SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
For safe instrument operation, you must read and understand the
instructions in this manual.
Warning notices
The symbols used for safety-related notices have the following
meaning:
WARNING
Severe injuries or damage to the equipment can occur if such
warnings are ignored.
Follow the prevention instructions carefully.
CAUTION
Damage to the equipment, minor injury, malfunctions or incorrect process can be the consequence if such warnings are ignored.
Follow the prevention instructions carefully.
A-96.730.101 / 1005193
Page 6
Chematest 30
Safety Instructions
Reagents
Electronics
WARNING
Health hazard
For safe handling of the reagents, you must read and understand the corresponding Material Safety Data Sheets (MSDS).
These can be downloaded from www.swan.ch.
CAUTION
Observe the following instructions during operation and when
charging the instrument:
Only charge the instrument using a standard type A USB
socket and the supplied USB cable.
Protect from heat and splash water during charging (not
IP67).
Do not expose the instrument to direct sunlight or other sourc-
es of heat, especially when charging.
CAUTION
The housing must not be opened except for the housing parts
specified in the maintenance chapter.
4A-96.730.101 / 100519
Page 7
Chematest 30
Parts and Controls
2.Parts and Controls
2.1.Instrument Overview
A
F
A
Charging socket
B
Cuvette adapter
C
Display
B
C
D
D
Keypad
E
Cuvette cap
F
Cuvette
E
A-96.730.101 / 1005195
Page 8
Chematest 30
A
B
C
D
E
Parts and Controls
2.2.Keys
APOWER
BARROW UP
CARROW DOWN
DEXIT
EENTER
to power on or off the instrument
to move up in a menu list and to increase digits
to move down in a menu list and to decrease digits
to exit a menu or command
to move back to the previous menu level
to open a selected sub-menu
to accept an entry
6A-96.730.101 / 100519
Page 9
Chematest 30
A
83%
Parts and Controls
2.3.Display
Once the instrument has started up, the main menu is displayed.
The main menu is organized as follows:
Main Menu
83%
Free chlorine
Comb. Cl
1 cuv.
2
pH Phenol red
B
C
A Quick start list
Provides quick access to the most important methods. By default, the
last three methods selected from the method catalogue are displayed.
The list is automatically adjusted each time a method is selected from
the method catalogue.
Alternatively, it is also possible to setup a fixed quick start list, see Initial
Setup, p. 8.
B Method catalogue
List of all available methods.
CMenu
Access to further instrument functions (e.g. display of stored measurements). Detailed descriptions of all menu functions can be found in
chapter Menu Explanations, p. 23.
Method Catalogue
Menu
SymbolsThe symbols displayed in the top right corner have the following
meaning:
Alarm active. See <Menu>/<Diagnostics> /<Alarms> for details.
Bluetooth activated
Battery status (remaining capacity in %)
Instrument is charging
A-96.730.101 / 1005197
Page 10
Chematest 30
Initial Setup
3.Initial Setup
3.1.Select language
Navigate to <Menu>/<Settings>/ <Miscellaneous> /<Language>
and select the applicable language from the list.
3.2.Set time and date
Set time and date under <Menu>/<Maintenance>/ <Set Time>.
3.3.Optional: Activate expert mode
The user can choose between two types of method guidance:
Expert mode inactive (default setting):
In this mode, the Chematest 30 guides the user step-by-step
through each measurement. The user has to confirm each step
before the next step is displayed.
Expert mode active:
The expert mode is intended for experienced users who are
already familiar with the respective measurement methods and
want a faster workflow. Only the most necessary information (e.g.
required reagents, reaction times) is displayed on the screen.
This setting can be changed under menu item <Menu>/ <Settings>/
<Miscellaneous>/<Expert mode>.
3.4.Optional: Set up user and/or ID list
Navigate to <Menu>/<Settings>/ <Identification> /<Attribute> and
select which attributes should be saved with each measurement
(none/user/ID/user and ID).
If applicable, enter the user and/or ID list under
<Menu>/<Settings>/ <Identification>/<ID List> and
<Menu>/<Settings> /<Identification>/ <List of Users>.
8A-96.730.101 / 100519
Page 11
Chematest 30
83%
[Method 1]
[Method 2]
Main Menu
[Method 3]
Method Catalogue
Menu
A
B
Initial Setup
3.5.Optional: Set up a fixed quick start list
By default, the instrument is configured to display the last three
methods selected from the method catalogue [B] in the quick start
list [A]. The list is automatically adjusted every time a method is selected from the method catalogue.
Alternatively, it is possible to set up a fixed quick start list. To do
this, proceed as follows:
1 Select the required methods from the method catalogue [B] in
2 Navigate to <Menu> / <Settings> / <Miscellaneous> / <Method
ABQuick start list
Method catalogue
reverse order: First number 3, then number 2, then number 1
(the measurement can be cancelled immediately after selecting
the method by pressing [Exit]).
save> and select "No".
The quick start list contains the three methods just selected
and will no longer change.
A-96.730.101 / 1005199
3.6.Further settings
Detailed descriptions of all instrument settings can be found in
chapter Menu explanation, subsection 4 Settings, p. 30.
Page 12
Chematest 30
Measuring
4.Measuring
4.1.Basic Rules for Photometric Determinations
The precision and repeatability of a photometric determination depends greatly on the operator's technique. Please observe the following rules:
Always use clean utensils
Residues from previous measurements can falsify the results. All
utensils such as syringes, cuvettes or similar and the cuvette cap
must be rinsed with clean water after each measurement. Rinse the
cuvette with the sample 2 to 3 times before each measurement.
Position cuvettes correctly
Cuvettes must always be positioned in the same way. Keep the cuvette free of fingerprints and dry the outside of the cuvette before
inserting it into the cuvette adapter.
Wipe off condensation
If condensation occurs on the cuvette (cold sample in a warm environment), wipe it off and carry out the measurement as quickly as
possible.
Observe proper sampling
Take the water samples about 10 cm below the water surface and
about 50 cm from the edge of the pool. Allow as little time as possible to elapse between taking the sample and making the measurement.
Observe proper dosing and mixing
Use the 10 ml syringe for exact dosing of the sample. Rinse the syringe several times with the sample. When filling the syringe, make
sure that there are no air bubbles and that the volume is 10 ml.
The order in which sample and reagents are added to the cuvette
varies from method to method. If the reagents are added to the cuvette after the sample, special care must be taken to ensure that
the sample and the reagents mix well. For this purpose, carefully
turn the closed cuvette over, but do not shake it.
10A-96.730.101 / 100519
Page 13
Chematest 30
Measuring
Observe reaction times
With most methods, the measurement can be carried out immediately after the reagents have been added and mixed with the sample. If a reaction time is required, this is indicated to the user on the
screen.
4.2.How To Use Reagents
Whenever possible, Swan supplies reagents in liquid form, as used
in professional laboratories. We believe that the disadvantage of
shorter shelf life is more than compensated for by the ease of use
compared to tablets or powders.
The OXYCON-DPD reagent is delivered in two small bottles to increase shelf life. One contains DPD as powder (DPD 1a), the other
contains the solvent (DPD 1b). Before use, fill the contents of DPD
1b into DPD 1a, close the bottle with the drop counter and shake
firmly until the DPD powder has completely dissolved. Write the
mixing date on the bottle. The prepared reagent can be stored at
room temperature for 2 months.
The reagents OXYCON START and OXYCON 2 have a shelf life of
at least 6 months. They will last a year if handled carefully and
stored at 5 °C. With the blister pack, all reagents can be taken out
of the case at once and placed in the refrigerator.
To dispense the necessary number of drops, hold the bottle at a 45°
angle. Before proceeding with the measurement, make sure that
the reagents are well mixed with the sample.
The Chematest 30 is calibrated against Swan's OXYCON reagents.
Accuracy may suffer severely, when reagents from other sources
are used.
A-96.730.101 / 10051911
Page 14
Chematest 30
Measuring
4.3.Notes on Individual Methods
Free chlorineWhen free chlorine is determined in the presence of cyanuric acid,
two forms of chlorine are measured: free chlorine and chlorine
bound to cyanuric acid. To determine the disinfection capacity of
the water, the concentration of cyanuric acid must also be measured and the result of the chlorine determination must be corrected as follows:
Cyanuric acid20 mg/l30 mg/l50 mg/l70 mg/l
Percentage of free
chlorine in the measured value
50%43%26%19%
Combined
chlorine
OzoneOzone decomposes in a very short time and is usually only present
The method catalogue of the Chematest 30 contains two options
for measuring combined chlorine: with one cuvette or with two cuvettes. Apart from the fact that one or two cuvettes are used, the
two methods are identical. Measuring with only one cuvette is faster and requires less reagents. However, some standards require
measurement with two cuvettes. The selection of the method
should be based on the site-specific requirements.
in traces. It is essential to rinse the cuvette three times with the
ozone-containing water before filling in the actual sample. Carry out
the further steps as quickly as possible, but work meticulously.
12A-96.730.101 / 100519
Page 15
Chematest 30
Measuring
4.4.Performing a Photometric Measurement
WARNING
Health hazard
For safe handling of the reagents, you must read and understand the corresponding Material Safety Data Sheets (MSDS).
These can be downloaded from www.swan.ch.
Explanation of
photometric
measurements
Starting a
measurement
The basic procedure is largely the same for all photometric measurements and is briefly outlined below. A photometric measurement consists of the following steps:
Step 1: Zero measurement
Step 2: The actual measurement
Step 3: Display and storage of the measured results. For de-
tails on the result screen, see example on 15.
The methods differ mainly in the following points:
Different reagents are required depending on the method.
Some methods require a reaction time. If this is the case for
the selected method, this will be indicated on the screen
during the measurement.
Some methods (e.g. combined chlorine) perform two different
measurements in a row and calculate the required values
from them.
To start a measurement, select the corresponding method from the
main screen. If it is not available there, select it from the method
catalogue.
The measurement itself is menu-driven. Follow the instructions on
the screen.
A-96.730.101 / 10051913
Page 16
Chematest 30
83%
Free chlorine
Place the adapter in
the Chematest and
close the cap.
------------------------------------------
<Enter> for zero
83%
Free chlorine
Measuring zero...
------------------------------------------
Wait for instructions
83%
Free chlorine
Add 5 drops of both:
Oxycon START and then
Oxycon DPD.
------------------------------------------
<Enter> to continue
83%
Free chlorine
Close the cap and
turn the adapter to mix.
Place in Chematest.
------------------------------------------
<Enter> to measure
83%
Free chlorine
Measuring...
------------------------------------------
Wait for instructions
Measuring
Example:
Measurement of
free chlorine
Step 1: Zero measurement
Free chlorine
Fill 10 ml of sample into
a cuvette. Dry outer part
and place it in the adapter.
------------------------------------------
<Enter> to continue
83%
Step 2: The actual measurement
Free chlorine
Remove adapter
with cuvette and
open the cap.
------------------------------------------
<Enter> to continue
83%
14A-96.730.101 / 100519
Page 17
Chematest 30
Measuring
Step 3: Display of the results
Free chlorine
Free chlorine
ID ID1
User User1
------------------------------------------
Free chlorine
ID ID1
User User1
------------------------------------------
Free chlorine
ID ID1
User User1
------------------------------------------
0.26
<Enter> to save
Please select
<Enter> to continue
ID
Please select
---------------
ID1
ID2
ID3
ID4
<Enter> to continue
User
Please select
------------------
User 1
User 2
User 3
User 4
<Enter> to continue
ppm
Pressing [Enter] saves the
83%
measurement in the data history
The measurement is initially
saved without ID and user.
NOTICE: Pressing [Exit]
discards the measurement.
83%
Depending on the
configuration, a selection
dialog for user and/or ID is
displayed or the summary
screen is displayed directly.
This can be set under
Attribute, p. 25.
Optional:
83%
– Select an ID from the list.
– Select a user from the list.
NOTICE: If you press [Exit] at
this point, you return to the main
screen, saving the measurement
83%
without user and ID.
Select "<Enter> to continue" and
Free chlorine
Please select
ID ID1
User User1
------------------------------------------
<Enter> to continue
Summary
Date, Time
Free chlorine
------------------------------------------
05.11.18 15:21
0.26 ppm
<Enter> to finish
83%
confirm with [Enter].
The ID and the user are
added to the measurement in
the data history.
Press [Enter] to return to the main
83%
screen.
A-96.730.101 / 10051915
Page 18
Chematest 30
Measuring
4.5.Automatic Shutoff
If the instrument is waiting for a user input (e.g. if the measurement
result is displayed) and no key is pressed for 10 minutes, the instrument shuts off automatically to save power. The automatic shutoff
has the same effect on the saving of measurement results as
pressing the [Exit] key. See Step 3: Display of the results, p. 15 for
details.
16A-96.730.101 / 100519
Page 19
Chematest 30
4
83%
Data History
Date, Time
Free chlorine
ID
User
20.03.19 07:59
0.26 ppm
ID 1
User 1
5
83%
Data History
Date, Time
Free chlorine
Total chlorine
Comb. chlorine
20.03.19 08:25
0.22 ppm
0.26 ppm
0.04 ppm
<Enter> for more
5
83%
Data History
ID
User
ID 1
User 1
<Enter> for more
Recalling Stored Measurement Data
5.Recalling Stored Measurement Data
Data historyPrevious measurements can be displayed under the menu item
<Menu> / <Data Storage> / <Data History>.
Scrolling
through the
data history
Press the ARROW DOWN and
ARROW UP keys to scroll through
the data history.
Two-page
entries
Storage
capacity
If the text “<Enter> for more” is
displayed on the bottom, the
selected entry consists of two
pages. Use the ENTER key to
toggle between the pages.
Up to 2700 measurements are memorized. Then the oldest measurement is deleted to save the newest one (circular buffer).
A-96.730.101 / 10051917
Page 20
Chematest 30
B
A
C
Maintenance
6.Maintenance
6.1.Photometer Verification
The verification kit contains two reference cuvettes (labeled “Low”
and “High”) with certified reference absorptions of approximately
0.3 and 1.5 and an additional cuvette for zero point determination.
The exact values are specified in the enclosed calibration certificate. The absorption values of the reference cuvettes simulate approximately 1 ppm and 4.5 ppm of free chlorine.
A
Zero
B
Low absorption
C
High absorption
Reference
values
Verification
procedure
Verification
history
Prior to performing the first verification and after each recertification, the reference values for the cuvettes “Low” and “High” need to
be set in menu <Settings>\<Sensors> \ <Photometer>. The reference values are specified in the enclosed calibration certificate.
Before each use, check the expiration date on the calibration certificate. If the expiration date has passed, send the verification kit to
Swan for recertification.
To start a verification, navigate to <Menu>/<Maintenance>/
<Verification>/ <Photometer>. Select cuvette “Low” and follow the
instructions on the screen. Repeat the same procedure with the cuvette “High”.
Can be reviewed in <Menu>/<Data Storage>/<Ver. History>.
Up to 32 verifications are memorized. Then the oldest verification is
deleted to save the newest one (circular buffer).
18A-96.730.101 / 100519
Page 21
Chematest 30
A-70.065.030
CHEMATEST
30
xxx Vers.: 1-A
Prod. No.:
Model:
Type:
Serial No.:
Made in Switzerland
B
A
B
Maintenance
6.2.Cleaning
Outside of
the housing
Photometer
compartment
If reagents drip onto the housing, wipe them off quickly with a soft
cloth moistened with water.
The photometer compartment can be easily cleaned if water or reagents have been spilled. Turn the unit to the rear, loosen the two
screws [A] and remove the cover [B]. Use a soft, lint-free cloth
moistened with water.
A ScrewsB Photometer cover
A-96.730.101 / 10051919
Page 22
Chematest 30
Maintenance
6.3.Setting Time and Date
The correct setting of the clock is important for the identification of
the stored measurements. It is therefore recommended to check
the date and time setting regularly.
The time and date can be set under the menu item <Menu> /
<Maintenance>/<Set Time>.
20A-96.730.101 / 100519
Page 23
Chematest 30
Troubleshooting
7.Troubleshooting
This chapter contains some hints to facilitate troubleshooting. For
detailed information on handling and cleaning parts, see Mainte-
nance, p. 18. For detailed information on programming the instru-
ment, see Menu Explanations, p. 23.
7.1.Device Errors
When this symbol is displayed at the top of the screen, a device error has occurred. Navigate to <Menu> /<Diagnostics>/<Alarms> /
<Pending errors> to see the error message.
ErrorDescriptionCorrective action
E001Bluetooth
E002IC ADC
E003Factory data
E004Invalid time
E017Events erased
– Switch the instrument on and off and check
if the error disappears.
– Call support
– Call support.
– Send the instrument back to Swan.
– Call support.
– Send the instrument back to Swan.
– Set time and date under <Menu> /
<Maintenance>/<Set Time>.
– No action necessary.
– This message informs that the measured
data have been deleted by calling the
function "Set to factory defaults”.
A-96.730.101 / 10051921
Page 24
Chematest 30
Troubleshooting
7.2.Errors During Measurement
Measurement
cannot be
started
Zero measure-
ment fails
Measurement
fails
Error message: Operation not possible! Consult the manual.
Possible causeCorrective action
Error E002, E003 or E004
is present
Error message: Operation not possible! Signal out of range.
Consult the manual.
Possible causeCorrective action
Extraneous light
Cuvette dirty
Residues of reagents
Lenses dirty
Turbid or bubble-containing sample
Error message: Operation not possible! Signal out of range.
Consult the manual.
Possible causeCorrective action
Extraneous light
Cuvette dirty
Concentration or pH value
outside the measuring
range of the instrument
Reagents expired
Measured value below
zero value
See Device Errors, p. 21.
Make sure that the cuvette cap is
properly closed.
Clean and rinse the cuvette.
Clean and rinse the cuvette.
Open the photometer
compartment and clean all
lenses, see Cleaning, p. 19.
Selecting <Menu> on the main screen displays the following submenus:
Menu 1 Data Storage
Menu 2 Diagnostics
Menu 3 Maintenance
Menu 4 Settings
The functions and settings of each menu are explained in the following sections.
1 Data Storage
1.1 Data History
Displays saved measurement data. Each record includes the following data:
Date, time
Measured value(s)
ID
User
If the fields ID and User contain the text "<Empty>", the assignment
of the attributes to the measurement was skipped by the user or deactivated under 4.2 Identification, p. 25.
Up to 2700 measurements are memorized. Then the oldest measurement is deleted to save the newest one (circular buffer).
1.2 Verification History
PhotometerDisplays previous photometer verifications. Each record includes
the following data:
Date, time
Reference value
Deviation
User
Verification passed/failed
Up to 32 verifications are memorized. Then the oldest verification is
deleted to save the newest one (circular buffer).
A-96.730.101 / 10051923
Page 26
Chematest 30
Menu Explanations
2 Diagnostics
2.1 Alarms
Pending errorsProvides the list of active errors with their status (active, acknowl-
Message ListShows the error history: Error code, date and time of issue and sta-
2.2 Identification
2.3 Sensors
PhotometerDisplays the raw values of the photo diode in volts.
BatteryDisplays information about the state of the battery.
2.4 Bluetooth
Device nameID that is visible to other Bluetooth-enabled devices.
StateStatus of Bluetooth communication.
MACHardware address of the Chematest 30, which serves as a unique
Fw verFirmware version of the Bluetooth module.
Hw verHardware version of the Bluetooth module.
edged). If an active error is acknowledged, it is moved to the Message List.
tus (active, acknowledged, cleared). 64 errors are memorized.
Then the oldest error is cleared to save the newest error (circular
buffer).
The meaning of each error message is explained in section Device
Errors, p. 21.
Designation: Designation of the instrument, e.g. Chematest 30.
S/N: Serial number of the instrument.
Version: Installed firmware version (e.g. V1.03-04/19).
Bootloader: Installed bootloader version (e.g. V.1.01).
Factory test: Test date of the instrument.
Disabled: Bluetooth is switched off.
Advertising: The instrument is searching for other devices.
Connected: A connection with another device has been
established.
identification feature during Bluetooth communication.
24A-96.730.101 / 100519
Page 27
Chematest 30
Menu Explanations
3 Maintenance
3.1 Simulation
For test purposes a measurement can be made with one or both
LEDs on. The results are displayed as raw values in volts. In case
of problems in the field, Swan Support may ask for these values.
LED GreenPerforms a test measurement with the green LED on.
LED IRPerforms a test measurement with the IR LED on.
Both LEDsPerforms three consecutive measurements:
both LEDs off
green LED on
IR LED on
3.2 Verification
PhotometerSee Photometer Verification, p. 18.
3.3 Set Time
Adjust date and time.
4 Settings
4.1 Sensors
PhotometerEnter the reference values of the verification cuvettes according to
4.2 Identification
AttributeSelect which attributes are to be queried:
ID ListEnter up to 10 IDs.
User ListEnter up to 10 user names.
the enclosed calibration certificate.
Range: 0.000–2.000
Optionally, each measurement can be assigned an ID (e.g. for
sampling points) and/or a user name. If activated, the ID and/or
user is queried at the end of each measurement.
Attribute
None
ID
User
ID&User
A-96.730.101 / 10051925
Page 28
Chematest 30
Menu Explanations
4.3 Units
DisinfectantsSet the unit for disinfectants (all except ozone).
Disinfectants
ppm
mg/l
OzoneSet the unit for ozone.
Ozone
ppb
µg/l
ppm
mg/l
4.4 Miscellaneous
LanguageSelect the language from the list (the choice of languages may vary
depending on the sales region):
Language
German
English
French
Spanish
Italian
Factory SettingThe instrument can be reset to factory default values in three differ-
ent ways:
Set defaults
No
Cal. + Ver.
Data History
Completely
Cal. + Ver.: Deletes the user calibration of all electrodes (not
relevant for Chematest 30) and all recorded verification
measurements. All other values are kept in memory.
Data History: Deletes all recorded measurements. All other
values are kept in memory.
Completely: Sets all settings and calibration values back to
default and deletes all recorded measurements and verifications.
26A-96.730.101 / 100519
Page 29
Chematest 30
Menu Explanations
Load FirmwareInitiates a firmware upload.
NOTICE: Firmware updates should be done by instructed
service personnel only.
Load Firmware
No
Yes
Expert modeHere you can choose between detailed instructions and expert
Method saveBy default, the quick start list on the main screen contains the last
PasswordPassword: Select a password different from 0000 to prevent unau-
mode for fast workflow.
Expert Mode
Inactive
Active
Inactive: The user is guided step by step through each
measurement.
Active: The steps are displayed in shortened form and the
instrument requires as few inputs from the user as possible.
three methods selected from the method catalogue. Alternatively, a
fixed quick start list can be set up, see Initial Setup, p. 8 for details.
Method save
No
Yes
No: As soon as "no" is selected, the quick start list no longer
changes. Select this setting if you want a fixed quick start list.
Yes: With this setting, the quick start list is automatically adjusted
every time a method is selected from the method catalogue.
thorized access to the following menus:
Maintenance
Settings
Each menu can be protected by a different password. If you forgot
the passwords, contact the closest SWAN representative.
A-96.730.101 / 10051927
Page 30
Chematest 30
Menu Explanations
4.5 Bluetooth
BluetoothActivates the Bluetooth connection to connect to an external app
(not yet available).
Bluetooth
Active
Inactive
28A-96.730.101 / 100519
Page 31
Chematest 30
Part Numbers
9.Part Numbers
9.1.Reagents
Reagent sets
Reagents
Part no.Product nameMethods
A-85.590.200Oxycon Pool
A-85.590.300Oxycon Chlor
A-85.590.400Oxycon Des
A-85.590.500Oxycon Ozone
A-85.580.300Oxycon CA
A-85.580.100Oxycon pH
Part no.Product nameDescription
A-85.510.100Oxycon StartBuffer solution
A-85.510.200Oxycon DPD
(1a + 1b)
A-85.510.300Oxycon 2Potassium iodide
Free chlorine
pH (phenol red)
Free chlorine
Total chlorine
Combined chlorine
Free chlorine
Free chlorine in presence
of chlorine-dioxide or
bromine
Chlorine-dioxide
Bromine
Iodine
Ozone
Ozone in presence of free
chlorine
Cyanuric acid
pH (phenol red)
DPD
A-85.580.200Oxycon GLReagent for masking of free
chlorine and ozone
Download
MSDS
A-96.730.101 / 10051929
The current Material Safety Data Sheets (MSDS) for the above listed reagents are available for downloading at www.swan.ch.
Page 32
Chematest 30
Part Numbers
9.2.Accessories
A-85.153.580Veri-Kit Transmission
9.3.Spare Parts and Consumables
A-70.065.204Carrying case for Chematest 30
A-70.065.630Cuvette adapter with cap
A-70.065.631Glass cuvettes (10-pack)
A-70.065.633Spare part set consisting of:
C-85.520.040Cleaning wipes for cuvette (100-pack)
A-70.065.600Brush for cuvette cleaning
C-88.917.200USB charging cable
Part no.Description
Part no.Description
1 dropper bottle 125 ml
3 disposable syringes 10 ml
3 disposable syringes 20 ml
C-89.511.010Blister pack for reagent bottlesl
30A-96.730.101 / 100519
Page 33
Chematest 30
Product Description
10.Product Description
10.1.List of Supported Methods
PhotometryThe Chematest 30 supports the following photometric methods:
Chlorine
Other
disinfectants
Other
parameters
MethodRangeIssued valuesDisplayed
Free chlorine 0– 10 ppmFree chlorineppm or mg/l
Total chlorine 0–10 ppmTotal chlorineppm or mg/l
Combined
chlorine using
one cuvette
Combined
chlorine using
two cuvettes
Free chlorine
in presence of
chlorinedioxide or
bromine
Ozone0–4000 ppbOzoneppb, µg/l,
Ozone in
presence of
free chlorine
Bromine0–23 ppmBromineppm, mg/l
Iodine0–35 ppmIodineppm, mg/l
Chlorinedioxide
Cyanuric acid 0 – 100 ppmCyanuric acidppm
pH (phenol
red)
0–10 ppmFree chlorine (fac),
total chlorine (tc),
combined chlorine
(cc)
0–10 ppmFree chlorine (fac),
total chlorine (tc),
combined chlorine
(cc)
0–10 ppmFree chlorineppm or mg/l
0–4000 ppbOzoneppb, µg/l,
0–19 ppmChlorine-dioxideppm, mg/l
6.5–8.0pHn/a
as
ppm or mg/l
ppm or mg/l
ppm or mg/l
ppm, mg/l
A-96.730.101 / 10051931
Page 34
Chematest 30
Product Description
10.2.Instrument Specification
HousingIP67 rated
Power supplyRechargeable lithium-ion battery.
Only charge the instrument using a standard type A USB socket
and the supplied USB cable. Protect from heat and splash water
during charging (not IP67).
Charging time: approximately 3 hours
Battery life: approximately 5000 measurements or one week of
standby operation
Automatic shutoff after 10 minutes of inactivity.
Conditions for
operation
Conditions for
storage:
Dimensions
and weight
Temperature: 0– 45 °C
Humidity: 0–100 % relative humidity (non-condensing)
Temperature: −20– 60 °C
For reagents, separate storage conditions apply. See label on
packaging.
Instrument:
10 x 7 x 22 cm, 390 g
Carrying case with content:
40 x 10 x 38 cm, 1.7 kg
32A-96.730.101 / 100519
Page 35
Chematest 30
H
I
G
A
B
C
D
E
F
Product Description
10.3.Scope of Supply
A
Carrying case
B
10 ml syringe
C
Dropper bottle for rinsing
water
D
Brush for cuvette cleaning
E
Wipes for cuvette cleaning
A-96.730.101 / 10051933
F
2 glass cuvettes
G
Blister pack for reagent
bottles
H
Chematest 30
I
USB charging cable (located
under the Chematest 30)
Page 36
Chematest 30
Regulatory Information
11.Regulatory Information
This product contains a third party Bluetooth module, which has
been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC Rules.
Contains FCC ID: T9JRN4020
This product meets all requirements of the relevant EU directives.
The declaration of conformity is available from Swan on request.
According to EU legislation, this product must not be disposed of
with household waste. Observe the local regulations for the disposal of electronic equipment.
34A-96.730.101 / 100519
Page 37
Chematest 30
Notes
12.Notes
A-96.730.101 / 10051935
Page 38
Chematest 30
Notes
36A-96.730.101 / 100519
Page 39
Chematest 30
Notes
A-96.730.101 / 10051937
Page 40
Chematest 30
SWAN
is represented worldwide by subsidiary companies
and distributors.
cooperates with independent representatives
all over the world.
SWAN Products
Analytical Instruments for:
High Purity Water
Feedwater, Steam and Condensate
Potable Water
Pool and Sanitary Water
Cooling Water
Waste Water and Effluents
Made in Switzerland
38A-96.730.101 / 100519