Swann AMI Oxytrace, AMI Oxytrace QED Operator's Manual

AMI Oxytrace AMI Oxytrace QED
Version 6.20 and higher
or’s Manual
Operat
A-96.250.531 / 300617
SWAN and its representatives maintain a fully trained staff of technical specialists around the world. For any technical question, contact your nearest SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG Studbachstrasse 13 8340 Hinwil Switzerland
Internet: www.swan.ch
E-mail: support@swan.ch
Document Status
Title:
ID:
Monitor AMI Oxytrace QED Operator’s Manual
A-96.250.531
Revision Issue
00 Jan. 2008 First Edition
01 Nov. 2008 Including AMI Oxytrace QED
02 July 2011 Chapter Quality Assurance added
03 Oct. 2013 New mainboard V2.4, update to FW Release 5.4
04 Sept. 2016 New mainboard V2.5, update to FW Release 6.20
© 2016, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved
subject to change without notice.
AMI Oxytrace QED
Table of Contents
1. Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1. Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2. General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2. Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.1. Description of the System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.2. Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.3. Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.4. Single Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.4.1 AMI Oxytrace Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
2.4.2 Sensor OXYTRACE G . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
2.4.3 QV-Flow PMMA OTG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
2.4.4 B-Flow SS316L OTG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3. Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.1. Installation Check List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.2. Mounting of Instrument Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.3. Connecting Sample Inlet and Outlet. . . . . . . . . . . . . . . . . . . . . . . 23
3.3.1 Swagelok Fitting Stainless Steel at Sample Inlet . . . . . . . . . . . . . 23
3.3.2 FEP Tube at Sample Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.4 Install the Swansensor Oxytrace G . . . . . . . . . . . . . . . . . . . . . . . 24
3.5. Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.5.1 Electrical Connection Diagram AMI Oxytrace . . . . . . . . . . . . . . . 27
3.5.2 Electrical Connection Diagram AMI Oxytrace QED . . . . . . . . . . . 28
3.5.3 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.6. Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.7. Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.7.1 Alarm Relay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
3.7.2 Relay 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.8. Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.8.1 Signal output 1 and 2 (current outputs) . . . . . . . . . . . . . . . . . . . . 33
3.9. Interface Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.9.1 Signal Output 3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.9.2 Profibus, Modbus Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.9.3 HART Interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
3.9.4 USB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.1. Establish Sample Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.2. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
A-96.250.531 / 300617 1
AMI Oxytrace QED
5. Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.1. Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.2. Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.3. Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.4. Changing Parameters and values. . . . . . . . . . . . . . . . . . . . . . . . . 40
6. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.1. Maintenance Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
6.2. Stop of Operation for Maintenance. . . . . . . . . . . . . . . . . . . . . . . . 41
6.3. Maintenance of the Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . 42
6.3.1 Electrolyte exchange. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.3.2 Clean Swansensor Oxytrace G and Flow Cell . . . . . . . . . . . . . . . 44
6.4. Maintenance of the Faraday electrode . . . . . . . . . . . . . . . . . . . . . 45
6.5. Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.6. Zero-Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.7. Faraday Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6.8. Quality Assurance of the Instrument. . . . . . . . . . . . . . . . . . . . . . . 51
6.8.1 Activate SWAN Quality assurance procedure . . . . . . . . . . . . . . . 52
6.8.2 Pre-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.8.3 Connect the sample lines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.8.4 Carry out comparison measurement. . . . . . . . . . . . . . . . . . . . . . . 55
6.8.5 Completion of the measurement. . . . . . . . . . . . . . . . . . . . . . . . . . 56
6.9. Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
6.10. Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
7. Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
8. Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
8.1. Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
8.2. Diagnostics (Main Menu 2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
8.3. Maintenance (Main Menu 3) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
8.4. Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
8.5. Installation (Main Menu 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
9. Program List and Explanations. . . . . . . . . . . . . . . . . . . . . . . . . 69
1 Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
10. Material Safety Data sheets . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
10.1. Reagents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
2 A-96.250.531 / 300617
AMI Oxytrace QED
11. Default Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
12. Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
13. Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
A-96.250.531 / 300617 3
AMI Oxytrace QED
Safety Instructions
AMI Oxytrace QED ­Operator’s Manual
This document describes the main steps for instrument setup, oper­ation and maintenance.

1. Safety Instructions

General The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety practices designed to minimize these risks.
If you carefully follow the information contained in this section, you can protect yourself from hazards and create a safer work environ­ment.
More safety instructions are given throughout this manual, at the respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Target
audience
OM Location The AMI Operator’s Manual shall be kept in proximity of the instru-
Qualification,
Training
Operator: Qualified person who uses the equipment for its intended purpose.
Instrument operation requires thorough knowledge of applications, instrument functions and software program as well as all applicable safety rules and regulations.
ment.
To be qualified for instrument installation and operation, you must:
read and understand the instructions in this manual as well as
the Material Safety Data Sheets.
know the relevant safety rules and regulations.
4 A-96.250.531 / 300617
AMI Oxytrace QED
Safety Instructions

1.1. Warning Notices

The symbols used for safety-related notices have the following sig­nificance:
DANGER
Your life or physical wellbeing are in serious danger if such warnings are ignored.
Follow the prevention instructions carefully.
WARNING
Severe injuries or damage to the equipment can occur if such warnings are ignored.
Follow the prevention instructions carefully.
CAUTION
Damage to the equipment, minor injury, malfunctions or incor­rect process can be the consequence if such warnings are ig­nored.
Follow the prevention instructions carefully.
Mandatory
Signs
A-96.250.531 / 300617 5
The importance of the mandatory signs in this manual.
Safety goggles
Safety gloves
AMI Oxytrace QED
Safety Instructions
Warning Signs The importance of the warning signs in this manual.
Electrical shock hazard
Corrosive
Harmful to health
Flammable
Warning general
Attention general
6 A-96.250.531 / 300617
AMI Oxytrace QED
Safety Instructions

1.2. General Safety Regulations

Legal
Requirements
Spare Parts
and
Disposables
Modifications Modifications and instrument upgrades shall only be carried out by
The user is responsible for proper system operation. All precautions must be followed to ensure safe operation of the instrument.
Use only official SWAN spare parts and disposables. If other parts are used during the normal warranty period, the manufacturer’s warranty is voided.
an authorized Service Technician. SWAN will not accept responsi­bility for any claim resulting from unauthorized modification or alter­ation.
WARNING
Risk of Electrical Shock
If proper operation is no longer possible, the instrument must be disconnected from all power lines, and measures must be taken to prevent inadvertent operation.
To prevent from electrical shock, always make sure that the
ground wire is connected.
Service shall be performed by authorized personnel only.Whenever electronic service is required, disconnect instru-
ment power and power of devices connected to. – relay 1, – relay 2, – alarm relay
WARNING
For safe instrument installation and operation you must read and understand the instructions in this manual.
WARNING
Only SWAN trained and authorized personnel shall perform the tasks described in this document.
A-96.250.531 / 300617 7
AMI Oxytrace QED
Product Description

2. Product Description

This manual describes the functions of the following monitors
AMI OxytraceAMI Oxytrace QED
Both monitors are mainly identical except that AMI Oxytrace QED includes a faraday verification.

2.1. Description of the System

Application AMI Oxytrace is used to measure low levels of oxygen in high-puri-
ty water. Especially in power plant water cycles (e.g. feedwater), a very low level of oxygen is needed to prevent corrosion.
Measuring
principle
Clark principle: The sensor consists of one noble metal electrode (e.g. platinum or
gold), a reference electrode (mostly Ag/AgCl) and optionally a met­al guard electrode.
The Clark-type electrode is the most widely used oxygen sensor for measuring oxygen dissolved in a liquid. The basic principle is that there is a cathode and an anode submersed in an electrolyte and a voltage is applied between the two parts. Oxygen enters the sensor through a permeable membrane by diffusion, and is reduced at the cathode according to
O
+ 4e- + 2 H2O --> 4 OH
2
This reaction creates a measurable current. There is a linear cor­relation between the oxygen concentration and the electrical cur­rent.
The guard electrode is on the same voltage level as the cathode but there is no current measurement. Oxygen which diffuses from the electrolyte to the cathode is consumed by the guard electrode. As a consequence, residual oxygen in the electrolyte will no more disturb the measurement signal and the response time to low oxy­gen levels will be shorter.
-
Temperature
compensa-
tion
The measuring signal depends on temperature, but is automatically compensated to 25 °C. The sample temperature is determined con­tinuously by a temperature sensor inside the oxygen electrode.
8 A-96.250.531 / 300617
AMI Oxytrace QED
Product Description
Faraday
Verification
40
35
30
25
20
15
10
Dissolved Oxygen [ppm]
5
For AMI Oxytrace QED only. When a direct current is passed through water by means of two
electrodes, electrolysis of the liquid takes place, according to the laws of Michael Faraday. Water is converted to molecular oxygen and hydrogen.
Thus, by controlling the current, a fixed, known amount of oxygen, which is independent of temperature and pressure, may be gener­ated. If the sample flow is known, an exact oxygen concentration increment can be generated at low levels. This increment is used to verify the flawless operation of the whole system. Thereby the re­sponse characteristics (incremental change and response time) of the sensor are taken into account. Sensor failures (loss of electro­lyte etc.) can be detected very easily as shown in the plot below.
operational sensor
defective sensor
0
30
60 90 120 150 180 210 240 270 300 330
Time [Sec]
Unusual sensor performance is recognized immediately and gives also a maintenance indication to the operator/user. The faraday verification is therefore an excellent QA/QC tool.
A-96.250.531 / 300617 9
AMI Oxytrace QED
Product Description
Signal
Outputs
Relay Two potential-free contacts programmable as limit switches for
Alarm Relay One potential free contact.
Input For potential-free contact to freeze the measuring value or to inter-
Safety
Features
Two signal outputs programmable for measured values (freely scaleable, linear or bilinear) or as continuous control output (control parameters programmable).
Current loop: 0/4–20 mA Maximal burden: 510 Third signal output available as an option.
measuring values, controllers or timer for system cleaning with au­tomatic hold function. Both contacts can be used as normally open or normally closed.
Maximum load: 1 A/250 VAC
Alternatively:
Open during normal operation, closed on error and loss of
power.
Closed during normal operation, open on error and loss of
power. Summary alarm indication for programmable alarm values and in­strument faults.
rupt control in automated installations (hold function or remote-off)
No data loss after power failure. All data is saved in non-volatile memory. Over voltage protection of in- and outputs.Galvanic sepa­ration of measuring inputs and signal outputs.The analyzer is facto­ry tested and ready for installation and operation.
Communica-
tion Interface
(optional)
USB Interface for logger downloadRS485 with Fieldbus protocol Modbus or Profibus DP.HART interface
10 A-96.250.531 / 300617
AMI Oxytrace QED
A
B
C
D
E
F
G
Product Description
Fluidics AMI Oxytrace
Swansensor oxygen combined with QV-flow PMMA OTG flow cell. The sample flows via sample inlet [G] through the flow regulating valve [D], where the flow rate can be adjusted. Then the sample flows into the measuring cell [C] were the Oxygen concentration and temperature of the sample is measured.
The sample leaves the measuring cell via flow sensor [B] through the sample outlet [E] and the drain funnel [F].
A
Oxygen sensor
B
Flow sensor
C
Flow cell
D
Flow regulating valve
A-96.250.531 / 300617 11
E
Sample outlet
F
Drain funnel
G
Sample inlet
AMI Oxytrace QED
A
B
C
D
F
E
G
H
Product Description
Fluidics AMI Oxytrace QED
Swansensor oxygen combined with QV-flow PMMA OTG flow cell. The sample flows via sample inlet [H] through the flow regulating valve [E], where the flow rate can be adjusted. Then the sample flows through the faraday electrode [D] into the measuring cell [C] were the Oxygen concentration of the sample is measured.
The sample leaves the measuring cell via flow sensor [B] through the sample outlet [F] and the drain funnel [G].
A
Oxygen sensor
B
Flow sensor
C
Flow cell
D
Faraday electrode
12 A-96.250.531 / 300617
E
Flow regulating valve
F
Sample outlet
G
Drain funnel
H
Sample inlet
AMI Oxytrace QED
Product Description

2.2. Technical Data

Power Supply Voltage:
Power consumption:
Electronics Aluminium with a protection degree of IP 66 / NEMA 4X
housing Ambient temperature:
Humidity: Display:
Sample
requirements
Flow cell and
connection
On-Site
requirements
Accuracy
Reproducibility
Flow rate: Temperature: Inlet pressure: Outlet pressure pH: Suspended solids:
Flow cell made of acrylic glass with built-in flow adjustment valve and digital sample flow meter
Sample inlet: Sample outlet:
Sample inlet:
Sample outlet:
± 1.5% of measured value or ± 0.2 ppb ±1% of measured value or ± 0.15 ppb
100–240 VAC (± 10%) 50/60 Hz (± 5%) or 24 VDC (± 10%) max. 30 VA
-10 to +50 °C 10–90% rel., non condensing backlit LCD, 75 x 45 mm
8–25 l/h 15–45 °C
0.2 to 1 bar pressure free not lower than pH 4 less than 10 ppm
1/4” Swagelok tube adapter for flexible tube 15x20 mm
Swagelok connection for stainless steel tube 1/4”. 15x20 mm (1/2”) hose nozzle which must end in a pressure free waste of suf­ficient capacity.
Measuring
Range
Temperature
Sample Flow
Measurement
A-96.250.531 / 300617 13
up to 60°C Resolution: 0.1°C
With Swan sample flow meter and alarm in case of insufficient sample flow.
AMI Oxytrace QED
AMI Oxytrace
850 mm / 33½”
13 mm / ½”
4 x dia. 10 mm / ⅜”
254 mm/ 10”
280 mm/ 11”
824 mm / 32
7
/
16
Product Description
Dimensions
(Oxytrace and
Oxytrace QED)
Panel:
Mounting hole distance Screws: Weight:
Stainless steel 280x850x 150 mm 254x824 8 mm diameter
12.0 kg
14 A-96.250.531 / 300617
AMI Oxytrace QED
A
B
C
D E
F G
H
I
Product Description

2.3. Instrument Overview

AMI Oxytrace
A
Panel
B
AMI Transmitter
C
Oxygen sensor
D
Flow sensor
E
Flow cell
A-96.250.531 / 300617 15
F
Flow regulating valve
G
Sample outlet
H
Drain funnel
I
Sample inlet
AMI Oxytrace QED
A
B
C
D
E F G H I
J
K
Product Description
AMI Oxytrace QED
A
Panel
B
AMI Transmitter
C
Faraday control unit
D
Oxygen sensor
E
Sample outlet
F
Flow sensor
G
Flow cell
H
Flow regulating valve
I
Faraday electrode
J
Drain funnel
K
Sample inlet
16 A-96.250.531 / 300617
AMI Oxytrace QED
Product Description

2.4. Single Components

2.4.1 AMI Oxytrace Transmitter

Electronic transmitter and controller for oxygen measurement.
18.5
7.5
180
165
140 103
4.5
Dimensions Width:
Height: Depth: Weight:
Specifications Electronics case:
Protection degree: Display: Electrical connectors:
A-96.250.531 / 300617 17
140 mm 180 mm 70 mm
1.5 kg Cast aluminum
IP 66 / NEMA 4X backlit LCD, 75 x 45 mm screw clamps
AMI Oxytrace QED
100
59
31.9
38.3
Product Description

2.4.2 Sensor OXYTRACE G

Sensor for the measurement of dissolved oxygen in ultra pure wa­ter. Precise oxygen measuring cell with integrated temperature sensor and guard electrode for faster initial response time after maintenance.
Technical data:Clark oxygen electrode
Cathode gold, anode silver, guard silver Zero current-free electrode system Robust 25 µm fluoropolymer diaphragm
Measuring
range
Temp. sensor:NT5K
Accuracy:0,3 % if calibration temperature = measuring temp.
Precision
Response time
Minimal flow
Operating temp.
Material
Protection
Weight
0–20 ppm O Automatic range switching
:
Range
0.1–9.99 ppb 10–199.9 ppb 200– 1999 ppb 2–20 ppm 0 – 200% saturation
1,5 % at ± 10 °C deviation to cal. temperature ± 1 % of reading or ±0.15 ppb
:
< 30 seconds (rising concentration)
t
:
90
50 cm/s Pressure resistance: 3 bar
:
max. 50 °C
:
shaft: polyacetal copolymer
:
cathode/anode/guard: gold/silver/silver membrane: fluoropolymer IP 68
:
150 g
:
(25 °C)
2
Resolution
0.01 ppb
0.1 ppb
1.0 ppb
0.01 ppm
18 A-96.250.531 / 300617
AMI Oxytrace QED
217.7
178.5
164
138
80
21
38
8
62
5.3
Product Description

2.4.3 QV-Flow PMMA OTG

Flow cell Flow cell made of acrylic glass with integrated flow sensor.
Sample temp. max. 45 °C
Inlet pressure max. 1 bar
Outlet pressure
Sample flow 6–25 l/h
Process con-
nection inlet
Outlet SERTO angle for 6 mm flexible tube
Dimensions see picture below
Pressure free
Swagelok connection for 1/4” tube
A-96.250.531 / 300617 19
AMI Oxytrace QED
Product Description

2.4.4 B-Flow SS316L OTG

Flow cell B-Flow SS316L OTG is made of stainless steel without flow sensor
and can be used for higher operating pressures and temperatures.
Operating temp.
Sensor max. 50 °C
Operating
pressure
Sensor max. 3 bar
Flow cell
connection
Dimensions see picture below
-10 to + 130 °C
max. 5 bar at 130 °C
2x female thread 1/8" ISO
52 40
126.1
128.6
99.5
20 A-96.250.531 / 300617
AMI Oxytrace QED
Installation

3. Installation

The first part of this chapter describes the preparing and placing of the system in position for use. The second part describes the wiring and the properties of the signal outputs and relay.
Restrictions
on use

3.1. Installation Check List

If exposed to weather, install a cover to protect the instrument from insolation and precipitation. See Technical Data, p. 13 for tempera­ture limitations.
Check Instrument’s specification must conform to the National Electrical
On site requirements
Installation Mount the instrument in vertical position. Display should be at eye
Electrical Connections
Swansensor
Oxytrace G
Code, all state and local codes, and all plant codes and standards for electrical equipment.
100–240 VAC ( 10%), 50/60 Hz ( 5%) or 24 VDC (±10%), isolated power outlet with ground connection and 30 VA Sample line with 8– 25 l/h and 0.2 – 3 bar. Waste line with pressure free drain.
level. Connect the sample and waste line, see Connecting Sample Inlet
and Outlet, p. 23.
Connect all external devices like limit switches, current loops and pumps, see Electrical Connections, p. 25. Connect the power cord, do not switch on power until all external devices are connected.
The Swansensor Oxytrace G is delivered with prefilled electrolyte chamber. A transport protection cap filled with water keeps the sensor wet during transport and storage. Before installation, remove the transport protection cap and clean the sensor sur­faces with water. Afterwards install the oxygen sensor and con­nect the cable. See Install the Swansensor Oxytrace G, p. 24.
A-96.250.531 / 300617 21
AMI Oxytrace QED
Installation
Tur n-o n power
Run-in Period Calibration
Open Sample Flow
Programming Program all parameters for external devices (interface, etc.). Set
Switch on power. First, the analyzer performs a self test, displays the firmware ver­sion and then starts normal operation
Leave the sensor at the air. The sensor has to run-in for at least 30 min, better 1 h. During this time it has to be at air with a dry and clean membrane and has to be connected to the powered instru­menty. Calibrate the sensor after this time, see Calibration, p. 47. and mount it into the flow cell afterwards.
Open the flow regulating valve to allow sample flow into the flow cell and into the waste. Sample must always overflow into waste
all parameters for instrument operation (limits, alarms).

3.2. Mounting of Instrument Panel

The first part of this chapter describes the preparing and placing of the system for use.
The instrument must only be installed by trained personnel. Mount the instrument in vertical position.For ease of operation mount it so that the display is at eye
level.
For the installation a kit containing the following installation
material is available:
– 4 Screws 8x60 mm
– 4 Dowels
– 4 Washers 8.4/24 mm
Mounting re-
quirements
The instrument is only intended for indoor installation. For dimensions see Dimensions (Oxytrace and Oxytrace QED), p.
14.
22 A-96.250.531 / 300617
AMI Oxytrace QED
12
3
9
6
ABCDE
F
Installation

3.3. Connecting Sample Inlet and Outlet

3.3.1 Swagelok Fitting Stainless Steel at Sample Inlet

Preparation Cut the tube to length and deburr it. The tube must be straight and
free from blemishes for approximately 1,5 x tube diameter from the end. Lubrication with lubricating oil, MoS2, Teflon etc. is recommended for the assembly and reassembly of bigger sized unions (thread, compression cone).
Installation 1 Insert the compression ferrule [C] and the compression
cone [D] into the union nut [B].
2 Screw on the union nut onto the body, do not tighten it.
3 Push the stainless steel pipe through the union nut as far as it
reaches the stop of the body.
4 Mark the union nut at 6 o’clock position.
5 While holding the fitting body steady, tighten the nut union 1¼
rotation using an open ended spanner.
A
Stainless steel tube
B
Union nut
C
Compression ferrule
D
Compression cone
E
Body
F
Tightened connection
A-96.250.531 / 300617 23
AMI Oxytrace QED
A
B
C
A
B
C
D
E
Installation

3.3.2 FEP Tube at Sample Outlet

1/2” Tube at waste funnel.
Connect the 1/2” tube [C] to the hose nozzle [B] and place it into a drain with atmospheric pressure.

3.4 Install the Swansensor Oxytrace G

The Swansensor Oxytrace G is delivered with prefilled electrolyte chamber [E]. A transport protection cap [B] filled with water [D] keeps the sensor wet during transport and storage. To install the sensor proceed as follows:
A
Waste funnel
B
Hose nozzle
C
1/2” tube
A
Fixing sleeve
B
Transport protection cap
C
Swansensor Oxytrace G
D
Water
E
Sensor cap filled with electrolyte
24 A-96.250.531 / 300617
1 Unscrew the fixing sleeve [A].
2 Remove the transport protection cap [B].
3 Clean the Swansensor Oxytrace G [C] with water.
4 Install the Swansensor Oxytrace G into the flow cell
5 Connect the sensor cable to the transmitter, see Electrical Con-
nection Diagram AMI Oxytrace, p. 27.
AMI Oxytrace QED
Installation

3.5. Electrical Connections

WARNING
Risk of electrical shock.
Do not perform any work on electrical components if the trans­mitter is switched on. Failure to follow safety instructions could result in serious injury or death.
Always turn off AC power before manipulating electric parts.Grounding requirements: Only operate the instrument from
an power outlet which has a ground connection.
Make sure the power specification of the instrument corre-
sponds to the power on site.
Cable
thicknesses
In order to comply with IP66, use the following cable thicknesses
ABC
PG 11 cable gland: cable Ø
A B
PG 7 cable gland: cable Ø
C
PG 9 cable gland: cable Ø
outer
3–6.5 mm
outer
4–8 mm
outer
5–10 mm
NOTICE: Protect unused cable glands
Wire  For Relays: Use max. 1.5 mm
end sleeves.
For Signal Outputs and Input: Use 0.25 mm
2
/ AWG 14 stranded wire with
2
/ AWG 23
stranded wire with end sleeves.
A-96.250.531 / 300617 25
AMI Oxytrace QED
Installation
WARNING
External Voltage.
External supplied devices connected to relay 1 or 2 or to the alarm relay can cause electrical shocks
Make sure that the devices connected to the following con-
tacts are disconnected from the power before resuming in-
stallation.
–relay 1
–relay 2
– alarm relay
WARNING
To prevent from electrical shock, do not connect the instrument to the power unless the ground wire (PE) is connected.
Do not connect unless specifically instructed to do so.
WARNING
The mains of the AMI Transmitter must be secured by a main switch and appropriate fuse or circuit breaker.
26 A-96.250.531 / 300617
AMI Oxytrace QED
Installation

3.5.1 Electrical Connection Diagram AMI Oxytrace

CAUTION
Use only the terminals shown in this diagram, and only for the mentioned purpose. Use of any other terminals will cause short circuits with possible corresponding consequences to material and personnel.
A-96.250.531 / 300617 27
AMI Oxytrace QED
Installation

3.5.2 Electrical Connection Diagram AMI Oxytrace QED

CAUTION
Use only the terminals shown in this diagram, and only for the mentioned purpose. Use of any other terminals will cause short circuits with possible corresponding consequences to material and personnel.
28 A-96.250.531 / 300617
Loading...
+ 68 hidden pages