Sunbeam 7620 User Manual

-
Instruction Manual
Table of Contents
Identification of Parts . . . . . . .
Notes for Proper Measurements ........... 3
DescriptionofDisplay
Preparation for Measurements ,............................................................ 7
Measurements . . ............ . . . . . . . . . . . . . . . . ...... 9,
Exhaust Velocity
More about Blood Pressure Measurements
Precautions
Specifications . . ........... 21
. . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
....................
Marks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
........
. . . . . . .
. . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
...
. .
.........
......
15,17,19
1
5
1 11
21
13
Identification of Parts
--._--.-..--
. Notes for Proper
1. Attach the arm cuff at the proper position of the arm as high as the level of th heart.
2. Do not. give vibration to the instrument during measurement, or proper measurement will not be achieved.
3. Perform measurement quietly in a relaxed posture.
4.
Do not wind the arm cuff over jacket or sweater sleeve,
or measurement cannot be done. Also, if the upper
arm is squeezed by the shirt sleeve rolled up to the up­per arm, proper measurement results will not be ob­tained.
5. Blood pressure naturally varies from time to time throughout the day.
6. Although such cases are rare, for subjectshaving an extremely weak pulse or irregular pulse, errors may result which prevent proper measurement. If such ab-
normal variations are noticed, consult your physician
or the sales representative -from which you purchased
this instrument.
Measurements
____
7.
This device your physician should you wish to use the device children.
3
is intended for adult use. Please consult
On
Description of
Display
Marks
Display
Measurement in
mark
v
V
Exhaust
A
Insufficient
pressure
#
Replace batteries
PUL
Err
Pulse
error
Exhaust speed
flashing (exhaust
speed error)
Corrective action
Adjust
exhaust speed to
to 5mmHg/s
2
Preparation for Measurements
1. Installing Batteries Remove the battery cover and insert batteries into the battery
compartment as shown, taking extreme care that the polarities and 8 are observed.
2. Attaching the Arm Cuff
Wind the arm cuff around the left upper arm tightly so that the
tom edge of the cuff is positioned one inch above the elbow joint.
@
bot-
3. Postures During Measurements Remain in a seated or reclining posture
during measurements.
Measurements
1. Turn Power ON
0
When power is turned ON, all display marks appear for approxi-
mately 1 second.
l
When the " If the until " 0
2. Attach the arm cuff to the left arm.
3. Pressurize the arm cuff with the rubber bulb.
l
Pressurize the arm cuff to approximately 30 - 40mmHg above
the estimated Systolic blood pressure to be measured.
0 A buzzer sounds when pressure reaches 160,
280mmHg.
4. When pressurization is completed, the automatic exhaust mechanism will gradually reduce arm cuff pressure, and the
-
mark wilt appear on the display indicating
ment is in progress.
@Pressure value appears on the left, and exhaust speed on the
right side of the display.
0 ""
is displayed the unit is ready for measurement.
V
mark appears, release air with the air release button
:
appears.
200
240 and
that
measure-
-7
(gradually
l
being
reduced)
Measurement
Mark
Accurate measurement cannot be taken when exhaust speed
is outside the 1 - 5 mmHg range.
l
Remain still during measurement.
@
When a pulse is detected, the pressure value moves to the
right side of the display. At this time the
9
the same rate as pulse rate, and the buzzer sounds.
*,
mark flashes at
. .
Measurements
@
When the measurement and buzzer sounds (beep), Systolic
pressure is displayed on the left and Diastolic pressure on the
right. 2 or 3 seconds later, pressure value and pulse appear alternate­ly on the display.
Flashing
0 When a measurement is made with insufficient pressure the
" Err "
mark appears. Pressurize to a value 30-40mmHg higher than that previously used and perform the measure­ment again.
Press tbe exhaust push button to completely release the air from within the cuff. Turn Power OFF
Press the power switch to turn power off.
Subsequent Measurements When a
measurement once and turn it on to restart the measurement. Automatic Power-OFF Funciion If this unit is left on after measurement an automatic power-off function turns the power off after approximately 2 min. 40 seconds.
needs to be made again, turn off the power
-_--11
Exhaust
1.
The exhaust velocity is the rate of gradual pressure drop from the
pressure applied manually using the rubber bulb.
2. The exhaust condition is displayed in the right portion of the display from the time of pressurization until the measurement mark starts flashing.
3. This unit is adjusted so that the exhaust velocity is l -5mm Hg/s
at 150 mmHg.
0 Note that if this exhaust velocity is not proper, correct
measurements are not possible.
Velocity
Exhaust Velocity Adjustment
0 Perform this adjustment procedure if the exhaust velocity is not
within the proper range as described above.
1. Apply the arm cuff to the left arm and
pressurize
proximately
suspected systolic pressure. Hold the
area@in
a screwdriver into groove
2. If the exhaust velocity is too fast,
turn to the left and if it is too slow, turn to the right so that the value shown in the right portion of the display is in the range 2
Hg/s.
When the adjustment is com-
pleted, reduce the pressure and re-
check. If the adjustment is not pro-
per after one try, repeat
times.
l Exhaust velocity will differ
slightly depending upon the size of the windlng the cuff.
to the
60mmHg
the figure, and insert
arm
and the method of
pressure ap-
above your
61
5mm-
several
More About Blood Pressure Measurements
How Is A Proper Measurement
Blood pressure varies according to the conditions time of measurement, so be sure to consider the following in order to obtain the most accurate measurements.
1.
In preparation for blood pressure measurement, the subject
should urinate and should
minutes prior to measurement.
2.
Shirts or other
be removed
3.
Exercise, eating and drinking, smoking, etc., prior to results.
4.5.The throughout the day. Measurement should be made regularly at the same time each day. Do not be too impressed by the results of one measurement. Keep a record of blood tell
a story.
6.
When making repeated measurement, the
arm
resulting in wide variations in measure
ment
values. For this reason, repeated
measurements should be made after a
rest period of approximately 5 to 10 minutes or the engorged
7.
In general, the
summer (when it is hot) and high in
winter (when it is cold).
6.
Emotional stresses may tend to cause
blood pressure to rise.
garments
prior to fitting the
measurement
blood
pressure
pressure variations.
becomes heavy with
af ter raising
condition.
blood
remain
which fit tightly on the upper
va r i es constantly
the
pressure is low in
arm
can affect the
Many readings
blood,
a r m
to relieve
Made?
prevailing
relatively stilt for 10 to 15
cuff
arm
at the
should
More About Blood Pressure Measurements
What is Blood Pressure?
The blood pressure will reach its highest levels in the large arteries
near the heart and drop off toward the peripheral areas of the cir-
culatory system.
The blood pressure varies in accordance with the beating of the
heart, and when the heart contracts, squeezing the blood out, the pressure inside the blood vessels is said to be systolic, and when the heart expands, sucking the blood back in, the pressure of the blood inside the blood vessels is said to be diastolic.
Hypertension, which is most common among adults and the old, if
left unattended, can cause many health problems including stroke,
heart attack, etc. It is therefore necessary to control the blood pressure to prevent it from becoming high, by reducing salt, dietary regimens, and by controlling thesubject’s activities. Even people who are born with high blood pressure can prevent the progress of high blood pressure by means of adequate control.
Why Is It a Good Thing to Measure
Blood Pressure at Home?
Having one’s blood pressure measured by a doctor in a hospital or a clinic, and group health checks, tend to stimulate nervousness in the
subject and may even create sufferers of high blood pressure. Also
blood pressures varies in accordance with a variety of conditions, and
so judgement is not possible on the basis of a single measurement.
The blood pressure measured first thing in the morning after get-
ting up, before taking any food, and with the subject still, is known as
the fundamental blood pressure. In practice it is rather difficult to
record the fundamental blood pressure, but to come as near as possi­ble to measuring the blood pressure in an environment that is close to this, it is useful to be able to take the measurement
Further, you could carry out your own blood pressure control with
your own blood pressure monitor at home to take blood pressure
readings on a regular basis.
17
at
home.
__~~_..
Precautions
1. This instrument is constructed of preci­sion components, and should not be sub.
jected to extremes in temperature,
humidity, direct sunlight, shock or dust.
2. Use a dry, soft cloth to clean the ment Do not use the thinner, alcohol, benzene, or wet dusters for the cleaning.
3. Avoid
tightly folding stage for the components.
long
periods, as such treatment may shorten the life of
the cuff or storing the hose in the
instru-
Specifications
Model Type
Display Measurement range
Accuracy
Pressurization
Depressurization
Deflation Power Sources Battery life Operating environment Storage environment Dimensions Weight Origin:
7620
Release Exhaust Value System Manual push button Type AA batteries Approx.
2
2o*c - +
Japan for Hanson
years with
4o(o”c. less than
.55OC, less
+ 10x - +
-
t44Cwx 7Wb 3qH)mrn
APProx 200g Made in pany Division of Sunbeam Carp Highway 45 North Shubuta, MS 39360
(4)
3 mm usage per day
85%
05% RH
Scale
RH
Com-
than
folded
FULL
This product is warranted for one year from the date of purchase
against defects in material and workmanship. If the product fails to operate during this period, return it prepaid to Hanson Scale Service Center 117 Central Industrial Row, Purvis, MS
ment without charge at the manufacturer’s option. This warranty
gives you specific legal rights, and you may have other rights which
vary. from state to state. warrantor.
Hanson Scale Company
Division of Sunbeam Corporation, Purvis,MS
If purchased in Canada, Return prepaid to:
Sunbeam Corporation (Canada)
1040 Islington Avenue
Toronto. Ontario
A Member Company of Allegherry International
Made in Japan.
CUSTOMER
We want to be sure you are completely satisfied with your new
chase. If you have any questions or experience any problems, please
call our toll free help line.
In U.S.A. only: phone:
1:00
p.m. -
In CANADA only: phone 1-800-268-2743 (112-800-268-2743 in B.C. only)
between
through Thursday) or 8:30 a.m.
Time.
We will
problems. It is
CALL BEFORE
ONE YEAR WARRANTY
39475,
for repair or replace.
39475
Limited.
M8Z
4R5
SERVICE HELP LINE
pur-
4:00
8:30
a.m.
give
you prompt assistance in resolving only questions or
1-800-621-8854
p.m. Central Standard Time, Monday through Friday.
12:30
p.m. and
very
important that you follow the instructions then.
R ETURNING
URNING THE PRODUCT FOR REPAIR.
between
1:30
12:00
p.m. (Friday) Eastern Standard
p.m. --
8:00
a.m.- 12 p.m. and
4:30
p.m. (Monday
31
Notes/Notas
Loading...