Thank you for purchasing this piece of Stuart
equipment. To get the best performance from
the equipment and for your own safety, please
read these instructions carefully before use.
Before discarding the packaging check that all
parts are present and correct.
General Description
The Stuart SCT1 temperature controller is the
ideal instrument for the accurate temperature
control of aqueous and oil based samples in the
laboratory. The SCT1 can be used in two
different modes; as a precise temperature
controller from 20 to 200°C or as a digital
thermometer from -4 to 325°C. With digital
setting and display of temperature, the SCT1
temperature controller is accurate to ±0.5°C.
The SCT1 is compatible with the following Stuart
hotplates and hotplate stirrers: US150, US152,
US152D, UC150, UC152, UC152D, SC162, and
CC162.
Safety Advice before use
If the equipment is not used in the
manner described in this manual and
with accessories other than those
recommended by Stuart the protection
provided might be impaired.
This equipment is designed to operate under the
following conditions: -
❖ For indoor use only
❖ Use in a well ventilated area
❖ Ambient temperature range +5°C to +40°C
❖ Altitude to 2000m
❖ Relative humidity not exceeding 80%
❖ Use with a minimum distance all around of
200mm from walls or other items.
Preparation for use
To use the SCT1 temperature controller with
either a metal top or ceramic top hotplate, place
the hotplate on a firm, level surface.
Open the jaws of the retort clamp on the rear of
the SCT1 temperature controller by turning the
clamp knob anticlockwise. The clamp can be
used with retort rods from 9.5mm to 13mm
diameter. Slide the SCT1 on to the retort rod and
tighten by turning the clamp knob clockwise.
NB: DO NOT OVER TIGHTEN
Connect the SCT1 controller plug to the DIN
probe socket at the rear of the hotplate (Figures
1 and 4).
Connect the probe plug to the probe socket at
the rear of the controller (Figures 2 and 4). To
ensure the best connection, tighten the locking
collar on the plug by turning clockwise.
Connect the mains supply lead to the IEC socket
at the rear of the hotplate and switch on at the
mains. Where the unit has a mains on/off switch,
switch the unit on using this switch.
To switch on the SCT1 temperature controller,
turn the heat control knob on the hotplate
clockwise (Figure 3).
NB: The controller power is supplied by the
hotplate
The SCT1 temperature controller and hotplate are
now ready for use. Correct communication
between the SCT1 temperature controller and
the hotplate is indicated by the illuminated amber
contact thermometer indicator on the front panel
of the hotplate.
NB.: The hotplate and controller will
perform an automatic self-test at switch
on, before the current temperature of the
probe is displayed. The factory default set
temperature is 20°C.
3
Operation
temperature
controller
SCT 1
C
Controller positioning
The SCT1 temperature controller is supplied with
a detachable temperature probe (Figure2). This
useful feature enables the SCT1 temperature
controller to be positioned away from damaging
fumes.
The temperature controller is factory set such that
the retort clamp is positioned for mounting on a
vertical retort rod. To mount the SCT1
temperature controller on to a horizontal retort
rod, ease back the clamp situated at the rear of
the controller sufficiently to clear the right-angled
raised location. Rotate clockwise by 90° until the
clamp snaps back into the location.
Probe positioning
For optimum temperature control, ensure the end
of the sensing probe is immersed at least 20mm
deep into the medium being controlled. The
temperature probe when not in use should be
placed in the probe holder located on the left
hand side of the SCT1 control panel. For safety
and to avoid contamination, always wipe the
probe clean prior to placing into holder.
Using the SCT1 temperature controller
The SCT1 can be used in two different modes,
either as a temperature controller (‘control
mode’) or as a digital thermometer (‘monitor
mode’).
To use the SCT1 in control mode
Ensure that the SCT1 temperature controller is
connected to the hotplate. Also ensure the
temperature probe is attached to the controller
and immersed in the medium to be controlled
before switching on at the mains.
US150, US152, US152D, UC150, UC152 and
UC152D
Turn the unit on using the On/Off switch at the
side of the unit. The SCT1 will power up in
monitor mode. To turn on the temperature
controller in control mode, turn the heat control
clockwise on the front panel of the hotplate. On
the digital models the LED display will show
“---“.
The SCT1 temperature controller in control mode
is indicated by the illuminated amber contact
thermometer indicator on the front panel of the
hotplate (Figure 3). The LED and display will be
visible on the SCT1 temperature controller panel
(Figure 5) and the current temperature of the
medium will be displayed.
Thermometer
LED
Increase
temperature
set point
Decrease
temperature
set point
Figure 5 - SCT1 temperature controller
panel
CC162 and SC162
A green LED will be visible on the front panel of
the hotplate. To turn on the temperature
controller, turn the heat control clockwise on the
front panel of the hotplate. The SCT1
temperature controller in control mode is
indicated by the illuminated amber contact
thermometer indicator on the front panel of the
hotplate (Figure 3).The LED and display will be
visible on the SCT1 temperature controller panel
(Figure 5) and the current temperature of the
medium will be displayed.
NB: Turning the heat control clockwise on
the hotplate turns the SCT1 temperature
controller into control mode and cannot be
used to set the hotplate temperature.
To set the temperature in controller mode
Press the + or - key once to instigate the
temperature set point mode. This is indicated by
a flashing LED on the SCT1 temperature
controller display. To select the temperature, use
the ‘+’ or ‘ –‘ keys to obtain the desired
temperature. To set the temperature, release the
key. The LED will cease flashing once the
temperature has been set. With the probe in the
sample, the thermometer LED will illuminate as
follows:
❖Green – temperature set point not reached
4
❖Amber – liquid at set point
❖Red – liquid hotter than set point
NB: To obtain accurate temperature control,
the minimum temperature set point must be
at least 15°C above the surrounding ambient
temperature.
To use the SCT1 in monitor mode
US150, US152, US152D, UC150, UC152 and
UC152D
To use the SCT1 controller in monitor mode
together with the hotplate, ensure the unit is
switched off using the mains control switch at
the side. Hold down both the ‘+’ and ‘ –‘ keys
simultaneously on the SCT1 temperature
controller and turn the mains switch to ‘ON’.
Once the digital display illuminates, release the
keys on the SCT1 temperature controller. Monitor
mode will be indicated by a red flashing LED at
the bottom right hand corner of the display of
the SCT1 temperature controller. The amber
control LED on the front panel of the hotplate
will NOT be illuminated in monitor mode.
CC162 and SC162
To use the SCT1 controller in monitor mode,
ensure the hotplate is switched off by turning the
heat control on the hotplate anticlockwise to the
‘OFF’ position. Hold down both the ‘+’ and ‘ –‘
keys simultaneously on the SCT1 temperature
controller, and turn the heat control ‘ON’. Once
the digital display illuminates, release the keys on
the SCT1 temperature controller. Monitor mode
will be indicated by a red flashing LED at the
bottom right hand corner of the display of the
SCT1 temperature controller. The amber control
LED on the front panel of the hotplate will NOT
be illuminated in monitor mode.
NB: When in monitor mode, the SCT1
temperature controller operates as a digital
thermometer only, and no longer controls
the temperature of the medium in use. The
hotplate temperature can now be manually
adjusted to the desired setting.
To change from monitor mode to control
mode
To change from monitor mode to controller
mode, press either the ‘+’ or’ –‘ key on the SCT1
temperature controller once. The unit will revert
to controller mod, and will be indicated by the
illuminated amber contact thermometer indicator
visible on the front panel of the hotplate. The red
LED on the bottom right hand corner of the SCT1
temperature controller display will extinguish
when the unit is in controller mode.
NB: For safety reasons, when reverting from
monitor to controller mode the SCT1
temperature controller will default to the set
point temperature of 20°C.
Troubleshooting
In the event of malfunction the SCT1
temperature controller can self-diagnose certain
fault conditions. In this case the display will show
a message in the form of ‘Err XX’ where XX is
either a single or double digit code, which
identifies the fault.
Error
code Fault condition Remedy
Err 1 Thermometer not
calibrated Return to factory
Err 2 Electronic fault Return to factory
Err 4 Probe plug removed Reconnect probe.
in control or monitor Continuing error
mode code indicates
hotplate fault return to factory
Err 8 Probe out of sample Replace probe in
when in control sample
mode
Err 16 Ambient temperature Switch off at
inside hotplate too 'mains' and allow
high (controller to cool
attached)
Servicing and Repair
WARNING: Ensure the hotplate is
disconnected from the mains electricity
supply before attempting maintenance or
servicing.
This equipment does not require routine
servicing. The only maintenance required is to
clean external surfaces with a damp cloth and
5
mild detergent. Spillages or splashes should be
cleaned up straightaway after isolating the
hotplate from the mains electricity supply.
Repairs
In the event of product failure it is recommended
that any repair is only undertaken by suitably
qualified personnel.
Only spare parts supplied by Bibby Scientific
or its agent should be used. Fitting of
non-approved parts may affect the
performance of the safety features of the
instrument.
For a comprehensive list of parts required by
service engineers conducting internal repairs, or
to receive a service manual, please contact the
Service Department of Bibby Scientific Ltd
quoting both the model and serial number.
For any other technical enquiries, please contact
the Technical Support Department:
E-mail: stuarthelp@bibby-scientific.com,
Tel: +44 (0)1785 810433.
Accessories
A range of accessories detailed below is available
for use with the SCT1 temperature controller.
Warranty
Bibby Scientific Ltd. warrants this equipment to
be free from defects in material and
workmanship, when used under normal
laboratory conditions, for a period of three (3)
years. In the event of a justified claim, Bibby
Scientific will replace any defective component or
replace the unit free of charge.
This warranty does NOT apply if damage is
caused by fire, accident, misuse, neglect,
incorrect adjustment or repair, damage caused by
installation, adaptation, modification, fitting of
non-approved parts or repair by unauthorised
personnel.
Temperature range ºC 20 to 200ºC
Accuracy, ºC ±0.5ºC
Resolution 1ºC
Dimensions, mm (w x d x h) 90 x 75 x 123
Net weight, kg 0.3 (inc. probe)
IP rating 54
Part code Description
SCT1/1 Probe holder
SCT1/2 PTFE probe
SR1 Retort rod
SCT1/MP Spare metal probe
6
Français
temperature
controller
SCT 1
C
heat-stir
US152
HOT
1
2
3
4
5
6
7
8
9
STIR
undergrad
off
Power
on/off
100
150
250
200
300°C
50
HEAT
HEAT
STIR
HOT
100
50
Poweron/off
heat-stir
US152
150
200
250
4
3
2
1
off
5
6
7
US152
HOT
1
2
3
4
5
6
7
8
9
STIR
undergrad
off
Poweron/off
100
150
250
200
300°C
50
HEAT
Poweron/off
100
150
250
200
300°C
50
HEAT
Figure 1
Vu de face
Support
de la
sonde
Fiche du contrôleur
Figure 3
Vue de face de la plaque chauffante
Témoin de
thermomètre
de contact
Témoin d'avertissement 'Hot'
Témoin d'agitation
activée
Sonde de
température
Figure 2
Vue arrière
Connecteur
de la sonde
Fiche de
la sonde
Bague de verrouillage
Figure 4
Vue arrière de la plaque chauffante
Pince à
mâchoires
Fixation
de tige
de statif
Contrôle de la
température
Connecteur de sonde DIN
Interrupteur
Contrôle de
l'agitation
marche/arrêt
d'alimentation
(sauf sur unité
d'agitateur
seul)
Douille
d'alimentation IEC
pour fiche
haute
température
5
Contrôleur de
température
SCT1 Manuel d'utilisation
Introduction
Nous vous remercions d'avoir fait l'acquisition de
cet équipement Stuart. Pour profiter au mieux de
ses performances, ainsi que pour votre sécurité,
lisez attentivement ces instructions avant toute
utilisation. Avant de jeter l'emballage, vérifiez
que tous les éléments sont présents et en bon
état.
Description générale
Le contrôleur de température Stuart SCT1 est
l'équipement idéal pour contrôler de manière
précise la température des échantillons aqueux
et à base d'huile, en laboratoire. Il peut s'utiliser
de deux manières différentes : en tant que
contrôleur de température de précision entre 20
et 200 °C ou en tant que thermomètre
numérique entre -4 et 325 °C. Grâce à son
réglage et à son affichage numériques de
température, il permet une précision de l'ordre
de ± 0,5 °C.
Le contrôleur de température SCT1 est
compatible avec les plaques chauffantes et les
agitateurs chauffants Stuart suivants : US150,
US152, US152D, UC150, UC152, UC152D,
SC162 et CC162.
Consignes de sécurité
avant utilisation
En cas de non respect des consignes
détaillées dans ce manuel ou
d'utilisation avec des accessoires
différents de ceux recommandés par Stuart,
la protection intégrée est susceptible de se
révéler inopérante.
Cet appareil est conçu pour fonctionner dans les
conditions suivantes :
❖ Utilisation en intérieur uniquement
❖ Utilisation dans un lieu correctement ventilé
❖ Température ambiante de +5 à +40 ºC
❖ Altitude maximale de 2000 m
❖ Taux d'humidité relative de 80 % maximum
❖ Ménager un espace minimum de 200 mm
autour de l'appareil par rapport aux murs et
aux autres équipements.
❖ Ne jamais déplacer ou transporter l'appareil
en cours d'utilisation ou lorsqu'il est connecté
à la plaque chauffante.
❖ Ne pas utiliser en atmosphère dangereuse ou
en présence de substances dangereuses.
Préparation à l'utilisation
Pour utiliser le contrôleur de température SCT1
avec une plaque chauffante avec surface de pose
en métal ou en céramique, placez la plaque
chauffante sur une surface plane et ferme.
Ouvrez les mâchoires de la pince, à l'arrière du
contrôleur de température SCT1, en faisant
tourner le bouton de la pince dans le sens
antihoraire. La pince peut s'utiliser avec des tiges
de statif d'un diamètre allant de 9,5 à 13 mm.
Faites glisser le SCT1 sur la tige de statif et serrez
la pince en faisant tourner son bouton dans le
sens horaire.
REMARQUE : PAS DE SERRAGE EXCESSIF.
Connectez la fiche du contrôleur SCT1 au
connecteur DIN, à l'arrière de la plaque
chauffante (figures 1 et 4).
Connectez la fiche de la sonde au connecteur
situé à l'arrière du contrôleur (figures 2 et 4).
Pour assurer un branchement optimal, serrez la
bague de verrouillage de la fiche en la faisant
tourner dans le sens horaire.
Branchez le cordon d'alimentation secteur au
connecteur IEC situé à l'arrière de la plaque
chauffante et connectez-le à l'alimentation
secteur. Si l'unité est dotée d'un interrupteur
marche/arrêt, mettez-la sous tension à l'aide de
cet interrupteur.
Pour allumer le contrôleur de température SCT1,
faites tourner le bouton de réglage de
température de la plaque chauffante dans le
sens horaire (figure 3).
6
REMARQUE : le contrôleur est alimenté par
temperature
controller
SCT 1
C
la plaque chauffante.
Le contrôleur de température SCT1 et la plaque
chauffante sont désormais prêts à l'emploi. Le
voyant orange du thermomètre à contact situé
en façade de la plaque chauffante s'allume pour
indiquer que la communication entre le
contrôleur de température SCT1 et la plaque
chauffante est correcte.
REMARQUE : la plaque chauffante et le
contrôleur effectuent un test automatique
au moment de leur mise sous tension, avant
d'afficher la température en cours de la
sonde. Par défaut, la température est réglée
sur 20 °C.
Utilisation
Installation du contrôleur
Le contrôleur de température SCT1 est fourni
avec une sonde de température amovible
(figure2). Cette caractéristique utile permet de le
tenir éloigné des émanations susceptibles de
l'endommager.
Par défaut, la pince à mâchoires du contrôleur
de température est positionnée de manière à
permettre un montage sur une tige de statif
verticale. Pour installer le contrôleur de
température SCT1 sur une tige de statif
horizontale, faites effectuer une rotation de la
pince située à l'arrière du contrôleur de manière
à présenter la partie à angle droit en haut. Faites
tourner la pince de 90° dans le sens horaire
jusqu'à ce qu'elle s'enclenche à nouveau.
Positionnement de la sonde
Pour une mesure optimale de la température,
assurez-vous que l'extrémité de la sonde se
trouve à une profondeur d'au moins 20 mm
dans l'échantillon contrôlé. Lorsqu'elle n'est pas
utilisée, la sonde de température doit être placée
dans le support de sonde situé à gauche du
panneau de commande du SCT1. Pour des
raisons de sécurité et pour éviter toute
contamination, essuyez toujours la sonde avant
de la placer dans son support.
Utilisation du contrôleur de température
SCT1
Le SCT1 peut s'utiliser de deux manières
différentes : en tant que contrôleur de
température (« mode contrôle ») ou en tant que
thermomètre numérique (« mode affichage »).
Utiliser le SCT1 en mode contrôle
Assurez-vous que le contrôleur de température
SCT1 est connecté à la plaque chauffante.
Assurez-vous également que la sonde de
température est connectée au contrôleur et
plongée dans l'échantillon à contrôler avant de
mettre l'équipement sous tension.
US150, US152, US152D, UC150, UC152 et
UC152D
Mettez l'unité sous tension à l'aide de
l'interrupteur marche/arrêt situé sur le côté. Le
SCT1 s'allume en mode affichage. Pour basculer
le contrôleur de température en mode contrôle,
faites tourner le bouton de réglage de la
température situé en façade de la plaque
chauffante dans le sens horaire. Sur les modèles
numériques, l'écran LED affiche le message
« --- ».
L'état actif du mode contrôle du SCT1 est
indiqué par l'allumage du voyant orange du
thermomètre à contact en façade de la plaque
chauffante (figure 3). Le voyant et l'affichage
sont visibles sur le panneau du contrôleur de
température SCT1 (figure 5) et la température de
l'échantillon est affichée.
DEL du
thermomètre
Augmentation
de la consigne
de température
Diminution de
la consigne de
température
Figure 5 : Panneau du contrôleur de
température SCT1
7
CC162 et SC162
Un voyant vert s'allume en façade de la
plaque chauffante. Pour mettre le contrôleur
de température sous tension, faites tourner
le bouton de réglage de température en
façade de la plaque chauffante dans le sens
horaire. Le mode contrôle du SCT1 est
indiqué par l'allumage du voyant orange du
thermomètre à contact en façade de la
plaque chauffante (figure 3). Le voyant et
l'affichage sont visibles sur le panneau du
contrôleur (figure 5) et la température de
l'échantillon est affichée.
REMARQUE : faire tourner le bouton de
réglage de température de la plaque
chauffante dans le sens horaire fait
basculer le contrôleur SCT1 en mode
contrôle, mais ne règle pas la
température de la plaque chauffante.
Pour régler la température en mode
contrôle
Appuyez une fois sur la touche + ou - pour
activer le mode Point de consigne de
température. L'activation est indiquée par un
voyant clignotant sur l'écran du contrôleur
SCT1. Pour sélectionner la température
souhaitée, utilisez la touche + ou -. Pour
valider la température, relâchez la touche. Le
voyant cesse de clignoter une fois la
température définie. Lorsque la sonde est
placée dans le liquide, le voyant du
thermomètre s'allume comme suit :
❖ Vert : point de consigne de température
non atteint
❖ Orange : température du liquide au point
de consigne
❖ Rouge : température du liquide
supérieure au point de consigne
Remarque : pour garantir la précision de
la mesure de la température, le point de
consigne de température minimum doit
être supérieur d'au moins 15 °C à la
température ambiante.
Utiliser le SCT1 en mode affichage
US150, US152, US152D, UC150, UC152 et
UC152D
Pour utiliser le contrôleur SCT1 en mode
affichage avec la plaque chauffante,
assurez-vous que l'unité est hors tension à
l'aide de l'interrupteur marche/arrêt situé sur
le côté. Appuyez simultanément sur les
touches + et - du contrôleur et, tout en les
maintenant enfoncées, réglez placez
l'interrupteur marche/arrêt sur marche
(« ON »). Une fois l'écran numérique allumé,
relâchez les touches du contrôleur de
température SCT1. Le mode affichage est
indiqué par un voyant rouge clignotant dans
l'angle inférieur droit de l'écran du
contrôleur. Le voyant de contrôle orange de
la façade de la plaque chauffante ne
s'allume PAS en mode affichage.
CC162 et SC162
Pour utiliser le contrôleur SCT1 en mode
affichage, assurez-vous que la plaque
chauffante est hors tension en faisant
tourner le bouton de contrôle de
température de la plaque chauffante dans le
sens antihoraire, jusqu'à la position arrêt
(« OFF »). Appuyez simultanément sur les
touches + et - du contrôleur de température
SCT1 et, tout en les maintenant enfoncées,
placez le bouton de contrôle de température
sur marche (« ON »). Une fois l'écran
numérique allumé, relâchez les touches du
contrôleur. Le mode affichage est indiqué
par un voyant rouge clignotant dans l'angle
inférieur droit de l'écran du contrôleur. Le
voyant de contrôle orange de la façade de la
plaque chauffante ne s'allume PAS en mode
affichage.
REMARQUE : lorsqu'il se trouve en mode
affichage, le contrôleur de température
SCT1 fonctionne uniquement en tant que
thermomètre numérique et ne contrôle plus
la température du liquide. La température
de la plaque chauffante peut alors être
réglée manuellement sur la valeur
souhaitée.
Passer du mode affichage au mode contrôle
Pour passer du mode affichage au mode
contrôle, appuyez une fois sur la touche + ou du contrôleur de température SCT1. L'unité
repasse en mode contrôle et ceci est indiqué par
le voyant orange du thermomètre à contact
8
Loading...
+ 22 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.