AMI INSPECTOR Resistivity
Default Values
A-96.250.781 / 171017 59
Alarm Relay: Alarm:
Resistivity Alarm high: .......................................................................... 200 M
Alarm low:........................................................................... 0.00 M
Hysteresis:.......................................................................... 1.00 M
Conductivity Alarm high: ..........................................................................2000 µS
Alarm low:.......................................................................... 0.000 µS
Hysteresis:..........................................................................10.00 µS
Delay: ......................................................................................5 Sec
Sample Flow; Flow Alarm........................................................... yes
Sample Flow; Alarm high.......................................................120 l / h
Sample Flow; Alarm low............................................................5 l/h
Sample Temp; Alarm high:....................................................... 90 °C
Sample Temp; Alarm low: .......................................................... 0 °C
Case temp. high: ..................................................................... 65 °C
Case temp. low:......................................................................... 0 °C
Relay 1 and 2 Function:.......................................................................... limit upper
Parameter:.................................................................... Meas. Value
Resistivity Setpoint: .............................................................................. 200 M
Hysteresis:............................................................................... 1 M
Conductivity Setpoint: ..............................................................................1000 µS
Hysteresis:..........................................................................10.00 µS
Delay: ....................................................................................30 Sec
If Function = Control upw. or dnw:
Parameter:....................................................................Meas. Value
Settings: Actuator: ..........................................................Frequency
Settings: Pulse Frequency: .............................................120 / min.
Resistivity Settings: Control Parameters: Setpoint: ........................... 200 M
Settings: Control Parameters: P-band: ................................ 1 M
Conductivity Settings: Control Parameters: Setpoint: ...........................1000 µS
Settings: Control Parameters: P-band: ...........................10.00 µS
Parameter:...................................................................Temperature
Settings: Actuator: ..........................................................Frequency
Settings: Pulse Frequency: .....................................
........120 / min.
Settings: Control Parameters: Setpoint: ...............................50 °C
Settings: Control Parameters: P-band: ..................................1 °C
Parameter:................................................................... Sample flow
Settings: Actuator: ..........................................................Frequency
Settings: Pulse Frequency: .............................................120 / min.
Settings: Control Parameters: Setpoint: ............................25.0 l/h
Settings: Control Parameters: P-band: ..................................1 l/ h