). See Figure 11.
115V, insert two (2) T10A, 5 x 20mm fuses
[Taber p/n 129735]
230V, insert two (2) T5A, 5 x 20mm fuses
[Taber p/n 128561]
NOTE:The factory default is set for 230V operation.
Aluminum Housing
The Abrasers are built with a rugged, compact
aluminum housing. This provides sufficient
weight and rigidity to ensure the instrument is
stable during testing. The overall dimensions of
the main units are:
Model 5135 27.9 cm x 40 cm x 25.4 cm
Model 5135 49.5 cm x 35.6 cm x 25.4 cm
(11” x 15.75” x 10”)
(19.5" x 14” x 10”)
Dual Voltage*
The 5135 and 5155 are dual voltage 115/230V,
60/50Hz switchable.
Abrading Wheels
The Taber Abraser is supplied with two types of
Genuine Taber abrading wheels, CS-10 and H-
18. For additional details on other available
wheels that may be used with this instrument,
see Characteristics of Abrading Wheels (page
21).
Auxiliary Port 1*
This feature is not currently used, but is being
supplied for future upgrades.
Figure 11
2. Slide the red voltage selector switch to the
appropriate voltage setting that the
instrument will be operated at.
NOTE:To remove the fuse holder, slide it out of the
switch selector. A small screwdriver may be used to
assist in the removal.
UWARNING:Prior to powering-up, ensure the
correct voltage is indicated per the detailed
instructions. Failure to set the correct voltage
configuration may result in damage to the Rotary
Platform Abrasion Tester’s electronic components
and will void any warranty.
* Denotes an improved feature
6 5135 / 5155 Operating Instructions ver 1.1
General Set-Up
Figure 12
1. The Abraser unit should be placed on a
flat, level and rigid surface.
your convenience. Discard the power
cord not used.
5. Turn the instrument on. The ON / OFF
power switch is located on the back of the
instrument immediately above the cord.
The instrument is now ready to operate.
UWARNING: Failure to connect the Rotary
Platform Abrasion Tester to a surge protector or
surge suppressor may result in damage to the
instrument’s electronic components.
2. Position the vacuum unit in a ventilated
area near the Abraser. Remove the motor
housing and install the vacuum cartridge
filter and filter bag.
3. Insert the stainless steel vacuum hose
elbow into the vacuum hose receptacle
found in the rear of the Abraser housing.
Connect the flexible rubber suction hose
to the vacuum unit and the elbow.
UWARNING:DO NOT place the vacuum in a closed
cabinet without adequate ventilation. Failure to
provide sufficient airflow may cause the motor to
overheat. Resulting in damage to the vacuum.
Powering-Up
1. Plug the female connector of the Abraser
line cord into the rear of the instrument.
2. Attach the male connector of the vacuum
power cord into the receptacle marked
“VACUUM” at the rear of the Abraser.
When operating at 115V, a patch cord is
used to connect the vacuum unit to the
Abraser.
3. If any accessories are to be used (Grit
Feeder or Wheel Refacer), connect the
male connector to the receptacle marked
“ACCESSORY” at the rear of the Abraser.
4. Connect the instrument to a 115 or 230
volt, 60 or 50-cycle circuit. Two power
cords (115V and 230V) are provided for
7 5135 / 5155 Operating Instructions ver 1.1
IMPORTANT
READ PRIOR TO USING INSTRUMENT
•
To activate the buttons on the membrane
switch, press firmly for approximately one (1)
second.
•
To access the MENU features or change the
MODE, the turntable must be stopped (the
Run or Reface Mode s atus indicated on the t
digital display must display “Stopped”).
1-Speed 4-Vacuum
2-Refacing 5-Outlet
3-Display 6-Reset
>Select Menu Option
1-Speed 4-Vacuum
2-Refacing 5-Outlet
3-Display 6-Reset
>0 Zero key for more
Alternating
Display
Screens
SETTING THE TEST PARAMETERS
Model 5155 Dual Abraser – The numerical
keypad is common to the entire instrument. The
digital display shows test cycles specific to each
turntable, and common functions to the entire
instrument (turntable speed and vacuum suction
level).
To change between the left or right turntables,
press either the START or STOP buttons on the
appropriate table. The last key depressed will
determine the side to which the data will be
entered, and is indicated by the word “KEYPAD”
displayed on the lower corners of the digital
display. For example, if the STOP key has been
depressed on the left side, any information
entered into the keypad will be for the left table.
Should the START or STOP key on the right
side be depressed, any values that are entered
will be for the right turntable.
Setting the Speed
Current Speed = *
*1- 60 Cycles/Minute
2- 72 Cycles/Minute
>Press 1 or 2
Current Speed = *
*1- 60 Cycles/Minute
2- 72 Cycles/Minute
>CLEAR key to return
Alternating
Display
Screens
Speed set to
72 Cycles/Minute
The speed of the Rotary Platform Abrasion
Tester is reported as CYCLES / MINUTE, where
one cycle is defined as a full rotation of the
turntable.
NOTE:The factory default speed is set at 60 RPM.
The speed of the Rotary Platform Abrasion
Tester can be selected as 60 or 72 rpm. The
tester will store the speed that was last selected
in memory and use this as a default until it is
changed. Speed is shown on the display.
The following instructions show how to change
the speed. To change speed, the instrument
must not be in operation (the Run or Reface
Mode status indicated on the digital display must
display “Stopped”).
1. Press the MENU button to display menu
options (see below for screen displays).
2. From the menu options, press 1 to select
the option to change speed.
3. Using the numeric keypad, enter “1” to
change to 60 cycles per minute or “2” to
enter 72 cycles per minute. The asterisk
(*) indicates the current speed selection.
4. Press START to begin testing.
8 5135 / 5155 Operating Instructions ver 1.1
NOTE:After pressing “1” or “2” in step 3, the display
screen will flash “Test speed set to ‘X’ Cycles/Minute”
to confirm your selection and will automatically return
to the main display screen.
NOTE:The turntables for the Model 5155 Dual
Abraser will rotate at the same rpm.
Setting the Reface Cycle Default
The Reface Mode allows you to reface the
wheels without adding test cycles to the
completed (remaining) test cycle count. This
convenient feature is ideal for any test that
requires the abrading wheels to be refaced at
intervals during the test.
The instrument is programmed with a default of
25 cycles. To change this value, follow the
instructions below.
1. Press the MENU button to display Menu
Options (see below for screen displays).
2. From the menu options, press 2 to select
Refacing.
3. Enter a new value, and press ENTER to
save.
3. Using the numeric keypad, enter “1” to
display Completed Cycles or “2” for
Remaining Cycles. The asterisk (*)
indicates the current counter display.
1-Speed 4-Vacuum
2-Refacing 5-Outlet
3-Display 6-Reset
>Select Menu Option
Change refacing
default cycles: 25
>Enter new total
>CLEAR key to return
Change refacing
default cycles: 50
>Enter new total
>ENTER key to save
Change refacing
default cycles: 50
>Enter new total
>CLEAR key to cancel
Alternating
Display
Screens
Default number of
Reface Cycles
set to
50 cycles
NOTE:After entering the new value in step 3 and
pressing ENTER, the display screen will flash “Default
number of Reface Cycles set to X cycles” to confirm
your selection and will automatically return to the main
display screen.
NOTE: If a new value is entered with the numeric
keypad or a Preset Cycles button when the Abraser is
in the Reface Mode, the number of reface cycles will
change. However, this DOES NOT change the default
refacing cycle value.
UWARNING: Do not exceed 50 cycles on an S-11
refacing disc. If more than 50 cycles are required,
replace the refacing disc.
Setting the Counter Display
The Rotary Platform Abrasion tester can be
programmed to show “Completed Cycles” (count
up) or “Remaining Cycles” (count down). The
following explains how to change the display:
1. Press the MENU button to display Menu
Options (see below for screen displays).
2. From the menu options, press 3 to select
Display.
1-Speed 4-Vacuum
2-Refacing 5-Outlet
3-Display 6-Reset
>Select Menu Option
Current Count Mode=*
*1- Completed Cycles
2- Remaining Cycles
>Press 1 or 2
display screen will flash “Counting mode set to display Remaining (Completed) Cycles” to confirm your
selection and will automatically return to the main
display screen.
Setting the Test Cycles
The Abraser can be preset to any number of test
cycles below the maximum value of 50,000.
The total cycle count shown on the display
screen defaults to the test duration value that
was last entered.
To reset the test cycles, depress the CLEAR
key, and then select the desired number of test
cycles using one of the two options discussed
below. The Abraser turntable will stop
automatically at the selected number of test
cycles. To begin testing a new specimen, press
the CLEAR key. It is not necessary to re-enter
the same value of test cycles at the beginning of
each test.
Option 1
new value may be entered with the numeric
keypad. After the desired number of test
cycles has been entered, press the ENTER or
START key. If an incorrect value is entered,
press the CLEAR key to return to the original
value.
Option 2
you conveniently enter values of 100, 500,
: With the Abraser in Run Mode, a
: The PRESET CYCLES buttons let
9 5135 / 5155 Operating Instructions ver 1.1
1000 or 2500. Press the 0 button to reset the
cycle count, and then select a preset cycle
button(s) to set the new test cycle. A
PRESET CYCLES button may be pressed
during operation or when a test is stopped, to
increase total cycles by that amount.
It is possible to increase (or decrease) test
duration beyond the original cycles. To override,
the number of turntable revolutions must be a
value greater than the completed cycles. Press
ENTER to accept the new value.
For the model 5155 Abraser, both turntables can
be preset to different values depending on the
nature of the test. Prior to entering a new value
for the test cycles, ensure the KEYPAD indicator
is shown on the correct turntable. To change
between the left or right turntables, press either
the START or STOP buttons on the appropriate
table. The ongoing cycle count for each
turntable will be displayed in the corresponding
digital readout.
NOTE: An abrasion cycle, or test cycles, is defined
as one complete revolution of the specimen in contact
with the abrading wheels.
NOTE: The instrument can be stopped at any time for
inspection and started up again (by depressing the
START key) without interrupting the counter reading in
the digital display.
NOTE:Entering a new value DOES NOT
automatically zero the completed cycle count. This
enables you to reprogram the total test cycles to
continue testing beyond the original number of cycles.
If you change total cycles to a value less than the
completed cycles, you must first press CLEAR to zero
the completed cycle count, enter the new value and
then press START.
Setting the Vacuum Level
The vacuum furnished with the Model 5135 and
5155 Abrasion Test Sets is equipped with a
variable vacuum control. For test control and
standardization of vacuum suction, the vacuum
level is calibrated to values of 50 – 100%.
Changing the vacuum level adjusts the speed of
the vacuum unit motor and the corresponding
vacuum suction for accurate standardization and
control of test procedures.
The vacuum level should be set high enough to
lift the abraded particles, but not lift flexible
samples. A vacuum level of 100% is
programmed into the Abraser. If you require
less than 100% vacuum, follow the instructions
below. The vacuum level will default to the
value that was last entered.
1. Press the MENU button to display Menu
Options (see below for screen displays).
2. From the menu options, press 4 to select
Vacuum.
3. Enter a new value between 50 and 100,
and press ENTER to save.
1-Speed 4-Vacuum
2-Refacing 5-Outlet
3-Display 6-Reset
>Select Menu Option
Change Vacuum Level
>Enter new level
>CLEAR key to return
Change Vacuum Level
>Enter new level
>ENTER key to save
Change Vacuum Level
>Enter new level
>CLEAR key to return
100%
90%
Alternating
Display
Screens
90%
Vacuum Level set to:
90%
NOTE: The factory default vacuum level is set at 100.
UWARNING:Do not use the vacuum pickup for
tests on wet materials. When conducting wet
tests, raise the vacuum nozzle, and disconnect the
electrical cord to the vacuum unit.
U WARNING:Do not place the v acuum in a closed
cabinet without adequate ventilation. Failure to
provide sufficient airflow may cause the motor to
overheat, resulting in damage to the vacuum.
Vacuum Only
The VACUUM ONLY key allows you to operate
the vacuum unit without the Abraser. This may
be necessary to clean the drive shaft and
specimen holder cavity or to properly operate
the Wheel Refacer.
If only the vacuum is needed, turn the Abraser
on then depress the VACUUM ONLY key. The
vacuum will operate at the vacuum level last
entered.
10 5135 / 5155 Operating Instructions ver 1.1
Vacuum Nozzle Clearance
The height of the vacuum pick-up nozzle is
adjustable and may be raised or lowered using
the knob found by the right abraser arm (Figure
13). Each full rotation of the adjustment knob
raises or lowers the nozzle by 1.27 mm
inch)
. To assist with vacuum nozzle adjustments,
a calibrated shim may be placed between the
test specimen and the nozzle.
(0.05
the start of the Grit Feeder at the same time a
test is commenced.
1. Press the MENU button to display Menu
Options (see below for screen displays).
2. From the menu options, press 5 to select
Outlet.
3. Using the numeric keypad, enter “1” for
Continuous Power or “2” for Power While
Testing. The asterisk (*) indicates the
current mode.
Figure 13
Many established test methods define the gap
between the test sample and vacuum pick-up
nozzle. For most flexible materials, the vacuum
nozzle should be no closer than 6.35 mm
. For rigid materials, the vacuum nozzle is
inch)
normally specified as a 1.59 mm
3.18 mm
The amount of vacuum suction is influenced by
the type of material being tested and height of
the nozzle above the specimen surface. The
intent of the vacuum system is to remove the
abraded particles BUT NOT LIFT THE TEST
SPECIMEN.
(0.125 inch) gap.
(0.0625 inch) to
(0.25
UWARNING: When testing flexible materials,
ensure the suction from the vacuum nozzle does
not lift the specimen. Any contact between the
specimen and vacuum pick-up nozzle during the
test, will cause additional wear that may influence
the test results.
1-Speed 4-Vacuum
2-Refacing 5-Outlet
3-Display 6-Reset
>Select Menu Option
Current Mode=*
*1- Continuous Power
2- Power While Test
>Press 1 or 2
Current Mode=*
*1- Continuous Power
2- Power While Test
>CLEAR key to return
Aux. Outlet Mode
Power While Testing
Set To:
U
WARNING: The use of the electrical receptacles
is restricted to the specified equipment only.
Alternating
Display
Screens
Accessory Receptacle
Incorporated into the Model 5135 and 5155
Abraser is an accessory receptacle to supply
power to either the Wheel Refacer or Grit
Feeder Attachment.
Through the MENU, you can choose to have
continuous power supplied to the receptacle or
ONLY while you test. When operating the
Wheel Refacer, select continuous power. This
allows you to reface wheels at any time,
provided the instrument is turned on. However
when operating the Grit Feeder, voltage should
be supplied ONLY when a test is being
conducted. This will enable you to coordinate
11 5135 / 5155 Operating Instructions ver 1.1
SPECIMEN SIZE
The width of the wear path is 12.7 mm (0.5 inch),
and located 31.75 mm
of the specimen. The size of the specimen may
vary depending on the material being evaluated.
For most rigid materials, a sample approximately
100 mm
(4.0 inch) square is recommended.
Flexible specimens are typically circular and
require the use of the clamp ring. If a mounting
card is used, the specimen should be
approximately 108 mm
no mounting card is used, a 135 mm
specimen is required such that the clamp ring
will grip the overlapped material.
If the specimen is less than the suggested size,
it is possible to position two or more specimens
on a mounting card to meet the size
requirement. If the edges are not flush together
or if the samples are of different thickness, the
edges may wear at a different rate than the rest
of the specimen.
The standard material thickness that can be
evaluated with the Taber Rotary Platform
Abraser is 6.35 mm
greater than 6.35 mm (0.25 inch) but less than
12.7 mm
(0.5 inch), the S-21 extension nut may
be used. Specimens with a thickness greater
than 12.7 mm
(0.5 inch) can be tested using the
arm height extension kit.
(1.25 inch) from the center
(4.25 inch) in diameter. If
(5.25 inch)
(0.25 inch). For materials
Figure 14
Mounting Sheets (S-37 and S-37-1)
These 108 mm (4.25 inch) sheets are coated on
both sides with pressure sensitive adhesive
overlaid with protective paper that is peeled off
just prior to use. Thinner than the S-36 and S36-1, mounting sheets are used to provide a
non-slip attachment of the specimen to the
surface of the turntable and are especially useful
for materials difficult to drill a center hole (such
as glass or ceramics). Mounting sheets are
available as square (S-37) or round (S-37-1),
and have a test record printed on both sides
enabling them to be used as a mounting card.
SPECIMEN MOUNTING OPTIONS
To expand the versatility of the Taber Abraser,
mounting options and interchangeable specimen
tables are available to evaluate materials not
easily tested with the standard configuration. All
tables are engineered with a ball plunger and
drive pin spring clutch that locks the table onto
the motor shaft. This prevents the table from
shifting during a test, but allows for it to be easily
removed for specimen inspection or to clean the
motor shaft cavity. To gain access to the
specimen holder, lift the suction nozzle on its
hinge and raise the abrading arms up and back
beyond their balance point.
Mounting Cards (S-36 and S-36-1)
These 108 mm (4.25 inch) rigid cards are coated
on one side with pressure sensitive adhesive.
Designed to hold textile and other flexible test
specimens tightly to the mounting surface, they
prevent wrinkling while undergoing testing. A
test record form is printed on the backside of the
card for permanent record. Mounting cards are
available as square (S-36) or round (S-36-1).
Figure 15
Figure 16
12 5135 / 5155 Operating Instructions ver 1.1
Specimen Holder (E100-125)
The specimen table equipped with the Abraser
(E100-125) has been designed to accommodate
most materials less than 6.35 mm
thick. Unless optional accessories are utilized, a
6.35 mm (0.25 inch) center hole must be punched
or drilled in the specimen so it can be securely
fastened to the threaded center post.
To secure rigid materials, use the clamp plate
and nut. For flexible materials, also use the
supplied E100-101 specimen clamp ring. A 108
mm
(4.25 inch) diameter rubber pad (S-19) is
provided to ensure a non-slip surface.
(0.25 inch)
Specimen Holder with Ring Clamp
(E140-14)
Rigid specimens that are slightly warped and
less than 6.35 mm
either dry or wet tested using the E140-14 Ring
Clamp Holder. Eight (8) screws are spaced
evenly on the clamp plate, and when tightened,
will correct for slightly warped conditions. This
holder can be furnished with (E140-14-S) or
without (E-140-14-NS) a center clamping screw.
Specimens should be 137 mm
diameter with a 6.35 mm
(0.25 inch) in thickness can be
(5.375 inch) in
(0.25 inch) center hole.
Figure 21
Figure 19
13 5135 / 5155 Operating Instructions ver 1.1
Figure 22
Textile Specimen Holder – Tensioning
Type (E140-15)
Textile Holder – Tensioning Type for Wet
or Dry Testing (E140-18)
Similar to the standard specimen holder, the
E140-15 Tensioning Textile Specimen Holder
has a slightly raised wear track. This gives
textile and other flexible specimens additional
tension when the clamp ring is drawn down over
the edge of the holder and the clamp plate is
tightened in the center recess. Requires a 134
mm (5.25 inch) diameter specimen with a 6.35 mm
(0.25 inch) center hole. This holder is not
recommended for use with rigid specimens or
with mounting cards.
NOTE: For proper mounting, place the specimen on
the holder and draw it taut with the clamp ring.
Tighten the ring, and then mount the clamp plate.
The E140-18 holder is designed to give an initial
stretch to woven fabrics so that they may be
tested while moist with minimum tendency to
wrinkle. A raised wear track provides additional
tension on the specimen, while a 9.53 mm
inch)
rim helps retain any liquid. Requires a
159mm
inch)
(6.25 inch) specimen with a 6.35 mm (0.25
center hole.
(0.375
NOTE: For proper mounting, place the specimen on
the holder, position the clamp ring and mark the holes.
Where each hole was marked, make a small cut such
that the screws may be aligned and mounted to the
holder. Partially tighten down the clamp ring. Draw
the fabric taut under the ring, tighten the screws, and
trim off excess material around the edge. Tighten the
clamp plate.
NOTE: When the fabric specimen is to be tested at
specific moisture content it must be conditioned to that
content before it is mounted on the holder.
Figure 24
Figure 27
Figure 25
14 5135 / 5155 Operating Instructions ver 1.1
Figure 28
Drive Pin Type Holder (E140-19)
Threaded Ring Specimen Holder (E140-21)
The E140-19 Drive Pin Holder is intended for
rigid, square specimens that do not have a
center-mounting hole. The specimen is adhered
to the holder with S-37 or S-37-1 mounting
sheets and opposing driving pins prevent it from
slipping while the test is being conducted.
NOTE: An example where this holder may be useful
is for glass or ceramic tile, which may be difficult to
drill a center hole.
The E140-21 Threaded Ring Specimen Holder
incorporates a clamp plate and flanged clamp
ring to test slightly warped, rigid specimens.
The ring is threaded to the body of the holder,
which securely clamps the outer edges of
specimens up to 137 mm
(5.375 inch) diameter.
Figure 30
Figure 31
Figure 33
Figure 34
15 5135 / 5155 Operating Instructions ver 1.1
Loading...
+ 37 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.