SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
This document describes the main steps for instrument setup, operation and maintenance.
1.Safety Instructions
GeneralThe instructions included in this section explain the potential risks
Targ et
audience
OM LocationThe AMI Operator’s Manual shall be kept in proximity of the instru-
Qualification,
Training
associated with instrument operation and provide important safety
practices designed to minimize these risks.
If you carefully follow the information contained in this section, you
can protect yourself from hazards and create a safer work environment.
More safety instructions are given throughout this manual, at the
respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Operator: Qualified person who uses the equipment for its intended
purpose.
Instrument operation requires thorough knowledge of applications,
instrument functions and software program as well as all applicable
safety rules and regulations.
ment.
To be qualified for instrument installation and operation, you must:
read and understand the instructions in this manual as well as
the Material Safety Data Sheets.
know the relevant safety rules and regulations.
A-96.250.321 / 1910153
Page 6
AMI Solicon4
Safety Instructions
1.1.Warning Notices
The symbols used for safety-related notices have the following significance:
DANGER
Your life or physical wellbeing are in serious danger if such
warnings are ignored.
Follow the prevention instructions carefully.
WARNING
Severe injuries or damage to the equipment can occur if such
warnings are ignored.
Follow the prevention instructions carefully.
CAUTION
Damage to the equipment, minor injury, malfunctions or incorrect process can be the consequence if such warnings are ignored.
Follow the prevention instructions carefully.
Mandatory
Signs
The importance of the mandatory signs in this manual.
Safety goggles
Safety gloves
4A-96.250.321 / 191015
Page 7
AMI Solicon4
Safety Instructions
Warning SignsThe importance of the warning signs in this manual.
Electrical shock hazard
Corrosive
Harmful to health
Flammable
Warning general
Attention general
A-96.250.321 / 1910155
Page 8
AMI Solicon4
Safety Instructions
1.2.General Safety Regulations
Legal
Requirements
Spare Parts
and
Disposables
ModificationsModifications and instrument upgrades shall only be carried out by
The user is responsible for proper system operation.
All precautions must be followed to ensure safe operation
of the instrument.
Use only official SWAN spare parts and disposables. If other parts
are used during the normal warranty period, the manufacturer’s
warranty is voided.
an authorized Service Technician. SWAN will not accept responsibility for any claim resulting from unauthorized modification or alteration.
WARNING
Risk of Electrical Shock
If proper operation is no longer possible, the instrument must be
disconnected from all power lines, and measures must be taken
to prevent inadvertent operation.
To prevent from electrical shock, always make sure that the
ground wire is connected.
Service shall be performed by authorized personnel only.
Whenever electronic service is required, disconnect instru-
ment power and power of devices connected to.
– relay 1,
– relay 2,
– alarm relay
WARNING
For safe instrument installation and operation you must read
and understand the instructions in this manual.
WARNING
Only SWAN trained and authorized personnel shall perform the
tasks described in this document.
6A-96.250.321 / 191015
Page 9
AMI Solicon4
Product Description
2.Product Description
2.1.Description of the System
This instrument is applicable for the measurement of the specific
conductivity in surface water, potable water and cooling water. The
complete system consists of the Transmitter AMI Solicon4, the conductivity sensor Swansensor Shurecon P and the flow cell
M-Flow PG.
Transmitter, flow cell and sensors are also available as single components.
Application
Range
Conductivity is a parameter for the total quantity of ions present in
the solution. It can be used for the controlling of:
the condition of waters
water purification
water hardness
completeness of ion analysis
Signal
Outputs
RelaysTwo potential-free contacts programmable as limit switches for
Alarm RelayOne potential free contact.
A-96.250.321 / 1910157
Two signal outputs programmable for measured values (freely scalable, linear, bilinear, log) or as continuous control output (control
parameters programmable).
Current loop:0/4 – 20 mA
Maximal burden:510 Ohm
Third signal output with the same specifications as option.
(Only possible if no communication interface is used.)
measuring values, controllers or timer for system cleaning with automatic hold function. Both contacts can be used as normally open
or normally closed.
Maximum load: 1 A/250 VAC
Alternatively:
Open during normal operation, closed on error or loss of
power.
Closed during normal operation, open on error or loss of
power.
Summary alarm indication for programmable alarm values and instrument faults.
Page 10
AMI Solicon4
Product Description
InputFor potential-free contact to freeze the signal outputs or to interrupt
control in automated installations (hold function or remote-off).
Communica-
tion interface
(optional)
Safety
Features
Tem per atu re
compensation
Measuring
principle
USB interface to store logger data.
RS485 interface with fieldbus protocol Modbus, Profibus DP or
No data loss after power failure. All data is saved in non-volatile
memory.
Over voltage protection of in- and outputs.
Galvanic separation of measuring inputs and signal outputs.
The analyzer is factory tested and ready for installation and operation.
The mobility of ions in water increase with higher temperature
which enlarges the conductivity. Therefore, the temperature is measured simultaneous by an integrated Pt1000 temperature sensor
and the conductivity is compensated to 25 °C.
When a voltage is set between two electrodes in an electrolyte solution, the result is an electric field which exerts force on the
charged ions: the positively charged cations move towards the negative electrode (cathode) and the negatively charged anions towards the positive electrode (anode). The ions, by way of capture
or release of electrons at the electrodes, are discharged and so a
current I flows through this cycle and the Ohms law V = I×R applies. From the total resistance R of the current loop, only the resistance of the electrolyte solution, respectively its conductivity
of interest.
The cell constant of the sensor is determined by the manufacturer
and is printed on the sensor label. If the cell constant has been programmed in the transmitter, the instrument measures correctly. No
calibration must be done, the sensor is factory calibrated.
Webserver connection via Modbus.
1
/R, is
8A-96.250.321 / 191015
Page 11
AMI Solicon4
A
C
B
E
F
D
G
Product Description
FluidicsThe flow cell M-Flow PG consists of the flow cell block [B] and the
calibration vessel [D].
The conductivity sensor [A] is fixed in the flow cell block [B] with a
threaded sleeve.
The sample enters at the sample inlet [F]. It flows through the del-
taT flow sensor [C] (if installed) and then through the flow cell block
into the calibration vessel [D], were the conductivity is measured.
The sample leaves the calibration vessel via flow cell block through
the sample outlet [E] and flows into the drain [G].
A
Conductivity sensor
B
Flow cell block
C
deltaT flow sensor (option)
D
Calibration vessel
A-96.250.321 / 1910159
E
Sample outlet
F
Sample inlet
G
Drain
Page 12
AMI Solicon4
Product Description
2.2.Instrument Specification
Power Supply
Electronics
housing
Sample
requirements
On-site
requirements
Measuring
Range
Accuracy
Voltage:
Power consumption:
Aluminium with a protection degree of IP 66 / NEMA 4X
Ambient temperature:
Limit range of operation:
Storage and transport:
Humidity:
Display:
100–240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
max. 30 VA
-10 to +50 °C
-25 to +65 °C
-30 to +85 °C
10–90% rel., non condensing
backlit LCD, 75 x 45 mm
4 to 15 l/h
up to 50 °C
up to 1 bar
pressure free
Flow celll with elbow hose nozzle 1/4”
diam.10 mm
or
deltaT sensor with hose nozzle
diam. 10 mm
1/2” hose nozzle for flexible tube
diam. 20 x 15 mm
2000 m above sea level
Resolution
0.01 S/cm
0.1 S/cm
1 S/cm
0.01 mS/cm
0.1 mS/cm
1 mS/cm
10A-96.250.321 / 191015
Page 13
AMI Solicon4
AMI Solicon4
850 mm / 33½”
13 mm / ½”
6 x dia. 6.5 mm / ¼”
254 mm/ 10”
280 mm/ 11”
412 mm /16 ¼”
824 mm / 32
7
/
16
”
Product Description
Dimensions
Panel:
Dimensions:
Screws:
Weight:
white PVC
280x850x 150 mm
5 mm or 6 mm diameter
9.0 kg
A-96.250.321 / 19101511
Page 14
AMI Solicon4
A
B
C
D
E
F
G
H
I
Product Description
2.3.Instrument Overview
F
Transmitter
A
Panel
B
Conductivity sensor
C
Flow cell block
D
Calibration vessel
E
12A-96.250.321 / 191015
Sample outlet
G
DeltaT flow sensor (option)
H
Drain
I
Sample inlet
Page 15
AMI Solicon4
Product Description
2.4.Single Components
2.4.1AMI Solicon4 Transmitter
Electronic transmitter and controller for conductivity measurement.
4 to 15 l/h
up to 50 °C
up to 1 bar @ 25 °C
Pressure-free outlet
below 0.5 mm
No strong acids and bases.
No organic solvents.
Width:
Front-to-back:
Height:
Panel mounting:
90 to 142 mm
105 mm
161 mm
3 screws M5
14A-96.250.321 / 191015
Page 17
AMI Solicon4
16.5
12
46
43
Sensor adapter
16 mm / PG 13.5
PG 13.5
120 ± 2
Product Description
2.4.3Swansensor Shurecon P
Sensor with integrated cable for the measurement of the specific
conductivity. Four electrode design with platinum electrodes and
built-in Pt1000 temperature sensor.
Specifications
Measuring range
Accuracy:
Temperature sensor:
Operating conditions:
0.1S/cm to 100mS / cm
± 1.5 % or ± 0.2 S/cm whichever is
greater
Pt 1000
> 50°C
Max. temperature short-time 90 °C
Max.pressure 10 bar at 25 °C.
Electrical connection:
Process connection:
A-96.250.321 / 19101515
Sensor with integrated cable
PG 13.5 mm
Page 18
AMI Solicon4
94
25
164
258
G 1”
thread
108
Product Description
2.4.4Swansensor Shurecon S
Sensor for the measurement of the specific conductivity. Four electrode design with integrated Pt 1000 temperature sensor.
Screw connector with cable must be ordered separately.
Specifications
Measuring range
Accuracy:
Temperature sensor:
Operating conditions:
0.1S/cm to 100mS / cm
± 1 % or ± 0.2 S/cm whichever is
greater
Pt 1000
Max. temperature: 120°C at 6.5 bar
Max. pressure: 12 bar at 20°C
Sterilizable at: 120°C / 5 bar / 30 min.
Plug M16 male (IP 67)
G 1" thread
20 mm
Electrical connection:
Process connection:
Space around sensor
tip:
16A-96.250.321 / 191015
Page 19
AMI Solicon4
250
80
70
12
4.5
175
163
Product Description
2.4.5Swansensor DeltaT
Calorimetric flow meter based on heat dissipation. For applications
in potable water, surface water treatment and effluent.
Specifications
Measuring range/Flow rate
Accuracy:
Response time t
90
:
Sample temperature:
A-96.250.321 / 19101517
Sample inlet and outlet:
Max. cable length:
0–40 l/h
± 20%
ca. 1 min
5–35 °C
for tubing diam. 10–11 mm
1 m
Page 20
AMI Solicon4
Installation
3.Installation
3.1.Installation Checklist Monitors
CheckInstrument’s specification must conform to your AC power rat-
ings. Do not turn on power until instructed to do so.
On site requirements
InstallationMount the instrument in vertical position. Display should be at
Electrical Wiring
Power-upTurn on the sample flow and wait until the flow cell is completely
Instrument
set-up
Run-in periodLet the instrument run continuously for 1 h.
CalibrationPerform a calibration if necessary, see Calibration, p. 42.
100–240 VAC (± 10%), 50/60 Hz (± 5%) or 24 VDC, isolated
(±10%) power outlet with ground connection and 30 VA.
Sample line with sufficient sample flow and pressure (seeInstru-
ment Specification, p. 10
eye level.
Connect sample and waste.
Connect all external devices like limit switches, current loops
and pumps.
Connect power cord; do not switch on power yet!
filled.
Switch on power.
Adjust sample flow.
Program all parameters for sensor and external devices (interface, recorders, etc.).
Program all parameters for instrument operation (limits, alarms).
18A-96.250.321 / 191015
Page 21
AMI Solicon4
AB
Installation
3.2.Mounting of Instrument Panel
To mount the instrument panel proceed according the following list:
The instrument must only be installed by trained personnel.
Mount the instrument in vertical position.
For ease of operation mount it so that the display is at eye
level.
For the installation a kit containing the following installation
material is available:
– 6 Screws 6 x60 mm
– 6 Dowels
– 6 Washers 6.4/12 mm
Mounting
requirements
3.3.Connecting Sample Inlet and Outlet
Sample inletUse a plastic tube (FEP, PA, or PE 10x12 mm) to connect the sam-
The instrument is only intended for indoor installation, dimensions
see Dimensions, p. 11.
ple inlet at the flow cell.
ABPlastic tube 10x12
Elbow hose nozzle
If a deltaT senor is installed, push the plastic tube over the hose
nozzle of the deltaT sensor.
Sample outletConnect the flexible tube 20x15 mm [C] to the 1/2” hose nozzle [B]
and place it into a drain with atmospheric pressure.
A
A
B
C
A-96.250.321 / 19101519
Waste funnel
B
1/2” Hose nozzle
C
Flexible tube 20x15 mm
Page 22
AMI Solicon4
A
Installation
3.4.Install Swansensor deltaT (Option)
The following description assumes that the installation of the deltaT
sensor takes place after commissioning of the monitor.
Install the deltaT sensor in vertical position with the sample inlet [F]
and cable gland looking downwards.
To ensure laminar flow, the sample inlet must not be restricted; e.g.
any fitting which creates turbulences.
B
C
D
E
F
A
Panel
B
Tube connection
C
Elbow hose nozzle
D
Hose nozzle at deltaT sensor outlet
E
deltaT sensor
F
Hose nozzle at deltaT sensor inlet
20A-96.250.321 / 191015
Page 23
AMI Solicon4
Installation
3.4.1Mount the deltaT Sensor and Connect the Sample Inlet
1 Stop operation according to chapter Stop of Operation for Main-
2 Mount the deltaT sensor [E] in vertical position to the panel [A].
3 Remove the plastic tube from the elbow hose nozzle at the flow
4 Connect the plastic tube to the hose nozzle [F] of the deltaT
5 Rotate the elbow hose nozzle of the flow cell by 180 degree.
6 Install the hose connection [B], enclosed in the installation kit,
tenance, p. 40.
cell.
sensor inlet.
from the hose nozzle [D] at the deltaT sensor outlet to the elbow
hose nozzle [C].
A-96.250.321 / 19101521
Page 24
AMI Solicon4
BA
Installation
3.4.2Connect the Sensor Cable to the Transmitter
Make sure the power specification of the instrument corresponds to
the power on site.
WARNING
Risk of electrical shock.
Do not perform any work on electrical components if the transmitter is switched on. Failure to follow safety instructions could
result in serious injury or death.
Always turn off AC power before manipulating electric parts.
Grounding requirements: Only operate the instrument from
an power outlet which has a ground connection.
1 Remove the plug [A] from the cable gland [B]
2 Open the AMI transmitter housing.
3 Feed the sensor cable through the cable gland [B] into the
transmitter housing.
4 Connect the cable to the terminals according to the connecting
diagram see Connection Diagram, p. 27.
5 Close the AMI transmitter housing.
6 Switch on power.
22A-96.250.321 / 191015
Page 25
AMI Solicon4
5.1.4.1
Flow
Flow measurementdeltaT
Q-Flow
Flow measurement
deltaT
none
5.1.4.1
Flow
Flow measurementdeltaT
Slope1.00
5.1.4
Standard 11 ppm
Flow
Standard 210 ppm
Sensors
Type of sensorNH4
No
Save ?
Yes
Installation
3.4.3Change Firmware Settings
1 Navigate to <Installation> <Sen-
sors> <Flow> <Flow measurement>.
2 Press [Enter]
3 Select <deltaT>
4 Press [Enter]
5 Press 2 x [Exit]
6 Press [Enter] to confirm with Yes
7 Press [Exit] until the display shows
the measuring values.
A-96.250.321 / 19101523
Page 26
AMI Solicon4
Installation
3.4.4Adjust the deltaT sensor
The deltaT flow sensor is factory calibrated at 20 °C (± 20 % accuracy).
The accuracy of the flow measurement depends on the ambient
temperature of the location where the deltaT sensor is installed. If
the ambient temperature is higher or lower than 20 °C, the deltaT
flow sensor can be adjusted. If an adjustment of the deltaT sensor
is necessary proceed as follows:
Run inAfter installation let the sensor run-in for at least 1h.
Determine the
flow rate
Adjust slope1 Navigate to <Installation/ Sensors / Flow>, choose <Slope> and
1 Put the sample outlet of the instrument for 10 min. into a mea-
suring cup with a sufficient volume.
2 To get the flow rate in l/h, multiply the amount of water con-
tained in the measuring cup with factor 6.
The flow rate in l/h results from the multiplication of the
amount of water after 10 min by 6.
press [Enter].
2 If the calculated flow rate is higher than the displayed flow rate
increase the Slope value.
3 If the calculated flow rate is lower than the displayed flow rate
decrease the Slope value.
4 Press [Exit] and save with [Enter].
5 Compare the calculated flow rate with the displayed flow rate.
If the flow rates are roughly equal, the adjustment is finished.
Else repeat step 1 to 5.
24A-96.250.321 / 191015
Page 27
AMI Solicon4
Installation
3.5.Electrical Connections
WARNING
Risk of electrical shock.
Do not perform any work on electrical components if the transmitter is switched on. Failure to follow safety instructions could
result in serious injury or death.
Always turn off AC power before manipulating electric parts.
Grounding requirements: Only operate the instrument from
an power outlet which has a ground connection.
Make sure the power specification of the instrument corre-
sponds to the power on site.
Cable
thicknesses
Wire For Power and Relays: Use max. 1.5 mm
In order to comply with IP66, use the following cable thicknesses
ABC
A
PG 11 cable gland: cable Ø
B
PG 7 cable gland: cable Ø
C
PG 9 cable gland: cable Ø
outer
3–6.5 mm
outer
4–8 mm
outer
5–10 mm
Note: Protect unused cable glands
2
/ AWG 14
stranded wire with end sleeves.
For Signal Outputs and Input: Use 0.25 mm
2
/ AWG 23
stranded wire with end sleeves.
A-96.250.321 / 19101525
Page 28
AMI Solicon4
Installation
WARNING
External Voltage.
External supplied devices connected to relay 1 or 2 or to the
alarm relay can cause electrical shocks.
Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming installation.
–relay 1
–relay 2
– alarm relay
WARNING
To prevent from electrical shock, do not connect the instrument
to the power unless the ground wire (PE) is connected.
Do not connect unless specifically instructed to do so.
WARNING
The mains of the AMI Transmitter must be secured by a main
switch and appropriate fuse or circuit breaker.
26A-96.250.321 / 191015
Page 29
AMI Solicon4
Installation
3.5.1Connection Diagram
CAUTION
Use only the terminals shown in this diagram, and only for the
mentioned purpose. Use of any other terminals will cause short
circuits with possible corresponding consequences to material
and personnel.
A-96.250.321 / 19101527
Page 30
AMI Solicon4
A
B
C
D
Installation
3.5.2Power Supply
WARNING
Risk of electrical shock
Do not perform any work on electrical components if the transmitter is switched on. Failure to follow safety instructions could
result in serious injury or death.
Always turn off AC power before manipulating electric parts.
Installation and maintenance of electrical parts must be per-
formed by professionals
A
Power supply connector
B
Neutral conductor, Terminal 2
C
Phase conductor, Terminal 1
D
Protective earth PE
Note: The protective earth wire (Ground) has to be connected
to the grounding terminal.
Installation
requirements
The installation must meet the following requirements.
Transmitter mains fuse 1,6 AT
Mains cable to comply with standards IEC 60227 or IEC
60245; flammable rating FV1
Mains equipped with an external switch or circuit-breaker
– near the instrument
– easily accessible to the operator
– marked as interrupter for AMI Solicon4
28A-96.250.321 / 191015
Page 31
AMI Solicon4
10
12
11
0V
Installation
3.6.Relay Contacts
3.6.1Input
Terminals 16/42
For programming see Program Overview, p. 48.
3.6.2Alarm Relay
Alarm output for system errors.
Error codes see Error List, p. 45.
TerminalsDescriptionRelay connection
1)
NC
Normally
Closed
10/11Active (opened) during normal
Note: Use only potential-free (dry) contacts.
Note: Max. load 1 A/250 VAC
NOTICE: With certain alarms and certain settings of the AMI
transmitter the alarm relay does not switch. The error, however,
is shown on the display.
operation.
Inactive (closed) on error and
loss of power.
1)
0V
11
10
12
NO
Normally
Open
12/11Active (closed) during normal
operation.
Inactive (opened) on error and
loss of power.
1) usual use
A-96.250.321 / 19101529
Page 32
AMI Solicon4
6
0V
7
6
0V
7
A
B
Installation
3.6.3Relay Contacts 1 and 2
Relay 1 and 2 can be configured as normally open or as normally
closed. Standard for both relays is normally open. To configure a
Relay as normally closed, set the jumper in the upper position.
Note: Rated load 1 AT / 250 VAC
NOTICE: Some error codes and the instrument status may
influence the status of the relays described below.
Relay
config.Terminals
Normally
Open
Normally
Closed
6/7: Relay 1
8/9: Relay 2
6/7: Relay 1
8/9: Relay 2
Jumper
pos.DescriptionRelay configuration
Inactive (opened) during normal operation
and loss of power.
Active (closed) when a
programmed function is
executed.
Inactive (closed) during
normal operation and
loss of power.
Active (opened) when a
programmed function is
executed.
ABJumper set as normally open (standard setting)
Jumper set as normally closed
For programming see Menu Installation Program List and Explana-
tions, p. 53.
30A-96.250.321 / 191015
Page 33
AMI Solicon4
A
BC
DE
AB
C
Installation
CAUTION
Risk of damage of the relays in the AMI Transmitter due to
heavy inductive load.
Heavy inductive or directly controlled loads (solenoid valves,
dosing pumps) may destroy the relay contacts.
To switch inductive loads > 0.1 A use an AMI relay box avail-
able as an option or suitable external power relays.
Inductive loadSmall inductive loads (max 0.1A) as for example the coil of a power
Resistive loadResistive loads (max. 1A) and control signals for PLC, impulse
ActuatorsActuators, like motor valves, are using both relays: One relay con-
relay can be switched directly. To avoid noise voltage in the
AMI Transmitter it is mandatory to connect a snubber circuit in parallel to the load.
A
AC or DC power supply
B
AMI Transmitter
C
AMI Relay box
D
Snubber
E
Power relay coil
pumps and so on can be connected without further measures
A
AMI Transmitter
B
PLC or controlled pulse pump
C
Logic
tact is used for opening, the other for closing the valve, i.e. with the
2 relay contacts available, only one motor valve can be controlled.
Motors with loads bigger than 0.1A must be controlled via external
power relays or an AMI relay box.
A
A
BC
AC or DC power supply
B
AMI Transmitter
C
Actuator
A-96.250.321 / 19101531
M
Page 34
AMI Solicon4
Installation
3.7.Signal Outputs
3.7.1Signal Output 1 and 2 (current outputs)
Signal output 1: Terminals 14 (+) and 13 (-)
Signal output 2: Terminals 15 (+) and 13 (-)
For programming see Program Overview, p. 48, Menu Installation
3.8.Interface Options
Note: Max. burden 510
If signals are sent to two different receivers, use signal isolator
(loop isolator).
A
AMI Transmitter
A
B
Slot for interfaces
C
Frontend PCB
D
Screw terminals
B
C
D
The slot for interfaces can be used to expand the functionality of
the AMI instrument with either:
an additional signal output
a Profbus or Modbus connection
a Hyperterminal interface
an USB Interface
32A-96.250.321 / 191015
Page 35
AMI Solicon4
A
Installation
3.8.1Signal Output 3
Requires the additional board for the third signal output 0/4 – 20 mA
PCB
The third signal output is installed in the upper holder on the main
board. You can operate either 3. signal output or communication interface, not both!
Terminal 38 (+) and 37 (-).
.
Note: Max. burden 510
.
A Third signal output 0/4 - 20 mA PCB installed on main board
A-96.250.321 / 19101533
Page 36
AMI Solicon4
PB-DPV1 V5.5
ON
OFF
81.420.020
Profibus
A
Installation
3.8.2Profibus, Modbus Interface
Terminal 37 PB, Terminal 38 PA
To connect several instruments by means of a network or to config-
ure a PROFIBUS DP or a MODBUS connection, consult the PROFIBUS/MODBUS manual. Use appropriate network cable.
Profibus, Modbus Interface PCB (RS 485)
A On - OFF switch
Note: The switch must be ON, if only one instrument is
installed, or on the last instrument in the bus.
3.8.3USB Interface
The USB Interface is used to store Logger data and for Firmware
up load. For detailed information see the corresponding installation
instruction.
USB Interface
34A-96.250.321 / 191015
Page 37
AMI Solicon4
Instrument Setup
4.Instrument Setup
4.1.Establish Sample Flow
1 Open sample flow tap.
2 Wait until the flow cell is completely filled.
3 Switch on power.
4.2.Programming
ProgrammingThe sensor characteristics are printed on the label of the sensor.
Program all sensor parameters in Menu 5.1.1.1,
<Installation> <Sensors> <Sensor parameters>:
Note: Cable length [m] Set the cable length to 0.0 m if the
sensors are installed in the flow cell on the AMI monitor.
Program all parameters for external devices (interface, recorders,
etc.). Program all parameters for instrument operation (limits,
alarms). See Program Overview, S. 48, for explanations, see Pro-
gram List and Explanations, S. 53.
A-96.250.321 / 19101535
Page 38
AMI Solicon4
Exit
Enter
BCDA
25.4°C
RUN
9 l/h
14:10:45
R1
525S
R2
1
Installation
Operation
Diagnostics
Messages
Maintenance
Main Menu
Enter
Exit
Operation
5.Operation
5.1.Keys
Ato exit a menu or command (rejecting any changes)
Bto move DOWN in a menu list and to decrease digits
Cto move UP in a menu list and to increase digits
Dto open a selected sub-menu
to move back to the previous menu level
to accept an entry
Program
Access, Exit
36A-96.250.321 / 191015
Page 39
AMI Solicon4
RUN
15:20:18
R1
R2
9.5 l/h24.8°C
µS
525
ABD
E
F
H
G
C
Operation
5.2.Display
A RUNnormal operation
B ERROR Error
C Keys locked, transmitter control via Profibus
D Time
E Process values
F Sample temperature
G Sample flow
H Relay status
HOLDinput closed or cal delay: Instrument on hold (shows
status of signal outputs).
OFFinput closed: control/limit is interrupted (shows status
of signal outputs).
Fatal Error
Relay status, symbols
upper/lower limit not yet reached
upper/lower limit reached
control upw./downw. no action
control upw./downw. active, dark bar indicates control intensity
motor valve closed
A-96.250.321 / 19101537
motor valve: open, dark bar indicates approx. position
timer
timer: timing active (hand rotating)
Page 40
AMI Solicon4
1
Messages
Operation
Maintenance
Diagnostics
Main Menu
Installation
1.1
Pending Errors
Messages
Message List
2.1
Interface
I/O State
Sample
Identification
Sensors
Diagnostics
3.1
Calibration
Simulation
Maintenance
Set Time 23.09.06 16:30:00
Process Cal.
4.1
Logger
Relay Contacts
Sensors
Operation
5.1
Interface
Miscellaneous
Relay Contacts
Sensors
Signal Outputs
Installation
Operation
5.3.Software Structure
Menu Messages 1
Reveals pending errors as well as an event history
(time and state of events that have occurred at an
earlier point of time).
It contains user relevant data.
Menu Diagnostics 2
Provides user relevant instrument and sample data.
Menu Maintenance 3
For instrument calibration, relay and signal output
simulation, and to set the instrument time.
It is used by the service personnel.
Menu Operation 4
User relevant parameters that might need to be
modified during daily routine. Normally password
protected and used by the process-operator.
Subset of menu 5 - Installation, but process-related.
Menu Installation 5
For initial instrument set up by SWAN authorized
person, to set all instrument parameters. Can be
protected by means of password.
38A-96.250.321 / 191015
Page 41
AMI Solicon4
5.1.2
Sensors
Sensor typeFOME
TemperatureNT5KStandards
Disinf.Free chlorine
4.4.1
Logger
Log interval30 min
Clear loggerno
4.1.3
Logger
Clear loggerno
Log interval30min
1 Hour
Interval.
5 min
30 min
10 min
4.1.3
Logger
Log interval10 min
Clear loggerno
4.1.3
Logger
Log interval
Clear loggerno
No
Save ?
Yes
5.3.1.1.1
Alarm High300 mS
Alarm Conductivity
Alarm Low0.00 S
Hysteresis1.00
S
Delay5 Sec
5.3.1.1.1
Alarm Conductivity
Alarm Low0.00
S
Hysteresis1.00
S
Delay5 Sec
Alarm High120 mS
Operation
5.4.Changing Parameters and values
Changing
parameters
The following example shows how to change the logger interval:
1 Select the parameter you want to
change.
2 Press [Enter]
3 Press [] or [] key to
highlight the required parameter.
4 Press [Enter] to confirm the selec-
tion or [Exit] to keep the previous
parameter).
The selected parameter is
highlighted but not saved yet.
5 Press [Exit].
Yes is highlighted.
6 Press [Enter] to save the new pa-
rameter.
The system reboots, the new
parameter is set.
Changing
values
A-96.250.321 / 19101539
1 Select the value you want to
change.
2 Press [Enter].
3 Set required value with [] or
[] key.
4 Press [Enter] to confirm the new
value.
5 Press [Exit].
Yes is highlighted.
6 Press [Enter] to save the new val-
ue.
Page 42
AMI Solicon4
Maintenance
6.Maintenance
6.1.Maintenance Schedule
Preventive maintenance frequency depends on water quality, on
the application, and on national regulations.
Control of certain set-point: Swimming pools, sanitary water:
WARNING
Stop operation before maintenance.
Stop sample flow.
Shut off power of the instrument.
Monthly
If required
Check sample flow.
Clean conductivity sensor
Perform a calibration
6.2.Stop of Operation for Maintenance
Stop sample flow.
Shut off power of the instrument.
40A-96.250.321 / 191015
Page 43
AMI Solicon4
A
C
D
B
E
Maintenance
6.3.Maintenance of the Sensor
A
Fixing sleeve
B
Conductivity sensor
C
Teflon washer
D
O-Ring
E
Flow cell block
6.3.1Remove the Sensor form the Flow Cell
To remove the sensor form the flow cell proceed as follows:
1 Unscrew and remove the fixing sleeve [A].
2 Pull the conductivity sensor [B] out of the flow cell block [B].
CleaningIf the sensor is contaminated, take a small brush and clean it with
water and detergents.
In case of heavy contamination with oil or grease, use ethanol to
clean it. Take a soft tissue and clean the tip of the sensor cautiously.
After each cleaning, the sensor must be rinsed with clean water.
6.3.2Install the Sensor into the Flow Cell
1 Make sure that the washer [C] and the O-ring [D] are in correct
position.
2 Push the sensor through the flow cell bock [E] into the flow cell.
3 Tighten the fixing sleeve [A] to fix the sensor.
A-96.250.321 / 19101541
Page 44
AMI Solicon4
3.1.5
Calibration
Clean the sensor
and place it in
standard solution
[Enter] to continue
3.1.5
Calibration
Sensor must have a
min. distance of 3 cm
from the beakers edge
[Enter] to continue
3.1.1
Calibration
Standard solution1.41 mS
Current Value10.07 S
Cell constant0.406 cm
-1
Progress
Maintenance
6.4.Calibration
Since the sensor is very reliable a calibration is normally not necessary.
A calibration is recommended if:
Reagent for
Calibration:
Calibration solution 1.413 mS/cm (25 °C) 1000 ml. Prepare according to DIN 38404 / ISO 7888: 1985 / EN 27888.
the
1 Stop the sample flow.
2 Navigate to menu <Maintenance> /<Calibration>.
3 Press [Enter] and follow the dialog on the Display.
4 Remove the sensor from the flow cell.
5 Clean the sensor carefully and rinse it with clean water, see
6 Use a beaker and fill it with one liter calibration solution.
7 Put the sensor into the beaker filled with calibration solution.
the cell constant is not known
the sensor is contaminated
the maintenance measurement shows a discrepancy.
Maintenance of the Sensor, p. 41.
The beakers diameter must be so large that the sensor has a
distance of min. 3 cm from its edge.
8 Wait at least 5 minutes to permit
temperature equilibration between
sensor and calibration solution.
9 Start the calibration procedure.
10 Press [Enter], to save the values if
the calibration was successful.
11 Install the sensor into the flow cell.
42A-96.250.321 / 191015
Page 45
AMI Solicon4
ABCDEF
Maintenance
6.5.Replacing Fuses
When a fuse has blown, find out the cause and fix it before replacing it with a new one.
Use tweezers or needle-nosed pliers to remove the defective fuse.
Use original fuses provided by SWAN only.
WARNING
External Voltage.
External supplied devices connected to relay 1 or 2 or to the
alarm relay can cause electrical shocks.
Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming installation.
– relay 1
– relay 2
– alarm relay
A
1.6 AT/250V Instrument power supply
B
1.0 AT/250V Relay 1
C
1.0 AT/250V Relay 2
D
1.0 AT/250V Alarm relay
E
1.0 AF/125V Signal output 2
F
1.0 AF/125V Signal output 1
A-96.250.321 / 19101543
Page 46
AMI Solicon4
Maintenance
6.6.Longer Stop of Operation
1 Stop sample flow.
2 Shut off power of the instrument.
3 Remove the sensor form the flow cell and dry it with a soft tis-
4 Empty and dry the flow cell.
sue.
44A-96.250.321 / 191015
Page 47
AMI Solicon4
25.4°C
HOLD
8 l/h
14:10:45
R1
525SR2
1
Installation
Operation
Diagnostics
Messages
Maintenance
Main Menu
1.1
Message List
Pending Errors
Messages
1.1.5
Pending Errors
Error CodeE002
Alarm low
[Enter] to Acknowledge
Error List
7.Error List
Error
Non-fatal Error. Indicates an alarm if a programmed value is exceeded.
Such Errors are marked E0xx.Fatal Error (blinking symbol)
Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect.
Fatal Errors are divided in the following two categories:
Errors which disappear if correct measuring conditions are re-
covered (i.e. Sample Flow low).
Such Errors are marked E0xx
Errors which indicate a hardware failure of the instrument.
Such Errors are marked E0xx
Error or fatal Error
Error not yet acknowledged.
Check Pending Errors 1.1.5 * and
take corrective action.
Press [ENTER].
A-96.250.321 / 19101545
Navigate to menu Messages.
Press [ENTER].
Navigate to menu Pending Errors.
Press [ENTER].
Press [ENTER] to acknowledge the
Pending Errors. The Error is reset and
saved in the Message List.
Page 48
AMI Solicon4
Error List
ErrorDescriptionCorrective action
E001Cond. Alarm high
E002Cond. Alarm low
E003Conc. Alarm high
E004Conc. Alarm low
E007 Sample Temp. high
E008 Sample Temp. low
E009Sample Flow high
E010Sample Flow low
E011Temp. shorted
E012Temp. disconnected
E013Case Temp. high
E014Case Temp. low
– check process
– check programmed value, see 5.3.1.1,
p. 62
– check process
– check programmed value, see 5.3.1.1,
p. 62
– check process
– check programmed value, see 5.3.1.5,
p. 64
– check process
– check programmed value, see 5.3.1.5,
p. 64
– check process
– check programmed value, see 5.3.1.3,
p. 63
– check process
– check programmed value, see 5.3.1.3,
p. 63
– check sample inlet pressure
– check programmed value, see
Signal OutputsSignal Output 1/ 2Parameter5.2.1.1/5.2.2.1*
5.2*5.2.1 /5.2.2*Current Loop5.2.1.2/5.2.2.2*
Function5.2.1.3/ 5.2.2.3*
ScalingRange Low5.2.x.40.10/ 11*
5.2.x.40Range High5.2.x.40.20 /21*
Relay ContactsAlarm RelayAlarm Conductivity Alarm High 5.3.1.1.1.1*
5.3*5.3.1*5.3.1.1*Alarm Low 5.3.1.1.1.25*
Hysteresis *5.3.1.1.1.35
Delay 5.3.1.1.1.45*
Sample FlowFlow Alarm5.3.1.2.1*
5.3.1.2*Alarm High5.3.1.2.2
Alarm Low5.3.1.2.35
Sample Temp.Alarm High5.3.1.3.1*
5.3.1.3*Alarm Low5.3.1.3.25*
Case Temp.highCase Temp. high 5.3.1.4.1*
5.3.1.4*Case Temp. low5.3.1.4.2*
A-96.250.321 / 19101551
Page 54
AMI Solicon4
Program Overview
Alarm Concentration Alarm High 5.3.1.1.5.1*
5.3.1.5*Alarm Low 5.3.1.1.5.25*
Hysteresis *5.3.1.1.5.35
Delay 5.3.1.1.5.45*
Relay 1/ 2Function5.3.2.1/*
5.3.2/ 5.3.3*Parameter5.3.2.20*
Setpoint5.3.2.300 *
Hysteresis5.3.2.400*
Delay5.3.2.50*
InputActive5.3.4.1*
5.3.4*Signal Outputs5.3.4.2*
Output/ Control5.3.4.3*
Fault5.3.4.4*
Delay5.3.4.5*
MiscellaneousLanguage 5.4.1*
5.4*Set defaults5.4.2*
Load Firmware5.4.3*
PasswordMessages5.4.4.1*
5.4.4*Maintenance5.4.4.2*
Operation5.4.4.3*
Installation5.4.4.4*
Sample ID5.4.5*
InterfaceProtocol5.5.1*(only with RS485
5.5*Device Address5.5.21*interface)
Baud Rate5.5.31*
Parity5.5.41** Menu numbers
52A-96.250.321 / 191015
Page 55
AMI Solicon4
Program List and Explanations
9.Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5Provides the list of active errors with their status (active, acknowl-
1.2 Message List
1.2.1Shows the error history: Error code, date / time of issue and status
2 Diagnostics
2.1 Identification
2.1.3Factory Test: Test date of the Instrument, Motherboard and Fron-
2.1.4Operating Time: Years / Days / Hours / Minutes / Seconds
2.2 Sensors
2.2.1Cond. Sensor
2.2.1.5Cal. History: In this menu the calibration values of the last
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
(active, acknowledged, cleared). 65 errors are memorized. Then
the oldest error is cleared to save the newest error (circular buffer).
In diagnostics mode, the values can only be viewed, not modified.
Designation: Designation of the instrument.
Version: Firmware of instrument (e.g. V6.00-10 / 15)
tend
Current value in µS
Raw value in µS
Cell Constant
Contamination
calibrations are saved.
Number: Numbers the calibrations in descending order.
Date, Time: Date and time of each calibration.
Cell constant: Shows the cell constant of the sensor in use.
Max. 64 data records are memorized. One process calibration corresponds to one data record.
A-96.250.321 / 19101553
Page 56
AMI Solicon4
Program List and Explanations
2.2.2Miscellaneous:
2.2.2.1Case Temp: Shows the current temperature in °C inside the trans-
2.3 Sample
2.3.1Sample ID: Shows the identification assigned to a sample. This
2.4 I/O State
2.4.1/2.4.2
mitter.
identification is defined by the user to identify the location of the
sample.
Temperature: Shows the current sample temperature in °C.
(Pt 1000): Shows the current sample temperature in Ohm.
Sample Flow: If Q-Flow is chosen
Shows the current sample flow in l/h
Raw Value: shows the sample flow in Hz.
Sample Flow: If deltaT is chosen
Shows the current sample flow in l/h
deltaT 1: Temperature measured at sample inlet of the
deltaT sensor
deltaT 2: Temperature measured at sample outlet of the
deltaT sensor
Shows current status of all in- and outputs.
Alarm Relay:Inactive or active
Relay 1 and 2:Inactive or active
Input:Open or closed
Signal Output 1 and 2:Actual current in mA
Signal Output 3 (option): Actual current in mA
2.5 Interface
Only available if optional interface is installed.
Review programmed communication settings.
54A-96.250.321 / 191015
Page 57
AMI Solicon4
Program List and Explanations
3 Maintenance
3.1 Calibration
Follow the commands on the screen. Save the value with the [Enter] key.
with the [] or [] key.
Press the [Enter] key.
Change the value or state of the selected item with the [] or
[] key.
Press the [Enter] key.
The value is simulated by the relay / signal output.
Alarm Relay:Inactive or active
Relay 1 and 2:Inactive or active
Signal Output 1 and 2:Actual current in mA
Signal Output 3 (option): Actual current in mA
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated values will be reset.
3.3 Set Time
Adjust date and time.
A-96.250.321 / 19101555
Page 58
AMI Solicon4
Program List and Explanations
4 Operation
4.1 Sensors
4.1.1Filter Time Constant: Used to damp noisy signals. The higher the
4.1.2Hold after Cal.: Delay permitting the instrument to stabilize again af-
4.2 Relay Contacts
4.3 Logger
4.3.1Log Interval: Select a convenient log interval. Consult the table be-
filter time constant, the slower the system reacts to changes of the
measured value.
Range: 5–300 Sec
ter calibration. During calibration plus hold-time, the signal outputs
are frozen (held on last valid value), alarm values, limits are not active.
Range: 5–6‘000 Sec
See Relay Contacts, p. 29
The instrument is equipped with an internal logger. The data can be
downloaded to a PC by SWAN Terminal if option “SWAN Terminal
interface” is installed or via USB stick if option “USB interface” is installed.
The logger can save approx. 1500 data records. The Records consists of: Date, time, alarms, measured value, measured value uncompensated, temperature, flow.
Range: 1 Second to 1 hour
low to estimate the max logging time. When the logging buffer is
full, the oldest data record is erased to make room for the newest
one (circular buffer).
Interval1 s5 s1 min5 min10 min30 min1 h
Time25 min2 h25 h5 d10 d31 d62 d
4.3.2Clear Logger: If confirmed with yes, the complete logger data is de-
leted. A new data series is started.
4.3.3If option USB interface is installed.
Eject USB Stick: With this function all logger data are copied to the
USB stick before the USB stick is deactivated.
Only visible it the optional USB interface is installed.
56A-96.250.321 / 191015
Page 59
AMI Solicon4
Program List and Explanations
5 Installation
5.1 Sensors
5.1.1Sensor parameters
5.1.1.1Cell Constant: Enter the cell constant printed on the sensor label.
5.1.1.2Temp. Corr: Enter the temperature correction printed on the sensor
label.
5.1.1.3Cable length: Enter the cable length. Set the cable length to 0.0 m if
the sensors are installed in the flow cell on the AMI monitor.
5.1.2Temp. Compensation
5.1.2.1Comp.: Available compensation models are:
none
Coefficient
non-linear DIN
5.1.3Flow
5.1.3.1Flow measurement: Select the type of flow sensor if a flow sensor
is installed.
Possible flow sensors: None; Q-Flow; deltaT.
5.1.3.2Slope: If flow measurement is set to deltaT.
The slope value is used to adjust the flow measurement of the
deltaT sensor if the ambient temperature is higher or lower
than 20 °C.
5.1.4Conc.
The menu <Concentration> (Conc.) allows the additional measurement of a known substance in the sample. The concentration of the
substance is calculated based on the conductivity of any of the following substances. The calculated value is displayed in %. As an
exception, TDS is displayed in mg/l.
5.2.1 and 5.2.2Signal Output 1 and 2: Assign process value, the current loop
range and a function to each signal output.
Note: The navigation in the menu <Signal Output 1> and
<Signal Output 2> is identical. For reason of simplicity only the
menu numbers of Signal Output 1 are used in the following.
5.2.1.1Parameter: Assign one of the process values to the signal output.
5.2.1.2Current Loop: Select the current range of the signal output.
Make sure the connected device works with the same current
range.
Available ranges: 0–20 mA or 4–20 mA
5.2.1.3Function: Define if the signal output is used to transmit a process
value or to drive a control unit. Available functions are:
Linear, bilinear or logarithmic for process values.
See As process values, p. 58
Control upwards or control downwards for controllers.
See As control output, p. 60
As process
values
The process value can be represented in 3 ways: linear, bilinear or
logarithmic. See graphs below.
[mA]
20
(0 - 20 [mA])
(4 - 20 [mA])
10 12
0 / 4
0.00.10.20.30.40.5
ABlinear
AB
X
X Measured value
bilinear
58A-96.250.321 / 191015
Page 61
AMI Solicon4
20
1
01234
101001’000 10’000
10 12
(0 - 20 [mA])
0 / 4
426
(4 - 20 [mA])
[mA]
X
Program List and Explanations
X Measured value (logarithmic)
5.2.1.40Scaling: Enter beginning and end point (Range low & high) of the
linear or logarithmic scale. In addition, the midpoint for the bilinear
scale.
Parameter Conductivity:
5.2.1.40.10Range low: 0 S–300 mS
5.2.1.40.20Range high: 0 S–300 mS
5.2.1.40.11Range low: -25 to + 270 °C
5.2.1.40.21Range high: -25 to +270 °C
5.2.1.40.12Range low: 0 –50 l / h
5.2.1.40.22Range high: 0 –50 l/ h
5.2.1.40.13Range low: 0 S–300 mS
5.2.1.40.23Range high: 0 S–300 mS
A-96.250.321 / 19101559
Parameter Temperature
Parameter Sample flow
Parameter Cond. uc:
Page 62
AMI Solicon4
Program List and Explanations
As control
output
Signal outputs can be used for driving control units. We distinguish
different kinds of controls:
P-controller: The controller action is proportional to the devia-
tion from the setpoint. The controller is characterized by the
P-Band. In the steady-state, the setpoint will never be
reached. The deviation is called steady-state error.
Parameters: setpoint, P-Band
PI-controller: The combination of a P-controller with an
I-controller will minimize the steady-state error. If the reset
time is set to zero, the I-controller is switched off.
Parameters: setpoint, P-Band, reset time.
PD-controller: The combination of a P-controller with a
D-controller will minimize the response time to a fast change
of the process value. If the derivative time is set to zero, the
D-controller is switched off.
Parameters: setpoint, P-Band, derivative time.
PID-controller: The combination of a P-, an I - and a D-con-
troller allows a proper control of the process.
Parameters: setpoint, P-Band, reset time, derivative time.
Ziegler-Nichols method for the optimization of a PID controller:
Parameters: Setpoint, P-Band, Reset time, Derivative time
Y
B
A
X
a
L
A
Response to maximum control output
B
Tangent on the inflection point
X
Time
Xp
Tn
Tv
= 1.2/a
= 2L
= L/2
The point of intersection of the tangent with the respective axis will
result in the parameters a and L.
Consult the manual of the control unit for connecting and programming details. Choose control upwards or downwards.
60A-96.250.321 / 191015
Page 63
AMI Solicon4
Program List and Explanations
Control upwards or downwards
Setpoint: User-defined process value for the selected parameter.
P-Band: Range below (upwards control) or above (downwards con-
trol) the set-point, within the dosing intensity is reduced from 100%
to 0% to reach the setpoint without overshooting.
5.2.1.43Control Parameters: if Parameters = Conductivity
5.2.1.43.10Setpoint
Range: 0 S–300 mS
5.2.1.43.20P-Band:
Range: 0 S–300 mS
5.2.1.43Control Parameters: if Parameters = Temperature
5.2.1.43.11Setpoint
Range: -25 to +270 °C
5.2.1.43.21P-Band:
5.2.1.43Control Parameters: if Parameters = Sample flow
5.2.1.43.12Setpoint
5.2.1.43.22P-Band:
5.2.1.43Control Parameters: if Parameters = Cond. uc.
5.2.1.43.13Setpoint
5.2.1.43.23P-Band:
5.2.1.43.3Reset time: The reset time is the time till the step response of a sin-
5.2.1.43.4Derivative time: The derivative time is the time till the ramp re-
5.2.1.43.5Control timeout: If a controller action (dosing intensity) is constantly
Range: -25 to +270 °C
Range: 0 –50 l/h
Range: 0 –50 l/h
Range: 0 S–300 mS
Range: 0 S–300 mS
gle I-controller will reach the same value as it will be suddenly
reached by a P-controller.
Range: 0–9’000 Sec
sponse of a single P-controller will reach the same value as it will
be suddenly reached by a D-controller.
Range: 0–9’000 Sec
over 90% during a defined period of time and the process value
does not come closer to the setpoint, the dosing process will be
stopped for safety reasons.
Range: 0–720 min
A-96.250.321 / 19101561
Page 64
AMI Solicon4
Program List and Explanations
5.3 Relay Contacts
5.3.1Alarm Relay: The alarm relay is used as cumulative error indicator.
Under normal operating conditions the contact is active.
The contact isinactive at:
Power loss
Detection of system faults like defective sensors or electronic
parts
High case temperature
Process values out of programmed ranges.
Note: Above definition assumes that the alarm relay is used as
normally closed (Terminals 10/11). If terminals 12/11 are used,
the alarm relay is normally open, then the above definition is
inverted. See also Alarm Relay, p. 29 and Connection Diagram,
p. 27
Program alarm levels, hysteresis values and delay times for the following parameters:
Alarm Conductivity
Sample Flow
Sample Temp.
Case Temp.
Alarm Concentration (visible if a Conc. parameter has been
selected)
5.3.1.1Alarm Conductivity
5.3.1.1.1Alarm High: If the measured value rises above the alarm high val-
5.3.1.1.25Alarm Low: If the measured value falls below the alarm low value,
5.3.1.1.35Hysteresis: Within the hyst. range, the relay does not switch. This
5.3.1.1.45Delay: Duration, the activation of the alarm relay is retarded after
ue, the alarm relay is activated and E001, is displayed in the message list.
Range: 0 S–300 mS
the alarm relay is activated and E002 is displayed in the message
list.
Range: 0 S–300 mS
prevents damage of relays contacts when the measured value fluctuates around the alarm value.
Range. 0 S–300 mS
the measuring value has risen above/fallen below the programmed
alarm.
Range: 0–28‘800 Sec
62A-96.250.321 / 191015
Page 65
AMI Solicon4
Program List and Explanations
5.3.1.2Sample Flow: Define at which sample flow an alarm should be issued.
5.3.1.2.1Flow Alarm: Program if the alarm relay should be activated if there
is a flow alarm. Choose between yes or no. The flow alarm will
always be indicated in the display, pending error list, saved in the
message list and the logger.
Available values: Yes or no
Note: Sufficient flow is essential for a correct measurement.
We recommend to program yes.
5.3.1.2.2Alarm High: If the measuring values rises above the programmed
5.3.1.2.35Alarm Low: If the measuring values falls below the programmed
5.3.1.3Sample Temp.
5.3.1.3.1Alarm High: If the measured value rises above the alarm high val-
5.3.1.3.25Alarm Low: If the measured value falls below the alarm low value,
5.3.1.4Case Temp.
5.3.1.4.1Case Temp. high: Set the alarm high value for temperature of elec-
5.3.1.4.2Case Temp.low: Set the alarm low value for temperature of elec-
value E009 will be issued.
Range: 9–20 l/h
value E010 will be issued.
Range: 5–8 l/ h
ue, the alarm relay is activated and E007, is displayed in the message list.
Range: 30–200 °C
the alarm relay is activated and E008 is displayed in the message
list.
Range: -10 to + 20 °C
tronics housing. If the value rises above the programmed value
E013 is issued.
Range: 30–75 °C
tronics housing. If the value falls below the programmed value
E014 is issued.
Range: -10 to +20 °C
A-96.250.321 / 19101563
Page 66
AMI Solicon4
Program List and Explanations
5.3.1.5Alarm Concentration: Visible if a Conc. parameter has been selected. TDS is displayed in mg/l, all other parameters are displayed
in %.
5.3.1.5.1Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E003, is displayed in the message list.
Range: 0.00%–99.90%
5.3.1.5.25Alarm Low: If the measured value falls below the alarm low value,
5.3.1.5.35Hysteresis: Within the hyst. range, the relay does not switch. This
5.3.1.5.45Delay: Duration, the activation of the alarm relay is retarded after
5.3.2 and 5.3.3Relay 1 and 2: The contacts can be set as normally open or nor-
the alarm relay is activated and E004 is displayed in the message
list.
Range: 0.00%–99.90%
prevents damage of relays contacts when the measured value fluctuates around the alarm value.
Range.0.00%–99.90 %
the measuring value has risen above or fallen below the programmed alarm.
Range: 0–28‘800 Sec
mally closed with a jumper. See Relay Contacts 1 and 2, p. 30. The
function of relay contacts 1 or 2 are defined by the user.
Note: The navigation in the menu <Relay 1> and <Relay 2> is
identical. For reason of simplicity only the menu numbers of
Relay 1 are used in the following.
1 First select the functions as:
- Limit upper/lower,
- Control upwards/downwards,
- Timer
- Fieldbus
2 Then enter the necessary data depending on the selected func-
tion.
64A-96.250.321 / 191015
Page 67
AMI Solicon4
Program List and Explanations
5.3.2.1Function = Limit upper/lower:
When the relays are used as upper or lower limit switches, program
the following:
5.3.2.20Parameter: select a process value
5.3.2.300Setpoint: If the measured value rises above respectively falls below
the set-point, the relay is activated.
ParameterRange
Conductivity0 S–300 mS
Temperature- 25 to +270 °C
Sample flow0– 50 l/h
Cond. uc0 S – 300 mS
5.3.2.400Hysteresis: within the hysteresis range, the relay does not switch.
This prevents damage of relay contacts when the measured value
fluctuates around the alarm value.
Parameter Range
Conductivity0 S–300 mS
Temperature- 25 to +270 °C
Sample flow0– 50 l/h
Cond. uc0 S – 300 mS
5.3.2.50Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed
alarm.
Range. 0–600 Sec
A-96.250.321 / 19101565
Page 68
AMI Solicon4
Program List and Explanations
5.3.2.1Function = Control upwards/ downwards:
The relays may be used to drive control units such as solenoid
valves, membrane dosing pumps or motor valves. When driving a
motor valve both relays are needed, relay 1 to open and relay 2 to
close the valve.
5.3.2.22Parameter: Choose on of the following process values.
Conductivity)
Temperature
Sample Flow
Cond. uc
5.3.2.32Settings: Choose the respective actuator:
Time proportional
Frequency
Motor valve
5.3.2.32.1
5.3.2.32.20Cycle time: duration of one control cycle (on/off change).
5.3.2.32.30Response time: Minimal time the metering device needs to react.
5.3.2.32.4Control Parameters
Actuator = Time proportional
Examples of metering devices that are driven time proportional are
solenoid valves, peristaltic pumps.
Dosing is controlled by the operating time.
Range: 0–600 sec.
Range: 0–240 sec.
Range for each Parameter same as 5.2.1.43, p. 61
5.3.2.32.1
5.3.2.32.21Pulse frequency: Max. pulses per minute the device is able to re-
Actuator = Frequency
Examples of metering devices that are pulse frequency driven are
the classic membrane pumps with a potential free triggering input.
Dosing is controlled by the repetition speed of dosing shots.
spond to. Range: 20–300/min.
66A-96.250.321 / 191015
Page 69
AMI Solicon4
Program List and Explanations
5.3.2.32.31Control Parameters
Range for each Parameter same as 5.2.1.43, p. 61
5.3.2.32.1
5.3.2.32.22Run time: Time needed to open a completely closed valve
5.3.2.32.32Neutral zone: Minimal response time in % of the runtime. If the re-