Swann AMI Solicon4 Operator's Manual

Page 1
AMI Solicon4
Version 6.00 and higher
s Manual
Operator’
A-96.250.321 / 191015
Page 2
Customer Support
SWAN ANALYTISCHE INSTRUMENTE AG Studbachstrasse 13 8340 Hinwil Switzerland
Internet: www.swan.ch
E-mail: support@swan.ch
Document Status
Title:
ID:
Monitor AMI Solicon4 Operator’s Manual
A-96.250.321
Revision Issue
00 April 2005 First Edition
01 April 2013 Update to FW-Release 5.30
02 October 2015 Update to FW-Release 6.00
© 2015, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved
subject to change without notice
Page 3
AMI Solicon4
Table of Contents
1. Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1. Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2. General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2. Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1. Description of the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2. Instrument Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3. Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4. Single Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.4.1 AMI Solicon4 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.4.2 Flow Cell M-Flow PG and M-Flow G1 . . . . . . . . . . . . . . . . . . . . . 14
2.4.3 Swansensor Shurecon P . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.4.4 Swansensor Shurecon S . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
2.4.5 Swansensor DeltaT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3. Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.1. Installation Checklist Monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.2. Mounting of Instrument Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.3. Connecting Sample Inlet and Outlet. . . . . . . . . . . . . . . . . . . . . . . 19
3.4. Install Swansensor deltaT (Option) . . . . . . . . . . . . . . . . . . . . . . . 20
3.4.1 Mount the deltaT Sensor and Connect the Sample Inlet . . . . . . . 21
3.4.2 Connect the Sensor Cable to the Transmitter . . . . . . . . . . . . . . . 22
3.4.3 Change Firmware Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.4.4 Adjust the deltaT sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.5. Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.5.1 Connection Diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.5.2 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.6. Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6.2 Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6.3 Relay Contacts 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.7. Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.7.1 Signal Output 1 and 2 (current outputs) . . . . . . . . . . . . . . . . . . . . 32
3.8. Interface Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.8.1 Signal Output 3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.8.2 Profibus, Modbus Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.8.3 USB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
A-96.250.321 / 191015 1
Page 4
AMI Solicon4
4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.1. Establish Sample Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.2. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5. Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.1. Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.2. Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.3. Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.4. Changing Parameters and values. . . . . . . . . . . . . . . . . . . . . . . . . 39
6. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.1. Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.2. Stop of Operation for Maintenance. . . . . . . . . . . . . . . . . . . . . . . . 40
6.3. Maintenance of the Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.3.1 Remove the Sensor form the Flow Cell . . . . . . . . . . . . . . . . . . . . 41
6.3.2 Install the Sensor into the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . 41
6.4. Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.5. Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.6. Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
7. Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
8. Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
8.1. Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
8.2. Diagnostics (Main Menu 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
8.3. Maintenance (Main Menu 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.4. Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.5. Installation (Main Menu 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
9. Program List and Explanations. . . . . . . . . . . . . . . . . . . . . . . . . 53
1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
10. Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
11. Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
12. Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
2 A-96.250.321 / 191015
Page 5
AMI Solicon4
Safety Instructions
AMI Solicon4 - Operator’s Manual
This document describes the main steps for instrument setup, oper­ation and maintenance.

1. Safety Instructions

General The instructions included in this section explain the potential risks
Targ et
audience
OM Location The AMI Operator’s Manual shall be kept in proximity of the instru-
Qualification,
Training
associated with instrument operation and provide important safety practices designed to minimize these risks.
If you carefully follow the information contained in this section, you can protect yourself from hazards and create a safer work environ­ment.
More safety instructions are given throughout this manual, at the respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Operator: Qualified person who uses the equipment for its intended purpose.
Instrument operation requires thorough knowledge of applications, instrument functions and software program as well as all applicable safety rules and regulations.
ment.
To be qualified for instrument installation and operation, you must:
read and understand the instructions in this manual as well as
the Material Safety Data Sheets.
know the relevant safety rules and regulations.
A-96.250.321 / 191015 3
Page 6
AMI Solicon4
Safety Instructions

1.1. Warning Notices

The symbols used for safety-related notices have the following sig­nificance:
DANGER
Your life or physical wellbeing are in serious danger if such warnings are ignored.
Follow the prevention instructions carefully.
WARNING
Severe injuries or damage to the equipment can occur if such warnings are ignored.
Follow the prevention instructions carefully.
CAUTION
Damage to the equipment, minor injury, malfunctions or incor­rect process can be the consequence if such warnings are ig­nored.
Follow the prevention instructions carefully.
Mandatory
Signs
The importance of the mandatory signs in this manual.
Safety goggles
Safety gloves
4 A-96.250.321 / 191015
Page 7
AMI Solicon4
Safety Instructions
Warning Signs The importance of the warning signs in this manual.
Electrical shock hazard
Corrosive
Harmful to health
Flammable
Warning general
Attention general
A-96.250.321 / 191015 5
Page 8
AMI Solicon4
Safety Instructions

1.2. General Safety Regulations

Legal
Requirements
Spare Parts
and
Disposables
Modifications Modifications and instrument upgrades shall only be carried out by
The user is responsible for proper system operation. All precautions must be followed to ensure safe operation of the instrument.
Use only official SWAN spare parts and disposables. If other parts are used during the normal warranty period, the manufacturer’s warranty is voided.
an authorized Service Technician. SWAN will not accept responsi­bility for any claim resulting from unauthorized modification or alter­ation.
WARNING
Risk of Electrical Shock
If proper operation is no longer possible, the instrument must be disconnected from all power lines, and measures must be taken to prevent inadvertent operation.
To prevent from electrical shock, always make sure that the
ground wire is connected.
Service shall be performed by authorized personnel only.Whenever electronic service is required, disconnect instru-
ment power and power of devices connected to. – relay 1, – relay 2, – alarm relay
WARNING
For safe instrument installation and operation you must read and understand the instructions in this manual.
WARNING
Only SWAN trained and authorized personnel shall perform the tasks described in this document.
6 A-96.250.321 / 191015
Page 9
AMI Solicon4
Product Description

2. Product Description

2.1. Description of the System

This instrument is applicable for the measurement of the specific conductivity in surface water, potable water and cooling water. The complete system consists of the Transmitter AMI Solicon4, the con­ductivity sensor Swansensor Shurecon P and the flow cell M-Flow PG. Transmitter, flow cell and sensors are also available as single com­ponents.
Application
Range
Conductivity is a parameter for the total quantity of ions present in the solution. It can be used for the controlling of:
the condition of waterswater purificationwater hardnesscompleteness of ion analysis
Signal
Outputs
Relays Two potential-free contacts programmable as limit switches for
Alarm Relay One potential free contact.
A-96.250.321 / 191015 7
Two signal outputs programmable for measured values (freely scal­able, linear, bilinear, log) or as continuous control output (control parameters programmable).
Current loop: 0/4 – 20 mA Maximal burden: 510 Ohm Third signal output with the same specifications as option.
(Only possible if no communication interface is used.)
measuring values, controllers or timer for system cleaning with au­tomatic hold function. Both contacts can be used as normally open or normally closed. Maximum load: 1 A/250 VAC
Alternatively:
Open during normal operation, closed on error or loss of
power.
Closed during normal operation, open on error or loss of
power.
Summary alarm indication for programmable alarm values and in­strument faults.
Page 10
AMI Solicon4
Product Description
Input For potential-free contact to freeze the signal outputs or to interrupt
control in automated installations (hold function or remote-off).
Communica-
tion interface
(optional)
Safety
Features
Tem per atu re
compensation
Measuring
principle
USB interface to store logger data.RS485 interface with fieldbus protocol Modbus, Profibus DP or
No data loss after power failure. All data is saved in non-volatile memory. Over voltage protection of in- and outputs. Galvanic separation of measuring inputs and signal outputs. The analyzer is factory tested and ready for installation and opera­tion.
The mobility of ions in water increase with higher temperature which enlarges the conductivity. Therefore, the temperature is mea­sured simultaneous by an integrated Pt1000 temperature sensor and the conductivity is compensated to 25 °C.
When a voltage is set between two electrodes in an electrolyte so­lution, the result is an electric field which exerts force on the charged ions: the positively charged cations move towards the neg­ative electrode (cathode) and the negatively charged anions to­wards the positive electrode (anode). The ions, by way of capture or release of electrons at the electrodes, are discharged and so a current I flows through this cycle and the Ohms law V = I×R ap­plies. From the total resistance R of the current loop, only the resis­tance of the electrolyte solution, respectively its conductivity of interest. The cell constant of the sensor is determined by the manufacturer and is printed on the sensor label. If the cell constant has been pro­grammed in the transmitter, the instrument measures correctly. No calibration must be done, the sensor is factory calibrated.
Webserver connection via Modbus.
1
/R, is
8 A-96.250.321 / 191015
Page 11
AMI Solicon4
A
C
B
E
F
D
G
Product Description
Fluidics The flow cell M-Flow PG consists of the flow cell block [B] and the
calibration vessel [D]. The conductivity sensor [A] is fixed in the flow cell block [B] with a
threaded sleeve. The sample enters at the sample inlet [F]. It flows through the del-
taT flow sensor [C] (if installed) and then through the flow cell block into the calibration vessel [D], were the conductivity is measured.
The sample leaves the calibration vessel via flow cell block through the sample outlet [E] and flows into the drain [G].
A
Conductivity sensor
B
Flow cell block
C
deltaT flow sensor (option)
D
Calibration vessel
A-96.250.321 / 191015 9
E
Sample outlet
F
Sample inlet
G
Drain
Page 12
AMI Solicon4
Product Description

2.2. Instrument Specification

Power Supply
Electronics
housing
Sample
requirements
On-site
requirements
Measuring
Range
Accuracy
Voltage:
Power consumption:
Aluminium with a protection degree of IP 66 / NEMA 4X
Ambient temperature: Limit range of operation: Storage and transport: Humidity: Display:
Flow rate: Temperature: Inlet pressure: Outlet pressure:
The analyzer site must permit connections to:
Sample inlet:
Sample outlet:
Max. Altitude:
Range
0.10 to 9.99 S/cm
10.0 to 99.9 S/cm 100 to 999 S/cm
1.00 to 9.99 mS/cm
10.0 to 29.9 mS/cm 30 to 100 mS/cm
± 0.5% of measured value
100–240 VAC (± 10%) 50/60 Hz (± 5%) or 24 VDC (± 10%) max. 30 VA
-10 to +50 °C
-25 to +65 °C
-30 to +85 °C 10–90% rel., non condensing backlit LCD, 75 x 45 mm
4 to 15 l/h up to 50 °C up to 1 bar pressure free
Flow celll with elbow hose nozzle 1/4” diam.10 mm or deltaT sensor with hose nozzle diam. 10 mm 1/2” hose nozzle for flexible tube diam. 20 x 15 mm 2000 m above sea level
Resolution
0.01 S/cm
0.1 S/cm 1 S/cm
0.01 mS/cm
0.1 mS/cm 1 mS/cm
10 A-96.250.321 / 191015
Page 13
AMI Solicon4
AMI Solicon4
850 mm / 33½”
13 mm / ½”
6 x dia. 6.5 mm / ¼”
254 mm/ 10”
280 mm/ 11”
412 mm /16 ¼”
824 mm / 32
7
/
16
Product Description
Dimensions
Panel: Dimensions: Screws: Weight:
white PVC 280x850x 150 mm 5 mm or 6 mm diameter
9.0 kg
A-96.250.321 / 191015 11
Page 14
AMI Solicon4
A
B
C
D
E
F
G
H
I
Product Description

2.3. Instrument Overview

F
Transmitter
A
Panel
B
Conductivity sensor
C
Flow cell block
D
Calibration vessel
E
12 A-96.250.321 / 191015
Sample outlet
G
DeltaT flow sensor (option)
H
Drain
I
Sample inlet
Page 15
AMI Solicon4
Product Description

2.4. Single Components

2.4.1 AMI Solicon4 Transmitter

Electronic transmitter and controller for conductivity measurement.
180
7.5
165
18.5
140 103
4.5
Dimensions
Width: Height: Depth: Weight:
Specifications
Electronics case: Protection degree: Display: Electrical connectors:
A-96.250.321 / 191015 13
140 mm 180 mm 70 mm
1.5 kg
Cast aluminum IP 66 / NEMA 4X backlit LCD, 75 x 45 mm screw clamps
Page 16
AMI Solicon4
60
102
23 23
161
10
33 38
90
142
9
Product Description

2.4.2 Flow Cell M-Flow PG and M-Flow G1

Flow cell M-Flow PG, made of PVC and acrylic glass with one
or Flow cell M-Flow G1, made of PVC and acrylic glass with one
13.5 mm sensor opening for Swansensor Shurecon P.
G 1” sensor opening for Swansensor Shurecon S.
Sample
connections
Inlet: Outlet:
G 1/4” thread G 1/4” thread
Each equipped with an elbow hose nozzle for 10 mm tube
Sample
conditions
Dimensions
Only valid for the flow cell without sensor.
Flow rate: Temperature: Inlet pressure: Sample outlet: Particle size:
4 to 15 l/h up to 50 °C up to 1 bar @ 25 °C Pressure-free outlet below 0.5 mm
No strong acids and bases. No organic solvents.
Width: Front-to-back: Height: Panel mounting:
90 to 142 mm 105 mm 161 mm 3 screws M5
14 A-96.250.321 / 191015
Page 17
AMI Solicon4
16.5
12
46
43
Sensor adapter 16 mm / PG 13.5
PG 13.5
120 ± 2
Product Description

2.4.3 Swansensor Shurecon P

Sensor with integrated cable for the measurement of the specific conductivity. Four electrode design with platinum electrodes and built-in Pt1000 temperature sensor.
Specifications
Measuring range Accuracy:
Temperature sensor: Operating conditions:
0.1S/cm to 100mS / cm ± 1.5 % or ± 0.2 S/cm whichever is greater Pt 1000 > 50°C Max. temperature short-time 90 °C Max.pressure 10 bar at 25 °C.
Electrical connection: Process connection:
A-96.250.321 / 191015 15
Sensor with integrated cable PG 13.5 mm
Page 18
AMI Solicon4
94
25
164
258
G 1” thread
108
Product Description

2.4.4 Swansensor Shurecon S

Sensor for the measurement of the specific conductivity. Four elec­trode design with integrated Pt 1000 temperature sensor. Screw connector with cable must be ordered separately.
Specifications
Measuring range Accuracy:
Temperature sensor: Operating conditions:
0.1S/cm to 100mS / cm ± 1 % or ± 0.2 S/cm whichever is greater Pt 1000 Max. temperature: 120°C at 6.5 bar Max. pressure: 12 bar at 20°C Sterilizable at: 120°C / 5 bar / 30 min. Plug M16 male (IP 67) G 1" thread 20 mm
Electrical connection: Process connection: Space around sensor tip:
16 A-96.250.321 / 191015
Page 19
AMI Solicon4
250
80 70 12
4.5
175
163
Product Description

2.4.5 Swansensor DeltaT

Calorimetric flow meter based on heat dissipation. For applications in potable water, surface water treatment and effluent.
Specifications
Measuring range/Flow rate Accuracy: Response time t
90
:
Sample temperature:
A-96.250.321 / 191015 17
Sample inlet and outlet: Max. cable length:
0–40 l/h ± 20% ca. 1 min 5–35 °C for tubing diam. 10–11 mm 1 m
Page 20
AMI Solicon4
Installation

3. Installation

3.1. Installation Checklist Monitors

Check Instrument’s specification must conform to your AC power rat-
ings. Do not turn on power until instructed to do so.
On site require­ments
Installation Mount the instrument in vertical position. Display should be at
Electrical Wir­ing
Power-up Turn on the sample flow and wait until the flow cell is completely
Instrument set-up
Run-in period Let the instrument run continuously for 1 h.
Calibration Perform a calibration if necessary, see Calibration, p. 42.
100–240 VAC (± 10%), 50/60 Hz (± 5%) or 24 VDC, isolated (±10%) power outlet with ground connection and 30 VA. Sample line with sufficient sample flow and pressure (see Instru-
ment Specification, p. 10
eye level. Connect sample and waste.
Connect all external devices like limit switches, current loops and pumps. Connect power cord; do not switch on power yet!
filled. Switch on power.
Adjust sample flow. Program all parameters for sensor and external devices (inter­face, recorders, etc.). Program all parameters for instrument operation (limits, alarms).
18 A-96.250.321 / 191015
Page 21
AMI Solicon4
AB
Installation

3.2. Mounting of Instrument Panel

To mount the instrument panel proceed according the following list:
The instrument must only be installed by trained personnel. Mount the instrument in vertical position.For ease of operation mount it so that the display is at eye
level.
For the installation a kit containing the following installation
material is available: – 6 Screws 6 x60 mm – 6 Dowels – 6 Washers 6.4/12 mm
Mounting
requirements

3.3. Connecting Sample Inlet and Outlet

Sample inlet Use a plastic tube (FEP, PA, or PE 10x12 mm) to connect the sam-
The instrument is only intended for indoor installation, dimensions see Dimensions, p. 11.
ple inlet at the flow cell.
ABPlastic tube 10x12
Elbow hose nozzle
If a deltaT senor is installed, push the plastic tube over the hose nozzle of the deltaT sensor.
Sample outlet Connect the flexible tube 20x15 mm [C] to the 1/2” hose nozzle [B]
and place it into a drain with atmospheric pressure.
A
A
B
C
A-96.250.321 / 191015 19
Waste funnel
B
1/2” Hose nozzle
C
Flexible tube 20x15 mm
Page 22
AMI Solicon4
A
Installation

3.4. Install Swansensor deltaT (Option)

The following description assumes that the installation of the deltaT sensor takes place after commissioning of the monitor.
Install the deltaT sensor in vertical position with the sample inlet [F] and cable gland looking downwards. To ensure laminar flow, the sample inlet must not be restricted; e.g. any fitting which creates turbulences.
B
C
D
E
F
A
Panel
B
Tube connection
C
Elbow hose nozzle
D
Hose nozzle at deltaT sensor outlet
E
deltaT sensor
F
Hose nozzle at deltaT sensor inlet
20 A-96.250.321 / 191015
Page 23
AMI Solicon4
Installation

3.4.1 Mount the deltaT Sensor and Connect the Sample Inlet

1 Stop operation according to chapter Stop of Operation for Main-
2 Mount the deltaT sensor [E] in vertical position to the panel [A].
3 Remove the plastic tube from the elbow hose nozzle at the flow
4 Connect the plastic tube to the hose nozzle [F] of the deltaT
5 Rotate the elbow hose nozzle of the flow cell by 180 degree.
6 Install the hose connection [B], enclosed in the installation kit,
tenance, p. 40.
cell.
sensor inlet.
from the hose nozzle [D] at the deltaT sensor outlet to the elbow hose nozzle [C].
A-96.250.321 / 191015 21
Page 24
AMI Solicon4
BA
Installation

3.4.2 Connect the Sensor Cable to the Transmitter

Make sure the power specification of the instrument corresponds to the power on site.
WARNING
Risk of electrical shock.
Do not perform any work on electrical components if the trans­mitter is switched on. Failure to follow safety instructions could result in serious injury or death.
Always turn off AC power before manipulating electric parts.Grounding requirements: Only operate the instrument from
an power outlet which has a ground connection.
1 Remove the plug [A] from the cable gland [B]
2 Open the AMI transmitter housing.
3 Feed the sensor cable through the cable gland [B] into the
transmitter housing.
4 Connect the cable to the terminals according to the connecting
diagram see Connection Diagram, p. 27.
5 Close the AMI transmitter housing.
6 Switch on power.
22 A-96.250.321 / 191015
Page 25
AMI Solicon4
5.1.4.1
Flow
Flow measurement deltaT
Q-Flow
Flow measurement
deltaT
none
5.1.4.1
Flow
Flow measurement deltaT
Slope 1.00
5.1.4
Standard 1 1 ppm
Flow
Standard 2 10 ppm
Sensors
Type of sensor NH4
No
Save ?
Yes
Installation

3.4.3 Change Firmware Settings

1 Navigate to <Installation> <Sen-
sors> <Flow> <Flow measure­ment>.
2 Press [Enter]
3 Select <deltaT>
4 Press [Enter]
5 Press 2 x [Exit]
6 Press [Enter] to confirm with Yes
7 Press [Exit] until the display shows
the measuring values.
A-96.250.321 / 191015 23
Page 26
AMI Solicon4
Installation

3.4.4 Adjust the deltaT sensor

The deltaT flow sensor is factory calibrated at 20 °C (± 20 % accu­racy).
The accuracy of the flow measurement depends on the ambient temperature of the location where the deltaT sensor is installed. If the ambient temperature is higher or lower than 20 °C, the deltaT flow sensor can be adjusted. If an adjustment of the deltaT sensor is necessary proceed as follows:
Run in After installation let the sensor run-in for at least 1h.
Determine the
flow rate
Adjust slope 1 Navigate to <Installation/ Sensors / Flow>, choose <Slope> and
1 Put the sample outlet of the instrument for 10 min. into a mea-
suring cup with a sufficient volume.
2 To get the flow rate in l/h, multiply the amount of water con-
tained in the measuring cup with factor 6.
The flow rate in l/h results from the multiplication of the amount of water after 10 min by 6.
press [Enter].
2 If the calculated flow rate is higher than the displayed flow rate
increase the Slope value.
3 If the calculated flow rate is lower than the displayed flow rate
decrease the Slope value.
4 Press [Exit] and save with [Enter].
5 Compare the calculated flow rate with the displayed flow rate.
If the flow rates are roughly equal, the adjustment is finished.
Else repeat step 1 to 5.
24 A-96.250.321 / 191015
Page 27
AMI Solicon4
Installation

3.5. Electrical Connections

WARNING
Risk of electrical shock.
Do not perform any work on electrical components if the trans­mitter is switched on. Failure to follow safety instructions could result in serious injury or death.
Always turn off AC power before manipulating electric parts.Grounding requirements: Only operate the instrument from
an power outlet which has a ground connection.
Make sure the power specification of the instrument corre-
sponds to the power on site.
Cable
thicknesses
Wire For Power and Relays: Use max. 1.5 mm
In order to comply with IP66, use the following cable thicknesses
ABC
A
PG 11 cable gland: cable Ø
B
PG 7 cable gland: cable Ø
C
PG 9 cable gland: cable Ø
outer
3–6.5 mm
outer
4–8 mm
outer
5–10 mm
Note: Protect unused cable glands
2
/ AWG 14
stranded wire with end sleeves.
For Signal Outputs and Input: Use 0.25 mm
2
/ AWG 23
stranded wire with end sleeves.
A-96.250.321 / 191015 25
Page 28
AMI Solicon4
Installation
WARNING
External Voltage.
External supplied devices connected to relay 1 or 2 or to the alarm relay can cause electrical shocks.
Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming in­stallation.
–relay 1 –relay 2 – alarm relay
WARNING
To prevent from electrical shock, do not connect the instrument to the power unless the ground wire (PE) is connected.
Do not connect unless specifically instructed to do so.
WARNING
The mains of the AMI Transmitter must be secured by a main switch and appropriate fuse or circuit breaker.
26 A-96.250.321 / 191015
Page 29
AMI Solicon4
Installation

3.5.1 Connection Diagram

CAUTION
Use only the terminals shown in this diagram, and only for the mentioned purpose. Use of any other terminals will cause short circuits with possible corresponding consequences to material and personnel.
A-96.250.321 / 191015 27
Page 30
AMI Solicon4
A
B
C
D
Installation

3.5.2 Power Supply

WARNING
Risk of electrical shock
Do not perform any work on electrical components if the trans­mitter is switched on. Failure to follow safety instructions could result in serious injury or death.
Always turn off AC power before manipulating electric parts.Installation and maintenance of electrical parts must be per-
formed by professionals
A
Power supply connector
B
Neutral conductor, Terminal 2
C
Phase conductor, Terminal 1
D
Protective earth PE
Note: The protective earth wire (Ground) has to be connected to the grounding terminal.
Installation
requirements
The installation must meet the following requirements.
Transmitter mains fuse 1,6 ATMains cable to comply with standards IEC 60227 or IEC
60245; flammable rating FV1
Mains equipped with an external switch or circuit-breaker
– near the instrument – easily accessible to the operator – marked as interrupter for AMI Solicon4
28 A-96.250.321 / 191015
Page 31
AMI Solicon4
10
12
11
0V
Installation

3.6. Relay Contacts

3.6.1 Input

Terminals 16/42 For programming see Program Overview, p. 48.

3.6.2 Alarm Relay

Alarm output for system errors. Error codes see Error List, p. 45.
Terminals Description Relay connection
1)
NC
Normally Closed
10/11 Active (opened) during normal
Note: Use only potential-free (dry) contacts.
Note: Max. load 1 A/250 VAC
NOTICE: With certain alarms and certain settings of the AMI
transmitter the alarm relay does not switch. The error, however, is shown on the display.
operation. Inactive (closed) on error and loss of power.
1)
0V
11
10
12
NO
Normally Open
12/11 Active (closed) during normal
operation. Inactive (opened) on error and loss of power.
1) usual use
A-96.250.321 / 191015 29
Page 32
AMI Solicon4
6
0V
7
6
0V
7
A
B
Installation

3.6.3 Relay Contacts 1 and 2

Relay 1 and 2 can be configured as normally open or as normally closed. Standard for both relays is normally open. To configure a Relay as normally closed, set the jumper in the upper position.
Note: Rated load 1 AT / 250 VAC
NOTICE: Some error codes and the instrument status may
influence the status of the relays described below.
Relay config. Terminals
Normally Open
Normally Closed
6/7: Relay 1 8/9: Relay 2
6/7: Relay 1 8/9: Relay 2
Jumper pos. Description Relay configuration
Inactive (opened) dur­ing normal operation and loss of power. Active (closed) when a programmed function is executed.
Inactive (closed) during normal operation and loss of power. Active (opened) when a programmed function is executed.
ABJumper set as normally open (standard setting)
Jumper set as normally closed
For programming see Menu Installation Program List and Explana-
tions, p. 53.
30 A-96.250.321 / 191015
Page 33
AMI Solicon4
A
BC
DE
AB
C
Installation
CAUTION
Risk of damage of the relays in the AMI Transmitter due to heavy inductive load.
Heavy inductive or directly controlled loads (solenoid valves, dosing pumps) may destroy the relay contacts.
To switch inductive loads > 0.1 A use an AMI relay box avail-
able as an option or suitable external power relays.
Inductive load Small inductive loads (max 0.1A) as for example the coil of a power
Resistive load Resistive loads (max. 1A) and control signals for PLC, impulse
Actuators Actuators, like motor valves, are using both relays: One relay con-
relay can be switched directly. To avoid noise voltage in the AMI Transmitter it is mandatory to connect a snubber circuit in par­allel to the load.
A
AC or DC power supply
B
AMI Transmitter
C
AMI Relay box
D
Snubber
E
Power relay coil
pumps and so on can be connected without further measures
A
AMI Transmitter
B
PLC or controlled pulse pump
C
Logic
tact is used for opening, the other for closing the valve, i.e. with the 2 relay contacts available, only one motor valve can be controlled. Motors with loads bigger than 0.1A must be controlled via external power relays or an AMI relay box.
A
A
BC
AC or DC power supply
B
AMI Transmitter
C
Actuator
A-96.250.321 / 191015 31
M
Page 34
AMI Solicon4
Installation

3.7. Signal Outputs

3.7.1 Signal Output 1 and 2 (current outputs)

Signal output 1: Terminals 14 (+) and 13 (-) Signal output 2: Terminals 15 (+) and 13 (-)
For programming see Program Overview, p. 48, Menu Installation

3.8. Interface Options

Note: Max. burden 510 If signals are sent to two different receivers, use signal isolator (loop isolator).
A
AMI Transmitter
A
B
Slot for interfaces
C
Frontend PCB
D
Screw terminals
B
C
D
The slot for interfaces can be used to expand the functionality of the AMI instrument with either:
an additional signal outputa Profbus or Modbus connectiona Hyperterminal interfacean USB Interface
32 A-96.250.321 / 191015
Page 35
AMI Solicon4
A
Installation

3.8.1 Signal Output 3

Requires the additional board for the third signal output 0/4 – 20 mA PCB
The third signal output is installed in the upper holder on the main board. You can operate either 3. signal output or communication in­terface, not both! Terminal 38 (+) and 37 (-). .
Note: Max. burden 510
.
A Third signal output 0/4 - 20 mA PCB installed on main board
A-96.250.321 / 191015 33
Page 36
AMI Solicon4
PB-DPV1 V5.5
ON
OFF
81.420.020
Profibus
A
Installation

3.8.2 Profibus, Modbus Interface

Terminal 37 PB, Terminal 38 PA To connect several instruments by means of a network or to config-
ure a PROFIBUS DP or a MODBUS connection, consult the PRO­FIBUS/MODBUS manual. Use appropriate network cable.
Profibus, Modbus Interface PCB (RS 485)
A On - OFF switch
Note: The switch must be ON, if only one instrument is installed, or on the last instrument in the bus.

3.8.3 USB Interface

The USB Interface is used to store Logger data and for Firmware up load. For detailed information see the corresponding installation instruction.
USB Interface
34 A-96.250.321 / 191015
Page 37
AMI Solicon4
Instrument Setup

4. Instrument Setup

4.1. Establish Sample Flow

1 Open sample flow tap.
2 Wait until the flow cell is completely filled.
3 Switch on power.

4.2. Programming

Programming The sensor characteristics are printed on the label of the sensor.
Program all sensor parameters in Menu 5.1.1.1, <Installation> <Sensors> <Sensor parameters>:
SW-xx-xx-xx ZK = 0.0417 Cell constant
SWAN AG Temperature correction
Enter the:
Cell constant [cm-1]  Temperature correction [°C]Cable length
Note: Cable length [m] Set the cable length to 0.0 m if the sensors are installed in the flow cell on the AMI monitor.
Program all parameters for external devices (interface, recorders, etc.). Program all parameters for instrument operation (limits, alarms). See Program Overview, S. 48, for explanations, see Pro-
gram List and Explanations, S. 53.
A-96.250.321 / 191015 35
Page 38
AMI Solicon4
Exit
Enter
BCDA
25.4°C
RUN
9 l/h
14:10:45
R1
525 S
R2
1
Installation
Operation
Diagnostics
Messages
Maintenance
Main Menu
Enter
Exit
Operation

5. Operation

5.1. Keys

A to exit a menu or command (rejecting any changes)
B to move DOWN in a menu list and to decrease digits
C to move UP in a menu list and to increase digits
D to open a selected sub-menu
to move back to the previous menu level
to accept an entry
Program
Access, Exit
36 A-96.250.321 / 191015
Page 39
AMI Solicon4
RUN
15:20:18
R1
R2
9.5 l/h 24.8°C
µS
525
AB D
E
F
H
G
C
Operation

5.2. Display

A RUN normal operation
B ERROR Error
C Keys locked, transmitter control via Profibus
D Time
E Process values
F Sample temperature
G Sample flow
H Relay status
HOLD input closed or cal delay: Instrument on hold (shows
status of signal outputs).
OFF input closed: control/limit is interrupted (shows status
of signal outputs).
Fatal Error
Relay status, symbols
upper/lower limit not yet reached upper/lower limit reached control upw./downw. no action
control upw./downw. active, dark bar indicates control intensity
motor valve closed
A-96.250.321 / 191015 37
motor valve: open, dark bar indicates approx. position timer timer: timing active (hand rotating)
Page 40
AMI Solicon4
1
Messages
Operation
Maintenance
Diagnostics
Main Menu
Installation
1.1
Pending Errors
Messages
Message List
2.1
Interface
I/O State
Sample
Identification Sensors
Diagnostics
3.1
Calibration
Simulation
Maintenance
Set Time 23.09.06 16:30:00
Process Cal.
4.1
Logger
Relay Contacts
Sensors
Operation
5.1
Interface
Miscellaneous
Relay Contacts
Sensors Signal Outputs
Installation
Operation

5.3. Software Structure

Menu Messages 1 Reveals pending errors as well as an event history
(time and state of events that have occurred at an earlier point of time). It contains user relevant data.
Menu Diagnostics 2 Provides user relevant instrument and sample data.
Menu Maintenance 3 For instrument calibration, relay and signal output
simulation, and to set the instrument time. It is used by the service personnel.
Menu Operation 4 User relevant parameters that might need to be
modified during daily routine. Normally password protected and used by the process-operator.
Subset of menu 5 - Installation, but process-related.
Menu Installation 5 For initial instrument set up by SWAN authorized
person, to set all instrument parameters. Can be protected by means of password.
38 A-96.250.321 / 191015
Page 41
AMI Solicon4
5.1.2
Sensors
Sensor type FOME
Temperature NT5K Standards
Disinf. Free chlorine
4.4.1
Logger
Log interval 30 min
Clear logger no
4.1.3
Logger
Clear logger no
Log interval 30min
1 Hour
Interval.
5 min
30 min
10 min
4.1.3
Logger
Log interval 10 min
Clear logger no
4.1.3
Logger
Log interval Clear logger no
No
Save ?
Yes
5.3.1.1.1
Alarm High 300 mS
Alarm Conductivity
Alarm Low 0.00 S Hysteresis 1.00
S
Delay 5 Sec
5.3.1.1.1
Alarm Conductivity
Alarm Low 0.00
S
Hysteresis 1.00
S
Delay 5 Sec
Alarm High 120 mS
Operation

5.4. Changing Parameters and values

Changing
parameters
The following example shows how to change the logger interval:
1 Select the parameter you want to
change.
2 Press [Enter]
3 Press [ ] or [ ] key to
highlight the required parameter.
4 Press [Enter] to confirm the selec-
tion or [Exit] to keep the previous parameter).
The selected parameter is highlighted but not saved yet.
5 Press [Exit].
Yes is highlighted.
6 Press [Enter] to save the new pa-
rameter.
The system reboots, the new parameter is set.
Changing
values
A-96.250.321 / 191015 39
1 Select the value you want to
change.
2 Press [Enter].
3 Set required value with [ ] or
[] key.
4 Press [Enter] to confirm the new
value.
5 Press [Exit].
Yes is highlighted.
6 Press [Enter] to save the new val-
ue.
Page 42
AMI Solicon4
Maintenance

6. Maintenance

6.1. Maintenance Schedule

Preventive maintenance frequency depends on water quality, on the application, and on national regulations.
Control of certain set-point: Swimming pools, sanitary water:
WARNING
Stop operation before maintenance.
Stop sample flow.Shut off power of the instrument.
Monthly
If required
Check sample flow.
Clean conductivity sensorPerform a calibration

6.2. Stop of Operation for Maintenance

Stop sample flow. Shut off power of the instrument.
40 A-96.250.321 / 191015
Page 43
AMI Solicon4
A
C
D
B
E
Maintenance

6.3. Maintenance of the Sensor

A
Fixing sleeve
B
Conductivity sensor
C
Teflon washer
D
O-Ring
E
Flow cell block

6.3.1 Remove the Sensor form the Flow Cell

To remove the sensor form the flow cell proceed as follows:
1 Unscrew and remove the fixing sleeve [A].
2 Pull the conductivity sensor [B] out of the flow cell block [B].
Cleaning If the sensor is contaminated, take a small brush and clean it with
water and detergents. In case of heavy contamination with oil or grease, use ethanol to
clean it. Take a soft tissue and clean the tip of the sensor cautious­ly.
After each cleaning, the sensor must be rinsed with clean water.

6.3.2 Install the Sensor into the Flow Cell

1 Make sure that the washer [C] and the O-ring [D] are in correct
position.
2 Push the sensor through the flow cell bock [E] into the flow cell.
3 Tighten the fixing sleeve [A] to fix the sensor.
A-96.250.321 / 191015 41
Page 44
AMI Solicon4
3.1.5
Calibration
Clean the sensor
and place it in
standard solution
[Enter] to continue
3.1.5
Calibration
Sensor must have a
min. distance of 3 cm
from the beakers edge
[Enter] to continue
3.1.1
Calibration
Standard solution 1.41 mS Current Value 10.07 S Cell constant 0.406 cm
-1
Progress
Maintenance

6.4. Calibration

Since the sensor is very reliable a calibration is normally not neces­sary. A calibration is recommended if:
Reagent for
Calibration:
Calibration solution 1.413 mS/cm (25 °C) 1000 ml. Prepare accord­ing to DIN 38404 / ISO 7888: 1985 / EN 27888.
the
1 Stop the sample flow.
2 Navigate to menu <Maintenance> /<Calibration>.
3 Press [Enter] and follow the dialog on the Display.
4 Remove the sensor from the flow cell.
5 Clean the sensor carefully and rinse it with clean water, see
6 Use a beaker and fill it with one liter calibration solution.
7 Put the sensor into the beaker filled with calibration solution.
the cell constant is not knownthe sensor is contaminated the maintenance measurement shows a discrepancy.
Maintenance of the Sensor, p. 41.
The beakers diameter must be so large that the sensor has a
distance of min. 3 cm from its edge.
8 Wait at least 5 minutes to permit
temperature equilibration between sensor and calibration solution.
9 Start the calibration procedure.
10 Press [Enter], to save the values if
the calibration was successful.
11 Install the sensor into the flow cell.
42 A-96.250.321 / 191015
Page 45
AMI Solicon4
ABCDEF
Maintenance

6.5. Replacing Fuses

When a fuse has blown, find out the cause and fix it before replac­ing it with a new one.
Use tweezers or needle-nosed pliers to remove the defective fuse. Use original fuses provided by SWAN only.
WARNING
External Voltage.
External supplied devices connected to relay 1 or 2 or to the alarm relay can cause electrical shocks.
Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming in­stallation.
– relay 1 – relay 2 – alarm relay
A
1.6 AT/250V Instrument power supply
B
1.0 AT/250V Relay 1
C
1.0 AT/250V Relay 2
D
1.0 AT/250V Alarm relay
E
1.0 AF/125V Signal output 2
F
1.0 AF/125V Signal output 1
A-96.250.321 / 191015 43
Page 46
AMI Solicon4
Maintenance

6.6. Longer Stop of Operation

1 Stop sample flow.
2 Shut off power of the instrument.
3 Remove the sensor form the flow cell and dry it with a soft tis-
4 Empty and dry the flow cell.
sue.
44 A-96.250.321 / 191015
Page 47
AMI Solicon4
25.4°C
HOLD
8 l/h
14:10:45
R1
525 SR2
1
Installation
Operation
Diagnostics
Messages
Maintenance
Main Menu
1.1
Message List
Pending Errors
Messages
1.1.5
Pending Errors
Error Code E002
Alarm low
[Enter] to Acknowledge
Error List

7. Error List

Error
Non-fatal Error. Indicates an alarm if a programmed value is ex­ceeded.
Such Errors are marked E0xx. Fatal Error (blinking symbol) Control of dosing devices is interrupted.
The indicated measured values are possibly incorrect. Fatal Errors are divided in the following two categories:
Errors which disappear if correct measuring conditions are re-
covered (i.e. Sample Flow low). Such Errors are marked E0xx
Errors which indicate a hardware failure of the instrument.
Such Errors are marked E0xx
Error or fatal Error Error not yet acknowledged. Check Pending Errors 1.1.5 * and take corrective action.
Press [ENTER].
A-96.250.321 / 191015 45
Navigate to menu Messages. Press [ENTER].
Navigate to menu Pending Errors. Press [ENTER].
Press [ENTER] to acknowledge the Pending Errors. The Error is reset and saved in the Message List.
Page 48
AMI Solicon4
Error List
Error Description Corrective action
E001 Cond. Alarm high
E002 Cond. Alarm low
E003 Conc. Alarm high
E004 Conc. Alarm low
E007 Sample Temp. high
E008 Sample Temp. low
E009 Sample Flow high
E010 Sample Flow low
E011 Temp. shorted
E012 Temp. disconnected
E013 Case Temp. high
E014 Case Temp. low
– check process – check programmed value, see 5.3.1.1,
p. 62
– check process – check programmed value, see 5.3.1.1,
p. 62
– check process – check programmed value, see 5.3.1.5,
p. 64
– check process – check programmed value, see 5.3.1.5,
p. 64
– check process – check programmed value, see 5.3.1.3,
p. 63
– check process – check programmed value, see 5.3.1.3,
p. 63
– check sample inlet pressure – check programmed value, see
5.3.1.2.2, p. 63
– check sample inlet pressure – Check flow regulating valve – check programmed value, see
5.3.1.2.35, p. 63
– Check wiring of sensor
– Check wiring of sensor
– check case/environment temperature – check programmed value, see
5.3.1.4.1, p. 63
– check case/environment temperature – check programmed value, see
5.3.1.4.2, p. 63
46 A-96.250.321 / 191015
Page 49
AMI Solicon4
Error List
Error Description Corrective action
E017 Control time-out
E018 Temp. out of Table
E019 Conc. out of Table
E024 Input active
E025 IC MK41T56
E026 IC LM75
E027 IC PCF8574
E028 EEProm Microcon
E029 EEProm Motherboard
E030 EEProm Frontend
E031 Cal. Recout
E032 Wrong Frontend
E033 Power-on
E034 Power-down
– Check control device or programming
in Installation, Relay contact, Relay 1/2
5.3.2 and 5.3.3, p. 64
– Check sample temperature
– See If Fault Yes is programmed in
Menu see 5.3.4, p. 69
– call service
– call service
– call service
– call service
– call service
– call service
– call service
– call service
– none, normal status
– none, normal status
A-96.250.321 / 191015 47
Page 50
AMI Solicon4
Program Overview

8. Program Overview

For explanations about each parameter of the menus see Program
List and Explanations, p. 53

8.1. Messages (Main Menu 1)

Menu 1 Messages is always accessible for everybody. No
password protection. No settings can be modified.
Menu 2 Diagnostics is always accessible for everybody. No
password protection. No settings can be modified.
Menu 3 Maintenance is for service: Calibration, simulation of
outputs and set time/date. Please protect with password.
Menu 4 Operation is for the user, allowing to set limits, alarm
values, etc. The presetting is done in the menu Installation (only for the System engineer). Please protect with password.
Menu 5 Installation: Defining assignment of all inputs and
outputs, measuring parameters, interface, passwords, etc. Menu for the system engineer. Password strongly recom­mended.
Pending Errors Pending Errors 1.1.5* * Menu numbers
1.1*
Message List Number 1.2.1*
1.2* Date, Time
48 A-96.250.321 / 191015
Page 51
AMI Solicon4
Program Overview

8.2. Diagnostics (Main Menu 2)

Identification Designation AMI Solicon4 * Menu numbers
2.1* Version V6.00-10/15
Factory Test Instrument 2.1.3.1*
2.1.3* Motherboard
Front End
Operating Time Years / Days / Hours / Minutes / Seconds 2.1.4.1*
2.1.4*
Sensors Cond. Sensor Current value
2.2* 2.2.1* Raw value
Cell constant
Contamination
Cal. History Number, 2.2.1.5.1*
2.2.1.5* Date, Time
Cell Constant
Miscellaneous Case Temp. 2.2.2.1*
2.2.2*
Sample Sample ID 2.3.1*
2.3* Temperature °C
(Pt1000) Ohm
Sample Flow l/h
Raw value Hz if Q-Flow
DeltaT 1 °C It deltaT sensor
DeltaT2 °C It deltaT sensor
I/O State Alarm Relay 2.4.1*
2.4* Relay 1 and 2 2.4.2*
Input
Signal Output 1 and 2
Interface Protocol 2.5.1* (only with RS485
2.5* Baud rate Interface)
A-96.250.321 / 191015 49
Page 52
AMI Solicon4
Program Overview

8.3. Maintenance (Main Menu 3)

Calibration Follow instructions 3.1.5* *Menu numbers
3.1*
Simulation Alarm Relay 3.3.1*
3.2* Relay 1 3.3.2*
Relay 2 3.3.3*
Signal Output 1 3.3.4*
Signal Output 2 3.3.5*
Set Time (Date), (Time)
3.3*

8.4. Operation (Main Menu 4)

Sensors Filter Time Const. 4.1.1*
4.1* Hold after Cal 4.1.2*
Relay Contacts Alarm Relay Alarm Conductivity Alarm High 4.2.1.1.1*
4.2* 4.2.1* 4.2.1.1* Alarm Low 4.2.1.1.25*
Hysteresis 4.2.1.1.35*
Delay 4.2.1.1.45*
If concentration Alarm Concentration Alarm High 4.2.1.2.1*
is chosen 4.2.1.2* Alarm Low 4.2.1.2.25*
Hysteresis 4.2.1.2.35*
Delay 4.2.1.2.45*
Relay 1/ 2 Setpoint 4.2.x.100*
4.2.2*/ 4.2.3* Hysteresis 4.2.x.200*
Delay 4.2.x.30*
Input Active 4.2.4.1*
4.2.4* Signal Outputs 4.2.4.2*
Output / Control 4.2.4.3*
Fault 4.2.4.4*
Delay 4.2.4.5*
Logger Log Interval 4.3.1*
4.3* Clear Logger 4.3.2* * Menu numbers
50 A-96.250.321 / 191015
Page 53
AMI Solicon4
Program Overview

8.5. Installation (Main Menu 5)

Sensors Sensor parameters Cell Constant 5.1.1.1* *Menu numbers
5.1* 5.1.1* Temp. Corr. 5.1.1.2*
Cable length 5.1.1.3*
Temp.Compensation Comp. none
5.1.2* 5.1.2.1* Coefficient
non-linear DIN
Flow Flow measurement none
5.1.3* 5.1.3.1* Q-Flow
deltaT
Conc. none
5.1.4* nitric acid
hydrochloric acid
sodium chloride
caustic soda
sulfuric acid
salinity
TDS as NaCl
TDS
Signal Outputs Signal Output 1/ 2 Parameter 5.2.1.1/5.2.2.1*
5.2* 5.2.1 /5.2.2* Current Loop 5.2.1.2/5.2.2.2*
Function 5.2.1.3/ 5.2.2.3*
Scaling Range Low 5.2.x.40.10/ 11*
5.2.x.40 Range High 5.2.x.40.20 /21*
Relay Contacts Alarm Relay Alarm Conductivity Alarm High 5.3.1.1.1.1*
5.3* 5.3.1* 5.3.1.1* Alarm Low 5.3.1.1.1.25*
Hysteresis * 5.3.1.1.1.35
Delay 5.3.1.1.1.45*
Sample Flow Flow Alarm 5.3.1.2.1*
5.3.1.2* Alarm High 5.3.1.2.2
Alarm Low 5.3.1.2.35
Sample Temp. Alarm High 5.3.1.3.1*
5.3.1.3* Alarm Low 5.3.1.3.25*
Case Temp.high Case Temp. high 5.3.1.4.1*
5.3.1.4* Case Temp. low 5.3.1.4.2*
A-96.250.321 / 191015 51
Page 54
AMI Solicon4
Program Overview
Alarm Concentration Alarm High 5.3.1.1.5.1*
5.3.1.5* Alarm Low 5.3.1.1.5.25*
Hysteresis * 5.3.1.1.5.35
Delay 5.3.1.1.5.45*
Relay 1/ 2 Function 5.3.2.1/*
5.3.2/ 5.3.3* Parameter 5.3.2.20*
Setpoint 5.3.2.300 *
Hysteresis 5.3.2.400*
Delay 5.3.2.50*
Input Active 5.3.4.1*
5.3.4* Signal Outputs 5.3.4.2*
Output/ Control 5.3.4.3*
Fault 5.3.4.4*
Delay 5.3.4.5*
Miscellaneous Language 5.4.1*
5.4* Set defaults 5.4.2*
Load Firmware 5.4.3*
Password Messages 5.4.4.1*
5.4.4* Maintenance 5.4.4.2*
Operation 5.4.4.3*
Installation 5.4.4.4*
Sample ID 5.4.5*
Interface Protocol 5.5.1* (only with RS485
5.5* Device Address 5.5.21* interface)
Baud Rate 5.5.31*
Parity 5.5.41* * Menu numbers
52 A-96.250.321 / 191015
Page 55
AMI Solicon4
Program List and Explanations

9. Program List and Explanations

1 Messages

1.1 Pending Errors
1.1.5 Provides the list of active errors with their status (active, acknowl-
1.2 Message List
1.2.1 Shows the error history: Error code, date / time of issue and status

2 Diagnostics

2.1 Identification
2.1.3 Factory Test: Test date of the Instrument, Motherboard and Fron-
2.1.4 Operating Time: Years / Days / Hours / Minutes / Seconds
2.2 Sensors
2.2.1 Cond. Sensor
2.2.1.5 Cal. History: In this menu the calibration values of the last
edged). If an active error is acknowledged, the alarm relay is active again. Cleared errors are moved to the Message list.
(active, acknowledged, cleared). 65 errors are memorized. Then the oldest error is cleared to save the newest error (circular buffer).
In diagnostics mode, the values can only be viewed, not modified.
Designation: Designation of the instrument. Version: Firmware of instrument (e.g. V6.00-10 / 15)
tend
Current value in µS Raw value in µS Cell Constant Contamination
calibrations are saved.
Number: Numbers the calibrations in descending order. Date, Time: Date and time of each calibration. Cell constant: Shows the cell constant of the sensor in use.
Max. 64 data records are memorized. One process calibration cor­responds to one data record.
A-96.250.321 / 191015 53
Page 56
AMI Solicon4
Program List and Explanations
2.2.2 Miscellaneous:
2.2.2.1 Case Temp: Shows the current temperature in °C inside the trans-
2.3 Sample
2.3.1 Sample ID: Shows the identification assigned to a sample. This
2.4 I/O State
2.4.1/2.4.2
mitter.
identification is defined by the user to identify the location of the sample.
Temperature: Shows the current sample temperature in °C. (Pt 1000): Shows the current sample temperature in Ohm. Sample Flow: If Q-Flow is chosen
Shows the current sample flow in l/h
Raw Value: shows the sample flow in Hz.
Sample Flow: If deltaT is chosen
Shows the current sample flow in l/h deltaT 1: Temperature measured at sample inlet of the deltaT sensor deltaT 2: Temperature measured at sample outlet of the deltaT sensor
Shows current status of all in- and outputs.
Alarm Relay: Inactive or active
Relay 1 and 2: Inactive or active
Input: Open or closed
Signal Output 1 and 2: Actual current in mA
Signal Output 3 (option): Actual current in mA
2.5 Interface
Only available if optional interface is installed. Review programmed communication settings.
54 A-96.250.321 / 191015
Page 57
AMI Solicon4
Program List and Explanations

3 Maintenance

3.1 Calibration
Follow the commands on the screen. Save the value with the [En­ter] key.
3.2 Simulation
To simulate a value or a relay state, select the
alarm relay,relay 1or 2 signal output 1or 2
with the [ ] or [ ] key. Press the [Enter] key. Change the value or state of the selected item with the [ ] or
[ ] key. Press the [Enter] key.
The value is simulated by the relay / signal output.
Alarm Relay: Inactive or active
Relay 1 and 2: Inactive or active
Signal Output 1 and 2: Actual current in mA
Signal Output 3 (option): Actual current in mA
At the absence of any key activities, the instrument will switch back to normal mode after 20 min. If you quit the menu, all simulated val­ues will be reset.
3.3 Set Time
Adjust date and time.
A-96.250.321 / 191015 55
Page 58
AMI Solicon4
Program List and Explanations

4 Operation

4.1 Sensors
4.1.1 Filter Time Constant: Used to damp noisy signals. The higher the
4.1.2 Hold after Cal.: Delay permitting the instrument to stabilize again af-
4.2 Relay Contacts
4.3 Logger
4.3.1 Log Interval: Select a convenient log interval. Consult the table be-
filter time constant, the slower the system reacts to changes of the measured value. Range: 5–300 Sec
ter calibration. During calibration plus hold-time, the signal outputs are frozen (held on last valid value), alarm values, limits are not ac­tive.
Range: 5–6‘000 Sec
See Relay Contacts, p. 29
The instrument is equipped with an internal logger. The data can be downloaded to a PC by SWAN Terminal if option “SWAN Terminal interface” is installed or via USB stick if option “USB interface” is in­stalled. The logger can save approx. 1500 data records. The Records con­sists of: Date, time, alarms, measured value, measured value un­compensated, temperature, flow. Range: 1 Second to 1 hour
low to estimate the max logging time. When the logging buffer is full, the oldest data record is erased to make room for the newest one (circular buffer).
Interval 1 s 5 s 1 min 5 min 10 min 30 min 1 h
Time 25 min 2 h 25 h 5 d 10 d 31 d 62 d
4.3.2 Clear Logger: If confirmed with yes, the complete logger data is de-
leted. A new data series is started.
4.3.3 If option USB interface is installed.
Eject USB Stick: With this function all logger data are copied to the USB stick before the USB stick is deactivated. Only visible it the optional USB interface is installed.
56 A-96.250.321 / 191015
Page 59
AMI Solicon4
Program List and Explanations

5 Installation

5.1 Sensors
5.1.1 Sensor parameters
5.1.1.1 Cell Constant: Enter the cell constant printed on the sensor label.
5.1.1.2 Temp. Corr: Enter the temperature correction printed on the sensor
label.
5.1.1.3 Cable length: Enter the cable length. Set the cable length to 0.0 m if
the sensors are installed in the flow cell on the AMI monitor.
5.1.2 Temp. Compensation
5.1.2.1 Comp.: Available compensation models are:
noneCoefficientnon-linear DIN
5.1.3 Flow
5.1.3.1 Flow measurement: Select the type of flow sensor if a flow sensor
is installed. Possible flow sensors: None; Q-Flow; deltaT.
5.1.3.2 Slope: If flow measurement is set to deltaT.
The slope value is used to adjust the flow measurement of the deltaT sensor if the ambient temperature is higher or lower than 20 °C.
5.1.4 Conc.
The menu <Concentration> (Conc.) allows the additional measure­ment of a known substance in the sample. The concentration of the substance is calculated based on the conductivity of any of the fol­lowing substances. The calculated value is displayed in %. As an exception, TDS is displayed in mg/l.
nonenitric acidhydrochloric acidsodium chloridecaustic sodasulfuric acidsalinityTDS as NaClTDS
A-96.250.321 / 191015 57
Page 60
AMI Solicon4
Program List and Explanations
5.2 Signal Outputs
5.2.1 and 5.2.2 Signal Output 1 and 2: Assign process value, the current loop
range and a function to each signal output.
Note: The navigation in the menu <Signal Output 1> and <Signal Output 2> is identical. For reason of simplicity only the menu numbers of Signal Output 1 are used in the following.
5.2.1.1 Parameter: Assign one of the process values to the signal output.
Available values:
ConductivityTemperatureSample flowCond. uc (uncompensated)
5.2.1.2 Current Loop: Select the current range of the signal output.
Make sure the connected device works with the same current range. Available ranges: 0–20 mA or 4–20 mA
5.2.1.3 Function: Define if the signal output is used to transmit a process
value or to drive a control unit. Available functions are:
Linear, bilinear or logarithmic for process values.
See As process values, p. 58
Control upwards or control downwards for controllers.
See As control output, p. 60
As process
values
The process value can be represented in 3 ways: linear, bilinear or logarithmic. See graphs below.
[mA]
20
(0 - 20 [mA])
(4 - 20 [mA])
10 12
0 / 4
0.0 0.1 0.2 0.3 0.4 0.5
ABlinear
AB
X
X Measured value
bilinear
58 A-96.250.321 / 191015
Page 61
AMI Solicon4
20
1 01234
10 100 1’000 10’000
10 12
(0 - 20 [mA])
0 / 4
426
(4 - 20 [mA])
[mA]
X
Program List and Explanations
X Measured value (logarithmic)
5.2.1.40 Scaling: Enter beginning and end point (Range low & high) of the linear or logarithmic scale. In addition, the midpoint for the bilinear scale.
Parameter Conductivity:
5.2.1.40.10 Range low: 0 S–300 mS
5.2.1.40.20 Range high: 0 S–300 mS
5.2.1.40.11 Range low: -25 to + 270 °C
5.2.1.40.21 Range high: -25 to +270 °C
5.2.1.40.12 Range low: 0 –50 l / h
5.2.1.40.22 Range high: 0 –50 l/ h
5.2.1.40.13 Range low: 0 S–300 mS
5.2.1.40.23 Range high: 0 S–300 mS
A-96.250.321 / 191015 59
Parameter Temperature
Parameter Sample flow
Parameter Cond. uc:
Page 62
AMI Solicon4
Program List and Explanations
As control
output
Signal outputs can be used for driving control units. We distinguish different kinds of controls:
P-controller: The controller action is proportional to the devia-
tion from the setpoint. The controller is characterized by the P-Band. In the steady-state, the setpoint will never be reached. The deviation is called steady-state error. Parameters: setpoint, P-Band
PI-controller: The combination of a P-controller with an
I-controller will minimize the steady-state error. If the reset time is set to zero, the I-controller is switched off. Parameters: setpoint, P-Band, reset time.
PD-controller: The combination of a P-controller with a
D-controller will minimize the response time to a fast change of the process value. If the derivative time is set to zero, the D-controller is switched off. Parameters: setpoint, P-Band, derivative time.
PID-controller: The combination of a P-, an I - and a D-con-
troller allows a proper control of the process.
Parameters: setpoint, P-Band, reset time, derivative time. Ziegler-Nichols method for the optimization of a PID controller: Parameters: Setpoint, P-Band, Reset time, Derivative time
Y
B
A
X
a
L
A
Response to maximum control output
B
Tangent on the inflection point
X
Time
Xp Tn Tv
= 1.2/a = 2L = L/2
The point of intersection of the tangent with the respective axis will result in the parameters a and L.
Consult the manual of the control unit for connecting and program­ming details. Choose control upwards or downwards.
60 A-96.250.321 / 191015
Page 63
AMI Solicon4
Program List and Explanations
Control upwards or downwards
Setpoint: User-defined process value for the selected parameter. P-Band: Range below (upwards control) or above (downwards con-
trol) the set-point, within the dosing intensity is reduced from 100% to 0% to reach the setpoint without overshooting.
5.2.1.43 Control Parameters: if Parameters = Conductivity
5.2.1.43.10 Setpoint Range: 0 S–300 mS
5.2.1.43.20 P-Band: Range: 0 S–300 mS
5.2.1.43 Control Parameters: if Parameters = Temperature
5.2.1.43.11 Setpoint Range: -25 to +270 °C
5.2.1.43.21 P-Band:
5.2.1.43 Control Parameters: if Parameters = Sample flow
5.2.1.43.12 Setpoint
5.2.1.43.22 P-Band:
5.2.1.43 Control Parameters: if Parameters = Cond. uc.
5.2.1.43.13 Setpoint
5.2.1.43.23 P-Band:
5.2.1.43.3 Reset time: The reset time is the time till the step response of a sin-
5.2.1.43.4 Derivative time: The derivative time is the time till the ramp re-
5.2.1.43.5 Control timeout: If a controller action (dosing intensity) is constantly
Range: -25 to +270 °C
Range: 0 –50 l/h
Range: 0 –50 l/h
Range: 0 S–300 mS
Range: 0 S–300 mS
gle I-controller will reach the same value as it will be suddenly reached by a P-controller. Range: 0–9’000 Sec
sponse of a single P-controller will reach the same value as it will be suddenly reached by a D-controller. Range: 0–9’000 Sec
over 90% during a defined period of time and the process value does not come closer to the setpoint, the dosing process will be stopped for safety reasons. Range: 0–720 min
A-96.250.321 / 191015 61
Page 64
AMI Solicon4
Program List and Explanations
5.3 Relay Contacts
5.3.1 Alarm Relay: The alarm relay is used as cumulative error indicator.
Under normal operating conditions the contact is active. The contact is inactive at:
Power lossDetection of system faults like defective sensors or electronic
parts
High case temperatureProcess values out of programmed ranges.
Note: Above definition assumes that the alarm relay is used as normally closed (Terminals 10/11). If terminals 12/11 are used, the alarm relay is normally open, then the above definition is inverted. See also Alarm Relay, p. 29 and Connection Diagram,
p. 27
Program alarm levels, hysteresis values and delay times for the fol­lowing parameters:
Alarm ConductivitySample FlowSample Temp.Case Temp.Alarm Concentration (visible if a Conc. parameter has been
selected)
5.3.1.1 Alarm Conductivity
5.3.1.1.1 Alarm High: If the measured value rises above the alarm high val-
5.3.1.1.25 Alarm Low: If the measured value falls below the alarm low value,
5.3.1.1.35 Hysteresis: Within the hyst. range, the relay does not switch. This
5.3.1.1.45 Delay: Duration, the activation of the alarm relay is retarded after
ue, the alarm relay is activated and E001, is displayed in the mes­sage list. Range: 0 S–300 mS
the alarm relay is activated and E002 is displayed in the message list. Range: 0 S–300 mS
prevents damage of relays contacts when the measured value fluc­tuates around the alarm value. Range. 0 S–300 mS
the measuring value has risen above/fallen below the programmed alarm. Range: 0–28‘800 Sec
62 A-96.250.321 / 191015
Page 65
AMI Solicon4
Program List and Explanations
5.3.1.2 Sample Flow: Define at which sample flow an alarm should be is­sued.
5.3.1.2.1 Flow Alarm: Program if the alarm relay should be activated if there
is a flow alarm. Choose between yes or no. The flow alarm will always be indicated in the display, pending error list, saved in the message list and the logger. Available values: Yes or no
Note: Sufficient flow is essential for a correct measurement. We recommend to program yes.
5.3.1.2.2 Alarm High: If the measuring values rises above the programmed
5.3.1.2.35 Alarm Low: If the measuring values falls below the programmed
5.3.1.3 Sample Temp.
5.3.1.3.1 Alarm High: If the measured value rises above the alarm high val-
5.3.1.3.25 Alarm Low: If the measured value falls below the alarm low value,
5.3.1.4 Case Temp.
5.3.1.4.1 Case Temp. high: Set the alarm high value for temperature of elec-
5.3.1.4.2 Case Temp. low: Set the alarm low value for temperature of elec-
value E009 will be issued. Range: 9–20 l/h
value E010 will be issued. Range: 5–8 l/ h
ue, the alarm relay is activated and E007, is displayed in the mes­sage list. Range: 30–200 °C
the alarm relay is activated and E008 is displayed in the message list. Range: -10 to + 20 °C
tronics housing. If the value rises above the programmed value E013 is issued. Range: 30–75 °C
tronics housing. If the value falls below the programmed value E014 is issued. Range: -10 to +20 °C
A-96.250.321 / 191015 63
Page 66
AMI Solicon4
Program List and Explanations
5.3.1.5 Alarm Concentration: Visible if a Conc. parameter has been se­lected. TDS is displayed in mg/l, all other parameters are displayed in %.
5.3.1.5.1 Alarm High: If the measured value rises above the alarm high val-
ue, the alarm relay is activated and E003, is displayed in the mes­sage list. Range: 0.00%–99.90%
5.3.1.5.25 Alarm Low: If the measured value falls below the alarm low value,
5.3.1.5.35 Hysteresis: Within the hyst. range, the relay does not switch. This
5.3.1.5.45 Delay: Duration, the activation of the alarm relay is retarded after
5.3.2 and 5.3.3 Relay 1 and 2: The contacts can be set as normally open or nor-
the alarm relay is activated and E004 is displayed in the message list. Range: 0.00%–99.90%
prevents damage of relays contacts when the measured value fluc­tuates around the alarm value. Range.0.00%–99.90 %
the measuring value has risen above or fallen below the pro­grammed alarm. Range: 0–28‘800 Sec
mally closed with a jumper. See Relay Contacts 1 and 2, p. 30. The function of relay contacts 1 or 2 are defined by the user.
Note: The navigation in the menu <Relay 1> and <Relay 2> is identical. For reason of simplicity only the menu numbers of Relay 1 are used in the following.
1 First select the functions as:
- Limit upper/lower,
- Control upwards/downwards,
- Timer
- Fieldbus
2 Then enter the necessary data depending on the selected func-
tion.
64 A-96.250.321 / 191015
Page 67
AMI Solicon4
Program List and Explanations
5.3.2.1 Function = Limit upper/lower:
When the relays are used as upper or lower limit switches, program the following:
5.3.2.20 Parameter: select a process value
5.3.2.300 Setpoint: If the measured value rises above respectively falls below
the set-point, the relay is activated.
Parameter Range
Conductivity 0 S–300 mS
Temperature - 25 to +270 °C
Sample flow 0– 50 l/h
Cond. uc 0 S – 300 mS
5.3.2.400 Hysteresis: within the hysteresis range, the relay does not switch.
This prevents damage of relay contacts when the measured value fluctuates around the alarm value.
Parameter Range
Conductivity 0 S–300 mS
Temperature - 25 to +270 °C
Sample flow 0– 50 l/h
Cond. uc 0 S – 300 mS
5.3.2.50 Delay: Duration, the activation of the alarm relay is retarded after
the measuring value has risen above/fallen below the programmed alarm. Range. 0–600 Sec
A-96.250.321 / 191015 65
Page 68
AMI Solicon4
Program List and Explanations
5.3.2.1 Function = Control upwards/ downwards:
The relays may be used to drive control units such as solenoid valves, membrane dosing pumps or motor valves. When driving a motor valve both relays are needed, relay 1 to open and relay 2 to close the valve.
5.3.2.22 Parameter: Choose on of the following process values.
Conductivity)TemperatureSample FlowCond. uc
5.3.2.32 Settings: Choose the respective actuator:
Time proportionalFrequencyMotor valve
5.3.2.32.1
5.3.2.32.20 Cycle time: duration of one control cycle (on/off change).
5.3.2.32.30 Response time: Minimal time the metering device needs to react.
5.3.2.32.4 Control Parameters
Actuator = Time proportional
Examples of metering devices that are driven time proportional are solenoid valves, peristaltic pumps.
Dosing is controlled by the operating time.
Range: 0–600 sec.
Range: 0–240 sec.
Range for each Parameter same as 5.2.1.43, p. 61
5.3.2.32.1
5.3.2.32.21 Pulse frequency: Max. pulses per minute the device is able to re-
Actuator = Frequency
Examples of metering devices that are pulse frequency driven are the classic membrane pumps with a potential free triggering input. Dosing is controlled by the repetition speed of dosing shots.
spond to. Range: 20–300/min.
66 A-96.250.321 / 191015
Page 69
AMI Solicon4
Program List and Explanations
5.3.2.32.31 Control Parameters
Range for each Parameter same as 5.2.1.43, p. 61
5.3.2.32.1
5.3.2.32.22 Run time: Time needed to open a completely closed valve
5.3.2.32.32 Neutral zone: Minimal response time in % of the runtime. If the re-
5.3.2.32.4 Control Parameters
5.3.2.1 Function = Timer:
5.3.2.24 Mode: Operating mode (interval, daily, weekly)
5.3.2.24
5.3.2.340 Interval: The interval can be programmed within a range
5.3.2.44 Run Time: Enter the time the relay stays active.
5.3.2.54 Delay: during run time plus the delay time the signal and control
5.3.2.6 Signal Outputs: Select operating mode of the signal output:
Actuator = Motor valve
Dosing is controlled by the position of a motor driven mixing valve.
Range: 5–300 Sec.
quested dosing output is smaller than the response time, no change will take place. Range: 1–20 %
Range for each Parameter same as 5.2.1.43, p. 61
The relay will be activated repetitively depending on the pro­grammed time scheme.
Interval
of 1–1440 min.
Range: 5–32400 sec.
outputs are held in the operating mode programmed below. Range: 0–6’000 Sec.
Cont.: Signal outputs continue to issue the measured value. Hold: Signal outputs hold the last valid measured value.
Measurement is interrupted. Errors, except fatal errors, are not issued.
Off: Signal outputs are switched off (set to 0 or 4 mA).
Errors, except fatal errors, are not issued.
A-96.250.321 / 191015 67
Page 70
AMI Solicon4
Program List and Explanations
5.3.2.7 Output/ Control: Select operating mode of the controller output:
Cont.: Controller continues normally.
Hold: Controller continues based on the last valid value.
Off: Controller is switched off.
5.3.2.24
5.3.2.341 Start time: to set the start time proceed as follows:
5.3.2.44 Run Time: see Interval
5.3.2.54 Delay: see Interval
5.3.2.6 Signal Outputs: see Interval
5.3.2.7 Output/ Control: see Interval
5.3.2.24
5.3.2.342 Calendar:
5.3.2.342.1 Start time: The programmed start time is valid for each of the pro-
5.3.2.342.2 Monday: Possible settings, on or off
5.3.2.342.8 Sunday: Possible settings, on or off
5.3.2.44 Run Time: see Interval
5.3.2.54 Delay: see Interval
5.3.2.6 Signal Outputs: see Interval
5.3.2.7 Output/ Control: see Interval
daily
The relay contact can be activated daily, at any time of a day.
1 Press [Enter], to set the hours.
2 Set the hour with the [ ] or [ ] keys.
3 Press [Enter], to set the minutes.
4 Set the minutes with the [ ] or [ ] keys.
5 Press [Enter], to set the seconds.
6 Set the seconds with the [ ] or [ ] keys.
Range: 00:00:00–23:59:59
weekly
The relay contact can be activated at one or several days, of a week. The daily starting time is valid for all days.
grammed days. To set the start time see 5.3.2.341, p. 68. Range: 00:00:00–23:59:59
to
68 A-96.250.321 / 191015
Page 71
AMI Solicon4
Program List and Explanations
5.3.2.1
5.3.4 Input: The functions of the relays and signal outputs can be de-
5.3.4.1 Active: Define when the input should be active:
5.3.4.2 Signal Outputs: Select the operation mode of the signal outputs
5.3.4.3 Output/Control: (relay or signal output):
Function = Fieldbus:
The relay will be switched via the Profibus input. No further param­eters are needed.
fined depending on the position of the input contact, i.e. no function, closed or open.
No: Input is never active.
When closed Input is active if the input relay is closed
When open: Input is active if the input relay is open
when the relay is active:
Cont.: Signal outputs continue to issue the measured
value.
Hold: Signal outputs issue the last valid measured value.
Measurement is interrupted. Errors, except fatal errors, are not issued.
Off: Set to 0 or 4 mA respectively. Errors, except fatal
errors, are not issued.
Cont.: Controller continues normally.
Hold: Controller continues on the last valid value.
Off: Controller is switched off.
5.3.4.4 Fault:
No: No message is issued in pending error list and the
alarm relay does not close when input is active.
Yes: Message E024 is issued and stored in the mes-
sage list. The Alarm relay closes when input is active.
5.3.4.5 Delay: Time which the instrument waits, after the input is deactivat-
ed, before returning to normal operation. Range: 0–6‘000 Sec
A-96.250.321 / 191015 69
Page 72
AMI Solicon4
Program List and Explanations
5.4 Miscellaneous
5.4.1 Language: Set the desired language.
Language German English French Spanish
5.4.2 Set defaults: Reset the instrument to factory default values in three
5.4.3 Load Firmware: Firmware updates should be done by instructed
different ways:
Set defaults no Calibration In parts Completely
Calibration: Sets calibration values back to default. All other
values are kept in memory.
In parts: Communication parameters are kept in memory. All
other values are set back to default values.
Completely: Sets back all values including communication
parameters.
service personnel only.
Load Firmware no yes
5.4.4 Password: Select a password different from 0000 to prevent unau-
5.4.4.1 Messages
5.4.4.2 Maintenance
5.4.4.3 Operation
5.4.4.4 Installation.
5.4.5 Sample ID: Identify the process value with any meaning full text,
thorized access to the following menus:
Each menu may be protected by a different password. If you forgot the passwords, contact the closest SWAN representative.
such as KKS number.
70 A-96.250.321 / 191015
Page 73
AMI Solicon4
Program List and Explanations
5.5 Interface
Select one of the following communication protocols. Depending on your selection, different parameters must be defined.
5.5.1 Protocol: Profibus
5.5.20 Device address: Range: 0–126
5.5.30 ID-Nr.: Range: Analyzer; Manufacturer; Multivariable
5.5.40 Local operation: Range: Enabled, Disabled
5.5.1 Protocol: Modbus RTU
5.5.21 Device address: Range: 0–126
5.5.31 Baud Rate: Range: 2
5.5.41 Parity: Range: none, even, odd
5.5.1 Protocol: Hyper Terminal
5.5.23 Baudrate: Range: 2 see separate RS232 Operation Manual for Logger download.
5.5.1 Protocol: USB-Stick:
Only visible if an USB interface is installed. No further settings are possible.
400–115 200 Baud
400–115 200 Baud
A-96.250.321 / 191015 71
Page 74
AMI Solicon4
Default Values

10. Default Values

Operation:
Sensors: Filter Time Const.:..................................................................10 Sec
Hold after Cal.:..................................................................... 300 sec
Relay Contacts Alarm Relay .................................................. same as in Installation
Relay 1/2 ...................................................... same as in Installation
Input.............................................................. same as in Installation
Logger: Logger Interval:.............................................................. 30 Minutes
Clear Logger: ................................................................................no
Installation:
Sensors Sensor Parameters; Cell Constant ...............................0.4000 cm
Sensor Parameters; Temp. corr. ...........................................0.00 °C
Sensor Parameters; Cable length............................................ 0.0 m
Temp. Compensation; Comp. ...................................................none
Flow; Flow measurement..........................................................none
Conc.: .......................................................................................none
Signal Output 1 Parameter:.................................................................... Conductivity
Current loop: ......................................................................4 – 20 mA
Function: ..................................................................................linear
Scaling: Range low: ........................................................... 0.000 µS
Scaling: Range high:............................................................ 100 mS
Signal Output 2 Parameter:....................................................................Temperature
Current loop: .....................................................................4 – 20 mA
Function: ..................................................................................linear
Scaling: Range low: ................................................................0.0 °C
Scaling: Range high:.............................................................50.0 °C
Alarm Relay: Alarm Conductivity:
Alarm high:........................................................................... 300 mS
Alarm low: ............................................................................ 0.00 µS
Hysteresis: .......................................................................... 1.00 mS
Delay:.......................................................................................5 Sec
Sample Flow (if Flow measurement = yes):
Flow Alarm.................................................................................. yes
Alarm high:........................................................................... 50.0 l/h
Alarm low: .............................................................................. 5.0 l/h
Sample Temp:
Alarm High: ............................................................................120 °C
Alarm Low:.................................................................................0 °C
-1
72 A-96.250.321 / 191015
Page 75
AMI Solicon4
Default Values
Case Temp. high:..................................................................... 65 °C
Case Temp. low: ........................................................................ 0 °C
Relay 1 and 2 Function:.......................................................................... limit upper
Parameter:.................................................................... Conductivity
Setpoint: ...............................................................................100 mS
Hysteresis:................................................................................1 mS
Delay: ....................................................................................30 Sec
If Function = Control upw. or dnw:
Parameter:.................................................................... Conductivity
Settings: Actuator: ..........................................................Frequency
Settings: Pulse Frequency: .............................................120 /min.
Settings: Control Parameters: Setpoint: ............................100 mS
Settings: Control Parameters: P-band: .................................1 mS
Settings: Control Parameters: Reset time: ...........................0 Sec
Settings: Control Parameters: Derivative Time: ...................0 Sec
Settings: Control Parameters: Control Timeout:................... 0 Min
Settings: Actuator: ................................................ Time proportional
Cycle time: ............................................................................ 60 s
Response time: ..................................................................... 10 s
Settings: Actuator ......................................................... Motor valve
Run time: ............................................................................... 60 s
Neutral zone: ...........................................................................5%
If Function = Timer:
Mode:.................................................................................... Interval
Interval: ................................................................................ 1 min
Mode: ........................................................................................daily
Start time:........................................................................ 00.00.00
Mode:..................................................................................... weekly
Calendar; Start time: ....................................................... 00.00.00
Calendar; Monday to Sunday:.................................................. Off
Run time: ...............................................................................10 Sec
Delay: ......................................................................................5 Sec
Signal output:............................................................................. cont
Output/Control: .......................................................................... cont
Input: Active............................................................................when closed
Signal Outputs........................................................................... hold
Output/Control .............................................................................. off
Fault..............................................................................................no
Delay .....................................................................................10 Sec
A-96.250.321 / 191015 73
Page 76
AMI Solicon4
Default Values
Miscellaneous Language:............................................................................. English
Set default:....................................................................................no
Load firmware: ..............................................................................no
Password: ........................................................... for all modes 0000
Sample ID: ....................................................................... - - - - - - - -
Interface Protocol:..................................................................... Hyperterminal
74 A-96.250.321 / 191015
Page 77
AMI Solicon4
Index

11. Index

A
Actuators . . . . . . . . . . . 31
Alarm Relay Application Range
. . . . . . . . 7, 29
. . . . . . . 7
C
Cable thicknesses . . . . . . 25
Calendar . . . . . . . . . . . 68
Calibration solution Cell constant
cell constant . . . . . . . . . . 8
Changing values
Checklist . . . . . . . . . . . 18
Cleaning
. . . . . . . . . . 41
Sensor
Concentration . . . . . . . . 57
. . . . . 42
. . . . . . . . . 35
. . . . . . . 39
D
Default Values . . . . . . . . 72
E
Error List . . . . . . . . . . . 45
F
Fluidics . . . . . . . . . . . . . 9
I
Inductive load . . . . . . . . 31
. . . . . . . . . . . . . 8, 29
Input
Instrument Overview . . . . 12
Interface . . . . . . . . . . . . 8
Interface RS485
. . . . . . . 34
M
Measuring principle . . . . . . 8
Measuring Range . . . . . . 10
Mounting
. . . . . . . . . . . 19
O
On-site requirements . . . . 10
P
Power Supply . . . . . . 10, 28
Programming
. . . . . . . . . 35
R
Relay Contacts 1 and 2 . . . 30
Relays. . . . . . . . . . . . . . 7
Resistive load . . . . . . . . 31
S
Safety Features . . . . . . . . 8
Sample flow, establish
Sample requirements . . . . 10
Sensor parameters . . . . . 57
Set Time
Setup . . . . . . . . . . . . . 35
Signal Output . . . . . . . . 58
Signal Outputs . . . . . . . 7, 32
Simulation . . . . . . . . . . 55
Software Specifications
. . . . . . . . . . . 55
Current loop
. . . . . . . . . . . 38
Swansensor deltaT
. . . 35
. . . . . . . 58
. . 17
T
Technical Data . . . . . . . . 12
Temperature compensation . 8
Terminals . . . . 27, 2930, 34
W
Wire . . . . . . . . . . . . . . 25
A-96.250.321 / 191015 75
Page 78
AMI Solicon4
Notes

12. Notes

76 A-96.250.321 / 191015
Page 79
AMI Solicon4
Notes
A-96.250.321 / 191015 77
Page 80
AMI Solicon4
SWAN
is represented worldwide by subsidiary companies and distributors.
cooperates with independent representatives all over the world.
SWAN Products
Analytical Instruments for:
High Purity Water
Feedwater, Steam and Condensate
Potable Water
Pool and Sanitary Water
Cooling Water
Waster Water and Effluents
Made in Switzerland
78 A-96.250.321 / 191015
Loading...