Suntex TC-7310-RS Operation Manual

TC-7310-RS
Intelligent
Turbidity/S.S.
Transmitter
Operation
Manual
Thank you for purchasing Suntex products. In order to continuously improve and enhance the transmitter’s function and capabilities, Suntex reserves the right to modify the content and icon display of the product without notice. When using this transmitter, please follow all instructions on installation and operation as described within this manual. Suntex Instruments Co., Ltd. is not liable for any direct or indirect loss or damages caused by improper usage of this product. If there are any omissions or mistakes, questions or concerns, regarding the product of this operation manual, please contact our staff. Thank you.
Precautions for Installation
Please read this operation manual thoroughly before installation to prevent incorrect wiring which may lead to instrument damage and/or safety issues.
In order to avoid electrical hazards, all wiring must be correctly connected and inspected
before connecting to power supply.
Meter installation site should be properly ventilated and kept from direct sunlight and high
temperature.
The signal cable requires a special coaxial cable material. Cables provided by Suntex are
strongly recommended. Please do not use normal electric wires.
Prevent power surge interference to the transmitter. Especially when using a three-phase
power system, make sure the device is properly grounded. If power surge interference occurs, separate the power supply of the transmitter from that of the controlled device (i.e. dosing machines, mixers, etc.), or install surge absorber to reduce power surges from all electromagnetic switches and power control device coils.
To protect the instrument, the internal relays must be connected to external power relays
with sufficient ampere capacity before connecting to external alarms or devices. (Please
refer to chapter 3.5 “Electrical Connection Illustration”)
Suntex logo is shown on the top right corner of the display during all operations. For function
illustration purposes, the logo is not shown in the figures presented in this operation manual.
CONTENTS
Precautions for Installation Brief Instructions 1
1. Specifications 4
2. Assembly and Installation
2.1 Transmitter Installation 5
2.2 Panel Mounting Illustration 5
2.3 Wall Mounting and Pipe Mounting Illustration 6
3. Overview of Turbidity/S.S. Transmitter TC-7310-RS
3.1 Rear Panel Illustration 7
3.2 Terminal Function Illustration 7
3.3 Terminal Function Description 8
3.4 Cable Wiring 8
3.5 Electrical Connection Illustration 9
4. Configuration
4.1 Front Panel Illustration 10
4.2 Keypad 10
4.3 LED Indicators 10
5. Operation
5.1 Measurement Mode 11
5.2 Setup Menu 11
5.3 Calibration Menu 11
5.4 Shortcuts 11
5.5 Default Values 11
5.5.1 Settings Default Values 11
5.5.2 Calibration Default Values 12
6. Measurement Display Mode
6.1 Text Mode 13
6.2 Real-Time Chart Mode 14
6.3 Trace Mode 15
6.4 Warning Symbols and Text 16
7. Settings
Settings Block Diagram 17
7.1 Setup Menu 19
7.2 Settings Security Code (Code) 20
7.3 Language 21
7.4 Measurement Mode/Unit (Mode) 22
7.5 Product Adjustment 24
7.6 Relay 1 25
7.7 Relay 2 26
7.8 Wiper 27
7.9 Electrode Zero 28
7.10 Clean 39
7.11 Analog Output 30
7.12 Date/Time (Clock) 31
7.13 RS-485 Communication 32
7.14 Sample Measurement Average (Digital Filter) 33
7.15 Backlight 34
7.16 Contrast 35
7.17 Logbook 36
7.18 Automatic Return (Return) 37
8. Calibration
Calibration Block Diagram 38
8.1 Calibration Setup Menu 39
8.2 Calibration Security Code (Code) 41
8.3 Calibration 42
8.3.1 Known Solution Calibration 42
8.4 Correction Factor 45
8.5 Automatic Return (Return) 46
9. MODBUS Protocol and Instructions
9.1 Communication Connection 47
9.2 MODBUS Name and Address Table 49
9.3 MODBUS Description Example 53
10. Error Messages (Error Code) 54
11. Maintenance 54
1
Brief Instructions
Description of Setup Settings (See Chapter 7 for Details)
Press and simultaneously to see the overview of the setup settings. Then press to modify setup settings. Press keypad according to the index bar on the bottom of the screen.
Index of Keypad
Keypad Index Bar Description
Return to previous level or action
Left or left page
Increase digit
Right or right page
Decrease digit
Confirm and proceed to next step
Setup Items
Function Icon Description
Measurement mode
1. Select from TC-100, TC-500, TC-3000, or TCS-1000
electrodes.
Mode
2. Select display mode from text, real-time or trace mode.
3. Measuring unit selection
TC-100, TC-500, TC-3000: NTU, FNU, or FTU. TCS-1000: ppm or mg/l
Product Adjustment. Modify the measuring value attained from
Product Adj.
the field based on the measuring value attained by sampling, so it is not necessary to take out the sensor for calibration.
Relay 1
First relay settings, select action off or Hi/Lo alarm
Relay 2
Second relay settings, select action off or Hi/Lo alarm
Wiper
2
Automatic wiper wash time settings, adjust detector window cleaning duration and interval
Electrode
Zero
Clean
Analog
Clock
RS-485
Digital Filter
Black-light
Sensor itself zero point correction, for sensor internal calibration Automatic wash time settings; adjust external sensor cleaning device ON and OFF duration
Current output corresponding to Turbidity/SS setting range
Time and date settings (An internal battery keeps the clock
running when disconnected from power. Replace with 3V CR2025/2032 lithium battery.)
RS485 serial interface (Modbus protocol)
Take 1~60 serial measurements, average continuously, and
display as the reading following stabilization
Backlight settings, set Auto/ON/OFF backlight, brightness, and
sensitivity
Contrast
Logbook
Return
Code
Language
Screen contrast settings
Event recorder logbook (50 data)
Measurement mode return settings
Setup mode security code. The setup code is precedential to
calibration code. A different security code for calibration mode can be set.
Available in English, Traditional Chinese, and Simplified
Chinese
3
Description of Calibration Settings (See Chapter 8 for Details)
Press and simultaneously to see the last calibration information. Press to make a new calibration or to modify calibration settings. Press keypad according to the index bar on the bottom of the screen.
Index of Keypad:
Keypad Index Bar Description
Return to previous level or action
Left or left page
Increase digit
Right or right page
Decrease digit
Confirm and proceed to next step
Calibration Items (Up to Five-Point Calibration)
Function Icon Description
Known Sol.
Calibration by known value solution
Correction
Calibration by correction factor (applicable for Turbidity and SS measurement value interconversion)
Code
Calibration mode security passcode
Return
Measurement mode return settings
Note
Due to the need for continuous improvement of the transmitter, we reserve the right to modify the icons and content. The icons and contents of the instrument are subject to change without notice.
0~100
NTU
(TC-100 sensor)
0.01
NTU
(TC-100 sensor)
4
1. Specifications
Model TC-7310-RS
Measuring Mode Turbidity/Suspended Solids
Measuring Unit
Turbidity: NTU, FTU, FNU
Suspended Solids: mg/l, ppm
0~500 NTU (TC-500 sensor)
Measuring Range
0~3000 NTU (TC-3000 sensor)
0~1000 ppm (TCS-1000 sensor)
0.1 NTU (TC-500 sensor)
Resolution
1 NTU (TC-3000 sensor)
0.1 ppm (TCS-1000 sensor)
Calibration Mode
Turbidity or S.S. standard solution up to 5 points calibration
and calibration by factor correction
Electrode Zero Sensor internal zero point correction activation function
Ambient Temp.
(Turbidity/S.S. sensor working temp.: 0~40°C)
0~50 °C
Storage Temp. -20~70 °C
Large LCM with sensor for backlight and contrast
Display
Chart mode: 3 min real-time dynamic graph
Text mode: Numerical display
Trace mode: Programmable 3 min to 4 week trend graph
Analog Output
Isolated DC 0/4~20 mA corresponding to main measurement,
Max. load 500
Serial Interface Isolated RS-485 (MODBUS RTU or ASCII)
Logbook 50 events
Contact RELAY contact, 240 VAC 0.5 A Max. (recommended)
Settings
Activate
Hi/Lo, Hi/Hi, Lo/Lo selectable, two limited programmable,
ON/OFF
Wash RELAY contact: ON 0~99 min 59 sec / OFF 0~999 hr 59 min
Auto, sensor internal function for fixed 30min per wipe (TC-100
Sensor Wiper
sensor), or 10 min per wipe (TC-500, TC-3000, TCS-1000 sensor)
Manual, setting from 2~29 min per wipe(TC-100), or 2~9 min per wipe
Voltage Output DC±12V
Power Supply 100 V~240 VAC ±10 %, 8W max., 50/60 Hz
Installation Wall or Pipe or Panel Mounting
Dimensions 144 mm x 144 mm x 115 mm (H×W×D), 1/2 DIN
Cut-off Dimensions 138 mm × 138 mm (H×W)
Weight 0.8 kg
Protection IP65 (NEMA 4X)
Note: The specifications are subject to change without notice.
138 mm
90 mm
5
2. Assembly and Installation
2.1 Transmitter Installation
The transmitter can be installed by panel mounting, wall mounting, or 2” pipe mounting. Panel Mounting:
Prepare a square hole of 138 mm x 138 mm on the panel box, and then insert the controller
directly into the hole. Insert the accessorial mounting bracket from the rear, and fix into the pickup groove.
2.2 Panel Mounting Illustration
138 mm
Hole size
Mounting bracket
Minimum
200mm
Minimum
200mm
Panel mounting illustration, fix with mounting brackets
Hole distances on the panel box
6
2.3 Wall Mounting and Pipe Mounting Illustration
Sun Shield (wall mounting, optional)
(Order No.: 8-35 + 8-35-2)
Sun Shield (pipe mounting, optional)
(Order No.: 8-35 + 8-35-1)
Pipe mounting illustration, fix with a U-shaped pipe clamp (optional).
(Order Number: 8-34)
Insert rubber plug into unused cable glands. Tighten up the cable gland to prevent water vapor infiltration.
4/20m
A
D-(A)
7
3. Overview of Turbidity/S.S. Transmitter TC-7310-RS
3.1 Rear Panel Illustration:
3.2 Terminal Function Illustration:
S+
Cal.
S-
V-
POWER
INPUT
POWER
AC
RS-485
D+(B) G
Check
V+
Shield
REL1
REL2
WASH/Cln
D-(A)
:
Shield
:
V+ :
Check
:
V-
:
S-
:
Cal.
:
8
3.3 Terminal Function Description
1 2
100~240 A
WASH: Wash relay contact for an external relay
4 5 6 7 8 9
10
11
12
13
14 15 16 17 18 19 20
REL2: Second alarm control, the contact for an external relay
REL1: First alarm control, the contact for an external relay
RS-485 output D+(A)
G: RS-485 output GND D+(B): RS-485 output D+(B) 4~20mA –Terminal: Turbidity/SS current output terminal -, for an
4~20mA +Terminal: Turbidity/SS current output terminal +, for an
Sensor signal shield wire
BROWN sensor wire (power + terminal)
GREEN sensor wire (self-test signal output terminal)
BLUE sensor wire (power - terminal)
BLACK sensor wire (signal - terminal)
RED sensor wire (calibration signal input terminal)
S+ : WHITE sensor wire (signal + terminal)
C: Power supply terminal
external recorder or PLC control
external recorder or PLC control
3.4 Cable Wiring
S+(White)
Cal.(Red)
S-(Black)
V-(Blue)
Check(Green)
V+(Brown)
Shield
9
3.5 Electrical Connection Illustration
100V~240VAC
Transmitter
Relay 1
Relay 2
WASH/Cln
100V~240VAC
Surge absorber
M
Cleaning device
Surge absorber
External relay
Surge absorber
Surge absorber
External relay
M
Dose feeder
Surge absorber
Surge absorber
External relay
Dose feeder
M
Note: The transmitter’s built-in miniature relays are required to be repaired and replaced by professional
technicians. External relays (Power Relay) must be connected to activate external devices to protect the instrument.
10
4. Configuration
4.1 Front Panel Illustration:
4.2 Keypad:
In order to prevent unauthorized operations, the transmitter utilizes multi-key and passcode functions to enter parameter and calibration setting modes. Descriptions of the key functions are as follows:
In parameter set-up mode, press this key to exit and return to measurement mode.
n calibration mode, press this key to exit and return t
I
o measurement mode.
1. In the parameter set-up mode and calibration mode, press this key to move left or
return to the previous page.
2. When adjusting values, press this key to increase the value.
1. In the parameter set-up mode and calibration mode, press this key to move right
or advance to the next page.
2. When adjusting values, press this key to decrease the value.
Key for confirmation; press this key to confirm data values or select parameter item
settings.
4.3 LED Indicators: WASH
RELAY1 RELAY2 B.L.
: Washing device operation indicator : Controlling of dose feeding operation indicator (Relay 1) : Controlling of dose feeding operation indicator (Relay 2)
: Light sensor, under automatic display backlight mode, the indicator will light up when the surrounding brightness changes
5. Operation
11
5.1 Measurement Mode:
After all electrical connections are secured and tested, connect the instrument to the power supply and turn it on. The transmitter will automatically enter measurement mode with the factory default settings or the previous user settings.
5.2 Setup Menu:
Please refer to the set-up instructions in Chapter 7. Press and simultaneously to enter set-up menu, and press to return to measurement mode.
5.3 Cal
ibration Menu:
Please refer to the calibration instructions in Chapter 8. Press and simultaneously to enter calibration menu, and press to return to measurement mode.
5.4 Shortcuts:
1. Under measurement mode, press and hold for 2 seconds to enter Logbook function.
Press to return to measurement mode.
2. Under measurement mode, press and hold for 2 seconds to switch from display mode to text mode, trace mode, or real-time chart display mode.
5.5 Default Values:
5.5.1 Settings Default Values:
Sensor model: TCS-1000
suring unit: mg/l
Mea Product adjustment: 0.0 mg/l Relay 1: High point alarm: AUTO, SP1=800.0 mg/l, Hys = 10.0 mg/l Relay 2: Low point alarm: AUTO, SP2= 200.0 mg/l, Hys = 10.0 mg/l Wash time: OFF Wiper: Manual, 9 minute/per wipe Analog current output: 4~20 mA, 0.0~1000.0 mg/l Date & Time: 2016/1/1 00:00:00 RS-485: RTU, 19200, Even, 1, ID: 001 Digital filter: 30 Backlight settings: OFF Contrast: 0 Logbook: None Auto return: Auto, 3 minutes Code settings: OFF
5.5.2 Calibration Default Values:
12
Calibration unit: mg/l Calibration point: No Cal. Correction factor: 1.0000 Auto return: Auto, 3 minutes Code settings: OFF
Note: The factory default calibration setting is “No Cal”, and the calibration value is “None”. This means
that the user has not yet calibrated the sensor with the transmitter. After every calibration, the calibration information display will be updated.
6. Measurement Display Mode
Main
measurement
Main measurement
Time and date
13
6.1 Text Mode
Text mode is the main measurement screen and shows the measurement value and unit, temperature compensation mode, temperature measurement and unit, time and date, as seen in the following illustration.
unit
6.2 Real-Time Chart Mode
14
Real-time chart mode shows a dynamic change of the measuring values during the last 3 minutes. Under setting mode, users are allowed to set corresponding Turbidity/SS measuring range (see section 7.4) to adjust the resolution of the curve. The smaller the range is being set, the higher resolution the display is. When the unit enters the setup or calibration mode, the real-time graphic will be re-updated after returning to measurement mode. When the measured value exceeds a set range of the upper and lower limit, the graphics will be presented in the upper and lower limits dotted line. Real-time chart mode display is shown as below. There are also real-time measurement value, & unit, and temperature value & unit which are displayed in the bottom of the screen. The timeline in real-time graphic is divided into 12 depicts, each of which describes a quarter minute (15 seconds).
3 minutes of main measurement
record curve
15s
Upper limit
Lower limit
The current
instantaneous value
Main measurement
value and unit
6.3 Trace Mode
15
Trace mode features value tracing on a graph with duration configurable from three minutes up to four weeks. The graph is composed of 60 record sets over time interval T. Each set of data (T/60) is displayed by calculating the mean, maximum, and minimum value of the previous 60 values. When a new data set is calculated, the set will be displayed on the right and thus shifting the previous sets to the left by one unit. If time (T) is set for 60 hours, then each displayed set will represent a calculation of the mean, maximum, and minimum values of the data recorded during the hour (T/60). The horizontal x-axis of the graph is divided into 12 marks, with each mark representing T/12, and five data sets between each mark. The user can set the Turbidity/SS graph range under Mode in Settings (see section 7.4). The smaller the range, the higher the display resolution. Trace mode shows the real-time measurement value and unit, and temperature value and unit at the bottom of the screen, as seen in the following illustration. Attention: If the time interval is reset, the trend in the data will not be retained, and will start a new trace record. Note: The time is displayed as (XX: XX) (hr: min); e.g. four weeks (672:00).
Upper limit
Lower limit
Record duration (T),
the negative sign is
to mean the record
time has passed.
Trend recording interval
(60 sets of data)
Main measurement
value and unit
The mean value of
the T/60 interval
T/12
The max. value of
the T/60 interval
The min. value of
the T/60 interval
The latest set of record
shown on the right sector
16
6.4 Warning Symbols and Text
1. When detector wiper is activated, the display will show “HOLD” and flash “Wiper Running”. At the same time, the transmitter will automatically turn off Relay 1 and Relay 2 function. After wiper cleaning is completed, both Relay 1 and Relay 2 will automatically returns normal status.
2. When the clean function is activated, the display will show “HOLD” and flash “Clean Running”. At the same time, the WASH indicator LED will light up, and the transmitter will automatically turn off Relay 1 and Relay 2 function. After cleaning is completed, both Relay 1 and Relay 2 will automatically returns normal status.
3. When Relay 1/Relay 2 Hi settings are activated, the display will flash “REL1-HI/ REL2-HI”, and RELAY1/RELAY2 indicator LED will light up. hen Relay 1/Relay 2 Lo settings are activated, the
play will flash “REL 1-Lo/ REL 2-Lo”, and the RELAY1/RELAY2 indicator LED will light up.
dis
4. When the Analog current output exceeds the upper/lower limit, the display will flash “mg/l-mA__
/mg/l-mA , or ppm-mA / ppm-mA
:Measurement mode
:Set-up mode
REL 1 high or
low point alarm
REL 2 high or
low point alarm
:Calibration mode
Relay function
on hold
Time and date
Cleaning is in
action
Analog output
over range alarm
Product Adjust ON
Measurement unit
Note: The “HOLD” warning text appears when clean function is activated, or when in setup menu, or in
alibration menu. Under HOLD status, the corresponding display and output are as follows:
c
1. Both Relay 1 and Relay 2 will cease from action. When entering settings menu or calibration
menu under cleaning status, the instrument will automatically halt the cleaning action.
2. The current output corresponding to measurement value remains at the last output value before
HOLD status.
3. The last signal output value from the RS-485 interface is kept at the last output value before
HOLD status.
Sensor detector
wiper is in action
7. Settings
17
Settings Block Diagram – Part 1
Settings Block Diagram – Part 2
18
Continued from
the previous page
Electrode Zero
Clean
Analog
Clock
RS-485
Digital
Filter
Back Light Return
Contrast
Logbook
Continued from
the previous page
Code
Clean
ON/OFF
Input Clean
ON Time
Input Clean
OFF Time
Input Clean
Hys. Time
Select
0/20 or 4/20 mA
Current Output
Input 0 or 4 mA
Ref. Value
Input 20 mA
Ref. Value
Clock
ON/OFF
ONAuto
Set Year
Set Date
Set Time
Select
RTU or
ASCII
Select
Baud Rate
Select Parity
Select
Stop Bit
Set Number
of Samples
Averaged
Auto ON OFF
Adj.
Bright-
ness
Adj.
Sensi-
tivity
Adj.
Bright-
ness
Adj.
Contrast
Display
Logbook
History
Auto Manual
Input
Return
Timer
Clean
Set ID
Number
Return to previous
level or action
7.1 Setup Menu
19
nde
r measurement mode, press the two keys and simultaneously to enter the
U current settings overview. Press to enter setup mode to modify the settings if necessary.
Press
Measurement (Mode)
P
ress or
Press or
Press
Press to confirm it
Setup Menu
20
7.2 Settings Security Code (Code)
After entering setup mode, select “Code” and press to enter code setting procedure.
The preset settings code is 1111. Note: The code for settings mode is at a higher security level than the code for calibration. Thus,
the code for setting mode can be used to unlock calibration mode.
When an incorrect passcode is entered, the display will show “Error Code”, press to re-enter, or press to exit.
Press to confirm it.
The first ‘0” of digits ‘0000” will start to flash. Press or to adjust the value, and t then press to confirm it and proceed to the next digit.
Press to confirm it.
Press or
Select to turn on or turn off code protection function. If turned on, please input a new passcode. The passcode will be required when you re-enter setup mode. Input the correct passcode to enter setup mode.
Press to confirm it.
Input new passcode
Press to confirm it.
Enter “Language” Setup
to
switch
to
switch
21
7.3 Language
After entering setup mode, select “Language” and press to enter system language selection. Select from English, Traditional Chinese or Simplified Chinese.
Press to confirm it.
Press or
Press or
Press to confirm it.
Enter “Mode/Unit” Setup
7.4 Measurement Mode/Unit (Mode/Unit)
Press or to
Press or to
Press or to
Press or to
22
f
A
ter entering setup mode, select “Mode/Unit” and press to configure
measurement settings: reference table, unit, and display mode.
For TC-100, TC-500, TC-3000 electrodes, selectable modes: NTU, FTU, FNU. For TCS-1000, selectable modes: mg/l, ppm.
Press to confirm it.
select the electrode.
Press to confirm it.
select measurement unit.
select measurement unit.
Press to confirm it.
select measurement unit.
Press to confirm it.
Measurement Display Settings
(continued on next page)
Measurement Display Settings
Press or to
Press or to
Press or to
Press or to
to adjust seconds
value, then press
to confirm and
proceed to adjust
Select NO to keep
previously recorded
data. Press to
23
select display mode.
select display mode.
Press to confirm it.
confirm it.
adjust maximum value.
adjust minimum value.
Select YES and press
Press to confirm it. Press to confirm it.
Press or _
minutes value.
Press to confirm it.
Press to confirm it.
Enter “Product Adjustment” Setup
7.5 Product Adjustment
reading
to
in the main display in
order to modify the reading of measurement
24
After entering setup mode, select “Code” and press to make fine adjustments to the measurement reading. Users are able to make sample reading adjustments based on actual sample measurement values without retrieving the sensor for calibration. A PDT icon will display on the main screen above the pH unit if the product is adjusted (see section 6.4).
Press to confirm it.
Users may compare the measurement with the reference value. Press or tune the adjustment value
Press to confirm it.
which will show in the bottom part of display.
Enter “Relay 1” Setup
7.6 Relay 1
on on SP
25
Enter setup of Relay 1. Select the item to turn the relay 1 function ON / OFF. If you select to
ur
n on relay 1, set relay 1 as “High set-point” alarm or “Low set-point” alarm. Set the value of
t set-point (SP) and Hysteresis (Hys.). Refer to the graph below for the relationship of the parameters (for high point alarms).
Turbidity/SS
Hys.
Press to confirm it.
Press or to select to activate Relay 1. If turned off, the menu will proceed to setup of Relay 2.
off
High Point Alarm Control
off
t (sec)
Press to confirm it.
Press or to select Relay 1 as Hi point or Lo point alarm.
Press or to adjust Hysteresis (Hys.) value
Press to confirm it.
Press to confirm it.
Press or to adjust Se
tting Point (SP) value.
Press to confirm it.
Press or to select to activate testing of Relay 1. If turned on, Relay 1 will begin testing, and RELAY 1 indication lamp will light up.
Press to confirm it.
Enter “Relay 2” Setup
7.7 Relay 2
SP
on on
26
Enter setup of Relay 2. Select the item to turn the relay 2 function ON / OFF. If you select to
ur
n on relay 2, set relay 2 as “High set-point” alarm or “Low set-point” alarm. Set the value of
t set-point (SP) and Hysteresis (Hys.). Refer to the graph below for the relationship of the parameters (for low point alarms).
Turbidity/SS
Hys.
Press to confirm it.
off
off
Press to confirm it.
Press or to select whether to turn on Relay 2 or not. If OFF, the guide menu will go to setup of Wiper.
Low Point Alarm Control
t (sec)
Press or to select Hi point or Lo point alarm of Relay 2.
Press or to adjust Hysteresis (Hys.) value
Press to confirm it.
Press or to adjust Setting Point (SP) value.
Press to confirm it.
Enter “Wiper” Setup
Press to confirm it.
Press or to select to activate testing of Relay 2. If turned on, relay 1 will begin testing, and RELAY 2 indication lamp will light up.
Press to confirm it.
7.8 Wiper
Press or to
Press or to
adjust wiper cleaning
27
f
ter entering setup mode, select “Wiper” and press to configure wiper settings: activation
A
mode and time interval. Select Auto to set automatic wiper cleaning according to the sensor
connected. Select manual to configure automatic cleaning interval. TC-100: Auto – Electrode default is 30 minute interval. Manual – Configure from 2 to 29 minutes. TC-500, TC-3000, TCS-1000: Auto – Electrode default is 10 minute interval Manual – Configure from 2 to 9 minutes.
Press to confirm it.
select Wiper mode.
Press to confirm it.
interval.
Press to confirm it.
Enter “Electrode Zero” Setup
7.9 Electrode Zero
The transmitter will
current
sent from the electrode
point
calibration is being done
ress
point
The transmitter will
current
sent from the electrode
point
28
f
ter entering setup mode, select “Electrode Zero” and press to perform zero-point
A
calibration. The user can determine if zero-point calibration is necessary from the deviation of
the electrode’s measurement values. If calibration procedure is necessary, the electrode must be
placed into distilled or deionized water for zero-point calibration.
Note: When performing zero-point calibration, factory default electrode output signals will be
altered.
Press to confirm it.
Select NO to exit zero-
calibration, and press
Select YES and p
display the 2mA
signaling that zero-
internally.
display the 4mA
signaling that zero-
calibration is complete.
Enter “Clean” Setup
7.10 Clean
on
on
on on
Relay
Contact
Timer
Settings
Measure
-
ment
or
or
29
f
ter entering setup mode, select “Clean” and press to activate and configure cleaning
A settings. Select “Auto” to turn on the function as automatic. Set the timer for the clean function to automatically turn cleaning on or off, and then set the Hysteresis value (Hys.). The relationship of these settings can be seen in the graph provided below.
Note: When the clean function is turned on, if any value is set to 0, the instrument will automatically turn off this function. When the clean function is activated under measurement mode, a “Clean Running” message will show on the top of the display. The measurement value will remain at the last measured value before cleaning. When entering settings menu or calibration menu while the clean function is activated, the instrument will automatically stop the cleaning action.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press or to select to activate cleaning. If turned off, the menu will proceed to setup of Analog
Press to confirm it.
Press to adjust the cleaning ON duration.
Press to confirm minutes and proceed to seconds.
Press to adjust the cleaning OFF duration.
Press to confirm minutes and proceed to seconds.
T
on
off
off
off
off
T
T
on
hys
Clean Timer Control
Press to confirm it.
Press to confirm it.
Enter “Analog” Setup
off
off
T
T
T
off
on
Press or to adjust Clean hysteresis (Hys.) time.
Press or to select
o activate cleaning test. If
t turned ON, the Clean function will begin and the “WASH” indicator will light up.
hys
T
off
7.11 Analog Output
30
f
ter entering setup mode, select “Analog” and press to configure analog output settings.
A Select 0~20 mA or 4~20 mA current output. Set the upper and lower limit relative to the desired measuring range. The smaller the range, the greater the current output resolution. When the measured value exceeds the higher range limit, the current will remain approximately 22 mA output. When the measured value exceeds the lower range limit, under 0~20 mA mode the current output will remain 0mA, under 4~20 mA mode the current output will remain approximately 2 mA. The exceptional output values can be used as an indication for determination failure. Under HOLD (measurement) status, the current output maintains the last output value before HOLD status. For the convenience of ensuring the current settings of an external recorder or of a PLC controller, the current output will be 0/4 mA or 20 mA under the analog output setup menu.
Press to confirm it.
Press or to select for current output range, 0-20 mA or 4-20 mA.
Press to confirm it.
Press or to set the range lower limit relative to 4 mA output signal.
Press to confirm it.
Press or to set the range lower limit relative to 0 mA output signal.
Press to confirm it.
Press or to set the range upper limit relative to 20 mA output signal.
Press to confirm it.
Enter “Clock” Setup
7.12 Date/Time (Clock)
ress
31
f
ter entering setup mode, select “Clock” and press to set or edit the date and time. If the
A clock is set to off, a clock will not be displayed under measurement mode. The calibration time of calibration records will also show “OFF” under calibration overview display. Note: The transmitter is able to clock in action even if the power supply is cut off due to an internal battery. If this fails, please replace the 3V CR2025 Li battery inside the transmitter.
Press to confirm it.
Press or to turn Clock on or off. If off, the guide menu will proceed directly to RS-485 setup.
Press to confirm it.
Press or to set th
e year.
Press to confirm it.
Press or to adjust the month. P to proceed to adjust the date.
Press to confirm it.
Enter “RS-485 Communication” Setup
Press to confirm it.
Press or to set the hour. Press to proceed to adjust the minute.
32
7.13 RS-485 Communication
After entering setup mode, select “RS-485” and press to configure RS-485 communication. According to the Modbus protocol, set the transmitting mode, parity, baud rate, stop bit, and ID number. For Modbus protocol details, please refer to chapter 9. When under hold status, the measurement signal output maintains the last output value before hold.
Press to confirm it.
Press or to select RTU or ASCII mode.
Press to confirm it.
Press or to select baud rate from 2400, 4800, 9600 or 19200.
Press to confirm it.
Press or to select parity check from Even, Odd or None.
Press to confirm it.
Press or to select stop bit as 1 bit or 2 bit.
Press to confirm it.
Press or to set the ID number of the transmitter. The valid value is from 1 to 247.
Press to confirm it.
Enter “Sample Measurement Average (Digital Filter)” Setup
33
7.14 Sample Measurement Average (Digital Filter)
After entering setup mode, select “Digital Filter” and press to configure transmitter digital filter. You may select the number of sample measurements to be averaged for each reading to increase the stability of the displayed measurement. The greater the number, the more stable the measurement value; the smaller the number, the more acute the measurement value.
Press to confirm it.
Press or to set the number of samples to be averaged.
Enter “Back Light” Setup
Press to confirm it.
34
7.15 Backlight
After entering setup mode, select “Backlight” and press to adjust display brightness (-2~2, dark ~ bright) and sensitivity of the brightness sensor (-2~2, insensitive ~ sensitive). Whether under OFF or AUTO mode, the touch-on function will activate the backlight when any button is pressed. If no buttons are pressed for 5 seconds, the display will return to the original backlight setting status.
ON setting: The backlight is always on. OFF setting: The backlight is off. When a button is pressed, it will enter touch-on status. Auto setting: Transmitter will activate or deactivate the backlight according to the ambient
lighting. When a button is pressed, it will enter touch-on status.
Press to confirm it.
Press or to select backlight mode.
ess or
Pr to select backlight mode.
Press to confirm it.
Press to confirm it.
Press or to select five backlight brightness levels from -2,
-1, 0, 1, 2.
Press or to select five backlight brightness levels from
-2, -1, 0, 1, 2.
Press to confirm it.
Press or to select five backlit sensitization levels from -2, -1, 0, 1, 2.
Press to confirm it.
Press to confirm it.
Press to confirm it.
Enter “Contrast” Setup
35
7.16 Contrast
After entering setup mode, select “Contrast” and press to adjust the display contrast (-2,
-1, 0, 1, 2, light to dark).
Press to confirm it.
Press or to select display contrast.
Press to confirm it.
Enter “Logbook” Setup
7.17 Logbook
36
f
ter entering setup mode, select “Logbook” and press to view event logbook. It is able to
A record and display up to 50 important events and error diagnosis records. The user can read important events via Logbook or Modbus. For detailed Modbus definitions of events please see to section 9.2, Modbus Address and Command Table.
Press to confirm it.
Press or to select different page.
Enter “Auto Return” Setup
Press to confirm it.
7.18 Automatic Return (Return)
37
f
ter entering setup mode, select “Return” and press to set the instrument to
A automatically exit the setup menu after a period of time without pressing any keys. “Manual Exit” requires the user to exit setup menu manually, while “Auto” will set the display to automatically exit the setup menu and return to measurement mode after a period of time without pressing any keys.
Press to confirm it.
Press or to select Auto return or manual exit.
Press to confirm it.
Press to confirm it.
Press or to adjust minutes, press to confirm and adjust seconds.
Press to confirm it.
Enter “Code” Setup
8. Calibration
38
Calibration Block Diagram
Return
Auto Manual
Input
Re
turn
Timer
Code
Input
P
asscode
OFF ON
New
Code
T
Select
C-100, TC-500, or TC-3000
Electrode
Known Sol.
TCS-1000
T
C-100
TC-500
TC-3000
Select First
B
uffer Range
Adjust First
Ca
l. Point
Display Current
from
Electrode
TCS-1000 Only
Select
Ca
l. Unit
TC-7310-RS
Ca
libration
Information
Select
TCS-1000
Electrode
Correction
F
actor
Adj. Correction
F
actor
Auto Display
C
al. Result
Press CAL
Select Second
B
uffer Range
Adjust Second
Ca
l. Point
Display Current
from
Electrode
Auto Display
C
al. Result
Press CAL
S
ame Procedure for Third and
Fourth Point
Select Fifth
B
uffer Range
Adjust Fifth
Ca
l. Point
Display Current
from Electrode
Return to previous
evel or action
l
Auto Display
al. Result
C
Press CAL
Select Turbidity/SS
CS-1000
T
Calibration Unit
NTUFTUFNU
TC-100 T
C-500
TC-3000
TCS-1000
Select Turbidity/SS
Calibration Unit
mg/lppm
8.1 Calibration Setup Menu
39
Under measurement mode, press and simultaneously to display Calibration Information. If you do not need to re-calibrate the measurement system, press to return to measurement mode. If you need to re-calibrate the system, press to enter calibration setup menu. (If the calibration time shows “OFF”, it represents that the clock function has been turned off.)
TCS-1000 Calibration Information
Press to confirm it
Measurement mode
Press and simultaneously
Pr
ess
Press to return to measurement mode
Press to confirm it.
Enter Calibration Setup Menu
TC-100 Calibration Information
40
Note: TC-100, TC-500, TC-3000 electrodes do not have correction factor function.
Press to confirm it
Measurement mode
Press and simultaneously
Pr
ess for next page.
Press to return
measurement mode
to
Press to confirm it.
Enter Calibration Setup Menu
8.2 Calibration Security Code (Code)
Enter
“Known Solution
” C
alibration
then press
41
e
lect the “Code” (passcode) after entering calibration setup mode to activate passcode
S function.
The preset calibration settings code is 1100.
When an incorrect code is entered, the display will show “Error Code”. Press to re-enter, or press to exit the calibration menu.
ess or
Pr
Press to confirm it.
The first “0” of digits “0000” will start to flash. Press or to adjust the value, to confirm it and proceed to the next digit.
Press to confirm it.
Press to confirm it.
Select to turn on or turn off code protection function. If turned on, please input a new code. A code will be required when you re-enter setup mode. Input the correct pass-cod to enter calibration setup menu.
Press to confirm it.
Press to confirm it.
Input new code
8.3 Calibration
Press or to
select the first point closest
to the known solution
Press or to
Press or to
standard solution value.
the sensor with
distilled water and place
for
42
s
each sensor has its unique characteristics, the transmitter is not calibrated to a sensor
A
by default (Cal. Point: No Cal). Calibration procedure must be performed by trained
personnel each time a new sensor is connected.
This instrument provides multi-point standard buffer solution (Known Sol.)
calibration allowing users to decide the number of points to calibrate to (max. 5 points).
When calibrating, users can select the following electrodes in setup mode for
calibration: Turbidity TC-100, TC-500, TC-3000, or Suspended Solids TCS-1000.
During calibration, the displayed information will be
slightly different based on the
sensor selected in setup mode.
During calibration, exceeding 80% of the sensor manufacturer’s suggested range is
not recommended.
For single point calibration, the reference value cannot be 0.
TCS-1000 electrode uses suspended solids (SS) standard solution, but is not easily
acquired. The transmitter is able to use turbidity standard solution to calibrate and
adjust the suspended solids measurement via a correction factor.
8.3.1 Known Solution Calibration
8.3.1.1 Suspended Solids (SS) TCS-1000 Electrode Calibration
During calibration, select SS unit: mg/l or ppm
or select Turbidity unit: NTU, FTU, or FNU
SS and Turbidity calibration each provides 5 preset standard solution values for
selection:
SS: 800.0, 400.0, 200.0, 100.0, 10.0 Turbidity: 1000, 800, 100, 20.0, 1.00
If a turbidity unit is selected for calibration, the s
ystem will procced directly to
correction factor adjustment following standard solution calibration.
The procedure for TCS-1000 using suspended solids standard solution for two-point
calibration is as follows:
value.
Press to confirm it.
Press to confirm it.
select the unit.
Press to confirm it.
adjust the Turbidity/S.S.
Rinse
into the first solution
measurement.
Press to confirm it.
The sensor will display
Press to directly
display the result or wait
to
Press to complete
point calibration
display the
Press to continue
Press or to
select the second point
closest to the known
Press to complete
Press or to
adjust the MLSS standard
solution value. Rinse the
sensor with distilled water
and place into the second
Press to directly
display the result or wait
to
received. Wait until the
43
the current received.
Wait until the value is
stable.
two-
and directly
result.
and proceed to multi-
po
int calibration
Press to confirm it.
for the transmitter
automatically confirm it.
Calibration results
overview
solution value.
single point calibration.
Press or to
o
ve between pages.
m
Press to confirm it.
solution for measurement.
Press to confirm it.
The transmitter will
display the current
value is stable.
Press to confirm it.
for the transmitter
automatically confirm it.
8.3.1.2 Turbidity TC-100, TC-500, TC-3000 Electrode Calibration
Press or to
st point closest
to the known solution
Press or to
adjust the Turbidity
standard solution value.
Rinse the sensor with
distilled water and place
into the first solution for
The sensor will display
Press to directly
display the result or wait
to
Press or to
select the second point
closest to the known
Press to complete
Press or to
adjust the Turbidity
standard solution value.
Rinse the sensor with
distilled water and place
into the second solution for
received. Wait until the
Press to directly
display the result or wait
to
Press to complete
point calibration
and directly display the
Press to continue
44
T
urbidity calibration each provides 5 preset standard solution values for selection:
TC-100: 80.0, 40.0, 20.0, 10.0, 1.0 TC-500: 400.0, 200.0, 100.0, 50.0, 5.0 TC-3000: 2400, 1200, 600, 300, 30
solution value.
Press to confirm it.
Press to confirm it.
single point calibration.
select the fir
value.
Press to confirm it.
measurement.
Press to confirm it.
The transmitter will
display the current
Press to confirm it.
measurement.
Press to confirm it.
value is stable.
Press to confirm it.
the current received.
Wait until the value is
stable.
for the transmitter
automatically confirm it.
two-
for the transmitter
automatically confirm it.
Press to confirm it.
result.
and proceed to multi-
po
int calibration
Known Solution Calibration
Display
Calibration Result
Rinse the sensor with distilled water
Suspended
Wait until the
factor
n
target
45
Calibration results
overview
Press or to
mo
ve between pages.
or
8.4 C
rection Factor
When calibrating TCS-1000 suspended solids electrode us
ing turbidity standard solution,
a correction factor needs to be applied to adjust the measured values for suspended solids
measurement values.
Once “Known Solution Calibration” is complete (refer to section 8.3.1.1 for TCS-1000
suspended solids electrode calibration procedure), correction factor adjustment screen will
follow. Users may also directly enter the correction factor adjustment screen from setup
mode.
Press to confirm it.
and place into known
Solids standard solution.
value is stable, then adjust the F-
using____ and until the mai
measurement value reflects the
Suspended Solids value.
Press to confirm it.
8.5 Automatic Return (Return)
46
f
ter entering setup mode, select “Return” and press to set the instrument to
A automatically exit the setup menu after a period of time without pressing any keys. “Manual Exit” requires the user to exit setup menu manually, while “Auto” will set the display to automatically exit the setup menu and return to measurement mode after a period of time without pressing any keys. Note: The return function for setup menu and calibration setup menu are independent settings.
Press to confirm it.
Press or to select Auto return or manual exit.
Press to confirm it.
Press to confirm it.
Press or to adjust minutes, press to confirm and adjust seconds.
Press to confirm it.
Enter “Calibration Code” Setup
9. MODBUS Protocol and Instructions
GND
Pull-low
47
9.1 Communication Connection
The transmitter’s RS-485 communication port features electronic isolation
protection, lightning protection, and provides an isolated ground terminal. It is able
to connect using standard twisted pair (segregated, double-stranded) cables. The
positive connection point D+(B) of all devices are connected together with one
strand of the double-stranded twisted-pair cable, and all the negative contacts D-(A)
with the other strand; the isolated wire shield must be connected to the ground
terminal (GND). In the laboratory, a standalone master-slave communication is
relatively simple. Therefore, standard cables can be used. However, for industrial
settings, strict engineering construction requirements must be taken into account for
proper installation. The wiring diagram is as follows:
V
Bus
Pull-high
re
sistor
R
pull-high
Terminator
R
T
D+(B)
GN
D Master
D-(A)
resistor R
Depends on site
co
low
nditions
pull-
Other
Mo
dbus
equipment
.....
TC-7310-RS
ID
: 02
TC-7310-RS
ID
: 01
Note:
1. The transmitter’s RS-485 interface is equipped with a protective ground terminal. When
ommunicating with the RS-485, the ground terminal must be used to eliminate safety risks.
c
2. Use 120 ohm impedance matching resistors at the signal cable terminals (D+(B) and D-(A)) of the
terminal devices to effectively reduce or eliminate signal reflection. Note: Three common causes of connection failures:
a. Open circuits: the signal cable has an open circuit. b. Short circuits: there is insufficient insulation between signal cables causing a short circuit. c. Idle-bus: there is no data transmission in the Controlbus.
The three conditions above can cause an unknown state of the Controlbus voltage, leading to the receiver receiving unexpected signals. Therefore, in complicated setups like multi-transmitter connections, the Controlbus requires a terminating resistor to avoid baud rate reflection, in addition to biasing circuits composed of pull-high and pull-low resistors, and voltage source (VBus). As seen in the diagram above, make sure the signal cable terminals D+(B) and D-(A) has free voltage in between to keep the signals complete. Some systems may not be suitable for a fixed resistor and VBus,
therefore users will need to be more aware of the settings. Some equipment parts (RS-485 converters
48
or modules) provide terminating, pull-high, pull-low resistors, and VBus settings. Please refer to the corresponding operation manuals before installation and connection. For VBus, the standard is 5 V, for pull-high and pull-low resistors, 1 K, and for terminating resistors, 120 K.
3. Without repeaters, the RS-485 network cannot exceed a maximum of 32 nodes. The maximum
transmission communication distance for RS-485 is up to 1200 meters. For long distance transmission, it is recommended to use cables that are specifically design for RS-485.
4. When communicating, all equipment within the network should be consistently kept in the same
transfer mode, baud rate, and parity. All devices must be under different addresses to prevent conflicts within the network communications.
5. The transmitter’s Modbus command can only access 50 registers. Abnormal information will be
returned if the maximum length is exceeded.
6. The idle time in which a slave instrument responds to a master machine is different according to
each model. Generally, the time needed should be longer than 0.5 seconds (some models may require a longer waiting-responding time, please note if the operation manual specifies).
9.2 MODBUS Name and Address Table
49
Modbus response table is as follows. As users communicate with transmitters through PLC or Man­machine Interface, check to see if the transmission of address subtracts 1 by default. If so, add 1 onto each address to match the table; e.g. for Turbidity/SS measurement logic address 0035H (16-bit) or 53 (10-bit), if the output signal is subtracted by 1 by the PLC or Man-machine Interface, users must input 0036H (16-bit) or 54 (10-bit) to match the appropriate logic address.
Function Code: 03H, 06, 10H Modbus Response (Setup Parameter)
Logic
Address
0000H None 0001H R Equipment’s ID
R/W
Item
Number of
Bytes
2 USHORT 1-247 1
Information
Type
Description of Data
Transmission
Default
Value
Note
0002H R
0005H R
0006H R
0007H R Parity 2 USHORT
0008H 0009H USHORT Minute
000AH USHORT Hour
R/W
000BH USHORT Day 000CH USHORT Month
Transmitter
Model
Communication
Protocol
Serial
Transmission
Speed
(Baud Rate)
Real-Time
Clock*
6 USHORT ASCII Code TC7310
0: RTU
2 USHORT
1: ASCII 0: 2400 1: 4800
2 USHORT
2: 9600 3: 19200 0: None
2: Odd
USHORT Second
12
0
3
1 1: Even
2016-01-
01,
00:00:00
000DH USHORT Year
000EH R/W Code Setting* 2 USHORT Code setting 1111
000FH R/W
USHORT 0: OFF
0010H
0011H 2 USHORT ON.S: 0-5999 0 Second 0012H 2 USHORT OFF.H: 0-999 0 Hour
R/W
0013H 2 USHORT OFF.M: 0-59 0 Minute 0014H 2 USHORT Hys.S: 0-9999 0 Second
Clean Relay*
2
USHORT 1: AUTO
Reserved
0
0015H
50
2 USHORT
0: OFF
1
1: AUTO
0016H 2 USHORT
0
1: Lo
0: Hi
R/W
Relay 1 *
0017H 4 FLOAT SP1 800.0 mg/l
0019H 4 FLOAT Hys1 10.0 mg/l
0: OFF
001BH
2 USHORT
1
1: AUTO 0: Hi
001CH 2 USHORT
1
1: Lo
R/W
Relay 2 *
001DH 4 FLOAT SP2 200.0 mg/l
001FH 4 FLOAT Hys2 10.0 mg/l
0: AUTO
0021H
2 USHORT
2 1: ON
Data
affected
by sign
byte
Data
affected
by sign
byte
R/W
Backlight
Brightness*
0022H 2
Backlight
0023H R/W
Sensitivity*
Sample Average
0024H R/W
of
Measurements
Product
0025H R/W
Adjustment
2: OFF SHORT 2: Highest Brightness SHORT 1: High Brightness SHORT 0: Standard
0 SHORT -1: Low Brightness SHORT -2: Lowest Brightness SHORT 2: Highest Sensitivity SHORT 1: High Sensitivity
2
SHORT 0: Standard
0 SHORT -1: Low Sensitivity SHORT -2: Lowest Sensitivity
2 USHORT 1-60 30
TC-100: -20~20NTU TC-500: -100~100NTU
4 FLOAT
0
TC-3000: -600~600NTU
0027H R/W
Wiper* 2 USHORT
TCS-1000: - 200~200mg/l 0MANUAL
0
1AUTO
0028H R/W Event Number*
51
2 USHORT 1~50 1
0029H R
002AH R USHORT Minute
002BH R USHORT Hour 002CH R USHORT Day
002DH R USHORT Month
002EH R USHORT Year 002FH R Event ID 2 USHORT 0~15 0
0030H R/W
Event Time 12
Wiper Act. Time
2 USHORT Minute 09
Interval
USHORT Second
2016-01-01
00:00:00
Note 1: Actions without * only supports function code 03H. Actions with * supports function code 03H, 06H, 10H. Note 2: FLOAT is a 32-bit IEEE 754 format. The above table is divided into two 16-bit register data transmissions.
The second 16-bit register (CC CD) will be transferred first, followed by the first 16-bit register (41 C8). Every 16-bit format is high-bit in the front and low-bit in the back. For example, if the Turbidity/SS measured value is 25.10NTU, the 16-bit of FLOAT data (hexadecimal) will be 41 C8 CC CD. The transmission order will be
CC CD 41 C8. For detailed description, please refer to section 9.3, Modbus example description. Note 3: USHORT represents unsigned short integer. Note 4: Event Logbook ID
If the user requires Modbus to retrieve information from the logbook, 0028H must be input first before reading 0029H~002F corresponding event. Logbook ID #1 represents the newest recorded event, ID #2 represents the last recorded event, and so on.
Event Definition Modbus ID #
Mea mode Measurement Mode 00
Set mode Setup Mode 01 Cal mode Calibration Mode 02
Power On Power on 03
Power Off Power off 04
NTU_mA Over Current over range 05
FTU_mA Over Current over range 06
Error 2 Slope over range 07 Error 3 Calibration value not stable 08 Error 4 Cannot identify standard solution 09 Error 5 Incorrect passcode 10 Error 6 Electrode not connected or not working 11
FNU_mA Over Current over range 12
Modbus Write Modbus input action 13
ppm_mA Over Current over range 14
mg/l_mA Over Current over range 15
Function Code: 03H Modbus Response (Measurement Parameter)
S
tatus
NTU
0050H
52
Logic
Address
R/W Item
Number of Bytes
Information
Type
Description of Data
Transmission
Default
Value
Note
0031H
0032H
0035H
0037H-
R
R
R
Measurement
Status
Sign Byte
Turbidity/SS
Measurement
2 USHORT
6 CHAR
4 FLOAT
Factory Reserved
1: Measurement
Function Code: 01H Modbus Response (Dispersion Parameter)
Logic
Address
0070H 0071H 0072H
Item BIT
Description Default Value Note
LO Alarm 1 Contact on 0 (Contact off)
Hi Alarm 1 Contact on 0 (Contact off)
mA too high 1
Contact on
0: Hold Status
FTU
FNU
ppm mg/l
Turbidity/SS
Measurement
0 (Contact off)
1
mg/l
ASCII
Code
0073H
mA too low 1 0074H 0075H Turbidity/SS out of range 0076H
0077H 0078H
RLY1 Action * 1 Contact on 0 (Contact off)
RLY2 Action* 1 Contact on 0 (Contact off)
Clean Action* 1 Contact on 0 (Contact off)
0079H-
008FH
1
Contact on
0 (Contact off)
Factory Reserved
Contact on 0 (Contact off)
Factory Reserved
9.2 MODBUS Name and Address Table
53
Modbus response table is as follows. As users communicate with transmitters through PLC or Man­machine Interface, check to see if the transmission of address subtracts 1 by default. If so, add 1 onto each address to match the table; e.g. for Turbidity/SS measurement logic address 0035H (16-bit) or 53 (10-bit), if the output signal is subtracted by 1 by the PLC or Man-machine Interface, users must input 0036H (16-bit) or 54 (10-bit) to match the appropriate logic address.
Function Code: 03H, 06, 10H Modbus Response (Setup Parameter)
Logic
Address
0000H None 0001H R Equipment’s ID
R/W
Item
Number of
Bytes
2 USHORT 1-247 1
Information
Type
Description of Data
Transmission
Default
Value
Note
0002H R
0005H R
0006H R
0007H R Parity 2 USHORT
0008H 0009H USHORT Minute
000AH USHORT Hour
R/W
000BH USHORT Day 000CH USHORT Month
Transmitter
Model
Communication
Protocol
Serial
Transmission
Speed
(Baud Rate)
Real-Time
Clock*
6 USHORT ASCII Code TC7310
0: RTU
2 USHORT
1: ASCII 0: 2400 1: 4800
2 USHORT
2: 9600 3: 19200 0: None
2: Odd
USHORT Second
12
0
3
1 1: Even
2016-01-
01,
00:00:00
000DH USHORT Year
000EH R/W Code Setting* 2 USHORT Code setting 1111
000FH R/W
USHORT 0: OFF
0010H
0011H 2 USHORT ON.S: 0-5999 0 Second 0012H 2 USHORT OFF.H: 0-999 0 Hour
R/W
0013H 2 USHORT OFF.M: 0-59 0 Minute 0014H 2 USHORT Hys.S: 0-9999 0 Second
Clean Relay*
2
USHORT 1: AUTO
Reserved
0
0015H
54
2 USHORT
0: OFF
1
1: AUTO
0016H 2 USHORT
0
1: Lo
0: Hi
R/W
Relay 1 *
0017H 4 FLOAT SP1 800.0 mg/l
0019H 4 FLOAT Hys1 10.0 mg/l
0: OFF
001BH
2 USHORT
1
1: AUTO 0: Hi
001CH 2 USHORT
1
1: Lo
R/W
Relay 2 *
001DH 4 FLOAT SP2 200.0 mg/l
001FH 4 FLOAT Hys2 10.0 mg/l
0: AUTO
0021H
2 USHORT
2 1: ON
Data
affected
by sign
byte
Data
affected
by sign
byte
R/W
Backlight
Brightness*
0022H 2
Backlight
0023H R/W
Sensitivity*
Sample Average
0024H R/W
of
Measurements
Product
0025H R/W
Adjustment
2: OFF SHORT 2: Highest Brightness SHORT 1: High Brightness SHORT 0: Standard
0 SHORT -1: Low Brightness SHORT -2: Lowest Brightness SHORT 2: Highest Sensitivity SHORT 1: High Sensitivity
2
SHORT 0: Standard
0 SHORT -1: Low Sensitivity SHORT -2: Lowest Sensitivity
2 USHORT 1-60 30
TC-100: -20~20NTU TC-500: -100~100NTU
4 FLOAT
0
TC-3000: -600~600NTU
0027H R/W
Wiper* 2 USHORT
TCS-1000: - 200~200mg/l 0MANUAL
0
1AUTO
0028H R/W Event Number*
55
2 USHORT 1~50 1
0029H R
002AH R USHORT Minute
002BH R USHORT Hour 002CH R USHORT Day
002DH R USHORT Month
002EH R USHORT Year 002FH R Event ID 2 USHORT 0~15 0
0030H R/W
Event Time 12
Wiper Act. Time
2 USHORT Minute 09
Interval
USHORT Second
2016-01-01
00:00:00
Note 1: Actions without * only supports function code 03H. Actions with * supports function code 03H, 06H, 10H. Note 2: FLOAT is a 32-bit IEEE 754 format. The above table is divided into two 16-bit register data transmissions.
The second 16-bit register (CC CD) will be transferred first, followed by the first 16-bit register (41 C8). Every 16-bit format is high-bit in the front and low-bit in the back. For example, if the Turbidity/SS measured value is 25.10NTU, the 16-bit of FLOAT data (hexadecimal) will be 41 C8 CC CD. The transmission order will be
CC CD 41 C8. For detailed description, please refer to section 9.3, Modbus example description. Note 3: USHORT represents unsigned short integer. Note 4: Event Logbook ID
If the user requires Modbus to retrieve information from the logbook, 0028H must be input first before reading 0029H~002F corresponding event. Logbook ID #1 represents the newest recorded event, ID #2 represents the last recorded event, and so on.
Event Definition Modbus ID #
Mea mode Measurement Mode 00
Set mode Setup Mode 01 Cal mode Calibration Mode 02
Power On Power on 03
Power Off Power off 04
NTU_mA Over Current over range 05
FTU_mA Over Current over range 06
Error 2 Slope over range 07 Error 3 Calibration value not stable 08 Error 4 Cannot identify standard solution 09 Error 5 Incorrect passcode 10 Error 6 Electrode not connected or not working 11
FNU_mA Over Current over range 12
Modbus Write Modbus input action 13
ppm_mA Over Current over range 14
mg/l_mA Over Current over range 15
Function Code: 03H Modbus Response (Measurement Parameter)
S
tatus
NTU
0050H
56
Logic
Address
R/W Item
Number of Bytes
Information
Type
Description of Data
Transmission
Default
Value
Note
0031H
0032H
0035H
0037H-
R
R
R
Measurement
Status
Sign Byte
Turbidity/SS
Measurement
2 USHORT
6 CHAR
4 FLOAT
Factory Reserved
1: Measurement
Function Code: 01H Modbus Response (Dispersion Parameter)
Logic
Address
0070H 0071H 0072H
Item BIT
Description Default Value Note
LO Alarm 1 Contact on 0 (Contact off)
Hi Alarm 1 Contact on 0 (Contact off)
MA too high 1
Contact on
0: Hold Status
FTU
FNU
ppm mg/l
Turbidity/SS
Measurement
0 (Contact off)
1
mg/l
ASCII
Code
0073H
MA too low 1 0074H 0075H Turbidity/SS out of range 0076H
0077H 0078H
RLY1 Action * 1 Contact on 0 (Contact off)
RLY2 Action* 1 Contact on 0 (Contact off)
Clean Action* 1 Contact on 0 (Contact off)
0079H-
008FH
1
Contact on
0 (Contact off)
Factory Reserved
Contact on 0 (Contact off)
Factory Reserved
9.3 Modbus Description Example (e.g.: Function Code 03H)
57
he
following description uses Turbidity/SS reading 0035H as an example. If the
T Turbidity/SS measured value is 25.10 NTU, confirm that the host and sub-machine communication format settings are correct. The host sends request commands according to the left column of the table below, then receives a corresponding response from the sub-machine according to the right column. This example shows the function code 03H as transmission data format; for other function codes, the same logic applies.
ASCII Mode:
Request
Response
Message Framing Hex Message Framing Hex ID, Address 01 ID, Address 01 Function code 03 Function code 03 Starting Address Hi 00 Byte Count 04 Starting Address Lo 35 Register value Hi CC No. of Registers Hi 00 Register value Lo CD No. of Registers Lo 02 Register value Hi 41 LRC C5 Register value Lo C8 LRC 56
RTU Mode:
Request
Response
Message Framing Hex Message Framing Hex ID, Address 01 ID, Address 01 Function code 03 Function code 03 Starting Address Hi 00 Byte Count 04 Starting Address Lo 35 Register value Hi CC No. of Registers Hi 00 Register value Lo CD No. of Registers Lo 02 Register value Hi 41 CRC Check Lo 0D Register value Lo C8 CRC Check Hi 45 CRC Check Lo 65 CRC Check Hi 5A
Note: FLOAT is a 32-bit IEEE 754 format. The above table is divided into two 16-bit
register data transfers. The back 16-bit register (CC CD) is transferred first, followed by the front 16-bit register (41 C8). Every 16-bit format is high-bit in the front and low-bit in the post. For example, if the Turbidity/SS reading is now 25.10 NTU, the 16-bit of FLOAT data (hexadecimal) is 41 C8 CC CD. However, the transmission order will be CC CD 41 C8.
10. Error Messages (Error Code)
58
Messages Reason Dispositions
Error2
Error3
Error4
Error5
Error6
Slope exceeds the upper or lower limit
The readout is unstable
during calibration
Standard solution cannot be
recognized
Incorrect passcode
ERROR CODE
No sensor connection or
sensor failure
Please check if the standard
solution has been re-used during
the calibration or is expired.
1. Please check and clear electrode
glass end of air bubbles,
maintain or replace electrode
and make another calibration.
1. Please replace and use new
standard solution.
2. Maintain or replace electrode
and make another calibration.
Re-enter passcode.
1. Please check if the sensor is
connected to the instrument.
2. Maintain the electrode or
replace a new electrode, and
make another calibration.
Serious error that does not
Error9
permit any further measurements
Please call service engineer.
11. Maintenance
Under normal operation and conditions, the transmitter does not need additional maintenance other
than scheduled cleaning and calibration of the electrode to ensure accurate and stable measurements.
SUNTEX INSTRUMENTS CO., LTD.
13F, No. 31, Lane 169, Kangning St., Xizhi Dist., New Taipei City, Taiwan (R.O.C.) Tel: 886-2-2695-9688 Fax: 886-2-2695-9693 e-mail: suntex@ms1.hinet.net
5040E TC-7310-RS/Technical data subject to alternations/ Quality Systems ISO 9001/201712
www.suntex.com.tw/en
Loading...