5.6.5 Move on Strobe Pulse Mode..................................................................................................24
5.6.6 Move on Sync, Inhibited by the Strobe Line Mode............................................................25
5.6.7 Edit a Filter Definition Mode ...............................................................................................27
5.6.7.1 General Information .................................................................................................................................................27
5.6.7.4 Programming a Filter Number (“Logical Filter”)\..............................................................................................29
5.6.7.5 Manual Adjustment of Attenuation .......................................................................................................................30
5.6.8 Setup Ring Buffer Mode........................................................................................................33
5.6.8.1 General Information .................................................................................................................................................33
5.6.8.2 Storing the FILTER NUMBER Sequence ............................................................................................................33
5.6.8.3 Viewing an Existing FILTER NUMBER Sequence ............................................................................................34
5.6.8.4 Running the RING BUFFER Sequence................................................................................................................35
The Lambda DG-4 is an illumination system designed for rapid change of wavelength and
comprises three subsystems: controller, optical path and 175-Watt Xenon arc lamp with
power supply. All three subsystems are conveniently housed in one cabinet. For the most
part this manual is organized around those subsystems to provide you with ready access to
information, as you need it. Separate documentation for the lamp and its power supply
(provided by PerkinElmer Optoelectronics) accompanies this manual.
In the next few pages you will find information on the safe use of this device, unpacking
instructions and quick-start (“Make it Glow”) directions to help you get acquainted with the
operation of the Lambda DG-4 (using the manual mode). Please take the time to read these
instructions to assure the safe and proper use of this instrument.
This manual is currently under construction. If there are any areas that you feel should be
covered in greater detail we would like to hear from you. Please contact our Technical
Support staff with your suggestions (415-883-0128 or info@sutter.com).
1
1.2
1.2 Lambda DG
Lambda DG----4 and Lambda DG
1.21.2
Lambda DGLambda DG
4 and Lambda DG----5555
4 and Lambda DG4 and Lambda DG
Throughout this manual, unless otherwise noted, all references and descriptions pertaining
to “Lambda DG-4” (or just “DG-4”) apply equally to the Lambda DG-5. The only difference
between the Lambda DG-4 and Lambda DG-5 is the set of light interference filters, as
follows:
• The Lambda DG-4 filter set consists of four 25mm (1-inch) diameter filters.
• The Lambda DG-5 filter set consists of three 18mm and two 25mm (1-inch) diameter
filters.
Both models of the Lambda DG series are calibrated at the factory according to each one’s
filter set arrangement.
1.3
1.3 Unpacking
Unpacking
1.31.3
UnpackingUnpacking
The Lambda DG-4 and associated hardware comes packed in a single carton. The following
is a list of the components found there. If you believe that any of these components are
missing or show obvious signs of damage from shipping, please contact the factory.
Parallel Interface Cable
Serial Interface Cable
Power Cord
4 Individual Filter holders with retaining rings, spacer rings and blanking discs
Neutral Density holder retaining ring, spacer ring and blanking disc
Replace fuse only with the same type Replace fuse only with the same type
5 x 20 mm glass tube
5 x 20 mm glass tube
5 x 20 mm glass tube5 x 20 mm glass tube
TTTTime delay
ime delay (medium)
ime delayime delay
RRRRating:
ating:
ating:ating:
If this instrument is equipped with
If this instrument is equipped with a
If this instrument is equipped with If this instrument is equipped with
삟 100
삟 220
If this instrument is equipped with a
If this instrument is equipped with a 300
If this instrument is equipped with a If this instrument is equipped with a
삟 100
삟 220
A spare fuse is provided, being located in the power input module. Please refer to
A spare fuse is provided, being located in the power input module. Please refer to
A spare fuse is provided, being located in the power input module. Please refer to A spare fuse is provided, being located in the power input module. Please refer to
Appendix A
Appendix A:
Appendix AAppendix A
instructions on ho
instructions on how to change the fuse.
instructions on hoinstructions on ho
(medium),
(medium) (medium)
100 –––– 120 VAC Operation: 4.0A, 125V fuse; or
120 VAC Operation: 4.0A, 125V fuse; or
100 100
120 VAC Operation: 4.0A, 125V fuse; or 120 VAC Operation: 4.0A, 125V fuse; or
120 VAC Operation: 5.0A, 125V fuse, or 120 VAC Operation: 5.0A, 125V fuse, or
220 –––– 240 VAC Operation: T
240 VAC Operation: T3.15A, 250V fuse.
220 220
240 VAC Operation: T 240 VAC Operation: T
: FUSE
FUSE REPLACEMENT
: :
FUSEFUSE
w to change the fuse.
w to change the fuse.w to change the fuse.
, time lag
time lag,,,, or
, ,
time lagtime lag
REPLACEMENT for more details on fuse ratings and for
REPLACEMENTREPLACEMENT
or ““““slow blow
slow blow””””))))
or or
slow blowslow blow
a 175
175----Watt lamp
a a
3.15A, 250V fuse.
3.15A, 250V fuse.3.15A, 250V fuse.
for more details on fuse ratings and for
for more details on fuse ratings and for for more details on fuse ratings and for
Watt lamp, replace with
175175
Watt lampWatt lamp
300----Watt lamp
Watt lamp, replace with
300300
Watt lampWatt lamp
, replace with
, replace with, replace with
, replace with
, replace with, replace with
Infrared Radiation: The infrared radiation (and ultraviolet radiation) generated by
Infrared Radiation: The infrared radiation (and ultraviolet radiation) generated by
Infrared Radiation: The infrared radiation (and ultraviolet radiation) generated by Infrared Radiation: The infrared radiation (and ultraviolet radiation) generated by
this lamp can cause significant skin burns and eye damage.
this lamp can cause significant skin burns and eye damage.
this lamp can cause significant skin burns and eye damage.this lamp can cause significant skin burns and eye damage.
Explosion: High internal pressure
Explosion: High internal pressure exists in any xenon arc lamp.
Explosion: High internal pressureExplosion: High internal pressure
High Voltage: High ignition voltages, which exist inside the cabinet, can be lethal.
High Voltage: High ignition voltages, which exist inside the cabinet, can be lethal.
High Voltage: High ignition voltages, which exist inside the cabinet, can be lethal.High Voltage: High ignition voltages, which exist inside the cabinet, can be lethal.
Ozone: Because the UV
Ozone: Because the UV----enhanced version of the lamp generates significant ozone, it
Ozone: Because the UVOzone: Because the UV
must be used in a well
must be used in a well----ventilated environment. For mos
must be used in a wellmust be used in a well
requirement to ventilate the ozone
requirement to ventilate the ozone----free lamp.
requirement to ventilate the ozonerequirement to ventilate the ozone
2.1
2.1 To Further Avoid Electrical Shock
To Further Avoid Electrical Shock and Fire
2.12.1
To Further Avoid Electrical ShockTo Further Avoid Electrical Shock
Always use the grounded power supply cord set provided to connect the unit to a
Always use the grounded power supply cord set provided to connect the unit to a
Always use the grounded power supply cord set provided to connect the unit to a Always use the grounded power supply cord set provided to connect the unit to a
grounded outlet (3
grounded outlet (3----prong). This is required to protect you from injury i
grounded outlet (3grounded outlet (3
an electrical hazard occurs.
an electrical hazard occurs.
an electrical hazard occurs.an electrical hazard occurs.
Do not disassemble the unit. The only user serviceable parts are the line fuse, the
Do not disassemble the unit. The only user serviceable parts are the line fuse, the
Do not disassemble the unit. The only user serviceable parts are the line fuse, the Do not disassemble the unit. The only user serviceable parts are the line fuse, the
xenon bulb, and filters. The line fuse is accessible from the outside of the unit. The
xenon bulb, and filters. The line fuse is accessible from the outside of the unit. The
xenon bulb, and filters. The line fuse is accessible from the outside of the unit. The xenon bulb, and filters. The line fuse is accessible from the outside of the unit. The
bulb and the filters are accessible v
bulb and the filters are accessible via separate panels on the side of the unit. Bulb
bulb and the filters are accessible vbulb and the filters are accessible v
replacement and filter installation/removal are covered in separate sections of the
replacement and filter installation/removal are covered in separate sections of the
replacement and filter installation/removal are covered in separate sections of the replacement and filter installation/removal are covered in separate sections of the
manual and should only be attempted with the power cord disconnected.
manual and should only be attempted with the power cord disconnected.
manual and should only be attempted with the power cord disconnected.manual and should only be attempted with the power cord disconnected.
To prevent fire or shock hazard do not expose the un
To prevent fire or shock hazard do not expose the unit to rain or moisture.
To prevent fire or shock hazard do not expose the unTo prevent fire or shock hazard do not expose the un
enhanced version of the lamp generates significant ozone, it
enhanced version of the lamp generates significant ozone, it enhanced version of the lamp generates significant ozone, it
ventilated environment. For most users, there is no
ventilated environment. For mosventilated environment. For mos
prong). This is required to protect you from injury in the event that
prong). This is required to protect you from injury iprong). This is required to protect you from injury i
exists in any xenon arc lamp.
exists in any xenon arc lamp. exists in any xenon arc lamp.
t users, there is no
t users, there is no t users, there is no
free lamp.
free lamp.free lamp.
and Fire----related
and Fire and Fire
ia separate panels on the side of the unit. Bulb
ia separate panels on the side of the unit. Bulb ia separate panels on the side of the unit. Bulb
related Injury
relatedrelated
Injury
Injury Injury
it to rain or moisture.
it to rain or moisture.it to rain or moisture.
n the event that
n the event that n the event that
Finally, the original manufacturer of the xenon lamp and power supply used in the
Finally, the original manufacturer of the xenon lamp and power supply used in the
Finally, the original manufacturer of the xenon lamp and power supply used in the Finally, the original manufacturer of the xenon lamp and power supply used in the
Lambda DG
Lambda DG----4 provides the additional safety information on the following page. It is
Lambda DGLambda DG
intended to amplify the information given above.
intended to amplify the information given above.
intended to amplify the information given above.intended to amplify the information given above.
4 provides the additional safety information on the following page. It is
4 provides the additional safety information on the following page. It is 4 provides the additional safety information on the following page. It is
Proper use and safe operating practices are the responsibility
of equipment manufacturers who incorporate the lamp into
equipment and users of such lamps and equipment. The
supplier of this lamp provides information on its products and
associated hazards, but it assumes no responsibility for aftersale operating and safety practices. All lamps are under
pressure and must be handled with care. Take appropriate
action through baffles, light shields, interlock switches or
other safeguards to protect personnel from harm due to
operation and/or failure of the lamp.
SAFE OPERATING INSTRUCTIONS
Do not operate this lamp except in accordance with proper
operating instructions and within recommended operating
specifications. Direct questions regarding lamp operation or
safety to your lamp supplier.
OPERATING HAZARDS
Read the following instructions and take all necessary precautions
I. EXPLOSION - The lamps are filled with xenon gas at very
high pressure. Lamps must be handled with the same care and
caution given any vessel containing these levels of pressure. A
hazard exists if the window or ceramic fractures and may
cause explosive mechanical failure. Face shields or proper
safety glasses are recommended during all handling
operations.
II. HIGH VOLTAGE - Ignition voltage of some lamp models
is very high and can be deadly. If portions of the circuit are
exposed, caution must be used in setup and operation of the
system. The input power must be disconnected from the power
source before attempting any service to the lamp.
III. INFRARED AND ULTRAVIOLET RADIATION - Do
not look directly at operating lamps orb reflected light.
Infrared and ultra violet radiation generated by the lamp can
cause skin burns and permanent eye damage.
CERMAX® Lamps
LAMP DISPOSAL
CERMAX lamps do not have reclaimable parts. Before
disposal, it is recommended to relieve a lamp’s gas pressure by
squeezing the tip-off with pliers until the gas escapes. If gas
pressure is not relieved, care should be taken to discard the
lamp in a landfill and not an incinerator.
SAFETY HAZARDS
The operation of lamps involves one or more of the following
hazards. In the absence of safe operating practices and
precautions, any one of these hazards could result in injury.
IV. OZONE - Some UV type lamps generate ozone, a toxic
gas, by virtue of the ultraviolet radiation. A lamp which gives
off ozone must be operated in a well ventilated area..
V. HOT SURFACES - Portions of the lamp can reach
temperatures of several hundred degrees centigrade and cause
serious burns if touched even after the lamp is turned off.
The optical path subsystem in the Lambda DG-4 is arranged so that 2 mirrors, each mounted
on a fast closed-loop servoed galvanometer, can route the collimated output of the lamp
through one of 4 optical paths (e.g., filter position 1 in the diagram below). Slide-in filter
holders allow 25 mm (or 1 inch) diameter filters to be introduced in each of the four optical
paths. An additional slide-in filter holder, intended for neutral density filters, is located in a
portion of the optical system common to all light paths. After passing through the filter
positions, the lamp output is focused on a 2-meter long UV (ultraviolet) transmitting light
guide, which acts as a conduit for the output to the microscope. At the output of the light
guide, an optical adapter collects the output and delivers it to the microscope.
Figure 3-1. Lambda DG-4 optical pathway.
The following instructions are meant to help you set up the Lambda DG-4 and verify that
the Lambda DG-4’s subsystems are operational using the manual mode of filter selection.
More detailed discussions of controller function, filter installation, the light source,
assignment of filter numbers and remote communications can be found in other sections of
this manual.
1. Connect the power cord to the socket on the back panel of the Lambda DG-4. Make
certain that the voltage selector next to the power cord socket has been properly preset to
the appropriate voltage (110 or 220). Do not turn on the power, yet.
2. Insert either end of the light guide into the port on the top panel of the Lambda DG-4
cabinet (see below) and tighten the thumbscrew to hold it in place. The top of the light
guide end sleeve should be flush with the top surface of the port fitting (see below). This
should result in maximal or near-maximal light output from the light guide. Minor
adjustments up from this position may improve output, and can be easily made once the
system is fully installed.
Figure 3-2. Light guide installation.
3. The other end of the light guide should be installed in the microscope using the
appropriate adapters (see the installation sheet provided with the specific adapter that
has been ordered) or directed away from anyone’s eyes and toward a non-reflective
surface. DO NOT LOOK DIRECTLY INTO THE LIGHT GUIDE!
4. A hose should be connected to the 4” fitting on the back of the Lambda DG-4 cabinet to
ventilate the ozone from the xenon lamp to the environment or to an ozone removal
device.
5. Identify the small access door used to install the neutral density FILTER HOLDER.
This access door is located near the top edge of the left side panel on the Lambda DG-4
cabinet (see below). Loosen the black knob on the access door and slide it up to install
the empty Neutral Density FILTER HOLDER (the neutral density FILTER HOLDER
is the one whose bottom surface is flat rather than grooved). Make certain that the end
nearest the filter cup goes in first. If you would like to install a neutral density filter at
this time please refer to the “Loading Filters” section of this manual.
6. Identify the large access door used to install the interference FILTER HOLDERS. This
access door is located in the middle of the left side panel on the Lambda DG-4 cabinet
(see above). Loosen the black knob on the access door and slide it up to install the
remaining four FILTER HOLDERS (the ones that have a groove on their bottom
surface). Install the two FILTER HOLDERS that contain blanking disks in FILTER
POSITIONS 3 and 4. If you would like to install interference filters at this time, please
refer to the “Loading Filters” section of this manual.
7. Make certain that the free end of the light guide is installed into the microscope or is
directed toward a non-reflective surface and away from anyone’s eyes. DO NOT LOOK
DIRECTLY INTO THE LIGHT GUIDE.
8. Turn on the power to the lamp subsystem using the switch labeled LAMP on the back
panel of the Lambda DG-4 cabinet. Due to the current draw during lamp ignition, it is
important to power up the LAMP first (before the MAIN switch is turned on) to avoid
damage to the circuitry of the other Lambda DG-4 subsystems. This same precaution
applies to the order in which you power up other electronic and computer equipment that
might be powered from or grounded to the same circuit as the one used for the Lambda
DG-4. GO DIRECTLY TO THE NEXT STEP! DO NOT OPERATE THE LAMP FOR
MORE THAN A FEW SECONDS WITHOUT TURNING ON THE MAIN POWER
SWITCH!
9. Within a few seconds of turning on the LAMP switch turn on the power to the optical
path and controller subsystems using the switch labeled MAIN on the back panel of the
Lambda DG-4 cabinet. The Lambda DG-4 will boot up and briefly display the
initialization screen followed by the MAIN MENU (shown on top of next page).
OFF= FILTER NUMBER, TURBO-BLANKING- OFF
OFF= FILTER NUMBER, TURBO-BLANKING- OFF
OFF= FILTER NUMBER, TURBO-BLANKING- OFFOFF= FILTER NUMBER, TURBO-BLANKING- OFF
INPUT = MANUAL MOVE ON COMMAND
INPUT = MANUAL MOVE ON COMMAND
INPUT = MANUAL MOVE ON COMMAND INPUT = MANUAL MOVE ON COMMAND
10. The Lambda DG-4 is under LOCAL (keypad) control upon start-up, as indicated by the
words “INPUT - MANUAL” on the MAIN MENU (see above). MOVE ON COMMAND is
the active operational mode at start-up, which means that filters are selected by pressing
the numeric keys corresponding to the desired FILTER NUMBER (see EDIT A FILTER
NUMBER definition of this value). The default FILTER NUMBER upon start-up is “0”.
This setting will disrupt the optical pathway within the Lambda DG-4 and prevent any
light output at the light guide port. The “0” FILTER NUMBER serves, then, as a
“shutter” whenever it is selected.
11. Press the “1” key followed by ENTER. This FILTER NUMBER is pre-programmed at
the factory to cause the light to be directed through the FILTER HOLDER that is loaded
into position 1 and then onto the light guide with maximal light output (i.e., minimal
attenuation).
12. Now experiment by selecting FILTER NUMBERS 0 through 4. Anytime you select a
single digit FILTER NUMBER you must complete the entry by pressing the ENTER key
or you must precede the single digit with a “0”. FILTER NUMBERS 0,1,2,3 and 4 are
pre-programmed to correspond with the closed “shutter” and the four FILTER
POSITIONS (with minimal attenuation), respectively.
Now that you have completed the “Make it Glow” section of the manual, you should be
familiar with the basic components and operation of the Lambda DG-4. To get set up for an
experimental application you will need to:
Install the LIGHT GUIDE into the OPTICAL HOLDER (if not already in place).
Install the OPTICAL HOLDER into the microscope using the appropriate adapter (if not
already in place).
Install the desired filters into the FILTER HOLDERS (if not already in place).
Program the desired FILTER NUMBERS.
Establish the desired interface between the Lambda DG-4 and your computer.
Each of these steps is discussed in the following sections of the manual.
The light output of the Lambda 10 DG4 is focused onto one end of a liquid light guide. The
output of the liquid light guide can then easily be coupled to your microscope or optical
system using correct mechanical adapter for your microscope. Please note that the light
guide has a minimum-bending radius of 40 mm (about 1.6 inches).
The liquid light guide should have black sleeves over the ferrules on both ends. These
ferrules, or “end sleeves”, are designed to slide into the mounting flange on the lambda DG-4
cabinet and the output adapter. Insert the end of the light guide into the light guide port
until the top of the black end sleeve is even with the top of the light guide port. Tighten the
locking screw on light guide port to secure the end of the light guide in position. This should
result in maximal or near-maximal light output from the light guide. Minor adjustments up
from this position may improve output; the adjustments can be easily made once the system
is fully installed.
The free end of the light guide slides into one end of the optical holder tube (shown below).
The other end of the optical holder tube has a UV grade fused silica lens system (in the
“short cup”). By sliding the light guide toward or away from these lenses, the light guide
output can be collimated or brought to a focus. The optical holder, in turn, slides into the
dovetail adapter, which has a “dovetail” collar on one end. This collar is used to fasten the
entire assembly to the microscope system using an adapter (not shown) that is specific to the
microscope make and model. Refer to the documentation included with each microscope
adapter for further installation instructions.
Figure 4-2. Light guide optical holder.
4.2
4.2 Loading Filters
Loading Filters
4.24.2
Loading FiltersLoading Filters
4.2.1
4.2.1 Filter Orientation
Filter Orientation
4.2.14.2.1
Filter OrientationFilter Orientation
The two sides of a filter are usually not identical due to differences in their surface coatings.
Some form of labeling will be present to differentiate the two sides. It is important to refer
to the filter manufacturer’s documentation to interpret the labeling and determine the
proper orientation relative to the light source. Remember, the light path inside the Lambda
DG-4 travels from BOTTOM TO TOP. Therefore, the filters should be oriented with their
“Lamp Side” down.
At the end of each filter holder is a filter cup assembly (shown below). There are at least two
items installed into the filter cup at the factory: a threaded retaining ring and a spacer ring
(to carry the force of the retaining ring to the unthreaded portion at the bottom of the cup).
In two of the filter cups, there will also be a blanking disc. Installation of filters into the
filter holders requires care and patience to avoid damaging the threads of the filter cups and
retaining rings as well as the filters.
Figure 4-3. Filter holder assembly.
The threaded retaining ring has two notches cut into it, 180 degrees apart. Place the
notched end of the brass spanner wrench into the notches on the retaining ring and carefully
unscrew it. If a retaining ring has been previously over-tightened, it is possible that the filter
cup will begin to unscrew from the slider. If this occurs, STOP. Turn the spanner wrench
around and, using the other end, insert it into the slots at the top of the cup. Then screw the
cup back into the slider. DO NOT push on the cup. Just screw it in until it seats firmly. It
should then be possible to remove the retaining ring.
Once the retaining ring is removed, the spacer (and blank, if present) should fall out. You
should become familiar with the assembly of this retaining system. In particular, you should
develop a feel for tightening the retaining ring without pushing on the wrench. This is
important because if you forcefully push on the filter holder while you are trying to load a
filter, you run the risk of scratching the filter.
Figure 4-4. Filter holder assembly with blanking disc.
Place the desired filter into the filter cup, first. With some 1” filters extra care must be
taken to avoid getting the filters jammed at an angle inside the filter cup.
Next, install the spacer ring into the filter cup. Unless the filter thickness precludes the use
of the spacer ring, always install the spacer to provide maximum separation between the
filter surface and end of the brass spanner wrench that is used to turn the retaining ring.
Finally, install the threaded retaining ring.
CAUTION: DO NOT OVER TIGHTEN THE RETAINING RING!
CAUTION: DO NOT OVER TIGHTEN THE RETAINING RING!
CAUTION: DO NOT OVER TIGHTEN THE RETAINING RING!CAUTION: DO NOT OVER TIGHTEN THE RETAINING RING!
Selection of a Filter PositionSelection of a Filter Position
In most cases, any given filter could be installed at any of the four FILTER POSITIONS.
However, some minor considerations may improve performance of the system. Adjacent
FILTER POSITIONS will require slightly less switching time. Therefore, if an experiment
will require rapid switching between two filters they should be located next to one another.
If separated by one or more filters, there will be brief transitions through the intervening
filter(s) unless the TURBO-BLANKING feature is enabled (see “Modes of Operation”).
While this feature eliminates the transmission of light at unwanted wavelengths, it does
further increase the time required for switching.
4.2.4
4.2.4 Blanking Discs
Blanking Discs
4.2.44.2.4
Blanking DiscsBlanking Discs
The blanking discs should only be removed from those positions that will contain a filter.
The discs are intended as a safety device for the occasion where you may be looking into the
microscope. We strongly urge that the discs be used. Please be careful and avoid the
possibility of direct visualization of high intensity light.
4.2.5
4.2.5 Neutral Density Filters
Neutral Density Filters
4.2.54.2.5
Neutral Density FiltersNeutral Density Filters
. Filter installation is the same for this filter holder as described above. The slide portion of
the neutral density filter holder, however, is mechanically different from those used for the
interference filters. As a result, the two types of holders may not be interchanged.
Identify the access doors used to install the neutral density FILTER HOLDER and the four
interference FILTER HOLDERS. The small access door located near the top edge of the left
side panel on the Lambda DG-4 cabinet (see below) is for the neutral-density FILTER
HOLDER. The larger door in the middle of the left side panel is fro the interference
FILTER HOLDER. Loosen the black knob on these access doors to slide them up and install
the appropriate FILTER HOLDER into the slot. The bottom of the neutral density FILTER
HOLDER is flat while the bottoms of the interference FILTER HOLDERS are grooved. For
this reason, these two types of FILTER HOLDERS are not interchangeable. Make certain
that the end nearest the filter cup goes in first.
Although the Lambda DG-4 is simple to use, proper operation depends on the user
understanding the design and characteristics of this instrument. This section provides the
detailed information needed to begin operation and to control the unit from its keypad.
5.1
5.1 Control Electronics
Control Electronics
5.15.1
Control ElectronicsControl Electronics
The control electronics for the filter changer are activated by the “MAIN” power switch on
the rear of the unit. You may operate the control electronics without the lamp on, but it is
best to turn the control electronics off before the lamp is started. Once the lamp has been
started, the control electronics may be turned on again.
5.2
5.2 Power
Power----up Sequence
5.25.2
PowerPower
There are two switches on the back of the Lambda DG-4 labeled LAMP and MAIN. The
LAMP switch will ignite and power the xenon lamp. The MAIN switch applies power to the
controller and the fan that cools all of the electronics (including the xenon lamp power
supply). The MAIN switch should be turned on immediately after the LAMP switch is
turned on. If you are only configuring the Lambda DG-4 controller or a computer interface,
you do not need to switch on the LAMP before turning on the MAIN switch. The following
message will be briefly displayed on the front panel when the MAIN switch is turned on:
After a few seconds the MAIN MENU will be displayed:
At this point, the Controller subsystem is ready for operation.
up Sequence
up Sequenceup Sequence
SUTTER FAST FILTER CONTROLLER PROGRAM
SUTTER FAST FILTER CONTROLLER PROGRAM
SUTTER FAST FILTER CONTROLLER PROGRAM SUTTER FAST FILTER CONTROLLER PROGRAM
OFF= FILTER NUMBER, TURBO-BLANKING- OFFOFF= FILTER NUMBER, TURBO-BLANKING- OFF
INPUT = MANUAL MOVE ON COMMAND
INPUT = MANUAL MOVE ON COMMAND
INPUT = MANUAL MOVE ON COMMAND INPUT = MANUAL MOVE ON COMMAND
When the unit is first turned on, the FILTER NUMBER
indicates that the filter selecting mirrors have been positioned to prevent lamp output from
the device. The status of the TURBO
OFF is the default setting for that feature when the controller is first powered up.
The second line of the display indicates if the unit is under local control and that the unit
will execute all commands as they are received.
The unit selects the PARALLEL MODE as the default remote interface at power on. If the
unit is under remote control at power on, the appropriate remote interface mode will be
automatically set and the display may change immediately to reflect any waiting computer
commands.