AMI Rescon
Default Values
A-96.250.461 / 070318 75
Scaling: Range low:................................................................ 0.0 °C
Scaling: Range high: ............................................................50.0 °C
Alarm Relay: Alarm:
Resistivity Alarm high: .......................................................................... 200 MΩ
Alarm low:........................................................................... 0.00 MΩ
Hysteresis:.......................................................................... 1.00 MΩ
Conductivity Alarm high: ..........................................................................2000 µS
Alarm low:.......................................................................... 0.000 µS
Hysteresis:..........................................................................10.00 µS
Delay: .......................................................................................... 5 s
Sample Flow; Flow Alarm........................................................... yes
Sample Flow; Alarm high.......................................................120 l/h
Sample Flow; Alarm low............................................................5 l/h
Sample Temp; Alarm high:.......................................................90 °C
Sample Temp; Alarm low:.......................................................... 0 °C
Case temp. high: .....................................................................65 °C
Case temp. low:......................................................................... 0 °C
Relay 1 and 2 Function:.......................................................................... limit upper
Parameter:.................................................................... Meas. Value
Resistivity Setpoint: .............................................................................. 200 MΩ
Hysteresis:............................................................................... 1 MΩ
Conductivity Setpoint: ..............................................................................1000 µS
Hysteresis:..........................................................................10.00 µS
Delay: ........................................................................................ 30 s
If Function = Control upw. or dnw:
Parameter:....................................................................Meas. Value
Settings: Actuator: .......................................................... Frequency
Settings: Pulse Frequency: ..............................................120/min
Resistivity Settings: Control Parameters: Setpoint:........................... 200 MΩ
Settings: Control Parameters: P-band: ................................ 1 MΩ
Conductivity Settings: Control Parameters: Setpoint:...........................1000 µS
Settings: Control Parameters: P-band: ...........................10.00 µS
Parameter:...................................................................Temperature
Settings: Actuator: .......................................................... Frequency
Settings: Pulse Frequency: .....................................
.........120/min
Settings: Control Parameters: Setpoint: ...............................50 °C
Settings: Control Parameters: P-band: .................................. 1 °C
Parameter:................................................................... Sample flow
Settings: Actuator: .......................................................... Frequency