24
8. Failure recovery
Failure phenomenon Possible causes Disposition
ERR1 The electrode coefficient bears a
large deviation when use the
standard solution in calibration.
Replace new standard solution for another
calibration. For others, please refer to the
disposition of measurement deviation in
the following.
ERR2 The measurement is unstable when
use the standard solution in
calibration.
Please refer to the disposition of unstable
measurement in the following.
The electrode is polluted. Clean and wash the electrode
The electrode coefficient (C) is
wrongly set.
Modify the electrode coefficient
The temperature coefficient (TC) is
wrongly set.
Modify the temperature compensation
coefficient.
The electrode is damaged or
broken-down.
Readjust the coefficient
There are air bubbles on the
electrode.
Churn up to remove the bubbles.
The battery is low. Replace with new battery.
Measurement deviation
Instrument fault Send to the original factory for repair.
The electrode is disturbed by the
bubbles.
Churn up to remove the bubbles.
The electrode is polluted. Clean and wash the electrode.
Electrode fault Replace the electrode.
The battery is low. Replace with new battery.
Unstable measurement
Instrument fault Send to the original factory for repair.
The electrode plugs is not inserted
into the required place.
Check the electrode plugs and reinsert it
into the required place.
The conductivity value is
zero or the special
resistivity value is out of
the range.
The electrode is broken. Replace the electrode.
Not soak deeply enough. Deepen the soaked length to ensure the
accuracy of temperature.
Deviation of standard comparison Use standard thermometer comparison to
correct the difference in temperature
The temperature sensor is broken Replace the electrode.
The battery is low. Replace with new battery.
Temperature deviation
Instrument fault Send to the original factory for repair.
The electrode plugs is not inserted
into the required place.
Check the electrode plugs and reinsert it
into the required place.
The temperature sensor is broken Replace the electrode.
Temperature is wrongly
displayed.
Instrument fault Send to the original factory for repair.