Summit BKRF1119B, BKRF1159SS, BKRF21SS Instruction Manual

TRACEABLE THERMOMETER/ALARM USER MANUAL
AL-3
Features
Thermometer simultaneously displays minimum, maximum and current temperatures The unit will provide a visual and audio alert when temperature rises above or falls below the
high and low set points
The min/max feature is designed to monitor and store the highest and lowest readings until the
memory is cleared
The temperature sensor is enclosed in a glycol-filled bottle, protecting it from rapid temperature
changes when refrigerator/freezer door is opened
Low battery alert function ( battery symbol flashes) Measuring temperature range -30 ~ 70 oC ( or -22 ~ 158 oF)  Accuracy : ± 0.5 oC (-10 ~ 10 oC or 14 ~ 50 oF), in other ranges ± 1 oC ( or ± 2 oF)  Sampling rate of 10 seconds Powered by a durable AAA lithium battery
SAVE THIS MANUAL FOR FUTURE REFERENCE
R1-041916 Page 1 of 4
Package Contents
Press the MODE/RESET button ONCE to reset Min/Max temperature record.
To display on the main screen:
Press ONCE to display high alarm temperature setting. Press TWICE to display low alarm temperature setting. Press THREE TIMES to display current temperature reading.
Press to change the temperature display scale from Fahrenheit (oF) to Celsius (oC) or Celsius to Fahrenheit
Main display screen
Display unit Temperature sensor in a glycol-filled bottle Instruction manual
Calibration certificate Lithium battery
*Remove the protective film that is applied to the LCD screen prior to using this product.
Parts and Controls/Features
*Press any key to turn off the alarm beeping sound.
Set Up
Battery Installation: Remove battery compartment cover located on the rear of the unit and remove
insulation strip from battery compartment. Following the polarity (+/-) diagram in the battery compartment, install the battery. Replace the battery cover. The unit will beep and all segments of the LCD will be activated.
R1-041916 Page 2 of 4
Loading...
+ 2 hidden pages