Swann AMI Rescon Operator's Manual

AMI Rescon
Version 6.20 and higher
A-96.250.461 / 070318
Operator’
s Manual
© 2016, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists around the world. For any technical question, contact your nearest SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG Studbachstrasse 13 8340 Hinwil Switzerland
Internet: www.swan.ch
E-mail: support@swan.ch
Document Status
Title:
Monitor AMI Rescon Operator’s Manual
ID:
A-96.250.461
Revision Issue
00 Sept. 2006 First Edition
01 June 2012 Update to Rev. 4.61
02 Jan. 2015 Update to Rev. 5.30, main board V2.4
03 April 2016 Update to Rev. 6.00, main board V2.5
AMI Rescon
A-96.250.461 / 070318 1
Table of Contents
1. Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.1. Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2. General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3. Restrictions for use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2. Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1. Description of the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.2. Instrument Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3. Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4. Single Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.4.1 Transmitter AMI Rescon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.4.2 Flow Cell QV-Flow and QV-HFlow SS316L 130 . . . . . . . . . . . . 14
2.4.3 Flow Cell B-Flow SS316L 130 . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.4.4 Swansensor RC U. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3. Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1. Installation Checklist Monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.2. Mounting of Instrument Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.3. Connecting Sample Inlet and Outlet. . . . . . . . . . . . . . . . . . . . . . . 19
3.3.1 Sample Inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.3.2 Sample Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.4. Install Sensor RCU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5. Connect the Resistivity Sensor Cable . . . . . . . . . . . . . . . . . . . . . 22
3.6. Connect the Flow Sensor Cable. . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.7. Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.7.1 Connection Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.7.2 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.8. Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.8.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.8.2 Alarm Relay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.8.3 Relay Contacts 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.9. Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.9.1 Signal Output 1 and 2 (current outputs) . . . . . . . . . . . . . . . . . . 30
3.10. Interface Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.10.1 Signal Output 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.10.2 Profibus, Modbus Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.10.3 HART Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.10.4 USB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
2 A-96.250.461 / 070318
AMI Rescon
4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.1. Establish Sample Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.2. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5. Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.1. Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.2. Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.3. Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.4. Changing Parameters and values. . . . . . . . . . . . . . . . . . . . . . . . . 38
6. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.1. Maintenance Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
6.2. Stop of Operation for Maintenance. . . . . . . . . . . . . . . . . . . . . . . . 39
6.3. Maintenance of Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.3.1 Clean Sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.4. Conductivity-Resistivity QC-Kit Test Plug. . . . . . . . . . . . . . . . . . . 43
6.4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.4.2 Carry out a transmitter check . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6.5. Fine Adjust . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.6. Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
6.7. Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
7. Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
8. Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
8.1. Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
8.2. Diagnostics (Main Menu 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
8.3. Maintenance (Main Menu 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8.4. Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8.5. Installation (Main Menu 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
9. Program List and Explanations. . . . . . . . . . . . . . . . . . . . . . . . . 56
1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
10. Default Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
11. Index. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
12. Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
AMI Rescon
Safety Instructions
A-96.250.461 / 070318 3
AMI Rescon - Operator’s Manual
This document describes the main steps for instrument setup, oper­ation and maintenance.
1. Safety Instructions
General The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety practices designed to minimize these risks.
If you carefully follow the information contained in this section, you can protect yourself from hazards and create a safer work environ­ment.
More safety instructions are given throughout this manual, at the respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Target
audience
Operator: Qualified person who uses the equipment for its intended purpose.
Instrument operation requires thorough knowledge of applications, instrument functions and software program as well as all applicable safety rules and regulations.
OM Location The AMI Operator’s Manual shall be kept in proximity of the instru-
ment.
Qualification,
Training
To be qualified for instrument installation and operation, you must:
read and understand the instructions in this manual as well as
the Material Safety Data Sheets.
know the relevant safety rules and regulations.
4 A-96.250.461 / 070318
AMI Rescon
Safety Instructions
1.1. Warning Notices
The symbols used for safety-related notices have the following sig­nificance:
DANGER
Your life or physical wellbeing are in serious danger if such warnings are ignored.
Follow the prevention instructions carefully.
WARNING
Severe injuries or damage to the equipment can occur if such warnings are ignored.
Follow the prevention instructions carefully.
CAUTION
Damage to the equipment, minor injury, malfunctions or incor­rect process can be the consequence if such warnings are ig­nored.
Follow the prevention instructions carefully.
Mandatory
Signs
The importance of the mandatory signs in this manual.
Safety goggles
Safety gloves
AMI Rescon
Safety Instructions
A-96.250.461 / 070318 5
Warning Signs The importance of the warning signs in this manual.
1.2. General Safety Regulations
Legal
Requirements
The user is responsible for proper system operation. All precautions must be followed to ensure safe operation of the in­strument.
Spare Parts
and
Disposables
Use only official SWAN spare parts and disposables. If other parts are used during the normal warranty period, the manufacturer’s warranty is voided.
Electrical shock hazard
Corrosive
Harmful to health
Flammable
Warning general
Attention general
6 A-96.250.461 / 070318
AMI Rescon
Safety Instructions
Modifications Modifications and instrument upgrades shall only be carried out by
an authorized Service Technician. SWAN will not accept responsi­bility for any claim resulting from unauthorized modification or alter­ation.
WARNING
Electrical Shock Hazard
If proper operation is no longer possible, the instrument must be disconnected from all power lines, and measures must be taken to prevent inadvertent operation.
To prevent from electrical shock, always make sure that the
ground wire is connected.
Service shall be performed by authorized personnel only.Whenever electronic service is required, disconnect instru-
ment power and power of devices connected to. –relay 1, –relay 2, – alarm relay
WARNING
For safe instrument installation and operation you must read and understand the instructions in this manual.
WARNING
Only SWAN trained and authorized personnel shall perform the tasks described in this document.
1.3. Restrictions for use
The AMI Rescon is designed for determination of specific conduc­tivity or resistivity in high purity water.
To avoid clogging the flow cell, prevent sand, oil or solids, from en­tering into the flow cell.
Sufficient sample flow is coercive for the correct function of the in­strument. To determine values near 18.18 MOhm-cm exactly, a sample flow of 70–100 l/h is needed.
AMI Rescon
Product Description
A-96.250.461 / 070318 7
2. Product Description
2.1. Description of the System
This instrument is applicable for the determination of specific resis­tivity or specific conductivity in high purity water.
Measuring
Principle
The resistivity of high purity water is determined with a sensor con­sisting of two metal electrodes. The characteristics of each sensor is expressed as cell constant. An alternating voltage (to minimize polarization effects) is applied to two electrodes. Depending on the concentration of ions in the sample, a signal results between the electrodes which is proportional to the resistivity of the water.
The measuring result is indicated as resistivity or conductivity. The conductivity measurement depends on temperature, as the mobility of ions increase with rising temperature. To eliminate these effects, the temperature is determined simultaneous with an incorporated NT5K temperature sensor. Several temperature compensation curves for different applications are available.
Signal
Outputs
Two signal outputs programmable for measured values (freely scal­able, linear, bilinear, log) or as continuous control output (control parameters programmable).
Current loop: 0/4 –20 mA Maximal burden: 510 Ω Third signal output available as an option. The third signal output
can be operated as a current source or as a current sink (selectable via switch).
Relays Two potential-free contacts programmable as limit switches for
measuring values, controllers or timer for system cleaning with au­tomatic hold function. Maximum load: 1 A/ 250 VAC
Alarm Relay One potential free contact.
Alternatively:
Open during normal operation, closed on error and loss of
power.
Closed during normal operation, open on error and loss of
power.
Summary alarm indication for programmable alarm values and in­strument faults.
8 A-96.250.461 / 070318
AMI Rescon
Product Description
Input One input for potential-free contact to freeze the measuring value
or to interrupt control in automated installations. Programmable as HOLD or OFF function.
Communica-
tion interface
(optional)
USB Interface for logger download.Third signal output (can be used in parallel to the USB interface)RS485 with Fieldbus protocol Modbus or Profibus DPHART interface
Safety
Features
No data loss after power failure. All data is saved in non-volatile memory. Over voltage protection of in- and outputs. Galvanic separation of measuring inputs and signal outputs.
Standard
flow cell
A QV-Hflow flow cell is supplied with the Monitor as standard.
USP Operat-
ing Mode
The USP Operating Mode implemented in the firmware of the AMI Rescon transmitter allows the measurement of pharmaceutical wa­ter according to the USP <645>. If the USP Operating Mode is set to <off> conductivity or resistivity standard measurement with automatic temperature compensation is performed.
If the USP Operating Mode is set to <on>, the temperature com­pensation function is disabled. The uncompensated measuring val­ues are compared with the values of an implemented table defined by USP (see table below). If the deviation of the measured values is too high, the Error 15 (USP Error) is issued.
Temperature [°C] Conductivity [S/cm] Temperature [°C] Conductivity [S/cm] 0 0.6 55 2.1 5 0.8 60 2.2 10 0.9 65 2.4 15 1.0 70 2.5 20 1.1 75 2.7 25 1.3 80 2.7 30 1.4 85 2.7 35 1.5 90 2.7 40 1.7 95 2.9 45 1.8 100 3.1 50 1.9
AMI Rescon
Product Description
A-96.250.461 / 070318 9
Conductivity
QC-Kit Test
Plug
The implemented Transmitter Test together with Conductivity QC­Kit Test Plug containing a traceable high precision resistor allows to perform a verification of he measuring electronics at any time.
Fluidics The flow cell QV-Hflow consists of the flow cell block [D], the flow
sensor [B] and the flow regulating valve [E]. The conductivity sensor RC-U [A] with integrated temperature sen-
sor is screwed into the flow cell block [D]. The sample enters at the sample inlet [F]. It flows through the flow
regulating valve [E], where the flow rate can be adjusted. Then the sample flows through the flow cell block [D] were the resistivity of the sample is measured.
The sample leaves the flow cell block via flow meter through the sample outlet [C].
A B C
Sensor RC-U Flow sensor Sample outlet
D E F
Flow cell block Flow regulating valve Sample inlet
A
B
C
D
E
F
10 A-96.250.461 / 070318
AMI Rescon
Product Description
2.2. Instrument Specification
Power Supply Voltage:
Power consumption:
100–240 VAC (± 10%) 50/60 Hz (± 5%) or 24 VDC (± 10%) max. 30 VA
Electronics Aluminium with a protection degree of IP 66 / NEMA 4X
housing Ambient temperature:
Limit range of operation: Storage and transport: Humidity: Display:
-10 to +50 °C
-25 to +65 °C
-30 to +85 °C 10–90% rel., non condensing backlit LCD, 75 x 45 mm
Measuring
range
Resistivity: Conductivity:
0.01–18.18 M-cm
0.055–1000 s/cm
Sample
requirements
Flow rate: Temperature Sample inlet pressure: Sample outlet pressure:
70– 100 l/h up to 95 °C up to 2 bar pressure free
On-site The analyzer site must permit connections to:
requirements Sample inlet:
Sample outlet:
Swagelok tube 1/4” adapter FEP flexible tube 6 mm
Max. Altitude: 2000 m above sea level
AMI Rescon
Product Description
A-96.250.461 / 070318 11
Dimensions Panel:
Mounting hole distance Screws: Weight:
280x850x180 mm, stainless steel 254x824 8 mm diameter
7.0 kg
AMI Rescon
850 mm / 33½”
13 mm / ½”
4 x dia.10 mm / ¼”
254 mm/ 10”
280 mm/ 11”
824 mm / 32
7
/
16
12 A-96.250.461 / 070318
AMI Rescon
Product Description
2.3. Instrument Overview
A B C D
Panel Transmitter Sensor RC-U Flow sensor
E F G H
Flow cell Sample inlet Flow regulating valve Sample outlet
A
B
C
D
E
F
H
G
AMI Rescon
Product Description
A-96.250.461 / 070318 13
2.4. Single Components
2.4.1 Transmitter AMI Rescon
Electronic transmitter and controller for conductivity measurement.
140
180
165
103
18.5
7.5
Dimensions Width:
Height: Depth: Weight:
140 mm 180 mm 70 mm
1.5 kg
Specifications Electronics case:
Protection degree: Display: Electrical connectors:
Cast aluminum IP 66 / NEMA 4X backlit LCD, 75 x 45 mm screw clamps
14 A-96.250.461 / 070318
AMI Rescon
Product Description
2.4.2 Flow Cell QV-Flow and QV-HFlow SS316L 130
Flow cell made of stainless steel with flow sensor for the connection to SWAN transmitter and with manual flow regulating valve. Con­nection to tube with Swagelok adapter.
For one sensor with ¾" NPT thread
33
38
44
139
28
135 151234
96
10
45
60 75
Technical data Sample inlet:
Sample outlet: Sample temperature: Sample flow, QV-Flow: Sample flow, QV-HFlow: Sample inlet pressure: Sample outlet pressure:
Swagelok G 1/4” thread Serto elbow for 6 mm flexible tube 0–60 °C 3–25 l/h 10–120 l/ h max. 15 bar at 50 °C pressure free
AMI Rescon
Product Description
A-96.250.461 / 070318 15
2.4.3 Flow Cell B-Flow SS316L 130
Flow cell made of stainless steel SS316L to connect to tubes. For one sensor with ¾" NPT thread, fitting length 89 mm.
15
26
33
40
7
17
130
214
96
10
45
60 75
Technical data Sample inlet:
Sample outlet: Sample temperature: Sample inlet pressure:
Swagelok G 1/8” thread Swagelok G 1/8” thread
-10–130 °C max. 10 bar at 130 °C
16 A-96.250.461 / 070318
AMI Rescon
Product Description
2.4.4 Swansensor RC U
Sensor for the measurement of the specific conductivity and specif­ic resistivity in high purity water.
Swansensor RC U with integrated cable
Swansensor RC U with Plug
28
156
89
231
89
Specifications
RC U-Sensor
Measuring range: respective: Operating temperature: Pressure: Accuracy (at 25 °C):
Temperature sensor: Thread:
0.055– 1000 S/cm
0.01–200 M
-10 to + 90 °C max. 10 bar at + 90 °C > +/- 0.5% up to 20 S/cm +/- 1% from 20 s/ cm to 1000 S/cm NT5K 3/4” NPT
AMI Rescon
Installation
A-96.250.461 / 070318 17
3. Installation
3.1. Installation Checklist Monitors
Check Instrument’s specification must conform to the National Electri-
cal Code, all state and local codes, and all plant codes and stan­dards for electrical equipment.
On site require­ments
100–240 VAC ( 10%), 50/60 Hz ( 5%) or 24 VDC (±10%), isolated power outlet with ground connection and 30 VA. Sample line with sufficient sample flow and pressure (see Instru-
ment Specification, p. 10).
Installation Mounting of Instrument Panel, p. 18
Connecting Sample Inlet and Outlet, p. 19
Connect sen­sors
Only if single components have been ordered.
Connect the Resistivity Sensor Cable, p. 22. Connect the Flow Sensor Cable, p. 22.
Electrical wir­ing
NOTICE: Do not switch on the Instrument until all electrical
connections are made.
Connect all external devices like limit switches, current loops
and pumps.
Power-up Establish Sample Flow, p. 33.
Switch on power. Adjust sample flow according to flow cell specifications.
Instrument set-up
Program all necessary parameters see Programming, p. 33
Program the flow cell type Program the measuring mode (resistivity or conductivity).Program all sensor data (cell constant, temperature correction,
cable length).
Program the required temperature compensation.Program all parameters for external devices (interface,
recorders, etc.).
Program all parameters for instrument operation (limits,
alarms).
Run-in period If the conductivity value of the sample is very low, the sensor
might need some time until the correct reading is displayed
18 A-96.250.461 / 070318
AMI Rescon
Installation
3.2. Mounting of Instrument Panel
The first part of this chapter describes the preparing and placing of the system for use.
The instrument must only be installed by trained personnel. Mount the instrument in vertical position.For ease of operation mount it so that the display is at eye
level.
For the installation a kit containing the following installation
material is available: – 4 Screws 8x 60 mm – 4 Dowels – 4 Washers 8.4/ 24 mm
Mounting re-
quirements
The instrument is only intended for indoor installation.
Instrument Specification, p. 10
AMI Rescon
Installation
A-96.250.461 / 070318 19
3.3. Connecting Sample Inlet and Outlet
3.3.1 Sample Inlet
Preparation For the sample inlet a stainless steel tube is connected to the Swa-
gelok fitting at the flow cell. Cut the tube to length and deburr it. The tube must be straight and free from blemishes for approximately 1,5 x tube diameter from the end. Lubrication with lubricating oil, MoS2, Teflon etc. is recommended for the assembly and reassembly of bigger sized unions (thread, compression cone).
Installation 1 Insert the compression ferrule [C] and the compression
cone [D] into the union nut [B].
2 Screw on the union nut onto the body, do not tighten it.
3 Push the stainless steel pipe through the union nut as far as it
reaches the stop of the body
4 Mark the union nut at 6 o’clock position.
5 While holding the fitting body steady, tighten the nut union 1¼
rotation using an open ended spanner.
3.3.2 Sample Outlet
FEP flexible tube 6 mm. Connect the tube to the serto elbow union and Insert it into an atmospheric drain of sufficient capacity. Max. tube length is 1.5 m. Do not connect longer tubes.
A B C
D
E F
Tube Union nut Compression ferrule Compression cone Body Tightened connection
12
3
9
6
ABCDE
F
20 A-96.250.461 / 070318
AMI Rescon
Installation
3.4. Install Sensor RCU
The following instruction applies for all, the flow cell and the pipe flanges. To install the sensor RC U into a flow cell or a pipe flange proceed as follows:
Install Sensor
into Flow Cell
Install Sensor
into a tube
Installation in a by-pass is recommended. Choose the installation point carefully, to ensure the sensor is always filled with water, even
if sample flow is interrupted. To avoid the formation of air bubbles inside the sensor make sure that the air vent holes are always sub­merged.
A B C
Teflon tape Sensor thread Flow cell
A
B
C
AMI Rescon
Installation
A-96.250.461 / 070318 21
Installation examples for tube installa-
tion
1 Wrap 7 turns of teflon tape around the sensor thread.
2 Screw the sensor into the flow cell or the pipe flange.
3 Tighten the sensor well with a monkey spanner.
A B C D E
Sensor Flange Air vent hole Tube Flow direction
A B C D E
Sensor Flange Air vent hole Flow direction Tube
AB C D E
AB C ED
22 A-96.250.461 / 070318
AMI Rescon
Installation
3.5. Connect the Resistivity Sensor Cable
WARNING
Electrical shock hazard!
Before opening the AMI Transmitter switch power off.
Use one of the cable glands to feed the sensor cable into the sen­sor housing.
1 Choose a suitable cable gland, see chapter Electrical Connec-
tions, p. 23.
2 Remove the plug [A] from the cable gland [B].
3 Open the AMI transmitter housing.
4 Feed the sensor cable through the cable gland [B] into the
transmitter housing.
5 Connect the cable to the terminals according to the connecting
diagram see Connection Diagram, p. 25.
6 Close the AMI transmitter housing.
7 Switch on power.
3.6. Connect the Flow Sensor Cable
If a flow cell with a flow sensor was ordered (QV-Flow) connect it to the AMI Transmitter. Proceed according to chapter Connect the Re-
sistivity Sensor Cable, p. 22.
BA
AMI Rescon
Installation
A-96.250.461 / 070318 23
3.7. Electrical Connections
WARNING
Risk of electrical shock.
Do not perform any work on electrical components if the trans­mitter is switched on. Failure to follow safety instructions could result in serious injury or death.
Always turn off AC power before manipulating electric parts.Grounding requirements: Only operate the instrument from
an power outlet which has a ground connection.
Make sure the power specification of the instrument corre-
sponds to the power on site.
Cable
thicknesses
In order to comply with IP66, use the following cable thicknesses
NOTICE: Protect unused cable glands
Wire  For Power and Relays: Use max. 1.5 mm
2
/ AWG 14
stranded wire with end sleeves.
For Signal Outputs and Input: Use 0.25 mm
2
/ AWG 23
stranded wire with end sleeves.
A B C
PG 11 cable gland: cable Ø
outer
5–10 mm
PG 7 cable gland: cable Ø
outer
3–6.5 mm
PG 9 cable gland: cable Ø
outer
4–8 mm
ABC
Loading...
+ 57 hidden pages