SUTO S 600 Instruction And Operation Manual

English
Instruction and operation manual
S 600
Portable compressed air purity analyzer
Dear Customer, Thank you for choosing our product. Please read the operating instructions in full and carefully observe them
before starting up the device. The manufacturer cannot be held liable for any damage which occurs as a result of non-observance or non­compliance with this manual.
Should the device be tampered with in any manner other than a procedure which is described and specified in the manual, the warranty is cancelled and the manufacturer is exempt from liability.
The device is designed exclusively for the described application. SUTO offers no guarantee for the suitability for any other purpose.
SUTO is also not liable for consequential damage resulting from the delivery, capability or use of this device.
2 S 600
Table of contents
1. Safety instructions.......................................................................4
2. Application.................................................................................8
3. Features.....................................................................................8
4. Technical data.............................................................................9
4.1 General data..........................................................................9
4.2 Electrical data .....................................................................10
4.3 Accuracy..............................................................................10
5. Dimensions...............................................................................11
6. Installation on site.....................................................................14
6.1 Installation requirements.......................................................14
6.2 Connection with the optional isokinetic sampling device.............17
6.3 Connection without the isokinetic sampling device....................18
6.4 Electrical connections............................................................19
6.5 Compressed air connections (inlet and outlet)..........................20
7. Setup and configuration.............................................................22
8. Operation.................................................................................22
8.1 Value view...........................................................................23
8.2 Graphic view........................................................................24
8.3 Main menu...........................................................................24
8.3.1 Files..............................................................................25
8.3.2 Service Info ...................................................................25
8.3.3 System Setting...............................................................26
8.4 Icons in the status bar...........................................................27
9. Guided measurement.................................................................27
9.1 Steps for guided measurement...............................................29
9.2 Reports for guided measurements...........................................33
10. Optional accessories.................................................................34
11. Maintenance............................................................................34
12. Disposal or waste.....................................................................34
13. Warranty................................................................................34
S 600 3

1. Safety instructions

1. Safety instructions
Please check if this instruction manual accords to the product type.
Please observe all notes and instructions indicated in this manual. It contains essential information which must be
observed before and during installation, operation and maintenance. Therefore this instruction manual must be read carefully by the technician as well as by the responsible user / qualified personnel.
This instruction manual must be available at the operation site of the flow sensor at any time. In case of any obscurities or questions, regarding this manual or the product, please contact the manufacturer.
WARNING!
Compressed air!
Any contact with quickly escaping air or bursting parts of the compressed air system can lead to serious injuries or even death!
Do not exceed the maximum permitted pressure range (see sensors label).
Only use pressure tight installation material.
Avoid that persons get hit by escaping air or bursting parts of the
instrument.
The system must be pressure less during maintenance work.
WARNING!
Voltage used for supply!
Any contact with energized parts of the
product, may lead to a electrical shock which can lead to serious injuries or even death!
Consider all regulations for electrical installations.
The system must be disconnected from any power supply during
maintenance work.
4 S 600
1. Safety instructions
WARNING!
Permitted operating parameters!
Observe the permitted operating parameters, any operation exceeding this parameters can lead to malfunctions and may lead to damage on the instrument or the system.
Do not exceed the permitted operating parameters.
Make sure the product is operated in its permitted limitations.
Do not exceed or undercut the permitted storage and operation
temperature and pressure.
The product should be maintained and calibrated frequently, at least
annually.
General safety instructions
It is not allowed to use the product in explosive areas.
Please observe the national regulations before/during installation
and operation.
Remarks
It is not allowed to disassemble the product.
Always check the compressed air connectors in terms of stability
and tightness.
ATTENTION!
Measurement values can be affected by malfunction!
The product must be installed properly and frequently maintained, otherwise it may lead to wrong measurement values, which can lead to wrong results.
Always observe the direction of the flow when installing the sensor. The direction is indicated on the housing.
Do not exceed the maximum operation temperature at the sensors tip.
Avoid condensation on the sensor element as this will affect the accuracy enormously.
S 600 5
1. Safety instructions
Storage and transportation
Make sure that the transportation temperature is between
-10 °C ... 70 °C.
For transportation it is recommended to use the packaging which comes with the sensor.
Please make sure that the storage temperature of the sensor is between -10 °C ... 50 °C.
Avoid direct UV and solar radiation during storage.
For the storage the humidity must be < 90%, no condensation.
ATTENTION!
Equipment may get damaged!
Please make sure, that your measuring point is free of excessive contamination/dirt. This should maintained before every measurement.
Observe the measuring point always before measurement if it is free of contamination like water drops, oil drops or other rough
contaminations.
Should water hit the inner electronics, the senors could be seriously damaged.
Check your measurement point with the enclosed test kit.
6 S 600
1. Safety instructions
ATTENTION!
Overpressure!
Remove always all protection caps before connecting the compressed air to the inlet.
S 600 7

2. Application

2. Application
The portable compressed air purity analyzer S 600 can measure, record and verify the quality parameters including particle quantity, dew point, temperature, pressure, oil vapor content for a compressed air system.
The S 600 is mainly used in industrial environments, and is not developed to be used in explosive areas. For the use in explosive areas, please contact the manufacturer.

3. Features

High-resolution 5” color touchscreen display and interface.
USB port for transferring the data to a memory stick.
Ethernet (Modbus / TCP and SUTO-Bus) interface for transferring
the data to SCADA systems.
Data logger with the capacity of 100 million values.
Integrated report generator for compressed air audits, used to
generate PDF files and copy them to a USB memory stick.
All-in-one portable hand carried measurement device.
TÜV approved measurement technology and quality assurance.
Multi-dew-point measurement system for a wide range of
measurement and a very high accuracy.
Latest PID sensor technology for oil vapor measurement.
All-in-One device measuring five parameters in a single device:
Particle counter, dew point/humidity, oil vapor, temperature and pressure (flow measurement as option).
Isokinetic sampling tube for particle measurement as option.
8 S 600

4. Technical data

4. Technical data

4.1 General data

Data logger Internal, 100-million values Parameter
Measuring parameter
Range Reference
Particle 0.3 ... 0.5 µm
0.5 ... 1.0 µm
1.0 ... 5.0 µm > 5.0 µm
Annex 1 / DIN 14644 (with isokinetic sampling device
DIN 8573) Dew point -100 °C ... +20 °C DIN 8573 Oil vapor 0.003 ... 10 mg/m3ZLG/
AIM 07120604 Pressure 3 ... 15 barg DIN 1301 Temperature 0 ... 50 °C DIN 60751
Reference settings ISO 1217, 20 °C 1000 mbar
DIN 1343, 0 °C 1013 mbar
Measurement principle
Parameter Principle
Particle quantity
Laser optical detection
Dew point Ceramic humidity sensor,
oscillating crystal Oil vapor PID Volume flow Thermal mass flow (Anemometer)
Medium Compressed air, non corrosive components Humidity of the
medium
< 40%, non condensation
Temp. of the medium
0 °C ... 50 °C
S 600 9
4. Technical data
Operation pressure 3 ... 15 barg Oil vapor sensor
durability
6,000 operating hours
Housing material PC + ABS, Aluminum Protection class IP 65 (cover closed) Dimension Please observe the drawings on the next page Display 5” color graphic display, 800 x 4800 Pixels with
touchscreen interface
Weight 6.80 kg

4.2 Electrical data

Power supply Mains supply adapter (AC/DC)
Input: 100 ... 240 VAC, 50/60 Hz, 1.4 A Output: 24 VDC, 2.5 A, 60 W max.

4.3 Accuracy

Accuracy
Parameter Accuracy
Particle quantity
50% @ 0.3 ... 0.4 µm (per JIS)
100% @ 0.4 ... 5.0 µm (per JIS) Dew point ± 2° C Oil vapor 5% of value ± 0.003 mg/m
3
Volume flow ± 2 % of value ± 0.3 % of range Temperature ± 0.1 K Pressure ± 0.08 bar
10 S 600

5. Dimensions

5. Dimensions
Dimensions S 600 in mm (cover closed):
S 600 11
5. Dimensions
Dimensions isokinetic sampling device (option) in mm:
12 S 600
5. Dimensions
3D view of the S 600 and the isokinetic sampling device:
S 600 13

6. Installation on site

6. Installation on site
Please make sure that all components listed below are included in your package.
Qty. Description Item No. 1 S 600 portable compressed air analyzer in a hand
carry case with handle and shoulder belt
P560 0600
1 USB OTG memory stick A554 0087 1 Operation and instruction manual PDF no P/N 1 Purge filter for pre-measurement (test kit) A554 0604 5 6mm Teflon hose adapter, stainless steal C219 0197 1 Power supply, 230 VAC / 24 VDC A554 0086 1 2 m Teflon hose, 6 OD x 4 ID mm, free adjustable C193 0002 1 1.5 m Teflon hose with quick connector A554 0003 The following items are included only if you have ordered the isokinetic
sampling device (A554 0600) 1 Isokinetic sampling device, including a flow sensor A554 0600 1 M12 connection cable for isokinetic sampling device A553 0134
If you need replacement materials for items in the preceding list or if you need further useful accessories, please contact the manufacturer or your local distributor.

6.1 Installation requirements

The device needs to be set up next to the measuring point. Please make sure that the device is put on a flat surface. In case you are using the additional isokinetic sampling devices, you must make sure that they are also put on a flat surface next to the S 600.
The tubes should be not bended too strong and be installed with a big curve radius to avoid turbulence in the air flow.
The isokinetic sampling device must be set up next to the S 600 to get a straight and short connection. Please observe the following chapter, connecting the isokinetic sampling device to the S 600.
Please connect the S 600 to the power during the measurement, and make sure that the device is not turned off or plugged off during the measurement because the data will then be lost and not saved.
14 S 600
6. Installation on site
ATTENTION!
If the device is not installed properly it may lead to wrong measurement results.
The device is designed to be operated indoors only. To use the device outdoors, please protect the device from direct sunlight and
rain.
ATTENTION!
Before you connect the device to your point of measurement, make sure that there is no rough contamination like water/oil drops or heavy dust. This may damage the sensor units. For this please use the purge filter test kit.
Before you start the measurement, check your point of measurement:
1. Connect the purge filter test kit onto your measuring point first. Open the purge valve on the test kit and purge air for a short period.
2. Check the filter in the test kit to see if it shows high contamination of water, oil or dust.
3. If the filter is contaminated severely, stop using the S 600 for measurement because this may lead to serious damage to the device. In case you are not sure, please contact the manufacturer.
S 600 15
6. Installation on site
ATTENTION!
Before connecting the compressed air, make sure that all protection caps are opened/removed!
Remove the protection caps from all parts of both sides of the instrument. If not removed, the device may get damaged and the risk of bursting parts under high pressure can even lead to personal injuries. All exhaust outlets will pass air during the measurement, if not, please contact the manufacturer.
ATTENTION!
Always use the 6 mm Teflon hose adapter to connect the teflon hoses to the S 600 and to the isokinetic sampling device! You may damage the device if not used.
Directly plugging and pulling the teflon hose more than once may lead to particle contamination, which can affect the measurement. To avoid this problem, please use the included adapter plugs and keep them on your tubes.
16 S 600
6. Installation on site

6.2 Connection with the optional isokinetic sampling device

The picture above shows the S 600 setup with the isokinetic sampling device connected. Please connect the isokinetic sampling device using the teflon hose delivered with the device. On the next page you can find the detailed description of the sampling device. Also you can find which outlets of the sampling device need to be connected to which input at the S 600.
The following diagram helps you better understand the setup connection.
S 600 17
6. Installation on site
1. Before connecting the S 600 to your compressed air, make sure ball valve 2 and 4 are closed.
2. Close the needle valve 3.
3. The connection to your compressed air system is achieved through the correct fitting (quick connector, teflon hose, etc.). Connect your compressed air system to the inlet 1 using the correct fitting.
4. Connect the outlet valve 2 to the inlet for dew point and oil vapor measurement at the S 600 using a teflon hose.
5. Connect the isokinetic outlet 4 with the inlet for the particle counting at the S 600.
6. Now open outlet valve 2 and 4 to start your measurementfollow the instructions on the screen.

6.3 Connection without the isokinetic sampling device

Connect your compressed air system using two teflon hoses to the corresponding inlets at the S 600. They are marked: Gas inlet Oil / Dew point measurement and Gas inlet Particle measurement.
18 S 600
6. Installation on site

6.4 Electrical connections

The S 600 offers three types of electrical connections. The power supply connector 1, the communication port for the isokinetic sampling device 2 and an Ethernet port 3 to communicate with
network devices.
ATTENTION!
Only use the power supply which comes with the
S 600!
S 600 19
6. Installation on site

6.5 Compressed air connections (inlet and outlet)

The S 600 offers two compressed air inputs on the right side of the housing. The inlets are shown on the picture above and are marked respectively to their functions: Gas inlet Oil / Dew point measurement and Gas inlet Particle measurement.
ATTENTION!
Permissible pressure!
Please observe the maximum permissible incoming pressure. It must be in between 3 and 15 bar overpressure. If the pressure exceeds this range, it will damage the device. If the pressure is too low, the volumetric flow will not be high enough which will lead to wrong results.
20 S 600
6. Installation on site
In the picture below shows the left side of the housing where the gas outlets can be found. The calibration outlets Gas outlet Dew point measurement and Gas outlet Oil vapor measurement are used to connect the internal sensors to references. These calibration outlets must be opened all the time during compressed air is connected. Also the other two outlets need to be opened before connected to the compressed air.
Note!
All gas outlets of the measurement device must stay opened during the complete usage of the device.
S 600 21

7. Setup and configuration

7. Setup and configuration
The S 600 is configured ex-works and ready to work out of the box. The S 600 provides a guided measurement procedure to guide you through device setup for each measurement parameter. All these setup settings are automatically saved into the device even after a power failure.
Note!
If you are facing problems setting up you device, contact the manufacturer or your local dealer for assistance.

8. Operation

After the S 600 gets powered, the initialization screen is displayed with an active progress bar.
During initialization, the device configures the sensors and runs initialization routines.
After the initialization is completed, the date of the last calibration is shown. Click OK to proceed.
22 S 600
8. Operation
The value screen as shown on the left appears.
Quick buttons or icon on the bottom bar are listed as follows:
Graphic:To switch to the graphic view.
Value: To switch to the value view.
Menu: To access the operation sub-menus.
The camera icon: To take a screenshot.

8.1 Value view

In this view, the S 600 shows all measured values in realtime.
To switch to the value screen, click Value in the bottom.
Note: During the first five minutes, the S 600 performs a purge process to ensure any remaining particles in the system are blown out. During this period, the counting numbers on the Particle pane appear green and blink.
S 600 23
8. Operation

8.2 Graphic view

The graphic view is pre-configured in the factory, and you do not need to change anything.
In case that you need to make changes, follow the instructions indicated in the following figure.

8.3 Main menu

Access the main menu by clicking Menu in the bottom bar. The screen with sub­menus is shown.
The main menu offers the access to the following sub-menus.
Guided Measurement
To start the guided measurements, which lead you through a complete measurement cycle. For more
24 S 600
8. Operation
information, see Chapter 9. Guided measurement.
Files To access the saved screenshots and the memory
usage as well.
Service info To show contact information for technical support.
System settings
To perform general settings on date, time and language and so on. To view information such as the serial number.
Communication To perform field bus settings and configure
communication parameters

8.3.1 Files

The files menu shows all screenshots for you to view, export, and delete them. You can also check the memory status.

8.3.2 Service Info

Shows contact information for technical support.
S 600 25
8. Operation

8.3.3 System Setting

You can perform general settings using this menu.
Password To set a password to protect the settings menu
from unauthorized access.
Back light To configure the brightness and the auto dimming
function of the screen.
Calibrate touch screen
To calibrate the touch screen if it does not respond to user inputs correctly or precisely, it can be calibrated.
Language To select the interface language. Date time To configure date and time.
26 S 600
8. Operation
Device info To view device information such as serial number. System update To perform a system update. Reset To restart the device (User settings will be
saved).

8.4 Icons in the status bar

USB memory stick is connected. By pressing the icon, the stick can be disconnected.
System errors occur. Press this icon to get further information.
Calibration is overdue, please. Contact the manufacturer of your local dealer.
The S 600 is connected to a PC by the USB cable.
Data logger status:
- STOP—Indicates that the data logger is not running.
- LOG—Indicates that the data logger is running.

9. Guided measurement

The S 600 provides a software-based guided measurement which takes you through the complete measurement. This leads to a simplified measurement process and prevents you from wrong measurements.
To start a guided measurement, do the following:
1. Click Menu > Guided measurement.
2. Select which type of measurement you want to perform.
S 600 27
9. Guided measurement
Monitoring with programmed time out: It starts a
measurement with a user-programmed period of measurement time. You can set the measurement time during the process of measurement preparation. The system will then, after finishing the programmed measurement duration, stop the measurement automatically and save the data. This mode is ideally used for audits where you must measure at several points. You can program for each point a duration of e.g. 2 hours and then you can compare the measurements.
Monitoring with manual stop: It starts the measurement
without a programmed stop time. You can click it to start the measurement and when you come back you can stop it. Then you can decide if you want to save or delete the data. This can be used to monitor changes in values.
3. Perform the guided measurement following the onscreen instructions. For more information, see section 9.1 Steps for guided measurement.
4. To view and manage the measurement files generated, click Report Manager. For more information, see section 9.2 Reports for guided measurements
28 S 600
9. Guided measurement

9.1 Steps for guided measurement

After you start a guided measurement, follow below steps to go through the whole process.
1. An overview is given about the selected measurement types. Click Yes to start.
2. The system asks if you are using the isokinetic sampling device, and this will affect the further steps and instructions. Select Yes if you have the isokinetic sampling device connected. Otherwise, click No.
3. Input your data, which will be then shown on the report.
S 600 29
9. Guided measurement
4. Select your compressed air class, which will then select the corresponding alarms.
- CLASS 0: To customize alarm settings.
5. If you selected CLASS 0, enter the limit values for each measurement channel, through which you define different alarm settings for the quality management.
6. Enter how long the measurement will take. The longer it takes, the more stabilized the values will be and the more exactly it will represent the system conditions.
30 S 600
9. Guided measurement
7. The system checks whether the pressure is in the valid range.
8. The system performs the self­calibration cycles and cleans the internal sensor components.
The following 3 steps only apply if you select Yes for the isokinetic sampling device, otherwise skip the next 3 steps
1. The S 600 is now ready for measurement. Please read the instructions shown on the screen carefully. Click Start.
S 600 31
9. Guided measurement
2. Follow the onscreen instructions to proceed.
3. Perform Isokinetic setup as instructed on the screen.
Now the device is well set up and starts to measure data. The remaining time is shown on the top left corner.
32 S 600
9. Guided measurement
During the measurement, you can see the Data logger status icon on the status bar switched from STOP to LOG. The remaining time is displayed in the upper left corner. Please wait. The system will stop the measurement automatically.
When the measurement is successfully completed, the left screen appears. You can choose discard or save the measurement data.

9.2 Reports for guided measurements

After performing guided measurements, you can view and manage measurement files through Guided Measurement > Report Manager.
In the report screen:
To view the
measurement results, click on the file (not the check box on the right). A window appears showing the PDF for your preview.
To copy, export , or delete files, select the
file check boxes, and then click the corresponding button at the bottom.
S 600 33

10. Optional accessories

10. Optional accessories
To purchase optional accessories, please contact the manufacturer or your local dealer:
Isokinetic sampling device, for particle sampling according to ISO
8573.
Teflon hoses and sorts of adapters.

11. Maintenance

Use a moist fabric to clean the device. For the use in GMP areas, the device must be disinfected through wipe disinfection. For more information, please contact the manufacturer or your local dealer.
ATTENTION!
Please dry the device after cleaning using a clean and dry fabric. Always take care, that the fabric for cleaning is not to wet as water could get into the device and lead to damage.

12. Disposal or waste

Electronic devices are recyclable material and do not belong in the household waste. The sensor, the accessories and its packing must be disposed according to your local statutory requirements. The dispose can also be carried by the manufacturer of the product, for this please contact the manufacturer.

13. Warranty

SUTO provides a warranty for this product of 24 months covering the material and workmanship under the stated operating conditions from the date of delivery. Please report any findings immediately and within the warranty time. If faults occur during the warranty time SUTO will repair or replace the defective unit, without charge for labour and material costs but there is a charge for other service such as transport and packing costs. Excluded from this warranty is:
Damage caused by:Improper use and non-adherence to the instruction manual.
34 S 600
13. Warranty
Use of unsuitable accessories.External influences (e.g. damage caused by vibration, damage
during transportation, excess heat or moisture).
Sensor lifetime, which is determined by the operating hours (6,000-hour sensor durability).
Filter capacity, which is determined by the operating hours (8,640­hour or 360-day lifetime).
The warranty is cancelled:
If you open the measurement instrument without a direct request written in this instruction manual.
If repairs or modifications are undertaken by third parties or unauthorized persons.
If the serial number has been changed, damaged or removed.
If the warranty sealing is removed or damaged.
Other claims, especially those for damage occurring outside the instrument are not included unless responsibility is legally binding.
Warranty repairs do not extend the period of warranty.
ATTENTION!
Batteries have a reduced warranty time of 12 months.
S 600 35
SUTO iTEC GmbH SUTO iTEC (ASIA) Co., Ltd.
Werkstr. 2 Room 10, 6/F, Block B, Cambridge Plaza 79426 Buggingen 188 San Wan Road, Sheung Shui, N.T. Germany Hong Kong
Tel: +49 (0) 7631 936889-0 Tel: +852 2328 9782 Fax: +49 (0) 7631 936889-19 Fax: +852 2671 3863 Email: sales@suto-itec.com Email: sales@suto-itec.asia Website: http://www.suto-itec.com Website: http://www.suto-itec.com
All rights reserved © Modifications and errors reserved.
S600_im_en_2019-2
36 S 600
Loading...