AMI Hydrogen QED
Default Values
A-96.250.791 / 040219 75
Sample Flow, Alarm low: ........................................................6.0 l/h
Sample Temp., Alarm High:..................................................... 50 °C
Sample Temp., Alarm Low:........................................................ 0 °C
Alarm Saturation; Alarm high..................................................120 %
Alarm Saturation; Alarm low ....................................................0.0 %
Alarm Saturation; Hysteresis......................................................2 %
Alarm Saturation; Delay............................................................... 5 s
Case temp. high: .....................................................................65 °C
Case temp. low:......................................................................... 0 °C
Relay 1 Function: .......................................................................... limit upper
Parameter:........................................................................ Hydrogen
Setpoint: ..........................................................................10.00 ppm
Hysteresis:........................................................................... 100 ppb
Delay: ........................................................................................ 30 s
Relay 2 Function: .......................................................................... limit upper
Parameter:....................................................................Temperature
Setpoint: ..................................................................................50 °C
Hysteresis:.................................................................................1 °C
Delay: ........................................................................................ 30 s
If Function = Control upw. or dnw:
Parameter:.................................................................... Meas. Value
Settings: Actuator: ..........................................................Frequency
Settings: Pulse Frequency: ..............................................120/min
Settings: Control Parameters: Setpoint: ....................... 10.00 ppm
Settings: Control Parameters: P-band: ............................100 ppb
Settings: Control Parameters: Reset time: ............................... 0 s
Settings: Control Parameters: Derivative Time: ....................... 0 s
Settings: Control Parameters: Control Timeout:................... 0 min
Settings: Act. Time prop.: Cycle time: ................................... 60 s
Settings: Act. Time prop.: Response time: ............................ 10 s
Settings: Act. Motor valve: Run time: .................................... 60 s
Settings: Act. Motor valve: Neutral zone: ................................5%
If Function = Timer:
Mode:.................................................................................... Interval
Interval: ................................................................................ 1 min
Mode: ........................................................................................daily
Start time:........................................................................ 00.00.00
Mode:..................................................................................... weekly
Calendar; Start time: ....................................................... 00.00.00
Calendar; Monday to Sunday:.................................................. Off
Run time: ................................................................................... 10 s
Delay: .......................................................................................... 5 s