Swiss Bionic iMRS One Operating Manual

intelligent Lifestyle
Operating Manual
2
Important Notes for Getting Started
In combination with the Omnium1 control device, the iMRS one represents the latest development and application standard in the eld of magnetic reso­nance stimulation systems for home use.
The iMRS one is a system for use at home. In medical use, it can also play an accompanying and supportive role for a number of conditions.
The iMRS one complies with the following guidelines and standards:
DIN EN ISO 13485
93/42/EEC Annex VI - Section 3 Medical Product Guidelines (MPG)
IEC 60601-1 Electrical Safety
IEC 60601-1-3 Tolerance of Electromagnetic Fields
This operating manual is a component of the scope of delivery.
It should be kept close at hand and remain with the system when sold.
Copyright
Copyright © 2014 Swiss Bionic Solutions Holding GmbH All rights reserved. No part of this manual, including the products and software described herein, may be reproduced, transferred, transcribed, stored in a retrieval system or translated into another language, without the implicit written permission of Swiss Bionic Solutions Holding GmbH. Documentation stored by the purchaser for backup purposes is excluded from this condition. This condition shall not apply for software that has been licensed under the General Public License (GPL) or other free open source licensing systems. Omnium1, the Omnium1 logo and the iMRS Logo are trademarks of Swiss Bionic Solutions Holding GmbH. All other trademarks are the property of their corresponding owners. The details of the content of this manual may deviate from the product or the associated software. All information in this document
may be changed without prior notication.
3
Table of Contents
1. Safety Instructions: Where You should be Careful ........................................4
1.1 Contra-indications ......................................................................................... 9
1.2 Side Effects ....................................................................................................9
2. Intended Use .................................................................................................10
2.1 Essential Performance Characteristics of the iMRS one ...........................10
3. Possible Applications ....................................................................................10
4. Scope of Delivery ........................................................................................... 11
4.1 Control Unit .................................................................................................12
5. Installation ..................................................................................................... 13
6. Activation ....................................................................................................... 14
7. Saving and Loading Pre-set Parameters ......................................................15
7.1 Saving ........................................................................................................... 15
7.2 Loading ........................................................................................................16
8. Starting an Application..................................................................................17
9. Quick Start Programs ....................................................................................18
10. Settings ........................................................................................................18
10.1 About .........................................................................................................18
10.2 Factory Settings ......................................................................................... 19
10.3 Users .......................................................................................................... 19
11. Updating ......................................................................................................19
12. Applicators ................................................................................................... 19
12.1 OmniMat .................................................................................................... 19
12.2 OmniPad .................................................................................................... 20
12.3 OmniSpot ................................................................................................... 20
13. Cleaning and Maintenance ......................................................................... 21
14. Maintenance & Error Message ...................................................................22
15. Identication & Technical Data ..................................................................23
15.1 Identication .............................................................................................23
15.2 Technical Data ...........................................................................................23
16. Warranty ......................................................................................................25
4
1. Safety Instructions: Where You should be Careful
If you feel dizzy, be careful when standing up!
There are no negative reports on the application of Magnetic-Resonance-Stim­ulation anywhere in the world. However, for reasons of safety, we recommend that people with orthostatic problems (dizziness when getting up) get up very slowly and carefully after application.
Avoid humidity:
This device may not be positioned in a damp or wet room!
The strength of the eld of magnetic resonance stimulation system also cor­responds to no more than 120 μT at its highest intensity setting. Thereby, it
is well under the values of popular electrical household appliances and well
under the recommended threshold of 400 μT for harmlessness according to
DIN 0848.
The fact that these values are in the range of the Earth’s owns natural magnetic
elds and the frequencies in the ionosphere (Schumann resonances) exclude
all possible danger.
Note about electromagnetic tolerance (EMC)
Medical, electric devices are subject to special precautions in terms of EMC and must be installed and placed in operation in accordance with the EMC instructions in the included documents. Electro-medical devices are particular
susceptible to the radio frequencies of portable and mobile communications equipment, such as cell phone phones and walkie-talkies.
The manufacturer only guarantees the compliance of the device with the EMC
requirement when used with the accessories listed in 14.2. The use of other
accessories may lead to increased emissions of electromagnetic interference or to reduced resistance to such interference. The accessories listed may only be used together with an Omnium1 control system from Swiss Bionic Solutions Schweiz GmbH. The device may not be operated in combination with other devices, nor be placed in a stack of, or located in the proximity, to such other devices. How­ever, if such an arrangement is necessary, the operation of the device must be
5
checked to ensure that it will operate as intended when stored in this manner.
The expansion of the magnetic eld from the applicators will have largely
abated at a distance of about 1.5 meters. People who are not receiving treat­ment should remain outside the indicated range during the course of an appli­cation.
We are legally obliged, in accordance with the EMC regulations for medical products, to provide you with the following information.
Guidelines and Manufacturer‘s Declaration: Electromagnetic Interfer­ence Emissions
The iMRS one is intended for operation in an ELECTROMAGNETIC ENVIRONMENT as shown below. The customer or user of the iMRS one should ensure that it is operated in such an environment.
Interference emission mea-
surements
Compliance Electromagnec Environment Guideline
RF emissions acc. to CISPR 11 Group 1 The iMRS one only uses RF energy for
its internal OPERATION. Its RF emission
is therefore very low and it is unlikely to interfere with neighboring electronic devices.
RF emissions acc. to CISPR 11 B The iMRS one is suitable for use in all
establishments including those in resi­denal, and similar areas that are direc­tly connected to the PUBLIC POWER GRID that also supplies buildings used for residenal purposes.
Emission of harmonic
frequencies according to IEC
61000-3-2
A
Emission of voltage uctua­ons or icker according to
IEC 61000-3-3
In compliance
6
Interference Immunity Tests
IEC 60601 Test Level
Compliance Level Electromagnetic Environment
- Guidelines
Discharge of static electricity (ESD) Acc. to IEC 61000­4-2
± 6 kV contact discharge ± 8 kV air dis­charge
± 6 kV contact discharge ± 8 kV air dis­charge
Floors should be made of wood or concrete or covered
in ceramic tiles. If the oor is
covered with synthetic mate­rial, the relative air humidity must be at least 30 %.
Fast-transient interference test/ Bursts acc. to IEC 61000-4-4
± 2 kV for power supply lines ± 1 kV for input and output lines
± 1 kV for power supply lines Not applicable
The quality of the supply
voltage should correspond to that in typical business or hospital surroundings.
Surges acc. to IEC 61000-4-5
± 1 KV voltage outer conduc­tor-outer conduc­tor ± 2 kV voltage outer conductor ­ground
± 1 KV voltage outer conduc­tor-outer conduc­tor Not applicable
The quality of the supply
voltage should correspond to that in typical business or hospital surroundings.
Voltage dips, short interruptions and
uctuations in the
supply voltage acc. to IEC 61000-4-11
< 5 % Ut < (> 95 % dip in Ut) for 1/2 period 40 % Ut (60 % dip in Ut) for 5 periods 70 % Ut (30 % dip in Ut) for 25 periods < 5 % Ut (> 95 % dip in Ut) for 5 sec
0 % Ut < (> 95 % dip in Ut) for 1/2 period 40 % Ut (60 % dip in Ut) for 5 periods 70 % Ut (30 % dip in Ut) for 25 periods 0 % Ut (> 95 % dip in Ut) for 5 sec
The quality of the supply
voltage should correspond to that in typical business or hospital surroundings. If the iMRS one user requires con­tinuous operation, even when interruptions in the power supply occur, supplying the iMRS one from an uninter­ruptible power supply or a battery is recommended.
Magnetic eld at a supply frequency
(50/60 Hz) acc. to IEC 61000-4-8
3 A/m 3 A/m Magnetic elds at power grid
frequency should correspond
to values typical for a busi­ness or hospital environment.
Comment: Ut is the AC power voltage before the application of the test level.
7
Guidelines and Manufacturer’s Declaration:
Electromagnetic Stability Interference
The iMRS one is intended for operation in an ELECTROMAGNETIC ENVIRONMENT as indicated below. The customer or iMRS user should ensure that it is operated in such an environment.
Interference Immunity Tests
IEC 60601 Test Level
Com­pliance Level
Electromagnetic Environment - Guidelines
Conducted RF interference acc. to IEC 610004­4-6 Radiated RF interference acc. to IEC 610004­4-3
3 V effective value 150 kHz up to 80 MHz 3 V/m 80 MHz up to 2.5 GHz
3V Portable and mobile radio devices, including
their cables, should not be used at a distance closer to the iMRS one than recommended, which has been calculated according to relevant
equation for the transmission frequency.
Recommended safe distance:
for 80 MHz up to 800 MHz for 800 MHz up to 2.5 GHz
Where P is the rated power of the transmitter in watts (W) according to the information from the transmitter manufacturer and d is the recom-
mended safe distance in meters (m). The eld
strength of stationary radio transmitters should
be investigated locally for all frequencies lower
than the compliance level
6
Interference is possible in proximity to devices that bear the following symbol.
NOTE 1 The higher frequency range applies at 80 MHz and 800 MHz. NOTE 2 These guidelines may not apply in all cases. The expansion of electromagnetic quanti­ties will be affected by absorption and reection from buildings, objects and people.
The eld strength of stationary transmitters, such as: the base stations of cordless telephones
and land mobile radio systems, amateur radio stations, AM/FM radio and television trans­mitters; cannot be determined in advance with theoretic precision. A study of the electro­magnetic phenomena of the location should be considered in order to determine the nature
of the ELECTROMAGNETIC SURROUNDINGS in terms of stationary transmitters. If the eld
strength measured at the location where the iMRS one will be used exceeds the COMPLIANCE LEVEL mentioned above, the iMRS one should be checked to verify its OPERATION in the man­ner intended. If unusual performance characteristics are observed, additional measures may necessary, such as changing the orientation or choosing a different location for the iMRS one.
The eld strength should be less than 3V/m over the frequency range 150 kHz to 80 MHz.
6)
National footnote: User here is meant in the sense of RESPONSIBLE ORGANISATION.
d = 1,2√P d = 1,2√P
d = 2,3√P
8
Recommended Safe Distances between Portable or Mobile RF Telecom­munication Devices and the iMRS one
The iMRS one is intended for operation in an ELECTROMAGNETIC ENVIRONMENT
in which radio frequency interference is controlled. The customer or the iMRS one
user can help with the avoidance of electromagnetic interference by maintaining the
minimum distance between portable and mobile radio frequency telecommunication
devices (transmitters) and the iMRS one, depending on the output power of the com­munication device as indicated below.
Nominal rating of the
transmitter W
Safe distance (m) dependent on transmitter frequency
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters whose maximum nominal output is not indicated in the table above, the recommended safe distance of d in meters (m) should be determined from the equation associated with the particular column, where P is the maximum nominal out­put of the transmitter in watts (W) according to information from the manufacturer.
NOTE 1: The higher frequency range applies at 80 MHz and 800 MHz.
NOTE 2: These guidelines may not apply in all cases. The expansion of electromagnetic
quantities will be affected by absorption and reection from buildings, objects and
people.
9
1. 1. Contraindications
Use of the iMRS one system is contraindicated for the following conditions:
Pregnancy
Epilepsy
Electronic implants such as pace makers or insulin pumps
(with the exception of approval by the consulting physician)
The iMRS one system may only be used with the approval of a health care practi­tioner and under medical supervision under the following conditions:
Presence of tumors
Serious cardiac arrhythmia
Acute attacks of hyperthyroidism
Extreme sensitivity to electromagnetic radiation
In principle:
Magnetic resonance stimulation does not replace medical therapy. Always consult your
doctor rst about any unfamiliar complaints.
1. 2. Side Effects
In therapeutic treatment of chronic cases, a so-called initial worsening (healing reac-
tion) arises in approximately 10% of the patients treated in the rst days or weeks of application, such as through an increase in the symptoms. This should be frequently
expected after prolonged medication, which should be interpreted as a side effect of the medicinally induced regulatory habit and of the transfer process to the activation of self-regulation.
A light itching on the body or a warm feeling may be felt in the prophylactic use. In exactly the same manner, bruises, cramps, strains, wounds and problems with the
bones, joints, teeth or jaw may make themselves known as light pain as a consequence
of the activation of the circulatory system. In all cases in which previously unnoticed physical reactions become noticeable as an accompaniment of the application, consul­tation with a doctor or therapist with experience in the application of magnetic reso-
nance stimulation is recommended for purposes of safe clarication.
10
2. Intended Use
With the iMRS one magnetic resonance stimulation system, weak, pulsating electro-
magnetic elds are used to activate various physiological processes in the body. This occurs through the magnetic eld pulses and the strength of these magnetic elds is no stronger than the average magnetic ux density of the Earth’s own magnetic eld.
Use other than as described in this manual may lead to damage to the device and unin-
tended health consequence and should therefore be avoided.
2.1. Essential Performance Characteristics of the
iMRS one
The essential performance characteristic of the iMRS one is the generation of specied magnetic eld pulses and the reinforcement of the body›s own energy elds.
3. Possible Applications
The iMRS one magnetic resonance stimulation system can be used:
For wellness purposes and to generally increase vitality and well-being.
For palliative purposes, such as the temporary relief of minor muscular aches and
pains.
For a temporarily increase of local blood circulation.
To relax muscles locally.
Important notice:
The iMRS one is regulated by the FDA as a therapeutic massager (21 CFR 890.5660). However, the iMRS one is not intended to diagnose, treat, cure or prevent any disease.
Those seeking treatment for a specic disease should consult a qualied integrative
physician prior to using our products.
11
4. Scope of Delivery
If one part should be damaged or missing, please contact your consultant at Swiss Bionic Solutions.
iMRS one Android software
(Pre-installed on the Omnium1)
OmniMat
OmniPad
OmniSpot
Not included with the iMRS one
basic set
D/A Converter 20-pin Cable
Operating Manual
12
4.1 Control Unit
The iMRS one will only function in connection with the Omnium1 control unit. The Omnium1 control device is included in the scope of delivery for all iMRS
one sets available. You will  nd the operating instructions for the Omnium1 in
the special manual that accompanies the control device.
Omnium1 Tablet
Charging unit
Earphones
Protective display  lm
Quick Start Guide
13
5. Installation
1. Take the individual parts out of the packaging. Place the full-body applica-
tor on a stable, even surface, such as on the oor, on a couch or under the
mattress (but not box-spring mattresses) of your bed (pay attention to the intensity adjustment, please). Heavily shaped and soft furniture, such as a sofa, chair, very soft surfaces should be avoided, since the pressure load on uneven supports may lead to damage to the built-in copper coils.
2. Aside from this, make sure that your environment is as free of electro-smog
as possible. There should not be any devices like: TVs, microwave ovens, ra­dio-based telephones and so forth in the immediate vicinity (approx. 1-2m) during an application.
3. Connect one of the supplied applicators to the Omnium1 control device as
follows:
Connect the supplied D/A converter with the 20-pin ribbon cable
Connect the second plug attached to the 20-pin ribbon cable with the
corresponding connection on the Omnium1
(see the Omnium1 operating
manual)
.
Connect the desired applicator with the D/A Converter.
4. Make sure that battery for your Omnium1 has been suciently charged.
The application will not be possible if the battery charge state drops below 5%. In that case, connect the power supply with the Omnium1
(see the
Omnium1 operating manual)
.
5. Check the system time selected (upper right) and set it to the current time,
if necessary
(see the Omnium1 operating manual)
. The built-in organ clock
will adjust automatically to this system clock during an application.
14
6. Activation
After you have completely installed the system and checked the battery charge state, activate the iMRS one app by tapping the program icon (1) on the Om­nium1 desktop (2).
Afterwards, the iMRS one initial screen will display the legally indicated cont­raindications. As soon as it has been acknowledged (by pressing ENTER), you will see the user interface for the iMRS one application.
2
1
15
7. 7. Saving and Loading Pre-set
Parameters
The iMRS one application provides you with the ability to save various users in advance. You can save all organ clocks with all time and intensity levels for each applicator in advance and load them as needed.
7.1 Saving
In order to create a new user, tap on the “ Settings” button on the user interface and then tap “Users”.
Tap the “Name” entry (4) and overwrite the placeholder. Then, tap on Next and the virtual keyboard will disappear
You can now start making settings for the  rst applicator with the  rst organ
clock setting. To do so, tap the circle next to Timer (3) and set the desired application duration from 1 to 60 minutes. It can be set in steps of 1 minute.
Afterwards, tap the circle next to Intensity (5) and set the desired intensity of
the magnetic  eld ( ux density). This setting has been divided into the levels:
Sensitive, 10, 25, 50, 100, 150, 200 and 400.
Tap on “Next” (6) to set the next organ clock. If you have set all four organ clocks (morning, afternoon, evening and night), you will automatically be taken to the next applicator.
2
3
4
5
6
1
16
If you have set all of the applicators, tap on “Save” (6) to save your entries per­manently on the Omnium1.
If you want to create another user, tap the “+” icon (2) above the list of users and repeat the process (in principle, any number of users desired can be cre­ated). If you created all of the users desired, you can tap “Back” (1) to return to the main screen for the iMRS one application.
7.2 Loading
Choose the user, once the system has successfully been started. To do that, tap on the User icon (2). A list of all of the created users (1) will appear. Tap on the desired user and start the application by tap the “Start/Stop” button (3). The iMRS one application will always detect automatically the connected appli­cator and use the pre-set user parameters from internal memory.
2
1
3
17
8. Starting an Application
Connect the D/A converter and the desired applicator (see Chapter 5, Installa-
tion) and start the iMRS one application (see Chapter 6, Activation). After the
system has been successfully started, chose the desired application duration. To do so, tap the circle next to Timer (1) and set the desired application dura­tion from 1 to 60 minutes. It can be set in steps of 1 minute.
Afterwards, tap the circle next to Intensity (4) and set the desired intensity of
the magnetic eld (ux density). This setting has been divided into the levels:
Sensitive, 10, 25, 50, 100, 150, 200 and 400.
The integrated organ clock will automatically adjust to the time set on your Omnium1. You can see the organ clock setting on the display in the upper right corner (3). However, if you want to change it manually, tap on the organ clock icon (2) and choose the desired organ clock setting (morning, afternoon, evening or night).
The magnetic eld application can be started by tapping the Start/Stop button
(5).
2
1
5
4
3
18
9. 9. Quick Start Programs
In addition to the parameters that can be manually set, the iMRS one appli-
cation provides  ve pre-set quick start programs (exclusive with the full body applicator). Tapping one button will su ce to activate the following programs:
Relaxation
Performance
Activation
Sleep
Regeneration
The magnetic  eld application will start immediately after you have touched one of the  ve quick start buttons. All of the necessary parameters have already been de ned.
10. Settings
Tapping on the “ Settings” button will open a sub-menu with various menu items. Tap the desired button to open a sub-menu item.
10.1 About
From this sub-menu item, you can see the version of the  rmware & hard-
ware, the serial number for the D/A converter (if it is connected), the version of the iMRS one software app as well as the serial number of the Omnium1 unit.
19
10.2 Factory Settings
By tapping the “Reset Factory Settings” button, you can restore the iMRS one app to its delivery state. Note: all of the stored users will be deleted.
10.3 Users
See Chapter 7, Saving and Loading Pre-set Parameters.
11. Updating
Whenever your Omnium1 is connected to the internet (via WIFI), your iMRS
one App automatically veries, whether a new software version is available. In this case a notication appears on your screen. Please klick on „YES“ followed by clicking on „Install“. The System will automatically update the App to the
newest version.
12. Applicators
Three applicators are available for selection for the iMRS one app.
12.1 OmniMat
The OmniMat full body applicator, which is being used as a general full body treatment. Three solid copper coil pairs with a different number of windings (intensity) have been built into the full body applicator. The copper coil pair at the head (cable connected to the applicator box) has the lowest number of
windings and thereby generates the lowest ux density. The middle pair of
copper coils has an already higher number of windings and the bottom pair
of coils (at the foot) has the highest number of windings with the high ux
density.
The full body applicator is divided by seams. It can be folded into three parts on the seams, however should not be bent or rolled in order to protect the copper coils.
The full body applicator has been covered with certied articial leather. The magnetic eld not only works directly above the full body applicator, but also spreads itself out in all directions. The eld strength amounts to less than or equal to 45 μT (micro-teslas) at the highest intensity setting on the full body
20
applicator. The recommended threshold for harmlessness in low-frequency magnetic elds is 400 μT in accordance with the standard that preceded DIN
0848. Horizontally, the expansion of the magnetic eld from the applicators
will have largely abated at a distance of about 1.5 meters.
This application can only be performed with the original full body applicator. If an applicator is not connected or the applicator is defective, an error message will appear on the Omnium1’s display. In both cases, the application will not allow itself to be started.
12.2 OmniPad
The OmniPad pillow applicator can be used for local applications, for example: a knee, a foot, a hand, a shoulder, the back and so forth. It has been covered
with certied articial leather. The material is easy to maintain, can be cleaned
and can be rinsed with mild disinfecting agents.
One pair of solid copper coils has been built into the OmniPad. It can be folded in the middle and has an extensible attachment belt.
Please note that the pillow applicator should not be completely covered by a blanket or plastic lm. Circulation of air must be ensured as a protective mea­sure against the formation of moisture.
The magnetic eld not only works directly above the pillow applicator, but also spreads itself out in all directions. The ux density of the pillow applica­tor is less than or equal to 70 μT at its highest intensity. The recommended threshold for harmlessness in low-frequency electromagnetic elds is 400 μT
in accordance with the standard that preceded DIN 0848. Horizontally, the
expansion of the magnetic eld from the applicators will have largely abated
at a distance of about 1.5 meters.
This application can only be performed with the original pillow applicator. If an applicator is not connected or the applicator is defective, an error message will appear on the Omnium1’s display. In both cases, the application will not allow itself to be started
12.3 OmniSpot
(Not included with the iMRS one basic set)
The OmniSpot applicator can be used for isolated applications, for example:
a knee, a foot, a hand and so forth. It has been covered with certied articial
21
leather. The material is easy to maintain, can be cleaned and can be rinsed with mild disinfecting agents.
Two solid copper coils have been built into the OmniSpot.
The magnetic eld not only works directly above the OmniSpot applicator, but also spreads itself out in all directions. The ux density of the pillow applica­tor is less than or equal to 120 μT at its highest intensity. The recommended threshold for harmlessness in low-frequency electromagnetic elds is 400 μT
in accordance with the standard that preceded DIN 0848. Horizontally, the
expansion of the magnetic eld from the applicators will have largely abated
at a distance of about 1.5 meters.
The placement of the coils across from each other will generate a so-called
Helmholtz effect (meaning a homogenous magnetic eld). It has a exible
attachment belt for simple and effective application.
13. Cleaning and Maintenance
Omnium1 Control Unit:
Please use a dry micro-ber cloth to clean ngerprints and skin oil from the
touchscreen. You can remove most remnants with a circular motion. For rough dirt, lightly moisten the cloth and clean the affected area with that. Note: Be very careful that water does not enter the housing, because that could lead to an irreparable defect in the control device.
Applicators:
The applicators should be cleaned at periodic intervals in order to maintain their visual appearance and avoid the collection of dirt and contaminants.
More frequent cleaning may be necessary depending on the frequency of
use and demand. Spots, dirt and any substances that attach themselves to the material should be immediately removed in order to avoid permanent staining. Use a mild soap in water solution or products recommended for
cleaning vinyl materials and articial leather and the removal spots from the
surface. Ultimately, use a moist white cloth for cleaning. Enamel, aggressive
or chemical cleaning or washing agents, liquids containing xylenes, acetone
or methyl-ethyl ketones (MEKs) will cause immediate damage and contribute to material exhaustion. The use of such agents will be at the user’s own risk.
22
14. Maintenance & Error Messages
The iMRS one system has been designed to be maintenance free.
Potential Error Messages:
Error Message Explanation
DA Converter not working properly The D/A converter is defective.
Low battery for DA convertor The battery is below the 5% charge state.
Error Coil Open Loop There is a defect or broken wire in the applicator.
Error No Applicator Connected An applicator has not been connected, or not been
connected properly.
Error No DA Converter Connected The D/A converter has not been connected, or not
been connected properly.
Goggle does not connect The OMNIBRAIN system has not been connected, or
not been connected properly (optional!).
Repairs and service should only be performed by the manufacturer or the
respective local oces of Swiss Bionic Solutions (you will nd the addresses at: www.swissbionic.com) or from your supporting Certied LifeStyle Consultant.
Manufacturer: Swiss Bionic Solutions Schweiz GmbH Bodenackerstr. 69 / E1 4657 Dulliken Switzerland
Phone: +41-62-2955951 Fax: +41-62-2955952 E-Mail: ch@swissbionic.com
23
15. Identication & Technical Data
15.1 Identication
Specicaon Plate for the iMRS one (D/A converter):
CE-Idencaon:
0483
The conformity according to EC direcves has been explained for the equipment described in this Operator›s Guide.
15.2. Technical Data for the iMRS one D/A Converter
Nr. Designation Values, unit, type and model
Control Unit
1 Device type
Magnec-Resonance-Smulaon
2 Type designation iMRS one
3 Nominal voltage 18 V
4 Nominal current 1,66 A
5 Max. voltage 30 VA
6 Avg. power consumption 9 W
7 HF frequency None
8 Mode of use
Connuous operaon
Typ: iMRS one V1.1
Hersteller/ Swiss Bionic Soluons Schweiz GmbH Vertrieb: Bodenackerstr. 69 4657 Dulliken, Schweiz
Betrieb nur mit Orginal-Zubehör
SN: 123456
®
Home
0483
®
Home
0483
0483
0483
®
Home
0483
®
Home
0483
Note: follow the instrucons in the operang manual
Applied part, Type B
Legacy Electrical Device Act
24
9 Construction Portable device
10 Protection class II
11 Application part Type B
12 Type of moisture protection Covered device
13 Fuse between the primary and sec-
ondary power supplies
Omnium1 Type BI20-180100-I power supply
14 Initial voltage at the applicator box Max. 20 V direct current
15 Boxes for connection with the Omni-
Mat, OmniPad and OmniSpot
M12, 5-pin
16 Case V0
17 Magnetic eld strength at the highest
intensity level (400%)
< 120 µT
18 Duration of use, selectable (using
automatic deactivation)
1-60 minutes
19 Weight 160 g
20 External dimensions 167 mm x 60 mm x 27 mm
21 Temperature (in use) +10°C bis +40°C
22 Temperature (stored) -20°C bis +45°C
23 Air humidity (in use) 30% - 75% RH (without condensation)
24 Air humidity (stored) 10% - 95% RH (without condensation)
25 Air pressure (in use) 700 - 1060 hPa
26 Air pressure (stored) 700 - 1060 hPa
Omnium1 Power Supply
1 Type BI20-180100-I
2 Power supply 100 V - 240 V~ / 50 Hz - 60 Hz / 500 mA
3 Output DC 18V 1A
4 Cable length 1,50 m
OmniMat Full Body Applicator
1 Coils 3 pairs of solid, uninsulated coils, copper
2 Covering Articial leather, disinfectable, foldable into 3
layers, do not roll or bend
3 Dimensions 170 x 58 x 2 cm
4 Cable length 2 m
OmniPad Pillow Applicator
1 Coils 1 pair of solid coils, copper
2 Covering Articial leather, disinfectble, do not roll or
bend
3 Dimensions 57 x 29 x 3 cm
4 Cable length 2 m
25
OmniSpot Applicator
1 Coils 2 solid copper coils
2 Covering Articial leather, disinfectble, do not roll or
bend
3 Dimensions 39 x 16 x 2,5 cm
4 Cable length 2 m
Intensity Measures values in µT OmniMat
whole body applicator
Measures values in µT OmniPad pillow applicator
Measures values in µT OmniSpot applicator
Foot Abdomen Shoulder
Sensitive 0,27 0,22 0,09 0,35 0,65
10 1,35 1,22 0,54 1,70 3,14
25 4,00 3,60 1,60 5,00 9,23
50 8,00 7,20 3,20 10,00 18,46
100 16,00 14,40 6,40 20,00 36,92
150 24,00 21,60 9,60 30,00 46,15
200 32,00 28,80 12,80 40,00 55,38
400 45,00 30,00 17,50 65,00 120,00
16. Warranty
The iMRS one magnetic resonance stimulation system is the result of inno­vative research and development work. Swiss Bionic Solutions provides the warranty on the applicators and the D/A converter for a period of 36 months from the date of purchase.
The warranty can only be maintained by the use of the system and the acces­sories as intended. For this reason, read the instruction in this operating man-
ual precisely. Warranty claims resulting from errors, damage or consequential
damage that result from non-compliance with the operating manual and the safety instruction shall not exist. The warranty shall only apply for the use with original iMRS one accessories. Save the purchase receipt for the system in order to demand warranty claims.
Version: 07/2014
26
27
28
Swiss Bionic Solutions Schweiz GmbH
Bodenackerstrasse 69 | 4657 Dulliken, Schweiz
Phone: +41 (62) 295 5951 | Fax: +41 (62) 295 5952
Email: ch@swissbionic.com
Swiss Bionic Solutions Deutschland GmbH
Am Sandfeld 17a | 76149 Karlsruhe, Deutschland
Phone: +49 (721) 464 7180 | Fax: +49 (721) 464 71818
Email: de@swissbionic.com
Swiss Bionic Solutions USA Inc.
1200 NE 7Th Ave Suite 7 | Fort Lauderdale, Florida 33304, USA
Phone: +1 (954) 766 4153 | Fax: +1 (954) 766 4156
Email: us@swissbionic.com
Swiss Bionic Solutions Canada Inc.
64 Skipper Lane | Oakville, ON, L6L 5Y8, Kanada
Phone: +1 (905) 465 0753 | Fax: +1 (1 866) 792 8182
Email: ca@swissbionic.com
Swiss Bionic Solutions Asia Ltd.
998 Canton Road | Mongkok | Kowloon | Hong Kong
Phone: +852 2337-8774
Email: asia@swissbionic.com
www.swissbionic.com
Loading...