8.2 Calibration principle: You may decide if you need to make a zero-point(offset: “oS” in
the display) calibration. If not, you may directly make a slope calibration (“SLP “in the
display.
OZONE (O3) Residual chlorine(Cl))
Zero-point
calibration
1. Not necessary (zero point accuracy of
the sensor< ±0.05ppm)
2. If in need, use distilled water or pure
water which contains no ozone to
clean it, and then drain out the water,
make zero point calibration with the
sensor in air.
1. Not necessary (zero point accuracy of
the sensor< ±0.05ppm)
2. If in need, use distilled water or pure
water which contains no chlorine to
clean it, and then drain out the water,
make zero point calibration with the
sensor in air.
Slope
calibration
Use spectrophotometer with ozone
reagent to make a measurement value
bench mark, and compare it with the
online measurement.
Use residual chlorine reagent (DPD-1)
to make a measurement value bench
mark, and compare it with the online
measurement.
8.3 Current type sensor calibration mode (ex: Ozone sensor)
1. Press to enter into calibration mode. According to 同時按
鍵,即可進入校正程式,利用按 或 鍵可視需要
選擇零點 “OFFSET 校正(顯示 oS)"後按 鍵進行步驟 3 或選擇斜率“Slope
校正(顯示 SLP)" 後按 鍵進行步驟 4。
2. 在校正時可隨時按 鍵離開校正程式,回到測量狀態。
3. 按 或 鍵輸入使顯示值與水樣臭氧值相同
(如右圖1.00ppm),接著
按 鍵,斜率校正
完畢,並
自動跳出校正程式,回
到測量狀態。
2. Press at any time under calibration mode to go back to
measurement mode.
1. Press to enter into calibration mode. According to
your need, press or to select zero-point (offset)
(displayed as “ oS ”), press to proceed step 3, or select
“slope calibration”(displayed as “SLP”), and then press to
proceed step 4.
+
3. When selecting zero-point (offset) calibration, please make
zero-point calibration in air after cleaning the sensor with
distilled or pure water which do not contain with chlorine. Wait
for display value becoming stable, press or to
adjust the value until it becomes 0.00. Then, press .
After zero-point calibration, the procedure will automatically
lead to slope calibration (displayed as “SLP”). If you do not
need to make a slope calibration, press to leave the
calibration mode and back to measurement mode.
4. When selecting slope calibration, put sensor into flow-through
chamber, and make the sample go into the chamber at least 15
minutes until the display value becomes stable. Leverage
spectrophotometer measurement as a bench mark value to
compare and adjust with the online value.