Swema SwemaTemp 21 Instructions Manual

SwemaTemp 21
Instructions
vers 1.17 KT20171113
SWEMA AB Pepparv. 27 SE-123 56 FARSTA Tel: +46 8 94 00 90 Fax: +46 8 93 44 93 E-mail: swema@swema.com Web Site: www.swema.com
Contents:
1. Introduction......................... 1
2. Technical data....................... 1
3. Start of instrument................ 1
4. Settings............................... 2
5. Temperature sensors............... 3
Read this before you start using the instrument!
It’s very important that you select Pt-100 mode if you have a Pt-100 sensor and Ni-100 mode if you have Ni-100 sensor, otherwise you will get inaccurate meter readings. See page 2 how to choose between Pt-100 and Ni-100.
1. Introduction
SwemaTemp 21 measures temperature with both Pt100 eller Ni100 sensors. SwemaTemp 21 is delivered with two easily replaceable AA batteries. Standard, alkaline or regular batteries can be used.
2. Technical data
Measuring range: -200…850°C; IEC 60751 for Pt100 and DIN 43760 for Ni100. One decimal resolution
Uncertainty (only SwemaTemp 21): At 0…50°C: ±0,15% read value, min ±0,3°C At -200…850°C: ±0,2% read value, min ±0,4°C See probes technical data for uncertainty of SwemaTemp 21 including probe.
The user should correct the measured values with the correction on the probe calibration protocol to obtain stated uncertainty of measu­rement.
95% coverage probability in non condensing, non moist air, <80%RH, non aggressive gases.
Battery: 2 x 1,5V IEC LR6, AA Size: 180x36x82mm Weight: 380g
3. Start of instrument
Press the HOLD button to freeze the value. A star will appear on the screen.
1.
Always turn off the instrument before removing the probe.
When starting the instrument (ON/OFF- button), the SwemaTemp 21 goes to measuring mode after six seconds.
If the voltage should drop below 2.4 V, the symbol for low battery is shown. If the voltage should drop below 1.6 V, the instrument can’t be turned on.
Press the LAMP button to light up the screen.