Suntex EC-4110-I, EC-4110-ICON Operation Manual

1
EC-4110-I EC-4110-ICON
ġ
Intelligent
ġ ġ
Transmitter
Operation
Manual
Precautions for installation Brief Instruction
1
1. Specifications
1.1 Specifications
4
1.2 Product packing 5
1.3 Measurement principle
6
2. Assembly and installation
2.1 Transmitter installation 8
2.2 Illustration of panel mounting 8
2.3 Illustration of wall mounting and pipe mounting
9
3. Overview of Conductivity Transmitter EC-4110-I/EC-4110-ICON
3.1 Illustration of rear panel 10
3.2 Illustration of terminal function
10
3.3 Description of terminal function 11
3.4 Wiring of cable 12
3.5 Circuit of cable
12
3.6 Illustration of electrical connection 13
3.7 Wiring diagram of extension cable
14
4. Configuration
4.1 Illustration of front panel
15
4.2 Keypad 15
4.3 LED indicators 15
5. Operation
5.1 Measurement mode
16
5.2 Set-up menu 16
5.3 Calibration menu
16
5.4 Shortcuts 16
5.5 Default value
5.5.1 Setting default value 16
5.5.2 Calibration default value
17
6. Measurement display mode
6.1 T ext mode
18
6.2 Real-time chart mode
19
6.3 Trace mode
20
6.4 Warning symbols and text
21
CONTENTS
7. Settings
Block diagram of settings
22
7.1 Entry of set-up menu 24
7.2 Security code of settings(Code)
25
7.3 Language 26
7.4 Measurement parameters(Mode) 27
7.4.1 Conductivity with temperature compensation 27
7.4.2 Absolute conductivity
29
7.4.3 Concentration 31
7.4.3.1 Defined concentration table
33
7.4.4 TDS 37
7.4.5 Salinity
38
7.5 Product adjustment 39
7.6 Temperature 40
7.7 Temperature compensation coefficient
41
7.8 Relay 1 43
7.9 Relay 2 44
7.10 Clean
45
7.11 Current output 46
7.12 RS-485 communication 48
7.13 Date/Time (Clock)
49
7.14 Sample average of measurements (Digital filter)
50
7.15 Backlight 51
7.16 Contrast
52
7.17 Logbook 53
7.18 Automatically back to measurement mode(Return) 54
8. Calibration
Block diagram of Calibration
55
8.1 Entry of calibration menu 56
8.2 Security password of calibration
57
8.3 Zero-point calibration
58
8.4 Cell constant calibration
59
8.5 Standard solution calibration(Std. solution)
60
8.5.1 KCl standard solution 60
8.5.1 NaCl standard solution
62
8.6 Automatically back to measurement mode(Return) 64
Thank you for purchasing Suntex products. In order to continually improve and enhance the transmitter’s function, Suntex reserves the right to modify the content and icon display of the product. The actual situation is subject to the instrument without notice. The operation manual is only provided for function and installation description, Suntex Instruments Co., Ltd. is not liable for any person or entity for any direct or indirect loss or damage due to improper usage of this product. If you have any questions or find omission, negligence or mistakes of the operation manual, please contact with our staff, thank you.
Precautions for installation
Wrong wiring will lead to breakdown or electrical shock of the instrument, please read this operation manual clearly before installation. zMake sure to remove AC power from the transmitter before wiring input, output connections,
and remove it before opening the transmitter’s housing.
zThe installation site of the transmitter should be good in ventilation and avoid direct
sunshine.
zThe material of signal cable should be special coaxial cable. Strongly recommend using our
coaxial cable. Do not use normal wires instead.
zAvoid electrical surge when using power. Especially when using three-phase power, use
ground wire correctly. If the power surges interference occurs, separate the power supply of transmitter from the control device, such as: dosing machines, mixers, etc. to make individual power supply for the transmitter; or set surge absorber to reduce the power surges at all electromagnetic switches and power control device coils.
zThe internal relay contact of the instruments is for alarm or control function. Due to safety,
please must connect to external relays which can stand enough ampere to make sure the safety operation of the instrument.˄Please refer to chapter 3.6ȾIllustration of electrical
connectionȿ˅
zThere a manufacturer logo usually shows in the display of transmitter, and the illustration of
each function in the manual is no longer expressed.
!
!
!
!
!
!
!
!
!
!
!
9. Modbus protocol and instructions 65
9.1 Communication connection
65
9.2 Modbus address table 67
9.3 Modbus example descriptions
71
10. Error messages (Error code)
72
11. Installation of cells
11.1 Appearance of cells
73
11.2 Installation approach
74
11.2.1 Flange installation 74
11.2.2 Tee installation
75
11.2.3 Immersion installation 76
12. Appendix
77
Brief Instruction
Description of set-up settings (see chapter 7 for details)
Press and simultaneously to see the overview of the set-up settings now. Then press I if you would like to modify set-up settings. Press keypad according to index of keypad on the screen.
Index of keypad
keypad Accordingly item Description
Back to upper layer
Choose leftward of change to left page
Increase digit
Choose rightward of change to right page
Decrease digit Confirm settings after modifications and then go through next
step
Selection of set-up items
keypad Accordingly item Description
Mode
Measurement mode, to choose Conductivity with temp. compensation (Cond.@tref), Concentration, TDS, Salinity or Absolute Conductivity (Cond.@tx) and measurement mode can be displayed in Text Mode, Real-time Mode, or Trace Mode.
Product Adj. Sample reading adjustment—for resistivity mode only.
Temperature
Temperature measurement and compensation, including MTC, PTC1000, PTC100, or NTC (4 types in total). MTC---Manual Temperature Compensation, PTC1000/PTC100/NTC--- Auto Temperature Compensation
Compensation
Temperature compensation setting, selection from linear(Lin.), non-linear(Non-Lin.), no compensation(Lin., 0.0%), 3 types
Relay 1 First relay setting, to choose action off or Hi/Lo alarm
к
Relay 2 Second relay setting, to choose action off or Hi/Lo alarm
Clean
Automatic wash time setting, to choose electrode clean
equipment’s ON and OFF duration
Current
Output
Current output according to Cond.@tref, Concentration, TDS, Salinity, or Cond.@tx setting range
RS-485 RS485 serial interface (Modbus protocol)
Clock
Clock setting (When out of power and reboot it, the
instrument’s time setting can maintain to the real time. If not, please replace the inner 3V CR2025 battery.)
Digital Filter
Take every serial 1~60 measurements, average them
continuously, and make it as the readings
Back Light
Backlight setting, to set Auto/ON/OFF backlight, brightness,
and sensitivity
Contrast Contrast of screen setting
Logbook Event recorder logbook (50 data)
Return Setting of returning to the measurement mode
Code
Security code of set-up mode. The set-up code is precedential to
calibration code, thus it can pass a different security code of calibration.
Language Available for English, Traditional Chinese, Simplified Chinese
л
Description of calibration settings (see chapter 8 for details)ġ ġġġġġġġġġġġġġ
Press and simultaneously to see the last calibration information. Then press if you would like to make a new calibration or modify setting of calibration. Press keypad according to index of keypad on the screen.
Index of keypad:
keypad Accordingly item Description
Back to upper layer
Choose leftward of change to left page
Increase digit
Choose rightward of change to right page
Decrease digit Confirm settings after modifications and then go through next
step
Selection of calibration items
keypad Accordingly item Description
Cell Constant
To adjust the instrument cell constant setting until the value goes the same with the given cell constant of the sensor
Std. Solution Use the appropriate standard solution to calibrate the system
Zero Zero-point calibration
Return Time interval setting of returning to the measurement mode
Code Security code of calibration mode.
Note
ġ
Due to the need for continuous improvement of the transmitter function, we reserve the right to modify the content and the icon of the function. The actual icons and contents are subject to the instrument without notice.
м
1. Specifications
1.1 Specifications
Model EC-4110-I EC-4110-ICON
Measuring modes Conductivity/TDS/Salinity/Temp./Concentration (EC-4110-ICON only)
Ranges
Conductivity
0.000 ȝS/cm~2000 mS/cm, Auto or Fixed
Salinity
0.0 ppt~70.0ppt
TDS 0ppm~19999ppm; 0.00~199.9ppt
Temp. PTC1000/PTC100: -30.0~200.0ÛC, NTC30K: -30.0~130.0ÛC
Concentration
NaCl_28%0-28 % HCl_18%0-18 % HCl_39%22-39 % HNO3_30%0-30 % HNO3_96%35-96 % NaOH_24%0-24 % NaOH_50%15-50 % H2SO4_37%0-37 % H2SO4_88%28-88 % H2SO4_99%89-99 % H3PO4_35%0-35 %
The boundary value of conc. conversion range of each solution can vary with changes of temp., if go over range, there twinkles an alarm on the display screen.
Self-defined table (up to 9 temp. data, and there are 9 corresponding points for cond. v.s. conc. at each temp.)(9 x 9 matrix)
Resolutions
Concentration
0.01 %
Conductivity
0.1 ȝS/cm, 0.01 / 0.1 / 1 mS/cm
Temp. 0.1ÛC
Accuracy
Conductivity
±1% (± 1Digit)
Temp.
±0.2qC (± 1 Digit),
Equipped with temperature error correction function
Temperature
Compensation
Automatic with NTC30K/ PTC1000/PTC100
Manual T emperature Compensation
Calibration mode
(1)Manual cell constant adjustment (2)Conductivity standard solution calibration (3)Zero-point calibration
Product Adjustment 0.7000~1.3000
Ambient Temp. 0~50 ÛC
Storage Temp. -20~70 ÛC
Cell Constant freely selectable 0.008~9.9999 cm-1
н
Temperature Coefficient
Linear temperature compensation from (0.00%~ 40.00%),
Non-Linear compensation, or No compensation
Compensation Temperature Freely set reference temperature 0~200 ÛC
Display Screen
Large LCM with sensitization sensor for auto/manual illumination
function and contract functio
n
Languages
English / Traditional Chinese / Simplified Chinese
Display
Text mode: Numerical display
Chart mode: 3 mins real-time dynamic graph
(
conc. and cond. only)
Trace mode: Set up from 3 mins to four weeks duration of the
measured value trend graph
(
conc. and cond. only)
Analog output
Isolated DC 0/4~20mA corresponding to main measurement,
max. load 500:
Logbook 50 event records
RS-485 interface Isolated (MODBUS RTU or ASCII)
Modbus communication
Offer measuring value, calibration data, product adjustment,
logbook, control parameter, etc or system setup
Settings
Contact RELAY contact, 240VAC 0.5A Max.(recommended)
Activate
Hi/Lo. Hi/Hi. Lo/Lo selectable two limited programmable,
ON/OFF
Wash RELAY contact ON 0~99min. 59sec. / OFF 0~999hr 59min.
Protection IP65 (NEMA 4X)
Power Supply 100V~240VAC±10%, 7W max., 50/60Hz
Installation Wall or Pipe or Panel Mounting
Dimensions 96m × 96mm × 132mm (HuWuD)
Cut off Dimensions 93 mm u 93 mm (HuW)
Weight 0.5Kg
Note: The specifications are subject to change without notice.
о
1.2 Product packing
1.2.1 Standard accessories
1. Transmitter
2. 8-201-PFA-10 sensor assembly(Including temperature sensor with PTC1000 and signal cable
for 10 meter)
1.2.2 Optional accessories If necessary, please order the spare parts as follows:
1. Installationplease refer to ch.11: Installation of Sensor Assembly
1.1 Flange Mounting
1. 8-201-PFA-10 sensor assembly
2. 5419059 Teflon gasket
3. 5420048 FEP sealing loop
4. 5329003 G3/4" hexagon nut
1.2 Tee installation
1. 5419061 stell tee(2"-3/4")
2. 5329003 G3/4" hexagon nut
1.3 Immersion installation
1. 5419060 immerison nut fitting(1"-3/4")
2. 7202-DO100 DO protection holder 1M
3. Threading type8-26 waterproof cap or coiling type8-09-6 circle junction box
2. Extension cable max. up to 50M (zero-point calibration, cell constant calibration or product
adjustment are required after installation)
1. 8-09-9 square junction box 9T
2. 5251023 signal cable, please advise specific length as ordered(M)
3. Standard calibration buffers(optional)
Note: Installation can be referred to chap. 11: Installation of Sensor Assembly
1.3 Measurement principle
1. EC-4110-I/EC-4110-ICON utililze conversion of electric field and magnetic field to measure
the conductivity in a sample solution. Sensor assembly is composed of Drive Coil and Receive Coil, whose induction is not directly contact with sample solution, belonging to inductive type. It is also composed of chemical-resistant materials, for example: the 8-201-PFA-10 is made up of PFA that can be applied to acidic, alkaline, salt solutions, effluents, etc. in online monitoring industrial environments.
2. EC-4110-I/EC-4110-ICON apply alternative signals of electric field to activate drive coil to
produce the magnetic field. After that, sample solution(containing ions in different conc.) turns into a conductive circuit, and then the signal can b e received on receive cell, turned into magnitude of current to measure the conductivity.
3. The temperature probes have three-wiring type PTC1000 and PTC100 and two-wiring type
NTC30K as reference. Users can select the spec according to sensor assembl y. ȾCond.@txȿ which presents conductivity is being measured at actual tmeperature Ⱦtxȿ, is regarded as absolute conductivity. Refer to temperature compensation modes (linear, non-linear, or off)
п
and reference temperature, “Cond.@tref” is to present conductivity based on desired ref. temperature.
4. EC-4110-I/EC-4110-ICON both have measurement of salinity and TDS.
5. EC-4110-ICON has built-in conc. conversion table including NaCl, HCl, HNO
3
, NaOH,
H
2SO4
, H3PO4, etc. Users can select the corresponding solution types and conc. range. When the conc. range is not applicable, there twinkles error message as an alarm during measurement. The absolute conductivity “Cond.@tx” can be turned into concentration of th e sample solution, and with compensation of ref. temperature it also can be treated as “Cond.@tref.”
6. EC-4110-ICON has self-defined conversion table between cond. and conc., and there are conc.
units %, ppm, or ppt as reference.
7. Self-defined table( max. up to 9 temperature data, and there includes 9 corresponding cond.
and conc. at each temp.) (9 x 9 matrix) can be input corresponding conductivity and concentration at the same temperature by users. The temp., cond., and conc. can be individually input in ascendant or descendant order. After that, the unit is going to check the data and correct somewhere the order goes wr ong. When users only have data of cond. and conc. at one kind of temp. for conversion, it is necessary to apply linear temperature compensation to input T.C. value.
8. Due to conductivity attained by electromagnetic induction not by contact type, cell constant is
to perform the geometric structure characteristic of a drive coil and a receive coil, and the signal cable of the sensor is labeled with cell constant made by lab when out of factory. The inductive signals can be influenced by the surroundings environments such as installation pipeline wall effect, metal(conductive) pipeline and plastic(insulation, non-conductive) pipeline, the distance from sensor assembly and the shell of pipeline. After installed, zero-point calibration in air and cell constant calibration are required, and there is no need to change the labeled cell constant since it can use span calibration to make product adjustment in the field directly.
9. It should restart zero-point calibration in air after each installation or maintenance, and the
surface of sensor assembly needs cleaning without any attachment, making it dry, and using cell constant or product adjustment for calibration.
10. The default measuring range is 2000mS/cm, and users can select 2000mS, 999.9mS, 99.99mS,
9.99mS, ȝ6, or AUTO, where AUTO means automatic switch unit.
р
2. Assembly and installation
2.1 T ransmitter installation:
This transmitter can be installed through panel mounting, wall mounting and 2” pipe mounting.
Installation of panel mounting:
Firstĭ prepare a square hole of 93 x 93mm on the panel box, and then insert the controller directly into the panel box. Insert the accessorial mounting bracket from the rear, and make it be fixed into pickup groove.
2.2 Illustration of panel mounting
93 mm
93 mm
Hole size
87 mm
87 mm
Hole distances on the panel box
Mounting bracket
Illustration of panel mounting, fixed with mounting bracketġ
112 mm
с
2.3 Illustration of wall mounting and pipe mounting
Installation of wall mounting Fixed with 4 x M4 screws
Sun Shield (Pipe mounting, Optional) ( Order No.: 8-35 + 8-35-1 + 8-35-3)
Sun Shield (Wall mounting, Optional) (Order No.: 8-35 + 8-35-2 + 8-35-3)
Insert the single hole rubber plug into the unused cable gland, and tighten up the cable gland to prevent from the penetration of water vapor.
т
3. Overview of Conductivity transmitter EC-4110-I/EC-4110-ICON
3.1 Illustration of rear panel
3.2 Illustration of terminal function
WASH/ClnREL2REL1
POWER
AC
INPUT
POWER
4/20m
A
RS-485
D+(B)
G
D-(A)
Drv Hi
Drv Lo
Drv S
h
Rev Hi Rev Lo Rev S
h
RTD A
RTD B
RTD B S
h
кй
3.3 Description of terminal function
炿烀
REL1: First alarm control, the contact for an
external rela
y
炿烁 炿烂
REL2: Second alarm control, the contact for an
external rela
y
炿烃 炿烄
WASH: Wash relay contact for an external relay
炿烅 炿烆
NC: None contact
炿烇
Power(NĪĻ100~240VAC Power supply terminal
炿烈
NC: None contact
烀炿
Power(L
ĪġĻ
100~240VAC Power supply terminal
烀烀 烀烁 烀烂
Drv Sh: Shield
烀烃
Drv Hi: Sensor receive coil terminal, High
烀烄
Drv Lo: Sensor receive coil terminal, Low
烀烅
Drv Sh: Shield
烀烆
RTD A: Three-wiring temp. probe terminal A
烀烇
RTD B: Three-wiring temp. probe terminal B
烀烈
RTD B Sh: Three-wiring temp. probe terminal B Shield
烁炿
4~20mA +terminal: Master measure current output terminal +, for an
external recorder or PLC control
烁烀
4~20mA – terminal: Master measure current output terminal -, for an
external recorder or PLC control
烁烁
D+(B): RS-485 output D+(B)
烁烂
G: RS-485 output GND
烁烃
D-(A): RS-485 output D-(A)
Temp. probe is PTC1000/PTC100(three-wiring diagram contacts 17 RTD A, 18 RTD B, and 19 RTD B Sh.) Temp. probe is NTC30K(two-wiring diagram contacts 17 RTD A and 19 RTD B Sh.) Temp. probe is PTC1000/PTC100(two-wiring diagram shorts with 18 RTD B and 19 RTD B Sh.)
Drv HiĻSensor drive coil terminal, High Drv Lo
Ļ
Sensor drive coil terminal, Low
кк
3.4 Wiring of cable
3.5 Circuit of cable
8-201-PFA-10 sensor assembly, 10M
Terminals Color of heat shrink tube Wire color Wire descriptions
Drv Hi
Red
Red Send H
Drv Lo
Brown Send L
Drv Sh
Clear Send SHIELD
Rev Hi
Black
Orange Receive H
Rev Lo
Black Receive L
Rev Sh
Clear Receive SHIELD
RTD A
Green
Yellow
PTC1000
RTD-A
RTD B
Green
PTC1000
RTD-B
RTD B Sh
Clear
PTC1000
RTD-B
кл
3.6 Illustration of electrical connection
Note: The transmitter built-in miniature relays is necessary to be repaired and replaced by professional technicians.
It is recommended to use an external relay (Power Relay) to activate the external equipments.
100V~240VAC
Relay 2Relay 1 WASH/Cln
100V~240VAC
Transmitter
Surge absorber
External relay
Surge absorber
External relay
Surge absorber
External relay
Cleaning device
Surge absorber
M
Dose feeder
Surge absorber
M
Dose feeder
Surge absorber
M
км
3.7 Wiring diagram of extension cable
When the length of signal cable needs extension(max. up to 50M), 8-09-9 square junction box 9T(there are 4 M4 fixed screws on the rear) and extension signal cable can be used. After installation, it needs re-calibration of zero-point of the sensor and cell constant, or product adjustments in field. Wiring diagram is as follows:
RED
BLACK
GREEN
RED
BLACK
GREEN
(8Ͳ201ͲPFAͲ10)10M
(ECͲ4110ͲI/ECͲ4110ͲICON)
11
12
13
14
15
16
17
18
19
8Ͳ09Ͳ9squarejunctionbox9T
Please inform specific length of extension signal cable as ordered.
кн
4. Configuration
ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ
4.1 Illustration of front panel
4.2 Keypad
In order to prevent inappropriate operation by others, before the parameter setting and calibration, the operation applies multi-keysĭġand coding protection if necessary. Description of
the key functions is in the following:
In the parameter set-up mode, pressing this key allows you exit parameter set-up mode
and back to Measurement mode.
Ļ In the Calibration mode, pressing this key allows you exit Calibration mode and back to
Measurement mode.
1. In the parameter set-up mode and Calibration mode, pressing this key to select
leftward or change to another page.
2. When adjusting value, press this key to increase the value.
1. In the parameter set-up mode and Calibration mode, pressing this key to select
rightward or change to another page.
2. When adjusting value, press this key to decrease the value.
Key for confirmation; pressing this key is essential when modifying data value or
selecting the parameter setting items in the window.
4.3 LED indicators: ACT
Washing operation relay(Clean) and dosing operation relay (Relay 1, Relay 2) indicator
B.L. : Light sensor; in the automatic display backlit mode, the lamp will light or go out as the
change of environmental brightness.
ко
5. Operation
5.1 Measurement mode:
ġġġAfter all electrical connections are finished and tested, connect the instrument to the power supply and turn it on. The transmitter will automatically entering measurement mode with the factory default settings or the last settings from user.
5.2 Set-up menu:
ġġġPlease refer to the set-up instructions in Chapter 7. Press and simultaneously to enter into set-up menu, and press to go press to back to measurement mode.
5.3 Calibration menu:
ġġġPlease refer to the calibration instructions in Chapter 8. Press and simultaneously to enter into calibration menu, and press to go back to measurement mode.
5.4 Shortcuts:
1. In the measurement mode, if selecting MTC for temperature compensation mode, you may press and to adjust MTC temperature value.
2. Under measurement mode, press continuously for 2 seconds to see the Logbook function directly. Press key to back to measurement mode.
3. Under measurement mode, press continuously for 2 seconds to switch the display mode from text mode, trace mode, and real-time chart display mode.
5.5 Default value:
5.5.1 Setting default value:ġġġġ ġġġġġġġġġġġġġġġġġġġġġġġġġġġ
Measurement mode: Conductivity@tref
Range: 2000mS
Temperature compensation: PTC1000 Temperature Coefficient: Lin, 2.00% Relay 1: High point alarm: AUTO, SP1= 1000mS, Hys.=10mS Relay 2: Low point alarm: AUTO, SP2 =100 mS, Hys.= 1.0 mS Wash time: OFF Analog current output (Cond/Res): 4~20 mA, 0~1000mS RS-485: RTU, 19200,EVEN, 1, ID: 1 Date & Time: 2014/1/1 00:00:00 Digital filter: 0 Backlight setting: Off Contrast: 0 Logbook: None Auto back: Auto, 3 minutes Code set-up: Off
5.5.2 Calibration default value:
ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ ġ
кп
Cal Type: No Cal Cal T emperature: None Cell Constant: 2.700
Zero: 0.0ȝS
Auto back: Auto, 3 minutes Code set-up: Off
Note: The factory default of calibration presetting is “No Cal”, and the cell constant setting is “2.700”. It means
that the user has not calibrated the sensor with the transmitter yet. When selecting standard solution or directly adjusting cell constant to finish calibration, the display shows cell constant of the cell. It should re-start a zero-point calibration in air after installation or maintenance, and the surface of sensor needs cleaning without attachment with cleaning solutions during calibration.
кр
6. Measurement display mode
6.1 Text mode
The normal mode is for digit display, the content is as the following illustration. It mainly includes main measurement value and unit, temperature measurement value and unit, temperature compensation mode, and clock display.
The first line is main measurement to correspond to desired parameter, and the second and the third line are auxiliary one to show related values but not necessary parameters. The fourth line is to show temperature basically.
Main measurement unit
Temperature
compensation Mode
(
MTC/ATC
)
Temperature
measurement & unit
Main measurement
Clock display
Temperature
Compensation mode
Third measuring value
Second measuring
value
Second measuring
value unit
Third measuring value
unit
Conc. type and range
кс
6.2 Real-Time Chart mode
Real-time chart mode is for dynamic display of real-time graphics. The duration is about three minutes of the recent changes in measured values of the curve. Users can set the mode to its corresponding Cond./Conc. measuring range (see section 7.4). The smaller the range is set, the higher resolution of the display is. When entering setup or calibration mode and returning to measurement mode, the real-time graphic will be re-updated. When the measured value exceeds a set range of the upper and lower limit, the graphics will be presented in the upper and lower limits dotted line. Real-time chart mode display is shown as below. There are also real-time measurement value, & unit, and temperature value & unit which are displayed in the bottom of the screen. The timeline in real-time graphic is divided into 12 depict, which is describe the range of representatives of each of 1 / 4 minutes (15 seconds).
3 minutes of main measurement record curve
Main measurement
value & unit
Temperature value
& unit
The current
instantaneous value
Upper limit
Lower limit
кт
6.3 Trace mode
The feature of the trace mode is the record duration which can be set by the user (range from three minutes, up to four weeks). The trend graphic records the measurements in the past T time. The trend is recorded by the 60 group structure. Hence, each group of units is recorded in T/60 time interval. The trend line is constructed by all value data which is calculated to the average (Mean Value), maximum (Max Value) and minimum (Min Value) form. When the latest T/60 record shows in the rightmost of the trend graphic, all the previous record will be moved to the left side of the graphic. For example, T is set to 60 hours, then each set of records will be calculated to the average, the maximum, the minimum values after one hour(T/60 = 1), each time interval. Timeline of trends which is divided into 12 depictions showed on the horizontal axis of the display is on behalf of each characterization interval T/12. So, every depiction has 5 (T/60) sets of records. Users can set the corresponding Cond./Conc. measuring range in its set-up menu(see section 7.4). The smaller the range is set, the higher resolution of the display is. The trace mode is shown as below. There are also real-time measurement value, & unit, and temperature value & unit which are displayed in the bottom of the screen. Attention: When the time interval has been reset, the trend in the data will not be retained, it will start a new trace record. Note: The time display format (XX: XX) (hr: min), for example, appear as four weeks (672:00).
T/12
Trend recording interval
Whole Time(60 data record sets)
Main measurement
value & unit
Temperature value & unit
Upper limit
Lower limit
The max. value of
the T/60 interval
The min. value of
the T/60 interval
The mean value
of the T/60
The latest set of record
shown on the right sector
Record
duration(T), the
negative sign is to
mean the record
лй
6.4 Warning symbols and text
1. When the clean function is activated, the display shows and twinkles the description, “Clean Running”. At the same time, the ACT indicator LED lights up, and the transmitter automatically turns off Relay 1 and Relay 2 function. After finishing cleaning, the Relay 1 and Relay 2 will automatically back to normal status.
2. When Relay 1/Relay 2 which is set in high setting point is in action, the display shows and twinkles the description, “REL 1-HI/REL 2-HI”, and ACT indicator LED lights up. When Relay 1/Relay 2 which is set in low setting point is in action, the display shows and twinkles the description, “REL 1-Lo/ REL 2-Lo”, and ACT indicator LED lights up.
3. When the Analog 1 current output exceeds the upper/lower limitation, the display twinkles ”
S-mA /
S-mA ” as an alarm.
Note: The “HOLD” warning text appears when clean function is activated, or when entering setup menu, or
when entering calibration menu. Under HOLD status, the corresponding display and output are as follows:
1. Both Relay 1 and Relay 2 cease from action. If enter setting menu or calibration menu under clean status, the instrument will stop clean status automatically.
2. The current output which is corresponding to measurement value remains at the last output value before HOLD status.
3. The last signal output value of RS-485 interface is kept at the last output value before HOLD status.
ˑ
Control function
on hold
Product adjust ON
Compensation unit
mS/cm炯tref
conductivity with
temperature comp.
mS/cm炯tx
absolute conductivity
: measurement mode : set-up mode : calibration mode
Relay 1 high or
low point alarm
Relay 2 high or low point alarm
Analog 1 output current
over range alarm
W
ash device in action
condition
лк
7. Settings
Block diagram of setting part 1
EC-4110-I
EC-4110-ICON
Overview
Mode Code Temperature
Input
Code
OFF ON
New
Code
MTC NTC PTC1000
Set
Temp.
Correct
Temp.
English
Traditi
onal
chinese
Simplif
ied
chinese
Return
Continue on next page
Language
Back to the previous
level / action
Product Adj.
Correct
SPAN
ON
Compensation
Linear
Comp.
Non-
Linear
Comp.
Input
linear
Comp.
Select
Product Adj.
ON or OFF
Set time
duration
Delete
all
YES
Select
Cond.@tref/Conc./TDS/
Salinity/Cond.@tx
Modes
Digital
Mode
Chart
Mode
Trace
Mode
Continue on next page
Relay 1
Cond.@tref
Cond.@tx
Mode
Conc.
Mode
TDS
Mode
Salinity
Mode
Select
Range
Set
Max
value
Set
Min.
value
Set
Max.
value
Set
Min.
Value
Input TC
Select
Solution
Select
Range
婧㔜
FACTOR
PTC100
No
Comp.
Input
TC
Value
Sub-display
ON or OFF
Sub-display
ON or OFF
Sub-display
ON or OFF
Sub-display
ON or OFF
лл
Loading...
+ 61 hidden pages