Sundyne Corporation warrants to Buyer for a period of twelve (12) months from the date of being placed in service (but not to exceed eighteen
(18) months after the date of shipment) that the equipment at the time of shipment will be free from defects of design, material and workmanship.
If any defects or malperformance occur during the warranty period, Sundyne’s sole obligation shall be limited to alteration, repair or replacement
at Sundyne’s expense, F.O.B. Factory, of parts or equipment, which upon return to Sundyne and upon Sundyne’s examination prove to be
defective. Equipment and accessories not manufactured by Sundyne are warranted only to the extent of and by the original manufacturers’
warranty. Sundyne shall not be liable for damage or wear to equipment caused by abnormal conditions, vibration, failure to properly prime or to
operate equipment without flow or caused by corrosives, abrasives or foreign objects. THE FOREGOING WARRANTY IS EXCLUSIVE AND
IN LIEU OF ALL OTHER WARRANTIES, WHETHER EXPRESSED OR IMPLIED INCLUDING ANY WARRANTY OF
MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. In no event shall Sundyne be liable for consequential or incidental
damages.
The following symbols may be found in the text
of this manual.
They have the following meanings:
WARNING: Text accompanied by this
symbol indicates that failure to follow
directions could result in bodily harm or
death.
ELECTRICAL HAZARD: Text
accompanied by this symbol indicates
that failure to follow directions could result
in electrical damage to equipment or
electrical shock.
Equipment and Safety Precautions
Sundyne Corporation manufactures centrifugal
pumps to exacting International Quality
Management System Standards (ISO 9001) as
certified and audited by Lloyd’s Register Quality
Assurance Limited. Genuine parts and
accessories are specifically designed and tested
for use with these products to ensure continued
product quality and performance. Sundyne
cannot test all parts and accessories sourced
from other vendors; incorrect design and/or
fabrication of such parts and accessories may
adversely affect the performance and safety
features of these products. Failure to properly
select, install or use authorized Sundyne pump
parts and accessories is considered misuse and
damage or failure caused by misuse is not
covered by Sundyne’s warranty. Additionally,
modification of Sundyne products or removal of
original components may impair the safety of
these products and their effective operation.
RECOMMENDED: Text accompanied by
this symbol indicates recommended
usage.
REMINDER: Text accompanied by this
symbol indicates a reminder to perform
an action.
EQUIPMENT USE ALERT: Text
accompanied by this symbol indicates
that failure to follow directions could result
in damage to equipment.
CAUTION
Sundyne pumps may handle hazardous,
flammable, and/or toxic fluids. Proper personal
protective equipment should be worn.
Precautions must be taken to prevent physical
injury. Pumpage must be handled and disposed
of in accordance with applicable environmental
regulations.
Note: Safety procedures must be applied
prior to any installation, maintenance,
or repair of a Sundyne pump. Failure to
follow safety precautions may lead to
injury!
1
Instruction and Operation Manual
Wearing Personal Protective
Equipment
To ensure safety, protective equipment must be
worn at all times when installing, performing
maintenance, or repairing equipment. The
following safety recommendations must be
adhered to for optimum safety:
• Safety glasses, with the minimum
requirement of side shields, must be worn
at all times.
• Steel-toed shoes must be worn when
lifting equipment greater than 15 pounds
(7 kg) or if pallet jacks or forklifts are
operated.
• Hearing protection is strongly
recommended at all times when noise
levels exceed 85 dB during an eight (8.0)
hour period.
Note: Chemical resistant gloves must be used
if chemicals are utilized (refer to Using
Chemicals for additional information).
Note: A dust mask respirator must be worn if
chemicals have warning labels
regarding fumes, dust, or mists.
When using more than one piece of protective
equipment, consider their compatibility. For
example, safety glasses will not interfere with
hearing protection equipment. Be sure to clean
all pieces of personal protective equipment
immediately after each use.
Once all work has been completed, the person
installing the lock or tag must remove it
according to company procedure.
Testing Equipment
Prior to performing a test on newly installed,
maintained, or repaired equipment; all personnel
in the immediate area must be warned.
Note: Follow company procedures prior to
equipment testing at all times.
Using Chemicals
Any chemicals to be used must be accompanied
by a relevant material safety data sheet (MSDS),
in accordance with government legislation. If
applicable, use chemical proof gloves.
Note: An eye wash station (or equivalent)
should be available in the event of
injury. If any hazardous or flammable
chemicals pass through the equipment,
a complete decontamination of the
equipment is required.
Protection from Falling
Fall protection and associated preventative
measures are required when working on
equipment located six feet or higher from the
ground.
Note: Follow company fall prevention
procedures prior to working on
equipment.
Using Forklifts
Any persons operating a forklift must have an
active recognized operator license.
Note: Before initializing forklift operation,
verify that the lift is in a safe operating
position.
Ensuring Electrical Safety
All electrical sources must be powered-off
before installation, service, or repair of
equipment occurs.
Note: Sundyne recommends that a Lock-
out/Tag-out program be followed prior
to altering the equipment. Locks or
2
tags must be provided to warn
employees that equipment is
temporarily unavailable.
Preventative Machine Guards
Preventative guards must remain in place on all
equipment.
Note: Only remove the guards while
performing maintenance or repair.
Replace the guards immediately after working
on the equipment and prior to start up.
EXPLOSION/FIRE HAZARD
Note: Never use an acetylene torch, open
flame, or heat to attempt to remove
parts that have seized together in
Sundyne equipment. Any residual
process gas or liquid that is flammable
can result in an explosion or fire with
potential for serious injury or death.
Instruction and Operation Manual
TRANSPORT AND STORAGE
Inspection
Immediately inspect your Sundyne product upon
receipt of the equipment. Check for any
damage, which may have occurred during
shipment. Notify the carrier and Sundyne
immediately if damage is evident.
Storing Your Pump Short-Term
If your Sundyne pump is not to be installed
immediately, protect it from exposure to
moisture and dust. Do not remove the factory
installed shipping covers for casing flanges and
Storing Your Pump Long-Term
In addition to the precautions in the short-term
section above, additional precautions are
required for long-term storage.
If your Sundyne pump will not be operated for a
period of time exceeding six months from the
date of shipment, long-term storage conditions
must be met to ensure minimum corrosion
damage to components.
Note: Sundyne does not accept liability for
equipment damaged during the storage
period. Sundyne does not guarantee
the quality of equipment during and
after the storage period.
Note: The input shaft on the pump may not
seal ports. Ensure that the shipping covers be
kept securely in place.
Note: Observe the storage instructions
Factors, which affect the quality of a Sundyne
pump, when stored, are:
Long-term storage methods must prevent
damaging conditions from making contact with
the internal components of the equipment. When
the equipment is stored in strong chemical
environments or near salt water, protection must
occur immediately upon receipt of the
equipment.
turn freely due to seal drag and speed
increasing gear meshes. If the input
shaft does turn freely, and if rotation is
“not smooth,” damage may have
occurred during shipping.
provided by the driver manufacturer.
• Humidity
• Temperature
• Surrounding chemicals
To ensure the original quality of the Sundyne
pump after storage, all components must be
inspected by an authorized Sundyne service
engineer. Components that are not
manufactured by Sundyne (except mechanical
seals) must be inspected by its own
manufacturer.
Note: Any inspection fees are the sole
responsibility of the purchaser.
3
Recommended Long-Term Storage
Procedures
Sundyne recommends that you do the following
to prevent damage to your pump during longterm storage:
1. Store your pump only in an indoor, climatecontrolled building. These conditions will
maintain constant temperature and humidity.
2. Perform inert gas purging of component
internals.
Instruction and Operation Manual
3. Ensure oil flooding of gearbox internals.
4. Use desiccant bags.
PRODUCT DESCRIPTION
Sundyne Centrifugal Pumps
Sundyne pumps provide high-energy
performance and competitive efficiencies in an
industrial quality, compact unit that is simple to
maintain. Sundyne pumps are single stage that
utilizes an integral gearbox. Designed to
increase the pressure of a continuous flow of
fluid by applying centrifugal action, Sundyne
pumps are most commonly used in HPI, CPI,
and Boiler Feed applications. Commonly applied
in refineries, petrochemical plants, and power
generation plants, Sundyne pumps are used in
high-head, low-to-medium flow processes. This
Note: Because long-term storage of
equipment is of a highly critical nature,
it is recommended that Sundyne be
contacted to provide more details on
the above procedures.
manual presents installation, servicing,
troubleshooting, maintenance and spare parts
information for the latest configuration of
Sundyne centrifugal pumps.
Note: Parenthetical numbers included in the
text correspond to item numbers on the
illustrated figures. The correct spare
part can be ordered for any generation
pump by referencing the item and serial
numbers.
4
Instruction and Operation Manual
INSTALLATION
Suction and Discharge Piping
Please adhere to the following best practices for
installing and maintaining suction and discharge
piping:
1. Install a suction (35-40 mesh) strainer and
clean the suction line prior to starting the
pump. This procedure will protect the
impeller from damage by mill scale, welding
slag, or other foreign particles during initial
startup.
Note: Sundyne Recommends installation of a
differential pressure instrument across
strainer to indicate strainer condition.
2. When installing piping to the pump, ensure
that all piping is supported independently
from the pump.
3. All piping must always line up with the pump
flanges.
Note: Never use force to position piping into
place at the flanged suction and
discharge connection locations. Failure
to have piping properly aligned may
impose excessive strains on the unit.
4. Sundyne recommends using a straight pipe
assembly of at least three times the length
of the pipe diameter.
Note: Carefully select the size of pipe and
fittings to be installed so that friction
losses will remain low.
5. Never use a suction pipe that is smaller in
diameter than the pump suction inlet.
6. Sundyne recommends installation of a
discharge check valve to prevent reverse
rotation.
7. Use block valves (both suction and
discharge) when isolating the pump during
shutdown. This practice will minimize
process leakage and prevent possible
reverse rotation from pump back-flow.
8. It is recommended that suction and
discharge pressure gauges be installed on
any pump that is not flow controlled. If no
flow measuring device is installed there is no
way to determine accurately where on its
curve the pump is operating.
Diffuser Cavity Vent
The diffuser cavity vent must be open to
atmosphere or to a safety drain with no
backpressure. See figure “Wet End
Components” for location.
5
Note: Process fluid may accumulate causing a
potentially hazardous situation if the
diffuser cavity is not properly vented.
Instruction and Operation Manual
Seal Environmental Control System
A seal environmental control system may be
required depending upon the pump seal
arrangement and application.
Always maintain the pump seal environment as
detailed on the specification sheet that
accompanies each unit.
Note: For most applications, a standard
control system can be obtained from
the factory.
Note: Diffuser Cavity Vent must be vented to
atmospheric drain.
Ensure that the specified seal environmental
control system is properly installed and that the
ports are open (or plugged) as indicated in
Figure 1.
Note: Port 1 must always be open or routed
to flare so that it is free to drain.
Figure 1. Seal Housing Port Identification
Note: Port 5 is the highest point in a vertically mounted pump and should be used for start up venting
only.
Liquid Buffer System
For double liquid seals and tandem liquid seals,
A liquid buffer system is used. Introduce the
buffer liquid into port 7, which will flow through
the seal cavity, and out from port 2.
Buffer flow should be 0.5 to 3 gpm (2 to 12
liters/min) with an inlet temperature of 60
o
F (16o to 49oC), and inlet pressure as
120
indicated on the pump specification sheet. The
liquid must be clean to 5 microns.
o
to
Mounting Vertical Units Without Stands
For all vertical units without stands, a mounting
base is recommended. The pump should be
mounted on a rigid foundation, secured in
position by one-inch diameter bolts. The bolts
should be installed in the foundation as shown
on the installation drawing. The length of the
bolts should be sufficient to extend at least ½inch above the nut.
6
Instruction and Operation Manual
Mounting Vertical Units with Stands (LMV)
Grouting of the base plate is required for all
vertical stand units. The top of the stand (driver
mounting surface) should be leveled by
shimming under the base prior to grouting the
Base Mounted (BMP) Units
Grouting of the base plate is required on all BMP
units. The base plate should be leveled prior to
grouting. After grout has been applied, it must
Driver and Coupling
Drivers are normally shipped separately from the
gearbox and pump. When a splined
interconnecting shaft is supplied, this shaft must
be lubricated at each end with one tube (5cc) of
anti-fretting compound (Sundyne Part Number
MP01AA10).
channels that are to be filled with grout through
the access holes. The nuts on the foundation
bolts should not be tightened until the grout has
set for at least 48 hours.
be allowed to set for at least 48 hours before
tightening foundation bolts.
Also available are solid shaft drivers coupled to
the gearbox with a flexible coupling. Drivers are
to be installed and maintained in accordance
with the manufacturer’s instructions.
Flexible Coupling for LMV Units Without a Vertical Stand
Note: Lock out the driver starting switch
before working on the coupling.
When installing flexible couplings, use those
supplied by Sundyne to ensure tolerance of
parallel and angular misalignment, and axial end
float. Use flexible disc couplings or gear type
couplings if not using those supplied by
Sundyne. Coupling installation for turbine drivers
is identical to that for motors.
The gearbox coupling hub is normally mounted
at the factory. The driver coupling hub is
mounted on all motors and turbines shipped
directly from Sundyne.
Driver Coupling is Not Mounted
If your product is received without the driver
coupling hub mounted, use the following
procedure when installing Falk or Thomas
couplings:
1. Measure the distance from the top surface
of the gearbox hub, to the datum face of the
driver adapter. This measurement is referred
to as dimension ”X”.
Figure 2. Dimension X
2. Determine the end gap (the distance between
each coupling hub) for the size of coupling
provided. Refer to the Coupling
Specifications tables in the Specifications
section of this manual for specific
measurements.
3. Subtract the end gap value from dimension X
to determine the distance from the driver
datum face to the coupling hub face This
value is referred to as dimension “Y.”
7
Instruction and Operation Manual
Figure 3. Dimension Y
4. Scribe the shaft to show dimension Y.
5. Ensure that the coupling hub bore, keyways,
and shaft are clean and free from burrs. Also
determine that the key fits in the keyways.
6. Heat the hub in an oil bath or oven to
approximately 250
o
F (121oC), or more if
necessary, so that the hub will slide onto the
motor shaft.
7. Position the hub at the scribed line on the
shaft.
8. Tighten the hub key set screw.
Note: Before the hub is installed onto the
flexible disk couplings, verify that the
coupling bolts and washers can be
assembled (Figure 4) from the motor
side of the hub when installed. If these
pieces do not assemble, insert short
bolts with bevel washers into the hub
flange before fitting them onto the shaft.
Figure 4. Assembly of Coupling Bolts and
Washers
Lube System
The internal lube oil system engineered for
Sundyne pumps consist of four major
components. They are:
• Gearbox sump
• Main lube pump
• Oil heat exchanger
• Oil filter
Gearbox Heat Exchanger
The standard heat exchanger is a shell and tube
water-cooled type. For optimum performance,
the following conditions must be met.
•Cool water must be provided to the
tube side at a maximum pressure of
150 psig (11 kg/cm
•Coolant flow must be controlled to
maintain a gearbox sump temperature
between 140°F and 160°F (60°to
71°C).
2
)(103.5Kpag).
The lube pump intakes oil from the sump and
passes it internally to an externally mounted
manifold. The oil is then passed through the
heat exchanger, the filter, and back into the
gearbox. Once the oil is passed through the
bearings, it then drains back into the sump.
•Maximum recommended temperature
is 180°F (82°C).
The optional air-cooled heat exchanger should
be controlled to maintain the same gearbox
sump temperatures as above.
Mount the heat exchanger lower than the oil
filter to prevent air pockets in the lube oil lines at
start up. Air pockets can cause oil starvation at
the bearings.
8
Instruction and Operation Manual
Note: The heat exchanger installation is a
Sundyne assembly and should not be
rearranged. The heat exchanger is
NEVER mounted higher than the filter.
Oil Manifolds
There are two standard oil manifold
configurations. MA01AA78 for units without heat
exchangers and MA01AA79 for units with heat
exchangers. For units purchased prior to 1991,
the standard manifold is MA01AA01.
Figure 5. Heat Exchanger Manifolds
MA01AA78 (left), MA01AA79 (Center),
MA01AA01 (right).
Figure 6. Heat Exchanger Mounting for
MA01AA01 (units prior to 1991)
The MA01AA01 Model requires a 1/4" sockethead pipe plug in the filter manifold when using
the gearbox heat exchanger. The plug must be
removed when the heat exchanger is not being
used.
Remote Heat Exchanger
All air-cooled heat exchangers as well as some
large water-cooled heat exchangers must be
mounted away from the gearbox.
Note: Interconnecting piping is the
purchaser's responsibility unless the
piping is included with packaged units.
Gearbox Sump
The gearbox sump holds approximately seven
U.S. quarts (6.6 liters) of oil, not including the oil
contained within the auxiliary piping and heat
exchanger. The oil level must always be
maintained as recommended by Sundyne.
Note: Sundyne recommends that the oil level
must be within ¼” of top of the round
sight glass when the machine is static
(refer to the mark labeled “MAX”).When
the pump is in operation,the level will
be approximately ¼” below the MAX
level, with bubbles filling the rest of the
glass. Do not overfill the gearbox.
All connecting piping, including fittings, must not
exceed 20' (6m). The minimum requirement of
all piping is 5/8" (16mm) inner diameter (I.D.)
tubing or piping. If pipe lengths exceed 20' (6m),
then the pipe diameter must be increased
accordingly.
Overfilling will cause overheating/
excessive foaming.
Gearbox Sump Heater (optional)
When gearbox oil temperature falls below –20
o
(-29
C) it becomes too viscous for proper lube
pump operation. A sump heater is required
when these conditions may exist. Two types of
sump heaters are recommended for these
conditions - electric or steam.
o
F
9
Instruction and Operation Manual
SundGard® Oil Filter
The specially designed oil filter is rated for 3
microns at a beta ratio of 200.
®
Note: Oil filters other than Sundgard
filters will void the Sundyne warranty.
-OEM
Main Lube Pump and Lube Oil Priming Kit
Note: Use the auxiliary lube oil priming pump
to circulate oil around the heater when
the main drive motor is not running.
The main lube pump is a positive displacement
gear type pump directly driven by the input shaft.
A pre-lube system is required on some pumps.
The kit consists of:
• Motor driven positive displacement pump
• Check valve
• Gages
• Necessary piping
The gearbox oil and filter should be changed
every six months. Refer to the Lube Oil
Specifications in the Specification section of this
manual for more information.
To start the pump, allow the pre-lube pump to
run for approximately 30 seconds. There should
be a minimum indication of 7 psig (0.5kg/cm
maintained for 30 seconds before starting the
main driver. If the oil piping has been drained,
allow several minutes of operation to bleed
trapped air from the system.
Note: Only allow shut down of pre-lube pump
after main driver is at full operating
speed.
Figure 7. Lube Oil Schematic
2
)
10
Instruction and Operation Manual
Oil Pressure
During normal operation the gearbox internal
lube pump will maintain oil pressure between 15
and 60 psig (1.0 and 4.2kg/cm
2
). This oil
pressure can vary depending on the bearing
configuration and characteristics of the oil being
used.
Note: Never operate the gearbox with oil
pressure less than 10 psig (0.7 kg/cm2)
11
Instruction and Operation Manual
COMMISSIONING, START UP, &
OPERATION
Pre-Commissioning
Familiarizing Yourself with the Pump
Before servicing and starting up the Sundyne
pump, carefully review all information on the
product, including:
• Specification sheets
• Outline drawings
• Performance curves
• Instruction and related manuals
• System P&ID/Process Flow Diagram
(Clients equipment)
• Control system and operational
philosophy/narrative (Client)
Familiarize yourself with the pump configuration
before starting and operating the pump.
Driver Instructions
Carefully follow all installation and starting
instructions provided by the driver manufacturer.
This information is included in the final data
package.
Verifying Auxiliaries
Before start up, verify that the following
auxiliaries are met:
• Check the utility connections
• Verify that the auxiliary piping conforms to
Sundyne standards, as indicated in the
detailed specifications
Installing a Seal Environmental Control
System
Install a system to control the seal environment.
Also, verify that port 1 is properly vented.
If required, install drain piping overhead to
ensure that the environment operates under
normal conditions. For more information,
contact Sundyne Corporation.
Checking Driver Rotation
If the driver is coupled, un-couple; then verify
that the direction of the driver rotates in the
same direction as the arrow stamped or cast on
the pump casing.
If the driver is splined, check the direction of the
motor fan.
Piping Connections
Verify that the following bolted or threaded
connections are tight:
• Pump flange bolts
• Seal environment piping and port
connections
• Cooling water connections to heat
exchanger (if applicable)
• Gearbox oil drain plug
• Pump case drain plug
• Verify all switch and instrument
connections
• Verify that all switch and instrument
settings are set to normal operating
standards
• Calibrate all measurement equipment,
such as flow meters, ampere meters, and
pressure meters, etc.
12
Instruction and Operation Manual
Pre-Start Up
Pressurizing the Fluid Loop
Pressurize the double seal buffer loop or
external seal flush, if applicable, prior to
admitting fluid into the pump casing.
Servicing the Gearbox
Fill the gearbox with lube oil up to a quarter inch
(¼) or 6 mm from the top of the oil level sight
glass.
Note: Prior to using lube oil, verify that it
conforms to acceptable lube oil
specification standards. Refer to the
SPECIFICATIONS section in this
manual for more information.
Under normal operation, the lube oil level will
lower about a quarter inch more than when idle.
Additionally bubbles will appear at the top of the
sight glass.
Operating Conditions
Note: Sundyne recommends that gearbox
lube oil be changed at least every six
months.
For requirement information about priming the
lube oil system, refer to the start-up section in
this manual.
Auxiliary Lube Pump
If your pump includes an auxiliary lubrication
pump, unlock the electrical circuit and move it to
the “hand” position. Check for oil leaks and
recheck the oil level.
Setting the Valves
To set the pump to the designated operating
point, start the pump with the suction valve in
the open position while throttling the discharge
valve.
Verifying Operating Conditions
Verify the following parameters against the
specifications on the specification sheet:
• Suction pressure
• Suction temperature
• Discharge pressure
• Total head
• Flow rate
• Power consumption
• Specific gravity
• Viscosity
• Net Positive Suction Head (NPSH)
The status of these conditions will significantly
alter performance of the pump if they are not in
accordance with the specification sheet.
Check with your Sundyne representative if the
operation conditions of your pump must run
under different parameters than indicated by the
specifications on the specification sheet.
Adjusting the Cooling Flow
If your model pump includes an installed heat
exchanger for the gearbox, adjust the cooling
flow to keep the temperature of the gearbox
sump at 140°-160°F (60°-71°C). Maximum
recommended temperature is 180ºF (82ºC).
13
Instruction and Operation Manual
Installation and Start-Up Checklist
Note: Lock out all switch gears, including main driver, auxiliary lubrication system and instrumentation
before working on this equipment.
This checklist is NOT intended to be inclusive. You must read and follow: instruction manuals, outline
drawings, specification sheets and curves for this equipment during installation, commissioning, and
operation. Your total satisfaction is our goal. Please call with any questions or comments. Be sure to have
the unit serial number that is imprinted on the gearbox nameplate, and request “Sundyne Field Service”.
Is all the information underlined above readily available?
Are the following bolted/threaded connections tight?
• Pump flange bolts?
• Seal environment piping and port connections?
• Cooling water connections to heat exchanger(s)(if applicable)?
• Gearbox oil drain plug?
• Pump case drain plug?
There are two types of connections between the motor and gearbox; a splined shaft or a
coupling. For splined connections, the splined shaft must be lubricated with the supplied spline
grease and the two o-rings installed prior to mounting the motor. It is recommended that the input
shaft be rotated by hand prior to mounting the motor. If the unit has a coupling, be sure the
coupling gap is correct and bolting between coupling halves is tight. This instruction manual
contains coupling set-up information. It is not necessary to align the coupling for run-out or
flatness as this is controlled by the rabbet fits on the gearbox and coupling adaptor.
Is a check valve installed in the discharge line?
Is port 1 open to atmosphere or piped to safety drain or flare or vent header? (Back pressure
must not exceed 5 psig). Sundyne recommends the installation of a check valve in the flare piping
to prevent backflow.
Note: A drip leg must be used if the port 1 connection rises from the seal housing.
Is the diffuser cavity vent open? (Plug must be removed.)
Note: Process fluid may accumulate causing a potentially hazardous situation if diffuser cavity is not
properly vented.
Are all other seal system ports identified and connected according to the outline drawings?
Is gearbox filled to within ¼” (6.35mm) of the top of the sight glass with the approved oil and the
breather fitting installed? Oil capacity is 7 quarts (6.6 liters). Is the needle valve on the gearbox
pressure gauge open? Removal of the vent plug below the fill/vent fitting will speed filling.
Has the oil filter, heat exchanger, and related piping been filled with oil (primed)?
Do process conditions, suction pressure, suction temperature, discharge header pressure, and
specific gravity agree with specification sheet information? DO NOT test the pump on water
unless it is designed for water. Check with your representative or Sundyne Corporation if you
must test on a different fluid than shown on the specification sheet.
If you have auxiliary lubrication pump, unlock the electrical circuit and start it in the “hand”
position. Check for oil leaks and recheck the oil level. If the process suction pressure exceeds
460 psig (32.3 kg/cm2g), the auxiliary lube pump should be running prior to and anytime the
suction is pressurized.
14
Instruction and Operation Manual
Prior to starting the unit, have you opened the suction valve fully and discharge throttled to allow
design flow, typically 40-50% open? Check the control valve to be sure it is functional. Inspect the
case drain, ports, and flanges for leaks. Has the pump been vented through port 6? Open both
supply and return valves supplying cooling water to the gearbox heat exchanger. Check suction
pressure to be sure it agrees with the specification sheet.
Unlock the main driver circuit and bump the motor. Rotation is CCW as viewed from the top end
of the motor. Is rotation correct? Once rotation is verified, run motor for 1 second on, 20 seconds
off. Do this several times until gearbox oil pressure gauge shows pressure, then start the main
driver. Oil pressure will be between 15-60 psig (1.1-4.2 kg/cm2g) depending on the type of
bearings in the gearbox. After commissioning, bumping the motor is not required.
If pressure control is being used, throttle the discharge valve immediately after start-up. Does the
discharge pressure agree with the specification sheet? If flow control is being used, adjust the
valve until flow agrees with the design value listed on the specification sheet.
Once the gearbox oil temperature has stabilized, adjust cooling water supply until the oil
temperature is 140-160°F (60-71°C) on units equipped with heat exchangers. Maximum
recommended temperature is 180°F (82°C).
Listen for any unusual noises or pressure fluctuations.
Note: If you have any questions or concerns about these procedures or the information supplied,
please call your representative or Sundyne Corporation.
Start-Up Procedures
Perform the following tasks to start the Sundyne
pump.
1. Run-in of pump: If the pump is to be run
under conditions which are considerably
different from those conditions listed on the
spec sheet (such as a change in specific
gravity, suction pressure, flow rate, etc.) the
factory should be consulted to ensure that
the run-in conditions are compatible with the
pump.
2. Check to ensure that the driver has been
serviced per instructions provided by the
driver manufacturer.
3. Auxiliaries - Check utility connections; verify
that auxiliary piping is per Sundyne
drawings; verify switch and instrument
connections and set points; calibrate flow
instruments and other transmitters.
4. Flushing screens should be installed in all
field assembled piping connections.
5. Check the pump specification sheet and
outline drawings for seal environment
requirements. Be sure seal housing port
piping is properly connected. If double seals
are used, buffer fluid must be pressurized
before suction pressure is applied to the
pump. Port 1 must be open. Maximum
allowable back pressure on Port 1 is 5 psig
(0.35 kg/cm²). Refer to figure 1.
6. Check to ensure the diffuser cavity vent is
vented.
7. Fill the gearbox with oil.
Remove the gearbox fill-vent plug and the
filter-breather cap from the fill opening on
the gearbox. Fill gearbox within ¼ inch
(6.4 mm) from top of oil level sight glass with
lube oil, which conforms to the specification
in Table 6. Where applicable, operate
auxiliary lube pump to fill heat exchanger
and filter. Add oil as necessary through fill
fitting until oil level stabilizes in sight glass.
The gearbox alone requires approximately 7
quarts (6.6 liters). Replace the filter
breather cap on the fill-opening fitting and
replace the fill-vent plug. If an auxiliary lube
pump is not used, remove the plug on top of
the oil filter (item 924g) manifold and fill the
oil filter and the heat exchanger with oil
(Figure 6).
8. Prime the lube oil system
The following actions must occur at the time
of initial pump installation and following
every re-installation after maintenance that
required the draining/removal of the gearbox
lubricating fluid.
Verify that gearbox lube oil pressure will be
achieved by priming the lube oil system and
expelling all of the air that is potentially
15
Instruction and Operation Manual
trapped. Priming can be achieved by either
operating the (optional) auxiliary lube oil
priming pump or by jogging the main driver
connected to the gearbox (oil pressure
should be observed by the second or third
jog - each of 2-3 seconds duration).
Note: Jogging is required for initial installation
or following re-installation after
maintenance and re-filling of gearbox
lubricant. Units remaining idle should
be jogged once a month to prevent the
bearings from brinelling and to prevent
internal rusting.
Jogging is also used to verify proper
direction of rotation for the main driver.
Jogging is a prudent, conservative activity
that can be quite useful to ensure long
service life of the Sundyne high-speed
products by providing a fluid film of lubricant
on the surfaces of bearings and gears.
After priming the lube oil system, check the
oil level in the gearbox sump, and add oil as
necessary.
Note: Never start the pump against a closed
discharge valve. Always check to
ensure that the discharge valve is
partially open
After priming the lube oil system as defined
above, the pump can now be routinely
started without the need of jogging. This
would include switching of main/stand-by
units, start-up of idle reserve units, start-up
of emergency units, etc. Non-operating
units should be started/used every 9-12
months on an alternating basis.
9. If an auxiliary lube system is installed, it
should be used in the following manner.
a) At the initial startup or after changing the
lube oil, run the pump for several
minutes to work any trapped air out of
the piping. Adjust the relief valve on the
auxiliary pump to provide 25 psi (1.76
2
kg/cm
b) The auxiliary system is intended to
provide oil before starting the main
driver. It should run for a minimum of 5
seconds at minimum pressure before
the main driver is started. Pressure
switches and time delays can be used if
automatic start sequences are desired.
c) After the start of the main driver, oil
pressure will be supplied by the main
lube pump inside the gearbox. An
increase in oil pressure should be
observed. Shut down the auxiliary pump
within two minutes of the main driver
start.
10. Adjust the heat exchanger cooling flow to
regulate the gearbox sump temperature
between 140
Approximately one hour may be required to
stabilize the temperature.
Maximum recommended temperature is
180ºF (82ºC).
) oil pressure to the system.
o
and 160oF (60o and 71oC).
Controlling the Pump During Startup
To ensure control of the pump during start up,
follow the start up procedures for your desired
configuration.
Single Operation
Start the pump with the suction valve open while
throttling the discharge valve. This will ensure
that the pump will reach the design flow
operating point.
If the process fluid is near its vapor pressure,
open the supply vessel seal cavity vent so that
the pump can fill with liquid.
16
Parallel Operation
To prevent back-flow, place check valves in the
discharge piping of each pump.
Note: Sundyne recommends installing
separate bypass loops around each
pump for additional operational
flexibility.
1. Start the first unit as described in the Single
Operation instructions.
2. Start the second unit with the bypass valve
set to maintain the flow above minimum flow.
Instruction and Operation Manual
A
y
2
A
3. Open the discharge valve on the second unit
so that the design flow of both units is
maintained.
Note: Do not operate the pumps at their peak
head capability.
Operation of Sundyne Pumps
Under normal operation, several factors must be
taken into consideration to ensure successful
pump operation. Experienced pump operators
will be aware of jeopardizing factors and their
effects.
Suction Conditions
Improper flow of liquid into the impeller is the
most common operational abuse of centrifugal
pumps. Two conditions must exist to prevent
turbulence at the eye of the impeller.
• Proper suction piping, see suction piping
section.
• Liquid reaching the impeller eye must
have enough vapor pressure to prevent
the fluid from flashing to a gas in the
impeller. If this condition occurs, it will
cause cavitation, which can damage the
impeller and inducer. When centrifugal
pumps cavitate the noise sounds like the
pump is “pumping gravel”. In high speed,
single stage pumps, this sound may not
be discernable. Cavitation can be
prevented by maintaining suction pressure
at a high enough level and suction
temperatures low enough to maintain Net
Positive Suction Head (NPSH) available
greater than Net Positive Suction Head
(NPSH
Minimum Flow Conditions
Vibration and noise will occur during operation of
centrifugal pumps if either of two conditions
exist:
• Internal flow separations
• Recirculation at low flow conditions
If the operator is noticing excessive noise or
vibration, operation must be suspended until the
cause is determined and corrected. Continued
use may cause damage to the pump.
Resonance in the discharge line can accentuate
noise, vibration, and damage to the pump,
) required by the pumps.
r
Sundyne recommends that separate flow
controls be used on each pump to provide a
lower minimum flow range than is achieved by
pressure control.
primarily when a control valve is located an
excessive distance downstream from the pump.
Entrained Gases
The head and capacity of centrifugal pumps will
be reduced by gas that is drawn in with the
liquid. Under normal operating conditions,
centrifugal pumps can tolerate up to 2% of gas
(by volume). Entrained gases can cause
damage to mechanical seals with the exception
of double seals. If you have entrained gas,
contact Sundyne for further instruction.
System Head Curve
The point of intersection between the system
curve and the pump characteristic curve
determines the flow or operation for the
centrifugal pump. For steady flow to occur, the
system curve must intersect the pump
characteristic curve at a significant angle. The
following diagram gives examples of satisfactory
and unsatisfactory angles of intersection.
Figure 8. Typical Operation
System (Head) Cur ve s
B
Good
HEAD
ngle of
Intersection
FLOW
Note: The curve for pump A has a significant
angle of intersection with system curves
D and E. The system curve D could
represent a system with the control
valve wide open while curve E could
represent the same system but with the
throttle valve closed to reduce flow from
flow 1 to flow 2. Pump curve B, on the
other hand, will provide only flow 2,
even with the control valve wide open
Valv e Partiall
E
1
D
Closed
Valve
Wide
Open
17
Instruction and Operation Manual
(curve D). When the control valve is
partially closed to create system curve
E, the curve E and lower pump curve B
are practically parallel. The lack of a
significant angle of intersection means
that the system is unstable, pump flow
is likely to fluctuate eratically and not
respond to control valve position.
Parallel Operation
Maximizing control is critical when operating
centrifugal pumps in parallel. One pump can
overpower the other in regards to head at a
lower total flow. If a simple, unrestricted manifold
connects two pumps at the discharge head, the
discharge head of one pump is imposed on the
other. All pumps will see the same discharge
head at a given time. This is demonstrated on
the following diagrams.
The characteristic curves of two pumps
designated A and B are demonstrated in the
Parallel Operation figure.
Since no two pumps will have exactly the same
performance, it is assumed that pump A
produces a slight amount more head than pump
B. The pumps are arranged with a common
manifold as shown in Parallel Units Common
Valve figure.
Figure 10. Parallel Units Common Valve
The pressure in the manifold is set at P
flow through pump A indicated as A
; the
1
on the
1
preceding curve. At the same time, the flow
through pump B is indicated as B
. However, if
1
the throttle valve is closed to cause the manifold
pressure P to rise to P
pump A and B are A
, then flows through
2
and B2 respectively. If the
2
throttle valve were closed even further, then
pump B would cease to flow entirely. Since
pump B would effectively be deadheaded, the
fluid in it would heat up and boil. During internal
boiling, it could encounter liquid slugging and
probable damage to the pump. Proper selection
of a control system can prevent this situation.
Figure 9. Parallel Operation
FLOW
18
Instruction and Operation Manual
MAINTENANCE
Disassembly of LMV-313
STEP 2
Install the anti-rotation device to prevent
impeller from spinning.
STEP 1
Remove the nuts from the pump casing
studs. Lift the gearbox and seal housing up
off of the pump casing. Remove o-rings 936B
and 936CB and backing ring 70B.
Note: Locking tools, part number TO01AK03
for splined or TO01AK02 for coupled
shafts, are available for purchase from
Sundyne Corporation.
STEP 3
Remove the inducer.
Note: All inducers have a left-handed thread.
When lifting, if the pump case and seal housing
are difficult to separate, use the three jacking
screws to break the seal.
19
Note: When working on a pump with a
tapered inducer, the inducer must be
removed before the diffuser can be
removed from the seal housing.
Instruction and Operation Manual
STEP 4
Remove the inducer housing, shim, and orings 936CA and 936C.
Note: When a tapered inducer is used, the
diffuser is bolted to the seal housing.
When a straight inducer is used, the
diffuser will remain in the pump case.
Removing the bolts from the inducer housing.
STEP 5
STEP 6
Remove the six diffuser mounting bolts.
Build stand shown, part number T-MT-215, is
available for purchase from Sundyne
Corporation.
STEP 7
Install 3/8” eyebolts into the diffuser for
lifting.
Remove the inducer stud and o-ring 936F.
STEP 8
Carefully lift the diffuser off the gearbox
housing.
20
Instruction and Operation Manual
STEP 9
Remove o-ring 936A from the diffuser cover.
Note: These instructions reflect a tandem seal
arrangement. If you have a single or
double seal arrangement, refer to the
seal arrangement drawings later in this
manual for differences.
STEP 10
STEP 12
Remove the impeller spacer and o-ring 936G.
STEP 13
Remove the lower seal rotating face and oring 936J.
Remove the impeller and o-ring 936G.
The impeller can be removed using two
screwdrivers to pry up gently on opposite sides.
STEP 11
Remove the diffuser cover and o-ring 936E.
STEP 14
Remove the lower seal retaining bolts and
remove the lower process seal.
21
Instruction and Operation Manual
STEP 15
Remove the lower sleeve and o-ring 936J.
STEP 16
Remove the upper rotating face and o-ring
936J.
STEP 18
Insert eyebolts into the seal housing for
lifting.
STEP 19
Lift off seal housing. Be careful not to
damage the upper process seal on the highspeed shaft. Remove two 936H o-rings.
STEP 17
Remove seal housing bolts.
22
Remove upper seal.
Instruction and Operation Manual
STEP 20
Remove upper sleeve and o-ring 936K.
STEP 21
Remove the gearbox seal retaining bolts.
STEP 23
Remove the thermal barrier gasket.
STEP 24
Remove the gearbox seal rotating face.
STEP 22
Remove the gearbox seal and o-ring 936P.
23
Use extraction tools to remove the rotating face.
STEP 25
If no maintenance rack is available, loosen
the nuts rather than the bolts.
Instruction and Operation Manual
STEP 26
Remove the alignment bolts.
Note: Alignment bolts are close tolerance and
must be removed first or damage may
occur to the aluminum housing. Damage
to the alignment bores can cause
increased vibration and reduce gearbox
life.
STEP 28
Remove the anti-rotation device.
STEP 29
Remove fill and vent cap from gearbox.
STEP 27
Remove the ½” gearbox housing bolts.
STEP 30
Remove the fill and vent fitting.
Use a pipe wrench to remove the fitting.
24
Instruction and Operation Manual
STEP 31
Lift off the upper housing.
Note: If necessary, a screwdriver (use cloth to
protect the aluminum) can be used in the
slots to assist in breaking the seal
between housings.
STEP 33
Remove the input shaft.
STEP 34
Remove the bearing plate. Turn over for high
speed bearing removal.
STEP 32
Tip the idler shaft out of the way by lifting it
up and out of the lower bearing race and tilt
it towards the outside of the gearbox.
25
Instruction and Operation Manual
STEP 35
Remove the upper journal bearing and thrust
washer, if supplied, from the bearing plate.
STEP 36
Remove the idler shaft and high-speed shaft
assembly by lifting straight up.
STEP 38
Remove the lower journal bearing.
STEP 39
Remove the lube pump from the bearing
plate. The longer two of the four bolts are
used to hold the pump in the bearing plate
and the shorter bolts hold the lube pump
together.
STEP 37
Remove the tilt pad bearing.
26
STEP 40
Remove all lube jets and plugs, flush with
solvent, then blow out all the lube passages.
Instruction and Operation Manual
Inspection and Cleaning and Repair
Inspecting All Bearings
Replace bearings if:
•They have been in operation for over
three years
• If rotation is not smooth
• If outside of inside diameters are worn
Note: Only replace bearings with
manufacturer’s approved replacement
bearings. Non-approved bearings may
jeopardize the mechanical integrity of
the gearbox and pump.
Note: Refer to the Specifications section of
the manual for all bearing and shaft
clearances.
High Speed Shaft
Inspect the High Speed Shaft at the thrust
washer and journal bearing contact areas.
Replace the shaft and gear assembly if:
•Outside diameter of shaft is less than
1.4960 inches
•If the shaft has bearing or washer
materials on it’s surface
new washer. If the tilting pads do not fit freely, or
if they show signs of metal pick-up or
overheating, install a new bearing assembly.
Note: The radial “free play” of the high speed
shaft can be as high as 0.011 inch
(0.28mm) due to the clearance in the
bearings. It is not possible to check for
shaft straightness while the gearbox is
assembled. To check straightness, the
shaft must be placed in V-blocks, on it’s
bearing journals, and have runout
measured at the impeller fit (0.0018
inch TIR max).
Gearbox Mechanical Seal
Carefully inspect the seals for abrasive particles,
excessive seal face wear and any binding of the
seal face washer.
Replace or rebuild a faulty mechanical seal.
Seals may be rebuilt by replacing the seal face
washer, wedge rings, o-ring, and springs. A seal
repair kit is available.
Replace or lap the seal rotating face if the wear
track is rough or worn to a depth greater than 2
helium light bands.
A combined total of 0.010 inch (0.25mm)
maximum may be removed from the surfaces of
the pump and gearbox seal rotating faces.
Excess material removal will result in incorrect
seal face loading causing increased seal
leakage.
Remove any high spots on the end surfaces of
the lower shaft sleeve and impeller hub to insure
that the seal rotating face will not be distorted by
clamping force of the impeller bolt.
Reassemble the seal, throttle bushing, if used
seal housing, and impeller using an o-ring repair
kit. All o-rings that were disturbed by
disassembly should be replaced. During
reassembly, carefully check the torque values
listed in Table 7.
• Shows signs of overheating
• Shows wear to a depth greater than
0.001” (0.03mm)
Inspect upper and lower thrust washers or tilting
pad bearing assembly. If metal is smeared into
radial lube grooves of the washer face, install a
27
The impeller may rub on the diffuser cover plate
(15) until o-rings (936D and 936E) are
compressed by tightening hex nuts (914A).
Check the gearbox input shaft for freedom of
rotation after the pump is assembled and all
bolts are tightened per Table 7.
Instruction and Operation Manual
Upper Shaft Sleeve
Ensure that there are no high spots on the end
surfaces of the shaft sleeve or the impeller hub.
High spots will distort the seal rotating face due
to the clamping force of the impeller bolt. Ensure
that shaft sleeve end faces are parallel within
0.0003” (0.0076mm).
Bearing and Shaft Clearances
Figure 11. Bearing and Shaft Clearances
28
Instruction and Operation Manual
Pump Assembly and Clearance Adjustments
General Clearance Adjustment
Information
If disassembly of the wet-end is only for minor
maintenance such as a seal change, the original
shims should still be satisfactory. If any of the
major components listed below have been
replaced, changes to the original shimming will
be necessary.
• Gearbox
• Seal housing
• Diffuser
• Pump case
Impeller Spacer Adjustments
Before proceeding, the gearbox should be fully
assembled. The seal housing (30) and the
thermal barrier gasket (87A) must be installed
also. The diffuser cover (15) and process
mechanical seals as well as all associated orings can be temporarily left off during this
procedure.
Position the exit flow path from the impeller
within the opening of the diffuser annulus by
changing the thickness of the impeller spacer
(158C). Make this adjustment with the thrust of
the output shaft away from the gearbox.
Position the gearbox vertically with the output
shaft up. Install the following components of the
rotating assembly as well as the original impeller
spacer (158C).
• Shaft sleeves
• Rotating faces
• Impeller
There are two methods for setting the impeller to
the diffuser. Both methods require the use of a
shaft loading tool. See Shaft Loading Tool
Installation figure for how to install the shaft
loading tool.
place the parts within tolerances for the position
specified in the Impeller Shim Adjustment figure.
Note: Some diffuser openings are much wider
than the impeller openings. The primary
tolerance is the 0.025 ± .020 inch
(0.635 ± 0.508 mm) setting. The
diffuser opening must not overlap the
impeller opening.
Method 2
If the proper tools for measuring dimensions of
“A” and “B” are not available, use this method.
Measure the dimensions for “C” and “D” and
transfer these dimensions to a template. Do this
by scribing position marks to indicate the diffuser
opening. Place the straight edge of the template
on the seal housing and visually adjust the
impeller opening to the scribe marks. To do this
it is necessary to leave the diffuser cover (15)
off. Otherwise it is necessary to put a notch in
the template. An example of a transfer template
can be seen in the Shaft Loading Tool
Installation figure.
After completing either of the two methods,
remove the shaft loading tool. Assemble the
following parts:
• Process mechanical seals
• O-rings
• Shaft sleeves
• Seal rotating faces
• Diffuser cover
• Impeller spacer
• Impeller
Inducer Housing Shim Adjustments
Reinstall the bolt from the shaft loading tool and
torque the impeller.
Remove the inducer housing (6) and shims
(158F) from the diffuser (13).
Method 1
Use a dial indicator and surface gauge to
measure dimensions “A” and “B” from the
Impeller Shim Adjustment figure. Record the
readings. Use a depth micrometer to measure
dimensions “C” and “D”. Compare “B” to “C” and
“A” to “D”. Select the spacer thickness that will
Place the diffuser (13) over the impeller allowing
it to rest on the seal housing surface. Make sure
that both surfaces are clean. If using a tapered
inducer, install o-ring (936A) and the diffuser
screws (906F). When using a straight inducer,
leave out o-ring (936A) temporarily.
29
Instruction and Operation Manual
Insert the inducer housing (6) into the diffuser
bore without the o-ring (936CA) and shims
(158F) installed. Install the crossover bar from
the shaft loading tool and torque the nut for the
proper preload. Allow firm contact between the
end of the inducer housing and the impeller. The
result should be a gap between the inducer
housing and the diffuser where the shims are
normally installed.
Measure the gap between the inducer housing
flange and the diffuser using feeler gauges.
Take readings from three points around the
flange. Average those three readings. Record
the average reading and add 0.010 inch (0.254
mm) to the gap reading. The result should
indicate the shim thickness to produce the
proper clearance. Confirm this number by
comparing it against the number in the Inducer
Housing Shim Adjustments figure.
The shim (158F) is made with 0.002 or 0.003inch (0.051 or 0.076 mm) thick laminations. Peel
off any layers that are not needed. Remove the
inducer housing (6) and install the shim (158F),
o-ring (936CA), and inducer housing (6) in the
diffuser.
Note: Proper shimming of the inducer housing
is critical. If not enough clearance is
allowed the result will be impeller
rubbing and hardware damage. If too
much clearance is allowed a loss of
performance will occur. A larger
clearance is preferred to hardware
damage.
Remove the shaft loading tool. If using a tapered
inducer, install and torque the inducer. For
proper torque values see the Torque Values
table. If using a straight inducer, remove the
diffuser assembly and install it in the pump case.
Install the inducer and continue with the pump
assembly.
Figure 12. Impeller Shim Adjustment
30
Instruction and Operation Manual
Figure 13. Shaft Loading Tool Installation
Figure 14. Inducer Housing Adjustment
31
Instruction and Operation Manual
Figure 15. Inducer Housing Shim Adjustment
32
Instruction and Operation Manual
Procedure for Checking the High-Speed Shaft Endplay and
Shoulder Height
The endplay of the shaft MUST be measured if
any of the following parts have been replaced:
• High speed shaft assembly
• Thrust washer
• Output housing
• Journal bearing
• Bearing plate
Sundyne recommends checking endplay during
any re-assembly.
To obtain correct shaft endplay shim spacers
must be installed as required.
Note: All parts must be dry and free of oil.
The output shaft assembly (A130) is dynamically
balanced. Under most circumstances it does not
require disassembly. Inspect the journal
bearings and thrust runners for discoloration
from overheating, also check for contact on the
gear teeth.
Shims (158) are used beneath both high speed
bearing assemblies (151A and B). The shims
function is to provide both axial float (end play)
and insure proper shaft extension. Both must be
checked with the output shaft thrust towards the
impeller. Dry bearings are required to make the
adjustments. Any presence of oil under the
thrust surfaces will produce inaccurate readings.
Step 1
Install the bearings (151A & B) without the shims
in the output housing and bearing plate.
Step 2
Place the output housing on a raised surface
with the open side up. Leave room for the input
shaft to project through.
Step 3
Install the output shaft (A130) into the output
housing bearing.
Step 4
Install the bearing plate. Be careful not to
damage the thrust washer (155B). Leave out the
idler shaft assembly (A140), gearbox gasket
(105), and the o-ring (936T) from the gearbox
split line. Install both of the alignment bolts
(909C). Do not install the gearbox input housing.
Clamp the bearing plate to the lower housing
using “C” clamps or short bolts. This will insure a
metal-to-metal fit.
Step 5
Turn the assembly over so the output shaft is
extended upward. Load the output shaft toward
the impeller (against bearing 151A). Measure
the shaft extension dimension. A standard depth
micrometer can be used if the gearbox seal
rotating face (51D) is installed on the shaft.
Determine the amount of shim to be used under
bearing (151A). Shaft shoulder height should be
1.203” ± .005”.
Step 6
Install a dial indicator to measure the shaft axial
float. This can be done inside or outside the
gearbox. Take the total float reading and
subtract the amount of shim to be used under
the bearing (151A), then subtract 0.015 inch
(desired end-float) from the total. The sum is the
required thickness of the shim to be placed
under the other bearing (151B).
Step 7
Shims supplied by Sundyne Corporation usually
come in thicknesses of 0.005” to 0.010”. Select
the shims that best obtain the shaft extension
and end float dimensions.
Note: The axial float should be 0.015 ± .002
inches. It is preferred to have slightly
more end float than slightly less end
float.
Step 8
Install the shims. Check both dimensions.
Repeat the procedure if necessary.
Step 9
After proper shimming has been completed the
gearbox can be reassembled. Do this by
reversing the disassembly procedures.
33
Instruction and Operation Manual
Figure 16. Output Shaft Shim Adjustment
Note: Shims (158) are adjusted to obtain noted shaft end float and noted shaft extension. Shaft
extension must be checked with thrust in noted direction. Measurements are made with dry (no
oil) bearings.
Item No. Part Name QTY
101A
102
151A
151B
155A
158
Gearbox Output Housing
Gearbox Bearing Plate
Standard Capacity Tilt Pad Thrust Bearing/Journal Bearing
Journal Bearing
High Capacity Tilt Pad Thrust Bearing
Shims (As Required)
34
1
1
1
1
1
1
Instruction and Operation Manual
Assembly Clearance Data Sheet
CUSTOMER
SERIAL NUMBER
DATE
ASSEMBLER
Shim above high speed shaft SP01AA04
Item 158
Shim below high speed shaft SP01AA04
Item 158
High speed shaft end play
0.015” ± 0.002”
High speed shaft extension
1.203” ± 0.005”
MINIMUM ALLOWABLE IMPELLER LABYRINTH CLEARANCE
0.010 + 0.005"
` - 0.000"
Impeller shim - SP01AK01
Item 158C
Inducer housing shim - SH06AK06
Item 158D
Diffuser Gap Maximum – 0.002”
Impeller Alignment
35
Instruction and Operation Manual
Figure 17. Service Check Points
36
Instruction and Operation Manual
Specialty Tools
Overhauling the Sundyne LMV 313 pump in the
field requires a couple of specialty tools that are
not generally needed while overhauling other
Sundyne models. These specialty tools are
recommended in order to simplify the overhaul
procedure. These tools also provide more
consistency in the setting of the high speed shaft
loading and clearances. The following section
will provide details on manufacturing these tools
in the field as well as the use of these tools.
Shaft Loading Tool
The shaft loading tool is used to preload the
shaft against the lower thrust bearing with an
approximate 1000 pounds (453.6 kg) end load.
This removes slack from the bearing stack-up. It
also allows for more consistency in adjusting the
impeller and inducer housing shims. See the
following figures for details.
Pumps using the tapered inducer configuration
do not require tool items 6, 7, and 8. This is due
to the diffuser in these machines being bolted to
the seal housing. To apply preload to the shaft,
lay tool detail 1 across the open end of the
inducer housing. (Refer to figure 19.)
It is vital that the mating surfaces of the diffuser
and the seal housing are clean. Any high spots
caused by nicks and dings must be removed so
that errors will not be introduced during
shimming and reassembly.
Figure 18. Shaft Loading Tool
37
Instruction and Operation Manual
Figure 19. Shaft Loading Tool Details
Item Description Qty.
1
1 x 3 rectangle
bar stock or
1 x 3 inch channel
(steel)
2 Same as item 1 2
3
Tension bolt
(loads impeller
and rotating
assembly)
4 Flat washer ¾” 2
5 Nut ¾-10 UNC 1
6
Bolt ¾-10 UNC,
8” long min.
threaded full
length. (Holds
diffuser against
seal housing. Not
required with
vaned diffuser.)
7
Bolt ½-12 UNC,
2” long (not
required with
vaned diffuser).
1
1
2
2
8 Flat Washer ½” 2
38
Instruction and Operation Manual
Impeller Locking Tool
The impeller must be kept from rotating when
attempting to torque the inducer. The inducer
must have 85-90 ft-lbs of torque (116-123 N-m)
applied to it. The method of applying the torque
varies depending on whether the inducer is
straight or tapered.
Straight Inducer
A strap wrench or hose clamp can be used to
hold the impeller in place with a straight inducer.
Fasten the tool around the impellers outside
diameter and block it against rotation while the
inducer is being torqued. The diffuser and
inducer housing are then installed over the
inducer and impeller.
Tapered Inducer
The tapered inducer must be installed only after
the diffuser and inducer housing are in place. Do
not lock the impeller by inserting a rod thru the
diffuser discharge into the impeller. Risk of
damaging the hardware is likely using this
method. The correct method is to lock the input
shaft. See the Input Shaft Locking Tool figure for
details on the suggested tools for both the
splined and coupled shaft configurations. The 90
ft-lb (123 N-m) inducer torque multiplied by the
gearbox input/output will result in the torque at
the input shaft when the inducer is tightened.
This is usually 3 to 4 times the inducer torque
and cannot be held by hand.
Figure 20. Input Shaft Loading Tool
Segment from SH05AA0-8 Spline Shaft
c
Steel Plate 2” wide x ½” Thick
d
½” or ⅝” hole for bolt into “tee” slot or weld pin to plate
e
1½” diameter schedule 40 pipe
f
¼” Square Keystock (weld both ends)
g
(Note: The coupled input shaft is 1½” diameter and uses ⅜” square key.)
39
Instruction and Operation Manual
Falk Steelflex Type Coupling Specifications
Table1. Falk Steelflex Type Coupling Specifications
Size
40T10 0.062 in.
50T10 0.062 in.
60T10 0.062 in.
70T10 0.062 in.
80T10 0.062 in.
Minimum Normal Maximum
(1.57 mm)
(1.57 mm)
(1.57 mm)
(1.57 mm)
(1.57 mm)
Falk Double Gear Type Coupling Specifications
Table 2. Falk Double Gear Type Coupling Specifications
* When using Teflon® o-rings, allow 15 minutes between torquing for the Teflon® to cold flow. Repeat torquing until
there is no change in torque.
Torque Values
44
Instruction and Operation Manual
Reassembling the LMV-313
STEP 1
Install all lube jets.
Install all plugs and install the lube pump in
the bearing plate.
STEP 3
Install lower journal bearing.
Note: Torque to 28-31 in-lbs.
Step 4
Install tilting pad thrust bearing.
STEP 2
Insert shim under lower journal bearing.
45
STEP 5
Install high-speed shaft assembly dry.
Instruction and Operation Manual
STEP 6
Install upper thrust washer onto the upper
journal bearing.
Note: Use a small amount of petroleum jelly to
hold washer in place.
STEP 7
Install shim.
STEP 8
Install bearing plate onto the lower housing
without gasket or o-ring.
STEP 9
Use alignment bolts to align the bearing
plate.
Re-install the upper journal bearing into the
bearing plate.
Note: Torque hex head screws to 35-40 in-lbs.
STEP 10
Clamp bearing plate to the lower gearbox
housing.
Note: “C” clamps can be utilized to perform
this task. Be sure to protect the housing
faces from any clamping damage.
46
Instruction and Operation Manual
STEP 11
Inspect endplay of the high-speed shaft.
Note: Endplay must fall within these
tolerances, (.015” +/- .002”)
Note: For complete instructions for checking
endplay and shaft shoulder height see
the Procedure for Checking HighSpeed Shaft End Play and Shoulder
Height in this manual.
STEP 13
Install idler shaft and high-speed shaft
assembly.
STEP 14
Install the bearing plate, gasket and o-ring.
STEP 12
With rotating face installed, check shaft
shoulder height using a depth micrometer.
Note: Tolerances for shaft shoulder height are
1.203 ± 0.005
Once the correct endplay is confirmed, install
the housing gasket and o-ring in the lower
housing. Pre-lubricate the shafts and bearings
from this point in assembly.
STEP 15
Install the input shaft.
47
Instruction and Operation Manual
STEP 16
Install upper gearbox housing.
STEP 17
Install the two alignment bolts, turning the
nuts but not the bolts. Then install the
remaining housing bolts.
STEP 19
Install the vent cap.
STEP 20
Install the anti-rotation tool.
STEP 18
Install the fill and vent fitting.
STEP 21
Install the gearbox seal rotating face,
gearbox seal, and o-ring 936P.
48
Instruction and Operation Manual
STEP 22
Install upper shaft sleeve and o-ring 936K.
Install o-ring 936J.
Note: If using Teflon or graphoil o-rings, do not
install any of the o-rings until all
measurements have been taken,
STEP 24
Install upper process seal and o-ring 936H
onto seal housing.
Note: Torque hex head screws to
95-102 in-lbs.
STEP 25
Install the seal housing.
Note: Before installing the seal housing insert
the bolts for the seal housing into holes
and temporarily install nuts on them to
hold in place. It is very difficult to insert
the bolts after the seal housing has
been installed onto the gearbox.
STEP 23
Install thermal barrier gasket.
Note: Torque seal housing bolts to 35-40 ft-
lbs.
49
Instruction and Operation Manual
STEP 26
Install upper process seal rotating face.
Note: Torque process seal hex head screws
to 95-102 in-lbs.
STEP 27
Install the lower process seal shaft sleeve.
Install two 936J o-rings, one on the top of the
sleeve and one on the bottom.
STEP 28
Install the lower process seal and 936H oring.
STEP 29
Install lower mating ring, o-ring 936G, and
impeller spacer 158C.
STEP 30
Install impeller o-ring 936G.
Note: Use petroleum jelly to hold the o-ring in
position.
50
Instruction and Operation Manual
STEP 31
Install the impeller and 936F o-ring.
Note: The diffuser cover is not installed for
this step.
STEP 32
Install the shaft-loading tool.
Note: Refer to the Specialty Tools section later
in this manual for additional information
on the shaft loading tool.
Apply anti-seize to the inducer stud.
Fully thread inducer stud into inducer
loading tool.
Note: Torque values are shown on the tool. 25
ft-lb for standard tilt pad thrust and 50 ftlb for high thrust glacier bearings.
Note: Inducer stud is left hand thread.
Thread inducer stud and inducer loading tool
into shaft end.
51
Instruction and Operation Manual
Torque shaft loading tool.
STEP 33
Measure the diffuser opening using a depth
micrometer. Transfer these readings to a
template.
Note: An Assembly Clearance Data Sheet is
provided later in this manual to record
your measurements.
STEP 35
Remove the impeller and shaft loading tool
and install the diffuser cover and o-ring
936E.
STEP 36
Note: Refer to Pump Assembly and Clearance
Adjustments section later in this manual
for additional information.
STEP 34
Use measurement of diffuser to ensure that
the openings will line up with the openings in
the impeller by changing the impeller
shimming as necessary.
Install impeller and o-ring 936A.
STEP 37
Install eyebolts into the diffuser.
52
Instruction and Operation Manual
STEP 38
Install the diffuser onto the seal housing.
Note: Offset bolt pattern.
STEP 39
Install diffuser mounting bolts.
STEP 41
Install the inducer housing, shims, and
o-rings 936C and 936CA.
Note: Ensure that the inducer tool doesn't
rotate while torque is applied on the
loading tool.
STEP 42
Install shaft loading tool to fully extend the
shaft, torque the shaft and then loosen tool.
STEP 40
Install soft solder over the steps of the
impeller.
STEP 43
Remove the shaft loading tool and the
inducer housing. Measure the pieces of
solder to determine if the amount of gap is
within tolerances (0.010" +.005"/-.000").
53
Instruction and Operation Manual
STEP 44
Install the correct shims to maintain the
tolerance in the last step. Then install the
inducer housing and o-rings 936CA and
936C.
Note: Return to step 26 if using Teflon or
graphoil o-rings.
STEP 45
Fully thread inducer stud into inducer and
install inducer into inducer housing.
STEP 47
Install T-hooks for lifting the gearbox.
STEP 48
Lift gearbox up and position over the pump
case. Install o-rings 936B, 936CB, and
backing ring 70B.
Note: Torque inducer stud to
85-90 ft-lbs.
STEP 46
Rotate the gearbox.
STEP 49
Lower the gearbox onto pump casing.
Note: Before operating the pump, ensure that
54
the diffuser cavity vent is open.
Instruction and Operation Manual
TROUBLESHOOTING
Gearbox & Pump Diagnostics
Several system factors may affect the
performance of the pump. These factors are:
• Temperature
• Specific gravity
• Suction pressure
• Driver speed
• Flow rate
• Control characteristics
These factors as well as internal problems must
be considered when analyzing pump system
performance. The following table gives
diagnostic information that can be useful when
analyzing gearbox and pump performance
problems.
Table 8. Gearbox and Pump Diagnostics
Situation/SymptomPossible Cause Investigative/Corrective Action
No flow, no pressure at
start-up.
Pump not completely filled with liquid. Bleed all vapor or air from port 6.
Allow more cool-down time if pumping low
temperature fluid.
Check suction line for air leak if suction
pressure is lower than atmospheric.
NPSH actually lower than NPSH requirement listed on
specification sheet.
Suction line blocked – check suction screen
and valve.
Excessive pressure drop through suction
piping.
Flow restricted by vapor pockets in high points
of suction line.
Suction tank level or pressure too low.
Entrained air or vapor in pumped fluid.
NPSH reduced by presence of more volatile
fluid in process fluid.
Failure of drive component, such as interconnecting
shaft or impeller key, or item missing from assembly.
Reverse direction of rotation. Direction of driver shaft rotation must be as
Insufficient total head.
Flow too high. Check total head and flow rate against
Wrong direction of driver shaft rotation. (It is possible
for the pump to develop greater than 50 percent
design total head in this condition).
NPSH actually lower than NPSH requirement listed on
specification sheet.
Excess clearance at impeller face (applicable to high
flow models, 313/333/343/348.
Flow too low, causing overheating of fluid resulting in
internal boiling and unstable pump operation.
55
Disassemble and inspect.
shown by arrow on pump casing. Note: Impeller
and driver rotate in the same direction.
performance curve.
Direction of driver shaft rotation must be as
shown by arrow on pump casing. Note: Impeller
and driver rotate in the same direction.
Refer to solutions listed under “No flow, no
pressure at start-up”.
Disassemble and check for clearance.
Increase through-flow rate.
Bypass part of pump discharge to supply tank.
Instruction and Operation Manual
Situation/SymptomPossible Cause Investigative/Corrective Action
Insufficient total head
cont.
Diffuser discharge throat partially plugged or
impeller damaged by passage of a solid particle.
Clean these areas of all obstructions and restore
surfaces to a smooth polished finish free of all
corrosion pitting. Edge of diffuser throat must be
sharp.
Driver overloaded.
Corrosion and/or erosion of diffuser throat (may
also be accompanied by corrosion/ erosion of
diffuser and cover surface adjacent to impeller).
Excessive recirculation from discharge to inlet. Check flow through external plumbing.
Process fluid specific gravity or viscosity different
from values shown on specification sheet.
Driver speed too low. Check speed against value listed on
Pressure gauges or flow meters in error Calibrate instrumentation.
Fluid specific gravity or viscosity higher than values
listed on specification sheet.
Electrical failure in electric driver. Check circuit breaker heater size and setting.
If edge of throat is no longer sharp and smooth
or has opened in size, head-rise may be
reduced. Opening of the inlet area of the throat
will result in higher flow rate and horsepower
consumption. Corrosion/erosion of diffuser and
cover surfaces will result in a significant
horsepower increase.
Pump o-ring (936C) damaged or missing.
Integral centrifugal separator orifice worn.
Check actual viscosity and specific gravity at
operating temperature. Viscosity higher than five
centipoise will cause reduced head and flow and
increased power consumption.
specification sheet.
Check actual viscosity and specific gravity
against value listed on specification sheet.
Check voltage and voltage balance between
phases.
Current for each phase should be balanced
within three percent.
Excessive discharge
pressure pulsations.
Mechanical failure in driver, gearbox or pump. Remove driver and check for freedom of rotation,
Corrosion pitting on surface of diffuser cover or
diffuser, adjacent to impeller blades. Head rise is
also reduced by this condition.
Flow rate too low. Increase flow rate through pump. Add bypass to
Insufficient NPSH available. Refer to solution for insufficient NPSH under “No
Defective flow control valve. Check control valve.
correct spacing
of pump and gearbox shaft assemblies.
Remove fluid end and search for any mechanical
failure.
Remove gearbox oil level sight glass and inspect
bottom of sump for wear particles. Bearings are
probably not damaged if no wear particles are
present.
Disassemble pump and inspect. Rough or
pitted surfaces can cause friction losses which
will significantly increase horsepower
consumption. Clean these areas of all
obstruction and restore surfaces to a smooth
polished finish. Check diffuser throat area at
the inlet; erosion or corrosion resulting in
roughness or increased area will increase
horsepower consumption. Note: A larger throat
size than design will allow a higher flow and
horsepower for a given head rise.
suction tank if necessary.
flow, no pressure at startup,” above.
56
Instruction and Operation Manual
Situation/SymptomPossible Cause Investigative/Corrective Action
Change of gearbox oil
from normal color to
milky pink or yellow
Shaft sleeve rubs on
inside diameter of seal.
Excessive gearbox oil
consumption.
High gearbox
temperature.
Gearbox oil contaminated with water or process
fluid.
Gearbox journal bearing failure. Install replacement exchange gearbox or
Low speed shaft seal (115) leakage. Check drain port for leakage. Replace shaft
High speed shaft mechanical seal (60C) leakage. Check upper gearbox housing drain port for
Leakage through heat exchanger into cooling fluid. Pressure test heat exchanger and replace if
High oil level. Shut down the unit and check oil level. Excessive oil foaming.
Inspect that seal housing port 1 and other seal
drains are open for unrestricted seal leakage
flow.
repair gearbox as outlined under
“Maintenance”
seal if required.
leakage. Replace shaft seal if required.
required.
temperature above 140°F, 60°C.
Check coolant flow and/or clean heat
exchanger.
Check oil level and adjust.
57
Instruction and Operation Manual
g
Pump Mechanical Seal Diagnostics
The following table contains diagnostic
information that is applicable to single seal,
Repair procedures for mechanical seals are
listed in this manual under Maintenance
double seal, and tandem seal equipped units.
.
Table 9. Pump Mechanical Seal Diagnostics
Situation/Symptom Possible Cause Investigative/Corrective Action
Sudden increase in seal
leakage.
Severe cavitation or loss of suction causing
vibration and bouncing of seal face.
Seal icing on low temperature pumps or icing
when handling fluids which vaporize at a
temperature of less than +32°F (0°C) at
atmospheric pressure
Solid particles in seal cavity or seal spring area
(seal faces usually have rough scratched
appearance).
Correct pump suction condition causing
cavitation. Bleed vapor from seal cavity and
restart.
Install double seal if loss of suction cannot be
prevented.
Quench with compatible fluid which will not
freeze at pump temperature through seal drain
port 2 or 7 to prevent ice formation on
atmospheric side of seal during start-up and in
running condition.
Use purge of dry nitrogen gas through ports
2 or 7.
Install double or tandem seal if ice is caused by
water in process fluid or supply external seal
flush of compatible fluid which does not contain
water.
Inspect for clogged integral centrifugal separator
orifices. Clean orifices if necessary (plan 31 if so
equipped.)
Supply external clean seal flush or double seal if
particles cannot be removed by separator.
Sudden increase in seal
leakage
cont.
Seal stationary face spring action is rough and
sticky.
Worn or damaged seal. Disassemble seal and rebuild or replace per
Wear pattern on seal rotating faces not uniform. Lightly lap surfaces of shaft sleeve and impeller
Wear pattern on stationary face smooth but not
uniform.
Edges of stationary face chipped and seal face
worn. (Vapor flashing in seal cavity will cause
excessive wear and/or cracking of rotating face.)
Seal rotating face cracked or broken. May be
caused by damage at assembly or thermal shock
caused by seal running dry.
If parts are corroded, replace with parts made
from compatible materials.
If formation of solids causes sticky seal analyze
fluid properties. Use external seal flush or double
seal arrangement.
instructions in maintenance section.
hub which contact rotating seal face to remove
hi
h spots. Install new seal faces.
Lap flat or replace seal.
Install seal cavity bypass to suction tank.
Prevent loss of pump suction.
Supply cool seal flush.
Install double seal.
Prevent loss of pump suction or supply
continuous external seal flush.
Install double seal.
58
Instruction and Operation Manual
Situation/Symptom Possible Cause Investigative/Corrective Action
Chemical attack of seal faces, seal parts or
o-rings.
Investigate fluid properties and determine
suitable materials for replacement.
Excessive radial high speed shaft movement.
Bent high speed shaft or severe out-of-balance.
Damage to mechanical seal secondary seal
(Teflon® wedge or U-cup or elastomer o-ring).
Loose stack-up of high-speed shaft attaching
components.
Check high speed shaft journal bearings and
replace if necessary.
Check if damage exists on impeller and/or
inducer which will indicate that a large particle
went through the pump.
Deposits on the impeller/inducer causing
unbalance.
Check for erosion and/or corrosion attack.
Install seal flush or double seal arrangement.
Check for correct impeller bolt/inducer torque.
Check for cold flow of Teflon® o-rings.
59
Instruction and Operation Manual
DRAWINGS & PARTS LISTING
Mechanical Seals
Table 10. Mechanical Seals
Lower
Process
Seal
60A 60-B 60C Mechanical Seal
61A 61B 61C Retainer, Seal
62A* 62B* 62C* Seal Face Washer
63A* 63B N/A Seal Spring Backup
64A* 64B* 64C* Seal Retaining Ring
65A 65B* 65C* Seal Spring
68A* 68B* N/A Teflon Wedge
69A* 69B* 69C* Secondary may be “U”
*Seal Repair Kits are Available and contain all parts marked with
a single asterisk.
Item 60C to be used on gearbox only.
Item No.
Upper
Process
Seal
Description
Gearbox
Seal
Assembly
Ring
Cup or O-Ring
To maximize seal performance consult the parts list for
correct seal configuration. For additional information,
please contact your area representative or the Sundyne
factory direct.
60
Instruction and Operation Manual
1 ½” gas seal for upper tandem position on light
hydrocarbon service only
Gearbox Seal Only
61
Instruction and Operation Manual
Wet-End Components
Figure 21. Wet-End Components
*Used only with tapered inducer configuration.
cVent must be open to atmosphere or to safety drain with no back pressure.
Item numbers can be cross-referenced to part numbers using parts list.
Note: Ensure that diffuser cavity vent (item 924B in the figure above) is open to atmosphere or to
safety drain with no back pressure. Failure to do so could result in serious injury or death.
62
Instruction and Operation Manual
Table 11. Wet-End Components
Item No. Part Name Qty. Item No. Part Name Qty.
1 Pump Housing 1 158C Impeller Spacer 1
2 Pump Impeller 1 158D Inducer Housing Shim AR
6 Inducer Housing, Alignment 1 905A Screw, Hex Head 4
9 Inducer 1 905G Screw, Hex Head 3
10 Impeller Stud 1 905L Screw, Hex Head 3
13 Diffuser Insert 1 906E Screw, Socket Head 4
14A Pin, Diffuser Alignment 1 906F* Screw, Socket Head 1
14B Pin, Housing Alignment 2 911 Stud, Pump Case 12